Castle. Model 233HC Vacuum and Gravity Steam Sterilizer for Healthcare Applications. data and specifications

Size: px
Start display at page:

Download "Castle. Model 233HC Vacuum and Gravity Steam Sterilizer for Healthcare Applications. data and specifications"

Transcription

1 Castle Model 233HC Vacuum and Gravity Steam Sterilizer for Healthcare Applications data and specifications PRODUCT A microcomputer-controlled automatic steam sterilizer with six preprogrammed cycles that are operator selectable. The cycles employ either pre-vacuum conditioning by a series of pressure/vacuum pulses or gravity (downward displacement) with positive pulsing conditioning. Both processes dynamically remove air. All cycles are microprocessor controlled and pre-set at the factory. Times and temperatures are adjustable. There are up to 12 cycles, 6 of which are configurable by our service representative during installation. Custom names, up to eight characters in length, can be designated for each configured cycle. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters. APPLICATION For general-purpose pre-vacuum or gravity steam sterilization of hospital supplies, the temperature range is from 230 F to 275 F (110 C to 135 C). The temperature range for liquid cycles is from 219 F to 275 F (104 C to 135 C). Typical applications include wrapped and unwrapped hard goods, gowns or towel packs and liquids in self-venting or unsealed containers. The liquid exhaust is microcomputer controlled for linear and consistent liquid cool-down rate, which is programmable. CHAMBER SIZES Width Height Length Volume 26'' (660 mm) 36'' (914 mm) 37'' (940 mm) 20 cu. ft. (568 L) 26'' (660 mm) 36'' (914 mm) 50'' (1270 mm) 27 cu. ft. (767 L) 26'' (660 mm) 36'' (914 mm) 61'' (1550 mm) 33 cu. ft. (936 L) DOOR OPERATION Manual Power INSTALLATION SELECTIONS Single Door Recessed Cabinet Double Door (50'' and 61'' only) Cabinet, recessed one end Recessed both ends CONTROL PANEL LOCATION Above door (as shown) Remote from unit, vertically mounted OPTIONAL SYSTEMS Utilities Control System. A seven-day timer for programmed startup and shutdown of the sterilizer is provided. The system shuts off water and steam to the unit to conserve energy. Cycles running beyond the programmed shutoff time will be completed. Uninterrupted Power Supply. Provides 115 V power for up to 30 minutes to complete a cycle in process. LOADING EQUIPMENT Loading Car Shall be constructed of tubular stainless steel frame for use as either a flat bed or adjustable shelf type by the addition or removal of sidebars or shelves. Eight adjustable shelving elevations are provided. One fixed bottom shelf and two full-length shelves and two sets of retaining bars are provided with each car. The 26'' (660 mm) wide chamber provides for additional space for a wider shelf, now at a nominal 24'' (610 mm). qty. Transfer Carriage Shall be constructed of tubular stainless steel. Casters shall be of high impact phenolic. Rear casters shall be swivel ball-bearing type, one fitted with a brake. Unit shall contain selfcentering device for alignment to sterilizer and have nylon front bumpers for protection. A locking mechanism shall engage the loading car when on the

2 carriage. A hand or knee lever shall disengage the loading car for transfer into sterilizer. A separate automatic latch mechanism shall engage the carriage to the sterilizer and a hand or knee lever shall disengage the carriage from the sterilizer. qty. Rack & Shelving For all chamber lengths. Two adjustable shelves shall be of corrosion resistant material and shall be capable of sliding out 50% of their length. qty. STANDARDS AND CODES The sterilizer shall comply with or meet the requirements of: ASME Code (Section VIII, Division 1) for Unfired Pressure Vessels Uniform Plumbing Code Intertek Testing Services ETL Listed to UL 544 Intertek Testing Services cetl Listed to CSA C22.2 No. 151 Seismic Anchoring Requirements per California Building Code MICROCOMPUTER CONTROLS Castle Sterilizers employ an Intel 80C-188 microprocessor on a dedicated controller (CPU). The controls consist of an operator switch panel, electronic input/output box, printer, mechanical chamber and jacket pressure gauges. When the supervisor password is enabled, a four-digit code must be entered to change cycle parameters. During a cycle, additional real time cycle information can be displayed by selecting the appropriate display screen. If the supervisor password control is disabled, cycle times and temperatures can be changed directly from the cycle select screen. Note: Selection of time and temperatures other than factory recommendations requires operator verification of the cycle efficacy. Controls are located above the door for convenience. A deflection barrier routes steam vapor and moisture away from the door and controls to assure the temperature of the components inside the control do not exceed their recommended ratings. If specified, the control panel can be located remotely from the sterilizer with up to 32.8 feet (10 m) of cable. The operator interface panel provides for on/off, cycle selection, door seal and unseal and status indicator lights. A two-line backlit LCD display identifies the cycle selected and provides descriptive status messages during the cycle. Audible and visible operator feedback is provided when a selection is made or a fault message is displayed. Temperature can be set, controlled and displayed in degrees Celsius or Fahrenheit, pressure in psia, BAR or kpa. Double door models have one printer and a complete operator interface at both ends of the sterilizer. The temperature of the discharge water is controlled by a temperature device to be less than 140 F (60 C). This switch also conserves water usage. Water in the chamber drain is continuously monitored during a cycle. If excess water is detected and can not be automatically corrected, a high water alarm alerts the operator. An RS 232 port is provided for serial communications for central data collection or remote service analysis. CYCLE DOCUMENTATION Using thermal paper for quiet operation, the printer documents cycle parameters and information. At cycle completion, a cycle performance record is printed. Paper is replaced by a drop in and quick feed method, and the printed strips can be either accumulated on an automatic take-up reel, or torn off for individual cycle storage. A last cycle duplicate print and paper feed switch is provided. The printer is located on the control panel and documents the following on a 200-dpi dot matrix printer (1.88'' wide print width): Process start time and date, total cycle time, sterilizer name and number Cycle parameters selected Cycle transition points, documenting time, temperature and pressure Process fault information messages with time of occurrence Cycle verification signature line CONFIGURABLE CYCLES During installation, our service representative can assign six of the 12 available preloaded cycles by using a software utility tool. Preloaded cycle selections include 5 Vacuum Cycles, 4 Gravity Cycles, 1 Bowie- Dick Test, 1 Leak Test and 1 Liquid Cycle*. Once cycles are assigned for each cycle selection switch, all parameters are held in memory, even if an extended power outage occurs. The factory-assigned cycles are: Two pre-vacuum cycles with factory set values of 3 minutes exposure at 275 F (135 C) and 16 and 3 minutes dry time respectively. Two gravity cycles with factory set values of 30 minutes exposure at 250 F (121 C) with 30 minutes dry time and 10 minute exposure at 275 F (135 C) with 30 minutes dry time. One Bowie-Dick Test Cycle at 273 F (132 C) for 3.5 minutes, with a momentary post vacuum phase to evacuate chamber steam. One vacuum leak test cycle at 268 F (131 C). *The liquid cycle, if used, is not for the sterilization of liquids intended for direct patient contact. PERFORMANCE When installed and connected to specified utility services, the system will provide accurate and repeatable performance. During the timed exposure phase, the temperature will be controlled at the drain by a sensor to 0.9 F (0.5 C) above the set point +/- 0.4 F (0.2 C). Temperature selectivity is in 1 F increments and timing functions are selectable in one-second

3 increments. Temperature control is by a continuous algorithm, a method called Proportional, Integrating, Derivative or PID. CYCLE PROGRESSION Prevac/Wrapped Goods (vacuum/pressure pulse preconditioning) a. Conditioning steam flows into the chamber for a timed period, followed by a series of pressure/ vacuum pulses to remove chamber air. b. Heat-Up to selected temperature. c. Exposure selected chamber temperature is attained and timed. d. Exhaust chamber vented below atmospheric pressure to a transition point. e. Dry a vacuum is created for the duration of the time selected, filtered air is admitted at the end of the drying time, chamber to atmospheric pressure. f. Cycle Complete a tone, light and display message will alert the operator. Gravity/Wrapped Goods (pressure pulse preconditioning) a. Conditioning steam flows into the chamber for a timed period, followed by a series of pressure pulses to remove chamber air. b. Heat-up to selected temperature. c. Exposure selected chamber temperature is attained and timed. d. Exhaust chamber vented to atmospheric pressure. e. Dry filtered air is drawn through chamber for the duration of time selected. (Either Gravity or Vacuum Dry is selectable; Vacuum Dry is recommended.) f. Cycle Complete a tone, display message and light alerts the operator. Gravity/Unwrapped Goods (3 minutes for nonporous items) a. Conditioning steam flows into chamber for a timed period to remove air. The 3-minute Flash cycle for nonporous items has a series of positive pulses for dynamic air removal. b. Heat-up to selected temperature. c. Exposure selected chamber temperature is attained and timed. d. Exhaust chamber vented to atmospheric pressure. e. Dry filtered air is drawn through chamber for the duration of time selected. f. Cycle Complete a tone, light and display message will alert the operator. Liquids gravity cycle with microcomputer-controlled linear exhaust. CONSTRUCTION Body shall be constructed of an inner shell, outer shell (jacket), backhead and monel gasket ring. Chamber shells are 0.250'' (6.35 mm) thick hot rolled carbon steel with an interior surface of nickel or monel clad, 0.025'' (0.635 mm) thick. All pressure vessel construction is per ASME code requirements for working pressures up to 45 psig (310 kpa). The monel gasket ring holds a continuous, one-piece silicone O-ring gasket, 1'' (25.4 mm) in diameter. Gasket is replaceable without tools. The body assembly is thermally insulated with foil-backed fiberglass insulation, and a steam baffle is provided to prevent condensation from wetting the load. A spare threaded opening permits passage of thermocouple leads to monitor interior and load temperatures. Door is nickel-clad steel 0.50'' (12.7 mm) thick with tensile strength of 70,000-85,000 psi ( MPa) and is dished for maximum strength. Locking mechanism consists of eight heavy steel clamps, which unseal the door before allowing it to open. The closure uses filament-wound PTFE bearings at all high-load points and is totally lubrication-free. A series of interlocking solid-state door switches and relays stop steam from entering the chamber with the door unsealed, and prevent door opening with pressure inside. DOOR OPERATION Power Door is operated by permanently lubricated gearmotors and controlled by solid-state components for reliability and safety. On cycle complete the door unseals to vent excess steam. Door design minimizes possibility of injury to personnel or damage to the mechanism if an obstruction is encountered on door opening. The door can be opened manually in case of an emergency. Manual Door sealing is accomplished by handactuated air cylinders. Two handles are provided for opening and closing the door. Solid-state switches confirm that the door is properly sealed. An auxiliary means of manual control is provided in case of emergency. PANELING Front panels shall be constructed of nominal 0.050'' (1.27 mm) thick 300 series stainless steel. Trim panels are built in to provide fit-up with recessing wall or optional cabinet. Complete door assembly and drive mechanisms are enclosed by removable stainless steel panels eliminating projections and making the unit easier to clean. The door is lined with 1'' (25.4 mm) foilbacked fiberglass insulation. WARRANTY Getinge/Castle warrants that each sterilizer is carefully tested, inspected and leaves the factory in proper working condition, free from visible defects. Sterilizers are warranted for one year from the start of the warranty, including parts and labor (excluding expendable parts). The ASME pressure vessel is further warranted to the original owner against structural failure for a period of 15 years from the date of initial operation. See our warranty pamphlet for complete details.

4

5

6 NOTICE Work by others Safe and efficient operation of this product is dependent upon the owner/user providing the services specified herein as well as any other normally accepted electrical, mechanical or plumbing interface between user s supply and this product. Getinge/Castle will not assume responsibility for problems that result from non-compliance with the above conditions. The following conditions and services are required by Getinge/Castle equipment and are to be provided by others. Service On unit Pipe size Pressure Range Flow rates max. connection To unit Dynamic At unit S1 = house steam 1-1/4'' [32] NPT psig 300 lbs./hr 1'' [25] NPT female [ kg/cm 2 ] [136 kg/hr] see note 4 S2 = clean steam 1'' [25] NPT psig 200 lbs./hr (Clean Steam option only) [ kg/cm 2 ] [136 kg/hr] 3/4'' [19] NPT female CW = cold water 1-1/4'' [32] NPT psig 11 gpm 1'' [25] NPT female [ kg/cm 2 ] [2.5 m 3 /Hr] see note 1 D = drain See note 2 Not applicable See note 2 2'' [51] ODT SV1 = Sterilizer vessel See note 6 Not applicable See note 6 Pressure relief valve vent 1'' [25] NPT female A = air 1/2'' [13] NPT psig <1 cfm 1/2'' [13] NPT female [ kg/cm 2 ] (Manual Door only) see note 7 SV2 = Sterilizer vessel See note 6 Not applicable See note 6 Pressure relief valve vent (Clean Steam option only) 1-1/4'' [32] NPT female It shall be the customer s responsibility to complete all electrical connections using properly sized wiring in accordance with the National Electrical Code and all applicable local codes. See the Getinge/Castle Installation Manual for specific instructions. Water and steam line shutoff valves and a disconnect switch must be provided at each sterilizer. Electrical Supply A dedicated supply with conveniently accessible disconnect switch is required for each sterilizer service listed below. Service Utility Max. current (amps) Consumption E = Power box 115V 50/60Hz 1PH 12 Amps 7 W/Hr 1/2'' conduit [13] (15 amp branch circuit) K = Junction box 240V 60Hz 3PH 7.8 Amp N/A 1/2'' conduit [13] 480V 60Hz 3PH 3.9 Amp Vacuum Pump Units (233LS) only

7 Notes To Architects & Contractors 1) Cold water: a) Cold-water quality use potable water with a hardness of grains/gal [8-170 ppm], and a temperature of less than 80 F. b) An optional water booster pump is available: For installations where water pressure is at least 20 psig [1.4 kg/cm 2 ] dynamic but less than 40 psig [2.8 kg/cm 2 ]. If a reduced pressure principle device is required by local code, which lowers the water pressure below the minimum specified. c) The optional booster pump requires mechanical electrical and plumbing hook-up by customer. A separate electrical service to the water booster pump junction box is necessary. Contact your local Getinge/Castle representative for Utility Data. 2) It shall be the customer s responsibility to provide a proper drainage system in accordance with applicable local codes. Temperature of drain water will not exceed 140 F [60 C] under normal operating conditions. If cold water supply is cut-off, temperature may exceed 200 F [93 C]. 3) It shall be the customer s responsibility to install the transformer that provides the control system with 120VAC. See transformer/universal export voltage option for available primary voltage configuration (one transformer is used with multiple input voltage capabilities). The transformer enclosure has an overall size of 10'' long x 7-1/4'' wide x 5-3/8'' deep [254 x 184 x 137]. 4) It shall be the customer s responsibility to provide condensate free steam between 97% and 100% saturated vapor. 5) Install the remotely located control panel; see drawing no. HS4053. Cable length from sterilizer to a remotely mounted control panel shall not exceed 32 feet [9754]. 6) Getinge/Castle recommends piping all chamber relief valves to a vented manifold outside the equipment service area. Caution must be exercised not to reduce the discharge capacity of the relief valve. Recommended piping practices for relief valve piping can be found in ASME Boiler and Pressure Vessel Code Section VIII, Div. 1, para. UG-135. Check local codes for special requirements. 7) Air compressor package roughing in drawing P/N HS ) Sealing Flange roughing in drawing P/N HS Weight and Crated Size Max. Weight lb [kg] Dimensions Unit Size Crated Uncrated Length Width Height 37'' [940] SD 2373 [1079] 2173 [988] 7'-10'' [2388] 4'-0'' [1219] 6'-10'' [2083] Single Door 2386 [1085] 2186 [994] 50'' [1270] 8'-10'' [2892] 4'-0'' [1219] 6'-10'' [2083] Double Door 2732 [1242] 2532 [1151] Single Door 2684 [1220] 2484 [1129] 61'' [1549] 9'-10'' [2997] 4'-0'' [1219] 6'-10'' [2083] Double Door 3019 [1372] 2819 [1281] Note: Assembled Sterilizer uncrated and off skid has max. dimensions of 3'-4'' wide x 5'-11 1/2'' height [1016 x 1816].

8 Getinge/Castle, Inc East Henrietta Road Rochester, New York U.S.A. Telephone (716) Facsimile (716) DS233HC 0101 Castle is a registered trademark. info@getingecastle.com Printed in U.S.A.

Sterilizer Model LS2020 Series Beta Star Life Science Products of R-V Industries

Sterilizer Model LS2020 Series Beta Star Life Science Products of R-V Industries SECTION 11630 - STERILIZERS PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1.2 REFERENCES 1. Small Steam Sterilizer A. The following industry, association and government codes and standards are cited

More information

AMSCO REMANUFACTURED MEDIUM STERILIZERS

AMSCO REMANUFACTURED MEDIUM STERILIZERS AMSCO REMANUFACTURED MEDIUM STERILIZERS APPLICATION Two configurations are available for sterilization of certain materials used in hospitals or laboratories. Gravity - designed for sterilization of liquids,

More information

AMSCO EVOLUTION STEAM STERILIZERS - NORTH AMERICA

AMSCO EVOLUTION STEAM STERILIZERS - NORTH AMERICA AMSCO EVOLUTION STEAM STERILIZERS - NORTH AMERICA APPLICATION This sterilizer is configured for prevacuum sterilization of heat- and moisture-stable materials used in healthcare facilities. These units

More information

BASIL 6000 TUNNEL CAGE WASHER

BASIL 6000 TUNNEL CAGE WASHER BASIL 6000 TUNNEL CAGE WASHER APPLICATION For thorough, efficient cleaning of cages, debris pans, bottles, feeder bowls, and miscellaneous items used in the care of laboratory animals. DESCRIPTION The

More information

Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION

Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION 5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512 616.656.2900 Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION The SKYTRON Integrity 270 sterilizers utilize gravity displacement

More information

SMALL STERILIZER SERIES 26 x 26 MODEL

SMALL STERILIZER SERIES 26 x 26 MODEL BETA STAR LIFE SCIENCE EQUIPMENT SMALL STERILIZER SERIES 26 x 26 MODEL ECO-FRIENDLY Reduce your facility s water footprint with Beta Star s EnviroVac Vacuum System. Mechanical and programmable conservation

More information

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 37.5" NORTH AMERICA

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 37.5 NORTH AMERICA AMSCO CENTURY MEDIUM STEAM STERILIZER 6 X 37.5" NORTH AMERICA APPLICATION Two configurations are available for sterilization of heat- and moisture-stable materials used in healthcare facilities: Prevacuum

More information

AMSCO CENTURY SMALL STERILIZERS DOMESTIC

AMSCO CENTURY SMALL STERILIZERS DOMESTIC AMSCO CENTURY SMALL STERILIZERS DOMESTIC APPLICATION Amsco Century Small Sterilizers are designed for sterilization of materials used in healthcare facilities. They are available in two configurations:

More information

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 26" (SCIENTIFIC)

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 26 (SCIENTIFIC) APPLICATION This sterilizer is offered in a prevacuum configuration and is designed for fast, efficient sterilization of heat- and moisturestable materials in scientific applications. The prevacuum configuration

More information

Control Operating Costs. Gain Process Efficiency.

Control Operating Costs. Gain Process Efficiency. S t e a m s t e r i l i z e r s Control Operating Costs. Gain Process Efficiency. Amsco Century Small and Amsco Millennium Steam Sterilizers Amsco Century Small and Amsco Millennium Steam Sterilizers The

More information

RHP120 RHP300 Steam Boiler Series

RHP120 RHP300 Steam Boiler Series Electra Steam, Inc. Phone: 540-662-3811 4407 Martinsburg Pike Fax: 540-665-8101 Clear Brook, VA 22624 email: Sales@reimersinc.com USA web: www.reimersinc.com HEATING POWER kw RHP120 RHP300 Steam Boiler

More information

Unfired Steam Generators (Series RB) Installation, Operation and Maintenance

Unfired Steam Generators (Series RB) Installation, Operation and Maintenance Unfired Steam Generators (Series RB) Installation, Operation and Maintenance Horizontal USG Vertical USG 534-EN Please read and save these instructions Disclaimers This Installation, Operation, and Maintenance

More information

LSII /51 STERILIZER MACHINE SPECIFICATION A REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH

LSII /51 STERILIZER MACHINE SPECIFICATION A REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH LSII-262639/51 STERILIZER MACHINE SPECIFICATION A-83-37630 REV DATE REVISION BY CHK APP 0 5/18/2016 RELEASED FOR INFORMATION JPK GBJ KWH Table of Contents 1.0 PART I GENERAL... 3 1.1 Vessel Volume... 3

More information

Project Name: Location: Engineer: Contractor: I. GENERAL

Project Name: Location: Engineer: Contractor: I. GENERAL Part I - General Part II - Product Part III - Installation Project Name: Location: Engineer: Contractor: I. GENERAL I. PRODUCT Rep: Date: A. Supply and install modulating and condensing boiler(s) as specified

More information

GIRTON CABINET WASHER MODEL 794A

GIRTON CABINET WASHER MODEL 794A DESCRIPTION GIRTON CABINET WASHER MODEL 794A The washer shall be a Girton Model 794A Cabinet Type Pressure Washer constructed of type-304 stainless steel, utilizing a reciprocating header spray system.

More information

Installation, Operation and Maintenance

Installation, Operation and Maintenance Bulletin No. 534 Installation, Operation and Maintenance Unfired Steam Generators (Series RB) Horizontal USG Vertical USG Disclaimers This Installation, Operation, and Maintenance Manual is intended to

More information

Suggested Specification: Benchmark 1.5 Low NOx Gas-Fired Hydronic Boiler(s)

Suggested Specification: Benchmark 1.5 Low NOx Gas-Fired Hydronic Boiler(s) Furnish and install as shown on AERCO International plans and operation and maintenance manuals, with all applicable codes and authorities having local, state and federal jurisdiction, Benchmark Series

More information

C. ASME Compliance: Fabricate and label water chiller heat exchangers to comply with ASME Boiler and Pressure Vessel Code: Section VIII, Division 1.

C. ASME Compliance: Fabricate and label water chiller heat exchangers to comply with ASME Boiler and Pressure Vessel Code: Section VIII, Division 1. SECTION 236426 - ROTARY-SCREW WATER CHILLERS PART 1 - GENERAL 1.1 SUMMARY A. This Section includes packaged, water cooled or air cooled as scheduled, electric-motor-driven, rotary-screw water chillers

More information

Hubbell NX500 High Efficiency Gas Condensing Water Heater

Hubbell NX500 High Efficiency Gas Condensing Water Heater Hubbell NX500 High Efficiency Gas Condensing Water Heater Suggested Specification Part I - General Part II - Product Part III - Installation Project Name: Location: Engineer: Contractor: I. GENERAL Rep:

More information

"ESDDR" Series Electric Humidifier

ESDDR Series Electric Humidifier ESDDR-1 Deionized, Demineralized, or Reverse Osmosis Water "ESDDR" Series Electric Humidifier The Electric Humidifier from PURE Humidifier Co. is loaded with features and options. All you need is deionized,

More information

AZTECA A LITRE STERILISER

AZTECA A LITRE STERILISER Product Description Azteca A-6615 is a fully automatic high-speed pre & post vacuum steam sterilizer. This model is an electrically heated sterilizer, which operates with saturated steam as a sterilizing

More information

WATER-TUBE POOL HEATERS

WATER-TUBE POOL HEATERS XTHERM, TYPE P - MODELS 1005A-2005A SUGGESTED SPECIFICATIONS Catalog No.: 6000.63B Effective: 3-01-15 Replaces: 9-01-09 WATER-TUBE POOL HEATERS PART 1 - GENERAL 1.1 SUMMARY A. Section includes condensing,

More information

INDIRECT FIRED, SEMI-INSTANTANEOUS STEAM & BOILER WATER HEATER. Compact Design. High Recovery. Energy Efficient. Stainless Alloy Construction

INDIRECT FIRED, SEMI-INSTANTANEOUS STEAM & BOILER WATER HEATER. Compact Design. High Recovery. Energy Efficient. Stainless Alloy Construction INDIRECT FIRED, SEMI-INSTANTANEOUS STEAM & BOILER WATER HEATER Compact Design High Recovery Energy Efficient Stainless Alloy Construction Factory Packaged Integral Safety System ASME Constructed & Stamped

More information

DIVISION FINNED WATER-TUBE WATER HEATERS

DIVISION FINNED WATER-TUBE WATER HEATERS HI DELTA, TYPE WH - MODELS 992C-2342C SUGGESTED SPECIFICATIONS Catalog No.: 3500.932A Effective: 04-01-17 Replaces: 10-31-16 DIVISION 23 52 33.13 FINNED WATER-TUBE WATER HEATERS - GENERAL 1.1 SUMMARY A.

More information

WATER-TUBE POOL HEATERS

WATER-TUBE POOL HEATERS XTHERM, TYPE P - MODELS 2505-4005 SUGGESTED SPECIFICATIONS Catalog No.: 6000.70 Effective: 09-15-16 Replaces: NEW WATER-TUBE POOL HEATERS - GENERAL 1.1 SUMMARY A. Section includes condensing, gas-fired

More information

AMSCO CENTURY MEDIUM STEAM STERILIZER 26" X 37.5" (SCIENTIFIC)

AMSCO CENTURY MEDIUM STEAM STERILIZER 26 X 37.5 (SCIENTIFIC) AMSCO CENTURY MEDIUM STEAM STERILIZER 26" X 37.5" (SCIENTIFIC) APPLICATION This sterilizer is offered in a prevacuum configuration and is designed for fast, efficient sterilization of heat- and moisturestable

More information

DIVISION FINNED WATER-TUBE WATER HEATERS

DIVISION FINNED WATER-TUBE WATER HEATERS DELTA LIMITED, TYPE WH - MODELS 399B-899B SUGGESTED SPECIFICATIONS Catalog No.: 3500.97B Effective: 04-01-17 Replaces: 11-15-08 DIVISION 23 52 33.13 FINNED WATER-TUBE WATER HEATERS - GENERAL 1.1 SUMMARY

More information

Revitalize Building Mechanical Systems (4619)

Revitalize Building Mechanical Systems (4619) SECTION 235216 FIRE-TUBE CONDENSING BOILERS PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 01 Specification

More information

Open the Door to Increased Productivity and Cost Efficiency

Open the Door to Increased Productivity and Cost Efficiency S t e a m s t e r i l i z e r s Open the Door to Increased Productivity and Cost Efficiency Amsco Century Medium Steam Sterilizers Amsco Century Medium Steam Sterilizers The Family of Sterilizers Designed

More information

Exclusively published and distributed by Architectural Computer Services, Inc. (ARCOM) for the AIA

Exclusively published and distributed by Architectural Computer Services, Inc. (ARCOM) for the AIA Page 236423-1 Copyright 2009 by The American Institute of Architects (AIA) Exclusively published and distributed by Architectural Computer Services, Inc. (ARCOM) for the AIA Modified by MSU Physical Plant

More information

XTHERM, TYPE WH - MODELS 1005A- 2005A DIVISION COMMERCIAL, WATER-TUBE, CONDENSING GAS DOMESTIC WATER HEATERS

XTHERM, TYPE WH - MODELS 1005A- 2005A DIVISION COMMERCIAL, WATER-TUBE, CONDENSING GAS DOMESTIC WATER HEATERS XTHERM, TYPE WH - MODELS 1005A- 2005A Catalog No.: 3500.801 Effective: 11-15-14 Replaces: NEW PART 1 - GENERAL 1.1 SUMMARY DIVISION 23 34 36.29 COMMERCIAL, WATER-TUBE, CONDENSING GAS DOMESTIC WATER HEATERS

More information

Trinity Tft High Efficiency Gas Condensing Boiler & Water Heater

Trinity Tft High Efficiency Gas Condensing Boiler & Water Heater Part I - General Part II - Product Part III - Installation Project Name: Location: Engineer: Contractor: I. GENERAL Rep: A. Supply and install modulating and condensing boiler(s) as specified herein. Date:

More information

MARK-COSTELLO WASTE STERILIZER SPECIFICATION SHEET

MARK-COSTELLO WASTE STERILIZER SPECIFICATION SHEET MARK-COSTELLO WASTE STERILIZER SPECIFICATION SHEET The Mark-Costello Waste Sterilzer is designed and manufactured in accordance with the following specifications: Vessel shall be built in accordance with

More information

AUTOCLAVE STEAM STERILIZERS (BENCH TOP)

AUTOCLAVE STEAM STERILIZERS (BENCH TOP) AUTOCLAVE STEAM STERILIZERS (BENCH TOP) Image: HS-1606VD Hanshin Bench Top Steam Sterilizer Small and easy to operate Three preset sterilization cycles Adjustable cycle presets Rapid heat up times Option

More information

Trinity Lx200 High Efficiency Gas Condensing Boiler & Water Heater

Trinity Lx200 High Efficiency Gas Condensing Boiler & Water Heater Part I - General Part II - Product Part III - Installation Project Name: Location: Engineer: Contractor: I. GENERAL Rep: Date: A. Supply and install modulating and condensing boiler(s)/water heater(s)

More information

RELIANCE 400 LABORATORY GLASSWARE WASHER

RELIANCE 400 LABORATORY GLASSWARE WASHER RELIANCE 400 LABORATORY GLASSWARE WASHER APPLICATION The Reliance 400 Laboratory Glassware Washer is designed for thorough cleaning of laboratory glassware, plastic and metal goods used in research, production

More information

ASME Packaged Electric Water Heater

ASME Packaged Electric Water Heater MODEL SH & H ASME Packaged Electric Water Heater 15-1600 KW - All Voltages & Phases, 80-5000 Gallon Capacity Features Reliable Packaged System Only high grade materials used in construction to ensure long

More information

Packaged Instantaneous Circulation Heater Full KW Selection In All Voltages, Single or Three Phase

Packaged Instantaneous Circulation Heater Full KW Selection In All Voltages, Single or Three Phase Features Heavy Duty Construction Constructed with high grade materials to ensure long operating life Turn-Key package is simple to specify and operate Factory wired electrical controls provide trouble-free

More information

FINNED WATER-TUBE POOL HEATERS

FINNED WATER-TUBE POOL HEATERS HI DELTA, TYPE P - MODELS 502C-902C SUGGESTED SPECIFICATIONS Catalog No.: 6000.601A Effective: 07-12-18 Replaces: 10-31-16 FINNED WATER-TUBE POOL HEATERS - GENERAL 1.1 SUMMARY A. Section includes gas-fired,

More information

SMALL STERILIZER SERIES LSII MODEL (20 x 20 )

SMALL STERILIZER SERIES LSII MODEL (20 x 20 ) BETA STAR LIFE SCIENCE EQUIPMENT SMALL STERILIZER SERIES LSII MODEL (20 x 20 ) ECO-FRIENDLY Reduce your facility s water footprint with Beta Star s EnviroVac Vacuum System. Mechanical and programmable

More information

Trinity Lx400 High Efficiency Gas Condensing Boiler & Water Heater

Trinity Lx400 High Efficiency Gas Condensing Boiler & Water Heater Part I - General Part II - Product Part III - Installation Project Name: Location: Engineer: Contractor: I. GENERAL Rep: Date: A. Supply and install modulating and condensing boiler(s)/water heater(s)

More information

High Capacity Instantaneous Steam Fired Water Heater

High Capacity Instantaneous Steam Fired Water Heater MODEL F High Capacity Instantaneous Steam Fired Water Heater Features Compact High Capacity Design Packaged System Water heater is factory assembled on an all welded steel mounting frame to save time and

More information

FINNED WATER-TUBE POOL HEATERS

FINNED WATER-TUBE POOL HEATERS HI DELTA, TYPE P - MODELS 992C-2342C SUGGESTED SPECIFICATIONS Catalog No.: 6000.611 Effective: 10-31-16 Replaces: NEW FINNED WATER-TUBE POOL HEATERS - GENERAL 1.1 SUMMARY A. Section includes gas-fired,

More information

CO 2. Incubators. Incubator, the first microprocessor-based system, and the first menu-driven control system, but we offer the

CO 2. Incubators. Incubator, the first microprocessor-based system, and the first menu-driven control system, but we offer the Radiant Warm Wall Compact Incubator Reduced Oxygen Incubator Automatic Water-Jacketed Incubator, stacked For more than 45 years, Lab-Line has been on the cutting edge of incubator development. Not only

More information

AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM (For Use Outside U.S.A. Only)

AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM (For Use Outside U.S.A. Only) AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM (For Use Outside U.S.A. Only) APPLICATION The Amsco V-PRO max Low Temperature Sterilization System is intended for use in the terminal sterilization

More information

PROVIDES TEPID WATER FOR EMERGENCY SHOWER

PROVIDES TEPID WATER FOR EMERGENCY SHOWER PROVIDES TEPID WATER FOR EMERGENCY SHOWER Model EMV FEATURES Heavy Duty Construction Hydrastone cement lining provides tank longevity Copper-silicon alloy tappings cannot rust or corrode High impact composite

More information

BASIL 6700 TUNNEL CAGE WASHER

BASIL 6700 TUNNEL CAGE WASHER BASIL 6700 TUNNEL CAGE WASHER APPLICATION For thorough, efficient cleaning of cages, debris pans, bottles, feeder bowls and miscellaneous items used in the care of laboratory animals. DESCRIPTION The Basil

More information

DIVISION COMMERCIAL, GRID-TYPE, FINNED-TUBE, GAS DOMESTIC WATER HEATERS

DIVISION COMMERCIAL, GRID-TYPE, FINNED-TUBE, GAS DOMESTIC WATER HEATERS MVB, TYPE WH - MODELS 503A-2003A SUGGESTED SPECIFICATIONS Catalog No.: 3500.952C Effective: 06-25-15 Replaces: 12-22-09 DIVISION 23 34 36.29 COMMERCIAL, GRID-TYPE, FINNED-TUBE, GAS DOMESTIC WATER HEATERS

More information

SECTION STEAM CONDENSATE PUMPS

SECTION STEAM CONDENSATE PUMPS PART 1 - GENERAL 1.1 DESCRIPTION SECTION 23 22 23 STEAM CONDENSATE PUMPS SPEC WRITER NOTES: 1. Delete between // ---- // if not applicable to project. Also delete any other item or paragraph not applicable

More information

STEAM STERILIZER SERIE S1000 (RBE)

STEAM STERILIZER SERIE S1000 (RBE) STEAM STERILIZER SERIE S1000 (RBE) INTRODUCTION: Of Turkish manufacturing, our sterilizers Series S1000 (RBE) are designed for the purpose of inactivating and sterilizing infectious waste generated in

More information

DIVISION FINNED WATER-TUBE BOILERS

DIVISION FINNED WATER-TUBE BOILERS MVB, TYPE H - MODELS 504A-2004A SUGGESTED SPECIFICATIONS Catalog No.: 2000.933C Effective: 06-25-15 Replaces: 12-21-09 DIVISION 23 52 33.13 FINNED WATER-TUBE BOILERS PART 1 - GENERAL 1.1 SUMMARY A. Section

More information

AMSCO C SERIES SMALL STEAM STERILIZERS DOMESTIC

AMSCO C SERIES SMALL STEAM STERILIZERS DOMESTIC AMSCO C SERIES SMALL STEAM STERILIZERS DOMESTIC APPLICATION AMSCO C Series Certified Pre-owned Small Steam Sterilizers are designed for sterilization of materials used in healthcare facilities. The sterilizers

More information

Commercial Semi-Instantaneous Water Heaters

Commercial Semi-Instantaneous Water Heaters FEATURES HWI Models A. O. Smith s HWI models are available for operation with steam or boiler water as the energy source. They are factory assembled with components sized, piped and checked at the factory

More information

BURNHAM COMMERCIAL 27 February 2008 MPC Multiple Pass Commercial Boiler Suggested Specifications

BURNHAM COMMERCIAL 27 February 2008 MPC Multiple Pass Commercial Boiler Suggested Specifications 2.1 GENERAL A. Reference is made to the clause entitled Buy American Act Balance of Payments Program Construction Materials FAR 52.225-9. Notwithstanding a bidder s right to offer identifiable foreign

More information

Platinous. temperature. and humidity. chambers. Top Quality Design. With an Easy-to-use. Touch-Screen. Programmer

Platinous. temperature. and humidity. chambers. Top Quality Design. With an Easy-to-use. Touch-Screen. Programmer Platinous temperature and humidity chambers Top Quality Design With an Easy-to-use Touch-Screen Programmer Platinous chambers are innovative, built with the busy user in mind. If you have ever used an

More information

ANSUL. Twin-Agent Fire Fighting/Securing Systems Specifications. Models 450/50-B, 450/100-B, 900/100-B INDEX SPECIFICATIONS

ANSUL. Twin-Agent Fire Fighting/Securing Systems Specifications. Models 450/50-B, 450/100-B, 900/100-B INDEX SPECIFICATIONS ANSUL Twin-Agent Fire Fighting/Securing Systems Specifications Models 450/50-B, 450/100-B, 900/100-B INDEX 1.0 GENERAL 1.1 References 1.2 Submittals 1.3 System Capabilities 1.4 Quality Control 1.5 Warranty,

More information

: 1.5kw/AC220V.50Hz (AC110V 60Hz) : 340*200*30mm (3piece) 400*200*30mm (3 piece)

: 1.5kw/AC220V.50Hz (AC110V 60Hz) : 340*200*30mm (3piece) 400*200*30mm (3 piece) TABLETOP (Flash) STEAM STERILIZER Semi Automatic Model Model : TTF20L / TTF25L Valid volume : 20L (φ250*420mm), 25L (φ250*520mm) Max. Working pressure : 0.22Mpa Max.Working temperature : 134C Adjustment

More information

YOUR PROCESS DEMANDS THE MOST DEPENDABLE TEMPERATURE CONTROL UNIT AVAILABLE.

YOUR PROCESS DEMANDS THE MOST DEPENDABLE TEMPERATURE CONTROL UNIT AVAILABLE. Temperature Control Units Water & Oil 0-00 F Portable Chillers 0 - F Central Chillers Packages & Modules 0 - F Pump Tank Stations Chilled or Tower Water 00-0 gallons Cooling Tower Cells 4-40 tons Filters

More information

Packaged Steam Fired Water Heater

Packaged Steam Fired Water Heater MODEL ST Packaged Steam Fired Water Heater Utilizes Steam To Heat Domestic Water, Semi-Instantaneous or Storage Type Heaters Features Reliable Packaged System Only high grade materials used in construction

More information

CNA-Series - Large Industrial Heaters Tankless Water Heating Solutions

CNA-Series - Large Industrial Heaters Tankless Water Heating Solutions 36-144 kw (122,800-491,300 BTUs) Certified Lead-Free Design New & Improved Pressure Drop Advantage Variable Temp Heat Exchanger NEMA 4 enclosure standard Independent Safeties ETL and cetl certified to

More information

Industrial Freezer Sales A division of IDS 5311 Derry Avenue Building D Agoura Hills CA Phone (818) Fax (818)

Industrial Freezer Sales A division of IDS 5311 Derry Avenue Building D Agoura Hills CA Phone (818) Fax (818) UCUF191S-86, UCUF241S-86 and UCUF281S-86 UPRIGHT ULTRACOLD FREEZERS Control and Alarm Features Manual defrost, adjustable control range from -50 C to -86 C. UL/CUL Listed. Factory preset at -80 C. Advanced

More information

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION IBAS LIGHTING CONTROL

DENVER PUBLIC SCHOOLS DESIGN AND CONSTRUCTION STANDARDS This Standard is for guidance only. SECTION IBAS LIGHTING CONTROL PART 0 A/E INSTRUCTIONS 0.01 DESIGN REQUIREMENTS A. Minimum lighting control (base scope of work) shall include all corridors. The lighting in all the corridors shall be turned on (if not already on) if

More information

DIVISION FINNED WATER-TUBE BOILERS

DIVISION FINNED WATER-TUBE BOILERS MVB, TYPE H - MODELS 503A-2003A SUGGESTED SPECIFICATIONS Catalog No.: 2000.932C Effective: 04-01-14 Replaces: 07-15-13 DIVISION 23 52 33.13 FINNED WATER-TUBE BOILERS PART 1 - GENERAL 1.1 SUMMARY A. Section

More information

SECTION PLUMBING EQUIPMENT

SECTION PLUMBING EQUIPMENT PART 1 GENERAL 1.01 SECTION INCLUDES A. Water heaters. B. Packaged water heating systems. C. Domestic water heat exchangers. D. Water storage tanks. E. Water softeners. F. Reverse osmosis equipment. G.

More information

AM SELECT DISHWASHER STANDARD FEATURES MODEL OPTIONS AT EXTRA COST ACCESSORIES VOLTAGE. Item # Quantity C.S.I. Section AM15

AM SELECT DISHWASHER STANDARD FEATURES MODEL OPTIONS AT EXTRA COST ACCESSORIES VOLTAGE. Item # Quantity C.S.I. Section AM15 AM SELECT DISHWASHER Item # Quantity C.S.I. Section 11400 AM SELECT DISHWASHER STANDARD FEATURES.74 gallons per rack final rinse water 58 racks per hour hot water sanitizing 65 racks per hour chemical

More information

SECTION COMPRESSED AIR SYSTEM. A. Pipe and pipe fittings, including valves, unions and couplings.

SECTION COMPRESSED AIR SYSTEM. A. Pipe and pipe fittings, including valves, unions and couplings. SECTION 15481 COMPRESSED AIR SYSTEM PART 1 - GENERAL 1.1 SECTION INCLUDES A. Pipe and pipe fittings, including valves, unions and couplings. B. Air compressor. C. After cooler. D. Refrigerated air dryer.

More information

STORAGE TANK INSTALLATION and OPERATION MANUAL

STORAGE TANK INSTALLATION and OPERATION MANUAL LST-I-O 100161529_2000014416 Rev F STORAGE TANK INSTALLATION and OPERATION MANUAL LOW LEAD CONTENT The information contained in this manual is intended for use by qualified professional installers, or

More information

High Pressure Conductivity Monitor

High Pressure Conductivity Monitor Product Data Sheet PDS 71-EDS880/1/rev.A August 2006 Model EDS880/1 High Pressure Conductivity Monitor COMPLETE SYSTEM FOR THE DETERMINATION OF CONDUCTIVITY IN HIGH PRESSURE SYSTEMS SENSOR INSTALLS DIRECTLY

More information

Product Data. Features/Benefits. 19XB Positive Pressure Storage System

Product Data. Features/Benefits. 19XB Positive Pressure Storage System Product Data 19XB Positive Pressure Storage System 19XB028 Copyright 1992 Carrier Corporation 19XB052 Carrier s Positive Pressure Storage (PPS) System ensures the critical conservation of refrigerants

More information

SECTION PLUMBING EQUIPMENT

SECTION PLUMBING EQUIPMENT PART 1 GENERAL 1.1 SECTION INCLUDES A. Water heaters B. Packaged water heating systems C. Water storage tanks D. Water softeners E. Pumps F. Circulators 1.2 REFERENCES SECTION 22 30 00 PLUMBING EQUIPMENT

More information

Portable CNA-Series - Large Industrial Heaters Tankless Water Heating Solutions

Portable CNA-Series - Large Industrial Heaters Tankless Water Heating Solutions Portable CNA-Series - Large Industrial Heaters Heavy duty industrial cart options simplify moving heater to various locations 36-72 kw (122,800-245,600 BTUs) Certified Lead-Free Design New & Improved Pressure

More information

Twin-Agent Fire Fighting/Securing Systems Specifications

Twin-Agent Fire Fighting/Securing Systems Specifications Twin-Agent Fire Fighting/Securing Systems Specifications Models 450/50, 450/100, 900/100 & 1350/200 INDEX 1.0 GENERAL 1.1 References 1.2 Submittals 1.3 System Capabilities 1.4 Quality Control 1.5 Warranty,

More information

SNA-SKID Series - Preassembled Safety Shower Heater Skid Tankless Water Heating Solutions

SNA-SKID Series - Preassembled Safety Shower Heater Skid Tankless Water Heating Solutions Pre-assembled skid system Back-to-Back heater models, SNA & SNAR High Flow Demand Safety Shower Applications 216-288 kw (737,000-983,000 BTUs) Temperature overshoot purge system NEMA 4 enclosure standard

More information

S19361F - Pre-Assembled Freeze-Protected Safety Shower Heater Skid Tankless Water Heating Safety Solutions

S19361F - Pre-Assembled Freeze-Protected Safety Shower Heater Skid Tankless Water Heating Safety Solutions Pre-assembled, freeze-protected safety fixture and SNA heater skid system Bradley Heat Trace S19-300 with SpinTec 54-144 kw (491,300 BTUs) Molded Duragrate and integrated shower pan Temperature overshoot

More information

Power Flame Incorporated

Power Flame Incorporated Power Flame Incorporated SUGGESTED SPECIFICATION FOR MODEL NVC ULTRA LOW NOx GAS BURNERS SUB 9 PPM NOx THE POWER TO MANAGE ENERGY 2001 South 21st Street, Parsons, Kansas 67357 Telephone: 620-421-0480,

More information

OPERATING MANUAL MODEL AIR 1500TM AIR DRYER

OPERATING MANUAL MODEL AIR 1500TM AIR DRYER OPERATING MANUAL MODEL AIR 500TM AIR DRYER Puregas, LLC 226-A Commerce St. Tel: 800-52-535 Broomfield, Colorado Fax: 303-657-2205 P/N P0255 REV A, 02/9/4 TABLE OF CONTENTS.0 GENERAL... 3 2.0 SPECIFICATIONS...

More information

TITAN SERIES Complete Central Chiller & Pump Tank Package 20 to 70 Tons 20 F to 70 F

TITAN SERIES Complete Central Chiller & Pump Tank Package 20 to 70 Tons 20 F to 70 F Temperature Control Units Water & Oil - F Portable Chillers Air & Water-Cooled - 7 F Central Chillers Air & Water-Cooled Packages & Modules - 7 F Pump Tank Stations Chilled or Tower Water - gallons Cooling

More information

Type 40HT Electrode Rebaking Ovens

Type 40HT Electrode Rebaking Ovens OPERATING INSTRUCTIONS Type 40HT Electrode Rebaking Ovens MODEL PART # DESCRIPTION (All 50-60 Cycles) TEMP RANGE** INSULATION CHAMBER SIZE CAPACITY WEIGHT & DIMENSIONS Type 40HT Series Model Type 40HT

More information

VHP ARD BIODECONTAMINATION SYSTEM

VHP ARD BIODECONTAMINATION SYSTEM VHP ARD BIODECONTAMINATION SYSTEM APPLICATION The VHP ARD Biodecontamination System is designed for portable biodecontamination* of clean, dry, sealed Enclosures.** The VHP 1000-ARD Biodecontamination

More information

SPECIFICATION MICROMIST FIRE SUPPRESSION SYSTEM WITH SHP-PRO CONTROL PANEL

SPECIFICATION MICROMIST FIRE SUPPRESSION SYSTEM WITH SHP-PRO CONTROL PANEL SECTION 1 - GENERAL CONDITIONS I. SCOPE: This specification outlines the requirements for a Fike Micromist Water Mist Fire Suppression System with automatic detection and control. The work described in

More information

SECTION ROTARY-SCREW WATER CHILLERS

SECTION ROTARY-SCREW WATER CHILLERS PART 1 GENERAL 1.01 SECTION INCLUDES A. Factory-assembled packaged chiller. B. Charge of refrigerant and oil. C. Controls and control connections. D. Chilled water connections. E. Electrical power connections.

More information

Centro-Matic Automated Lubrication Systems System Controls

Centro-Matic Automated Lubrication Systems System Controls Selecting the right controls for your automated lubrication system is one of the last steps in the design process. Several different models may be chosen to control power-operated pumps, depending on the

More information

HOT AIR OVEN. NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

HOT AIR OVEN.   NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. HOT AIR OVEN NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation. HOT AIR OVEN LHAO-100 SERIES LABOCON HOT AIR OVEN LHAO-100 SERIES Labocon

More information

Integrated Process Solutions. Series 48 Heaters. heater jackets sizes 1 1/2'' to 4'' (nw 40 to nw 100)

Integrated Process Solutions. Series 48 Heaters. heater jackets sizes 1 1/2'' to 4'' (nw 40 to nw 100) Integrated Process Solutions w w w. m k s i n s t. c o m Series 48 s heater jackets sizes 1 1/2'' to 4'' (nw 40 to nw 100) Series 48 heaters have been designed for safety, flexibility, power conservation,

More information

ESTEEM AUTOCLAVE SLIDING SINGLE DOOR:-

ESTEEM AUTOCLAVE SLIDING SINGLE DOOR:- ESTEEM AUTOCLAVE SLIDING SINGLE DOOR ESTEEM AUTOCLAVE SLIDING SINGLE DOOR:- ESTA-108A For Sterilization of glass wares / containers, vessels, linen, rubber articles culture media etc. Sliding doors with

More information

AUTOMATIC STERILIZERS

AUTOMATIC STERILIZERS AUTOMATIC STERILIZERS MIDMARK DELIVERS WHAT YOU NEED IN INSTRUMENT PROCESSING. Our full line of sterilizers are designed to address all of your critical cleaning needs. In fact, we re sure you ll find

More information

9100 SERIES FLOOR LOADING CART & UTENSIL WASHER/DISINFECTOR

9100 SERIES FLOOR LOADING CART & UTENSIL WASHER/DISINFECTOR CUSTOMER: REFERENCE: 9100 SERIES FLOOR LOADING CART & UTENSIL WASHER/DISINFECTOR PRODUCT SPECIFICATION PRODUCT The Getinge 9100 Washer/Disinfectors are a series of fully automatic, large capacity, floor

More information

Datasheet. Dry-pipe Vacuum Sprinkler System, Electric Release Self contained unit. Features. FIREFLEX VACTEC description

Datasheet. Dry-pipe Vacuum Sprinkler System, Electric Release Self contained unit. Features. FIREFLEX VACTEC description FIREFLEX VACTEC description This FIREFLEX VACTEC integrated fire protection system consists of a vacuum system trim totally pre-assembled, prewired and factory tested. All electrical and mechanical components

More information

Model GW1020 Laboratory Glassware Washer General Specifications

Model GW1020 Laboratory Glassware Washer General Specifications Model GW1020 Laboratory Glassware Washer General Specifications 1.0 General Description The Model GW1020 Laboratory Glassware Washer is a high capacity spray cabinet washer specifically designed to clean

More information

SAMPLE SPECIFICATION FOR RIELLO ARRAY MODULATING BOILER

SAMPLE SPECIFICATION FOR RIELLO ARRAY MODULATING BOILER SAMPLE SPECIFICATION FOR RIELLO ARRAY MODULATING BOILER PART 1 GENERAL 1.01 RELATED DOCUMENTS A. ANSI Z21.13 American National Standard for Gas-Fired Low Pressure Steam and Hot Water Boilers B. ASME Section

More information

Call us today! (800)

Call us today! (800) 500 TON AIR-COOLED CHILLER PA-ACCTR-500 Call us today! (800)341-4297 The 500-Ton portable chiller package features a Trane RTAC chiller unit. Trane leads the industry in providing some of the most reliable,

More information

Design and Construction Standards SECTION PLUMBING EQUIPMENT

Design and Construction Standards SECTION PLUMBING EQUIPMENT SECTION 15450 PLUMBING EQUIPMENT PART 1 GENERAL 1.1 SECTION INCLUDES: A. Water heaters. B. Packaged water heating systems. C. Water storage tanks. D. Water softeners. E. Pumps. F. Circulators. 1.2 REFERENCES

More information

VENTLESS DOOR-TYPE DISHWASHER

VENTLESS DOOR-TYPE DISHWASHER Item # Quantity C.S.I. Section 11400 VENTLESS STANDARD FEATURES Internal condensing system minimizes water vapor Does not require a vent hood Energy recovery Sense-A-Temp 70 F rise electric booster heater.74

More information

FN 032/055/120 DRY HEAT STERILIZERS / OVENS FN 300/400/500 DRY HEAT STERILIZERS / OVENS KD 200/400/700 OVENS EV 018 VACUUM OVEN

FN 032/055/120 DRY HEAT STERILIZERS / OVENS FN 300/400/500 DRY HEAT STERILIZERS / OVENS KD 200/400/700 OVENS EV 018 VACUUM OVEN FN 032/055/120 DRY HEAT STERILIZERS / OVENS Three different sizes: 32, 55 and 120 liters. Temperature range: Ambient Temperature +5 C / 250 C. Designed for sterilization, drying and heating purposes. User

More information

GB Condensing Gas Boiler

GB Condensing Gas Boiler Engineering Submittal Sheet Boiler Overview C US Engineering Specifications There shall be provided and installed a quantity of Buderus GB12/90 condensing hot water boiler(s), suitable for induced draft

More information

SPECIFICATION: AlarmLine TM INTELLIGENT LINEAR HEAT DETECTION

SPECIFICATION: AlarmLine TM INTELLIGENT LINEAR HEAT DETECTION 1. GENERAL SPECIFICATION 1.1 The contractor shall provide an Intelligent AlarmLine Linear Heat Detection System to perform the following functionality: A. Fire Alarm, Supervisory and Trouble-Event Initiation

More information

DIVISION FINNED WATER-TUBE BOILERS

DIVISION FINNED WATER-TUBE BOILERS MVB, TYPE H - MODELS 2503-4003 SUGGESTED SPECIFICATIONS Catalog No.: 2000.934A Effective: 04-01-14 Replaces: 06-01-13 DIVISION 23 52 33.13 FINNED WATER-TUBE BOILERS PART 1 - GENERAL 1.1 SUMMARY A. Section

More information

NC SERIES. Horizontal Steam Sterilizers

NC SERIES. Horizontal Steam Sterilizers NC SERIES Horizontal Steam Sterilizers NC 300/430/570/710 HORIZONTAL STEAM STERILIZERS Our continuous investments in research and development have enabled us to develop our horizontal steam sterilizer

More information

SECTION DOMESTIC WATER HEATERS

SECTION DOMESTIC WATER HEATERS SECTION 223500 PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 01 Specification Sections, apply to

More information

DIVISION FINNED WATER-TUBE BOILERS

DIVISION FINNED WATER-TUBE BOILERS HI DELTA, TYPE H - MODELS 992C-2342C SUGGESTED SPECIFICATIONS Catalog No.: 2000.953 Effective: 10-31-16 Replaces: NEW DIVISION 23 52 33.13 FINNED WATER-TUBE BOILERS - GENERAL 1.1 SUMMARY A. Section includes

More information