DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

Size: px
Start display at page:

Download "DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific."

Transcription

1 DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Revision F Part of Thermo Fisher Scientific

2

3 DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Revision F

4

5 2011 Thermo Fisher Scientific Inc. All rights reserved. Microsoft and Windows are either trademarks or registered trademarks of Microsoft Corporation in the United States and/or other countries. HART is a registered trademark of the HART Communication Foundation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Thermo Fisher Scientific Inc. (Thermo Fisher) makes every effort to ensure the accuracy and completeness of this manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result of errors or omissions. Thermo Fisher reserves the right to make changes to the manual or improvements to the product at any time without notice. The material in this manual is proprietary and cannot be reproduced in any form without expressed written consent from Thermo Fisher.

6 This page intentionally left blank.

7 Revision History Revision Level Date Comments Initial release. A Name change. B Per ECO C Per ECO D Per ECO E Per ECO F Per ECO Thermo Fisher Scientific DensityPRO+ Installation Guide v

8 This page intentionally left blank.

9 Contents Safety Information & Guidelines... ix Safety Considerations...ix Warnings, Cautions, & Notes...ix Chapter 1 Introduction Associated Documentation Chapter 2 Chapter 3 Chapter 4 Handling & Storage ESD Procedures Unpacking, Inspection, & Storage Installation General Licensing Guidelines General Power Options The Transmitter Installing Transmitter Boards AC Power Supply The Gauge Head Mounting Configurations Mounting Instructions Dual Chain Mount Pipe Saddle (Cradle) Mount Pipe Spool Mount Wiring Preparation General Wiring Procedures For Scintillation Detectors For Ion Chamber Detectors For the Transmitter EMC Requirement for ATEX Version of 1400S Transmitters Procedure Detector Type Selection Power Supply Wiring Power to the Scintillation Detector Thermo Fisher Scientific DensityPRO+ Installation Guide vii

10 Contents DC Power AC Power Power to the Ion Chamber Detector Power to the Transmitter AC Power DC Power Detector Signal Wiring Optional Wiring Relays Current Output Contact Closure (Switch) Inputs Temperature Compensation Flow Input Serial Communications Optional RS485 Wiring RS485 Termination RS232 Wiring HART Communications Chapter 5 Appendix A Appendix B Appendix C Appendix D Getting Help Contact Information Returning Equipment for Service Ordering Information...A-1 Specifications...B-1 Drawings... C-1 Adding Ferrite Clamps to 1400S Transmitter...D-1 Procedure... D-1 Index...INDEX-1 viii DensityPRO+ Installation Guide Thermo Fisher Scientific

11 Safety Information & Guidelines This section contains information that must be read and understood by all persons installing, using, or maintaining this equipment. Safety Considerations Warnings, Cautions, & Notes Failure to follow appropriate safety procedures or inappropriate use of the equipment described in this manual can lead to equipment damage or injury to personnel. Any person working with or on the equipment described in this manual is required to evaluate all functions and operations for potential safety hazards before commencing work. Appropriate precautions must be taken as necessary to prevent potential damage to equipment or injury to personnel. The information in this manual is designed to aid personnel to correctly and safely install, operate, and/or maintain the system described; however, personnel are still responsible for considering all actions and procedures for potential hazards or conditions that may not have been anticipated in the written procedures. If a procedure cannot be performed safely, it must not be performed until appropriate actions can be taken to ensure the safety of the equipment and personnel. The procedures in this manual are not designed to replace or supersede required or common sense safety practices. All safety warnings listed in any documentation applicable to equipment and parts used in or with the system described in this manual must be read and understood prior to working on or with any part of the system. Failure to correctly perform the instructions and procedures in this manual or other documents pertaining to this system can result in equipment malfunction, equipment damage, and/or injury to personnel. Warnings, cautions, and notes are used throughout this manual to alert users to potential hazards or important information. Failure to heed the warnings and cautions in this manual can lead to injury or equipment damage. Warning Warnings notify users of procedures, practices, conditions, etc. which may result in injury or death if not carefully observed or followed. The triangular icon displayed with a warning depends on the type of hazard (general, electrical, radiation). Thermo Fisher Scientific DensityPRO+ Installation Guide ix

12 Safety Information & Guidelines Warnings, Cautions, & Notes Caution Cautions notify users of operating procedures, practices, conditions, etc. which may result in equipment damage if not carefully observed or followed. Note Notes emphasize important or essential information or a statement of company policy regarding an operating procedure, practice, condition, etc. Tip Tips may also be used in this manual. They are suggestions or things to consider that will help you use the instrument or this manual. x DensityPRO+ Installation Guide Thermo Fisher Scientific

13 Chapter 1 Introduction Refer to the Density PRO+ User Guide (p/n ) for a description of instrument functions and features, ordering information, and product specifications. Associated Documentation In addition to this guide, the following documents must be read and understood by all persons installing, using, or maintaining this equipment: DensityPRO+ User Guide (p/n ) Gamma Radiation Safety (p/n ) Thermo Fisher Scientific DensityPRO+ Installation Guide 1-1

14 This page intentionally left blank.

15 Chapter 2 Handling & Storage This chapter addresses procedures for handling electrostatic discharge (ESD) sensitive equipment, as well as procedures for unpacking, inspecting, and storing of the system. Caution This system is an ESD sensitive instrument. Use proper ESD protective equipment and procedures. Failure to comply with ESD procedures can result in circuit damage. ESD Procedures The instrument contains electronic components that can be damaged from discharges of static electricity. Ordinarily, handling the circuit boards by their edges will not damage the circuits. Caution Do not touch the circuit board components. Observe the following when installing, setting up, servicing, troubleshooting, or repairing the instrument: 1. Use an antistatic bag. Most instrument subassemblies are shipped in a special antistatic bag. When not installed, keep the assembly in the bag as often as possible. 2. Remove ESD sensitive subassemblies only under the following conditions: a. When at a designated static-free workstation or when the bag is grounded at a field site. b. After the conductive area of the container has been neutralized. c. After making firm contact with an antistatic mat and / or firmly gripping a grounded individual. 3. Personnel handling ESD sensitive devices should be neutralized to a static-free workstation by means of a grounding wrist strap that is connected to the station or to a good grounding point at the field site. Thermo Fisher Scientific DensityPRO+ Installation Guide 2-1

16 Handling & Storage Unpacking, Inspection, & Storage 4. Do not allow clothing to make contact with ESD sensitive devices. 5. Avoid touching edge connectors and components. 6. Avoid partially connecting ESD sensitive devices. These devices, especially the power supply connector, can be damaged by floating leads. 7. Ground test equipment. 8. Avoid static charges during troubleshooting. Unpacking, Inspection, & Storage All personnel involved in the packing, shipping, or receiving of hazardous material must be trained in accordance with the United States Department of Transportation (DOT) and OSHA hazardous materials regulations or in accordance with the Canadian Nuclear Safety Commission (CNSC) regulations. Note Inspection, adjustment, installation, and maintenance of the instrument must be performed by experienced personnel only. 1. Upon receipt, inspect the instrument for damage that may have occurred while in transit. If there is evidence of rough handling or damage, file a damage claim with the transportation company immediately. Notify Thermo Fisher and / or your sales representative as soon as possible. 2. Carefully inspect the packing material prior to discarding it to ensure that all equipment and instruction paperwork has been removed. 3. Use the original packing material and container for storage if necessary. 4. If storing the instrument, the storage environment should be protected, free from extremes of temperatures and high humidity, and fall within the environmental constraints listed in the specification appendix. 2-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

17 Chapter 3 Installation Read Gamma Radiation Safety (p/n ) prior to installing the equipment. Tip Copies of the drawings referenced in this manual are provided in Appendix C. General The Thermo Scientific DensityPRO+ integrated gamma density system consists of three components: 1. One of the following transmitter models: Thermo Scientific 1400A transmitter with non-metallic enclosure (NEMA 4X) Thermo Scientific 1400A transmitter with cast aluminum enclosure (NEMA 7/ NEMA 4) Thermo Scientific 1400S transmitter with stainless steel enclosure (NEMA 4X/IP65) 2. One of the following detector models: Thermo Scientific 9701 or 9702 ion chamber detector Thermo Scientific 9719B or 9720B scintillation detector 3. One of the following source head models: Thermo Scientific 5190, 5191, 5176, 5200, 5201, 5202, 5203, 5204, or 5211 Note In this manual, the generic use of transmitter or transmitters implies the 1400A and 1400S transmitters. The specific model name will be used only when necessary. Note The combination of the detector and the source head is referred to as the gauge head. Thermo Fisher Scientific DensityPRO+ Installation Guide 3-1

18 Installation Licensing Licensing Warning The instrument is a nuclear device regulated by federal and / or state authorities. You are responsible for knowing and following the pertinent safety and regulatory requirements. Refer to Gamma Radiation Safety (p/n ) for a summary of these requirements. Warning Moving or removing an installed source housing or any assembly that includes a source housing requires a person who is specifically licensed to install and commission Thermo Scientific source heads. Guidelines In the United States, your general license permits you to own and install all of the instrument s components, including the source head. However, you may not commission the instrument (remove the lock and open the source housing shutter for the first time) without a specific license authorizing radiation commissioning of the instrument. In Canada, you are only allowed to remove the instrument from the shipping container if your CNSC license has a condition authorizing mounting / dismounting of devices. For assistance obtaining a license or commissioning / decommissioning the instrument, contact Thermo Fisher. Warning Do not install the system in any hazardous area other than those approved. Refer to the equipment tag for the specific approvals applicable to the configuration of your instrument. Warning Do not apply power to the instrument in any hazardous area unless the safety ground is properly wired inside the instrument and the cover is properly installed. General Review the following guidelines when planning gauge installation. Also refer to the installation layout drawing in the drawing appendix. 1. Correct power source is available. See Power Options later in this chapter. 2. Operating temperature range: a. Transmitter: -40 C to +60 C (-40 F to +140 F) b. Scintillation or ion chamber detector: -30 C to +60 C (-20 F to +140 F) 3-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

19 Installation Guidelines 3. The transmitter should be mounted so that the display is easily seen and the keypad on the transmitter (if applicable) is easily accessible. 4. The transmitter should not be mounted near high voltage, high current, corrosive vapors, corrosive liquids, or corrosive esters. 5. There should be enough clearance to install and service the gauge head. Refer to the appropriate drawings for your type of gauge and mount (Appendix C). 6. The source housing should be positioned so that the radioactive source identification tag is visible. The source housing tag should be upright. 7. The gauge should not be mounted where process overflow or other material can collect in the beam path. The source shutter mechanism must be kept free of debris. 8. The maximum cable length between the transmitter and the detector varies with the gauge of the wire, but should not exceed 5000 feet (1525 m) for an ion chamber detector or 1000 feet (300 m) for a scintillation detector. Power Options 1400A/1400S transmitters 115/230 Vac ( Vac), 50/60 Hz, 12 VA 24 Vdc (20 28 Vdc), 12 VA 9701/9702 ion chamber detector 115/230 Vac ± 10%, 50/60 Hz, 500 VA 9719B/9720B scintillation detector 24 Vdc ± 20%, 12 VA, at detector input 115/230 Vac ± 15%, 50/60 Hz, 12 VA Thermo Fisher Scientific DensityPRO+ Installation Guide 3-3

20 Installation The Transmitter The Transmitter Warning Do not install the unit in any hazardous area other than those approved. Refer to the equipment tag for the specific approvals applicable to the configuration of your instrument. Follow all instructions on the installation drawings. Warning Do not apply power to the unit in any hazardous area unless the safety ground is properly wired inside the unit and the cover is properly installed. Warning Use proper lifting procedures during installation to avoid injury. Refer to the proper dimensional drawing for the transmitter mounting dimensions: : 1400A transmitter with non-metallic enclosure (NEMA 4X) : 1400A transmitter with cast aluminum enclosure (NEMA 7/ NEMA 4) : 1400S transmitter with stainless steel enclosure (NEMA 4X/IP65) Mount the transmitter in one of the following two ways: With four screws: Place one screw through each of the four holes in the transmitter enclosure s flange and tighten. With four nuts, four bolts, and eight washers: Place one bolt through each of the four holes in the transmitter enclosure s flange, with one washer on each side. Screw on the nuts, and tighten the bolts. 3-4 DensityPRO+ Installation Guide Thermo Fisher Scientific

21 Installation The Transmitter Installing Transmitter Boards Warning Remove all power from the unit prior to opening the transmitter enclosure. Electrocution can result if power is present. The following procedure applies to all transmitter boards except for the AC power supply. Refer to AC Power Supply later in this chapter. 1. Verify that power has been removed from the transmitter. 2. Remove the screws on the faceplate (cover) and open it. 3. If replacing a board, remove the board that is currently installed: a. Unplug any screw-in terminal connectors that have been wired from the board. b. Remove the board retaining screw in the upper card guide. c. Firmly but carefully push down on the top edge of the board to disengage the board connector from the transmitter. Pull up and out on the bottom edge of the board to remove it. 4. Align the board with the upper and lower card guide slots, and insert the board into the slots. 5. Line up the 3-pin by 16-pin male connector on the board with the corresponding connector on the back of the transmitter. 6. Firmly but carefully push on the board until the connector is fully seated. 7. Insert and tighten the board retaining screw in the upper card guide. 8. Connect the screw-terminal connector(s) to the corresponding replacement board connectors. 9. Refer to Chapter 4 for wiring procedures. 10. Close the transmitter and tighten the cover screws. Thermo Fisher Scientific DensityPRO+ Installation Guide 3-5

22 Installation The Transmitter AC Power Supply Warning If a power supply fails, the input capacitor might still be charged at a high voltage (up to 400 volts). Caution Do not remove the four screws that secure the power supply to the mounting bracket. The power supply and mounting bracket are placed as a single unit. Note The old power supply assembly is not field repairable. 1. Follow the procedure in step 3 above to remove all other transmitter boards (CPU, PCB 1, PCB 2, PCB 3, and PCB 4). 2. Remove the two hold-down screws from the base of the power supply mounting bracket. These screws secure the power supply mounting bracket to the main board. 3. Disconnect cables from the power supply (AC, DC, and Protective Safety Ground). Cut the tie wrap that secures the toroid to the power supply. 4. If replacing the power supply, remove the existing power supply and mounting bracket assembly. Replace it with the new supply. 5. Connect the cables to the replacement power supply (see Power to the Transmitter in Chapter 4 for wiring instructions). 6. Tighten the two hold-down screws to secure the power supply to the main board. (A screw-holding screwdriver facilitates installation.) 7. Reinstall the transmitter boards. 3-6 DensityPRO+ Installation Guide Thermo Fisher Scientific

23 Installation The Gauge Head The Gauge Head Warning Installation must be in accordance with local and national electric codes for the area classifications. Warning The source housing handle must be in the OFF position during installation. Warning Do not reach inside the source housing at any time during installation. Warning Use proper lifting procedures during installation to avoid injury. Mounting Configurations The optimum gauge mounting configuration depends on the application; however, the source head and the detector are typically mounted together on the opposite sides of a pipe. The following mounting options are supported for the gauge: Dual chain Pipe saddle (cradle) mount Pipe spool with gauge head pre-installed The chain mount is the most frequently used mounting configuration, as it allows the unit to be mounted on a range of pipe sizes using the same mounting hardware. Note Moving the gauge to a pipe with a different diameter may require changing the radiation source size and/or reconfiguring the detector (ion chamber) electronics. Contact Thermo Fisher for assistance. In addition to the guidelines listed earlier in this chapter, note the following when planning gauge head installation. Whichever mounting configuration is used, it is important to align the beam path (the centerline of the source housing) as closely as possible with the centerline of the detector housing. Be sure to mount the gauge head securely, since any movement or change in alignment can affect the gauge s calibration. Thermo Fisher Scientific DensityPRO+ Installation Guide 3-7

24 Installation The Gauge Head For best performance the beam must pass through a representative cross-section of the process material being measured. The gauge measures only the material that passes through the beam. Prevent suspended solids from settling out of the measured area by mounting the gauge head on a vertical section of pipe. For mounting on a horizontal pipe: Position the beam path at a 30- to 45-degree angle from vertical. This position tends to average density variations caused by settling, while reducing the effect of any trapped gases or solids that accumulate in the top or bottom of the pipe. If the process material is a solution, a light slurry, or a single-phase liquid that will not separate, the gauge head can be mounted such that the beam is horizontal. If the process material is a slurry, position the gauge head as far as possible from any elbows, tees, or valves (these tend to separate suspended solids). Position the beam path in the plane of the upstream elbow so the measurement includes any uneven distribution caused by the fitting. Figure 3 1. Gauge installation examples 3-8 DensityPRO+ Installation Guide Thermo Fisher Scientific

25 Installation The Gauge Head Mounting Instructions The following sections provide installation details for the various mounting configurations. Refer to the following drawings for detector dimensions : 9719B scintillation detector, NEMA 4 housing : 9720B scintillation detector, explosion proof housing : 9701 ion chamber detector, NEMA 4 housing : 9702 ion chamber detector, explosion proof housing Warning Use proper lifting procedures during installation to avoid injury. Caution Do not over tighten the bolts. Dual Chain Mount For the dual chain mount, two mounting chains are wrapped around the pipe with their ends engaging keyhole-shaped openings in the source head mounting plate. The detector housing is then clamped to the two chains center links (opposite the source mounting plate) by tensioning bolt assemblies. Details of the chain mount depend on the detector type, but the installation is similar for both types. Refer to the following drawings for mounting details : 9719B scintillation detector : 9720B scintillation detector : 9701/9702 ion chamber detector Figure 3 2. Dual chain mount Thermo Fisher Scientific DensityPRO+ Installation Guide 3-9

26 Installation The Gauge Head 1. Use the tables below to determine the number of chain links to thread through the keyhole slots. Count the links and mark the ones that will be engaged at each end of the two chains (four links in all). For scintillation detectors, use chain p/n For ion chamber detectors, use chain p/n Table 3 1. Dual chain mounting for 9719B and 9720B scintillation detectors Pipe Size (inches) Links from Center Links from End Links from Center Links from End Table 3 2. Dual chain mounting for 9701 and 9702 ion chamber detectors Pipe Size (inches) Links from Center Links from End Links from Center Links from End 3-10 DensityPRO+ Installation Guide Thermo Fisher Scientific

27 2. For 9701/9702 ion chamber detectors, partially assemble the two tensioning bolts (hardware kit ) in the following order: a. 1 tensioning bolt (large) b. 1 flat washer: Place onto bolt next to bolt head. c. 1 spacer ring d. 2 Belleville washers with concave sides together e. 1 special nut: Thread nut securely onto the bolt end. Installation The Gauge Head 3. For 9719B/9720B scintillation detectors, partially assemble the two tensioning bolts (hardware kit ) in the following order: a. 1 tensioning bolt (large) b. 1 chain: Place center ring of chain over the bolt head. c. 1 bar nut: Position the guard tube towards the bolt head. d. 1 special nut: Thread the nut securely onto the bolt end. After the special nut is fully threaded, thread the bar nut back down the bolt so that it rests against the special nut. 4. From the pipe side of the source mounting plate, thread one end of a chain through the plate s top keyhole opening until you reach one of the four end links you marked in step 1. Engage this link by sliding it into one of the keyhole slots. 5. Thread one end of the other chain through the bottom keyhole and engage the marked link as you did with the top chain. 6. Position the source housing mounting plate on the pipe. If necessary, use a lift, hoist, or other means to hold it in position. 7. Pass the center part of the upper and lower chains around the pipe and thread their ends through the corresponding keyholes. Engage the marked links in the empty slots. Note All four chain ends must engage at the same link. If you make an adjustment, adjust all four chain ends by the same amount. Also make sure there are no twists in the chain as you wrap it around the pipe. Otherwise the gauge head will not be aligned correctly. Thermo Fisher Scientific DensityPRO+ Installation Guide 3-11

28 Installation The Gauge Head Note You may need to adjust chain links (thread more or fewer links through each slot) to allow for pipe size tolerance, insulation, etc. Start with approximately 10 cm (4 inches) of play. 8. Position the detector on the pipe, opposite the source mounting plate. If necessary, use a lift, hoist, or other means to hold it in position. 9. To secure the detector and the source head mounting plate on the pipe: a. Position the upper chain, including the tensioning bolt assembled earlier, over the detector s top mounting arm. b. Place a spacer ring and two Belleville washers with concave sides together onto the pilot end of the special nut. c. Insert the pilot end of the chain tensioning bolt into the hole in the top detector mounting arm and tighten the chain tensioning bolt finger-tight. d. Repeat the previous three steps for the lower chain. 10. Alternately and uniformly tighten the chain tensioning bolts until the spacers are just touching the mounting arms. 11. Place the loose ends of the upper chain over the top edge of the source mounting plate so they will not interfere with the source housing installation. 12. Position the source housing so its four mounting holes engage the four studs on the mounting plate. If necessary, use a lift, hoist, or other means to hold it in position. 13. Use the provided lock washers and nuts to secure the source housing. Pipe Saddle (Cradle) Mount A pipe saddle, or cradle, mount consists of two identical mounting plates that are bolted together on opposite sides of the process pipe. Refer to drawing 85726N and follow the steps below to install a gauge head with a pipe saddle mount DensityPRO+ Installation Guide Thermo Fisher Scientific

29 Installation The Gauge Head Figure 3 3. Pipe saddle mount 1. Assemble the two halves of the saddle mount onto the process pipe using the supplied nuts, bolts, and washers. Thread the nuts and bolts together, but do not tighten them. 2. Adjust the saddle mount halves so the mounting plates are parallel and even with each other. Then tighten the bolts evenly so the clamp ends are the same distance apart on both sides of the pipe. 3. Bolt the detector and source housings to the mounting plates on either side of the pipe saddle. 4. Tighten all bolts securely so the gauge head components cannot shift positions. Thermo Fisher Scientific DensityPRO+ Installation Guide 3-13

30 Installation The Gauge Head Pipe Spool Mount A pipe spool is normally a 30-inch section of pipe with either mounting plates or a complete gauge head already installed. Often, a length of the required pipe is sent to Thermo Fisher to be converted into a pipe spool mount. Caution Use correct pipe fitting techniques suitable for the pipe being used and the process material that the pipe will handle. 1. Assemble the pipe spool into the existing pipe. 2. Position the source housing so its identification tag is upright. 3. If the detector and source housings are not already installed, bolt them to the mounting plates on either side of the pipe saddle. 4. Tighten all bolts securely DensityPRO+ Installation Guide Thermo Fisher Scientific

31 Chapter 4 Wiring Wiring should be performed in the following order: 1. Power to the transmitter and detector 2. The detector to the transmitter 3. Optional wiring: a. Relay contacts b ma current output(s) c. Contact closure inputs d. Optional temperature compensation board to the transmitter e. Serial communications Preparation Review the following carefully prior to connecting any wiring. Warning Remove all power from the unit before making any connections. Electrocution can result if power is present. Warning All wiring must be done by qualified individuals in accordance with applicable codes such as the NEC (National Electric Code) ANSI/NFPA 70 specifications or the Canadian Electrical Code Part 1. Only approved conduit, boxes, and fittings may be used. Warning Do not apply power to the unit in any hazardous area unless the safety ground is properly wired inside the unit and the cover is properly installed. Thermo Fisher Scientific DensityPRO+ Installation Guide 4-1

32 Wiring General Wiring Procedures Warning For hazardous location installations, the cable entries must be sealed per the installation layout drawing For non-hazardous location installations, the cable entries into the enclosures must be sealed to prevent passage of gas or vapors. The surrounding atmosphere or liquids should not affect the sealing compound. The minimum thickness of the sealing compound should be 16 mm (5/8 ). Warning If metal conduit is used, the conduit must be grounded. Warning Before wiring, verify that the source shutter is in the closed (OFF) position. The installation layout drawing provides general guidance for routing the cables between the detector and the transmitter. The drawings below provide instructions on wiring the detector and transmitter : Installation wiring, 1400A/1400S transmitter : Installation wiring, 9719B/9720B scintillation detector : Installation wiring, 9701/9702 ion chamber detector General Wiring Procedures This section provides general instructions for detector and transmitter wiring. Wiring details, such as power supply connections, are provided in the remainder of this chapter. For Scintillation Detectors Use the following procedure when wiring 9719B and 9720B scintillation detectors: 1. Make sure all source shutters are in the OFF position. 2. Make sure all power to the gauge is turned off. 3. Remove the housing access cover: For the 9719B detector, remove the bolts that secure the cover to the housing. For the 9720B detector, loosen the screw on the cover retaining bracket and slide the bracket off of the housing cover. Unscrew the housing access cover (two lugs are provided on the top of the cover to aid in the removal of the cover). 4-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

33 Wiring General Wiring Procedures 4. Remove the cable conduit plugs only from the hole(s) that will be used. As shown in the installation layout drawing , lay one conduit for the signal cables (and DC power input, if applicable), and a second conduit for the AC power input. 5. Pull each cable through the correct conduit fitting and into the enclosure. Leave approximately 150 mm (six inches) for strain relief. Secure the conduit, making sure it is completely sealed. 6. Connect the input power and the signal wires as follows. Reference the installation wiring drawing a. Remove the plug-in screw-terminal connector from the board connector. The connector should be tightly seated. To remove the connector, brace the board with your hand, being careful not to touch any of the circuits or components. b. Use a 1/8-inch screwdriver to loosen the screws on the plug-in connector. c. Insert the wires and make connections as shown in the installation wiring drawing. d. Tighten the screws in the connector to secure the wires and plug the screw-terminal connector back into the corresponding onboard connector. e. The optional temperature compensation board uses a spring terminal block connector. For each wire to be connected, use a flat screwdriver to push and hold the appropriate tab down. Insert the wire, and then release the tab to secure the wire. 7. When the detector wiring is complete, replace and secure the detector housing cover. Thermo Fisher Scientific DensityPRO+ Installation Guide 4-3

34 Wiring General Wiring Procedures For Ion Chamber Detectors Use the following procedure when wiring 9701 and 9702 ion chamber detectors. You will need a 1/8-inch bladed screwdriver. Note While wiring the ion chamber detector, locate the Hi-Meg label on the detector circuit board. A large resistance value will be written on this label (for example, 4.7 x 1011). Write down this value along with the detector s serial number. The Hi-Meg value will be needed during the gauge setup to fine tune the gauge. 1. Make sure all source shutters are in the OFF position. 2. Make sure all power to the gauge is turned off. 3. Remove the housing access cover: For the 9701 detector, remove the bolts that secure the cover to the housing. For the 9702 detector, loosen the screw on the cover retaining bracket and slide the bracket off of the housing cover. Unscrew the housing access cover (two lugs are provided on the top of the cover to aid in the removal of the cover). 4. Remove the foam insulator and the two brass-colored centering plates. 5. As shown in the installation layout drawing , the AC power and signal wiring should not be run in the same conduit for more than three meters (ten feet). If required, use a conduit splitter and lay one conduit for the signal cables and a second conduit for the AC power wiring. 6. Pull each cable through the conduit and into the enclosure. Leave approximately 150 mm (six inches) for strain relief. Secure the conduit, making sure it is completely sealed. 7. Connect the input power, the ion chamber signal wires, and the temperature compensation board wires, if applicable, as shown in the installation wiring diagram DensityPRO+ Installation Guide Thermo Fisher Scientific

35 Wiring General Wiring Procedures Note Connect the AC wiring Earth ground to the internal safety ground terminal (on the inside of the detector housing) as shown in the wiring diagram. Do not connect the shield from the signal cable at the detector end. 8. Position the two brass-colored centering plates over the circuit board with the smaller plate on top of the larger plate. Be careful to not pinch any wiring. Thread in and finger-tighten the four screws that secure the centering plates. 9. Insert a medium-sized flat blade screwdriver into the stepped (jagged) opening in the larger centering plate. Twist the screwdriver to push the two plates apart until the detector is locked in position. Tighten the four screws to secure the centering plates and the detector. 10. When the wiring is complete, replace and secure the detector housing cover. For the Transmitter EMC Requirement for ATEX Version of 1400S Transmitters In order for the 1400S transmitter to meet electromagnetic emission requirements for CE-EMC approvals, ferrite clamps must be installed within the transmitter enclosure on all power and signal wiring going into and out of the transmitter housing. One ferrite clamp is installed on the wires of the AC power supply input. Five additional ferrite clamps are provided with the EMC ferrite clamps kit (p/n ). The installation procedure accompanies the kit. It is also provided in Appendix D. Note The ferrite clamps are only used in the ATEX version of the 1400S transmitter. They are not required for the CSA version or the plastic 1400A transmitter. Thermo Fisher Scientific DensityPRO+ Installation Guide 4-5

36 Wiring General Wiring Procedures Procedure The transmitter has slots for the following boards: Optional AC power supply CPU board PCB 1: Standard current I/O / DC power board PCB 2: Optional current I/O board with two contact inputs and two optional relays PCB 3: Optional current I/O board with two contact inputs and two optional relays PCB 4: Voltage/current pulse interface (VPI) board (p/n ), used to interface with the detector Follow this general procedure for wiring boards installed in the transmitter, RS232 or RS485 serial ports, relay contacts, contact closure inputs, 4 20 ma current output(s), and detector signal cables to the transmitter. You will need a 1/8-inch bladed screwdriver. Note CSA certified fittings must be used to maintain CSA rating for the enclosure. 1. Ensure power is removed, and open the enclosure cover. 2. Remove the plugs only from the cable conduit holes you will use. The holes are labeled on the transmitter enclosure. 3. Pull the cable through the fitting on the end of the enclosure. Leave approximately 6 inches (150 mm) of stress relief. Secure the conduit, ensuring it is completely sealed. 4. For ATEX versions of the 1400S: Install a ferrite clamp on all wiring. You will need kit Refer to Appendix D. 5. Connect a cable to a PCB as follows: a. Remove the plug-in screw terminal connector from the board connector. If the connector is tight, brace the board with your hand to remove the screw terminal connector, but be careful to not touch any circuit components. b. Loosen the terminal screws on the connector. Insert the wires into the connector and make connections as shown in the appropriate installation wiring drawing. 4-6 DensityPRO+ Installation Guide Thermo Fisher Scientific

37 c. Tighten the terminal screws to secure the wires. Wiring Power Supply Wiring d. Plug the screw-terminal connector back into the corresponding onboard connector. 6. When all connections are complete, close and secure the transmitter cover. Detector Type Selection Three, two-pin headers (JP1, JP2, and JP3) are located along the edge of the VPI board near the top of the board. A jumper should be installed on one (and only one) of these headers to indicate the detector type as follows: JP1: Ion chamber detector JP2: Scintillation detector SGD mode JP3: Scintillation detector PRO mode Caution Remove power from the transmitter before changing jumper settings. Failure to do so may result in damage to the VPI board. Power Supply Wiring Power to the Scintillation Detector When connecting power to the 9719B/ 9720B scintillation detectors, refer to installation wiring diagram Note To meet the requirements of CSA , an external switch or circuit breaker must be installed to allow the power source to be disconnected from the gauge. In addition, protective bonding (grounding) must always be provided, even if a DC power source is used. DC Power The 9719B/9720B scintillation detectors are designed to operate on 24 Vdc ± 20%. The input connector for the DC source voltage wiring is located on the interface/ adapter board. Note To meet the requirements of CSA , the input DC terminals shall be supplied from an SELV (Safety Extra Low Voltage) source. Connections are shown in the table below. Thermo Fisher Scientific DensityPRO+ Installation Guide 4-7

38 Wiring Power Supply Wiring Table 4 1. DC power wiring for 9719B/9720B scintillation detectors User Power Supply (+24Vdc, 0.5 A) Detector + J1-2 - J1-1 AC Power With the AC power board installed, the scintillator detector may be operated using either 115 or 230 Vac ± 15%. The AC supply voltage is set by the 115/230 Vac selector switch. The selector switch is located on the AC power board, below the connectors and just above the transformer. To access the selector switch, it is necessary lift the detector chassis several inches out of the housing. If both AC and DC input power are supplied to the detector, the detector will draw power from whichever source provides the higher DC voltage. Caution Applying 230 Vac with the selector switch in the 115 Vac position will damage the equipment. Caution For reliable operation and to maintain safety approval, the F1 fuse on the AC power board must only be replaced with an approved fuse. Reference installation wiring diagram Table 4 2. AC power wiring for 9719B/9720B scintillation detectors AC Power Supply Detector N J17-1 L J17-2 Note Connect the AC wiring Earth ground to the internal safety ground terminal as shown in the wiring diagram. 4-8 DensityPRO+ Installation Guide Thermo Fisher Scientific

39 Wiring Power Supply Wiring Power to the Ion Chamber Detector The 9701/9702 ion chamber detectors may be operated using either 115 or 230 Vac ± 10%. The maximum input power requirement is 500 VA. When connecting power to the 9701/9702 detectors, refer to installation wiring drawing Note Connect the AC wiring Earth ground to the internal safety ground terminal provided on the inside of the detector housing as shown in the wiring diagram. Caution Applying 230 Vac with the selector switch in the 115 Vac position will damage the equipment. The ion chamber detector is shipped wired for either 115 or 230 Vac operation. If necessary, the detector s operating voltage can be changed by following the procedure below. 1. The two red wires should be installed on the marked push-on tabs (RED1 and RED2). 2. Install the four heater wires (two white wires and two black wires) on the push-on tabs as follows: a. One white wire on the Wh tab. b. One black wire on the Bk tab. c. For 115 Vac operation: - One white wire on the Wh/115 tab. - One black wire on the Bk/115 tab. d. For 230 Vac operation: - One white wire on the Wh/230 tab. - One black wire on the tab Bk/230 tab. 3. Set the 115/230 Vac selection switch (located next to the transformer) to the desired voltage. Caution For reliable operation and to maintain safety approval, the F1 fuse on the AC power board must only be replaced with an approved fuse. Reference installation wiring diagram Thermo Fisher Scientific DensityPRO+ Installation Guide 4-9

40 Wiring Power Supply Wiring Power to the Transmitter When connecting power to the 1400A/1400S transmitters, refer to installation wiring drawing Note To meet the requirements of CSA , an external switch or circuit breaker must be installed to allow the power source to be disconnected from the gauge. In addition, protective bonding (grounding) must always be provided even if a DC power source is used. AC Power With AC power, the 1400A and 1400S transmitters are designed to operate on Vac, at an input frequency of 50/60 Hz. The same system configuration can be directly wired to either 115 or 230 Vac. The maximum input power requirement is 12 VA when all options are present. Connections are shown in the table below. Table 4 3. AC power wiring for 1400A/1400S transmitters AC Power Supply L N Transmitter J1-L J1-N DC Power Note To meet the requirements of CSA , the input DC terminals must be supplied from an SELV (Safety Extra Low Voltage) source. With 24 Vdc power, the 1400A and 1400S transmitters are designed to operate on Vdc. The maximum input power requirement is 12 VA when all options are present. Connections are shown in the table below. Table Vdc power wiring for 1400A/1400S transmitters +24Vdc Power Supply +24Vdc +24Vdc return Transmitter J5-8 (+24Vdc input) J5-9 (GND) 4-10 DensityPRO+ Installation Guide Thermo Fisher Scientific

41 Wiring Detector Signal Wiring Detector Signal Wiring All signals from the detector are wired to the VPI board that is installed in the transmitter slot PCB 4. Refer to the appropriate installation wiring drawings for the detailed requirements : Installation wiring, 1400A/1400S transmitter : Installation wiring, 9719B/9720B scintillation detector : Installation wiring, 9701/9702 ion chamber detector Caution Remove power from the transmitter before changing jumper settings. Failure to do so may result in damage to the VPI board. Signal wiring for the 9719B/9720B scintillation detectors are shown in the tables below. Use AWG individually shielded cable (1000 ft maximum). Table 4 5. Wiring for scintillation detector data and references pulses Scintillation Adaptor Board VPI Board in 1400A/1400S Transmitter J2-1 (DATA (H)) J1B-15 (+PULSE DATA) J2-2 (SHIELD) J1B-21 (SHIELD) J2-3 (DATA (L)) J1B-16 (-PULSE DATA) J2-4 (REF (H)) J1B-17 (+PULSE REF) J2-5 (SHIELD) J1B-21 (SHIELD) J2-6 (REF (L)) J1B-18 (-PULSE REF) Table 4 6. Wiring for scintillation detector HV pulses or duty cycle mode Scintillation Adaptor Board VPI Board in 1400A/1400S Transmitter J3-1 (INC (H)) J1B-19 (+HV INCREASE) J3-2 (SHIELD) J1B-24 (SHIELD) J3-3 (INC (L)) J1B-20 (-HV INCREASE) J3-4 (DEC (H)) J1B-22 (+HV DECREASE) J3-5 (SHIELD) J1B24 (SHIELD) J3-6 (DEC (L)) J1B-23 (-HV DECREASE) Thermo Fisher Scientific DensityPRO+ Installation Guide 4-11

42 Wiring Optional Wiring Note The transmitter is used in duty cycle mode unless it is wired to a D series scintillation detector. The duty cycle mode is set by connecting jumper P3 pin 1 and 2 on the scintillation adapter board and shorting JP3 on the VPI board. Note Install two jumpers on P2. Jumper pin 1 to pin 3 and pin 2 to pin 4. Signal wiring for the 9701/9702 ion chamber detectors is shown below. Table 4 7. Wiring for ion chamber detector signal 9701/9702 Ion Chamber VPI Board in 1400A/1400S Transmitter J2-3 (+) J1A-8 (+ION CHAMBER) J2-4 (-) J1A-0 (-ION CHAMBER) Optional Wiring For optional wiring, refer to installation wiring drawing Relays I/O boards installed in transmitter slots PCB 1 through PCB 3 can have two relays each, allowing for a maximum of six relays. Connector J1 is for relays and is only installed on the PCB if relays were specified in the order. Relay contacts are Form C SPDT, isolated 8 A, 22 Vac. You can assign process alarms to control the relays. Refer to the DensityPRO+ user guide for details. Current Output One current output (I/O) board is supplied as standard and installed in the PCB 1 slot. Two additional I/O boards are optionally available are installed the transmitter s PCB 2 and PCB 3 slots. The current output is normally shipped from the factory configured as an isolated self-powered output. You can reconfigure the current output as isolated loop-powered by removing the jumper from pins 3 and 4 of the J5 connector. In the loop-powered configuration, the user must supply +24 Vdc loop power to pin 3 of J5. Pin 2 of J5 is Iout+. The current output is programmable between 3.8 and 20.5 ma. Refer to the DensityPRO+ user guide for details. The minimum output in the selfpowered configurations is approximately 0.4 ma. The minimum output current in the loop-powered configuration is approximately 3 ma. Maximum load is 800 ohms DensityPRO+ Installation Guide Thermo Fisher Scientific

43 Wiring Serial Communications Contact Closure (Switch) Inputs Two contact closure inputs are included on each I/O board. They are dry contact inputs between ground and switch 1 and ground and switch 2. The gauge can be configured to execute a command or other function upon a user-provided contact opening or closing. Refer to the DensityPRO+ user guide for details on assigning commands to the contact closure inputs. Temperature Compensation Temperature compensation circuitry with 2-wire or 3-wire 100 ohm platinum RTD is optionally available. For 9701/9702 scintillation detectors, connections are made at J19. For 9719B/9720B ion chamber detectors, connections are made from the temperature compensation board. For the 1400A/1400S transmitter, connections are made at J1A of the VPI board installed in the PCB 4 slot. Reference the appropriate installation wiring diagrams for specific wiring information. Flow Input Serial Communications There is a provision on the VPI board for a user-provided 4 20 ma flow input (pins 6 and 7 of connector J1A). Refer to the DensityPRO+ user guide for details on how to configure the gauge to use the flow input signal. The transmitter provides one RS232 single-drop serial interface. One RS485 multi-drop serial interface is optionally available. Screw-terminal connectors for both ports are located on the CPU board. The optional RS485 connector (J6) includes the +8 Vdc required to power the Thermo Scientific HHT as well as the +DATA and DATA connections. An RJ-11 (phone jack) connector is also provided for the RS485 port. The HHT can be connected directly to the RJ-11 connector. Both the RS232 and optional RS485 ports are always active and provide independent access to the measurement readings and software functions. The setup menus, however, can only be accessed by one port at a time. For information on configuring communications, refer to the DensityPRO+ user guide. Thermo Fisher Scientific DensityPRO+ Installation Guide 4-13

44 Wiring Serial Communications Optional RS485 Wiring Connecting a PC serial port (COM) to the optional RS485 port on the gauge requires an RS485/RS232 converter (p/n ). The J6 connector on the transmitter CPU is for the RS485 connections. Refer to installation wiring drawing and make the RS485 connections as follows. 1. Connect +DATA and DATA on the RS485 connector (J6) to the corresponding connections on the RS485/RS232 converter. 2. Connect the RS485/RS232 converter to the PC using a standard DB9 serial cable. RS485 Termination If you are having trouble using another device on the RS485 chain, verify that it is properly terminated for its position on the chain. To terminate a device, connect a 120 resistor between its RS485 +/ data terminals. Never terminate more than the first and last device in the chain. The transmitter includes a three-pin header (W4), located near the bottomfront edge of the CPU board next to the RS485 connector (J6), to terminate the RS485 connection. For a single unit installation or for a multiple unit installation where the transmitter is the last unit in line, the W4 jumper should be placed on pins 1 and 2 to terminate the RS485 line. For a multiple unit installation where the transmitter is the not last unit in line, the W4 jumper should be placed on pins 2 and 3 so that the RS485 line is not terminated at the transmitter. RS232 Wiring COM1 or COM2 on a PC can be connected directly to the transmitter RS232 port. This connection requires five wires: ground, TX, RX, RTS, and CTS. The RS232 port connector (J8) is located on the transmitter CPU board. Connections for standard DB9 and DB25 PC serial port connectors are listed in the table below. Table 4 8. RS232 connections RS232 Connector (J8) on Transmitter CPU PC Serial Port DB9 Connector J8-1 (RTS) Pin 8 Pin 5 J8-2 (TX) Pin 2 Pin 3 J8-3 (CTS) Pin 7 Pin 4 J8-4 (RX) Pin 3 Pin 2 J8-5 (GND) Pin 5 Pin 7 PC Serial Port DB25 Connector 4-14 DensityPRO+ Installation Guide Thermo Fisher Scientific

45 Wiring HART Communications HART Communications The HART communication protocol is supported over the 4 20 ma current output with an optional daughter board. You communicate with the gauge using the standard 275 or 375 Field Communicator. Refer to the DensityPRO+ HART operation manual (p/n ). Thermo Fisher Scientific DensityPRO+ Installation Guide 4-15

46 This page intentionally left blank.

47 Chapter 5 Getting Help Contact Information The local representative is your first contact for support and is well equipped to answer questions. You can also contact Thermo Fisher directly for application assistance. Process Instruments 1410 Gillingham Lane Sugar Land, TX USA +1 (800) (713) direct +1 (713) 4573 fax A-101, 1CC Trade Tower Senapati Bapat Road Pune Maharashtra, INDIA +91 (20) (20) fax 14 Gormley Industrial Avenue Gormley, Ontario L0H 1G0 CANADA +1 (905) (905) fax Ion Path, Road Three Winsford, Cheshire CW7 3GA UNITED KINGDOM +44 (0) (0) fax Unit , 7/F Tower West Yonghe Plaza No. 28 Andingmen East Street, Beijing CHINA +86 (10) (10) fax Returning Equipment for Service Contact Thermo Fisher for specific instructions. Please have the following available: Specific information about the problem. A contact name and phone number, in case we need more information. A purchase order authorizing repairs or a request for a quote. Thermo Fisher Scientific DensityPRO+ Installation Guide 5-1

48 This page intentionally left blank.

49 Appendix A Ordering Information Table A B/9720B scintillation detectors Part Number Description Scintillation adapter board (transmitter interface) AC power supply board -- Fuse-F1, 250V, 0.125A (1/8A SB), 3AG Temperature compensation board Triangle board Table A /9702 ion chamber detectors P/N Description Preamp/power supply boards -- Fuse-F1, 250V, 5A, Slo-Blow, 3AG Table A A transmitter P/N Description Power supply assembly CPU board I/O board, 24 Vdc input power, no relays I/O board, 24 Vdc input power, 2 relays I/O board, no DC input power, no relays I/O board, no DC input power, 2 relays -- I/O board fuse-f1, 250V, 0.75A, Slo-Blow HART communication option VPI board RS485 board Thermo Fisher Scientific DensityPRO+ Installation Guide A-1

50 Ordering Information Table A S transmitter P/N Description HLT-MTXR-SS 1400S stainless steel transmitter Kit, NDM CPU board with EPROM for density VPI board AC power supply OTO545B HART interface module I/O board, 24 Vdc input power, no relays I/O board, 24 Vdc input power, 2 relays I/O board, no DC input power, no relays I/O board, no DC input power, 2 relays RS485 board Kit, EMC ferrite clamps for 1400S transmitter (for ATEX units only) A-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

51 Appendix B Specifications Results may vary under different operating conditions. Table B 1. Performance specifications Operating principle Accuracy Stability Gamma radiation from the source passes through the process pipe and material. The amount of radiation reaching the detector decreases as the process level increases. ± 1% of span Drift less than 1.0% absolute radiation change per 6 months Table B /9702 ion chamber detector specifications Power supply Operating temperature range 115/230 Vac ±10%, 50/60 Hz, 500 VA -30 C to +60 C (-20 F to +140 F) Approvals Non-explosion proof (9701): NEMA 4, Type 4 Explosion proof (9702): Class I, Div. 1, Groups B, C, D Class II, Div. 1, Groups E, F, G Class III NEMA 4, Type 4 Dimensions Non-explosion proof (9701): ~8 in diameter x 15.5 in tall (200 mm diameter x 400 mm tall) Explosion proof (9702): ~8 in diameter x 19.5 in tall (200 mm diameter x 500 mm tall) Weight Non-explosion proof (9701): 58 lb (26.3 kg) Explosion proof (9702): 100 lb (45.4 kg) Thermo Fisher Scientific DensityPRO+ Installation Guide B-1

52 Specifications Table B B/9720B scintillation detector specifications Power options Operating temperature range Approvals Dimensions Wiring entry CE-EMC surge protection board junction box 24 Vdc ± 20%, 12 VA, at detector input 115/230 Vac ±15%, 50/60 Hz, 12 VA -30 C to +60 C (-20 F to +140 F) Non-explosion proof (9719B): Class II, Div. 1, Groups E, F, G Class III, Div. 1 NEMA 4, Type 4 Explosion proof (9720B): Class I, Div. 1, Groups B, C, D Class II, Div 1, Groups E, F, G Class III NEMA 4, Type 4 CE-ATEX II 2 G Ex d IIC T5 Gb CE-EMC compliance contingent upon installation of EMC protection kit as illustrated on drawing Non-explosion proof (9719B): ~6.5 in diameter x 14 in tall (165 mm diameter x 360 mm tall) Explosion proof (9720B): ~6.5 in diameter x 15 in tall (165 mm diameter x 380 mm tall) (2) 3/4 NPT female conduit ports. Use only suitably approved cable glands and blanking plugs. CE-ATEX II 2 G Ex d IIC T6 Gb -40 C<=Ta<=60 C Warning: Do not open when an explosive atmosphere may be present. B-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

53 Specifications Table B A transmitter specifications Power options Operating temperature Approvals Dimensions Weight 24 Vdc input (20 28 Vdc), 12 VA 115/230 Vac input ( Vac), 50/60 Hz, 25 VA -40 C to +60 C (-40 F to +140 F) Non-explosion proof (plastic enclosure): Class I, Div. 2, Groups A, B, C, D Class II, Div. 2, Groups E, F, G Class III NEMA 4X, Type 4X Explosion proof enclosure: Class I, Div. 1, Groups C, D Class II, Div. 1, Groups E, F, G Class III NEMA 4 / 7, Type 4 / 7 Non-explosion proof (plastic enclosure): 12 x 8.5 x 8.25 in (305 x 216 x 210 mm) Explosion proof enclosure: 14 x 15.5 x 10.5 in (356 x 394 x 267 mm) Non-explosion proof (plastic enclosure): 7.5 lb (3.4 kg) Explosion proof: 60 lb (27.2 kg) Circuit protection Meets IEC 801-1, -2, and -3 Current outputs and relay outputs are isolated Display Serial interface Current output Contact closure inputs Relays Four-line backlit display; up to eight readouts displayed simultaneously RS232, full duplex communication with remote terminal, printer, PC, or host computer Optional RS485, half-duplex party-line communication to host computer or HHT One standard with two additional optionally available Standard configuration: Isolated, self-powered, 700-ohm maximum load Alternate configuration: Isolated, loop-powered, 24 Vdc nominal supply voltage, 800-ohm maximum load Two per I/O board Two optional per I/O board; six maximum. Form C SPDT, isolated 8 A, 220 Vac Thermo Fisher Scientific DensityPRO+ Installation Guide B-3

54 Specifications Table B S transmitter Power options Operating temperature Approvals 24 Vdc (20 28 Vdc), 12 VA 115/230 Vac ( Vac), 50/60 Hz, 12 VA -40 C to +60 C (-40 F to +140 F) NEMA 4X / IP65 ATEX II 3 G Ex na IIC T6 (-40 C Ta 60 C) DC version, EPSILON 08 ATEX 2387: II 3 G Ex na nc IIC T4 (-20 C Ta 50 C) AC version, EPSILON 08 ATEX 2387 CSA Class I, Div. 2, Groups A, B, C, and D; Class II, Div. 2, Groups F and G; Class III Enclosure type 4X, Temperature code T3C / T4 CE compliant Low Voltage Directive compliant EMC Directive compliant Dimensions See drawing Display Serial outputs Current output Contact closure inputs Four-line backlit display; up to eight readouts displayed simultaneously RS232 full duplex RS485 half-duplex One standard with two additional optionally available Standard configuration: Isolated, self-powered, 800-ohm maximum load Alternate configuration: Isolated, loop-powered, 24 Vdc nominal supply voltage, 800-ohm maximum load Two per I/O board B-4 DensityPRO+ Installation Guide Thermo Fisher Scientific

55 Appendix C Drawings Note Information presented in this chapter has been regenerated from original drawings. Every effort is made to maintain document accuracy. However, in order to enhance legibility, the documents may have been restructured, and some information may have been intentionally excluded. Therefore, the drawings within this guide may not be an exact duplicate of the original drawings. Note Drawings in this manual are included for reference only and may not be the current version. Contact the factory if you need a copy of the latest revision. Table C 1. Installation layout & wiring diagrams Drawing # Rev. Description Page A Installation layout, 1400A transmitter to 9719B, 9720B, 9701, 9702 detectors (2 sheets) D Installation wiring, 9719B/9720B scintillation detectors to 1400A transmitter (3 sheets) C 4 C B Installation wiring, 1400A/1400S transmitter (3 sheets) C A Installation wiring, 9701/9702 ion chamber detectors to 1400A transmitter (1 sheet) C B RS485 installation and wiring (1 sheet) C 13 Table C 2. Transmitter dimensional drawings Drawing # Rev. Description Page F 1400A transmitter dimensional diagram, NEMA 4X enclosure (2 sheets) D 1400A transmitter dimensional diagram, NEMA 4/NEMA 7 enclosure (1 sheet) C 14 C A 1400S transmitter dimensional diagram (1 sheet) C 17 Thermo Fisher Scientific DensityPRO+ Installation Guide C-1

56 Drawings Table C 3. Detector dimensional drawings Drawing # Rev. Description Page B Installation drawing, detector assembly, flat mount, NEMA 4 housing (2 sheets) B Installation drawing, detector assembly, flat mount, explosion proof housing (2 sheets) Mounting dimensions, 9702 ion chamber detector, explosion proof housing (1 sheet) Mounting dimensions, 9701 ion chamber detector, NEMA 4 housing (1 sheet) C 18 C 20 C 22 C 23 Table C 4. Source housing dimensional drawings Drawing # Rev. Description Page B Mounting dimensions 5200 source housing, 100 mci or less (1 sheet) C Mounting dimensions 5201 source housing, 100 mci or less (1 sheet) D Mounting dimensions 5202 source housing, 500 mci or less (1 sheet) C Mounting dimensions 5203 source housing, 2000 mci or less (2 sheets) C Mounting dimensions 5204 source housing, 8000 mci or less (2 sheets) C 24 C 25 C 26 C 27 C 29 Table C 5. Gauge head installation drawings Drawing # Rev. Description Page B Installation drawing, gauge head, two-chain mount, NEMA 4 housing (3 sheets) B Installation drawing, gauge head, two-chain mount, explosion proof housing (3 sheets) A Installation drawing, gauge head, two-chain mount, 9701/9702 ion chamber detector (1 sheet) C 31 C 34 C 37 C-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

57 Drawings Table C 5. Gauge head installation drawings, cont. Drawing # Rev. Description Page 85726N E Installation drawing, gauge head, pipe saddle mount (1 sheet) C C Fabrication details, 1 to 3 Z-pipe section (2 sheets) C B Mounting dimensions, 1 to 3 Z-pipe mount (1 sheet) C A Fabrication details, 4 Z-pipe section (1 sheet) C A Mounting dimensions, 4 Z-pipe mount (1 sheet) C 43 Table C 6. Source housing options Drawing # Rev. Description Page E Mounting instructions, remote manual actuator for source housings (2 sheets) C 44 Thermo Fisher Scientific DensityPRO+ Installation Guide C-3

58 Drawings Figure C : Installation layout, 1400A transmitter to 9719B, 9720B, 9701, 9702 detectors (sheet 1 of 2) C-4 DensityPRO+ Installation Guide Thermo Fisher Scientific

59 Drawings Figure C : Installation layout, 1400A transmitter to 9719B, 9720B, 9701, 9702 detectors (sheet 2 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-5

60 Drawings Figure C : Installation wiring, 9719B/9720B scintillation detectors to 1400A transmitter (sheet 1 of 3) C-6 DensityPRO+ Installation Guide Thermo Fisher Scientific

61 Drawings Figure C : Installation wiring, 9719B/9720B scintillation detectors to 1400A transmitter (sheet 2 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-7

62 Drawings Figure C : Installation wiring, 9719B/9720B scintillation detectors to 1400A transmitter (sheet 3 of 3) C-8 DensityPRO+ Installation Guide Thermo Fisher Scientific

63 Drawings Figure C : Installation wiring, 1400A/1400S transmitter (sheet 1 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-9

64 Drawings Figure C : Installation wiring, 1400A/1400S transmitter (sheet 2 of 3) C-10 DensityPRO+ Installation Guide Thermo Fisher Scientific

65 Drawings Figure C : Installation wiring, 1400A/1400S transmitter (sheet 3 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-11

66 Drawings Figure C : Installation wiring, 9701/9702 ion chamber detectors to 1400A transmitter (sheet 1 of 1) C-12 DensityPRO+ Installation Guide Thermo Fisher Scientific

67 Drawings Figure C : Optional RS485 installation and wiring (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-13

68 Drawings Figure C : 1400A transmitter dimensional diagram, NEMA 4X enclosure (sheet 1 of 2) C-14 DensityPRO+ Installation Guide Thermo Fisher Scientific

69 Drawings Figure C : 1400A transmitter dimensional diagram, NEMA 4X enclosure Thermo Fisher Scientific DensityPRO+ Installation Guide C-15

70 Drawings Figure C : 1400A transmitter dimensional diagram, NEMA 4 / NEMA 7 enclosure (sheet 1 of 1) C-16 DensityPRO+ Installation Guide Thermo Fisher Scientific

71 Drawings Figure C : 1400S transmitter dimension diagram (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-17

72 Drawings Figure C : Installation drawing, detector assembly, flat mount, NEMA 4 housing (sheet 1 of 2) C-18 DensityPRO+ Installation Guide Thermo Fisher Scientific

73 Drawings Figure C : Installation drawing, detector assembly, flat mount, NEMA 4 housing (sheet 2 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-19

74 Drawings Figure C : Installation drawing, detector assembly, flat mount, explosion proof housing (sheet 1 of 2) C-20 DensityPRO+ Installation Guide Thermo Fisher Scientific

75 Drawings Figure C : Installation drawing, detector assembly, flat mount, explosion proof housing (sheet 2 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-21

76 Drawings Figure C : Mounting dimensions, 9702 ion chamber detector, explosion proof housing (sheet 1 of 1) C-22 DensityPRO+ Installation Guide Thermo Fisher Scientific

77 Drawings Figure C : Mounting dimensions, 9701 ion chamber detector, NEMA 4 housing (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-23

78 Drawings Figure C : Mounting dimensions 5200 source housing, 100 mci or less (sheet 1 of 1) C-24 DensityPRO+ Installation Guide Thermo Fisher Scientific

79 Drawings Figure C : Mounting dimensions Model 5201 source housing, 100 mci or less (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-25

80 Drawings Figure C : Mounting dimensions Model 5202 source housing, 500 mci or less (sheet 1 of 1) C-26 DensityPRO+ Installation Guide Thermo Fisher Scientific

81 Drawings Figure C : Mounting dimensions Model 5203 source housing, 2000 mci or less (sheet 1 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-27

82 Drawings Figure C : Mounting dimensions Model 5203 source housing, 2000 mci or less (sheet 2 of 2) C-28 DensityPRO+ Installation Guide Thermo Fisher Scientific

83 Drawings Figure C : Mounting dimensions Model 5204 source housing, 8000 mci or less (sheet 1 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-29

84 Drawings Figure C : Mounting dimensions Model 5204 source housing, 8000 mci or less (sheet 2 of 2) C-30 DensityPRO+ Installation Guide Thermo Fisher Scientific

85 Drawings Figure C : Installation drawing, gauge head, single chain mount, explosion proof housing (sheet 1 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-31

86 Drawings Figure C : Installation drawing, gauge head, single chain mount, explosion proof housing (sheet 2 of 3) C-32 DensityPRO+ Installation Guide Thermo Fisher Scientific

87 Drawings Figure C : Installation drawing, gauge head, single chain mount, explosion proof housing (sheet 3 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-33

88 Drawings Figure C : Installation drawing, gauge head, two-chain mount, explosion proof housing (sheet 1 of 3) C-34 DensityPRO+ Installation Guide Thermo Fisher Scientific

89 Drawings Figure C : Installation drawing, gauge head, two-chain mount, explosion proof housing (sheet 2 of 3) Thermo Fisher Scientific DensityPRO+ Installation Guide C-35

90 Drawings Figure C : Installation drawing, gauge head, two-chain mount, explosion proof housing (sheet 3 of 3) C-36 DensityPRO+ Installation Guide Thermo Fisher Scientific

91 Drawings Figure C : Installation drawing, gauge head, two-chain mount, 9701/9702 ion chamber detector (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-37

92 Drawings Figure C N: Installation drawing, gauge head, pipe saddle mount (sheet 1 of 1) C-38 DensityPRO+ Installation Guide Thermo Fisher Scientific

93 Drawings Figure C : Fabrication details, 1 to 3 Z-pipe section (sheet 1 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-39

94 Drawings Figure C : Fabrication details, 1 to 3 Z-pipe section (sheet 2 of 2) C-40 DensityPRO+ Installation Guide Thermo Fisher Scientific

95 Drawings Figure C : Mounting dimensions, 1 to 3 Z-pipe mount (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-41

96 Drawings Figure C : Fabrication details, 4 Z-pipe section (sheet 1 of 1) C-42 DensityPRO+ Installation Guide Thermo Fisher Scientific

97 Drawings Figure C : Mounting dimensions, 4 Z-pipe mount (sheet 1 of 1) Thermo Fisher Scientific DensityPRO+ Installation Guide C-43

98 Drawings Figure C : Mounting instructions, remote manual actuator for source housings (sheet 1 of 2) C-44 DensityPRO+ Installation Guide Thermo Fisher Scientific

99 Drawings Figure C : Mounting instructions, remote manual actuator for source housings (sheet 2 of 2) Thermo Fisher Scientific DensityPRO+ Installation Guide C-45

100 This page intentionally left blank.

101 Appendix D Adding Ferrite Clamps to 1400S Transmitter Note The ferrite clamps are only used in the ATEX version of the 1400S transmitter. They are not required for the CSA version or the plastic 1400A transmitter. If needed, the ferrite clamps kit may be ordered separately. In order for the 1400S transmitter to meet electromagnetic emission requirements for CE-EMC approvals, ferrite clamps must be installed within the transmitter enclosure on all power and signal wiring going into and out of the transmitter housing. Procedure The ferrite clamp is part of the EMC ferrite clamps kit (p/n ). The kit consists of the following: Five ferrite clamps (p/n ) At least one clamp key The procedure The clamp key is used to remove the ferrite clamp from the wire during wiring changes. Figure D 1. Ferrite clamps A ferrite clamp will need to be installed on all wiring that enters or exits the transmitter enclosure. This may be a shielded twisted pair or a bundle of individual wires. The limitation is that the wires through one clamp cannot exceed 5 mm in diameter. Also, all of the clamps MUST be installed within the enclosure. Thermo Fisher Scientific DensityPRO+ Installation Guide D-1

102 Adding Ferrite Clamps to 1400S Transmitter Procedure Figure D 2. Ferrite clamps installed within enclosure As shown in Figure D 2 the ferrite clamp can be placed around a shielded twisted pair cable or a bundle of individual cables, as long as the diameter of the bundle does not exceed 5 mm (a little over 3/16 ). The ferrite clamp kit (p/n ) contains five clamps. Depending on your wiring scheme, you may need additional clamps. They can be ordered using part number For example, in a Thermo Scientific DensityPRO+ system with a scintillation detector, flow input, and three outputs, there will be eight shielded twisted pair cables and one power cable. This system would require a total of nine ferrite clamps. So you would need either two kits or one kit and four additional clamps to be EMC compliant. Note All ATEX version 1400S transmitters shipped after August 2009 will contain one ferrite clamp kit and will have one ferrite clamp installed on the AC voltage input wiring. Note If the locking mechanism of the clamp should fail, a plastic wire tie can be used to secure the clamp around the wire. D-2 DensityPRO+ Installation Guide Thermo Fisher Scientific

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific.

DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N Part of Thermo Fisher Scientific. DensityPRO+ Gamma Density System with Remote Transmitter Installation Guide P/N 717818 Revision D Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter Installation

More information

Continuous Gamma Level System Installation Guide P/N

Continuous Gamma Level System Installation Guide P/N 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F Part of Thermo Fisher Scientific 4790 Continuous Gamma Level System Installation Guide P/N 717830 Revision F 2009 Thermo Fisher

More information

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN

LevelPRO Measurement Systems. Continuous Gamma Level System Installation Guide PN LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 LevelPRO Measurement Systems Continuous Gamma Level System Installation Guide PN 1-0702-040 2014 Thermo Fisher

More information

Troubleshooting Thermo Scientific LevelPRO Gauges

Troubleshooting Thermo Scientific LevelPRO Gauges Troubleshooting Thermo Scientific LevelPRO Gauges Key Words Scintillation detector Current board Relay output Communications Level indication Purpose A chapter in the user guide for the Thermo Scientific

More information

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C

DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N Part of Thermo Fisher Scientific. Revision C DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819 Revision C Part of Thermo Fisher Scientific DensityPRO+ Gamma Density System with Remote Transmitter User Guide P/N 717819

More information

FLT93 Installation, Operation and Troubleshooting Guide

FLT93 Installation, Operation and Troubleshooting Guide FLT93 Installation, Operation and Troubleshooting Guide Pre-Installation A. To get the best results from the instrument, the instrument should be mounted 20 pipe diameters downstream from any valve, pipe

More information

Nuclear Density Gauges

Nuclear Density Gauges Nuclear Density Gauges DensityPRO DensityPRO+ DensityPRO-C ConsistencyPRO CutPRO 3680 Authorised Distributor www.transmin.com.au Thermo Scientific Nuclear Density Gauges Energy Attenuation Gauges The expandable

More information

Sapphire. Ultrasonic Point Level System User Guide P/N Part of Thermo Fisher Scientific. Revision C

Sapphire. Ultrasonic Point Level System User Guide P/N Part of Thermo Fisher Scientific. Revision C Sapphire Ultrasonic Point Level System User Guide P/N 900-000068 Revision C Part of Thermo Fisher Scientific Sapphire Ultrasonic Point Level System User Guide P/N 900-000068 Revision C 1992 Thermo Fisher

More information

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE DSGH Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE Revision History Revision History Version of manual Description Date 1.0 Initial release 051025 1.1

More information

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D

DensityPRO C. Gamma Density System User Guide P/N Part of Thermo Fisher Scientific. Revision D DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D Part of Thermo Fisher Scientific DensityPRO C Gamma Density System User Guide P/N 900-748300 Revision D 2007 Thermo Fisher Scientific

More information

THERMAL BUILDING SOLUTIONS EN-TraceTekTTSIM1A-IM-H /16

THERMAL BUILDING SOLUTIONS EN-TraceTekTTSIM1A-IM-H /16 TraceTek TTSIM-1A TraceTek Sensor Interface Module with Relay Installation/OPERATION Instructions Approvals and Certifications TYPE NM General Signaling Equipment 76LJ Only AC versions are UL listed and

More information

Victoreen Primalert 35 Area Radiation Monitor

Victoreen Primalert 35 Area Radiation Monitor Victoreen 05-437 Primalert 35 Area Radiation Monitor Operators Manual March 2005 Manual No. 126011 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

SAFETY INFORMATION AND WARNINGS

SAFETY INFORMATION AND WARNINGS This manual refers to the Model SST-3 control panel manufactured since October 31, 2013, which uses a universal (100 277 VAC; 50/60 Hz) power supply. Older units use a voltage-specific power supply and

More information

MODEL 8144-RD CLOCK SELECTOR/DISTRIBUTION AMPLIFIER INSTRUCTION MANUAL. SPECTRACOM CORPORATION 95 Methodist Hill Drive, Suite 500 Rochester, NY 14623

MODEL 8144-RD CLOCK SELECTOR/DISTRIBUTION AMPLIFIER INSTRUCTION MANUAL. SPECTRACOM CORPORATION 95 Methodist Hill Drive, Suite 500 Rochester, NY 14623 MODEL 8144-RD CLOCK SELECTOR/DISTRIBUTION AMPLIFIER INSTRUCTION MANUAL SPECTRACOM CORPORATION 95 Methodist Hill Drive, Suite 500 Rochester, NY 14623 PHONE 585-321-5800 FAX 585-321-5218 REVISIONS, IF ANY,

More information

Duct and Rough Service Carbon Monoxide Sensor

Duct and Rough Service Carbon Monoxide Sensor Product Identification and Overview Duct and Rough Service Carbon Monoxide Sensor BAPI s Carbon Monoxide Sensor offers enhanced electrochemical sensing with outstanding accuracy at low concentrations.

More information

Greentrol Automation, Inc.

Greentrol Automation, Inc. Model GA-100 Technical Manual GreenAlert 100 Series Greentrol Automation, Inc. Installation, Operation and Maintenance Technical Manual GA-100 Document Name: TM_GA-100_R1C TM_GA-100_R1C Part Number: 960--0000

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS... SECTION Table of Contents SECTION 1 INTRODUCTION...................... 1.1 Description............................... 1.2 Features................................. 1.3 Models..................................

More information

Informer100 Speaker Models I-IP100AC and I-IP100DC

Informer100 Speaker Models I-IP100AC and I-IP100DC Informer100 Speaker Models I-IP100AC and I-IP100DC IP-Enabled High-Powered Indoor/Outdoor Speaker Description, Specifications, and Manual 25500387 Rev. A3 0319 Printed in U.S.A. Copyright 2017-2019 Federal

More information

DETECTOR TEST LOG. Detector Identification Information DHX-501 INTELLIGENT AIR DUCT SMOKE DETECTOR HOUSING INSTALLATION AND MAINTENANCE INSTRUCTIONS

DETECTOR TEST LOG. Detector Identification Information DHX-501 INTELLIGENT AIR DUCT SMOKE DETECTOR HOUSING INSTALLATION AND MAINTENANCE INSTRUCTIONS TEST LOG Detector Identification Information Manufacturer and Serial Date Detector Model: Number: Installed: Description of Detector Location: Test Results and Maintenance Data Date Test Test Maintenance

More information

21-light Remote Annunciator. Owner s Manual

21-light Remote Annunciator. Owner s Manual 21-light Remote Annunciator Owner s Manual Annunciator Description... Inside Font Cover Detailed Specifications... 1 Environmental Specifications... 1 Power Supply Requirements... 1 Communication With

More information

TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS

TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS TTSIM-2 TRACETEK SENSOR INTERFACE MODULE WITH LCD AND RELAY INSTALLATION/OPERATION INSTRUCTIONS GENERAL INFORMATION Please read these instructions and keep them in a safe place. These instructions must

More information

4100S. The operation and functions described in this manual are available from Software Version 4100S onwards.

4100S. The operation and functions described in this manual are available from Software Version 4100S onwards. 4100S The operation and functions described in this manual are available from Software Version 4100S-023-013 onwards. Specifications: Item Specification Details Enclosure Steel IP30 Dimensions H x W x

More information

Table of Contents. What to Expect with. Mounting Options. Tools Needed

Table of Contents.   What to Expect with. Mounting Options. Tools Needed Table of Contents www.hunterfan.com What to Expect with Your Installation Congratulations on purchasing your new Hunter ceiling fan! It will provide comfort and performance in your home or office for many

More information

Ronan Model X96S Series Density System

Ronan Model X96S Series Density System Ronan Model X96S Series Density System For more than 27 years Ronan has provided the process control industry with unsurpassed products in the harshest environments. With a philosophy of designing new

More information

3MAVR60XXD Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 3MAVR60XXD

3MAVR60XXD Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 3MAVR60XXD Owner s Guide and Installation Manual 3MAVR60XXD Series Fan CUL Model NO. : 3MAVR60XXD Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

Beacon 410A Gas Monitor Operator s Manual

Beacon 410A Gas Monitor Operator s Manual Beacon 410A Gas Monitor Operator s Manual Part Number: 71-0397 Revision: F Released: 12/5/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be

More information

Installing a Fast Oven Upgrade Kit

Installing a Fast Oven Upgrade Kit Agilent 6850 Series II Network GC System Accessory G3350B There are three Fast Oven upgrade kits: 120 V, 200 V, and 230 V. The following table indicates what each kit contains: Description Quantity 120

More information

3MAVR60XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MAVR60XX

3MAVR60XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MAVR60XX Owner s Guide and Installation Manual 3MAVR60XX Series Fan UL Model NO. : 3MAVR60XX Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea

Lightning Warning System: LWS MK-II. User's Guide. KOREA EMI Technologies Co.,Ltd , Jangam-ri, Majang-myeon, Icheon-si, Gyeonggi-do, Korea User's Guide CONTENTS FEATURES WHAT YOU NEED TO GET STARTED INSTALLATION DETAILS Installing the Antenna Installing the External Alert Siren The main console OPERATING PROCEDURE Power On Routine Push Buttons

More information

DIGITAL METERS Large Display Temperature Meters Instruction Manual

DIGITAL METERS Large Display Temperature Meters Instruction Manual DIGITAL METERS Large Display Temperature Meters PD755 PD756 PD757 Handles Thermocouple & RTD Inputs with Simplicity J, K, T, E, R, & S Thermocouples 100 Ω Platinum RTD (0.00385 or 0.00392 curve) Large

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS The SensorNet I/O Module provides attachment for up to 32 alarms and up to 16 switched devices. An RS-232 or RS-422 communications port is also provided for receiving serial alarms.

More information

Eldex Column Heater. Operator s Manual

Eldex Column Heater. Operator s Manual Eldex Eldex Column Heater Operator s Manual Eldex Laboratories, Inc. 30 Executive Court Napa, CA 94558 Tel: (707) 224-8800 Fax: (707) 224-0688 www.eldex.com Rev. B: 081700 2000Eldex Laboratories, Inc.

More information

Installation and Operating Instructions

Installation and Operating Instructions The Leader of RF and Sonic Level For Assistance Call 1-800-527-6297 Outside North America 1-215-674-1234 Installation and Operating Instructions Series 504-1200 Ultrasonic VeriGAP Switch using 404-1200

More information

Installing the Cisco ONS FAP-4 Fuse Alarm Panel

Installing the Cisco ONS FAP-4 Fuse Alarm Panel Installing the Cisco ONS 15454-FAP-4 Fuse Alarm Panel Product Number: 15454-FAP-4= This document explains how to install, test, operate, and maintain the Cisco ONS 15454-FAP-4 fuse alarm panel. This document

More information

TTSIM-1A. TraceTek Sensor Interface Module with Relay. Installation/Operation Instructions. Installation Items (not supplied) Tools Required.

TTSIM-1A. TraceTek Sensor Interface Module with Relay. Installation/Operation Instructions. Installation Items (not supplied) Tools Required. TTSIM-1A TraceTek Sensor Interface Module with Relay Installation Items (not supplied) General Information Installation/Operation Instructions Please read these instructions and keep them in a safe place.

More information

WAILEA OWNER S MANUAL

WAILEA OWNER S MANUAL WAILEA OWNER S MANUAL The blades in each pack are matched for equal weight to assure smooth fan operation. If more than one fan is being installed, be careful not to mix blades from different cartons.

More information

installation and operation manual for Hunter Ceiling Fans

installation and operation manual for Hunter Ceiling Fans For Your Records and Warranty Assistance Model Name: Catalog/Model No.: Serial No.: Date Purchased: Where Purchased: For reference also attach your receipt or a copy of your receipt to the manual. installation

More information

Flame Rod Manual. Publication Rev A

Flame Rod Manual. Publication Rev A Publication INTRODUCTION This manual contains information for the Flame Rod Flame Detection system from Forney Corporation, 16479 North Dallas Parkway, Suite 600 Addison, TX 75001. All personnel should

More information

NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions

NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions NGC-UIT-ORD User interface terminal for Raychem NGC systems Installation Instructions Description The NGC-UIT-ORD is a panel mounted display used in conjunction with other approved Raychem control and

More information

INSTRUCTION MANUAL (ATEX/IECEx/SIL2) BExBG05D-SIL Flameproof Xenon SIL 2 Beacons For use in Flammable Gas and Dust Atmospheres

INSTRUCTION MANUAL (ATEX/IECEx/SIL2) BExBG05D-SIL Flameproof Xenon SIL 2 Beacons For use in Flammable Gas and Dust Atmospheres INSTRUCTION MANUAL (ATEX/IECEx/SIL2) BExBG05D-SIL Flameproof Xenon SIL 2 Beacons For use in Flammable Gas and Dust Atmospheres 1) Introduction The BExBG05D-SIL is a flameproof beacon which is certified

More information

3MNLR72XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MNLR72

3MNLR72XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MNLR72 Owner s Guide and Installation Manual 3MNLR72XXD Series Fan UL Model NO. : 3MNLR72 Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm Operating & Maintenance Manual Alert-4 Ethernet LCD Master Alarm w w w. a m i c o. c o m Contents User Responsibility 4 Introduction 4 Features 5 Description of the Alarm 5 Shipment Details 5 The Alarm

More information

3SKYR56XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 3SKYR56(XXXXX)

3SKYR56XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 3SKYR56(XXXXX) Owner s Guide and Installation Manual 3SKYR56XXD Series Fan ETL Model NO. : 3SKYR56(XXXXX) Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER

More information

SLATE. Base Module INSTALLATION INSTRUCTIONS R8001A1001

SLATE. Base Module INSTALLATION INSTRUCTIONS R8001A1001 SLATE Base Module R8001A1001 INSTALLATION INSTRUCTIONS Scan for more information Application SLATE brings configurable safety and programmable logic together into one single platform. The platform can

More information

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual We manage heat MANUAL Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual 1850 N Sheridan Street South Bend, Indiana 46628 (574) 233-1202 or (800) 234-4239

More information

Remote Vacuum Sensor and Variable Speed Vacuum Pump Control Manual. Version Date - June Part Number

Remote Vacuum Sensor and Variable Speed Vacuum Pump Control Manual. Version Date - June Part Number Innovation In and Out of Parlour Remote Vacuum Sensor and Variable Speed Vacuum Pump Control Manual Version - 1.0 Date - June 2016 Part Number - 39-0038 Index Manual Version... 3 About the Remote Vacuum

More information

Juna Ceiling Fan JU54. Installation Instructions Instrucciones de instalación Instructions d installation

Juna Ceiling Fan JU54. Installation Instructions Instrucciones de instalación Instructions d installation Juna Ceiling Fan JU54 Installation Instructions Instrucciones de instalación Instructions d installation Read and Save These Safety Precautions. Turn off electricity at main switch before wiring or servicing

More information

CP CP

CP CP CP-8161-333 CP-8161-433 Electronic Programmable Controller Six Stage, Dual Setpoint General Instructions APPLICATION Electronic six stage programmable controller with proportional output for heating, cooling

More information

5LCM52XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : AC-552AL

5LCM52XX Series Fan. Owner s Guide and Installation Manual. UL Model NO. : AC-552AL Owner s Guide and Installation Manual 5LCM52XX Series Fan UL Model NO. : AC-552AL Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME: RETAILER

More information

4100U City and Relay Cards Installation Instructions

4100U City and Relay Cards Installation Instructions 4100U City and Relay Cards Installation Instructions Introduction This publication describes the installation procedure for the following: 4100-6031/6032 City Circuit Cards 4100-6033 Alarm Relay Card Inspecting

More information

3AVOR56XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 3AVOR56XXXXX

3AVOR56XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 3AVOR56XXXXX Owner s Guide and Installation Manual 3AVOR56XXD Series Fan ETL Model NO. : 3AVOR56XXXXX Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER

More information

Public Safety DAS Annunciator Panel

Public Safety DAS Annunciator Panel Public Safety DAS Annunciator Panel 120 VAC Models: 1221-A, 1221-B, 1221-C Revision D 91117 48 VDC Models: 1221-A-48, 1221-B-48, 1221-C-48 24 VDC Models: 1221A-24, 1221-B-24, 1221-C-24 CAUTION: (Read This

More information

Installation Instructions

Installation Instructions Installation Instructions Full Size Tumble Action Washers Before beginning installation, carefully read these instructions. This will simplify the installation and ensure the washer is installed correctly

More information

Instruction Manual. B Issue B. next Pump Accessories

Instruction Manual. B Issue B. next Pump Accessories Instruction Manual B811-00-880 Issue B next Pump Accessories This page has been intentionally left blank. Contents Section Page 1 Introduction... 1 Contents gea/0086/06/10 1.1 Scope and definitions...

More information

3DIR52XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3DIR52XXD

3DIR52XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3DIR52XXD Owner s Guide and Installation Manual 3DIR52XXD Series Fan UL Model NO. : 3DIR52XXD Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

ANNUNCIATOR FIXED MODULE

ANNUNCIATOR FIXED MODULE 12 Clintonville Road Northford, CT 06472 Phone: 203-484-7161 Fax: 203-484-7118 THE ANNUNCIATOR FIXED MODULE Installation Manual for the AFM-16ATF and AFM-32AF Annunciator Modules Document # 15970 3/27/95

More information

SC-6 Six Supervised Control Module

SC-6 Six Supervised Control Module INSTALLATION AND MAINTENANCE INSTRUCTIONS SC-6 Six Supervised Control Module SPECIFICATIONS Normal Operating Voltage: Stand-By Current: Alarm Current: Temperature Range: Humidity: Dimensions: Accessories:

More information

TILGHMAN CEILING FAN. LISTED For Damp Location E ITEM # MODEL #WCK52LMW5N WCK52NWZ5N. Español p. 20 ATTACH YOUR RECEIPT HERE

TILGHMAN CEILING FAN. LISTED For Damp Location E ITEM # MODEL #WCK52LMW5N WCK52NWZ5N. Español p. 20 ATTACH YOUR RECEIPT HERE ITEM #0294980 0294981 TILGHMAN CEILING FAN Harbor Breeze is a registered trademark of LF, LLC. All Rights Reserved. MODEL #WCK52LMW5N WCK52NWZ5N Español p. 20 ATTACH YOUR RECEIPT HERE Serial Number Purchase

More information

Dryer. User manual DV22K6800** DV22K A-00_EN (US)_ indd :15:41

Dryer. User manual DV22K6800** DV22K A-00_EN (US)_ indd :15:41 Dryer User manual DV22K6800** DV22K6800-03650A-00_EN (US)_151211.indd 1 2015-12-11 7:15:41 Before installation Read through the following instructions before installing the dryer, and keep this manual

More information

Snifter ATEX22 VERSION. User Manual. Distributor

Snifter ATEX22 VERSION. User Manual. Distributor Snifter ATEX22 VERSION User Manual Distributor Version 1.4 09/09/2009 Table of Contents 1. INTRODUCTION... 3 1.1. Safety... 3 1.2. Product overview... 4 1.3. How does it work?... 4 2. INSTALLATION... 5

More information

5YK60XXX Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 5YK60XXX

5YK60XXX Series Fan. Owner s Guide and Installation Manual. CUL Model NO. : 5YK60XXX Owner s Guide and Installation Manual 5YK60XXX Series Fan CUL Model NO. : 5YK60XXX Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP

INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP RV Products A Division of Airxcel, Inc. P.O. Box 4020 Wichita, KS 67204 1976-360 (1-02) PP TABLE OF CONTENTS 1. Warnings......................................................

More information

Table of Contents. What to Expect with. Tools Needed. Mounting Options. Blades

Table of Contents.   What to Expect with. Tools Needed. Mounting Options. Blades Table of Contents 1.888.830.1326 Ceiling Bracket 30 inches Ladder Downrod 3 Wiring 10 Operation, Maintenance & Cleaning Light Kit 11 Troubleshooting??? 17 15 1 13 5 Blades 9 Switch Housing 12 Canopy 6

More information

52 BAYTOWNE CEILING FAN

52 BAYTOWNE CEILING FAN 52 BAYTOWNE CEILING FAN Owner s Manual Models #20501 If a problem cannot be remedied or you are experiencing difficulty in installation, please contact the Service Department: 1-877-459-3267, 9 a.m.- 5

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

Pipe Freeze Protection Control SCFP-CO-F130 Installation and Operation Manual

Pipe Freeze Protection Control SCFP-CO-F130 Installation and Operation Manual MANUAL Pipe Freeze Protection Control SCFP-CO-F130 Installation and Operation Manual Model FPT 130 Single Point Freeze Protection Heat Trace Control Table of Contents SCFP-CO-F130 Overview... 3 Installation...

More information

TOUCHDOWN 48 CEILING FAN

TOUCHDOWN 48 CEILING FAN TOUCHDOWN 48 CEILING FAN MODEL #50205 Español p. 20 Questions, problems, missing parts? Before returning to your retailer, call our customer service department at 1-877-361-3883, Monday - Thursday, 8 am

More information

Installation Instructions Dual Fuel Ranges

Installation Instructions Dual Fuel Ranges Installation Instructions Dual Fuel Ranges E30DF74EPS E36DF76EPS E48DF76EPS 5995447082 2 Safety IMPORTANT SAFETY INSTRUCTIONS Safety Precautions Do not attempt to install or operate your unit until you

More information

OPTISENS TUR 2000 Technical Datasheet

OPTISENS TUR 2000 Technical Datasheet OPTISENS TUR 2000 Technical Datasheet Sensor for turbidity measurement in water and wastewater Rugged design for harsh applications Integrated transmitter with direct 4...20 ma output Near infrared light

More information

RK-05 Combustible Gas Transmitter Operator s Manual

RK-05 Combustible Gas Transmitter Operator s Manual 65-2405RK-05 Combustible Gas Transmitter Operator s Manual Part Number: 71-0180RK Revision: 0 Released: 2/16/11 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction

More information

Table of Contents. What to Expect with. Tools Needed. Mounting Options. Blades

Table of Contents.  What to Expect with. Tools Needed. Mounting Options. Blades www.hunterfan.com Table of Contents Ceiling Bracket 30 inches Downrod Ladder 3 Wiring Operation, Maintenance & Cleaning Light Kit 11 Troubleshooting??? 16 15 1 12 10 9 5 Blades Canopy 6 4 PÁGINA 2 7 feet

More information

SAVE THESE INSTRUCTIONS

SAVE THESE INSTRUCTIONS Built-In Dishwasher Dishwashers Write the model and serial numbers here: Model # Serial # You can find them on the tub wall just inside the door or. the lower part of back. Installation Instructions DDW1802W

More information

MAYFIELD CEILING FAN LISTED E ITEM # MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE.

MAYFIELD CEILING FAN LISTED E ITEM # MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE. Harbor Breeze is a registered trademark of LF, LLC. All Rights Reserved. ITEM #0331094 0331096 MAYFIELD CEILING FAN MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE Serial Number Purchase

More information

52 STRATHMERE CEILING FAN

52 STRATHMERE CEILING FAN 52 STRATHMERE CEILING FAN Owner s Manual Models #20341 If a problem cannot be remedied or you are experiencing difficulty with installation, please contact the Service Department: 1-877-459-3267, 9 a.m.-

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

14PRR62XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 14PRR62(X)

14PRR62XXD Series Fan. Owner s Guide and Installation Manual. ETL Model NO. : 14PRR62(X) Owner s Guide and Installation Manual 14PRR62XXD Series Fan ETL Model NO. : 14PRR62(X) Attach sales receipt to this card and retain as your proof of purchase DATE OF PURCHASE: MODEL NUMBER: RETAILER NAME:

More information

TECH SHEET DO NOT DISCARD PAGE 1. If unsuccessful entry into diagnostic mode, actions can be taken for specific indications:

TECH SHEET DO NOT DISCARD PAGE 1. If unsuccessful entry into diagnostic mode, actions can be taken for specific indications: TECH SHEET DO NOT DISCARD PAGE 1 WARNING Electrical Shock Hazard Disconnect power before servicing. Replace all parts and panels before operating. Failure to do so can result in death or electrical shock.

More information

Installation Guide. For Models: E-MLV36BNK4LK1 E-MLV36ESP4LK1 E-MLV36WW4LK1 READ THESE INSTRUCTIONS AND AND SAVE THEM FOR FUTURE USE

Installation Guide. For Models: E-MLV36BNK4LK1 E-MLV36ESP4LK1 E-MLV36WW4LK1 READ THESE INSTRUCTIONS AND AND SAVE THEM FOR FUTURE USE READ THESE INSTRUCTIONS AND AND SAVE THEM FOR FUTURE USE Federal regulations require ceiling fans with light kits manufactured or imported after January 1, 2009, to limit total wattage consumed by the

More information

ELSTON 52 CEILING FAN

ELSTON 52 CEILING FAN ELSTON 52 CEILING FAN MODEL #10290 Español p. 21 Questions, problems, missing parts? Before returning to your retailer, call our customer service department at 1-877-361-3883, Monday - Thursday, 8 am -

More information

MODELS 304GC, 304GCX, 314GC, and 314GCX

MODELS 304GC, 304GCX, 314GC, and 314GCX 2562109A REV. A1 0317 Printed in U.S.A. MODELS 304GC, 304GCX, 314GC, and 314GCX INSTALLATION AND SERVICE INSTRUCTIONS FOR SELECTONE Limited Warranty This product is subject to and covered by a limited

More information

48 TAYLOR CEILING FAN

48 TAYLOR CEILING FAN 48 TAYLOR CEILING FAN Owner s Manual Models #20554 If a problem cannot be remedied or you are experiencing difficulty in installation, please contact the Service Department: 1-877-459-3267, 9 a.m.- 5 p.m.

More information

52 BARSTOW CEILING FAN

52 BARSTOW CEILING FAN 52 BARSTOW CEILING FAN Owner s Manual Models #20519 If a problem cannot be remedied or you are experiencing difficulty with installation, please contact the Service Department: 1-877-459-3267, 9 a.m.-

More information

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL

MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL TRACON MODEL FPT-130 SINGLE POINT FREEZE PROTECTION HEAT TRACE CONTROL TABLE OF CONTENTS FPT 130 Overview... 2 Installation... 3 Power Source and Load Connections... 4 Temperature Sensor... 5 External

More information

COMSPHERE 48 Vdc CENTRAL OFFICE POWER UNIT

COMSPHERE 48 Vdc CENTRAL OFFICE POWER UNIT COMSPHERE 48 Vdc CENTRAL OFFICE POWER UNIT INSTALLATION GUIDE Document No. 3000-A2-GB41-50 October 2002 COMSPHERE 48 Vdc Central Office Power Unit COMSPHERE 48 Vdc Central Office Power Unit Installation

More information

Table of Contents What to Expect with Your Installation. Top Housing. Ceiling Plate. Tools Needed.

Table of Contents What to Expect with Your Installation. Top Housing. Ceiling Plate. Tools Needed. Table of Contents Congratulations on purchasing your new Hunter ceiling fan! It will provide comfort and performance in your home or office for many years. This installation and operation manual contains

More information

Analog Room Pressure Monitor RPC Series

Analog Room Pressure Monitor RPC Series Description The Room Pressure Monitor is used to measure differential pressure in the range of 0.125 to 1"wc or 30 to 250 Pa. It combines precision high sensitivity silicon sensing capabilities and the

More information

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg. WARD OXYGEN Installation, Operation, and Maintenance Instructions MEGA medical electronic gas alarm MEGA WARD MEGA medical electronic gas alarm WARD POWER SUPPLY Medical Electronic Gas Alarm OXYGEN SYSTEM

More information

RKA-LEL Sample Draw Combustible Gas Detector

RKA-LEL Sample Draw Combustible Gas Detector 35-3000RKA-LEL Sample Draw Combustible Gas Detector Specifications Table 1 lists specifications for the sample draw combustible gas detector. Table 1: Specifications Target Gas Input Power Combustible

More information

ProStar Ceiling Fan PS52. Installation Instructions Instrucciones de instalación Instructions d installation

ProStar Ceiling Fan PS52. Installation Instructions Instrucciones de instalación Instructions d installation ProStar Ceiling Fan PS52 Installation Instructions Instrucciones de instalación Instructions d installation Read and Save These Safety Precautions. Turn off electricity at main switch before wiring or

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TT-1343 5/06b INSTALLATION INSTRUCTIONS Original Issue Date: 8/03 Model: Automatic Transfer Switches Equipped with Series 1000 Programmable Controller Market: ATS Subject: Remote Annunciator Kits GM28938-KP1,

More information

Owner s Guide and Installation Manual

Owner s Guide and Installation Manual For Your Records and Warranty Assistance For reference, also attach your receipt or a copy of your receipt to the manual. Model Name Type 2 Models Owner s Guide and Installation Manual Model No. Catalog

More information

BB TM 508/509 ANTI TWO-BLOCK ALARM SYSTEM HYDRAULIC CRANE INSTALLATION. Greer Company Crane Systems. Greer Company

BB TM 508/509 ANTI TWO-BLOCK ALARM SYSTEM HYDRAULIC CRANE INSTALLATION. Greer Company Crane Systems. Greer Company BB TM 508/509 ANTI TWO-BLOCK ALARM SYSTEM HYDRAULIC CRANE INSTALLATION GREER COMPANY 1918 East Glenwood Place, Santa Ana, CA 92705 Tel: (714) 259-9702 FAX (714) 259-7626 BB TM 508/509 ANTI TWO-BLOCK ALARM

More information

/9308 External Battery Chargers Installation Instructions

/9308 External Battery Chargers Installation Instructions 4081-9306/9308 External Battery Chargers Installation Instructions Introduction This publication describes the installation procedure for an External Battery Charger used with the 4100U or 4100ES System.

More information

Colibri. Automatic Tank Gauge CL6 Series Installation Guide

Colibri. Automatic Tank Gauge CL6 Series Installation Guide Colibri Automatic Tank Gauge CL6 Series Installation Guide Manual # Revision Date Changes from Previous Revision 000-2153 C Jan. 2013 Pipe compound brand change, Updated Control drawing. Franklin Fueling

More information

INSTALLATION INSTRUCTIONS FOR 6536 SERIES TWO TON PACKAGED HEAT PUMP

INSTALLATION INSTRUCTIONS FOR 6536 SERIES TWO TON PACKAGED HEAT PUMP INSTALLATION INSTRUCTIONS FOR 6536 SERIES TWO TON PACKAGED HEAT PUMP TABLE OF CONTENTS 1. Warnings...2 2. Component Match-Up...2 3. Unit Depiction Figures...3 4. Blower Performance Data...5 5. General

More information

SONAR PROCESS MONITORING SYSTEM SUPPLEMENT FOR ATEX CLASS I, ZONE 2 SAFETY

SONAR PROCESS MONITORING SYSTEM SUPPLEMENT FOR ATEX CLASS I, ZONE 2 SAFETY SONAR PROCESS MONITORING SYSTEM SUPPLEMENT FOR ATEX CLASS I, ZONE 2 SAFETY Table of Contents 1 SCOPE...1-1 2 MARKINGS...2-1 2.1 Transmitter [TB8-xx-xx-xx-03] for use with Sensor Head [SH-xxx-xx-xA-xxx-03]:...2-1

More information

1. SAFETY RULES WARNING WARNING. 8. Avoid placing objects in the path of the blades.

1. SAFETY RULES WARNING WARNING. 8. Avoid placing objects in the path of the blades. 1 1. SAFETY RULES 1. To reduce the risk of electric shock, insure electricity has been turned off at the circuit breaker or fuse box before beginning. 2. All wiring must be in accordance with the National

More information

52 SAN LUCAS CEILING FAN

52 SAN LUCAS CEILING FAN 52 SAN LUCAS CEILING FAN Owner s Manual Models #20551 If a problem cannot be remedied or you are experiencing difficulty in installation, please contact the Service Department: 1-877-459-3267, 9 a.m.-

More information

Camarillo 52 Ceiling Fan

Camarillo 52 Ceiling Fan Owner s Manual Camarillo 52 Ceiling Fan Part # 269263, 269259, 269287 Model # 32091, 32092, 32087 Exclusively Distributed by: HD Supply Facilities Maintenance, Ltd. Atlanta, GA 30339 2017 Made in China

More information