Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Size: px
Start display at page:

Download "Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc."

Transcription

1 Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Jake Zaragoza Site: Big Bend NP (BIBE-KB) Service Date(s): 2/21/217 to 2/22/217 Site Operator: Lydia Smith Subject: Semiannual Visit All site visit and calibration forms are attached, detailing the pre- and post-maintenance calibrations and test results. Calibration acceptance criteria are defined as the thresholds at which corrective action is required. Data acceptance criteria are defined as the thresholds at which the quality of the ambient data is questionable and may require invalidation. This report is not complete without the inclusion of the calibration form worksheets. SUMMARY OF FINDINGS GASEOUS POLLUTANT SUPPORT EQUIPMENT Zero-Air System (Werther / Panther Compressor PC7/4E): Maintenance - No maintenance for the pump or compressor was required at this time. The media for the zero-air system was replaced. Station Temperature: QAM style temperature sensor Pre-Maintenance Testing - The sensor was compared against a certified temperature transfer standard. The sensor was found to be responding within calibration acceptance criteria. Maintenance - No maintenance was performed. Post-Maintenance Testing - The sensor was compared against a certified temperature transfer standard. The sensor was found to be responding within calibration acceptance criteria. GASEOUS POLLUTANT ANALYZERS Ozone Analyzer (Thermo 49c): Pre-Maintenance Testing - The ozone analyzer was compared against a Level 2 ozone transfer standard. The ozone analyzer was found to be responding within calibration acceptance criteria. Maintenance - Cleaned fan and cells for analyzer. Analog output was adjusted. The lamp setting on the analyzer was increased to 49% and the span on the analyzer was calibrated. The fan was cleaned before the analog out and line loss test. Post-Maintenance Testing - The ozone analyzer was compared against a Level 2 ozone transfer standard. The ozone analyzer was found to be responding within calibration acceptance criteria.

2 Ozone Station Reference (Thermo 49c): Pre-Maintenance Testing - The Level 3 ozone station reference was recertified using a Level 2 ozone transfer standard. The ozone station reference was found to be responding within calibration acceptance criteria. Maintenance - Cleaned fan and cells of the station reference. Analog output was adjusted. The lamp setting on the station reference was increased to 5%. The logger correction factors were updated. The fan and cells were cleaned before the analog out and line loss test. Post-Maintenance Testing - The ozone station reference was compared against a Level 2 ozone transfer standard. The ozone station reference was found to be responding within calibration acceptance criteria. METEOROLOGICAL SENSORS Ambient Temperature (Climatronics 193): Pre-Maintenance Testing - The sensor was compared against a certified temperature transfer standard in three water baths controlled at temperatures between and 5 degrees Celsius. The sensor was confirmed to be responding within calibration acceptance criteria. The aspirator fan was functioning correctly. Maintenance - No maintenance was performed. Relative Humidity (Rotronic MP61): Pre-Maintenance Testing - The sensor was collocated with a certified relative humidity transfer standard. The sensor was confirmed to be responding within calibration acceptance criteria. The aspirator fan was functioning correctly. Maintenance - The sensor was replaced with a recently serviced Rotronic MP61A. Post-Maintenance Testing - The sensor was collocated with a certified relative humidity transfer standard. The sensor was confirmed to be responding within calibration acceptance criteria. Wind Speed (Climatronics 175): Pre-Maintenance Testing - The sensor was challenged with a certified anemometer drive. The sensor was confirmed to be responding within calibration acceptance criteria. The starting threshold test for the sensor was within acceptance criteria. Maintenance - The sensor was replaced with a recently serviced Climatronics wind speed sensor. Post-Maintenance Testing - The sensor was challenged with a certified anemometer drive. The sensor was confirmed to be responding within calibration acceptance criteria. The starting threshold test for the sensor was within acceptance criteria. Wind Direction (Climatronics 176): Pre-Maintenance Testing - The reference alignment for the sensor was checked using a compass. The reference alignment for the sensor was confirmed to be within acceptable limits. The accuracy of the sensor was tested by comparison to a reference. The sensor was confirmed to be responding within calibration acceptance criteria. The linearity of the sensor was within acceptable limits. The starting threshold test results were outside of acceptance criteria. Maintenance - The sensor was replaced with a recently serviced Climatronics wind direction sensor.

3 Post-Maintenance Testing - The reference alignment for the sensor was checked using a compass. The reference alignment for the sensor was confirmed to be within acceptable limits. The accuracy of the sensor was tested by comparison to a reference. The sensor was confirmed to be responding within calibration acceptance criteria. The linearity of the sensor was within acceptable limits. The starting threshold test for the sensor was within acceptance criteria. Solar Radiation (Li-Cor Pyranometer): Pre-Maintenance Testing - The sensor was collocated with a certified solar radiation transfer standard. The sensor was confirmed to be responding within calibration acceptance criteria. The sensor was found to be level and clean. Maintenance - The transfer standard could not be collocated immediately next to the site sensor, but was placed on a tripod instead. Post-Maintenance Testing - No post-maintenance checks were performed. The sensor was confirmed to be responding within calibration acceptance criteria. Precipitation (Climatronics 158): Pre-Maintenance Testing - The sensor was challenged using a known volume of water. The sensor was found responding within calibration acceptance criteria. The sensor was found to be level, but was not clean. The functionality of the heater was not assessed. Maintenance - The level was adjusted very slightly. Post-Maintenance Testing - The sensor was challenged using a known volume of water. The sensor was confirmed to be responding within calibration acceptance criteria. PARTICULATE MONITORS AND SAMPLERS CASTNET Filter Pack Flow: Pre-Maintenance Testing - A leak check on the system was performed and results were within acceptable limits. The flow was checked using a certified flow standard measuring flow in standard conditions. The measured flow was found to be outside of calibration acceptance criteria. Maintenance - The MFC was unable to control the flow. The flow box was swapped with a new style Alicat/Teflon filter box. The As left data was used to calculate new high/low output scaling factors. Post-Maintenance Testing - The flow was checked using a certified flow standard measuring flow in standard conditions. The measured flow was confirmed to be within calibration acceptance criteria. ADDITIONAL COMMENTS Lydia serves as the site operator and was able to meet with me during the visit. Lydia does a good job of keeping the site operational and accurate.

4 SITE INFORMATION ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 LATITUDE North Deg Min LONGITUDE West 13 1 Sec CALCULATE-> Decimal Decimal Deg Min Sec --CALCULATE-> ELEVATION Meters Feet --CALCULATE-> Feet --CALCULATE-> Meters Photo Documentation Completed? Yes No N/A Cellular Phone Coverage Cellular Phone Coverage Protocol? Carrier? # of Bars? Signal Strength? 1X Verizon 3 -X dbm -X dbm DAY TIME IN TIME OUT 2/21/217 9:3 8:3 2/22/217 9:3 4:3 SITE STANDARDS PM Flow Reference Please verify site standards used by the site operator MANUFACTURER MODEL SERIAL # Calibration Expiration Date NOTES:

5 TEMPERATURE / VERTICAL TEMPERATURE DIFFERENCE SYSTEM VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT National Park Service FIELD SPECIALIST Jake Zaragoza DATE 2/21/217 SITE NAME Big Bend National Park - K-Bar Ranch Road DATE OF LAST VISIT 7/12/216 MANUFACTURER MODEL Temperature Reference Eutechnics 44 SERIAL NUMBER 342 EXPIRATION DATE 1/3/218 2m Temperature Sensor Manufacturer Climatronics 193 Serial Number EPA149 List sensors according to height on tower, from highest to lowest. Manufacturer Serial Number Climatronics 193 EPA149 Temp. Deltas DATA ACCEPTANCE CRITERIA (<=) Ambient Temperature Difference ( o C) Vertical Temperature Difference ( o C) 1. N/A 2m Temperature Bath Temp ( o C) DAS Difference MAX ABS Difference.6 MAX ABS Difference Aspirator fan functional 2m? Yes No N/A Yes No N/A Yes No N/A Yes No N/A Each sensor was verified against its data channel? Yes No N/A Each Temperature Difference = Upper - Lower? Yes No N/A Bath Temp ( o C) MAX ABS Difference MAX ABS Difference NOTES: No maintenance performed.

6 ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road STATION TEMPERATURE SENSOR VERIFICATION & CALIBRATION Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 Temperature Reference MANUFACTURER MODEL SERIAL NUMBER Eutechnics EXPIRATION DATE 1/3/218 CALIBRATION ACCEPTANCE CRITERIA (<=) Temperature Difference ( o C) 1. Temperature Temperature Reference ( o C) DAS ( o C) Difference Reference ( o C) DAS ( o C) Difference NOTES: No maintenance was performed.

7 ABBR. BIBE-KB CLIENT SITE NAME RELATIVE HUMIDITY SENSOR VERIFICATION & CALIBRATION National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 RH SENSOR REFERENCE MANUFACTURER Rotronic MODEL HG2S3 SERIAL NUMBER EXPIRATION DATE 8/5/217 Manufacturer Serial Number Rotronic MP Manufacturer Serial Number Rotronic MP61A FIELD CALIBRATION ACCEPT. CRITERIA (<=) Relative Humidity Difference (%) 5% Hour Relative Humidity (%) Relative Humidity (%) STD DAS Difference V Hour STD DAS Difference 23.6% 26.2% 2.7% % 35.3% 1.1% 22.6% 25.1% 2.6% % 26.9% 2.3% 19.2% 21.8% 2.7% % 21.1% 2.9% 15.8% 18.2% 2.5% % 18.5% 2.6% 14.4% 16.6% 2.2% % 17.1% 2.3% Average 2.5% Average 2.2% V Aspirator fan functional? Yes No N/A NOTES:

8 WIND SPEED SENSOR VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE 2/21/217 DATE OF LAST VISIT 7/12/216 MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE Wind Speed Reference RM Young 1882A/1883A CA2241 1/26/218 Wind Speed Torque Gauge RM Young 1831 NA Manufacturer and Climatronics / Manufacturer and Climatronics / Heavy Duty Aluminum Heavy Duty Aluminum Sensor Serial # 4559 Sensor Serial # 1697 Cups Serial # 1967 Cups Serial # 1967 DATA ACCEPTANCE CRITERIA (<=) Wind Speed Difference (m/s).25 Wind Speed Difference (%) Wind Speed Threshold (m/s vs % Diff.) 5.% 5. if wind speed <= 5 m/s if wind speed > 2 m/s 2. Select UNITS m/s Motor Speed (rpm) Target Speed Wind Speed DAS Difference.231 N/A N/A N/A % % % % % Starting Threshold Torque <=.3 g-cm TORQUE.1 NO ACTION REQUIRED Heater sleeve functional? Yes No N/A Wind Speed Motor Speed (rpm) Target Speed DAS Difference..238 N/A N/A N/A % % % % % Starting Threshold Torque <=.3 g-cm TORQUE.1 NO ACTION REQUIRED NOTES:

9 WIND DIRECTION SENSOR VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT National Park Service FIELD SPECIALIST Jake Zaragoza DATE 2/21/217 SITE NAME Big Bend National Park - K-Bar Ranch Road DATE OF LAST VISIT 7/12/216 Direction Alignment Reference Direction Linearity Reference Direction Torque Gauge MANUFACTURER MODEL SERIAL NUMBER Brunton Transit ARS Jig 45 Degree Disc NA RM Young 1831 NA EXPIRATION DATE Manufacturer & Sensor Serial # Vane Serial # Climatronics Manufacturer & Climatronics ? Sensor Serial # Vane Serial # 3498 Local Magnetic Declination (degrees) Method NOAA ACCEPTANCE CRITERIA (<=) 6.7 Cross-arm Alignment Error (degrees) (CAL) 2 Total Align. Diff (degrees) (DATA) 5 Sensor Linearity (degrees) (CAL) 5 Mag. Dec. from NOAA (deg/min/sec) Landmarks From the North From the South From the East From the West. Degrees Is the Reference Alignment intended to be N-S? YES Is the Reference Alignment intended to be N-S? YES Reference Alignment (degrees) 2. Reference Alignment (degrees) 2. SENSOR ALIGNMENT N-S Reference Degrees DAS From the North From the South 18 From the East 9 From the West 27 Total Alignment MAX ABS Diff OR Difference SENSOR ALIGNMENT N-S Reference Degrees DAS From the North From the South 18 From the East 9 From the West 27 Total Alignment MAX ABS Diff Difference SENSOR ALIGNMENT SENSOR ALIGNMENT Landmark Degrees DAS Difference Landmark Degrees DAS Difference From the North From the North From the South 18 From the South 18 From the East 9 From the East 9 From the West 27 From the West 27 Total Alignment MAX ABS Diff Total Alignment MAX ABS Diff OR SENSOR ALIGNMENT SENSOR ALIGNMENT X Reference Degrees DAS Difference X Reference Degrees DAS Difference Align with Ref (N) Align with Ref (S) Perp with Ref (E) Perp with Ref (W) Total Alignment MAX ABS Diff 4.1 Total Alignment MAX ABS Diff 3. OR OR SENSOR LINEARITY Point DAS MAX Difference Difference N/A SENSOR LINEARITY Point DAS Difference N/A MAX Difference 2 NO ACTION REQUIRED NO ACTION REQUIRED Starting Threshold TORQUE Starting Threshold Torque <= 6. g-cm 7. Torque <= 6. g-cm ACTION REQUIRED TORQUE 4. NO ACTION REQUIRED Heater sleeve functional? Yes No N/A NOTES: No sensor S/N was found on the 'as found' sensor.

10 ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road SOLAR RADIATION SENSOR VERIFICATION & CALIBRATION Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 Solar Radiation Reference MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE LiCor Li-2X-L PY /17/217 MULTIPLIER 2. Manufacturer Serial Number Translator Logger Type ESC LiCor Pyranometer PY92131 Climatronics 354 High Input (V) 5. Low Input (V). High Output Low Output. Manufacturer LiCor Pyranometer Serial Number PY92131 Translator Climatronics 354 Logger Type High Input (V) 5. ESC Low Input (V). High Output Low Output. CALIBRATION ACCEPTANCE CRITERIA (<=) Difference from CTS (%) 5% DATA ACCEPTANCE CRITERIA (<=) Difference from CTS (%) 1% Solar Radiation Hour CTS (W/m 2 ) DAS (W/m 2 ) Difference % % % % V DARK RESPONSE DAS (W/m 2 ) 1 MEAN ABS % DIFF 3.6% Sensor found clean? Yes No Sensor found level? Yes No Solar Radiation Hour CTS (W/m 2 ) DAS (W/m 2 ) Difference V DARK RESPONSE DAS (W/m 2 ) MEAN ABS % DIFF NOTES:

11 PRECIPITATION SENSOR VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 MANUFACTURER MODEL SERIAL NUMBER Precipitation Reference Novalynx Drip Bottle NA EXPIRATION DATE Manufacturer Serial Number Climatronics Manufacturer Serial Number Climatronics CALIBRATION ACCEPTANCE CRITERIA (<=) Difference from Input Volume (%) 5% DATA ACCEPTANCE CRITERIA (<=) Difference from Input Volume (%) 1% X Reference Chart Input Volume (ml) Manufacturer Climatronics Climatronics G-H 158 Diameter (in.) mm/tip ml/tip Met One Met One RM Young Texas Electronics TR-525I-HT DAS target Conversions Value Units Value 1. inch mm 1. Units mm inch Reference (ml) Target (mm) Precipitation DAS (mm) Difference % Heater functional? Yes No N/A Sensor found level? Yes No Sensor found clean? Yes No Precipitation Reference (ml) Target (mm) DAS (mm) Difference % NOTES: Cleaned tipping bucket. Adjusted level slightly.

12 CASTNET FLOW SYSTEM VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 MFC High Flow Reference MANUFACTURER MODEL SERIAL NUMBER Bios Definer 22-H EXPIRATION DATE 1/9/217 MFC Manufacturer Serial Number Apex AX-MC-1SLPM-D-PCV65/5M MFC Manufacturer Serial Number Alicat MC-1SLPM-D-PCV Pump Manufacturer Thomas 17CA18 Serial Number Flow Target 3. ENTER FLOWS IN SLPM Pump Manufacturer Serial Number Thomas 17CA MFC Display Linear MFC VDC Linear Regression Regression Date 2/21/217 Setpoint Slope Setpoint Slope Intercept Intercept ESC Logger Scaling High Input (V) Low Input (V) High Output (LPM) Low Output (LPM) Operational Checks Vacuum DAS Flow MFC Flow Flow Standard Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Difference Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % FAIL 3 (target) 3. SLPM % FAIL 4 (high) 3.5 SLPM % FAIL 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM) -.7 *Note: A leak is present if the difference between the zero and leak test value is greater than.1 lpm Recalibration REQUIRED *Note: A calibration is required if the difference between the transfer standard and the DAS value is greater than +/- 2.% at the target setting or greater than +/- 2.5% at either the low or high point NEW CALIBRATION FACTORS MFC Display Linear MFC VDC Linear Regression Regression ESC Logger Scaling Setpoint Slope Intercept Setpoint Slope Intercept High Input (V) Low Input (V) High Output (LPM) Correlation.281 Correlation.629 Low Output (LPM).256 Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Difference Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % 3 (target) 3. SLPM % 4 (high) 3.5 SLPM % 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM). *Note: A leak is present if the difference between the zero and leak test value is greater than.1 lpm Operational Checks Vacuum 2. DAS Flow 3. MFC Flow 3. Flow Standard 3. NOTES: As found with old box. Unable to solve flow issue onsite. At one point I had only the pump connected, all else disconnected, and the flow still could not be adjusted. Rebuilt pump and swapped the box. The 'as left' was with the new box, it was used to get new logger values. They are on the 2nd excel sheet.

13 CASTNET FLOW SYSTEM VERIFICATION & CALIBRATION ABBR. BIBE-KB CLIENT National Park Service FIELD SPECIALIST Jake Zaragoza DATE 2/21/217 SITE NAME Big Bend National Park - K-Bar Ranch Road DATE OF LAST VISIT 7/12/216 MANUFACTURER MODEL SERIAL NUMBER EXPIRATION DATE MFC High Flow Reference Bios Definer 22-H /9/217 MFC MFC Manufacturer Alicat Manufacturer Alicat MC-1SLPM-D-PCV65 MC-1SLPM-D-PCV65 Serial Number Serial Number Pump Pump Manufacturer Thomas Manufacturer Thomas 17CA18 17CA18 ENTER Serial Number Serial Number FLOWS IN Flow Target 3. SLPM MFC Display Linear MFC VDC Linear Regression Regression ESC Logger Scaling Date 2/21/217 High Input (V) 5. Setpoint 3. Setpoint Low Input (V). Slope Slope High Output (LPM) 5.85 Intercept Intercept Low Output (LPM) -.4 Operational Checks Vacuum 2. DAS Flow 3. MFC Flow 3. Flow Standard 3. Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Difference Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % 3 (target) 3. SLPM % 4 (high) 3.5 SLPM % 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM). *Note: A leak is present if the difference between the zero and leak test value is greater than.1 lpm Recalibration Not Required *Note: A calibration is required if the difference between the transfer standard and the DAS value is greater than +/- 2.% at the target setting or greater than +/- 2.5% at either the low or high point NEW CALIBRATION FACTORS MFC Display Linear MFC VDC Linear Regression Regression ESC Logger Scaling Setpoint 3.6 Setpoint 3.5 High Input (V) 5. Slope 1.4 Slope 1.3 Low Input (V). Intercept -.17 Intercept -.13 High Output (LPM) 5.1 Correlation 1. Correlation 1. Low Output (LPM) -.13 Target Flow Standard Rotometer MFC Display MFC Volts DC DAS Display Difference Value % Pump Off. SLPM Leak Test. SLPM SLPM % 2 (low) 2.5 SLPM % 3 (target) 3. SLPM % 4 (high) 3.5 SLPM % 5 4. SLPM % MAX > 4. SLPM % Leak Test Results Flow (SLPM).1 *Note: A leak is present if the difference between the zero and leak test value is greater than.1 lpm Operational Checks Vacuum DAS Flow 3. MFC Flow 3. Flow Standard 3. NOTES: The 'as left' from the previous sheet was used as the 'as found' here to get new logger values. The 'as found' logger values were listed on the new box.

14 OZONE ANALYZER VERIFICATION & CALIBRATION () ABBR. BIBE-KB CLIENT National Park Service FIELD SPECIALIST Jake Zaragoza DATE 2/21/217 SITE NAME Big Bend National Park - K-Bar Ranch Road DATE OF LAST VISIT 7/12/216 Logger Slope Correction.9952 Logger Intercept Correction CALIBRATION ACCEPTANCE CRITERIA (<=) TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE Mean Absolute Difference (%) 3% Manufacturer Thermo Manufacturer Thermo Manufacturer Thermo Maximum Absolute Difference (%) 3% Linearity % of Fullscale 1% 49i 49c 49c DATA ACCEPTANCE CRITERIA (<=) Serial Number Serial Number Serial Number Mean Absolute Difference (%) 1% Coefficient 1. Coefficient 1.12 Coefficient 1. Maximum Absolute Difference (%) 1% Background. Background.2 Background. Linearity % of Fullscale 2% Cell A Intensity (Hz) 9326 Cell A Intensity (Hz) 7973 Cell A Intensity (Hz) SLOPE INT Cell B Intensity (Hz) Cell B Intensity (Hz) 6396 Cell B Intensity (Hz) 8355 Transfer Standard Correction Factors Flow A (lpm).789 Flow A (lpm).658 Flow A (lpm).664 Flow B (lpm).815 Flow B (lpm).682 Flow B (lpm).673 Full Scale (ppb) 25 Pressure (mmhg) Pressure (mmhg) Pressure (mmhg) 69.9 Linear Regression Plot Bench Temp ( C) 34.3 Bench Temp ( C) 37.2 Bench Temp ( C) 38.2 Bench Lamp Temp ( C) 54.2 Bench Lamp Temp ( C) 56.4 Bench Lamp Temp ( C) 56.7 Ozone Lamp Temp ( C) 68.2 Ozone Lamp Temp ( C) 71.2 Ozone Lamp Temp ( C) NA Ozone Source Yes Ozone Source No Transfer Standard Level LEVEL 2 Transfer Standard Level LEVEL 3 O3 LAMP TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE POINT TARGET % or mv Display Corrected DAS Diff %Diff Linearity % DAS Diff %Diff Linearity % ZERO N/A.1%. -.2 N/A.% %.5% % -.3% % -.2% %.9% % -.8% % -1.% ACTION %.% %.1% %.3% %.3% ZERO N/A.1%. -.3 N/A.% Mean ABS % Diff 1.1% Mean ABS % Diff 1.3% Max ABS % Difference 2.7% Max ABS % Difference 3.5% ACTION Analog Test Analyzer DAS Reference DAS Slope 1.1 Slope 1.16 Zero Full Scale Line Loss Test Analyzer Display Span (site tubing) 26 Span (bypass tubing) % P/F -.5% Y-Intercept -.6 Y-Intercept -.26 Correlation.9999 Correlation.9999 Residuals Plot y = 1.94x +.89 Transfer Standard (ppb) R² =.9998 NOTES: Replaced zero air media. Cleaned fans and L3 cells before analog test and line loss test. Cleaned analyzer cells post analog and line loss test. Adjusted analog out for both L3 and analyzer. Increased lamp setting on analyzer to 49%. Increased lamp setting to 5.3% on L3. Lamps for both instruments may need replacing during next SAM. Updated logger correction factors. Calibrated analyzer. Residual (ppb) Analyzer DAS (ppb) Regression Concentration Value (ppb)

15 ABBR. BIBE-KB CLIENT National Park Service FIELD SPECIALIST Jake Zaragoza DATE 2/21/217 SITE NAME Big Bend National Park - K-Bar Ranch Road DATE OF LAST VISIT 7/12/216 Logger Intercept Correction CALIBRATION ACCEPTANCE CRITERIA (<=) TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE Mean Absolute Difference (%) 3% Manufacturer Thermo Manufacturer Thermo Manufacturer Thermo Maximum Absolute Difference (%) 3% Linearity % of Fullscale 1% 49i 49c 49c DATA ACCEPTANCE CRITERIA (<=) Serial Number Serial Number Serial Number Mean Absolute Difference (%) 1% Coefficient 1. Coefficient 1.26 Coefficient 1. Maximum Absolute Difference (%) 1% Background. Background.2 Background. Linearity % of Fullscale 2% Cell A Intensity (Hz) 9315 Cell A Intensity (Hz) Cell A Intensity (Hz) 172 Cell B Intensity (Hz) Cell B Intensity (Hz) 9476 Cell B Intensity (Hz) Transfer Standard Correction Factors.9867 Flow A (lpm).789 Flow A (lpm).658 Flow A (lpm).664 Flow B (lpm).813 Flow B (lpm).688 Flow B (lpm).677 Full Scale (ppb) 5 Pressure (mmhg) Pressure (mmhg) Pressure (mmhg) 69.9 Linear Regression Plot Bench Temp ( C) 32.1 Bench Temp ( C) 32.7 Bench Temp ( C) 33.9 Bench Lamp Temp ( C) 54.1 Bench Lamp Temp ( C) 56.1 Bench Lamp Temp ( C) 56.7 Ozone Lamp Temp ( C) 68.2 Ozone Lamp Temp ( C) 71 Ozone Lamp Temp ( C) NA Transfer Standard Level LEVEL 2 Transfer Standard Level POINT TARGET Zero Full Scale 1 OZONE ANALYZER VERIFICATION & CALIBRATION () Ozone Source Yes Ozone Source Logger Slope Correction.9964 No LEVEL 3 O3 LAMP TRANSFER STANDARD AMBIENT ANALYZER STATION REFERENCE % or mv Display Corrected DAS Diff %Diff Linearity % DAS Diff %Diff Linearity % ZERO N/A.%. -.3 N/A.1% % -.1% %.% %.% %.1% %.1% %.% 4 1 NA %.% % -.1% %.% % -.2% ZERO N/A.%. -.4 N/A.1% Mean ABS % Diff 1.9% Mean ABS % Diff.5% Max ABS % Difference 2.5% Max ABS % Difference 1.6% Analog Test Analyzer DAS Reference DAS Slope 1.26 Slope.999 Y-Intercept -.53 Y-Intercept Correlation 1. Correlation Residuals Plot Residual (ppb) Analyzer DAS (ppb) SLOPE INT y =.981x Transfer Standard (ppb) R² = Regression Concentration Value (ppb) NOTES: Forgot to write down lamp level for 4th point.

16 CALIBRATION AND VERIFICATION STANDARDS ABBR. BIBE-KB CLIENT SITE NAME National Park Service FIELD SPECIALIST Big Bend National Park - K-Bar Ranch Road Jake Zaragoza DATE DATE OF LAST VISIT 2/21/217 7/12/216 MANUFACTURER MODEL SERIAL # Calibration Expiration Date Ozone Transfer Standard Thermo 49i /2/218 MFC High Flow Reference Bios Definer 22-H /9/217 MFC Low Flow Reference Temperature Reference Eutechnics /3/218 AT/RH Sensor Reference Rotronic HG2S /5/217 Barometric Pressure Reference Wind Speed Reference (high rpm) RM Young 1882A/1883A CA2241 1/26/218 Wind Speed Reference (low rpm) Wind Speed Torque Gauge RM Young 1831 NA Wind Direction Alignment Reference Brunton Transit Wind Direction Linearity Reference ARS Jig 45 Degree Disc NA Wind Direction Torque Gauge RM Young 1831 NA Solar Radiation Reference Multiplier 2. LiCor Li-2X-L PY /17/217 UV Radiation Reference Multiplier Precipitation Reference Volume 945 ml Novalynx Drip Bottle NA Voltage Measurement Reference Voltage Source PM Flow Standard #1 PM Flow Standard #2 PM Flow Standard #3 PM Flow Standard #4 PM Temperature Standard #1 PM Temperature Standard #2 PM Temperature Standard #3 PM Temperature Standard #4 PM Barometric Pressure Standard #1 PM Barometric Pressure Standard #2 PM Barometric Pressure Standard #3 PM Barometric Pressure Standard #4 TEOM MTV Standard HiVol Direct Flow Standard HiVol Orifice Plate Orifice Manometer Stagnation Manometer

17 -5 ppb scale

18 -1 ppb scale

19

20

21

22

23

24

25

26

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Will Yahr Site: Big Bend NP - (BIBE-KB) Service Date(s): 7/26/217

More information

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Mike Slate Site: Canyonlands National Park (CANY-IS) Service Date(s):

More information

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc.

Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Semiannual Maintenance and Calibration Report Prepared by Air Resource Specialists, Inc. Client: National Park Service Field Personnel: Mike Farinacci Site: Island in the Sky (CANY-IS) Service Date(s):

More information

MP 601A Humidity Probe INSTRUCTION MANUAL

MP 601A Humidity Probe INSTRUCTION MANUAL MP 601A Humidity Probe INSTRUCTION MANUAL CONTENTS DESCRIPTION... 3 OPERATION... 3 WIRING... 5 INSTALLATION... 7 MAINTENANCE...7 CALIBRATION BASICS... 8 CALIBRATION PROCEDURE... 9 SPECIFICATIONS... 11

More information

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection Procedure No: SOP-020 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, and to

More information

HOBO U30 Remote Monitoring Systems

HOBO U30 Remote Monitoring Systems onset HOBO U30 Remote Monitoring Systems Web-based environmental monitoring All electronics housed within a research-grade, tamperproof enclosure Setup is quick and easy with plug-and-play sensors GSM

More information

I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B. For Vantage Pro and Vantage Pro Plus

I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B. For Vantage Pro and Vantage Pro Plus I NTEGRATED S ENSOR S UITE WITH F AN-ASPIRATED R ADIATION SHIELD ADDENDUM, REV B For Vantage Pro and Vantage Pro Plus The Vantage Pro Integrated Sensor Suite (ISS) with the Fan-Aspirated Radiation Shield

More information

CMT MANUAL 6. NMC SENSORS

CMT MANUAL 6. NMC SENSORS CMT MANUAL 6. NMC SENSORS TABLE OF CONTENTS INDEMNITY... 5 DOCUMENT INFORMATION... 5 ACKNOWLEDGEMENTS... 5 6.1 INTRODUCTION... 6 6.2 TEMPERATURE AND HUMIDITY SENSORS... 7 6.2.1. TEMPERATURE SENSOR... 7

More information

Air Resource Specialists, Inc.

Air Resource Specialists, Inc. Air Resource Specialists, Inc. Optec NGN-2A Nephelometer Support Systems Price List July 2006 FULLY INTEGRATED OPTEC NGN-2A MONITORING AND DATA LOGGING SUPPORT SYSTEMS (costs do not include the Optec NGN-2A

More information

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF 2015 SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF FINAL APPROVAL REVISION #: 0 ISSUE DATE: December 4, 2015 PAGE: 1 of 13 This Methods

More information

Model OEM-2 INSTRUCTIONS FOR USE

Model OEM-2 INSTRUCTIONS FOR USE ECO SENSORS, INC. 3-03.2 OEM OZONE CONTROLLER Model OEM-2 INSTRUCTIONS FOR USE GENERAL The model OEM-2 is a system to control ozone generators and alarms based on an adjustable ozone concentration set

More information

UV Sentry. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific Fugitive Emissions Quantification

UV Sentry. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific Fugitive Emissions Quantification UV Sentry Open Path Multi-gas UVDOAS Analyzer Ammonia Benzene Carbon Disulfide Chlorine Cost Effective Perimeter & LDAR Monitoring The UV Sentry offers perimeter monitoring with among the lowest cost per

More information

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu)

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) PART 1 - GENERAL A. RELATED DOCUMENTS SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) Federal, State, and local requirements as applicable. Attached Permit Attached drawings B. SUMMARY 1. Provide

More information

MODEL GFC7000EU CO 2 ANALYZER

MODEL GFC7000EU CO 2 ANALYZER MODEL GFC7000EU CO 2 ANALYZER Teledyne Analytical Instruments 16830 Chestnut St. City of Industry, Ca. 91748 TEL: 626-961-9221 FAX: 626-961-2538 TEL: 626-934-1500 E-MAIL: ai-customerservice@teledyne.com

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE OPTEC NGN-2 NEPHELOMETER AUDIT PROCEDURES (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4700 DATE JANUARY 1995 AUTHORIZATIONS

More information

Instruction Manual for FORMALDEHYDE GAS DETECTOR MODEL FP-30/FP-40

Instruction Manual for FORMALDEHYDE GAS DETECTOR MODEL FP-30/FP-40 Instruction Manual for FORMALDEHYDE GAS DETECTOR MODEL FP-30/FP-40 This instrument is sold and serviced by RKI Instruments, Union City, CA For questions, parts, or service please contact RKI at (800) 754-5165.

More information

Model 6650SP. Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer

Model 6650SP. Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer Model 6650SP Filber Optic-Based UV-Fluorescence Oil-in-Water Analyzer 6650SP - OVERVIEW The Model 6650SP is the next generation in-water Analyzer. The 6650SP incorporates the most advanced, state-of-the-art

More information

FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST

FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST AHRI CERTIFICATION PROGRAM FOR AIR-COOLED WATER-CHILLING PACKAGES USING THE VAPOR COMPRESSION CYCLE ACCL Participant: Date of Test: Witness Test

More information

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications LI-830 and LI-850 Gas Analyzers For Continuous Monitoring Applications Easy Operation The LI-830 CO2 and LI-850 CO2/H2O Analyzers are high-performance monitoring solutions that give accurate, stable readings.

More information

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications

LI-830 and LI-850 Gas Analyzers. For Continuous Monitoring Applications LI-830 and LI-850 Gas Analyzers For Continuous Monitoring Applications Easy Operation The LI-830 CO2 and LI-850 CO2/H2O Analyzers are high-performance monitoring solutions that give accurate, stable readings.

More information

Integrating Nephelometer - Visibility and Particulate Monitoring. Summary

Integrating Nephelometer - Visibility and Particulate Monitoring. Summary Summary This paper investigate the correlation between the Aurora 1000 integrating Nephelometer single wavelength green source 525 nm and the Thermo Fisher 1400A/B TEOM, Met-One 1020 BAM, and FRM sampler.

More information

SENTRY-MS Mono-static Multi-gas Open Path Analyzer

SENTRY-MS Mono-static Multi-gas Open Path Analyzer SENTRY-MS Mono-static Multi-gas Open Path Analyzer Air Monitoring Where You Need it... The Cerex Sentry-MS Multi-gas Open Path Analyzer is specifically designed for rapid deployment for a variety of indoor

More information

metro B6012 VIBRATIONS SIMPLIFIED... Product brochure Easiest portable vibration analyzer and balancer TECHNOLOGIES PVT. LTD.

metro B6012 VIBRATIONS SIMPLIFIED... Product brochure Easiest portable vibration analyzer and balancer TECHNOLOGIES PVT. LTD. metro B6012 VIBRATIONS SIMPLIFIED... Product brochure Easiest portable vibration analyzer and balancer TECHNOLOGIES PVT. LTD. Product brocure 1 is reliable, fast and easiest to use tool designed for routine

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) STANDARD OPERATING PROCEDURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE OPTICAL DATA REDUCTION AND VALIDATION (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4400 DATE NOVEMBER 2004 AUTHORIZATIONS TITLE

More information

Using the Model 1500 Portable Chilled Mirror Hygrometer

Using the Model 1500 Portable Chilled Mirror Hygrometer APPLICATION NOTE 2016_0120 MODEL 1500 Rev C Using the Model 1500 Portable Chilled Mirror Hygrometer The Model 1500 is a portable, multi-function, optical chilled mirror hygrometer designed to accurately

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES AUTHORIZATIONS TITLE NAME SIGNATURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES AUTHORIZATIONS TITLE NAME SIGNATURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE SITE SELECTION FOR OPTICAL MONITORING EQUIPMENT (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4050 DATE FEBRUARY 1994 AUTHORIZATIONS

More information

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009 STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009 Introduction Honeywell s STT 3000 family of microprocessor based transmitters covers the full

More information

C-NRPP Quality Assurance Guidance for Radon Test Devices

C-NRPP Quality Assurance Guidance for Radon Test Devices C-NRPP Quality Assurance Guidance for Radon Test Devices Table of Contents: 1. Purpose 2. Definitions 3. Quality Assurance Plan 4. Calibration 5. Spikes/Performance Tests 6. Duplicates 7. Blanks 8. Cross-Checks

More information

RPM1600 Series Room Pressure Monitors

RPM1600 Series Room Pressure Monitors RPM1600 Series Room Pressure Monitors Technical Bulletin LB-RPM1611-0, LB--0 Code No. LIT-12012228 Issued October 2017 Refer to the QuickLIT website for the most up-to-date version of this document. How

More information

Global Water Instrumentation, Inc.

Global Water Instrumentation, Inc. Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 8 Tipping Bucket RG600 02/05/10 00-714 - 1 - Congratulations on your purchase

More information

Smoke Particulate Monitors: 2006 Update

Smoke Particulate Monitors: 2006 Update United States Department of Agriculture Forest Service Technology & Development Program Smoke Particulate Monitors: 2006 Update 2500 Watershed, Soil, and Air December 2006 0625 2842 MTDC Smoke Particulate

More information

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements Specifications COMMANDER 150 Universal Process Indicator High visibility 6-digit LED display the clearest view of your process status 0.1% measurement accuracy precise indication of process measurement

More information

Integrated Sensor Suite with Fan-Aspirated Radiation Shield

Integrated Sensor Suite with Fan-Aspirated Radiation Shield Integrated Sensor Suite with Fan-Aspirated Radiation Shield For Vantage Pro2 and Vantage Pro2 Plus Installation Instructions Addendum The Vantage Pro2 Integrated Sensor Suite (ISS) with the Fan-Aspirated

More information

Three categories of instruments

Three categories of instruments Instrument Descriptions Part 2 of 2 Three categories of instruments are discussed in this report: optical real-time, continuous particulate monitors, mass quantifying real-time continuous particulate monitors,

More information

Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification

Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification Application of a Reagentless Direct Conductivity Online TOC Analyzer for Clean-in-Place Verification By Paul Farrell, Emek Blair (Hach Company) and Torben Andersen (McFlusion) ABSTRACT This paper describes

More information

ANEMOMETER 4403-RF-BAT

ANEMOMETER 4403-RF-BAT ANEMO 4403 Radio impulsive output Anemometer - Wireless Robust wind speed sensor,extremely resistant and flexible, Suitable for many different application. Stainless steel bearings and high quality technical

More information

SITE REPORT FOR CARE (CANADA) (Submitted by Rodica Nitu) Summary and purpose of document

SITE REPORT FOR CARE (CANADA) (Submitted by Rodica Nitu) Summary and purpose of document WORLD METEOROLOGICAL ORGANIZATION COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION INTERNATIONAL ORGANIZING COMMITTEE (IOC) FOR THE WMO SOLID PRECIPITATION INTERCOMPARISON EXPERIMENT (SPICE) Fourth

More information

LIQUIDATOR Electronic Ionization System

LIQUIDATOR Electronic Ionization System LIQUIDATOR Electronic Ionization System Operation and Installation Manual LIQUITECH LIQUIDATOR Series Electronic Controller Unit LIQUITECH Model QLTF Flow Cell LiquiTech, Inc. 421 Eisenhower Lane South

More information

Ammonia Leak Detector Manual

Ammonia Leak Detector Manual Ammonia Leak Detector Manual Document No. 70-PHW-1023-R2.5 No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical,

More information

ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF

ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF ENVIRONMENTAL MONITORING IN PREPARATION FOR THE INSTALLATION OF A GREEN ROOF Andrew Martin School of Civil Engineering, Purdue University, West Lafayette, Indiana ABSTRACT Green roofs are becoming an increasingly

More information

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 SAE Alternate Refrigerant Cooperative Research Program Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000 To: Alternate Refrigerant Task Force Members From: Ward Atkinson

More information

PRODUCT CONFORMITY CERTIFICATE

PRODUCT CONFORMITY CERTIFICATE PRODUCT CONFORMITY CERTIFICATE This is to certify that the GC 866 PID airtoxic (Model A73022) manufactured by: Chromatotec / airmotec 15, Rue d Artiguelongue Saint -Antoine 33240 Val de Virvée France has

More information

SimpleVIS Viscometer Reference Guide P

SimpleVIS Viscometer Reference Guide P SimpleVIS Viscometer Reference Guide P82.0090 Run Status Indicator Lamps Sample/Solvent Port Waste Bottle Dust Cover Temperature Calibration Port Touch Screen Display RS-232 Serial Port Exhaust Muffler

More information

Model 1062 Environment Monitor

Model 1062 Environment Monitor Model 1062 Environment Monitor Continuously Monitors Environment Temperature Humidity Barometric Pressure Line Voltage Line Frequency Applications Manufacturing The 1062A Environment Monitor will ensure

More information

DF-340E PROCESS ANALYZERS APPLICATIONS FEATURES

DF-340E PROCESS ANALYZERS APPLICATIONS FEATURES PROCESS ANALYZERS DF-340E The DF-340E is a Coulometric sensor based oxygen analyzer suitable for outdoor installation and designed to measure trace oxygen and percent level in pure and multigas backgrounds.

More information

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors

Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors Calculated Energy Savings for Disabling Anti- Sweat Door Heat on Glass Display Case Doors ET Project Number: Product Manager: Sachin Andhare Pacific Gas and Electric Company` Prepared By: Danielle Geers

More information

S304 and S305 Dust Emission Monitors. User Manual. Distributor

S304 and S305 Dust Emission Monitors. User Manual. Distributor S304 and S305 Dust Emission Monitors User Manual Distributor Version 6.4 30/5/2012 Table of Contents 1. INTRODUCTION... 3 1.1 Safety... 3 1.2 Product overview... 4 1.3 Principle of operation... 4 2. INSTALLATION...

More information

THEORY OR OPERATION 2 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 OPTIONS 11

THEORY OR OPERATION 2 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 OPTIONS 11 Table of Contents INTRODUCTION 1 THEORY OR OPERATION 2 DESCRIPTION 5 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 SPECIFICATIONS 10 OPTIONS

More information

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications: Industrial Chilled Mirror Hygrometer Integral frost/dew point measuring head Internal sample pump Optimal Response Injection System ForceFrost TM function Ice-Test user calibration verification User configurable

More information

The Participant will select one of the two Approval Categories (below) and submit the appropriate documentation.

The Participant will select one of the two Approval Categories (below) and submit the appropriate documentation. ACCL Participant: Date of Inspection: Test Stand Inspection Report Number: Approval Results: FORM ACCL-PC4 ACCL FACILITY APPROVAL PROCEDURE AND CHECKLIST AHRI CERTIFICATION PROGRAM FOR AIR-COOLED (ACCL)

More information

TOOLS AND INSTRUMENTS

TOOLS AND INSTRUMENTS TOOLS AND INSTRUMENTS MANNIX INSTRUMENTS DELUXE DIGITAL POCKET THERMOMETER DIGITAL ANEMOMETER F/C Switch Protective Cover Auto Power Off ON/OFF Switch Data Hold Simultaneous Display of Air Velocity and

More information

Table of Contents. Page Quick Set-up Instructions 1. Theory of Operation 2. Specifications 3. TM 1003W Controls and Connections 4

Table of Contents. Page Quick Set-up Instructions 1. Theory of Operation 2. Specifications 3. TM 1003W Controls and Connections 4 Table of Contents Page Quick Set-up Instructions 1 Theory of Operation 2 Specifications 3 TM 1003W Controls and Connections 4 1. Power Source Connector 5 2. Alarm Output Connector 5 3. Variation Detection

More information

Revision: None DCN No. First Issue Date: November 30, 1998 Product: Standard DCT, 120-volts

Revision: None DCN No. First Issue Date: November 30, 1998 Product: Standard DCT, 120-volts Figure 1Document: SPEC-112 Revision: None DCN No. First Issue Date: November 30, 1998 Product: Standard DCT, 120-volts construction requirements of UL approved Class H (180 C) insulation system. Transformers

More information

Model SC3004. Transducer Display and Signal Conditioning Unit DESCRIPTION. Fourteen channel chassis FEATURES

Model SC3004. Transducer Display and Signal Conditioning Unit DESCRIPTION. Fourteen channel chassis FEATURES Transducer Display and Signal Conditioning Unit DESCRIPTION The SC series models are self-calibrating microprocessor-based signal conditioners. Indicators are available with several different types of

More information

Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20

Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20 Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20 PRODUCT DATA FEATURES Solid or split core loop-powered current transmitters Fast response time Integral DIN rail mounting flange

More information

Ventostat Wall Mount. Amphenol Advanced Sensors. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. Features:

Ventostat Wall Mount. Amphenol Advanced Sensors. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. Features: Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in a compact low cost package. Patented

More information

Pump-Down Controller MODEL 4052

Pump-Down Controller MODEL 4052 Pump-Down Controller 4-20mA Input/Scalable Output Seal Fail Monitoring Duplex Pump Alternation Hand-Off-Auto Controls Dual Run-time Meters RS-485/Modbus Communications DESCRIPTION The Model 4052 Pump-Down

More information

200 SERIES ADVANCED PHOTO/THERMAL/IR DETECTOR PRODUCT SPECIFICATION

200 SERIES ADVANCED PHOTO/THERMAL/IR DETECTOR PRODUCT SPECIFICATION GENERAL DESCRIPTION This product is the photo/thermal/ir variant of 2 Series Advanced, the redesigned 2 Series all SMD adopting the new chamber and a new style of covers. It works with the 2 Series extended

More information

HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL

HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL HTO-45 HUMIDITY TEMPERATURE PROBE INSTRUCTION MANUAL -2- CONTENTS DESCRIPTION...3 OPERATION...3 INSTALLATION (Tips for fixed probe installation)...4 MAINTENANCE...6 CALIBRATION...6 SPECIFICATIONS...8 PLEASE,

More information

19 mm Series Low Cost, Stainless Steel, Isolated Pressure Sensors

19 mm Series Low Cost, Stainless Steel, Isolated Pressure Sensors 19 mm Series Low Cost, Stainless Steel, Isolated Pressure Sensors DESCRIPTION Honeywell s stainless steel 19C, 19U, and 19 Vacuum Gage Series sensors were developed for pressure applications that involve

More information

BUREAU OF METEOROLOGY

BUREAU OF METEOROLOGY BUREAU OF METEOROLOGY DEPARTMENT OF THE ENVIRONMENT AND HERITAGE INSTRUMENT TEST REPORT 2002_664 Evaluation of Desiccants Page 1 of 7 1. AIM This technical note sets out the process and methodology employed

More information

Series Temperature Controller Instruction Sheet

Series Temperature Controller Instruction Sheet Series Temperature Controller Instruction Sheet Thank you very much for purchasing DELTA A Series. Please read this instruction sheet before using your A series to ensure proper operation and please keep

More information

Measurement Solutions for Refrigeration Technology.

Measurement Solutions for Refrigeration Technology. Measurement Solutions for Refrigeration Technology. Digital manifold, leak detection, vacuum measurement, and lots more. 01_Brochure_testo552_MASTER.indd 1 30.01.2013 11:27:13 Testo refrigeration technology

More information

KANE-ALERT-CO 2 Monitor with Temperature

KANE-ALERT-CO 2 Monitor with Temperature KANE-ALERT-CO 2 Monitor with Temperature Operating Instructions Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel: +44 (0) 1707 375550 Fax: +44 (0) 1707 393277

More information

Temperature Controllers

Temperature Controllers Model TEC-4100 1/4 DIN Model TEC-4100 1/4 DIN Temperature Controller Ordering Code: Power Input BOX 1 4 = 90-250 VAC 5 = 11-26 VAC / VDC TEC-4100- Configurable for 4 Programmable Outputs and NEMA 4X/IP65

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) FEATURES AND SPECIFICATIONS MOLYTEK 2702 AND 3702 RECORDERS 2702 3702 REV: 03302004 Standard Specifications for Model

More information

Meridian wiredhart. HART Field Device Specification Goldmine Rd. Monroe, NC USA

Meridian wiredhart. HART Field Device Specification Goldmine Rd. Monroe, NC USA HART Field Device Specification Meridian wiredhart 4320 Goldmine Rd. Monroe, NC 28110 USA HART is a registered trademark of the HART Communication Foundation TABLE OF CONTENTS 1. Introduction...5 1.1 Scope...5

More information

NOVALYNX CORPORATION MODEL 200-WS-21-A DUAL SET POINT WIND ALARM INSTRUCTION MANUAL

NOVALYNX CORPORATION MODEL 200-WS-21-A DUAL SET POINT WIND ALARM INSTRUCTION MANUAL NOVALYNX CORPORATION MODEL 200-WS-21-A DUAL SET POINT WIND ALARM INSTRUCTION MANUAL REVISION DATE: JAN 2005 Receiving and Unpacking Carefully unpack all components and compare to the packing list. Notify

More information

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support.

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support. AutomationDirect 1/16 DIN Series Technical Support We strive to make our manuals the best in the industry. We rely on your feedback to let us know if we are reaching our goal. If you cannot find the solution

More information

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories October 2018 C-NRPP/PNCR-C 1 (855) 722-6777 c-nrpp.ca Canadian

More information

2017 Company Overview

2017 Company Overview 2017 Company Overview WHO IS SERVOFLO SERVOFLO IS A STOCKING DISTRIBUTOR PROVIDING TECHNICAL AND LOGISTICAL SERVICES FOR: Pressure sensors Mass Flow sensors Humidity sensors Temperature Sensors Micropumps

More information

Audit/Twinning Report

Audit/Twinning Report Audit/Twinning Report Station name: Danum Valley Date: November 26 302017 Auditor: Dr. Thomas Tuch (WMO GAW WCCAP, TROPOS) Responsible station staff: Maznorizan Bt Mohamad, Mohan Kumar A/L Sammathuria

More information

Performance of R-22, R-407C and R-410A at Constant Cooling Capacity in a 10

Performance of R-22, R-407C and R-410A at Constant Cooling Capacity in a 10 Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2000 Performance of R-22, R-407C and R-410A at Constant Cooling Capacity in

More information

Models TX-KE and TX-KP Toxic Gas Transmitters

Models TX-KE and TX-KP Toxic Gas Transmitters Models TX-KE and TX-KP Toxic Gas Transmitters Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Dr. Lake Zurich, IL. 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 www.pureairemonitoring.com

More information

SX Series Silicon Pressure Sensors 0 psi to 1 psi to 0 psi to 150 psi

SX Series Silicon Pressure Sensors 0 psi to 1 psi to 0 psi to 150 psi SX Series Silicon Pressure Sensors 0 psi to 1 psi to 0 psi to 150 psi DESCRIPTION The SX Series pressure sensors provide the lowest cost components for measuring pressures up to 150 psi. These sensors

More information

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual Read instructions carefully before operating this device. ❶ This device is not to be used for Human Life Support applications. ❷ To avoid possible

More information

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9. Humidity sensors Behavior in different environments Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.29 CZECH TECHNICAL UNIVERSITY IN PRAGUE CERN Content Content...

More information

Monthly Operating Report (MOR) Addendum for UV: Calculated Dose Approach

Monthly Operating Report (MOR) Addendum for UV: Calculated Dose Approach PWS Name:. Monthly Operating Report (MOR) Addendum for UV: Calculated Dose Approach PWSID:. Reporting Period:. Reactor Total Flow Through (MG) of Off-Specification Events Off-Specification Data Total Off-Specification

More information

0-10 V Wall & Duct Humidistat Installation Instructions

0-10 V Wall & Duct Humidistat Installation Instructions 0-10 V Wall & Duct Humidistat Installation Instructions This document covers the operation and installation instructions for the following Nortec Digital Humidistats: Part #: Description: 1510142 0-10

More information

Module Features are-configurable, no module jumpers to set

Module Features are-configurable, no module jumpers to set December 2011 PACSystems* RX3i Isolated Thermocouple Input Module, 6 Channels, IC695ALG306 Isolated Thermocouple Input Module, 12 Channels, IC695ALG312 Isolated Thermocouple Input module IC695ALG306 provides

More information

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING 2 INSTALLATION. Premier Series

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING 2 INSTALLATION. Premier Series Premier Series MODEL 5000 USER MANUAL 5-2 Day Programmable Single Stage Heat/Cool Digital Thermostat Compatible with low voltage single stage gas, oil or electric heating or cooling systems, including

More information

Model 5190-E Temperature/Humidity Probe

Model 5190-E Temperature/Humidity Probe Model 5190-E Temperature/Humidity Probe User s Manual Rev. D All Weather Inc. 1165 National Drive Sacramento, CA 95834 USA 800.824.5873 www.allweatherinc.com Copyright 2010, All Weather, Inc. All Rights

More information

TEST REPORT #65. Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A) in a R-410A Scroll Compressor

TEST REPORT #65. Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A) in a R-410A Scroll Compressor Air-Conditioning, Heating, and Refrigeration Institute (AHRI) Low-GWP Alternative Refrigerants Evaluation Program (Low-GWP AREP) TEST REPORT #65 Compressor Calorimeter Test of Refrigerant L-41-2 (R-447A)

More information

WeatherLink for Alarm Output Addendum

WeatherLink for Alarm Output Addendum WeatherLink for Alarm Output Addendum Introduction This Streaming Data Logger is designed to provide an electrical interface between a Vantage Pro or Vantage Pro2 weather station console or Weather Envoy

More information

Toxic Gas Detection. Model F12. Toxic Gas Detection

Toxic Gas Detection. Model F12. Toxic Gas Detection Toxic Gas Detection Model F12 Toxic Gas Detection Smart Sensor Technology The Series F12 Gas Transmitter is an Intrinsically Safe (IS) version of our explosion-proof D12 transmitter. In its standard form,

More information

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES NEPHELOMETER MAINTENANCE (IMPROVE PROTOCOL) AUTHORIZATIONS TITLE NAME SIGNATURE

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES NEPHELOMETER MAINTENANCE (IMPROVE PROTOCOL) AUTHORIZATIONS TITLE NAME SIGNATURE QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES TITLE NEPHELOMETER MAINTENANCE (IMPROVE PROTOCOL) TYPE STANDARD OPERATING PROCEDURE NUMBER 4100 DATE OCTOBER 1996 AUTHORIZATIONS TITLE NAME SIGNATURE

More information

IQ/OQ Protocol Installation Qualification/ Operation Qualification

IQ/OQ Protocol Installation Qualification/ Operation Qualification IQ/OQ Protocol Installation Qualification/ Operation Qualification ZZ FreeZone Freeze Dry Systems (To be used with some FreeZone Systems manufactured prior to August, 2004- see model number chart inside)

More information

2-wire Passive Intelligent Loop (LED) Display Meter

2-wire Passive Intelligent Loop (LED) Display Meter Features Two-wire passive intelligent loop current (LED) digital meter 2-wire Passive Intelligent Loop (LED) Display Meter Embedded 2-wire Passive Programmable 4-20mA Loop (Four Digits LED) Display Meter

More information

The following report is prepared and published by:

The following report is prepared and published by: The following report is prepared and published by: Josh Dennis, Thermal Process Engineer josh.dennis@readingthermal.com of 7 CORPORATE BLVD.SINKING SPRING, PA 19608 USA T: 610-678-5890 F: 610-693-6262

More information

WATCHMAN Model LBW-WATCHMAN Ammonia Leak Detector

WATCHMAN Model LBW-WATCHMAN Ammonia Leak Detector WATCHMAN Model LBW-WATCHMAN Ammonia Leak Detector IMPORTANT READ THIS FIRST....3 CAUTIONS... 3 AVOIDING NUISANCE ALARMS... 3 STANDARD FEATURES... 4 AVAILABLE OPTIONS... 4 PARTS DESCRIPTION... 5 FRONT PANEL

More information

Section B Field Instruments

Section B Field Instruments 91-00-01-01 Section B Field Instruments Contents Product Type Literature No. Overview 30-00-02-01 34 ST 3000 Series 100e Smart Pressure s SMV 3000 Smart Multivariable Model STM125 Series 100e Models Series

More information

Pump-Up Controller MODEL 4062

Pump-Up Controller MODEL 4062 Pump-Up Controller 4-20mA Input/Scalable Output Seal Fail Monitoring Duplex Pump Alternation Hand-Off-Auto Controls Dual Run-time Meters RS-485/Modbus Communications DESCRIPTION The Model 4062 Pump-Up

More information

HIn3550 Magnetic Field Monitor User's Manual

HIn3550 Magnetic Field Monitor User's Manual HIn3550 Magnetic Field Monitor User's Manual Copyright 1993 by Holaday Industries, Inc. Manual #600053 10/97 $12.50 Revision Record Manual #600053 HIn3550 Magnetic Field Monitor Revision Description Date

More information

Free Chlorine Measuring System

Free Chlorine Measuring System Product Data Sheet PDS 71-FCL/rev.H September 2008 Model FCL Free Chlorine Measuring System COMPLETE SYSTEM INCLUDES sensor, connecting cable, analyzer, and flow controller. CONTINUOUS ph CORRECTION eliminates

More information

IMR 5000 IMR 400. CEMS Operating Manual IMR USA S/N: Sample IMR Sensors Inside IMR 400. Gas Dryer Inside. IMR Environmental Equipment Inc.

IMR 5000 IMR 400. CEMS Operating Manual IMR USA S/N: Sample IMR Sensors Inside IMR 400. Gas Dryer Inside. IMR Environmental Equipment Inc. CEMS Operating Manual IMR USA S/N: Sample IMR 5000 IMR 400 IMR 400 IMR 5000 Measure MESURE PURGE Purge CALIBRATE Calibrate OFF OFF RESET Measure ON ON Calibrate RESET PUMP Gas Dryer Inside PUMP Sensors

More information

Series Digital Controller Instruction Sheet

Series Digital Controller Instruction Sheet 216/3/11 Series Digital Controller Instruction Sheet Thank you very much for purchasing DELTA DTC Series Temperature Controller. Please read this instruction sheet before using your DTC series to ensure

More information

Infrared Carbon Dioxide Sensor Module CM1107-Dual Beam

Infrared Carbon Dioxide Sensor Module CM1107-Dual Beam Infrared Carbon Dioxide Sensor Module CM1107-Dual Beam CM1107 Specification Working temperature -10 ~+50, 5~95%RH Introduction Dual beam (single light source, dual wavelength) CM1107 CO2 gas sensor module

More information

19 mm Series. Pressure Sensors Low-Cost, Stainless Steel, Isolated Sensors WARNING WARNING. Sensing and Control

19 mm Series. Pressure Sensors Low-Cost, Stainless Steel, Isolated Sensors WARNING WARNING. Sensing and Control FEATURES Low cost Rugged, isolated stainless steel package Small size Reliable semiconductor technology Calibrated and temperature compensated Absolute and gage pressures Vacuum compatible, isolated sensors

More information

Description (Data Point Name) Unit Size

Description (Data Point Name) Unit Size AC-PRO-II Communications Modbus Register Map Document Revision 2.0 Firmware v2.0 AC-PRO-II: Registers 7000 7019: Output Coils 7020 7099: Information 7107 7199: User settings (can be set using Modbus Communications)

More information

Executive Summary 4G WiMax Radio Access Cabinet

Executive Summary 4G WiMax Radio Access Cabinet Executive Summary 4G WiMax Radio Access Cabinet From Napkin to Network 4 Development Team-Goals Partner in Design, Concept and Manufacturing of Next Gen-Radio Access Cabinet Design a Cabinet Environment

More information