ANGELUS (Ver 2.03) Patient Monitoring System

Size: px
Start display at page:

Download "ANGELUS (Ver 2.03) Patient Monitoring System"

Transcription

1 ANGELUS (Ver 2.03) Patient Monitoring System

2 CONTENT Contents of Manual 1 Meaning of Symbols used in this manual 2 Section 1. GENERAL Introduction Composition of unit Panel Description Basic Operation Concepts General Safety Information 28 Section 2. Preparations Preparation Flowchart Installation conditions Installation of ANGELUS Power Basic Operation of ANGELUS Description of Safety Symbols & Units Symbols 41 Section 3. Changing System Setup Settings Short-cut menu for easy use List and Explanation of the Setup settings Angelus System settings SpO2 Settings NIBP Settings Temperature Settings Trend Settings 72 Section 4. Necessary Settings before Monitoring Changing Date and Time 74 Section 5. Monitoring Screen Safety Precautions for Monitoring Overview 88

3 Section 6. Alarm Function What is an alarm Silencing alarm Suspending all alarms before occurrence Alarm Changeable Range Alarm Type Alarm Settings 93 Section 7. SpO2 Monitoring General Preparing for SpO2 Monitoring Monitoring SpO Changing SpO2 Settings 103 Section 8. NIBP Monitoring General Preparing for NIBP Measurement Measuring and Monitoring NIBP Error Message on NIBP Changing NIBP Settings 119 Section 9. Temperature Monitoring Preparing for temperature monitoring Changing Temperature Settings 138 Section 10. Maintenance Battery Pack SpO NIBP Temperature Cleaning and Disinfection the unit Yearly Inspection 151 Section 11. Specification 152

4 Contents of Manual This section provides a general overview of the unit and how to operate it. If you have not used an Angelus patient monitor before, read this section first. The patient monitor, Angelus, have several connectors for SpO2, NIBP, Temperature. This Manual contains all the information needed to operate our Patient Monitor, Angelus, which has been designed by giving most consideration on users convenience, and it provides exact information on the conditions of the patients including the measured values such as oxygen saturation in blood (SpO2), blood pressure (NIBP) and temperature. This manual is written to help users effectively operate Angelus. Each feature of this unit could be used more effectively, if relevant medical books referenced for clearer understanding of clinical meaning and pathology. In case some problems occur during the operation of our units, please contact local agent. 1

5 Meaning of Symbols used in this manual Symbols are used to specially emphasize the agreed details as follows. Users need to surely follow all the caution and notes listed in this Manual. In the event that the unit is damaged due to misuse or negligence by a user, the manufacturer or authorized dealers shall not be responsible for any damage or loss to the unit. Warning The title Warning is used to inform the users of possible causes that could inflict injury, death, or properly damage to the patients. Caution The title Caution is used to inform the users of possible causes that could inflict injury to the patients although it might not severe enough to cause deaths. Note The title Note is used to inform the users of items that are of importance in terms of installation, operation, or maintenance of the Unit although the failure does not inflict the bodily harm to the patients. Dangerous warning should never be included under the Note signal. 2

6 Section 1. GENERAL 1.1 Introduction General Warranty Period Request for Services How to Contact us Composition of Unit Panel Description Front & Operation Panel LED Indicators Rear Panel Side Panel Basic Operation Concepts Screen Displays General Safety Information Cautions for Electrical Safety Keeping & Storing Safety 29 3

7 1.1.1 General 1.1 Introduction Thank you for using Angelus (Patient Monitor). Please read this manual carefully before operating this unit for proper handling and maintenance, and get familiar with all the functions and features of this unit for safe handling and reliable performance. Please, try to fully understand the features and operation and maintenance methods of this unit prior to actual operation by thoroughly and carefully reading this manual. This is the only way to insure safe operation and reliable performance of the unit for a long time. KTMED Inc. only provides the reliable units to our customers. All of assembling, extending, adjusting or repairing of this unit shall be carried out only by the service personnel authorized by our local agent and KTMED Inc. Electrical connection or installation location has to comply with relevant regulations. Operate the unit as directed by this manual. This unit is useful to monitor unconscious or emergency patients. (It measures SpO2, Blood Pressure, and Temperature.) And this unit is restricted to one patient at a time. To insure the safety of patients, use only the parts and accessories recommended by us. In the event that this Unit needs to be used by connecting with other unit that is not listed in this Manual, be sure to notify us or our authorized dealers about this intent. It can be used while moving of patient with battery or DC. 4

8 1.1.2 Warranty Period Our Units are produced through out strict quality control, management, packed, and transported to insure the best installation. So you are recommended to do not transport or move the unit without KTMED s packing as you wishes. If not, it can be damaged by delivery. The warranty period for this unit is specified to be 2 years. (But accessories are specified to be one year.) Please contact with KTMED for service. When it malfunctions during the course of normal operation while under the Warranty Period, it will be repaired without any cost to the consumer at our company. We recommend that you have to check regular calibrations at least 24 months. In this case, please contact with KTMED. Clinical accuracy characteristics (manufacturer of the unit) are available from KTMED. When it malfunctions during the warranty period, please inform the service personnel on basic facts, such as model name, unit No., purchase date, and the brief description of the problem. If the defects in units or workmanship for two years from the purchased date, manufacturer will repair or replace your unit with new or rebuilt parts without any charges except below cases, After warranty period Defects from operating or handling carelessness Defects from a natural calamity Defects from repair or disassembled the unit by unauthorized person 5

9 1.1.3 Request for Services The services for KTMED Inc. units can be performed only by the personnel authorized by the Customer Service Dept. of KTMED or our local agent. In the event that the unit is serviced or tried to be serviced by an unauthorized third party during the warranty period, the warranty shall be nullified. The local agent is obliged to provide the services requested by the customers. When any abnormalities are found in the unit or a danger to bodily harm exists, the unit has to be repaired fast and adequately by authorized personnel. At time of unit malfunctions, please take following steps. Immediately contact our local agent which provides the repairing services for us as a Contractor. Please be prepared to inform the receiver about the manufacturer s label, serial number, date of original purchasing and brief description of malfunction or enquiry before making call. KTMED or local agent will make available on request circuit diagrams, component part lists, descriptions, calibration instructions or other information which will assist your appropriately qualified technical personnel to repair those parts of unit which are designated by KTMED Inc repairable. The customer is responsible for any freight charges, when unit is shipped to KTMED Inc. for servicing (this includes all customs charges). Our local agent will first try to analyze and solve the problem while talking to you on the phone if possible. 6

10 1.1.4 How to Contact us Contact us at following phone No. and address for various services and unit supplies. We are at your service all the time. Do not hesitate to call us any inquiries at all. KTMED Inc. Tel: Fax: Service Request When request for repairing service, please find out the unit s model name, unit serial No. and brief description of problem in advance, and then contact our Customer Service Dept. 7

11 1.2 Composition of unit 1 Angelus main unit: 1ea 2 Power cord (110V /220V): 1ea 3 NIBP cuff for Child: 1ea 4 NIBP cuff for adult: 1ea 5 NIBP cuff hose: 1ea 6 SpO2 reusable sensor for adult: 1ea 7 SpO2 extension cable: 1ea 8 IV pole clamp: 1ea 8

12 1.3 Panel Description Front& Operation panel This section provides an overview of the operator controls, indicators and symbols. Detailed use of each control is described in other sections of this manual. 1 Malfunction Big LED 2 3 Digit Dual Color FND 3 LED indicators 4 4 Digit FND 5 10 BAR LED 6 Speaker 7 [Start/ Stop] button 8 [Trend] button 9 [[Setup/ Menu] button 10 [Esc/silence] button 9

13 1.3.2 LED indicators LOW BATTERY: When batteries are low, the LOW BATTERY indicator will be steadily illuminated and audio-able alarm goes off. With connecting AC power, alarm will be stop automatically NO SIGNAL: When SpO2 sensor can not read the finger or pulse rate, the NO SIGNAL will be steadily illuminated. In this case, condition of SpO2 sensor and patient should be checked. ALARM SILENCE: When alarm sound occurs, press this button, and then alarm sound doesn t go off. FINGER FAULT: The FINGER FAULT is steadily illuminated until SpO2 sensor recognizes and read the finger signal. SENSOR FAULT: The SENSOR FAULT is steadily illuminated until SpO2 sensor is connected with ANGELUS correctly. ADULT: When NIBP is set in adult mode, the ADULT is steadily illuminated. NEONATE: When NIBP is set in neonate mode, the NEONATE is steadily illuminated. 10

14 1.3.3 Rear panel 1 IV Pole Clamp Holder 2 Internal Battery Cover 3 Battery Cover Fixing Bolt 4 Description of Specifications 5 Earth 6 External AC power Input Source 7 Hand Holder 11

15 1.3.4 Side panel 1 Hand Holder 2 IV POLE CLAMP HOLDER 3 SpO2 SENSOR Socket 4 NIBP Air Coupler 5 Temperature Sensor Socket 12

16 1.4 Basic Operation Concepts Screen displays Following are the screens available on the Angelus. For details about the individual screens see the appropriate section. The shadow of the previous screen may remain for a few minutes after changing the screen. Check the following items in the unit and measuring sensors, before operating the unit. Check whether there exists any mechanical danger. Inspect the cable which connects to outside, and power cord, and accessories. Check whether the unit and sensors are connected exactly. 13

17 1 Initial screen If you press located on the front of the unit, the unit will turn itself on. If you press once more, the unit will turn itself off. When turning on the unit, version information with melody will be displayed for a few seconds as below picture. If the unit is not turned on or indicates low battery on the LED of status LED with an alarm sound, the unit needs to be recharged with an AC power source. After the initial display, the unit is ready to measure SpO2, Pulse Rate, Blood Pressure, and Temperature. Angelus Ver The unit indicates pulse rate, SpO2, blood pressure, temperature numerical value If the unit is not turned on or battery lamp on red with alarm sound, the unit needs to be recharged with an AC power source. 2 Display when you press the Setup/Menu button 1) Press [Set up] button for more than 2 seconds and press [Up/Down] button to change the set values. 14

18 2) In this mode, these UP/DOWN, Confirm, Esc/silence buttons are used for setting the unit s parameters. [UP/DOWN] buttons are used for scrolling the menu. [Confirm] button is used for selecting one of them. [Esc/silence] button is used for ESC button to exit. 3 Set up the System Volume 15

19 Further information can fully be explained in the System volume set up in section

20 4 Set up the Alarm Alarm Sound Volumes Alarm Interval Further information can fully be explained in the Alarm set up in section and Alarm settings in section

21 5 Set up the Time Year Month Date AM/PM 18

22 Hour Minute Second Further information can fully be explained in the Changing Date and Time in section

23 6 Set up the Alarm Limit Pulse rate alarm limit - High limit - Low limit Further information can fully be explained in the Changing the Pulse rate Alarm limit in section

24 SpO2 alarm limit - High limit - Low limit Further information can fully be explained in the Changing the SpO2 Alarm limit in section

25 NIBP alarm limit - Systolic high limit - Systolic low limit 22

26 - MAP high limit - MAP low limit 23

27 - Diastolic high limit - Diastolic low limit Further information can fully be explained in the Changing the NIBP Alarm limit settings in section

28 Temperature alarm limit - Temperature high limit - Temperature low limit Further information can fully be explained in the Changing the Temperature Alarm limit in section

29 7 Set up the Blood Pressure (NIBP) NIBP Mode or Further information can fully be explained in the Selecting the cuff type in section NIBP Interval Further information can fully be explained in the Selecting the interval time in section NIBP Cuff Exhaust 26

30 or Further information can fully be explained in Selecting the Exhaust type in section Set up the Temperature Temperature Unit Further information can fully be explained in the Changing the temperature unit in section

31 1.5 General Safety Information Cautions for Electrical Safety Before operation of the Unit, make sure to check following items: Never use this unit in the presence of any flammable anesthetic gas, concentrated oxygen or hyperbaric oxygen. Failure to follow this warning may result in explosion. Never use the unit in a high-pressure oxygen medical care tank. Failure to follow this warning may cause explosion or fire. Whether power supply is appropriate.(ac 100~240V, 50/60Hz) Confirm it before connecting to an outlet. Whether connections (power line or selected unit) are properly made to the unit. Whether the unit is properly grounded. Inaccurate measurement may occur when the sensor sites are exposed to excessive amounts of patient movement. To ensure patient safety, the user should use only the parts and accessories are supplied or recommended by KTMED Inc. If other company s sensors are used for Angelus, it may damage to the monitor. Note The Unit should be placed far from generator, X-ray unit, broadcasting unit, or transmitting wires, so as to prevent the electrical noises from being generated during the operation. When these units are near the unit, it can be inaccurate measurements. 28

32 1.5.2 Keeping & Storing Safety Please read this manual carefully before operating. Don t operate or store the Unit under the following environments. Store the damp location, and operate the Unit with wet hand. The place where the unit is exposed to direct sunlight. The place where the unit is closed to electrical heating apparatuses. The place where the temperature fluctuation is rather big. (The temperature falls below -20 C or exceed 60 C.) The place where moisture level could go up considerably or where air is not ventilated properly. (Humidity is above 75 %.) The place where the unit is exposed to excessive shocks or vibrations. The place where the unit is exposed to chemical or explosive gas. Disassembly of Unit should be done only by the authorized personnel; otherwise, we will not be obliged to provide any services at all. Make sure that this unit prevents dust and especially metal debris from penetrating in. The place where the atmospheric pressure falls below 70KPa (700mbar) or exceed 106KPa (1060mbar). The place where the unit is exposed to heavy density oil vapor. The place where the unit is exposed to salty atmosphere. If this unit is used with other electrical unit such like ultrasound surgical at the same time, it maybe breakdown. Use sensor and cable that have connected with unfit voltage or damaged ones. Use while testing of MRI or CT etc, it is harmful to patient and might be occurred unreasonable results. Standard operation conditions are as follows. Temperature: 0 ~ 50 C Humidity: 10~ 95% Standard storage and transportation conditions are as follows. Temperature: -20~ 60 C Humidity: 10~ 95% 29

33 Section 2. Preparations 2.1 Preparation Flowchart Installation conditions Warnings Cautions Installation of ANGELUS Power connection Connection of Measuring Sensor with Accessory Power AC or DC Power Source Selection External Power Internal Power Power Source Battery Specifications Basic Operations of ANGELUS How to Operate ANGELUS Description of Safety Symbols & Unit Symbols 41 30

34 2.1 Preparation Flowchart You may not need to do all these. 1) Install the unit and do the procedures in section 2. 2) Check or change any initial settings on the Setup screen. These items usually do not need to be changed. Refer to Section 3. 3) Check or change the necessary settings before monitoring in section 4. Date and time System setup (Sound Volume) 4) Prepare the unit (SpO2 sensors, NIBP cuffs, Temp sensors, etc.) for monitoring individual parameter and check or change the settings for each parameter. Refer to section 7, section 8, and section 9. 31

35 2.2 Installation conditions Warnings Never use this unit in the presence of any flammable gas, concentrated oxygen or hyperbaric oxygen. Failure to follow this warning may result in explosion. Connect only the specified instruments to the connector or connectors marked by following the specified procedure. Otherwise, electrical leakage current may harm the patient and operator Cautions Avoid collision when moving the monitor on a cart. Strong impact may damage the monitor. Avoid location where the monitor is sprinkled with liquids. Avoid direct sprinkling, spray or moist air from inspirator or a humidifier. If fluids are accidentally spilled on the unit, take the unit out of service and check for damage. Avoid exposing the unit to direct sunlight. And do not place the unit in dusty area. Connect the power cord to an AC outlet which can supply enough AC current to the monitor. The unit can not function properly with low current. Do not place the unit in an MRI examination room. The monitor may not function properly, or noise from the unit may interfere with the MRI. And do not use an electrical blanket. It may affect monitoring. When there is any problem on the unit, turn off the power immediately and disconnect the power cord from the AC outlet. Take the Monitor out of service and check for damage. 32

36 2.3 Installation of Angelus Power Connection Put the power cord into the power cable of Angelus. (Put the power cord to No. 6) Select the power source between an external AC and DC. When you use an external AC power source, connect power cord to AC INPUT located at the rear of the unit. In the case of an external DC power source, connect the power, 12~24V 3A, to its connector of the rear part. 33

37 2.3.2 Connection of Measuring Sensor Connect the relevant parameter sensors to the appropriate connector. If you want to know further information, refer to section 7, 8, and 9. Warning Only use the SpO2 sensors, NIBP cuffs, Temperature probes supplied with the unit. If not, it may cause some form of harmful effects to the patient or its user. For more detailed information concerning this situation, please contact KTMED or local agent. 34

38 2.4 Power AC or DC Power Source Selection The unit can operate on either battery or AC power. Select the power source between an external AC and DC. When you use an external AC power source, connect power cord to AC INPUT located at the rear of the unit. In the case of an external DC power source, connect the power, 12~24V 3A, to its connector of the rear part. For example, when transferring a patient, the unit automatically switches to battery power. The battery is charged when the power cord is plugged into an AC outlet and the AC current is supplied to the monitor. The battery is also charged during monitoring. The monitor can operate for about 2 hours with a fully charged battery pack when: Charged and used in normal temperature (about 25 ) Recorder is not used No alarm occurs No NIBP measurement 35

39 2.4.2 External Power Warning Only use the provided power cord. Using other power cords may result in electrical shock or other injury to the patient and operator. When the provided power cord cannot be used, operate the monitor on battery power. 1 Power Connection Put the power cord into the power cable of Angelus. (You put the power cable to No.6) Connect the provided power cord to the AC SOURCE connector on the rear panel of the unit and plug the cord into an AC outlet. When the AC power is supplied to the unit, the AC power lamp on the front panel lights and a buzzer sounds. 36

40 2.4.3 Internal power Angelus has a battery pack which is composed of 6 pieces of the NIMH, 1.2V. This unit can be operated about 2hours with a fully charged battery. When you connect the unit with a direct DC power source, the batteries are automatically charged. It takes about 6 hours to fully recharge a battery. There would be nothing happened even if the unit is connected to DC power with fully charged batteries at the same time. The condition of the internal power source is indicated by battery lamp. Battery pack is charged fully, green light comes on. While battery pack is discharged, red light comes on. If you want to recharge batteries, use a DC power source. If battery pack is recharged enough time, but red light still comes on, then the battery pack should be replaced. You have to contact to KTMED or local agency. Replacement internal of batteries Battery pack is fixed with a connector Note You should use the batteries supplied by KTMED only. Normally, the battery should be changed at least every year. 37

41 2.4.4 Power Source 1 AC power source When you connect AC power to the unit, the lamp on the front panel will be blinking. Use AC 100~240V, 50/60Hz. If the unit is supplied out of these ranges of power, the unit might be damaged. AC power is supplied to the unit, the green LED AC power light in the front panel is turned on. And when the unit is not in use, it is automatically changed to auto recharge mode. 2 DC power source This unit is possible to be used in a car with DC power. Connect DC power with DC INPUT connector on the bottom side. In this case, Power should be DC 12~ 24V, 3A Battery Specifications When the unit is turned on while AC power is cut off, the power is supplied from the battery and the LED battery power indicator at the upper section of the unit is turned on. In case battery power is running out. When the battery capacity goes below 12V, the alarm maybe go off, and battery lamp will blink for a few minute, and then the power will be turned off. Therefore, AC power has to be supplied soon in order to prevent the power shortage. Recharging period: It takes about 6 hours. Continuous Operation Time per full recharge: 2hours. 38

42 2.5 Basic Operations of Angelus Followings are the screens available on the Angelus. For details about the individual screens, see the appropriate section. Check the following items in the unit and measuring sensors, before operating the unit. Check whether there exists any mechanical danger. Inspect the cable which connects to outside, and power plug, and accessories. Check whether the unit and sensor is connected exactly. 39

43 2.5.1 How to operate Angelus Note: If you press [Setup/Menu] button more than 2 seconds, you can start System mode. If you want to use short-cut menu, press [setup/menu] button as usual. Also, if you want to know short-cut menu, refer to the Short-cut menu for easy use in section ) If you press located on the front of the unit, the unit will turn itself on. If you press once more, the unit will turn itself off. 2) If you want to change the set values while in use, use [Setup] button, [Up/Down] button, and [Confirm] button to change the set values. 3) The unit should be connected with the measuring sensor. 4) If the unit is not turned on or indicates low battery on the LED of status LED with an alarm sound, the unit needs to be recharged with an AC power source. After the initial display, the unit is ready to measure SpO2, Pulse Rate, Blood Pressure and Temperature. 40

44 2.6 Description of Safety Symbols & Unit Symbols The International Electro-technical Commission (IEC) has established a set of symbols for medical electrical unit regarding connections or warnings of potential hazards. The classification and symbols are shown below; Symbol Description Attention, consult ACCOMPANYING DOCUMENTS Date of manufacture Serial Number Type BF Equipment Equipotential terminal Power stand-by button Defibrillation-Proof Type BF applied parts Temp SpO2 Alarm SET UP Body Temperature connector Oxygen Saturation connector Alarm Button for changing to system setup 41

45 KTMED s model Angelus is classified as; Ordinary unit without protection against ingress of water. Unit not suitable for use in presence of a flammable anesthetic mixture by standard of IEC (Safety of Electric Medical Unit) Unit that complies with class B for noise-emission, Level B for noise-immunity, per standard of IEC (Electromagnetic Compatibility Requirements) Applied parts : BF applied parts. Protection against harmful ingress of water: ordinary IPXO. Degree of safety in the presence of flammable anesthetics mixture with air or with oxygen or with nitrous oxide. Not suitable for use in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide. 42

46 Section 3. Changing System Setup Settings 3.1 Short-cut menu for easy use NIBP Set up with [Confirm] button Selecting the NIBP cuff type Changing the NIBP Interval time Alarm Limit set up with [Setup] button Changing Pulse rate Alarm Limit Changing SpO2 Alarm Limit Changing NIBP Systolic Alarm Limit Changing NIBP Diastolic Alarm Limit Changing NIBP Mean Alarm Limit Changing Temperature Alarm Limit Alarm Silence System Menu List and Explanation of the Setup settings List of all settings Menu button ANGELUS SYSTEM Settings System Volume Setup Alarm Setup Clock Setup SpO2 Settings NIBP Settings Temperature Settings Changing the temperature unit Trend Settings View mode Erase mode 72 43

47 This section explains how to change setting after the unit is installed and before monitoring. The initial setting on the Setup screen must be changed before monitoring. Change these settings while monitoring interrupts monitoring. All other settings can be changed at any time without interrupting monitoring. This section also explains how to initialize the system. This procedure returns all settings to the factory default settings and deletes all stored data memory. 44

48 3.1 Short-cut Menu for easy use If you want to use short-cut menu, press [setup/menu] button as usual. These are 2 short-button menus for easy manipulation. We use existing menu buttons for this operation NIBP Set up with [Confirm] button Users can setup the NIBP Mode and NIBP Interval easily. For manipulation, the users should use [Confirm] button and [Up/Down] button Selecting the NIBP cuff type 1) Press the [Confirm] button, the Angelus NIBP Adult menu is displayed. 2) Press the [Up/Down] button, then select Adult or Neonate. 3) Press the [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 45

49 Changing the NIBP Interval time 1) Press [Confirm] button one more time, NIBP Interval menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value. (Interval time range: Off(0)/1/3/4/5/10/15/30/45/60/90/120/240 minute) 3) Press the [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 46

50 3.1.2 Alarm Limit set up with [Setup] button Users can setup Alarm Limit in each parameter easily. For manipulation, the users should use [Setup] button and [Up/Down] button. When the users press [Setup] button, the Pulse Alarm Limit (High) menu is displayed. If the users press this button step by step, Pulse Alarm Limit (Low) is displayed. The next is SpO2 High/Low, and the last is temperature. In other words, manipulation process is step by step Changing Pulse rate Alarm Limit 1) Press [Setup] button continuously until Pulse rate high Limit menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value from 26 to

51 3) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 4) Press [Setup] button continuously until Pulse rate low Limit menu is displayed. 5) Press the [Up/Down] button, then choose appropriate value from 25 to ) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 48

52 Changing SpO2 Alarm Limit 1) Press [Setup] button continuously until SpO2 high Limit menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value from 36 to ) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 49

53 4) Press [Setup] button continuously until SpO2 low Limit menu is displayed. 5) Press the [Up/Down] button, then choose appropriate value from 35 to 99. 6) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 50

54 Changing NIBP Systolic Alarm Limit 1) Press [Setup] button continuously until NIBP Systolic high Limit menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value from 31 to 300. (In case of neonate mode, alarm limit become from 16 to 150) 3) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 51

55 4) Press [Setup] button continuously until NIBP Systolic low Limit menu is displayed. 5) Press the [Up/Down] button, then choose appropriate value from 30 to 299. (In case of neonate mode, alarm limit become from 15 to 149) 6) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 52

56 Changing NIBP Diastolic Alarm Limit 1) Press [Setup] button continuously until NIBP Diastolic high Limit menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value from 31 to 300. (In case of neonate mode, alarm limit become from 16 to 150) 3) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 53

57 4) Press [Setup] button continuously until NIBP Diastolic low Limit menu is displayed. 5) Press the [Up/Down] button, then choose appropriate value from 30 to 299. (In case of neonate mode, alarm limit become from 15 to 149) 6) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 54

58 Changing NIBP Mean Alarm Limit 1) Press [Setup] button continuously until NIBP Mean high Limit menu is displayed. 2) Press the [Up/Down] button, then choose appropriate value from 31 to 300. (In case of neonate mode, alarm limit become from 16 to 150) 3) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 55

59 4) Press [Setup] button continuously until NIBP Mean low Limit menu is displayed. 5) Press the [Up/Down] button, then choose appropriate value from 30 to 299. (In case of neonate mode, alarm limit become from 15 to 149) 6) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 56

60 Changing Temperature Alarm Limit 1) Press [Setup] button continuously until Temperature high Limit menu is displayed. 2) Press the [Setup] button, and then choose appropriate value from 15.1 to 45. (In the temperature display LED) 3) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 57

61 4) Press [Setup] button continuously until Temperature low Limit menu is displayed. 5) Press the [Setup] button, and then choose appropriate value from 15.0 to (In the temperature display LED) 6) Press [ESC] button or 10 seconds later (automatically), setup is stored and returned to the measuring mode. 58

62 3.1.3 Alarm Silence Although the measurement value is out of alarm limit, if users don t want to use Alarm sound, the users can disable Alarm sound. However, if there is an error in the sensor or system, the alarm sound is on. a) Alarm On Setup Press [Alarm Silence] button for 2 seconds. Then, Alarm LED is flashed. In this mode, the alarm sound is not generated even if the measurement value is out of alarm limit. b) Alarm On Setoff Press [Alarm Silence] button for 2 seconds. Then, flashed Alarm LED is off. In this mode, the alarm sound is returned to normal mode System Menu Press the Setup button for 2 seconds, you can enter System Menu. In this mode, users can setup all parameters described in the manual. Note: If you want to use short-cut menu, press [setup/menu] button as usual. If you press [Setup/Menu] button more than 2 seconds, you can start System mode. Further information can fully be explained in the SpO2 monitoring in section 7, NIBP monitoring in section 8, Temperature monitoring in section 9. 59

63 3.2 List and Explanation of the Setup Settings List of all settings Various parameters are user selectable from set up menu. The menu of ANGELUS is constructed as below explanation. The first menu is categorized as Angelus System, SpO2 Setup, NIBP Setup and Temperature Setup. Under each category, the user can set up the parameters Menu button We have 4 menus for setting the unit s working parameters. If you press [Setup/Menu] button more than 2 seconds, you can see the menu mode. In this mode, these [Up/Down], [Confirm], [Esc/silence] buttons are used for setting the unit s parameters. [Up/Down] buttons are used for scrolling the menu. [Confirm] button is used for selecting one of them. [Esc/silence] button is used for ESC button to exit. The menu button allows to you: 1 Angelus SET: There are three modes. System volume: Adjust the Angelus sound volume. Alarm Setup: Adjust the alarm volume. Clock Setup: Clock setup the same way as on a typical watch. 2 NIBP SET: There are adult and neonate modes. You can choose according to the patient age. Furthermore, it can be chosen interval time, alarm limit value, and exhaust type. MODE: There are adult and neonate modes. You can choose a proper type for each patient. INTERVAL: Adjust the NIBP interval time of automatic NIBP measurement. Systole: Adjust the Systolic blood pressure alarm limit values. Diastole: Adjust the Diastolic blood pressure alarm limit values. MAP: Adjust the Mean of blood pressure alarm limit values. Exhaust: Choose linear or step type. 60

64 3 SpO2 SET: You can adjust the SpO2 value and pulse rate alarm limit values. Pulse rate limit: Adjust the pulse rate alarm limit value. SpO2 limit: Adjust the SpO2 alarm limit value. 4 Temperature SET: There are Celsius and Fahrenheit modes. You can choose a desired mode. Furthermore, it can be chosen alarm limit value. Temperature unit: Choose Celsius or Fahrenheit. Temperature limit: Adjust the Temperature alarm limit value. 61

65 List of all settings Menu Sub menu 1 Setup/Menu Angelus SET SpO2 SET NIBP SET Temper ature SET Sub menu 2 Sub menu 3/Value Value System Volume 0~9 Alarm Setup Alarm volume 0~9 Alarm Interval 1~60 Clock Setup Year 0~99 Month 1~12 Date 1~31(29,30) Hour 1~12 Minute 0~59 Second 0 Pulse rate limit High limit 26~255 LOW limit 25~254 SpO2 limit High limit 36~100 LOW limit 35~99 NIBP mode ADULT/NEONATE NIBP Interval OFF(0)~ 60 Systole High limit(neonate) 16~150 LOW limit(neonate) 15~149 High limit(adult) LOW limit(adult) MAP High limit(neonate) 16~150 LOW limit(neonate) 15~149 High limit(adult) LOW limit(adult) Diastole High limit(neonate) 16~150 LOW limit(neonate) 15~149 High limit(adult) LOW limit(adult) Celsius High limit 15.1~45.0 LOW limit 15.0~44.9 Fahrenheit High limit 59.1~113.0 LOW limit 59.0~

66 3.3 Angelus SYSTEM settings Set up Variation range Comments System 0/1/2 9 Adjustment of system volume volume Alarm 1/2 9 Adjustment of alarm volume volume Alarm Off(0)/1/ 59/60 Adjustment of alarm interval time/minute Interval Clock Year: 00~99 Month: 1~12 Date: 1~31 Adjustment of Clock AM/PM Hour: 1~12 Minute: 0~59 Second: 0 Esc/Silence EXIT Return to the previous screen Angelus system setup MAP Angelus System System volume Alarm set up Clock set up 1. To select desired mode, press up/down button. 2. Then press confirm button. 63

67 3.3.1 System Volume Set up System volume setup MAP System volume 0~9 1. To select desired volume level, press up/down button. 2. Then press confirm button. 1) Press Setup/Menu button, and using [Up/Down] button, move to Angelus System mode, then press [Confirm] button as below picture. 2) To select System volume mode, using [Up/Down] button then press [Confirm] button. 64

68 3) Using [Up/Down] button, set the volume level from 0 to 9, then press [Confirm] button. 65

69 3.3.2 Alarm Set up When an alarm sounds, you need to identify the alarm. And act appropriately, according to the cause of the alarm. Identify which monitor is in alarm Check the patient s condition Identify the cause of the alarm Silence the alarm, if necessary When alarm condition id eliminated check that the alarm reset Alarm setup MAP Alarm set up Alarm volume Alarm interval 1~9 Off(0)~60 1. To select desired mode, press up/down button. 2. Then press confirm button. 3. To select desired value, press up/down button. 4. Then press confirm button. Further information can fully be explained in the Alarm volume setup in section 6.6.1, Alarm interval time setup in section

70 3.3.3 Clock Set up You can set the date and time to be displayed on the screen. Once it has been set, the internal clock retains the setting even when the system is turned off. Clock setup MAP Clocksetup Year Month Date AM/PM Hour Minute Second 00~99 1~12 1~31 AM/PM 1~12 0~ To select desired mode, press up/down button. 2. Then press confirm button. 3. To select desired value, press up/down button. 4. Then press confirm utton. Further information can fully be explained in the Changing date and time in section

71 3.4 SpO2 Settings When the ANGELUS is first powered on, the alarm limits are set to the factory default values. The following table lists alarm limit range for the SpO2 value and pulse rate. Parameter High Low SpO2 Value Pulse Rates To change an existing alarm limit, perform the following procedures should be performed. SpO2 Setup MAP Angelus SpO2 Pulse Alarm limit SpO2 Alarm limit 1. Press setup/menu button more than 2 seconds. 2. To select desired mode, press up/down button. 3. Then press confirm button. Further information can fully be explained in the SpO2 Monitoring in section 7. 68

72 3.5 NIBP Settings When you adjust the cuff type, NIBP measuring interval time, and cuff exhaust type, you should use this mode. Set up Selection Range Comments Mode Adult/ neonate Cuff type Exhaust Linear/Step Exhaust type Interval Off(0)/1/2/3/ 60min Set up the time interval when measured automatically Esc/Silence EXIT Return to the previous screen NIBP Setup MAP Angelus NIBP Mode 1. Press setup/menu button more than 2 seconds. 2. To select desired mode, press up/down button. Interval Systole Alarm Limit Diastole Alarm Limit Mean Alarm Limit Exhaust Further information can fully be explained in the NIBP Monitoring in section 8. 69

73 3.6 Temperature settings Temperature can be monitored by attaching the probe to the patient and using the TEMP connector in the monitor. Temperature setup MAP Angelus Temperature Unit Alarm Limit(c) Alarm Limit (F) Celsius High Limit High Limit Fahrenheit Low Limit Low Limit 1. To select desired mode, press up/down button. 2. Then press confirm button Temperature unit type Set Up Selection Comments Range Celsius Celsius Scale for the measuring temperature, in which water freezes at 0 degrees and boils at 100 degrees. It is represented by the symbol C. Fahrenheit Fahrenheit Scale for the measuring temperature, in which water freezes at 32 degrees and boils at 212 degrees. It is represented by the symbol F. Esc/Silence EXIT Return to the previous screen 70

74 Temperature Unit setting MAP Temperature Unit Celsius Fahrenheit 1. To select desired unit mode, press up/down button. 2. Then press confirm button. Further information can fully be explained in the Temperature Monitoring in section

75 3.7 Trend Settings Trend function allows users to check the stored data on a patient as well as the longterm variation of patient data, showing the long-term trend of patient s condition at any time. (But, printout is possible only when a printer is equipped.) Press [TREND] button which on the below side of front panel. After selecting TREND section, there are two modes. (View, Erase modes) Set up Selection Range Comments View If you want to see the trend mode, you choose this mode Erase CANCEL/ DONE If you want to erase the measurement record, you choose done. If not, you choose cancel. Exit EXIT Return to the previous screen View mode To select a View mode, pressing [Up/Down] button then pressing [Confirm] button. DONE/CANCEL mode will be displayed on the screen. Select DONE mode to finish setting up TREND. If you want to see the stored data, press [Up/Down] button Erase mode When monitoring the patient is no longer required, delete the data. The patient data on the review screens are deleted, when you choose ERASE mode. 1) After selecting TREND section, press [confirm] button. After selecting TREND mode, press [Up/Down] button. ERASE mode will be displayed on the screen. Selecting DONE mode can delete measurement record. 72

76 Section 4. Necessary Settings before Monitoring 4.1 Changing Date and Time Year Month Date AM/PM Hour Minute Second 83 73

77 4.1 Changing Date and Time When the monitor turns on, the current time is displayed in the lower right corner of the screen Year 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Following menu will appear on the screen for setting up year of the unit. 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 3) To select the value for desired the YEAR by pressing [Up/Down] button, press [Confirm] button to complete the setup. 74

78 4) Using [Up/Down] button, set year from 01 to 99. Then press [Confirm] button Month 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up month of the unit. 75

79 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 3) To select the value for desired the Month by pressing [Up/Down] button, then press [Confirm] button to complete the setup. 4) Using [Up/Down] button, set month from 1 to 12. Then press [Confirm] button. 76

80 4.1.3 Date 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up date of the unit. 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 77

81 3) To select the value for desired the Date by pressing [Up/Down] button, then press [Confirm] button to complete the setup. 4) Using [Up/Down] button, set date from 1 to 31. Then press [Confirm] button AM/ PM 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up AM or PM of the unit. 78

82 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 3) To select between AM and PM, by pressing [Up/Down] button. Then press [Confirm] button to complete the setup. 79

83 4.1.5 Hour 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up hour of the unit. 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 80

84 3) To select the value for desired the Hour by pressing [Up/Down] button, then press [Confirm] button to complete the setup. 4) Using [Up/Down] button, set hour from 1 to 12. Then press [Confirm] button Minute 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up minute of the unit. 81

85 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 3) To select the value for desired the Minute by pressing [Up/Down] button, then press [Confirm] button to complete the setup. 4) Using [Up/Down] button, set minute from 0 to 59. Then press [Confirm] button. 82

86 4.1.7 Second 1) Press [Setup/Menu] button more than 2 seconds. Using [Up/Down] button, move to Angelus System mode. Then press [Confirm] button. Then following menu will appear on the screen for setting up second of the unit. 2) Using [Up/Down] button, move to Clock Setup mode. Then press [Confirm] button. 83

87 3) To select the value for desired the Second by pressing [Up/Down] button, then press [Confirm] button to complete the setup. 4) Using [Up/Down] button, set second only 0 Then press [Confirm] button. 84

88 Section 5. Monitoring Screen 5.1 Safety Precautions for Monitoring Using an Electro Surgery Unit Using a Defibrillator Overview Monitoring Screen Trend mode (Review Screen) 88 85

89 This section explains how to monitor the patient s data. Before monitoring the patient: Prepare the patient and equipment Before monitoring a new patient, follow the flowchart Preparation in section 2 Read the safety precautions in the Safety Precautions for Monitoring in section 5.1 In this section: Overview gives general information for all monitoring. 86

90 5.1 Safety Precautions for Monitoring Before beginning monitoring, observe the following safety precautions and the safety precautions in section 7, 8, 9 for SpO2, NIBP, and Temperature monitoring parameters Using an electro surgery Unit Warning Electrosurgical units (ESU) emit a lot of RF interference. If the unit is used with an ESU, RF interference may affect the monitor operation. Place the unit as far as possible from the ESU. Locate them on opposite sides of the operating table, if possible. Connect the monitor and ESU to different AC outlets located as far as possible from each other Using a Defibrillator Warning To avoid the risk of serious electrical burn, shock, or other injury during defibrillator, all users must keep clear of the bed and must not touch the patient or any equipment connected to the patient. If you draw out sensor or extension cable indifferently, these are broken or had short circuit. 87

91 5.2 Overview Monitoring screen When you first begin monitoring, a monitoring screen appears. The monitoring screen displays value for Pulse rate, SpO2, Blood pressure, and Temperature parameter. Any time you press the button on the front panel, the monitoring screen appears. The parameters on the monitoring screen depend on the measured parameter. When the monitor is turned on, it enters Setup mode while the monitor is waiting for the probe to be attached to the patient. The SpO2, NIBP, Temperature monitoring starts when the sensors are connected to the connector on the monitor and sensors are attached to the patient Trend mode (Review screen) The trend displays the stored data. You can display a stored data of the past hours on the trend window, the list of vital sign data on the screen. 88

92 Section 6. Alarm Function 6.1 What is an alarm Silencing alarm Suspending all alarms before occurrence Alarm Changeable Range Alarm Type Alarm Settings Alarm volume Setup Alarm Interval Time Setup 96 89

93 6.1 What is an alarm Angelus provides some alarm functions for the users. Users are recommended to carefully read this manual to understand and respond appropriately when the alarm goes off. Alarm informs error or information. When the monitor detects an abnormal patient condition, it can change an alarm sound, alarm lamp indication (malfunction lamp or power lamp). You can set each individual alarm condition. There are several types of alarm. The different alarm types are fully explained in the Alarm Types section. 6.2 Silencing alarm Several alarms occur together and if the [ESC/Silence] button is pressed, all alarms are silenced. When a SpO2 alarm, NIBP alarm, Pulse rate alarm, and Temperature alarm are silenced, the alarm resumes after the alarm silenced ends. When an alarm other than a vital signs alarm is silenced, the alarm indication dose not resume after the alarm silence ends. 90

DINAMAP ProCare Monitor. Operation Manual

DINAMAP ProCare Monitor. Operation Manual DINAMAP ProCare Monitor Operation Manual Contents Introduction About the DINAMAP ProCare Monitor...7 Indications...7 Contraindications...7 Warnings... 8 Cautions...9 Product Compliance...10 Symbols...11

More information

M9500 Patient Monitor User s Manual

M9500 Patient Monitor User s Manual M9500 Patient Monitor User s Manual Guangdong Biolight Meditech Co., Ltd. Address: Innovation First Road, Technology Innovation Coast, Jinding, Zhuhai, P.R.CHINA Tel: +86-756-3399900 Fax: +86-756-3399989

More information

OPERATOR S MANUAL. For YM1000 Vital Signs Monitor. EU Representative TECNOMED 2000 S.L. Valencia, Madrid Spain

OPERATOR S MANUAL. For YM1000 Vital Signs Monitor. EU Representative TECNOMED 2000 S.L. Valencia, Madrid Spain OPERATOR S MANUAL For YM1000 Vital Signs Monitor EU Representative TECNOMED 2000 S.L. Valencia, 25 28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650 1 Donghwa ri, Munmak

More information

SpotOn Temperature Monitoring System. SpotOn. Temperature Monitoring System. Model 370 Installation and Service Manual

SpotOn Temperature Monitoring System. SpotOn. Temperature Monitoring System. Model 370 Installation and Service Manual 3 SpotOn Temperature Monitoring System Model 370 Installation and Service Manual Please forward to the Biomedical Engineering Department 3M SpotOn Temperature Monitoring System Revision History Revision

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2009 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate,

More information

WARM LINE. IV Fluid Warmer. Product Code: J1448. JORGENSEN LABORATORIES PHONE FAX

WARM LINE. IV Fluid Warmer. Product Code: J1448. JORGENSEN LABORATORIES PHONE FAX WARM LINE IV Fluid Warmer Product Code: J1448 JORGENSEN LABORATORIES PHONE 800-525-5614 FAX 970-663-5042 info@jorvet.com Warm Line IV Fluid Warmer Portable and light weight Only warms IV fluids not transfusion

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series Copyright 2005 Welch Allyn. All rights are reserved. No one is permitted to reproduce

More information

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain OPERATOR S MANUAL Patient Monitor M20 EU representative TECNOMED 2000 S.L. Valencia, 25-28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1 Donghwa-ri, Munmak-eup, Wonju-si,

More information

Manufacturer s Responsibility Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of t

Manufacturer s Responsibility Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of t Preface Thank you for using M9500 patient monitor. In order to enable you to skillfully operate Monitor as soon as possible, we provide this user s manual with delivery. When you install and use this instrument

More information

REFRACTION TABLE HRT-7000

REFRACTION TABLE HRT-7000 1 User Manual REFRACTION TABLE HRT-7000 HUVITZ REFRACTION TABLE HRT-7000 2 Notice This product might cause a malfunction due to the electromagnetic waves of the cell phones, walkie-talkie, and wireless

More information

Atlas Monitor Outline Train the Trainer

Atlas Monitor Outline Train the Trainer Atlas Monitor Outline Train the Trainer A. ATLAS Monitor 1. Atlas Monitor is not designed for use on children younger than 3 years old. 2. Weight: 9.5lbs to 13.2 lbs, depending on configuration 3. Variable

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2014 Welch Allyn. All rights are reserved. To support the intended use of the product

More information

QAM.L L550 series. Patient Monitor Manual QAM.L

QAM.L L550 series. Patient Monitor Manual QAM.L QAM.L550.1112.2 L550 series Patient Monitor Manual QAM.L550.0915.3 Preface This manual has been provided to give detailed descriptions on DARAY s L550 patient monitor, covering the performance, operation

More information

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor OPERATOR S MANUAL NPB-4000/4000C Patient Monitor Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Nellcor Puritan Bennett, Durasensor, C-LOCK, Oxisensor II, Dura-Y, and the Nellcor

More information

End-Tidal CO2 User Guide CAPNO3 T NIBP. SpO2. Ver Charmcare Co., Ltd.

End-Tidal CO2 User Guide CAPNO3 T NIBP. SpO2. Ver Charmcare Co., Ltd. End-Tidal CO2 User Guide CAPNO3 T NIBP EtCO2 ECG SpO2 Ver 2.00 Charmcare Co., Ltd. -1- Table of Contents Indications for use statement 4 Warranty 4 Battery Replacement 4 Contact Us 5 Definition of Terms

More information

Tesla M3. MRI Patient Monitoring System. Instructions for Use. Version 4.0

Tesla M3. MRI Patient Monitoring System. Instructions for Use. Version 4.0 Instructions for Use Version 4.0 Tesla M3 MRI Patient Monitoring System 0120 1 MIPM Mammendorfer Institut für Physik und Medizin GmbH, herein after called MIPM. Printed in Germany Subject to change without

More information

INSTRUCTION MANUAL FOR

INSTRUCTION MANUAL FOR INSTRUCTION MANUAL FOR Cool Block The 1st Edition Yamato Scientific Co.,, Ltd. 2-1-6, Nihonbashi Honcho, Chuo-ku Tokyo 103-8432 JAPAN Congratulations on your selection of Yamato Scientific s Cool Block

More information

Warmette Blanket Cabinet Warmette Dual Cabinet. User Manual

Warmette Blanket Cabinet Warmette Dual Cabinet. User Manual Warmette Blanket Cabinet Warmette Dual Cabinet User Manual Natus and Warmette are registered trademarks of Natus Medical Incorporated. 2013 Natus Medical Incorporated. All rights reserved. This manual

More information

Model 900. Pulse Oximeter User s Manual

Model 900. Pulse Oximeter User s Manual Model 900 Pulse Oximeter User s Manual MEDIAID INC. 17517 Fabrica Way Suite H, Cerritos, CA 90703 USA Phone: 714-367-2848; Fax: 714-367-2852 Website: www.mediaidinc.com POX010-900 Contents Table of Contents

More information

CITY MULTI Control System and Mitsubishi Air Conditioners Wireless Remote Controller PAR-SL100A-E

CITY MULTI Control System and Mitsubishi Air Conditioners Wireless Remote Controller PAR-SL100A-E CITY MULTI Control System and Mitsubishi Air Conditioners Wireless Remote Controller PAR-SL100A-E OPERATION MANUAL FOR USER For safe and correct use, please read this operation manual thoroughly before

More information

Digital Refrigerant System Analyzer. MAXMIN Psi kpa Bar MPa C F R TIME P T UNITS

Digital Refrigerant System Analyzer. MAXMIN Psi kpa Bar MPa C F R TIME P T UNITS Digital Refrigerant System Analyzer INSTRUCTION MANUAL ENGLISH LOW START PRS T-LOW SUPERHEAT MAXMIN kpa Bar MPa R TIME P T HIGH END PRS T-HIGH SUBCOOL VAC 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com

More information

Veterinary Patient Warming System Controller User Manual

Veterinary Patient Warming System Controller User Manual Veterinary Patient Warming System Controller User Manual Contents Introduction... 3 Indications for use... 3 Contraindications... 3 Warnings... 3 Precautions... 4 Proper Use and Maintenance... 5 Initial

More information

Tesla M3. MRI Patient Monitoring System. Instructions for Use. Version 3.1

Tesla M3. MRI Patient Monitoring System. Instructions for Use. Version 3.1 Instructions for Use Version 3.1 Tesla M3 MRI Patient Monitoring System 0120 1 MIPM Mammendorfer Institut für Physik und Medizin GmbH, herein after called MIPM. Printed in Germany Subject to change without

More information

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models:

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models: Cardell Touch Veterinary Vital Signs Monitor For Models: 8013-001 8013-002 8013-003 8013-004 User s Guide TP200 Rev. A 003-2840-00 Rev. AA2 (6/20/18) Software Version V4.0.X Product Information Dealer:

More information

CALIBRATABLE 7-FUNCTION THERMO- HYGROMETER USER S MANUAL

CALIBRATABLE 7-FUNCTION THERMO- HYGROMETER USER S MANUAL CALIBRATABLE 7-FUNCTION THERMO- HYGROMETER USER S MANUAL EP8710 Please read this manual carefully and thoroughly before using this product. TABLE OF CONTENTS Introduction.................... 2 4 Key Features.......................

More information

DK-11 Countertop Water System. Owner's Manual

DK-11 Countertop Water System. Owner's Manual DK-11 Countertop Water System Owner's Manual Atmospheric Water Systems, Inc. 189 Cross Street, Suite 110 San Luis Obiospo, CA 93401 phone: 805-858-9280 fax: 805-549-9790 email: cs@aws-h2o.com www.aws-h2o.com

More information

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor GE Healthcare Quick Guide Dash 2500 Patient Monitor Hardkeys Main Menu Hide menu and view entire waveform screen Display the large numeric vital signs Review or change alarm limits for all operating parameters

More information

Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input

Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with Thermocouple Input User Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Digi-Sense TC9000 Advanced PID and On/Off Temperature Controller with

More information

1 Appendix - Product Specifications

1 Appendix - Product Specifications 1 Appendix - Product Specifications WARNING The patient monitor may not meet its performance specification if stored or used outside the manufacturer s specified temperature and humidity range. 1.1 Safety

More information

5580 S. Nogales Highway Tucson, AZ Phone: (800) Fax: (520) ANAPOD Instructions For Use

5580 S. Nogales Highway Tucson, AZ Phone: (800) Fax: (520) ANAPOD Instructions For Use 5580 S. Nogales Highway Tucson, AZ 85706 Phone: (800) 975-7987 Fax: (520) 294-6061 www.westmedinc.com TM ANAPOD Instructions For Use PN 75054, Rev 02 ii TABLE OF CONTENTS Servicing: Beyond the replacement

More information

Design Manual Installation Operation Maintenance

Design Manual Installation Operation Maintenance Design Manual Installation Operation Maintenance Model FT194 UV/IR Portable Flame Detector Test Lamp 23282 Mill Creek Drive, Suite 215 Laguna Hills, CA 92653 USA +1.949.583.1857 Phone +1.949.340.3343 Fax

More information

Digital Marine Exhaust Temperature Alarm

Digital Marine Exhaust Temperature Alarm Digital Marine Exhaust Temperature Alarm Model: SM007D/S INTRODUCTION COMPONENTS Marine water cooled exhaust systems are designed to withstand temperatures of up to about 120 C. However the exhaust gases

More information

Propaq LT Outline Train the Trainer

Propaq LT Outline Train the Trainer Train the Trainer A. Propaq LT and LTR Propaq LT Standalone monitor Propaq LTR Monitor that has the wireless Acuity feature enabled 1. Lightweight (Less than 2 pounds) 2. Uses rechargeable lithium ion

More information

USER GUIDE. Software Version 1.11

USER GUIDE. Software Version 1.11 USER GUIDE USER GUIDE Software Version 1.11 Foreword Precaution: The Aquamantys Pump Generator is for use only by qualified medical personnel properly trained in the use of electrosurgical equipment,

More information

About this Manual. Statement. P/N: Release Date: August 2010 Copyright EDAN INSTRUMENTS, INC All rights reserved.

About this Manual. Statement. P/N: Release Date: August 2010 Copyright EDAN INSTRUMENTS, INC All rights reserved. About this Manual P/N: 01.54.110221-13 Release Date: August 2010 Copyright EDAN INSTRUMENTS, INC. 2008-2010. All rights reserved. Statement This manual will help you understand the operation and maintenance

More information

ARGUS VCM Vital Compact Monitor

ARGUS VCM Vital Compact Monitor Vital Compact Monitor Art. No.: 2.540041 Rev. a *2.540041* Service Handbook Contents Figures Tables Notice This document contains proprietary information that is protected by copyright. All Rights Reserved.

More information

USER MANUAL. Version: 1.0. ROBE Show Lighting s.r.o. Czech Republic

USER MANUAL. Version: 1.0. ROBE Show Lighting s.r.o. Czech Republic USER MANUAL Version: 1.0 ROBE Show Lighting s.r.o. Czech Republic www.robe.cz Table of contents 1.Introduction 4 2.Cautions 4 3. Unpacking & Inspection 5 4. Setup 5 5. Operation 5 5.1.Control Module Operation

More information

Operating Instructions Air Conditioner

Operating Instructions Air Conditioner F566113 Operating Instructions Air Conditioner Indoor Unit CS-E18HKR CS-E1HKR CS-E4HKR CS-E8HKR Outdoor Unit CU-E18HKR CU-E1HKR CU-E4HKR CU-E8HKR Before operating the unit, read these operating instructions

More information

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X Vital Signs Monitor 300 Series Directions for Use Software Version 1.0X Copyright 2003, Welch Allyn Protocol, Inc. Welch Allyn is a registered trademark of Welch Allyn, Inc. Protocol is a trademark of

More information

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium.

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium. Patient Monitor HBP-2070 Operator s Manual EU representative OBELIS S.A Bd. Général Wahis, 53 1030 Brussels, Belgium Local distributor Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1

More information

13500 BTU RV Air Conditioner Installation and Operating Instructions

13500 BTU RV Air Conditioner Installation and Operating Instructions 13500 BTU RV Air Conditioner Installation and Operating Instructions Model No.:HCB-R135-A-ID HCB-R135-A-OD 3092402 V140804 Thank you for choosing a Soleus Air RV Air Conditioner. This owner s manual will

More information

IMAGE PRESS & IMAGE PORCELAIN OVEN. 120V, 50/60 Hz Models and 230V, 50/60 Hz Models. Operator s Manual

IMAGE PRESS & IMAGE PORCELAIN OVEN. 120V, 50/60 Hz Models and 230V, 50/60 Hz Models. Operator s Manual IMAGE PRESS & IMAGE PORCELAIN OVEN 120V, 50/60 Hz Models and 230V, 50/60 Hz Models Operator s Manual TABLE OF CONTENTS Congratulations.. 3 Warranty. 3 Safety Notice. 4 Electrical Safety Notes.. 5 Knowledge

More information

Electric Fireplace Insert with Heater

Electric Fireplace Insert with Heater Electric Fireplace Insert with Heater OWNER S MANUAL Model No. FP406R-Q PLEASE SAVE THESE INSTRUCTIONS WARNING: READ THIS OWNER S MANUAL CAREFULLY BEFORE USE. PET OWNERS WARNING: The health of some small

More information

Operating Instructions Air Conditioner

Operating Instructions Air Conditioner F56579 Operating Instructions Air Conditioner Indoor Unit Outdoor Unit CS-E18GKR CS-E1GKR CS-E4GKR CS-E8GKR CU-E18GKR CU-E1GKR CU-E4GKR CU-E8GKR Before operating the unit, read these operating instructions

More information

Time Now Corporation. Invisible Clock - General Information. Alarm Switch. Timing Functions - Your Invisible Clock has four timing modes

Time Now Corporation. Invisible Clock - General Information. Alarm Switch. Timing Functions - Your Invisible Clock has four timing modes Time Now Corporation Invisible Clock - General Information U.S. Patent Pending This section provides important general information about your Invisible Clock. The next section Invisible Clock Operation

More information

User Instruction Manual

User Instruction Manual Read Entire Manual before Operating Device User Instruction Manual PLEASE NOTE: It is important that the user read the entire manual before operating this device. Table of Contents Page Symbol Definitions...

More information

TidalGuard HD. Handheld CO 2 / SpO 2 and CO 2 monitors. Operating Manual Rev. H (8/15)

TidalGuard HD. Handheld CO 2 / SpO 2 and CO 2 monitors. Operating Manual Rev. H (8/15) TidalGuard HD Handheld CO 2 / SpO 2 and CO 2 monitors Operating Manual 003-2275-00 Rev. H (8/15) Table of Contents Table of Contents Service Information... iv Preface... vi 1. Safety... 1-1 1.1 Safety

More information

MD-MINI Dry Bath. Instruction Manual

MD-MINI Dry Bath. Instruction Manual MD-MINI Dry Bath Instruction Manual Catalog No. MD-MINI www.majorsci.com service@majorsci.com Version 01B Revision on: 2013.07.15 Packing list MD-MINI: 1x MD-MINI Dry Bath 1x MD-MINI Lid 1x Power Cord

More information

Owner's Manual TABLE OF CONTENTS

Owner's Manual TABLE OF CONTENTS 40MAQ High Wall Ductless System Sizes 09 to 36 Owner's Manual TABLE OF CONTENTS PAGE A NOTE ABOUT SAFETY... 2 GENERAL... 2 PART NAMES... 3 FUNCTION BUTTONS... 4 DISPLAY PANELS... 5 REMOTE CONTROL... 6

More information

OPERATION MANUAL. StableTemp Economy Mechanical Convection Oven. Model : ,-90,-91,-92,-93,-94

OPERATION MANUAL. StableTemp Economy Mechanical Convection Oven. Model : ,-90,-91,-92,-93,-94 OPERATION MANUAL StableTemp Economy Mechanical Convection Oven Model : 52412-89,-90,-91,-92,-93,-94 This operation manual describes the important subjects on maintaining the product s functions, safety,

More information

Portable Air Conditioner 6,000 BTU 8,000 BTU 10,000 BTU

Portable Air Conditioner 6,000 BTU 8,000 BTU 10,000 BTU Portable Air Conditioner 6,000 BTU 8,000 BTU 10,000 BTU OPERATING INSTRUCTIONS PCR-06-01 PCR-08-01 PCR-10-01 3058080 V170223 PURCHASE INFORMATION Thank you for choosing a Chigo Portable Air Conditioner.

More information

INTEGRATED VOICE EVACUATION SYSTEM VX-3000 SERIES

INTEGRATED VOICE EVACUATION SYSTEM VX-3000 SERIES PLEASE READ THIS MANUAL FIRST! INTEGRATED VOICE EVACUATION SYSTEM VX-3000 SERIES Thank you for purchasing TOA s Integrated Voice Evacuation System. Please carefully follow the instructions in this manual

More information

OWNER S MANUAL. R 410A Ductless Split System Air Conditioner and Heat Pump

OWNER S MANUAL. R 410A Ductless Split System Air Conditioner and Heat Pump R 410A Ductless Split System Air Conditioner and Heat Pump Models DLC4(A/H) Outdoor Unit, DLF4(A/H) Indoor Unit Sizes 9K, 12K, 18K, 24K, 30K and 36K Please read the operating instructions and safety precautions

More information

CLEO VERSATILITY IN VITAL SIGNS

CLEO VERSATILITY IN VITAL SIGNS CLEO VERSATILITY IN VITAL SIGNS CLEO PORTABLE weighs less than 3 lbs TOUCHSCREEN easy and intuitive to use MOBILE backup battery powered The Cleo is a new and intuitive approach to patient vital signs

More information

BIOVAC 500 SMOKE EVACUATOR

BIOVAC 500 SMOKE EVACUATOR BIOVAC 500 SMOKE EVACUATOR INSTRUCTION MANUAL REF 909105 and 909105-05 Wallach Surgical Devices 95 Corporate Drive Trumbull, CT 06611 U.S.A. Phone: (203) 799-2000 Fax: (203) 799-2002 IMLP011 Rev. B 2/09

More information

Operating Instructions

Operating Instructions Operating Instructions Air Conditioner Model No. Indoor Unit CS-PC9QKH CS-PC12QKH CS-PC18QKH CS-PC24QKH CS-PC28QKH Outdoor Unit CU-PC9QKH CU-PC12QKH CU-PC18QKH CU-PC24QKH CU-PC28QKH Operating Instructions

More information

GENANO AIR PURIFICATION TECHNOLOGY

GENANO AIR PURIFICATION TECHNOLOGY GENANO AIR PURIFICATION TECHNOLOGY GENANO 310 AIR PURIFICATION DEVICE OPERATING INSTRUCTION This manual is for the Genano 310 log. Version. GENANO LTD Kimmeltie 3 FI- 02110 Espoo FINLAND Tel + 358 (0)

More information

5) Do not start or stop the unit by inserting or pulling out the power plug.

5) Do not start or stop the unit by inserting or pulling out the power plug. 3058080 V170306 PURCHASE INFORMATION Thank you for choosing a Soleus Air Portable Air Conditioner. This Owner s Manual will provide you with valuable information necessary for the proper care and maintenance

More information

VS2000V Veterinary Vital Signs Monitor Operation Manual

VS2000V Veterinary Vital Signs Monitor Operation Manual VS2000V Veterinary Vital Signs Monitor Operation Manual - English Version 1.0, July 2011 2011 UTECH Co., Ltd. All rights reserved. Table of Contents Table of Contents Warranty and Service Information...

More information

Computer Room Guard Model VM Manual and Installation Instructions

Computer Room Guard Model VM Manual and Installation Instructions Computer Room Guard Model VM500-8 Manual and Installation Instructions For units purchased since December 2004 Index Page General Description 3 Installation, Wiring Diagram 3-5 Accessing the Computer Room

More information

CR13000-PAC OWNER S MANUAL. Portable Air Conditioner. PORTABLE AIR CONDITIONER with HEAT PUMP. Operating and Servicing Instructions

CR13000-PAC OWNER S MANUAL. Portable Air Conditioner. PORTABLE AIR CONDITIONER with HEAT PUMP. Operating and Servicing Instructions OWNER S MANUAL Operating and Servicing Instructions CR13000-PAC PORTABLE AIR CONDITIONER with HEAT PUMP Portable Air Conditioner Questions or concerns? For assistance, please call Customer Service ClimateRight

More information

IMPORTANT SAFETY INSTRUCTIONS

IMPORTANT SAFETY INSTRUCTIONS WARNING TO PREVENT FIRE OR SHOCK HAZARD, DO NOT USE THIS PLUG WITH AN EXTENSION CORD, RECEPTACLE OR OTHER OUTLET UNLESS THE BLADES CAN BE FULLY INSERTED TO PREVENT BLADE EXPOSURE. TO PREVENT FIRE OR SHOCK

More information

Inverter Split-type Room Air Conditioner

Inverter Split-type Room Air Conditioner OWNER S MANUAL Inverter Split-type Room Air Conditioner Please read the operating instructions and safety precautions carefully and thoroughly before installing and operating your room air conditioner.

More information

Table of Contents. English

Table of Contents. English OM-E0799E 000 English Thank you for purchasing VIVA ace Motor Kit. Please read this Operation Manual and the VIVA ace Basic Set Operation Manual carefully before use for operating instructions and care

More information

FastGene Plate Centrifuge. Instruction Manual. Catalog No. NG

FastGene Plate Centrifuge. Instruction Manual. Catalog No. NG FastGene Plate Centrifuge Instruction Manual Catalog No. NG040 www.nippongenetics.eu info@nippongenetics.eu Version 01 20.01.2015 Content -1x Plate Centrifuge -2x Adapter for 96 well plates -1x Power Cord

More information

Product Specifications

Product Specifications 1.Safety Specifications 1.1Product Classification Class Type of protection against electric shock Specification Class I, with internal power supply Degree of protection against electric shock Anti-defibrillation:

More information

Installation and Operating Manual

Installation and Operating Manual Installation and Operating Manual SR868C6 System Regulator for Solar Thermal Systems Display Panel Illustration Pos. Button on display panel Button description 1 Green lamp Power indication lamp 2 On/Off

More information

Owner s Manual. Model 8800 Universal Communicating Thermostat. Includes Operating Instructions and Warranty Information

Owner s Manual. Model 8800 Universal Communicating Thermostat. Includes Operating Instructions and Warranty Information Model 8800 Universal Communicating Thermostat Owner s Manual Includes Operating Instructions and Warranty Information READ AND SAVE THESE INSTRUCTIONS 61000762A 8800 Tstat Owners.indd 1 3/28/11 4:19:57

More information

Portable Air Conditioner with Heat Pump Technology Operating Instructions. Model No.: HCB-P13HP-D. Reference No.: BPD13HP V140217

Portable Air Conditioner with Heat Pump Technology Operating Instructions. Model No.: HCB-P13HP-D. Reference No.: BPD13HP V140217 Portable Air Conditioner with Heat Pump Technology Operating Instructions 3092402 Model No.: HCB-P13HP-D Reference No.: BPD13HP V140217 Thank you for choosing a Soleus Air Portable Air Conditioner with

More information

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version This document is a guideline only to be used as an aid to comprehensive Inservice training. NAME: HOSPITAL: DATE: VALIDATOR:

More information

Multiple use Digital Thermometer With 7 Day Memory

Multiple use Digital Thermometer With 7 Day Memory Owner s Manual Computemp 5 Multiple use Digital Thermometer With 7 Day Memory Congratulations. You are now the owner of one of the most remarkable and fascinating ways available to observe temperature

More information

Tissue-Tek TEC Plus Cryo Console OPERATING MANUAL

Tissue-Tek TEC Plus Cryo Console OPERATING MANUAL Tissue-Tek TEC Plus Cryo Console OPERATING MANUAL Sakura Finetek Japan Co., Ltd., Tokyo 103-0023, Japan Sakura Finetek U.S.A., Inc., Torrance, CA 90501, USA Sakura Finetek Europe B.V., Zoeterwoude, Netherlands

More information

OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36

OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36 OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36 TABLE OF CONTENTS PAGE SAFETY PRECAUTIONS... 2 GENERAL... 2 INDOOR UNIT PART NAMES... 3 REMOTE CONTROL PART NAMES...

More information

Hades 2. Digital Recycle and Lighting Timer with High Temp Shut-Off Instruction Manual. Notes: 1 Square = Foot/Feet

Hades 2. Digital Recycle and Lighting Timer with High Temp Shut-Off Instruction Manual. Notes: 1 Square = Foot/Feet Notes: 1 Square = Foot/Feet Hades 2 Digital Recycle and Lighting Timer with High Temp Shut-Off Instruction Manual VANCOUVER, WASHINGTON U.S.A. VANCOUVER, WASHINGTON U.S.A. www.titancontrols.net Revision

More information

RGR150 USER S MANUAL. Wireless Rain Gauge with Thermometer and Clock

RGR150 USER S MANUAL. Wireless Rain Gauge with Thermometer and Clock RGR150 manual-final-091908:layout 1 9/19/08 8:59 AM Page 1 RGR150 USER S MANUAL Wireless Rain Gauge with Thermometer and Clock INTRODUCTION Thank you for selecting this Wireless Rain Gauge. This device

More information

OWNER'S MANUAL R-410A Duct Free Split System Air Conditioner and Heat Pump

OWNER'S MANUAL R-410A Duct Free Split System Air Conditioner and Heat Pump R-10A Duct Free Split System Air Conditioner and Heat Pump Product Family: DFF(A/H)H, DFC(A/H) Please read the operating instructions and safety precautions carefully and thoroughly before installing and

More information

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry Cardell Insight Veterinary Monitor For Models: 8014 Blood Pressure 8015 Blood Pressure, Pulse Oximetry User Guide 003-2981-00 Rev. B (2/21/17) Software Version 1.2V Product Information Dealer: Date of

More information

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English Operator s Manual Avant 2120 Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor 0123 English CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

More information

OMNI Express VS VERSATILITY IN VITAL SIGNS

OMNI Express VS VERSATILITY IN VITAL SIGNS OMNI Express VS VERSATILITY IN VITAL SIGNS OMNI Express VS TOUCHSCREEN easy and intuitive to use PORTABLE weighs less than 6lbs MOBILE backup battery powered The Omni Express VS is a new intuitive approach

More information

The door can be opened with a RF card. The door can be opened with a transit RF card. Up to 20 cards be registered.

The door can be opened with a RF card. The door can be opened with a transit RF card. Up to 20 cards be registered. Product Features The door can be opened with a RF card. The door can be opened with a transit RF card. Up to 20 cards be registered. The door can also be opened with a password. Once the user s password

More information

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual

Portable Oxygen Monitor OX-07. (Type B) <No Gas alarm type> Operating Manual PT0E-0922 Portable Oxygen Monitor OX-07 (Type B) Operating Manual Request for the Customers Read and understand this operating manual before using this gas monitor. Use it in accordance

More information

Multiparameter. er Monitor DM

Multiparameter. er Monitor DM d e n t a l a n d m e d i c a l e q u i p m e n t User manual Multiparameter er Monitor DM ENGLISH Contents 1 Introduction...1 1.1 General Information...1 1.2 Screen Display...2 1.3 Button Functions...5

More information

Digital Room Thermostat

Digital Room Thermostat Digital Room Thermostat Instruction Manual Model No RT500 PRODUCT COMPLIANCE This product complies with the essential requirements of the following EC Directives: Electro-Magnetic Compatibility directive

More information

HANDHELD PULSE OXIMETER

HANDHELD PULSE OXIMETER Attendant HANDHELD PULSE OXIMETER Owner s Manual Please keep and refer to this Owner s Manual. Thank you for purchasing an Attendant Handheld Pulse Oximeter from Direct Supply Equipment & Furnishings.

More information

MODEL 3X Series User s Manual

MODEL 3X Series User s Manual MODEL 3X Series User s Manual Pulse Oximeter with options for 24 hour Memory, Printer & USB / Infrared Ports MEDIAID INC. 17517 Fabrica Way Suite H Cerritos, CA 90703 USA (Tel) 714-367-2848 (Fax) 714-367-2852

More information

MaxiCompressor. Limited Warranty. High Performance 50 PSI Compressor 501-S

MaxiCompressor. Limited Warranty. High Performance 50 PSI Compressor 501-S Limited Warranty Global Medical Holdings (GMH) warrants the MaxiCompressor for 1 year from the date of purchase due to faulty parts or workmanship. This warranty is limited to the original purchaser of

More information

14,000 BTU Portable Air Conditioner 14,200 BTU Heat Pump Model Number: LX-140 Operating Instructions. Model No. LX Soleus Air International

14,000 BTU Portable Air Conditioner 14,200 BTU Heat Pump Model Number: LX-140 Operating Instructions. Model No. LX Soleus Air International 14,000 BTU Portable Air Conditioner 14,200 BTU Heat Pump Model Number: LX-140 Operating Instructions 3092402 Model No. LX-140 2012 Soleus Air International Thank you for choosing a Soleus Air LX-140 Portable

More information

Multi-parameter Patient Monitor Operator s Manual

Multi-parameter Patient Monitor Operator s Manual Multi-parameter Patient Monitor Operator s Manual Contents Disclaimer Statement...4 Chapter One General Introduction to Product...6 1.1 General introduction to patient monitor...7 1.2 Display Interface...12

More information

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English Operator s Manual Avant 2120 Avant 2120 Pulse Oximeter and NIBP Monitor 0123 English 1 CAUTION! CAUTION! Federal law (USA) restricts this device to sale by or on the order of a physician. Read this entire

More information

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use Welch Allyn Vital Signs Monitor 6000 Series Directions for use 2010 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the

More information

CASMED 740 Vital Signs Monitor

CASMED 740 Vital Signs Monitor CASMED 740 Vital Signs Monitor User s Manual 2 THE CASMED 740 VITAL SIGNS MONITOR FEATURES CASMED 740-1 Non-Invasive Blood Pressure and Pulse Rate. CASMED 740-2 Non-Invasive Blood Pressure, Pulse Rate

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

MyKi Care Thermometer (MCT 1.0) USER MANUAL

MyKi Care Thermometer (MCT 1.0) USER MANUAL Precautions for use Any form of modification to this device is forbidden; Don t use the device together with MRI or CT equipment; Explosion danger: don t use the device in flammable anesthetic gas; Don

More information

Operating Instructions Air Conditioner

Operating Instructions Air Conditioner alleru-buster F566894 Operating Instructions Air Conditioner Indoor Unit CS-RE9JKR CS-RE1JKR Outdoor Unit CU-RE9JKR CU-RE1JKR ENGLISH ~ 7 Before operating the unit, read these operating instructions thoroughly

More information

V6 Vital Signs Monitor User s Manual

V6 Vital Signs Monitor User s Manual V6 Vital Signs Monitor User s Manual Product Information Product Model: V6 Product Name: Vital Signs Monitor Manufacturer: GuangdongBiolightMeditech Co.,Ltd. After Service Contact Information: Address:

More information

Kinsa Smart Ear ThermometerTM Instructions For Use

Kinsa Smart Ear ThermometerTM Instructions For Use ENGLISH Kinsa Smart Ear ThermometerTM Instructions For Use Thank you for purchasing the Kinsa Smart Ear Thermometer, a professionally accurate instrument for fast and easy temperature taking in the ear.

More information

EasyTronic III MANUAL SERVICE

EasyTronic III MANUAL SERVICE rev.6 EasyTronic III MANUAL SERVICE General characteristics: Power supply 24 Vac ±15% Max consumption at 24Vac 300mA Relay outputs 6 Maximum relay current 8 A res. Serial standard RS232 2 Serial standard

More information

Infusion Pump Caddo 16B Learning Material Ver. 1.1 An ISO 9001:2008 company 94, Electronic Complex, Pardesipura Indore - 452 010 India Tel: +91-731 4211100 Fax: +91-731-2555643 e mail: info@scientech.bz

More information

Dry Bath Heating Systems

Dry Bath Heating Systems Dry Bath Heating Systems Instruction Manual N2400-4001 Dry Bath Heating System, Single Block, Digital N2400-4002 Dry Bath Heating System, Dual Block, Digital www.starlabgroup.com DECLARATION OF CONFORMITY

More information

Portable Air Conditioner with Heat Pump Technology PH4-10R-01, PH4-12R-01, & PH4-14R-01 Operating Instructions

Portable Air Conditioner with Heat Pump Technology PH4-10R-01, PH4-12R-01, & PH4-14R-01 Operating Instructions Portable Air Conditioner with Heat Pump Technology PH4-10R-01, PH4-12R-01, & PH4-14R-01 Operating Instructions 3092402 2006 Soleus Air International Thank you for choosing a Soleus Air Portable Air Conditioner

More information