PENTAX VIDEO DUODENOSCOPE ED-3490TK

Size: px
Start display at page:

Download "PENTAX VIDEO DUODENOSCOPE ED-3490TK"

Transcription

1 Only For the Americas INSTRUCTIONS FOR USE (REPROCESSING) PENTAX VIDEO DUODENOSCOPE ED-3490TK This instructions for use describes the recommended procedures for the reprocessing and maintenance of the duodenoscope after its use. For the inspection and directions for use, please refer to the separate operation instructions for use.

2 Intended Use The ED-3490TK, Video Duodenoscope is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract and Biliary Tract. The Upper Gastrointestinal Tract and Biliary Tract includes but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Common Bile Duct, Hepatic Duct and Cystic Duct. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations. Sterility Statement The endoscope identified in this IFU is a reusable semicritical device. Since the endoscope is packaged non-sterile, it must be cleaned and high-level disinfected or sterilized BEFORE initial use. Prior to each subsequent procedure, it must be subjected to appropriate cleaning and either high-level disinfection or sterilization processes. Contraindication Please consult regional and national health authority recommendations and requirements regarding protocols to follow in order to reprocess and/or destroy endoscopes that will be used or have been determined to have been used (post procedure) on patients afflicted with Creutzfeldt-Jacob Disease (CJD or vcjd). Notes Read this IFU before operating, and save this book for future reference. Failure to read and thoroughly understand the information presented in this IFU, as well as those developed for ancillary endoscopic equipment and accessories, may result in serious injury including infection by cross contamination of the patient and/or user. Furthermore, failure to follow the instructions in this IFU and the companion IFU may result in damage to, and/or malfunction of, this equipment. This IFU describes the recommended procedures for the reprocessing and maintenance of the equipment after its use. It is the responsibility of each medical facility to ensure that only well-educated and appropriately trained personnel, who are competent and knowledgeable about the endoscopic equipment, antimicrobial agents/processes, and hospital infection control protocol are involved in the reprocessing of these medical devices. Current infection control guidelines require that G.I. endoscopes and other semi-critical medical devices that normally come into contact with intact mucus membranes, such as in the gastrointestinal tract, must at least be cleaned and high-level disinfected before patient use. Only the user can determine if an instrument has undergone appropriate infection control procedures prior to each clinical use. It must be recognized that infection control practices involve many complex and often controversial issues which are constantly evolving. PENTAX strongly recommends that user remain informed of the latest federal and local regulations, and it encourages users to follow infection control guidelines developed by various organizations for health care professionals. If you have any questions regarding the information in this IFU or concerns pertaining to the safety and/or use of this equipment, please contact your local PENTAX representative. Conventions Throughout this IFU, the following conventions will be used to indicate a potentially hazardous situation which, if not avoided; WARNING : could result in serious adverse reactions and potential safety hazards. CAUTION : may result in adverse reactions for safe and effective use of the device. NOTE : may result in property-damage. Also advises owner/operator about important information on the use of this equipment. Prescription Statement Federal (U.S.A) law restricts this device to sale by or on the order of a physician or other appropriately licensed medical professional.

3 Symbols on Marking Symboles distinctifs Symbol for MANUFACTURER Symbol for DATE OF MANUFACTURE Attention, consult instructions for use Attention, consulter le manuel d utilisation Type BF applied part (Safety degree specified by IEC 6060-) Partie appliquée de type BF (niveau de sécurité spécifié par la norme CEI 6060-) Symbol for Authorised Representative in the European Union この CE マーキングは EC 指令への適合宣言マークです The CE marking assures that this product complies with the requirements of the EC directive for safety. Das CE Zeichen garantiert, daß dieses Produkt die in der EU erforderlichen Sicherheitsbestimmungen erfüllt. Le logo CE certifie que ce produit est conforme aux normes de sécurité prévues par la Communauté Européenne. II marchio CE assicura che questo prodotto è conforme alle direttive CE relative alla sicurezza. La marca CE asegura que este producto cumple todas las directivas de seguridad de la CE.

4

5 TABLE OF CONTENTS NOMENCLATURE... Video Duodenoscopes... PROCEDURE FLOWCHART...3 CARE AFTER USE General Application Important Instructions Internal Channels of Video Duodenoscope Endoscope Pre-Cleaning Leak Testing Cleaning High-Level Disinfection Sterilization Endoscope components and accessories Cleaning High-Level Disinfection Sterilization...48 General Endoscope components and accessories 2 POST REPROCESSING AND STORAGE SERVICING APPENDIX PENTAX Medical Compatible Reprocessing Systems/Agents...53 Reprocessing procedures of Video Duodenoscope Model ED-3490TK address the following internal channel systems. Video Duodenoscopes Suction Channel Air/Water Feeding Channel ED-3490TK Y Y Y N : YES : NO

6 NOMENCLATURE Video Duodenoscopes ED-3490TK RUBBER INLET SEAL (OF-B90) AIR/WATER FEEDING VALVE (OF-B88) SUCTION CONTROL VALVE (OF-B20) BUTTON BUTTON 2 INSTRUMENT CHANNEL INLET CONTROL BODY Alternate view BUTTON BUTTON 2 BUTTON 4 BUTTON 3 MODEL DESIGNAT UP/DOWN DEFLECTION LOCK BUTTON 4 BUTTON 3 Model Name Minimum Instrument Channel Width RIGHT/LEFT DEFLECTION LOCK Functions similar to Up/Down lock CANNULA/FORCEPS ELEVATOR CONTROL KNOB RIGHT/LEFT DEFLECTION CONTROL KNOB UP/DOWN DEFLECTION CONTROL KNOB UMBILICAL CORD STRAIN RELIEF BOOT ED-3490TK Minimum Instrument Channel Width (ID φmm) Video Duodenoscopes Minimum Instrument Channel Width (IDømm) ED-3490TK 4.2

7 DISTAL END STRAIN RELIEF BOOT INSERTION TUBE SECTION BENDING SECTION INSERTION TUBE ASSEMBLY ION AIR/WATER PORT PVE CONNECTOR Alternate view VENTING CONNECTOR VENTILATION CAP OF-C5 PVE CONNECTOR ELECTRICAL CONTACTS PVE SOAKING CAP OE-C9 Feedback Terminal LIGHT GUIDE PLUG SUCTION NIPPLE 2

8 PROCEDURE FLOWCHART Pre-Cleaning Preparation Wiping the insertion tube Aspiration of detergent solution through suction channel Flushing the air/water channel with air Transport to cleaning room Leak Testing Preparation Cleaning Cleaning the all external surfaces Brushing the suction channel and cylinder Filling the suction channel with detergent solution Soaking entire scope in detergent solution Rinsing with clean potable water Drying by wiping with a new lint-free gauze High-Level Disinfection Preparation Immersing the endoscope in disinfectant Filling channels with disinfectant Special disinfection handing procedure for elevator Soaking in disinfectant Rinsing Drying Sterilization 3

9 CARE AFTER USE -. General NOTE This Instructions for Use (IFU) has been written in accordance with 2CFR Part 80, ISO 7664, and national guidelines on reprocessing of medical products. General --. Application WARNING Reprocessing may affect device functionality. Prior to use, always inspect the endoscope, components, and accessories according to their respective IFU for proper functionality to determine that they are appropriate for patient use. Components and Accessories for Video Duodenoscopes TYPE MODEL Video Duodenoscope Name Number ED-3490TK Suction Control Valve OF-B20 Y Endoscope Component Air/Water Feeding Valve OF-B88 Y Inlet Seal OF-B90 Y Accessory Bite Block OF-Z5 Y PVE Soaking Cap OE-C9 Y Ventilation Cap OF-C5 Y Reprocessing Accessory Cleaning Adapter OF-B53 Y Cleaning Adapter OF-G7 Y Cleaning Brush CS-C9S Y Cleaning Brush CS602T Y Y N : YES : NO 4

10 --2. Important Instructions General WARNING Reusable Medical Devices that are initially supplied non-sterile require the end user to high level disinfect or sterilize them prior to initial use and to reprocess them correctly according to this IFU after each subsequent use. Proper care of the device after each clinical diagnostic or therapeutic procedure is extremely important. Immediately (within one hour) after the completion of a procedure, the endoscope, its removable components, and accessories should be precleaned and mechanically cleaned with a detergent solution specified in this IFU. Generally, if they are not precleaned within 5 minutes and mechanically cleaned within one hour after the conclusion of the procedure, dried blood, mucus, or other patient debris may cause damage to the devices or interfere with the ability of the user to properly reprocess them. The use of detergent immediately after each procedure to dissolve and remove organic contaminants and proteinaceous debris is essential to the proper care and maintenance of the endoscope. Prior to disinfection or sterilization, all instruments and components must be meticulously cleaned. Failure to do so can result in incomplete or ineffective disinfection and sterilization. Always inspect reprocessed endoscopes and accessories prior to use according to their respective Instructions for Use (IFU). WARNING are semicritical devices that require cleaning and at least highlevel disinfection. Use only legally marketed solutions and/or automated endoscope reprocessors(aers) for which efficacy validation testing including rinsing validation testing after high level disinfection with PENTAX products has been performed by their manufacturer. A list of legally marketed solutions/systems that have been determined to be materially compatible with PENTAX brand products is contained in this manual. 5

11 D R L F L U F R D D R L CAUTION During the reprocessing process, always wear protective equipment (e.g., gloves, gowns, face masks, etc.) to minimize the risk of cross contamination. Contact the manufacturer and follow local regulations regarding safe use, appropriate handling, and disposal of cleaning and disinfection solutions, including alcohol and rinse water. Safety Data Sheets available from the cleaning and disinfection solution manufacturers should be consulted to provide guidance to end users about formulation, hazards, chemical and physical properties, first aid, handling and storage, stability, precautions, disposal, etc. To avoid damaging the endoscope, do NOT twist, rotate or excessively bend the strain reliefs [(), (2)] during inspection, clinical use, reprocessing, or any handling activity. Be particularly cautious regarding the insertion tube strain relief [()]. When wiping the insertion tube and the umbilical cord, never apply excessive force or torque to the strain reliefs, the insertion tube, or the umbilical cord. General () (2) F U F U F F Do NOT Twist or Rotate Do NOT Bend Figure.. 6

12 NOTE General This IFU contains detailed recommendations on the manual reprocessing of PENTAX endoscopes using PENTAX supplied cleaning/disinfecting adapters. AERs may also be used to reprocess flexible endoscopes. However, only those AERs should be used whose manufacturers provide device-specific instructions and have validation data to support each AER claim with respect to PENTAX instruments. AER manufacturers should be consulted for their specific claims including, but not necessarily limited to: a) the ability of the AER to provide a cleaned and high-level disinfected (or sterilized) endoscope and endoscope components (e.g., valves), b) the identification of any special feature (internal channel) or endoscope component that cannot be reprocessed and therefore requires manual reprocessing, c) the microbial quality of the rinse water, d) the inclusion of an automated alcohol rinse cycle, e) the inclusion of a terminal drying cycle that removes the majority of water from within endoscope channels, f) maintenance procedures for water filter replacement and/or decontamination of the filtration system to ensure water of suitable quality, g) compliance with local regulations and/or guidelines. NOTE PENTAX flexible endoscopes should not be exposed to temperatures in excess of 40 F (60 C) during either reprocessing or storage. NOTE All of the steps in the validated reprocessing protocol described in this manual are intended to be performed in rapid succession and as a single, continual procedure. There should be no breaks in between steps of the protocol that are of sufficient duration to permit the endoscope to dry to such an extent that dislodged debris and/or microbial contaminants would be permitted to dry onto any endoscope surface. In the event that drying of the endoscope occurs due to an excessive break in the reprocessing procedure, the procedure should be completely repeated, beginning with the first precleaning step. 7

13 --3. Internal Channels of Video Duodenoscope The following internal schematic diagram is designed to help users better understand the intricate construction of PENTAX endoscopes. Please note that all solution entrance ports and flow pathways are illustrated below. (0) (2) () (3) (7) (4) (4) (5) (6) (8) (9) () Air/WaterNozzle (2) Water Feeding Channel (3) Air Feeding Channel (4) Inlet Seal (5) Instrument Channel Inlet (6) Cleaning Adapter OF-B53 (7) Suction Channel (8) Air/Water Cylinder (9) Suction Cylinder (0) PVE Soaking Cap (ATTACH) () Ventilation Cap (REMOVE) (2) Luer-Slip Syringe with Cleaning/ Disinfecting Solution (3) Suction Nipple (4) Cleaning Adapter OF-G7 General (3) () (2) Figure..2 NOTE The elevator wire/channel is completely sealed and therefore not exposed to potential contaminants and/or patient debris. Since the elevator wire/channel is fully enclosed and not exposed to patient material, it does not require special reprocessing. However, the actual elevator mechanism (front, back and base) must be cleaned and thoroughly reprocessed. Be aware that all recessed areas around the elevator mechanism must be thoroughly cleaned using detergent solution and a cleaning brush (CS-C9S). Cleaning instructions are provided within this IFU. 8

14 -2. Endoscope Video Duodenoscope ED-3490TK can be subjected to the following cleaning, disinfection, and sterilization process. Cleaning Sterilization Duodenoscope High-Level Liquid Steam Model Manual Ultrasonic Disinfection Chemical Sterilization Sterilization ED-3490TK Y N Y Y N Y N : Yes : No -2-. Pre-Cleaning WARNING During reprocessing, always wear protective equipment (e.g., gloves, gowns, face masks, etc.) to minimize the risk of cross contamination. Pre-cleaning is intended to remove visible debris from the endoscope immediately after its withdrawal from the patient, in order to increase the effectiveness of the subsequent cleaning procedure. are generally soiled with debris such as blood, tissues, and mucus when they are withdrawn from a patient. When such debris dries, it cannot be adequately removed in the subsequent cleaning procedure. It should be noted that pre-cleaning cannot substitute for the mechanical cleaning process. Always mechanically clean the endoscope after precleaning. During pre-cleaning, never wipe the insertion tube with alcohol or disinfecting solution. These solutions may fix organic contaminants and proteinaceous debris to the instrument and have an adverse effect on endoscope functionality and proper reprocessing. When using detergent, use only legally marketed brands that have been tested and found to be materially compatible by PENTAX (see APPENDIX). When injecting detergent solution through an internal channel, ensure that the solution exits freely from the endoscope distal end. If not, the channel might be blocked. Never use an endoscope with a blocked channel. Contact your local PENTAX facility to arrange for repair of the device. CAUTION Immediately after use, the metal light guide plug and electrical contacts/ pins of the endoscope may be HOT. To avoid burns, do not touch these areas immediately after use. For safer handling after a procedure, grasp the PVE connector housing. In order to prevent damage to the endoscope, do not place any removable components other than the items described in section -2-- of this Instructions for Use in the container used to transport the endoscope. In order to avoid damaging the endoscope, never subject it to suction in excess of 66kPa (9.57psi). 9

15 NOTE If the use of detergent solution is not permitted in the clinical procedure room, remove the endoscope from the procedure room and perform precleaning immediately in the reprocessing area Items required Endoscope component Inlet seal (OF-B90) Suction control valve (OF-B20) Air/water feeding valve (OF-B88) Other equipment Protective equipment such as gloves, gowns, face masks, etc., to minimize the risk of cross contamination. Detergent solution, Endozime (Ruhof Corporation) Water bottle and Video processor External suction source 500 ml basin Lint-free gauze 50 ml luer slip plastic syringe Preparation ) Wear personal protective equipment. 2) Prepare and fill a 500 ml basin with 400 ml ml of detergent solution per manufacturer's instructions (temperature, concentration). In the case of Endozime, add 30 ml of Endozime concentrate to 3.8 L ( gallon) of clean potable water at 20 C~30 C (68 F~86 F) Wiping the insertion tube and flushing the elevator The term Lower Position used in this document is depicted in Figure.2. Always be sure to visually check elevator positioning. Do not depend solely upon the operation and/ or position of the elevator control lever. Lower Position Figure.2. 0

16 F D R L U ) Turn off the lamp switch of the processor used for the duodenoscope. 2) Set the elevator to Lower Position by operating the elevator control lever on the control body. 3) Immediately after removing the endoscope from the patient, gently wipe the entire length of the insertion tube from control body to distal end three times using lint-free gauze soaked with detergent solution. Figure.2.2 4) Place the distal end of endoscope into a basin. 5) While the elevator is lowered, using a 50 ml syringe filled with detergent, spray 50 ml of detergent onto the rear of the elevator chamber. Figure Aspirating detergent solution through the suction channel ) Attach inlet seal (OF-B90) and suction control valve (OF-B20) to the endoscope. 2) Connect suction tube from external suction source to the endoscope suction nipple. () () Suction Tube (2) Suction Nipple (2) Figure.2.4 3) Turn on the external suction source.

17 4) Place the distal end of endoscope into a basin, and aspirate the detergent solution through the suction channel by pressing suction control valve (OF-B20) for 0 seconds. (2) () () OF-B90 (2) OF-B20 (3) Detergent Solution (3) Figure.2.5 5) Remove the distal end from the detergent solution, and press the suction control valve (OF-B20) for 0 seconds. 6) Turn off the external suction source. 7) Disconnect the suction tube from the endoscope suction nipple Flushing air/water channel with air ) Attach inlet seal (OF-B90) and air/water feeding valve (OF-B88) to the endoscope 2) Connect the air/water feeding tube of the water bottle to the endoscope air/water port. () () Air/water Feeding Tube (2) Air/water Port (2) Figure.2.6 3) Place the distal end of endoscope into a basin. 4) Set the lever on the water bottle to the drain position. 2

18 5) With the air pump of video processor ON and set to the HIGHEST pressure setting, flush the air channel with air by covering the top of air/water feeding valve (OF-B88) for 0 seconds. Then, discharge all water in the water channel by pressing the button of air/water feeding valve (OF-B88) for 0 seconds. (2) () () OF-B90 (2) OF-B88 Figure.2.7 6) Turn off the air pump of the video processor, and disconnect the air/water feeding tube of the water bottle from the endoscope air/water port Transport to cleaning room ) Turn off the power of the video processor, and detach the endoscope PVE connector from the video processor. 2) Transport the pre-cleaned endoscope to the cleaning room in a closed container. 3

19 -2-2. Leak Testing Before reprocessing and/or immersion in any fluids, PENTAX endoscopes should be tested for the loss of integrity in their watertight construction by using either PENTAX brand leakage testers (SHA-P2 or SHA-P5) or a leakage tester that is sold by PENTAX. For specific details on PENTAX leak detection procedures, please refer to the instructions supplied with PENTAX leakage testers or a leakage tester that is sold by PENTAX. CAUTION Various types of manual and automated endoscope leakage testers exist. Some are stand-alone units, and others may be integrated into an AER. PENTAX does not evaluate leakage testers that are not sold by PENTAX to verify their specific product claims with respect to their effectiveness to accurately detect leaks and/or their compatibility with PENTAX endoscopes. Insufficient pressures may reduce the likelihood of accurate leak detection, especially if the endoscope s distal bending section is not flexed during testing. Also, excessive pressures may adversely affect the endoscope, especially if pressurization occurs during automated reprocessing at elevated temperatures. PENTAX accepts no responsibility for use of leakage testers that it does not sell. Users should check with the leakage tester manufacturer and confirm their specific product claims, including compatibility with PENTAX endoscopes at various temperatures and their ability to detect leaks with/ without fluid immersion and with/without flexing of the endoscope s distal bending section. 4

20 -2-3. Cleaning WARNING During reprocessing, always wear protective equipment (e.g., gloves, gowns, face masks, etc.) to minimize the risk of cross contamination. In order to ensure thorough cleaning, be sure to perform all cleaning steps. The effectiveness of each cleaning step will influence the effectiveness of subsequent steps. Failure to properly follow the cleaning steps described may result in incomplete or ineffective cleaning, disinfection and/or sterilization of the endoscope and/or endoscope components and accessories, resulting in a potential risk of cross-contamination. Immediately (within one hour) after the completion of a procedure, the endoscope and its components should be thoroughly and carefully cleaned with detergent solution. If the endoscope and its components are left uncleaned for an excessive time after use, dried blood, mucus or other patient debris may cause damage or interfere with the ability of the user to properly reprocess the device. For cleaning, use only legally marketed detergents that have been tested according to the instructions of the manufacturer and found to be materially compatible by PENTAX (see APPENDIX). Fresh detergent solution must be used for each endoscope that is reprocessed. After using operational/cleaning accessories (e.g., forceps,needles, snares, brushes etc.) with the endoscope, carefully check that all accessories are intact and that no parts have fallen off and become lodged within the endoscope's instrument/suction channel. Furthermore, ensure that any therapeutic devices (e.g., clips, stents, etc.) passed through the channel are accounted for after use. If the channel becomes blocked or clogged due to the accumulation of debris, an accessory that cannot be removed, or other cause, do NOT attempt to correct the blockage or continue to use the endoscope. In such a case, contact your local PENTAX Medical service facility to have the endoscope repaired. The use of an endoscope with a blocked internal channel may result in ineffective reprocessing and/or introduction of debris and/or device components into a patient during a subsequent procedure, posing a risk of cross-contamination. CAUTION PVE soaking cap (OE-C9) must be properly secured over the electrical contacts. Failure to do so can result in water invasion and damage to the endoscope. If an endoscope is cleaned without the PVE soaking cap attached, do not use the endoscope, and contact your local PENTAX service facility or sales representative. The ventilation cap (OF-C5) must be taken OFF during reprocessing. Failure to do so can result in damage to the endoscope. If an endoscope is cleaned with the ventilation cap attached, do not use the endoscope, and contact your local PENTAX service facility or sales representative. During cleaning, never twist, rotate, or bend the insertion portion and umbilical cord excessively. Never subject the endoscope to ultrasonic cleaning methods. In order to prevent damage to the endoscope, do not place any objects other than the reprocessing accessories listed in section of this Instructions for Use with the endoscope when immersing the endoscope in a cleaning basin. 5

21 Items required Endoscope component Inlet seal (OF-B90) Reprocessing accessory PVE soaking cap (OE-C9) Cleaning brush (CS602T) Cleaning brush (CS-C9S) Cleaning Adapter (OF-B53) Cleaning Adapter (OF-G7) Other equipment Protective equipment such as gloves, gowns, face masks, etc., to minimize the risk of cross contamination. Detergent solution, Endozime (Ruhof Corporation) Clean potable water Basin sufficient in size to immerse the entire endoscope (at least 50 cm in length x 40 cm in width x 5 cm in depth, or 9.7 inches in length x 5.8 inches in width x 5.9 inches in depth) Lint-free gauze 50 ml luer slip plastic syringe Preparation ) Wear personal protective equipment. 2) Attach PVE soaking cap (OE-C9) to the endoscope. 3) Detach ventilation cap (OF-C5) from the endoscope. (2) () OE-C9 (2) OF-C5 () Figure.2.8 4) Fill a basin with a sufficient volume of detergent solution to completely immerse the endoscope. Prepare the detergent in accordance with the manufacturer s instructions (temperature, concentration). In the case of Endozime, add 30 ml of Endozime concentrate to 3.8 L ( gallon) of clean potable water at 20 C~30 C (68 F~86 F). 6

22 Cleaning all external surfaces CAUTION Do not squeeze or severely bend the insertion tube. Do not use any abrasive materials. Be careful to avoid damage to the distal lenses. The terms Lower Position and Upper Position" used in this document are depicted in Figure.2.9. Always be sure to visually check elevator positioning. Do not depend solely upon the operation and/or position of the elevator control lever. Lower Position Upper Position Figure.2.9 ) Fully immerse the endoscope with the components attached in the detergent solution. 2) Detach inlet seal (OF-B90), suction control valve (OF-B20), and air/water feeding valve (OF-B88) from the endoscope. Open the cap of inlet seal (OF-B90) in the detergent solution. (2) (3) () () OF-B90 (2) OF-B20 (3) OF-B88 Figure.2.0 3) Suction control valve (OF-B20) and air/water feeding valve (OF-B88) are reprocessed separately from endoscope. (see Section -3 Endoscope components and accessories ) 4) While still immersed in the detergent solution, wipe the entire surface of endoscope and inlet seal (OF-B90) two times in one direction with a lint-free gauze. 7

23 Brushing and flushing of Elevator 5) Set the elevator to the Upper Position by operating the elevator control lever on the control body. 6) Hold cleaning brush (CS-C9S) with fingers on brush shaft near the small brush head. 7) Perform steps a) through c) in order to brush the front of elevator chamber and elevator recess area with cleaning brush (CS-C9S). Repeat nine more times (total ten times). a) Place the smaller brush head of cleaning brush (CS-C9S) underneath the front opposite of lens side of the elevator chamber. b) Brush to the lens side. c) Brush to the front scraping out the debris in the lens side corner and wall of the elevator chamber. c b a Figure.2. 8) Clean the debris from the brush head by rubbing with fingers. 9) Similarly perform step a) through c) in order to brush from the front lens side. Repeat nine more times. (total ten times) a) Place the smaller brush head of cleaning brush (CS-C9S) underneath the front lens side of the elevator chamber. b) Brush to the opposite of lens side. c) Brush to the front scraping out the debris in the front opposite of lens side corner and wall of the elevator chamber. a b b Figure.2.2 0) Clean the debris from the brush head by rubbing with fingers. 8

24 ) While the elevator is raised, place the smaller brush head of cleaning brush (CS-C9S) into the space behind the elevator. Brush to the front scraping out the debris in the back of the elevator. Repeat nine times. (total ten times) Back Figure.2.3 2) Clean the debris from the brush head by rubbing with fingers. 3) While the elevator is raised, place the smaller brush head of cleaning brush (CS-C9S) into the front area of elevator chamber. Brush to the front scraping out the debris in the back of the front area of elevator chamber. Repeat nine times. (total ten times) Front Figure.2.4 4) Clean the debris from the brush head by rubbing with fingers. 5) While the elevator is raised, using a 50 ml syringe filled with detergent solution without air, spray 50 ml of detergent onto the front of the Elevator Chamber. Figure.2.5 9

25 6) Set the elevator to Lower Position by operating the elevator control lever on the control body. 7) Perform steps a) through c) in order to brush the back area of elevator chamber and elevator recess area with cleaning brush (CS-C9S). Repeat nine more times (total ten times). a) Place the smaller brush head of cleaning brush (CS-C9S) into the opposite side of lens side of the back area of elevator chamber. b) Brush to the lens side. c) Brush to the front scraping out the debris in the right corner of the back area of elevator chamber. c b a Figure.2.6 8) Similarly perform steps a) through c) in order to brush from the lens side. Repeat nine more times. (total ten times) a) Place the smaller brush head of cleaning brush (CS-C9S) underneath the lens side of the back of elevator chamber. b) Brush to the opposite of lens side. c) Brush to the front scraping out the debris in the opposite of lens side corner and wall of the back area of elevator chamber. a b c Figure

26 9) Place the smaller head of cleaning brush (CS-C9S) on the elevator. In order to clean the upper surface of the elevator, brush back and forth ten times. Figure ) While the elevator is lowered, using a syringe filled with detergent without air, spray 50 ml of detergent onto the rear of the elevator chamber. Figure.2.9 2) Visually inspect the entire surface of the endoscope to insure that no soil is present, paying special attention to areas such as the distal end, elevator, elevator chamber, and control body, which are the most likely regions to retain visible soil. Figure ) If any soil is still present on the distal end and control body, use cleaning brush (CS-C9S) to gently brush until all soil has been removed. 23) If any soil is still present on the elevator and elevator chamber, repeat steps 5-20 until all soil has been removed. 2

27 Brushing the suction channel and cylinder WARNING Do not use cleaning brushes other than those specified in this instruction for use. Failure to do so can result in endoscope damage or incomplete or ineffective cleaning. Cleaning brush (CS602T) is provided non-sterile for one time use. Never reuse the brush on more than one instrument. Prior to use, ensure that cleaning brushes are not damaged (e.g., kinked shaft or bent or missing bristles). In order to prevent the dissemination of patient debris left in the endoscope channel into the environment, always withdraw brushes slowly. CAUTION In order to avoid damage to the endoscope distal end, never attempt to insert a cleaning brush into the endoscope distal tip. Do not insert cleaning brush (CS602T) into the suction control valve cylinder. The brush head of cleaning brush (CS602T) could become stuck within the suction control valve cylinder. Never apply excessive pressure to introduce or withdraw the brush. This can result in damage to the endoscope and/or the brush. Some manufacturers cleaning brushes/devices have been found to damage PENTAX endoscopes and/or create the need for service, as they can become lodged ( stuck ) inside various lumens of PENTAX endoscopes. The likelihood of endoscope damage or servicing increases if a cleaning device does not have a protective tip (or contains any sharpedged surface), if its flexible shaft uses a flimsy plastic material that is not firm enough to allow for easy accessory advancement, and/or if the proper sequence and/or direction of channel cleaning is not followed as described in the PENTAX IFU. To prevent excessive friction between brush and channel, do NOT tightly coil the insertion tube and umbilical cord to a diameter of less than 30cm. NEVER attempt to pass the cleaning brush through a fully angulated endoscope. Failure to follow these instructions can result in endoscope or brush damage. 22

28 Suction channel cleaning ) Insert the large brush head of cleaning brush (CS-C9S) into the instrument channel inlet until it cannot be advanced further. Repetitively move the brush back and forth while twisting it left and right for 0 seconds in order to scrub the entire inner surface of the instrument channel inlet. () (2) () CS-C9S (2) Instrument Channel Inlet Figure.2.2 2) Withdraw cleaning brush (CS-C9S) from the inlet, and clean the debris from the brush head by rubbing with fingers. 3) Insert the blue tip of cleaning brush (CS602T) into the opening of the suction nipple, and gently pass the brush until it appears in the suction cylinder. (3) () CS602T (2) Suction Nipple (3) Suction Cylinder (2) () Figure ) Grasp the blue tip of the brush shaft, and gently pull the brush from the suction cylinder until the brush heads exit the suction cylinder. Clean debris from the brush head by rubbing with fingers. 23

29 5) Insert the blue tip of cleaning brush (CS602T) into the opening at the bottom of the suction cylinder on the control head, and gently advance the brush until it exits the distal end of the endoscope. () (2) () CS602T (2) Suction Cylinder (3) Distal End (3) Figure ) Grasp the blue tip of the brush shaft, and gently pull the brush from the distal end of the endoscope until the brush heads exits the distal end. Clean debris from the brush head by rubbing with fingers. Cylinder cleaning 7) Insert the large brush head of cleaning brush (CS-C9S) into the suction cylinder until it cannot be advanced further. Repetitively move the brush back and forth while twisting it left and right for 0 seconds in order to scrub the entire inner surface of the cylinder. () () CS-C9S (2) Suction Cylinder (2) Figure ) Withdraw the brush from the cylinder, and clean debris from the brush head by rubbing with fingers. 24

30 9) Insert the large brush head of cleaning brush (CS-C9S) into the air/water cylinder until it cannot be advanced further. Repetitively move the brush back and forth while twisting it left and right for 0 seconds in order to scrub the entire inner surface of the cylinder. () () CS-C9S (2) Air/water Cylinder (2) Figure ) Withdraw the brush from the cylinder, and clean debris from the brush head by rubbing with fingers. ) After brushing the endoscope, dispose of cleaning brush (CS602T), which is a single-use accessory Filling the channel with detergent solution WARNING While injecting detergent solution through the channels, avoid the introduction of air. The presence of air bubbles can prevent contact of the detergent solution with channel surfaces. Always fully immerse the endoscope while flushing detergent solution into the endoscope channel. It is imperative that the cleaning adapter (OF-B53) be securely attached to the endoscope. Failure to properly connect and secure the cleaning adapter can result in ineffective and incomplete reprocessing. CAUTION In order to avoid damage to the endoscope, never apply excessive force if resistance is encountered while flushing the detergent solution into the channels. Always immerse the endoscope, components, and accessories completely in the detergent solution during cleaning. 25

31 Attaching components and accessories to the endoscope ) Follow the steps below in order to attach cleaning adapter (OF-B53) to the air/ water cylinder and suction cylinder. This adapter (seals) the air/water and suction cylinders to permit unidirectional flow of solution through these delivery/aspiration systems. [a] Raise the locking cap and slide the side covers (marked with ) downward. [b] Align the cylinder guide over the air/water and suction cylinder ports. [c] Slide the cleaning adaptor (OF-B53) forward. [d] Holding the side covers (marked with ), push down and slide the locking cap under the locking tab to secure. [e] Secure the locking cap with the locking tab on the base. [a] () (2) () Rubber Seal (2) Locking Cap (3) Base (3) [b] (2) () Cylinder Guide (2) Cylinder Ports (3) Cylinder Inlets () (3) [c] [d] (2) () End of Locking Cap (2) Locking Tab () [e] () Locking Tab () () Figure

32 2) Attach inlet seal (OF-B90) and cleaning adapter (OF-G7) to the endoscope. () (2) () OF-B53 (2) OF-B90 (3) OF-G7 (3) Figure.2.27 Filling channels with detergent solution 3) Attach a 50 ml syringe filled with the detergent solution without air to cleaning adaptor (OF-G7), and inject 250 ml of the detergent solution into the air/water channel. () OF-G7 () Figure ) Check to confirm that detergent solution flows out from the air/water nozzle(s) on distal end. 5) Attach a 50 ml syringe filled with detergent solution without air to the suction nipple, and inject 350 ml of the detergent solution into the suction channel. () Suction Nipple () Figure

33 6) Attach a 50 ml syringe filled with detergent solution without air to the inlet seal (OF-B90), and inject 50 ml of detergent solution into the suction channel. () () OF-B90 Figure ) Check to confirm that detergent solution flows out from the suction channel opening on the distal end. 8) After injecting detergent solution into all channels, detach inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) from the endoscope, and leave them to soak along with the endoscope in the detergent solution until the next step. Open the cap of inlet seal (OF-B90). () (2) () OF-B53 (2) OF-B90 (3) OF-G7 (3) Figure

34 Soaking in detergent solution WARNING The detergent solution must remain in contact with ALL internal channels, external endoscope surfaces, and components for the time period recommended by the manufacturer of the detergent. During immersion, detach all components and accessories (except the soaking cap) from the endoscope to ensure contact of all endoscope surfaces with the detergent solution. Avoid introduction of air during the flushing process. Confirm that no air bubbles exit from the channel openings at the endoscope distal tip. The presence of the air bubbles can prevent contact of the detergent with channel surfaces. CAUTION NEVER subject the endoscope to an ultrasonic cleaning method. ) While fully immersing the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7), ensure that there are no air bubbles on all external surfaces. If any air bubbles are detected, flush them away with detergent solution using a 50 ml syringe. 2) Soak the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7) under conditions (temperature, concentration, time) specified by the detergent manufacturer. In the case of Endozime, the soaking time is 3 minutes. Figure.2.32 Purging detergent solution from all channels 3) After soaking, attach inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) to the endoscope, and close the cap of inlet seal (OF-B90). (see Figure.2.27, p27) 4) Attach a 50 ml syringe filled with air to cleaning adapter (OF-G7), and flush the air/water channel with 250 ml of air to purge residual detergent solution. (see Figure.2.28, p27) 5) Attach a 50 ml syringe filled with air to the suction nipple, and flush the suction channel with 350 ml of air to purge residual detergent solution. (see Figure.2.29, p27) 6) Attach a 50 ml syringe filled with air to the inlet seal (OF-B90), and flush the suction channel with 50 ml of air to purge residual detergent solution. (see Figure.2.30, p28) 7) Remove the endoscope (with inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) attached) from the detergent solution. 29

35 Rinsing WARNING It is important that all internal channels, external endoscope surfaces, and components be thoroughly rinsed with clean water to remove residual detergent solution. Failure to do so can result in ineffective or incomplete disinfection and sterilization. First rinse ) Place the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF- G7) into a basin of clean potable water that is of sufficient volume to completely immerse the endoscope. 2) Detach inlet seal (OF-B90) and cleaning adapters (OF-B53, and OF-G7) from the endoscope (see Figure.2.3, p28). Open the cap of inlet seal (OF-B90). 3) Wipe all exterior surfaces of the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7) two times with a lint-free gauze in order to remove residual detergent solution. 4) While still completely immersed in clean potable water, grasp the distal end, control body of endoscope, and PVE connector with two hands, and agitate it under the clean potable water by moving it from side to side repeatedly for 20 seconds. Figure ) Grasp inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) with a hand, and agitate them under the clean potable water by moving them from side to side repeatedly for 20 seconds. 6) Attach inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) to the endoscope, and close the cap of inlet seal (OF-B90). (see Figure.2.27, p27) 7) Attach a 50 ml syringe filled with clean potable water without air to cleaning adapter (OF-G7), and flush the air/water channel with 250 ml of water. (see Figure.2.28, p27) 8) Attach a 50 ml syringe filled with clean potable water without air to the suction nipple, and flush the suction channel with 350 ml of water. (see Figure.2.29, p27) 9) Attach a 50 ml syringe filled with clean potable water without air to the inlet seal (OF- B90), and flush the suction channel with 50 ml of water. (see Figure.2.30, p28) 0) Remove the endoscope (with inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) attached) from the clean potable water. 30

36 ) Attach a 50 ml syringe filled with air to cleaning adapter (OF-G7), and flush the air/water channel with 250 ml of air. (see Figure.2.28, p27) 2) Attach a 50 ml syringe filled with air to the suction nipple, and flush the suction channel with 350 ml of air. (see Figure.2.29, p27) 3) Attach a 50 ml syringe filled with air to the inlet seal (OF-B90), and flush the suction channel with 50 ml of air. (see Figure.2.30, p28) Second rinse 4) Fill a basin with clean potable water, and repeat steps - 3 in order to perform a second complete rinse. 5) Detach inlet seal (OF-B90) and cleaning adapters (OF-B53 and OF-G7) from the endoscope. (see Figure.2.3, p28) Drying ) Gently wipe and dry all external surfaces of the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7) with a new lint-free gauze Visual inspection NOTE Ideally, work surfaces used for inspection require,000-2,000 lux illumination (ANSI/AMMI ST79). ) Under adequate lighting, visually inspect the entire surface of the endoscope and accessories to insure that no debris is present, paying special attention to the surfaces of the duodenoscope elevator, recessed areas around the elevator, and the surfaces of the elevator chamber. If any soil is still present on the duodenoscope and accessories, remove it by wiping and/or brushing the area(s) in question in detergent solution with lint-free cloths and/or the appropriate brush(es) specified in this IFU until it has been completely removed. After complete removal of residual soil has been attained, perform all rinsing steps that normally follow cleaning. 3

37 -2-4. High-Level Disinfection Prior to high-level disinfection, the end user should confirm that the high level disinfectant concentration is greater than the minimum effective concentration (MEC) of reused disinfectant as specified in the manufacturer s instructions. WARNING During the reprocessing process, always wear protective equipment (e.g., gloves, gowns, face masks, etc.) to minimize the risk of cross contamination. Prior to disinfection, it is imperative that any solutions previously used in the cleaning process be thoroughly rinsed and dried. Failure to do so can result in ineffective or incomplete disinfection. For high-level disinfection, use an appropriate disinfecting solution according to the instructions of the disinfectant s manufacturer (temperature, concentration, time). Adhere to the instructions to accomplish effective and complete disinfection. The endoscope may be damaged if exposed to a disinfectant under conditions other than those specified by the disinfectant manufacturer. Use only a legally marketed disinfectant that has been tested according to the instructions provided by the disinfectant manufacturer and found to be materially compatible by PENTAX. It is imperative that ALL internal surfaces of the channels be in contact with the disinfecting solution for the time period and at the temperature recommended by the manufacturer of the disinfectant. Ideally, all final rinses should be performed with sterile water, clean potable water, or water that meets the requirements of the health care facility. Regardless of the quality of the rinse water used, it is essential to perform a final alcohol rinse followed by forced air in order to completely dry the endoscope channels and prevent bacterial colonization and/or infections associated with waterborne microorganisms. The basin that is used to perform disinfectant immersion should be thoroughly cleaned prior to filling it with disinfectant solution. Prior to disinfection, the endoscope and accessories must be meticulously cleaned. Failure to do so can result in incomplete or ineffective disinfection. CAUTION Prior to high level disinfection, attach PVE soaking cap (OE-C9). Failure to do so can result in water invasion and damage to the endoscope. If the endoscope is disinfected without the PVE soaking cap attached, do not use the endoscope, and contact your local PENTAX service facility or sales representative. Prior to disinfection, detach the ventilation cap (OF-C5). Failure to do so can result in damage to the endoscope. If the endoscope is disinfected with the ventilation cap attached, do not use the endoscope, and contact your local PENTAX service facility or sales representative. During disinfection, never twist, rotate or bend the insertion tube and umbilical cord excessively. In order to prevent damage to the endoscope, do not place any objects other than the reprocessing accessories described in section of this Instructions for Use with the endoscope when immersing it in the disinfection basin. 32

38 Items required Endoscope component Inlet seal (OF-B90) Reprocessing accessory PVE soaking cap (OE-C9) Cleaning adapter (OF-B53) Cleaning adapter (OF-G7) Other equipment Protective equipment such as gloves, gowns, face masks, etc., to minimize the risk of cross contamination. Disinfecting solution, CIDEX Activated Dialdehyde Solution (Johnson & Johnson) Sterile water (preferred) or clean potable water 70-90% medical grade ethyl or isopropyl alcohol Basin sufficient in size to immerse the entire endoscope (at least 50 cm in length x 40 cm in width x 5 cm in depth, or 9.7 inches in length x 5.8 inches in width x 5.9 inches in depth). Sterile gauze 50 ml luer slip syringe Preparation ) Wear personal protective equipment. 2) Attach PVE soaking cap (OE-C9) to the endoscope. 3) Ensure that ventilation cap (OF-C5) is detached from the endoscope. (2) () OE-C9 (2) OF-C5 () Figure ) Prepare a basin of sufficient volume capacity to fully immerse the endoscope with CIDEX Activated Dialdehyde Solution in accordance with the disinfectant manufacturer s instructions for concentration and temperature Immersing the endoscope in disinfecting solution ) Fully immerse the endoscope, inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7) in the disinfecting solution. Open the cap of inlet seal (OF-B90). 2) Wipe all external surfaces of the endoscope with a sterile gauze two times in order to thoroughly contact disinfecting solution to the surfaces. Ensure that all surfaces are in contact with disinfecting solution. 33

39 3) Lower the elevator by operating the elevator control lever on the control body Filling the channel with disinfecting solution WARNING When filling endoscope channels with disinfectant, avoid the introduction of air. The presence of air bubbles can prevent contact of the disinfectant with channel surfaces. Avoid introduction of air during the flushing process. Confirm that no air bubbles exit from the channel openings at the endoscope distal tip. The presence of the air bubbles can prevent contact of the disinfectant with channel surfaces. Always immerse the endoscope while filling endoscope channels with disinfectant. The terms Lower Position, Middle Position, and Upper Position" used in this document are depicted in Figure.2.35 Always be sure to visually check elevator positioning. Do not depend solely upon the operation and/or position of the elevator control lever. Lower Position Middle Position Upper Position Figure.2.35 Always immerse the endoscope, components, and accessories in disinfecting solution during disinfection. ) Attach inlet seal (OF-B90), and cleaning adapters (OF-B53 and OF-G7) to endoscope, and close the cap of inlet seal (OF-B90). () (2) () OF-B53 (2) OF-B90 (3) OF-G7 (3) Figure

EG-2990i, EG-2790i EG-1690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i10, EG29-i10

EG-2990i, EG-2790i EG-1690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i10, EG29-i10 Only For the Americas INSTRUCTIONS FOR USE (REPROCESSING) PENTAX VIDEO UPPER G.I. SCOPES EG-2990i, EG-2790i EG-690K, EG-2490K EG-2790K, EG-2990K EG-3490K, EG-3890TK EG27-i0, EG29-i0 Product Overview These

More information

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II)

ON TRACK Reprocessing In-service / Competency for CF/PCF/GIF Endoscopes (EVIS, EXERA, EXERA II) THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title.

Attendance Sheet. (List each model reviewed during this session) Print Name Signature Title. Print Name Signature Title. THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT

OPERATION MANUAL. (Cleaning, Disinfection and Storage) Endoscope ED-530XT US Endoscope ED-530XT OPERATION MANUAL (Cleaning, Disinfection and Storage) Thank you for purchasing our product. Read this manual carefully before use to avoid unexpected accidents and to take full advantage

More information

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE Single use soft brush (MAJ-1888) INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF TYPE Q180V Accessories: Water resistant cap (MH-553) Biopsy valve (MB-358) Air/water valve (MH-438) Suction cleaning

More information

2015 Myco Industries, Inc Fax:

2015 Myco Industries, Inc Fax: Endoscope Cleaning and Handling It is highly important that you manually clean the endoscope immediately after it is removed from the patient, BEFORE it is automated or manually disinfected. Cleaning your

More information

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician. Reprocessing Manual Fuse 1C Colonoscope Part Number FSE-056-EN-2.00 Released August, 2013. This manual includes complete reprocessing instructions for the Fuse 1C Colonoscope system. Please refer to the

More information

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE XP160 OLYMPUS GIF TYPE 160 OLYMPUS GIF TYPE Q160 OLYMPUS GIF TYPE Q160Z OLYMPUS GIF TYPE 1TQ160 OLYMPUS GIF TYPE XTQ160 EVIS EXERA COLONOVIDEOSCOPE

More information

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use. Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1 ENGLISH Intended Use The Welch Allyn Rigid Reusable & Single Use Sigmoidoscopes and Anoscopes are used to illuminate and

More information

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. Flexible Nasopharyngoscope Operator Manual Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc. 1. Introduction The flexible nasopharyngoscope is sold in unsterile condition as a reusable

More information

Instructions for Use Spinal Punches IFU

Instructions for Use Spinal Punches IFU Indications for use Spinal punches are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column. Contraindiciations Instruments should not

More information

Reusable Laryngoscope Blades and Handles

Reusable Laryngoscope Blades and Handles Manufacturer: Starkling Medizintechnik Method: Re-processing instructions Symbol: N.A. Device(s): 1 Green System Reusable Laryngoscope Blade with integrated Light Guide, Solid Stainless Steel Construction

More information

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية 1. POLICY STATEMENT: Cleaning, decontamination and sterilization method should be done in CSSD service facility not in the end user department side. 2. PURPOSE: To ensure the proper way of collections,

More information

When you sterilize an instrument set, you know what the standards are.

When you sterilize an instrument set, you know what the standards are. Sterilization & Infection Control The importance of cleaning in earnest A regular column on sterilization and infection control issues. When you sterilize an instrument set, you know what the standards

More information

2. Push the drain connector into the disinfectant removal port until it clicks.

2. Push the drain connector into the disinfectant removal port until it clicks. Chapter 3 Inspection Before Use 2. Push the drain connector into the disinfectant removal port until it clicks. When connecting the drain connector to the disinfectant removal port, do not push the connector

More information

CLEANING INSTRUCTIONS

CLEANING INSTRUCTIONS Reusable Instrument Care Manual Pagoda Pedicle Screw System Manufacturer Ortho Development Corporation 12187 South Business Park Drive Draper, UT 84020 Phone 801-553-9991 Fax 801-553-9993 www.orthodevelopment.com

More information

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) Fiber Optic Cables Instructions for Use and Processing Instructions DFU-950-0031-00 Fiber Optic Cables Made in USA Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) 737 7611

More information

MODEL 7000 SUCTION UNIT

MODEL 7000 SUCTION UNIT MODEL 7000 SUCTION UNIT OPERATOR S MANUAL Caution Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices

More information

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications TABLE OF CONTENTS MANUAL PRE-CLEANING High Speed Handpieces 2

More information

I.F.U Re/Processing Reusable Medical Devices

I.F.U Re/Processing Reusable Medical Devices I.F.U Re/Processing Reusable Medical Devices Distributed by: MANUFACTURER: GA 26-09-001 -EN /20130719 Contents 1 Overview of preparation methods... 3 2 Safety and responsibility... 4 3 Explanation of symbols...

More information

INSTRUCTION MANUAL V8F145

INSTRUCTION MANUAL V8F145 MEDIZINTECHNIK FÜR TIERÄRZTE INSTRUCTION MANUAL V8F145 MEDIZINTECHNIK FÜR TIERÄRZTE Eltastr. 8 78532 Tuttlingen Phone: +49/7461/96580 55 Fax: 1 +49/7461/96580-91 www.eickemeyer.de export@eickemeyer.de

More information

Cleaning & Sterilization

Cleaning & Sterilization Handpieces & Attachments Instruments Cleaning & Sterilization System 7 System 6 System 5 CD4 CD3 SABO 2 SABO RemB CORE TPS Processing Instructions User/Patient Safety all contribute to the efficacy of

More information

Instructions for Use: Fiberscope and LED Light Source

Instructions for Use: Fiberscope and LED Light Source Instructions for Use: and Brand Name of Product Generic Name of Product Product Code Number(s) Intended Use Range of Applications for Product Key Specifications of Product, FBR-001, FBR-002, FBR-ULB-100,

More information

OLYMPUS AMERICA, INC. ENDOSCOPE GROUP TWO CORPORATE CENTER DRIVE MELVILLE, NEW YORK TEL (631) March 3, 2003

OLYMPUS AMERICA, INC. ENDOSCOPE GROUP TWO CORPORATE CENTER DRIVE MELVILLE, NEW YORK TEL (631) March 3, 2003 OLYMPUS AMERICA, INC. ENDOSCOPE GROUP TWO CORPORATE CENTER DRIVE MELVILLE, NEW YORK 11747-3157 TEL (631) 844-5000 March 3, 2003 Re: GI Endoscopes with Water Features Dear Olympus Customer: Over the years,

More information

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013 Video Laryngoscope Operator s Manual ( Instructions for use ) Doc no: 100-160-000 Revision 1.1 8th January 2013 Aircraft Medical Ltd 2012 www.aircraftmedical.com Contents 1 Introduction 2 1.1 Description

More information

Operating / User Manual

Operating / User Manual UniARM Surgical Support System Operating / User Manual Version 1.1 MITAKA KOHKI Co., Ltd. 1. Introduction Introduction Thank you for purchasing the UniARM for use in surgical procedures. To use the device

More information

Product Catalogue 2016

Product Catalogue 2016 Product Catalogue 2016 1985 The Reusable Bottle 2004 The Disposable Bottle Top 2015 THE ANA - The latest evolution The ANA- The future of water irrigation systems Independent lab reports show that at the

More information

X99 Series. Dental Handpiece. Instruction For Use

X99 Series. Dental Handpiece. Instruction For Use X99 Series Dental Handpiece Instruction For Use 0413 Table of Contents Symbols... 2 Introduction... 3-4 Before Use... 5 Product description... 6 Technical Specifications... 7 Operation... 8-11 Hygienic

More information

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution S168-544-OO1E REV.A Schuco byallied MEDICAL ASPIRATOR MnrInI -33OA USER'S MANUAL A Caution Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by

More information

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing Instructions for Use. Description and Intended Use Integra LifeSciences Corporation reusable surgical device sets consist of various

More information

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts Modular Slide Lock Handle Assembly 1 Movable Handle 2 Fixed Handle 3 Handle Access For Cleaning 4 Window 5 Handle Screw 6 Keeper 7 Nonconductive Luer Port Cap (shown covering the Luer Port) 8 Shaft Insulation

More information

Table of Contents. English

Table of Contents. English OM-E0799E 000 English Thank you for purchasing VIVA ace Motor Kit. Please read this Operation Manual and the VIVA ace Basic Set Operation Manual carefully before use for operating instructions and care

More information

Torrent irrigation pump Reorder No Operating and Maintenance Manual

Torrent irrigation pump Reorder No Operating and Maintenance Manual Torrent irrigation pump Reorder No. 00711569 Operating and Maintenance Manual 00732292 Rev. D Issued Date: August 2018 Warning: An issued or revision date for these instructions is included for the user

More information

FLASH PAK STERILIZATION CONTAINER SYSTEM

FLASH PAK STERILIZATION CONTAINER SYSTEM INSTRUCTIONS FOR USE FLASH PAK STERILIZATION CONTAINER SYSTEM Description The Flash Pak Sterilization Container System consists of a family of rigid reusable containers that provide an effective sterilization

More information

ENDOSCOPE DECONTAMINATION : Where are we now!

ENDOSCOPE DECONTAMINATION : Where are we now! ENDOSCOPE DECONTAMINATION : Where are we now! Christina Bradley Laboratory Manager Hospital Infection Research Laboratory City Hospital Dudley Road, Birmingham CHOICE OF DECONTAMINATION METHOD Depends

More information

ASPIRE Laboratory Aspirator

ASPIRE Laboratory Aspirator ASPIRE Laboratory Aspirator USER MANUAL Rev 2/14/18 Accuris Instruments / Benchmark Scientific Ph: (908) 769-5555 E-mail: info@accuris-usa.com (C) Benchmark Scientific, 2018 THE ACCURIS ASPIRE LABORATORY

More information

CORRECT ENDOSCOPE REPROCESSING

CORRECT ENDOSCOPE REPROCESSING CORRECT ENDOSCOPE REPROCESSING Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham HIRL Birmingham 1 Inadequate decontamination procedures and

More information

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS CLEANING AND HANDLING OF WRIGHT INSTRUMENTS 150824-0 The following languages are included in this packet: English (en) Deutsch (de) Nederlands (nl) Français (fr) Español (es) Italiano (it) Português (pt)

More information

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only V. Mueller en V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA Rx Only V. Mueller, CareFusion and the CareFusion logo are trademarks or registered trademarks of CareFusion Corporation

More information

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI ULTRASONIC SCALER PIEZO MINI CONTENTS XI - SYMBOLS 1. INTRODUCTION 1 Alternating current Type BF device 2. WARNINGS 1 3. PRESENTATION 1 3.1 Presentation 1 3.2 Technical description 2! Warning, please refer

More information

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY SUBJECT: ULTRASONIC CLEANER MONITORING DAILY DEPARTMENT: CPD, CSSD, SPA, MDRD, SPD, dental offices, medical offices and department using

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ General Information Use The Air drill System is a pneumatic powered system used for many applications orthopedic and trauma surgery. To ensure proper operation of the air drill, use only original attachment

More information

TBJ Model SPCS B Scope Pre-Cleaning Sink. Product Specification. Product Description

TBJ Model SPCS B Scope Pre-Cleaning Sink. Product Specification. Product Description TBJ Model SPCS 30-48-1B Scope Pre-Cleaning Sink Product Specification Product Description The TBJ Model SPCS 30-48-1B work sink is specifically designed for use in GI Labs and Sterile Processing Departments

More information

ALLERGEN MIXING ASSISTANT

ALLERGEN MIXING ASSISTANT ALLERGEN MIXING ASSISTANT User Manual 9495 SW Locust Street Suite E Portland, OR 97223 (503) 379-0110 Fax: (503) 715-1378 www.xtractsolutions.com SM026R2 Welcome to Xtract Solutions Thank you for purchasing

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 16 Symbols on product packages! Caution, general warning symbol Caution, see documentation

More information

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Irrespective of where the decontamination is performed or who it is performed by the

More information

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope,

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope, Page 1 of 9 Legend 1 Ocular 2 Focussing ring 3 Working channel for flexible instruments 4 Bending protection 5 Deflectable, distal end 6 Lever for control of distal end 7 Fixation thread for pressure compensation

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

Aesculap Implant Systems

Aesculap Implant Systems Aesculap Implant Systems Aesculap Implant Systems Orthopedics USA Instructions for Use/Technical Description Page 1 of 14 Symbols on product packages Caution, general warning symbol Caution, see documentation

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

SERVICE PROCEDURES. Mattress 2 Pipe Overlay or Replacement

SERVICE PROCEDURES. Mattress 2 Pipe Overlay or Replacement Mattress 2 Pipe Overlay or Replacement Contents 2 pipe Mattress Mattress 3 Air Leak Test 4 Air Cell Replacement 5 Air Pipe Connector Replacement 6 Troubleshooting 7 Safety 9 Care & Cleaning 10 Definition

More information

CER-1 / CER-2 (MV-1 / MV-2) (CLM-5001X / CLM-5002VX)

CER-1 / CER-2 (MV-1 / MV-2) (CLM-5001X / CLM-5002VX) CER-1 / CER-2 (MV-1 / MV-2) (CLM-5001X / CLM-5002VX) Endoscope Reprocessor User / Service Manual 2016 Medivators Inc. All rights reserved. Printed in the United States of America The following trademarks

More information

SERRES SUCTION BAG SYSTEM MANUAL

SERRES SUCTION BAG SYSTEM MANUAL SERRES SUCTION BAG SYSTEM MANUAL INSTRUCTIONS FOR USE Please read this manual carefully before installing or using the Serres Suction Bag System. A B C 1 C6 C4 C5 C2 C1 C3 3.1 3.2 1 2 3 3.3 3.4 2.1 2.2

More information

CPAP Washbasin Manual

CPAP Washbasin Manual CPAP Washbasin Manual www.mtts-asia.com CPAP-WB-MAN-01-REV3.02EN Company information MEDICAL TECHNOLOGY TRANSFER AND SERVICES CO., LTD No. 26, Lane 41, An Duong Vuong street, Tay Ho district, Hanoi, Vietnam

More information

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files Cautions Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or with a high chlorine

More information

Requests for Daily Care and Maintenance

Requests for Daily Care and Maintenance Inkjet printer TS500-800 Requests for Daily Care and Maintenance This machine is a precision machine equipped with extremely fine mechanisms. Especially, the nozzle surface of the heads from which the

More information

Requests for Daily Care and Maintenance

Requests for Daily Care and Maintenance Inkjet printer SIJ-320UV Requests for Daily Care and Maintenance This machine is a precision machine equipped with extremely fine mechanisms. Especially, the nozzle surface of the heads from which the

More information

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily.

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily. Suggested daily policy for monitoring the performance of an sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items with the LumCheck Cleaning verification test SUBJECT: Monitoring

More information

Cleaning Instructions for Symbol IP54 Devices

Cleaning Instructions for Symbol IP54 Devices Cleaning Instructions for Symbol IP54 Devices Purpose: The purpose of this document is to demonstrate the proper methods and solutions to clean Symbol mobile computer IP54 devices as well as their respective

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ s 65-201-00-65-202-00 01 Stopcock left 02 Spring Cap 03 Cannula CLEANING / STERILIZATION Immediately after use the instruments have to be soaked in a combined disinfection and cleaning solution in order

More information

Operator s Manual. Histology Bath

Operator s Manual. Histology Bath Operator s Manual Histology Bath 110-827 05.09.12 Table of Contents Introduction... 2 General Safety Information... 2 Safety Recommendations... 3 Unpacking Your Histology Bath...4 Contents...4 Components

More information

THE BASICS OF DECONTAMINATION

THE BASICS OF DECONTAMINATION THE BASICS OF DECONTAMINATION Objectives Review the basic factors that impact cleaning and decontamination of surgical devices Present the fundamentals of manual and automatic cleaning Review various cleaning

More information

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

*smith&nephew VERSAJET HYDROSURGERY SYSTEM Product Manual Manuel D utilisation Bedienungshandbuch Manual Del Producto Manuale Prodotto Producthandleiding *smith&nephew VERSAJET HYDROSURGERY SYSTEM VERSAJET Hydrosurgery System Product Manual Contents

More information

DRAFT. Module 4. Decontamination & Environmental Cleaning

DRAFT. Module 4. Decontamination & Environmental Cleaning Module 4. Decontamination & Environmental Cleaning Learning Objectives 1. Define the 3 steps of decontamination 2. List and describe the process for cleaning and disinfecting 3. Discuss safe management

More information

PJ130S High Pressure Cleaner

PJ130S High Pressure Cleaner PJ130S High Pressure Cleaner Operator Manual Original instructions WARNING: Do not use the machine without reading the instruction sheet. Item Description OVERVIEW 1. High pressure washer 2. On/off switch

More information

PRO 2000i LOW LEVEL DEPOSITOR

PRO 2000i LOW LEVEL DEPOSITOR DEPOSITORS AND AUTOMATED CAKE PRODUCTION SYSTEMS PRO 2000i LOW LEVEL DEPOSITOR OPERATION AND SPARE PARTS MANUAL Serial No. PR2L- (Please quote this number when ordering spares, and making service calls)

More information

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices 71381339 REVA CLEANING AND STERILIZATION BRO.indd 1 5/21/12 3:41 PM 71381339 REVA CLEANING AND STERILIZATION

More information

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization 1. WARNINGS AND PRECAUTIONS 1.1. Caution should be exercised when handling instruments with sharp points, or cutting/drilling

More information

Use and Care Guide. Over the Patient Units and Swing Mount Units

Use and Care Guide. Over the Patient Units and Swing Mount Units Over the Patient Units and Swing Mount Units Table of Contents Page General Information 2 Operation 3-5 Dental Unit Adjustments 6-10 Cleaning, Disinfecting and Sterilization 11-12 Maintenance 13-15 Page

More information

Processing instructions for the Ligamys suture forceps

Processing instructions for the Ligamys suture forceps Processing instructions for the Ligamys suture forceps Table of contents 1 Scope of application 4 2 Warnings and precautions 4 5 3 Restrictions 6 4 Processing instructions 7 4.1 Processing immediately

More information

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS) 1 INTRODUCTION This paper is intended to give general

More information

Enteral Pump Instructions for Use To Assemble When the low

Enteral Pump Instructions for Use To Assemble When the low Enteral Pump Instructions for Use Before initial use on battery power and after extended storage periods, the pump must be plugged into an AC power source for a minimum of 12 hours (pump may be operated

More information

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System This document contains Instructions for Use regarding the processing of the SterilContainer S prior to sterilization

More information

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741

More information

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0 x HANDS-ON MODULE 5 CLEANING FLOW-i V3.0 Before cleaning BEFORE CLEANING CLEANING PROCEDURES OVERVIEW Cleaning procedure when a bacterial/viral filter IS used Cleaning procedure when NO bacterial/viral

More information

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations Model G32-S Model G32-E Disinfection Soak Stations Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482 Website: WWW.CIVCO.COM Copyright

More information

SOLUSCOPE SERIES 3 DESIGNED WITH THE FUTURE IN MIND FEATURES FLEXIBLE ENDOSCOPE DISINFECTOR / STERILISER

SOLUSCOPE SERIES 3 DESIGNED WITH THE FUTURE IN MIND FEATURES FLEXIBLE ENDOSCOPE DISINFECTOR / STERILISER SOLUSCOPE SERIES 3 FLEXIBLE ENDOSCOPE DISINFECTOR / STERILISER FEATURES Sterile Cycle HLD Cycle Continuous Leak Test Channel Flow Monitoring Pre Disinfection Wash Cycle Fast Cycle Time Barcode Tracking

More information

Instructions. Portable Absorb Phlegm. Apparatus

Instructions. Portable Absorb Phlegm. Apparatus Instructions On Portable Absorb Phlegm Apparatus Please carefully read the instructions before attempting to operate this apparatus. Product Features 1. General Portable absorb phlegm apparatus designed

More information

Cleaning and disinfecting

Cleaning and disinfecting Cleaning and disinfecting The goal of infection control is to prevent the transmission of infection to patients/subjects and staff during pulmonary function testing. Cleaning and disinfecting instructions

More information

Harvey Chemiclave 5000 Owner And Operator Guide

Harvey Chemiclave 5000 Owner And Operator Guide Is Happy To Present You With This Free Copy Of The Harvey Chemiclave 5000 Owner And Operator Guide Be sure to look to us when it is time to order parts and supplies for your Chemiclave Please Note: Even

More information

INSTRUCTION MANUAL. Endo-Jet 2000 (EJ-2) PAULDRACH medical GmbH Porschestrasse Garbsen GERMANY

INSTRUCTION MANUAL. Endo-Jet 2000 (EJ-2) PAULDRACH medical GmbH Porschestrasse Garbsen GERMANY INSTRUCTION MANUAL Endo-Jet 2000 (EJ-2) PAULDRACH medical GmbH Porschestrasse 22 30827 Garbsen GERMANY Phone: +49-5131-4965-0 Fax: +49-5131-4965-19 www.pauldrach-medical.de info@pauldrach.de Dear customer,

More information

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018

HOMES AND SENIORS SERVICES. APPROVAL DATE: April 2004 REVISION DATE: March 2016 REVISION DATE: March 2017 REVIEW DATE: November 2018 Page 1 of 6 PURPOSE: All equipment used for resident care will be appropriately cleaned, disinfected and/or sterilized to minimize the risk of transmission of infection by reducing micro-organism contamination

More information

PRE-INSTALLATION PROCEDURE

PRE-INSTALLATION PROCEDURE PRE-INSTALLATION PROCEDURE Required with each BLUPURA Water Treatment System. 1 each External Filter 1 each Chiller 1 each CO 2 Regulator 1 each External Filter 1 each Water Pressure Regulator 1 each 4

More information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information Health Information U S E R G U I D E Home Respiratory Equipment Cleaning Guide This document is intended for Patients and Families for use at home The goal of the cleaning guide is to properly handle,

More information

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

IMPORTANT INFORMATION ON THE DEPUY SYNTHES For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE DEPUY SYNTHES MatrixNEURO RECONSTRUCTION MESH, STERILE MatrixNEURO PREFORMED MESH, STERILE GP2949-B-CAN DESCRIPTION The DePuy Synthes

More information

INSTRUCTIONS SUCTION PUMP. CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician.

INSTRUCTIONS SUCTION PUMP. CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. INSTRUCTIONS SUCTION PUMP KV -5 CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 0086 IMPORTANT Please read this manual carefully before attempting to use the

More information

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes

Operator s Manual. Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Model G14TC-3 Disinfection Soak Station for Transesophageal Ultrasound Probes Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona, IA 52247 USA Tel: 1-800-445-6741 Fax: 1-877-329-2482

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN VETERINARY EQUINE PRACTICE THEATRE AND STERILE PROCEDURES WALLS/TILES - Wash all surfaces from top to bottom using F10CLXD, rinse with - Wash off visible dirt (e.g. blood) using F10CLXD CHAIRS - Trigger spray using F10CL and wipe

More information

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/ HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks 65-227-01-65-240-12 1 Safety and responsibility Before using the instrument

More information

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15 Recommended Care, Cleaning and Sterilization Instructions for Reusable Instruments & Accessories Symmetry Surgical Inc. 3034 Owen Drive, Antioch, TN 37013 USA 1-800-251-3000 Fax: 1-615-964-5566 www.symmetrysurgical.com

More information

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M Operator s Guide Autoclavable Internal Handles with Integrated Paddles 9650-0550 Rev. M The issue date for the Autoclavable Internal Handles with Integrated Paddles Operator s Guide (REF 9650-0550 Rev.

More information

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories Table of Contents 1. Intended Use Statement... 3 2. Proper Use... 3 3. General Information...

More information

- 2 - IMPORTANT SAFETY REMINDERS IT IS IMPORTANT THAT YOU:

- 2 - IMPORTANT SAFETY REMINDERS IT IS IMPORTANT THAT YOU: USER MANUAL IMPORTANT SAFETY REMINDERS This cleaner should only be used for domestic cleaning, as described in this user guide. Please ensure that this guide is fully understood before operating the appliance.

More information

Washer disinfectors. MLC Series. for transfer trolleys, sterile containers and bulky items

Washer disinfectors. MLC Series. for transfer trolleys, sterile containers and bulky items Washer disinfectors MLC Series for transfer trolleys, sterile containers and bulky items MLC Series Washer disinfectors for transport trolleys, sterile containers and bulky items Saving time is a must!

More information

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see  for demo video Keeping you insafehands to avoid needlestick injuries Astek Innovations Ltd Astek House I Atlantic Street I Altrincham Cheshire I WA14 5DH I United Kingdom Telephone +44 (0)161 942 3900 I Fax +44 (0)161

More information

CER-1 & CER-2 Installation & Testing Instructions

CER-1 & CER-2 Installation & Testing Instructions CER-1 & CER-2 Installation & Testing Instructions i Read all installation instructions completely before installing this device. 1 2 3 INTRODUCTION This document details the installation procedure for

More information

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS

PROFESSIONAL USER GUIDELINES FOR PRACTICE INFECTION CONTROL AND HYGIENE IN COMPANION ANIMAL VETERINARY CLINICS/HOSPITALS GENERAL - BEST PRACTICE CLEANING EQUIPMENT Sweep or rinse away loose debris. On floors, use a dust-attracting mop, not a broom Where possible, make use of vacuum (Clean bag daily) Remove spills (feaces,

More information

The Principles, The Process, and The Results:

The Principles, The Process, and The Results: The Principles, The Process, and The Results: Presented by: Fred Alston, CSPDT Eastern Region Clinical Sales and Support Manager April 21, 2017 I am an employee of Healthmark Industries Fraser, Michigan

More information

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times

Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times Rapid Heat Sterilizers with 6, 8, and 12 Minute Sterilization Cycle Times USER MANUAL MODELS: COX 115V-N COX 220V-N COX 115V-R COX 220V-R 2364 Leicester Road, P.O. Box 175, Leicester, NY 14481 Phone (585)

More information