VITALOGIK= 4000 / 6000 Series Configured Patient Monitors SERVICE MANUAL FDM REV.A

Size: px
Start display at page:

Download "VITALOGIK= 4000 / 6000 Series Configured Patient Monitors SERVICE MANUAL FDM REV.A"

Transcription

1 VITALOGIK= 4000 / 6000 Series Configured Patient Monitors SERVICE MANUAL FDM REV.A September 2013

2 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices Manufacturer s Name : Mennen Medical Ltd. 4 Hayarden Street, Yavne, P.O. Box 102, Rehovot, , Israel Tel.: Fax: European Representative : Charter-Kontron Limited Unit 18 Avant Business Centre 21 Denbigh Road Milton Keynes MK1 1DT England Tel.: Fax: US Representative: Mennen Medical Corp. 290 Andrews Road Feasterville-Trevose, PA Phone Fax Publication No.FDM REV.A Revision: September 2013 Copyright Mennen Medical Ltd All RIGHTS RESERVED Registered trademarks are the intellectual property of their respective holders.

3 VitaLogik 4000/6000 Service Manual TABLE OF CONTENTS Chapter 1: System Description Overview Labeling Symbols Field Repairs Statement Training and Qualification System Features and Capabilities Monitor Inputs Controls and Outputs System Specification Chapter 2: Hardware Description VitaLogik Overview Main Units CPU Card Main Storage Main Power Supply OEM Power Supply Unit Rechargable Battery Splitter Board Front Panel Front End Board Inputs (CIPAM) ECG / RSP BP / CO Temp NIBP SPO2 (Nellcor) SPO2 (Masimo) Microstream UIM - Universal Input Remote Keypad (Optional) Chapter 3: Setting Up the System Introduction Accessing the System Setup Menu Date & Time Setup General Settings General Units Mennen Medical i

4 VitaLogik 4000/6000 Service Manual Timeouts Hospital Name Setting the General Settings Panel Alarm Volume and Controls Enable Alarm Off Silence Time SpO2 Alarm Tone Priority Level Setting the Alarm Volume & Control Panel Sound Event Default Alarm Limits Event Setup Vital Sign (VS) Setup Accessing the Vital Signs Setup Dialog Panels ECG Setup Respiration Setup BP Setup NIBP Setup SpO2 Setup Temperature Setup CritiCool Setup EtCO2 Setup EtCO2 Calibration Cardiac Output (CO) Setup Transcoutaneous Gas (TcGas) Setup CO2 (from Anesthetic Gas) Setup O2 Setup N2O Setup Anesthetic Agent Setup Ventilator Setup Parameter Hierarchy Monitor Profiles Setup Showing/Hiding the Monitor profile on the Main Screen Changing the Position of an Monitor Profile Clearing a Monitor profile Admit by Default Report Setup Naming a Report Automatic Reports Network Setup Recorder Setup Trends Setup Quick Keys Setup Tabular Charts Naming a UDC Editing a UDC Clearing a UDC From the UDC List Software Version Enmove Interface Setup ii Mennen Medical

5 VitaLogik 4000/6000 Service Manual Change Password Check Disk Permission Editor Remove Saved Patients Touch Screen Calibration Demo Activation Copy Configuration Utility Chapter 4: System Installation Overview System Acceptance Safety Precautions Grounding the System Equipotential Grounding Electrosurgery, Diathermy and Defibrillation Protection Mounting of Monitors Accidental Wetting of the Equipment General Hazards Site Survey Unpacking and Inspection Setting Up the Basic System Connecting and Wiring the LAN Cables Network Connections Mounting Options Mounting the Switch and Patch Panel Mounting and Connecting the LAN Hub and the Patch Panel Mounting the VitaLogik Monitor on the Wall Chapter 5: Upgrading Software Upgrading Using Microsoft Windows Utilities General Required Hardware Required Software Upgrading VitaLogik 4000/ Copying the Tar file from the Laptop to VitaLogik 4000/ Upgrading VitaLogik 4000/6000 Software Chapter 6: Maintenance and Cleaning Periodic Testing and Preventive Maintenance Cleaning the VitaLogik 4000/6000 Bedside Monitor Replacing the VitaLogik 4000/6000 Battery Calibration and Preventive Maintenance Error Messages and Troubleshooting Error Messages Trouble Shooting VitaLogik 4000/6000 Bedside Monitor Troubleshooting Vital Signs Mennen Medical iii

6 VitaLogik 4000/6000 Service Manual Appendix A: Spare Parts Spare Parts VitaLogik 4X00... A-1 Spare Parts VitaLogik A-6 Appendix B: Drawings External Connectors... B-2 Patient Input Connectors... B-3 VitaLogik 6000 Assemblies... B-8 VitaLogik Top Assembly... B-8 VitaLogik 6000 Chassis ASSY with Touch Screen... B-10 BEZEL Front Panel VL 6X00 ASSY... B-12 Mechanical Connector SUB ASSY 15" for FOR VL6XXX Invasive and Non Invasive... B-14 Rear Cover ASSY 15"... B-16 VitaLogik 6000 Recorder Cover ASSY... B-18 Bottom Cover VitaLogik 6000 ASSY... B-20 Plate, Battery Cover, Plate ASSY... B-22 PCBA VitaLogik ASSY... B-24 VitaLogik 4X00 Assemblies... B-26 VitaLogik 4X00 Top ASSY (Internal Wiring Diagram)... B-26 VL4X00 Chassis ASSY... B-28 BEZEL Front Panel VL 4X00 ASSY... B-30 Mechanical Connector SUB ASSY 15" for FOR VL 4X00 Invasive and Non Invasive... B-32 Rear Cover VL4XXX ASSY... B-34 Bottom Cover VitaLogik 4000 ASSY... B-36 Plate, Battery Cover, Plate ASSY... B-38 PCBA, VitaLogik ASSY.... B-40 Appendix C: CerebraLogik Introduction... C-1 Indications for Use... C-1 Compliance... C-2 CerebraLogik Storage... C-2 Icons... C-2 Patient Safety... C-3 CerebraLogik Label... C-4 CerebraLogik - Waveforms and Display Modes... C-4 Quality Bar and Graph... C-5 CerebraLogik Full Disclosure... C-5 Operating VitaLogik Monitor with CerebraLogik... C-5 Preparing for Operation... C-5 Electrode Location... C-6 Technical Alarm... C-7 CerebraLogik Service... C-7 CerebraLogik Drawing... C-7 iv Mennen Medical

7 VitaLogik 4000/6000 Service Manual Appendix D: Additional Information and Technical Assistance Appendix E: Laser Printer Setup Appendix F: Test Procedures Appendix G: Acceptance Report Mennen Medical v

8 VitaLogik 4000/6000 Service Manual vi Mennen Medical

9 CHAPTER 1: SYSTEM DESCRIPTION Overview Labeling The VitaLogik 4000/6000 Series Patient monitoring system is a compact monitor that suits a wide range of patient applications. The VitaLogik 4000/6000 series offer solutions to both intermediate and critical care monitoring. It is mains or battery operated and can be used both as a stationary, or as transportable monitor. It is offered with an optional touch screen for fast and easy operation. The VitaLogik 4000/6000 is easily adapted to a diverse range of monitoring requirements including: Adult, Pediatric, Neonatal and patient transportation. Each monitor can be customized to meet the user's needs. As computerized monitors, the VitaLogik 4000/6000 series, allow the users to take advantage of data connectivity, embedded data storage capabilities and high resolution display. Figure 1-1: VL4X00 Label Mennen Medical 1-1

10 System Description VitaLogik 4000/6000 Service Manual Symbols Figure 1-2: VL6000 Label Following is a short description of the meaning of the various symbols which appear on the VitaLogik 4000/6000 Series. Symbol Description Alternating Current Equipotential Danger High Voltage Attention, consult accompanying documents (Service to be performed by qualified technician) Type BF applied part Type BF applied par - Defibrillator-proof Type CF applied part - direct cardiac application defibrillator-proof 1-2 Mennen Medical

11 VitaLogik 4000/6000 Service Manual Symbol Description Refer to instruction manual/ booklet Restricts the sale and use of this instrument to qualified medical personnel only. Not for use in explosive gas environment Electrical and electronic equipment - Dispose according to local regulation Date of manufacture CE Mark Mennen Medical 1-3

12 System Description VitaLogik 4000/6000 Service Manual Field Repairs Statement Repair of Mennen Medical equipment in the field should be performed by service engineers, authorized by Mennen Medical. Repairs should be performed on a board level only. The list of replaceable component and boards is available in the spare part list. System testing should be performed using the test form available in the Service manual. Training and Qualification Mennen Medicalor it's authorized distributer provides training for technicians and sales personal as per the intended use of the device or system. The scope of the training is part of the agreement between Mennen Medical or it's authorized distributer and/or the end user. Servicing of Mennen Medical equipment is allowed only to persons trained and qualified to service the equipment efficiently and safely. System Features and Capabilities The VitaLogik 4000/6000 displays a wide range of vital sign clinical parameters. It includes: ECG/Respiration multi-lead for monitoring ECG and respiration. Dual (four optional) BP inputs for monitoring two invasive blood pressures. CO/2TMP connector to measure either thermodilution Cardiac Output or two temperatures. Temp connector. NIBP input for monitoring non-invasive blood pressure using the oscillometric method. Pulse Oximetry. Microstream End Tidal CO2 (EtCO2) for monitoring CO2 during exhalation (EtCO2) and inhalation (inco2) and Respiration Rate. Serial Input for communication and interface with an external, auxiliary device. The VitaLogik 4000/6000 s main monitoring features include: Continuously updated vital sign parameter data displayed as numeric values and continuous waveforms. Clinical and technical alarm detection as well as alarm notification in audio and visual formats. 1-4 Mennen Medical

13 VitaLogik 4000/6000 Service Manual Monitor Display Features Easy access to collected data in the form of tabular charts and graphical trends as well as various types of clinical reports. Patient data output to a recorder or a printer, if connected. You can interact with the VitaLogik 4000/6000: Using interactive panels These panels include menus and dialog panels that enable you to configure system and monitoring parameters. Responding to Events Fixed keys on the front panel enable you to respond quickly to events such as alarms. Reviewing Clinical Data Charts, Trends, Full Disclosure and Overview enable you to view and compare data in different formats and over selected time periods. The VitaLogik 4000/6000 can display up to eight traces, depending on how the vital sign display is configured. For monitor control procedures, refer to the User s Manual. All clinical data collected from the inputs is stored in the flash memory. This data includes waveforms, vital signs, trends, charts and beat-to-beat data. The VitaLogik 4000/6000 can store at least 6 hours of patient waveform and 10 days of numeric data for review purposes. The front panel of the main processing unit contains 5 fixed keys and the Quicknob. These features enable you to interact with the system. A remote keypad provides direct access, via the orange keys, to the menus of Vital signs, Patient data, and Setup, via the green keys, to Event, Print, Record, and Freeze and by the yellow key to Timer. This allows the user to reach these functions without opening the Main menu. For a detailed description of the functions and controls of the main processing unit, see Chapter 6 Controls and Functions of the VitaLogik 4000/6000 User Manual. The main screen display is divided into four areas: Global Header Area Patient Area Patient Display Area QuicKeys For a description of all monitor display features and functions, Chapter 6 Controls and Functions of the VitaLogik 4000/6000 User Manual. Mennen Medical 1-5

14 System Description VitaLogik 4000/6000 Service Manual Inputs The Vital Signs inputs are housed on the left side panel. They include: 12 Lead ECG/RSP Dual BP CO/2TMP Temp NIBP SpO2 EtCO2 MicroStream Controls and Outputs Video output for external display External keyboard and mouse Remote keypad Analog output EtCO2 gas outlet Figure 1-3: The VitaLogik 4000/6000 Remote Keypad System Specification See the following pages. 1-6 Mennen Medical

15 VitaLogik 4000/6000 Service Manual Mennen Medical 1-7

16 System Description VitaLogik 4000/6000 Service Manual 1-8 Mennen Medical

17 CHAPTER 2: HARDWARE DESCRIPTION This chapter includes the following sections: VitaLogik Overview CPU Card Main Storage Main Power Supply OEM Power Supply Unit Rechargable Battery Splitter Board Front Panel Front End Board Inputs (CIPAM) ECG / RSP 2BP / CO Temp NIBP SPO2 (Nellcor) SPO2 (Masimo) Remote Keypad (Optional) WARNING!!! Electrical shock hazard! Disconnect the main A/C power supply cable from the A/C power socket before service!! If power must remain connected for troubleshooting while the unit is uncovered, use extreme caution. When you have completed a test, immediately unplug the unit. Do not wear rings, bracelets or metal objects that may come in contact with live power contacts. Mennen Medical 2-1

18 Hardware Description VitaLogik 4000/6000 Service Manual VitaLogik Overview The following electronic block diagram describes the main units and internal interfaces for the VitaLogik 4000/6000 system. Figure 2-1: Electronics block diagram of main units and internal interfaces of VitaLogik 4000/ Mennen Medical

19 VitaLogik 4000/6000 Service Manual Main Units VitaLogik 4000/6000Main CPU (see CPU Card ) Front End ( see Front Panel ) Power Splitter ( see Rechargable Battery ). User Interface ( see Front Panel ) Recorder - Optional built in recorder for 25mm/Sec 3 WF printing. L-Iion battery- 16.8Volt 5.2A L-ion battery. Medical Power Supply - "XP" medical grade power supply for output voltage of 24Volt/100W AC Line filter - AC line filter for medical uses. LCD for VL6000- Built in NEC 15" / SHARP 15" XGA display with LVDS interface. for VL4x00- Built in NEC 12.1" / SHARP 12.1" XGA display with LVDS interface. Inverter - The inverter provides high voltage for the NEC LCD. The light level is controlled by the power splitter board via specific signal. Note: SHARP display does not require an inverter. Touch Screen - ElO 12.1" / 15" IT which communicates with the CPU via a serial interface. The power splitter board provides the required 5 Volts and the communication signal routine between the touch controller and the CPU. Main Storage - Flash memory card (see Main Storage ). Mennen Medical 2-3

20 Hardware Description VitaLogik 4000/6000 Service Manual Figure 2-2: VitaLogik 6000 Internal Wiring Diagram 2-4 Mennen Medical

21 VitaLogik 4000/6000 Service Manual Figure 2-3: VitaLogik 4000 Internal Wiring Diagram CPU Card The main CPU board is a single board computer, custom designed according to VitaLogik 4000/6000 specifications. The board includes the following basic features and technologies: VIA Core Fusion processor - 533MHz On-board 128MB Sdram memory On-board sound blaster and audio amplifier Compact Flash interface Mennen Medical 2-5

22 Hardware Description VitaLogik 4000/6000 Service Manual 4 COM ports LPT VGA and LVDS interfaces USB 1.0 Main Storage 2 Port LAN 10/100BT BIOS Setup configured for Mennen Medical Ltd. applications. The CPU runs the Mennen Medical Ltd. Patient Monitoring application under QNX4.25 operating system which is located in the external flash card. The main storage consists of a 1 GB, 8 GB or 16 GB flash memory which stores the QNX operation system with PHOTON window manager and the VitaLogik system application software. The flash memory stores all of a patient s clinical data (vital signs and waveforms). These files are retrieved when the user enters the data review panels. Figure 2-4: Flash Card Size Main Power Supply The main power supply is composed of power supply module (AC to DC), DC-to- DC Splitter board, and a rechargargable battery. OEM Power Supply Unit "XP" medical grade power supply for output voltage of 24Volt/100W. The main power supply supplies DC voltage to the splitter board. Note: There is no field adjustment or repair. In case of malfunction, replace with a new power supply supplied by Mennen Medical Ltd Mennen Medical

23 VitaLogik 4000/6000 Service Manual Rechargable Battery Splitter Board The unit is supplied with one 16.8Volt 5.2A L-ion battery (two battery optional). The unit can operate for 3 hours with one battery and 6 hours with two batteries. The AC input voltage is VAC. The Splitter Board handles the followings: DC/DC Converter; Input: +24V Output:+12V, +8.25V, +5V Batteries charger and switching between AC to DC. Panel; power and User Interface Remote control; power and user interface LCD; power and light level Touch screen; Power and control Recorder; power (8.25V) and LPT interface connector Shutdown the system Power Splitter Board FE Power Connector (DC OUT) 24V IN -Main Power Connector CPU Power Connector (DC OUT) 4 C om1 to CPU Com-1 1 LPT to CPU LPT Jtag Connector (MM use only) Remote CTL Connector 5 2 C om2 to CPU Com-2 Recorder Connector 3 Panel Connector FAN & UIM power Connector (DC OU T) 6 Touch Screen Connector LCD Connector Figure 2-5 Power Splitter Board Mennen Medical Ltd. PN Mennen Medical 2-7

24 Hardware Description VitaLogik 4000/6000 Service Manual Connectors: FE (Front End - "Cipam") Power Connector provides +12Volt for FE board via U6 switch regulator (item 1 in Figure 2-5). The power circuit is controlled by main controller which turns the power ON/OFF according the monitor state. Maximum quiescent current for full options is ~800mA. Main Power Connector +24Volt input from main power supply. CPU (VitaLogik Main board) Power Connector provides +5Volt for VitaLogik 4000/6000 main board via U2 switch regulator (item 2 in Figure 2-5). The power circuit is controlled by main controller which turns On/Off the power according to the monitor state. Maximum quiescent current is ~3500mA. COM1 - RS232 UART interface between power splitter board and VitaLogik 4000/6000 main board. All user interface commands are transmitted to/from the VitaLogik 4000/6000 CPU via this interface LPT - parallel interface to VitaLogik 4000/6000 main board. This interface is using for built in recorder option. COM2 - RS232 UART interface between touch screen controllers, to VitaLogik 4000/6000 main board. Recorder Connector - parallel interface to built in recorder. This interface include dedicated 8.25Volt power circuits (item 3 in Figure 2-5). Panel Connector - Interface connection to panel board. LCD Connector- interface connector to LCD inverter. This interface includes dedicated 12Volt power circuits (item 6 in Figure 2-5) which controlled by the main controller. The interface includes control line for back light power consumption which can control the light value for AC connection and battery use. Touch Screen - RS232 UART interface between touch screen controller to VitaLogik 4000/6000 main board. The power splitter provides power lines (+5v) for the touch controller and route the data line to/from Com2. Fan & UIM Connector - The interface include 12Volt for the fan system which is controlled by the main controller according to the system state. The interface also includes optional 5Volt lines for UIM power connection. Remote Control - The interface includes a RS232 UART level and a 5Volt power line to the remote control unit. The communication with the power splitter is done by a specific UART channel to the main controller and goes to VitaLogik 4000/6000 CPU via COM1. TAG Interface - interface connector for development and manufacturing procedures. 2-8 Mennen Medical

25 VitaLogik 4000/6000 Service Manual Main Units: Main Controller IC controller, handles the followings: Power Circuits Charger Panel- User Interface Remote control LCD light level ImportantI The board software version notation is SPB:X.X. Upgrading the software should only be performed by Mennen Medical Ltd.. Q1 - Main power switch module which converts main 24Volt input to 14Volt (item 1 in Figure 2-5) Q2-12Volt Power switch circuits for LCD inverter (item 6 in Figure 2-5). Q3-12Volt Power switch circuits for FE ("CIPAM") board (item 4 in Figure 2-5) Q4-5Volt Power switch circuits for VitaLogik 4000/6000 main board (item 2 in Figure 2-5) Q5-8.5Volt Power switch circuits for built in recorder and panel board (item 3 in Figure 2-5) Charger - handle the L-ion battery charging. Power Switch (S2) - automatic power source control for power circuits (AC or Battery). Front Panel Figure 2-6: Mennen Medical Ltd. Panel Board PN: The front panel board includes the following: Mennen Medical 2-9

26 Hardware Description VitaLogik 4000/6000 Service Manual Upper LEDs alarm connector - 4 wire connector plug to Mennen Medical Ltd. LED board PN: The interface includes power lines and three colors (yellow/red/cyan) LED signals. Panel Connector - 20 pin connector for interfacing to the power splitter board. This interface includes the following: DC Power Optical switch signals NIBP, Print and Silence switches Main and ESC switches Main power ON/OFF switch LED indicators for the following items: On/Off switch AC line Battery status Alarm led status Optical switch Connector Following is a basic block diagram of the information flow. Figure 2-7: Basic Block Diagram 2-10 Mennen Medical

27 VitaLogik 4000/6000 Service Manual Front End Board Inputs (CIPAM) The Front End is a single acquisition board (PN: XXX) charged with acquiring and analyzing all monitor signals and transfer them to the main computer via a LAN interface for displaying, saving etc.. The Front End board includes the interfaces for the following parameters: 3/5/10 Lead ECG Respiration 4 x Invasive blood pressure. SpO2 - Masimo/Nellcore Cardiac Output Temperature x 2 NIBP EtCO2 UIM x 2 Aux channels Note: Parameter details and specifications can be found in the VitaLogik 4000/ 6000 User Manual. Mennen Medical 2-11

28 Hardware Description VitaLogik 4000/6000 Service Manual ECG / RSP 2BP / CO Figure 2-8: Data Aqusition Block Diagram The ECG is measured using a multi-lead cable via electrodes attached to the patient s chest. Monitoring the ECG produces a continuous waveform of cardiac electrical activity to enable an accurate assessment of a patient s current physiological condition. With the ECG, you can use a 3, 5 or 10-lead electrode ECG cable set to display up to seven selectable ECG leads in up to three channels. The Dual BP and CO/2TMP input is housed in the VitaLogik 4000/6000 front panel, where it occupies a single slot. It supports two channels that enable it to 2-12 Mennen Medical

29 VitaLogik 4000/6000 Service Manual simultaneously monitor two BPs as well as Cardiac Output or two body temperatures. Blood Pressure Principle of Operation Blood pressure usually is considered to be the pressure in the brachial arm artery at heart level. There is a maximum and minimum pressure that occurs in this artery due to the heart action. The maximum pressure, called systolic pressure, occurs at the completion of contraction of the heart. The minimum pressure, called diastolic pressure, occurs at the completion of the resting (repolarization) phase of the heart. Pulse pressure is the difference between systolic and diastolic pressures. As one of the physiological variables that can be quite readily measured, blood pressure is considered a good indicator of the status of the cardiovascular system. The indirect (noninvasive) method of blood pressure measurement has certain disadvantages in that it does not provide a continuous recording of pressure variations and its practical repetition rate is limited. Furthermore, only systolic and diastolic arterial pressure readings can be obtained, with no indication of the details of pressure waveform. The indirect method is also somewhat subjective and often fails when the blood pressure is very low (as would be the case when a patient is in shock). Direct (invasive) blood pressure measurement, on the other hand, do provide a continuous readout of the blood pressure waveform and are considerably more accurate than the indirect method. They require, however, that a blood vessel be punctured in order to introduce the sensor. Blood pressure is measured via a fluid-filled catheter and dome pressure transducer. The active element of a typical transducer consists of a thin membrane and resistive element, which changes in resistance as the membrane deflects with pressure. Impressing a fixed voltage across the resistive element derives the blood pressure signal. The amount of current that flows through the variable resistance element, once sealed, is considered the blood pressure signal. The blood pressure signal than pass through a low pass filter to cut noise signals, than to the analog to digital converter (ADC). The ADC serial data output pass to the P&SB for further processing and data storage. Cardiac Output Principle of Operation Cardiac output is perhaps the most important parameter of the entire circulatory system because it determines the blood flow to all issues of the body. Diseases of the heart can result in a decrease of cardiac output leading to an inadequate nutrition of the cells of the body. Cardiac output may be expressed as the product of heart rate and volume of blood pumped per beat of the heart, assuming a consistent stroke volume: Cardiac Output = (Heart Rate) (Stroke Volume) Mennen Medical 2-13

30 Hardware Description VitaLogik 4000/6000 Service Manual Cardiac output or blood flow is directly proportional to mean blood pressure and inversely proportional to peripheral resistance. Cardiac Output by Thermodilution An exact amount of a solution of known temperature is injected into the right atrium or superior vena cava and the resultant change in blood temperature is detected by the thermistor in the pulmonary artery. Cardiac output is inversely proportional to the integral temperature change. The following equation is used to compute cardiac output by the thermodilution method: ( 108. )( Ct)( 60) Vi( Tb Ti) CO.. Tb() t dt 0 Temp NIBP Where: C.O.= Cardiac Output (ml/min) 1.08= Correlation constant Ct= Correction for injectate temperature rise (from the table) 60= Seconds/minute Vi= Volume of injectate Tb= Initial blood temperature (C) Ti= Initial injectate temperature (C) The ADC converts the analog signals from the CO catheter and the injection bath to digital signals. The ADC serial data output pass to the P&SB for further processing and data storage. Temperature is detected either through the Temp. input or from the CO/2 Temp. input when Cardiac Output (CO) is not being monitored. Non-Invasive blood pressure is detected in the NIBP. Pressure is measured using a cuff and is derived by the oscillometric method. The cuff is automatically inflated and systolic, diastolic and mean pressures are derived as the cuff deflates. Adults and neonates may be monitored. The VitaLogik bedside monitor can be configured to take NIBP readings in one of three modes: manual, automatic and stat mode. In the manual mode, a single NIBP measurement is taken as required. In the automatic mode, measurements are repeated at regular, pre-set intervals. In the stat mode, blood pressure is measured as many times as possible during a period of five minutes Mennen Medical

31 VitaLogik 4000/6000 Service Manual Readings can be logged in the patient s chart. Note: During NIBP operation, isolated readings may be displayed which indicate a significant change in blood pressure values. This may have been caused by a change in the patient s actual pressure or by a movement artifact, muscle tremors, or a weak pulse. Take a manual ausculatory blood pressure to verify the patient s pressure before administering any medication. Note: The width of the cuff should be 50% of the limb circumference or 2/3 of the upper arm length. The inflation part of the cuff should be long enough to encircle 50-80% of the limb. Note: Inflation pressures used in adult NIBP monitoring are not appropriate for neonates. Also, neither the pediatric nor the adult cuff is suitable for neonates. Limitations of the Oscillometric Method The patient s actual condition at the time of measurement has a direct bearing on the reliability of that measurement. When taking an NIBP measurement, a regular arterial pressure pulse is sought. When a patient s condition makes that pulse hard to detect, it may take longer to get a reading and the reading may be unreliable. In some cases, the patient s condition may make it impossible to take a reading. The following conditions may interfere with the oscillometric method: Heart-Lung Machine - It is impossible to take an NIBP reading when a patient is connected to a heart-lung machine. Arrhythmia - An irregular heartbeat caused by cardiac arrhythmia may cause the NIBP reading to be unreliable or impossible to perform. Again, measurement time is increased. Changes in Pressure - If a patient s blood pressure changes rapidly over the period of time during which the arterial pressure pulses are being analyzed for the reading, the measurement will be unreliable and may be impossible to obtain. Patient Movement - Patient movement, shivering, tremors, and convulsions may interfere with detection of partial pressure pulses. NIBP measurements may be unreliable or impossible to perform. In any case, measurement time is increased. Heart Rate Extremes - An NIBP reading cannot be taken if the patient s heart rate is less than 40 BPM or more than 300 BPM. Mennen Medical 2-15

32 Hardware Description VitaLogik 4000/6000 Service Manual Severe Shock - Reduced blood flow to the extremities due to shock or hypothyroid causes reduced pulsations of the arteries. This causes the NIBP readings taken under such conditions to be unreliable. Limitations for Neonates - Due to the small size of arteries in premature babies and in neonates, it becomes more difficult to sense the arterial pulse. To achieve optimal results in Neonatal NIBP: 1. Use the shortest practical tube length possible between the NIBP input and the Cuff. Use approximately 1.5 meters as a practical compromise. 2. Do not tighten the Cuff on the baby s limb. Leave a space of approximately 5 mm between the Cuff and the limb. Preparatory Checklist Make sure that the NIBP monitoring is activated. Select the correct cuff size for the patient. Cuff width should be either 40% of the arm s circumference or 2/3 the upper arm length. If you are unsure whether the cuff is large enough, use a larger cuff. Connect the fitting at the end of the pressure hose to the corresponding fitting on the cuff and the monitor. Ensure that the cuff is not open and that fittings are tight to prevent air leaks. No reading can be obtained when cuff is open. Note: 1. An incorrect cuff size, especially a normal cuff on obese patients, may result in incorrect readings. 2. Always take care to avoid compression or restriction of pressure tubes. WARNING!!! Do not perform NIBP measurements on patients with sickle-cell anemia or any condition involving skin damage. Do not apply the cuff to a limb where an intravenous infusion or catheter is inserted. The infusion is slowed or blocked during cuff inflation, and tissue damage may result around the catheter. Due to the risk of hematoma in the limb where the cuff is placed, use clinical judgment when deciding whether or not to perform automatic blood pressure readings on patients with severe blood clotting disorders Mennen Medical

33 VitaLogik 4000/6000 Service Manual It is important when measuring blood pressure in neonatal patients, to make sure that the VitaLogik is correctly configured for neonatal readings, due to the lower overpressure level used. Leakage Test The NIBP leakage test procedure is described in Addendix D of the VitaLogik 4000/6000 User Manual. WARNING!!! Never perform a leakage test when the cuff is attached to a patient SPO2 (Nellcor) SPO2 (Masimo) The SPO 2 measures the patient s arterial oxygen saturation and pulse rate. Arterial oxygen saturation is the percentage of oxygenated hemoglobin in relation to the total hemoglobin. For example, if 95 percent of the hemoglobin molecules in the arterial red blood cells combine with oxygen, the blood has an oxygen saturation of 95 percent. The Sp02 numeric value represents the percentage of hemoglobin molecules that have combined with oxygen molecules to form oxyhemoglobin. The SPO 2 measures oxygen saturation using the Pulse Oximetry method. This continuous, non-invasive method measures the light absorption in the patient s tissue (for example, using a finger probe) to a receiver on the other side. Absorption by tissue and venous blood is constant. However, the blood flow in the arterioles is pulsate and, therefore, varies with time. Measuring light absorption during pulsation can derive oxygen saturation of the arterial blood. The VitaLogik provides continuous monitoring of arterial oxygen saturation expressed as a percentage. Moment-to-moment variations in capillary filling are presented as a continuous pulse waveform. The pulse rate is derived from the pulse waveform. The SpO2 input measures the patient s arterial oxygen saturation and pulse rate. Arterial oxygen saturation is the percentage of oxygenated hemoglobin in relation to the total hemoglobin. For example, if 95 percent of the hemoglobin molecules in the arterial red blood cells combine with oxygen, the blood has an oxygen saturation of 95 percent. The SpO2 numeric value represents the percentage of hemoglobin molecules that have combined with oxygen molecules to form oxyhemoglobin. The SpO2 input measures oxygen saturation using the Pulse Oximetry method. This continuous, non-invasive method measures the light absorption in the patient s tissue (for example, using a finger probe). Mennen Medical 2-17

34 Hardware Description VitaLogik 4000/6000 Service Manual Microstream Absorption by tissue and venous blood is constant. However, the blood flow in the arterioles is pulsatile and, therefore, varies with time. Measuring light absorption during pulsation can derive oxygen saturation of the arterial blood. The VitaLogik provides continuous monitoring of arterial oxygen saturation expressed as a percentage. Moment-to-moment variations in capillary filling are presented as a continuous pulse waveform. The pulse rate is derived from the pulse waveform. The End Tidal CO2 (EtCO2) input together with the Oridion sidestream CO2 control) sensor provides CO2 measurement data to the MPU through the front panel. Principles of Operation Intended Use VitaLogik s End Tidal CO2 (EtCO2) input uses Non-Dispersive Infrared Spectroscopy (NIDS) with single beam, single detector and fully chopped ratiometric measurement with spinning gas cell. The EtCO2 input uses infrared (IR) light to measure the concentration of CO2 in a sample of respiratory gas. It then measures the inspiratory and end tidal (end expiratory) CO2 levels (inco2 and EtCO2) and converts the data to percentages or mmhg. Changes in carbon dioxide levels are displayed graphically as a CO2 waveform. The MicroStream EtCO2 uses sidestream monitoring with low sampling volume of 50 ml/min. In Sidestream monitoring, a tube is attached to the patient s airway circuit. Respiratory gas is then drawn into the tube and sent to an IR sensor in a separate measuring device. This method can be used on patients who are not intubated. The unit performs automatic barometric pressure compensation to adjust the CO2 reading for changes due to pressure fluctuations. The barometric pressure is determined continuously in the mainstream mode and during power-up in the sidestream mode. The system incorporates an internal pressure transducer that measures barometric pressure continuously. The EtCO2 value is given in mmhg or as a percentage. The effects of humidity are eliminated by a Nafion dryer tube and a filter in the sampling set. The unit automatically senses when a sampling tube is installed and begins drawing a gas sample for sidestream operation. The VitaLogik s End Tidal CO2 (EtCO2) Microstream, input is intended for use with both intubated and non-intubated patients. The input uses real-time data to calculate and display the numerical values for End-tidal carbon dioxide (EtCO2), inspired carbon dioxide (inco2) and Respiration rate Mennen Medical

35 VitaLogik 4000/6000 Service Manual Warning!! Airway Adapter The airway adapter is intended for use on a single patient and should be discarded after use. Do not attempt to clean with air or Oxygen. Microstream User Calibration Calibration Interval Calibration should be performed at 1200 operating hours after first factory calibration and then annually or after 4000 operating hours, whichever comes first. When the operating hours criterion is detected the monitor will display, in the EtCO2 area, the message: "Calibration Alert" Calibration Personnel and Equipment Calibration should be performed by qualified service personnel. Calibration must be performed with a manufacturer authorized Calibration kit containing 5% CO 2 gas and the connecting means. A manufacturer approved Calibration kit can be purchased from Scott Medical (part number ORFBD). The kit consists of: Calibration Gas containing 5% CO 2, 21% O 2 Tubing Adapter Calibration FilterLine Start calibration only after the device has been operating for at least 20 minutes in a normal operating mode (not in Standby mode) and connected to a FilterLine. Note: Prior to calibration, verify that the FilterLine supplied with the Calibration Kit is firmly attached. Figure 2-9: Calibration kit Mennen Medical 2-19

36 Hardware Description VitaLogik 4000/6000 Service Manual Calibration Procedure 1. In System Setup select EtCO2 Calibration. This will open the Calibration panel with the last calibration date at the bottom of the panel 2. Set the gas concentration to 5% 3. Connect the calibration gas and click on Start Calibration. 4. The message "Calibration in Progress" will appear 5. At the end of the process the message "Calibration has succeeded" will be displayed. 6. Stop the gas flow. Note: If the gas flow is irregular or if the gas concentration is not same as set a Calibration failure message will be displayed on the panel UIM - Universal Input Figure 2-10: Calibration has Failed message Two Universal inputs are located at the left side of the monitor. The UIM input enables connection of vendor devices to the VitaLogik monitor Table 1: Vendor device communication parameters Device Baud Rate No. of Bits Parity Stop bits Interface cable p/n Anesthesia gas analyzer No Ventilator Drager Evita Even Mattise Ventilator none Mennen Medical

37 VitaLogik 4000/6000 Service Manual Table 1: Vendor device communication parameters BIS none Remote Keypad (Optional) The remote keypad and QuicKnob enable the user to operate the VitaLogik system through an external cable connected to the Remote port. The Remote port is located on the rear panel of the unit and enables performing additional functions. Figure 2-11: Remote Keypad General View The Remote keypad consists of 12 fixed keys, QuicKnob, buzzer, MCU and RS-232 interface (see in Figure 2-11:). The connection to the unit includes power for the Remote keypad that is provided via the Splitter board (+5V). Mennen Medical 2-21

38 Hardware Description VitaLogik 4000/6000 Service Manual 2-22 Mennen Medical

39 CHAPTER 3: SETTING UP THE SYSTEM This chapter includes instructions for configuring the following VitaLogik 4000/6000 parameters: Store Diagnostic File Date & Time Setup General Settings Alarm Volume & Controls Disable Touch Screen (in Touch screen option) Sound Event Default Alarm Limits Event Setup NIBP Leakage test Vital Sign (VS) Setup Parameter Hierarchy Monitor Profiles Admit by Default Report Setup Automatic Reports Network Setup Recorder Setup Trends Setup Quick Keys Tabular Charts Software Version Mennen Medical 3-1

40 Setting Up the System VitaLogik 4000/6000 Service Manual Change Password Check Disk Permission Editor Remove Saved Patients Touch Screen Calibration Demo Mode Build Master Configuration Copy Configuration From Introduction System Setup enables configuration of global controls, such as setting the date and time or defining the default language for the VitaLogik 4000/6000. It is also used to define default values for all monitored vital signs. The VitaLogik 4000/6000 is configured at the factory with factory default values which can be changed as needed by your hospital or department. Access to the System Setup menu is limited to authorized personnel such as a System Administrator or Hospital Biomedical Engineer possessing a pre-defined password. Accessing the System Setup Menu From the System Setup menu, you can access the following dialog panels: Store Diagnostic File Use this panel to store a diagnostic file in the event of a malfunction or failure of the VitaLogik 4000/6000. In case of failure download the diagnostic files and send them to the Service dept. of Mennen Medical. Date & Time Setup Use this panel to set the date and time and to set time display formats for the VitaLogik 4000/6000 bedside monitor. See Date & Time Setup. General Settings Use this dialog panel to define general information, such as the hospital name and parameters controlling general panel behavior. See General Settings. Disable/Enable Touch Screen Monitors and Ensemble equipped with Touch Screen have a toggle to disable the touch screen feature for cleaning 3-2 Mennen Medical

41 VitaLogik 4000/6000 Service Manual Alarm Volume & Controls Use this panel to define general alarm-related parameters, such as alarm volume, color scheme, etc. See Alarm Volume and Controls. Sound Event Use this panel to assign audio messages to various events. See Sound Event. Default Alarm Limits Use this panel to set new default alarm limits for all monitored vital signs. See Default Alarm Limits. Event Strip Use this panel to prepare or modify a list of Event Strip Labels, and to define the response of the VitaLogik 4000/6000 to the activation of the Mark Event key. Leakage Test Use this panel to perform Leakage test of the NIBP module and cuff. The test is password protected (after Event Strip) Vital Sign (VS) Setup Use this panel to set default parameters and units for each vital sign available for monitoring in the VitaLogik 4000/6000 bedside monitor. These include alarm parameters and other parameters specific to the monitored vital sign. See Vital Sign (VS) Setup. Parameter Hierarchy Use this panel to present the Hierarchy of VS parameters. See Parameter Hierarchy. Monitor Profiles Use this panel to preset and label monitoring profiles. See Monitor Profiles Setup. Admit By Default Use this option to return the VitaLogik 4000/6000 to default parameters. See Admit by Default. Report Setup Use this dialog panel to define and edit report formats. See Report Setup. Automatic Reports Use this panel to set the time and type of automatic report printout on the network printer. See Automatic Reports. Network Setup Mennen Medical 3-3

42 Setting Up the System VitaLogik 4000/6000 Service Manual Use this panel to configure devices, such as monitors, printers, etc. on the network. See Network Setup. Recorder Setup Use to define default recording options. See Recorder Setup. Trends Setup Use to set the scales of the graphic trend for each vital sign. See Trends Setup. Quick Keys Use this panel to prepare QuicKeys for single button access to monitoring functions. See Quick Keys Setup. Tabular Charts Use this panel to create User Defined Charts, customized to the specific needs of a unit or patient. See Tabular Charts. Software Version The Software Version panel displays the software version of all inputs and manages loading of new software to the vital signs inputs. See Software Version Change Password Use this panel to define a new password for entering the System Setup menu. See Change Password. Check Disk Check disk is a technical procedure resets all parameters to factory default settings.. This panel is for use by authorized service personnel only. See Check Disk. Permission Editor Use this panel to define which VitaLogik 4000/6000 menu options are available. See Permission Editor. Remove Saved Patients Use this panel to access a list of all the patient data saved by the VitaLogik 4000/ 6000 and to remove some or all of the saved files. See Remove Saved Patients. Touch Screen Calibration Note: Monitors and Ensemble equipped with Touch Screen have a key to start Touch screen calibration See Touch Screen Calibration. In the System Setup menu, some of the dialog panels can be accessed only when the VitaLogik 4000/6000 is not currently monitoring a patient. In this case, if a patient is being monitored, he/she must first be discharged, 3-4 Mennen Medical

43 VitaLogik 4000/6000 Service Manual meaning that monitoring of that patient must be stopped. On the other hand, some of the dialog panels can be accessed only when a patient is currently being monitored. Demo Mode See Demo Activation. Build Master Configuration Use this key to store a configuration for future copying to different Envoy monitors. Copy Configuration From Use this key to copy a configuration from a master configuration monitor. Note: See Table 3-1 below for the access capabilities in the dialog panel: Table 3-1: Access Capabilities in the Dialog Panel Patient Discharged Patient Monitored Store Diagnostic File Date & Time Setup ++ General Settings ++ Enable Touch Screen Alarm Volume & Controls Sound Event Default Alarm Limits ++ Event Setup Leakage Test CO2 Calibration ++ Vital Sign (VS) Setup ++ Parameter Hierarchy ++ Monitor Profiles ++ Admit By Default ++ Report Setup Automatic Reports Network Setup Mennen Medical 3-5

44 Setting Up the System VitaLogik 4000/6000 Service Manual Table 3-1: Access Capabilities in the Dialog Panel Recorder Setup Trends Setup Quick Keys ++ Tabular Charts Software Version ++ Change Password Check Disk ++ Permission Editor Remove Saved Patients Touch Screen Calibration Demo Mode ++ - Build Master Configuration ++ Copy Configuration From ++ To discharge a patient: 1. Press the Setup key to access the ADT menu and select Discharge Patient. 2. A window appears informing you that all patient data will be erased and asking you to confirm patient discharge. 3. Press Discharge to stop patient monitoring and erase patient data. The VitaLogik 4000/6000 bedside monitor is now ready to admit a new patient or Press Cancel to cancel the discharge, close the panel and return to the main screen display. Once the patient has been discharged, all defined Monitor Profile Admit Keys appear on the now blank display screen. You are now ready to access the System Setup menu. To access the System Setup menu: 1. Press Setup button, the Setup menu is displayed. 2. Turn the QuicKnob until the System Setup button is highlighted and press. The System Password panel appears requesting the password. 3-6 Mennen Medical

45 VitaLogik 4000/6000 Service Manual Figure 3-1: The System Password Dialog Box 3. Enter the correct password. The System Setup menu is displayed. Date & Time Setup This panel enables you to set the date and time and to set time display formats for the VitaLogik 4000/6000 bedside monitor. Time Server (implemented in VitaLogik 4000/6000 SW Version 3.4). The Date and Time setting of the VitaLogik 4000/6000 monitor can be set on each monitor individually, or, if the monitors are connected to a LAN, by one device defined as the Time Server. Only ONE device on the network can function as the Time Server. If one of the devices is set to become the Time Server, then the Date and Time setting on all other devices will be disabled and the time can be changed only on the Server device. The Time Server sends a time signal every 30 seconds to all the devices on the network. Mennen Medical 3-7

46 Setting Up the System VitaLogik 4000/6000 Service Manual VitaLogik 4000/6000 monitors that are in the Discharge mode, reset their time to the time sent by the server. VitaLogik 4000/6000s that are in the process of monitoring a patient, will not respond to the Time Server. The Enguard always responds to the Time Server and resets its own time, but this will not affect the device that the Enguard is viewing. The Ensemble always responds to the Time Server and will reset its own time, but this will not affect the device/s it is viewing. To create a Time Server: 1. Select System setup (password protected) in the Setup menu. 2. Select Date & Time Setup in the System setup menu. 3. In the Time Server Master, set the Time server Master to ON (Green). 4. Use the Current Date & Time box to change the time and/or date. Note: If there is a Time Server Master on the network, then: All other devices will show the Device ID and the Device name of the Time Server Master on the Date & Time panel. The Current Date & Time box will show the date & time, but will be dimmed and not accessible. The Date and Time Setup Panel is divided into 3 major areas (note - the function Daylight Saving Date is not active): ²Current Date and Time As the name implies, the current date and time are set in this area. ²Shift End Time Specify the end times for each work shift according to what is standard in your department or hospital. The end time defined for the shift also indicates the start time of the following shift. For example, if the Day shift ends at 15:00, this indicates that the Evening Shift begins at 15:00. The Date and Time Setup panel provides space for three shifts: Day, Evening and Night. ²Display Format In the Display Format area, you define how the time and the date appear on the main screen display. 3-8 Mennen Medical

47 VitaLogik 4000/6000 Service Manual Figure 3-2: The Date & Time Setup Panel To set the date and time display formats: 1. Select the Date textbox. A calendar panel is displayed. 2. Using the QuicKnob, select the day, month and year. Turn the QuicKnob until the monthly calendar is highlighted and press. Now the calendar is selected. Begin turning the QuicKnob again to move through the days of the month. When the correct day is highlighted, press the QuicKnob. Turn the QuicKnob until one of the arrows on the month indicator is highlighted. The arrow on the right, when selected, advances the months, the arrow on the left goes back. Press to select the correct month. Turn the QuicKnob until the year is highlighted and press to select. Turn the QuicKnob clockwise to advance in 10-year steps. Press and turn the QuicKnob to advance in one-year steps. Turn the QuicKnob counterclockwise in the same manner to go back. 3. Select Cancel Changes to close the panel without applying the changes or Select Accept to apply the changes and return to the Date and Time Setup panel. 4. Select the Time area. An Hour/Minute panel is displayed. Select Hours/Minutes and turn the QuicKnob to set the time. Turning the QuicKnob clockwise goes forward in time. Turning the QuicKnob counter-clockwise goes back. Mennen Medical 3-9

48 Setting Up the System VitaLogik 4000/6000 Service Manual Once you have set the minutes/hours, you must press the QuicKnob again to go on to the next parameter. 5. Under Shift End Time, select Day, Evening, or Night. The Hour/Minute panel is displayed. Enter the shift end time in the same way you set the time for the Time area. 6. Under Display Format: Select Date Format. A drop-down list is displayed. There are three available options: ddmmmmyy, ddmmyyyy, or mmddyy. The factory default is mmddyy. Select Separator. The Keyboard panel is displayed. Select a character to be used by the system as a separator between elements of a date (such as. or / ). Only a single character can be selected. 7. Under Display Leading Zeros, select the Date or Time checkbox to activate display of a leading zero (for example, 01/01/98, or 01:30). Clear the checkbox to deactivate. Select the 24 Hour Display checkbox to activate display of time in a 24- hour format (00:00-24:00). Clear the checkbox to have time displayed in a 12-hour format (12A.M. AND 12 P.M.) When the 24 Hours Display checkbox is cleared, the Suffix (am/pm) checkbox is enabled. Select to display the suffix AM or PM after the time when time is displayed in 12-hour format. Clear the checkbox to remove the suffix. 8. Select Discard Changes to close the panel without applying any changes, or Select Main Screen to apply the changes and close the panel. General Settings Use this dialog panel to set parameters that: affect general panel behavior (timeouts, etc.) define general information, such as the name of the hospital, set the interface language for the panels, default measurement units, etc. set the Remote View panel, between full panel and 1/3 panel deactivate the fan 3-10 Mennen Medical

49 VitaLogik 4000/6000 Service Manual Figure 3-3: The General Settings Dialog Panel The General Settings panel is divided into five areas; General, Units, Timeouts, Hospital Name and and Special Features General Language Frequency Under the General area, you can configure the following parameters: Defines the language used for display purposes and as the interface language on the dialog panels. You can choose from a provided list of languages - currently, English and French are supported. The factory default language is English. Defines the monitor frequency in Hertz (Hz). You can select either 50 or 60Hz. Freeze Time Screen Orientation By pressing the Freeze button on the front panel, a user can stop updating of all waveform information displayed on the main screen for a period of time which is defined here. The available freeze times are 10, 20, or 30 seconds. Defines the location in which numerical values and waveforms are displayed on the main screen. There are two options available; Right-Left or Left-Right. Mennen Medical 3-11

50 Setting Up the System VitaLogik 4000/6000 Service Manual Units Under the Units area, you can define the default measurement units used for patient s height and weight. Height The units of measurement available are; CM (centimeters), or IN (inches). The factory default is CM. Weight Timeouts Hospital Name Available units of measurement are; Kilograms (Kg), Grams (Gr), Pounds (Lb), or Pounds and Ounces (Lb & Oz.). The factory default is Kg. Under the Timeouts area, you can define the period of time which elapses from the last time an action was performed in a panel or menu to the time it is automatically closed by the system and disappears from the main screen display. The timeout can be defined for each type of panel and for menus. The following panels can be configured for timeout; Vertical Panels, Horizontal Panels, Full Panels, Confirmation Panels, Keyboard Panels and Menus. The available timeout options for all types of panels are; None (panel or menu will not automatically close) 1 min., 2 min., 4 min., or 10 min. The factory default timeout for all panel types is 2 min. Under Hospital Name, you can enter the name of the hospital. This name will then appear in the Header area on the main screen display. Setting the General Settings Panel The General Settings panel can be accessed from the System Setup menu, or, if you are currently in another System Setup panel, from the drop-down menu list to the right of the back menu button ( System Setup ) on the top of the screen. To define General Settings: 1. Under General, select the required Language, Frequency, Freeze Time, and Screen Orientation. 2. Under Charts, select the Default Frequency. 3. Under Units, select the required default measurement units for the patient s height and weight Mennen Medical

51 VitaLogik 4000/6000 Service Manual 4. Under Timeouts, select the required timeouts for the various panels and menus. 5. Under Hospital Name, select the Hospital Name text box. The alphanumeric keyboard appears. Using the QuicKnob, enter the name of the hospital and close the keyboard. 6. Select Discard Changes to close the panel without applying any changes or select Main Screen to apply your changes and return to the blank screen. Alarm Volume and Controls Use this panel to define general alarm-related parameters, such as alarm volume, color scheme, etc. These parameters are defined for each alarm priority level, enabling the alarm response for each alarm priority to be unique. For a more detailed description of Alarms, including alarm priority and notification, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operating Manual. Figure 3-4: The Alarm Volume & and Controls Panel The panel is divided into two major areas. Parameters common to all alarms are located on the left of the panel. These parameters include; Alarm Color Scheme, Suspension Time, Silence Time, and SpO2 Alarm Tone. On the right side of the panel, parameters specific to each alarm priority are set. These fields include; File Name, Repeat Every (sec), End and Init Volume, Delay, Step and Increase Mennen Medical 3-13

52 Setting Up the System VitaLogik 4000/6000 Service Manual Enable Alarm Off Silence Time SpO2 Alarm Tone Priority Level File Name Intensity. These parameters remain disabled until an alarm priority is selected. A description of each Alarm Volume and Control panel field follows. Depending on the regulation of each country or institution, the option to be able to put alarms off can be allowed by activating Enable Alarm Off or can be prevented by not activating Enable Alarm Off. When the Alarms Off key on the VitaLogik 4000/6000 front panel is pressed, all detection of clinical alarms is disabled for a pre-defined period. Technical alarms are still detected and visually displayed in the appropriate area on the main screen. Under Suspension Time, you can select the period of time that Alarms Off is in effect when the Alarms Off button is pressed. The available options are; 30, 60, 90, and 120 seconds. Alarms Off can be configured to be permanent (where allowed by law). When the Enable Alarms Off checkbox is selected, the option Permanent is added to the available options under Suspension Time. The factory default suspension time is 60 seconds. When pressed, the Silence key terminates any audio alarms active at that moment for a predefined period which you can set under Silence Time. The available options are; 30, 60, 90, and 120 seconds. The factory default is 30 seconds. Defines whether SpO2 alarms are indicated with the same standard tone as other alarms of the same priority or whether they are indicated with a tone unique to SpO2. The available options are Standard, or Unique. Defines the alarm priority level for which you wish to define alarm volume parameters. The available priority levels are; C1, C2, C3, C4, Low HR, High HR, T1, T2, and SpO2. When you select an alarm priority level, the following fields are enabled: Defines the.wav file (sound file) which supplies the alarm tone for the selected alarm priority. The available options are; C1.WAV, C2.WAV, C3.WAV, T1.WAV, T2.WAV and SpO2.WAV. Each file corresponds to the default alarm priority of that name. Additional *.WAV are available for user selection Mennen Medical

53 VitaLogik 4000/6000 Service Manual CAUTION! If you select for C1, C2, C3, C4 a different *.wav file than the default, it is recommended to use the Inter-burst Interval per the following standard requirement: Table 3-1: Priority Standard Priority C1 C2 Inter-burst Interval 0 to 5 sec 2.5 to 30 sec C3 >C2 C4 > or = C3 Repeat Every (sec) Defines the time period after which an alarm is repeated (every 10 seconds, every 20 seconds, etc.). The available options are; 10, 20, 30, 40, 50, and 60 seconds. You can press the Play button to sample the selected alarm tone. The factory default values are as follows: C1, C2, C3 and C4-20 seconds T1-30 seconds T2-60 seconds SpO2-5 seconds Note: Note:A repeat time of 0 seconds indicates that the alarm tone is continuous. End and Init Volume Delay (sec) Sets the volume of the alarm tone when the alarm is first triggered (Initial Volume) and the maximum volume the alarm can reach until responded to (End Volume). The alarm volume gets progressively louder until it reaches the maximum level, then remains at that maximum level until someone responds. End and Init Volume is set on a scale of Sliders are provided to set the volume. The left slider controls the end (maximum) volume and the right slider controls the init (initial) volume. Sets the time delay (in seconds) until the alarm increases in volume. Delay is set on a scale of Mennen Medical 3-15

54 Setting Up the System VitaLogik 4000/6000 Service Manual Step (vol) Increase Intensity Sets the increase in alarm volume after each delay. Step is set on a scale of 0-20 (in units of 4). Select this checkbox to enable increase in alarm volume. If this option is disabled, the alarm tone remains at its initial level of volume. Setting the Alarm Volume & Control Panel You can access the Alarm Volume & Control panel from the System Setup menu. To set Alarm Volume and Controls: 1. Select the required Alarm Color Scheme. 2. Select or clear the Enable Alarm Off checkbox, as required. 3. From the drop-down menu, select the required Suspension Time for the alarm. 4. From the Silence Time drop-down menu, select the alarm silence time. 5. From the SpO2 Alarm Tone drop-down menu, select the type of alarm tone for SpO2 alarms. 6. From the Priority Level drop-down menu, select the alarm priority level to be configured. 7. Print the Message you would like to receive in the event of an external device alarm and press accept. 8. From the File Name drop-down menu, select the alarm tone.wav file for the selected priority level. 9. Set the Repeat Every (sec) scale by moving the slider to the required time period. Select Play to hear a sample of the alarm tone. Select Stop to end the playback of the alarm tone. 10. Under End and Init Volume, set the left slider to the maximum alarm volume and the right slider to the initial alarm volume. 11. Set the Delay slider to the required delay (in seconds). 12. Set the Step slider to the required number of steps for each alarm volume increase. 13. Select the Increase Intensity checkbox to activate this function. Clear the checkbox to deactivate Mennen Medical

55 VitaLogik 4000/6000 Service Manual 14. Press Discard Changes to close the panel without applying any changes made or Press Main Screen to apply the changes and exit to the main screen. Sound Event Sound events are audio messages that are heard when certain events occur. Sound files are supplied for the following events: When performing cardiac output, the message Inject now is heard at injection time (CO Inject.wav). When performing Wedge Pressure procedure (PCWP), the audio message Deflate balloon is heard when it is time to deflate the balloon (PCWPTimeOut.wav). When starting up the computer (StartUp.wav). NIBP Success (NIBP Success.wav). NIPB Fail (NIPB Fail.wav). Use the Sound Event panel to assign sound files to events at the VitaLogik 4000/6000, and to define the volume of the sound. Figure 3-5: The Sound Event Panel To configure sound events: 1. From the System Setup menu, select Sound Event. The Sound Event panel is displayed. Mennen Medical 3-17

56 Setting Up the System VitaLogik 4000/6000 Service Manual 2. In the Event list, select the event to which you want to assign an audio message. 3. Under the File Name list appear all sound files, including technical and clinical alarms. Select the sound file that will supply the sound for the selected event. 4. Under Volume, set the required volume, and click Play to test the sound file. 5. Click Main Screen to save settings and close the dialog box. Note: Factory default setting is for Volume = 60 for Inject Now and Deflate Balloon, and Volume = 0 for all other events. Default Alarm Limits Each vital signs available for monitoring appears as a button on the Default Alarm Limits panel. The current alarm limits for each vital sign are also displayed on the button. Select a vital sign. The appropriate scale is displayed where you can set new default alarm limits for the selected vital sign. You can select Fixed or Dynamic Alarm Limits. When Dynamic Alarm Limits is selected, the alarm limits change according to the patient s vital signs, when Set to Default is activated Mennen Medical

57 VitaLogik 4000/6000 Service Manual Figure 3-6: The Default Alarm Limits Panel To set default alarm limits: 1. Turn the QuicKnob until the required vital sign is highlighted and press. The appropriate options and scale are displayed on the right side of the dialog panel. 2. Under Default Type, select Fixed or Dynamic. 3. Select Pressure Type (for BPs). 4. Select the required limit (high or low), highlight the required slider with the QuicKnob and turn the QuicKnob clockwise or counter-clockwise to move the slider to the required alarm limit. Press in the QuicKnob to set the limit. 5. When done, press the Escape button on the front panel to return to the Vital Sign list. 6. Repeat steps 1-5 until you have set alarm limits for all required vital signs. Mennen Medical 3-19

58 Setting Up the System VitaLogik 4000/6000 Service Manual 7. Select Discard Changes to close the panel without applying any changes, or Select Main Screen to apply the changes and close the panel. Event Setup Use the Event Setup panel to create or modify Event labels, and to determine the response to activation of the Mark Event key. Figure 3-7: The Event Setup Panel To Insert a new label: 1. Select the Event Strip List in the Mark Event panel - it turns green 2. Select a Label 3. Select Insert - this opens the virtual keyboard on the VitaLogik 4000/6000 screen 4. Prepare a label and press Accept 5. The new label is added above the selected label (item 2) To Delete a label: 1. Select the Event Strip List in the Mark Event panel - it turns green 2. Select a Label 3. Select Delete this deletes the label To Edit a label: 1. Select the Event Strip List in the Mark Event panel - it turns green 3-20 Mennen Medical

59 VitaLogik 4000/6000 Service Manual 2. Select a Label 3. Select Edit - this opens the virtual keyboard on the VitaLogik 4000/6000 screen 4. Edit the label and press Accept 5. The new label will replace selected label To set the response to the Mark Event Key: Three response options are available: Chart only Chart + Strip Chart + Strip + Label Chart A row of vital signs at the time of Mark Event activation is added to the charts. The row is by a green dot. Chart + Strip In addition to the above, an Event Strip is stored in the monitor and is labeled as Manual Event. The Event strip starts 10 seconds before the time at which you press the Mark Event Key, and ends 10 seconds after the time at which you pressed the above key. Chart + Strip + Label NIBP Leakage Test In addition to the above, a list of labels is displayed on the monitor screen. The user can now select one of the labels for the event. The event timing depends on the time of button activation and not on the time of the labeling. Use the Leakage Test to test the leakage of the NIBP module including the patient cuff. To test NIBP Leakage 1. To start the test, disconnect the cuff from the patient, rap the cuff around itself 2. Press Leakage Test --- A password panel will open Mennen Medical 3-21

60 Setting Up the System VitaLogik 4000/6000 Service Manual 3. Enter the password The test panel will open. The test panel contains a warning message : Disconnect Patient from Cuff before Testing 4. Be sure that the cuff is not connected to a patient and press Test 5. The module will inflate the cuff to a pressure of about 200 mmhg for adult cuff, or about 120 mmhg for neonatal cuff. 6. After that all internal valves will be closed and the leakage will be tested for total time of 4 minutes 7. The test results are displayed as : Pressure after 1 minute of inflation and Pressure Difference that is the pressure drop between the Initial pressure at 1 minute and the final pressure 3 minutes later. Note: If you assume that the leakage is in internal parts of the module, start the test with a cuff connected, and at about 1 minute close the cuff tube near the pressure outlet using a surgical clamp. Vital Sign (VS) Setup The Vital Signs Setup menus enable you to set default parameters for each vital sign available for monitoring in the VitaLogik 4000/6000 bedside monitor. These include alarm parameters and other parameters specific to the monitored vital sign. The Vital Signs Setup menu enables you to access the following dialog panels: ECG Setup Respiration Setup BP Setup NIBP Setup SpO2 Setup Temperature Setup CritiCool Setup EtCO2 Setup CO Setup cgas Setup CO2 Setup O2 Setup N2O Setup Agent Setup Ventilator Setup SvO Mennen Medical

61 VitaLogik 4000/6000 Service Manual Accessing the Vital Signs Setup Dialog Panels A dialog panel is provided for each available vital sign. You can access these dialog panels from the Vital Sign Setup menu. To access the Vital Sign Setup menu: 1. In the System Setup menu, select Vital Signs Setup. The menu is displayed. 2. Select the vital sign for which you want to set defaults. The appropriate dialog panel is displayed. ECG Setup This dialog panel enables you to set ECG default parameters. These parameters include: Alarm priorities for Heart Rate, Fault and Pacemaker alarms Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Heart Rate alarm priority can be set at a level of either C-1 or C-2. The factory setting is C-1. Alarm Latching For a complete description of Alarm Latching, (see User Guide for a full description of this feature. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Arrhythmia Package Enables you to determine the level of Arrhythmia analysis. The available levels are: OFF, Basic (includes Asystole, Vfib, V-tach, Run, PVC, and Pause) and Extended (includes Bradycardia, Supraventricular Tachycardia, SV Irregular Rhythm, Idioventricular Rhythm, Couplets of PVCs, Triplets of PVCs, Bigeminy, Trigeminy, R on T, PVC, Vfib, V-Tach, Multifocal PVCs, Interpolated PVCs, Asystole, Run, Pause, Premature Artrial Contraction, and Sinus Tachycardia). Arrhythmia Priority Enables you to set default priority levels for all available Arrhythmia alarms. The available alarm priorities are as follows: Mennen Medical 3-23

62 Setting Up the System VitaLogik 4000/6000 Service Manual Table 3-2: Arrhythmia Alarm Priorities Arrhythmia C-1 C-2 C-3 Default Priority Asystole X C-1 Ventricular Fibrillation (VFIB) X C-1 Ventricular Tachycardia (VTACH) X C-1 Runs of PVC beats (RUN) X X C-2 Bigeminy (BGM) X X X C-2 PVC beats per minute (PVC) X X X C-2 Pause Event (PAUSE) X X X C-3 Bradycardia (BRADY) X C-1 Tachycardia (TACHY) X C-1 Irregular Rhythm (IRREGU- LAR) Idioventricular Rhythm (IDIO- VENT) X X X C-2 X X X C-2 Couplets of PVCs (COUPLET) X X X C-2 Triplets of PVCs (TRIPLET) X X X C-2 Trigeminy (TRIGEM) X X X C-2 R on T X X X C-2 Premature Atrial Contraction (PAC) X X X C-2 Atrial Fibrillation (AFIB) X X X C-2 Supraventricular Tachycardia (SVTACH) Multi-focal PVCs (MULTIFO- CAL) X X X C-2 X X X C-2 Interpolated PVCs (INTPVC) X X X C-2 Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 3, 5, or 10 seconds. The factory default is set to 5 seconds Mennen Medical

63 VitaLogik 4000/6000 Service Manual Default Color Defines the color in which ECG vital signs and waveforms appear on the main screen display. You have a choice of 16 colors. The factory default color is yellow. Figure 3-8: The ECG Setup Panel To define ECG default parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select ECG Setup. The ECG Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the appropriate drop-down list, set the priority levels for Heart Rate, Faults and Pacer. From the Latching drop-down list, set alarm latching to On or Off. 4. Under Arrhythmia, select the Arrhythmia Package that you want. 5. Under Arr Priority: From the Arr Name drop-down list, select the type of arrhythmia for which you want to set alarm priority. From the Priority drop-down list, select the priority level for the selected Mennen Medical 3-25

64 Setting Up the System VitaLogik 4000/6000 Service Manual Respiration Setup type of arrhythmia. 6. ST analysis: Toggle on (green) if you wish ST to be functional, or off to prevent activation of ST measurement. 7. OR monitoring: Toggle on (green) if you wish to enable Electrosurhgery filter (ESF) on the top displayed ECG lead, 8. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 9. Select ECG Default Color. A panel displaying all available colors appears. Select the required color. The panel closes. 10. Select Discard Changes to close the ECG Setup panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. This dialog panel enables you to set Respiration default parameters. These parameters include: Alarm priorities for Resp Rate, Apnea and alarms Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Resp Rate, Apnea, and alarm priority can be set at a level of C-1, C-2, or C-3. The factory defaults are as follows: Resp Rate: C-2 Apnea: C-2 Alarm Latching For a complete description of Alarm Latching, see the User Guide. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 3, 5, or 10 seconds. The factory default is set to 5 seconds Mennen Medical

65 VitaLogik 4000/6000 Service Manual Apnea Reset Defines the default number of breaths the system uses to reset the apnea timer. The available options are; 2 breaths, 3 breaths, and 4 breaths. Default Color Defines the color in which Respiration vital signs values and waveforms appear on the main screen display. You have a choice of 16 colors. The factory default color is green. Figure 3-9: The Respiration Setup Panel To define Respiration default parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select Respiration Setup. The Respiration Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the appropriate drop-down list, set the default priority levels for Resp Rate and Apnea. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. From the Apnea Reset drop-down list, select the number of breaths required to reset the Apnea timer. 6. Select Respiration Default Color. A panel displaying all available colors appears. Select the required color. The panel closes. Mennen Medical 3-27

66 Setting Up the System VitaLogik 4000/6000 Service Manual BP Setup 7. Select Discard Changes to close the Respiration Setup panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. This dialog panel enables you to set BP default parameters. These parameters include general default values applicable to all BPs and default values applicable to a specific BP site (such as ART, CVP, etc.) General BP Default Parameters Alarm priorities for BP, and Fault alarms Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. BP alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Fault alarm priority can be set to T-1, or T-2. The factory default is T-1. Alarm Latching See User Guide for a full description of this feature. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 3, 5, or 10 seconds. The factory default is set to 5 seconds. Auto Scale Automatically sets the range of the BP waveform on the main screen display. After Zero enables you to specify if autoscaling is done automatically after the BP is zeroed and in which cases. The available options are; Always, Never, and Only After First Zero. Under Type, you can define the type of autoscaling implemented. The available options are; Normal, and Only Upper Limit (meaning that the lower scale limits should be kept). BP Site-Specific Default Parameters Defaults per Site Name 3-28 Mennen Medical

67 VitaLogik 4000/6000 Service Manual Defines the site for which the site-specific parameters are being set. Units Determines the units used to measure the selected BP. The available options are; mmhg, Kpa, and cm H 2 0. The factory default is mmhg. Numeric Formats Determines the numeric format in which the selected BP is displayed. When units are defined as mmhg, the available options are: SSS/DDD MMM, SSS/DDD mmm, and SSS/DDD. The factory default is SSS/DDD MMM. When units are defined in Kpa, the available options are: SS.S/DD.D, and MM.M. The factory default is SS.S/DD.D MM.M. Manual Scale Defines the default high and low scale limits for use when implementing manual scaling. The available options are as follows: High Scale - 250, 300 and 350 Low Scale: -15, -10, and 0 BP Default Color Defines the color in which BP vital signs values and waveforms appear on the main screen display. You have a choice of 16 colors. The factory default colors are as follows: ART/UA - Pink PAP - Light Blue CVP - Orange All others - Grey or Silver Mennen Medical 3-29

68 Setting Up the System VitaLogik 4000/6000 Service Manual Figure 3-10: The BP Setup Dialog Panel Displaying the Color Selection Panel To define default BP parameters 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select BP Setup. The BP Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the appropriate drop-down list, set the default priority levels for BP and Fault alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. Under Auto Scale: From the After Zero drop-down list, select Always, Never, or Only After First Zero. Under Type, select Normal, or Only Upper Limit Mennen Medical

69 VitaLogik 4000/6000 Service Manual NIBP Setup 6. From the Defaults Per Site Name drop-down list, select the BP for which you want to set site-specific default values. 7. From the Units drop-down list, select the type of units used to measure the selected BP. 8. From the Numeric Formats drop-down list, select the required format in which to display the BP reading. 9. Under Manual Scale; from the High/Low Scale drop-down lists, select the required upper/lower limits for manual scaling. 10. Select BP Default Color. A panel displaying all available colors appears. Select the required color. The panel closes. 11. Select Discard Changes to close the BP Setup panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. This dialog panel enables you to set NIBP default parameters. These parameters include setting alarm priority for both NIBP and Pulse Rate alarms, defining the default cuff type for the monitor, setting reading intervals and specifying the measurement units used to display NIBP values. Alarm priorities for NIBP Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. NIBP alarm priority can be set to C-1, C-2, or C-3. The factory default is C-1. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Cuff Type Defines the default cuff type to be used by the VitaLogik 4000/6000. The available cuff types are Adult or Neonatal. The factory default is Adult. Mennen Medical 3-31

70 Setting Up the System VitaLogik 4000/6000 Service Manual Reading Intervals Use to set defaults for T and Q intervals. Refer to Chapter 11, NIBP Monitoring in the VitaLogik 4000/6000 Operation Manual for a complete description of this concept. Units Determines the units used to measure NIBP. The available options are; mmhg, Kpa, and cm H 2 0. The factory default is mmhg. Numeric Formats Determines the numeric format in which the NIBP reading is displayed. When units are defined as mmhg, the available options are; SSS/DDD MMM, SSS/DDD mmm, and SSS/DDD. The factory default is SSS/DDD MMM. When units are defined in Kpa, the available options are: SS.S/DD.D, and MM.M. The factory default is SS.S/DD.D NIBP Default Color Defines the color in which NIBP values appear on the main screen display. You can choose from 16 available colors. Figure 3-11: The NIBP Setup Dialog Panel To set NIBP default parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed Mennen Medical

71 VitaLogik 4000/6000 Service Manual SpO2 Setup 2. From the Vital Signs menu, select BP Setup. The BP Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the appropriate drop-down list, set the default priority levels for NIBP alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. Under Cuff Type: Select the Adjustable Cuff Size checkbox to enable user access to the Cuff Type panel. Clear the checkbox Adjustable Cuff Type to remove the Cuff Type menu item from the NIBP menu. Select the default cuff type, Adult or Neonatal. 5. Under Default Reading Intervals, set the required parameters for T and Q intervals. 6. From the Units drop-down list, select the type of units used to measure NIBP. 7. From the Numeric Formats drop-down list, select the required format in which to display the NIBP reading. 8. Select NIBP Default Color. A panel displaying all available colors appears. Select the required color. 9. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Note: It is possible to simplify the use of NIBP timing by selecting either Single T or Single Q. If this is selected only single timing mode will be available. This dialog panel enables you to set SpO2 default parameters. These parameters include setting alarm priority for SpO2, Pulse Rate and Fault alarms, activating or deactivating alarm latching, defining Alarm Detection delay and disabling SpO2 alarms during NIBP readings. Alarm priorities for SpO2, Pulse Rate, and Fault alarms Alarm priority affects the order of audible and visual alarm notification. Mennen Medical 3-33

72 Setting Up the System VitaLogik 4000/6000 Service Manual SpO2 and Pulse Rate alarm priority can be set to C-1, C-2, or C-3. The factory default for both parameters is C-1. Fault alarm priority can be set to T-1 or T-2. The factory default is T-1. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Response Time When an alarm is detected, a pre-defined response time allows the system to check in order to avoid false alarms. The delay can be set to 3, 6, or 12 seconds. The factory default is set to 6 seconds. SpO2 Default Color Defines the color in which SpO2 values appear on the main screen display. You can choose from 16 available colors. Figure 3-12: The SpO2 Setup Dialog Panel To set SpO2 default parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed Mennen Medical

73 VitaLogik 4000/6000 Service Manual 2. From the Vital Signs menu, select BP Setup. The BP Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the appropriate drop-down list, set the default priority levels for SpO2, Pulse Rate, and Fault alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. Select the Disable SpO2 Alarms During NIBP checkbox to stop detection of SpO2 alarms (if monitored) while taking an NIBP reading. Clear the checkbox to reactivate SpO2 alarms. 5. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 6. Select SpO2 Default Color. A panel displaying all available colors appears. Select the required color. 7. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Temperature Setup This dialog panel enables you to set Temperature default parameters. These parameters include setting Temperature alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default color for Temperature values. Temperature Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Temperature alarm priority can be set to C-1, C-2, C-3 or C-4. The factory default is C-4. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Mennen Medical 3-35

74 Setting Up the System VitaLogik 4000/6000 Service Manual Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 3, 5, or 10 seconds. The factory default is set to 5 seconds. Units Defines the default units used to measure and display temperature. The available options are Celsius or Fahrenheit. The factory default is Celsius. Temperature Default Color Defines the color in which Temperature values appear on the main screen display. You can choose from 16 available colors. Figure 3-13: The Temperature Dialog Panel To set default Temperature parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select Temperature Setup. The Temperature Setup dialog panel is displayed. 3. Under Extended Alarm Response: From the Temp drop-down list, set the default priority level for temperature alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. From the Units drop-down list, select the required default measurement unit (Celsius or Fahrenheit) Mennen Medical

75 VitaLogik 4000/6000 Service Manual CritiCool Setup 6. Select Temperature Default Color. A panel displaying all available colors appears. Select the required color. 7. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. This dialog panel enables you to set CritiCool Pro default parameters. These parameters include neonatal and adult default settings and temperature units. in Figure 3-14: show the CritiCool setup panel. EtCO2 Setup Figure 3-14: CritiCool Setup. This dialog panel enables you to set EtCO2 default parameters. These parameters include setting EtCO2 alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for EtCO2 values. EtCO2 Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Mennen Medical 3-37

76 Setting Up the System VitaLogik 4000/6000 Service Manual EtCO2 alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 0 seconds. Units Defines the default units used to measure and display EtCO2. The available options are % or mmhg. The factory default is %. EtCO2 Default Color Defines the color in which EtCO2 values appear on the main screen display. You can choose from 16 available colors Mennen Medical

77 VitaLogik 4000/6000 Service Manual Figure 3-15: The EtCO2 Dialog Panel To set default EtCO2 parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select EtCO2 Setup. The EtCO2 Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for EtCO2 alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. From the Units drop-down list, select the required default measurement unit (% or mmhg). 6. Select EtCO2 Temperature Default Color. A panel displaying all available colors appears. Select the required color. 7. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Mennen Medical 3-39

78 Setting Up the System VitaLogik 4000/6000 Service Manual EtCO2 Calibration Figure 3-16: EtCO2 Calibration panel For Microstream EtCO2 we have to perform span calibration with standard CO2 concentration after 1200 hours of first use and then once every year. To perform the calibration use gas mixture 5%, 21%D2m Bal. N2 (Scott Medical Products 6141 Easton Road Box 310, Plumsteadville, PA 18949, (215) Fax: (215) ). To perform the calibration 1. In System Setup, select EtCO2 Calibration (see in Figure 3-16:). 2. Set the calibration value to 5.0% 3. Open the gas flow and select "Start Calibration" 4. After several seconds the calibration will end 5. Stop Gas flow. Cardiac Output (CO) Setup CO Default Color This dialog panel enables you to set the CO default color. Defines the color in which CO values appear on the main screen display. You can choose from 16 available colors Mennen Medical

79 VitaLogik 4000/6000 Service Manual Figure 3-17: The CO Dialog Panel To set default CO parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select CO Setup. The CO Setup dialog panel is displayed. 3. Select CO Default Color. A panel displaying all available colors appears. Select the required color. 4. Select Discard Changes to close panel without applying any changes -or- Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Transcoutaneous Gas (TcGas) Setup This dialog panel enables you to set TcGas default parameters. These parameters include setting TcGas alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for TcGas values. TcGas Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Mennen Medical 3-41

80 Setting Up the System VitaLogik 4000/6000 Service Manual TcGas alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. TcGas technical fault priority can be set as T1 or T2. The factory default is T2. The Sensor Temperature alarm priority is set T2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 3, 5, or 10 seconds. The factory default is set to 5 seconds. Units Defines the default units used to measure and display temperature. The available options are mmhg or kpa for Tc PO2/ Tc Co2. The factory default is mmhg. The sensor temperature unit options are Celsius or Fahrenheit. The factory default is Celsius. TcGas Default Color Defines the color in which TcGas values appear on the main screen display. You can choose from 16 available colors. Figure 3-18: The TcGas Dialog Panel 3-42 Mennen Medical

81 VitaLogik 4000/6000 Service Manual To set TcGas parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select TcGas Setup. The TcGas Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for TcGas alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. From the TcGas drop-down list, select the required default measurement unit (mmhg or kpa). Select the required default measurement unit (Celsius or Fahrenheit) from the Sensor Temperature units drop down list 6. Select TcGas Default Color. A panel displaying all available colors appears. Select the required color. 7. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. CO2 (from Anesthetic Gas) Setup This dialog panel enables you to set CO2 default parameters. These parameters include setting CO2 alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for CO2 values. CO2 Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. CO2 alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Mennen Medical 3-43

82 Setting Up the System VitaLogik 4000/6000 Service Manual Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 0 seconds. Units Defines the default units used to measure and display CO2. The available options are % or mmhg. The factory default is mmhg. CO2 Default Color Defines the color in which CO2 values appear on the main screen display. You can choose from 16 available colors. Figure 3-19: The CO2 Dialog Panel To set default CO2 parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select CO2 Setup. The CO2 Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for Co2 alarms. From the Latching drop-down list, set default alarm latching to On or Off Mennen Medical

83 VitaLogik 4000/6000 Service Manual 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. From the Units drop-down list, select the required default measurement unit (% or mmhg). 6. Select CO2 Default Color. A panel displaying all available 7. colors appears. Select the required color. 8. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. O2 Setup This dialog panel enables you to set O2 default parameters. These parameters include setting O2 alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for O2 values. O2 Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. O2 alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 0 seconds. O2 Default Color Defines the color in which O2 values appear on the main screen display. You can choose from 16 available colors. Mennen Medical 3-45

84 Setting Up the System VitaLogik 4000/6000 Service Manual To set default O2 parameters: Figure 3-20: The O2 Dialog Panel 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select O2 Setup. The O2 Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for O2 alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds). 5. Select O2 Default Color. A panel displaying all available 6. colors appears. Select the required color Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu Mennen Medical

85 VitaLogik 4000/6000 Service Manual N2O Setup This dialog panel enables you to set N2O default parameters. These parameters include setting N2O alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for N2O values. N2O Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. N2O alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 0 seconds. N2O Default Color Defines the color in which N2O values appear on the main screen display. You can choose from 16 available colors. Figure 3-21: The N2O Dialog Panel Mennen Medical 3-47

86 Setting Up the System VitaLogik 4000/6000 Service Manual To set default N2O parameters: 1. 1.From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed From the Vital Signs menu, select N2O Setup. The N2O Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for N2O alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds) Select N2O Default Color. A panel displaying all available 6. colors appears. Select the required color Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Anesthetic Agent Setup This dialog panel enables you to set Anesthetic Agent default parameters. These parameters include setting Anesthetic Agent alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for Anesthetic Agent values. Anesthetic Agent Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Anesthetic Agent alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay 3-48 Mennen Medical

87 VitaLogik 4000/6000 Service Manual When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 10 seconds. Anesthetic Agent Default Color Defines the color in which Anesthetic Agent values appear on the main screen display. You can choose from 16 available colors. Figure 3-22: The Anesthetic Agent Dialog Panel To set default Anesthetic Agent parameters: 1. 1.From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed From the Vital Signs menu, select Anesthetic Agent Setup. The Anesthetic Agent Setup dialog panel is displayed. Under Extended Alarm Response: Set the default priority level for Anesthetic Agent alarms. From the Latching drop-down list, set default alarm latching to On or Off From the Anesthetic Agent drop down list, select either one of the 5 Anesthetic Agents or select Automatic for automatic selection of the Agent. 4. From the Alarm Detection Delay drop-down list, select the required delay time (in seconds) Select Agent Default Color. A panel displaying all available 6. colors appears. Select the required color. Mennen Medical 3-49

88 Setting Up the System VitaLogik 4000/6000 Service Manual 7. 7.Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Ventilator Setup This dialog panel enables you to set Ventilator default parameters. These parameters include setting Ventilator alarm priority activating or deactivating alarm latching, defining Alarm Detection delay and selecting the default units and color for Ventilator values. Ventilator Alarm Priority Alarm priority affects the order of audible and visual alarm notification. See User Guide for a full description of this feature. Ventilator alarm priority can be set to C-1, C-2, or C-3. The factory default is C-2. Alarm Latching For a complete description of Alarm Latching, refer to Chapter 7, Alarms, in the VitaLogik 4000/6000 Operation Manual. The available options are Off, Visual (latching visual alarms only), and On (both visual and audible alarms). The factory default is Off. Alarm Detection Delay When an alarm is detected, a pre-defined delay period allows the system to check in order to avoid false alarms. The delay can be set to 0, 3, 5, or 10 seconds. The factory default is set to 5 seconds. Ventilator Default Color Defines the color in which Anesthetic Agent values appear on the main screen display. You can choose from 16 available colors Mennen Medical

89 VitaLogik 4000/6000 Service Manual Figure 3-23: The Ventilator Dialog Panel To set default Ventilator parameters: 1. From the System Setup menu, select Vital Signs Setup. The Vital Signs menu is displayed. 2. From the Vital Signs menu, select Ventilator Setup. The Ventilator Setup dialog panel is displayed. 3. Under Extended Alarm Response: Set the default priority level for Ventilator alarms. From the Latching drop-down list, set default alarm latching to On or Off. 4. From the Alarm Detection Delay drop-down list, select the required delay 5. time (in seconds). 6. Select Ventilator Default Color. A panel displaying all available 7. colors appears. Select the required color. 8. Select Discard Changes to close the panel without applying any changes or Select Main Screen to apply the changes and exit the panel. Note: When you exit a vital signs setup panel, you return to the main screen. To continue System Setup, you must press the System Setup button again to access the System Setup menu. Mennen Medical 3-51

90 Setting Up the System VitaLogik 4000/6000 Service Manual Parameter Hierarchy Note:This dialog panel is active only when a patient is being monitored. The Parameter Hierarchy panel enables you to set the waveform and big number display on the monitor screen, from top for high hierarchy to bottom for lower hierarchy. Figure 3-24: The Parameter Hierarchy Setup Panel To set the Parameter Hierarchy panel 1. From the System Setup menu, select Parameter Hierarchy. The Parameters list menu is displayed. 2. From the Parameters list menu, select a parameter. 3. To promote a parameter, select Promote 4. To demote a parameter, select Demote 5. When the required hierarchy is obtained, select main screen The display will then be reorganized according to the new hierarchy Mennen Medical

91 VitaLogik 4000/6000 Service Manual Monitor Profiles Setup When admitting a patient, the admitting staff member must define which vital signs to monitor and how those parameters appear on the main screen display. In order to simplify this process, the monitor can be configured with pre-defined monitor profiles which include all definitions necessary for admitting various types of patients, such as adult, pediatric, geriatric, etc. Monitor profiles make it possible to prepare standard patient monitoring profiles allowing staff to connect a patient quickly without having to first configure the VitaLogik 4000/6000 (or at least to reduce configuration to a minimum). Up to 12 Monitor profiles can be configured for use in the VitaLogik 4000/6000. This includes 11 user-defined monitor profiles and one default Admit as previous key that saves the settings defined for the previous patient monitored by the unit. Only those Monitor profiles defined for use in the unit appear on the display screen. You can set up Monitor profiles while monitoring an actual patient, or you can start the VitaLogik 4000/6000 without actually connecting a patient, set up monitoring parameters and then save them as a Monitor profile. Figure 3-25: The Monitor Profiles Setup Dialog Panel Monitor profiles are defined in the Monitor profiles panel which can be accessed from the Setup menu. The keys are displayed on the panel in the order in which they appear on the VitaLogik 4000/6000 screen when admitting a patient. Any defined Monitor profiles are labeled with their defined name and monitoring parameters. To set up a new Monitor profile: 1. Set up the monitoring parameters you want according to the instructions in the Vital Signs Monitoring section. Mennen Medical 3-53

92 Setting Up the System VitaLogik 4000/6000 Service Manual 2. Press the Setup key on the front panel and select Monitor profiles from the Setup menu. The Monitor profile panel is displayed. 3. Press Select Key. 4. Select an available key. A frame appears around the key and the buttons along. Figure 3-26: The Monitor Profiles Selection Setup To set up a new Monitor profile: 1. Set up the monitoring parameters you want according to the instructions in the Vital Signs Monitoring section. 2. Press the Setup key on the front panel and select Monitor profiles from the Setup menu. The Monitor profile panel is displayed. 3. Press Select Key. 4. Select an available key. A frame appears around the key and the buttons along the bottom of the panel become available. 5. Click Save. To name a new Monitor profile: 1. Press Select Key in the Monitor profiles panel 2. Turn the QuicKnob and press an available key 3. Press Label. The alphanumeric keyboard panel appears. 4. Enter the text for the label and press Accept. Showing/Hiding the Monitor profile on the Main Screen Once you have defined a Monitor profile, you can also determine whether it is available for use when admitting a new patient. A green bullet appears next to those Monitor profiles that are set to appear on the main screen Mennen Medical

93 VitaLogik 4000/6000 Service Manual To show or hide the Monitor profile: 1. In the Monitor profiles panel, press Select Key. 2. Select the required Monitor profile. A frame appears around the key indicating that it is selected. If no bullet appears in the label, press Show/Hide. The green bullet appears and the Monitor profile will appear on the main screen the next time you want to admit a patient. If a green bullet appears in the label, press Show/Hide. The bullet disappears and the Monitor profile is removed from the main screen. Changing the Position of an Monitor Profile The Monitor profiles are arranged on the Monitor profile panel in the positions where they will appear on the main screen. You can move the keys to the most convenient position. To move a Monitor profile: 1. In the Monitor profiles panel, press Select Key. 2. Turn the QuicKnob and press the key you want. A frame appears around the key, indicating that it is selected. 3. Press Right to move the selected Monitor profile clockwise or press Left to move the key counterclockwise. Clearing a Monitor profile A previously defined Monitor profile may no longer suit your monitoring needs. In that case you can simply clear the Monitor profile from the VitaLogik 4000/6000. To clear a Monitor profile: 1. In the Monitor profiles panel, press Select Key. 2. Turn the QuicKnob until you reach the key you want to clear and press. A frame appears around the selected key. 3. Press Clear. The Monitor profile is removed from the Monitor profiles panel and will no longer appear on the main screen when admitting a patient. Mennen Medical 3-55

94 Setting Up the System VitaLogik 4000/6000 Service Manual Admit by Default This panel enables you to admit a patient by a factory set default monitor profile. You can only use this option after the patient has been discharged. Figure 3-27: The Admit by Default Setup Panel To admit by Default: 1. 1.Discharge the patient 2. 2.Select System Setup 3. 3.From the System Setup menu, select Admit by Default 3-56 Mennen Medical

95 VitaLogik 4000/6000 Service Manual Report Setup In order to review patient data, the VitaLogik 4000/6000 generates reports. You can name and define the format for up to 10 types of reports. Reports are configured in the Report Setup dialog panel. All available reports appear in the Reports List. You can configure and edit the format for each report, rename it, and clear the report from the Report List. All reports can be printed and placed in the patient s records. For example, if you want to set up a report for Hemodynamics, you might first name a report Hemodynamics and then, in Edit Report, set up the format for the Hemodynamics report. For each report, specify whether the report type is Charts or Trends. Naming a Report Figure 3-28: The Report Setup Dialog Panel By default, reports are named Report 1, Report 2, etc. You can change the name of any report according to what is accepted practice in your hospital or department. To rename a report: 1. In the System Setup menu, select Report Setup. 2. From the Reports List, select a report. 3. Select the Report Name textbox. The alphanumeric keyboard panel is displayed. Enter the new report name and close the panel. The new report name now appears on the Reports List. Mennen Medical 3-57

96 Setting Up the System VitaLogik 4000/6000 Service Manual Editing a Report When you edit a report, you define its type, duration and layout. You can select from one of two report types: Charts, or Trends. Charts A chart documents a patient s condition as observed at regular intervals in table format, much in the same way as a nurse records data in a bedside chart. The VitaLogik 4000/6000 supports the following set of clinical charts for the purpose of data acquisition, clinical calculations and review. Hemodynamics Respiration Oxygenation Renal Clearance User Defined Charts (with the name as given in Tabular Charts) The format of the User Defined chart can be edited. All other charts have a fixed format. Figure 3-29: Setting Up a Chart-Type Report To edit the format of a Vital Signs chart-type report: 1. From the System Setup menu, select Report Setup. 2. From the Report List, select a report. 3. Select Edit Report. The Edit Report dialog panel is displayed Mennen Medical

97 VitaLogik 4000/6000 Service Manual 4. In the Type drop-down list, select Chart. 5. From the Duration drop-down list, select the time period during which data is collected for the chart. 6. From the Paper Size drop-down list, select the size of the paper you want the report printed out on. Available options are; Letter, Legal, Executive, A4, and A3. 7. From the Paper Orientation drop-down list, select Portrait (vertical orientation) or Landscape (horizontal orientation). 8. From the Chart Type drop-down list, select Vital Signs. 9. Press Select All Filters for all available filters to be included in the chart or Press Unselect All Filters and select only those filters you want to include in the chart. The available filters are; Alarm, Manual, Cardiac Output, PWedge, Apnea Alarm, NIBP, and CVP. 10. Select Discard Changes to close the panel without applying any changes or select Main Screen to apply your changes and return to the blank screen. Figure 3-30: An Example of a Charts Report Showing All Filters Mennen Medical 3-59

98 Setting Up the System VitaLogik 4000/6000 Service Manual Trends Trends allows you to review data in a graphic representation rather than just a list of numbers. Trends represents changes in vital signs over time. Comparing the patterns of change gives an indication of the patient s progress. For each trend-type report, up to four axes can be presented. Each axis includes a left vital sign parameter and a right vital sign parameter for 8 parameters in total. In the Edit Report panel, you can see a preview of the report as you configure it. Each parameter appearing in the report can be configured to appear in a different color. The available colors are; red, black, green, and blue. Scaling can be configured as fixed or dynamic. You can also define that the data appear on a grid background to make reading the graph easier. Figure 3-31: The Report Setup Dialog Panel Displaying a Trend Report To edit the format of a trends-type report: 1. From the System Setup menu, select Report Setup. 2. From the Report List, select a report. 3. Select Edit Report. The Edit Report dialog panel is displayed. 4. In the Type drop-down list, select Trends. 5. From the Duration drop-down list, select the time period during which data is collected for the chart Mennen Medical

99 VitaLogik 4000/6000 Service Manual 6. From the Paper Size drop-down list, select the size of the paper you want the report printed out on. Available options are; Letter, Legal, Executive, A4, and A3. 7. From the Paper Orientation drop-down list, select Portrait (vertical orientation) or Landscape (horizontal orientation). 8. Select the number of axes to appear in the trend report. Up to eight axes can be presented. Each axis contains a left parameter and a right parameter. 9. Select an axis and define its left and right parameters: From the Vital Sign drop-down list, select the vital sign for the left or right parameter. Under Parameter Color, select the required color. From the Scaling drop-down list, select Fixed, or Dynamic. When Fixed Scaling is selected, turn the QuicKnob and select the min./ max scaling values, as required. Select or clear the Grid checkbox to activate or deactivate display of the trended data on a grid. 10. Select Discard Changes to close the panel without applying any changes or select Main Screen to apply your changes and return to the blank screen. Clearing a Report From the Reports List To clear a report from the Reports List: 1. From the System Setup menu, select Report Setup. 2. From the Report List, select the report you want to remove from the list. 3. Press Clear Report. The Report is removed from the Reports List and the undefined Report# appears in its place. Automatic Reports This function is only available if a printer was selected in Network Setup. This panel enables you to activate the automatic printout of reports at a preselected time daily. Mennen Medical 3-61

100 Setting Up the System VitaLogik 4000/6000 Service Manual Figure 3-32: The Automatic Reports Setup Panel To set Automatic Reports Setup Panel 1. Select System Setup 2. From the System Setup menu, select Automatic Reports 3. Select the report for automatic printing 4. Select the invoke time, either at the end of one of the three shifts, or at a user defined time. Note: Each selected report can be printed automatically up to three times every day Mennen Medical

101 VitaLogik 4000/6000 Service Manual Network Setup The Network Setup panel is used to name a device and to name and define which unit each device is associated with. You can associate a device with each of the units located on the network, or you can create a new unit and assign to it a device. The panel also displays the device ID number and IP address as defined during manufacturing. If a printer is located on the LAN, you can enable the printer for the device. If the monitor is connected to a Data Base (D.B.) the panel will show the Data Base Device ID and IP address. Monitors equipped with Wireless LAN board, can be activated from the Network Setup panel. "Wireless Network ID" and "WAP Pass Phase" can be downloaded from the panel to the wireless board. This provides a protected wireless network to Mennen Monitors. Figure 3-33: The Network Setup Dialog Panel To configure Network Setup: 1. In the System Setup menu, click Network Setup. 2. Click in the Device Name text box and select a device, or enter the name of the device with the alphanumeric keypad. 3. Click in the Device Unit/Group text box and select a unit from the list of units located on the network. To create a new unit, select the Manual option and using the alphanumeric keypad type the name of the new unit or group. Mennen Medical 3-63

102 Setting Up the System VitaLogik 4000/6000 Service Manual 4. Click the Printer checkbox if you wish to use the printer and select the printer to be used. 5. Select Discard Changes to close the panel without applying any changes, or Select Main Screen to apply the changes and close the panel. Recorder Setup If the VitaLogik 4000/6000 is connected to a network, it can be set up to send a recording to an in-built (optional) recorder or a defined list of recorders or printers located on the network. The VitaLogik 4000/6000 supports three recording types, each of which can be configured to send recordings to different recorders or printers, as is convenient for the user. These recording types are: Manual Recording A manual recording is initiated by the user at the bedside. The user can select a real-time or delayed recording and can select from the following recording durations: 20, 30, 45, and 60 sec. or Continuous. Alarm Recording A recording triggered by an alarm documenting the ECG waveforms, all VS values and the waveform of the clinical parameter which triggered the alarm (when not ECG). When using a dual channel recorder, the first channel is always ECG1. For the second channel, you can select ECG2 or any other parameter that produces a waveform. Selective Recording In a selective recording, the user defines which waveforms to record. The following recording durations can be selected: 20, 30, 45, or 60 sec. Each recording type has its own characteristics. If the VitaLogik 4000/6000 is on a network and a number of different recorders at different locations are available, you can assign specific recorders to specific recording types. For example, manual alarms can be sent to a local recorder located next to the VitaLogik 4000/6000, while alarm and all-waveform recordings can be sent to a Central Nurse Station Mennen Medical

103 VitaLogik 4000/6000 Service Manual Figure 3-34: The Recorder Setup Dialog Panel Trends Setup This dialog panel enables you to set the scales to be used in the graphic trend for each vital sign. Figure 3-35: The Trends Setup Dialog Panel Mennen Medical 3-65

104 Setting Up the System VitaLogik 4000/6000 Service Manual Two scaling options are available:fixed and Dynamic. If the Fixed option is selected, the scales of the Trend graph will be fixed and independent of the input data. In the Dynamic option, the scale of the trend graph will depend on the maximal and minimal parameter values from the patient admission. To modify Trends Setup: 1. Select Trends Setup from the System Setup menu - a large panel will open. 2. On the left of the panel open the "Vital Sign" window - a list of vital signs appears. 3. Select a parameter - for example RSP. 4. In the Scaling window select either "Fixed" or "Dynamic" 5. If you select Fixed, set the Min (minimum) and the Max (maximum) limits of the trend scale. 6. If you select "Dynamic" the Min and Max values will be dimmed and will show the maximal and minimal limits of the dynamic scaling. 7. Click on Accept to accept the selected settings. Quick Keys Setup Note:This dialog panel is active only when a patient is being monitored. The Quick Keys bar is a row of up to 12 user-defined function keys that appear on the bottom of the main screen display. This panel provides you with quick access to your most frequently used functions and menus. For example, if you want to access the Charts panel, instead of pressing the Data Review fixed key on the front panel and then selecting Charts from the Data Review menu, you can simply turn the QuicKnob until the Charts Quick Key is highlighted and press to display the Charts panel. Any function can be configured as a Quick Key. Quick Keys are configured by your hospital biomedical engineer or System Administrator in System Configuration. Note: Quick Keys disappear from the screen every two minutes. To redisplay the Quick Keys, simply turn the QuicKnob. Figure 3-36: The Quick Keys Bar Mennen Medical

105 VitaLogik 4000/6000 Service Manual Figure 3-37: The Quick Keys Setup Panel To assign a Quick Key 1. Select System Setup 2. From the System Setup menu, select Quick Keys 3. Use the Quicknob to select one of the Quick Keys this opens a menu with a green background 4. Press and rotate the Quicknob to open the menu item and open the sub menus 5. Select the function that you wish to assign press the Quicknob to finish the assignment. Tabular Charts A chart documents a patient s condition as observed at regular intervals in table format, much in the same way as a nurse records data in a bedside chart. The VitaLogik 4000/6000 supports a set of clinical charts for the purpose of data acquisition, clinical calculations and review (For more details, see Report Setup) The VitaLogik 4000/6000 also enables to create User Defined Charts (UDF), customized to specific needs of a unit or patient. This is done in the Tabular Charts panel. Up to 10 User Defined charts can be defined in System Setup. You can define the parameters that appear in the UDC, edit the format of a UDC, rename it, and clear a UDC from the User Defined Charts List. All UDCs can be printed and placed in the patient s records. Mennen Medical 3-67

106 Setting Up the System VitaLogik 4000/6000 Service Manual Naming a UDC Figure 3-38: The Tabular Charts Dialog Panel By default, User Defined Charts are named Tabular Chart 1, Tabular Chart 2, etc. You can change the name of any tabular chart according to what is accepted practice in your hospital or department. To rename a Tabular chart: 1. In the System Setup menu, select Tabular Charts. The Tabular Charts panel opens. 2. From the User Defined Charts List, select a tabular chart. 3. Select the UDC Name textbox. The alphanumeric keyboard panel is displayed. Enter the new UDC name and close the panel. The new chart name now appears in the User Defined Charts List. Editing a UDC Editing a User Defined Chart includes the following procedures: Defining the parameters to appear in the chart. The parameters can be selected from any of the available parameters associated with the VitaLogik 4000/6000 predefined charts. Defining the order in which the parameters will appear in the chart. Defining an alarm column for display of clinical alarms for the selected parameters Mennen Medical

107 VitaLogik 4000/6000 Service Manual Defining filters the conditions according to which entries are added to the chart. The available filters are: Manual - Chart entries based on a manually selected event. Alarm - Chart entries when an alarm is triggered. Cardiac Output - A chart entry each time a cardiac output reading occurs. PCWP - A chart entry each time a pulmonary wedge reading occurs. NIBP - A chart entry each time a NIBP reading occurs. CVP - A chart entry each time a VCP reading from the PAP channel occurs. Apnea - A chart entry each time an apnea event occurs. Time Interval - Chart entries based on a defined time interval. To edit a User Defined Chart: 1. From the System Setup menu, select Tabular Charts. The Tabular Charts panel opens. 2. From the User Defined Charts List, select a chart and click the Edit UDC button. An editing dialog panel appears, displaying all available charts supported by the VitaLogik 4000/6000, and the parameters associated with each. Figure 3-39: Editing a Tabular Chart 3. Click the arrow beneath a category (chart name), and from the list of parameters available for the category select the required parameters for the UDC. The selected parameters appear on the bar above. Mennen Medical 3-69

108 Setting Up the System VitaLogik 4000/6000 Service Manual 4. To define the order of parameters in the UDC, click on a selected parameter and then click the Move left or Move right buttons until the parameter reach the required location. If you want to remove any of the selected parameters, select the parameter and click Remove Parameter. 5. To display alarms, click the Insert alarm button. 6. Click Accept to apply your changes and return to the Tabular Charts panel. Note: If you want to return to the Tabular Charts panel without applying any changes, click Back to UDC List. If you want to move to the next or to the previous UDC in the list, click Next UDC or Previous UDC accordingly. 7. To set filters, in the Tabular Charts panel click the Define Filter button. A window opens displaying all available filters. Figure 3-40: Defining the Filters 8. Press Select All Filters for all available filters to be included in the chart or Press Unselect All Filters and select only those filters you want to include in the chart. 9. Click Accept to apply the changes and return to the Tabular Charts panel. The new UDC appears in the User Defined Charts list Mennen Medical

109 VitaLogik 4000/6000 Service Manual Clearing a UDC From the UDC List To clear a UDC from the UDC List: 1. From the System Setup menu, select Tabular Charts. 2. From the User Defined Charts List, select the UDC you want to remove from the list. 3. Press Clear UDC. The UDC is removed from the User Defined Charts List and the undefined Tabular Chart# appears in its place. Software Version The Software Version panel displays the software version of the bedside monitor, the module rack, and all modules located in the Module Rack. It also allows you to download the software for the modules. Figure 3-41: The Software Version Panel To download module software: 1.Click the Update BEP Software button. A window with a list box containing software versions appears. 2. Select the new version to update. Mennen Medical 3-71

110 Setting Up the System VitaLogik 4000/6000 Service Manual Figure 3-42: Selecting the Software Version 4. Click the Update CIPAM Software button. A window with a list box containing software versions appears. 5. Select the new version and update. 6. Press the QuicKnob. A message appears, requesting you to confirm the upgrade of the software. 7. Click Yes to download the software. 8. Select Main Screen to close the panel. Enmove Interface Setup To use the VitaLogik 4000/6000 with the Ensemble EN Central Station,the monitor has to receive a Network Number in the common network of Enmoves and VitaLogik 4000/6000s, To set the Network Number enter:system Setup >>Enmove Interface Setup and give the monitor a Network Number between 1 and 64. The Network Number is displayed in System Setup >>Network Setup under the heading:vitalogik 4000/6000 to MCS interface CAUTION! Each of the monitors VitaLogik 4000/6000 s and Enmove s must have a different Network Number. Note:All VitaLogik 4000/6000 monitor with software version 3.6 and up can be monitors on Ensemble EN Central Nurse Station 3-72 Mennen Medical

111 VitaLogik 4000/6000 Service Manual Change Password The Change Password option allows you to change the password that enables access to the System Setup menu. Access to the System Setup menu is limited to authorized personnel, such as a System Administrator or Hospital Biomedical Engineer. To change the password: 1. In the System Setup menu, click Change Password. The Change Password panel opens. Figure 3-43: Changing the Password 2. Enter the new password (3 digits) and click Accept. 3. A panel appears requesting you to type the password again. Type the password again and click Accept. Mennen Medical 3-73

112 Setting Up the System VitaLogik 4000/6000 Service Manual Check Disk When running Check Disk, all patient information on the disk is erased. Warning! Avoid using this option. To run Check Disk: In the System Setup menu, select Check Disk. Permission Editor In the Permission Editor panel you can define which VitaLogik 4000/6000 functions are available. By default, access to all functions is available. However, you can configure certain functions to be unavailable. For example, you can decide that at a certain station recording of vital signs data cannot be performed. The Permission Editor displays the entire VitaLogik 4000/6000 menu in tree format, and allows you to define, for each menu option, whether it is available or unavailable for the current VitaLogik 4000/6000. The following modes can be set for each menu option: On - select this checkbox to activate the menu option. The option appears in the appropriate panel and can be accessed. Off - select this checkbox to remove the menu option from the appropriate panel. The option cannot be accessed. Dimmed - select this checkbox when you want the menu option to appear dimmed on the panel. The option cannot be accessed Mennen Medical

113 VitaLogik 4000/6000 Service Manual Figure 3-44: The Permission Editor Panel To define permission for a menu option: 1. From the System Setup menu, select Permission Editor. The Permission Editor panel appears displaying a menu tree with all VitaLogik 4000/6000 menu options. 2. In the menu tree, click the plus (+) icon that appears to the left of a menu item to access the menu level below that listed on the screen. Any item from any level can be selected when assigning permission. 3. Select the required menu item, and then select the permission status for the item. 4. Repeat step 3 until defining permission for all the menu option that you want. 5. Click Main Screen to apply changes and close the panel. Remove Saved Patients Use this panel to view the list of patients with data saved in the VitaLogik 4000/ 6000 monitor. To delete a patient from the list 1. Select the patient row - this opens a list of the following options: Delete - to delete the patient from the list Sort the list, or Close Mennen Medical 3-75

114 Setting Up the System VitaLogik 4000/6000 Service Manual Note: The data of a patient that is deleted from the list is LOST and cannot be retrieved. Touch Screen Calibration Select the Touch Screen Calibration item. A blank screen will appear. Follow the instruction and touch the screen corners. At the end of calibration press Main Screen. Demo Activation The VitaLogik 4000/6000 monitor has a Demo Mode where signals stores in the system can be viewed and almost all features of the monitor can be demonstrated The following vital signs data is used in Demo Mode: 12 lead ECG, 80 Bpm, ST (-2.0) - (1.5), Filter = ST Respiration RA-LL 19 BPM, ART, PAP, waveforms and vital signs SpO2 96%, Pulse EtCO2, 0 to 5% (0 to 39 mmhg), Respiration 17 BPM Temp: Temp1-35 C, Temp2-31.2C, Temp3-36.8C, Temp C, Temp3- Temp4-0.2C Not Available: Cardiac Output, NIBP, Arrhythmia Patient Clinical Data (Waveforms, Vital Signs, Bit-to-Bit & QRS) source will be two minutes of recorded signals data played in a continuous loop. To start Demo Mode 1. Discharge the patient. 2. Enter System Setup (see To access the System Setup menu:) 3. Select Demo Mode from System Setup menu. 4. Enter password Admit a patient 6. Monitor goes into Demo Mode. To exit Demo Mode Discharge 3-76 Mennen Medical

115 VitaLogik 4000/6000 Service Manual Copy Configuration Utility The Copy Configuration utility (available from V5.2) enables transfering monitor configurations from onevitalogik 4000 or VitaLogik 6000 monitor to other VitaLogik 4000 or VitaLogik 6000 monitors on the same network. CAUTION! CAUTION! You can transfer configurations between monitors only when they are of the same software version. You can only transfer configurations between monitors which are the same model (VitaLogik 4000 or VitaLogik 6000 to VitaLogik 4000 or VitaLogik 6000, not from VitaLogik 4000 or VitaLogik 6000 to Envoy or from Envoy to VitaLogik 4000 or VitaLogik 6000). Copying a configuration includes the following steps: 1. Building a reference monitor configuration 2. Storing the configuration on the reference monitor 3. Copying the configuration to each monitor you want to copy the configuration to. To build the reference monitor configuration: 1. Admit a patient without inserting a name or any demographic information. 2. Set the monitoring profiles, the alarm limits and display options. 3. Discharge the patient. To store the configuration on the reference monitor: 1. Enter System Setup. 1. Select Build Master Configuration from System Setup menu. The monitor will restart and store the configuration files To copy the configuration to a monitor: 1. On Main menu select > Setup ' A - D - T ' > Discharge Patient 2. On Main menu select > Setup > System Setup > Copy Configuration from. 3. Choose the source (reference) monitor's unit. 4. Choose the source (reference) monitor from the list. Are you Sure" panel will appear. 5. Select "Copy" The monitor will restart and will come up with the new configuration. Mennen Medical 3-77

116 Setting Up the System VitaLogik 4000/6000 Service Manual 6. Verify that all the changes including the new profiles, permissions, quick keys are displayed. CAUTION! If Alarm Watch was part of the configuration, be sure to remove the target monitor watching itself and if required to add the reference monitor to the watched monitors list. Example : The reference monitor was BED 1, and it was watching BED 2 Upon transfer BED 2 is part of the list of beds watched by BED 2 which means that BED 2 watches itself, while BED 1 is not watched by BED 2. You need to remove BED 2 from the list of beds watched by BED 2, and to add BED 1 to the watched bed list. Note: If Enmove interface is used, enter "Enmove Interface Setup" panel of each of the target monitors and set it to the proper Enmove network number Mennen Medical

117 CHAPTER 4: SYSTEM INSTALLATION This chapter includes the following sections: Overview System Acceptance Site Survey Unpacking and Inspection Setting Up the Basic System Mounting Options Overview The basic VitaLogik 4000/6000 system usually consists of more than one VitaLogik bedside monitor, recorder, laser printer and local area network (LAN). The installation includes the following stages: Site survey Unpacking and inspection Setting up the basic system Mennen Medical 4-1

118 System Installation VitaLogik 4000/6000 Service Manual 4 Pair Cable Interface Switch Lan Cable Enguard Enesemble Enscirbe Laser Printer Figure 4-1: VitaLogik System Basic Setup In addition, various options exist for mounting the VitaLogik bedside monitor see Mounting of Monitors. System Acceptance Fill the Device Acceptance Check list and send it to Mennen Medical Service department for future reference. The form can be found in - Appendix G Safety Precautions Before installing the system, you must consider the following safety precautions: System grounding Equipotential grounding Electrosurgery, diathermy and defibrillation protection Monitor Mounting Accidental wetting of the equipment 4-2 Mennen Medical

119 VitaLogik 6000 Service Manual Patient connections General hazards Grounding the System To protect patients and hospital personnel, the VitaLogik 4000/6000 system must be grounded. Accordingly, the system is equipped with a detachable 3 wire cable which grounds the instrument to the power line ground (protective earth) when plugged into an appropriate 3 wire receptacle. If a 3 wire receptacle is not available, consult the hospital electrician. WARNING!!! WARNING!!! Do not use a 3 wire to 2 wire adapter with this instrument. If it is not evident from the instrument specifications whether a particular instrument combination is hazardous or not. For example, due to summation of leakage currents, the user should consult the manufacturers concerned or an expert in the field to ensure that the necessary safety of all instruments concerned will not be impaired by the combination. Equipotential Grounding Protection class 1 instruments are already included in the protective grounding (protective earth) system of the room by way of grounding contacts in the power plug. For internal examination on the heart or the brain, the VitaLogik must have a separate connection to the equipotential grounding system. One end of the equipotential grounding cable (potential equalization conductor) is connected to the equipotential grounding terminal on the instrument s rear panel and the other end to one point of the equipotential grounding system. The equipotential grounding system assumes the safety function of the protective grounding conductor if ever there is a break in the protective grounding system. Examinations in or on the heart (or brain) should only be carried out in medically used rooms incorporating an equipotential grounding system. Check each time before use that the instrument is in perfect working order. The cable connecting the patient to the instrument must be dry and clean. Note: Only VitaLogik 4000/4500 has an Equipotential grounding terminal. WARNING!!! When performing examinations in or on the heart (or brain), the conductive parts of electrodes and associated connectors, including the neutral electrode, should not contact other conductive parts including earth. Mennen Medical 4-3

120 System Installation VitaLogik 4000/6000 Service Manual Electrosurgery, Diathermy and Defibrillation Protection Neither electrosurgery and diathermy instruments nor defibrillation will damage the monitor. If the correct electrodes are used and applied according to the manufacturer s instructions, the screen display will recover within 10 seconds of defibrillation. The inputs have floating inputs are defibrillator-proof. WARNING!!! Do not touch the patient, bed or instrument during defibrillation. Mounting of Monitors Mount the monitors on Mennen Medical Ltd. wall mounts. CAUTION! Mennen Medical Ltd. is responsible for monitor's wall mount and shelves if and only if they were installed by Mennen Medical Ltd. WARNING!!! Mennen Medical Ltd. is not responsible for the stability of the wall mounts and shelves unless checked by Mennen Medical Ltd. at least once a year. Note: It is the responsibility of the end user (hospital) to ensure that the stability of the wall mounts or shelves is checked at least once a year. Accidental Wetting of the Equipment General Hazards If, for whatever reason, the equipment accidentally gets wet, remove all patient connections and turn the unit off. Thoroughly wipe all wet areas and then allow sufficient time to ensure that the equipment is completely dry before proceeding. It is the responsibility of the hospital, its consultants and/or contractors to ensure that the wall is adequately safe for mounting the bedside monitor. This includes selecting the appropriate fasteners and installing them properly. In constructions and remodeling work where the wall covering may be partially or entirely removed, a 16 gauge steel stud must be used for attaching the wall channel. The following mounting instructions are provided as guideline information only. It is recommended that the hospital establish a policy for periodical check of the wall mounting structure for proper installation. WARNING!!! Do not place any portion of the monitor directly over a patient s bed. To ensure easy reach and handling of the bedside monitor, all controls should not be higher than 182 cm (72 ) above the floor. 4-4 Mennen Medical

121 VitaLogik 6000 Service Manual Site Survey Performing a site survey ensures that all environmental, electrical, and cabling requirements are taken into consideration. This ensures that the actual installation proceeds as smoothly as possible and that the installed equipment will run reliably. Perform a site survey every time an VitaLogik bedside monitor is installed. The hospital Biomedical Engineer and Facilities Engineer should participate in any discussion, which must include: Reviewing details and drawings of hospital layout and architecture. Planning of all cable ducts, raceways, runs, wall plates, pull boxes and other interconnection details. Establishing length of cable runs. Reviewing AC power requirements, locations and types of outlets. Discussing the type of emergency (backup) AC power supply to be used, time of transfer (periodic test), degree of transient introduced into the power line (how long is power interrupted?), frequency and voltage of emergency power. Reviewing environmental considerations, such as ambient temperature, humidity control, potential static charge, and expected tolerances in the area planned for the VitaLogik. Measure AC power disturbances over a period of at least 24 hours with a power line analyzer. It is important to perform this test over the period of emergency power transfer to measure the severity of transients during the transfer. Environmental Requirements Locate the VitaLogik out of direct sunlight and in a clean air atmosphere. Avoid dusty areas or areas where there are airborne dirt particles. Avoid static build-up and discharge, which may result in disruptions of operation and loss of data. Reasonable environmental parameters are: Temperature 5 to 40 C Relative Humidity 10% to 93% (non-condensing) The following table describes the equipment physical dimensions and power consumption. To complete the site survey please fill the following form and send it with the purchase order. Mennen Medical 4-5

122 System Installation VitaLogik 4000/6000 Service Manual Unpacking and Inspection Before unpacking the VitaLogik 4000/6000 bedside monitor, inspect visually the packaging for any signs of damage or abuse, which may have resulted during shipment. If any signs of external damage are visible, you have to contact your nearest Mennen Medical Ltd. agent before proceeding to unpack the unit. Failure to do so could invalidate the warranty. Unpack carefully the package and check all components against the packing list to confirm their availability. If any damage appears or any part is missing, contact your Mennen Medical Ltd. agent immediately. Do not attempt to set up the system if any component is damaged. IMPORTANT! Transfer the packaging materials to recycling as per the regulation of the import country. Setting Up the Basic System The VitaLogik basic system consists of the following parts: the LAN switch and patch panel, and the nurse station. These should be set up according to system description. Take care when handling the components. Proper installation of the VitaLogik 4000/6000 is divided into the following steps: Connecting and wiring the LAN cables Mounting the LAN outlets Mounting and connecting the LAN hub and the patch panel Mounting the VitaLogik monitor on the wall Connecting the interconnection cables for each device Mounting the nurse station devices Note: If a hospital has special installation requirements, these may call for special planning and materials that must be furnished by that hospital and are not covered by this manual. Connecting and Wiring the LAN Cables To connect and wire the LAN cables: 1. After determining the location of the nurse station and the VitaLogik bedside monitors, for each VitaLogik bedside monitor, place a standard shielded twisted pair LAN cable from the nurse station to the bedside monitor. Leave 20cm (7.8 ) spare cable on the bedside monitor side, and 120cm (47.2 ) spare cable on the nurse station side. 4-6 Mennen Medical

123 VitaLogik 6000 Service Manual 2. For each VitaLogik bedside monitor, connect the four wires (two Rx and two Tx wires, according to the connections drawing) to the LAN outlet. Figure 4-2: LAN Cables Wiring Diagram Mennen Medical 4-7

124 System Installation VitaLogik 4000/6000 Service Manual Network Connections Network type 100-Base-TX Ethernet network Infrastructure installation Figure 4-3: System Overview The distance between switch to devices will not exceed 100m. The distance between connected switches (not in stack) will not exceed 100m. Cables, patch panel and wall plates will be with accordance to Category 5E standard (TIA/EIA 568A). Note: This will cover us up to bandwidth of 100mb/sec. Maximum bandwidth needed to NGMS network can be calculated roughly as follows: 0.5mb/sec * (# of active devices connected). This is rarely achieved and the actual bandwidth is half of this. For design purposes, the maximum rate must be considered. As long as switches used for the active equipment, the total bandwidth calculation is not important. Patch panels and switches will be located in a standard 19 communication racks. 4-8 Mennen Medical

125 VitaLogik 6000 Service Manual Cable specifications 100BaseTX is based on the American National Standards Institutes (ANSI) Twisted Pair-Physical Medium Dependent (TP-PMD) specification. The ANSI TP- PMD supports UTP and STP cabling. 100BaseTX uses the 100BaseX signaling scheme over 2-pair Category 5 UTP or STP. The following are Mennen Medical Ltd. s network cable specification: Characteristics Cable 100BaseTX Category 5E Type 1 and 2 STP Number of pairs 4 pairs * Connector Maximum segment length Maximum network diameter ISO 8877 (RJ-45) connector 100 meters 200 meters * Only 2 pairs are in use (in Figure 4-4:) Figure 4-4: Network Connections To connect network devices such as desktops, printers etc., use a straight-through cables. To connect a remote switch to another switch, use a roll-over cable. You can identify a roll-over cable by comparing the two modular ends of the cable. Holding the cables side-by-side, with the tab at the back, the wire connected to the pin on the outside of the left plug should be the same color as the wire connected to the pin on the outside of the right plug. (See Figure) If your cable came from Mennen Medical Ltd., pin 1 will be white on one connector, and pin 8 will be white on the other (a roll-over cable reverses pins 1 and 8, 2 and 7, 3 and 6, and 4 and 5). Mennen Medical 4-9

126 System Installation VitaLogik 4000/6000 Service Manual Communication Rack Arrangement Figure 4-5: Communication Rack Arrangment All communication racks will be 19 standard. Each patch panel must include at least 16 ports. To avoid spaghetti syndrome in communication racks, make sure that the jumper cables (connects patch panel ports to switches) will be no more than 50cm long. In large installations, to avoid using jumper cables longer than 50cm, a careful thought must be taken regarding the arrangement of the patch panels and switches inside the communication rack. In most installations you will find a block of patch panels and underneath a block of switches. This kind of arrangement forces you to use jumper cables of different lengths. In that case arrangement of several blocks is more logical (see example) 4-10 Mennen Medical

127 VitaLogik 6000 Service Manual Active equipment installation # Conditions Solution Quantities 1 The number of devices connected to each department/site is up to 24 # of departments is one or more There is no interaction between departments/site Use switches. (# of switches) = (# of departments) 2 The number of devices connected is more than 24 All of the devices are within the distance boundaries There isn t a clear division of departments / sites 3 The number of devices connected in each department is up to 24 All of the devices are within the distance boundaries and located in a radius of less than 100m from the location of the switches. There is a clear division of departments. 4 The number of devices connected in each department or site is up to 24 or more It is impossible to locate all/part of departmental switches/stack in on place There is a clear division of departments or sites. # of remote departments / sites is up to 3. 5 Same as #4 except that number of departments/sites is more than 3. Use a stack of switches. In the stack, the switches are connected between them with a very fast bus (2.4Gbit/sec). In order to achieve this, every switch must be equipped with a cascade card (item # 2). Also needed is a special cable to connect the switches (item # 3). Use a stack of switches. In the stack, the switches are connected between them with a very fast bus (2.4Gbit/sec). In order to achieve this, every switch must be equipped with a cascade card (item # 2). Also needed is a special cable to connect the switches (item # 3). In that case, connect each department to a different switch. Connect each department to a different switch. If possible, cascade departmental switches. Connect remote switches with Ethernet cables (normal, not cross). Make sure that the distance won t exceed 100m. the number of devices that can be connected is equal to: 24 * (# of switches) (# of remote switches x 2). If it isn t possible to connect all remote switches to the same switch, they can be cascaded with Ethernet cables. Same as #4 except that all departmental switches/stacks will be connected to a center switch. (# of switches and cards) = INT((# of devices) / 24) + 1 (# of cascade cables) = (# of switches) 1 (# of switches and cards) = (# of departments) (# of cascade cables) = (# of switches) 1 Use formulas from case # 2 or 3 as needed. Same as #4. Add 1 to the total number of switches Mennen Medical 4-11

128 System Installation VitaLogik 4000/6000 Service Manual Equipment used ITEM Manufacturer Model P/N 1 Nortel Networks T switch AL2012B14 2 Nortel Networks Cascade card for BayStack 4xx AL Nortel Networks Cascade cable for BayStack 4xx (1m) AL Nortel Networks MDA card with 2 fiber ports AL Mounting Options Mounting the Switch and Patch Panel After placing the LAN cables, mount the LAN outlets on its place according to the hospital requirements. The following three types of LAN outlets are available: Wall outlet mounted on the wall surface. Wall outlet mounted behind the wall surface. Wall outlet mounted together with other outlets on an aluminum plate. Mounting and Connecting the LAN Hub and the Patch Panel To mount the LAN hub and the patch: 1. Mount the LAN hub and the patch panel inside the enclosure and place them under the nurse station desk (or at any other convenient place). Secure them if necessary. 2. At the nurse station side, connect each LAN cable to the patch panel according to the wiring installation diagram (in Figure 4-2:). 3. For each VitaLogik bedside monitor, connect a bridge cable from the patch panel to the LAN hub. Mounting the VitaLogik Monitor on the Wall To properly mount the VitaLogik bedside monitor, refer to the wall mounting instructions supplied with your wall mounting kit. It is the responsibility of the hospital, its consultants, and/or contractors to determine that the wall is adequate to safely mount instrumentation. This includes the selection of appropriate fasteners and the proper installation Mennen Medical

129 VitaLogik 6000 Service Manual Wall Mount for VL4X cm Item Number part number Description QTY WALL CHANNEL 19' ARM ADAPTER PLATE for Envoy/VL5XXX NUT-HEX 1/4-20 THD SCREW M5X20 PH,PHILIPS SS DIN 7985A A2 4 Figure 4-6: VitaLogik Mounting Diagram Figure 4-7: VitaLogik 4000/6000 Mounting Views Mennen Medical 4-13

130 System Installation VitaLogik 4000/6000 Service Manual 4-14 Mennen Medical

131 CHAPTER 5: UPGRADING SOFTWARE This chapter describes the technique employed to support VitaLogik software upgrades. The proposed method attempts to resolve the various possible installation scenarios, as experienced by personnel servicing the VitaLogik 4000/6000 with the various tools. To upgrade VitaLogik 4000/6000 you need to upgrade using Microsoft Windows utilities. In addition the following is also described: Upgrading Main software of VitaLogik 4000/6000. Upgrading the inputs and/or Front panel Note: You should check the system after upgrading! Note: Send to Mennen Medical Ltd. Service Department, the p/n and the serial numbers of any device that you upgraded. Upgrading Using Microsoft Windows Utilities General The VitaLogik 4000/6000 software consists of: QNX operating system VitaLogik Application software The following instructions cover all the procedures needed to achieve a successful upgrade of one VitaLogik 4000/6000. Mennen Medical 5-1

132 Upgrading Software VitaLogik 4000/6000 Service Manual Required Hardware Personal computer Modem Network Card with RJ45 connector Network Patch Cable for a connection to the switch or Network Cross Connect cable for a direct connection to a VitaLogik 4000/6000. Required Software Windows 95 or 98 Internet browser software (AOL, Internet Explorer, etc. WS_FTP95 or compatible program Winzip VitaLogik 4000/6000 Software Installation Process This section describes the technique employed to support VitaLogik 4000/6000 software upgrades. The proposed method attempts to resolve the various possible installation scenarios, as experienced by personnel servicing the VitaLogik with the various tools. There are two main procedures for upgrading: replacing hard disk or Compact Flash Upgrade by computer. WARNING!!! Only a person authorized by Mennen Medical Ltd. for performing such a procedure should perform replacing Hard Disk / Compact Flash or Upgrading by software installation. It is recommended to connect the patient to another monitor until the upgrading procedure will complete. You should test the system after upgrading! Record the serial number and part number and report to Mennen Medical Ltd. Service department 5-2 Mennen Medical

133 VitaLogik 4000/6000 Service Manual Upgrading VitaLogik 4000/6000 WARNING!!! Before upgrading VitaLogik 4000/6000, make sure there is no patient connected to the VitaLogik 4000/6000. To change the IP Address of your laptop 1. Connect your computer to the VitaLogik 4000/6000 network. 2. Select the Network Neighborhood icon and press the right button of your mouse 3. Select Properties 4. In the Configuration area, select the TCP/IP that is bound to your network card. 5. Click on Properties 6. Select IP Address tab 7. Change the following properties: IP Address = (250 or number higher than 180) Subnet Mask = Click on OK button twice to exit and save the new settings. 9. To take effect, the PC should be rebooted. Copying the Tar file from the Laptop to VitaLogik 4000/6000 To copy the Tar file from a Laptop to VitaLogik 4000/ On your PC : Using WS_FTP95 program, create a new session called Envoy Update. 2. In the Host Name field type the IP address of the Envoy / VitaLogik you wish to connect to. 3. Use root as User ID and go4it as Password. Check the Save Password box to save the setting for the next update. Mennen Medical 5-3

134 Upgrading Software VitaLogik 4000/6000 Service Manual 4. In the Remote System field, select the Envoy.tar file (old) under /UPDATE directory and click on the Delete button to delete it. 5. Select the tar file (Local System field) and transfer it into the Envoy / VitaLogik (Remote System field) into /UPDATE directory). When the transfer is completed, click on the Rename button and rename the file you have just transferred from Envoy.tar.xxx to Envoy.tar Upgrading VitaLogik 4000/6000 Software The following steps use the new version of the tar file to create a new Envoy / VitaLogik 4000/6000 software version. IMPORTANT! The following procedure is the upgrade procedure for software versions lower than 5.4a to 5.4a and from 6.0 to software version above 6.0. Software upgrade from versions lower than 6.0 to 6.0 and higher, requires replacing the flash memory. To upgrade VitaLogik 4000/6000 Software: 1. Discharge the monitor, admit, and wait for 10 sec and discharge the monitor again. 2. On your PC Open MS-DOS prompt 3. Type in the following commands: telnet <VitaLogik IP address > (for example: telnet ) login as root 5-4 Mennen Medical

135 VitaLogik 4000/6000 Service Manual Run the command # uflags sc A window, requesting confirmation for the software update, will pop-up. Press Start. The VitaLogik 4000/6000 reboots and the new software is installed. 4. Wait till the VitaLogik 4000/6000 system builds the new version and scan checks the disk automatically ( This may take up to 5 minutes) At the end of this procedure you will get the normal VitaLogik 4000/6000 monitor screen 5. At the end of the VitaLogik 4000/6000 update, in the discharge mode, select Setup, System Setup and open Software Version panel. Update the last BEP and CIPAM software version by pressing one by one on Update BEP Software and on Update CIPAM Software buttons. Note: At the end of the software update, check the VitaLogik 4000/6000 functionality and confirm that all the parameters work properly. Notes: 1. If the VitaLogik 4000/6000 is not in discharge mode, the confirmation window will be displayed once the patient is discharged. 2. If the button (start) is not pressed within two minutes, the update will start automatically. Table 5-1: Software Versions Tables Mennen P/N Name Hardware version Current software Update To Comments Recommended VitaLogik 4X / 020/110/120/ 210/220 VitaLogik Any Any V5.4A If Ensemble is connected to V.L. you should upgrade the ensemble to the same version VitaLogik Cipam Any V2.5r VitaLogik BEP Any V2.03 Transport SPB -- V1.3 VitaLogik 6X00 Mennen Medical 5-5

136 Upgrading Software VitaLogik 4000/6000 Service Manual Table 5-1: Software Versions Tables Mennen P/N Name Hardware version Current software Update To Comments Recommended / 020/110/120/ 210/220 VitaLogik Any Any V5.4A If Ensemble is connected to V.L. you should upgrade the ensemble to the same version VitaLogik Cipam Any V2.5r VitaLogik BEP Any V2.03 Transport SPB -- V6.5 VitaLogik 4X00 Ensemble / 180/280 Ensemble ALL Any V5.4A Upgrading central force you to upgrade the monitors that are connected to. Enguard / 190/290 Enguard ALL Any V5.4A Upgrading central force you to upgrade the monitors that are connected to. Enscribe Enscribe ALL Any V2.17 Laser Printer: HP4000N, HP4050N, HP4100N,HP2420N, HP3005N, IBM4100N, HP P2055DN 5-6 Mennen Medical

137 CHAPTER 6: MAINTENANCE AND CLEANING Periodic Testing and Preventive Maintenance Once every year perform a full monitor test, according to the Test procedure (see Appendix F Test Procedures ). For VitaLogik 4000/6000 with the EtCO2 option you also need to calibrate EtCO2 Every 3 years: Replace the CPU ventilator Replace the main battery Every 4 years: Replace the backup battery on the computer board Replace the Quicknob (optical switch) Every 5 years : Replace the main ventilator Note: Annual tests and preventive maintenance must be performed by Mennen Medical authorized service person Cleaning the VitaLogik 4000/6000 Bedside Monitor The VitaLogik 4000/6000 bedside monitor can be cleaned with most common hospital cleaning solutions and detergents. Be careful not to use caustic detergents or ammonia or acetone-based cleaning solutions. These may damage the bedside monitor. To clean the VitaLogik 4000/6000 bedside monitor and accessories: 1. Switch the monitor off and disconnect from the power supply before cleaning. Mennen Medical 6-1

138 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual 2. Dust the monitor regularly. Clean with a lint-free cloth or sponge dampened with cleaning solution. Abrasive scouring powders and pads should be avoided to prevent damage to the monitor. 3. Clean the display screen with non-abrasive glass cleaners, such as Ajax or Windex. Use a lint-free cloth. A paper towel may damage the screen. Before cleaning a Touch Screen disable the Touch functionality. To disable Touch enter System Setup (password protected) and press the "Touch Screen" key. At the end of the cleaning re-activate the Touch functionality. 4. Clean patient cables and all exposed surfaces with mild soap solution only. Using alcohol or any type of concentrated cleaning solution may impair patient cable flexibility. 5. Remove any adhesive used to attach the cable to the patient. To remove adhesive residue, use a plaster remover solution made up of one-third alcohol, two-thirds water, or use a commercial tape remover such as Scholl Double Seal tape remover. Do not use strong solvents (acetone, straight alcohol, ammonia, etc.). These will damage the cable. 6. Do not autoclave patient leads, transducer components or other sensors. Autoclaving permanently damages these instruments. If you need to sterilize accessories, we recommend EtCO2 gas sterilization. Replacing the VitaLogik 4000/6000 Battery To replace the battery: 1. Turn the Monitor to view the lower cover and turn the latches to "0" position. 6-2 Mennen Medical

139 VitaLogik 4000/6000 Service Manual 2. Remove the lower cover. 3. Turn the battery cover latch at 90º 4. Remove the battery cover 5. Disconnect the battery connector and remove the battery. 6. Replace the battery and connect the battery connector. 7. Close the battery cover by turning the latch 90º and pushing it. 8. Close the lower plastic cover by turning the latch 90º and pushing it. 9. Replace the battery and connect the battery connector. 10. Close the battery cover by turning the latch 90º and pushing it. 11. Close the lower plastic cover by turning the latch to 1 and pushing it. Mennen Medical 6-3

140 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual Calibration and Preventive Maintenance Mennen Medical Ltd. Ltd. recommends that the VitaLogik 4000/6000 be checked at least once a year by qualified Mennen Medical Ltd. personnel for proper calibration of the unit. The exterior of the unit should be periodically checked for any signs of damage or abuse. Units or parts which show any signs of damage should be immediately referred to a qualified technician. Before commencing monitoring on a patient: check for any mechanical damage. check all external leads and accessories. check all functions of the instrument needed to monitor the patient and ensure that the instrument is in good working order. Do not use the VitaLogik 4000/6000 for any monitoring procedure on a patient if you identify features which demonstrate impaired functioning on the instrument. Contact the hospital biomedical engineer, or Mennen Medical Ltd. service engineer. Full performance checks, including safety checks, must be performed by qualified service personnel at least once a year and after every repair. The synchronization of the monitor and defibrillator should be checked as often as required by the Hospital Procedures Policy, and must be tested at least every three months by qualified service personnel. The grounding and equipotential resistances of the monitor and power cables must be tested at least every 3 months. All checks which require the instrument to be opened must be made by qualified service personnel. Safety and maintenance checks can also be made by Mennen Medical Ltd. personnel. Your local Mennen Medical Ltd. agent will be glad to give you information about service contracts. 6-4 Mennen Medical

141 VitaLogik 4000/6000 Service Manual Error Messages and Troubleshooting Error Messages If you have any problems, or the setup procedure does not function as described, check the appropriate table, as follows: In normal monitor use most of the specific faults are associated by error message that can guide you in locating and correcting the problemthe list below shows all the error messages that the monitor will present. Table 6-1: Error Messages Vital Sign Alarm Message Cause - All Vital Signs (parameter) "Cable Out Cable out, Faulty cable, Lead V1 Fault" when more Faulty internal connector CIPAM board (parameter) "Artifact (parameter) "Fault ECG "RL Fault RL or more than one lead fault "Lead xx Fault "All Leads Fault Lead xx disconnected Fault on more than "Pacer not sensed" "Pacer non function" "Pacer non capture" BP "Zero Failure "Transducer Fault Unstable transducer/transducer failure NIBP Read Failed (3) Zero pressure detected Read Failed (9) Read Failed (11) Read Failed (12) Read Failed (13) Read Failed (14) Read Failed (15) Hardware failure Pump failure (adult cuff used in neonatal mode) Diastolic BP too low to measure Motion artifact Inflation not sufficiently above systolic point Abnormal oscillometric waveform detected (may be due to motion) Mennen Medical 6-5

142 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual Table 6-1: Error Messages Vital Sign Alarm Message Cause - Read Failed (16) Read Failed (17) Read Failed (18) Read Failed (19) Read Failed (20) Read Failed (21) Arrhythmic pulse detected Measurement time-out (may be due to motion) Pulse rate too fast (may be due to motion) Cuff over-pressure sensed Loose cuff Neonatal cuff used in adult mode RESP Sp02 "RSPECG Coincidence "SpO2 Relocate Probe "SpO2 Wrong Probe "SpO2 Probe Disconnected "SpO2 Low Signal "SpO2 Low Light "SpO2 C-Lock mode EtCO2 Microstream Tube Disconnected Check that the sampling tube is connected Calibration Alert Replace tube Purge Auto Zero Calibrate the EtCO2 Replace the tube Wait for the end of purge Wait for the end of Auto zero 6-6 Mennen Medical

143 VitaLogik 4000/6000 Service Manual Table 6-1: Error Messages Vital Sign Alarm Message Cause - Anesthesia Gas Analyzer Ventilator Cable Out No Comm Device Error Error S Error F Error PS CO2 Check Sampling Line Calibration Zero Cal. In progress Span Cal. In progress Zero Calibration Failure Initialize Check Sampling line Tube Disconnected No Water-trap found Safety Valve Open Disconnect O2 Inlet Air Inlet Low Battery Valve Leak Inoperable Gas Supply Error Communication Error Cable between Andros and monitor not connected Communication Loss: cable disconnected or device does not transmit System Error Fan Error Power Supply Error Sampling delivery system error System is performing auto calibration Wait for end of calibration Zero Calibration Failure Making communication with device Replace Water trap Mennen Medical 6-7

144 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual Table 6-1: Error Messages Vital Sign Alarm Message Cause - BIS Poor SQI Poor Low SQI No BISx Sensor not connected Check sensor Sensor fault Sensor expired No more uses (for SRS) Sensor Invalid Low Signal quality BISx not connected No sensor Check Electrodes Replace sensor Replace sensor Replace SRS sensor Replace with avalid sensor 6-8 Mennen Medical

145 VitaLogik 4000/6000 Service Manual Trouble Shooting VitaLogik 4000/6000 Bedside Monitor If you have any problems, or the VitaLogik 4000/6000 does not function as described, check the table to identify the problem and possible corrective actions: Table 6-2: VitaLogik 4000/6000 Bedside Monitor faults Problem Problem Cause Corrective Action Monitor is not turning ON: 1.No leds 2.No Display 3.No Fan 1.Burned Fuses 2.Battery Failure 3.Power Splitter Failure 1.Check fuses on AC Plug 2.Replace Battery 3.Replace Power Splitter board Monitor is not running: 1.Front leds are working 2.Display is ON without any display 1.Power Splitter Failure 2.SBC Failure 1.Replace Power Splitter board 2.Replace SBC Monitor is running without front leds indications 1.Power Splitter Failure 1.Disconnect AC line and battery connection for 3 sec and plug them again 2.Replace Power Splitter Monitor is running without LCD display (VGA is working) 1.Bios Setting 2.LCD inverter cable 3.LCD inverter board 4.LCD cable 5.SBC failure 6.Power Splitter Failure 7.LCD failure 1.Set the bios to Mennen Medical Ltd. default in bios application 2.Replace the LCD inverter cable 3.Replace the LCD inverter board 4.Replace the LCD cable 5.Replace SBC 6.Replace Power Splitter board 7.Replace LCD Monitor is running without waveforms and vital sign. "Front end Failure" message is appearing 1.LAN cross cable failure. 2.Front End failure 3.CF failure 4Power Splitter Failure 5.SBC Failure 1.Replace LAN cable 2.Replace Front End board 3.Replace CF card 4.Replace Power Splitter board 5.Replace SBC No Battery charging 1.Battery failure 2.AC line failure 3.Power Splitter failure 4.Power Supply failure 1.Replace battery 2.Check AC fuses 3.Replace Power Splitter Board 4.Replace power supply Front Panel not working (Knob + Switches) 1. Front Panel Flat Cable connections 2. Power Splitter 3.Front Panel Board 1.Check connections between Front Panel to Power Splitter board and power splitter board to SBC (Com1,2) 2.Replace Power Splitter board 3.Replace Front Panel Mennen Medical 6-9

146 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual Table 6-2: VitaLogik 4000/6000 Bedside Monitor faults Problem Problem Cause Corrective Action Touch Screen not working 1.Application settings 2.Cable connections 3.Touch Controller Failure 4.Touch Plate failure 1.Check the setting for touch display in application (should be at enable state) 2.Check cables between Power Splitter board to touch controller and Com cable from power splitter to SBC 3.Replace touch controller 4.Replace touch plate Knob is not working (all other switches are working) 1.Knob failure 1.Replace Knob MCU Remote not working 1.Back DB9 connection setting at production 2.Remote failure 3.Power Splitter failure 1.Verify that the back DB9 connection is set to MCU remote by connection to power splitter board. 2.Replace Remote 3.Replace Power Splitter board Upper LEDs are not working 1.LED PCBA cable connection 2.LED PCBA failure 3.Front Panel failure 1.Check cable between LED PCBA to front panel board. 2.Replace Led PCBA 3.Replace Front Panel board No Audio 1.Application setup. 2.Speaker failure 3.SBC failure 1.Check the Alarm volume settings 2.Replace Speaker 3.Replace SBC Recorder Not working 1.Application settings 2.Cables connections 3.Power Splitter 4.Recorder 5.SBC 1.Verify that the application configured to use the internal recorder 2.Check cable and connections between Power splitter to SBC and Recorder 3.Replace Power Splitter in case of not having front panel leds indications 4.Replace Recorder 5.Replace SBC NIBP not working 1.NIBP Pump 2.NIBP Cable 3.Front End board 1.Replace pump 2.Replace NIBP Flat Cable 3.Replace Front End board EtCO2 Not Working 1.EtCo2 Module failure 2.EtCO2 cable failure 3.Front End failure 1.Replace Module 2.Replace EtCO2 Cable 3.Replace Front End board 6-10 Mennen Medical

147 VitaLogik 4000/6000 Service Manual Table 6-2: VitaLogik 4000/6000 Bedside Monitor faults Problem Problem Cause Corrective Action SpO2 Not measuring 1.SpO2 type mismatch 2.SpO2 Sensor failure 3.SpO2 cable failure 4.SpO2 FE connector failure 5.SpO2 boar failure 6.Front end failure 1.Verify SW version the SPO2 type 2.Replace Sensor 3.Replace cable 4.Replace SPO2 Panel connector 5.Replace SpO2 board 6.Replace Front End ECG,IBP,CO,TEMP failures UIM Not working 1.Parameter cable or sensor failure 2.Parameter FE connector failure 3.Front End Failure 1.Application setting 2.Interface cable 3.Target system failure 4.Internal cables failure 5.Front End failure 1.Replace parameter cable and sensor sets. 2.Replace parameter FE connector. 3.Replace Front End board 1.Verify the UIM setting in the application 2.Replace interface cable to target device 3.Replace target device 4.Replace VitaLogik internal UIM cable 5.Replace Front End Application is not running stack or cyclic restarting 1.Mennen Medical Ltd. configuration files 2.Compact flash 1.Update the application 2.Replace CF card Mennen Medical 6-11

148 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual Troubleshooting Vital Signs ECG/Respiration Monitoring Table 6-3: ECG/Respiration monitoring fault Problem Probable Cause Corrective Action No display 1. The system is not plugged in and switched on. 2. The ECG/Resp cable is not inserted correctly. 3. The Display is broken. 1. Check that the system is plugged in and switched on. 2.Check that the ECG cable is correctly inserted. 3. Call Mennen Medical Ltd. Service. ECG Cable Out message instead of HR/Resp numerics. 1. The ECG cable is not connected. 1. Check that the ECG cable is connected. Lead fault or All lead fault 1. One or more electrodes are detached. 2. No electrolyte gel on the electrodes. 3. Strong solvent used for cleaning skin. 4. A 3 Lead set is used and channels 2 and 3 are not switched off. 1. Check that the electrodes are correctly attached. 2. Check that there is sufficient gel on the electrodes. 3. Change solvents used for cleaning skin in line with your Hospital Procedure Policy. 4. Ensure that channels 2 and 3 are switched off or change to 5-lead set. ECG waveform is present but the Resp channel shows a flat line. Straight line on one or more ECG waveforms 1. One of the Resp electrodes is loose 2. No electrolyte gel on the electrodes. Lead fault 1. Check that the Resp electrodes are properly attached. 2. Check that there is sufficient gel on the electrodes. Reconnect the faulty lead 6-12 Mennen Medical

149 VitaLogik 4000/6000 Service Manual Response to Single Lead Fault Table 6-4: Responst to Single Lead Fault Lead Fault Active leads Straight lines LA II, avr, avl, avf I, III, all V RA III, avr, avl, avf I, II, all V LL I, avr, avl, avf II, III, all V RL V1 Vb (b = 2 to 6) I, II, III, avr, avl, avfm V2 to V6 I, II, III, avr, avl, avfm V2 to V6 I, II, III, avr, avl, avfm V1 to V6 (without Vb) V1 V1 Vb Invasive Blood Pressure Monitoring Table 6-5: Invasive Blood Pressure Monitoring fault Problem Probable Cause Corrective Action Abnormally low pressure No Pressure Noise or fling in the pressure waveform Dampened arterial waveform Resonating arterial waveform Transducer level higher than heart. Loose connections. If the monitor is switched on and there is no pressure, a transducer or connector wire may be broken. 1. Movement of PAP catheter tip. 2. Motion of pressure tubing. 1. Thrombus formation or blood left in catheter following blood sampling. 2. Large air bubble in tubing. 3. Catheter tip against the vessel wall. 4. Catheter kinking or arteriospasm. 5. Tubing too long or pliant. 6. Improper sequence of stopcock operation. 7.Defective transducer and/or amplifier. 1. Small air bubbles to the in, or very close transducer. 2. Tubing too long. Check patient and transducer positions. Check and tighten connections. 1. Replace transducer and cable. 2. Check stopcocks. 1. Reposition catheter (according to hospital procedures policy). 2. Make sure the patient or ventilator tubing is not touching the pressure tubing. Use syringe to withdraw air or particles in catheter, then flush the line with fresh solution. Use syringe to withdraw air in tubing, then flush line with fresh solution. Reposition the catheter to relieve spasm. Shorten the tubing (recommended length 3-4ft.) or replace with large diameter stiff tubing. Flush line, re-zero and recalibrate. Replace transducer and have the monitor checked. 1. Carefully flush the transducer and tubing when setting up system. Debubble carefully after each flush. 2. Shorten tubing. Mennen Medical 6-13

150 Maintenance and Cleaning VitaLogik 4000/6000 Service Manual EtCO2 Problem Probable Cause Corrective Action Airway Adapter occluded (intubated patient) Dirty/foggy adapter Incorrect adapter/sensor connection Sensor on top of circuit 1. Check adapter for cleanliness and fog. An airway adapter may fog when placed in a warm circuit. Wait a few minutes for the fogging to disappear. 2. Check that the adapter/sensor are correctly connected. Sidestream occlusion Kinks in line. Watertrap obstructed Check for kinks in the sampling line. If no kinks are evident, disconnect the sampling line from the water-trap. If message clears, replace the sampling line. If message remains, replace the watertrap. If message persists, contact your Mennen Medical Ltd. service representative. Error message Device fault detected If Mainstream is being activated, check that the sensor in use is the correct sensor recommended for use with the unit. Repeat start-up procedure. If the error message persists, contact your Mennen Medical Ltd. service representative. Exhaust occluded Kinked exhaust line, kinked internal pneumatics, bad pump Disconnect the exhaust port con-nection. If message persists, contact your Mennen Medical Ltd. service representative Mennen Medical

151 APPENDIX A: SPARE PARTS Appendix A includes recommended spare parts for VitaLogik 4X00 (see Table A-1) and VitaLogik 6000 (see Table A-2). Note: See comments in Table A-1 for annual replacement recommendation. Spare Parts VitaLogik 4X00 Table A-1: Spare Parts for VitaLogik 4X00 Part Number Description QTY. Repair Exchange Comments BATT Li-Ion battery pack 14.8V/5.2AH, Batt. for CPU Board Fuse slow blow 5x20 mm 2A 250 Vac P/S VAC to +24CDC 1 NO NO Replace after 3 years 1 NO NO Replace after 4 years 2 NO NO 1 NO Yes Optical Switch 2 NO NO Replace after 4 years Processor Card 1 NO yes Antenna for Wireless PCBA Wireless Assy Board (EU) Fan for CPU Board 1 NO NO 1 NO Yes 5 NO NO Replace after 3 years Mennen Medical A-1

152 VitaLogik 4000/6000 Service Manual Table A-1: Spare Parts for VitaLogik 4X00 Part Number Description QTY. Repair Exchange Comments Nibp Module 1 NO Yes SpO2 Nellcor OEM Board- NELL OEM Massimo board MS PCBA Carmel With 2ibp/NIBP & Spo2 Nellcor PCBA Carmel With 4ibp/NIBP & Spo2 Nellcor PCBA Carmel With 2IBP/NIBP & Spo2 Masimo PCBA Carmel With 4IBP/NIBP & Spo2 Masimo PCBA Carmel With 2ibp/NIBP & Spo2 Nellcor PCBA Carmel With 4ibp/NIBP & Spo2 Nellcor PCBA Carmel With 2IBP/NIBP & Spo2 Masimo PCBA Carmel With 4IBP/NIBP & Spo2 Masimo ECG connector assy. For vl SpO2 connector assy. Masimo SPO2 Connector Assy. Nellcor CO connector assy. For vl TEMP. Input Connector 1 NO Yes For Nellcor Monitor Type 1 NO Yes For Masimo Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Masimo Monitor Type 1 Yes Yes For Masimo Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Masimo MS2011 Monitor Type 1 Yes Yes For Masimo MS2011 Monitor Type 1 NO NO 1 NO NO For Masimo Monitor Type 1 NO NO For Nellcor Monitor Type 1 NO NO 1 NO NO A-2 Mennen Medical

153 VitaLogik 4000/6000 Service Manual Table A-1: Spare Parts for VitaLogik 4X00 Part Number Description QTY. Repair Exchange Comments SpO2 connector assy. MS2011 Masimo SPO2 Connector Assy. Nellcor Flash Card 1GB Include s.w Flash Card 8GB Include s.w 1 NO NO For Masimo MS2011 Monitor Type 1 NO NO For Nellcor Monitor Type 1 NO Yes 1 NO Yes Main Fan Assy. 1 NO NO Replace after 4Y Recorder OEM 1 NO Yes Option Cable - CPU board to LCD (LVDS) Cable - Splitter board to Touch Screen Cont. Board Cable - Splitter board to LCD Inverter board Cable - Splitter board to LPT Cable - Splitter board to front panel PCB assy Cable - Splitter board to Recorder Cable - Splitter to com.1/com Cable Assy. Speaker to CPU Cable - LAN cross, CIPAM to cpu board Cable - AC input to PSU Cable - DC, PSU to Splitter board 1 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 1 NO NO 0 NO NO 0 NO NO 0 NO NO Mennen Medical A-3

154 VitaLogik 4000/6000 Service Manual Table A-1: Spare Parts for VitaLogik 4X00 Part Number Description QTY. Repair Exchange Comments Cable - DC, splitter to CPU Cable - DC, Splitter to CIPAM Cable - Remote port to splitter Cable - DC, Splitter to CIPAM 0 NO NO 1 NO NO 1 NO NO 1 NO NO Cable Analog out 1 NO NO Cable - front panel PCB assy to upper leds (C.P.C.) Cable - wire for ground pin to PSU Cable-Y for Keyboard and mouse 1 NO NO 1 NO NO 1 NO NO Cable UIM 1 NO NO PCBA Power Splitter Board PCBA Front board (switches), PCBA, Touch Controller 1 NO Yes 1 NO NO 1 NO Yes Knob VL4XXX 3 NO NO Replace after 4 years Rear Cover VL4XXX Handle VL4XXX, Rubber for front key Plate Blank (instead of Touch Screen) Front Panel For VL NO NO 1 NO NO 1 NO NO 1 NO NO 1 NO NO A-4 Mennen Medical

155 VitaLogik 4000/6000 Service Manual Table A-1: Spare Parts for VitaLogik 4X00 Part Number Description QTY. Repair Exchange Comments Front panel V.L Rear Cover VL4XXX Assy " Internal LCD Display IBP connector assy. 1 NO NO 1 NO N0 1 NO Yes 1 NO NO Mennen Medical A-5

156 VitaLogik 4000/6000 Service Manual Spare Parts VitaLogik 6000 Note: See comments in Table A-2 for annual replacement recommendations. Table A-2: Spare Parts for VitaLogik 6000 Part Number Description QTY. Repair Exchange Comments BATT Li-Ion battery pack 14.8V/5.2AH, Batt. for CPU Board Fuse slow blow 5x20 mm 2A 250 Vac P/S VAC to +24CDC 1 NO NO Replace after 3 years 1 NO NO Replace after 4 years 2 NO NO 1 NO Yes Optical Switch 2 NO NO Replace after 4 years Processor Card 1 NO yes Antenna for Wireless PCBA Wireless Assy Board (EU) Fan for CPU Board 1 NO NO 1 NO Yes 5 NO NO Replace after 3 years Nibp Module 1 NO Yes SpO2 Nellcor OEM Board- NELL OEM Massimo board MS PCBA Carmel With 2ibp/NIBP & Spo2 Nellcor PCBA Carmel With 4ibp/NIBP & Spo2 Nellcor PCBA Carmel With 2IBP/NIBP & Spo2 Masimo 1 NO Yes For Nellcor Monitor Type 1 NO Yes For Masimo Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Masimo Monitor Type A-6 Mennen Medical

157 VitaLogik 4000/6000 Service Manual Table A-2: Spare Parts for VitaLogik 6000 Part Number Description QTY. Repair Exchange Comments PCBA Carmel With 4IBP/NIBP & Spo2 Masimo PCBA Carmel With 2ibp/NIBP & Spo2 Nellcor PCBA Carmel With 4ibp/NIBP & Spo2 Nellcor PCBA Carmel With 2IBP/NIBP & Spo2 Masimo PCBA Carmel With 4IBP/NIBP & Spo2 Masimo ECG connector assy. For vl SpO2 connector assy. Masimo SPO2 Connector Assy. Nellcor CO connector assy. For vl TEMP. Input Connector SpO2 connector assy. MS2011 Masimo SPO2 Connector Assy. Nellcor Flash Card 1GB Include s.w Flash Card 8GB Include s.w 1 Yes Yes For Masimo Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Nellcor Monitor Type 1 Yes Yes For Masimo MS2011 Monitor Type 1 Yes Yes For Masimo MS2011 Monitor Type 1 NO NO 1 NO NO For Masimo Monitor Type 1 NO NO For Nellcor Monitor Type 1 NO NO 1 NO NO 1 NO NO For Masimo MS2011 Monitor Type 1 NO NO For Nellcor Monitor Type 1 NO Yes 1 NO Yes Main Fan Assy. 1 NO NO Replace after 4Y Recorder OEM 1 NO Yes Option Cable - Splitter board to Touch Screen Cont. Board 0 NO NO Mennen Medical A-7

158 VitaLogik 4000/6000 Service Manual Table A-2: Spare Parts for VitaLogik 6000 Part Number Description QTY. Repair Exchange Comments Cable - Splitter board to LPT Cable - Splitter board to Recorder Cable - Splitter to com.1/com Cable - LAN cross, CIPAM to cpu board Cable - AC input to PSU Cable - DC, PSU to Splitter board Cable - DC, splitter to CPU Cable - DC, Splitter to CIPAM Cable - Remote port to splitter Cable - DC, Splitter to CIPAM 0 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 0 NO NO 1 NO NO 1 NO NO 1 NO NO Cable Analog out 1 NO NO Cable-Y for Keyboard and mouse Cable UIM RJ45 to 2*9PIN Cable UIM Flat Conn to 2*9Pin PCBA, Touch Controller Flash Card 16GB Include s.w Cable, CPU Board to LCD (LVDS) Cable, Spliter to front pcba 1 NO NO 1 NO NO 1 NO NO 1 NO Yes 1 NO NO 0 NO NO 0 NO NO A-8 Mennen Medical

159 VitaLogik 4000/6000 Service Manual Table A-2: Spare Parts for VitaLogik 6000 Part Number Description QTY. Repair Exchange Comments Cable, Recorder flat Cable Cable, Spliter To LCD Sharp Cable Speaker Assy PCBA Power Splitter Board PCBA Front board (switches), 0 NO NO 0 NO NO 1 NO NO 1 Yes Yes 1 NO NO Knob VL4XXX 3 NO NO Replace after 4Y Bezel for VL6X Plate, 15' Blank Plate for VL6X LED 15' Internal Display IBP connector assy. 1 NO NO 1 NO NO 1 NO NO 1 NO NO Mennen Medical A-9

160 VitaLogik 4000/6000 Service Manual A-10 Mennen Medical

161 APPENDIX B: DRAWINGS Appendix B includes figures and drawings that will assist you in disassembling and assembling VitaLogik 4000/6000 during maintanence and repair. External Connectors VitaLogik 6000 Assemblies VitaLogik Top Assembly VitaLogik 6000 Chassis ASSY with Touch Screen BEZEL Front Panel VL 6X00 ASSY Mechanical Connector SUB ASSY 15" for FOR VL6XXX Invasive and Non Invasive Rear Cover ASSY 15" VitaLogik 6000 Recorder Cover ASSY Bottom Cover VitaLogik 6000 ASSY Plate, Battery Cover, Plate ASSY PCBA VitaLogik ASSY VitaLogik 4X00 Assemblies VitaLogik 4X00 Top ASSY (Internal Wiring Diagram) VL4X00 Chassis ASSY BEZEL Front Panel VL 4X00 ASSY Mechanical Connector SUB ASSY 15" for FOR VL 4X00 Invasive and Non Invasive Rear Cover ASSY 15" Bottom Cover VitaLogik 4000 ASSY Plate, Battery Cover, Plate ASSY PCBA, VitaLogik ASSY. Mennen Medical B-1

162 VitaLogik 4000/6000 Service Manual External Connectors B-2 Mennen Medical

163 VitaLogik 4000/6000 Service Manual Patient Input Connectors Table 1: Connector ECG / RESP PCB Z1 Description Color Panel # Green 1 C2 BLUE & YELLOW 7 2 C1 GREEN 9 3 RA RED 2 4 LL BLACK 8 5 LA BLUE 1 6 COM YELLOW 3 7 SHIELD GRAY 4 8 RL BROWN 10 9 C6 WHITE C5 BLUE & WHITE 5 11 C4 BLUE & GREEN 6 12 C3 PINK CID VIOLET Table 2-1: SpO2 Masimo MS-11 PCB J18 Description Color Panel # Blue 1 Anode BLUE 1 2 Cat GREEN 9 3 INNER SH RED 2 4 OUTER SH BROWN 10 5 GND YELLOW 3 6 CID GRAY 4 7 IRCAT WHITE 11 8 RED CAT WHITE & BLUE 5 Mennen Medical B-3

164 VitaLogik 4000/6000 Service Manual Table 2-2: SpO2 Masimo MS-2011 PCB J18 Description Color Panel # Blue 1 Anode Blue 1 2 Cathode Green 9 3 Inner Shield Red 2 4 Outer Shield Brown 10 5 GND Yellow 3 6 CID Gray 4 7 IR LED cathode White 11 8 Red LED Cathode White&Blue 5 9 Sensor Common Black Sensor Return Pink 12 Table 3: SpO2 Nellcor PCB J18 Description Color Panel # Blue 1 Anode BLUE 1 2 Cat GREEN 6 3 INNER SH RED 2 4 OUTER SH BROWN 7 5 GND YELLOW 3 6 CID GRAY 4 7 IRCAT WHITE 11 8 RED CAT WHITE & BLUE 5 B-4 Mennen Medical

165 VitaLogik 4000/6000 Service Manual Table 4: IBP ( 1, 2 ) PCB (J11,J15) Description Color Panel # Red 1 CID BLUE 1 2 BP1+ RED 2 3 BP1- YELLOW 3 4 NC GRAY 4 5 EXC BROWN 5 6 GND GREEN 6 7 GND BLACK 7 8 GND BROWN 8 Mennen Medical B-5

166 VitaLogik 4000/6000 Service Manual Table 5: IBP CONNECTRORS ( 1, 2 ) ; ( 3, 4 ) PCB (J2,J4) ; (J11,J15) Description Color Panel # Violet 1 CID RED 2 2 BP1+ BROWN 5 3 BP1- GREEN EXC BLUE 1 6 GND YELLOW CID RED 2 2 BP2+ BLACK 7 3 BP2- WHITE EXC GRAY Table 6: Temp 1 PCB J6 Discretion Color Panel # Black 1 TEMP1 BLUE 1 2 PID1 RED 2 3 GND YELLOW B-6 Mennen Medical

167 VitaLogik 4000/6000 Service Manual Table 7: Temp 2 PCB J6 Discretion Color Panel # Black TEMP2 GRAY 1 5 PID2 BROWN 2 6 GND GREEN 3 Table 8: CO / 2 Temp PCB J1 Discretion Color Panel # Orange 1 CID BLUE 1 2 GND RED 2 3 INJ YELLOW 3 4 GND GRAY 4 5 NC BROWN 5 6 REF GREEN 6 7 BLOOD BLACK 7 8 WHITE 8 Mennen Medical B-7

168 VitaLogik 4000/6000 Service Manual VitaLogik 6000 Assemblies VitaLogik Top Assembly B-8 Mennen Medical

169

170 VitaLogik 6000 Chassis ASSY with Touch Screen VitaLogik 4000/6000 Service Manual B-10 Mennen Medical

171 REVISIONS ZONE REV. DESCRIPTION DATE APPROVED F A 11/30/2009 RELEASED FOR PRODUCTION F E E RECORDER (P/N 781OPT400) D D OPTION 780OPT400 - WITH RECORDER (see sheet #5) C C B B P/N VL 6XXX Recorder Cover ASSY NOTES: 1. OPTION VL4XXX RECORDER A 781OPT400 - see sheet #5. 2. OPTION TOUCH SCREEN B 781OPT012 - see sheet #6. A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION 8/27/2009 8/27/2009 THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: SCALE: SHEET 1 OF 1:2.5 VL6000 CHASSIS ASSY SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: B REV. A ±X.X ±X.X ±X.X ±X.X A 6

172 F X F X4 X8 X8 ITEM NO. DESCRIPTION Default/QTY. 1 15" INTERNAL LCD 1 2 CHASSIS 15" 1 3 SPACER K30 S6x CPAM BRACKET 1 5 SPEAKER 5WATT 1 E 6 RING FOR VL6XXX SPEAKER 1 7 LINE CORD RETAINER 15" 1 8 RECORDER HOLDER 1 22 X6 15 X2 24 X4 9 PLATE, BATTERY COVER PLAT ASSY 15" 2 E 10 BATTERY SPACER 1 11 INVERTER, 15" LCD INVERTER PCA 1 12 PCB HSB-6681-A10-E2-MENNEN 1 13 AC INPUT MODULE-ASSY 1 14 COVER, VL6XXX CPU AND RECORDER COVER X4 X4 15 F E SCREW LOCK ASSY RING FOR VL 6XXX FAN 1 17 VL6XXX Recorder Cover ASSY 1 18 FAN ASSY 1 19 SCREW #4-40x3/8 HX WH SLOT 4 20 Vesa 100 BRACKET 1 21 SPACER K30 S6x16 2 D D 22 SCREW M3x6 PH, PHILLIPS SS DIN 7985A A WASHER-FLAT M3 DIN 125A-A WASHER-SPLIT LOCK M3 SS DIN 127B A WAY D TYPE SOCKET CONNECTOR SOLDER TERMINATION 2 26 PCB POWER SPLITTER BOARD 1 27 P/S AC VAC to +24V DC 1 C X4 33 X4 30 X X2 28 SCREW M3x10 PHILLIPS SS DIN 7985A A Stand off M3x25 S FLAT HEAD CROSS RECESS SCREW M3x PAN HEAD SCREW,TORX,WN1452 K30x ANALOG CABLE 1 33 REMOTE CABLE 1 34 FUSE 2 31 X7 31 C B X X4 16 X4 X4 3 X B A X2 X2 X2 X X X X2 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DESIGN DRAWIN CHECK PROJ. ENG. Q.C DATE BY ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION 8/27/2009 8/27/2009 THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VL6000 CHASSIS ASSY SCALE: SHEET 2 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: B REV. A ±X.X ±X.X ±X.X ±X.X A 6

173 F F E E D D 1 C C B 7 16 B A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VL6000 CHASSIS ASSY ASSY STEP #1 SCALE: SHEET 3 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: B REV. A ±X.X ±X.X ±X.X ±X.X A 6

174 F F E 12 E 70 s/n D D 3 21 C C 20 B B A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VL6000 CHASSIS ASSY ASSY STEP #2 SCALE: SHEET 4 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: B REV. A ±X.X ±X.X ±X.X ±X.X A 6 4

175 F F 39 E E D D A P/N 781OPT400 C C B B A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VL6000 CHASSIS ASSY WITH RECORDER SCALE: SHEET 5 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: REV. A B ±X.X ±X.X ±X.X ±X.X A 6

176 F F E B P/N E D D C C B B A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VL6000 CHASSIS ASSY WITH TOUCH SCREEN SCALE: SHEET 6 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: B REV. A ±X.X ±X.X ±X.X ±X.X A 6

177 VitaLogik 4000/6000 Service Manual BEZEL Front Panel VL 6X00 ASSY B-12 Mennen Medical

178 REVISIONS ZONE REV. DESCRIPTION DATE APPROVED F B A 11/30/2009 RELEASED FOR PRODUCTION F E E D D A (2 : 1) C 13 C 1 4 x13 B ITEM NO. DESCRIPTION Default/QTY. 1 SW OPTICAL ENCODER, 24 POS x12 x12 x12 A2 A1 5 x13 B 2 KNOB for VL 6XXX 1 3 BEZEL FOR VL6X SCREW M3x6 FH PHILLIPS SS DIN 965 A PLATE, TOUCH SCREEN SUPPORT 13 6 SEALING TOUCH SCREEN 1 A1 A2 NOTES: T.S OP012. NON T.S OP014. A 7 SCREW M3x6 PH, PHILLIPS SS DIN 7985A A WASHER-SPLIT LOCK M3 SS DIN 127B A WASHER-FLAT M3 DIN 125A-A PCBA FOR UPPER LEDS 1 11 PCBA FRONT BOARD LIGHTGUIDES KIT 1 13 RUBBER FOR FRONT KEY 1 14 CABLE FOR TOP LEDs HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION 6/21/2009 6/21/2009 THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: BEZEL FRONT PANEL VL 6X00 ASSY SCALE: SHEET 1 OF 1:5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: REV. A B ±X.X ±X.X ±X.X ±X.X A A 1

179 VitaLogik 4000/6000 Service Manual Mechanical Connector SUB ASSY 15" for FOR VL6XXX Invasive and Non Invasive B-14 Mennen Medical

180 REVISIONS REV. DESCRIPTION DATE APPROVED 1 RELEASED FOR PROTOTIPE 6/21/ ADD ISOLATIONPLATE ITEM #9 7/3/2009 A RELEASED FOR PRODUCTION 11/30/2009 D D B C C RED WIRE GREEN WIRE B A B NOTES: A B SPO2 CONNECTOR, SEE UNIT TOP ASSY. FOR ETCO2 OPTION. ITEM NO. DESCRIPTION QTY. 6 1 PLUG, AIR CONNECTOR, 3mm HOSE TAIL 1 2 SPO2 CONNECTOR ASSY 1 A 3 TEMP. CONNECTOR ASSY 1 4 PLATE ETCO2 SUPPORT 2 5 SCREW M2.5x6 PH, PHILLIPS SS DIN 7985A A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A 6 IBP CONNECTOR ASSY. FOR CFE 2 7 CO CONNECTOR ASSY 1 8 PLATE, INPUTS CONNECTORS SUPPORT 1 9 ISOLATION PLATE 1 B ETCO CONNECTOR ASSY 1 P:\Mennen Monitor 15 inch \Mennen 15inch PRODUCTION\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY 6/21/2009 6/21/2009 DESCRIPTION: MECHANICAL CONNECTOR SUB ASSY 15" FOR VL6XXX INVASIVE AND NON INVASIVE SCALE: 1:2 SHEET 1 OF 1 SIZE A3 PART No: /650/680 BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A

181 VitaLogik 4000/6000 Service Manual Rear Cover ASSY 15" B-16 Mennen Medical

182 REV. DESCRIPTION DATE SIGN A RELEASED FOR PRODUCTION 11/30/2009 D x4 x6 D C C B B ITEM NO. DESCRIPTION Default/QTY. 1 COVER, VL6XXX REAR COVER 1 2 Self-adhesive, Polyurethane Bumper BS FRAME, VL6XXX FRONT FRAME 1 4 FLAT HEAD CROSS RECESS SCREW M3x8 6 A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A P:\Mennen Monitor 15 inch \Mennen 15inch PRODUCTION\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY 8/20/2009 8/20/2009 DESCRIPTION: SCALE: 1:2.5 SHEET 1 OF REAR COVER ASSY 15" 1 SIZE A3 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A

183 VitaLogik 4000/6000 Service Manual VitaLogik 6000 Recorder Cover ASSY B-18 Mennen Medical

184 REVISIONS ITEM NO. DESCRIPTION QTY. ZONE REV. DESCRIPTION DATE APPROVED 1 COVER, VL6XXX RECORDER COVER 1 2 RECORDER COVER BRACKET 1 A B RELEASED FOR PRODUCTION 11/30/ SCREW M3x6 PH, PHILLIPS SS DIN 7985A A2 4 4 WASHER-SPLIT LOCK M3 SS DIN 127B A2 4 5 WASHER-FLAT M3 DIN 125A-A2 4 D D 1 2 C C x4 x4 x4 B B A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A P:\Mennen Monitor 15 inch \Mennen 15inch PRODUCTION\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY 8/26/2009 8/26/2009 DESCRIPTION: SCALE: 1:1 SHEET 1 OF VL6XXX Recorder Cover ASSY PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER SIZE N10 1 A3 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A B

185 VitaLogik 4000/6000 Service Manual Bottom Cover VitaLogik 6000 ASSY B-20 Mennen Medical

186 REVISIONS ZONE REV. DESCRIPTION DATE APPROVED A RELEASED FOR PRODUCTION 11/30/ D D C 2 C 3 B B ITEM NO. DESCRIPTION QTY. 1 BOTTOM COVER VITALOGIK 60XXX 1 2 SCREW QUITCK LOCK FEMALE (RoSH) 2 3 SCREW QUICK LOCK MALE (RoHS) 2 A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A P:\Mennen Monitor 15 inch \Mennen 15inch PRODUCTION\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY 4/30/2009 4/30/2009 DESCRIPTION: BOTTOM COVER VL6XXX ASSY PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER SCALE: 1:1 SIZE N10 SHEET 1 OF 1 A3 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A

187 VitaLogik 4000/6000 Service Manual Plate, Battery Cover, Plate ASSY B-22 Mennen Medical

188 REVISIONS ZONE REV. DESCRIPTION DATE APPROVED 1 RELEASED FOR PRODUCTION 5/7/ CHANGE DIMENSION 5/11/ NEW DESIGN 7/16/2009 D A RELEASED FOR PRODUCTION 11/22/2009 D C C R3 typ B B (x2) FOR SET A NOTES: 1. REMOVE BURRS AND BREAK SHARP CORNERS. 2. BENDING RADIUS min AVAILABLE. 3. FOR DETAILS SEE THE ATTACHED SOLID MODEL FILE. P:\Mennen Monitor 15 inch \Mennen 15inch PDR\ HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY 4/30/2009 4/30/ DESIGN DRAWIN CHECK PROJ. ENG. Q.C THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: SCALE: 1:1 SHEET 1 OF AL 5052 H32 1.5mm THK. FINISH: PLATE, BATTERY COVER PLATE PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER SIZE N10 1 A3 SURFACE ROUGH. DIMENSIONS: mm 1 ALODINE TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A B A

189 VitaLogik 4000/6000 Service Manual PCBA VitaLogik ASSY B-24 Mennen Medical

190 REVISIONS REV. DESCRIPTION DATE SIGN. A RELEASED FOR PRODUCTION 20/02/2008 B CHANGE DRAWING 11/5/2008 D D C C B B A NOTES: 1. (*) NUMBER OF TOP ASSY: "A" ETCO2 OPTION. 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: SCALE: 1:2 SHEET 1 OF 3 PCBA, V.L. Assy. SIZE A3 FINISH: PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B A

191 ITEM NO. PART NUMBER DESCRIPTION /QTY PCBA, VL Mother Board for V.L SpO2 Nellcor OEM Board-NELL SPACER FF M3 L=7mm NYLON SCREW M3x6MM NY WASHER FLAT NYLON 3.2x7x0.5 Nylon-6 2 D D NIBP Module (CAS OEM)-ND 12VDC 9600 Baud Rate SCREW 4-14x1/4 PLT-30 PH PHIL STL WASHER - FLAT NYLON # SPACER M3 6x16 POM+30% GLASS FIBERS NIBP Cable Assy.(60mm) SCREW M3x5 PH, PHIL SS DIN 7985A A2 1 "A" 641 OPT 200 ETCO2 OPTION C C 5 26 B B A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A * ASSY PCBA, V.L. Assy. with NIBP & Neiicon DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: SCALE: 1:2 SHEET 2 OF 3 PCBA, V.L. Assy. SIZE A3 PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B

192 ITEM NO. PART NUMBER DESCRIPTION /QTY PCBA, VL Mother Board for V.L SPO2 OEM BOARD SPACER 8mm NIBP Module (CAS OEM)-ND 12VDC 9600 Baud Rate SCREW 4-14x1/4 PLT-30 PH PHIL STL 22 D D SPACER M3 6x16 POM+30% GLASS FIBERS NIBP Cable Assy.(60mm) WASHER - FLAT NYLON # C C 641 OPT 200 ETCO2 OPTION B B * ASSY PCBA, V.L. Assy. with NIBP & Massimo A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: SCALE: 1:2 SHEET 3 OF 3 PCBA, V.L. Assy. SIZE A3 PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B

193 VitaLogik 4X00 Assemblies VitaLogik 4000/6000 Service Manual VitaLogik 4X00 Top ASSY (Internal Wiring Diagram) B-26 Mennen Medical

194 6 ITEM NO. DESCRIPTION QTY. 1 PCBA, V.L. Assy. 1 2 MAINFRAME SUB ASSY FOR VL4XXX 1 3 BEZEL FRONT PANEL V.L. 4X00 ASSY 1 4 MECHANICAL CONNECTOR SUB ASSY 1 5 CONNECTORS COVER INPUT 1 6 SPO2 CONNECTOR 1 10 SCREW M3x8 PH, PHILLIPS SS DIN 7985A A SCREW M4x8 PH, PHILLIPS SS DIN 7985A A SCREW M4x10 PH, PHILLIPS SS DIN 7985A A SCREW M4x8 FH, PHILLIPS SS DIN 965 A REAR COVER VL4XXX ASSY 1 15 BOTTOM COVER VL4XXX ASSY 1 17 ETCO DOOR 1 18 UIM PORT DOOR-SET 2 20 LABLE P/N AND S/N VITALOGIK 1 21 SPO2 LABLE 1 22 BKT-LINE CORD RETAINER 1 5 C REVISIONS REV. DESCRIPTION DATE SIGN. A RELEASED FOR PRODUCTION 20/02/2008 B ADD SHEET #2 25/06/2008 C CHANGE OF MATERIALS 06/11/2008 F F E 6 4 E D D C C B B NOTES: 1. (*) NUMBER OF TOP ASSY: A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: VITALOGIK 4X00 TOP ASSY DESCRIPTION: SCALE: SHEET 1 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: REV. C ±X.X ±X.X ±X.X ±X.X A 2 B

195 F F E E D D C C B B A HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: VITALOGIK 4X00 TOP ASSY (INTERNAL WIRING DIAGRAM) SCALE: SHEET 2 OF 1:2.5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: XXX BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: REV. C ±X.X ±X.X ±X.X ±X.X A 2

196 VitaLogik 4000/6000 Service Manual VL4X00 Chassis ASSY B-28 Mennen Medical

197

198

199

200

201

202

203

204

205

206

207

208 VitaLogik 4000/6000 Service Manual BEZEL Front Panel VL 4X00 ASSY B-30 Mennen Medical

209 REVISIONS REV. DESCRIPTION DATE SIGN. A RELEASED FOR PRODUCTION 24/02/2008 B CHANGE DESIGN 06/11/2008 F F E E B D D C 11 C B 1 12 A B ITEM NO. DESCRIPTION QTY. 1 SW OPTICAL ENCODER, 24 POS 1 2 SCREW M3x6 PH, PHILLIPS SS DIN 7985A A CABLE, FRONT PANEL PCBA TO UPPER LEDS 1 4 PCBA Front board (switches) 1 5 PCBA FOR UPPER LEDS 1 NOTES: 1. A T. S OPT012 NON T.S OPT014. A 6 PLATE, TOUCH SCREEN SUPPORT 10 7 UPPER LED HOUSING (SET OF LEFT & RIGHT)(C.P.C.) 1 8 BUTTON PAD - COLOR OPTTION #1 1 9 SEALING TOUCH SCREEN 1 10 FRONT PANEL V.L.4000 (Non Invasive) 1 11 KNOB V.L. 45XXX 1 12 FOOT-STABILIZING 2 13 SCREW M3x6 FH PHILLIPS SS DIN 965 A B 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DATE BY DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: BEZEL FRONT PANEL V.L. 4X00 ASSY SCALE: SHEET 1 OF 1:5 SIZE A2 FINISH: SURFACE ROUGH. 1 PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER X DIMENSIONS: mm TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: HOLES LOCATION: ANGLES: REV. B ±X.X ±X.X ±X.X ±X.X A 1

210 VitaLogik 4000/6000 Service Manual Mechanical Connector SUB ASSY for FOR VL 4X00 Invasive and Non Invasive B-32 Mennen Medical

211 REVISIONS REV. DESCRIPTION DATE APPROVED A RELEASED FOR PRODUCTION 19/02/2008 B CHANGE QTY ITEM #5 10/13/2008 D D B B C C RED WIRE GREEN WIRE B B A B NOTES: 1. A CONNECTOR FROM THIS PLACE SEE IN A TOP ASSY. 2. ITEM "B" FOR 681OPT200 ONLY. 7 B ITEM NO. DESCRIPTION QTY. 1 PLUG, AIR CONNECTOR, 3mm HOSE TAIL 1 A 2 ECG CONNECTOR ASSY 1 3 TEMP. CONNECTOR ASSY 1 4 PLATE, INPUTS CONNECTORS SUPPORT 1 5 PLATE ETCO2 SUPPORT 2 6 SCREW M2.5x6 PH, PHILLIPS SS DIN 7985A A IBP CONNECTOR ASSY. FOR CFE 2 8 CO CONNECTOR ASSY 1 B ETCO CONNECTOR ASSY 1 P:\MENNEN MONITOR PROTOTYPE-verB\ B 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: MECHANICAL CONNECTOR SUB ASSY FOR VL4XXX INVASIVE AND NON INVASIVE SCALE: 1:2 SHEET 1 OF 1 SIZE A3 FINISH: PART No: /450/480 BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B A

212 VitaLogik 4000/6000 Service Manual Rear Cover VL4XXX ASSY B-34 Mennen Medical

213 REVISIONS REV. DESCRIPTION DATE APPROVED A RELEASED FOR PRODUCTION 19/02/ D D C 3 C 13 ITEM NO. DESCRIPTION QTY. 1 SPEAKER AND HARNESS ASSY 1 2 BRACKET HANDLE 1 3 BRACKET SCREW HANDEL SUPPORT 2 4 AXLE (HANDLE SUPPORT) 2 B 12 5 REAR COVER VL4XXX 1 6 HANDLE VL4XXX 1 7 CAP SCREW HANDLE VL4XXX 2 8 FOOT-STABILIZING 2 B 11 9 SCREW M3x6 PH, PHILLIPS SS DIN 7985A A SCREW M4x10 PH, PHILLIPS SS DIN 7985A A WASHER SPLIT LOCK M4 SS DIN 127 B A WASHER-FLAT M4 SS DIN 125A A SCRW M4x6 PH, PHILLIPS SS DIN 7985A A2 4 1 A 8 M:\MENNEN MONITOR PROTOTYPE-verB\ HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: SCALE: 1:5 SHEET 1 OF 1 REAR COVER VL4XXX ASSY SIZE A3 FINISH: PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A A

214 VitaLogik 4000/6000 Service Manual Bottom Cover VitaLogik 4000 ASSY B-36 Mennen Medical

215 REVISIONS REV. DESCRIPTION DATE APPROVED A RELEASED FOR PRODUCTION 19/02/2008 D D C C B B ITEM NO. DESCRIPTION QTY. 1 BOTTOM COVER VL4XXX 1 2 SCREW QUICK LOCK MALE (RoHS) 2 3 SCREW QUITCK LOCK FEMALE (RoSH) 2 A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A M:\MENNEN MONITOR PROTOTYPE-verB\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: BOTTOM COVER VL4XXX ASSY PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER SCALE: 1:2 SIZE N10 SHEET 1 OF 1 A3 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A

216 VitaLogik 4000/6000 Service Manual Plate, Battery Cover, Plate ASSY B-38 Mennen Medical

217 ZONE REV. DESCRIPTION DATE APPROVED A RELEASED FOR PRODUCTION 09/03/ D D 3 1 C C B B ITEM NO. PART NUMBER DESCRIPTION QTY PLATE, BATTERY COVER PLATE SCREW QUICK LOCK MALE (RoHS) SCREW QUITCK LOCK FEMALE (RoSH) 1 A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A M:\MENNEN MONITOR PROTOTYPE-verB\ DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: PLATE, BATTERY COVER PLAT ASSY PART No: BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER SCALE: 1:1 SIZE N10 SHEET 1 OF 1 A3 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. A

218 VitaLogik 4000/6000 Service Manual PCBA, VitaLogik ASSY. B-40 Mennen Medical

219 REVISIONS REV. DESCRIPTION DATE SIGN. A RELEASED FOR PRODUCTION 20/02/2008 B CHANGE DRAWING 11/5/2008 D D C C B B A NOTES: 1. (*) NUMBER OF TOP ASSY: "A" ETCO2 OPTION. 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. MATERIAL: DESCRIPTION: SCALE: 1:2 SHEET 1 OF 3 PCBA, V.L. Assy. SIZE A3 FINISH: PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 TOLERANCES,mm UNLESS OTHERWISE SPECIFIED LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B A

220 ITEM NO. PART NUMBER DESCRIPTION /QTY PCBA, VL Mother Board for V.L SpO2 Nellcor OEM Board-NELL SPACER FF M3 L=7mm NYLON SCREW M3x6MM NY WASHER FLAT NYLON 3.2x7x0.5 Nylon-6 2 D D NIBP Module (CAS OEM)-ND 12VDC 9600 Baud Rate SCREW 4-14x1/4 PLT-30 PH PHIL STL WASHER - FLAT NYLON # SPACER M3 6x16 POM+30% GLASS FIBERS NIBP Cable Assy.(60mm) SCREW M3x5 PH, PHIL SS DIN 7985A A2 1 "A" 641 OPT 200 ETCO2 OPTION C C 5 26 B B A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A * ASSY PCBA, V.L. Assy. with NIBP & Neiicon DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: SCALE: 1:2 SHEET 2 OF 3 PCBA, V.L. Assy. SIZE A3 PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B

221 ITEM NO. PART NUMBER DESCRIPTION /QTY PCBA, VL Mother Board for V.L SPO2 OEM BOARD SPACER 8mm NIBP Module (CAS OEM)-ND 12VDC 9600 Baud Rate SCREW 4-14x1/4 PLT-30 PH PHIL STL 22 D D SPACER M3 6x16 POM+30% GLASS FIBERS NIBP Cable Assy.(60mm) WASHER - FLAT NYLON # C C 641 OPT 200 ETCO2 OPTION B B * ASSY PCBA, V.L. Assy. with NIBP & Massimo A 4 HAYARDEN st. YAVNE. P.O.B. 102, REHOVOT 76100, ISRAEL TEL: FAX: MATERIAL: FINISH: THIS PUBLICATION CONTAINS PROPRIETARY INFORMATION OF MENNEN MEDICAL LTD. IT IS SUPPLIED TO THOSE CONCERNED WITH THE OPERATION AND MAINTENANCE OF THE EQUIPMENT COVERED THEREIN AND MAY NOT BE DICLOSED TO OTHERS WITHOUT THE EXPRES WRITTEN PERMISSION OF MENNEN MEDICAL LTD. TOLERANCES,mm UNLESS OTHERWISE SPECIFIED A DESIGN DRAWIN CHECK PROJ. ENG. Q.C ILYA B. ILYA B. ELI M. RELEASE FOR PRODUCTION DATE BY DESCRIPTION: SCALE: 1:2 SHEET 3 OF 3 PCBA, V.L. Assy. SIZE A3 PART No: XXX* BREAK ALL SHARP EDGES R0.1 Min. OR CHAMFER N10 SURFACE ROUGH. DIMENSIONS: mm 1 LENGTHS: HOLES: +0.1 HOLES LOCATION: 0.2 ANGLES: 0.5 REV. B

222 VitaLogik 4000/6000 Service Manual B-42 Mennen Medical

223 APPENDIX C: CEREBRALOGIK Introduction The CerebraLogik system is a VitaLogik module designed to monitor EEG and display long term Amplitude Integrated EEG (aeeg). The VitaLogik monitor uses the CerebraLogik front end amplifier to record two channels of EEG. The CerebraLogik is a stand alone unit to be attached to a basinet or patient bed. Its interface cable is connected to the UIM input of the monitor, and can be activated via the UIM menu. The EEG display shows real time EEG with history of EEG compressed as Amplitude Integrated EEG aeeg, showing 3 hours of EEG. CAUTION! It is not allowed to connect the interface cable to any equipment, other than VitaLogik monitors. Indications for Use The intended use of the CerebraLogik is to monitor the state of the brain by acquisition of EEG signals, and display the stored EEG in a compressed form of Amplitude Integrated EEG - aeeg and in conjunction with other clinical data. WARNING!!! EEG monitoring is intended for use in hospitals only, and under the direct supervision of a licensed healthcare practitioner trained in its proper use. CAUTION! The dual channel EEG monitoring does not replace full EEG examination. WARNING!!! CerebraLogik is not defibrillator proof; in case of defibrillation; disconnect the EEG electrodes from the CerebraLogik.. Mennen Medical C-1

224 Compliance VitaLogik 4000/6000 Service Manual VitaLogik 4000/6000 bedside monitor is also designed to comply with the international EMC requirements for medical electrical equipment IEC CerebraLogik Storage CAUTION! If the CerebraLogik is stored for a long period of time, it is recommended to store it on it's original packaging. Icons The following icons appear on the CerebraLogik adjacent to the electrode input connectors. Table 1: Icons Attention, consult accompanying documents (Service to be performed by qualified technician.) Type BF applied part Refer to instruction manual/ booklet Restricts the sale and use of this instrument to qualified medical personnel only. Not for use in explosive gas environment Electrical and electronic equipment - Dispose according to local regulation Date of manufacture CE Mark C-2 Mennen Medical

225 VitaLogik 4000/6000 Service Manual Patient Safety Patient safety is assured by double isolation between the EEG input and the electrical power. The double isolation prevents summation of leakage between the EEG electrodes and other applied parts. CAUTION! CerebraLogik is not intended for use during surgery and electrosurgery. Figure C-1: CerebraLogik Module CerebraLogik is equipped with sockets for 5mm DIN connectors, for use with disposable cup or needle electrodes with wires. Use Grass Safelead Surface or Subdermal needle electrodes or equivalent. The sockets are marked as Left +, Left, Right +, Right, REF. The five sockets build 2 channels of EEG in the following matter: The two Left inputs comprise Channel 1, marked as "Left" and the two Right inputs comprise Channel 2, marked as "Right". Both channels use the REF as a Reference point. Both channels are differential with high common mode rejection, thus preventing interference from devices such as pacemakers and peripheral stimulators. Each EEG channel can be displayed on the VitaLogik together with its corresponding aeeg channel. Note: The DIN connectors have protected female plugs, thus preventing the possibility that the plug touchs any conductor other than the CerebraLogik sockets. Mennen Medical C-3

226 VitaLogik 4000/6000 Service Manual WARNING!!! Use of electrodes that do not use DIN connectors is forbidden and may jeopardize the patient safety. CAUTION! Conductive parts of ELECTRODES, including the NEUTRAL ELECTRODE, should not come in contact with other conductive parts including earth. CerebraLogik Label Figure C-2: CerebraLogik label CerebraLogik - Waveforms and Display Modes VitaLogik equipped with CerebraLogik, can display two sections for each channel in parallel: real time EEG and aeeg waveforms. The real time EEG is shown under the correlating aeeg of each channel. The aeeg shows up to 3 hours with the current time at the right side of the panel. During the recording you can see the historical EEG waveforms in parallel as full disclosure of the specific patient in a special "History panel". The display can be configured in a variety of modes: EEG real time waveform display EEG real time with aeeg Very big numbers with one EEG and aeeg display EEG waveform plus one hour aeeg EEG and aeeg History panel C-4 Mennen Medical

227 VitaLogik 4000/6000 Service Manual Quality Bar and Graph The EEG signal quality is shown by a vertical resistance bar to the right of the EEG waveform. The historical quality is shown on the aeeg panel by a superimposed resistance line, with a scale of 0 to 5. Where zero (0) represents good quality and 5 worse quality or faulty electrode. CerebraLogik Full Disclosure Figure C-3: aeeg + EEG real time CerebraLogik records the EEG and aeeg during the entire time the patient is connected to the VitaLogik and to the CerebraLogik electrodes. You can export the full data at the end of the procedure or, if desired, sections of CerebraLogik recording can be exported before the patient is dismissed. You can choose the sections manually. Viewing of the exported data is done on any PC using the CerebraLogik Viewer software by Mennen Medical. Operating VitaLogik Monitor with CerebraLogik Preparing for Operation To prepare the CerebraLogik for operation 1. Connect the CerebraLogik to one of the UIM inputs of the VitaLogik monitor 2. From the Vital Signs menu select the relevant UIM channel 3. Select CerebraLogik and activate Mennen Medical C-5

228 VitaLogik 4000/6000 Service Manual 4. Attach the EEG scalp electrodes 5. Secure the CerebraLogik to a convenient location near the patient's head using the attachment clip. Electrode Location The two channels are marked as Left and Right. You can use either one or both channels, providing that a reference electrode is connected on the scalp. If two channels are used, a third Cross channel that is the difference between the Left and Right channel is also available for display. in Figure C-4: shows the electrode location for single or dual channel aeeg as recommended by El-Dib et al (2009)*. A B Figure C-4: Electrode location for aeeg (El-Dib et al (2009). A: Single channel electrode placement (P3-P4); B: Dual Channel Placement (C3-P3 and C4-P4) 1 1.* El-Dib M, Change T, Tsuchida TN, Clancy RR. Amplitude-Integrated electroencephalograpy in neonates Pediatr Neurol 2009;41: C-6 Mennen Medical

229 VitaLogik 4000/6000 Service Manual Technical Alarm If any of the electrodes is not connected, a technical alarm is created and the quality bar appears full. A label with the faulty electrode is presented to the right of the Realtime or History panels. For example: EEG Lead L (+) off, or EEG Lead R ( - ) for the Left positive electrode, or the Right negative electrode failure respectively. If more than one electrode is not connected or the reference electrode is not connected, the following message is displayed: EEG Check Electrodes. CerebraLogik Service CerebraLogik service is limited to: Replacing the complete module Replacing plastic parts Note: For repairs, send the module to Mennen Medical. CerebraLogik Drawing See the following pages for CerebraLogik drawing. Mennen Medical C-7

230

Menntor X7 SERVICE MANUAL. Modular Patient Monitors FDM

Menntor X7 SERVICE MANUAL. Modular Patient Monitors FDM Menntor X7 Modular Patient Monitors SERVICE MANUAL FDM-781-000 Rev.A Sept. 2014 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices Manufacturer s Name : Mennen Medical Ltd.

More information

OMNI TOUCH SCREEN PATIENT MONITOR

OMNI TOUCH SCREEN PATIENT MONITOR OMNI TOUCH SCREEN PATIENT MONITOR OMNI Intuitive Designed for a fast paced work environment, the Infinium Omni patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

Omni (K) PATIENT MONITOR

Omni (K) PATIENT MONITOR Omni (K) PATIENT MONITOR Omni (K) Intuitive Designed for a fast paced work environment, the Infinium Omni (K) patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

ENSEMBLE Central Nurse Station USER MANUAL

ENSEMBLE Central Nurse Station USER MANUAL ENSEMBLE Central Nurse Station USER MANUAL DDM-555-001 Rev C. Nov. 2014 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices Manufacturer s Name : Mennen Medical Ltd. 4 Hayarden

More information

DPM Summary of Features and Benefits

DPM Summary of Features and Benefits Summary of Features and Benefits PAT I E N T M O N I TO R 12.1" TFT display, configurable up to 8 waveforms, enables maximum data analysis of your most critical parameters Standard features include 3/5-lead

More information

T1 BEDSIDE MONITOR GUIDELINE

T1 BEDSIDE MONITOR GUIDELINE APPENDIX A T1 BEDSIDE MONITOR GUIDELINE 5.23 Software Version This document is a guideline only to be used as an aid to comprehensive Inservice training. NAME: HOSPITAL: DATE: VALIDATOR: A. OVERVIEW 1.

More information

SECTION L MONITORING EQUIPMENT

SECTION L MONITORING EQUIPMENT BREATH-ALERT Y1090 BREATH-ALERT MK4 MONITOR The Breath-Alert Mk4 respiratory monitor is a simple yet sensitive and reliable device for monitoring breathing. A heat-sensitive sensor is placed in the breath

More information

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version This document is a guideline only to be used as an aid to comprehensive Inservice training. NAME: HOSPITAL: DATE: VALIDATOR:

More information

Atlas Monitor Outline Train the Trainer

Atlas Monitor Outline Train the Trainer Atlas Monitor Outline Train the Trainer A. ATLAS Monitor 1. Atlas Monitor is not designed for use on children younger than 3 years old. 2. Weight: 9.5lbs to 13.2 lbs, depending on configuration 3. Variable

More information

SPECIFICATION FOR MULTIPARAMETER MONITOR

SPECIFICATION FOR MULTIPARAMETER MONITOR SPECIFICATION FOR MULTIPARAMETER MONITOR Eight channel high resolution colour TFT display with 12.1" screen size and auto spacing User Selectable Display formats and waveform colours. ECG monitoring -

More information

OMNI Express VS VERSATILITY IN VITAL SIGNS

OMNI Express VS VERSATILITY IN VITAL SIGNS OMNI Express VS VERSATILITY IN VITAL SIGNS OMNI Express VS TOUCHSCREEN easy and intuitive to use PORTABLE weighs less than 6lbs MOBILE backup battery powered The Omni Express VS is a new intuitive approach

More information

Product Specifications

Product Specifications 1.Safety Specifications 1.1Product Classification Class Type of protection against electric shock Specification Class I, with internal power supply Degree of protection against electric shock Anti-defibrillation:

More information

QAM.L L550 series. Patient Monitor Manual QAM.L

QAM.L L550 series. Patient Monitor Manual QAM.L QAM.L550.1112.2 L550 series Patient Monitor Manual QAM.L550.0915.3 Preface This manual has been provided to give detailed descriptions on DARAY s L550 patient monitor, covering the performance, operation

More information

PULSE OXIMETER PACKAGE

PULSE OXIMETER PACKAGE instrumentation and software for research PULSE OXIMETER PACKAGE CANL-425SV-A USER S MANUAL DOC-095 Rev. 1.6 Copyright 2012 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478

More information

General Skills. yes. Device controls, buttons, indicators, and connectors

General Skills. yes. Device controls, buttons, indicators, and connectors HeartStart MRx Skills Checklist This checklist assesses qualified medical personnel s understanding of MRx operation after instructor-based training. Perform a quick skills assessment using the General

More information

Vitalmax 4000 CL Product Specifications

Vitalmax 4000 CL Product Specifications Page 1 of 5 Vitalmax 4000 CL 510 Garden Avenue North, Clearwater, FL 33755 Phone: (727) 442-8118 Fax: (727) 443-7257 A. Mechanical Description Size 10" H x 13" W x 8"D (25 cm x 33 cm x 20 cm) Weight 18

More information

PM-9000 Express Portable Patient Monitor

PM-9000 Express Portable Patient Monitor FEATURES 12.1" color TFT with maximum 8 waveforms Rich in ECG software OxyCRG software available 40 seconds full-disclosure waveforms review Drug dose calculation Bed to bed view Networking capacity Multi-Gas/O2

More information

901-M / 901-ML. Pulse Oximetry System. Technical Information and Service Manual

901-M / 901-ML. Pulse Oximetry System. Technical Information and Service Manual 901-M / 901-ML Pulse Oximetry System Technical Information and Service Manual CONTENTS 1. PRELIMINARY REMARK...... 4 2. FUNCTION DESCRIPTION......... 5 2.1. INTENDED USE...5 2.2. CONTRAINDICATION...5 2.3.

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series Copyright 2005 Welch Allyn. All rights are reserved. No one is permitted to reproduce

More information

CLEO VERSATILITY IN VITAL SIGNS

CLEO VERSATILITY IN VITAL SIGNS CLEO VERSATILITY IN VITAL SIGNS CLEO PORTABLE weighs less than 3 lbs TOUCHSCREEN easy and intuitive to use MOBILE backup battery powered The Cleo is a new and intuitive approach to patient vital signs

More information

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor GE Healthcare Quick Guide Dash 2500 Patient Monitor Hardkeys Main Menu Hide menu and view entire waveform screen Display the large numeric vital signs Review or change alarm limits for all operating parameters

More information

V6 Vital Signs Monitor User s Manual

V6 Vital Signs Monitor User s Manual V6 Vital Signs Monitor User s Manual Product Information Product Model: V6 Product Name: Vital Signs Monitor Manufacturer: GuangdongBiolightMeditech Co.,Ltd. After Service Contact Information: Address:

More information

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor

OPERATOR S MANUAL. NPB-4000/4000C Patient Monitor OPERATOR S MANUAL NPB-4000/4000C Patient Monitor Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Nellcor Puritan Bennett, Durasensor, C-LOCK, Oxisensor II, Dura-Y, and the Nellcor

More information

Propaq LT Outline Train the Trainer

Propaq LT Outline Train the Trainer Train the Trainer A. Propaq LT and LTR Propaq LT Standalone monitor Propaq LTR Monitor that has the wireless Acuity feature enabled 1. Lightweight (Less than 2 pounds) 2. Uses rechargeable lithium ion

More information

TidalGuard HD. Handheld CO 2 / SpO 2 and CO 2 monitors. Operating Manual Rev. H (8/15)

TidalGuard HD. Handheld CO 2 / SpO 2 and CO 2 monitors. Operating Manual Rev. H (8/15) TidalGuard HD Handheld CO 2 / SpO 2 and CO 2 monitors Operating Manual 003-2275-00 Rev. H (8/15) Table of Contents Table of Contents Service Information... iv Preface... vi 1. Safety... 1-1 1.1 Safety

More information

Quick start guide. English

Quick start guide. English Quick start guide English Read your entire Astral User Guide or Clinical Guide before use. The Astral device 1. Adapter port Can be fitted with single limb adapter, single limb leak adapter or double limb

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2009 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate,

More information

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry Cardell Insight Veterinary Monitor For Models: 8014 Blood Pressure 8015 Blood Pressure, Pulse Oximetry User Guide 003-2981-00 Rev. B (2/21/17) Software Version 1.2V Product Information Dealer: Date of

More information

BeneVision Central Monitoring System. Quick Reference Guide

BeneVision Central Monitoring System. Quick Reference Guide BeneVision Central Monitoring System Quick Reference Guide NOTE: The Quick Reference Guide is not intended as a replacement to the Operating Instructions. Prior to operating equipment, the user must be

More information

VS2000V Veterinary Vital Signs Monitor Operation Manual

VS2000V Veterinary Vital Signs Monitor Operation Manual VS2000V Veterinary Vital Signs Monitor Operation Manual - English Version 1.0, July 2011 2011 UTECH Co., Ltd. All rights reserved. Table of Contents Table of Contents Warranty and Service Information...

More information

qube mini Overview

qube mini Overview D a t a S h e e t qube mini Overview The 91389 qube mini is a compact, robust, and light-weight patient monitor with extensive clinical usability. Key features of the qube mini monitor include: Eye-level

More information

5/ 5i/ 5c/ 7/ 7i/ 7c/ 5 elite/ 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite

5/ 5i/ 5c/ 7/ 7i/ 7c/ 5 elite/ 5i elite/ 5c elite/ 7 elite/ 7i elite/ 7c elite Summary The élance 5 / 5i / 5c / 7 / 7i / 7c /5 elite/ 5i elite / 5c elite / 7 elite / 7i elite /7c elite are integrated vital signs monitors with widescreen resistive touchscreen display. The monitors

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2014 Welch Allyn. All rights are reserved. To support the intended use of the product

More information

Connex Integrated Wall System 2.2 Software Version

Connex Integrated Wall System 2.2 Software Version Connex Integrated Wall System 2.2 Software Version Clinical Inservice Presentation MC12931 1 Table of Contents Device Overview & Intended Use Power ON Power OFF Home Screen Navigation Spot Profile Intervals

More information

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain OPERATOR S MANUAL Patient Monitor M20 EU representative TECNOMED 2000 S.L. Valencia, 25-28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1 Donghwa-ri, Munmak-eup, Wonju-si,

More information

740 SELECT. Multi-Parameter Monitor

740 SELECT. Multi-Parameter Monitor Multi-Parameter Monitor User Manual Addendum - CO2 Parameters This User Manual Addendum describes the features and operations of the Multi-Parameter monitor: Software Version 2.2 or above. 21-22-0333 Rev

More information

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models:

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models: Cardell Touch Veterinary Vital Signs Monitor For Models: 8013-001 8013-002 8013-003 8013-004 User s Guide TP200 Rev. A 003-2840-00 Rev. AA2 (6/20/18) Software Version V4.0.X Product Information Dealer:

More information

OVEN INDUSTRIES, INC.

OVEN INDUSTRIES, INC. OVEN INDUSTRIES, INC. OPERATING MANUAL Model 5C7-252 TEMPERATURE CONTROLLER With PLC Inputs Introduction Thank you for purchasing our controller. The Model 5C7-252 is an exceptionally versatile unit and

More information

MODEL 3X Series User s Manual

MODEL 3X Series User s Manual MODEL 3X Series User s Manual Pulse Oximeter with options for 24 hour Memory, Printer & USB / Infrared Ports MEDIAID INC. 17517 Fabrica Way Suite H Cerritos, CA 90703 USA (Tel) 714-367-2848 (Fax) 714-367-2852

More information

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N Nellcor Bedside SpO 2 Patient Monitoring System, PM100N In-Service Presentation Nellcor Bedside SpO 2 Patient Monitoring System is CE marked and commercially av ailable in European Union countries. Nellcor

More information

Menntor X7 Specification

Menntor X7 Specification Specification is a modular monitor consisting of a Host Monitor a Multi Parameter Module, and two Single Parameters Modules. Host Monitor Specification Hardware and Parameters is a modular monitor with

More information

ANGELUS (Ver 2.03) Patient Monitoring System

ANGELUS (Ver 2.03) Patient Monitoring System ANGELUS (Ver 2.03) Patient Monitoring System CONTENT Contents of Manual 1 Meaning of Symbols used in this manual 2 Section 1. GENERAL 3 1.1 Introduction 4 1.2 Composition of unit 8 1.3 Panel Description

More information

DINAMAP ProCare Monitor. Operation Manual

DINAMAP ProCare Monitor. Operation Manual DINAMAP ProCare Monitor Operation Manual Contents Introduction About the DINAMAP ProCare Monitor...7 Indications...7 Contraindications...7 Warnings... 8 Cautions...9 Product Compliance...10 Symbols...11

More information

Patient Monitoring System YM5500. Service Manual

Patient Monitoring System YM5500. Service Manual Patient Monitoring System YM5500 Service Manual SERVICE MANUAL For YM5500 Patient Monitor Mediana Co.,Ltd. 231-6 Nonhyun-Dong, Kangnam-Ku, Seoul, Korea Tel: (82) 2 542 3375 Fax: (82) 2 542 7447 YM5500

More information

M9500 Patient Monitor User s Manual

M9500 Patient Monitor User s Manual M9500 Patient Monitor User s Manual Guangdong Biolight Meditech Co., Ltd. Address: Innovation First Road, Technology Innovation Coast, Jinding, Zhuhai, P.R.CHINA Tel: +86-756-3399900 Fax: +86-756-3399989

More information

Ultraview SL2400 With Wireless-LAN-Option

Ultraview SL2400 With Wireless-LAN-Option Ultraview SL2400 With Wireless-LAN-Option Spacelabs Healthcare Connecting Innovation with Care Spacelabs Healthcare is fully committed to making strong connections: between clinicians and their patients;

More information

This is to certify that the optical output of the: SENSORNET LR-DTS Temperature Sensing System: (Variants Mark 1, Mark 2a and Mark 2b)

This is to certify that the optical output of the: SENSORNET LR-DTS Temperature Sensing System: (Variants Mark 1, Mark 2a and Mark 2b) Optical Services Ltd. This is to certify that the optical output of the: SENSORNET LR-DTS Temperature Sensing System: (Variants Mark 1, Mark 2a and Mark 2b) Meets the requirements stated in IEC 60825-1:2001

More information

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis)

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis) Quick Reference Guide STAN S31 fetal heart monitor Important! Read the STAN S31 User Manual. 1. 2. 3. 1 2 3 4 5 6 Transducer connectors 4. Function buttons Power switch 5. Information panel Fetal ECG (FECG)

More information

SERVICE MANUAL. Patient Monitor M30. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

SERVICE MANUAL. Patient Monitor M30. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain SERVICE MANUAL Patient Monitor M30 EU representative TECNOMED 2000 S.L. Valencia, 25-28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1 Donghwa-ri, Munmak-eup, Wonju-si,

More information

Transport Pro Patient Monitor Operator s Manual Software Version 2.1

Transport Pro Patient Monitor Operator s Manual Software Version 2.1 GE Healthcare Transport Pro Patient Monitor Operator s Manual Software Version 2.1 Transport Pro English 2024598-003 (CD) 2024579-020A (paper) 2007 General Electric Company. All rights reserved. The information

More information

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium.

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium. Patient Monitor HBP-2070 Operator s Manual EU representative OBELIS S.A Bd. Général Wahis, 53 1030 Brussels, Belgium Local distributor Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1

More information

CARDELL Veterinary Vital Signs Monitor

CARDELL Veterinary Vital Signs Monitor CARDELL Veterinary Vital Signs Monitor Models 9403, 9404 and 9405 User s Manual Manufactured for: CARDELL Veterinary Vital Signs Monitors IMPORTANT: This manual addresses all parameters of the CARDELL

More information

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1

Capnostream 20p Bedside Capnography Monitor. The only integrated, complete picture of oxygenation and ventilation 1 Capnostream 20p Bedside Capnography Monitor The only integrated, complete picture of oxygenation and ventilation 1 Smart Capnography and Pulse Oximetry Technology Smart Capnography is a suite of algorithms

More information

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X Vital Signs Monitor 300 Series Directions for Use Software Version 1.0X Copyright 2003, Welch Allyn Protocol, Inc. Welch Allyn is a registered trademark of Welch Allyn, Inc. Protocol is a trademark of

More information

The information in this document is subject to change without notice.

The information in this document is subject to change without notice. Operator s Manual NOTICE The information in this document is subject to change without notice. Welch Allyn makes no warranty of any kind on this material, including but not limited to, the implied warranties

More information

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English

Operator s Manual. Avant Pulse Oximeter and NIBP Monitor. English Operator s Manual Avant 2120 Avant 2120 Pulse Oximeter and NIBP Monitor 0123 English 1 CAUTION! CAUTION! Federal law (USA) restricts this device to sale by or on the order of a physician. Read this entire

More information

User s Reference Manual Volume 1 System Information

User s Reference Manual Volume 1 System Information Philips M1165/66/67/75/76/77A CMS Patient Monitoring System and Philips M1205A V24 and V26 Patient Monitor User s Reference Manual Volume 1 System Information PHI Part Number M1046-9220L Printed 02/2003

More information

SERVICE MANUAL YM6000 Patient Monitor

SERVICE MANUAL YM6000 Patient Monitor SERVICE MANUAL YM6000 Patient Monitor EU representative TECNOMED 2000 S.L. Valencia, 25 28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650 1 Donghwa ri, Munmak eup, Wonju

More information

GE Healthcare. Dash Variable-Acuity Monitoring

GE Healthcare. Dash Variable-Acuity Monitoring GE Healthcare Dash Variable-Acuity Monitoring The RIGHT information. Right now. In your busy schedule, having accurate patient information at your fingertips is critical. And there is no time for false

More information

DPM 4 Patient Monitor. Operator s Manual

DPM 4 Patient Monitor. Operator s Manual DPM 4 Patient Monitor Operator s Manual I Intellectual Property Statement Mindray DS USA, Inc. (hereinafter called Mindray DS) owns the intellectual property rights to this product and this manual. This

More information

Instructions for Use. Infinity Vista. Infinity Patient Monitoring Series

Instructions for Use. Infinity Vista. Infinity Patient Monitoring Series Instructions for Use Infinity Vista Infinity Patient Monitoring Series Infinity Configured Monitoring Series Infinity Vista Instructions for Use WARNING: For a full understanding of the performance characteristics

More information

CNAP MONITOR 500 HD QUICK GUIDE. This Quick Guide DOES NOT REPLACE the user manual!

CNAP MONITOR 500 HD QUICK GUIDE. This Quick Guide DOES NOT REPLACE the user manual! CNAP MONITOR 500 HD QUICK GUIDE This Quick Guide DOES NOT REPLACE the user manual! Version 07/2016 V1.5.5 EN 2 CONTENT SET UP MEASUREMENT NAVIGATION FAQs FUNCTIONS: FAST ACCESS Hemodynamic Parameters...5

More information

Multi-Parameter Patient Monitor

Multi-Parameter Patient Monitor KIZLON Multi-Parameter Patient Monitor Email: info@kizlonmedical.com Website: www.kizlonmedical.com Multi-Parameter Patient Monitor KPM-A1 Series s of Portable Patient Monitor KPM-A1 Series has color TFT

More information

PATIENT MONITORING SYSTEMS

PATIENT MONITORING SYSTEMS PATIENT MONITORING SYSTEMS PATIENT MONITORING SYSTEMS The benefits of HEYER Scalis to meet the market situation HEYER Scalis function & features make it fit to match the requirement of standard applications

More information

SureSigns VS4 Vital Signs Monitor

SureSigns VS4 Vital Signs Monitor SureSigns VS4 Vital Signs Monitor INSTRUCTIONS FOR USE Release A.04 English Notice Proprietary Information This document contains proprietary information, which is protected by copyright. Copyright Copyright

More information

Model 900. Pulse Oximeter User s Manual

Model 900. Pulse Oximeter User s Manual Model 900 Pulse Oximeter User s Manual MEDIAID INC. 17517 Fabrica Way Suite H, Cerritos, CA 90703 USA Phone: 714-367-2848; Fax: 714-367-2852 Website: www.mediaidinc.com POX010-900 Contents Table of Contents

More information

Dryer Controller M720

Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 2.00 Software version 2.00 Manual M720 Dryer controller Page 1 of 60 Document history Preliminary version: - Created in April, 2009 Hardware Version 2.00,

More information

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual

MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual MICROCAPSTAR End-Tidal CO 2 Analyzer Instruction Manual Read instructions carefully before operating this device. ❶ This device is not to be used for Human Life Support applications. ❷ To avoid possible

More information

M8000VET /M9000VET Veterinary Monitor User s Manual

M8000VET /M9000VET Veterinary Monitor User s Manual M8000VET /M9000VET Veterinary Monitor User s Manual Guangdong Biolight Meditech Co., Ltd. Address: Innovation First Road, Technology Innovation Coast, Jinding, Zhuhai, P.R.CHINA Tel: +86-756-3399900 Fax:

More information

Manufacturer s Responsibility Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of t

Manufacturer s Responsibility Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of t Preface Thank you for using M9500 patient monitor. In order to enable you to skillfully operate Monitor as soon as possible, we provide this user s manual with delivery. When you install and use this instrument

More information

MRI Patient Monitoring System (Model )

MRI Patient Monitoring System (Model ) MRI Patient Monitoring System (Model 865214) INSTRUCTIONS FOR USE Revision C English A Specifications Specifications PATIENT SAFETY GENERAL Conforms to UL STD 60601-1. Certified to CAN/CSA STD C22.2 No.

More information

Mark 25 Ultrapure Water Conductivity Analyzer

Mark 25 Ultrapure Water Conductivity Analyzer Martek Instruments, Inc. Mark 25 Ultrapure Water Conductivity Analyzer Instruction Manual WARRANTY POLICY Unless otherwise stated, MARTEK INSTRUMENTS, INC. warrants this equipment to be free from defects

More information

Operator s Manual. LifeSense LS1-9R. Vital Signs Monitor. Capnography/Pulse Oximeter

Operator s Manual. LifeSense LS1-9R. Vital Signs Monitor. Capnography/Pulse Oximeter Operator s Manual Vital Signs Monitor Capnography/Pulse Oximeter LifeSense LS1-9R CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner. Consult Instructions

More information

OPERATOR S MANUAL. For YM1000 Vital Signs Monitor. EU Representative TECNOMED 2000 S.L. Valencia, Madrid Spain

OPERATOR S MANUAL. For YM1000 Vital Signs Monitor. EU Representative TECNOMED 2000 S.L. Valencia, Madrid Spain OPERATOR S MANUAL For YM1000 Vital Signs Monitor EU Representative TECNOMED 2000 S.L. Valencia, 25 28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650 1 Donghwa ri, Munmak

More information

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English Operator s Manual Avant 2120 Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor 0123 English CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

More information

Tissue Oximeter. Operator s Manual

Tissue Oximeter. Operator s Manual Tissue Oximeter Operator s Manual ViOptix, Incorporated 39655 Eureka Drive Newark, CA 94560 Phone: 510-226-5860 Fax: 510-226-5864 Website: www.vioptix.com About This Manual Trademarks T.Ox Tissue Oximeter

More information

Quick setup guide. English

Quick setup guide. English Quick setup guide English The Astral device 1. Adapter port Can be fitted with single limb adapter, single limb leak adapter or double limb adapter (Astral 150 only) 2. Handle 3. Inspiratory port (to

More information

according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC

according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC The CNAP Monitor 500 meets the requirements of -mark 0408 according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC The product starting with the serial number

More information

PowerRouter application guideline

PowerRouter application guideline PowerRouter application guideline Software installation tool - version 3.4 Before operating the PowerRouter, you may initialize the PowerRouter by using the PowerRouter software installation tool. The

More information

1 Appendix - Product Specifications

1 Appendix - Product Specifications 1 Appendix - Product Specifications WARNING The patient monitor may not meet its performance specification if stored or used outside the manufacturer s specified temperature and humidity range. 1.1 Safety

More information

Adaptive CyCLO Technical and HMI User Guide. CyCLO User Guide. Version th December 2017 REV

Adaptive CyCLO Technical and HMI User Guide. CyCLO User Guide. Version th December 2017 REV CyCLO User Guide Version 2.00 19 th December 2017 REV 2.00 1 Contents 1. Hardware... 3 1.1. Introduction... 3 1.2. Electrical Specification... 3 1.3. Board Overview... 4 1.4. Electrical Installation...

More information

PATIENT MONITORING SYSTEMS

PATIENT MONITORING SYSTEMS PATIENT MONITORING SYSTEMS Patient Monitoring Systems DESIGN FEATURES HEYER VizOR is our premium solution for sophisticated patient monitoring applications. VizOR comes in its standard configuration with

More information

User Manual. Dryer Controller M720

User Manual. Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 1.00 Software version 1.00 Preliminary version Manual M720 Dryer controller Page 1 of 42 Document history Preliminary version: - Created in April, 2009

More information

PATIENT MONITORING SYSTEMS

PATIENT MONITORING SYSTEMS PATIENT MONITORING SYSTEMS Patient Monitoring Systems DESIGN FEATURES HEYER VizOR is our premium solution for sophisticated patient monitoring applications. VizOR comes in its standard configuration with

More information

OMNI. Patient Monitor USER S MANUAL

OMNI. Patient Monitor USER S MANUAL OMNI Patient Monitor USER S MANUAL Ver 1.0 2010 Infinium Medical, Inc. All rights reserved. Issued Date: Sep 30,2010 Infinium Medical Inc. Website: www.infiniummedical.com Address: 12151 62nd St North

More information

CASMED 740 Vital Signs Monitor

CASMED 740 Vital Signs Monitor CASMED 740 Vital Signs Monitor User s Manual 2 THE CASMED 740 VITAL SIGNS MONITOR FEATURES CASMED 740-1 Non-Invasive Blood Pressure and Pulse Rate. CASMED 740-2 Non-Invasive Blood Pressure, Pulse Rate

More information

DINAMAP PRO Series V2 Monitor. Operation Manual

DINAMAP PRO Series V2 Monitor. Operation Manual PRO Series 100-400V2 Operation Manual DINAMAP PRO Series 100-400V2 Monitor Operation Manual 1 DINAMAP PRO 400V2 Monitor 2 DINAMAP PRO Series 100-400V2 Monitor Operation Manual This manual is for DINAMAP

More information

PowerLogic ION Setup Meter Configuration Software Configuration Guide

PowerLogic ION Setup Meter Configuration Software Configuration Guide PowerLogic ION Setup Meter Configuration Software Configuration Guide 70002-0293-03 12/2010 Conventions Used in this Manual This section describes the symbols and terminology used in this guide. Symbols

More information

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual Veterinary Portable Multi-parameter Patient Monitor Operation Manual Content Chapter 1 Introduction... 1-1 1.1 General Information... 1-3 1.2 Screen Display... 1-4 1.3 Button Functions and Basic Operation...

More information

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A i.c³ User Guide For Helmer i.series Ultra-Low Freezers 360175-A/A Document History Revision Date CO Supersession Revision Description A 18 APR 2014* 9275 n/a Initial release. * Date submitted or change

More information

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use

Welch Allyn Vital Signs Monitor 6000 Series. Directions for use Welch Allyn Vital Signs Monitor 6000 Series Directions for use 2010 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the

More information

BEDSIDE MONITORING Dash 3000/4000 V5

BEDSIDE MONITORING Dash 3000/4000 V5 BEDSIDE MONITORING Dash 3000/4000 V5 Quick Reference Guide 2004 General Electric Company PN: 0304-CS-QRGD Rev. A 2018408-010 BEDSIDE MONITORING Dash 3000/4000 V5 Quick Reference Guide Components Display

More information

Multiparameter. er Monitor DM

Multiparameter. er Monitor DM d e n t a l a n d m e d i c a l e q u i p m e n t User manual Multiparameter er Monitor DM ENGLISH Contents 1 Introduction...1 1.1 General Information...1 1.2 Screen Display...2 1.3 Button Functions...5

More information

OPERATOR S MANUAL. NELLCOR N-200 Pulse Oximeter

OPERATOR S MANUAL. NELLCOR N-200 Pulse Oximeter OPERATOR S MANUAL NELLCOR N-200 Pulse Oximeter To contact Nellcor s representative: In the United States, call 1-800-NELLCOR or 510 463-4000; outside of the United States, call Nellcor s local representative.

More information

Instructions for Use Vista 120

Instructions for Use Vista 120 Instructions for Use Vista 120 To properly use this medical device, read and comply with these Instructions for Use. Patient Monitor Software 1.n Release date April 25 2011 Screen images Schematic renderings

More information

B40 Patient Monitor. Affordable clinical excellence. Advanced clinical parameters. Ease of use for fast decision-making. Performance and reliability

B40 Patient Monitor. Affordable clinical excellence. Advanced clinical parameters. Ease of use for fast decision-making. Performance and reliability B40 Patient Monitor Affordable clinical excellence Patients with acute, life-threatening conditions need the best possible care. The B40 Monitor from GE Healthcare provides a continuous flow of quality

More information

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas User s Manual TIGER S EYE E-Series Mark V Jockey TIGERFLOW Systems, Inc. 4034 Mint Way Dallas, Texas 75237 214-337-8780 www.tigerflow.com TABLE OF CONTENTS Introduction... 4 Sequence of Operation... 5

More information

PM-9000Vet Veterinary Portable Multi-Parameter Patient Monitor. Operator s Manual

PM-9000Vet Veterinary Portable Multi-Parameter Patient Monitor. Operator s Manual PM-9000Vet Veterinary Portable Multi-Parameter Patient Monitor Operator s Manual CE Marking The product bears CE mark indicating its conformity with the provisions of the Council Directive 84/539/EEC

More information

SureSigns VS4. Vital Signs Monitor Technical Data Sheet

SureSigns VS4. Vital Signs Monitor Technical Data Sheet SureSigns VS4 Vital Signs Monitor Technical Data Sheet The SureSigns VS4 is a vital signs monitor that measures blood pressure, pulse rate, oxygen saturation (SpO 2 ), and temperature. The VS4 monitor

More information

ARGUS VCM Vital Compact Monitor

ARGUS VCM Vital Compact Monitor Vital Compact Monitor Art. No.: 2.540041 Rev. a *2.540041* Service Handbook Contents Figures Tables Notice This document contains proprietary information that is protected by copyright. All Rights Reserved.

More information