Model gffoz Process Ozone Sensor (In-Line)

Size: px
Start display at page:

Download "Model gffoz Process Ozone Sensor (In-Line)"

Transcription

1 Date ECO# Rev Description Originator 9/15/ First Publication DB 10/8/ Miscelaneous corrections DB 10/28/ Minor changes BM 5/1/ Added CE compliance BM TM IN USA, INCORPORATED 87 CRESCENT ROAD NEEDHAM, MA USA TEL: FAX: Model gffoz Process Ozone Sensor (In-Line) (software versions 1.02 and higher)

2 Contents CONTENTS... 2 FIGURES... 3 TABLES... 3 GENERAL PRECAUTIONS... 4 GFFOZ SENSOR DESCRIPTION... 6 GFFOZ SPECIFICATIONS... 8 INSTALLATION AND CONNECTIONS... 9 Mounting...9 Pneumatic Connections and Configuration...9 Electrical Connections to the IN USA, INC. Controller...12 OPERATION Initial Power-on and Warm-up...13 Zeroing the Sensor...13 MAINTENANCE Removing the Sensor Cover...15 Accessing the UV Lamp...15 CONTACTING IN USA, INC About the Company...17

3 Figures FIGURE 1: GFFOZ SENSOR...6 FIGURE 2: GFFOZ DIMENSIONS AND MOUNTING HOLE PATTERN...6 FIGURE 3: IN-LINE, FULL-FLOW CONFIGURATION...10 FIGURE 4: SIDE STREAM CONFIGURATIONS...11 FIGURE 5: UV LAMP ACCESS...16 Tables TABLE 1: ZEROING PHASES...14

4 General Precautions WARNINGS: 1. Ozone (O 3 ) is a toxic gas. High concentrations of ozone are dangerous and harmful to humans. Take reasonable steps to avoid exposure. The current maximum 8-hour exposure limit for ozone is 0.1 ppm (according to U.S. OSHA). 2. Install appropriate safety monitoring equipment wherever high concentrations of ozone are used. IN USA, INC. manufactures several ozone monitors for workplace safety applications. 3. Materials in contact with high concentrations of ozone should be suitable for such use. 316L Stainless, Teflon, Chemraz and Kynar are recommended. 4. When performing any maintenance to the unit, make sure all AC power is disconnected from the unit. 5. The ozone sensor contains an ultraviolet (UV) lamp. Never remove the lamp from its housing while the unit is powered on. Eye exposure to UV light is extremely dangerous. 6. Certain components may be hot to the touch. Please allow proper cooling time before working with these components. 7. Never attempt to open ozone catalyst canisters (if supplied). The content of the canisters can be hazardous if not handled properly. 8. Use only IN USA, INC. recommended spare parts. Substitution parts could result in damage to the equipment, may create hazardous conditions and will void the warranty. CAUTIONS: 1. Read the operating manual before operating the unit. 2. Do not subject the unit to extreme physical or thermal shock. 3. Use care in handling the unit and any of its components.

5 GENERAL NOTES 1. IN USA, AFX, gffoz, SCI and Excellence in Instrumentation are Trade Marks of IN USA, INC. 2. This document is Copyright Protected. 3. IN USA, INC. reserves the right to make changes to the product covered in this manual to improve performance, reliability or manufacturability. Make sure that this Manual is used with the original Product it was shipped with. 4. Although every effort has been made to ensure accuracy of the information contained in this manual, IN USA assumes no responsibility for inadvertent errors. Contents of the manual are subject to change without notice. 5. IN USA assumes no responsibility for the use of any measuring schemes described herein. 6. This product is not intended or recommended by IN USA for use in (a) medical therapy or physical therapy of any kind whether as a direct or adjunct part of such therapy, including, without limitation, life support (i.e., critical medical) applications or (b) any nuclear facility applications. IN USA will not knowingly sell this product for use in such applications. Use of the IN USA product in connection with medical or like treatment cannot be reasonably expected to produce accurate monitorings of therapy or treatment and may cause failure of the life support device or significantly affect its safety or effectiveness. Use by any direct purchaser or after-market purchaser in such applications whether or not known to IN USA shall absolve IN USA of any responsibility or liability to such purchaser (s) or to any person (s) subjected to or affected by such use knowingly or unknowingly.

6 gffoz Sensor Description IN USA, INC's. gffoz* sensor is designed for measuring process ozone in the gas phase at concentrations up to 400 g/m 3. The sensor can be installed directly in-line in a full-flow stream at flow rates up to 15 SLPM or more. The gffoz is clean-ready for high ultra-pure applications. The gffoz is a state-of-the-art smart sensor incorporating IN USA, INC's. innovative dual optical path sample sensing technology (patent pending) to provide unparalleled performance. Calibration data is stored on-board, making sensor replacement easy. gffoz sensors must be connected to one of IN USA, INC's. Sensor Control Interface controllers (series "SCI"). Some of these controllers interface to multiple sensors and can be equipped with ozone generator servo controls. Please refer to the controller's operating manual for details. SCREWS FOR COVER (QTY 2) Figure 1: gffoz Sensor 15 PIN "D" CONNECTOR POWER ON INDICATOR LED INLET PORT OUTLET PORT *gffoz = gas phase, Full Flow OZone sensor. Figure 2: gffoz Dimensions and Mounting Hole Pattern

7

8 gffoz Specifications Measuring Principle Application Pressure and Temperature UV Light Source UV absorption; dual optical path sample sensing (patent pending) Continuous in-situ measurement of high concentrations of ozone in gas phase Automatic compensation for pressure and temperature Low pressure mercury vapor lamp Units of Measure % weight, g/m 3, g/nm 3 Measuring Range Repeatability Sample Flow Sample Pressure Sample Ports Wetted Materials Dimensions (W x H x D) Maintenance Compliance g/m 3 (0-26% weight in O 2 at NTP) 1.0 g/m 3 or 1% of reading, including zero drift (higher precision available) 0.2 to 15 SLPM. Pressure drop at 10 SLPM is less than 5 mb 40 psia operating; 80 psia proof 1/4" VCR standard Electropolished 316L, sapphire. Compatible with semiconductor processes 5.1" x 5.8" x 2.1" [130mm x 146 mm x 54mm] UV lamp replacement every months. No other scheduled maintenance CE

9 Installation and Connections Mounting The gffoz sensor has threaded mounting holes on its bottom plate. The hole pattern is shown in Figure 2, page 6. Use #8-32 screws for mounting. Pneumatic Connections and Configuration IMPORTANT!!: Gas flow should be as indicated by the gas flow direction arrows on the sensor. The female VCR fitting is the gas inlet. The male VCR fitting is the gas outlet. To avoid leaks, tighten the VCR fittings per the manufacturer's recommendation. Protect the VCR fitting faces from scratches and contamination. Always use new stainless steel gaskets - never use other materials. Although the gffoz is primarily designed for in-line measurement as shown in Figure 3, it can also be installed in a side-stream configurations, as shown in Figure 4 (by-pass mode and return mode). When used in a by-pass mode, the sensor's outlet should be vented to a catalytic module to destroy all traces of ozone. Contact IN USA, INC. for the appropriate catalytic module. NOTE: The exhaust from the catalyst in by-pass mode can be vented to ambient. However, it should not be vented into a clean-room environment. A particulate filter is recommended on the sensor inlet in order to protect the sensor's optical components and prolong the life of the sensor and of the UV lamp.

10 Figure 3: In-Line, Full-Flow Configuration PARTICULATE FILTER (TYP µm) gffoz gffoz SENSOR FULL FLOW OZONE SENSOR PATENT PENDING FLOW OZONE LINE CAUTION!!: When installed in-line, the sensor operates at the pressure of the ozone supply line. Do not exceed the maximum operating pressure of the sensor.

11 PATENT PENDING PATENT PENDING Figure 4: Side Stream Configurations ZERO REFERENCE GAS gffoz BY-PASS MODE FULL FLOW OZONE SENSOR FLOW EXHAUST SIDE-STREAM SAMPLE GAS ( L/MIN) FILTER (TYP µm) IN USA OZONE CATALYST FLOW RESTRICTOR OR FIXED ORIFICE ISOLATION VALVE MAIN OZONE GAS STREAM (AT LINE PRESSURE) SIDE-STREAM SAMPLE GAS ( L/MIN) gffoz FULL FLOW OZONE SENSOR FLOW RETURN MODE FILTER (TYP µm) FILTER (TYP µm) FLOW RESTRICTOR OR FIXED ORIFICE ISOLATION VALVE MAIN OZONE GAS STREAM (AT LINE PRESSURE)

12 In side-stream configurations, the use of a fixed orifice flow restrictor (as shown in Figure 4) is recommended in order to achieve the desired side-stream flow rates. Side-stream flow rates are usually set as low as possible, and flow restrictors can be selected such that they produce the desired flow at the process gas pressure. Metering or needle valves are also acceptable. CAUTION!!: When installed in the "return" mode, the sensor operates at the pressure of the ozone supply line. Do not exceed the maximum operating pressure of the sensor. As shown in Figure 4, the by-pass mode can easily incorporate an external zero reference gas line to allow for convenient "on-demand" zeroing of the sensor. Electrical Connections to the IN USA, INC. Controller As mentioned earlier, the gffoz sensor must be connected to an IN USA, INC. "SCI" series controller using the cable supplied by IN USA, INC. The connection on the gffoz is through a 15 pin male "D" connector located on top of the unit (see Figure 1, page 6).

13 Operation Initial Power-on and Warm-up In order to achieve maximum performance and stability, it is strongly recommended to allow an overnight warm up period on initial startup or on a "cold" start. This warm up period should preferably be done while the unit is purging with ozone-free gas (typically oxygen). After this warm up period, the unit can be zeroed as explained below. NOTE: It is recommended that the sensor be powered on at all times (24 hours/day, 7 days/week). This will insure peak performance, minimal zero drift, and maximum system life. Zeroing the Sensor Because of its advanced, patent-pending design, the gffoz is a highly zero-stable instrument. However, as with any similar device, the sensor must be zeroed periodically with an ozone-free reference gas. The interval between zeros depends upon operating conditions, but it is recommended that the zero point be verified at least once every 4 weeks.

14 Zeroing always involves the following steps: Table 1: Zeroing Step Purging with an ozone-free reference gas (typically oxygen) Zeroing Purpose To get rid of ozone gas inside the optical chamber prior to zeroing To make internal parameter adjustments to reflect the true zero of the sensor The zeroing phase is initiated through the sensor controller (SCI), either remotely or via its keypad. Please refer to the operating manual for the controller. The recommended ozone-free reference (purge) gas is oxygen. Oxygen is readily accessible given that it is used as feed-gas for ozone generators. Nitrogen and clean dry air may also be used. On start-up, the unit should be warmed up and purged with a reference gas overnight prior to zeroing. Thereafter, if the unit is operating continuously, a reference for about 15 minutes is adequate.

15 Maintenance The gffoz sensor is designed to operate continuously with minimal maintenance requirements. Only the UV lamp will need to be replaced on a periodic basis (every months). Lamp replacement does not change calibration and can be done while the sensor is on line and while gas is flowing through it. After lamp replacement, re-zero unit following the after start-up zeroing procedure. The use of particulate filters on the gas inlet of the monitor is recommended in order to minimize the risk of optical component soiling. Removing the Sensor Cover The UV lamp can be replaced without disconnecting pneumatic lines and without interrupting the flow of gas through the sensor. The lamp is accessed by removing the cover with the two thumb screws as shown in Figure 1 on page 6 REMINDER: SAFETY FIRST!! Disconnect power to the sensor (power off) before removing the cover. Accessing the UV Lamp Once the cover is removed as described above, unplug the lamp and slide it up and out of its housing by loosening the lamp set-screw (see Figure 5 below ). To reinstall the lamp, follow the aforementioned steps in reverse order. NOTE: Some lamps may have marks indicating preferred rotational position. If so, install the lamp accordingly.

16 Figure 5: UV Lamp Access UV LAMP PLUG UV LAMP UV LAMP SET-SCREW

17 Contacting IN USA, INC. IN USA, INC. technical and service personnel are ready to help and can be reached by mail, , telephone, or fax. Address: IN USA, INC. 87 Crescent Road Needham, MA U.S.A. Telephone: (USA) Fax: (USA) (attn:customer service) About the Company IN USA, INC. manufactures instrumentation and control systems. The company is the world's leading manufacturer of ozone monitoring equipment. IN USA, INC. provides customized, optical-based monitoring systems which include sophisticated process control hardware and software.

OG5000 Ozone Generator Model OG5000B-A. User Manual

OG5000 Ozone Generator Model OG5000B-A. User Manual OG5000 Ozone Generator Model OG5000B-A Copyright 2017 610-0082-01H DCN7785 Teledyne API 25 September 2017 Contents CONTENTS... 1 FIGURES... 2 TABLES... 2 1. GENERAL PRECAUTIONS... 3 Safety... 5 Safety

More information

AC-Series Ozone Generators. User Manual

AC-Series Ozone Generators. User Manual AC-Series Ozone Generators User Manual Copyright 2017 610-0191-01D DCN7808 Teledyne API 22 December 2017 AC Series Ozone Generators User Manual Contents 1 SAFETY... 1 1.1 General Precautions... 1 1.2

More information

SERIES IN-2000 SINGLE AND MULTI-CHANNEL GASEOUS OZONE ANALYZERS

SERIES IN-2000 SINGLE AND MULTI-CHANNEL GASEOUS OZONE ANALYZERS TM INSTRUMENTATION R IN USA, INCORPORATED 87 CRESCENT ROAD NEEDHAM, MA 02194 USA TEL: 617-444-2929 FAX: 617-444-9229 SERIES IN-2000 SINGLE AND MULTI-CHANNEL GASEOUS OZONE ANALYZERS OPERATING MANUAL VERSION

More information

SEC Signature Process Gas Analyzer

SEC Signature Process Gas Analyzer SEC Signature Process Gas Analyzer Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 email sensor@minn.net

More information

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu)

SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) PART 1 - GENERAL A. RELATED DOCUMENTS SAMPLE SPECIFICATIONS CONTINUOUS MONITORING SYSTEM (Insitu) Federal, State, and local requirements as applicable. Attached Permit Attached drawings B. SUMMARY 1. Provide

More information

PUV, PIR and PFO Series BROCHURE

PUV, PIR and PFO Series BROCHURE PUV, PIR and PFO Series BROCHURE JUNHO 2013 ABB Multiwave TM process photometers PUV3402, PIR3502 and PFO3372 Measure vapor and liquid samples in IR, NIR, UV and VIS regions Process photometers PUV3402,

More information

RK Carbon Monoxide Transmitter Operator s Manual

RK Carbon Monoxide Transmitter Operator s Manual 65-2335RK Carbon Monoxide Transmitter Operator s Manual Part Number: 71-0177RK Revision: 0 Released: 4/12/11 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual

More information

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING FOR OPERATION FROM 115 AC POWER Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783

More information

DFM. flow pressure temperature MULTI PARAMETER DIGITAL MASS FLOW METERS. Totalizer. Interface

DFM. flow pressure temperature MULTI PARAMETER DIGITAL MASS FLOW METERS. Totalizer. Interface Multi Parameter flow meters provide accurate data on three different fluid parameters: flow pressure temperature The flow rate can be displayed in volumetric flow or mass flow engineering units for standard

More information

Model OEM-2 INSTRUCTIONS FOR USE

Model OEM-2 INSTRUCTIONS FOR USE ECO SENSORS, INC. 3-03.2 OEM OZONE CONTROLLER Model OEM-2 INSTRUCTIONS FOR USE GENERAL The model OEM-2 is a system to control ozone generators and alarms based on an adjustable ozone concentration set

More information

451 Black Forest Road Hull, IA Ph: (712) Fax: (712) TS g/hr Turnkey Ozone Generation System

451 Black Forest Road Hull, IA Ph: (712) Fax: (712) TS g/hr Turnkey Ozone Generation System 451 Black Forest Road Hull, IA 51239 Ph: (712) 439-6880 Fax: (712) 439-6733 TS-40 40 g/hr Turnkey Ozone Generation System Installation and Operation Manual TS-40 Manual Cautions, Warnings and Hazards

More information

Using the DewMaster Deluxe Dew Point/ Humidity Hygrometer

Using the DewMaster Deluxe Dew Point/ Humidity Hygrometer JAN 0 ::0 AM SERVOLOCK APPLICATION NOTE 0_0_ Rev Using the Deluxe Dew Point/ Humidity Hygrometer The Model is a versatile, multi-function, optical chilled mirror hygrometer designed to continuously measure

More information

RK CO 2 Transmitter Operator s Manual

RK CO 2 Transmitter Operator s Manual 65-2397RK CO 2 Transmitter Operator s Manual Part Number: 71-0185RK Revision: B Released: 7/18/14 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual before

More information

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR

SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR SAFETY MANUAL SIL 2 Certified Model FP-700 Combustible Hydrocarbon Gas Sensor Version 2.0 1 SAFETY CERTIFIED MODEL FP-700 COMBUSTIBLE GAS DETECTOR This manual addresses the specific requirements and recommendations

More information

CT-7 Series Toxic Gas Transmitter Operator s Manual

CT-7 Series Toxic Gas Transmitter Operator s Manual 65-2341 CT-7 Series Toxic Gas Transmitter Operator s Manual Part Number: 71-0424 Revision: P1 Released: 6/8/17 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction

More information

P/N CC-100. Coolant Circulator. Operation manual

P/N CC-100. Coolant Circulator. Operation manual P/N 4411-0019 CC-100 Coolant Circulator Operation manual Manual Version 1 Revision B July 1, 1998 Table of Contents Chapter 1 Introduction... 5 Description...5 Specifications...5 Chapter 2 Warnings...

More information

Safe Operation Manual. Honeywell IQ Force Gas Detector

Safe Operation Manual. Honeywell IQ Force Gas Detector Safe Operation Manual Honeywell IQ Force Gas Detector Honeywell Analytics 800-663-4164 403-248-9226 Fax 403-575-3708 03JUN2012 P/N 50105843-047 Version 01 http://www.honeywell.com HONEYWELL IQ Force PERSONAL

More information

RK/ RK Oxygen Detector Operator s Manual

RK/ RK Oxygen Detector Operator s Manual 65-2494RK/65-2497RK Oxygen Detector Operator s Manual Part Number: 71-0179RK Revision: 0 Released: 2/16/11 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating detector.

More information

A measured step forward

A measured step forward A measured step forward Operations & Maintenance Manual TM JESCO UV 405-412-440-450-480-550-80 2 LCD LCD PLUS Contents 1. Introduction...3 2. General Principles and Safety Instructions...4 3. Instructions

More information

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors

Quick Start Guide LIQ-QSG-Hx438, Rev A June Rosemount Hx438. Amperometric Steam Sterilizable Dissolved Oxygen Sensors Quick Start Guide LIQ-QSG-Hx438, Rev A June 2017 Rosemount Hx438 Amperometric Steam Sterilizable Dissolved Oxygen Sensors Safety Information WARNING! Do not exceed temperature and pressure limitations

More information

VMS-3. Part Number: Ultraviolet Systems VMS-3. Installation, Operation & Maintenance

VMS-3. Part Number: Ultraviolet Systems VMS-3. Installation, Operation & Maintenance VMS-3 Part Number: 95-0031 Ultraviolet Systems VMS-3 Installation, Operation & Maintenance The following are the types of flags used in this technical manual. They designate safety related items and important

More information

SEC Millenium Carbon Dioxide Gas Detector

SEC Millenium Carbon Dioxide Gas Detector SEC Millenium Carbon Dioxide Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email sales@sensorelectronic.com

More information

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL

LABORATORY ULTRA ZERO AIR GENERATOR. ZAC-ULT Series USER MANUAL LABORATORY ULTRA ZERO AIR GENERATOR ZAC-ULT Series USER MANUAL 166 Keystone Drive Montgomeryville, PA 18936 Telephone: 215-641-2700 Fax: 215-641-2714 Email: mtgmmville@matheson-trigas.com INT-0261-XX rev

More information

RK-02 Multi Point Detector Operator s Manual

RK-02 Multi Point Detector Operator s Manual 65-2485RK-02 Multi Point Detector Operator s Manual Part Number: 71-0237RK Revision: A Released: 11/26/14 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual

More information

RK-05 Combustible Gas Transmitter Operator s Manual

RK-05 Combustible Gas Transmitter Operator s Manual 65-2405RK-05 Combustible Gas Transmitter Operator s Manual Part Number: 71-0180RK Revision: 0 Released: 2/16/11 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction

More information

VIQUA is an ISO9001:2008 registered company specializing in the design, manufacture and sale of ultraviolet systems for:

VIQUA is an ISO9001:2008 registered company specializing in the design, manufacture and sale of ultraviolet systems for: PRO50 520374-R_RevB Table of Contents Preface... ii Contact Information... ii About VIQUA a Trojan Technologies Business... ii Scope.... ii 1.0 Project & System Description... 1 1.1 Project Description...

More information

RK Multi Point Detector Operator s Manual

RK Multi Point Detector Operator s Manual 65-2480RK Multi Point Detector Operator s Manual Part Number: 71-0198RK Revision: C Released: 11/26/14 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual before

More information

HUM-M2 Humidity Sensor Module Description

HUM-M2 Humidity Sensor Module Description HUM-M2 Humidity Sensor Module Description Humidity Sensor Modules are designed for environment monitoring and controlling in industrial, commercial and other buildings. These transmitters can be used for

More information

THERMAPHASE INSTALLATION AND OPERATING INSTRUCTIONS

THERMAPHASE INSTALLATION AND OPERATING INSTRUCTIONS Page 1 of 10 THERMAPHASE INSTALLATION AND OPERATING INSTRUCTIONS Purpose of Manual The purpose of this manual is to provide operating, servicing and repair instructions for the Summit standard models ThermaPhase

More information

RK Hydrogen Sulfide Transmitter Operator s Manual

RK Hydrogen Sulfide Transmitter Operator s Manual 65-2331RK Hydrogen Sulfide Transmitter Operator s Manual Part Number: 71-0176RK Revision: B Released: 11/26/14 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction

More information

Combustible Gas Sample-Draw Detector Operator s Manual

Combustible Gas Sample-Draw Detector Operator s Manual 35-3001-01 Combustible Gas Sample-Draw Detector Operator s Manual Part Number: 71-0282RK Revision: A Released: 8/3/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual Read instructions carefully before operating this device. ❶ This device is not to be used for Human Life Support applications. ❷ To avoid possible

More information

Models TX-KE and TX-KP Toxic Gas Transmitters

Models TX-KE and TX-KP Toxic Gas Transmitters Models TX-KE and TX-KP Toxic Gas Transmitters Instruction Manual PureAire Monitoring Systems, Inc. 557 Capital Dr. Lake Zurich, IL. 60047 Phone: 847-726-6000 Fax: 847-726-6051 Toll-Free: 888-788-8050 www.pureairemonitoring.com

More information

EN - english. Instructions for installation and operation. Mobile humidity meter DP 207

EN - english. Instructions for installation and operation. Mobile humidity meter DP 207 EN - english Instructions for installation and operation Mobile humidity meter DP 207 Dear customers, Thank you very much for deciding in favour of the mobile humidity measuring instrument DP 207. Please

More information

Model Super TJE. Ultra Precision Pressure Transducer DESCRIPTION FEATURES

Model Super TJE. Ultra Precision Pressure Transducer DESCRIPTION FEATURES Ultra Precision Transducer DESCRIPTION Model Super TJE is one of the most accurate industrial pressure transducers available today. The Super TJE features 0.05 % accuracy and zero temperature error of

More information

VMUS-4 Ozone Generator

VMUS-4 Ozone Generator VMUS-4 Ozone Generator Installation and Operation Manual pg. 1 Cautions, Warnings and Hazards Refer to the manual of the ozone generating system first, to assure proper location of all ozone equipment.

More information

AUTOMOTIVE REFRIGERANT ANALYZER OPERATION MANUAL. Analyzer Part Number: 68941

AUTOMOTIVE REFRIGERANT ANALYZER OPERATION MANUAL. Analyzer Part Number: 68941 AUTOMOTIVE REFRIGERANT ANALYZER OPERATION MANUAL Analyzer Part Number: 68941 Table of Contents FOR YOUR SAFETY:... 3 IDENTIFIER WARNINGS/GENERAL CAUTIONS... 3 WELCOME... 5 1 INTRODUCTION AND OVERVIEW...

More information

Operating Instructions Compur Monitox plus

Operating Instructions Compur Monitox plus Operating Instructions Compur Monitox plus Subject to changes without notice. Copyright Compur Monitors München Editor: Compur Monitors GmbH & Co. KG Weissenseestr. 101 D-81539 München Content Page Monitox

More information

APC BC300 Series 40kW 208/450/480V User Guide

APC BC300 Series 40kW 208/450/480V User Guide APC BC300 Series 40kW 208/450/480V User Guide Copyright 2002 APC Denmark ApS This manual is subject to change without notice and does not represent a commitment on the part of the vendor Thank You Thank

More information

M7100 Pressure Transducer

M7100 Pressure Transducer SPECIFICATIONS Performance standard on and off highway engine and vehicle OEMs Rugged for heavy equipment and outdoor use Designed specifically for high volume applications Stainless steel wetted surfaces

More information

SIMRAD GD10PE IR GAS DETECTOR

SIMRAD GD10PE IR GAS DETECTOR SIMRAD GD10PE IR GAS DETECTOR Operating Manual P3404E 1st edition Jan. 2004 www.simrad-optronics.com CONTENTS 1. PRODUCT DESCRIPTION... 5 1.1 General description... 5 1.2 Construction... 5 1.3 Application

More information

Owner s Manual. Filterless UV Air Purifier. This Manual Includes:

Owner s Manual. Filterless UV Air Purifier. This Manual Includes: Filterless UV Air Purifier Item No. 203758 Owner s Manual This Manual Includes: Warranty Safety Warnings Operating Instructions Replacement Directions Troubleshooting Tips Cleaning and Maintenance Tips

More information

ToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual

ToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual ToxiRAE II PGM-1100 Series Personal Toxic Gas Monitor User Manual 045-4003-000, Revision B, March 2005 - READ BEFORE OPERATING - This manual must be carefully read by all individuals who have or will have

More information

FLEXIBLE, AFFORDABLE, HIGH SPECIFICATION OXYGEN GAS ANALYSIS DESIGNED FOR DEMANDING HAZARDOUS AREA PROCESS CONTROL APPLICATIONS.

FLEXIBLE, AFFORDABLE, HIGH SPECIFICATION OXYGEN GAS ANALYSIS DESIGNED FOR DEMANDING HAZARDOUS AREA PROCESS CONTROL APPLICATIONS. SERVOTOUGH FLEXIBLE, AFFORDABLE, HIGH SPECIFICATION OXYGEN GAS ANALYSIS DESIGNED FOR DEMANDING HAZARDOUS AREA PROCESS CONTROL APPLICATIONS Flexible Can be used in hazard rated locations including Zone

More information

Using the Model 1500 Portable Chilled Mirror Hygrometer

Using the Model 1500 Portable Chilled Mirror Hygrometer APPLICATION NOTE 2016_0120 MODEL 1500 Rev C Using the Model 1500 Portable Chilled Mirror Hygrometer The Model 1500 is a portable, multi-function, optical chilled mirror hygrometer designed to accurately

More information

Hydrocarbon Gas Detector

Hydrocarbon Gas Detector SEC Millenium Hydrocarbon Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email: sales@sensorelectronic.com

More information

HIGH RANGE DIGITAL DEW POINT METER & MONITOR Model #8074 IMS # (TED1-8074)

HIGH RANGE DIGITAL DEW POINT METER & MONITOR Model #8074 IMS # (TED1-8074) HIGH RANGE DIGITAL DEW POINT METER & MONITOR Model #8074 IMS #116075 (TED1-8074) 2004 IMS Co. October, 2004 IMS Company 10373 Stafford Road Chagrin Falls, OH 44023-5296 Telephone: (440) 543-1615 Fax: (440)

More information

GAS MONITOR Model Model

GAS MONITOR Model Model Sierra Monitor Corporation 1991 Tarob Court, Milpitas, CA 95035 (408) 262-6611 (800) 727-4377 (408) 262-9042 - Fax E-mail: sierra@sierramonitor.com Web Site: www.sierramonitor.com GAS MONITOR Model 2350-00

More information

Mercury Vapor Monitor VM-3000

Mercury Vapor Monitor VM-3000 Member of the envea Group Mercury Vapor Monitor VM-3000 Continuous mercury measurement in air and other gases Direct reading mercury vapor meter Proven reliable detection method: UV absorption (CVAAS)

More information

Bench Series Ovens Models 21 / 31 / 51 Operating Manual

Bench Series Ovens Models 21 / 31 / 51 Operating Manual Heat cycle HEATER SWITCH RECIRC. FAN Quincy Lab, Inc. Bench Series Ovens Models 21 / 31 / 51 Operating Manual Standard Contents (1) Bench Series Oven (2) Adjustable chrome wire shelf (4) Shelf brackets

More information

Operation and Maintenance Manual

Operation and Maintenance Manual Operation and Maintenance Manual GDS-48 Remote Bridge Sensor for Combustible Gases AUTHORIZED DISTRIBUTOR: GasDetectorsUSA.com - Houston, Texas USA sales@gasdetectorsusa.com - 832-615-3588 CAUTION: FOR

More information

SLT-Series Installation and maintenance instructions

SLT-Series Installation and maintenance instructions Installation and maintenance instructions MODELS UVG SLT1, UVG SLT2, UVG SLT30, UVG SLT40, UVG SLT75, UVG SLT80, UVG SLT125, UVG SLT172 ATTENTION: Please read this manual carefully and follow the instructions.

More information

HFID Heated Total Hydrocarbon Analyzer Module

HFID Heated Total Hydrocarbon Analyzer Module Product Data Sheet PDS 103-685.A01 October, 2003 NGA 2000 HFID Heated Total Hydrocarbon Analyzer Module Heated analyzer designed to operate from 200 F to 400 F Eliminates loss of hydrocarbons due to condensation

More information

DF-340E PROCESS ANALYZERS APPLICATIONS FEATURES

DF-340E PROCESS ANALYZERS APPLICATIONS FEATURES PROCESS ANALYZERS DF-340E The DF-340E is a Coulometric sensor based oxygen analyzer suitable for outdoor installation and designed to measure trace oxygen and percent level in pure and multigas backgrounds.

More information

Apex E Quick Installation Guide

Apex E Quick Installation Guide ELEMENTAL SCIENTIFIC INC. Apex E Quick Installation Guide Apex E Quick Installation Guide ã Elemental Scientific Inc. 2440 Cuming Street Omaha, NE 68131 USA Phone 402.991.7800 Fax 402.991.7799 Table of

More information

VM-3000 Mercury Vapor Monitor

VM-3000 Mercury Vapor Monitor MERCURY INSTRUMENTS Analytical Technologies VM-3000 Mercury Vapor Monitor Continuous mercury measurement in air and other gases Direct reading mercury vapor meter Proven reliable detection method: UV absorption

More information

5New features & EXtreme design, now with HART Protocol and DuraSource Technology. Ultima X Series Gas Monitors

5New features & EXtreme design, now with HART Protocol and DuraSource Technology. Ultima X Series Gas Monitors Ultima X Series Gas Monitors Versatile fixed instruments provide continuous monitoring of many hazardous gases using catalytic, electrochemical, and infrared gas detection methods. HART Protocol is now

More information

RK/ RK Carbon Monoxide Detector Operator s Manual

RK/ RK Carbon Monoxide Detector Operator s Manual 65-2496RK/65-2499RK Carbon Monoxide Detector Operator s Manual Part Number: 71-0156RK Revision: 0 Released: 2/16/11 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

85 Uncompensated SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage

85 Uncompensated SPECIFICATIONS FEATURES APPLICATIONS STANDARD RANGES. 316L SS Pressure Sensor Small Profile 0-100mV Output Absolute and Gage FEATURES Weldable and Threaded Process Fittings -40ºC to +125ºC Operating Temperature ±0.1% Pressure Non Linearity Solid State Reliability APPLICATIONS Medical Instruments Process Control Fresh & Waste

More information

DRY AIR SYSTEMS, INC Metro Boulevard Maryland Heights, Missouri (314) fax (314)

DRY AIR SYSTEMS, INC Metro Boulevard Maryland Heights, Missouri (314) fax (314) DRY AIR SYSTEMS, INC. 2655 Metro Boulevard Maryland Heights, Missouri 63043 (314) 344-1114 fax (314) 344-0677 HD SERIES DRIERS TABLE OF CONTENTS WHY AN AIR DRYER 3 WHAT IS A DESICCANT AIR DRYER 3 Desiccant

More information

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies.

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies. GE Sensing & Inspection Technologies Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in

More information

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies.

Ventostat Wall Mount. Telaire Wall Mount CO 2. , Humidity and Temperature Transmitters. GE Sensing & Inspection Technologies. GE Sensing & Inspection Technologies Ventostat Wall Mount Telaire Wall Mount, Humidity and Temperature Transmitters Features: Patented, Absorption Infrared Gas sensing engine provides high accuracy in

More information

HTU-500 Ozone Generator

HTU-500 Ozone Generator HTU-500 Ozone Generator Installation and Operation Manual pg. 1 Cautions, Warnings and Hazards Refer to the manual of the ozone generating system first, to assure proper location of all ozone equipment.

More information

TOUCHSTONE Dual Variable Speed System. Installation & Operating Manual with Troubleshooting and Configuration Guide

TOUCHSTONE Dual Variable Speed System. Installation & Operating Manual with Troubleshooting and Configuration Guide TOUCHSTONE Dual Variable Speed System CIDU-315-03-02-01 Configuration #1 Dual Variable Speed Pump/Blower Configuration #2 Variable Speed Pump Configuration #3 Variable Speed Blower Installation & Operating

More information

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION

SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION SEC SIGNATURE ETO MONITOR TRAINING SENSOR ELECTRONICS CORPORATION SEC SIGNATURE ETO MONITOR SIGETO1420597 PRODUCT INTRODUCED TO ETHYLENE OXIDE STERILIZATION INDUSTRY IN 2001. OVER 850 UNITS IN FIELD WORLDWIDE

More information

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature Critical Environment Technologies Canada Inc. Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature OPERATION MANUAL REV: A JUNE-2-2008 #145, 7391 Vantage Way Delta, BC V4G 1M3 Canada Phone:

More information

9 W, 15 W and 25 W Industrial Fiber Laser Systems IFL Series

9 W, 15 W and 25 W Industrial Fiber Laser Systems IFL Series COMMERCIAL LASERS 9 W, 15 W and 25 W Industrial Fiber Laser Systems IFL Series Key Features Diffraction-limited light delivered through fiber termination module Single transverse mode, collimated beam

More information

GCS Multipoint Carbon Dioxide Gas Concentration Sampler

GCS Multipoint Carbon Dioxide Gas Concentration Sampler Electronic Control Manual 16 Sensors and Stats Section S Technical Bulletin Issue Date 0101 GCS-000- Multipoint Carbon Dioxide Gas Concentration Sampler Product Overview Page *3 Installation 5 Mounting

More information

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL

GASGUARD Cl 2. Chlorine Sensor OPERATING & INSTALLATION MANUAL GASGUARD Cl 2 Chlorine Sensor OPERATING & INSTALLATION MANUAL Operating and Installation Manual 2 Table of Contents GasGuard Cl 2 Operating and Instruction Manual General description. 4 Installation..

More information

RK-05 Carbon Monoxide Detector Operator s Manual

RK-05 Carbon Monoxide Detector Operator s Manual 65-2433RK-05 Carbon Monoxide Detector Operator s Manual Part Number: 71-0189RK Revision: 0 Released: 5/17/11 RKI Instruments, Inc. www.rkiinstruments.com WARNING Read and understand this instruction manual

More information

BALLAST REPLACEMENT. Contents WARNING INSTALLATION MANUAL

BALLAST REPLACEMENT. Contents WARNING INSTALLATION MANUAL BALLAST REPLACEMENT INSTALLATION MANUAL WARNING Turn off power at main source before making any electrical connections or servicing the unit. UV light produced by this unit is harmful to the eyes and may

More information

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL

GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL GASGUARD H 2-1% Hydrogen Sensor OPERATING & INSTALLATION MANUAL 2 Table of Contents GasGuard H 2-1% General description. 4 Installation.. 4 Locating the sensor. 4 Installation guidelines 5 Wiring.. 6 Operation..

More information

Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual

Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual 35-3001-07 Combustible Gas/Hydrogen Sulfide Sample-Draw Detector Operator s Manual Part Number: 71-0411 Revision: 0 Released: 8/7/17 www.rkiinstruments.com WARNING Read and understand this instruction

More information

M5800 Pressure Transducer with Rotatable Display

M5800 Pressure Transducer with Rotatable Display Pressure Transducer with Rotatable Display SPECIFICATIONS Wide Temperature Range Compact Variety of Pressure Ports and customized Electrical Configurations FEATURES CE Compliance Reverse Polarity Protection

More information

Black Body Calibration Source User Manual

Black Body Calibration Source User Manual Black Body Calibration Source Surface Plate Type with External Controller Model CBB18GC- 2 WD1069 Rev B 10/23/17 Wahl Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853

More information

Carbon Dioxide Sample-Draw Detector Operator s Manual

Carbon Dioxide Sample-Draw Detector Operator s Manual 35-3001-05-03 Carbon Dioxide Sample-Draw Detector Operator s Manual Part Number: 71-0413 Revision: 0 Released: 8/4/17 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Constant climate chambers

Constant climate chambers KBF series (E5.2) Constant climate chambers Constant climate chambers The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions.

More information

Dräger X-am 7000 Multi-Gas Detection

Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 is the solution for the simultaneous and continuous measurement of up to five gases. It is the ideal companion in a variety of applications where the

More information

SIL 2. Ultima X Series Gas Monitors CERTIFIED

SIL 2. Ultima X Series Gas Monitors CERTIFIED Ultima X Series Gas Monitors 5 Now with SIL 2 CERTIFIED Products Versatile fixed instruments provide continuous monitoring of many hazardous gases using catalytic, electrochemical, and infrared gas detection

More information

LMV52 Control. Specifications. Document No. LV February 3, Product Description. Sample Specification

LMV52 Control. Specifications. Document No. LV February 3, Product Description. Sample Specification LMV52 Control Product Description February 3, 2017 The LMV52 is a microprocessor-based burner management system with matching system components for the control and supervision of forced draft burners.

More information

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL

GASGUARD O 2. Oxygen Sensor OPERATING & INSTALLATION MANUAL GASGUARD O 2 Oxygen Sensor OPERATING & INSTALLATION MANUAL 2 Table of Contents General description. 4 Installation.. 4 Locating the sensor. 4 Installation guidelines 5 Wiring.. 6 Operation.. 7 Start-up...

More information

Instruction Sheet DANGER

Instruction Sheet DANGER VIQUA Lamp and Quartz Sleeve Replacement Models: D4, E4, F4, D4 Premium, Plus Models Discontinued Models: A, B, C, D, E, F, B4, C4 Overview UV Lamp replacement: The amount of UV light created by the UV

More information

1.2.3 Returning Instrument for Repair.

1.2.3 Returning Instrument for Repair. 2 Contents Installation....Page 5 Electrical and I/O...Page 8 Operating Principle...Page 11 Specifications.......Page 12 Operation.Page 14 Menu Screens......Page 15 Installation Tips and Troubleshooting.

More information

F-1100 Series Single Turbine Meters Suitable for pipe sizes 1 ¼ and larger.

F-1100 Series Single Turbine Meters Suitable for pipe sizes 1 ¼ and larger. ONICON Insertion Turbine Flow Meters Provide an accurate, cost effective option for flow measurement of relatively clean liquids such as those found in closed loop HVAC systems. Each meter is built to

More information

ENVAIR lab CO2 Incubator

ENVAIR lab CO2 Incubator ENVAIR lab CO2 Incubator Laboratory Incubator Air-jacketed CO2 Incubator Introduction C02 Incubators are widely used in scientific research to grow and maintain cell cultures. A ENVAIR lab CO2 Incubator

More information

U86B. Urea Dosing SPECIFICATIONS FEATURES APPLICATIONS

U86B. Urea Dosing SPECIFICATIONS FEATURES APPLICATIONS SPECIFICATIONS Mountable with O-ring seal Stainless Steel wetted surfaces Amplified ASIC calibrated Absolute, sealed gage Cable option Analog output SENT Output Optional (contact factory for details) FEATURES

More information

ULTRAMAX GOLD UMX REMOTE LAMP AIR DISINFECTION SYSTEMS w/ ODOR REDUCING TECHNOLOGY (PCO) OWNERS GUIDE

ULTRAMAX GOLD UMX REMOTE LAMP AIR DISINFECTION SYSTEMS w/ ODOR REDUCING TECHNOLOGY (PCO) OWNERS GUIDE ULTRAMAX GOLD UMX REMOTE LAMP AIR DISINFECTION SYSTEMS w/ ODOR REDUCING TECHNOLOGY (PCO) OWNERS GUIDE Models: UMX-1200T/ UMX-1900T / UMX-1202T/ UMX-1902T 120-240VAC Power Input with Remote Mount Lamps

More information

New Features TDLS 200. The Process TDLS Solution. Tunable Diode Laser Spectroscopy (TDLS) Analyzer.

New Features TDLS 200. The Process TDLS Solution. Tunable Diode Laser Spectroscopy (TDLS) Analyzer. The Process TDLS Solution New Features TDLS 200 Tunable Diode Laser Spectroscopy (TDLS) Analyzer Bulletin 11Y01B01-01E-A www.yokogawa.com/an/index.htm The first Tunable Diode Laser analyzer Innovative

More information

Constant climate chamber with humidity

Constant climate chamber with humidity KBF series (E5.2) Constant climate chambers Constant climate chamber with humidity The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate

More information

Constant climate chambers

Constant climate chambers KBF series (E5.2) Constant climate chambers Constant climate chambers The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions.

More information

U5300 Industrial Pressure Transducer

U5300 Industrial Pressure Transducer FEATURES Heavy Industrial CE Approval 10 V/m EMI Protection Reverse Polarity Protection on Input Short Circuit Protection on Output ±0.1% Accuracy ±0.5% Total Error Band Compact Outline -40 C to +125 C

More information

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL S-SCORT II Model 15006 SSCOR, INC. 11064 Randall Street Sun Valley, CA 91352 USA Telephone +1-818-504-4054 800-434-5211 Fax +1-818-504-6032 www.sscor.com Email:

More information

D5100 Industrial Differential Pressure Transducer

D5100 Industrial Differential Pressure Transducer FEATURES Heavy Industrial CE Approval (amplified only) As Low As ±0.1% Pressure Non Linearity Rugged Construction: Can Withstand 50g Shock/20g Vibration Up to -40 C to +125 C Operating Temperature Range

More information

SprintIR -W CO2 Sensor

SprintIR -W CO2 Sensor Datasheet SprintIR -W CO2 Sensor Measurement range: 0-5%; 0-20%; 0-100% Battery applications Low power Fast response Long lifetime Tel: +44 (0) 1236 781 900 info@gassensing.co.uk www.gassensing.co.uk Benefits

More information

Model 100 Duct Mount Kit

Model 100 Duct Mount Kit D ETCON, INC. Model 100 Duct Mount Kit Model 100 DM-100 H2S 100 PPM CAL PORT Operator s Installation and Instruction Manual Covers all versions of the 100 Series Sensor DETCON, Inc. 4055 Technology Forest

More information

GPR-1500 DA PPM Oxygen Analyzer

GPR-1500 DA PPM Oxygen Analyzer Technical Specifications * Accuracy: Analysis: Application: < 2% of FS range under constant conditions 0-100, 0-1000 PPM, 0-1%, 0-25% (CAL) FS Auto-ranging or manual lock on a single range Oxygen analysis

More information

RK/ RK Oxygen Detector Operator s Manual

RK/ RK Oxygen Detector Operator s Manual 65-2502RK/65-2510RK Oxygen Detector Operator s Manual Part Number: 71-0074RK Revision: A Released: 3/1/11 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating detector.

More information

GATEKEEPER GAS PURIFICATION SYSTEM DX SERIES FOR 20, 30, 50 AND 60 NM 3 /HR

GATEKEEPER GAS PURIFICATION SYSTEM DX SERIES FOR 20, 30, 50 AND 60 NM 3 /HR ELECTRONIC MATERIALS GATEKEEPER GAS PURIFICATION SYSTEM DX SERIES FOR 20, 30, 50 AND 60 NM 3 /HR Continuous flow, point-of-use purified CO 2 gas at a low cost of ownership continuous flow, point-of-use

More information

Dräger Polytron Fixed Gas Detector. Dräger Polytron 5700 Infrared DrägerSensor PIR 7000 to detect combustible gases and vapors

Dräger Polytron Fixed Gas Detector. Dräger Polytron 5700 Infrared DrägerSensor PIR 7000 to detect combustible gases and vapors Dräger Polytron 5700 Fixed Gas Detector Dräger Polytron 5700 Infrared DrägerSensor PIR 7000 to detect combustible gases and vapors Dräger Polytron 5700 Including e-box for increased safety wiring The Dräger

More information

RK / RK Methane Detector Operator s Manual

RK / RK Methane Detector Operator s Manual 61-1006RK / 61-0192RK Methane Detector Operator s Manual Part Number: 71-0228RK Revision: 0 Released: 10/29/18 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information