Section. 3: CADD Pump. Updated: June 20, Section Author(s): sstafford & jhval

Size: px
Start display at page:

Download "Section. 3: CADD Pump. Updated: June 20, Section Author(s): sstafford & jhval"

Transcription

1 Section 3: CADD Pump Updated: June 20, 2012 Section Author(s): sstafford & jhval

2 Section 3: CADD Pump 2 Section 3: CADD Pump Field Guide

3 Section Contents CADD Prizm Diagram... 5 CADD Pump Quick Reference... 7 Programming Screens... 7 Alarms & Troubleshooting... 7 History and Recordkeeping Functions... 8 Programming / Setup Sequence... 9 Changing the Cassette CADD Post Test FORM: Continuous Medication Infusions Field Guide Section 3: CADD Pump 3

4 4 Section 3: CADD Pump Field Guide

5 Model 6100 and 6101 Ambulatory Infusion Pump Click here to View the full Clinical Information CADD Prizm Diagram A. Green indicator light: blinks when pump is running B. Amber indictor light: blinks when pump is stopped or alarming C. Polemount bracket recess (back of pump) D. Power jack (for power pack or AC adapter) E. Data in/out jack (for remote dose cord, modem cable or printer cable) F. Battery compartment (side of pump) G. Cassette (part of reservoir or administration set that attaches to pump) H. Air Detector Field Guide Section 3: CADD Pump 5

6 6 Section 3: CADD Pump Field Guide

7 CADD Pump Quick Reference CADD Prizm VIP Model 6100 and 6101 Ambulatory Infusion Pump Programming Screens Reservoir Volume: Volume of fluid in filled reservoir Continuous Rate: Continuous rate of drug delivery in ml/hr Milliliters Given: Displays total amount delivered since date and time shown Air detectors caused multiple errors in the field, so most companies, including QualMed, do not use them on the Prizm pumps. Alarms & Troubleshooting The CADD-Prizm pump signals with audible alarms and messages on the display when a condition requires your attention. In addition to referring to this table, use the? key while the alarm is occurring for information about the alarm and corrective action. Reservoir Volume Low Level of fluid in the reservoir is low. Prepare to install a new reservoir. Reservoir Volume Zero Reservoir volume is 0.0 ml. Press NEXT to stop alarm. Install new reservoir if appropriate. 9-volt Battery Low 9-volt battery is low but pump is operable. Change battery soon. 9-volt Battery Depleted 9-volt battery is depleted and cannot operate pump; install fresh battery immediately. High Pressure Possible blockage downstream (between pump and patient), clamp or kink in fluid path. Find cause and remove blockage to continue. Or, press STOP/START to stop pump and silence alarm; remove blockage and restart pump. Upstream Occlusion (Model 6101 only) Fluid is not flowing from the reservoir to the pump. Check for a kink, a closed clamp or an air bubble in tubing between reservoir and pump. Press STOP/START to silence alarm for two minutes; remove obstruction and restart pump. Error Detected Pump fault occurred. Close tubing clamp, remove pump from service. 3 Beeps Every 5 Minutes; Stopped on Main Screen Pump is stopped. Start pump if appropriate. Field Guide Section 3: CADD Pump 7

8 History and Recordkeeping Functions The pump may be running or stopped to view functions or to print. To View Event Log To Clear Dose Counters To Clear Units Given 1. Press OPTIONS. 2. Press Y or N until Event Log option appears. 3. Press ENTER to display Event Log. 4. Press N to page back through Event Log entries or Y to page forward. 5. When finished, press NEXT. A demand dose must be programmed and the pump must be in LL0 or LL1. 1. Stop the pump. 2. Press NEXT repeatedly until Dose Counters screen appears. 3. Press ENTER to clear and update date and time. Pump must be in LL0 or LL1. 1. Stop the pump. 2. Press NEXT repeatedly until Milliliters Given, Milligrams Given or Micrograms Given screen appears. 3. Press ENTER to clear and update date and time. WARNING: This Quick Reference Card should be used by clinicians only. Do not permit patients to have access to this card, as the information would allow access to all programming and operating functions. Assistance with the CADD-Prizm pump is available 24 hours a day by calling in the USA and Canada. Important to do. This is a common omission and can be frustrating in the field. 8 Section 3: CADD Pump Field Guide

9 Programming / Setup Sequence To go to the next programming screen without changing a displayed setting, press NEXT. During programming, the pump may ask you to confirm certain settings; press Y or N as appropriate. These screens are dependent on other settings and may not appear. Battery must be installed; pump must be stopped and in Lock Level On Main Screen, press NEXT. Enter Reservoir Volume 2. Press Y or N to select desired Reservoir Volume. 3. Press ENTER. Enter Units 4. Press Y or N to select desired units. 5. Pres ENTER. *Enter Concentration 6. Press Y or N to select desired Concentration. 7. Press ENTER. Enter Continuous Rate 8. Press Y or N to select desired Continuous Rate (select the upper limit if the program will be adjusted in LL1). 9. Press ENTER. Enter Demand Dose *Enter Demand Dose Lockout Time *Enter Max Doses Per Hour *Clear Dose Counters Clear Units Given Check Air Detector Status *Insert New Patient Marker Review Program If Operating in LL1 with Upper Limits, Decrease Continuous Rate or Demand Dose 10. Press Y or N to select desired amount of patient activated demand dose (select the upper limit if program will be adjusted in LL1). 11. Press ENTER 12. Press Y or N to select desired time between Demand Doses. 13. Press ENTER. 14. Press Y or N to select desired number of Demand Doses per hour. 15. Press ENTER. 16. Press ENTER to clear the Dose Counters; this updates the date and time markers. 17. Press ENTER to clear the Dose Counters; this updates the date and time markers. 18. If air detector is attached, check the Air Detector status, then press NEXT. 19. Press ENTER to insert a New Patient Marker in the Event Log. 20. Press Y (this clears the previous Extended History). 21. Press NEXT repeatedly to review program. If pump will be operated in LL1 to allow adjustment of continuous rate or demand dose (up to the maximum entered in LL0): 22. Change lock level to LL Press NEXT to go to Continuous Rate or Demand Dose screen. 24. Press N to scroll down to starting value. 25. Press ENTER. Follow steps for attaching cassette, priming fluid path, starting pump. Field Guide Section 3: CADD Pump 9

10 Changing the Cassette To Remove Used Cassette 1. Stop the pump. 2. Close all tubing clamps. 3. Use key to unlock used cassette. 4. Use side of key or coin to unlatch used cassette. 5. Disconnect tubing from patient. 6. Remove and discard used cassette. To Attach New Cassette 1. Close tubing clamp. 2. Insert new cassette hooks into pump s hinge pins. 3. Place pump with cassette on firm surface and push down on top of pump. 4. Use coin or side of key to hatch cassette. 5. Use key to lock cassette. 6. The mark on latching button must line up with solid dot on side of pump Priming the Fluid Path Using the Prime Option 7. Gently twist and pull on cassette to make sure it is attached properly. 8. When Cassette Locked appears, press NEXT. 9. If Reset Reservoir to ml? appears, press Y if appropriate. 10. When Prime Tubing? appears, open tubing clamp and press Y if appropriate. Follow priming instructions on pump s screen. When primed, close tubing clamp. 11. If applicable, thread tubing through Air Detector. 12. If AutoLock not in use/change Lock Level from LL0? appears, press Y. Change the lock level to LL1 or LL2 (see To Change Lock Level). 13. When Start the Pump? appears, connect new tubing to patient s access site, open tubing clamp, and press Y. The priming feature is available either: a) when you attach a cassette (above), or b) in the Options Menu. To access the Prime option in the Options Menu: Pump must be in LL0 or LL1. 1. Stop the pump 2. Press OPTIONS. 3. Prime should be displayed; press ENTER. 4. Follow priming instructions on pump screen. 10 Section 3: CADD Pump Field Guide

11 To Reset Reservoir Volume Without Changing Cassette 1. Stop the pump. 2. Press NEXT to display Reservoir Volume screen. 3. Press ENTER. 4. If Message appears asking you to reset the Reservoir Volume, press Y. Field Guide Section 3: CADD Pump 11

12 12 Section 3: CADD Pump Field Guide

13 CADD Post Test 1) The code to change/clear settings on the CADD pump is. The pump needs to be running or stopped (circle one) to perform this task. 2) The code to administer a clinician bolus on the CADD pump is. The pump needs to be running or stopped (circle one) to perform this task. 3) How do you ensure the concentration of the medicine is correct before administering a clinician bolus via the CADD pump? 4) What lock level should the CADD pump be in for narcotic administration? 5) A nurse is working on a Saturday and has been called to replace a foley catheter on a patient who has a CADD pump running. Does the nurse need to clear the pump (boluses and total amount) if the patient s pain is well-controlled and the only reason for the visit was to change the foley catheter? Yes or No (Circle correct answer) 6) Please check the instances when a nurse should double-check the CADD settings with another RN/infusion pharmacist or medical director and chart to that effect. When: increasing an oral medicine with a doctor s order. changing a narcotic cassette and not changing any settings. changing any settings on a CADD pump. starting nebulizers. changing any settings on an intrathecal pump. 7) How often should a CADD pump be cleared (bolus and total amount)? 8) What is the maximum amount of time a bag or cassette can hang before being changed? 9) How many extra bags/cassettes should be in a pt s home at all times? 10) How often should you change the battery? 11) What should you do if a pt with a CADD pump dies? 12) When do you start a new infusion flowsheet? 13) How do you prime a line using the CADD pump? Does the pump need to be running or stopped (circle one) to perform this task. Field Guide Section 3: CADD Pump 13

14 14 Section 3: CADD Pump Field Guide

15 FORM: Continuous Medication Infusions Forms can be found in the Forms Room located on the 7 th floor Field Guide Section 3: CADD Pump 15

16 16 Section 3: CADD Pump Field Guide

Description of the CADD-Prizm VIP Pump

Description of the CADD-Prizm VIP Pump Basic Operations General Description and Basic Operations Section 1 Description of the CADD-Prizm VIP Pump The Deltec CADD-Prizm ambulatory drug delivery pump provides measured drug therapy to patients

More information

Clinician Information

Clinician Information Clinician Information CADD-Prizm VIP Ambulatory Infusion Pump Model 6100 and 6101 This booklet provides guidelines for the use, care and maintenance of the CADD-Prizm VIP Ambulatory Infusion Pump. The

More information

Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery.

Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery. Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery. The CADD-Prizm VIP (Variable Infusion Profile) pump is designed for A NEW CADD PUMP BENCHMARK FOR AMBULATORY INFUSION CARE. We

More information

CADD-Prizm PCS II Ambulatory Infusion Pump Model Technical Manual

CADD-Prizm PCS II Ambulatory Infusion Pump Model Technical Manual CADD-Prizm PCS II Ambulatory Infusion Pump Model 6101 Technical Manual 1 Table of Contents 1. Introduction... 1 Limited Warranty...1 Exposure to Radiation or Magnetic Resonance Imaging (MRI)...1 2. CADD

More information

Administration of AMBISOME Using the GemStar Pump

Administration of AMBISOME Using the GemStar Pump 1 1-800-862-2731 Administration of AMBISOME Using the GemStar Pump Drug Name: Volume and Rate: Schedule: FLUSHING: Dextrose Ambisome Dextrose Heparin KEY POINTS: 1. Always wash your hands with an antibacterial

More information

Syringe Pump. User Manual

Syringe Pump. User Manual Syringe Pump User Manual Specifications Length Height Width Weight Waterproof Rating Battery 30 cm 13 cm 12.5 cm 1.8 kg IPX3 Rechargeable Li Polymer 7.4 V 1900 mah 6 Hour Battery Life Running 3 Hours When

More information

HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION

HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION HOMEPUMP ECLIPSE* Ambulatory Infusion System PATIENT INFORMATION IMPORTANT PATIENT INFORMATION Patient name Medication (name and dose) Emergency contact phone numbers Health care provider contact Information

More information

Alaris Syringe Module v9

Alaris Syringe Module v9 Cool Gray 7U Text- PMS 294U + QUICK REFERENCE GUIDE Operator Precautions: To ensure proper operation of the Alaris System (formerly known as Medley System ), user must be familiar with related features,

More information

Home Choice cycler for Peritoneal Dialysis: Set up, Disconnect, & Disposal Updated 2015

Home Choice cycler for Peritoneal Dialysis: Set up, Disconnect, & Disposal Updated 2015 Home Choice cycler for Peritoneal Dialysis: Set up, Disconnect, & Disposal Updated 2015 1 Table of Contents Table of Contents... 2 Home Choice Cycler for Dialysis Quick Guide... 3 Have questions not answered

More information

Sapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design

Sapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design Sapphire TM Ambulatory Infusion System Bring Home The Future Of Infusion Care > Intuitive, easy-to-use touch screen design > Rugged, durable construction for home care infusion needs > Engineered to provide

More information

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10 OPERATOR S MANUAL CONTENTS I. INTRODUCTION... pg 2 FRONT HOUSING PARTS LOCATION (Figure 1)... pg 4 REAR HOUSING PARTS LOCATION (Figure 2)... pg 5 II. CONTROL PANEL... pg 6 KEYPAD (Figure 3)... pg 6 III.

More information

Table of Contents. kangaroo 924. English. General Information... Operating Procedures... Battery... Beginning Administration...

Table of Contents. kangaroo 924. English. General Information... Operating Procedures... Battery... Beginning Administration... Table of Contents English SECTION I General Information... SECTION II Operating Procedures... Battery... Beginning Administration... SECTION III Alarms... SECTION IV Cleaning... Important User Information...

More information

Alaris Syringe Module v9

Alaris Syringe Module v9 Cool Gray 7 U Text - PMS 294U QUICK REFERENCE GUIDE WARNINGS: To ensure proper operation of the Alaris System, you must be familiar with related features, disposables, administration sets, set-up, and

More information

Enteral Pump Instructions for Use To Assemble When the low

Enteral Pump Instructions for Use To Assemble When the low Enteral Pump Instructions for Use Before initial use on battery power and after extended storage periods, the pump must be plugged into an AC power source for a minimum of 12 hours (pump may be operated

More information

Mini Rythmic PN+ Set-up Guide

Mini Rythmic PN+ Set-up Guide Mini Rythmic PN+ Set-up Guide The only dedicated portable Home Parenteral Nutrition system to meet all the requirements of both adult and paediatric patients. MODE - Enter Code 0112 Volume Over Time Intermittent

More information

Alaris Syringe module FAQs

Alaris Syringe module FAQs Alaris Syringe module FAQs 1. Is this the first syringe pump from CareFusion? No, Alaris products from CareFusion have a longstanding tradition of building syringe infusion pumps, going back almost 20

More information

ACTIV.A.C. User Manual THERAPY SYSTEM. For Patients and Clinicians. Rx Only

ACTIV.A.C. User Manual THERAPY SYSTEM. For Patients and Clinicians. Rx Only ACTIV.A.C. THERAPY SYSTEM User Manual For Patients and Clinicians Rx Only DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY KCI HEREBY DISCLAIMS ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION

More information

OPERATOR S MANUAL. The Belmont Rapid Infuser, FMS Boston Road Billerica, MA 01821, USA US/Canada Worldwide

OPERATOR S MANUAL. The Belmont Rapid Infuser, FMS Boston Road Billerica, MA 01821, USA US/Canada Worldwide OPERATOR S MANUAL The Belmont Rapid Infuser, FMS2000 780 Boston Road Billerica, MA 01821, USA 866-663-0212 US/Canada 978-663-0212 Worldwide The Belmont RAPID INFUSER, FMS2000 OPERATOR S MANUAL Belmont

More information

VENTURI Compact 300 NPWT Systems Choice of Therapy Modes Advanced Predictive Low Vacuum Alarm

VENTURI Compact 300 NPWT Systems Choice of Therapy Modes Advanced Predictive Low Vacuum Alarm VENTURI Compact 300 NPWT Systems Negative Pressure Wound Therapy (NPWT) is a mechanical wound care treatment that uses controlled negative pressure to assist and accelerate wound healing. VENTURI Compact

More information

Duodopa. Important Risk Minimisation Information Patient Pocket Guide EDUCATIONAL MATERIAL FOR RISK MINIMISATION

Duodopa. Important Risk Minimisation Information Patient Pocket Guide EDUCATIONAL MATERIAL FOR RISK MINIMISATION Duodopa Important Risk Minimisation Information Patient Pocket Guide Index Duodopa Patient Pocket Guide Introduction to Duodopa.............................. 1 The Duodopa System................................

More information

v.a.c.simplicity TM Quick Reference Guide Power Connection Canister Battery Charging Light Control Panel

v.a.c.simplicity TM Quick Reference Guide Power Connection Canister Battery Charging Light Control Panel v.a.c.simplicity TM Quick Reference Guide Power Connection Battery Charging Light Control Panel Canister Please consult the user manual for more information. Important Safety Information Accompanies this

More information

TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP

TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP TPCA TM PATIENT CONTROLLED ANALGESIA SYRINGE PUMP The TPCA Ambulatory Patient Controlled (PCA) Syringe Pump with Lockbox has a library of protocols offering MediGuard ; a Dose Error Reduction System (DERS)

More information

Thopaz+ Quick Reference Guide. Precious life Progressive care

Thopaz+ Quick Reference Guide. Precious life Progressive care Thopaz+ Quick Reference Guide Precious life Progressive care How to prepare Thopaz+ for use single double 0.3 l 0.8 l 2 l Select single or double (small, medium or large) and ONE of the following canisters

More information

M E D I C A L D E V I C E S

M E D I C A L D E V I C E S Intuitive Generation M E D I C A L D E V I C E S Follow your intuition... Ease of use allows more time for your patients To meet the demands of a changing clinical environment, Injectomat Agilia has been

More information

Cl inician Training Packet

Cl inician Training Packet Cl inician Training Packet VOCSN Clinician Training Checklist Clinician Training Packet Photocopy and complete this checklist as a record. Provide a hands-on demonstration of each item, and ensure the

More information

T34TM AMBULATORY SYRINGE PUMP

T34TM AMBULATORY SYRINGE PUMP T34TM AMBULATORY SYRINGE PUMP The T34 Ambulatory Syringe Pump is the safety standard in ambulatory syringe pumps, offering a comprehensive range of features designed to maximise safe and simple administration

More information

bibag System User s Guide

bibag System User s Guide 2008K@home bibag System User s Guide 2008K@home bibag System User s Guide For software version 4.30 and greater Copyright 2010-2014, Fresenius USA, Inc. All Rights Reserved This document contains proprietary

More information

Rocket Quick Guide. Rocket PSU Portable Suction Unit

Rocket Quick Guide. Rocket PSU Portable Suction Unit Rocket Quick Guide Rocket PSU Portable Suction Unit Contents Changing the PSU and changing the clock time...3 Attaching the suction unit to the chest drain bottle...4-5 Operating instructions...6-8 Changing

More information

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE SILENCE Directions for Use Alaris System Supports Guardrails Suite MX (v8) August 2005 ALARM MONITOR STANDBY ALARM INFUSE STANDBY ALARM INFUSE STANDBY %SpO 2 RATE (ml/h) RATE (ml/h) PULSE (BPM) OPTIONS

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

Syringe Pump Caddo 15B Learning Material Ver. 1.1 An ISO 9001:2008 company 94, Electronic Complex, Pardesipura Indore - 452 010 India Tel: +91-731 4211100 Fax: +91-731-2555643 e mail: info@scientech.bz

More information

Blood/Fluid Warmer. The blood/fluid warmer is used to provide safe and effective warming of blood, blood products, and liquids.

Blood/Fluid Warmer. The blood/fluid warmer is used to provide safe and effective warming of blood, blood products, and liquids. Covenant Health Compassiara1e rote led by Catholic m laes Blood/Fluid Warmer Neonatal Nursery Policy & Procedures Manual Policy Group: Intravenous Approved by: Gail Cameron Director, Maternal, Neonatal

More information

Smoke Detector and Siren. Manual

Smoke Detector and Siren. Manual Smoke Detector and Siren Manual 004001 Smoke Detector and Siren Manual Quick Start... 2 Product Description... 2 Installation Guidelines... 3 Behavior within the Z-Wave Network... 3 Operating the Device...

More information

Wireless Central Monitoring Unit

Wireless Central Monitoring Unit 433-CMU-40/60 Instructions Wireless Central Monitoring Unit 1 3 4 10 21 23 24 Caregiver Pocket Pager CordLess Bed and Chair Sensor Pads Wireless Motion Sensor CordLess Floor Mats Wireless Nurse Call Button

More information

Bravo. Troubleshooting Guide

Bravo. Troubleshooting Guide Bravo Troubleshooting Guide Table of Contents Table of Contents 2 Bravo Alarm Code by Alarm Code Indication 3 Alarm Codes A022, A023, A024, E020 & E021 Door Locking Problems 4-9 Alarm Code E010 10 Alarm

More information

Original Perfusor & Infusomat

Original Perfusor & Infusomat Original Perfusor & Infusomat Your solution for every application System capable, user-friendly, reliable. fm system workstations The entire spectrum of fluid management One system for all requirements

More information

UPS BATTERY TEST. Testing the UPS Battery (Verification Test) Locking models: 13/19/22-USGF, 13/19/22- USGR, 43/44-UDGR, 43/44-UDGF

UPS BATTERY TEST. Testing the UPS Battery (Verification Test) Locking models: 13/19/22-USGF, 13/19/22- USGR, 43/44-UDGR, 43/44-UDGF 30 Armstrong Ave. Georgetown. Ont, L7G 4R9, Canada Tel: 800 800 5706 or 905 702 1441 Fax: 905 702 1442 www.minusforty.com SUBJECT: MODELS: UPS BATTERY TEST Testing the UPS Battery (Verification Test) Locking

More information

PWM. Solar Charge controller with Ethernet. Solar Smart PWM 20Amp. Hardware Description : Release : 19 June 2014

PWM. Solar Charge controller with Ethernet. Solar Smart PWM 20Amp. Hardware Description : Release : 19 June 2014 Solar Charge controller with Ethernet Release : 19 June 2014 Hardware Version : Version 1 Firmware version 1 PC Application Software : Version 1.0.0.0 Hardware Description : The Solar Smart regulator was

More information

(12) Patent Application Publication (10) Pub. No.: US 2014/ A1

(12) Patent Application Publication (10) Pub. No.: US 2014/ A1 (19) United States US 20140O394.46A1 (12) Patent Application Publication (10) Pub. No.: US 2014/0039446 A1 Day (43) Pub. Date: Feb. 6, 2014 (54) (71) (72) (21) (22) (60) PATIENT CARE SYSTEM FOR CRITICAL

More information

VICTORIAN RESPIRATORY SUPPORT SERVICE AUSTIN HEALTH

VICTORIAN RESPIRATORY SUPPORT SERVICE AUSTIN HEALTH VICTORIAN RESPIRATORY SUPPORT SERVICE AUSTIN HEALTH GUIDELINES FOR THE OPERATION OF ASTRAL 100/150 SIZE: Width: 21.5cm Height: 9.3cm Depth: 28.5cm Weight:3.2kg The ResMed Astral is a portable positive

More information

Smart-Wall TV Kit - Installation & Operation Instructions -

Smart-Wall TV Kit - Installation & Operation Instructions - If any parts are missing or damaged, contact your Dealer before starting installation. DO NOT install a damaged kit. Be sure to also reference the appliance Installers Guide. Approvals The flexible duct

More information

Tube Feeding by Pump UHN. Information for patients and families. Read this book to learn:

Tube Feeding by Pump UHN. Information for patients and families. Read this book to learn: Tube Feeding by Pump UHN Information for patients and families Read this book to learn: how to use your feeding tube how to keep it clean common problems what you can do about them Please visit the UHN

More information

Infusion Pump Caddo 16B Learning Material Ver. 1.1 An ISO 9001:2008 company 94, Electronic Complex, Pardesipura Indore - 452 010 India Tel: +91-731 4211100 Fax: +91-731-2555643 e mail: info@scientech.bz

More information

Directions for Use Alaris System (with PC Unit, Model 8000)

Directions for Use Alaris System (with PC Unit, Model 8000) ALARM Directions for Use Alaris System (with PC Unit, Model 8000) Supports Guardrails Suite MX with Guardrails Point-Of-Care software and v9 Operating System software. February 2007 ALARM MONITOR STANDBY

More information

Directions for Use. Alaris System (with Alaris PC Unit, Model 8015) September 2009

Directions for Use. Alaris System (with Alaris PC Unit, Model 8015) September 2009 Directions for Use Alaris System (with Alaris PC Unit, Model 8015) Supports Guardrails Suite MX with Guardrails Point-Of-Care software and v9 Operating System software. September 2009 ALARM MONITOR STANDBY

More information

Ventricular Assist System

Ventricular Assist System HeartWare Ventricular Assist System Patient Manual CAUTION: Investigational device. Limited by United States law to investigational use. Unless otherwise noted, designates a registered trademark of HeartWare,

More information

Oxygen Concentrator Instruction Booklet

Oxygen Concentrator Instruction Booklet 4701 N Superior Dr Muncie, IN 47303 Phone 765-282-1000; Fax 765-286-3351 Store Hours: 8:30am 5:00pm Oxygen Concentrator Instruction Booklet No Smoking Signs Should Always Be Posted in a Visible Location

More information

Hi-Tech, Low Cost Stanley Solutions. Product Brochure

Hi-Tech, Low Cost Stanley Solutions. Product Brochure Hi-Tech, Low Cost Stanley Solutions Product Brochure System Overview UMP - the leader in fall monitoring systems Innovative benefits, economical prices UMP offers the largest selection of sensor pads on

More information

Quick-Start Guide Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input Models and

Quick-Start Guide Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input Models and Quick-Start Guide Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input THE STANDARD IN PRECISION MEASUREMENT 1 Getting Started: 7 6 2 1 Front 5 4 3 1) RUN/STOP Button

More information

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

ZX1e ZX2e ZX5e. Document No Issue 01 user manual ZX1e ZX2e ZX5e Document No. 996-130 Issue 01 user manual MORLEY-IAS ZX2E/ZX5E Fire Alarm Control Panels Table of Contents 1 INTRODUCTION... 4 1.1 NOTICE... 4 1.2 WARNINGS AND CAUTIONS... 4 1.3 NATIONAL

More information

AutoSet AutoSet for Her Elite Escape Auto Escape. H5i H5i for Her. Welcome Guide. English. POSITIVE AIRWAY PRESSURE DEVICEs HEATED HUMIDIFIER

AutoSet AutoSet for Her Elite Escape Auto Escape. H5i H5i for Her. Welcome Guide. English. POSITIVE AIRWAY PRESSURE DEVICEs HEATED HUMIDIFIER AutoSet AutoSet for Her Elite Escape Auto Escape POSITIVE AIRWAY PRESSURE DEVICEs H5i H5i for Her HEATED HUMIDIFIER Welcome Guide English Welcome Thank you for choosing the S9 Series or H5i. Before operating

More information

Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN Rev 1

Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN Rev 1 Covidien Articulating Absorba Tack Foil Pouch Leak Tester Owner s Manual IPE SN 12497 Rev 1 MANUFACTURER: INNOVATIVE PRODUCTS & EQUIPMENT, INC. 5 PROGRESS AVE TYNGSBORO, MASSACHUSETTS 01879 U.S.A Revision

More information

Rev C May GE Interlogix. Part No: R. CareGard. User Guide

Rev C May GE Interlogix.   Part No: R. CareGard. User Guide g 466-1936 Rev C May 2003 GE Interlogix www.ge-interlogix.com Part No: 60-883-95R CareGard User Guide FCC Notices FCC Part 15 Information to the User Changes or modifications not expressly approved by

More information

Introduction. OM-THA2

Introduction. OM-THA2 Introduction. The OM-THA2 is a multi-function product that monitors Temperature, Humidity and Dew Point, provides alarms for out of range conditions, and continuously logs data. It consists of a base unit

More information

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved BRAVO AUTOCLAVES Installation Notes BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved For questions, technical support, service and inquires, contact: Manufactured

More information

Watchguard WGAP864 User Manual

Watchguard WGAP864 User Manual Watchguard WGAP864 User Manual v1.0 Issued September 2016 1 2 Table of Contents Glossary... 5 1. Introduction to your Watchguard WGAP864... 6 2. Before Operating your Alarm System... 6 3. Understanding

More information

Portable Air Conditioner Instruction Manual

Portable Air Conditioner Instruction Manual Portable Air Conditioner Instruction Manual WA-1240AE1: 12,000BTU (cooling only) WA-1240H1: 12,000BTU (cooling & heating) WA-1420E1: 14,000BTU (cooling only) WA-1420H1: 14,000BTU (cooling & heating) Thank

More information

S9 Escape. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE

S9 Escape. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE S9 Escape POSITIVE AIRWAY PRESSURE DEVICE Welcome Guide English Welcome Thank you for choosing the S9 Escape. Before operating this device, please read the entire Welcome and Information Guides. DC power

More information

VPAP COPD. H5i. Welcome Guide. Making quality of care easy. Respiratory Care Solutions. English NONINVASIVE VENTILATOR HEATED HUMIDIFIER

VPAP COPD. H5i. Welcome Guide. Making quality of care easy. Respiratory Care Solutions. English NONINVASIVE VENTILATOR HEATED HUMIDIFIER VPAP COPD NONINVASIVE VENTILATOR H5i HEATED HUMIDIFIER Welcome Guide English Respiratory Care Solutions Making quality of care easy Respiratory Care Solutions Making quality of care easy Welcome Thank

More information

U ser's Guide PC6010

U ser's Guide PC6010 User's Guide PC6010 Quick Reference Guide This manual is for Basic and Advanced users. Each of these types of user can access a different set of functions. The and symbols next to the title of each procedure

More information

MULTI REACH Stick Vacuum

MULTI REACH Stick Vacuum MULTI REACH Stick Vacuum USER GUIDE 2151 SERIES IMPORTANT SAFETY INSTRUCTIONS IMPORTANT: READ ALL INSTRUCTIONS BEFORE USE. RETAIN INSTRUCTIONS FOR FUTURE REFERENCE. Do not use in rain or leave outdoors

More information

Installation Instructions EFS Hand Washing Station

Installation Instructions EFS Hand Washing Station Installation Instructions EFS Hand Washing Station Both water and soap are sequenced and dispensed from outlet. Override button allows for non potable, water-only dispensing. Lighted icons indicate the

More information

User Guide. HTM Nurse Call. In the event of requiring assistance, please contact your Installer/Supplier.

User Guide. HTM Nurse Call. In the event of requiring assistance, please contact your Installer/Supplier. HTM 08-03 Nurse Call User Guide In the event of requiring assistance, please contact your Installer/Supplier. Alarm Radio Monitoring Ltd Brunel Road, Leominster, Herefordshire. HR6 0LX tel: 01568 610 016

More information

USER MANUAL For Clinicians. Rx Only

USER MANUAL For Clinicians. Rx Only USER MANUAL For Clinicians Do not discard. Please retain this user manual for future reference. For additional copies, in the US, visit www.acelity.com, www.veraflo.com and www. vaculta.com or contact

More information

THANK YOU FOR VOTING TEXECOM THANK YOU FOR VOTING TEXECOM USER GUIDE. Security Control Panels 1 INS167

THANK YOU FOR VOTING TEXECOM THANK YOU FOR VOTING TEXECOM USER GUIDE. Security Control Panels 1 INS167 Veritas User Guide Quick Reference Guide THANK YOU FOR VOTING TEXECOM THANK YOU FOR VOTING TEXECOM USER GUIDE Security Control Panels 1 INS167 Overview Premier Series User Guide 1. Overview Introduction

More information

Version 1.03 January-2002 USER S MANUAL

Version 1.03 January-2002 USER S MANUAL Version 1.03 January-2002 1 USER S MANUAL 2 Version 1.03 January-2002 System Details CUSTOMER:...... PHONE:... FAX:... INSTALLED BY:...... PHONE:... FAX:... MAINTENANCE & SERVICE:...... PHONE:... FAX:...

More information

Syringe Pump. Syringe Pump. Product Model: ZD-50C6 Syringe Pump. ZD-50C6-H Syringe Pump

Syringe Pump. Syringe Pump. Product Model: ZD-50C6 Syringe Pump. ZD-50C6-H Syringe Pump Medical Group Syringe Pump Syringe Pump Purge/Bolus ZD-50C6 Syringe Pump Infusion Master infusion tube disconnect(option). ZD-50C6-H Syringe Pump Infusion Master Infusion tube disconnect(option). Atmosphere

More information

THA2 TEMPERATURE/ HUMIDITY/ DEWPOINT ALARM WITH LOGGING CAPABILITY. 99 Washington Street Melrose, MA Phone Toll Free

THA2 TEMPERATURE/ HUMIDITY/ DEWPOINT ALARM WITH LOGGING CAPABILITY. 99 Washington Street Melrose, MA Phone Toll Free 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com THA2 TEMPERATURE/ HUMIDITY/ DEWPOINT ALARM WITH LOGGING CAPABILITY Introduction.

More information

Temporary wireless battery operated fire system from Channel Safety Systems

Temporary wireless battery operated fire system from Channel Safety Systems Temporary wireless battery operated fire system from Channel Safety Systems CHANNEL SAFETY SYSTEMS t: 0845 884 7000 Petersfield Business Park f: 0845 884 6000 Bedford Road Petersfield Hampshire e: sales@channelsafety.co.uk

More information

TBR 40 1 GD Blood Refrigerator Instruction Manual. for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD. (40 x 500ml Blood Bag Capacity)

TBR 40 1 GD Blood Refrigerator Instruction Manual. for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD. (40 x 500ml Blood Bag Capacity) for BLOOD STORAGE REFRIGERATORS MODEL TBR 40 1 GD (40 x 500ml Blood Bag Capacity) REVISION 10/10/07 PLEASE READ CAREFULLY BEFORE OPERATION Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 9

More information

INSTRUCTIONS CONTROL PRODUCTS INITIAL SET-UP. Innovative Technologies in Custom Electronic Design & Manufacturing

INSTRUCTIONS CONTROL PRODUCTS INITIAL SET-UP. Innovative Technologies in Custom Electronic Design & Manufacturing INSTRUCTIONS CONTROL PRODUCTS Innovative Technologies in Custom Electronic Design & Manufacturing Thank you for purchasing the minialarm Compact Security System. The minialarm is a self contained security

More information

Testing the System. Battery Test. Dialer Test. Fire Drill Test (Code + [#] + 69) One-Man Fire Walk-Test (Code + [#] + 68)

Testing the System. Battery Test. Dialer Test. Fire Drill Test (Code + [#] + 69) One-Man Fire Walk-Test (Code + [#] + 68) F A 1 7 0 0 c Testing the System Battery Test When AC power is present, the FA1700C runs a brief battery test every 60 seconds to determine if there is a battery connected, and runs an extended battery

More information

TV2 Room Pressure Monitor Quick Start Guide

TV2 Room Pressure Monitor Quick Start Guide TV2 Room Pressure Monitor Quick Start Guide 1. The Care and Use guide for your TV2 Room Pressure Monitor can be downloaded here: www.e2di.com/usersguides.html Please print out the Users Guide. Although

More information

7. Troubleshooting. Description and Corrective Action. Low Pressure

7. Troubleshooting. Description and Corrective Action. Low Pressure 7. Troubleshooting On detection of a fault condition, the pump provides a visual-only warning followed by an audible and visual alarm if the fault is not cleared. Warning onditions The visual-only warning

More information

Automatic Phone-Out Home Monitoring Systems

Automatic Phone-Out Home Monitoring Systems Automatic Phone-Out Home Monitoring Systems Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are intended to monitor homes, cabins, and other premises

More information

Added password for IP setup page : Password must be in IP format!

Added password for IP setup page : Password must be in IP format! NETWORK POWER MONITOR Release : 21 August 2014 Hardware Version : Version 7 Firmware version 1.00 PC Application Software : Version (latest)...2 Added password for IP setup page : Password must be in IP

More information

SERVICE MANUAL WASHING MACHINE

SERVICE MANUAL WASHING MACHINE Website:http://www.LGEservice.com E-mail:http://www.LGEservice.com/techsup.html WASHING MACHINE SERVICE MANUAL! CAUTION READ THIS MANUAL CAREFULLY TO DIAGSE PROBLEMS CORRECTLY BEFORE SERVICING THE UNIT.

More information

Achieva. Ve nt i l at or with Flow Acceleration Limit and Apnea Backup. User s Manual

Achieva. Ve nt i l at or with Flow Acceleration Limit and Apnea Backup. User s Manual Achieva User s Manual Ve nt i l at or with Flow Acceleration Limit and Apnea Backup 840664790 Ventilator System Service Manual 4-070089-00 Rev. A (10/98) Achieva, Nellcor Puritan Bennett, and Puritan-Bennett

More information

MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE

MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE MODEL 5800A CARBON MONOXIDE (CO) MONITOR AND ALARM SAMPLE-DRAWING TYPE 11//07 A153-021500-1 INSTRUCTION MANUAL - MODEL 5800A CO ALARM I INTRODUCTION This MST Model 5800A Carbon Monoxide (CO) Alarm is a

More information

MULTI Hand Vacuum USER GUIDE 1985 SERIES

MULTI Hand Vacuum USER GUIDE 1985 SERIES MULTI Hand Vacuum USER GUIDE 1985 SERIES IMPORTANT SAFETY INSTRUCTIONS IMPORTANT: READ ALL INSTRUCTIONS BEFORE USE. RETAIN INSTRUCTIONS FOR FUTURE REFERENCE. Do not use in rain or leave outdoors while

More information

DI220 Using the Optional Docking Station The maintenance and operating functions of the Micro IV can be automated with the Micro IV Docking Station.

DI220 Using the Optional Docking Station The maintenance and operating functions of the Micro IV can be automated with the Micro IV Docking Station. DI220 Using the Optional Docking Station The maintenance and operating functions of the Micro IV can be automated with the Micro IV Docking Station. The following steps outline the correct sequence to

More information

VPAP Auto VPAP ST VPAP S. H5i. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE HEATED HUMIDIFIER

VPAP Auto VPAP ST VPAP S. H5i. Welcome Guide. English POSITIVE AIRWAY PRESSURE DEVICE HEATED HUMIDIFIER VPAP Auto VPAP ST VPAP S POSITIVE AIRWAY PRESSURE DEVICE H5i HEATED HUMIDIFIER Welcome Guide English Welcome Thank you for choosing the S9 VPAP Series or H5i. Before operating these devices, please read

More information

MINIMED 670G SYSTEM CHANGING THE MINIMED MIO INFUSION SET QUICK REFERENCE GUIDE

MINIMED 670G SYSTEM CHANGING THE MINIMED MIO INFUSION SET QUICK REFERENCE GUIDE START HERE Wash your hands. Select Reservoir & Tubing. Press. Select Options. Select New Reservoir. Remove the infusion set you have been using by loosening the adhesive and pulling away from body. Remove

More information

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010

URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010 . 600 N. Field Drive Lake Forest, IL 60045 (224) 212 2000 www.icumed.com 15 June 2017 URGENT: MEDICAL DEVICE CORRECTION Plum 360 Infusion System List Number: 30010 Dear Valued Plum 360 Infusion System

More information

User s Manual. i-warm Fluid Warmer

User s Manual. i-warm Fluid Warmer i-warm Fluid Warmer User s Manual Midmark Corporation 60 Vista Drive, Box 286 Versailles, OH 45380-0286 Customer Service and Technical Service TEL: 1-800-MIDMARK (1-800-643-6275) www.midmark.com i-warm

More information

Quick start guide. English

Quick start guide. English Quick start guide English Read your entire Astral User Guide or Clinical Guide before use. The Astral device 1. Adapter port Can be fitted with single limb adapter, single limb leak adapter or double limb

More information

S9 AutoSet & S9 Elite

S9 AutoSet & S9 Elite S9 AutoSet & S9 Elite POSITIVE AIRWAY PRESSURE DEVICEs Welcome Guide English Welcome Thank you for choosing the S9 AutoSet or S9 Elite. Before operating these devices, please read the entire Welcome and

More information

VPAP Adapt. H5i. Welcome Guide. English. adaptive servo-ventilator

VPAP Adapt. H5i. Welcome Guide. English. adaptive servo-ventilator VPAP Adapt adaptive servo-ventilator H5i HEATED HUMIDIFIER Welcome Guide English Welcome Thank you for choosing the VPAP Adapt or H5i. Before operating the device, please read the entire Welcome and Information

More information

MedReady Set-up, Helpful Hints, and Solutions for Special Needs

MedReady Set-up, Helpful Hints, and Solutions for Special Needs MedReady Set-up, Helpful Hints, and Solutions for Special Needs MODELS AVAILABLE: MedReady with no monitoring option 1400 or 1500 MedReady with transmitter to monitor with a medical alert 1400M or 1520

More information

Endoscope Reprocessor Installation Instructions Endoscope Reprocessing System

Endoscope Reprocessor Installation Instructions Endoscope Reprocessing System Endoscope Reprocessor Installation Instructions Endoscope Reprocessing System Introduction The following document outlines the installation procedure for the CER 1 and CER 2 reprocessor(s). Please refer

More information

CER-1 & CER-2 Installation & Testing Instructions

CER-1 & CER-2 Installation & Testing Instructions CER-1 & CER-2 Installation & Testing Instructions i Read all installation instructions completely before installing this device. 1 2 3 INTRODUCTION This document details the installation procedure for

More information

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL SE R IA L N O. M O D E L N O. Tem p. Range 1 - = > 60 o F 2 - = 40-60 o F 3 - = < 40 o F G as Type 1 - R-11 2 - R-12 3

More information

Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc CONTENTS. A. About this manual...2

Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc CONTENTS. A. About this manual...2 Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc Introduction CONTENTS A. About this manual...2 B. Manufacturer s Responsibility....2 C. Warranty.. 2 D. Unpacking and Accessories.2

More information

V.A.C.RX4 USER MANUAL THERAPY SYSTEM. Rx Only

V.A.C.RX4 USER MANUAL THERAPY SYSTEM. Rx Only V.A.C.RX4 THERAPY SYSTEM USER MANUAL Rx Only DISCLAIMER OF WARRANTY AND LIMITATION OF REMEDY KCI HEREBY DISCLAIMS ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANTY OF

More information

Elite 16D Version 16 Zone Controller Arrowhead Alarm Products Ltd. Operating Guide. Proudly Designed and Manufactured in New Zealand

Elite 16D Version 16 Zone Controller Arrowhead Alarm Products Ltd. Operating Guide. Proudly Designed and Manufactured in New Zealand 6 Elite 16D Version 16 Zone Controller Arrowhead Alarm Products Ltd Operating Guide 1 Proudly Designed and Manufactured in New Zealand CONTENTS Page No. INTRODUCTION 3 About your Alarm 3 OPERATING YOUR

More information

Left Ventricular Assist Devices

Left Ventricular Assist Devices Indications Evaluation of patient with implanted Left Ventricular Assist Device. If the patients device is not a HeartMate II contact OHSU consult line. HeartMate II The current LVAD deployed is the HeartMate

More information

Maintenance Manual PC6010. WARNING This manual contains information on limitations regarding product use and function

Maintenance Manual PC6010. WARNING This manual contains information on limitations regarding product use and function WARNING This manual contains information on limitations regarding product use and function and information on the limitations as to liability of the manufacturer. The entire manual should be carefully

More information

JUNO-NET. Fire Alarm Control Panel OPERATION & MAINTENANCE MANUAL

JUNO-NET. Fire Alarm Control Panel OPERATION & MAINTENANCE MANUAL Fire Alarm Control Panel & MANUAL CONTENTS SECTION -. Description of the fire control panel fascia.... Alarm.... Reset the system.... Sound and silence the alarms.... Read the Fire, Fault, Test and Disabled

More information

Ice Lined Refrigerator User Operation & Maintenance manual

Ice Lined Refrigerator User Operation & Maintenance manual Ice Lined Refrigerator User Operation & Maintenance manual www.newmeditech.com 1 Contents Topics Page No. 1.1 About this manual 3 1.2 Warnings 3 1.3 Working Safely 4 1.4 General Safety 4 1.5 Electrical

More information

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide

D8024, D9024, D10024 Analog Fire Alarm Control Panels Programming Guide System Reset Trou ble Silence Ala rm Silence Manual Ala rm ENTER NO YES Letters Numb ers Keyword Radionics System Reset Trouble Silence Alarm Silence Manual Alarm ENTER NO YES Le ters Numbers Keyw ord

More information