Operating instructions for moisture analyzers - UX 3001 (UX 3011, UX 3031, UX 3081) series

Size: px
Start display at page:

Download "Operating instructions for moisture analyzers - UX 3001 (UX 3011, UX 3031, UX 3081) series"

Transcription

1 NPA Basic information Intended use ULTRA X moisture analysers are used to quickly measure moisture and dry matter in solid, viscous and fluid substances by means of thermo gravimetric analysis. A sample is weighed while being dried by infra-red heat. Common features for all ULTRA X devices: Short measurement, gentle and even drying of samples, high reproducibility Automatic or -controlled measurement shut-off UX 3011/ 3011Q/ 3011HQ Temperature-controlled infra-red heater 250 W/ 375W Temperature range: C/ 360 C/ 600 C Stainless steel drying tray 111 mm diameter UX 3031 Special device for gypsum industry 2 temperature-controlled heaters for separate measurement of free and chemically bound moisture in gypsum products Temperature range: C Stainless steel drying tray 111 mm diameter UX 3081/ 3081WQ Special device for large samples, e.g. in the chipboard industry 2 temperature-controlled ceramic heaters Temperature range: C/ 300 C Stainless steel drying tray 245 x 120 mm info@apinstruments.de, Internet: Seite 1

2 Contents Search term Page Accessories, splied 4 and 18 Assigning device ID 9 Automatic mode 7 Automatic shut-off 7 Buttons 6 Calibrating the balance 10 Care and maintenance 16 Changing the print ribbon 5 Communication 17 Connectors 6 Consumables 13 Data interface 13 Date and 10 Dription of interface 13 Display 6 Features 1 ID number 9 Indicator 6 Installation site 4 Intended use 1 Loading the printing paper 5 Maintenance and care 16 Meaning of error messages and remedy 15 Measuring moisture 12 Menu 6 and 14 Menu lock 11 Menu tings 7 Packaging 4 Parameter ting 7 Printer options 9 Printer self-test 16 Safety instructions 3 Setting the measurement unit 7 Setting the temperature 7 Start- 4 Technical data 17 Timer 8 Transport 4 Warranty 4 info@apinstruments.de, Internet: Seite 2

3 Safety instructions Always transport the device in an right position and protect it from impact. Always operate the device on a level surface, using a vibration-free base, Avoid draughts. Never touch the (luminous) heater when it is hot, as there is a risk of injury!! Always disconnect the plug from the mains sply before performing any work on the device (e. g. changing the heater)! Repairs should only be carried out by a qualified technician.! Do not place any flammable substances next to the device. Fire hazard! Please take extra care when using flammable samples! You may need to contact the manufacturer for advice.! Please ensure that no sample spills drip over the edge of the tray sport plate. Use a vacuum cleaner to carefully remove particles. Only use accessories specifically designed for use with ULTRA X devices. Do not move the device while it is measuring, as this could cause measuring errors. Do not leave any substances in the weighing pan, if they are not for measurement. Goods should only be shipped in the original packaging using the original packaging material. ULTRA X moisture analysers are stand-alone devices and not suitable for use in continuous production runs. Any such use is at your own risk. info@apinstruments.de, Internet: Seite 3

4 Warranty a&p instruments undertakes to repair faulty devices within the warranty period, free-of-charge, if the defect is a manufacturing fault. This warranty covers the repair of spare parts and working hours required. It does not cover normal wear-and-tear and soiling and staining of the device. a&p instruments will not accept any freight costs. This warranty does not apply to: improper use of the device, use of the device for purposes other than those stated by us, mechanical damage or damage caused by fluids or sample matter, incorrect or wrong electrical wiring, mechanical damage of the load cell e.g. as a result of being overloaded. Transport, packaging, installation site Please check the packaging and device for any damage on delivery. Please retain the original packaging in case you need to return the goods to a&p instruments. Please only use the original packaging when returning the goods. Accessories splied: 3081 Tray carrier 2 trays Sample tongs Mains cable Start- für x Place the device on a stable, vibration-free surface, as level as possible and protected from draughts. The most suitable locations are the corners of a room, rooms with only one entrance and heavy tables on concrete floors (not for ULTRA X 3081) First adjust the heater to the correct height. To do this, place the level gauge next to the reflector pan on the device. Now adjust the height of the reflector so that its lower edge is in line with the top surface of the rod. Quartz heaters should be 10 mm higher. To do this, slightly loosen (never completely unscrew) the knurled screw on the back of the reflector holder. Retighten the knurled screw manually. 65 mm 75 mm 1. 3 C Now swing the heater to the right or backwards to its rearmost position. Do not use force to twist the heater beyond its stop point, as this could damage the device. UX info@apinstruments.de, Internet: Seite 4

5 1. 4. Now place one of the two drying trays splied on the tray carrier using one or two s of tongs. The notch in the reflector tray acts as an access gap for easy positioning of the tray Check that the voltage indicated on the type plate matches the voltage at the installation site The power cable is inserted into the device s power jack and a suitable earthed mains sply. After the power sply has been connected, the device can be switched on or off using the main switch on the rear cover. After the device has been switched on, it may take to 30 secs before the weight readout is displayed In order to prevent weighing system measuring errors, the device should be switched on 20 minutes before being used to allow for heat compensation. If weight values vary in the display s milligram range during this period, this is considered normal within this phase. If the device is switched off using the on/ off button, it does not require a warm- period if switched on again with the same button ULTRA X moisture analysers are splied, so that in most cases substances can be measured immediately. If you are not satisfied with the result, the measurement parameters can be differently: Loading the printing paper Remove the top cover. To do this, press the printer cover. Insert the paper into the metal slot beneath the printer ribbon and press and hold the linefeed button until approx. 20 mm has been inserted. Now thread the paper through the slot in the printer cover and secure the printer cover. Changing the printer ribbon Remove the printer cover. To do this, press on the rear of the printer. Remove the used ribbon and insert a new ribbon. You may need to turn the scroll wheel in the direction of the arrow until the ribbon cartridge is positioned correctly. The ribbon must be in front of the paper. Now secure the printer cover. info@apinstruments.de, Internet: Seite 5

6 Operation Display and buttons mode value (weight) unit T on re off P1 P2 P3 T on re off Tare button for ting the Button for accessing and unassigned weight readout back to zero scrolling through the menu P1 Button for reting the Button for accessing and unassigned processor program scrolling through the menu P2 Button for exiting Button for confirming entries unassigned P3 menu levels The tare button has 2 functions: Press briefly s back to zero for small weight discrepancies to approx. 16 g. Press and hold weight is tare e.g. an aluminum foil tray. Any weight lower than the one displayed when the device is switched on cannot be back to zero. Connector: Power sply Menu Interface RS 232 Print Conection (only 3081) for separat Printer Power sply switch USB-Interface Factory tings: Temperature:105 C (45 C / 360 C) Automatic shut-off with parameters: Weight loss 5 mg/ Start of query 3 Min/ Query interval 45 sec Quartz heater 12mg/ 3 Min/ 15 sec Buzzer switched on when measuring process is complete Menu tings The device s options menu can be accessed using the or button. In this instruction manual, it is assumed that the button follows the menu item sequence. The menu is operated using a standard structure. UP or DOWN SET ESC Set parameters Confirmation and if required, return to previous menu level Return to previous menu level and exit from menu info@apinstruments.de, Internet: Seite 6

7 Temperature control.. Value is stored.. Value is stored: Automatic shut-off or r The automatic system dries the sample until a constant weight is reached and then switches off the heater. The following tings can be defined here: Parameters can behere: 1. Weight loss in mg a value should be defined here which corresponds to approx. 0.1 % of the original sample weight e.g. 10 g original sample weight 10 mg If the weight loss is <= the mg value displayed here within the interval period shown below, thedevice stops measuring, if the minimum for the query has expired 2. Query interval in sec an interval should be selected here, so that the period for drying the sample is not too short, but also not too long. A period of 20 to 30 secs is normally adequate. info@apinstruments.de, Internet: Seite 7

8 3. Start of query in min Specify the drying period required for the samples without the automatic system being activated. The factory- 3 minutes are usually adequate. This value should be increased for extremely low temperatures, so that the device does not switch itself off too early. This affects the UX 3031 in particular, when free moisture is being measured at 45 C. The next menu option or r is now displayed. Values for the infra-red heater and quartz heater (UX3031 only) are entered separately. Both heaters can be controlled separately via an automatic ting or the r. It is possible to select a drying of to 180 minutes. as shown for infra-red heaters. Measurement unit During measurement, the device optionally displays: % moisture % dry matter (TS) % moisture Atro ( ), i.e. moisture related to the dry matter Dry matter (TS) in g/1000g Weight loss g (V) Residual weight g (R) info@apinstruments.de, Internet: Seite 8

9 Printer options The number of prints for devices without a printer must be to 1! Up to 9 No printout or printout every secs. Data output for external printers (other manufacturer) Linefeed enables an additional LF to be sent via the interface (factory ting no LF). Device ID number Enter a number 9999, which you can 4x use to identify the device info@apinstruments.de, Internet: Seite 9

10 Date and Use / buttons to the date and then press. Use / buttons to the month and then press. Use / buttons to the year and then press. Use / buttons to the hour and then press Use / buttons to the minutes and then press Data and will appear on the printout Buzzer Use / and buttons to enable or disable the buzzer and thenpress. Calibrating the balance Tray carrier and tray must be on the balance! Following the menu instructions, use a 500 g weight F1 to calibrate the balance. Tray being used, must be empty mode value (weight) unit mode value (weight) unit mode value (weight) unit If the display shows switch the power switch off and on, and try it again. info@apinstruments.de, Internet: Seite 10

11 Language The first two letters on the second line stand for Automatic (A) and Timer (T).They show the parameters for each heater. Combinations: AA = Automatic quartz heater /Automatic infra-red heater TA = Timer quartz heater/ Automaticinfra-red heater, TT = Timer quartz heater/ Timer infra-red heater, AT = Automatic quartz heater/ Timer infra-red heater. You can also lock the menu options to prevent accidental changes to drying parameters. The CAL menu option is still available when the menu is locked.. The menu can be locked and unlocked in the following way: Press and hold the buttonfor 5 secs: mode value (weight) unit Position Enter 5-digit 5x underlined pin etc. mode value (weight) unit or Position Enter 5-digit 5x underlined pin etc. If you forget your PIN, you can obtain a new one from a&p instruments. info@apinstruments.de, Internet: Seite 11

12 Moisture analysis process The device is connected to the mains sply and switched on with a drying tray. A weight value is displayed. Set balance to zero using tare button. Display: g. (may take to 4 secs) UX ) 1. Evenly distribute approx.10 g sample material in the drying tray and wait for the weight readout to stop. 2) 2. Swing the heater over the sample (for UX 3081 use heater switch) After a short, the heater switches itself on and the drying phase begins. Header data is printed in advance for devices with a built-in printer. The sample is dried until it has reached a constant weight; the heater is then switched off automatically. The and measurement value is printed for devices with a printer. If you do not wish to make any measurements, please always ensure that the heater has been swung back to its rearmost position! 3. Measurement values are displayed for the entire length of the heater is positioned above the sample. During the measuring process, the device should not be moved as this causes measurement errors. The built-in printer prints the following at the start of the measurement process: Model, device ID date,, heater temperature, r, initial weight And once the measurement process is complete, after the heater has been swung back: Measurement value, measurement, date, substance signum Please ensure drying trays have cooled before using for measurement. If the measurement process ends before it is complete, the display shows: with For substances which are difficult to remove from the tray, we recommend using aluminium foil trays, which can be made quickly and cheaply with the ULTRA X foil press. info@apinstruments.de, Internet: Seite 12

13 Data interface Weighing data, temperature and drying data are also shown on the display via a standard serial interface. Depending on the model, a separate UX 3091 printer or a computer can be connected to the device. Interface dription Model: V24 RS 232 serial data transmission and flash player Transmission rate: Number of bits Parity check Handshake 9600 baud 8 data bits 1 stop bit none none Datenkabel Feuchtemesser - DataChannel ca. 2 m PC Data transmission using a flash player If the computer receiving the data is connected to the internet, the system automatically installs a driver and creates a COM interface, as soon as the computer is connected to the moisture analyser. If this is not the case, the driver has to be manually installed in the device manager. A special software is required for transmitting data to a PC: ULTRA X DataChannel Data can be exported in four different data formats:. xls;.html;.xml;.txt. The software is easy to install and operate: Connect devices, place CD in the PC, the installation is self-explanatory. Accessories: Order no.: Use: Stainless steel drying tray Sample carriers UX 3011/ 3031 (2 trays splied) AV drying tray, large Sample carriers UX 3081(2 trays splied) Tray tongs for lifting the tray (splied) Mains cable H Mains connection (splied) Consumables: Printing paper 58 mm I for built-in printer Ribbon I for built-in printer info@apinstruments.de, Internet: Seite 13

14 Overview of menu via the UP button M e n u Menu language Language, GERMAN Menu CAL Calibrate?, press <> or <> Calibrate, If balance empty, press <> Please wait! Add 500 g weight, press <> Please wait! Remove weight, CAL done! Menu buzzer Buzzer enabled / disabled Menu date + Date. Day. Month. Year, Time : Hours : Minutes Menu ID Machine ID 9999 Menu Data output Data output, no LF/ with LF Menu Printer Printer, Number of print copies Print copies, _ print copies Printer, Interval Printout, every secs Menu unit unit, moisture Menu Auto/ Timer, Quarz Auto/ Timer, Quarz, Automatic Automatic, weight loss mg weight loss mg, mg Automatic, Interval secs Interval secs, secs Automatic, Start mins Start mins, mins <> Menu Auto/ Timer, IR Auto/ Timer, IR Timer Timer, Min Menu Temperature control Temp. control, I heater 170 C Temp. control, Q heater 360 C (only for ULTRA X 3031 model) info@apinstruments.de, Internet: Seite 14

15 Potential errors Error Reason Remedy Display blank : Initial load too small or large, Tray carrier and / or tray missing Place empty drying dray on centre of tray carrier Location with too much disturbance Place on a vibration-free surface Draught Protect from draughts Area around tray carrier is dirty Carefully clean area around tray carrier Balance is faulty Balance without power Balance is not connected to control card Balance is faulty Balance is continuing to run Location with too much disturbance Draught Area around tray carrier is dirty Balance is faulty Drying paramenters cannot Menu is locked be entered Place on a vibration-free surface Protect from draughts Carefully clean area around tray carrier Unlock menu lock, contact the person responsible or a&p instruments Menu lock cannot be unlocked 5-digit pin for unlocking menu lock is incorrect at a.ukena@apinstruments.de Balance cannot be tared Weight within minus range Switch off device, place empty tray on balance and switch on again Weight difference less than 16 mg or more than 16 mg Less than 16 mg to zero, press button briefly More than 16 mg tare, press and hold Balance cannot be calibrated: Initial load too small or large, tray carrier and / or tray missing. Place empty drying dray on centre of tray carrier, Location with too much disturbance Place on a vibration-free surface Draught Protect from draughts Area around tray carrier is dirty Carefully clean area around tray carrier Calibration process has stopped Switch power-switch off and on and try it again Balance ting is incorrect Balance is faulty Heater will not switch on: Heater is faulty Change heater Temperature probe is faulty Infrared heater will not S202S01 relay is faulty switch off Measurement not Shut-off tings incorrect Check shut-off tings completing: Timer tings incorrect Check r tings info@apinstruments.de, Internet: Seite 15

16 Error Reason Remedy No printout Printer is faulty or no data transmission No linefeed Linefeed button or printer is faulty No connection to printer Printout distorted Printing unit or printer board is faulty Printer not printing a strip in the print graphic Printing unit is not installed correctly Printing unit or printer board is faulty Printout too light Ribbon is empty Change ribbon Printing unit is dirty Care and maintenance Please ensure that no pollutants enter the device. Always keep the area around the tray carrier clean. Do not use compressed air to clean the device. Do not overload the balance, max. weight 750 g. Protect device from moisture and impact. For ideal care and maintenance, we recommend taking out a service contract. Printer self-test If your printout is incorrect, you will need to run a printer self-test: Disconnect the device from the mains sply using the main switch or plug. Press and hold the linefeed button and connect the device to the mains, then release the linefeed button. The printer will print a self-test: If this printout is correct, the printing unit and printer board are working but the data transmission is not functioning. If the printout is incorrect, there may be several reasons for this error. info@apinstruments.de, Internet: Seite 16

17 Technical data Balance: Weight range Resolution Sample volume Weighing tray For moisture analysis: Initial weight Heater capacity to 750 g 0.1 g max. 95 cm³ 110 mm round/ 245 x 120 mm any weight from approx. 5 g, recommended from 10 g Infrared heater 250 watts Quartz heater 375 watts Ceramic heater 2x 250 watts (3081)/ 2x 400 watts (3081 WQ) Automatic shut-off Automatic drying to a constant weight, using 3 factors for I heater and Q heaters, which can be separately: Query start, query interval, weight loss/ query interval Temperature Infrared heater adjustable from 40 C to 170 C Quartz heater adjustable from 100 C to 360 C Ceramic heater adjustable from 40 C to 180 C (3081)/ 120 C to 360 C (3081QW) Measuring units Reading accuracy 0.1% Communication: % moisture % dry matter % ATRO moisture g solid matter/ kg Data terminal V24 RS 232 Data output all weight and measuring data meet GLP requirements with a start, optional variable interval printout, date,, total measuring etc. Using ULTRA X DataChannel software can be transmitted to a PC as a For operating the device:.xls file in Microsoft Excel,.html file,.xml file and.txt file. Mains connection 230 V 2%/ + 15 % Mains frequency Hz Power input 250 watts 850 watts, depending on the model Dimensions approx. B 385 x D 275 x H 420 mm Weight approx. 9 kg/ 15 kg (3081 model) Technical data subject to change without notice ULTRA X moisture analysers are stand-alone devices and not suitable for use in continuous production runs. Any such use is at your own risk! info@apinstruments.de, Internet: Seite 17

18 Accessories: Order no.: Use: Stainless steel drying tray Sample carriers UX 3011/ 3031 (2 trays splied) AV drying tray, large Sample carriers UX 3081 (2 trays splied) Tray tongs for lifting the tray (splied) Mains cable H Mains connection (splied) Calibration weight F1 500 g for calibrating the balance Aluminium foil 130 mm x 0.1 mm Aluminium foil 130 mm x 0.3 mm Aluminium foil 130 mm x 260 mm H for placing on the drying tray H for placing on the drying tray H for placing on the drying tray Foil press for shaping aluminium foil Software DataChannel for transmitting data to a PC Consumables: Printing paper 58 mm I for built-in printers Ribbon I for built-in printers Infrared heater 250 W H Spare part Quartz heater 375 W Spare part Ceramic heater 250 W Spare part Ceramic heater 400 W Spare part Ceramic heater 600 W Spare part info@apinstruments.de, Internet: Seite 18

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460)

TS400. Operating Manual. Test Station for Microtector II Series (G450/G460) Operating Manual TS400 Test Station for Microtector II Series (G450/G460) GfG GESELLSCHAFT FÜR GERÄTEBAU MBH KLÖNNESTRASSE 99 44143 DORTMUND, Germany TEL. +49 / (0)2 31 / 5 64 00 0 FAX +49 / (0)2 31 /

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.goodforgas.com GfG Products for Increased Safety Congratulations

More information

Operating instructions Electronic Moisture Analyser

Operating instructions Electronic Moisture Analyser KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Electronic Moisture Analyser

More information

Operating instructions Electronic Moisture Analyser

Operating instructions Electronic Moisture Analyser KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating instructions Electronic Moisture Analyser

More information

Operations Manual TS400. Test Station for G450/G460 Gas Detector

Operations Manual TS400. Test Station for G450/G460 Gas Detector TS400 Test Station for G450/G460 Gas Detector Operations Manual 1194 Oak Valley Dr, Ste 20, Ann Arbor MI 48108 USA (800) 959-0329 (734) 769-0573 www.gfg-inc.com GfG Products for Increased Safety Congratulations

More information

Halogen Moisture Analyzer

Halogen Moisture Analyzer Halogen Moisture Analyzer About the Halogen Moisture Analyzer Mode XY-105MW XY-102MW XY-100MW Specification 110g/5mg 110g/2mg 110g/1mg Readability 5mg 2mg 1mg Tray size Φ 90 Heating mode Halogen Lamp Display

More information

Mark 25 Ultrapure Water Conductivity Analyzer

Mark 25 Ultrapure Water Conductivity Analyzer Martek Instruments, Inc. Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual WARRANTY POLICY Unless otherwise stated, MARTEK INSTRUMENTS, INC. warrants this equipment to be free from defects

More information

Advance Optima Module Magnos 17

Advance Optima Module Magnos 17 Advance Optima Module Magnos 17 Service Manual 43/24-1002-0 EN Table of Contents Page Chapter 1: Description of functions 1-1 Chapter 2: Module variants and components 2-1 Chapter 3: Analyzer variants

More information

USER MANUAL FOR OPERATING SYSTEM

USER MANUAL FOR OPERATING SYSTEM P2262 ALARM PANEL USER MANUAL FOR OPERATING SYSTEM 21765-07 September 1999 Associated Controls (Aust) PTY. LTD. 29 Smith Street, Hillsdale, NSW, 2036. PH (02) 9311 3255, FAX (02) 9311 3779 Page 1 of 177

More information

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA 99336 www.jmcvr.com 1-509-586-9893 Table of Contents TABLE OF FIGURES...1 OVERVIEW OF THE HRC CAPABILITIES...2 INSTALLATION AND

More information

Infrared Moisture Determination Balance FD-610. Operating Manual

Infrared Moisture Determination Balance FD-610. Operating Manual Infrared Moisture Determination Balance FD-610 Operating Manual Infrared Moisture Determination Balance Safety Notes Improper use of the infrared moisture tester in violation of the following safety notes

More information

Moisture analysis in minutes - not hours. Metal, Ceramic, Halogen or Quartz The Choice is Yours

Moisture analysis in minutes - not hours. Metal, Ceramic, Halogen or Quartz The Choice is Yours Moisture analysis in minutes - not hours Metal, Ceramic, Halogen or Quartz The Choice is Yours The right analyzer for any application Foods, chemical, pharmaceutical products, building materials or animal

More information

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS... SECTION Table of Contents SECTION 1 INTRODUCTION...................... 1.1 Description............................... 1.2 Features................................. 1.3 Models..................................

More information

SCAN200E USER S MANUAL

SCAN200E USER S MANUAL SCAN200E USER S MANUAL Code No. 2071 1052 rev. 1.4 Code No. 2071 1052 Rev. 1.4 Page 2/16 SCAN200E User s Manual Foreword This manual is for SCAN200E Controller running software version 2.03 or later. We

More information

Number of instruction: LMI-11-02/01/08/ENG-C. Moisture analyzer MBC series

Number of instruction: LMI-11-02/01/08/ENG-C. Moisture analyzer MBC series Number of instruction: LMI-11-02/01/08/ENG-C Moisture analyzer MBC series JANUARY 2008 2 1. TECHNICAL PARAMETERS... 5 2. BASIC INFORMATION... 6 2.1. Balance appropriation... 6 2.2. Usage conditions...

More information

OPERATING MANUAL. EchoTherm PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS IN35, IN45 and IN55-12VDC. DOCUMENT NUMBER IN35-03 Revised 15 May 2015

OPERATING MANUAL. EchoTherm PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS IN35, IN45 and IN55-12VDC. DOCUMENT NUMBER IN35-03 Revised 15 May 2015 OPERATING MANUAL EchoTherm PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS IN35, IN45 and IN55-12VDC DOCUMENT NUMBER IN35-03 Revised 15 May 2015 TORREY PINES SCIENTIFIC, INC. 2713 Loker Ave. West Carlsbad,

More information

Instruction Manual MB90 Moisture Analyzer

Instruction Manual MB90 Moisture Analyzer Instruction Manual Moisture Analyzer EN-1 Table of Contents 1. GETTING TO KNOW YOUR MOISTURE ANALYZER... 4 1.1 Introduction... 4 1.2 Overview of the Moisture Analyzer... 4 1.3 What is a Moisture Analyzer?...

More information

FIRERAY 5000 range USER GUIDE

FIRERAY 5000 range USER GUIDE FIRERAY 5000 range USER GUIDE 0044-003-04 IMPORTANT PLEASE NOTE: The beam path MUST be kept clear of obstructions at all times! Failure to comply may result in the Detector initiating a Fire or Fault signal.

More information

IMR 5000 IMR 400. CEMS Operating Manual IMR USA S/N: Sample IMR Sensors Inside IMR 400. Gas Dryer Inside. IMR Environmental Equipment Inc.

IMR 5000 IMR 400. CEMS Operating Manual IMR USA S/N: Sample IMR Sensors Inside IMR 400. Gas Dryer Inside. IMR Environmental Equipment Inc. CEMS Operating Manual IMR USA S/N: Sample IMR 5000 IMR 400 IMR 400 IMR 5000 Measure MESURE PURGE Purge CALIBRATE Calibrate OFF OFF RESET Measure ON ON Calibrate RESET PUMP Gas Dryer Inside PUMP Sensors

More information

KERN MPK Version GB

KERN MPK Version GB KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operation manual Personal scales KERN MPK Version

More information

725B Configuration Software Manual

725B Configuration Software Manual 725B Configuration Software Manual REV DATED DESCRIPTION AUTHOR APPROVED 0 09-03-10 First Issue P.Cartmell Page 1 of 80 SECTION 1 - SOFTWARE INSTALLATION... 5 725B ConfigurationSoftware Installation...

More information

portable gas analyser

portable gas analyser www..com portable gas analyser The largest of s analysers equipped with electrochemical cells. It can fit even up to 7 EC cells and up to 3 sensors. GA-60 has a large (320*240), graphical LCD with backlighting.

More information

OPERATOR S MANUAL MODEL AP15/AP15-1/AP15-2 PC-ALARM PANEL

OPERATOR S MANUAL MODEL AP15/AP15-1/AP15-2 PC-ALARM PANEL ap15_2manual04/22/13 Page 1 4/22/2013 1 Serial Number : Option: OPERATOR S MANUAL MODEL AP15/AP15-1/AP15-2 PC-ALARM PANEL Micro Seven, Inc. 1095-K N.E. 25th Hillsboro, OR 97124 U.S.A. phone: 503-693-6982

More information

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual Read instructions carefully before operating this device. ❶ This device is not to be used for Human Life Support applications. ❷ To avoid possible

More information

TECHNICAL MANUAL CVM 20 C 5005 CV/04-99 GB

TECHNICAL MANUAL CVM 20 C 5005 CV/04-99 GB Summary 1 CONNECTIONS... 3 1.1 TEMPERATURE PROBES...3 1.2 LOW VOLTAGE DIGITAL INPUTS...3 1.3 LIVE DIGITAL INPUTS...4 1.4 RELAY OUTPUTS...5 2 POWER SUPPLY... 6 3 SERIAL CONNECTIONS... 6 4 SOFTWARE... 7

More information

Moisture analyzer PMC series

Moisture analyzer PMC series Number of instruction: LMI-01-01/10/11/USA Moisture analyzer PMC series MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 600 Radom, Bracka 28 Street, POLAND phone +48 48 38 48 800, phone/fax.

More information

User Manual Watt Drop-In Induction Range. Model: ID18DB 08/2017. Please read and keep these instructions. Indoor use only.

User Manual Watt Drop-In Induction Range. Model: ID18DB 08/2017. Please read and keep these instructions. Indoor use only. 3100 Watt Drop-In Induction Range Model: ID18DB 08/2017 Please read and keep these instructions. Indoor use only. www.avantcoequipment.com 1 Specifications Model No... ID18DB Power Rating... Front: 500-1300W,

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

Manual v1 Page 1 of 19. Manual. Laboratory autoclave

Manual v1 Page 1 of 19. Manual. Laboratory autoclave Manual 20.2.2015 v1 Page 1 of 19 Manual Laboratory autoclave Autoclave using steam for the treatment of material and laboratory processes e.g. sterilization, as an aid in research and development CertoClav

More information

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH

M-Sens 2. Online-Moisture Meter for Solids. Operating Instructions. SWR engineering Messtechnik GmbH EN M-Sens 2 Operating Instructions Online-Moisture Meter for Solids SWR engineering Messtechnik GmbH CONTENTS Page 1. System Overview......................................................................

More information

Operating Manual Bathroom Balance

Operating Manual Bathroom Balance KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating Manual Bathroom Balance KERN MGB Version

More information

Operating manual. a Look Solutions 1 product

Operating manual. a Look Solutions 1 product Operating manual a Look Solutions 1 product Set of Equipment supplied 1 Unique 2 1 tank lid with quick connector, silicon tube and air valve filter 1 canister of UNIQUE-fluid, 2l 1 manual As an option:

More information

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs User Manual Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs Models 89800-03 and 89800-04 THE STANDARD IN PRECISION MEASUREMENT Table of Contents

More information

IMR 5000 Operating Manual S/N: SAMPLE For information only! IMR Environmental Equipment Inc.

IMR 5000 Operating Manual S/N: SAMPLE For information only! IMR Environmental Equipment Inc. IMR 5000 Operating Manual S/N: For information only! IMR Environmental Equipment Inc. INTRODUCTION 3 SAFETY INSTRUCTIONS 3 SYSTEM DESCRIPTION - IMR 5000 GAS ANALYSIS SYSTEM 4 Function chart IMR 5000 -

More information

Refrigerated Incubator Model and Operating Instructions

Refrigerated Incubator Model and Operating Instructions Refrigerated Incubator Model 165000 and 165000-2 Operating Instructions N2400379 - Rev. 1 08May2018 1 Contents 1. SAFETY...3 1.1. EMF INTERFERENCE...4 1. PRODUCT INFORMATION...5 1.1 INTRODUCTION...5 2.

More information

Sierra Model 951 Digital Flo-Box. Instruction Manual Part Number IM-951 Revision A 09-05

Sierra Model 951 Digital Flo-Box. Instruction Manual Part Number IM-951 Revision A 09-05 Sierra Model 951 Digital Flo-Box Instruction Manual Part Number IM-951 Revision A 09-05 5 Harris Court, Building L Monterey, CA 93940 (831) 373-0200 (800) 866-0200 Fax (831) 373-4402 www.sierrainstruments.com

More information

MOISTURE ANALYZER PMC SERIES

MOISTURE ANALYZER PMC SERIES User manual no.: IMMU-12-03-04-14-USA MOISTURE ANALYZER PMC SERIES BALANCES AND SCALES RADWAG 26 600 Radom, Bracka 28, POLAND Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10 Email: office@radwagusa.com

More information

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP

M-Sens 2. Online moisture measurement for solids. Operating Instructions. SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP EN M-Sens 2 Operating Instructions Online moisture measurement for solids SWR engineering Messtechnik GmbH PART OF THE ENVEA GROUP CONTENTS Page 1. System overview.............................................

More information

ECOLOG TN2, TN3-P, TN4, TN4-L, TP2, TP4-L,TH1, TH2

ECOLOG TN2, TN3-P, TN4, TN4-L, TP2, TP4-L,TH1, TH2 TN2, TN3-P, TN4, TN4-L, TP2, TP4-L,TH1, TH2 1. Product Overview 1.1 Display Large LCD display for measured values and status Time Alarm (will stay lit until Upper / Lower limit value (except for TN2) the

More information

KERN MFB Type MGE Version GB

KERN MFB Type MGE Version GB KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating manual Electronic bathroom scales KERN

More information

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585)

2364 Leicester Road, P.O. Box 175, Leicester, NY Phone (585) Fax (585) Dry Heat Sterilizers with 60 Minute Sterilization Cycle Times MODEL 3100 MODEL 2100 2364 Leicester Road, P.O. Box 175, Leicester, NY 14481 Phone (585) 382-3223 Fax (585) 382-9481 www.cpac.com December

More information

Oakton TEMP 9500 Advanced Multiparameter Controller

Oakton TEMP 9500 Advanced Multiparameter Controller Oakton TEMP 9500 Advanced Multiparameter Controller Models: 89800-03 & 89800-04 Oakton Instruments 625 E Bunker Ct. Vernon Hills, IL 60061, USA 1-888-4OAKTON (1-888-462-5866) info@4oakton.com Contents

More information

ADOS Biogas 905. Multi-channel Gas Analyser. A D O S GmbH Instrumentation and Control Trierer Str FRG

ADOS Biogas 905. Multi-channel Gas Analyser. A D O S GmbH Instrumentation and Control Trierer Str FRG ADOS Biogas 905 Multi-channel Gas Analyser Issue: 05/2006 A D O S GmbH Instrumentation and Control Trierer Str. 23-25 @ 52078 Aachen @ FRG TEL. : (0241) 9769-0 FAX : (0241) 9769-16 E-Mail : info@ados.de

More information

Service Instructions

Service Instructions BASIC EXCELLENT PERFECT U Universal ovens I Incubators S Sterilisers Service Instructions Mo Tu We Th Fr Sa Su t3 on off h t2 t1 t4 loop 4 3 2 1 STERI DEFRO C MIN MAX C IN 1 IN 2 OUT IN 1 IN 2 OUT % rh

More information

Table of Contents SECTION PAGE

Table of Contents SECTION PAGE Table of Contents SECTION PAGE SECTION 1 INTRODUCTION................... 1.1 Description.............................. 1.2 Features................................ 1.3 Models.................................

More information

Warranty Registration

Warranty Registration WARRANT Y AND LIMITS OF LIABILIT Y Vulcain Inc. warrants to the original purchaser that its product, and the component parts thereof, will be free from defects in workmanship and materials for a period

More information

SyringeHeater.com. Syringe Heater. Thermo-Kinetic Heat Clamping. Model: HEATER-KIT-1LG HEATER-KIT-5SP

SyringeHeater.com. Syringe Heater. Thermo-Kinetic Heat Clamping. Model: HEATER-KIT-1LG HEATER-KIT-5SP SyringeHeater.com SyringePump.com Syringe Heater Thermo-Kinetic Heat Clamping Model: HEATER-KIT-1LG HEATER-KIT-5SP CAUTION RISK OF ELECTRIC SHOCK CAUTION: SURFACE OF HEATING PADS WILL GET VERY HOT WARNING

More information

Dryer Moisture Systems Inc. Dryer Master M 2. Product Manual

Dryer Moisture Systems Inc. Dryer Master M 2. Product Manual Dryer Moisture Systems Inc. Dryer Master M 2 Revision 1. January, 27 Revision 1.1 March, 27 Revision 1.2 June, 27 Revision 1.22 July, 27 Revision 1.22.1 January, 28 Revision 1.22.2 April, 28 Revision 1.23

More information

OPERATING MANUAL MOLECULAR DIMENSIONS, INC. PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS MD5-601 AND MD5-603

OPERATING MANUAL MOLECULAR DIMENSIONS, INC. PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS MD5-601 AND MD5-603 OPERATING MANUAL MOLECULAR DIMENSIONS, INC. PROGRAMMABLE DIGITAL CHILLING INCUBATOR MODELS MD5-601 AND MD5-603 DOCUMENT NUMBER DM5-601/603-1 January 21, 2011 MOLECULAR DIMENSIONS, INC. United Kingdom Office

More information

Tovatech LLC Phone (973) Rutgers St, Maplewood, NJ 07040

Tovatech LLC Phone (973) Rutgers St, Maplewood, NJ 07040 Moisture content is a quality control criterion signifying the amount of moisture in a product. Laws as well as industry trade associations define the correct moisture content for a broad range of products

More information

Operating instructions Page 12

Operating instructions Page 12 Operating instructions Page 12 Refrigerator with explosion-proof interior container Read the operating instructions before switching on for the first time 7082 271-00 LKEXv 910 Disposal notes The appliance

More information

testo 315-1/2 Instruction manual 99 Washington Street Melrose, MA Phone Toll Free

testo 315-1/2 Instruction manual 99 Washington Street Melrose, MA Phone Toll Free testo 315-1/ Instruction manual en 99 Washington Street Melrose, MA 0176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Contents Foreword.................................................................

More information

VISY-X. Technical Documentation. VISY-SoftView V 1.7 User Guide for VISY-Command GUI and VISY-View Touch. Edition: Version: 6 Art.

VISY-X. Technical Documentation. VISY-SoftView V 1.7 User Guide for VISY-Command GUI and VISY-View Touch. Edition: Version: 6 Art. Technical Documentation VISY-X VISY-SoftView V 1.7 User Guide for VISY-Command GUI and VISY-View Touch Edition: 2015-04 Version: 6 Art. No: 350026 FAFNIR GmbH Bahrenfelder Str. 19 22765 Hamburg, Germany

More information

VIBRA-series: VIBRA, VIBRA +

VIBRA-series: VIBRA, VIBRA + DATASHEET VIBRA-series (VIB.00100/..120/..130) VIBRA-series: VIBRA, VIBRA + Profound VIBRA-series Vibrations from pile driving, construction, road or rail traffic, demolition work and blasting can create

More information

Installation Manual Tracker 221

Installation Manual Tracker 221 Installation Manual Tracker 221 KSR KUEBLER AG Im Kohlstatterfeld 17 D-69439 Zwingenberg / Germany www.ksr-kuebler.com Contents About this Installation Manual...4 Description Tracker 220-series...4 Installing

More information

Operating Instructions. Moisture Analyzer MJ33

Operating Instructions. Moisture Analyzer MJ33 Operating Instructions Moisture Analyzer MJ33 Overview of your MJ33 Moisture Analyzer 2 1 18 MJ33 C L A S S I C L i g h t 11 17 12 16 15 14 13 19 1 20 24 2:54 88 C 23 22-12.48 % MC 21 8 10 7 9 5 25 26

More information

Rev B, 9/2/2009. Kodiak Chiller Overview

Rev B, 9/2/2009. Kodiak Chiller Overview 930-0001 Rev B, 9/2/2009 Kodiak Chiller Overview Presentation Outline Phone: 781-933-7300 Lytron Technical Support Contact Information 3 Introduction 4 Part I: Unpacking 5 Part II: Installation 7 Part

More information

LCF Touch Modbus Datasheet Application Security Advice Caution

LCF Touch Modbus Datasheet Application Security Advice Caution LCF Touch Modbus Electronic Fan Coil Thermostat with Touch Display (Flush mounting) Datasheet Subject to technical alteration Issue date: 3.11.214 Application Modern design flush mounting fan coil room

More information

User Manual. Digi-Sense TC9600 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

User Manual. Digi-Sense TC9600 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs User Manual Digi-Sense TC9600 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs Models 89800-13 and 89800-14 THE STANDARD IN PRECISION MEASUREMENT Table of Contents

More information

Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM

Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM Algo-Tec 6500/6600 INTERACTIVE ADDRESSABLE FIRE CONTROL SYSTEM (1-4 LOOPS) USER MANUAL Protec Fire Detection plc, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT, ENGLAND +44 (0) 1282 717171 www.protec.co.uk

More information

and Channel indicator and monitoring unit B Operating Instructions 2.00/

and Channel indicator and monitoring unit B Operating Instructions 2.00/ and Channel indicator and monitoring unit B 70.0202 Operating Instructions 2.00/00375947 Contents 1 Instrument functions... 3 1.1 Ordering details... 3 1.2 Functions of the LMD-96... 4 1.3 Functions of

More information

OPERATING MANUAL MOLECULAR DIMENSIONS, INC. DIGITAL CHILLING INCUBATOR MODELS MD5-600, MD5-602, AND MD5-604

OPERATING MANUAL MOLECULAR DIMENSIONS, INC. DIGITAL CHILLING INCUBATOR MODELS MD5-600, MD5-602, AND MD5-604 OPERATING MANUAL MOLECULAR DIMENSIONS, INC. DIGITAL CHILLING INCUBATOR MODELS MD5-600, MD5-602, AND MD5-604 DOCUMENT NUMBER DM5-601/603-1 January 21, 2011 MOLECULAR DIMENSIONS, INC. United Kingdom Office

More information

Operating manual. Viper S USA. a Look Solutions 1 product

Operating manual. Viper S USA. a Look Solutions 1 product Operating manual Viper S a Look Solutions 1 product USA Set of Equipment supplied 1 Viper S 1 25.3 oz tank incl. lid with Quick Connector 1 manual Please check whether all the products you ordered are

More information

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY S E C T I O N 10 C O N T E N T S N A V I G A T I O N Introduction.................... 10.2 Portable Meters................. 10.3 Transmitter..................... 10.9 Probes......................... 10.10

More information

Installation and operating instructions. Temperature difference controller 6 inputs, 3 outputs, integrated data logger for SD card

Installation and operating instructions. Temperature difference controller 6 inputs, 3 outputs, integrated data logger for SD card SOLARTHERMIE - SOLAR THERMAL - SOLAR TÉRMICO - SOLAIRE THERMIQUE - SOLARE TERMICO Installation and operating instructions Temperature difference controller 6 inputs, 3 outputs, integrated data logger for

More information

INSTRUCTIONS FOR USE OF COMBINED BOILER INTENDED FOR COMBUSTION OF BOTH PELLETS AND SOLID FUEL ABC COMBO

INSTRUCTIONS FOR USE OF COMBINED BOILER INTENDED FOR COMBUSTION OF BOTH PELLETS AND SOLID FUEL ABC COMBO INSTRUCTIONS FOR USE OF COMBINED BOILER INTENDED FOR COMBUSTION OF BOTH PELLETS AND SOLID FUEL ABC COMBO .Technical specifications Boiler power DESCRIPTION Water content in a boiler Required draft Supply

More information

Infrared Moisture Balance (FD-800)

Infrared Moisture Balance (FD-800) Dual Temperature Control Moisture Analyzer (FD-800) FD-800 is premium infrared moisture determination balance which employs a revolutionary dual temperature sensor method to enable high precision moisture

More information

73 Series Spectrophotometer. Accessory fitting and operation manual

73 Series Spectrophotometer. Accessory fitting and operation manual 73 Series Spectrophotometer Accessory fitting and operation manual 061 672 Safety Please read this information carefully prior to installing or using this equipment. 1. The unit described in this manual

More information

SC-F3G User Manual 1.0

SC-F3G User Manual 1.0 SC-F3G User Manual 1.0 Table of Contents 1. Introduction... 3 2. Functions... 3 3. Features... 3 4. Package Contents... 3 5. Device Configuration... 4 6. Status LED signals... 5 7. Before You Start...

More information

Sirius Technologies, Inc.

Sirius Technologies, Inc. Model#: ST501 CO2/Temp/RH Indoor Air Quality Meter Users Manual Page 1 TABLE OF CONTENTS FEATURES & SPECIFICATIONS... 3 WHAT S IN THE BOX... 7 START UP AND GENERAL OPERATION... 8 Recording and Datalogging...

More information

VR60 GLASS FROSTERS INSTALLATION AND OPERATING MANUAL PLEASE LEAVE WITH OPERATOR. VR60 Glass Froster. Series 1

VR60 GLASS FROSTERS INSTALLATION AND OPERATING MANUAL PLEASE LEAVE WITH OPERATOR. VR60 Glass Froster. Series 1 VR60 GLASS FROSTERS INSTALLATION AND OPERATING MANUAL PLEASE LEAVE WITH OPERATOR VR60 Glass Froster Series 1 Imperial Machine Company Limited Unit 1, Abbey Road Wrexham Industrial Estate Wrexham LL13 9RF

More information

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time Operating instructions Page 14 Refrigerator Read the operating instructions before switching on for the first time 7085 039-00 LKv 5710 Content Disposal notes... 14 Description of the appliance... 14 Safety

More information

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers. ModSync Sequencing System Installation & Operation Manual For use with Fulton Steam Boilers. Revision 3.0 8/21/2008 - 2 - Table of Contents Introduction Page 4 Features Page 4 Sequence of Operation Page

More information

Operating instructions Page 14

Operating instructions Page 14 Operating instructions Page 14 Refrigerator with explosion-proof interior container Read the operating instructions before switching on for the first time 7085 479-00 LKexv Inhalt Disposal notes... 14

More information

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English

PROCESS ANALYSERS. SERVOFLEX MiniMP (5200) Gas Analyser. Operator Manual. Part Number: A Revision: 8 Language: UK English PROCESS ANALYSERS SERVOFLEX MiniMP (5200) Gas Analyser Operator Manual Part Number: 05230001A Revision: 8 Language: UK English This page intentionally blank. LIST OF CONTENTS Section Page 1 DESCRIPTION

More information

Operating Instructions. Moisture Analyzer HE53

Operating Instructions. Moisture Analyzer HE53 Operating Instructions HE53 Table of Contents 1 Introduction 3 1.1 Conventions and Symbols Used in these Operating Instructions 3 2 Safety Information 4 2.1 Definition of Signal Warnings and Symbols 4

More information

EasyTronic III MANUAL SERVICE

EasyTronic III MANUAL SERVICE rev.6 EasyTronic III MANUAL SERVICE General characteristics: Power supply 24 Vac ±15% Max consumption at 24Vac 300mA Relay outputs 6 Maximum relay current 8 A res. Serial standard RS232 2 Serial standard

More information

Element. Controller. Manual for use and maintenance. Element. Climate Controller. Ag/MIS/UMGB /17 Rev 1.0 P/N:

Element. Controller. Manual for use and maintenance. Element. Climate Controller. Ag/MIS/UMGB /17 Rev 1.0 P/N: Manual for use and maintenance Element Controller Element Climate Controller Ag/MIS/UMGB-2439-02/17 Rev 1.0 P/N: 110584 Element Manual for use and maintenance Original instructions This manual for use

More information

OPERATING MANUAL. EchothermJ HPLC COLUMN CHILLER/HEATER MODEL CO30. DOCUMENT NUMBER CO30-03 Revised February 10, 2010

OPERATING MANUAL. EchothermJ HPLC COLUMN CHILLER/HEATER MODEL CO30. DOCUMENT NUMBER CO30-03 Revised February 10, 2010 OPERATING MANUAL EchothermJ HPLC COLUMN CHILLER/HEATER MODEL CO30 DOCUMENT NUMBER CO30-03 Revised February 10, 2010 TORREY PINES SCIENTIFIC, INC. 2713 Loker Ave. West Carlsbad, CA 92010 Telephone: (760)-930-9400

More information

KERN MGD Version GB

KERN MGD Version GB KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: info@kern-sohn.com Phone: +49-[0]7433-9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating manual Electronic bathroom scales KERN

More information

Ambient Temperature/ Relative Humidity & Dew Point Temperature Sensors to USB Output. Model LFS108B

Ambient Temperature/ Relative Humidity & Dew Point Temperature Sensors to USB Output. Model LFS108B Ambient Temperature/ Relative Humidity & Dew Point Temperature Sensors to USB Output Model LFS108B Copyright 2018 - LLC. All rights reserved. Table of Contents Quick Start... 3 1- Introduction... 4 2-

More information

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas User s Manual TIGER S EYE E-Series Mark V Jockey TIGERFLOW Systems, Inc. 4034 Mint Way Dallas, Texas 75237 214-337-8780 www.tigerflow.com TABLE OF CONTENTS Introduction... 4 Sequence of Operation... 5

More information

Part 3 Troubleshooting

Part 3 Troubleshooting Part Troubleshooting What is in this part? This part contains the following chapters: Chapter See page Troubleshooting 2 Error Codes: Hydro-box 7 Error Codes: Outdoor Units Error Codes: System Malfunctions

More information

ALGM. Moisture Analyser

ALGM. Moisture Analyser ALGM Moisture Analyser User Manual ALGM_72_14.01_EN_U Content: 1. General description... 3 2. Completeness... 3 3. Security rules... 4 4. Environment protection... 5 5. Technical data... 5 6. Keys and

More information

Medical Air circulated Refrigerator TC (2 independent temperature ranges), also in accordance with DIN 58345

Medical Air circulated Refrigerator TC (2 independent temperature ranges), also in accordance with DIN 58345 External Dimensions: B = 530 mm D = 640 mm H = 1600 mm Inside Dimension: Useful space on top Useful space at the bottom B = 450 mm B = 450 mm D = 535 mm D = 535 mm H = 530 mm H = 530 mm Capacity: 280 l

More information

OPERATION MANUAL. THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

OPERATION MANUAL. THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS OPERATION MANUAL THE JENCO MODEL 6230 MICROCOMPUTER BASED ph/mv/temp PORTABLE METER JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS GENERAL INTRODUCTION The model 6230 is a precise instrument

More information

MB23 and MB25 Moisture Analyzers

MB23 and MB25 Moisture Analyzers SERVICE MANUAL MB23 and MB25 Moisture Analyzers Ohaus Corporation 7 Campus Drive, Suite 310, Parsippany, NJ 07054 USA (973) 377-9000 SERVICE MANUAL MB23 and MB25 Moisture Analyzers The information contained

More information

HRX Technical Manual. Version 1.2

HRX Technical Manual. Version 1.2 HRX 5000 Technical Manual Version 1.2 Contents: Specification...2 Connectors...5 RS-485 Network Connectors (J6 and J7)...5 RS-232 to Printer (J19)...6 RS-232 to PC (J8)...7 TCP/IP...8 Power (J21)...9 Fire

More information

Turbimax CUE21 / CUE22

Turbimax CUE21 / CUE22 Technical Information Turbimax CUE21 / CUE22 Turbidimeter for on-line measurement Application On-line continuous monitoring of clean water: Drinking water Treated process water Your benefits Versions with

More information

OILTESTER - Deep-frying Oil Tester. Instruction manual

OILTESTER - Deep-frying Oil Tester. Instruction manual OILTESTER - Deep-frying Oil Tester Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Safety and the environment... 4 2.1. About this document... 4 2.2. Ensure safety... 5 2.3. Protecting the

More information

CLEANROOM MONITOR CR3A Network - Installation Instructions

CLEANROOM MONITOR CR3A Network - Installation Instructions CLEANROOM MONITOR CR3A Network - Installation Instructions INTRODUCTION The CR3 Series Cleanroom Monitor, was developed specifically to allow for monitoring of confined spaces with accuracy and reliability.

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 E25.0.01.6C-14 Data logger for temperature as of version V7.0 Operating Manual EASYLOG 40K EASYLOG 40KH EASYLOG 40KH-E300 EASYLOG 40KH-E600 EASYLOG 40KH-GOF EASYLOG 40K EASYLOG 40KH WEEE-Reg.-Nr. DE93889386

More information

21-light Remote Annunciator. Owner s Manual

21-light Remote Annunciator. Owner s Manual 21-light Remote Annunciator Owner s Manual Annunciator Description... Inside Font Cover Detailed Specifications... 1 Environmental Specifications... 1 Power Supply Requirements... 1 Communication With

More information

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications Instruction Manual Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications Manual PCT-3000 Revision 9/2014 The PCT-3000 series control console incorporates

More information

TEIS Induction Module. 4 cooking zones hob with TAP user interface

TEIS Induction Module. 4 cooking zones hob with TAP user interface TEIS Induction Module 4 cooking zones hob with TAP user interface Introduction New development of a 3 zones or 4 zones Induction hob (for 3 zones the 140mm zone is not used) TEIS induction module = The

More information

ProSens Dust Measurement

ProSens Dust Measurement EN Operating Instructions ProSens Dust Measurement SWR engineering Messtechnik GmbH CONTENTS Page 1. Introduction.......................................................................... 3 1.1 Safety............................................................................

More information

FIRECLASS Networking. Addressable Fire Alarm Control Panels From Software version 21. Product Application and Design Information FC-A-FCNET-A

FIRECLASS Networking. Addressable Fire Alarm Control Panels From Software version 21. Product Application and Design Information FC-A-FCNET-A FIRECLASS Networking Addressable Fire Alarm Control Panels From Software version 21 Product Application and Design Information FC-A-FCNET-A Doc. version 1 23. March 2012 FIRECLASS. Hillcrest Business Park,

More information

Omniguard 4: Features

Omniguard 4: Features OMNIGUARD 4 The clear choice for monitoring and documenting Vacuum and Pressure in a containment area. Ideal for Asbestos, Lead and Mold abatement and Clean Room monitoring. Omniguard 4: Features The Omniguard

More information

This document hosted by: Instruction Manual MB120 Moisture Analyzer

This document hosted by:  Instruction Manual MB120 Moisture Analyzer This document hosted by: www.oldwillknottscales.com Instruction Manual Moisture Analyzer Table of Contents 1. GETTING TO KNOW YOUR MOISTURE ANALYZER... 1 1.1 Introduction... 1 1.2 Overview of the Moisture

More information