INSTRUCTION MANUAL V8F145

Size: px
Start display at page:

Download "INSTRUCTION MANUAL V8F145"

Transcription

1 MEDIZINTECHNIK FÜR TIERÄRZTE INSTRUCTION MANUAL V8F145 MEDIZINTECHNIK FÜR TIERÄRZTE Eltastr Tuttlingen Phone: +49/7461/ Fax: 1 +49/7461/ export@eickemeyer.de

2 INSTRUCTION MANUAL V8F145 Thank you for purchasing the Eickemeyer V8F145 Flexible Veterinary Endoscope. The 145cm endoscope is designed to provide direct visualization. This highly flexible instrument allows for comfortable, safe passage through anatomical curves. However, its flexibility also makes it a delicate instrument requiring proper handling to ensure satisfactory performance. Familiarize yourself with this instrument before using it in clinical procedures. Read the following instructions carefully. This unit is not sterile upon receipt. Follow standard disinfection or sterilization practice before initiating clinical use. DO NOT STEAM STERILIZE THE VETERINARY ENDOSCOPE. WARNING: The user of this instrument should be thoroughly informed with the contents of this manual prior to using the instrument. The operator should also be knowledgeable of the procedure and related hazards availing himself/herself of adequate training in the use of this instrument prior to use. Failure to achieve adequate training and familiarization with the manual contents could result in damage to the instrument and/or injury to the patient and/or user. TABLE OF CONTENTS INTRODUCTION 1.0 Changing the Biopsy Inlet Cap 4.2 Guidelines for Handling the Endoscope MM and VIDEO PHOTOGRAPHY 5.0 Inspecting the Endoscope 1.2 CLEANING PROCEDURES 6.0 OPERATING THE ENDOSCOPE 2.0 Checking for Watertight Integrity 6.1 Preparing the Light Source 2.1 Cleaning the Endoscope 6.2 Connecting the Light Guide, Air/Water Line Cleaning the Accessories 6.3 and Suction Tubing 2.2 Alternate Cleaning Procedure for Field Use 6.4 Eyepiece Adjustment 2.3 DISINFECTING 7.0 Eyecup Installation 2.4 Recommended Cold Disinfecting Solutions 7.1 Deflecting the Distal End 2.5 Disinfecting the Endoscope 7.2 2

3 Suction 2.6 ETO STERILIZATION 8.0 Insufflation 2.7 STORAGE 9.0 Irrigation 2.8 PARTS LIST 10.0 PREPARING FOR USE 3.0 REPAIRS, SERVICE 11.0 Holding the Endoscope 3.1 SPECIFICATIONS 12.0 Procedural Consideration 3.2 TROUBLESHOOTING 13.0 USE OF ACCESSORIES 4.0 WARRANTY 14.0 Biopsy Forceps Introduction The V8F145 endoscope is a highly flexible fiber optic instrument designed to allow atraumatic introduction for internal visual examination in animals. Its slim diameter and longer shaft combine to make this a versatile instrument for many veterinary applications. 1.1 GUIDELINES FOR HANDLING THE ENDOSCOPE The V8F145 is a delicate optical instrument that must be properly handled to ensure satisfactory performance and prevent accidental damage. Light and image are transmitted through thousands of fragile glass fibers that run the length of the flexible insertion shaft. These fibers will break under extreme flexing or bending and cause black spots to appear in the viewing field. Possibility of fiber breakage can be greatly reduced by adhering to the following precautions. DO NOT DEFLECT THE DISTAL BENDING PORTION BY HAND. DO NOT BEND THE INSERTION TUBE OR LIGHT GUIDE SHARPLY. DO NOT FORCE THE INSTRUMENT AGAINST RESISTANCE. DO NOT STORE THE INSTRUMENT IN ITS CARRYING CASE. DO NOT ALLOW THE DISTAL TIP TO COME IN SHARP CONTACT WITH ANY HARD OBJECT. DO NOT ATTEMPT SELF-REPAIR. ALL REPAIRS MUST BE DONE BY AN AUTHORIZED REPRESENTATIVE. 1.2 INSPECTING THE ENDOSCOPE Visually inspect the endoscope for damage or wear before and after every use. Pay particular attention 3

4 to the insertion shaft, observing for cuts, holes or protrusions. Irrigate the operating channel using a 30cc syringe to check for obstructions or damage (Refer to Section 2.8). Check the optics by viewing printed material through the fiberscope at a distance of ½ inch while focusing the eyepiece (Refer to Section 2.3). You should be able to obtain a sharply focused image. Rotate the control knobs to assure that they move freely with no roughness or sticking. Be sure that both brakes are in the FREE position (see Section 3.0, 4C). Operate the control knob brakes to assure that the bending section is stabilized when the brakes are engaged. Check that the knobs will rotate freely when the brakes are released. If any irregularities are found, do not use the instrument. Contact your Vet-Vu customer service representative for service instructions. 2.0 OPERATING THE ENDOSCOPE 2.1 PREPARING THE AIR / LIGHT SOURCE The Vet-Vu 1185U 150-watt light source is recommended for use with the endoscope. Refer to the illuminator instruction manual for complete operating instructions. Set the illuminator on a firm, flat surface. Plug the power cord into the appropriate electrical outlet. 2.2 CONNECTING THE LIGHTGUIDE, AIR / WATER LINE AND SUCTION TUBING Screw lightguide adapter onto lightguide. Insert lightguide into lightguide connection on illuminator. Secure with thumbscrew. Attach color-coded ends of air / water line to matching air / water connections on umbilical. Attach remaining end to water bottle. (Water bottle fitting is designed for self-orientation and cannot be reversed. See Figure 2.) Tighten knurled nut. Attach one end of suction tubing to suction connector on umbilical and attach other end to suction pump. (Suction pump an optional accessory and not shown in Figure 2.) NOTE: To fill or empty water bottle easily, remove bottle from illuminator and unscrew bottle from cap. 2.3 EYEPIECE ADJUSTMENT 4

5 Rotate focus adjustment ring while viewing printed material until image is in the best possible focus. The orientation pointer in the optical image indicates the upward direction of the shaft. 2.4 EYECUP INSTALLATION Stretch molded eyecup over eyepiece of endoscope. Using the rubber eyecup permits more comfortable visualization by eliminating ambient light between the eye and eyepiece. The eyecup will also prevent the eyepiece from scratching eyeglasses. 2.5 DEFLECTION OF THE DISTAL END Rotate large control knob (labeled U/D ) in clockwise direction to turn distal end downward. Rotate knob in counterclockwise direction to produce upward deflection. Turn small control knob (labeled L/R ) clockwise to deflect tip to the right and counterclockwise to deflect to the left. Do not exert excessive force on the deflection control knobs. Such force may cause damage to the endoscope and may cause patient injury. Always release the control knob brakes before attempting to deflect, advance, or withdraw the instrument. 2.6 SUCTION Turn suction pump on and depress red suction button on control handle to suction. Mounting the biopsy inlet cap allows suction with the biopsy forceps in place in the biopsy channel. The cap seals around the forceps for this purpose. 2.7 INSUFFLATION Depress the Power and Air buttons located on the front panel of the 1185U illuminator. The air pump will start and air will begin to flow from the vent on top of the blue air / water valve. To insufflate, cover the air control vent with a finger. DO NOT DEPRESS. To minimize patient discomfort, use insufflation sparingly. Over-insufflation can cause excessive cramping, spasm, and possible patient injury. Remove air by applying intermittent suction while withdrawing the endoscope. 5

6 2.8 IRRIGATION Water feeding and / or lens cleaning is accomplished by depressing the air / water button (illuminator power and air buttons must be depressed for automatic water), or by attaching a filled syringe to the secondary lens inlet and gently injecting. Figure 3 Irrigation through secondary irrigation inlet Combining irrigation and insufflation, allows cleansing of the lens during a procedure, without removing the instrument from the patient. Intermittent use of suction during the lens cleaning process will remove excess air or water. NOTE: To refill water bottle, lift bottle from holder, hold cap and unscrew bottle from cap. Fill bottle, screw bottle into cap. DO NOT TURN CAP. Replace bottle into holder. 3.0 PREPARING FOR USE 1. Prepare air / light source as in Connect lightguide, water bottle, and suction tubing as in Adjust focus eyepiece as in Check operations of deflection control knobs. a. Inspecting Up / Down Controls Release up/down brake. Rotate large knob slowly in clockwise direction and tip should rotate downward. Turn knob slowly in counterclockwise direction to produce upward deflection. b. Inspecting Right /Left Controls Release Left / right brake. Rotate small knob slowly in clockwise direction and tip should bend to the right. Turn the knob slowly counterclockwise to produce deflection to the left. c. Inspecting Functioning of the Articulation Brakes Release articulation brakes and articulate the scope. Engage brake by moving brake control counterclockwise to confirm bending section is stabilized when brake is in place. Knobs and brakes should move smoothly. 6

7 5. Inspect insertion tube visually for any dents, bulges, protrusions, cuts, or other irregularities. Pass your fingers over entire tube and bending section to check for any protruding objects, cuts, or loose coverings, or any other irregularities. 6. Confirm the proper function of the air feed. With air pump on, immerse distal tip of scope in small basin of water. Cover air control button with finger; air should bubble out through air nozzle. 7. Confirm the proper functioning of the water feed. a. Depress air/ water button. Water should flow from water nozzle. Note: At the first depression, it will take a few seconds before water is emitted. b. Use a syringe; inject clear water through water inlet. Water should flow from water nozzle on the distal tip. 8. Confirm proper functioning of the suction. With the scope properly connected and suction pump turned on, again immerse the tip of the scope in a small basin of water. Depress suction button. Water should aspirate through the instrument and into collection bottle of the suction apparatus. 9. Check biopsy channel for obstructions by passing forceps through channel as described in Section 4.1 See section 4.0 for further information on the use of the biopsy forceps and maintenance of the biopsy inlet cap. 3.1 HOLDING THE ENDOSCOPE Hold the control handle in left hand so your thumb, index, and middle fingers can operate the larger control knob easily. The suction button and the air / water control button may be depressed by the index or third finger. The smaller outside knob may be operated entirely with fingers of the left hand, or with thumb and fingers of the right hand. It will occasionally be desirable to operate both knobs simultaneously, necessitating a two handed technique. 3.2 PROCEDURAL CONSIDERATIONS 1. Apply a water-soluble lubricant (i.e. K-Y jelly or equivalent) to distal section of insertion tube. Do not get any lubricant on the objective lens. It will cloud the image. Do not use oils, ointments or other substances having a petroleum base or containing Vaseline. These substances may cause deterioration of the material used in the endoscope construction. 7

8 Keep lumen in view at all times. If loss of orientation occurs, withdraw scope slightly until lumen is once again in view. Red-out, a diffused red image in the eyepiece caused by the scope tip coming in direct contact with the wall, may occur during the procedure. Withdraw the scope slightly until lumen is once more in view. CAUTION: ADVANCING THE SCOPE WHILE THERE IS A RED- OUT COULD RESULT IN PERFORATION. Use insufflation sparingly. Over insufflation can cause excessive cramping, spasm, and possible patient injury. To remove excess air, apply intermittent suction as the endoscope is withdrawn. 3. During the procedure, if the view through the instrument becomes obscured due to contamination of objective lens, it can usually be cleared by a combination of water feed and insufflation. Excess water or air may be removed by simultaneously depressing the suction button. 4. Suctioning may be performed when the biopsy forceps are in place in the channel and the biopsy inlet cap mounted. The cap seals around the forceps for this purpose. 5. Tip deflection is described in Section 2.5. If tip deflection ceases to function properly, stop examination immediately. Release brakes. Return the knobs to their neutral position. Align the U and the L with the control buttons and carefully withdraw the instrument while observing through it. 6. Before withdrawing the endoscope, be sure that the brakes are in the FREE position. 4.0 USE OF ACCESSORIES 4.1 BIOPSY FORCEPS NOTE: Use only Vet-Vu recommended biopsy forceps. Non-compliance will void the warranty. To use biopsy forceps, grip forceps with 2 3 cm protruding away from fingers. Lubricate with water to ease passage. Insert distal end into biopsy inlet (the biopsy inlet cap provides a rubber seal). Pass forceps through channel. Advance using short strokes to avoid kinking. 8

9 NOTE: Easiest passage is obtained when the endoscope tip is not articulated. To operate forceps, insert thumb into forceps handle and hold grip between index and middle fingers. When grip is pushed forward, cups open; when grip is pulled back, cups close. Be sure that the cups are closed before inserting the biopsy forceps into the biopsy inlet and before withdrawing the forceps from the endoscope. If resistance in encountered in passing a forceps or brush through the channel, do not attempt to force them. Aspirate cc s of water through the channel. Clean the biopsy channel with a cleaning brush, and then attempt to insert the forceps again. If resistance to passage continues, contact Vet-Vu for service instructions. 4.2 CHANGING THE BIOPSY INLET CAP Replace biopsy inlet cap when ability to suction with biopsy forceps in place is reduced or when water or air is observed to leak from the biopsy inlet MM AND VIDEO PHOTOGRAPHY The V8F145 endoscope is suitable for both 35mm and video photography. Camera or video adapters are necessary to attach eyepiece of scope to lens of camera. Contact Vet-Vu customer service for information on available adapters and equipment. 6.0 CLEANING PROCEDURES Scrupulous cleaning of the endoscope and its accessories after use greatly contributes to the good working condition of the instrument and to the effectiveness of disinfecting and / or sterilization practices. Several general precautions must be observed. The endoscope must be cleaned immediately after every use. If material is allowed to dry within the instrument, partial or total obstruction of the channels may occur. Do not steam autoclave the endoscope or any of its accessories. 9

10 6.1 CHECKING FOR WATERTIGHT INTEGRITY BEFORE IMMERSING THE ENDOSCOPE, CHECK FOR WATERTIGHT INTEGRITY. Connect the leakage tester to the ETO vent. Remove the biopsy inlet cap. Insufflate leakage tester until gauge reads 160mm HG or until pressure valve sounds. Discontinue insufflation. Wait 10 seconds. If pressure gauge does not drop more than 6mm HG, instrument is watertight and may be immersed. Depress red lever on bottom of leak tester to relieve pressure inside the instrument. If pressure in the gauge does drop, do not immerse the instrument. Contact Vet-Vu customer service representative for instructions. 6.2 CLEANING THE ENDOSCOPE Supplies needed: Large Basin Channel Cleaning Tube Mild Soap Solution* 30cc Syringe Tap Water Cotton-tipped Applicators 70% Isopropyl Alcohol Silicone Lubricant Gauze Pads Protective Gloves Cleaning Brush *Any non-abrasive protein breaking soap solution may be used. DO NOT STEAM AUTOCLAVE THE ENDOSCOPE OR ANY OF ITS ACCESSORIES. 1. Detach eyecup if used. Saturate a gauze pad with mild soap solution. Gently wipe scope handle, shaft, lightguide, eyecup, and tubing with pad. AVOID pulling and stripping motions that may damage the shaft and lightguide sheathing. Rinse shaft thoroughly with clean water. 2. Alternately feed clean water and air several times into the secondary lens irrigation inlet using the irrigation tube and syringe. Use clean water only or the channel may become clogged. 3. Connect endoscope to suction pump (See Section 2.2). Aspirate approximately 100 cc s of soap solution through the suction channel. 4. Remove biopsy inlet cap, suction button, and air / water button by pulling straight up. Place in cleaning solution. NOTE: ONLY IMMERSE AFTER WATERTIGHT INTEGRITY OF THE INSTRUMENTHAS BEEN TESTED AND CONFIRMED (REFER TO SECTION 6.1). 10

11 5. Brush entire suction biopsy channel several times. Always insert wire end of brush first and gently pull through. (See Figures 6 & 7). a. Pass brush through biopsy channel inlet until it emerges at distal end of shaft. b. Pass brush through biopsy channel beginning at distal end of shaft until it passes through handle and appears in suction valve housing. c. Again, pass brush through biopsy channel beginning at distal end of shaft until it emerges from distal end of shaft. d. Pass brush through suction valve housing into umbilical and continue until it emerges from the suction connection of umbilical. 6. Attach channel-cleaning tube to biopsy inlet. Dip free end of channel cleaning tube and shaft into cleaning solution. Aspirate cc of solution by covering suction control housing. Repeat with clear water to rinse. 7. Turn off suction device and remove suction line and channel cleaning tube. 8. Immerse entire instrument in cleaning solution. Gently scrub all external surfaces. 9. Gently clean distal tip of instrument to remove any matter. It may be necessary to clean the air water nozzle with a soft bristled brush to remove any matter lodged in the port. Do not attempt to remove matter from the sprayer with a needle or other sharp object. Damage to the objective end components may occur. 10. Remove the instrument; place in clean water and rinse. 11. Using the cleaning brush, gently wash and rinse all valves. 12. Reconnect suction tube to orange section tube connector on umbilical. While holding control section out of the water, turn on the suction device. Aspirate water for approximately 10 seconds. 13. Remove entire instrument from the water. Continue to aspirate air for approximately 30 seconds to dry channel. Turn off suction device and disconnect the line. 14. Dry all external surfaces of the instrument. 11

12 15. Flush all channels with 70% isopropyl alcohol. 16. Empty water bottle and reconnect air / water lines. Replace air / water valve in handle. With air pump turned on, cover valve to dry airline. Depress air / water valve until all moisture is removed from water line. 17. Turn off light source and disconnect instrument. 18. Carefully clean distal lens and eyepiece using a cotton tip applicator dipped in 70% isopropyl alcohol. Always wipe away from the air / water nozzle. 19. Dry the suction valve and forceps inlet valve and reattach to instrument. A light coating of silicone lubricant may be applied on valve stem to help maintain smooth operation. 6.3 CLEANING THE ACCESSORIES 1. Thoroughly clean biopsy forceps and brushes in cleaning solution with a soft bristled brush. Take particular care to remove all debris such as blood, stool, or mucous from biopsy cups and needle. 2. Rinse thoroughly in clean water. Dry completely. Compressed air may be used. 3. Forceps may be lubricated with mineral oil. Open and close forceps several times after lubricating. 4. Accessories may be disinfected or ETO gas sterilized. 6.4 ALTERNATE CLEANING PROCEDURE FOR FIELD USE NOTE: This procedure is provided as an alternative when suction is unavailable and field conditions do not permit cleaning the endoscope in the manner described above. The following procedure is only a TEMPORARY cleaning measure; the instrument should be thoroughly cleaned as described above as soon as it is returned to the office or hospital. Rinse off insertion tube thoroughly with clear water. Be particularly careful to remove debris from distal tip. Remove forceps inlet valve and replace with cleaning tube. Connect a large syringe to irrigation tube. Flush channel with at least 100 cc of water. Pass channel-cleaning brush along entire length of channel. Flush channel again with another 100 cc of water. Inject 30 cc of clear water through secondary lens inlet. Follow with air. 12

13 7.0 DISINFECTING Thorough mechanical cleaning and drying of the endoscope and accessories are prerequisites to disinfecting. Only those disinfecting solutions recommended by Vet-Vu for use on endoscopes may be used. Exposures to the disinfectant should be limited to minutes maximum. Please read section RECOMMENDED COLD DISINFECTING SOLUTIONS FOR VET-VU ENDOSCOPES. Glutaraldehyde Iodophor Other Banicide Wescodyne Control III Metricide Cidex (NOT CIDEX 7 OF CIDEX PLUS) Wavicide-01 Zorbicide NOTES: For the most current list of disinfecting solutions, contact Vet-Vu customer service. These materials are considered safe to use on all Vet- Vu endoscopes, if used according to manufacturer s instructions for disinfecting. These solutions are very aggressive, and failure to follow the disinfecting cycles precisely, including complete cleaning and rinsing, may result in lens pitting, adhesive degradation or other complications over time. Vet-Vu is not responsible for damage when solutions other than the above are used. Vet-Vu makes no claims regarding the effectiveness of cold disinfecting solutions. Follow the manufacturer s recommendations for DISINFECTING only. DO NOT USE concentrates, heat, and / or extended cycles; any one of these may damage the fiberscope material. Efficacy of disinfectants may vary. Consult the manufacturer for information regarding the specific properties of a given product. 7.2 DISINFECTING THE ENDOSCOPE The following supplies are needed: All Channel Irrigator Disinfecting Solution Clean Tap Water Channel Cleaning Tube Basin Syringe Protective Gloves Water Bottle Remove control buttons by pulling straight up, and caps, and place in disinfectant. Connect the all channel irrigator as follows (Figure 9). 13

14 Situate valve adapter on one leg of the irrigator into suction valve housing. Situate valve adapter on second leg of irrigator into air-water valve housing. (Note: both legs of the allchannel irrigator are identical.) 1. Immerse entire scope in disinfectant. 2. Fill a 30 cc syringe with disinfectant and connect to luer-lock fitting on all channel irrigator. Depress plunger until all disinfectant has expelled from syringe. Repeat procedure until no bubbles come from any of scope channels. 3. Disconnect the all channel irrigator and allow the instrument to remain in disinfecting solution for the recommended period of time. 4. Remove instrument from disinfectant and place in clean water. 5. Reattach all channel irrigator. 6. Reattach the all-channel irrigators. 7. Repeat Step 4, using clean water in place of disinfectant. 8. Remove instrument from water. Fill syringe with air and purge all water from instrument. 9. Rinse outside of endoscope under running water. Remove all channel irrigator while rinsing. 10. Rinse all valves and caps thoroughly. Dry and lubricate with silicone lubricant. 11. Replace air / water button. Connect fiberscope to empty water bottle and turn on air supply. By alternately covering and depressing the air / water button, thoroughly dry the air / water lines. 12. Replace suction button and biopsy inlet cap. Connect endoscope to suction pump and aspirate to completely dry the channel. 13. Rinse all channels with 70% isopropyl alcohol and air dry. 14. Dry outside surfaces of instrument. Use a cottontipped applicator, dry the distal tip, eyepiece and valve housings. 14

15 8.0 ETO STERILIZATION (ETHYLENE OXIDE STERILIZATION ONLY) Thorough mechanical cleaning and drying of the endoscopes and accessories are also prerequisites to sterilization. Both may be gas sterilized following the sterilizing equipment manufacturer s procedures. 1. THOROUGHLY CLEAN INSTRUMENT AS DESCRIBED ABOVE. 2. DRY INSTRUMENT AND CHANNEL. 3. ATTACH ETO VENT CAP TO INSTRUMENT. 4. DO NOT EXCEED 130 F (54 C) OR 10 PSI WHEN STERILIZING INSTRUMENT. 5. REMOVE BIOPSY INLET CAP, SUCTION BUTTON, AND AIR / WATER BUTTON PRIOR TO STERILIZING. STERILIZE THSE COMPONENTS ALONG WITH YOUR ENDOSCOPE. 6. DO NOT AUTOCLAVE THESE INSTRUMENT OR ACCESSORIES. 7. EACH INSTITUTION MUST ESTABLISH THE EFFICACY OF ITS STERILIZATION AND AERATION PROCEDURES. 9.0 Storage The endoscope can be stored on a flat, protected surface in a clean, dry area such as a cabinet, or hung on a wall mounted unit. Never store the endoscope or its accessories in the carrying case PARTS LIST 10.1 COMPONENTS Part No. COMPONENTS Part No. V8F145 Veterinary Endoscope U illuminator Biopsy inlet cap Plastic Syringe 30cc Channel Cleaning Brushes Carrying Case Molded Eyecup Channel Irrigator Channel Cleaning Tube Leakage Tester Water Bottle Umbilical Water Bottle Assembly REPLACEMENT PARTS Part No. EJA Lamps for 1185U (2) AB Vet-Vu Lightguide Adapter ACCESSORIES Refer to current price list REPAIRS, SERVICE 15

16 For information on currently available accessories or specified replacement parts for the endoscope and other products or service of this instrument, contact Vet-Vu customer service. Do not attempt any self-repair of this instrument. For any abnormalities in appearance or function, contact your Vet-Vu customer service department. The warranty on this endoscope does not apply if the endoscope has been misused in any way, or has been altered or repaired by other than an authorized Vet-Vu Service Center. Complete warranty information is defined on page 17 of this manual and is contained on the warranty card packed with this endoscope. Quality assurance standards for the endoscope industry allow for three broken image fibers in a new instrument. AUTHORIZED SERVICE CENTERS VET-VU / SWISS PRECISION PRODUCTS, INC. ONE WATER STREET SPENCER, MA TEL: FAX: For international service, contact one of the following: INSPECTION OPTICS WALTER EICKEMEYER Garston Bridge Eltaster 8 Garsdale, Frome Tuttlingen Somerset, BA11 1RU GERMANY Tel / Fax: Tel: Fax: OPTOMED JORGEN KRUUSE A/S 3 Avenue Du Canada Parc Tech DK 5290 Marslev Bat Alpha Les Ulis Cedex DENMARK FRANCE Tel: Tel: Fax: Fax:

17 12.0 SPECIFICATIONS Size 95 Working Length 145cm Shaft Diameter 8.0mm Distal End Diameter 8.0mm Biopsy Channel 2.5mm End Tip Deflection Field of View Up 180 Depth of Field 3-100mm Down 120 Eyepiece Magnification 35X Right/Left 100 Insufflation Automatic Air Irrigation/Lens Cleaning Automatic or Manual Immersible Yes Photo Capability Automatic or Manual Warranty One Year *Because we continually upgrade out products, specifications are subject to change without notice. Illustrations may be slightly different than actual. symptom possible cause solution Image not clear --Eyepiece not adjusted to user s eyesight Turn focus adjusting ring until fiber pattern is in sharp focus. --Distal objective obscured Rinse lens by injecting water through lens irrigation inlet. --Lenses dirty Remove scope, clean distal objective and eyepiece with cotton swab and alcohol. --Water damage to lens Return to Eickemeyer for evaluation/repair. Inadequate Suction --Suction valve blocked Remove valve, clean valve and housing with cotton swabs, alcohol, lubricate O rings with petroleum jelly and replace. --Rubber valve at forceps inlet is worn Replace forceps inlet valve. 17

18 --Suction channel blocked Return to Eickemeyer for evaluation/repair. --Suction pump not on or tubing disconnected Verify good connections between suction pump and scope. Turn pump on. No Insufflation --Air pump not turned on Turn pump on; test scope for air flow --Air tubing not connected Connect green air tubing to air outlet on illuminator/air pump. --Air nozzle blocked Soak tip of scope in warm soapy water for 2-3 min. Gently inject warm water through green line while covering air valve opening. --Water bottle cap not tightly secured Tighten cap. Cannot irrigate --Water bottle empty Fill water bottle. --Water nozzle blocked Soak tip of scope in warm soapy water for 2-3 min. Gently inject warm water through green line while covering air valve opening. --Water bottle cap not secured tightly Tighten cap. Suction button sticks --Dirty valve stem Remove valve, clean valve and housing with cotton swab and alcohol. Lubricate O rings with petroleum jelly. Replace. Forceps inlet valve --Rubber valve at forceps inlet is worn Replace forceps inlet valve. 18

19 spitting or leaking Insufficient articulation --Loose articulation knob Return to Eickemeyer for evaluation/repair. --Degree of articulation less than specified Return to Eickemeyer for evaluation/repair. Inadequate illumination --Illuminator not turned on. Turn illuminator on. Adjust light intensity. --Damage to light transmitting fibers Return to Eickemeyer for evaluation/repair. Illuminator not working --Not plugged into electrical outlet Plug into grounded outlet --Lamp is burned out Replace lamp according to instructions. --Fuse is burned out Replace fuse according to instructions. Eickemeyer s Service Center is available during normal business hours (Eastern Time) to offer assistance should the need arise. Call WARRANTY For one year from the date of purchase, Eickemeyer will repair or replace this product, at its option, for shipping charges only, if defective in workmanship or material. Return it properly packaged to Eickemeyer. An Extended warranty is available. The extended warranty (up to $250.00) must be purchased within 30 days of purchase date. Contact Eickemeyer customer service for details. THIS WARRANTY DOES NOT APPLY IF THE PRODUCT HAS BEEN MISUSED IN ANY WAY OR HAS BEEN ALTERED OR REPAIRED BY OTHER THAN AUTHORIZED EICKEMEYER SERVICE REPRESENTATIVES. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES IMPLIED OR EXPRESSED. ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. NO ONE IS AUTHORIZED TO MAKE ANY OBLIGATIONS FOR EICKEMEYER NOT IN ACCORDANCE WITH THE ABOVE. EICKEMEYER SHALL UNDER NO CIRCUMSTANCES BE LIABLE FOR SPECIAL, INCIDENTIAL, OR CONSEQUENTIAL DAMAGES FROM ANY ALLEGED NEGLIGENCE, BREACH OF WARRANTY, STRICT LIABILITY OR ANY OTHER THEORY ARISING OUT OF OR RELATING TO THE DESIGN, MANUFACTURE, USE OR HANDLING OF THE PRODUCT. 19

Instructions for Use: Fiberscope and LED Light Source

Instructions for Use: Fiberscope and LED Light Source Instructions for Use: and Brand Name of Product Generic Name of Product Product Code Number(s) Intended Use Range of Applications for Product Key Specifications of Product, FBR-001, FBR-002, FBR-ULB-100,

More information

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II)

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II) THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title. THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XP160 OLYMPUS GIF TYPE 160 OLYMPUS GIF TYPE Q160 OLYMPUS GIF TYPE Q160Z OLYMPUS GIF TYPE 1TQ160 OLYMPUS GIF TYPE XTQ160 EVIS EXERA COLONOVIDEOSCOPE

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT US Endoscope ED-530XT OPERATION MANUAL (Cleaning, Disinfection and Storage) Thank you for purchasing our product. Read this manual carefully before use to avoid unexpected accidents and to take full advantage

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. Flexible Nasopharyngoscope Operator Manual Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. 1. Introduction The flexible nasopharyngoscope is sold in unsterile condition as a reusable

More information

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see  for demo video Keeping you insafehands to avoid needlestick injuries Astek Innovations Ltd Astek House I Atlantic Street I Altrincham Cheshire I WA14 5DH I United Kingdom Telephone +44 (0)161 942 3900 I Fax +44 (0)161

More information

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician. Reprocessing Manual Fuse 1C Colonoscope Part Number FSE-056-EN-2.00 Released August, 2013. This manual includes complete reprocessing instructions for the Fuse 1C Colonoscope system. Please refer to the

More information

Reusable Laryngoscope Blades and Handles

Reusable Laryngoscope Blades and Handles Manufacturer: Starkling Medizintechnik Method: Re-processing instructions Symbol: N.A. Device(s): 1 Green System Reusable Laryngoscope Blade with integrated Light Guide, Solid Stainless Steel Construction

More information

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use. Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1 ENGLISH Intended Use The Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes and Anoscopes are used to illuminate and

More information

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) Fiber Optic Cables Instructions for Use and Processing Instructions DFU-950-0031-00 Fiber Optic Cables Made in USA Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) 737 7611

More information

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations Model G32-S Model G32-E Disinfection Soak Stations Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482 Website: WWW.CIVCO.COM Copyright

More information

Requests for Daily Care and Maintenance

Requests for Daily Care and Maintenance Inkjet printer TS500-800 Requests for Daily Care and Maintenance This machine is a precision machine equipped with extremely fine mechanisms. Especially, the nozzle surface of the heads from which the

More information

IMPORTANT SAFETY INSTRUCTIONS WARNING: When using electric appliances, basic

IMPORTANT SAFETY INSTRUCTIONS WARNING: When using electric appliances, basic INSTRUCTIONS PERTAINING TO; RISK OF FIRE, ELECTRIC SHOCK OR HARM TO PERSONS IMPORTANT SAFETY INSTRUCTIONS WARNING: When using electric appliances, basic precautions should always be followed, including

More information

MODEL 7000 SUCTION UNIT

MODEL 7000 SUCTION UNIT MODEL 7000 SUCTION UNIT OPERATOR S MANUAL Caution Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices

More information

CONTENTS. Fuji Limited 2 Year Warranty

CONTENTS. Fuji Limited 2 Year Warranty Fuji Limited 2 Year Warranty Fuji Industrial Spray Equipment Ltd. ( Fuji ) provides a 24 month limited warranty on the product to the original purchaser effective from the date of purchase against defects

More information

Margo Moo TM and Digger Dog TM Steam Inhaler Guidebook

Margo Moo TM and Digger Dog TM Steam Inhaler Guidebook Margo Moo TM and Digger Dog TM Steam Inhaler Guidebook Item 40-750-000 Item 40-751-000 Please read this guidebook completely before operating this unit. TOLL-FREE CUSTOMER CARE HELPLINE: 1-800-622-4714

More information

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

K Specifications. Max. Water Volume

K Specifications. Max. Water Volume K 2.35 High Pressure Washer Operator Manual Overview... 2 Precautions... 2-4 Assembly Instructions... 4 Operating Instructions... 5 Using the Accessories... 6 Working with Detergents... 7 Shut Down and

More information

Enteral Pump Instructions for Use To Assemble When the low

Enteral Pump Instructions for Use To Assemble When the low Enteral Pump Instructions for Use Before initial use on battery power and after extended storage periods, the pump must be plugged into an AC power source for a minimum of 12 hours (pump may be operated

More information

2015 Myco Industries, Inc Fax:

2015 Myco Industries, Inc Fax: Endoscope Cleaning and Handling It is highly important that you manually clean the endoscope immediately after it is removed from the patient, BEFORE it is automated or manually disinfected. Cleaning your

More information

DZ5080 US Index01 19/09/07 10:50 Page 2

DZ5080 US Index01 19/09/07 10:50 Page 2 www.rowentausa.com 2 1 13 12 3b* 3a 3c 11 3 3d* 10 9 4 5 6 7 8 DESCRIPTION 1 Burst of steam button 2 Spray button 3 Variable steam control a Dry ironing b Steam ironing c Self Clean d Autosteam * 4 Water

More information

Use and Care Guide. Over the Patient Units and Swing Mount Units

Use and Care Guide. Over the Patient Units and Swing Mount Units Over the Patient Units and Swing Mount Units Table of Contents Page General Information 2 Operation 3-5 Dental Unit Adjustments 6-10 Cleaning, Disinfecting and Sterilization 11-12 Maintenance 13-15 Page

More information

ASPIRE Laboratory Aspirator

ASPIRE Laboratory Aspirator ASPIRE Laboratory Aspirator USER MANUAL Rev 2/14/18 Accuris Instruments / Benchmark Scientific Ph: (908) 769-5555 E-mail: info@accuris-usa.com (C) Benchmark Scientific, 2018 THE ACCURIS ASPIRE LABORATORY

More information

FOR VETERINARY USE ONLY

FOR VETERINARY USE ONLY OtoPet-USA Earigator Setup and Use Instructions FOR VETERINARY USE ONLY OtoPet-USA is a division of DermaZoo TM 8909 Iverleigh Court, Potomac, MD 20854 Office: 301-365-0192 Toll Free: 866-374-0897 Fax:

More information

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741

More information

INSTRUCTIONS FOR OPERATION AND CARE OF

INSTRUCTIONS FOR OPERATION AND CARE OF INSTRUCTIONS FOR OPERATION AND CARE OF LITTLE Dipper Compact Paraffin Bath Model LD-2 800-782-7706 626-968-6681 www.whitehallmfg.com Whitehall Manufacturing is a Member of Acorn Engineering s Family of

More information

English. Your Model Number is: HD1500 IMPORTANT

English. Your Model Number is: HD1500 IMPORTANT Your Model Number is: HD1500 IMPORTANT Attention Valued Customer: The serial number of your machine and date of purchase is necessary information to facilitate warranty claims and the ordering of replacement

More information

Requests for Daily Care and Maintenance

Requests for Daily Care and Maintenance Inkjet printer SIJ-320UV Requests for Daily Care and Maintenance This machine is a precision machine equipped with extremely fine mechanisms. Especially, the nozzle surface of the heads from which the

More information

K-SELECT Use & Care Guide. Get the most from your new Keurig brewer

K-SELECT Use & Care Guide. Get the most from your new Keurig brewer K-SELECT Use & Care Guide Get the most from your new Keurig brewer Small Cup MAX. Before You Brew 1 2 Plug & Place Rinse & Fill Remove packing tape from brewer and plug into a grounded outlet. Place a

More information

K-COMPACT Use & Care Guide. Get the most from your new Keurig brewer

K-COMPACT Use & Care Guide. Get the most from your new Keurig brewer K-COMPACT Use & Care Guide Get the most from your new Keurig brewer Before You Brew 1 Plug & Place 2 Rinse & Fill Remove packing tape from brewer and plug into a grounded outlet. Place a large mug (10

More information

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

EG-2990i, EG-2790i EG-1690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i10, EG29-i10

EG-2990i, EG-2790i EG-1690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i10, EG29-i10 Only For the Americas INSTRUCTIONS FOR USE (REPROCESSING) PENTAX VIDEO UPPER G.I. SCOPES EG-2990i, EG-2790i EG-690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i0, EG29-i0 Product Overview These

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

QuickSteamer USER S GUIDE 1950 SERIES QUICKSTEAMER 2050 SERIES QUICKSTEAMER POWERBRUSH. Safety Instructions. Product View. Assembly/Accessories 6-9

QuickSteamer USER S GUIDE 1950 SERIES QUICKSTEAMER 2050 SERIES QUICKSTEAMER POWERBRUSH. Safety Instructions. Product View. Assembly/Accessories 6-9 QuickSteamer USER S GUIDE 1950 SERIES QUICKSTEAMER 2050 SERIES QUICKSTEAMER POWERBRUSH 3 Safety Instructions 4 Product View 5 Assembly/Accessories 6-9 Operations / Special Tools 10 Routine Maintenance

More information

K Specifications. Max. Water Volume

K Specifications. Max. Water Volume K 2.75 High Pressure Washer Operator Manual Overview... 2 Precautions... 2-4 Assembly Instructions... 4 Operating Instructions... 5 Using the Accessories... 6 Working with Detergents... 7 Shut Down and

More information

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI ULTRASONIC SCALER PIEZO MINI CONTENTS XI - SYMBOLS 1. INTRODUCTION 1 Alternating current Type BF device 2. WARNINGS 1 3. PRESENTATION 1 3.1 Presentation 1 3.2 Technical description 2! Warning, please refer

More information

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications TABLE OF CONTENTS MANUAL PRE-CLEANING High Speed Handpieces 2

More information

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU-20B CompriCart TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the Unit...................2 -

More information

Installation & Operating Guide

Installation & Operating Guide 5-036 HOT WATER TANK Installation & Operating Guide Read all instructions thoroughly. Keep this guide for future reference. Proof of purchase is required for Warranty. Staple receipt or proof of purchase

More information

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution S168-544-OO1E REV.A Schuco byallied MEDICAL ASPIRATOR MnrInI -33OA USER'S MANUAL A Caution Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by

More information

HH80C HYDRAULIC CRIMP TOOL

HH80C HYDRAULIC CRIMP TOOL HH80C HYDRAULIC CRIMP TOOL DATASHEET SEE PAGE 8 FOR IMPORTANT INFORMATION CONCERNING LIMITED WARRANTY, AND LIMITATION OF LIABILITY INTRODUCTION: The HH80C is a hand actuated hydraulic crimp tool designed

More information

Ultrasonic Cool Mist Humidifier

Ultrasonic Cool Mist Humidifier Model No.: LV450CH Ultrasonic Cool Mist Humidifier Questions or or Concerns? Please contact us Monday Mon-Fri 9:00AM-5:00PM - Friday am - PST 5:00 pm PST at support@levoit.com (888) 726-8520 email or at

More information

Pet. Easy Guide. VAX Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire, WR9 OQH, UK - website:

Pet. Easy Guide. VAX Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire, WR9 OQH, UK   - website: V-027P Rapide Pet User Guide 10/7/07 10:05 Page 14 EEC STATEMENT OF COMPLIANCE Manufacturer/EEC importer: VAX Limited, hereby on our own responsibility declare that the products V-027/V-027PT/V-027P/CCW-701

More information

Operating / User Manual

Operating / User Manual UniARM Surgical Support System Operating / User Manual Version 1.1 MITAKA KOHKI Co., Ltd. 1. Introduction Introduction Thank you for purchasing the UniARM for use in surgical procedures. To use the device

More information

QuickSteamer TM U S E R S G U I D E 1770/1970 SERIES QUICKSTEAMER 2080 SERIES QUICKSTEAMER POWERBRUSH Safety Instructions.

QuickSteamer TM U S E R S G U I D E 1770/1970 SERIES QUICKSTEAMER 2080 SERIES QUICKSTEAMER POWERBRUSH Safety Instructions. QuickSteamer TM U S E R S G U I D E 1770/1970 SERIES QUICKSTEAMER 2080 SERIES QUICKSTEAMER POWERBRUSH 3 Safety Instructions 4 Product View 5 Assembly 6-8 Operations 8-9 Maintenance 10 Troubleshooting 10

More information

Wax Base Heater & Dispenser

Wax Base Heater & Dispenser Wax Base Heater & Dispenser Service Manual Models: IDWB2/0900, IDWB2/0775, IDWB3/0900, IDWB3/0775, IDWB4/0900, IDWB4/0775 Introduction............................................................................

More information

CMU701/ 705 OPERATIONS & INSTRUCTION MANUAL

CMU701/ 705 OPERATIONS & INSTRUCTION MANUAL CMU701/ 705 OPERATIONS & INSTRUCTION MANUAL MANUFACTURED BY 2519 Fairmont Road Morgantown, West Virginia 26501 PHONE: (304)-983-2642 FAX: (304)-983-2643 www.imswv.com 1 TABLE OF CONTENTS: 1.) Introduction

More information

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE NOVA Pump & Pad (On/Off Switch & Pressure Adjustable) Item # APP-2001 IMPORTANT SAFEGUARDS When using electrical products, especially when children are present,

More information

Electric Pressure Washer

Electric Pressure Washer Electric Pressure Washer S1900 039-8575-4 OWNER S MANUAL If you have any questions or comments, visit our website at www.simoniz.ca or call our toll free hotline 1-866-4SIMONIZ (1-866-474-6664) PLEASE

More information

CF1400 DO NOT RETURN PRODUCT TO PLACE OF PURCHASE. IF YOU HAVE ANY QUESTIONS, CALL OUR TOLL FREE HOTLINE:

CF1400 DO NOT RETURN PRODUCT TO PLACE OF PURCHASE. IF YOU HAVE ANY QUESTIONS, CALL OUR TOLL FREE HOTLINE: CF1400 DO NOT RETURN PRODUCT TO PLACE OF PURCHASE. IF YOU HAVE ANY QUESTIONS, CALL OUR TOLL FREE HOTLINE: 1-866-902-9690 8:00 am to 5:00 pm, C.S.T., Monday - Friday www.forcetoolsusa.com TABLE OF CONTENTS

More information

Use & Care Guide. Get the most from your new Keurig brewer

Use & Care Guide. Get the most from your new Keurig brewer Use & Care Guide Get the most from your new Keurig brewer ADD DESCALE MAX. ADD DESCALE ADD DESCALE ADD DESCALE Before You Brew 1 Plug & Place Rinse & Fill 2 Remove packing tape from brewer and plug into

More information

MONSTER FLOOR (SC1Z1) INSTRUCTION MANUAL. CUSTOMER SERVICE: (AUS) (NZ) Model #: SC1Z1 AUS, NZ

MONSTER FLOOR (SC1Z1) INSTRUCTION MANUAL. CUSTOMER SERVICE: (AUS) (NZ) Model #: SC1Z1 AUS, NZ Steam Booster MONSTER FLOOR (SC1Z1) INSTRUCTION MANUAL CUSTOMER SERVICE: 1800 663 866 (AUS) 0800 666 783 (NZ) Model #: SC1Z1 AUS, NZ IMPORTANT SAFE T SAVE THESE INSTRUCTIONS When using an electrical appliance,

More information

Instructions for Use Spinal Punches IFU

Instructions for Use Spinal Punches IFU Indications for use Spinal punches are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Contraindiciations Instruments should not

More information

OPERATOR S MANUAL SF-12 AND SF12A HI VOLUME WATER HEATER

OPERATOR S MANUAL SF-12 AND SF12A HI VOLUME WATER HEATER OPERATOR S MANUAL SF-12 AND SF12A HI VOLUME WATER HEATER MODELS SF12-A PICTURED Page 1 of 12 Table of Contents Safety Precautions Section 1. Section 2. Section 3. Section 4. Section 5. Section 6. Systems

More information

Instruction Sheet DANGER

Instruction Sheet DANGER VIQUA Lamp and Quartz Sleeve Replacement Models: D4, E4, F4, D4 Premium, Plus Models Discontinued Models: A, B, C, D, E, F, B4, C4 Overview UV Lamp replacement: The amount of UV light created by the UV

More information

SM088 STEAM MOP INSTRUCTION MANUAL. For Authentic Monster Replacement Parts Call Model #: SM088 G-B-R

SM088 STEAM MOP INSTRUCTION MANUAL. For Authentic Monster Replacement Parts Call Model #: SM088 G-B-R SM088 STEAM MOP INSTRUCTION MANUAL Call 888.896.8786 Model #: SM088 G-B-R IMPORTANT SAFE T SAVE THESE INSTRUCTIONS When using an electrical appliance, basic safety precautions should always be observed,

More information

1200 EVERYDAY PRESSURE WASHER. Instruction Manual. Model: PW If you have any questions or comments, please call our toll free

1200 EVERYDAY PRESSURE WASHER. Instruction Manual. Model: PW If you have any questions or comments, please call our toll free 1200 EVERYDAY PRESSURE WASHER Instruction Manual Model: PW0501400 If you have any questions or comments, please call our toll free 1-866-924-7626 PLEASE KEEP THE ORIGINAL BOX, PACKAGING AND RECEIPT. BEFORE

More information

QuickSteamer USER S GUIDE 1950 SERIES 6-8. Safety Instructions. Product view. Assembly. Operations / Special Tools. Routine Maintenance

QuickSteamer USER S GUIDE 1950 SERIES 6-8. Safety Instructions. Product view. Assembly. Operations / Special Tools. Routine Maintenance QuickSteamer USER S GUIDE 1950 SERIES 3 Safety Instructions 4 Product view 5 Assembly 6-8 Operations / Special Tools 9 Routine Maintenance 10 Troubleshooting 10 Consumer Services 11 Replacement Parts 12

More information

CONSTANT SUCTION UNIT

CONSTANT SUCTION UNIT CONSTANT SUCTION UNIT OPERATOR/MAINTENANCE M A N U A L Squire-Cogswell/Aeros Instruments Inc 1111 Lakeside Drive Gurnee, IL 60031-4099 1-800-662-5822/847-855-0800 Fax: 847-855-6218 www.aerosinstruments.com

More information

PORTABLE DOG BATH TOOL

PORTABLE DOG BATH TOOL TM PORTABLE DOG BATH TOOL USER GUIDE 1842 SERIES More of a video person? Look for this icon and go online for an instructional how-to video at BISSELL.com/videos IMPORTANT SAFETY INSTRUCTIONS READ ALL

More information

Electric Pressure Washer

Electric Pressure Washer Electric Pressure Washer S1700 039-8574-6 Owner s Manual If you have any questions or comments, visit our website at www.simoniz.ca or call our toll free hotline 1-866-4SIMONIZ (1-866-474-6664). PLEASE

More information

PRESSURE WASHER. OWNER'S MANUAL Model VPW49HA 120V., 60 Hz., 13 Amps.

PRESSURE WASHER. OWNER'S MANUAL Model VPW49HA 120V., 60 Hz., 13 Amps. WASHER OWNER'S MANUAL Model VPW49HA 120V., 60 Hz., 13 Amps. EURO-PRO Operating LLC U.S.: 94 Main Mill Street, Door 16 Canada: 4400 Bois-Franc Plattsburgh, NY 12901 St-Laurent, QC H4S 1A7 Tel.: 1 (800)

More information

P.N User Manual STAINLESS STEEL ELECTRIC KETTLE 5964/5965 Series MODELS

P.N User Manual STAINLESS STEEL ELECTRIC KETTLE 5964/5965 Series MODELS User Manual STAINLESS STEEL ELECTRIC KETTLE MODELS 5964/5965 Series www.oster.ca P.N. 166019 IMPORTANT SAFEGUARDS When using electrical appliances, basic safety precautions should always be followed to

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS Siphonic Ultra-High Efficiency Two-Piece Toilets HOptimum Single Flush Models 88AA / 88CA / 88DA HOption Dual Flush Models 886 / 88 / 889 Thank you for selecting American Standard

More information

INSTRUCTIONS FOR OPERATION AND CARE OF

INSTRUCTIONS FOR OPERATION AND CARE OF INSTRUCTIONS FOR OPERATION AND CARE OF LITTLE Splint Compact Splint Pan Model LS-1 Whitehall Manufacturing is a Member of Acorn Engineering s Family of Companies. 6900-191-001 July 2010 Introduction...3

More information

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE Procedure Number: II.01.200a Effective Date: 09/14/99 Review and Update: 09/2003 TITLE: ADVIA 120 ROUTINE MAINTENANCE PURPOSE AND PRINCIPLE: The ADVIA 120

More information

APW PSI PRESSURE WASHER

APW PSI PRESSURE WASHER APW5001 1300 PSI PRESSURE WASHER 1300 PSI PRESSURE WASHER Please carefully read this manual which has been specifically designed to provide all the necessary information for the correct use of the machine,

More information

HC12W COOL MIST ULTRASONIC HUMIDIFIER OWNER S MANUAL. Read and save these instructions.

HC12W COOL MIST ULTRASONIC HUMIDIFIER OWNER S MANUAL. Read and save these instructions. COOL MIST ULTRASONIC HUMIDIFIER HC12W OWNER S MANUAL Read and save these instructions. 2 BECAUSE YOU RE NOT LIKE EVERYONE ELSE And neither are we. Always at the forefront of our industry, our goal is to

More information

English Manual for Jewelry Casting and Investing The professional choice

English Manual for Jewelry Casting and Investing The professional choice English Manual for Jewelry Casting and Investing The professional choice QUICK START GUIDE: For thorough instructions, please review attached instructions. However, below, you will find a simplified start-up

More information

ULTRASONIC HUMIDIFIER

ULTRASONIC HUMIDIFIER To Buy: Visit www.sylvane.com or call (800) 934-9194 For Product Support: Contact Sunpentown at 1-800-330-0388 ULTRASONIC HUMIDIFIER [ADORABLE MONKEY] To prolong the life of this humidifier, using distilled

More information

COMPACT GARMENT STEAMER

COMPACT GARMENT STEAMER COMPACT GARMENT STEAMER 1103901405-2 - 06/07 www.rowentausa.com 8 1 2 3 4 5 a b 9 6 10 7 11 * * * 12 * 15 13 16 14 * *Depending on model: Accessories are specific to certain models or available as an optional

More information

WARRANTY TABLE OF CONTENTS TECHNICAL SPECIFICATIONS. Product Warranty Always keep your receipt as proof of purchase. Technical specifications...

WARRANTY TABLE OF CONTENTS TECHNICAL SPECIFICATIONS. Product Warranty Always keep your receipt as proof of purchase. Technical specifications... 4 TABLE OF CONTENTS Technical specifications... 1 Warranty... 2 Important precautions Read first!... 3 Safety features and electrical power information... 5 Parts and diagram list... 6 Assembly and operation

More information

Gas Connector Installation Instructions

Gas Connector Installation Instructions Gas Connector Installation Instructions 6015 Enterprise Drive Export, PA 15632 USA 1-800-DORMONT (367-6668) (724) 733-4800 Fax (724) 733-4808 www.dormont.com Safety Precautions - Read warnings carefully

More information

2. Push the drain connector into the disinfectant removal port until it clicks.

2. Push the drain connector into the disinfectant removal port until it clicks. Chapter 3 Inspection Before Use 2. Push the drain connector into the disinfectant removal port until it clicks. When connecting the drain connector to the disinfectant removal port, do not push the connector

More information

18 INCHES, OSCILLATING STAND FAN

18 INCHES, OSCILLATING STAND FAN To Buy: Contact Sylvane at 800-934-9194 or visit www.sylvane.com For Product Support: Contact Soleus Air at (888) 876-5387 18 INCHES, OSCILLATING STAND FAN OWNER S MANUAL Model # FSM-45 PLEASE READ AND

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 700 INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model ABM - 700 M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: 800-831-1525 Mazeppa, MN 55956 Fax : (507)843-4953

More information

Garment Steamer Model SC620

Garment Steamer Model SC620 OWNER S MANUAL Garment Steamer Model SC620 For information or assistance: EURO-PRO Corporation 178 West Service Rd. Champlain, NY 12919 1-800-798-7398 www.euro-pro.com IMPORTANT SAFETY INSTRUCTIONS When

More information

AUTO SPEED BLENDER INSTRUCTION MANUAL MODEL: BLCLMB1

AUTO SPEED BLENDER INSTRUCTION MANUAL MODEL: BLCLMB1 AUTO SPEED BLENDER INSTRUCTION MANUAL MODEL: BLCLMB1 1 TABLE OF CONTENTS IMPORTANT SAFEGUARDS 3 POLARIZED PLUG 4 PRODUCT DIAGRAM 5 USAGE INSTRUCTIONS PREPARING YOUR AUTO SPEED BLENDER FOR USE 6 HOW TO

More information

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL S-SCORT II Model 15006 SSCOR, INC. 11064 Randall Street Sun Valley, CA 91352 USA Telephone +1-818-504-4054 800-434-5211 Fax +1-818-504-6032 www.sscor.com Email:

More information

Instruction Sheet DANGER

Instruction Sheet DANGER VIQUA Lamp and Quartz Sleeve Replacement Models: D4, E4, F4, D4 Premium, Plus Models Discontinued Models: A, B, C, D, E, F, B4, C4 Overview UV Lamp replacement: The amount of UV light created by the UV

More information

WINDSHIELD REPAIR MANUAL SUPER CINCH

WINDSHIELD REPAIR MANUAL SUPER CINCH WINDSHIELD REPAIR MANUAL FOR SUPER CINCH WWW.GlassProSystems.Com 1-800-395-7380 COPYRIGHT Glass Pro Systems 2006 1 Table of Contents 1. Cover Page 2. Introduction 3. Introduction, continued 4. Parts of

More information

K 2.97 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS

K 2.97 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS K 2.97 M High Pressure Washer Operator Manual Overview... 2 Important Precautions... 3-4 Assembly Instructions... 4 Operating Instructions... 5 GFCI Instructions... 6 Using the Accessories... 6 Working

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE

INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE Heritage VorMax Right Height Elongated 2-piece Toilet Model 205AA104 Thank you for selecting American Standard - the benchmark of fine quality for over 100

More information

HW-17 Record Cleaning Machine Setup and Instruction Manual

HW-17 Record Cleaning Machine Setup and Instruction Manual HW-17 Record Cleaning Machine Setup and Instruction Manual VPI Industries, Inc., 77 Cliffwood Ave. #3B, Cliffwood, NJ 07721 Phone: 732-583-6895, Email: Sales@vpiindustries.com http://www.vpiindustries.com

More information

Requests for Daily Care and Maintenance

Requests for Daily Care and Maintenance Inkjet printer UJF-7151 plus Requests for Daily Care and Maintenance This machine is a precision machine equipped with extremely fine mechanisms. Especially, the nozzle surface of the heads from which

More information

Instruction Leaflet Read instructions before operating. Retain for future reference.

Instruction Leaflet Read instructions before operating. Retain for future reference. pure indoor living ULTRASONIC HUMIDIFIER MODEL: BUL6100-CN Instruction Leaflet Read instructions before operating. Retain for future reference. Questions? Comments? Call 1-800-253-2764 in North America

More information

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model ABM - 715 INSTALLATION and INSTRUCTION MANUAL for QUALITY AIR BREATHING SYSTEMS Model ABM - 715 M A R T E C H S E R V I C E S C O M P A N Y P.O. Box 7079 OFFICE: 800-831-1525 Mazeppa, MN 55956 Fax : (507)843-4953

More information

Natural Mist Humidifier. Vicks. (Model V3500) IMPORTANT! READ AND SAVE THESE INSTRUCTIONS. Use and Care Manual

Natural Mist Humidifier. Vicks. (Model V3500) IMPORTANT! READ AND SAVE THESE INSTRUCTIONS. Use and Care Manual IMPORTANT! READ AND SAVE THESE INSTRUCTIONS Vicks Natural Mist Humidifier (Model V3500) Use and Care Manual If you have any questions about the operation of your Vicks Natural Mist Humidifier, call our

More information

Instruction Manual. Cheese Warmer with Pump

Instruction Manual. Cheese Warmer with Pump Instruction Manual Cheese Warmer with Pump 10700 Medallion Drive, Cincinnati, Ohio 45241-4807 USA 2013 Gold Medal Products Co. Part No. 38298 SAFETY PRECAUTIONS Page 2 INSTALLATION INSTRUCTIONS Inspection

More information

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

*smith&nephew VERSAJET HYDROSURGERY SYSTEM Product Manual Manuel D utilisation Bedienungshandbuch Manual Del Producto Manuale Prodotto Producthandleiding *smith&nephew VERSAJET HYDROSURGERY SYSTEM VERSAJET Hydrosurgery System Product Manual Contents

More information

P.N ELECTRIC KETTLE User Manual Series MODEL BVSTKT308

P.N ELECTRIC KETTLE User Manual Series MODEL BVSTKT308 TM User Manual ELECTRIC KETTLE MODEL BVSTKT308 Series www.oster.ca P.N. 162703 IMPORTANT SAFEGUARDS When using electrical appliances, basic safety precautions should always be followed to reduce the risk

More information

CONGRATULATIONS! TABLE OF CONTENTS

CONGRATULATIONS! TABLE OF CONTENTS CONGRATULATIONS! Thank you for choosing Mytee Products for your supply needs. By doing so, you have joined thousands of your fellow cleaning professionals in choosing quality, service, and value in your

More information

Owner s. 18 Misting MF18W. To view

Owner s. 18 Misting MF18W. To view Owner s Manual MF18W 18 Misting Fan IMPORTANT SAFETY INFORMATION: Please carefully read this manual before attempting to install or use this misting fan. For your safety, comply with all safety instructionss

More information

Because you re not like everyone else. EVAPORATIVE COOLER EC220W OWNER S MANUAL

Because you re not like everyone else. EVAPORATIVE COOLER EC220W OWNER S MANUAL Because you re not like everyone else. EVAPORATIVE COOLER EC220W OWNER S MANUAL 2 BECAUSE YOU RE NOT LIKE EVERYONE ELSE. And neither are we. Always at the forefront of our industry, our goal is to offer

More information

TQ Series. Electronic Control Pump Instruction Manual. 50Hz. ISO 9001 Certified Walrus Pump Co., Ltd.

TQ Series. Electronic Control Pump Instruction Manual. 50Hz. ISO 9001 Certified Walrus Pump Co., Ltd. TQ Series Electronic Control Pump Instruction Manual 50Hz ISO 9001 Certified Walrus Pump Co., Ltd. EC Declaration of Conformity Manufacturer: Walrus Pump Co., Ltd. Address: No. 83-14, Dapiantou, Sanjhih

More information

CLEANING INSTRUCTIONS

CLEANING INSTRUCTIONS Reusable Instrument Care Manual Pagoda Pedicle Screw System Manufacturer Ortho Development Corporation 12187 South Business Park Drive Draper, UT 84020 Phone 801-553-9991 Fax 801-553-9993 www.orthodevelopment.com

More information