ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

Size: px
Start display at page:

Download "ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE"

Transcription

1 ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE Procedure Number: II a Effective Date: 09/14/99 Review and Update: 09/2003 TITLE: ADVIA 120 ROUTINE MAINTENANCE PURPOSE AND PRINCIPLE: The ADVIA 120 Hematology System is a fully automated diagnostic instrument with a throughput of 120 samples per hour (CBC/diff). The analyzer uses whole blood samples to provide Complete Blood Count, differential count, reticulocyte count and body fluid count. PROCEDURE: 1. Assure that instrument, computer, and printer are all on. 2. Perform START UP by following Starting each Shift procedure. Go to Help. Click Quick Guide then Starting each Shift. Follow the procedures for Emptying the waste container, Emptying the overflow bottle, Checking reagents, and Obtaining the background count. 2.1 Procedure for Emptying the waste container: a. Make sure that the analyzer is not sampling. b. Cancel all the Auto Cycles, Auto Wash, Auto Rinse, Auto Standby, and Auto Startup. c. Disconnect the level-sensor switch (1). d. On the waste removal assembly, turn the mode selector knob (2) from NORMAL to EMPTY. e. The waste in the container should start to empty. It will take between two and five minutes to completely empty the container. When you see air bubbles in the discharge line (3), the container is empty. II a Advia 120 Maintenance 9/2003 Page 1 of 9

2 f. Once the container is empty, reconnect the level-sensor switch (1) and turn the mode selector knob (2) back to the NORMAL setting. g. Reset any auto cycles that you cancelled in step Procedure for Emptying the Overflow bottle: Visually check the fluid level in the small, overflow bottle located to the right of the RBC sample and sheath pumps. If it has any liquid in it, empty the bottle, then clean the vent lines and chambers Procedure for Cleaning the Vent Lines and Chambers: Note: Clean one vent line or chamber at a time and analyzer mode should be Ready to Run. a. Remove the tubing from the overflow bottle that leads to the vent of the Pathway or chamber that you want to clean. Or, if there is no overflow tube installed, attach a 12-inch piece of tubing to the vent. b. Put the free end of the tube into a beaker of 25% solution of household bleach and water. EZ KLEEN MUST be used to clean the perox vacuum shuttle chamber and vent line! c. Open the valve that will create a vacuum to aspirate the bleach solution from the beaker through the vent line and chamber to the main waste line at the bottom of the UFC block.. To open a valve, on the Utilities menu, click the Exerciser tab. Click Valves from the list on the left. Click the appropriate valves. The appropriate valves are: Baso chamber open V13 Perox vacuum shuttle chamber open V18 and V17 (USE EZ KLEEN ONLY) Retic chamber open V65 RBC chamber open V10 RBC/baso/retic vacuum shuttle chamber open V22 and V21 Hemoglobin chamber open V36 d. Allow alternate segments of household bleach solution and air to run through the line and the chamber until they are clean. e. Immerse the free end of the tube into a beaker of deionized water to rinse the line and chamber. f. Remove the tube from the beaker and close the valve. g. Exit the Exerciser. h. Replace the tube into the overflow bottle, or if it was not an overflow tube, remove it from the vent opening Procedure for checking reagents: a. Use the Startup tab or the Reagent Log tab to check the supply of all reagents except ADVIA 120 DEFOAMER. Visually check the supply of ADVIA 120 Defoamer. b. If you need to replace reagents, use the Reagent Log tab. II a Advia 120 Maintenance 9/2003 Page 2 of 9

3 2.5. Procedure for Obtaining the background counts: Run a background count cycle to obtain a BASO WBC background count, a platelet background count, and an HGB Trans (Hemoglobin baseline Transmission). a. On the Routine Operations menu, click Startup. b. Click Refresh. c. The background results are color coded. Green - Within range Red - Out of range If any result is out of range, click Refresh on the Startup tab to run another background count cycle. If any result is still unacceptable, perform a system wash. 3. Perform any scheduled maintenance that is due. Go to System Log then click on Service Log. Click Scheduled. Perform all maintenance that is checked. When done click on Update Log. Weekly or after 2000 samples 3.1 Procedure for Washing the RBC/baso/retic flowcell: a. Prepare 25% solution of household bleach and water. b. On the Utilities menu, click te Hydraulics tab. Click Flowcell Wash and select RBC/baso/retic flowcell. c. Hold the 25% bleach solution under the open-tube sample probe and click Start in the window. d. Do not push the aspiration paddle. The vacuum will aspirate the solution. e. Hold the beaker until the wash block starts to move down, approximately 80 seconds. f. The flowcell cleaning will continue approximately 2.5 minutes longer. 3.2 Procedure for Washing the Perox flowcell: g. Prepare a beaker with EZ Kleen. h. On the Utilities menu, click te Hydraulics tab. Click Flowcell Wash and select Perox flowcell. i. Hold the beaker ofez Kleen under the open-tube sample probe and click Start in the window. j. Do not push the aspiration paddle. The vacuum will aspirate the solution. k. Hold the beaker until the wash block starts to move down, approximately 80 seconds. l. The flowcell cleaning will continue approximately 2.5 minutes longer. 3.3 Procedure for cleaning shear-valve and aspiration pathways in the UFC Anaylzer mode should be Ready to Run. a. Prepare a beaker of bleach and another beaker with water. b. On the Utilities menu, click the Exerciser tab. Click Syringe Pumps on the left hand side. Click on the image of the valve II a Advia 120 Maintenance 9/2003 Page 3 of 9

4 under Selector valve to Automated Closed then to Manual Closed then to Open. c. Click the valves button on the left. Click valve V72 to close. Closed valve Opened Valve d. Open valves V1, V47, and V74. e. Hold a beaker of household bleach under the open-tube sample probe until 5 ml is aspirated. f. Repeat step e using 5ml of water. g. Close valve V74 and click V72 to open and make sure that V73 is closed. h. Repeat steps e and f. i. Close V74 and V72, and open V73. j. Repeat steps e and f. k. Close valves V1, V47, and V73. l. Click on the Analyzer Status tab to exit. 3.4 Procedure for Cleaning the Shear Valve Analyzer mode should be in Standby. a. Prepare beaker with bleach and fill a wash bottle with DI water. b. Place paper towels directly under the shear valve to prevent any fluid from dripping down into the analyzer. c. Remove the knurled nut by turning counterclockwise. d. Remove the compression spring. e. Remove the rotor by holding the shear valve with one hand and with the other hand rotate the rotor until it can be pulled forward and off the shaft. f. Remove the front shear face by gently rotating the front face until loose, then pull forward and remove. DO NOT use sharp objects or excessive force to remove the shear face. g. Place the front shear face in the beaker of bleach for 10 minutes, then rinse thoroughly with DI water. h. Rinse the rear shear face by using the DI water bottle. Be sure to use paper towels to catch the dripping water. DO NOT wipe either of the shear faces with paper towels. II a Advia 120 Maintenance 9/2003 Page 4 of 9

5 i. Put the shear valve back together. Be sure to align the front shear valve with the smaller loops at 9 and 11 o clock positions and the large loop at 5 o clock. 3.4 Inspect the centering collars for salt buildup. Clean if necessary Procedure to clean the Autosampler centering collar a. Turn off the analyzer b. Tilt the front cover down. c. Remove the sample line from the bottom of the needle base. d. Loosen the two thumb screws and tilt the autosampler assembly forward. e. Pull the spring-loaded knob and turn a ¼. f. Remove the centering collar by pulling it up and out. g. Place the red needle cover over the needle. h. Remove the three lines from the collar. i. Place the collar in a beaker filled with 25% bleach and soak for 5 minutes. j. Remove any remaining residue using a cotton swab then rinse with DI water. k. Use a stylet or heavy fishing line to clean the three ports and the center bore of the collar. l. Attach a piece of tubing to a syringe, then flush each port with DI water. m. Reconnect the three tubes to the collar. n. Remove the needle cover and replace the collar. Be careful not to bend the needle. Turn the spring loaded knob back to its original position. Carefully pull on the collar to check the collar is put in correctly. If the collar can be removed check the spring loaded knob. o. Reposition the autosampler assembly and that it is seated correctly over the guide pins, tighten the thumb screws. p. Reconnect the sample line to the base of the collar. q. Close the analyzer cover and turn the power on Procedure for cleaning the Manual closed-tube centering collar. a. Turn off the analyzer b. Open the front cover. II a Advia 120 Maintenance 9/2003 Page 5 of 9

6 c. Pull the centering collar up. d. Remove the tubing from the collar e. Place the red needle cover over the needle. f. Place the centering collar in a beaker filled with 25% bleach solution for 5 minutes. g. Remove any remaining residue with a cotton swab and rinse with DI water. h. Use a stylet or heavy fishing line to clean the port and center bore of the collar. i. Attach a piece of tubing with a syringe to the port and flush with DI water. j. Reconnect the tubing to the collar. k. Remove the red needle cover and replace the collar, being careful not to touch the needle. l. Close the analyzer cover and turn the power on. 3.5 Inspect the perox-cap vent hole for buildup and clean if necessary. 3.6 Procedure for Shutting down the system a. On the Routine Operations menu, Log OFF. c. Click Shut down NT. Wait 1-2 minutes while the software shuts down. d. When the message appears It is now safe to turn off your computer, turn the analyzer off then switch the computer off. e. Leave the system shut down for 10 minutes f. Turn the computer on. g. When the computer displays the Begin Logon message, press Ctrl, Alt, and Delete at the same time to log on to Windows NT. h. Enter the operator name and password in the Logon Information and click OK. i. Click on Advia 120 j. Turn on analyzer. Monthly or after 8,000 samples 3.7 Procedure for inspecting and cleaning the syringes and plungers. a. Turn the analyzer off. b. Turn the thumb wheel counterclockwise until the plunger is lowered about ½ inch. c. Disconnect the tubing from the fittings, then disconnect the fittings from the syringe head. d. Remove the nut and washer at the top of the syringe and slide the syringe down and remove. e. If inspecting and cleaning the 50 ul syringe, remove the small plastic bushing inside the syringe by using a paper clip to pop the bushing out. f. Remove the plunger from the syringe. II a Advia 120 Maintenance 9/2003 Page 6 of 9

7 g. Inspect the syringe. Clean with 25% bleach solution and rinse with water if dirty. h. Inspect the plunger. Clean with a kim-wipe and 25% bleach solution and rinse with water if dirty. i. If the plunger appears to be worn or seems to be loose when replacing in the syringe, replace the plunger. (See procedure 3.10 ) j. Wet the plunger with saline and insert into the barrel of the syringe. k. Replace the bushing into the syringe (50-ul only). l. Insert the bottom end of the syringe into the slot at the bottom of the carriage. m. Slide the syringe up and into the mounting hole in the frame with the input port facing left on the RBC/baso/retic channel and facing right on the perox channel. n. Install the washer and the nut. o. Reconnect the tubing to the syringe fittings. 3.8 Procedure for replacing the autosampler and manual sampler needle. a. Turn off the analyzer b. Remove the centering collar. c. Place the needle cover over the needle firmly and turn counterclockwise to loosen. d. Discard the needle along with its cover. e. Clean the needle base using a cotton swab soaked in DI water. f. Remove the clear plastic end from the cover of a new needle. g. Screw the needle clockwise into the needle base. h. Remove the red cover from the needle and save for next installation. 3.9 Procedure for replacing the sheath filters Analyzer mode should be in Ready to Run Note: The filters are not interchangeable. The perox sheath filter is green and the RBC/baso filter is clear. a. Pull the perox sheath filter located on the left side toward you to remove it from its mounting. b. Disconnect the reagent line attached to the barbed fitting located at the top of the filter. c. Disconnect the luer fitting from the connector at the output port located at the bottom of the filter. DO NOT disconnect the reagent line leading from the luer fitting. d. Replace the filter. e. Attach the luer fitting to the connector located at the bottom of the filter first, then connect the reagent line to the barbed fitting at the top of the filter f. Insert the filter back into the mounting clip. g. Repeat the procedure for the RBC/baso sheath filter on the right side of the instrument. II a Advia 120 Maintenance 9/2003 Page 7 of 9

8 h. Prime reagent lines 5 times. Every two months or after 16,000 samples 3.10 Procedure for replacing the 50-ul syringe plunger a. Follow procedure 3.7 to remove the plunger. b. Loosen the set screws on the metal bushing on the plunger stem using a hex wrench. c. Remove the plunger and discard. Keep the metal bushing. d. Insert the pin gauge into the top side of the syringe. e. Wet the plunger with saline and insert into the syringe barrel until it touches the pin gauge. f. Replace the small plastic bushing into the syringe. g. Replace the metal bushing onto the plunger stem with the narrow collar of the bushing toward the plunger. h. Push the metal bushing up against the syringe and tighten the set screws. i. Remove the pin gauge from the syringe. j. Reinstall the sample syringe. Every 6 months 3.11 Procedure to replace the 1000-ul syringe plungers. Note: 1000-ul syringe plungers do not have plastic bushings a. Follow procedure Procedure for cleaning the air-circulation filter. Note: Facing the back of the instrument the air filter is located on the bottom right side. a. Slide the filter out of its frame. b. Use canned air to remove excess dust from the filter. c. Flush the filter with DI water. d. Allow it to air dry. e. Replace the filter 4. Run Quality Control materials. Follow Quality control procedure (procedure number II ). 5. At the end of each shift, perform Ending Each Shift procedure. Go to Help menu then click on Quick Guide then Ending Each Shift. Follow procedures for Cleaning the perox chamber, Washing the RBC/Baso/Retic flowcel, Washing the system, Performing End of Day procedure and Logging Off. 5.1 Procedure for Washing the system: a. On the Utilities menu, click Hydraulic Functions. b. Click System Wash, select 3 or 5 for the number of cycles, then click start. 5.2 Procedure for performing the End of Day: a. On Customize menu, click System Setup, then click Tools Modify. II a Advia 120 Maintenance 9/2003 Page 8 of 9

9 b. When the tools list appears, double-click End of Day. c. Select the SID Reset check box to perform a SID RESET. d. At the end of the month, select the QC Close Out check box to transfer the Daily QC file into Cumulative QC file. Use the list box to select close out for all controls or a specific control. e. Select the Purge Database check box. f. Note values in Total Samples and to be purged fields. In the to be purged field, type the number of record, if any, you want deleted from the All complete file. g. Click Ok to confirm and start. 5.3 Procedure for Logging Off: a. On the Routine Operations menu, click Log On/Off. b. Click Log Off. REAGENTS AND EQUIPMENT: 1. CBC Timepac (catalog no. T ) 2. Diff Timepac (catalog no. T ) 3. AutoRetic (600 test) (catalog no. T ) 4. DEFOAMER (catalog no. T ) 5. EZ KLEEN (catalog no. T ) 6. Sheath Rinse 20-L (catalog no. T ) 7. Bayer Advia 120 instrument. REFERENCES: 1. ADVIA 120 Hematology System Workbook, Customer Training Manual, and Operator s Guide. REVIEW DATES AND SIGNATURES: Chief Medical Officer: Associate Chief Medical Officer: II a Advia 120 Maintenance 9/2003 Page 9 of 9

FACSCalibur Startup Procedures

FACSCalibur Startup Procedures FACSCalibur Startup Procedures Always turn on the FACSCalibur before turning on the computer when acquiring data. This enables the computer to recognize that the cytometer is connected. When analyzing

More information

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days

Sanitise Syrup Lines & Valves. Taylor PH61 Cleaning every 14 days Taylor PH61 Cleaning every 14 days Drain the Syrup Lines Remove each syrup feed tube from syrup bottle and let excess syrup drain from the feed tube back into the syrup bottle. When flow of syrup from

More information

Administration of AMBISOME Using the GemStar Pump

Administration of AMBISOME Using the GemStar Pump 1 1-800-862-2731 Administration of AMBISOME Using the GemStar Pump Drug Name: Volume and Rate: Schedule: FLUSHING: Dextrose Ambisome Dextrose Heparin KEY POINTS: 1. Always wash your hands with an antibacterial

More information

PrimePour TM Operation Manual

PrimePour TM Operation Manual COT2000 SERIES PrimePour TM Operation Manual READ BEFORE USE. 840247001 7/14 For more Good Thinking visit www.commercial.hamiltonbeach.com 840247001_v18.indd 1 IMPORTANT SAFETY INSTRUCTIONS WARNING When

More information

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II)

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II) THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL E

MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL E MULTIWASH III MICROPLATE WASHER OPERATION AND MAINTENANCE MANUAL 8618-01 E DECLARATION OF CONFORMITY We, TriContinent Scientific, Inc. declare under our sole responsibility that the following products:

More information

MacDonald Steel Limited

MacDonald Steel Limited DELUXE FILLER / CROWNER MODEL - DFC4 1 REV 00 CAUTION / WARNING READ MANUAL BEFORE OPERATING MACHINE DO NOT EXCEED 50 PSI ON Co2 OR BEER LINES DO NOT EXCEED 100 F (35 C) ON Co2 OR BEER LINES DO NOT HOSE

More information

Operator s Manual. The FREEDOM ROCKER. Contents. Congratulations!

Operator s Manual. The FREEDOM ROCKER. Contents. Congratulations! Operator s Manual The FREEDOM ROCKER Congratulations! Congratulations on your purchase of a Next Advance Freedom Rocker! We re sure that once you use it you won t want to return to manual processing. Please

More information

I N ST R UC T I ON MODEL 2612 & 2712 SLICERS MODELS 2612 ML ML FORM (12-98) 2612 SLICER

I N ST R UC T I ON MODEL 2612 & 2712 SLICERS MODELS 2612 ML ML FORM (12-98) 2612 SLICER I N ST R UC 2612 SLICER T I ON S MODEL 2612 & 2712 SLICERS MODELS 2612 ML-104829 2712 ML-104822 701 S. RIDGE AVENUE TROY, OHIO 45374-0001 FORM 34141 (12-98) Installation, Operation, and Care of MODEL 2612

More information

CEREC MC XL Premium Package / MC XL / MC X / MC and inlab MC XL. Cleaning and Maintenance

CEREC MC XL Premium Package / MC XL / MC X / MC and inlab MC XL. Cleaning and Maintenance CEREC MC XL Premium Package / MC XL / MC X / MC and inlab MC XL Cleaning and Maintenance Contents 03 04 04 05 06 07 07 07 08 08 09 09 09 09 10 12 15 Maintenance Schedule Water System Changing the Filter

More information

Cleaning the internal optics surfaces

Cleaning the internal optics surfaces Cleaning the internal optics surfaces The face of the fiber-optic assembly and the lens are directly exposed to sample fluids inside the analyzer s flow cell. Automatic rinse cycles use high-pressure rinse

More information

Pet. Easy Guide. VAX Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire, WR9 OQH, UK - website:

Pet. Easy Guide. VAX Ltd., Kingswood Road, Hampton Lovett, Droitwich, Worcestershire, WR9 OQH, UK   - website: V-027P Rapide Pet User Guide 10/7/07 10:05 Page 14 EEC STATEMENT OF COMPLIANCE Manufacturer/EEC importer: VAX Limited, hereby on our own responsibility declare that the products V-027/V-027PT/V-027P/CCW-701

More information

Symphony Quick Start Manual Document: Revision: A Distek Inc. Quick Start Manual

Symphony Quick Start Manual Document: Revision: A Distek Inc. Quick Start Manual Symphony Quick Start Manual Document: 3871-0002 Revision: A Quick Start Manual Height Adjustment 2 1 Height Adjustment The apparatus height is set and configured for baskets and paddles purchased at Distek,

More information

Standard Operating Manual

Standard Operating Manual Standard Operating Manual EVG 101 Spray Coater Version 1.1 Page 1 of 24 Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size 2.3 Spin speed 2.4 Spin ramp

More information

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title. THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

Agilent 6400 Series. Maintenance Guide. Agilent Technologies

Agilent 6400 Series. Maintenance Guide. Agilent Technologies Agilent 6400 Series Triple Quad LC/MS Maintenance Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic

More information

CBT LW MAINTENANCE GUIDE

CBT LW MAINTENANCE GUIDE CBT LW MAINTENANCE GUIDE PICTOGRAMS Each Signifier displayed here is specific to this User Manual. Menu Previous Advance Note Tip Example Powder Feeder Mixing Bowl Weigh Scale CBP Tanks Control Panel PSD

More information

Nilfisk Inc Winnetka Avenue North Minneapolis, MN REV.03( ) VF80189

Nilfisk Inc Winnetka Avenue North Minneapolis, MN REV.03( ) VF80189 Nilfisk Inc. 9435 Winnetka Avenue North Minneapolis, MN 55445 www.usviper.com REV.03(05-) VF8089 SAFETY PRECAUTIONS This machine is intended for commercial use. It is constructed for use in an indoor

More information

Model # VWK-2 INSTRUCTION MANUAL

Model # VWK-2 INSTRUCTION MANUAL Model # VWK-2 INSTRUCTION MANUAL Ver. March 2015 ALA Scientific Instruments Inc. 60 Marine Street Farmingdale, NY 11735 Tel. # 631.393.6401 Fax: # 631.393.6407 E-mail: support@alascience.com www.alascience.com

More information

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011 QUICK REFERENCE Ion OneTouch Template Kit Publication Part Number 4468008 Revision E Date 31 October 2011 Note: For safety and biohazard guidelines, refer to the Safety section in the Ion OneTouch Template

More information

INSTALLATION. Glass Panel Doors (select models) CAUTION

INSTALLATION. Glass Panel Doors (select models) CAUTION Location Do not install refrigerator near oven, radiator or other heat source. If not possible, shield refrigerator with cabinet material. Do not install where temperature falls below 55 F (13 C) or rises

More information

OPERATION and MAINTENANCE MANUAL

OPERATION and MAINTENANCE MANUAL MULTIWASH III MICROPLATE WASHER OPERATION and MAINTENANCE MANUAL P/N 8618-01 B DECLARATION OF CONFORMITY We, TriContinent Scientific, Inc. declare under our sole responsibility that the following products:

More information

Enteral Pump Instructions for Use To Assemble When the low

Enteral Pump Instructions for Use To Assemble When the low Enteral Pump Instructions for Use Before initial use on battery power and after extended storage periods, the pump must be plugged into an AC power source for a minimum of 12 hours (pump may be operated

More information

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see  for demo video Keeping you insafehands to avoid needlestick injuries Astek Innovations Ltd Astek House I Atlantic Street I Altrincham Cheshire I WA14 5DH I United Kingdom Telephone +44 (0)161 942 3900 I Fax +44 (0)161

More information

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved

BRAVO AUTOCLAVES. Installation Notes. BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved BRAVO AUTOCLAVES Installation Notes BRAVO 1 Quick Start Guide SD-402 Rev. 1.0 Copyright 2012 SciCan Ltd. All rights reserved For questions, technical support, service and inquires, contact: Manufactured

More information

918 Clock Gauge Installation, Operation, and Maintenance Instructions

918 Clock Gauge Installation, Operation, and Maintenance Instructions 98 Clock Gauge Installation, Operation, and Maintenance Instructions The 98 Clock Gauge is designed to measure liquid level in an aboveground storage tank. The gauge mounts on top of the tank and is activated

More information

ICED CAPPUCCINO MACHINE Taylor MODEL: 390

ICED CAPPUCCINO MACHINE Taylor MODEL: 390 4.8.6.3 ICED CAPPUCCINO MACHINE Taylor MODEL: 390 LOCATION: Front of House WHEN CLEANED: Daily, Weekly, Monthly, Seasonal TOOLS/SUPPLIES REQUIRED: Salmon Cloth 2 or 3-Compartment Sink Sink Detergent Sanitizer

More information

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A CLEAN MIST ULTRASONIC HUMIDIFIER Instruction Manual and Warranty Information IM0036A READ AND SAVE THESE INSTRUCTIONS Contents Safety Instructions... 1 Unpacking / Specifications... 2 Parts & Contents...

More information

SINGLE SPEED SYRINGE PUMP

SINGLE SPEED SYRINGE PUMP instrumentation and software for research SINGLE SPEED SYRINGE PUMP PHM-100 USERS MANUAL DOC-083 Rev. 1.6 Copyright 2014 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone:

More information

Operation and Maintenance Manual BLIZZARD50 CS6-50-VD 120V/60HZ

Operation and Maintenance Manual BLIZZARD50 CS6-50-VD 120V/60HZ Operation and Maintenance Manual BLIZZARD50 CS6-50-VD 120V/60HZ CS6-50-VD Operation & Maintenance Manual Table of Contents: 1.0 Introduction 1 2.0 Unpacking your COOL- SPACE 1 3.0 Set-up of COOL-SPACE

More information

SOP: LFA SOP For Automatic Capsule Filling Machine

SOP: LFA SOP For Automatic Capsule Filling Machine SOP For Automatic Capsule Filling Machine Alastair Sanderson 31/08/16 1 Objective To provide the operating procedure for Automatic Capsule Filling Machine. 2 Scope Includes the operating procedure for

More information

5800 Refrigerator Compressor Capacitor/Relay Kit

5800 Refrigerator Compressor Capacitor/Relay Kit 5800 Refrigerator Compressor Capacitor/Relay Kit Instruction Sheet 60-4702-009 Revision B, January 25, 2016 Overview The refrigerator compressor relay and capacitor are components of the Teldyne Isco 4700

More information

I N ST R UC T I ON. MODEL HAB500 AroMatic TM BREAD SLICERS MODEL FORM (4-99) AROMATIC BREAD SLICER 701 S. RIDGE AVENUE TROY, OHIO

I N ST R UC T I ON. MODEL HAB500 AroMatic TM BREAD SLICERS MODEL FORM (4-99) AROMATIC BREAD SLICER 701 S. RIDGE AVENUE TROY, OHIO I N ST R UC AROMATIC BREAD SLICER T I ON S MODEL HAB500 AroMatic TM BREAD SLICERS MODEL HAB500 ML-104845 701 S. RIDGE AVENUE TROY, OHIO 45374-0001 FORM 34198 (4-99) TABLE OF CONTENTS GENERAL..........................................................

More information

ULTRASONIC HUMIDIFIER WITH ANTIBACTERIAL TANK & BASE Instruction Manual and Warranty Information IM0012A READ AND SAVE THESE INSTRUCTIONS

ULTRASONIC HUMIDIFIER WITH ANTIBACTERIAL TANK & BASE Instruction Manual and Warranty Information IM0012A READ AND SAVE THESE INSTRUCTIONS ULTRASONIC HUMIDIFIER WITH ANTIBACTERIAL TANK & BASE Instruction Manual and Warranty Information IM0012A READ AND SAVE THESE INSTRUCTIONS Contents Safety Instructions... 1 Specifications / Parts & Contents...

More information

FEI Nova 200 Dual Beam system maintenance

FEI Nova 200 Dual Beam system maintenance David Haviland A 1(6) FEI Nova 200 Dual Beam system maintenance 1 General This document describes regular maintenance of the FEI Nova 200 Dual Beam system at Albanova Nanofabrication Facility. All work

More information

TECHNICAL INFORMATION T 15xx Dryers

TECHNICAL INFORMATION T 15xx Dryers TECHNICAL INFORMATION T 15xx Dryers 2010 Miele USA Table of Contents 1.0 2.0 3.0 T 15xx Dryers Construction and Design... 5 1.1 Appliance Overview Vented Models... 5 1.2 Appliance Overview Condenser Models...

More information

User information. Cleaning/descaling and sanitizing

User information. Cleaning/descaling and sanitizing User information How the dispenser works Follett s 110 series automatic-load ice and water dispensers are equipped with Follett s 425 lb (193 kg)/day ice machine. In the continuous icemaking process, water

More information

MODEL MC-UL592 MC-UL594 POWER SOURCE. 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H)

MODEL MC-UL592 MC-UL594 POWER SOURCE. 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H) Order Number: PMMA091040CE Vacuum Cleaner MC-UL592/MC-UL594 SPECIFICATION MODEL MC-UL592 MC-UL594 POWER SOURCE 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H) 1200-1300

More information

Multi-Mode Electrode pro

Multi-Mode Electrode pro Multi-Mode Electrode pro Manual 8.110.8018EN / 2015-06-30 Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com Multi-Mode Electrode pro 6.1246.120

More information

A Issue 3. USA Distributor:

A Issue 3. USA Distributor: 2 USA Distributor: 3 4 5 6 Table of Contents COMPANY INFORMATION...3 Contact Addresses...3 SAFETY INFORMATION...4 GENERAL SAFETY...4 SYMBOLS...5 CHEMICAL SAFETY...6 ENVIRONMENT...6 WARRANTY STATEMENT...6

More information

CONSTANT SUCTION UNIT

CONSTANT SUCTION UNIT CONSTANT SUCTION UNIT OPERATOR/MAINTENANCE M A N U A L Squire-Cogswell/Aeros Instruments Inc 1111 Lakeside Drive Gurnee, IL 60031-4099 1-800-662-5822/847-855-0800 Fax: 847-855-6218 www.aerosinstruments.com

More information

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL S-SCORT II Model 15006 SSCOR, INC. 11064 Randall Street Sun Valley, CA 91352 USA Telephone +1-818-504-4054 800-434-5211 Fax +1-818-504-6032 www.sscor.com Email:

More information

Operating Manual for. Ultrawash Plus. Absorbance. Plate Washing. Luminescence. Automated Processing. Software

Operating Manual for. Ultrawash Plus. Absorbance. Plate Washing. Luminescence. Automated Processing. Software Absorbance Plate Washing Luminescence Automated Processing Operating Manual for Ultrawash Plus Software DYNEX Technologies, Inc. 14340 Sullyfield Circle, Chantilly, VA 20151-1683 Phone: (703) 631-7800

More information

Bravo Display Warning Messages

Bravo Display Warning Messages Bravo Display Warning Messages The sterilizer periodically reminds the user about recommended routine maintenance operations that must be carried out in order to ensure the proper operation of the device.

More information

5 Maintenance. A November

5 Maintenance. A November Maintenance 5 Maintenance Contents Cleaning frequency chart...5-2 Cleaning tools and materials...5-2 Opening the scanner cover...5-2 Cleaning procedures...5-3 Vacuuming the output tray and input elevator...5-4

More information

ACQUITY UPLC H-Class Sample Manager - Flow Through Needle Operator s Overview and Maintenance Information

ACQUITY UPLC H-Class Sample Manager - Flow Through Needle Operator s Overview and Maintenance Information ACQUITY UPLC H-Class Sample Manager - Flow Through Needle Operator s Overview and Maintenance Information Revision B Copyright Waters Corporation 2010 All rights reserved Copyright notice 2010 WATERS CORPORATION.

More information

Altra Series Dampener

Altra Series Dampener Crestline TM Altra Series Dampener Installation Instructions Heidelberg MO X88-66 10/97 Rev-A GENERAL INFORMATION ATTENTION CRESTLINE ALTRA SERIES TM DAMPENER OWNER! Accel Graphic Systems provides parts

More information

Explorer CleaningPartsDirect.com Carpet Extractor. Operator and Parts Manual. Home Find... Go To.. Model No.: Can.

Explorer CleaningPartsDirect.com Carpet Extractor. Operator and Parts Manual. Home Find... Go To.. Model No.: Can. Explorer 1500 Carpet Extractor Model No.: 608808 609231 Can. Operator and Parts Manual CleaningPartsDirect.com 662-393-3045 NOBLES 12875 RANSOM STREET HOLLAND MI 49424 U.S.A. CUSTOMER SERVICE: 1-800-365-6625

More information

BEFORE OPERATING THE MACHINE: WARNING

BEFORE OPERATING THE MACHINE: WARNING BEFORE OPERATING THE MACHINE: Read the manual carefully and completely before attempting to operate the unit. This manual has important information for the use and safe operation of the machine. Keep this

More information

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU-20B CompriCart TABLE OF CONTENTS: Introduction.........................1 Package Contents....................1 Setting Up the Unit...................2 -

More information

Installation. Leveling

Installation. Leveling Your refrigerator was packed carefully for shipment. Remove and discard shelf packaging and tape. Do not remove the serial plate. Location Do not install refrigerator near oven, radiator or other heat

More information

University of Minnesota Nano Fabrication Center

University of Minnesota Nano Fabrication Center Equipment name: Spray/Puddle Developer Badger name: develop spinner CEE-4 Model: 200X Location: Bay 2 Revisionist: Paul Kimani Revision #: 1 Date: April 26, 2013 Description The develop spinner CEE-4 spinner

More information

THERMOSTAT RETROFIT INSTRUCTIONS

THERMOSTAT RETROFIT INSTRUCTIONS With these instructions: You will remove the two existing thermostats from the dishmachine and will replace with a single new solid-state thermostat. TOOLS REQUIRED: Adjustable Wrench Philips Screwdriver

More information

Using the Lely Robot Sampler CanWest DHI Robot Manual

Using the Lely Robot Sampler CanWest DHI Robot Manual Using the Lely Robot Sampler 4 Lely Check List First time start-up settings done? Is sampler level? Is milk line connected through sampling valve? Is air pressure to the shuttle correct? Are there enough

More information

Page 1 of 18. Part# /5/2013

Page 1 of 18. Part# /5/2013 Part# 1002655-06 8/5/2013 This manual contains important information concerning the installation and operation of the gun washers listed above. Read manual thoroughly and keep for future reference INSTRUCTIONS

More information

TECHNICAL INFORMATION Touchtronic Clothes Dryers

TECHNICAL INFORMATION Touchtronic Clothes Dryers TECHNICAL INFORMATION Touchtronic Clothes Dryers Includes: T1302, T1303, T1322, T1329ci T1403 & T1405 2004 Miele This page intentionally left blank. Table of Contents GENERAL INFORMATION A. Warning and

More information

PRO 2000i LOW LEVEL DEPOSITOR

PRO 2000i LOW LEVEL DEPOSITOR DEPOSITORS AND AUTOMATED CAKE PRODUCTION SYSTEMS PRO 2000i LOW LEVEL DEPOSITOR OPERATION AND SPARE PARTS MANUAL Serial No. PR2L- (Please quote this number when ordering spares, and making service calls)

More information

Clean Water Made Easy. CWS Time Clock Softener Installation & Start Up Guide. Questions?

Clean Water Made Easy. CWS Time Clock Softener Installation & Start Up Guide. Questions? Clean Water Made Easy www.cleanwaterstore.com CWS Time Clock Softener Installation & Start Up Guide Thank you for purchasing a Clean Water System! With proper installation and a little routine maintenance

More information

FILLING & CLEANING GUIDE

FILLING & CLEANING GUIDE FILLING & CLEANING GUIDE 2017 VENDORS EXCHANGE INTERNATIONAL, INC. Dear Valued Customer, Thank you for purchasing the Wittenborg 9100. This manual outlines filling and cleaning steps to keep your machine

More information

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description.

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description. Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit Operating Instructions 1 Product Description The Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip

More information

CyTOF Autosampler 5 User Manual

CyTOF Autosampler 5 User Manual CyTOF Autosampler 5 User Manual CyTOF Autosampler 5 User Manual This is Class I, cord-connected equipment. For use in commercial, industrial or business enviroments. For Research Use Only. Not for use

More information

BAXTER: FACTOR VII 600 IU (Plasma Derived)

BAXTER: FACTOR VII 600 IU (Plasma Derived) BAXTER: FACTOR VII 600 IU (Plasma Derived) STEP 1 Bring recombinant factor VII to room temperature before using. Choose a clean, dry, well-lit work area and lay out the following supplies Contained within

More information

Falex Thermal Fowling Tester (FT2)

Falex Thermal Fowling Tester (FT2) 450 Falex Thermal Fowling Tester (FT2) * SAE Unit * OPERATION & MAINTENANCE MANUAL Version 1.1 Falex Corporation 1020 Airpark Drive Sugar Grove, IL 60554 www.falex.com Phone: (630) 556-3669 FT2 (F450)

More information

How to Fix a Leaky Toilet:

How to Fix a Leaky Toilet: How to Fix a Leaky Toilet: First, determine the type of toilet you have. Pressurized Toilet: Inside this toilet is a sealed tank. When water is fed from the water line, the air inside the tank is compressed.

More information

Home Medical Equipment Oxygen

Home Medical Equipment Oxygen Home Medical Equipment Oxygen OhioHealth Home Medical Equipment 7708 Green Meadows Drive, Suite D Lewis Center, Ohio 43035 614-566-0850 or 1-800-300-7075 OXYGEN CONCENTRATOR 1. The On/Off switch is on

More information

Owner's Manual. WS Series. Water Softener

Owner's Manual. WS Series. Water Softener Owner's Manual WS-165-150 Series Water Softener Table of Contents WHAT'S INCLUDED 3 OPERATING CONDITIONS 4 ASSEMBLY INSTRUCTIONS 6 FLUSHING THE WATER LINES 13 MASTERPROGRAMMING 14 PROGRAMMING KEYAND GENERAL

More information

VARIABLE RATE INFUSION PUMP

VARIABLE RATE INFUSION PUMP instrumentation and software for research VARIABLE RATE INFUSION PUMP PHM-100VS-AC USER S MANUAL DOC-302 Rev. 1.0 Copyright 2016 All Rights Reserved P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343

More information

Consumable Replacement and Cleaning Instructions

Consumable Replacement and Cleaning Instructions Consumable Replacement and Cleaning Instructions 1 Consumables and Replacement Cycles To maintain high-quality scanned images, it s important to clean your scanner regularly and replace the consumables

More information

UNITY 2 TM. Air Server Series 2 Installation Manual. Version September 2012

UNITY 2 TM. Air Server Series 2 Installation Manual. Version September 2012 UNITY 2 TM Air Server Series 2 Installation Manual Version 1.12 September 2012 1. Installation...2 1.1. Tools required...2 1.2. Procedure...2 1.3. Requirement for a sample pump (U-ASPMP1 / U-ASPMP2)...8

More information

930 Compact IC Flex. 930 Compact IC Flex Oven SeS/PP/Deg. Manual Short Instructions EN /

930 Compact IC Flex. 930 Compact IC Flex Oven SeS/PP/Deg. Manual Short Instructions EN / 930 Compact IC Flex 930 Compact IC Flex Oven SeS/PP/Deg Manual Short Instructions 8.930.8120EN / 2017-07-31 Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com

More information

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX

STANDARDD OPERATING PROCEDURE FOR THE LUMINEX The Flow Cytometry Facility STANDARDD OPERATING PROCEDURE FOR THE LUMINEX All users must follow the SOP & fill in the logbook. If you are unsure about anything please do not hesitate to ask for help. Before

More information

Replacement Guide: TCRU/ORU Model Name: Pro C901S/Pro C901

Replacement Guide: TCRU/ORU Model Name: Pro C901S/Pro C901 Replacement Guide: TCRU/ORU Model Name: Pro C90S/Pro C90 Read this manual carefully before using this machine and keep it handy for future reference. How to Read This Manual Introduction This manual contains

More information

8 lb Lightweight Vacuum Model CK LW 13/1

8 lb Lightweight Vacuum Model CK LW 13/1 TORNADO INDUSTRIES 7401 W. LAWRENCE AVENUE CHICAGO, IL 60706 (708) 867-5100 FAX (708) 867-6968 www.tornadovac.com Operations & Maintenance Manual For Commercial Use Only 8 lb Lightweight Vacuum Model CK

More information

52 DorsetTM. Instruction Manual. Basic Function Wall Control System Included. A Kichler Decor ceiling fan

52 DorsetTM. Instruction Manual. Basic Function Wall Control System Included. A Kichler Decor ceiling fan Basic Function Wall Control System Included 52 DorsetTM II A Kichler Decor ceiling fan Kichler Lighting 7711 East Pleasant Valley Road P.O. Box 318010 Cleveland, Ohio 44131-8010 Customer Service 866.558.5706

More information

Dacor Technical Service

Dacor Technical Service Attention: This manual is just a section from the complete Wall Oven Service Manual. If you find that you require the complete service manual, which includes exploded views and parts, use and care information

More information

Eldex Column Heater. Operator s Manual

Eldex Column Heater. Operator s Manual Eldex Eldex Column Heater Operator s Manual Eldex Laboratories, Inc. 30 Executive Court Napa, CA 94558 Tel: (707) 224-8800 Fax: (707) 224-0688 www.eldex.com Rev. B: 081700 2000Eldex Laboratories, Inc.

More information

Installation Instructions

Installation Instructions Aluminum Blinds Installation Instructions Echelon, Traditions & Integra Aluminum Blinds 2 Metro Aluminum Blinds *Click on any page to return to the Table of Contents* Echelon, Traditions & Integra Aluminum

More information

Caution PROTECT YOUR FACTORY WARRANTY!

Caution PROTECT YOUR FACTORY WARRANTY! 6 Cleaning Caution PROTECT YOUR FACTORY WARRANTY! The recommended cleaning procedures must be performed to maintain your factory warranty. Allow only Zebra authorized technicians to service the Printer

More information

MINIMED 670G SYSTEM CHANGING THE MINIMED MIO INFUSION SET QUICK REFERENCE GUIDE

MINIMED 670G SYSTEM CHANGING THE MINIMED MIO INFUSION SET QUICK REFERENCE GUIDE START HERE Wash your hands. Select Reservoir & Tubing. Press. Select Options. Select New Reservoir. Remove the infusion set you have been using by loosening the adhesive and pulling away from body. Remove

More information

kéìíê~=iqd=qìåöëíéå=dêáåçéê= lééê~íáçåë=j~åì~ä=

kéìíê~=iqd=qìåöëíéå=dêáåçéê= lééê~íáçåë=j~åì~ä= NJUMMJVQQJOOQP ïïïk~êåjòçåékåçã kéìíê~iqdlééê~íáçåëj~åì~ä m^dbnlcnm kéìíê~iqdqìåöëíéådêáåçéê lééê~íáçåëj~åì~ä ^o`jwlkbk`lj IfåÅ. OMVNi~ëm~äã~ëaêáîÉpíÉK` `~êäëä~çi`^vomnnjnrrn qéäwtsmkvpnknrmm UMMKVQQKOOQP

More information

FD 430 Envelope Sealer

FD 430 Envelope Sealer FD 430 Envelope Sealer 1/2018 OPERATOR, MAINTENANCE, & PARTS MANUAL REV. 1 2 CONTENTS 1 SAFETY INSTRUCTIONS... 5 1.1 Symbols and reference key... 5 1.2 Basic safety precautions... 6 1.3 Safety advice...

More information

Zoom HT Microplate Washer. The Fastest Microplate Washer For The Life Sciences

Zoom HT Microplate Washer. The Fastest Microplate Washer For The Life Sciences Zoom HT Microplate Washer The Fastest Microplate Washer For The Life Sciences Advanced Tools for The Life Sciences www.titertek-berthold.com Zoom HT Microplate Washer Zoom HT is fast It takes only 17 seconds

More information

Standard Operation Procedures - Inferno 4500 Daily Startup

Standard Operation Procedures - Inferno 4500 Daily Startup Daily Startup 1. Turn on the ventilation hoods and lights in the morning. 2. Plug the grease reservoir into the outlet on the side of the oven marked ALWAYS ON. 3. Position the hand lever from CLEAN to

More information

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro 2006/01 Content Important Information 1. Notes... 3 1.1 CE - Labeling... 3 1.2

More information

Printer Startup and Demo Build The purpose of this section is to execute the following:

Printer Startup and Demo Build The purpose of this section is to execute the following: Printer Startup and Demo Build The purpose of this section is to execute the following: Power the printer on. Installing new material cartridge. Removing material cartridges. Procedure to empty waste bag.

More information

ALLERGEN MIXING ASSISTANT

ALLERGEN MIXING ASSISTANT ALLERGEN MIXING ASSISTANT User Manual 9495 SW Locust Street Suite E Portland, OR 97223 (503) 379-0110 Fax: (503) 715-1378 www.xtractsolutions.com SM026R2 Welcome to Xtract Solutions Thank you for purchasing

More information

STOP! Read this notice. STOP! Read before proceeding. These instructions are for gas conversion of the Evergreen boiler only.

STOP! Read this notice. STOP! Read before proceeding. These instructions are for gas conversion of the Evergreen boiler only. Series Identification Read the boiler rating label to determine the size and series number. CONDENSING GAS BOILER 220/299/300/399 Gas Conversion Instruction for Evergreen 220, 299/300 and 399 boilers Natural

More information

52 Lacey LED. Instruction Manual. 6 Speed DC Wall Control System

52 Lacey LED. Instruction Manual. 6 Speed DC Wall Control System 6 Speed DC Wall Control System 52 Lacey LED HIGH EFFICIENCY DC MOTOR Kichler Lighting 7711 East Pleasant Valley Road P.O. Box 318010 Cleveland, Ohio 44131-8010 Customer Service 866.558.5706 8:30 AM to

More information

Heat Exchanger Block Replacement Instructions

Heat Exchanger Block Replacement Instructions Series 1-4 Gas-fired water boiler Heat Exchanger Block Replacement Instructions Ultra-80 S1-4 Heat Exchanger Block Replacement Kit, Part No. 383-500-773 Ultra-105 S1-4 Heat Exchanger Block Replacement

More information

CEILING FAN OWNER'S MANUAL

CEILING FAN OWNER'S MANUAL Style that revolves around you. CEILING FAN OWNER'S MANUAL Vail with DC motor 10/15 WARNING: Read and follow these instructions carefully and be mindful of all warnings shown throughout. GENERAL INSTALLATION

More information

Designed in the UK. Rapide Ultimate. User Guide Vax Rapide Ultimate

Designed in the UK. Rapide Ultimate. User Guide Vax Rapide Ultimate Designed in the UK Rapide Ultimate User Guide Vax Rapide Ultimate VCWRU Contents Contents 1 Machine Overview 2 General Safety Information 3 Getting Started 5 How To Use 6 Looking After Your Machine 9 FAQs

More information

Replacing All Seals. This procedure explains how to replace all seals in the pump head assembly of the Alliance 2695 and 2795 Separations Modules.

Replacing All Seals. This procedure explains how to replace all seals in the pump head assembly of the Alliance 2695 and 2795 Separations Modules. SSPR10023721 - NEW DOC. NO. REV. 01 1 OF 5 This procedure explains how to replace all seals in the pump head assembly of the Alliance 2695 and 2795 Separations Modules. When to Perform Replace all seals:

More information

JOHN DEERE GATOR HPX/XUV 2 PASSENGER HEATER INSTALLATION INSTRUCTIONS (p/n: 9PH20S30)

JOHN DEERE GATOR HPX/XUV 2 PASSENGER HEATER INSTALLATION INSTRUCTIONS (p/n: 9PH20S30) P. 1 of 12 JOHN DEERE GATOR HPX/XUV 2 PASSENGER HEATER INSTALLATION INSTRUCTIONS (p/n: 9PH20S30) Item: Qty: Description: 1 2 1 x 1 x 5/8 Tee Fitting 2 2 Plastic Snap-in Hose Grommet 3 4 1-1/2" Hose Clamps

More information

541D19 SERIES. Technical Manual. A Division of Aquion Partners L.P.

541D19 SERIES. Technical Manual. A Division of Aquion Partners L.P. 541D19 SERIES Technical Manual A Division of Aquion Partners L.P. Table of Contents Introduction... Page 1 Technical Specifications... Page 2 Flow Diagrams... Page 3 Injector & Flow Control Selection Injector...

More information

Model UVGMD-AL MULTIVERSA CAN PACKAGING MACHINE Operators Manual

Model UVGMD-AL MULTIVERSA CAN PACKAGING MACHINE Operators Manual Model UVGMD-AL MULTIVERSA CAN PACKAGING MACHINE Operators Manual MODEL UVGMD-AL INTRODUCTION Model UVGMD-AL offers your choice of atmospheric, vacuum only, vacuum then gas, or multiflush double seaming.

More information

MINIMED 670G SYSTEM CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE

MINIMED 670G SYSTEM CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE START HERE Wash your hands. Press. Select Options. Select Reservoir & Tubing. Select New Reservoir. Remove the infusion set you have been using by loosening the adhesive and pulling away from body. 6 Remove

More information

One Less Thing to Forget...

One Less Thing to Forget... Installation Guide One Less Thing to Forget... Sign up now for cartridge auto-renew and save 10%! Stop stressing about when to purchase your next nuvoh2o replacement cartridge. Sign up for our auto-renew

More information

Crestline Dampening System

Crestline Dampening System Crestline Dampening System Installation Instructions Hamada 500, 600, 700, E-47, SU47 Series Parent Press X88-25 01/2001 Rev-C 2593 GENERAL INFORMATION ATTENTION CRESTLINE DAMPENER OWNER! Accel Graphic

More information

TC-710 RECYCLING BALER

TC-710 RECYCLING BALER OPERATING AND MAINTENANCE INSTRUCTIONS TC-710 RECYCLING BALER READ INSTRUCTIONS THOROUGHLY BEFORE OPERATING 3451 S. 40th Street Phoenix, AZ 85040 602.437.5020 800.223.4540 www.tsissg.com info@tsissg.com

More information

Model Series. PumpAgents.com - buy pumps and parts online QUIET - FLUSH ELECTRIC TOILET. Model Series FEATURES SPECIFICATIONS INSTALLATION

Model Series. PumpAgents.com - buy pumps and parts online QUIET - FLUSH ELECTRIC TOILET. Model Series FEATURES SPECIFICATIONS INSTALLATION PumpAgents.com - Click here for Pricing/Ordering Model 37045-Series QUIET - FLUSH ELECTRIC TOILET FEATURES Very quiet flush cycle - like a household toilet Single button flush actuator - with dual function

More information