Troubleshooting Thermo Scientific LevelPRO Gauges

Size: px
Start display at page:

Download "Troubleshooting Thermo Scientific LevelPRO Gauges"

Transcription

1 Troubleshooting Thermo Scientific LevelPRO Gauges Key Words Scintillation detector Current board Relay output Communications Level indication Purpose A chapter in the user guide for the Thermo Scientific LevelPRO gauge is dedicated to troubleshooting. Providing the material from that chapter in this stand-alone document will help users keep troubleshooting procedures easily accessible. Note: This technical bulletin is to be used as a quick reference for troubleshooting the LevelPRO gauge. It is not meant to be used as a replacement for the installation and user guides. Note: Troubleshooting must be done only by qualified personnel. Note: Do not use the HART communication system for troubleshooting. You must use either the Thermo Scientific Model 9734 Handheld Terminal (HHT) or a computer with RS232/RS485 converter and the Thermo Scientific TMTComm communication software to access the technical troubleshooting capabilities of the LevelPRO system. The LevelPRO system is ESD sensitive. Use proper ESD protective equipment and procedures. Failure to comply with ESD procedures can result in circuit damage. The Scintillation Detector Note: Enable the Service Only Items in the Special Functions menu before proceeding with this section. If you suspect a problem with the scintillation detector, you can check the detector output by accessing the Internal Value of Sensor Signal item in the Snapshot item of the Diagnostics menu. The direct access code (DAC) for this item is If the tank is at least partially empty, this value is the count rate in counts/second (cps) and should be much larger with the source shutter on than with the shutter off (typically more than ten times larger). As a further check, empty the tank and compare the Internal Value of Sensor Signal item to the STD Value in Use item, which is displayed in the Sensor Standardization menu. The DAC for this item is If the gauge was working properly when it was last standardized, the two values should be similar. The match will not be exact because the STD value is filtered and corrected for background radiation and source decay. If the displayed signal value is not behaving correctly, check the gauge head for misalignment, buildup, or wear, and check for debris in the beam path and faulty shutter operation before testing the detector signals. Use a long handled brush to remove debris in the beam path to ensure that no part of your body, including your hands, enters the radiation beam path.

2 If you can eliminate these other factors, it is likely that the cause is related to the CPU board. Verify that the cable connections between the detector board and CPU board are securely seated. If the problem persists, contact Thermo Fisher Scientific Technical Support. 1. Locate the J11 connector on the CPU board. The blow-away jumper setting for normal operation is shown in Figure 1. Remove all power from the unit prior to opening the enclosure. Electrocution can result if power is present. The Current Board If you suspect a problem with the current output, attach an ammeter in series with the current output load and verify the current output at various levels. Use the commands from the User Service & Related Items submenu within the Security, Service, and Diagnostic Functions menu group to hold the current output at specific levels for testing. Figure 1. Blow-away jumper in normal operation position 2. With power removed, remove the blow-away jumper, rotate it 90 clockwise, and insert it onto the pins as shown in Figure 2. The Relay If you suspect a problem with a relay output, attach a continuity tester to the suspected relay output and use the Test Relays commands in the Factory Service & Related Items menu to test relay operation. Communication Problems If the gauge will not communicate at all, try hard booting the system. Note: This procedure will erase memory and return all parameters to the factory defaults. Remove power from the gauge prior to performing this procedure. Figure 2. Blow-away jumper in erase memory position 3. Apply power to the gauge for approximately 30 seconds. 4. Remove power from the gauge, and return the jumper to its original position. 5. Apply power to the gauge. All parameters will be set to factory defaults.

3 Level Indication If the gauge indication is 100% and will not change, review this section carefully and then follow the steps provided. Equipment Required For this procedure, you will need: One Model 9734 HHT or a PC with RS232/RS485 converter and TMTComm software installed One multimeter Site Preparation You will need to access the gauge while it is in the housing and in the radiation field coming from the radioactive source on the opposite side of the vessel. The source shutter will need to be ON and the level in the vessel should be as low as possible, without adversely affecting the process. If you cannot lower the level of the process material in the tank to a point well below the top of the detector sensitive length you may not be able to perform all the required tests. If you are in a classified area, a HOT WORK permit may be necessary. Procedure the transition from the empty radiation levels to the lower radiation levels where the beam is being blocked by the remaining material in the vessel. If you have access to the original survey document completed when the gauge was installed, you will also be able to verify whether there is any significant reduction in radiation field readings that cannot be accounted for by normal source decay. This may indicate a problem with a buildup of process material on the vessel walls. If a radiation survey meter is not available or it is not feasible to visually confirm the material level in the vessel, the indication of the vessel level must be accepted. 3. Connect to the LevelPRO detector. If you have to remove the detector housing lid to connect the communication device to the gauge, verify that the red power LED on the CPU board is lit (see Figure 3). The LED indicates that the CPU is getting the 5 Vdc required for operation. If the gauge is powered by 120 Vac, this voltage will be present while accessing the gauge. Proceed with caution. Do not come in contact with this voltage. 1. Verify that the radioactive source shutter is in the ON position. 2. Verify that the vessel is not actually full of material. Figure 3. Power LED on the CPU board If the system is AC powered and the LED is not lit, check the fuse on the power supply board (Figure 4), which is a amp sloblow. If the system is DC powered and the LED is not lit, check the F1 fuse located on the CPU board (Figure 5). If a radiation survey meter is available, use it to find the actual level. Do this by placing the survey meter detector between the vessel and detector. Starting at the top of the gauge measurement area, move the probe slowly down the length of the detector until you see

4 If the sensor signal does not change or is zero when you cycle the source shutter, the detector signal is getting lost or corrupted somewhere between the scintillation crystal and the CPU. The problem may be related to the photomultilplier tube (PMT), the triangle board (p/n ), the cabling (p/n ), or the CPU (p/n ). Figure amp slo-blow fuse on AC powered systems 5. The next step is to turn on the diagnostic measurements feature. Use DAC Press the right arrow once to change the selection from Do not enable (Figure 7) to Do enable (Figure 8) to enable diagnostic measurements and then exit. Do not enable diag measurements Change to Do Continue as is. Figure 7. Diagnostics measurements disabled Figure 5. F1 fuse on CPU board for DC powered systems 4. Using DAC , bring up the Internal Value of Sensor Signal item. Depending on application, source size, vessel wall, etc., the count rate should be in the thousands. The value shown in Figure 6 is for example only. Internal value of sensor signal 2.057E4 Figure 6. Example internal value of sensor signal Do enable diag measurements Change to Do not Continue as is. Figure 8. Diagnostic measurements enabled If the measurements do not show up on the screen, enter command 72, and they will appear. With the diagnostic measurements turned on, the display will toggle between two screens (Figures 9 and 10). Along with the level reading, four new values are displayed. These values are explained in the next section. While the detector value is still displayed, close the shutter on the radioactive source. The displayed count rate should go down towards zero. Open the source shutter and the count rate should go back up to where you started. If this occurs, the detector is working properly. The problem may be programming related % 0000 ERRSIG 1081 REF (cps) For setup, press Figure 9. Diagnostics measurement screen % 9175 VD 2215 CTLSIG For setup, press Figure 10. Diagnostic measurement screen 2

5 Diagnostic Measurement Values In addition to the level reading, four values are displayed on the diagnostic measurement screens. ERRSIG: This value may also be shown as HVCorr, depending on the version of software in the instrument. ERRSIG is a relative value representing the length of the control pulses for high voltage control. This number should be at or near 0. If it is staying under 10, the detector is under normal control and working correctly. If this number is running from 10 to 50, the detector may be damaged (often from being exposed to temperatures over 140 F for some period of time), and its indication should be suspect. If this number is running from 50 to 256 (the highest possible value), the gauge cannot control the high voltage. It is likely that the detector has failed. CTLSIG: This value may also be shown as HVCtrl, depending on the version of software in the instrument. This is a relative value representing the amount of control time the CPU is using to control the detector. The value can be from 10 to The normal control range for a new detector is from 2000 to Every detector will have a different value, but the value should always be steady. If the value is ramping up or down, then the detector is faulty and not under control. REF(cps): This value is the REF count rate in cps. It should be 1/8 of the VD count rate. The CPU will adjust the CTLSIG until the VD gets to 8 times the REF count rate. If it is unable to reach this 8:1 ratio, the CTLSIG will ramp up to 8000, go back to 10, and ramp up again. If the REF(cps) is a low number like 5, the reference counts are getting lost somewhere between the PMT and the CPU. Recorded Values The values listed in Table 1 are useful to Thermo Fisher Scientific Technical Support when troubleshooting the LevelPRO system. Please have these values readily available when calling for assistance. To recall these values, press EXIT, enter the DAC, and press the down arrow. Record the value shown. Press EXIT again to return to the main screen. The first four values {ERRSIG through REF(cps)} were accessed in step 5 of the above procedure. VD: This value is the DATA count rate in cps. It should be the same value as the internal value of the sensor signal (DAC ) recalled in step 4 of this procedure. If this value is above 9999, the display will show >9999. To display this value, you will need to enable the R&D test items by entering DAC The value will display in scientific notation. Be aware that other values will also change to scientific notation and lose their names. Therefore, it may not be desirable to leave this feature enabled very long. This value should increase with a lower process level and decrease with a raised process level.

6 Table 1. Recorded values for troubleshooting Parameter Name/DAC Description Value ERRSIG Error from stable control CTLSIG Control signal from CPU VD Data count rate REF(cps) Reference count rate Standardization (zero) value Snapshot of detector signal Software version Time constant in seconds Weeks since standardization Source half-life In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. Canada +1 (905) (905) fax China +86 (10) (10) fax India +91 (20) (20) fax United Kingdom +44 (0) (0) fax United States +1 (800) (713) (713) fax HART is a registered trademark of the HART Communication Foundation Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819 Revision C Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819

More information

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL AM-806 Waste/Door Radiation Monitor OPERATIONS MANUAL Section Contents 1. SPECIFICATIONS 2. MAIN COMPONENTS 3. THEORY OF OPERATION 4. OPERATING INSTRUCTIONS 5. INSTALLATION Page 2 1.0 SPECIFICATIONS SENSITIVITY

More information

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 2014 Thermo Fisher

More information

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific. DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N 717818 Revision F Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter Installation

More information

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE Revision History Revision History Version of manual Description Date 1.0 Initial release 051025 1.1

More information

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D Part of Thermo Fisher Scientific DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D 2007 Thermo Fisher Scientific

More information

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER

MAINTENANCE & TROUBLESHOOTING GUIDE LEAK ALARM CHANNEL DRY OIL WATER AUX ALARM HIGH LOW CRITICAL WATER TANK LEVEL ALARM MODEL LDE-740 ADVANCE PAPER PNEUMERCATOR Liquid Level Control Systems Electronic Systems Excluding LC2000 And TMS Series MAINTENANCE & TROUBLESHOOTING GUIDE 400 300 500 FUEL LEVEL LEAK MONITOR 1 200 600 OIL/GAS NORMAL WATER PNEUMERCATOR

More information

RELEASE DEVICES RELEASE DEVICE-WPS MODELS A/B INSTALLATION MANUAL U.L. LISTED CANADIAN LISTED CSFM: :100 GENERAL DESCRIPTION:

RELEASE DEVICES RELEASE DEVICE-WPS MODELS A/B INSTALLATION MANUAL U.L. LISTED CANADIAN LISTED CSFM: :100 GENERAL DESCRIPTION: RELEASE DEVICES MADE IN THE U.S.A. RELEASE DEVICE-WPS MODELS A/B INSTALLATION MANUAL U.L. LISTED CANADIAN LISTED CSFM: 7300-1418:100 GENERAL DESCRIPTION: S/N: The "WPS" World Power Series, Time Delay Release

More information

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific. DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N 717818 Revision D Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter Installation

More information

Continuous Gamma Level System Installation Guide P/N

Continuous Gamma Level System Installation Guide P/N 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F Part of Thermo Fisher Scientific 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F 2009 Thermo Fisher

More information

ND4000 Manual Marine Hydraulics

ND4000 Manual Marine Hydraulics ND4000 Manual Marine Hydraulics Version 1.0 May 2012 Page 1 Contents Line Counter Components... 3 Proximity Sensors... 3 Connecting the Proximity Sensors to the SSCD... 3 Fuse... 3 USB... 4 Computer Display...

More information

Firefly II Multi Voltage RELEASE DEVICES. Firefly II & IIB-MV INSTALLATION MANUAL

Firefly II Multi Voltage RELEASE DEVICES. Firefly II & IIB-MV INSTALLATION MANUAL Firefly II Multi Voltage RELEASE DEVICES Firefly II & IIB-MV INSTALLATION MANUAL GENERAL DESCRIPTION: S/N The The Cookson Company Firefly II-MV time delay release devices are designed for use on rolling

More information

Ronan Model X96S Series Density System

Ronan Model X96S Series Density System Ronan Model X96S Series Density System For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing new

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TT-1343 5/06b INSTALLATION INSTRUCTIONS Original Issue Date: 8/03 Model: Automatic Transfer Switches Equipped with Series 1000 Programmable Controller Market: ATS Subject: Remote Annunciator Kits GM28938-KP1,

More information

FIRERAY 5000 range USER GUIDE

FIRERAY 5000 range USER GUIDE FIRERAY 5000 range USER GUIDE 0044-003-04 IMPORTANT PLEASE NOTE: The beam path MUST be kept clear of obstructions at all times! Failure to comply may result in the Detector initiating a Fire or Fault signal.

More information

SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE U.S.A. SS90-WPS MODEL A/B INSTALLATION MANUAL

SOLID STATE SECURITIES, INC. THE SS90 SERIES RELEASE DEVICES MADE IN THE U.S.A. SS90-WPS MODEL A/B INSTALLATION MANUAL S S S S S S LISTED L * PATENT PENDING S S 90 OLID TATE ECURITIES, INC SOLID STATE FAIL-SAFE UNIT U 99Y9 RESET RELEASING DEVICE SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE

More information

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications Instruction Manual Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications Manual PCT30006, Revision 9/20/2016 The PCT30006 control enclosure incorporates

More information

SERIES VAC Microprocessor-Based Direct Spark Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION. Export Information (USA)

SERIES VAC Microprocessor-Based Direct Spark Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION. Export Information (USA) R SERIES 35-70 120 VAC Microprocessor-Based Direct Spark Ignition Control F-35-70 November 2015 FEATURES Safe start with DETECT-A-FLAME flame sensing technology Custom pre-purge and inter-purge timings

More information

AQ255 and AQ257 Series Expansion Zoning Panels

AQ255 and AQ257 Series Expansion Zoning Panels AQ55 and AQ57 Series Expansion Zoning Panels PRODUCT DATA AQ574B AQ558B FEATURES The AQ55/AQ57 Series Expansion Zoning Panels have the following features: Availability of 4 Expansion Zoning Panels. Zone

More information

RC4000 Radiation Detection System

RC4000 Radiation Detection System RC4000 Radiation Detection System LAURUS Systems Inc. 3460 Ellicott Center Drive Suite 101 Ellicott City, MD 21043 Tel. +1 (410) 465-5558 Fax. +1 (410) 465-5257 Web Site: www.laurussystems.com Email: radcomm.sales@laurussystems.com

More information

SAFETY INFORMATION AND WARNINGS

SAFETY INFORMATION AND WARNINGS This manual refers to the Model SST-3 control panel manufactured since October 31, 2013, which uses a universal (100 277 VAC; 50/60 Hz) power supply. Older units use a voltage-specific power supply and

More information

User Manual. Dryer Controller M720

User Manual. Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 1.00 Software version 1.00 Preliminary version Manual M720 Dryer controller Page 1 of 42 Document history Preliminary version: - Created in April, 2009

More information

MVP-D-TEK Vehicle Loop Detector 9 Volts DC to 240 Volts AC. Operating Instructions

MVP-D-TEK Vehicle Loop Detector 9 Volts DC to 240 Volts AC. Operating Instructions MVP-D-TEK Vehicle Loop Detector 9 Volts DC to 240 Volts AC Operating Instructions We have designed the new MVP-D-TEK vehicle loop detector with the following objectives in mind: 1. Compact package to allow

More information

Instruction Manual PCS-400-XY. Pump Controller. Y=2: 4-20 ma Y=6: 0-10 VDC X=5: 1-5 VDC. 4 Pump Control Relays 1 Alarm Relay 1 Signal Fail

Instruction Manual PCS-400-XY. Pump Controller. Y=2: 4-20 ma Y=6: 0-10 VDC X=5: 1-5 VDC. 4 Pump Control Relays 1 Alarm Relay 1 Signal Fail Manufacturers of Process Controls and Instrumentation Instruction Manual Model: Function: Input: PCS-400-XY Pump Controller X=2: 4-20 ma X=5: 1-5 VDC X=6: 0-10 VDC X=7: Output: Power: 4 Pump Control Relays

More information

IMPORTANT. PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times!

IMPORTANT. PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times! USER GUIDE English IMPORTANT PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times! Failure to comply may result in the Detector initiating a Fire or Fault signal. Contents

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

MADE IN THE U.S.A. MODELS A/B INSTALLATION MANUAL

MADE IN THE U.S.A. MODELS A/B INSTALLATION MANUAL RELEASE DEVICES MADE IN THE U.S.A. MODELS A/B INSTALLATION MANUAL U.L. LISTED CANADIAN LISTED CSFM: 7300-1418:100 GENERAL DESCRIPTION: S/N: These Time Delay Release Devices are U.L. Listed, Canadian Listed,

More information

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications Instruction Manual Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications Manual PCT-3000 Revision 9/2014 The PCT-3000 series control console incorporates

More information

THE "SS90" SERIES RELEASE DEVICES MADE IN THE U.S.A. MODELS A/B INSTALLATION MANUAL

THE SS90 SERIES RELEASE DEVICES MADE IN THE U.S.A. MODELS A/B INSTALLATION MANUAL S S S S S S OLID TATE ECURITIES, INC S S 90 SOLID STATE FAIL-SAFE UNIT * PATENT PENDING LISTED 99Y9 RESET U L RELEASING DEVICE SOLID STATE SECURITIES, INC. THE "SS90" SERIES RELEASE DEVICES MADE IN THE

More information

Net/X Thermostat Network Installation and Troubleshooting Guide

Net/X Thermostat Network Installation and Troubleshooting Guide Net/X Thermostat Network Installation and Troubleshooting Guide Introduction This document contains basic procedures that can be used when installing the Net/X Thermostat network. It is highly recommended

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas User s Manual TIGER S EYE E-Series Mark V Jockey TIGERFLOW Systems, Inc. 4034 Mint Way Dallas, Texas 75237 214-337-8780 www.tigerflow.com TABLE OF CONTENTS Introduction... 4 Sequence of Operation... 5

More information

Motorised Infrared Optical Beam Smoke Detector. User Guide

Motorised Infrared Optical Beam Smoke Detector. User Guide Motorised Infrared Optical Beam Smoke Detector User Guide EN 1. General Information 50cm 50cm 8-100m Ensure clear line of sight from Detector to Reflector Mount on solid surfaces (structural wall or girder)

More information

Thermo Scientific Ramsey TM Series Sensing Speed Conditions on Rotating Shafts and Machinery

Thermo Scientific Ramsey TM Series Sensing Speed Conditions on Rotating Shafts and Machinery Thermo Scientific Ramsey TM Series 60-200 Sensing Speed Conditions on Rotating Shafts and Machinery The Thermo Scientific Ramsey Series 60-200 monitors under-speed, overspeed and zero-speed conditions

More information

Dryer Controller M720

Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 2.00 Software version 2.00 Manual M720 Dryer controller Page 1 of 60 Document history Preliminary version: - Created in April, 2009 Hardware Version 2.00,

More information

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg. WARD OXYGEN Installation, Operation, and Maintenance Instructions MEGA medical electronic gas alarm MEGA WARD MEGA medical electronic gas alarm WARD POWER SUPPLY Medical Electronic Gas Alarm OXYGEN SYSTEM

More information

Motorised Infrared Optical Beam Smoke Detector. User Guide

Motorised Infrared Optical Beam Smoke Detector. User Guide Motorised Infrared Optical Beam Smoke Detector User Guide EN 1. General Information 50cm 50cm 8-100m Ensure clear line of sight from Detector to Reflector Mount on solid surfaces (structural wall or girder)

More information

MSPA-MP METAPACK SERVICE MANUAL. Visual step-by-step guide to easily identify & correct technical problems! Gecko Electronics Inc.

MSPA-MP METAPACK SERVICE MANUAL. Visual step-by-step guide to easily identify & correct technical problems! Gecko Electronics Inc. MSPA-MP METAPACK SERVICE MANUAL Gecko Electronics Inc. Visual step-by-step guide to easily identify & correct technical problems! Table of Contents Power & Ground Check Tools and Parts 3 Electrical Wiring

More information

OVEN INDUSTRIES, INC.

OVEN INDUSTRIES, INC. OVEN INDUSTRIES, INC. OPERATING MANUAL Model 5C7-252 TEMPERATURE CONTROLLER With PLC Inputs Introduction Thank you for purchasing our controller. The Model 5C7-252 is an exceptionally versatile unit and

More information

S/601 Controller Installation Manual

S/601 Controller Installation Manual Checked Version Release date QA V4.6.5 PC1 EN 02.07.2012 2011 S/601 Controller Installation Manual FLOW CONTROL PC1 - Pump controller Introduction Thank you for using the S/601 flow, batch and pump control

More information

ACSI AO ELECTRIC LATCH RETRACTION CONTROLLER INSTALLATION INSTRUCTIONS I.D. 1089, REV. F

ACSI AO ELECTRIC LATCH RETRACTION CONTROLLER INSTALLATION INSTRUCTIONS I.D. 1089, REV. F II 1400-2 ACSI 1406-04-AO ELECTRIC LATCH RETRACTION CONTROLLER INSTALLATION INSTRUCTIONS I.D. 1089, REV. F INSTALLATION Transformer Model BE31763001 by Basler Electric, Transformer Model 2010028 by Galaxy

More information

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea User's Guide CONTENTS FEATURES WHAT YOU NEED TO GET STARTED INSTALLATION DETAILS Installing the Antenna Installing the External Alert Siren The main console OPERATING PROCEDURE Power On Routine Push Buttons

More information

DE20 INSTRUCTION MANUAL

DE20 INSTRUCTION MANUAL DE20 INSTRUCTION MANUAL Now UL Listed! SET POINT DE20 SET YOUR DISTRIBUTOR: Chemical Distributors, Inc. 80 Metcalfe Street - Buffalo, NY 14206 Phone: 800-777-2436 716-856-2300 Fax: 716-856-7115 sales@cdibuffalo.com

More information

1. Quick start-up guide

1. Quick start-up guide 1. Quick start-up guide This quick guide leads the way to install, start-up and configure necessary parameters in the normal cases. 1. PREPARING INSTALLATION Install fiber optic cable inside conduit. This

More information

Operating Instructions

Operating Instructions Operating Instructions DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement Document ID: 31392 Nuclear Revision history Version Description Date 1.0 Initial release. Formerly

More information

General Purpose IO Technical Manual

General Purpose IO Technical Manual General Purpose IO Technical Manual Revision 1.06 8 November 2013 Pakton Technologies PAE222 GPIO Manual.docx Page 1 of 21 Revision 1.06 Last updated 8/11/2013 Table of Contents INTRODUCTION...3 Scope

More information

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA 99336 www.jmcvr.com 1-509-586-9893 Table of Contents TABLE OF FIGURES...1 OVERVIEW OF THE HRC CAPABILITIES...2 INSTALLATION AND

More information

Analog Room Pressure Monitor RPC Series

Analog Room Pressure Monitor RPC Series Description The Room Pressure Monitor is used to measure differential pressure in the range of 0.125 to 1"wc or 30 to 250 Pa. It combines precision high sensitivity silicon sensing capabilities and the

More information

Installation, Operating and Maintenance Manual

Installation, Operating and Maintenance Manual STATUS ZONES CONTROLS FIRE FAULT DISABLED FIRE 1 2 3 4 5 6 7 8 TEST FAULT DISABLED 1 5 BUZZER SILENCE RESET 1 2 TEST 2 6 LAMP TEST 3 SUPPLY 3 7 SYSTEM FAULT 4 8 SOUNDERS ACTIVATE/ SILENCE 4 FAULTS INSTRUCTIONS

More information

Contamination Meter Users Manual

Contamination Meter Users Manual RDS-80 Contamination Meter U Users Manual Version 1.01 LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - www.laurussystems.com NOTE This document and the information herewith are copyrighted

More information

Module Features are-configurable, no module jumpers to set

Module Features are-configurable, no module jumpers to set December 2011 PACSystems* RX3i Isolated Thermocouple Input Module, 6 Channels, IC695ALG306 Isolated Thermocouple Input Module, 12 Channels, IC695ALG312 Isolated Thermocouple Input module IC695ALG306 provides

More information

Apollo Intelligent Auto Aligning Beam Detector. User Guide

Apollo Intelligent Auto Aligning Beam Detector. User Guide Apollo Intelligent Auto Aligning Beam Detector User Guide EN 1. General Information 50cm 50cm 8-100m Ensure clear line of sight from Detector to Reflector Mount on solid surfaces (structural wall or girder)

More information

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS... SECTION Table of Contents SECTION 1 INTRODUCTION...................... 1.1 Description............................... 1.2 Features................................. 1.3 Models..................................

More information

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm Operating & Maintenance Manual Alert-4 Ethernet LCD Master Alarm w w w. a m i c o. c o m Contents User Responsibility 4 Introduction 4 Features 5 Description of the Alarm 5 Shipment Details 5 The Alarm

More information

Fire Control Panel FS5100

Fire Control Panel FS5100 Fire Control Panel FS5100 INSTRUCTION MANUAL Revision 6/02.11 Contents 1. Introduction... 5 2. Terminology... 5 3. Function... 7 4. Technical data... 7 4.1. Modules... 7 4.1.1. Type of modules... 7 4.1.2.

More information

FlashGuard 3000B Dual Lighting System Troubleshooting Guide

FlashGuard 3000B Dual Lighting System Troubleshooting Guide FlashGuard 3000B Dual Lighting System Troubleshooting Guide Table of Contents Section Flashhead (Strobe) Troubleshooting Flowchart 1 Multiple Strobe Troubleshooting Flowchart 2 Sidelight Troubleshooting

More information

ORP (mv) Controller Model: ORP-XP2. UP DOWN ENTER Time & Date Main Menu Alarm Reset

ORP (mv) Controller Model: ORP-XP2. UP DOWN ENTER Time & Date Main Menu Alarm Reset Instruction Manual ORP (mv) Controller Model: ORP-XP2 Power Control Alarm Comms UP DOWN ENTER Time & Date Main Menu Alarm Reset Supplied by: Convergent Water Controls Pty Ltd 2/4 Huntley Street, PO Box

More information

PL-1 Thermostat User s Guide

PL-1 Thermostat User s Guide Table of Contents Introduction... 1 PL-1 Thermostat User s Guide Version 2.0 Firmware Version 2.0 Conventions... 2 Front Panel Layout... 3 Thermostat Modes... 4 Thermostat Mode Programming... 5 Set Point

More information

NISSAN INTEGRATED HOMELINK TRANSMITTER

NISSAN INTEGRATED HOMELINK TRANSMITTER Classification: Reference: Date: EL99-012a NTB99-067a November 7, 2003 NISSAN INTEGRATED HOMELINK TRANSMITTER APPLIED VEHICLES: 1996-2004 Pathfinder - with HomeLink Option 1997-2004 Maxima - with HomeLink

More information

SERIES VAC Microprocessor-Based Direct Spark Ignition Control with Inducer Blower Relay FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION

SERIES VAC Microprocessor-Based Direct Spark Ignition Control with Inducer Blower Relay FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION R SERIES 35-61 24 VAC Microprocessor-Based Direct Spark Ignition Control with Inducer Blower Relay F-35-61 August 2015 FEATURES Safe start with DETECT-A-FLAME flame sensing technology Custom pre-purge

More information

Introduction... 3 Setup & Suggestions... 4 Basic Use... 4 Setting Time... 4 Setting Temperature... 5 Setting Height / Pressure...

Introduction... 3 Setup & Suggestions... 4 Basic Use... 4 Setting Time... 4 Setting Temperature... 5 Setting Height / Pressure... Table of of Contents Contents... 2 Introduction... 3 Setup & Suggestions... 4 Basic Use... 4 Setting Time... 4 Setting Temperature... 5 Setting Height / Pressure... 6 Aligning the pedestal... 6 Guidelines

More information

ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS

ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS II 1400-3 ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS Features: For use with doors or groups of doors that must interlock with each other to regulate control of accessing one area from another,

More information

Ronan Model X96S Series Weighing System

Ronan Model X96S Series Weighing System Ronan Model X96S Series Weighing System For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing new

More information

IJ3000 IJ3000 Flushing System N Revision B Installation Instructions Page 1 of 10

IJ3000 IJ3000 Flushing System N Revision B Installation Instructions Page 1 of 10 IJ3000 IJ3000 Flushing System 5760-029N Revision B Installation Instructions Page 1 of 10!! WARNING: Do not turn on until all plumbing and electrical connections have been made. WARNING: The flushing system

More information

Sierra Model 951 Digital Flo-Box. Instruction Manual Part Number IM-951 Revision A 09-05

Sierra Model 951 Digital Flo-Box. Instruction Manual Part Number IM-951 Revision A 09-05 Sierra Model 951 Digital Flo-Box Instruction Manual Part Number IM-951 Revision A 09-05 5 Harris Court, Building L Monterey, CA 93940 (831) 373-0200 (800) 866-0200 Fax (831) 373-4402 www.sierrainstruments.com

More information

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION IBAS LIGHTING CONTROL

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION IBAS LIGHTING CONTROL PART 0 A/E INSTRUCTIONS 0.01 DESIGN REQUIREMENTS A. Minimum lighting control (base scope of work) shall include all corridors. The lighting in all the corridors shall be turned on (if not already on) if

More information

Universal Gas Detector

Universal Gas Detector Universal Gas Detector Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Drive Lake Zurich, Illinois 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 pureairemonitoring.com

More information

Ethernet General Purpose

Ethernet General Purpose Ethernet General Purpose Technical Manual Revision 1.03 8 November 2013 Pakton Technologies IO PAE224 Ethernet GPIO Manual.docx Page 1 of 22 Revision 1.03 Last updated 8/11/2013 Table of Contents INTRODUCTION...3

More information

RES-210 (-211, -220, -221) / RES-420

RES-210 (-211, -220, -221) / RES-420 RESISTR GB Replacing RES-2xx/ Replacement Instructions The RESISTR temperature controller can be used as an alternative to the following controllers of the series 200 : RES-210: Analog temperature display,

More information

Revision November 2013 JVA Technologies. Ethernet General Purpose IO Technical Manual

Revision November 2013 JVA Technologies. Ethernet General Purpose IO Technical Manual Revision 1.03 8 November 2013 JVA Technologies Ethernet General Purpose IO Technical Manual www.jva-fence.com.au Table of Contents INTRODUCTION...3 Scope and Purpose...3 Glossary...3 SPECIFICATIONS...4

More information

Waters 2795 Separations Module

Waters 2795 Separations Module S U P P L E M E N T A L R E L E A S E N O T E S Waters 2795 Separations Module Software Update Version 2.05 Firmware Update Version 2.05 Information Version 2.05 firmware incorporates improved checking

More information

Spa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert)

Spa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert) Spa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert) P.N. 7876C (export) February 12, 2015 For Spas equipped with BP2100, BP6013 controllers and Spa Touch panel. Spa

More information

Waters 2695D Separations Module

Waters 2695D Separations Module S U P P L E M E N T A L R E L E A S E N O T E S Waters 2695D Separations Module Firmware Update Version 2.04d Firmware Update Version 2.04d Information This firmware update incorporates improved connectivity

More information

Pump-Up Controller MODEL 4062

Pump-Up Controller MODEL 4062 Pump-Up Controller 4-20mA Input/Scalable Output Seal Fail Monitoring Duplex Pump Alternation Hand-Off-Auto Controls Dual Run-time Meters RS-485/Modbus Communications DESCRIPTION The Model 4062 Pump-Up

More information

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs User Manual Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs Models 89800-03 and 89800-04 THE STANDARD IN PRECISION MEASUREMENT Table of Contents

More information

HT-2 / 9600 Series Control Contents

HT-2 / 9600 Series Control Contents HT-2 / 9600 Series Control Contents Tools & Parts Tools Required Parts Required Error Messages 3 Flashing Dots Pressure or Flow Switch Not Activated Pressure or Flow Switch Activated Temperature Sensor

More information

CAUTION. Installation Manual

CAUTION. Installation Manual System Controller SC-301-6M CAUTION Installation Manual In order to use this product safely, read this installation manual carefully and follow the installation instructions. Potential dangers from accidents

More information

SERIES VAC Microprocessor-Based Intermittent Pilot Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION

SERIES VAC Microprocessor-Based Intermittent Pilot Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION R SERIES 35-703 120 VAC Microprocessor-Based Intermittent Pilot Ignition Control F-35-703 July 2016 FEATURES Safe start with DETECT-A-FLAME flame sensing technology Custom pre-purge and inter-purge timings

More information

Spa Touch Control Panel with 2000, 2100 controllers. (Spa Owner s Manual insert)

Spa Touch Control Panel with 2000, 2100 controllers. (Spa Owner s Manual insert) Spa Touch Control Panel with 2000, 2100 controllers (Spa Owner s Manual insert) P.N. 7876B February 11, 2015 For Spas equipped with BP2000, BP2100 controllers and Spa Touch panel. Spa Touch Control Panel

More information

SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE -

SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE - SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN VC3ED ML-137013 - NOTICE - This Manual is prepared for the use of trained Vulcan Service Technicians and should not be used by those not properly

More information

ITC 2000 HEATING CONTROL CABINET. Installation & Operation Manual

ITC 2000 HEATING CONTROL CABINET. Installation & Operation Manual ITC 2000 HEATING CONTROL CABINET Installation & Operation Manual VTI, Incorporated 24 McMillan Way Newark, DE 19713 Phone (302) 738 0500 FAX (302) 738 6594 Revision Level 0.03 Manual No. 90003314 June,

More information

Waters 2695 Separations Module

Waters 2695 Separations Module S U P P L E M E N T A L R E L E A S E N O T E S Waters 2695 Separations Module Firmware Update Version 2.04 Firmware Update Version 2.04 Information This firmware update incorporates improved checking

More information

Macurco DVP-120 Detection and Ventilation Control Panel

Macurco DVP-120 Detection and Ventilation Control Panel Macurco DVP-120 Detection and Ventilation Control Panel User Instructions IMPORTANT: Keep these User Instructions for reference. TABLE OF CONTENTS 1 Introduction 4 1.1 General Information 4 1.2 Features

More information

DYGIZONE GJD910 Lighting Controller & Enunciator

DYGIZONE GJD910 Lighting Controller & Enunciator DYGIZONE GJD910 Lighting Controller & Enunciator MASTER WIRING IDENTIFICATION Power up to the DygiZone and you will see: All the LED s (red,yellow,green and blue buttons) will flash All the LCD icons will

More information

Transmission Cooling System J Operation Manual. TRANSFLOW Transmission Cooling System Service Tool J

Transmission Cooling System J Operation Manual. TRANSFLOW Transmission Cooling System Service Tool J Transmission Cooling System J-45096 Operation Manual TRANSFLOW Transmission Cooling System Service Tool J-45096 1 Safety Precautions WARNING: TO PREVENT PERSONAL INJURY AND/OR DAMAGE TO EQUIPMENT: ALLOW

More information

Pump-Down Controller MODEL 4052

Pump-Down Controller MODEL 4052 Pump-Down Controller 4-20mA Input/Scalable Output Seal Fail Monitoring Duplex Pump Alternation Hand-Off-Auto Controls Dual Run-time Meters RS-485/Modbus Communications DESCRIPTION The Model 4052 Pump-Down

More information

AOYUE 968A+ INT. Deluxe Repairing System INSTRUCTION MANUAL

AOYUE 968A+ INT. Deluxe Repairing System INSTRUCTION MANUAL AOYUE INT 968A+ Deluxe Repairing System INSTRUCTION MANUAL Thank you for purchasing Aoyue INT 968A+ Repairing System. It is important to read the manual before using the equipment. Please keep manual in

More information

RCS Residential Control Systems Inc.

RCS Residential Control Systems Inc. RCS Residential Control Systems Inc. Model TZ16 Z-Wave Communicating Thermostat with Rev P HVAC Control Unit INSTALLATION AND OPERATION MANUAL DCN: 141-00882 Rev 02 5/18/06 This manual applies to the following

More information

Generator Condition Monitor (GCM-X)

Generator Condition Monitor (GCM-X) UTILITY SYSTEMS Generator Condition Monitor (GCM-X) Installation and Operation Manual GA0353P01 Rev. G Contents Specifications... 6 Application... 7 Principle of Operation... 7 System Description... 9

More information

Duct and Rough Service Carbon Monoxide Sensor

Duct and Rough Service Carbon Monoxide Sensor Product Identification and Overview Duct and Rough Service Carbon Monoxide Sensor BAPI s Carbon Monoxide Sensor offers enhanced electrochemical sensing with outstanding accuracy at low concentrations.

More information

Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E

Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E Thermo Fisher Scientific Messtechnik GmbH Frauenauracher Straße 96 91056 Erlangen Germany www.thermo.com info.rmp.erlangen@thermofisher.com

More information

CAREL IR33+ CONTROLLER

CAREL IR33+ CONTROLLER CAREL IR33+ CONTROLLER Programming Instructions The Carel IR33+ controller has a two-level menu system for programming the operation of the case; Level 1 and Level 2 program settings. Unlike IR33, the

More information

D A R A - 4. Data Aire Relay Auto-Changeover

D A R A - 4. Data Aire Relay Auto-Changeover D A R A - 4 Data Aire Relay Auto-Changeover WARNING: If adding the DARA-4 panel to existing units that did not include a DARA-4 panel when originally purchased it will be necessary to add a relay or relays.

More information

Model 1000 H 2 S E-Chem/ CO 2 Process Analyzer

Model 1000 H 2 S E-Chem/ CO 2 Process Analyzer 24VDC OUT 2.1A 18-36 VDC IN Model 1000 H 2 S E-Chem/ CO 2 Process Analyzer + + SAMPLE FLOW AIR FLOW SAMPLE FLOW Operator s Installation and Instruction Manual Detcon Model 1000 H 2 S/CO 2 Analyzer consists

More information

Safe T Net 210. Instruction Manual Part Number Mar2010

Safe T Net 210. Instruction Manual Part Number Mar2010 Safe T Net 210 Instruction Manual Part Number 71-0012 31Mar2010 2007 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available

More information

Dry Contact Probe DCP (#30008)

Dry Contact Probe DCP (#30008) Dry Contact Probe DCP (#30008) The Networked Robotics DCP probe enables network data collection of the state of any normally open or normally closed switch. This product can sense alarm contact outputs

More information

T3/E3 Fiber Optic Line Drivers

T3/E3 Fiber Optic Line Drivers MT618A-ST-R3 January 2004 MT619A-ST-R2 MT620AE-R2 T3/E3 Fiber Optic Line Drivers CUSTOMER SUPPORT INFORMATION Order toll-free in the U.S.: Call 877-877-BBOX (outside U.S. call 724-746-5500) FREE technical

More information

Table of Contents. Part I SAFETY MESSAGE TO INSTALLERS OF EMERGENCY WARNING EQUIPMENT 2 Part II Vehicle Operation 4. Part III Safety Messages 13

Table of Contents. Part I SAFETY MESSAGE TO INSTALLERS OF EMERGENCY WARNING EQUIPMENT 2 Part II Vehicle Operation 4. Part III Safety Messages 13 Manual I SS2000 TouchScreen Installation Guide Table of Contents Foreword 0 Part I SAFETY MESSAGE TO INSTALLERS OF EMERGENCY WARNING EQUIPMENT 2 Part II Vehicle Operation 4 1 Installing the... SS2000 TouchScreen

More information

SECTION 1: STARTING 1.1 UNPACKING AND INSPECTION

SECTION 1: STARTING 1.1 UNPACKING AND INSPECTION SECTION 1: STARTING 1.1 UNPACKING AND INSPECTION Upon receipt of the laser system, inspect the shipping containers for gross external damage. Unpack the unit and inspect for internal damage. If any damage

More information

SERIES VAC Microprocessor-Based Hot Surface Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION AGENCY CERTIFICATIONS

SERIES VAC Microprocessor-Based Hot Surface Ignition Control FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION AGENCY CERTIFICATIONS SERIES 35-65 24 VAC Microprocessor-Based Hot Surface Ignition Control F-35-65 August 2015 FEATURES Safe start with DETECT-A-FLAME flame sensing technology Custom pre-purge and inter-purge timings* 120/240

More information