Processing instructions for the Ligamys suture forceps

Size: px
Start display at page:

Download "Processing instructions for the Ligamys suture forceps"

Transcription

1 Processing instructions for the Ligamys suture forceps

2

3 Table of contents 1 Scope of application 4 2 Warnings and precautions Restrictions 6 4 Processing instructions Processing immediately after use Preparation of the detergents Instructions for manual pre-cleaning Instructions for machine cleaning and disinfection 4.5 Inspection and maintenance Packaging of individual instruments Sterilisation instructions Storage instructions 15 5 Effectiveness of processing 16 6 Useful life / Number of processing cycles 16 7 Detection of damage and its causes Customer service information / Address 22 3

4 1 Scope of application 2 Warnings and precautions Mathys Ltd Bettlach, as the manufacturer of surgical instruments which can be reprocessed, is obligated to provide processing instructions. These processing instructions which were created based on the requirements of ISO 17664, apply specifically to Ligamys suture forceps of Mathys Ltd Bettlach. The chemical parameters of the process as well as the equipment listed in the following processing instructions are recommendations resulting from the findings of the validation process for instrument processing of Mathys Ltd. The processor is responsible for ensuring that the processing that is actually performed with the individual equipment, process chemicals and personnel in the processing facility achieves the desired results. The same chemicals, parameters, and same technical equipment as in the validation method of Mathys Ltd do not necessarily have to be used. Equivalent or alternative products may be used with which a successful validated cleaning and sterilisation process can be proven as a result. If your hospital has a method that is already established and validated and which is proven to lead to the desired result, there is no need for your hospital to change the method. Personnel coming into contact with potentially contaminated or contaminated surgical instruments must implement the generally acknowledged precautions. Special care is required when handling instruments with pointed parts or sharp edges. Personal protection equipment (overall, mask, safety goggles, visor, gloves, shoes, shoe covers, etc.) is necessary in order to avoid contact with contaminated or potentially contaminated materials, instruments and products. Low-foaming detergents should be used for manual cleaning procedures in order to ensure visibility of the instruments. When cleaning manually with brushes, it is recommended that the instruments should be held constantly below the surface of the cleaning solution. This will ensure that no aerosols will form and avoid splashes that could spread contaminating substances. In order to avoid the accumulation of residues of detergents, all detergents must be removed completely from the product surfaces. 4

5 Heavy objects may not be placed on top of delicate products. Do not allow contaminated instruments to dry before reprocessing them. This is important, since all the steps described below for cleaning and sterilisation will be facilitated where blood, bodily fluids, bone and tissue residues as well as saline solution or disinfectants are not allowed to dry onto the surfaces of used instruments. time recommended by the manufacturers should be observed under all circumstances. Mathys Ltd Bettlach recommends using detergents with a ph of < 9.5. For machine cleaning, the indications provided by the machine and detergent manufacturers must be complied with. When loading the cleaning trays and baskets, it should be ensured that instruments do not damage each other and that they can be easily rinsed all around. The chloride and iodide ions contained in detergents and disinfectants cause pitting corrosion. For this reason, contact of the instruments with these ions must be kept as brief as possible. Rinse the instruments thoroughly with regular water and then distilled water in order to remove all residues. Never leave wet instruments to stand. Rather, dry them immediately. The condensation of moisture resulting from sterilisation can be avoided by extending the drying phase. Strong acid or alkaline detergents or overly high doses can attack and destroy the protective passive layer of the instruments or markings. The concentration and exposure 5

6 3 Restrictions Patients who, in view of prion diseases such as Transmissible Spongiform Encephalopathy (TSE) or variant Creutzfeldt-Jakob Disease (vcjd) and the associated infections, are considered to be a risk must therefore be operated on using disposable instruments whenever possible. When using instruments that can be reprocessed, it is recommended that the entire cleaning procedure be repeated at least twice. WARNING: It is of the utmost importance that alkaline detergents should be completely neutralised and rinsed thoroughly off the instruments. The suture forceps should be carefully inspected after cleaning. In doing so, the functionality of the forceps should be checked. It should be ensured that the tip of the cannula of the suture forceps is sufficiently sharp for use and that it was not damaged during cleaning. A guide wire must be used to check whether the cannula is completely free If in your opinion this is no longer the case, contact your local Mathys partner. Mathys Ltd Bettlach recommends choosing a cleaning agent with added enzymes to decompose blood, bodily fluids and tissues. Please note that some enzyme solutions are intended specifically for decomposing faecal matter or other organic contaminating substances, and are therefore not suitable for cleaning surgical instruments. There are no restrictions preventing repeated processing of Mathys Ltd Bettlach s instruments, since this has a limited impact on the useful life of the instruments. Their useful life is defined as a rule by wear and damage resulting from use. The use of hard water ( dh value > 14) should be avoided. For initial rinsing, soft tap water is suitable, followed by thoroughly rinsing with distilled water (conductivity < 5 microsiemens) in order to remove all residues. Ordinary water often contains high concentrations of minerals (e. g. chalk or silicic acid) which appear as sharply delimited stains on the instrument surface. TIP: Instruments should never be left wet but dried immediately. 6

7 4 Processing instructions 4.1 Processing immediately after use First of all, any remaining bodily fluids and tissues must be wiped off the instruments with a lint-free disposable cloth. Then the instruments must be placed in a bowl with distilled water or covered with damp cloths so that tissues and bodily fluids do not dry out. Saline solution, blood, bodily fluids, tissues, bone residues or other organic particles must be removed as soon as possible from instruments before cleaning. Optimum cleaning is ensured if the instruments are cleaned within 30 minutes after use, in order to minimise the risk of substances and materials drying on their surfaces. In order to prevent contamination, used instruments must be taken to the Main Procurement Department in closed or covered containers. TIP: Soaking instruments in proteolytic enzyme solutions facilitates cleaning, particularly in the case of instruments featuring complex shapes and areas difficult to access (e. g. cannulated and tubular designs, and so on). The enzyme solutions break down the substances containing proteins and thus prevent materials containing blood and proteins from drying on the instruments. WARNING: Always strictly follow the manufacturer s instructions for preparing and using the solutions. 7

8 4.2 Preparation of the detergents The concentration, temperature and exposure time for the cleaning agents recommended by the manufacturers must be observed under all circumstances to achieve optimal cleaning. Treatment with a neutralising agent and thorough rinsing are absolutely necessary after treatment with alkaline agents. The Ligamys suture forceps must be treated in a twostage cleaning procedure, first a manual cleaning followed by machine cleaning, in order to obtain the necessary cleaning results. When using dry detergents in powder form, it is necessary to make sure that the detergents are completely dissolved before use to avoid discolouration or corrosion of the instruments. Very dirty cleaning solutions (containing blood and / or that have become cloudy) must be replaced with freshly prepared solutions. 8

9 4.3 Instructions for manual pre-cleaning No. Step Equipment / Medium 1 Mechanical removal of all visible contaminations with a synthetic brush* under running water (cold) until all visible residue has disappeared. Synthetic brush Running water, cold 2 Flush the suture-guiding cannula of the suture forceps with 50 ml cleaning solution that contains enzymes. Ensure that liquid visibly flows out of the tip of the cannula (see Fig. 1). If the cannula is blocked by fragments of tissue, the roller housing must be folded up and the cannula must be flushed by directly applying a syringe (see Fig. 2). Once again, ensure that liquid visibly flows out of the tip of the cannula. If there is no flow of liquid, the suture forceps should not be used and must be replaced. Plastic syringe without Luer lock deconex TWIN PH10 and deconex TWIN ZYME (Bohrer Chemie AG), each 1 % (v / v) in distilled water 3 The pre-rinsed suture forceps are treated with ultrasound in the combined cleaning agent / enzyme solution in an ultrasonic batch (30 60 khz) for 30 minutes at approx. 37 C. It should be ensured that the forceps are completely covered with the cleaning agent. Ultrasound bath (30 60 khz) deconex TWIN PH10 and deconex TWIN ZYME (Bohrer Chemie AG), each 1 % (v / v) in distilled water 4 Rinse the suture forceps well under running water. Running water, cold 5 Flush the suture-guiding cannula of the suture forceps with 50 ml tap water. 50 ml plastic syringe without Luer lock Running water, cold 9

10 6 Flush the suture-guiding cannula of the suture forceps with 50 ml distilled water. 50 ml plastic syringe without Luer lock Running water, cold 7 The suture forceps should be checked visually for any residues or damage. If residues are still visible, these should be removed using a synthetic brush under running water and manual cleaning steps 3 to 7 should be repeated. * Brushes must be decontaminated and sterilised or disposed after use. Do NOT use steel brushes. Visual inspection Possibly synthetic brush Running water, cold Fig. 1: Flushing the cannula through the roller housing Fig. 2: Direct flushing of the cannula after folding up the roller housing. To fold up: Twist and pull on the knurled metal knob 10

11 4.4 Instructions for machine cleaning and disinfection It is recommended that cleaning be performed using professional cleaning and disinfection equipment, e. g. Miele G7836CD. The cleaning basket must have a Luer lock attachment so that the suture forceps can be inserted. No. Step Conditions / Medium 1 The suture forceps with the roller housing folded up are secured to the cleaning basket using a Luer lock adapter (see Fig. 3). Ensure that the forceps are not damaged when tilted or by the rotating parts of the washing machine. 2 Pre-rinse 1 Water quality: Duration: Temperature: 3 Cleaning process Dosage: Cleaning basket with Luer lock attachment Water quality: Duration: Temperature: 4 Intermediate rinse I Water quality: Duration: Temperature: Mains water 2 minutes Cold 0.5 % deconex TWIN PH10 at 30 C and 0.2 % deconex TWIN ZYME at 40 C (Bohrer Chemie AG) Distilled water 10 minutes 55 C Mains water 2 minutes 50 C 11

12 5 Intermediate rinse II Water quality: Duration: Temperature: 6 Rinsing and thermal disinfection Water quality: Duration: Temperature: 7 Drying Duration: Temperature: 8 The suture forceps should be checked visually for any residues or damage. If residues are visible, the entire manual and machine process should be repeated. Visual inspection Distilled water 2 minutes 50 C Distilled water 10 minutes > 90 C 15 minutes 115 C Fig. 3: Machine cleaning via Luer lock adapter 12

13 4.5 Inspection and maintenance In order to make sure that all dirt has been removed, top priority must be given to checking each suture forceps carefully. Should any dirt be discovered sticking to any instruments, the entire cleaning / disinfection process must be repeated without delay. The suture forceps should be checked for damage. In particular, the tip of the cannula must be sharp and the forceps must be able to be closed and opened without any resistance. Apply maintenance products specifically: The joint mechanisms identified with a blue arrow in Fig. 4 should be treated with small drops of paraffin / white mineral oil-based lubricant according to the EU Pharmacopoeia, e. g. with InstruCare from Laboratorium Dr. Deppe GmbH (without silicone, biocompatibility guaranteed, can be steamsterilised and is steam-permeable, physiologically safe according to DAB and 31 LMBG). Be aware of the expiry date indicated by the manufacturer for the stored concentrations diluted for use. Damage or signs of wear that might have a negative impact on the functioning of the instrument must be reported to your local Mathys partner. He / she is responsible for deciding whether to repair and / or replace instruments. Fig. 4: Joint mechanisms to be treated 13

14 4.6 Packaging of individual instruments 4.7 Sterilisation instructions The suture forceps can be sterilised using steam sterilisation. It is recommended to place the suture forceps in the instrument set holder and pack the entire basket either in a special bag or container suitable for steam sterilisation (e. g. paper bag according to DIN EN 868-5, heat- and self-sealing transparent bags, or other equivalent material). WARNING: If sterilisation fleece is used, this must be free of any residues of detergents. Mathys Ltd Bettlach advises against re-usable fleece. The hospital is responsible for the in-house procedures for implementing all the steps specified above in order to guarantee complete penetration of the sterilising steam through the instruments. The hospital must furthermore implement protective measures with regard to sharp or potentially dangerous places on instruments. Instructions given by the manufacturer of the sterilising equipment must always be followed. If several instrument sets are sterilised in one and the same sterilisation cycle, the maximum capacity of the sterilising equipment as specified by the manufacturer may not be exceeded. For optimum sterilisation the instrument sets must be properly prepared and packed tidily in the instrument trays and containers provided for this purpose. Only if this is done can the steam spread and penetrate so as to reach all the surfaces. In case of steam sterilisation, it must be ensured that the product is completely dry after sterilisation. 14

15 4.8 Storage instructions National recommendations / guidelines for sterilising instruments must be followed in all cases. The following steps at least must be taken (e. g. Zirbus Autoclave LSA 346): The following sterilisation steps are recommended: No. Step Conditions 1 Vacuum Pressure: < 120 mbar 2 Steam Duration: Temperature: Duration: Temperature: 3 Dry Pressure: Duration: 18 minutes 134 C 3 minutes * 134 C < 120 mbar 10 minutes 4 Cool Temperature: at room temperature The packaged and sterile instruments must be stored in a dry and cool room, protected from dust, insects, vermin and direct sunlight. The room may be accessed by authorised personnel only. The equipment for storage and transport must be designed in such a way as to avoid any muddling, overloading or falling of items. Sterile medical products may never be stored directly on the floor. The instruments will be used in order of arrival of the goods, and the sterile packaging must be checked very thoroughly to ensure that it is intact before it is opened. WARNING: Should the packaging or a sterile fleece have ecome torn, punctured, visibly damaged or damp, the instrument set must be repackaged and re-sterilised. Also, if there are signs on a sterilisation container that the lid seals or filters have been opened or damaged, the instrument set must be re-sterilised and the sterile filter replaced. In the case of reusable filters, these must undergo a thorough visual inspection. * Process for UK 15

16 5 Effectiveness of processing 6 Useful life Number of processing cycles The processing method recommended in the present processing instructions has been reviewed by an external laboratory. The results with regard to both limiting and guidance values for protein residues meet the requirements of the guideline by DGKH, DGSV and AKI for machine cleaning and thermal disinfection processes for medical products ( Assessment, 3 rd edition, 2008). The useful life of the Ligamys suture forceps is defined as a rule by wear and damage resulting from use. The user is responsible for checking optimal functionality (cutting sharpness, cleanliness, corrosion) before each use. 16

17 7 Detection of damage and its causes Defects and their causes as well as correct trouble-shooting in the event of damage are shown in the following table. Defect Cause Examination and measure Bent or broken cannula tips Improper handling Overpacked trays Can be successfully bent back Continued use Not possible to bend back or broken Do not use instrument any longer and discard Complete opening of the jaws not possible Improper handling Overload during use Overpacked trays Do not use instrument any longer and discard Correct Incorrect 17

18 Defect Cause Examination and measure Complete closing of the jaws not possible Improper handling Overload during use Overpacked trays Do not use instrument any longer and discard Correct Incorrect Tube bent Improper handling Overload during use Overpacked trays Can be successfully bent back Continued use Not possible to bend back or broken Do not use instrument any longer and discard 18

19 Defect Cause Examination and measure Knurled knob blocked Improper handling Tissue residues in the roller housing Successful removal of tissue residues with positive functional check Continued use Functional check negative Do not use instrument any longer and discard Luer connection damaged Improper handling Overpacked trays Damage due to being dropped Connection can no longer be attached Do not use instrument any longer and discard 19

20 Defect Cause Examination and measure Labelling no longer legible Worn off due to use Fading due to the cleaning process, cleaning agents and sterilisation procedure Article number, lot and CE marking can still be read Continued use Article number, batch or CE marking can no longer be read Do not use instrument any longer and discard Corrosion Unsuitable cleaning agents Excessive time spent in cleaning agents and disinfectants Improper processing Do not use instrument any longer and discard 20

21 Defect Cause Examination and measure Organic residues Unsuitable cleaning agents and disinfectants Residues allowed to dry for too long prior to processing Insufficient brushing, rinsing during manual pre-cleaning Repeat the processing If repeat processing is not successful Do not use instrument any longer and discard Water spots (lime or silicate deposits) Poor water quality Rinsing with distilled water not performed Penetration of cleaning agents containing silicate Repeat the processing 21

22 8 Customer service information address Mathys Ltd Bettlach Robert Mathys Strasse 5 P. O. Box 2544 Bettlach, Switzerland Tel Fax info@mathysmedical.com 22

23

24 Mathys Ltd Bettlach Robert Mathys Strasse Bettlach P. O. Box Switzerland Tel Fax info@mathysmedical.com Item no Mathys Ltd Bettlach

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15 Recommended Care, Cleaning and Sterilization Instructions for Reusable Instruments & Accessories Symmetry Surgical Inc. 3034 Owen Drive, Antioch, TN 37013 USA 1-800-251-3000 Fax: 1-615-964-5566 www.symmetrysurgical.com

More information

I.F.U Re/Processing Reusable Medical Devices

I.F.U Re/Processing Reusable Medical Devices I.F.U Re/Processing Reusable Medical Devices Distributed by: MANUFACTURER: GA 26-09-001 -EN /20130719 Contents 1 Overview of preparation methods... 3 2 Safety and responsibility... 4 3 Explanation of symbols...

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ Arthroscopy irrigation cannula Arthroscopy irrigation cannulas 65-201-00-65-202-00 01 Stopcock left 02 Spring Cap 03 Cannula CLEANING / STERILIZATION Immediately after use the instruments have to be soaked

More information

CLEANING INSTRUCTIONS

CLEANING INSTRUCTIONS Reusable Instrument Care Manual Pagoda Pedicle Screw System Manufacturer Ortho Development Corporation 12187 South Business Park Drive Draper, UT 84020 Phone 801-553-9991 Fax 801-553-9993 www.orthodevelopment.com

More information

Instructions for Use Spinal Punches IFU

Instructions for Use Spinal Punches IFU Indications for use Spinal punches are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Contraindiciations Instruments should not

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ s 65-201-00-65-202-00 01 Stopcock left 02 Spring Cap 03 Cannula CLEANING / STERILIZATION Immediately after use the instruments have to be soaked in a combined disinfection and cleaning solution in order

More information

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS INSTRUCTIONS FOR USE & STERILIZATION PMS PRECISION MEDICAL SPECIALITIES GMBH Kreuzstrasse 5 DE-78532 Tuttlingen PO Box 4607 DE-78511 Tuttlingen Fon: +49 74 61 13 131 Fax: +49 74 61 76 698 info@pms-tuttlingen.de

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS CAUTION: The following instructions are for all Sklar Coated stainless steel instruments. Read instructions prior to use. Improper

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks 65-227-01-65-240-12 1 Safety and responsibility Before using the instrument

More information

Operating / User Manual

Operating / User Manual UniARM Surgical Support System Operating / User Manual Version 1.1 MITAKA KOHKI Co., Ltd. 1. Introduction Introduction Thank you for purchasing the UniARM for use in surgical procedures. To use the device

More information

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية 1. POLICY STATEMENT: Cleaning, decontamination and sterilization method should be done in CSSD service facility not in the end user department side. 2. PURPOSE: To ensure the proper way of collections,

More information

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES Contents: Pagina 1 Introduction...2 2 Treatment instructions...2 3 Cleaning...2 3.1 Place of use...4

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches Caution Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine content.

More information

RAP-hip Hip stem extraction set

RAP-hip Hip stem extraction set Cleaning, sterilisation and maintenance RAP-hip Hip stem extraction set C Distributor: Marketed by: SAFRIMA AG Unterworbenstrasse 51 CH-3252 Worben Switzerland Phone: +41 32 387 05 91 Fax: +41 32 387 05

More information

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray Equipment Cleaning and Sterilization Table of Contents 1. WARNINGS AND PRECAUTIONS... 3 2. INSTRUMENT PREPARATION... 3 3. CAVITY PREPARATION

More information

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) 1 INTRODUCTION This paper is intended to give general

More information

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) Fiber Optic Cables Instructions for Use and Processing Instructions DFU-950-0031-00 Fiber Optic Cables Made in USA Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) 737 7611

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ General Information Use The Air drill System is a pneumatic powered system used for many applications orthopedic and trauma surgery. To ensure proper operation of the air drill, use only original attachment

More information

Recommendations for cleaning, decontamination and sterilisation. devices

Recommendations for cleaning, decontamination and sterilisation. devices Recommendations for cleaning, decontamination and sterilisation of MEDACTa INTERNATIONAL orthopedic devices 1 This document was prepared to provide cleaning, decontaminating and sterilising instructions

More information

RECOMMENDATIONS FOR CLEANING DEVICES

RECOMMENDATIONS FOR CLEANING DEVICES RECOMMENDATIONS FOR CLEANING D E C O N TA M I N AT I O N A N D S T E R I L I Z AT I O N O F M E D A C T A I N T E R N A T I O N A L O R T H O P E D I C DEVICES 1 This document was prepared to provide cleaning,

More information

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany Important information You receive a high-quality product by purchasing this sterile container system. lts proper handling and use will be described in the following. Please read this user manual thoroughly

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 14 Symbols on product packages Caution, general warning symbol Caution, see documentation

More information

THE BASICS OF DECONTAMINATION

THE BASICS OF DECONTAMINATION THE BASICS OF DECONTAMINATION Objectives Review the basic factors that impact cleaning and decontamination of surgical devices Present the fundamentals of manual and automatic cleaning Review various cleaning

More information

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications TABLE OF CONTENTS MANUAL PRE-CLEANING High Speed Handpieces 2

More information

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY Sonic Irrigator SA Automated Cannulated Instrument Reprocessing Sonic Irrigation Twin Cannulated Pulse Flush Systems Load performance optimisation Independent process monitoring EN ISO 15883/HTM 2030 compliant

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Irrespective of where the decontamination is performed or who it is performed by the

More information

Jennings Street School

Jennings Street School Infection Prevention and Control and Protocols Rationale Following standard infection control precautions can minimise the risk of norovirus outbreaks caused by person-to-person transmission in any institution

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

2015 Myco Industries, Inc Fax:

2015 Myco Industries, Inc Fax: Endoscope Cleaning and Handling It is highly important that you manually clean the endoscope immediately after it is removed from the patient, BEFORE it is automated or manually disinfected. Cleaning your

More information

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing Instructions for Use. Description and Intended Use Integra LifeSciences Corporation reusable surgical device sets consist of various

More information

Cleaning & Sterilization

Cleaning & Sterilization Handpieces & Attachments Instruments Cleaning & Sterilization System 7 System 6 System 5 CD4 CD3 SABO 2 SABO RemB CORE TPS Processing Instructions User/Patient Safety all contribute to the efficacy of

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 16 Symbols on product packages! Caution, general warning symbol Caution, see documentation

More information

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only V. Mueller en V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA Rx Only V. Mueller, CareFusion and the CareFusion logo are trademarks or registered trademarks of CareFusion Corporation

More information

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG Guide Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG Acceptance Cleaning, Disinfection Inspection, Packing Use T Sterilization t Storage Return

More information

REUSABLE ENDOSCOPY INSTRUMENTS (thermo resistant)

REUSABLE ENDOSCOPY INSTRUMENTS (thermo resistant) IMPORTANT ADVICES PLEASE READ CAREFULLY AND KEEP THE INSTRUCTIONS FOR FUTURE REFERENCE REUSABLE ENDOSCOPY INSTRUMENTS Field of application The following instructions are for all reusable endoscopy instruments

More information

Cleaning and Sterilisation for Burs, accessories and filling material

Cleaning and Sterilisation for Burs, accessories and filling material Cleaning and Sterilisation for Burs, accessories and filling material EN FOR DENTAL USE ONLY CLEANING AND STERILISATION PROCEDURE FOR BURS, ACCESSORIES AND FILLING MATERIAL 1) FOREWORD Devices that are

More information

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS CLEANING AND HANDLING OF WRIGHT INSTRUMENTS 150824-0 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt)

More information

When you sterilize an instrument set, you know what the standards are.

When you sterilize an instrument set, you know what the standards are. Sterilization & Infection Control The importance of cleaning in earnest A regular column on sterilization and infection control issues. When you sterilize an instrument set, you know what the standards

More information

The Do's and Don ts in the Instrument Sterilisation Process

The Do's and Don ts in the Instrument Sterilisation Process The Do's and Don ts in the Instrument Sterilisation Process Just like a skilfully placed filling, the dental sterilisation process requires fine attention to detail. One overlooked step during this critical

More information

Important information

Important information Version SE_023827 AK Date March 2015 Important information (with Cleaning and Sterilization Instructions) Important information 2 Basic Instructions on the Use of Synthes Implants and Instruments for Orthopedics

More information

Instructions for Use (GA16)

Instructions for Use (GA16) Instructions for Use (GA16) Instruments and Accessories for Arthroscopy Endoscopy Laparoscopy Resectoscopy Urology Hysteroscopy made from stainless steels as per DIN EN ISO 7153-1 and from high quality

More information

DRAFT. Module 4. Decontamination & Environmental Cleaning

DRAFT. Module 4. Decontamination & Environmental Cleaning Module 4. Decontamination & Environmental Cleaning Learning Objectives 1. Define the 3 steps of decontamination 2. List and describe the process for cleaning and disinfecting 3. Discuss safe management

More information

Cleaning and Sterilization for Instruments & Posts

Cleaning and Sterilization for Instruments & Posts Cleaning and Sterilization for Instruments & Posts EN FOR DENTAL USE ONLY CLEANING AND STERILIZATION PROCEDURE FOR ENDODONTIC FILES, HAND INSTRUMENTS, PINS AND POSTS, STAINLESS STEEL DRILLS, STAINLESS

More information

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician. Reprocessing Manual Fuse 1C Colonoscope Part Number FSE-056-EN-2.00 Released August, 2013. This manual includes complete reprocessing instructions for the Fuse 1C Colonoscope system. Please refer to the

More information

PERFORM Operating Document

PERFORM Operating Document PERFORM Operating Document Use and Maintenance of Tuttnauer Table-Top Autoclave PC-POD-CA-006-v03 Revision History Version Reason for Revision Date 01 New POD 13-Aug-13 02 POD Section 3.1.3 and 3.2 revised

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

Processing Instructions. Cleaning so easy.

Processing Instructions. Cleaning so easy. EN Processing Instructions. Cleaning so easy. 2 A good decision. Congratulations on your new tiologic easyclean Wash-Tray. Machine treatment of the entire instrument set is an essential component of quality

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. Flexible Nasopharyngoscope Operator Manual Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. 1. Introduction The flexible nasopharyngoscope is sold in unsterile condition as a reusable

More information

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices In BC Health Authorities THIS DOCUMENT IS INTENDED TO DESCRIBE BEST PRACTICES HEALTH

More information

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION CFSA CSSD Conference Durban ICC Wednesday 8 th March 2017 Using technology to improve patient safety THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION Helen Loudon Independent Infection

More information

WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS

WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS Always with you Contents Introduction 3 Levels of cleanliness 5 Cleaning and disinfection processes 6 Washer-disinfector for human-waste containers

More information

Manual Cleaning? SOBECC Round Table: Manual and Automated Cleaning Methods th July 2012

Manual Cleaning? SOBECC Round Table: Manual and Automated Cleaning Methods th July 2012 Manual Cleaning? SOBECC Round Table: Manual and Automated Cleaning Methods 11.40-12.30 27 th July 2012 Chirurgie-Instrumenten-Arbeitsgruppe Berlin Dr. med. Dipl.-Ing. Thomas W. Fengler What is so manual

More information

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106 MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES SUBJECT: Infection Control POLICY NO: 106 PURPOSE: The activities of collecting, processing and delivering an adequate supply of clean linens

More information

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G Cavitator Ultrasonic Cleaner Instruction Manual Model 20G Made in China 1333 South Claudina Street Anaheim, CA 92805 Toll Free: (800) 854 9305 Tel: (714) 533 2221 FAX: (714) 635 7539 Web site: www.mettlerelectronics.com

More information

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

IMPORTANT INFORMATION ON THE DEPUY SYNTHES For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE DEPUY SYNTHES MatrixNEURO RECONSTRUCTION MESH, STERILE MatrixNEURO PREFORMED MESH, STERILE GP2949-B-CAN DESCRIPTION The DePuy Synthes

More information

STERITITE INSTRUCTIONS FOR USE

STERITITE INSTRUCTIONS FOR USE STERITITE INSTRUCTIONS FOR USE If a rigid container system is used, the manufacturers instructions regarding set preparation and assembly should be followed. -ANSI/AAMI ST79: 2010 & A1:2010 Notes from

More information

Reusable Laryngoscope Blades and Handles

Reusable Laryngoscope Blades and Handles Manufacturer: Starkling Medizintechnik Method: Re-processing instructions Symbol: N.A. Device(s): 1 Green System Reusable Laryngoscope Blade with integrated Light Guide, Solid Stainless Steel Construction

More information

X99 Series. Dental Handpiece. Instruction For Use

X99 Series. Dental Handpiece. Instruction For Use X99 Series Dental Handpiece Instruction For Use 0413 Table of Contents Symbols... 2 Introduction... 3-4 Before Use... 5 Product description... 6 Technical Specifications... 7 Operation... 8-11 Hygienic

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS GENERAL - BEST PRACTICE CLEANING EQUIPMENT Sweep or rinse away loose debris. On floors, use a dust-attracting mop, not a broom Where possible, make use of vacuum (Clean bag daily) Remove spills (feaces,

More information

The Principles, The Process, and The Results:

The Principles, The Process, and The Results: The Principles, The Process, and The Results: Presented by: Fred Alston, CSPDT Eastern Region Clinical Sales and Support Manager April 21, 2017 I am an employee of Healthmark Industries Fraser, Michigan

More information

ASPIRE Laboratory Aspirator

ASPIRE Laboratory Aspirator ASPIRE Laboratory Aspirator USER MANUAL Rev 2/14/18 Accuris Instruments / Benchmark Scientific Ph: (908) 769-5555 E-mail: info@accuris-usa.com (C) Benchmark Scientific, 2018 THE ACCURIS ASPIRE LABORATORY

More information

Cavitator Ultrasonic Cleaner Instruction Manual

Cavitator Ultrasonic Cleaner Instruction Manual Cavitator Ultrasonic Cleaner Instruction Manual Made in China 1333 South Claudina Street Anaheim, CA 92805 Toll Free: (800) 854 9305 Tel: (714) 533 2221 FAX: (714) 635 7539 Web site: www.mettlerelectronics.com

More information

Electric steam steriliser

Electric steam steriliser Table of Contents Introduction... 2 Safety instructions and important notes.... 3 Description.... 6 Preparing for sterilisation.... 7 Operation.... 7 Cleaning the appliance and removal of limescale deposits...

More information

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com BioSonic Ultrasonic Cleaning Systems biosonic.coltene.com 2018 BioSonic An essential part of your Infection Control System Three Decades of Trusted Technology BioSonic UC150 BioSonic UC125 BioSonic UC50DB

More information

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope,

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope, Page 1 of 9 Legend 1 Ocular 2 Focussing ring 3 Working channel for flexible instruments 4 Bending protection 5 Deflectable, distal end 6 Lever for control of distal end 7 Fixation thread for pressure compensation

More information

INSTRUCTIONS CertoClav EL

INSTRUCTIONS CertoClav EL High Pressure-LABORATORY AUTOCLAVE Autoclave using steam for the treatment of material and laboratory processes e.g. sterilization As an aid in research and development Thermometer Lid handle with locking

More information

Cavitator Ultrasonic Cleaners

Cavitator Ultrasonic Cleaners Cavitator Ultrasonic Cleaners Instruction Manual (for model # 18G) 1333 South Claudina Street Anaheim, CA 92805 Toll Free: (800) 854 9305 Tel: (714) 533 2221 FAX: (714) 635 7539 Web site: www.mettlerelectronics.com

More information

LipoCollector II Operating Instructions

LipoCollector II Operating Instructions 1 of 26 LipoCollector II Operating Instructions ISO 13485 Operating instructions for LipoCollector II Item No. 650100 All rights reserved, particularly the right to reproduce and distribute as well as

More information

Reprocessing of surgical instruments

Reprocessing of surgical instruments Reprocessing of surgical instruments Reprocessing of surgical instruments Reprocessing Reprocessing Content 1 2 5... Introduction 6 7... Pre-treatment Microsoft Tag Readers Thanks to the Microsoft Tag

More information

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE Tested for optimal cleaning and performance ciso 15883 VALIDATED 2 GETINGE CLEAN GETINGE CLEAN THE ANSWER TO YOUR CLEANING PROBLEMS What does the word

More information

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUBJECT: ULTRASONIC CLEANER MONITORING DAILY DEPARTMENT: CPD, CSSD, SPA, MDRD, SPD, dental offices, medical offices and department using

More information

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032

MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES. SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032 MARSHALLTOWN MEDICAL AND SURGICAL CENTER ENVIRONMENTAL SERVICES SUBJECT: Room Cleaning Occupied Room (All Areas) PROCEDURE NO: 032 PURPOSE: To provide a systematic approach to remove the accumulation of

More information

O-PEP Hygiene-Beileger

O-PEP Hygiene-Beileger O-PEP Hygiene-Beileger 2013 PARI GmbH Spezialisten für effektive Inhalation, 018D5002-A 02/13 Hygienic re-use in the professional setting - 13 - en Information as of: 2013-02. We reserve the right to make

More information

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts Modular Slide Lock Handle Assembly 1 Movable Handle 2 Fixed Handle 3 Handle Access For Cleaning 4 Window 5 Handle Screw 6 Keeper 7 Nonconductive Luer Port Cap (shown covering the Luer Port) 8 Shaft Insulation

More information

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018 Page 1 of 6 PURPOSE: All equipment used for resident care will be appropriately cleaned, disinfected and/or sterilized to minimize the risk of transmission of infection by reducing micro-organism contamination

More information

Instructions. Portable Absorb Phlegm. Apparatus

Instructions. Portable Absorb Phlegm. Apparatus Instructions On Portable Absorb Phlegm Apparatus Please carefully read the instructions before attempting to operate this apparatus. Product Features 1. General Portable absorb phlegm apparatus designed

More information

CODE OF PRACTICE. No 11 CLEANING OF EQUIPMENT AND PREMISES

CODE OF PRACTICE. No 11 CLEANING OF EQUIPMENT AND PREMISES 1. MANAGEMENT CODE OF PRACTICE No 11 CLEANING OF EQUIPMENT AND PREMISES 1.1 Everyone in the food chain must be personally committed to ensuring that satisfactory standards are achieved. Requisite standards

More information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information Health Information U S E R G U I D E Home Respiratory Equipment Cleaning Guide This document is intended for Patients and Families for use at home The goal of the cleaning guide is to properly handle,

More information

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XP160 OLYMPUS GIF TYPE 160 OLYMPUS GIF TYPE Q160 OLYMPUS GIF TYPE Q160Z OLYMPUS GIF TYPE 1TQ160 OLYMPUS GIF TYPE XTQ160 EVIS EXERA COLONOVIDEOSCOPE

More information

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices 71381339 REVA CLEANING AND STERILIZATION BRO.indd 1 5/21/12 3:41 PM 71381339 REVA CLEANING AND STERILIZATION

More information

Operating Instructions

Operating Instructions Operating Instructions Elmasonic E Ultrasonic Cleaning Units english Elma Schmidbauer GmbH Gottlieb-Daimler-Str. 17 D-78224 Singen Phone +49 (0)7731 / 882-0 Fax +49 (0)7731 / 882-266 info@elma-ultrasonic.com

More information

Corporate. Laundry SOP. Document Control Summary

Corporate. Laundry SOP. Document Control Summary Corporate Laundry SOP Document Control Summary Status: Version: Author/Title: Owner/Title: Approved by: Ratified: Related Trust Strategy and/or Strategic Aims Implementation Date: Review Date: Key Words:

More information

Söring Ultrasonic Generator Sonoca 300

Söring Ultrasonic Generator Sonoca 300 0-0,9 bar 0 100 % 0 150 ml 0 100 % Schlauch befüllen Filling hose User Manual Söring Ultrasonic Generator Sonoca 300 Absaugung Aspiration Ultraschall Ultrasound Spülung Irrigation Information Pump on/off

More information

STERIS Instrument Cleaning Chemistries

STERIS Instrument Cleaning Chemistries STERIS is REACH Compliant. What is REACH? REACH is a European Union regulation concerning the Registration, Evaluation, Authorisation and restriction of Chemicals. It came into force on 1st June 2007 and

More information

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily.

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily. Suggested daily policy for monitoring the performance of an sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items with the LumCheck Cleaning verification test SUBJECT: Monitoring

More information

CPAP Washbasin Manual

CPAP Washbasin Manual CPAP Washbasin Manual www.mtts-asia.com CPAP-WB-MAN-01-REV3.02EN Company information MEDICAL TECHNOLOGY TRANSFER AND SERVICES CO., LTD No. 26, Lane 41, An Duong Vuong street, Tay Ho district, Hanoi, Vietnam

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS GENERAL - BEST PRACTICE CLEANING EQUIPMENT Sweep or rinse away loose debris. On floors, use a dust-attracting mop, preferably not a broom Where possible, make use of vacuum (Clean bag daily) Remove spills

More information

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories Table of Contents 1. Intended Use Statement... 3 2. Proper Use... 3 3. General Information...

More information

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use. Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1 ENGLISH Intended Use The Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes and Anoscopes are used to illuminate and

More information

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060)

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060) Instructions for use BA Optima Air motors BA602 (BA640081) BA604 (BA640060) Contents 1. Introduction....3 5 2. Safety notes...6 9 3. Product description... 10 11 4. Operation... 12 13 Assembly/Removal,

More information

Technical Data Monograph. Reliance Vision Single Chamber Washer/Disinfector

Technical Data Monograph. Reliance Vision Single Chamber Washer/Disinfector Technical Data Monograph Reliance Vision Single Chamber Washer/Disinfector Table of Contents 1. Introducing: the Reliance Vision Single Chamber Washer/Disinfector...3 2. Environmentally Conscious Design...3

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE THEATRE AND STERILE PROCEDURES WALLS/TILES - Wash all surfaces from top to bottom using F10SCXD, rinse with - Wash off visible dirt (e.g. blood) using F10SCXD CHAIRS - Trigger spray using F10SC and wipe

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE THEATRE AND STERILE PROCEDURES WALLS/TILES - Wash all surfaces from top to bottom using F10CLXD, rinse with - Wash off visible dirt (e.g. blood) using F10CLXD CHAIRS - Trigger spray using F10CL and wipe

More information

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see  for demo video Keeping you insafehands to avoid needlestick injuries Astek Innovations Ltd Astek House I Atlantic Street I Altrincham Cheshire I WA14 5DH I United Kingdom Telephone +44 (0)161 942 3900 I Fax +44 (0)161

More information

Commit 2 Clean TM/MC. Bloodborne Pathogens Awareness Program

Commit 2 Clean TM/MC. Bloodborne Pathogens Awareness Program Commit 2 Clean TM/MC Bloodborne Pathogens Awareness Program Commit 2 Clean TM/MC Bloodborne Pathogens Awareness Program Table of Contents Cleaning Minor Spills on Hard Surfaces... 2 Cleaning Major Spills

More information

Instrument Cleaning Chemistries Automated Washing Process

Instrument Cleaning Chemistries Automated Washing Process Instrument Cleaning Chemistries Automated Washing Process Procedure Pre-cleaning Prolystica Enzymatic Presoak and Cleaner Klenzyme Enzymatic Presoak and Cleaner ZolvStat Multi-Enzymatic Detergent Pre-Klenz

More information