Translation of the original operating manual. Constant climate chambers with program control

Size: px
Start display at page:

Download "Translation of the original operating manual. Constant climate chambers with program control"

Transcription

1 Operating Manual Translation of the original operating manual KBF (E6) Constant climate chambers with program control Model Model version Art. No. KBF 115 KBF V , KBF V , KBF 240 KBF V , KBF V , KBF 720 KBF V , KBF V , KBF 1020 KBF V , KBF V , KMF (E6) Constant climate chambers with enlarged temperature and humidity range with program control Model Model version Art. No. KMF 115 KMF V , KMF V , KMF 240 KMF V , KMF V , KMF 720 KMF V , KMF V , BINDER GmbH Address: Post office box 102, Tuttlingen, Germany Phone: Fax: Internet: info@binder-world.com Service Hotline: Service Fax: Service service@binder-world.com Service Hotline USA: or x3 Service Hotline Asia Pacific: or Service Hotline Russia and CIS: Issue 07/2017 Art. N

2 Content 1. SAFETY Legal considerations Structure of the safety instructions Signal word panel Safety alert symbol Pictograms Word message panel structure Localization / position of safety labels on the chamber Type plate General safety instructions on installing and operating the chambers Intended use Operating instructions Measures to prevent accidents Resistance of the humidity sensor against harmful substances CHAMBER DESCRIPTION Chamber overview Instrument panel Lateral control panels Rear view with water connections COMPLETENESS OF DELIVERY, TRANSPORTATION, STORAGE, AND INSTALLATION Unpacking, and checking equipment and completeness of delivery Guidelines for safe lifting and transportation Storage Location of installation and ambient conditions INSTALLATION AND CONNECTIONS Spacer for wall distance Wastewater connection Freshwater supply Automatic fresh water supply via water pipe Manual fresh water supply via external freshwater can (option) Connection kit for connecting the chamber to the water main Safety kit: Hose burst protection device with reflux protection device (available via BINDER INDIVIDUAL customized solutions) Electrical connection Connection of the voltage changer (option for KBF) FUNCTIONAL OVERVIEW OF THE MB2 CHAMBER CONTROLLER Operating functions in normal display Display views: Normal display, program display, chart-recorder display Controller icons overview Operating modes Controller menu structure Main menu Settings submenu Service submenu Principle of controller entries Performance during and after power failures Performance when opening the door START UP Turning on the chamber KBF + KMF (E6) 07/2017 page 2/155

3 6.2 Controller settings upon start up Turning on/off humidity control SET-POINT ENTRY IN FIXED VALUE OPERATING MODE Set-point entry for temperature, humidity, and fan speed through the Setpoints menu Direct setpoint entry for temperature and humidity via Normal display Special controller functions via operation lines TIMER PROGRAM: STOPWATCH FUNCTION Starting a timer program Performance during program delay time Stopping a running timer program Pausing a running timer program Cancelling a running timer program Performance after the end of the program TIME PROGRAMS Starting an existing time program Performance during program delay time Stopping a running time program Pausing a running time program Cancelling a running time program Performance after the end of the program Creating a new time program Program editor: program management Deleting a time program Section editor: section management Add a new program section Copy and insert or replace a program section Deleting a program section Value entry for a program section Section duration Set-point ramp and set-point step Special controller functions via operation lines Setpoint entry Tolerance range Repeating one or several sections within a time program Saving the time program WEEK PROGRAMS Starting an existing week program Cancelling a running week program Creating a new week program Program editor: program management Deleting a week program Section editor: section management Add a new program section Copy and insert or replace a program section Deleting a program section Value entry for a program section Set-point ramp and set-point step modes Weekday Start time Setpoint entry Special controller functions via operation lines NOTIFICATION AND ALARM FUNCTIONS Notification and alarm messages overview Notifications Alarm messages KBF + KMF (E6) 07/2017 page 3/155

4 Messages concerning the humidity system State of alarm Resetting an alarm, list of active alarms Tolerance range settings Activating / deactivating the audible alarm (alarm buzzer) TEMPERATURE SAFETY DEVICES Over temperature protective device (class 1) Overtemperature safety controller class Safety controller modes Setting the safety controller Message and measures in the state of alarm Function check Temperature safety device class 3.3 (option) Temperature safety device class Temperature safety device class USER MANAGEMENT Authorization levels and password protection Log in Log out User change Password assignment and password change Password change Deleting the password for an individual authorization level New password assignment for service or admin authorization level when the password function was deactivated Activation code GENERAL CONTROLLER SETTINGS Selecting the controller s menu language Selecting the temperature unit Setting date and time Display configuration Adapting the display parameters Touchscreen calibration Network and communication Serial interfaces Ethernet Web server USB menu GENERAL INFORMATION Service contact page Current operating parameters Event list Technical chamber information CHART RECORDER DISPLAY Views Show and hide legend Switch between legend pages Show and hide specific indications History display Setting the parameters HUMIDIFICATION / DEHUMIDIFICATION SYSTEM Function of the humidifying and dehumidifying system KBF + KMF (E6) 07/2017 page 4/155

5 18. DEFROSTING AT REFRIGERATING OPERATION OPTIONS Communication software APT-COM 3 DataControlSystem (option) RS485 interface (option) Data logger kits (option) Analog outputs for temperature and humidity (option) Zero-voltage relay alarm outputs for temperature and humidity (option) Water protected internal socket (option for KBF not for UL chambers) Additional flexible Pt 100 temperature sensor (available via BINDER INDIVIDUAL customized solutions) Object temperature display with flexible Pt 100 temperature sensor (option) External freshwater and wastewater cans (option) Mounting the freshwater can Mounting the wastewater can Mounting with wastewater recycling BINDER Pure Aqua Service (option) MAINTENANCE, CLEANING, AND SERVICE Maintenance intervals, service Cleaning and decontamination Cleaning Decontamination Sending the chamber back to BINDER GmbH DISPOSAL Disposal of the transport packing Decommissioning Disposal of the chamber in the Federal Republic of Germany Disposal of the chamber in the member states of the EU except for the Federal Republic of Germany Disposal of the chamber in non-member states of the EU TROUBLESHOOTING TECHNICAL DESCRIPTION Factory calibration and adjustment Over current protection Definition of usable volume KBF Technical Data KMF technical data Equipment and options (extract) Spare parts and accessories (extract) Dimensions size Dimensions size Dimensions size Dimensions size CERTIFICATES AND DECLARATIONS OF CONFORMITY EU Declaration of Conformity for KBF EU Declaration of Conformity for KMF Certificate for the GS mark of conformity of the Deutsche Gesetzliche Unfallversicherung e.v. (German Social Accident Insurance) DGUV PRODUCT REGISTRATION CONTAMINATION CLEARANCE CERTIFICATE For chambers located outside USA and Canada For chambers located in USA and Canada KBF + KMF (E6) 07/2017 page 5/155

6 Dear customer, For the correct operation of the chambers, it is important that you read this operating manual completely and carefully and observe all instructions as indicated. Failure to read, understand and follow the instructions may result in personal injury. It can also lead to damage to the chamber and/or poor equipment performance. 1. Safety This operating manual is part of the components of delivery. Always keep it handy for reference. The device should only be operated by laboratory personnel especially trained for this purpose and familiar with all precautionary measures required for working in a laboratory. Observe the national regulations on minimum age of laboratory personnel. To avoid injuries and damage observe the safety instructions of the operating manual. Failure to observe the safety instructions. Serious injuries and chamber damage. WARNING Observe the safety instructions in this operating manual. Carefully read the complete operating instructions of the chambers. 1.1 Legal considerations This operating manual is for informational purposes only. It contains information for installing, start-up, operation and maintenance of the product. Note: the contents and the product described are subject to change without notice. Understanding and observing the instructions in this operating manual are prerequisites for hazard-free use and safety during operation and maintenance. In no event shall BINDER be held liable for any damages, direct or incidental arising out of or related to the use of this manual. This operating manual cannot cover all conceivable applications. If you would like additional information, or if special problems arise that are not sufficiently addressed in this manual, please ask your dealer or contact us directly by phone at the number located on page one of this manual Furthermore, we emphasize that the contents of this operating manual are not part of an earlier or existing agreement, description, or legal relationship, nor do they modify such a relationship. All obligations on the part of BINDER derive from the respective purchase contract, which also contains the entire and exclusively valid statement of warranty administration. The statements in this manual neither augment nor restrict the contractual warranty provisions. 1.2 Structure of the safety instructions In this operating manual, the following safety definitions and symbols indicate dangerous situations following the harmonization of ISO and ANSI Z Signal word panel Depending on the probability of serious consequences, potential dangers are identified with a signal word, the corresponding safety color, and if appropriate, the safety alert symbol. DANGER Indicates an imminently hazardous situation that, if not avoided, will result in death or serious (irreversible) injury. KBF + KMF (E6) 07/2017 page 6/155

7 WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious (irreversible) injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (reversible) injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in damage to the product and/or its functions or of a property in its proximity Safety alert symbol Use of the safety alert symbol indicates a risk of injury. Observe all measures that are marked with the safety alert symbol in order to avoid death or injury Pictograms Warning signs Electrical hazard Hot surface Explosive atmosphere Stability hazard Lifting hazard Scalding hazard High humidity Danger of frost Risk of corrosion and / or chemical burns Harmful substances Biohazard Pollution Hazard Mandatory action signs Mandatory regulation Read operating instructions Disconnect the power plug Lift with several persons Lift with mechanical assistance Environment protection Wear protective gloves Wear safety goggles KBF + KMF (E6) 07/2017 page 7/155

8 Prohibition signs Do NOT touch Do NOT spray with water Do NOT climb Information to be observed in order to ensure optimum function of the product Word message panel structure Type / cause of hazard. Possible consequences. Instruction how to avoid the hazard: prohibition Instruction how to avoid the hazard: mandatory action. Observe all other notes and information not necessarily emphasized in the same way, in order to avoid disruptions that could result in direct or indirect injury or property damage. 1.3 Localization / position of safety labels on the chamber The following labels are located on the chamber: Pictograms (warning signs) Risk of injury (on outer door, only KBF-240V and KMF-240V). Observe the safety instructions in the operating manual. Hot surface (inner glass door above the glass door handle) Burning and scalding hazard (chamber rear) Service label KBF + KMF (E6) 07/2017 page 8/155

9 Figure 1: Position of labels on the chamber front (KBF-240V and KMF-240V) Figure 2: Position of labels on the chamber rear Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER Service for these replacements. KBF + KMF (E6) 07/2017 page 9/155

10 1.4 Type plate The type plate sticks to the left side of the chamber, bottom right-hand. Nominal temp. 70 C 2,10 kw / 9,5 A Max. operating pressure 15 bar 158 F V / 50 Hz R 134A 0,17 kg IP protection V / 60 Hz Contains fluorinated Safety device DIN N ~ greenhouse gases covered Class 3.1 by the Kyoto Protocol Art. No Project No. Built 2017 Constant climate chamber BINDER GmbH Im Mittleren Ösch Tuttlingen / Germany KBF 240 E6 Figure 3: Type plate (example KBF 240 regular chamber ) Serial No Made in Germany Indications of the type plate (example) Information BINDER Manufacturer: BINDER GmbH KBF 240 Model designation Constant climate chamber Device name Serial No Serial no. of the chamber Built 2017 Year of construction Nominal temperature 70 C / 158 F Nominal temperature IP protection 20 IP type of protection acc. to standard EN Temp. safety device DIN Temperature safety device acc. to standard DIN 12880:2007 Class 3.1 Class of temperature safety device Art. No Art. no. of the chamber Project No. --- Optional: Special application acc. to project no. 2,10 kw Nominal power 9,5 A Nominal current V / 50 Hz Nominal voltage range +/-10% V / 60 Hz at the indicated power frequency 1 N ~ Current type Max. operating pressure 15 bar Max operating pressure in the refrigerating system (15 bar / 218 PSI) R 134A - 0,17 kg Refrigerant type and filling weight Contains fluorinated greenhouse gases covered by the Kyoto Protocol Symbol on the type plate Information CE conformity marking Electrical and electronic equipment manufactured / placed on the market in the EU after 13 August 2005 and be disposed of in separate collection according to Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). KBF + KMF (E6) 07/2017 page 10/155

11 Symbol on the type plate Information GS mark of conformity of the Deutsche Gesetzliche Unfallversicherung e.v. (DGUV), Prüf- und Zertifizierungsstelle Nahrungsmittel und Verpackung im DGUV Test (German Social Accident Insurance (DGUV), Testing and Certification Body for Foodstuffs and Packaging Industry in DGUV Test). The chamber is certified according to Customs Union Technical Regulation (CU TR) for the Eurasian Economic Union (Russia, Belarus, Armenia, Kazakhstan Kyrgyzstan). 1.5 General safety instructions on installing and operating the chambers With regard to operating the chambers and to the installation location, please observe the DGUV guidelines on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120 or ZH 1/119, issued by the employers liability insurance association) (for Germany. BINDER GmbH is only responsible for the safety features of the chamber provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts. To operate the chamber, use only original BINDER accessories or accessories from third-party suppliers authorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories. Danger of overheating. Damage to the chamber. CAUTION Do NOT install the chamber in unventilated recesses. Ensure sufficient ventilation for dispersal of the heat. Do not operate the chambers in hazardous locations. Explosion hazard. Danger of death. DANGER Do NOT operate the chamber in potentially explosive areas. KEEP explosive dust or air-solvent mixtures AWAY from the chamber. The chambers do not dispose of any measures of explosion protection. Explosion hazard. Danger of death. DANGER Do NOT introduce any substance into the constant climate chamber which is combustible or explosive at working temperature. NO explosive dust or air-solvent mixture in the inner chamber. Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature inside the chamber must lie below the flash point or below the sublimation point of the charging material. Familiarize yourself with the physical and chemical properties of the charging material, as well as the contained moisture constituent and its behavior with the addition of heat energy and humidity. KBF + KMF (E6) 07/2017 page 11/155

12 Familiarize yourself with any potential health risks caused by the charging material, the contained moisture constituent or by reaction products that may arise during the temperature process. Take adequate measures to exclude such risks prior to putting the chamber into operation. Electrical hazard. Danger of death. DANGER The chamber must NOT become wet during operation or maintenance. The chambers were produced in accordance with VDE regulations and were routinely tested in accordance to VDE (IEC ). During and shortly after operation, the temperature of the inner surfaces almost equals the set-point. CAUTION The glass doors, the glass door handles, and the inner chamber will become hot during operation. Danger of burning. Do NOT touch the glass doors, the inner surfaces or the charging material during operation. WARNING Stability hazard. Danger of injury. Damage to the chamber and the charging material. Housing cover breakaway. Do NOT climb on the lower housing cover. Do NOT load the lower housing cover and the door with heavy objects while the chamber door is open. KBF + KMF (E6) 07/2017 page 12/155

13 1.6 Intended use Constant climate chambers series KBF and KMF are suitable for exact conditioning of harmless materials. A mixture of any component of the charging material with air must NOT be explosive. The operating temperature must lie below the flash point or below the sublimation point of the charging material. Any component of the charging material must NOT be able to release toxic gases. Other applications are not approved. The chambers are not classified as medical devices as defined by the Medical Device Directive 93/42/EEC. Following the instructions in this operating manual and conducting regular maintenance work (chap. 20) are part of the intended use. Explosion or implosion hazard. Danger of poisoning. Danger of death. DANGER Do NOT introduce any substance combustible or explosive at working temperature into the chamber, in particular no energy sources such as batteries or lithium-ion batteries. NO explosive dust or air-solvent mixture in the inner chamber. Do NOT introduce any substance which could lead to release of toxic gases. The charging material shall not contain any corrosive ingredients that may damage the machine components made of stainless steel, aluminum, and copper. Such ingredients include in particular acids and halides. Any corrosive damage caused by such ingredients is excluded from liability by BINDER GmbH. WARNING: If customer should use a BINDER chamber running in non-supervised continuous operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two chambers, if this is feasible. In case of foreseeable use of the chamber there is no risk for the user through the integration of the chamber into systems or by special environmental or operating conditions in the sense of EN :2010. For this, the intended use of the chamber and all its connections must be observed. 1.7 Operating instructions Depending on the application and location of the chamber, the operator of the chamber must provide the relevant information for safe operation of the chamber in a set of operating instructions. Keep these operating instructions with the chamber at all times in a place where they are clearly visible. They must be comprehensible and written in the language of the employees. KBF + KMF (E6) 07/2017 page 13/155

14 1.8 Measures to prevent accidents The operator of the chamber must observe the following rule: Betreiben von Arbeitsmitteln. Betreiben von Kälteanlagen, Wärmepumpen und Kühleinrichtungen (Operation of work equipment. Operation of refrigeration systems, heat pumps and refrigeration equipment) (GUV-R 500 chap. 2.35) (for Germany). The manufacturer took the following measures to prevent ignition and explosions: Indications on the type plate See operating manual chap Operating manual An operating manual is available for each chamber. Overtemperature monitoring The chamber is equipped with a temperature display, which can be read from outside. The chamber is equipped with an additional safety controller (temperature safety device class 3.1 acc. to DIN 12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding. Safety, measurement, and control equipment The safety, measuring, and control equipment is easily accessible. Electrostatic charge The interior parts are grounded. Non-ionizing radiation Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electrical equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent magnets. If persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field source to implant) of 30 cm, an influence of these implants can be excluded with high probability. Protection against touchable surfaces Tested according to EN ISO :2008. Floors See operating manual chap. 3.4 for correct installation Cleaning See operating manual chap Examinations The chamber has been inspected by the Deutsche Gesetzliche Unfallversicherung e.v. (DGUV) (German Social Accident Insurance (DGUV) (German Social Accident Insurance (DGUV), Testing and Certification Body for Foodstuffs and Packaging Industry in DGUV Test) and bears the GS mark. KBF + KMF (E6) 07/2017 page 14/155

15 1.9 Resistance of the humidity sensor against harmful substances The following list of harmful substances refers only to the humidity sensor and does not include any other materials incorporated in the chamber or prohibited substances in relation to explosion protection. Some gases - especially clean gases - do not have any influence on the humidity sensor. Others do have a very small influence, whereas others may influence the sensor to a larger extent. The following gases do not influence the sensor and the humidity measurement: Argon (Ar), carbon dioxide (CO 2 ),helium (He), hydrogen (H 2 ), neon (Ne), nitrogen (N 2 ), nitrous oxide (N 2 O), oxygen (O 2 ) The following gases do not or to a minor extent influence the sensor and the humidity measurement: Butane (C 4 H 10 ), ethane (C 2 H 6 ), methane (CH 4 ), natural gas propane (C 3 H 8 ) The following gases do not, or to a minor extent influence the sensor and the humidity measurement, provided that the indicated loads are not exceeded: Maximum work place threshold limit value Tolerated concentration with permanent load Substance Formula ppm mg/m 3 ppm mg/m 3 Ammonia NH Acetone CH 3 COCH Benzene Chlorine Cl Acetic acid CH 3 COOH Ethyl acetate CH 3 COOC 2 H Ethanol C 2 H 5 OH Ethylene glycol HOCH 2 CH 2 OH Formaldehyde HCHO Isopropanol (CH 3 ) 2 CHOH Methanol CH 3 OH Methyl ethyl ketone C 2 H 5 COCH Ozone O Hydrochloric acid HCl Hydrogen sulphide H 2 S Nitrogen oxides NOx Sulphur dioxide SO Toluol C 6 H 5 CH Xylene C 6 H 4 (CH 3 ) These values are to be considered as approximate values. The sensor resistance largely depends on the temperature and humidity conditions during the time of exposure to harmful substances. Avoid simultaneous condensation. Tolerated error of measurement: ± 2 %r.h. The maximum work place threshold limit value is one that can be regarded as harmless for humans. Vapors of oil and fat are dangerous for the sensor because they may condensate at the sensor and thus prevent its function (insulating layer). For similar reasons it is not possible to measure smoke gases. KBF + KMF (E6) 07/2017 page 15/155

16 2. Chamber description The constant climate chambers KBF and KMF are equipped with a multifunctional microprocessor display controller with 2-channel technology for temperature and humidity plus a digital display accurate to onetenth of a degree resp. 0.1% r.h. With its comprehensive program control functions, the display program controller MB2 permits the high precision performance of temperature and humidity cycles. With its microprocessor controlled humidifying and dehumidifying system the KMF is a high-precision constant climate chamber. The KBF completely meets the requirements for climatic chambers of the stipulated stability and durability tests for pharmaceutical products: Stability tests acc. to ICH guideline CPMP/ICH/2736/99 (Q1A) The KMF completely meets the requirements of the stipulated stability and durability tests for industrial products. The APT.line preheating chamber system guarantees high level of spatial and time-based temperature precision, thanks to the direct and distributed air circulation into the interior. The fan supports exact attainment and maintenance of the desired temperature accuracy. Furthermore, it permits simulating exactly and over long periods constant conditions for other applications such as sample conditioning for material testing of paper, textiles, plastics, building materials, etc. A resistance humidifying system humidifies the air. For this purpose, use deionized (demineralized) water. The option BINDER Pure Aqua Service allows using the chamber with any degree of water hardness. The inner chamber, the pre-heating chamber and the interior side of the doors are all made of stainless steel V2A (German material no , US equivalent AISI 304). The housing is RAL 7035 powdercoated. All corners and edges are also completely coated. All chamber functions are easy and comfortable to use thanks to their clear arrangement. Major features are easy cleaning of all chamber parts and avoidance of undesired contamination. The efficient program controller is equipped with a multitude of operating functions, in addition to recorder and alarm functions. Programming of test cycles is easily accomplished via the modern MB2 touch screen controller and is also possible directly with a computer via Intranet in connection with the communication software APT-COM 3 DataControlSystem (option, chap. 19.1). The constant climate chamber KMF comes equipped with an Ethernet serial interface for computer communication. In addition, the BINDER communication software APT-COM 3 permits networking up to 30 chambers and connecting them to a PC for controlling and programming, as well as recording and representing temperature and humidity data. For further options, see chap The chambers size 240, 720, and 1020 are equipped with four castors. Both front castors can be easily locked via the attached brakes. KBF: temperature range: 0 C / 32 F up to 70 C / 158 F, humidity range: 10% r.h. to 80% r.h. KMF: temperature range -10 C / 14 F up to +100 C / 212 F, humidity range: 10 % r.h. to 98 % r.h. For the control ranges of temperature and humidity, see climatic diagrams (chap. 17). KBF + KMF (E6) 07/2017 page 16/155

17 2.1 Chamber overview (A) (B) (C) (D) Figure 4: Constant climate chamber KBF / KMF size 240 (A) (B) (C) (D) Instrument box Door handle Outer door Refrigerating machine and humidity generation module 2.2 Instrument panel 5,7" controller display with touchscreen USB interface Pilot lamp Figure 5: Instrument panel with MB2 program controller and USB interface KBF + KMF (E6) 07/2017 page 17/155

18 2.3 Lateral control panels (1) (2) (5) (3) (6) (8) (4) (7) (9) right side left side (optional) Figure 6: Lateral control panels at the sides of the refrigerating / humidity generation module with optional equipment (1) Main power switch (2) DIN socket for additional Pt 100 sensor (available via BINDER INDIVIDUAL customized solutions) (3) DIN socket for analog outputs (option) (4) RS485 interface (5) DIN socket for switching contacts (option for KMF) (6) DIN socket for zero-voltage relay alarm output (option) (7) Ethernet interface (8) Temperature safety device class 3.1 (part of option Safety device class 3.3 ) (9) Temperature safety device class 3.2 (part of option Safety device class 3.3 ) KBF + KMF (E6) 07/2017 page 18/155

19 2.4 Rear view with water connections (10) (11) (12) (13) (14) Figure 7: Rear view of the chamber with water connections (10) Socket for optional freshwater can (chap ) (11) Power cable (12) Purging outlet for humidifying module for service purpose only (13) Freshwater connection IN with screw thread ¾ for hose ½, with union nut (14) Wastewater connection OUT with hose olive for hose ½ KBF + KMF (E6) 07/2017 page 19/155

20 3. Completeness of delivery, transportation, storage, and installation 3.1 Unpacking, and checking equipment and completeness of delivery After unpacking, please check the chamber and its optional accessories, if any, based on the delivery receipt for completeness and for transportation damage. Inform the carrier immediately if transportation damage has occurred. The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This has no impact on the function and performance of the chamber. Please remove any transportation protection devices and adhesives in/on the chamber and on the doors and remove the operating manuals and accessory equipment. Sliding or tilting of the chamber. Damage to the chamber. Risk of injury by lifting heavy loads. CAUTION Do NOT lift or transport the chamber using the door, the handle, or the lower housing. Lift chambers size 115 from the pallet at the four lower corners with the aid of four people Lift chambers size 240 from the pallet at the four lower corners with the aid of six people or with a fork lifter. Set the fork lifter only from the front or rear in the middle of the chamber. Lift the chambers sizes 720 and 1020 from the pallet using technical devices (fork lifter). Set the fork lifter only from the front or rear in the middle of the chamber. Do NOT set the fork lifter from the chamber side. If you need to return the chamber, please use the original packing and observe the guidelines for safe lifting and transportation (chap. 3.2). For disposal of the transport packing, see chap Note on second-hand chambers (Ex-Demo-Units): Second-hand chambers are chambers that were used for a short time for tests or exhibitions. They are thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly. Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before commissioning the chamber. KBF + KMF (E6) 07/2017 page 20/155

21 3.2 Guidelines for safe lifting and transportation The front castors of KMF 240, 720 and 1020 can be blocked by brakes. After operation, please observe the guidelines for temporarily decommissioning the chamber (chap. 21.2). Please move the chambers with castors only when empty and on an even surface, otherwise the castors may be damaged. Sliding or tilting of the chamber. Damage to the chamber. Risk of injury by lifting heavy loads. CAUTION Transport the chamber in its original packaging only. For moving or shipping, secure the constant climate chamber with transport straps. Do NOT lift or transport the chamber using the door, the handle, or the lower housing. Lift chambers size 115 at the four lower corners with the aid of 4 people Lift chambers size 240 at the four lower corners with the aid of 6 people or with a fork lifter. Set the fork lifter only from the front or rear in the middle of the chamber. Lift the chambers sizes 720 and 1020 using technical devices (fork lifter). Set the fork lifter only from the front or rear in the middle of the chamber. Do NOT set the fork lifter from the chamber side. You can order transport packing for moving or shipping purposes from BINDER service. Permissible ambient temperature range during transport: If the steam humidifying system has NOT been emptied: +3 C / 37.4 F to +60 C / 140 F. After BINDER Service has emptied the steam humidifying system: -10 C / 14 F to +60 C / 140 F. With temperatures below +3 C / 37.4 F, water must be completely removed from the humidifying system. CAUTION Transport below +3 C / 37.4 F with filled steam humidifying system. Freezing in the steam generator. Damage to the chamber. Contact BINDER Service before any transportation below +3 C / 37.4 F. 3.3 Storage Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for temporary decommissioning (chap. 21.2). Permissible ambient temperature range during storage: If the steam humidifying system has NOT been emptied: +3 C / 37.4 F to +60 C / 140 F. After BINDER Service has emptied the steam humidifying system: -10 C / 14 F to +60 C / 140 F KBF + KMF (E6) 07/2017 page 21/155

22 With temperatures below +3 C / 37.4 F, water must be completely removed from the humidifying system. CAUTION Storage below +3 C / 37.4 F with filled steam humidifying system. Freezing in the steam generator. Damage to the chamber. Contact BINDER Service before any transportation below +3 C / 37.4 F. Permissible ambient humidity: max. 70 % r.h., non-condensing CAUTION Condensation by excess humidity. Danger of corrosion on the housing after operating at humidity values > 70 % r.h. for a long period. Dry the chamber completely before shut-down: Set the humidity to 0 % r.h. To enable dehumidification, the humidifying and dehumidifying system must be activated (deactivated operation line Humidity off, chap. 7.3 and setting Control on, chap. 6.3). Set the temperature set point to 60 C / 140 F for approx. 2 hours (Manual mode). Only then, shut down the chamber at the main power switch (1) and close the tap of the water supply. When after storage in a cold location you transfer the chamber to its warmer installation site, condensation may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and is completely dry. 3.4 Location of installation and ambient conditions Set up the constant climate chamber on a flat, even surface, and in a well-ventilated, dry location and align it using a spirit level. The site of installation must be capable of supporting the chamber s weight (see technical data, chap. 23.5). The chambers are designed for setting up inside a building (indoor use). Danger of overheating. Damage to the chamber. CAUTION Do NOT set up chamber in non-ventilated recesses. Ensure sufficient ventilation for dispersal of the heat. Permissible ambient temperature range during operation: +18 C / 64.4 F to +32 C / 89.6 F. At elevated ambient temperature values, fluctuations in temperature can occur. The ambient temperature should not be substantially higher than the indicated ambient temperature of +22 C +/- 3 C / 71.6 F ± 5.4 F to which the specified technical data relate. Deviations from the indicated data are possible for other ambient conditions. Lower values of the temperature range indicated in the technical data are valid at an ambient temperature of max. 25 C / 77 F. With each degree of ambient temperature >25 C / 77 F, the refrigeration power decreases by 1.5 K. KBF + KMF (E6) 07/2017 page 22/155

23 Permissible ambient humidity: 70 % r.h. max., non-condensing When operating the chamber at temperature set-points below ambient temperature, high ambient humidity may lead to condensation on the chamber. Installation height: max m / 6562 ft. above sea level. A water tap (1 bar to 10 bar) is necessary for the installation of the humidification system (chap. 4.3). If no suitable in-house water connection is available, you can manually supply water by filling the freshwater can (option, chap. 19.9). To avoid any possible water damage, provide a floor drain at the location of the device. Select a suitable installation site to avoid any consequential damage by splashing water. When placing several chambers of the same size side by side, maintain a minimum distance of 250 mm / 9.84 in between each chamber. Wall distances: rear 100 mm / 3.9 in, sides 160 mm / 6.29 in. Spacing above the chamber of at least 100 mm / 3.9 in must also be accounted for. Danger by stacking. Damage to the chambers. CAUTION Do NOT place the chambers on top of each other. To completely separate the chamber from the power supply, you must disconnect the power plug. Install the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger. For the user there is no risk of temporary overvoltages in the sense of EN :2010. With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several times a year. Avoid any conductive dust in the ambiance according to the chamber layout complying with pollution degree 2 (IEC ). Do not install or operate the chamber in potentially explosive areas. Explosion hazard. Danger of death. DANGER Do NOT operate the chamber in potentially explosive areas. KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber. After turning off the chamber, you must close the tap of the water supply. Install the chamber in a way that the freshwater supply is easily accessible. In case of a prolonged temporal decommissioning: Leave the chamber door open or remove the access port plugs. KBF + KMF (E6) 07/2017 page 23/155

24 4. Installation and connections 4.1 Spacer for wall distance Please fix both spacers with the supplied screws at the chamber rear. This serves to ensure the prescribed minimum distance to the rear wall of 100 mm / 3.94 in. Figure 8: Spacer for wall distance Figure 9: Chamber rear with mounted spacers 4.2 Wastewater connection Fasten the wastewater hose to the wastewater connection OUT (14) on the rear of the chamber (olive 14 mm). Observe the following points: You can use a part of the supplied water hose as a drainage hose. In case another hose is used, it has to be permanently resistant against at least 95 C / 203 F. Mount the wastewater hose with a maximum positive inclination of 1 m and a maximum total length of 3 m. Protect the chamber end of the drainage hose with one of the supplied hose clamps. Reliably prevent sucking back of wastewater. The end of the wastewater hose must not be immersed in liquids. This can be ensured e.g., by free discharge. Waste water is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off only when required, thus there is no continuous waste water flow. Protect the waste water supply with the supplied hose clamps. KBF + KMF (E6) 07/2017 page 24/155

25 4.3 Freshwater supply Connect the waste water pipe before connecting the chamber to a freshwater pipe or filling the freshwater can (option, chap. 19.9). You can supply the chamber with freshwater via a water pipe or by manually filling a freshwater can (option, chap. 19.9). Water intake temperature NOT below +5 C / 41 F and not exceeding 40 C / 104 F. CAUTION Calcification of the humidifying system. Damage to the chamber. Operate the chamber with deionized (demineralized) water only. Types of suitable water quality: Deionized water from a water treatment installation already existing at the customer's site. Conductivity from 1 µs /cm up to a maximum of 20 µs/cm. (Water, which is in equilibrium with the CO 2 in the air, and has a conductivity below 1 µs/cm (ultrapure water), may cause acid corrosion due to its low ph.) Water treated by the optional water treatment system BINDER Pure Aqua Service (disposable system). A reusable measuring equipment to assess the water quality is included (chap ). BINDER GmbH is NOT responsible for the water quality at the user s site. Any problems and malfunctions that might arise following use of water of deviating quality is excluded from liability by BINDER GmbH. The warranty becomes void in the event of use of water of deviating quality Automatic fresh water supply via water pipe An enclosure inside the chamber contains the connection kit for freshwater and wastewater. Install the freshwater connection using either the enclosed water hose or another pressure-resistant one. To accomplish this, remove the cover of the freshwater connection IN (13) on the rear of the chamber. Protect both ends of the hose with two of the four supplied hose clamps. Before turning on the chamber, check the connection for leaks. Water supply is automatically effected via the freshwater connection IN (13). As the appliance only lets in water when required, there is no continuous water flow. Supply pressure 1 to 10 bar when connecting to a water pipe Water type: deionized (demineralized) water Water intake temperature NOT below +5 C / 41 F and not exceeding 40 C / 104 F. The water intake should be provided with a shut-off slide or water-tap. For the water supply, fix the delivered adapter with hose olive on the thread at the rear of the chamber. Protect the water supply at one side with the supplied hose clamp. KBF + KMF (E6) 07/2017 page 25/155

26 4.3.2 Manual fresh water supply via external freshwater can (option) If no house water connection with suitable water is available, you can manually supply water by filling a freshwater can (option, volume: 20 liters / 0.71 cu.ft. You can attach the freshwater can on the rear of the chamber or place it next to the chamber (chap. 19.9). To guarantee humidification during 24 hours even at high humidity set-points with manual water supply, we recommend filling the freshwater can (option) daily at the end of the day Connection kit for connecting the chamber to the water main A safety kit against flooding caused by burst water hoses is enclosed with the constant climate chamber. It consists of the following: Hose burst protection device 2 hose nozzles with screwing 4 hose clamps 6m water hose, divisible for the feed hose and drain Protection principle of hose burst protection: Whenever a strong water flow of about 18 l / min. occurs, e.g. caused by a burst water hose, a valve automatically cuts off the water supply, which can be heard as a clicking noise. The water supply now remains shut until it is manually released. Assembly: Screw the hose burst protection device onto a water tap with a G¾ inch right turning thread connection. The connection is self-sealing. Establish the connection between the safety kit and the chamber with a part of the supplied hose. Protect both ends of the hose by the supplied hose clamps. We recommend connecting the hose as the last step in order to avoid twisting the hose while screwing on the safety kit. Open the water tap slowly in order to avoid actuating the hose burst protection device. Figure 10: Assembly of the connection kit Release of the reflux protection device: In case the burst protection device has interrupted the water supply, first find the reason and remove it as necessary. Close the water tap. Release the valve by a half left-turn of the upper knurled part. You can hear the release of the valve as a clicking noise. Tighten the burst protection device against the water tap by a right turn. Open the water tap slowly afterwards. KBF + KMF (E6) 07/2017 page 26/155

27 Maintenance of the assembly of the hose burst protection device: Calcification can impair valve function. We recommend an annual inspection by a skilled plumber. The plumber should demount the safety kit to check the valve by hand for function, calcification or blockage. CAUTION Danger of calcification. Impairment of valve function. Have a plumber inspect the valve annually. Remove calcifications by citric acid or acetic acid solutions. Continue by testing the function and tightness of the mounted chamber Check: Quickly open the water tap while there is no chamber connected the valve should cut off the water flux without any delay Safety kit: Hose burst protection device with reflux protection device (available via BINDER INDIVIDUAL customized solutions) A safety kit with a reflux protection device is available for protection of the drinking water system, and against flooding caused by burst water hoses. Protection principles: Whenever a strong water flow of about 18 l / min. occurs, e.g. caused by a burst water hose, a valve automatically cuts off the water supply, which can be heard as a clicking noise. The water supply now remains shut until it is manually released. A possible endangering of the drinking water system depends on the risk potential of the charging material. Under unfavorable conditions (e.g. decreasing pressure inside the tap water system), drained off charging material could be sucked out of the chamber via the steam generator into the tap water system and therefore contaminate the drinking water. The safety kit with reflux protection device provides security in case of short-term utilization of substances with low risk potential. When using substances bearing a higher risk potential, install a pipe disconnector to assure absolute protection. It is the user s responsibility to prevent (according to national standards) any reflux of contaminated water from getting into the drinking water system. Assembly: The standard supplied parts hose burst protection device, hose nozzle with screwing are not needed. Screw the pre-mounted assembly of the hose burst protection and reflux protection devices onto a water tap with a G¾ inch right turning thread connection. The connection is self-sealing. Establish the connection between the safety kit and the chamber with a part of the supplied hose. Protect both ends of the hose with the supplied hose clamps. We recommend connecting the hose as the last step in order to avoid twisting it while screwing on the safety kit. Open the water tap slowly in order to avoid actuating the hose burst protection device. KBF + KMF (E6) 07/2017 page 27/155

28 Figure 11: Assembly of the safety kit (hose burst protection and reflux protection devices) Release of the reflux protection device: In case the hose burst protection device interrupts the water supply, first find the reason and remove it as necessary. Close the water tap. Release the valve by a half left-turn of the upper knurled part. You can hear the release of the valve as a clicking noise. Tighten the burst protection device against the water tap by a right turn. Open the water tap slowly afterwards. Maintenance of the assembly of hose burst protection and reflux protection devices: Calcification can impair the function of both valves. We recommend an annual inspection by a skilled plumber. The plumber should remove the safety kit with the reflux protection device to check both valves by hand for proper function and calcification or blockage. Danger of calcification. Impairment of valve function. CAUTION Have a plumber inspect the two valves annually. Remove calcifications by citric acid or acetic acid solutions. Continue by testing the function and tightness of the mounted chamber. Check: Quickly open the water tap while there is no chamber connected the valve should cut off the water flux without any delay. KBF + KMF (E6) 07/2017 page 28/155

29 4.4 Electrical connection The chambers are supplied ready for connection. They come with a fixed power connection cable of at least 1800 mm / in in length. Model version Art. No. (x = 0 or 1) Power plug Voltage +/-10% at the indicated power frequency Current type Chamber fuse KBF V KBF V KBF V KBF V 9x x x x Shock-proof plug V at 50 Hz V at 60 Hz 1N~ 16 Amp KBF V KBF V KBF V KBF V 9x x x x NEMA 6-20P V at 50Hz V at 60Hz 2~ 16 Amp KMF V KMF V KMF V 9x x x Shock-proof plug V at 50 Hz V at 60 Hz 1N~ 16 Amp KMF V KMF V KMF V 9x x x NEMA 6-20P V at 50Hz V at 60Hz 2~ 16 Amp The domestic socket must also provide a protective conductor. Make sure that the connection of the protective conductor of the domestic installations to the chamber s protective conductor meets the latest technology. The protective conductors of the socket and plug must be compatible! Prior to connection and start-up, check the power supply voltage. Compare the values to the specified data located on the chamber s type plate (left chamber side, bottom right-hand, see chap. 1.4). When connecting, please observe the regulations specified by the local electricity supply company and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker. Pollution degree (acc. to IEC ): 2 Installation category (acc. to IEC ): II Danger of incorrect power supply voltage. Damage to the equipment. CAUTION Check the power supply voltage before connection and start-up. Compare the power supply voltage with the data indicated on the type plate. See also electrical data (chap. 23.5). To completely separate the chamber from the power supply, you must disconnect the power plug. Install the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger. Remark when operating the chamber with a power frequency of 60 Hz: WARNING High leakage current. Electrical hazard. Earth connection is essential before connecting supply. Check socket before inserting plug. KBF + KMF (E6) 07/2017 page 29/155

30 When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5 mamp is possible. If grounding through the power cable is insufficient or missing, the leakage current may flow through the user s body. Correct installation of the professional grade power socket provided by the user safely avoids this. Before connecting the chamber to the socket, please check its grounding contact type plug for appropriate construction and if it is undamaged. 4.5 Connection of the voltage changer (option for KBF) The voltage changer enables the constant climate chamber to operate at a power frequency of 115 Volt. It is packed separately and supplied together with the constant climate chamber. The voltage changer is supplied with a fixed power connection cable with a NEMA 5-20P plug. It is protected against excess-current with an internal over-current release category B16A. The connection is made by the customer. Sliding or tilting of the voltage changer. Damage to the voltage changer. Risk of injury by lifting heavy loads. CAUTION Lift the voltage changer at both carrying handles from the pallet with two persons. Do not install the voltage changer in the exhaust air flow at the rear of the constant climate chamber. For the installation of the voltage changer next to the constant climate chamber, provide a wall distance the alternating climate chamber of approx. 0.4 m / 1.3 ft. Danger of overheating. Damage to the voltage changer. CAUTION Do NOT install the voltage changer in unventilated recesses. Ensure sufficient ventilation for dispersal of the heat. (A) (B) (C) (A) (B) (C) (D) (E) Carrying handle Pilot lamp (green) Power switch Connection socket for KBF Power cable (D) (E) Figure 12: Voltage changer (front) KBF + KMF (E6) 07/2017 page 30/155

31 To establish the electrical connection of the constant climate chamber with the voltage changer, proceed in the following order: 1. Connect the power cable of the constant climate chamber to the connection socket (D) of the voltage changer 2. Establish the power connection of the voltage changer. The socket must provide a protective conductor. 3. Turn on the voltage changer at the power switch (C) (position I ). The green pilot lamp (B) lights up. 4. Turn on the constant climate chamber with the main power switch (1) in the lateral control panel Position 0 = off Position I = on Figure 13: Power switch of the voltage changer Dimensions of the voltage changer Width mm 255 Depth (without door handles) mm 360 Depth (incl. cable and door handles) mm 450 Height mm 300 Length of the connection cable to wall socket mm 172 Lateral wall clearance of the constant climate chamber to set up the voltage changer (minimum) mm 400 Electrical connection data of the voltage changer Input side V 115 A 26,9 Output side (to the chamber) V 214 A 13,0 Power frequency Hz 50 / 60 KBF + KMF (E6) 07/2017 page 31/155

32 5. Functional overview of the MB2 chamber controller The MB2 chamber controller controls following parameters inside the chamber: Temperature in C Relative humidity in % r.h. Fan speed in % For the control ranges of temperature and humidity, see climatic diagrams (chap. 17). You can enter the desired set point values in fixed value operation mode directly on the display surface or via the setpoint menu. For program operation the controller offers programming week and time programs. In addition there is a timer program available (stopwatch function). The controller offers various notifications and alarm messages with visual and audible indication and remote alarms via , an event list (trace file) and the graphical display of the measuring values in the in der chart recorder view. The MB2 program controller permits programming temperature and humidity cycles, and specifying the fan speed and special controller functions for each program section. You can enter values or programs directly at the controller or use the APT-COM 3 DataControlSystem software (option) specially developed by BINDER. Operating mode Temperature values Humidity values Actual values Set-point values Functional icons Figure 14: Normal display of the MB2 program controller (sample values) KBF + KMF (E6) 07/2017 page 32/155

33 5.1 Operating functions in normal display Current operating mode Text list for information icons Date, time, authorization level of the logged-in user, memory Quick setpoint entry Continue to next screen Back to Normal display Information Program start Setpoint entry Event list Display of active alarms Access to main menu Figure 15: Operating functions of the MB2 controller in normal display (example values) KBF + KMF (E6) 07/2017 page 33/155

34 5.2 Display views: Normal display, program display, chart-recorder display Press the Change view icon to toggle between normal display, program display and chartrecorder display. Press the Normal display icon to return from program display and chart recorder display back to Normal display. Normal display (actual values / setpoint values) Program display (example: time program) Chart recorder display KBF + KMF (E6) 07/2017 page 34/155

35 5.3 Controller icons overview Navigation icons in Normal display Icon Signification Function Main menu Alarm Event list Setpoint setting Program start Program pause Program cancelling Information Normal display Change view Access from Normal display to the main menu Access from Normal display to the list of active alarms Access from Normal display to the event list Access from Normal display to the setpoint entry menu: setpoint entry for Fixed value operation, turning on/off humidity control, safety controller settings Start a previously entered time or week program, continue a paused time program Pause a running time program Cancel a running time or week program Information on program operation, setpoints, actual values, and the safety controller Return from program display or chart recorder display to Normal display Toggle between Normal display, program display, and chart recorder display Functional icons in individual menus Icon Signification Function Back Update Confirm Close Reset alarm Change keyboard Edit Return from each menu to Normal display Update the event list and alarm messages Take over the entries and exit the menu / continue menu sequence. Exit the menu / cancel menu sequence. Entries are not taken over. When terminating a menu sequence, an information window appears, which must be confirmed. Acknowledge the alarm and mute the buzzer. Change between uppercase and lower case characters, digits and special characters Edit settings of time and week programs KBF + KMF (E6) 07/2017 page 35/155

36 Functional icons in the chart recorder display Icon Signification Function Show legend Hide legend Switch legend Show indications Hide indications History display Curve selection Search Zoom Show scroll buttons Hide scroll buttons Show legend Hide legend Switch between legend pages Show indication Door open (B2) Hide indication Door open (B2) Pause chart recorder and change to history display. Data recording continues. Go to Curve selection submenu in the history display Go to Search submenu in the history display to select the required instant Go to Zoom submenu in the history display to select the zoom factor Show scroll buttons in the history display to scroll to an instant Hide scroll buttons in the history display to scroll to an instant Information icons referring to chamber conditions Icon Text information Condition Idle mode Temperature range Humidity range Door open Humidity off Controller is in Idle mode Current actual temperature value outside the tolerance range Current actual humidity value outside the tolerance range Chamber door is open The humidification / dehumidification system is turned off Information icon in the touchscreen calibration menu Symbol Information Remaining time to touch the display when calibrating the touchscreen. KBF + KMF (E6) 07/2017 page 36/155

37 5.4 Operating modes The MB2 program controller operates in the following operating modes: Idle mode The controller is not functional, i.e., there is no heating or refrigeration and no humidification or dehumidification. The fan is off. The chamber approximates ambient values. You can activate and deactivate this operating mode with the Idle mode control contact in Fixed value operating mode (chap. 7.3), time program operation (chap ) and week program operation (chap ). Fixed value operating mode The controller operates as a fixed-point controller, i.e., set-points for temperature, humidity, and fan speed can be defined, which are then maintained until the next manual change (chap. 7.1). Timer program operation Stopwatch function: during an entered duration the controller constantly equilibrates to the setpoints entered in Fixed value operation mode. Time program operation An entered time program for temperature and humidity is running. The controller offers 25 program memory places with 100 program sections each. The total number of program sections of all programs is unlimited Week program operation An entered week program for temperature and humidity is running. The controller offers 5 program memory places with 100 switching points each. The switching points can be distributed over all days of the week. KBF + KMF (E6) 07/2017 page 37/155

38 5.5 Controller menu structure Use the navigation icons in the screen footer in Normal display to access the desired controller functions. The available functions depend on the current authorization level Service, Admin or User (chap. 13.1). This is selected either during login or can be available without password protection. Main menu: program settings, further information, Service submenu. The Settings submenu allows general configuration of the controller. chap List of active alarms chap. 11 Access to the event list chap Setpoint entry for Fixed value operation, turning on/off humidity control, safety controller settings Start/ pause/ cancel an already entered, respectively a running time program or start / cancel an already entered, respectively a running week program chap. 7, 6.3, 12.2 chap. 9.1, 9.2, 10.1 Unless noted otherwise, the figures show the functional range, which is available for the user with Admin authorization level. KBF + KMF (E6) 07/2017 page 38/155

39 5.5.1 Main menu The main menu provides access to the general configuration of the controller as well as to program entry and the user administration. Additionally there are support functions like a contact page or the display calibration depending on the available angle. Press the Main menu icon to access the main menu from Normal Display. Press the Back icon to return from each setting menu to Normal Display. The main menu provides the following functions and submenus. User management: login and logout, password management chap. 13 Chamber information chap Settings submenu (not visible for user with User authorization level) Program entry submenu for time and week programs chap. 14 chap. 9 and 10 Service submenu chap BINDER Service contact page chap Calibrating the touch screen chap Back to Normal Display Settings submenu Settings of many general controller functions and network settings (chap. 14). Available only for users with Service and Admin authorization level Service submenu Access to service data, controller reset to factory settings (chap ) Available only for users with Service and Admin authorization level. Full functional range only for BINDER Service (users with Service authorization level). Programs submenu Access to the controller s program functions (chap. 8, 9, 10) KBF + KMF (E6) 07/2017 page 39/155

40 5.5.2 Settings submenu The Settings submenu is available for users with Service or Admin authorization level. It serves to enter date and time, select the language for the controller menus and the desired temperature unit and to configure the controller s communication functions. Path: Main menu > Settings Setting the temperature unit, menu language chap. 14.1, 14.2 Setting date and time chap Setting the display brightness, continuous operation and screen saver Settings for the measurement chart: storage interval, storage values, minimum and maximum values Setting the tolerance range and delay time for tolerance range alarm Configuration of the optional RS485 interface, setting of the device address chap chap chap chap Entry of the MAC address and IP address chap Password protection for web server access chap Configuration of the server, assignment of addresses chap Back to main menu Service submenu The Service submenu is available for users with Service or Admin authorization level. When loggedin with Admin authorization level the user will find information to tell the BINDER Service in service case. Path: Main menu > Service Serial number of the chamber, setup version of the controller software chap No function Information for BINDER Service Reset to factory settings Back to main menu (view with Admin authorization level) KBF + KMF (E6) 07/2017 page 40/155

41 5.6 Principle of controller entries In the selection and entry menus there are icons displayed in the footers which you can use to take over the entry or cancel it. Selection menu (example) Entry menu (example) After completing the settings there are the following possibilities: Press the Confirm icon to take over the entries and exit the menu or continue the menu sequence. Press the Close icon to exit the menu or cancel the menu sequence without taking over the entries. When terminating a menu sequence, an information window appears, which must be confirmed. 5.7 Performance during and after power failures During a power failure, all controller functions are shut down. The zero-voltage relay alarm output (option, chap. 19.5) is switched to alarm position for the whole duration of the power failure. After the power returns, all functions return to the same status the chamber had before power failure.. The controller continues to function in the original operating mode it was in previously before the power failure occurred. Performance after power failure in Idle mode Control is deactivated Performance after power failure in Fixed value operation mode All functions return to the same status the chamber had before power failure. The set-points are immediately resumed. Performance after power failure during time program operation The program is resumed at the point where the interruption occurred with the latest set-points reached during the program run. Performance after power failure during week program operation The week program continues with the values corresponding to the current time. Power failure and power return are noted in the event list (chap. 15.3). If during power failure an alarm has occurred (tolerance range, safety controller, temperature safety device class 3.3 (option), confirm the alarm. See chap KBF + KMF (E6) 07/2017 page 41/155

42 5.8 Performance when opening the door When you open the door the fan starts running with minimum speed. After 60 seconds from opening the door, heating, refrigeration, humidification, dehumidification and fan turn off. After closing the door, heating, refrigeration, humidification, dehumidification and fan turn on again. 6. Start up 6.1 Turning on the chamber After connecting the supply lines (chap. 4), turn on the chamber by its main power switch (1). The lit pilot lamp shows the chamber is ready for operation. When the main power switch is turned on and yet the controller display is dark, the display is in stand-by mode. Press on the touchscreen to activate it. Open the water-tap for freshwater supply. Alternatively, fill the freshwater can (option, chap. 19.9). The humidifying and dehumidifying system must be activated (deactivated operation line Humidity off, chap. 7.3 and setting Control on, chap. 6.3). After the first turning on of the chamber or after an interruption of the power supply the relative humidity will increase after a delay of about 20 minutes. During this period, the relative humidity can drop considerably. Warming chambers may release odors in the first few days after commissioning. This is not a quality defect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day and in a well-ventilated location. WARNING: If customer should use a BINDER chamber running in non-supervised continuous operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two chambers, if this is feasible. 6.2 Controller settings upon start up The window Language selection enables the language selection, in case that it s activated in the Start-up menu. Afterwards occurs a request of the time zone and the temperature unit. The controller will function in the operating mode, which was active before the last shut-down. It controls temperature and humidity in fixed value operating mode to the last entered values and in the program mode to the set points achieved beforehand. KBF + KMF (E6) 07/2017 page 42/155

43 Locked operation Provided that the user administration has been activated by the assignment of passwords for the different authorization types, the controller operation is first locked after turning on the unit, recognizable by the closed lock icon in the header. In the locked view the controller provides all display functions. No setting functions are available. The setpoints are shaded (light grey) in normal display. Changing them by direct entry in the fixed value operating mode is not possible. The functional icons for setpoint entry and program start in the footer are without function. After turning on the unit, user log-in is required to operate the controller (chap. 13.2) Operation without user log-in / without password-protection If the password function has been deactivated, after turning on the unit without user log-in there are those controller functions available, which correspond to the highest authorization level without a password protection. There is no lock icon in the header. 6.3 Turning on/off humidity control Turning off humidity control is required when operating the chamber without water connection in order to avoid humidity alarms. For further information see chap. 17. Press the Setpoint setting icon to access the Setpoint setting menu from Normal display. Setpoints menu. Select Control on/off. You can turn humidity control (humidification and dehumidification) on or off. If the Humidity checkbox is marked, humidity control is active. Mark / unmark the checkbox to change the setting. KBF + KMF (E6) 07/2017 page 43/155

44 7. Set-point entry in Fixed value operating mode In Fixed value operating mode you can enter a temperature set-point, a humidity set-point, the fan speed, and the switching-state of up to 16 operation lines. All settings made in Fixed value operating mode remain valid until the next manual change. They are saved also when turning off the chamber or in case of toggling to Idle Mode or Program Mode. Temperature KBF Temperature KMF Humidity KBF Humidity KMF Setting ranges -5 C / 41 F up to 70 C / 158 F. -15 C / 5 F up to 100 C / 212 F 0 % r.h. up to 80 % r.h. 0 % r.h. up to 10 % r.h. Fan speed 40% up to 100 % Control ranges 0 ºC / 32 F up to 70 C / 158 F without humidity 10 C / 50 F up to 70 C / 158 F with humidity -10 C / 14 F up to 100 C / 212 F without humidity 10 C / 50 F up to 90 C / 194 F with humidity 10 % r.h. to 80 % r.h. see climatic diagrams, chap % r.h. to 98 % r.h. see climatic diagrams, chap. 17. Reduce the fan speed only if required, because the spatial distribution of temperature and humidity will also be reduced. Technical data refers to 100% fan speed. For the control range of temperature and relative humidity, see the temperature / humidity diagrams chap. 17). With set-point type Limit, adapt the safety controller (chap. 12.2) or the temperature safety device class 3.3 (option, chap.12.3) always when you changed the temperature set-point. Set the safety controller set-point or the set-point of temperature safety device class 3.3 (option) by approx. 2 C to 5 C above the controller temperature set-point. Recommended setting: Set-point type Offset with safety controller set-point 2 C. When operating without humidity by setting Control off (chap. 6.3), the humidity tolerance range function is automatically deactivated. When operating without humidity by activated operation line Humidity off (chap. 7.3), set the humidity tolerance range to 0 in order to avoid tolerance range alarms (chap. 11.4). KBF + KMF (E6) 07/2017 page 44/155

45 7.1 Set-point entry for temperature, humidity, and fan speed through the Setpoints menu Press the Setpoint setting icon to access the Setpoint setting menu from Normal display. Setpoints menu. Select Fixed value operation setpoints to access the individual parameters. Select the field Temperature and enter the desired temperature setpoint. KBF setting range: -5 C up to 70 C, KMF setting range: -15 C up to 100 C. Confirm entry with Confirm icon. Select the field Humidity and enter the desired humidity setpoint. KBF setting range: 0% r.h. up to 80% r.h., KMF setting range: 0% r.h. up to 100% r.h. Confirm entry with Confirm icon. Select the field Fan and enter the desired fan speed setpoint. Setting range: 40% up to 100% fan speed. Confirm entry with Confirm icon. When entering a value outside the setting range, the message: Value outside of limits! (Min: xxx, Max: xxx) appears (xxx is a wildcard for the limits of the respective parameter). Press the Confirm icon and repeat the entry with a correct value. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 45/155

46 7.2 Direct setpoint entry for temperature and humidity via Normal display Alternatively you can also enter the setpoints directly via Normal display. Normal display. Select the setpoint you want to change. Example: Temperature entry menu. Enter the desired setpoint and confirm entry with Confirm icon. 7.3 Special controller functions via operation lines Press the Setpoint setting icon to access the Setpoint setting menu from Normal display. You can define the switching state of up to 16 operation lines (control contacts). They are used to activate / deactivate special controller functions. Operation line Humidity off serves to turn off the humidity. Operation line Idle mode activates / deactivates the operating mode Idle mode. The other operation lines are without function. Use the Setpoints menu to configure the operation lines. Setpoints menu. Select the field Functions on/off. Functions on/off entry menu. Mark / unmark the checkbox to activate / deactivate the desired function. Activated operation line: switching status 1 (On) Deactivated operation line: switching status 0 (Off) KBF + KMF (E6) 07/2017 page 46/155

47 The operation lines count from right to left. Example: Activated operation line Humidity off = Deactivated operation line Humidity off = Timer program: stopwatch function During an entered duration the controller constantly equilibrates to the setpoints entered in Fixed value operation mode (temperature, humidity, fan speed, configuration of the operation lines). This duration can be entered as a Timer program. During the program runtime, any setpoint changes do not become effective; the controller equilibrates to the values which were active during program start. 8.1 Starting a timer program In Normal display press the Program start icon to access the Program start menu. Program start menu In the field Program type select Timer program. Select the field Program duration and enter the desired program duration. Press the Confirm icon. Select the field Program start and enter the desired start time of the program in the Program start entry menu. Press the Confirm icon. The program delay time until program start begins. Normal display. Information on the bottom of the screen indicates the currently running program and the time already passed. The grey bar shows how much time of the whole time is elapsed Performance during program delay time During the configured program delay time until program start, the controller equilibrates to the current setpoints of Fixed value operation mode. Modifications of these setpoints are possible but become effective only after the timer program is finished. When the configured moment for program start is reached, the program delay time ends and the program starts running. The controller equilibrates to the values which had been active during program start KBF + KMF (E6) 07/2017 page 47/155

48 8.2 Stopping a running timer program Pausing a running timer program Press the Program pause icon to interrupt the program. The program is paused. The program runtime stops running down, the time display flashes. There are the following options: Press the Program start icon to continue the program Press the Cancelling icon to cancel the program Cancelling a running timer program Press the Program cancelling icon to cancel the program. A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be cancelled. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. 8.3 Performance after the end of the program After the end of the program the message Device changes to fixed value operation mode appears on the screen. Press the Confirm icon. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. KBF + KMF (E6) 07/2017 page 48/155

49 9. Time programs The MB2 program controller permits programming time programs with real-time reference. It offers 25 program memory positions with up to 100 program sections each. For each program section you can enter a temperature set-point, fan speed, section duration, type of temperature transition (ramp or step) and the tolerance range. If the safety controller has been set to limit mode, check the setting of the safety controller when changing the temperature set-point, (chap. 12.2). Reduce the fan speed only if required, because the spatial distribution of temperature and humidity will also be reduced. Technical data refers to 100% fan speed. Programming remains saved in case of a power failure or after turning off the unit. Path: Main menu > Programs> Time program 9.1 Starting an existing time program In Normal display press the Program start icon to access the Program start menu. Program start menu In the field Program type select the setting Time program. In the field Program select the desired program. Select the field Program start and enter the desired program start time in the Program start entry menu. Press the Confirm icon. The program delay time until program start begins. The program end is adapted automatically depending on the entered program duration. After completing the settings, press the Confirm icon to take over the entries and exit the menu. The program starts running. If instead you press the Close icon to exit the menu without taking over the entries, the program will not start. Normal display. Information on the bottom of the screen indicates the currently running program and the time already passed. The grey bar shows how much time of the whole time is elapsed. If program duration has been set to infinite, the grey bar is not displayed. KBF + KMF (E6) 07/2017 page 49/155

50 9.1.1 Performance during program delay time During the configured program delay time until program start, the controller equilibrates to the current setpoints of Fixed value operation mode. Modifications of these setpoints are effective. When the configured moment for program start is reached, the program delay time ends and the program starts running. 9.2 Stopping a running time program Pausing a running time program Press the Program pause icon to interrupt the program.. The program is paused. The program runtime stops running down, the time display flashes. There are the following options: Press the Program start icon to continue the program Press the Cancelling icon to cancel the program Cancelling a running time program Press the Program cancelling icon to cancel the program. A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be cancelled. After confirming the message, the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. 9.3 Performance after the end of the program After the end of the program the message Device changes to fixed value operation mode appears on the screen. Press the Confirm icon. As long as the message has not been confirmed, the setpoint of the last program section remains effective. Program the last section as desired. If e.g. heating, refrigeration, humidification an dehumidification shall turn off, activate operation line Idle mode in the last program section. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. KBF + KMF (E6) 07/2017 page 50/155

51 9.4 Creating a new time program Path: Main menu > Programs > Time program Time program menu: overview of the existing programs. Select an empty program place. Enter the program name and, if desired, additional program information in the corresponding fields. Press the Confirm icon. The program view opens (chap. 9.5). 9.5 Program editor: program management Path: Main menu > Programs > Time program Time program menu: overview of the existing programs. Select an existing program (example: program 3) or create a new program (chap. 9.4). The program view opens. Program view (example: program 3). If a new program has been created, there is just one program section. There are the following options: Select a program section to open the section editor (chap. 9.6) Press the Edit icon to open the program editor KBF + KMF (E6) 07/2017 page 51/155

52 Program editor: Edit program menu Select the desired function and press the Confirm icon. The program editor offers following options: Change the program name Copy program Replace program: Replacing an new or an existing program with the copied program. This menu point is visible only after a program has been copied. Delete program Add new section To add a new section, select Add new section and press the Confirm icon. The program view opens. Program view. A new section is always added at the very bottom (example: section 2) Deleting a time program Path: Main menu > Programs > Time program In the Time program menu select the program to be deleted. The program view opens. In the program view press the Edit icon to open the program editor In the program editor select Delete program and press the Confirm icon. The program is deleted. The controller returns to the program view. KBF + KMF (E6) 07/2017 page 52/155

53 9.6 Section editor: section management Path: Main menu > Programs > Time program Select the desired program. Program view. Select the desired program section (example: section 1) Section view (example: section 1). There are the following options: Select a parameter to enter or modify the according value (chap. 9.7) Press the Edit icon to open the program editor Section editor: Edit section menu Select the desired function and press the Confirm icon. The section editor offers following options: Copy section Replace section: Replacing an existing section with the copied section. This menu point is visible only after a section has been copied. Insert section: Adding the copied section. This menu point is visible only after a section has been copied. Delete section Add new section KBF + KMF (E6) 07/2017 page 53/155

54 9.6.1 Add a new program section Section editor: Edit section menu. Select Add new section and press the Confirm icon. Then select whether to insert the new section before or after the current section. Press the Confirm icon. The new section opens Copy and insert or replace a program section Program view. Select the program section to be copied (example: section 1) Section view (example: section 1). Press the Edit icon to open the section editor. Section editor: Edit section menu Select Copy section and press the Confirm icon. The current section (example: section 1) is copied. The controller returns to the section view. Section view (example: section 1). Press the Close icon to change to the program view, if you want to select another section to be replaced or before or after which the copied section shall be inserted KBF + KMF (E6) 07/2017 page 54/155

55 or Press the Edit icon to open the section editor if you want the current section to be replaced or the copied section to be inserted before or after it Program view. Select the section to be replaced or before or after which the copied section shall be inserted (example: section 2) and press the Confirm icon. Section view (example: section 1). Press the Edit icon to open the section editor Select Replace section to replace the selected section with the copied section or Select Insert section to additionally add the copied section. In this case select whether to insert it before or after the selected section. Section editor: Edit section menu Press the Confirm icon Deleting a program section In the program view select the program section to be deleted. The section view opens. In the section view press the Edit icon to open the section editor In the section editor select Delete section and press the Confirm icon. The section is deleted. The controller returns to the section view. KBF + KMF (E6) 07/2017 page 55/155

56 9.7 Value entry for a program section Path: Main menu > Programs > Time program Select the desired program and section. The section view gives access to all parameters of a program section. You can enter or modify the values. Program name and section number Section duration Type of setpoint transition: ramp or step Operation lines Repeating one or several sections within a program Temperature setpoint Temperature tolerance range: minimum and maximum Humidity setpoint Humidity tolerance range: minimum and maximum Fan speed The setting and control ranges for the individual parameters are the same as for Fixed value operating mode (chap. 7) Section duration Section view (partial view). Select the field Duration indicating the time. Duration entry menu. Enter the desired section duration with the arrow keys and press the Confirm icon. Setting range: 0 up to 99 hours 59 min 59 sec. KBF + KMF (E6) 07/2017 page 56/155

57 9.7.2 Set-point ramp and set-point step You can define the type of temperature and humidity transitions for each individual program section. Ramp mode: Gradual changes of temperature and humidity The set-point of a given program section functions as the section s start temperature. During the section s duration, the set-point gradually passes to the set-point of the subsequent program section. The actual value follows the continually changing set-point. If the last program section is in ramp mode and the setpoint shall change within this section, then you must program an additional section (with the shortest possible section duration) to provide the target temperature of the last program section. Otherwise, the setpoint would remain constant during the section s duration. Programming in the ramp mode allows all kinds of temperature and humidity transitions: Gradual changes of temperature and humidity The setpoint changes its value gradually during the entered section duration. The actual value follows the continually moving set-point at any time. Program sections with constant temperature and humidity The setpoints (initial values) of two subsequent program sections are identical; so the temperature and humidity remain constant during the entire duration of the first program section. Sudden changes of temperature and humidity Steps can be programmed in ramp mode as temperature or humidity changes (ramps) that occur during a very short interval. If the duration of this transitional program section is very short (minimum entry 1 sec), the temperature or humidity change will proceed rapidly within the minimum amount of time. Step mode: Sudden changes of temperature and humidity The set-point of any program section functions as the section s target value. At the start of the program section, the unit heats up or cools down and humidifies/dehumidifies the chamber with the maximum speed to reach the entered value; and then it holds it for the remaining section time. Therefore the setpoint temperature remains constant for the section s duration. These changes occur rapidly within the minimum amount of time (minimum entry: 1 second). Programming in the step mode allows only two kinds of temperature and humidity transitions: Programming gradual changes of temperature and humidity (ramps) is impossible in the step mode Program sections with constant temperature and humidity The setpoints (target values) of two subsequent program sections are identical; so the temperature and humidity remain constant during the entire duration of the first program section. Sudden changes of temperature and humidity The entered setpoint of the section is reached as fast as possible and then held constant for the remaining section duration. Selecting the setting Ramp or Step Section view (partial view). In the field Course select the desired setting Ramp or Step. KBF + KMF (E6) 07/2017 page 57/155

58 Ramp and Step mode example (representation of a temperature course) W/ C t/min. Corresponding program table Section No. Duration [hh:mm:ss] Temperature [ C] Humidity [% rh] Fan [%] Ramp or Step 1 00:10: xxxx xxxx Step 2 00:20: xxxx xxxx Step 3 00:10: xxxx xxxx Step 4 00:20: xxxx xxxx Step 5 00:10: xxxx xxxx Ramp 6 00:30: xxxx xxxx Ramp 7 00:30: xxxx xxxx Ramp 8 00:00: xxxx xxxx Ramp Special controller functions via operation lines You can define the switching state of up to 16 operation lines (control contacts). They are used to activate / deactivate special controller functions. Operation line Humidity off serves to turn off the humidity. Operation line Idle mode activates / deactivates the operating mode Idle mode. The other operation lines are without function. Use the Section editor to configure the operation lines. Section view. Select the field Functions on/off. Functions on/off entry menu. Mark / unmark the checkbox of the desired function to activate / deactivate it and press the Confirm icon. The controller returns to the section view. KBF + KMF (E6) 07/2017 page 58/155

59 Section view indicating the operation lines. Activated operation line: switching status 1 (On) Deactivated operation line: switching status 0 (Off) The operation lines count from right to left. Example: Activated operation line Humidity off = Deactivated operation line Humidity off = Setpoint entry Select the field Temperature and enter the desired temperature setpoint. KBF setting range: -5 C up to 70 C, KMF setting range: -15 C up to 100 C. Confirm entry with Confirm icon. The controller returns to the section view. Select the field Humidity and enter the desired humidity setpoint. KBF setting range: 0% r.h. up to 80% r.h., KMF setting range: 0% r.h. up to 100% r.h. Confirm entry with Confirm icon. The controller returns to the section view. Select the field Fan and enter the desired fan speed setpoint. Setting range: 40% up to 100% fan speed. Confirm entry with Confirm icon. The controller returns to the section view Tolerance range You can specify a temperature and humidity program tolerance range for each program section with different values for the tolerance minimum and maximum. When the actual value exceeds the given threshold, the program is interrupted. This is indicated on the display (see below). When the actual temperature is situated again within the entered tolerance limits, the program automatically continues. Therefore, the duration of the program may be extended due to the programming of tolerances. Programming of tolerances may extend program duration. An entry of for the tolerance minimum means minus infinite and an entry of for the tolerance maximum means plus infinite. Entry of these values will never lead to program interruption. The entry of 0 for the tolerance minimum and/or maximum deactivates the respective tolerance function. When requesting rapid value transitions, we recommend not programming tolerance values in order to enable the maximum heating-up, cooling-down, humidification or dehumidification speed. KBF + KMF (E6) 07/2017 page 59/155

60 Section view, showing the temperature tolerance band Select the field Tolerance band min and enter the desired lower tolerance band value. Setting range: to Confirm entry with Confirm icon. The controller returns to the section view. Select the field Tolerance band max and enter the desired upper tolerance band value. Setting range: to Confirm entry with Confirm icon. The controller returns to the section view. Set the tolerance ranges for other parameters accordingly, if desired. If one of the actual values (temperature and/or humidity) is outside the program tolerance range the whole program course is interrupted. During this program interruption time the controller equilibrates to the set-points of the current section. The screen header indicates Program pause (tolerance band). The program runtime indication flashes and does not proceed any further. When the temperature or humidity values are back within the entered program tolerance range, the program continues automatically Repeating one or several sections within a time program You can repeat several subsequent sections together. It is not possible to define the start section the same time also as the target section, therefore you cannot repeat a single individual section. Enter the desired number of repetitions in the field Number of repetitions and the number of the section to start the repetition cycle with in the field Start section for repetition To have sections repeated infinitely, enter the number of repetitions as -1. The selected sections are repeated as many times as selected. Then the program continues. Section view, showing the repetition function Select the field Number of repetitions and enter the desired number of repetitions. Setting range: 1 to 99, and -1 for infinite. Confirm entry with Confirm icon. The controller returns to the section view. Select the field Start section for repetition and enter the section number, at which the repetition should start. Setting range: 1 up to the section before the currently selected section. Confirm entry with Confirm icon. The controller returns to the section view. KBF + KMF (E6) 07/2017 page 60/155

61 9.7.7 Saving the time program Section view. After the all desired values of the program section have been configured, press the Confirm icon to take over the programming. The controller changes to the program view. Program view. Press the Confirm icon to take over the programming. The controller changes to the Normal display. To save the programming it is absolutely required to press the Confirm icon. Otherwise all settings will be lost! There is no confirmation prompt! KBF + KMF (E6) 07/2017 page 61/155

62 10. Week programs The MB2 program controller permits programming week programs with real-time reference. It offers 5 week program places in total with up to 100 shift points for each week program. Path: Main menu > Programs> Week program 10.1 Starting an existing week program In Normal display press the Program start icon to access the Program start menu. Program start menu. In the field Program type select the setting Week program. In the field Program select the desired program. There are no further settings available in the Program start menu for week programs, as they are needed only for time programs. After completing the settings, press the Confirm icon to take over the entries and exit the menu. The program starts running. If instead you press the Close icon to exit the menu without taking over the entries, the program will not start. After starting the week program, the previously entered week program setpoints are active and will be equilibrated according to the current time. Information on the bottom of the screen indicates the currently running program Cancelling a running week program Press the Program cancelling icon to cancel the program. A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be cancelled. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. KBF + KMF (E6) 07/2017 page 62/155

63 10.3 Creating a new week program Path: Main menu > Programs > Week program Week program menu: overview of the existing programs. Select an empty program place. Enter the program name and, if desired, additional program information in the corresponding fields. Press the Confirm icon. The program view opens. Program view. For the first section no weekday is specified. Therefore the section is first marked in red and cannot be saved. KBF + KMF (E6) 07/2017 page 63/155

64 10.4 Program editor: program management Path: Main menu > Programs > Week program Week program menu: overview of the existing programs. Select an existing program (example: program 1). Program view (example: program 1). If a new program has been created, there is just one program section. There are the following options: Select a program section to open the section editor (chap. 10.5) Press the Edit icon to open the program editor Program editor: Edit program menu. Select the desired function and press the Confirm icon. The program editor offers following options: Change program name. This menu also offers to configure the ramp / step mode setting (chap ). Copy program Replace program: Replacing an new or an existing program with the copied program. This menu point is visible only after a section has been copied. Delete program Add new section KBF + KMF (E6) 07/2017 page 64/155

65 To add a new section, select Add new section and press the Confirm icon. The program view opens. Program view. With a new section no weekday is specified. Therefore the section is first marked in red and cannot be saved. A new section is always added at the very bottom (example: section 3). When the section start is specified the sections are automatically arranged in the correct chronological order Deleting a week program Path: Main menu > Programs > Week program In the Week program menu select the program to be deleted. The program view opens. In the program view press the Edit icon to open the program editor In the program editor select Delete program and press the Confirm icon. The program is deleted. The controller returns to the program view. KBF + KMF (E6) 07/2017 page 65/155

66 10.5 Section editor: section management Path: Main menu > Programs > Week program Select the desired program. Program view. Select the desired program section (example: section 1) Section view (example: section 1). There are the following options: Select a parameter to enter or modify the according value (chap. 10.6) Press the Edit icon to open the program editor Section editor: Edit section menu Select the desired function and press the Confirm icon. The section editor offers following options: Copy section Replace section: Replacing an existing section with the copied section. This menu point is visible only after a section has been copied. Insert section: Adding the copied section. This menu point is visible only after a section has been copied. Delete section Add new section KBF + KMF (E6) 07/2017 page 66/155

67 Add a new program section Section editor: Edit section menu. Select Add new section and press the Confirm icon. Program view. With a new section no weekday is specified. Therefore the section is first marked in red and cannot be saved. A new section is always added at the very bottom (example: section 3). When the section start is specified the sections are automatically arranged in the correct chronological order Copy and insert or replace a program section Section editor: Edit section menu Select Copy section and press the Confirm icon. The current section (example: section 1) is copied. The controller returns to the program view. Program view Select the section to be replaced or before or after which the copied section shall be inserted (example: section 2). Press the Confirm icon The controller returns to the section editor KBF + KMF (E6) 07/2017 page 67/155

68 Select Replace section to replace the selected section with the copied section or Select Insert section to additionally add the copied section. Press the Confirm icon. If you selected Insert section the sections are automatically arranged in the correct chronological order. Section editor: Edit section menu Deleting a program section In the program view select the program section to be deleted. The section view opens. In the section view press the Edit icon to open the section editor In the section editor select Delete section and press the Confirm icon. The section is deleted. The controller returns to the section view Value entry for a program section Path: Main menu > Programs > Week program Select the desired program and section. The setting and control ranges for the individual parameters are the same as for Fixed value operating mode (chap. 7) Set-point ramp and set-point step modes The explanation of the settings Ramp or Step is given in chap You can define the type of temperature and humidity transitions for the entire week program. Select the desired program and press the Edit icon to open the program editor. In the program editor select the Change program name function and press the Confirm icon. KBF + KMF (E6) 07/2017 page 68/155

69 Change program name menu. In the field Course select the desired setting Ramp or Step and press the Confirm icon Weekday In the field Weekday select the desired weekday. Section view. With Daily selected, this section will run every day at the same time Start time Section view. Select the field Moment. Entry menu Moment. Select with the arrow keys the desired start moment of the section and press the Confirm icon. KBF + KMF (E6) 07/2017 page 69/155

70 Setpoint entry Select the field Temperature and enter the desired temperature setpoint. KBF setting range: -5 C up to 70 C, KMF setting range: -15 C up to 100 C. Confirm entry with Confirm icon. The controller returns to the section view. Select the field Humidity and enter the desired humidity setpoint. KBF setting range: 0% r.h. up to 80% r.h., KMF setting range: 0% r.h. up to 100% r.h. Confirm entry with Confirm icon. The controller returns to the section view. Select the field Fan and enter the desired fan speed setpoint. Setting range: 40% up to 100% fan speed. Confirm entry with Confirm icon. The controller returns to the section view Special controller functions via operation lines You can define the switching state of up to 16 operation lines (control contacts). They are used to activate / deactivate special controller functions. Operation line Humidity off serves to turn off the humidity. Operation line Idle mode activates / deactivates the operating mode Idle mode. The other operation lines are without function. Select the desired program and section. You can set the operation lines in the Functions on/off field. For details please refer to chap KBF + KMF (E6) 07/2017 page 70/155

71 11. Notification and alarm functions 11.1 Notification and alarm messages overview Notifications Notifications are indicated by information icons displayed in the screen header in Normal display An information icon serves as an indication of a certain condition. If this condition persists, in some cases an alarm will be triggered after a fix or configurable interval. As long as the condition persists, the information icon therefore continues to be displayed also in state of alarm. If during alarm the conditions ends, e.g., if during a tolerance range alarm the actual value returns to within the tolerance range, the information icon disappears, whereas the alarm will continue until manual acknowledgement. Press the flash icon next to the information icon to access the corresponding text information. Normal display showing the text information. The currently valid information texts are highlighted in black (example: Idle mode) Condition Information icon Text information Start after condition occurred The controller is in Idle mode (chap. 5.4). Idle mode immediately The current actual temperature value is outside the tolerance range (chap. 11.4) The current actual humidity value is outside the tolerance range (chap. 11.4) The humidification / dehumidification system is turned off (via operation line and/or by setting Control on/off ) or Temperature setpoint below 0 C or above 95 C Temperature range Humidity range Humidity off immediately immediately immediately Chamber door open Door open immediately Notifications are not shown in the event list. KBF + KMF (E6) 07/2017 page 71/155

72 Alarm messages Condition The current actual temperature value is outside the tolerance range (chap. 11.4) The current actual humidity value is outside the tolerance range (chap. 11.4) Alarm message Temperature range Humidity range Start after condition occurred after configurable time after configurable time Zero-voltage relay alarm output (option) time as alarm start time as alarm start Open chamber door Door open after 5 minutes ---- Power failure immediately Exceeded setpoint of the safety controller class 3.1 Exceeded maximum or minimum temperature (option temperature safety device class 3.3) Temperature sensor defective Safety controller temperature sensor defective Safety controller immediately ---- Temp. safety device e.g or <-<-< or >->-> Safety controller sensor immediately ---- immediately ---- immediately ---- Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown in the event list Messages concerning the humidity system Condition and measures The humidity module is defective. Contact BINDER service The humidity module cannot fill up. In case of freshwater supply via water pipe: The water tap is closed, or the chamber is defective (e.g. inlet valve of humidity module). In case of freshwater supply via freshwater can (option, chap. 19.9): Water can is empty. Humidification is turned off. In case of refrigerating operation, the interior is strongly dehumidified. When the water supply is functional again, the humidity system restarts, or the chamber is defective. The humidity module cannot empty the condensate tank. Wastewater tube obstructed. Check the length and location of the wastewater tube. If appropriate contact BINDER service. Purging is required soon. Turn off and on again the humidity switch to start purging. After a successful purging the notification resets automatically. Maintenance of the humidity system is required. Contact BINDER service. Message Humidity system Freshwater supply Wastewater Humidity system purging req. Humidity module service Start after condition occurred immediately immediately immediately after predefined time (approx. 3-5 months depending on use) after predefined time (approx. 1 year) Messages concerning the humidity system are shown in the event list. KBF + KMF (E6) 07/2017 page 72/155

73 When operating the chamber without water connection, turn off humidity control in the setpoints menu (chap. 6.3) in order to avoid humidity alarms State of alarm 1. Visual indications in Normal display: alarm message, screen header flashing in red color 2. Audible alert, if the buzzer is enabled (chap. 11.5). 3. Switching the zero-voltage relay alarm output (option, chap. 19.5) to transmit the alarm e.g., to a central monitoring system. (c) (a) Normal display in state of alarm (example). (a) Screen header flashing in red color and showing the alarm message (b) Alarm icon on the bottom of the screen: change to the list of active alarms and alarm acknowledgement (b) (c) If applicable, information icon in the screen header. Indication of a certain condition 11.3 Resetting an alarm, list of active alarms Normal display in state of alarm (example). Press the Alarm icon List of active alarms. Press the Reset alarm icon. KBF + KMF (E6) 07/2017 page 73/155

74 Pressing the Reset alarm icon mutes the buzzer for all active alarms. The icon then disappears. Acknowledging while the alarm condition persists: Only the buzzer turns off. The visual alarm indication remains on the controller display. The alarm remains in the list of active alarms. When the alarm condition has ended, the visual alarm indication is automatically cleared. The alarm is then no longer in the list of active alarms. Acknowledging after the alarm condition has ended: The buzzer and the visual alarm indication are reset together. The alarm is then no longer in the list of active alarms. The zero-voltage relay alarm output resets together with the alarm Tolerance range settings In this menu you can set the deviation between the actual value and setpoint which that shall cause a tolerance range alarm. This function only activates after the set-point has been reached once. Path: Main menu > Settings > Various Submenu Various. Select the field Range alarm delay and enter the time in minutes, after which the range alarm shall be triggered. Setting range: 15 min to 120 min. Confirm entry with Confirm icon. Select the field Temperature range and enter the desired value for the temperature range. Setting range: 2 C to 10 C. Confirm entry with Confirm icon. Select the field Humidity range and enter the desired value for the humidity range. Setting range: 5% r.h. to 20% r.h. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. If there are actual values outside the tolerance range the following information icons for the corresponding parameter are displayed: Icon Signification Information Temperature range Humidity range The temperature value is outside the tolerance range The humidity value is outside the tolerance range If the condition persists, an alarm is triggered after the configured interval ( range alarm delay ). It is visually indicated in Normal display. If the alarm buzzer is activated (chap. 11.5) there is an audible alert. The zero-voltage relay alarm output (option, chap. 19.5) switches to transmit the alarm. The alarm is shown in the list of active alarms (chap. 11.3). KBF + KMF (E6) 07/2017 page 74/155

75 11.5 Activating / deactivating the audible alarm (alarm buzzer) Path: Main menu > Settings > Chamber Chamber submenu (example). In the field Audible alarm select the desired setting off or on and press the Confirm icon. 12. Temperature safety devices 12.1 Over temperature protective device (class 1) The chamber is equipped with an internal temperature safety device, class 1 acc. to DIN 12880:2007. It serves to protect the chamber and prevents dangerous conditions caused by major defects. If a temperature of approx. 110 C / 230 F is reached, the over temperature protective device permanently turns off the chamber. The user cannot restart the device again. The protective cut-off device is located internally. Only a service specialist can replace it. Therefore, please contact an authorized service provider or BINDER Service Overtemperature safety controller class 3.1 The chambers are regularly equipped with an electronic overtemperature safety controller (temperature safety device class 3.1 according to DIN 12880:2007). The safety controller is functionally and electrically independent of the temperature control system. If an error occurs, it performs a regulatory function. With option temperature safety device class 3.3 (chap. 12.3), the safety controller is not used. It must be set to the maximum limit value (KBF: 70 C / 158 F, KMF: 100 C / 212 F). Please observe the DGUV guidelines on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120 or ZH 1/119, issued by the employers liability insurance association) (for Germany). The overtemperature safety controller serves to protect the chamber, its environment and the contents from exceeding the maximum permissible temperature. In the case of an error, it limits the temperature inside the chamber to the entered safety controller set-point. This condition (state of alarm) is indicated visually and additionally with an audible alert if the buzzer is enabled (chap. 11.5). The alarm persists until the chamber cools down below the configured safety controller setpoint. Check the setting regularly and adjust it following changes of the set-point or charge. The safety controller only activates after the set-point has been reached once. KBF + KMF (E6) 07/2017 page 75/155

76 Safety controller modes You can select between Limit (absolute) and Offset (relative) safety controller mode Limit: Absolute maximum permitted temperature value This setting offers high safety as a defined temperature limit will not be exceeded. It is important to adapt the safety controller set-point after each modification of the temperature set-point. Otherwise, the limit could be too high to ensure efficient protection, or, in the opposite case, it could prevent the controller from reaching an entered set-point outside the limit range. Offset: Maximum overtemperature above any active temperature set point. The maximum temperature changes internally and automatically with every set-point change. This setting is recommended for program operation. It is important to check the safety controller setpoint and safety controller mode occasionally, as it does not offer a fix, independent limit temperature value, which would never be exceeded. Example: Desired temperature value: 40 C, desired safety controller value: 45 C. Possible settings for this example: Temperature set point Safety controller mode Safety controller set-point 40 C Limit (absolute) 45 C Offset (relative) 5 C Setting the safety controller Press the Setpoint setting icon to access the Setpoint setting menu from Normal display. Setpoints menu. Select the field Safety controller to access the settings. In the field Mode select the desired setting Limit or Offset. Select the corresponding field Limit or Offset according to the selected mode and enter the desired safety controller setpoint. Confirm entry with Confirm icon. KBF + KMF (E6) 07/2017 page 76/155

77 Regularly check the safety controller setting for set-point type Limit or Offset in Fixed value operating mode according to the entered set-point temperature value in program mode according to the highest temperature value of the selected temperature program Set the safety controller set-point by approx. 2 C to 5 C above the desired temperature setpoint. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries Message and measures in the state of alarm The state of alarm is indicated visually in Normal display by the alarm message Safety controller alarm and the screen header flashing in red color. If the buzzer is enabled (chap. 11.5) there is an additional audible alert (chap. 11.2). The alarm remains active until it is acknowledged on the controller and the inner temperature falls below the set safety controller setpoint. Then the heating is released again. Normal display with safety controller alarm. Press the Alarm icon List of active alarms. Press the Reset alarm icon Function check Check the safety controller at appropriate intervals for its functionality. It is recommended that the authorized operating personnel should perform such a check, e.g., before starting a longer work procedure. KBF + KMF (E6) 07/2017 page 77/155

78 12.3 Temperature safety device class 3.3 (option) With the option over/under temperature protective device (temperature safety device class 3.3 acc. to DIN 12880:2007) the chamber is equipped with two additional safety devices (class 3.1 and class 3.2). The combination of the safety devices is regarded as a safety device class 3.3. The temperature safety device, class 3.3, serves to protect the constant climate chamber, its environment and the contents from exceeding the maximum permissible temperature. Please observe the DGUV guidelines on safe working in laboratories (formerly BGI/GUV-I 850-0, BGR/GUV-R 120 or ZH 1/119, issued by the employers liability insurance association) (for Germany). With safety device class 3.1 a maximum value for the temperature is set that the chamber will not exceed due to the regulatory function of the safety device class 3.1. This protection against excessively high temperatures can, for example, serve to protect the constant climate chamber, its environment and the material under treatment from excess temperatures. With safety device class 3.2 a minimum value for the temperature is set that the chamber will not fall below due to the regulatory function of the safety device class 3.2. This protection against excessively low temperatures can, for example, serve to protect sensitive loads from under cooling. Both safety devices are functionally and electrically independent of the temperature control system. If an error occurs, they perform regulatory function. Safety devices class 3.1 (8) and class 3.2 (9) are located in the left lateral control panel. With option temperature safety device class 3.3, the safety controller (chap. 12.2) must be set to the maximum limit value (KBF: 70 C / 158 F, KMF: 100 C / 212 F). (8) Excess temperature Set-point (9) Insufficient temperature Figure 16: Temperature safety device class 3.3 KBF + KMF (E6) 07/2017 page 78/155

79 Temperature safety device class 3.1 (8) (8a) If you turn the control knob (8) to its end-stop (position 10), the safety device class 3.1 protects the appliance. If you set the temperature a little above the set-point, it protects the charging material. If the safety device class 3.1 has taken over control, identifiable by the red alarm lamp (8a) lighting up, the message Temp. safety device on the controller will be displayed and the buzzer will sound, then proceed as follows: Reset the buzzer by pressing the RESET key on the controller Disconnect the chamber from the power supply Have an expert examine and rectify the cause of the fault. Start up the chamber again Setting: To check the response temperature of the safety device class 3.1, turn on the chamber and set the desired set point at the temperature controller. The sections of the scale from 1 to 10 correspond to the temperature range from 0 C / 32 F to 120 C / 248 F and serve as a setting aid. Turn the control knob (8) of the safety device using a coin to its end-stop (position 10) (chamber protection). When the set point is reached, turn back the control knob (8) until its trip point (turn it counter-clockwise). The trip point is identifiable by the red alarm lamp (8a), the message Temp. safety device on the controller display, and the buzzer sounds. Reset the buzzer with the RESET key on the controller. The optimum setting for the safety device is obtained by turning the control knob clockwise by approximately two scale divisions, which shuts off the red alarm lamp (8a). Figure 17: Setting safety device class 3.1 Check the setting regularly and adjust it following changes of the set-point or charge. Function check: Check the temperature safety device class 3.1 at appropriate intervals for its functionality. It is recommended that the authorized operating personnel should perform such a check, e.g., before starting a longer work procedure. KBF + KMF (E6) 07/2017 page 79/155

80 Temperature safety device class 3.2 (9) (9a) The safety device class 3.2 is equivalently set to a minimum temperature the chamber will not fall below. This protection against prohibited low temperatures can, for example, serve to protect sensitive cultures from cooling down too much. If the control knob (9) is turned to its minimum (position 1), the safety device class 3.2 has no effect. If it is set to a temperature somewhat lower than that selected by means of the controller, it functions as a protective device for the material under treatment. If the temperature safety device class 3.2 has assumed regulation, identifiable by the red alarm lamp (9a) lighting up, the message Temp. safety device on the controller display, and the buzzer sounds, please proceed as follows: Reset the buzzer with the RESET key on the controller. Disconnect the chamber from the power supply. Have an expert examine and rectify the cause of the fault. Start up the chamber again Setting: To check the response temperature of the safety device class 3.2, put the chamber into operation and set the desired set point at the temperature controller. The sections of the scale from 1 to 10 correspond to the temperature range from -40 C / -40 F to +160 C / 320 F and serves as a setting aid. Turn the control knob (9) of the safety device by means of a coin to position 1 (thermostat without effect). When the set point is reached, reset the safety device to its trip point (turn it clockwise). The trip point is identifiable by the red alarm lamp (9a), the message Temp. safety device on the controller display, and the buzzer sounds. Reset the buzzer with the RESET key on the controller. The optimum setting for the safety device is obtained by turning the control knob counter-clockwise by approximately two scale divisions, which shuts off the red alarm lamp (9a). Figure 18: Setting safety device class 3.2 Check the setting regularly and adjust it following changes of set-point or charge. Function check: Check the temperature safety device class 3.2 at appropriate intervals for its functionality. It is recommended that the authorized operating personnel should perform such a check, e.g., before starting a longer work procedure. KBF + KMF (E6) 07/2017 page 80/155

81 13. User management 13.1 Authorization levels and password protection The available functions depend on the current authorization level Master, Service, Admin or User. The authorization levels are hierarchical: Every authorization includes all functions of the next lower level. Master authorization level Highest authorization level, only for developers Extensive authorization for controller operation and configuration, outputs/inputs, alarm settings, parameter sets and operating ring display All passwords can be changed in the log out submenu (chap. 13.3). Service authorization level Authorization level only for BINDER service Extensive authorization for controller operation and configuration, access to service data The passwords for Service, Admin and User authorization levels can be changed in the log out submenu (chap. 13.3). Admin authorization level Expert authorization level, for the administrator Authorization for controller configuration and network settings and for operating those controller functions required for operating the chamber. Restricted access to service data. Password (factory setting): 2. The passwords for Admin and User authorization levels can be changed in the log out submenu (chap. 13.3). User authorization level Standard authorization level for the chamber operator Authorization for operating the controller functions required for operating the chamber. No authorization for controller configuration and network settings. The Settings and Service submenus of the main menu are not available. Password (factory setting): 1 The password for the User authorization level can be changed in the log out submenu (chap. 13.3). As soon as a password has been assigned for an authorization level, the access to this level and the related controller functions are only available after log-in with the appropriate password. If for an authorization level no password is assigned, the related controller functions of this level are available for every user without login. If passwords have been assigned for all authorization levels, access to the controller functions is locked without login. KBF + KMF (E6) 07/2017 page 81/155

82 Operation after user login At user login, the authorization level is selected and confirmed by entering the respective password. Following user login, controller operation is available, recognizable by the open-lock icon in the header. The available controller functions correspond to the user s authorization level. Password protection activated for all levels: operation without user login is locked If passwords have been assigned for all authorization levels, the controller is locked without registration of a user. As long as no user is registered, controller operation is locked, recognizable at the closed-lock icon in the header. This requires that the user management has been activated by the assignment of passwords for the individual authorization levels. Password protection for at least one level deactivated: operation without user login is possible If passwords have not been assigned for all authorization levels, after turning on the chamber there are those controller functions available, which correspond to the highest authorization level without password protection. No lock icon is shown in the display header. User login is neither required nor possible. To activate the password protection and user login, perform new password assignment (chap ). KBF + KMF (E6) 07/2017 page 82/155

83 Information window To check the authorization level of the user currently logged-in, select in Normal display the arrow far right in the display header. The information window shows date and time, the controller s free memory space and under Authorization the authorization level of the current user. If passwords have been assigned for all authorization levels, a user without login (password entry) has no authorization. There are only viewing functions available. Display when all authorization levels are password protected and no user has logged in: No authorization level is displayed. If passwords have been assigned only for some of the authorization levels, a user without login (password entry) has access to the functions of the highest authorization level without password protection. Display when only some of the authorization levels are password protected (example: no protection for the User and Admin levels) and no user has logged in: The user s effective authorization (due to lack of password protection) is shown. Example: user with Admin authorization. If passwords have been assigned for some or all of the authorization levels, user login (password entry) provides the authorization for the corresponding password-protected level. Display when at least some of the authorization levels are password protected and a user has logged in. The user s authorization (by password entry) is shown. Example: user with Admin authorization. KBF + KMF (E6) 07/2017 page 83/155

84 13.2 Log in Path: Main menu > User > Log in Controller without a user logged-in Selection of user type (example) All selection possibilities are password protected Controller with logged-in user After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Controller with deactivated password KBF + KMF (E6) 07/2017 page 84/155

85 13.3 Log out Path: Main menu > User > Log out User logoff with Admin authorization Controller with logged-in user Controller without a user loggedin User logoff with Admin authorization Controller with logged-in user Controller without a user loggedin 13.4 User change If the password function has been deactivated (chap ) this function is not available. Path: Main menu > User > User change Controller with logged-in user KBF + KMF (E6) 07/2017 page 85/155

86 User selection (example) All selection possibilities are password protected Controller with loggedin user 13.5 Password assignment and password change This function is not available for a user logged-in with User authorization Password change A logged-in user can change the passwords of his current level and of the next lower level(s). Example: A user with Admin authorization can change the passwords for the Admin and User authorization levels. Path: Main menu > User > Password User logged-in with Service or Admin authorization KBF + KMF (E6) 07/2017 page 86/155

87 Selection of the authorization level (example: view with Admin authorization) Enter desired password. If desired, press the Change keyboard icon to access other entry windows. In the Keyboard switch window you can select different keyboards to enter uppercase and lowercase letters, digits, and special characters. All types of characters can be combined within one single password. Example: access the digit entry window Entry of digits To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation (sample picture). For each character of the password, the required keyboard appears automatically. Then press the Confirm icon. KBF + KMF (E6) 07/2017 page 87/155

88 Deleting the password for an individual authorization level A user logged-in with Service or Admin authorization can delete the passwords of his current level and of the next lower level(s). To do this no password is entered during a password change. Path: Main menu > User > Password Controller with logged-in user (e.g. with Admin authorization) Select the authorization level for which the password shall be deleted. Do NOT enter anything in the Password screen. Press the Confirm icon. The password is deleted. Do NOT enter anything in the Confirm password screen. Press the Confirm icon. KBF + KMF (E6) 07/2017 page 88/155

89 New password assignment for service or admin authorization level when the password function was deactivated If the password protection for an authorization level has been deactivated, i.e., no password is assigned, no login for this level is possible. Therefore access to this authorization level is available without login. If the password for the Service or Admin authorization has been deleted (chap ), a new password can be assigned for the current level and the next lower level(s) without user login. Example: The password for the Admin authorization level was deleted, therefore every user without login has full access to the functions of the Admin authorization level. If access to this level shall become password protected again, the user can assign a new password for the Admin authorization level with the Password function. Path: Main menu > User > Password Controller with deactivated password for Service or Admin authorization Select the authorization level, for which you want to assign a password. (Example: Admin authorization) Enter the desired password. If desired, press the Change keyboard icon to access other entry windows. To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation. For each character of the password, the required keyboard appears automatically. Then press the Confirm icon. KBF + KMF (E6) 07/2017 page 89/155

90 13.6 Activation code Certain functions of the controller can be unlocked with a previously generated activation code. The activation code enables access to functions available only in the Service authorization level by users without a Service authorization. Such functions include e.g., adjustment or extended configurations. The activation code is available in authorization levels. Path: Main menu > User> Activation code Controller with loggedin user Activation code menu. Select the first of the four entry fields. Activation code entry window. Enter the first four characters of the activation code and press the Confirm icon. Select the next of the four entry fields and proceed accordingly until the entire code has been entered. The available functions are indicated by marked checkboxes. Example: Extended configurations available. Activation code menu with entered code (sample view). Press OK to take over the entry Under Expiration date the date of expiry of the code is displayed. KBF + KMF (E6) 07/2017 page 90/155

91 14. General controller settings Most of the general settings can be accessed in the Settings submenu, which is available for users with Service or Admin authorization level. It serves to enter date and time, select the language for the controller menus and the desired temperature unit and to configure the controller s communication functions Selecting the controller s menu language The MB2 program controller communicates by a menu guide using real words in German, English, French, Spanish, and Italian. Path: Main menu > Settings > Chamber Chamber submenu. Select the desired language. Chamber submenu. Select if there shall be a language query after restarting the chamber and press the Confirm icon Selecting the temperature unit Following start-up of the chamber: Or later: Path: Main menu > Settings > Chamber Select the desired temperature unit and press the Confirm icon. KBF + KMF (E6) 07/2017 page 91/155

92 Change of the temperature unit between C and F. If the unit is changed, all values are converted accordingly C = degree Celsius F= degree Fahrenheit 0 C = 31 F 100 C = 212 F Conversion: [value in F] = [value in C] 1, Setting date and time Following start-up of the chamber after language selection: Select the time zone and configure the daylight saving time switch. Or later: Path: Main menu > Settings > Date and time Date and time submenu. Select the field Date / time. Date / time entry menu. Enter date and time and press the Confirm icon. KBF + KMF (E6) 07/2017 page 92/155

93 Date and time submenu. In the field Daylight saving time switch select the desired setting Automatic or Inactive. Date and time submenu. Select the desired time zone and press the Confirm icon. Date and time submenu. Select the desired start of the daylight saving time. Date and time submenu. Select the desired end of the daylight saving time and press the Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries Display configuration Adapting the display parameters This function serves to configure parameters like display brightness and operating times. Path: Main menu > Settings > Display > Display Display submenu. KBF + KMF (E6) 07/2017 page 93/155

94 Select the field Brightness. Move the grey slide to the left or right to define the brightness of the display left = darker (minimum value: 0) right = brighter (maximum value: 100) Press the Confirm icon. Select the field Wait time for screen saver and enter the desired waiting time for the screen saver in seconds. Setting range: 10 sec up to sec. During the waiting time the display is off. Confirm entry with Confirm icon. In the field Activate continuous operation select the desired setting Yes or No. Select the field Begin continuous operation (possible only if continuous operation is activated) and enter the time with the arrow keys. Confirm entry with Confirm icon. Select the field End continuous operation. (only possible if continuous operation is activated) and enter the time with the arrow keys. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries Touchscreen calibration This function serves to optimize the display for the user s individual angular perspective. Path: Main menu > Calibrate touchscreen Normal display. Select Calibrate touchscreen and follow the instructions on the display. You need to touch all four corners of the touchscreen to calibrate it. Appropriate boxes appear successively in each corner. The clock icon shows how much time there is left to touch the currently activated box. If the box is not touched withing this period, calibration is aborted and the display changes to Normal display. After completing the calibration, i.e., touching all four boxes, the display changes to Normal display. KBF + KMF (E6) 07/2017 page 94/155

95 14.5 Network and communication For these settings at least the Admin authorization level is required Serial interfaces The chamber is optionally equipped with a serial RS485 interface. This menu allows to configure the communication parameters of the RS485 interface. The device address is required to recognize chambers with this interface type in a network, e.g. when connecting it to the optional communication software BINDER APT-COM 3 DataControlSystem (chap. 19.1). In this case do not change the other parameters. Path: Main menu > Settings > Serial interfaces Serial interfaces submenu. Select the desired setting in the field Baud rate. Select the desired setting in the field Data format. Select the field Minimum response time and enter the desired minimum response time. Confirm entry with Confirm icon. Select the field Device address and enter the device adress. Factory setting is 1. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 95/155

96 Ethernet Configuration Path: Main menu > Settings > Ethernet Ethernet submenu. In the field IP address assignment select the desired setting Automatic (DHCP) or Manual. With selection Manual you can enter the IPaddress, the subnet mask and the standard gateway manually. Select DNS device name and enter the DNS device name. Confirm entry with Confirm icon. In the field DNS server address select the desired setting Automatic or Manual. With selection Manual you can enter the DNS server address manually. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 96/155

97 Display of MAC address Path: Main menu > Device info > Ethernet Ethernet submenu (example) Web server This controller menu serves to configure the web server. Then you can enter the chamber s IP-address in the Internet. The IP address is available via Chamber information > Ethernet. The BINDER web server opens. Enter the user name and password which have been assigned for the web server in the controller menu. This enables online access to the controller display, to see e.g., the event list or error messages. In this view no settings can be changed. Path: Main menu > Settings > Web server Webserver submenu. In the field Password active select the desired setting Yes or No. Select the field User name and enter the desired user name. Confirm entry with Confirm icon. Select the field Password and enter the desired password. Confirm entry with the Confirm icon. Select the field Automatic log out after and enter the time in minutes after which the webserver shall log out automatically. Setting range: 0 min to min. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 97/155

98 As soon as an alarm was triggered, an is sent to the configured address. Path: Main menu > Settings > address entry: submenu. Select the desired address field and enter the address. You can use the Keybord change icon for entry. Confirm entry with Confirm icon. server settings: submenu. Select the field server to access the settings In the field Authentication select the desired setting None or SMTP auth. With the setting SMTP auth, you can enter a password under password. Select the field user name and enter the desired user name. Confirm entry with Confirm icon. Select the field SMTP mail server URL and enter the SMPT mail server URL. Confirm entry with Confirm icon. Select the field sender and enter the desired sender. Confirm entry with Confirm icon. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 98/155

99 14.6 USB menu When you insert a USB-stick, the USB menu opens. Depending on the user s authorization level, different functions (highlighted in black) are available for the logged-in user. Available functions with User authorization level Available functions with Admin authorization level Available functions with Service authorization level KBF + KMF (E6) 07/2017 page 99/155

100 15. General information 15.1 Service contact page Path: Main menu > Contact 15.2 Current operating parameters Press the Information icon to access the Info menu from Normal display. Info menu. Select the desired information. Select Program operation to see information on a currently running program. KBF + KMF (E6) 07/2017 page 100/155

101 Select Setpoints to see information on the entered setpoints and operation lines. Select Actual values to see information on the current actual values. Select Safety controller to see information on the safety controller status Event list The Event list displays status information and errors of the current day. It enables to view the last 100 events or defective conditions of the chamber. Press the Event list icon to access the event list from Normal display. Event list Press the Update icon to update the event list. Attention: Following a modification of the language setting (chap. 14.1) or the storage interval of the chart recorder (chap. 16.2) the Event list is cleared. KBF + KMF (E6) 07/2017 page 101/155

102 15.4 Technical chamber information Path: Main menu > Device info Chamber name and setup Versions of CPU, I/O module and safety controller Information on digital and analog inputs and outputs and phase angle outputs Information on modbus analog and digital inputs Information on Ethernet connection, MAC address display for BINDER Service for BINDER Service for BINDER Service chap Back to main menu KBF + KMF (E6) 07/2017 page 102/155

103 16. Chart recorder display This view offers graphic representation of the measurement course. Data representation imitates a chart recorder and allows recalling any set of measured data at any point of time taken from the recorded period Views Press the Change view icon to access the pen recorder display Show and hide legend Show legend Hide legend Press the Show legend icon to display the legend on the right side of the display Legend shown on the right side of the display Switch between legend pages Switch legend Press the Switch legend icon to switch between the legend pages Switching between the legend pages 1 st page 2 nd page KBF + KMF (E6) 07/2017 page 103/155

104 Show and hide specific indications Show indications Hide indications Press the Show indications icon to display the indication Door open (B2). Indication Door open displayed History display History display Press the History display icon to change to the history display. History display. The chart recorder is paused. Data recording continues in the background. Move the central red line by tapping and holding to the desired position. The legend at the right side shows the values of the current line position. Then further icons appear: KBF + KMF (E6) 07/2017 page 104/155

105 History display: Curve selection Curve selection Press the Curve selection icon to access the Curve selection submenu. Curve selection submenu. Select the curves to be displayed by checking the checkbox of the corresponding parameter. Press the Confirm icon History display: Search the required instant Search Press the Search icon to access the Search submenu. Search submenu. Select the required instant by entering its date and time and press the Confirm icon KBF + KMF (E6) 07/2017 page 105/155

106 History display: Zoom function Zoom Press the Zoom icon to access the Zoom submenu. Zoom submenu. Select the zoom factor and press the Confirm icon History display: Show and hide scroll buttons to scroll to an instant Show scroll buttons Hide scroll buttons Press the Show scroll buttons icon to access the Page selection submenu. Page selection submenu. Scroll buttons are shown on the left and on the right. Use them to move along the timeline. KBF + KMF (E6) 07/2017 page 106/155

107 16.2 Setting the parameters This menu allows setting the storage interval, the type of values to be shown and the scaling of the temperature and humidity charts. Path: Main menu > Settings > Measurement chart Measurement chart submenu. Select the field Storage interval and enter the desired storage interval. Confirm entry with Confirm icon. The available presentation depends on the pre-selected storage rate. Factory setting: 60 seconds. This means the higher the storage rate, the more precisely but shorter the data representation will be. In the field Storage values select the desired value type to be displayed. For scaling the representation select the desired minimum and maximum temperature or humidity value and enter the desired values. Temperature display range: -20 C up to 110 C. Humidity display range: 0% r.h. up to 100% r.h. Confirm each entry with Confirm icon. CAUTION Setting the storage rate or rescaling (minimum and/or maximum) will clear the measured-value memory and the event list. Danger of information loss. Change the storage rate or scaling ONLY if the previously registered data is no longer needed. After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. KBF + KMF (E6) 07/2017 page 107/155

108 17. Humidification / dehumidification system The chamber is equipped with a capacitive humidity sensor. This results in a control accuracy of up to +/- 3 % r.h. of the set point. The temperature-humidity diagrams (Figure 19) show the possible working ranges for humidity. In the setpoints menu you can turn humidity control (humidification and dehumidification) on or off with the setting Control on/off (chap. 6.3). With humidity control turned off, the humidification module cools down. After activation it will take up to 20 minutes until the humidification function is fully available again. This setting is required when operating the chamber without a water connection in order to avoid humidity alarms. Operation line Humidity off serves to turn off the humidification / dehumidification system in Fixed value operation (chap. 7.3, time program operation (chap ) and week program operation (chap ). This allows configuring the disconnection for individual program sections. When the humidification / dehumidification system is turned off via operation line it remain on standby (filled and heated). Therefore it is immediately available after turning on. The preset temperature and humidity values should be situated within the optimum range (hatched range in Figure 19).Only within this area will the chamber not be exposed to excessive moisture due to condensation. In the short-term set points outside the optimum range can also be targeted. The control accuracies of ± 3 % r.h., however, cannot be guaranteed in this case. KMF: When operating the chamber with activated humidity, humidity control turns off automatically at temperature set-points below 0 C / 32 F or above 95 C / 203 F. The information icon Humidity off is displayed in the screen header in Normal display. When the temperature setpoint is set back to the range from 0 C / 32 F to 95 C / 203 F, humidity control turns on again and the information icon Humidity off disappears. % r.f % r.f C B A A B C KBF temperature-humidity diagram KMF temperature-humidity diagram Figure 19: Temperature-humidity diagrams Range A: Control range of temperature and relative humidity, condensation free range Range B: Discontinuous range (no continuous operation, up to 24 h) Range C: In this range, condensation in the inner chamber is possible C KBF + KMF (E6) 07/2017 page 108/155

109 Heat emission of electrical devices connected inside the chamber may modify the temperature and humidity range. The chambers are equipped with a door heating system to prevent condensation in the door area. If the set points for temperature or humidity are outside the optimum range, condensation can arise in the door area. Condensation by excess humidity. CAUTION Danger of corrosion on the housing after operating at humidity values > 70 % r.h. for a long period. Dry the appliance completely before shut-down: Set the humidity to 0 % r.h. The humidity system must be activated. Set the temperature set point to 60 C / 140 F for approx. 2 hours (Manual mode). Only then, shut down the chamber at the main power switch (1) and close the water supply tap. Having turned off the chamber by the main power switch (1), always close the water supply tap. If you operate the chamber at high humidity and then immediately turn off the chamber, the internal wastewater collector may overflow due to the condensate. This may lead to the emergence of water at the chamber. CAUTION Overflow of the internal wastewater tank due to condensate. Emergence of water at the chamber. Following high humidity operation, do NOT directly turn off the chamber. Pump off the condensate before shut-down: Set the humidity to 0 % r.h. The humidity system must be activated. Operate the chamber for at least 2 hours. Only then, shut down the chamber at the main power switch (1) and close the water supply tap. KBF + KMF (E6) 07/2017 page 109/155

110 17.1 Function of the humidifying and dehumidifying system Humidifying system The humidifying and dehumidifying system is located in the humidity generation module. In a cylindrical container with a volume of about 2 liters an electrical resistance heating evaporates water. The water content is kept exactly at the boiling point, and thus steam can be immediately generated in sufficient quantity for rapid humidity increases or for compensation of humidity losses, e.g. by door openings. Condensation forming on the outer walls of the useable volume is led through a water drain in the outer chamber into the wastewater can which is pumped off automatically to the wastewater pipe when required. Freshwater You can supply the chamber with freshwater via a water pipe or by manually filling a freshwater can (option, chap. 19.9). You can mount the can on the rear of the chamber or place it next to the chamber. In order to ensure accurate humidifying, observe the following points with regard to the freshwater supply: Supply pressure 1 to 10 bar when connecting to a water pipe Water type: deionized (demineralized) water To ensure humidification during 24 hours even at high humidity set-points with manual water supply, we recommend filling the freshwater can (option) at the end of each day. Water intake temperature NOT below +5 C / 41 F and not exceeding 40 C / 104 F. BINDER GmbH is NOT responsible for the water quality provided by the customer. Any problems and malfunctions that might arise following use of water of deviating quality is excluded from liability by BINDER GmbH. Automatic fresh water supply via water pipe With this type of supply, the humidity system is continuously functional. Manual fresh water supply via freshwater can (option, chap. 19.9) With this type of supply, the humidity system is functional only if the water can is sufficiently filled. Check the filling level daily. The water reserve in the can is sufficient for a period, which may last between one and several days, depending on the humidity demand (entered humidity set-point and number of door openings). Waste water The condensation water from the interior is collected in an internal can with a volume of approx. 0.5 liters. It is pumped off via the waste water pipe. Dehumidifying system When the humidity system is activated, the chamber dehumidifies as needed in order to reach the entered humidity set-point inside the control range of temperature and relative humidity (Figure 19). Dehumidification occurs in case of need by means of defined dew point undershoot of several evaporators of the refrigeration system. The condensate which forms is carried away as waste water. If the humidity system is turned off while there are descending temperature curves, then operation of the refrigeration system may cause dehumidification of the charging material. For error indications concerning water supply and humidity system, see chap and 22. KBF + KMF (E6) 07/2017 page 110/155

111 18. Defrosting at refrigerating operation BINDER constant climate chambers are very diffusion-proof. To ensure high temperature precision there is no automatic cyclic defrosting device. The DCT refrigerating system largely avoids icing of the evaporation plates. However, at very low temperatures the moisture in the air can condense on the evaporator plates leading to icing. Always close the door properly. Operation with temperature set-points > +5 C / 41 F at an ambient temperature of 25 C / 77 F: The air defrosts the ice cover automatically. Defrosting is continually performed. Operation with temperature set-points < +5 C / 41 F: Icing on the evaporator is possible. Defrost the chamber manually. With temperature set-points < +5 C / 41 F, regularly defrost the chamber manually: Set the humidity to 0 % r.h. The humidity system must be activated. Set the temperature to 40 C / 104 F (Manual Mode). Let the chamber operate for about 30 minutes with the door closed. Too much ice on the evaporator is noticeable by reduced refrigerating performance. When turning off the chamber following prolonged refrigerating operation below +5 C / 41 F, there is danger of overflowing due to uncontrolled defrosting of icing on the evaporator. CAUTION Uncontrolled defrosting of icing on the evaporator. Danger of overflowing. After several days of refrigerating operation below +5 C / 41 F: Do NOT directly turn off the chamber. Manually defrost the chamber (see description above). Then, shut down the chamber at the main power switch (1) and close the tap of the water supply. Keep removed the access port plugs. KMF: Operation with temperature set-points below 0 C / 32 F: While operating the chamber with set-points below < 0 C / 32 F condensation is possible at the inner surface of the door around the door gasket. In case of heavy condensation, check tightness of the door gasket. After one or two days operation at a set-point < 0 C / 32 F a thin ice layer can cover the inner chamber door, the front margins of the inner kettles and may be the glass window.. The amount depends of the ambient temperature and humidity. This does not influence the proper function of the refrigerating system. Refrigerating performance decreases while operating the chamber at temperatures < 0 C / 32 F due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once a week. KBF + KMF (E6) 07/2017 page 111/155

112 19. Options 19.1 Communication software APT-COM 3 DataControlSystem (option) The chamber is regularly equipped with an Ethernet interface (4) that can connect the BINDER communication software APT-COM 3 DataControlSystem. The actual temperature and humidity values are given at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers with RS 422 interface can be cross-linked. The MAC Address is indicated in the Device info controller menu (chap ). For further information on networking please refer to the operating manual of the BINDER communication software APT-COM RS485 interface (option) With this option, the chamber is equipped with an additional 2-wire RS485 serial interface (7) that can connect the BINDER communication software APT-COM 3 DataControlSystem. The actual temperature and humidity values are given at adjustable intervals. For further information, please refer to the operating manual of the BINDER communication software APT-COM Data logger kits (option) BINDER Data Logger Kits offer an independent long-term measuring system for temperature and humidity, available for different temperature ranges. According to the selected kit, the Data Logger can measure and record also the ambient temperature and humidity values via a second multi-function sensor. BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a real-time function. Measurement data are recorded in the Data Logger and can be read out after the measurement via the RS232 interface of the Data Logger. It offers a programmable measuring interval and permits storing up to measuring values. Reading out is done with the Data Logger evaluation software. You can give out a combined alarm and status protocol directly to a serial printer. KBF: Data Logger Kit TH 70: Multi-function sensor for chamber temperature and humidity: Temperature range -40 C / -40 F up to 70 C / 158 F, humidity range 0% r.h. up to 100% r.h. KBF: Data Logger Kit TH 70/70: Multi-function sensor for chamber temperature and humidity: Temperature range -40 C / -40 F up to 70 C / 158 F, humidity range 0% r.h. up to 100% r.h. Multi-function sensor for ambient temperature and humidity: Temperature range -40 C / -40 F up to 70 C / 158 F, humidity range 0% r.h. up to 100% r.h. KMF: Data Logger Kit TH 100: Multi-function sensor for chamber temperature and humidity: Temperature range -40 C / -40 F up to +100 C / 212 F, humidity range 0% r.h. up to 100% r.h. KMF: Data Logger Kit TH 100/70: Multi-function sensor for chamber temperature and humidity: Temperature range -40 C / -40 F up to +100 C / 212 F, humidity range 0% r.h. up to 100% r.h. Multifunction sensor for ambient temperature and humidity: Temperature range -40 C / -40 F up to 70 C / 158 F, humidity range 0% r.h. up to 100% r.h. For detailed information on installation and operation of the BINDER Data Logger, please refer to the mounting instructions Art. No and to the original user manual of the manufacturer, supplied with the data logger. KBF + KMF (E6) 07/2017 page 112/155

113 19.4 Analog outputs for temperature and humidity (option) With this option the chamber is equipped with analog outputs 4-20 ma for temperature and humidity. These outputs allow transmitting data to external data registration systems or devices. The connection is realized as a DIN socket (3) in the right lateral control panel as follows: ANALOG OUTPUT 4-20 ma DC PIN 1: Temperature PIN 2: Temperature + PIN 3: Humidity PIN 4: Humidity + Temperature range: -10 C / 14 F to +100 C / 212 F Humidity range: 0 % r.h. to 100 % r.h. A suitable DIN plug is enclosed. Figure 20: Pin allocation of DIN socket (3) for option analog outputs 19.5 Zero-voltage relay alarm outputs for temperature and humidity (option) The chamber equipment with optional zero-voltage relay outputs for temperature and humidity (option) permits the transmission of alarms to a central monitoring system. Connection is established via a DIN socket (6) located on the right lateral control panel. Figure 21: Pin configuration of the DIN socket (6) Temperature contact 1 Pin 1: Pin 2 Pin 2: Make Humidity contact 1 Pin 3: Pin 2 Pin 4: Make In case of a temperature alarm, pins 1 and 2 are open; with humidity alarm, pins 3 and 4 are open. This happens simultaneously with the alarm message shown on the controller display. In case of power failure, both contacts are open. Maximum loading capacity of the switching contacts: 24V AC/DC - 2,5A Electrical hazard. Danger of death. DANGER Damage to switching contacts and connection socket. Do NOT exceed the maximum switching load of 24V AC/DC 2.5A. Do NOT connect any devices with a higher loading capacity. A temperature and humidity alarm message will remain visible on the controller display during the whole time of the alarm transmission via the zero-voltage relay outputs. As soon as the cause of the alarm is rectified, you can reset the alarm transmission via the zero-voltage relay outputs together with the alarm message on the controller. KBF + KMF (E6) 07/2017 page 113/155

114 In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the duration of the power failure. Afterwards, both contacts will close automatically. When using the communication software APT-COM 3 DataControlSystem (option, chap. 19.1) via the Ethernet interface of the constant climate chamber for data acquisition, the alarm is not automatically transmitted to the APT-COM protocol. Set the tolerance limits for recording limit value excesses separately in APT-COM Water protected internal socket (option for KBF not for UL chambers) The internal socket is splash proof. IP system of protection V 1N ~ Hz Charge max. 500 W Maximum permitted operating temperature with this option: 90 C / 194 F. Exceeding the permitted maximum temperature. Electrical hazard. Danger of death. Damage to the internal socket. WARNING Do NOT exceed the temperature set-point of 90 C / 194 F. Set the safety controller to 90 C / 194 F. With optional temperature safety device, class 3.3, set the mechanical thermostat class 3.1 to 90 C / 194 F. Heat emission of electrical devices connected inside the chamber may modify the temperature and humidity range. CAUTION Risk of short circuit. Damage to the chamber. Use the supplied plug only (IP protection type 67). Plug it in and tighten it by screwing to secure contact. If the socket is not used, close the screw lid and turn it to secure. KBF + KMF (E6) 07/2017 page 114/155

115 19.7 Additional flexible Pt 100 temperature sensor (available via BINDER INDIVIDUAL customized solutions) An additional flexible temperature sensor Pt100 allows measuring the temperature of the charging material by means of an independent measuring system utilizing Pt 100 entry. The Pt 100 sensor s top protective tube can be immersed into liquid substances Pt 100 Sensor PIN 1: Pt 100 PIN 2: Pt 100 PIN 3: Pt DIN socket (5) A suitable DIN plug is enclosed DIN plug Figure 22: Pin configuration of the DIN socket (5) in the right lateral control panel Technical data of the Pt100 sensor: Three-wire technique Class B (DIN EN 60751) Temperature range up to 320 C / 608 F Stainless steel protective tube with a length of 45 mm / 1.78 in, material no Object temperature display with flexible Pt 100 temperature sensor (option) The object temperature display enables the determination of the actual temperature of the charging material during the whole process. The object temperature is measured via a flexible Pt100 temperature sensor and can be viewed on the controller display. You can immerse the sensor top protective tube of the flexible Pt 100 into liquid substances. Normal display with object temperature display (sample values) The object temperature data are put out together with the data of the temperature controller and can be documented by the communication software APT-COM (option, chap. 19.1) developed by BINDER. Technical data of the Pt100 sensor: Three-wire technique Class B (DIN EN 60751) Temperature range up to 320 C / 608 F Stainless steel protective tube with a length of 45 mm / 1.78 in, material no KBF + KMF (E6) 07/2017 page 115/155

116 19.9 External freshwater and wastewater cans (option) If no suitable in-house water connection is available, you can manually supply water by filling the optional external freshwater can. There is an additional external water can for the waste water. Volume: 20 liters / 0.71 cu.ft. The cans are placed in holding devices. You can affix them directly at the rear of the chamber or place them next to the chamber. Figure 23: Rear view KMF with installed external water cans (option) Mounting the freshwater can (1) Fixing (if required) Hang the can with its holding device on its 4 carriers. You can install it either at the left or the right side. Cable connection (15) Freshwater hose connection (16) Figure 24: Freshwater can (option) KBF + KMF (E6) 07/2017 page 116/155

117 (2) Cable connections Connect the plug of the cable to the socket (10) at the rear of the chamber. The socket (10) is marked with a sticker: Socket (10) (14) (13) Figure 25: Connections at the chamber rear (3) Hose connections Plug the freshwater hose into the hose connection (16) above the freshwater can and secure it with a hose clamp. You can use a part of the standard supplied water hose. Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwater connection IN (13) at the rear of the chamber. When the freshwater can is empty, the message Freshwater supply will be displayed on the controller (chap ), the buzzer sounds, and the humidification module turns off. After acknowledging the alarm, the humidification module tries to fill up and start operating. To guarantee humidification during 24 hours even at high humidity set-points with manual water supply, we recommend filling the freshwater can (option) at the end of each day. KBF + KMF (E6) 07/2017 page 117/155

118 Mounting the wastewater can (1) Fixing (if required) Hang the can with its holding device on its 4 carriers at the free space next to the freshwater can. (2) Hose connections Plug the wastewater hose to the hose connection (17) of the wastewater can and secure it with a hose clamp. You can use a part of the standard supplied water hose. Hose connection of the wastewater can (17) Figure 26: Wastewater can (option) Plug the free hose edge to the wastewater connection OUT (14) at the rear of the chamber and secure it with a hose clamp. You can remove the wastewater can with its holding device for emptying (disconnect the hose first before emptying). Overflow of the wastewater can. Damage to the surrounding. CAUTION Empty the wastewater can in a timely manner before it is full. Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the wastewater can. KBF + KMF (E6) 07/2017 page 118/155

119 Mounting with wastewater recycling When the chamber interior is clean, you can reuse the wastewater from the chamber. Connect the wastewater connection OUT (14) of the chamber with the freshwater hose connection (18) of the freshwater can. The wastewater can is not used in this case. CAUTION Soiling of the vapor humidification system. Damage to the chamber. Reuse wastewater ONLY with a clean chamber interior. In case of soiling / contamination of the interior, conduct the wastewater to the wastewater connection or use the wastewater can. (1) Fixing of the freshwater can (if required) Hang the can with its holding device on its 4 carriers. You can install it either at the left or the right side. (2) Cable connections of the freshwater can Connect the plug of the cable to the socket (10) at the rear of the chamber as described in chap (3) Hose connections Plug the wastewater hose into the hose connection (18) of the freshwater can and secure it with a hose clamp. You can use a part of the standard supplied water hose. Plug the free hose edge to the wastewater connection OUT (14) at the rear of the chamber and secure it with a hose clamp. Hose connection of the freshwater can (18) Figure 27: Freshwater can (option) Bringing a source of humidity into the inner chamber may increase wastewater production. Regularly check the filling level of the freshwater can. KBF + KMF (E6) 07/2017 page 119/155

120 19.10 BINDER Pure Aqua Service (option) The optional BINDER water treatment system (disposable unit) is available to treat tap water. The lifetime of the unit depends on water quality and the amount of treated water used. The measuring equipment to assess the water quality is reusable. For detailed information on operating the water treatment system BINDER Pure Aqua Service and its function, please refer to the operating manual delivered with BINDER Pure Aqua Service. 20. Maintenance, cleaning, and service 20.1 Maintenance intervals, service Electrical hazard. Danger of death. DANGER The chamber must NOT become wet during operation or maintenance works. Do not remove the rear panel of the chamber. Before conducting maintenance work, turn off the chamber at the main power switch and disconnect the power plug. General maintenance work must be conducted by licensed electricians or experts authorized by BINDER. Maintenance work at the refrigeration system must only be conducted by qualified personnel who underwent training in accordance with EN 13313:2010 (e.g. a refrigeration technician with certified expert knowledge acc. to regulation 303/2008/EC). Follow the national statutory regulations. Ensure regular maintenance work is performed at least once a year and that the legal requirements are met regarding the qualifications of service personnel, scope of testing and documentation. All work on the refrigeration system (repairs, inspections) must be documented. The warranty becomes void if maintenance work is conducted by non-authorized personnel. Have conducted regular maintenance work on the steam humidifier at least once a year. The operating behavior and the maintenance intervals of the humidifier essentially depend on the available water quality and the amount of steam produced in the meantime. We recommend cleaning the condensers at least twice a year. A qualified technician must perform cleaning. Replace the door gasket only when cold. Otherwise, the door gasket may become damaged. With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several times a year. KBF + KMF (E6) 07/2017 page 120/155

121 We recommend taking out a maintenance agreement. Please consult BINDER Service: BINDER telephone hotline: +49 (0) BINDER fax hotline: +49 (0) BINDER hotline: service@binder-world.com BINDER service hotline USA: or x3 (toll-free in the USA) BINDER service hotline Asia Pacific: or BINDER service hotline Russia and CIS BINDER Internet website BINDER address BINDER GmbH, post office box 102, Tuttlingen, Germany International customers, please contact your local BINDER distributor Cleaning and decontamination Clean the chamber after each use to avoid potential corrosion damage by ingredients of the test material. Electrical hazard. Danger of death. DANGER Do NOT spill water or cleaning agents over the inner and outer surfaces. Before cleaning, turn off the chamber at the main power switch and disconnect the power plug. Completely dry the appliance before turning it on again Cleaning Disconnect the chamber from the power supply before cleaning. Disconnect the power plug. The interior of the chamber must be kept clean. Thoroughly remove any residues of test material. Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents: Exterior surfaces inner chamber racks door gaskets Instrument panel Zinc coated hinge parts rear chamber wall Standard commercial cleaning detergents free from acid or halides. Alcohol-based solutions. We recommend using the neutral cleaning agent Art. No Standard commercial cleaning detergents free from acid or halides. We recommend using the neutral cleaning agent Art. No Standard commercial cleaning detergents free from acid or halides. Do NOT use a neutral cleaning agent on zinc coated surfaces. Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER service. KBF + KMF (E6) 07/2017 page 121/155

122 We recommend using the neutral cleaning agent Art. No for a thorough cleaning. Any corrosive damage that may arise following use of other cleaning agents is excluded from liability by BINDER GmbH. Any corrosive damage caused by a lack of cleaning, is excluded from liability by BINDER GmbH. Danger of corrosion. Damage to the chamber. CAUTION Do NOT use acidic or chlorine cleaning detergents. Do NOT use a neutral cleaning agent on other kind of surfaces e.g., the zinc coated hinge parts or the rear chamber wall. For surface protection, perform cleaning as quickly as possible. After cleaning completely remove cleaning agents from the surfaces with a moistened towel. Let the chamber dry. Soapsuds may contain chlorides and must therefore NOT be used for cleaning. With every cleaning method, always use adequate personal safety controls. Following cleaning, leave the chamber door open or remove the access port plugs. The neutral cleaning agent may cause health problems in contact with skin and if ingested. Follow the operating instructions and safety hints labeled on the bottle of the neutral cleaning agent. Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes. Contact with skin, ingestion. CAUTION Skin and eye damage due to chemical burns. Do not ingest. Keep away from food and beverages. Do NOT empty into drains. Wear protective gloves and goggles. Avoid skin contact. KBF + KMF (E6) 07/2017 page 122/155

123 Decontamination The operator must ensure that proper decontamination is performed in case a contamination of the chamber by hazardous substances has occurred. Disconnect the chamber from the power supply prior to chemical decontamination. Pull the power plug. Do not use decontamination agents that may cause a hazard due to reaction with components of the device or the charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER service. You can use the following disinfectants: Inner chamber Standard commercial surface disinfectants free from acid or halides. Alcohol-based solutions. We recommend using the disinfectant spray Art. No For chemical disinfection, we recommend using the disinfectant spray Art. No Any corrosive damage that may arise following use of other disinfectants is excluded from liability by BINDER GmbH. With every decontamination method, always use adequate personal safety controls. In case of contamination of the interior by biologically or chemically hazardous material, there are two possible procedures depending on the type of contamination and charging material: 1. Spray the inner chamber with an appropriate disinfectant. Before start-up, the chamber must be absolutely dry and ventilated, as explosive gases may form during the decontamination process. 2. If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave. In case of eye contact, the disinfectant spray may cause eye damage due to chemical burns. Follow the operating instructions and safety hints labeled on the bottle of the disinfectant spray. Recommended precautions: To protect the eyes use sealed protective goggles. Eye contact. Eye damage due to chemical burns. Do NOT empty into drains. Wear protective goggles. CAUTION After using the disinfectant spray, allow the chamber to dry thoroughly, and aerate it sufficiently. KBF + KMF (E6) 07/2017 page 123/155

124 20.3 Sending the chamber back to BINDER GmbH If you return a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number (RMA number) that has previously been issued to you. An authorization number will be issued after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us. The authorization number will be issued following receipt of the information below: BINDER product type and serial number Date of purchase Name and address of the dealer from which you bought the BINDER product Exact description of the defect or fault Complete address, contact person and availability of that person Exact location of the BINDER product in your facility A contamination clearance certificate (chap. 26) must be faxed in advance The authorization number must be applied to the packaging in such a way that it can be easily recognized or be recorded clearly in the delivery documents. For security reasons we cannot accept a chamber delivery if it does not carry an authorization number. Return address: BINDER GmbH Abteilung Service Gänsäcker Tuttlingen Germany 21. Disposal 21.1 Disposal of the transport packing Packing element Material Disposal Straps to fix packing on pallet Plastic Plastic recycling Wooden transport box (option) Non-wood (compressed matchwood, IPPC standard) Wood recycling with metal screws Metal Metal recycling Pallet Solid wood (IPPC standard) Wood recycling with foamed plastic stuffing PE foam Plastic recycling Transport box Cardboard Paper recycling with metal clamps Metal Metal recycling Top cover Cardboard Paper recycling Edge protection Styropor or PE foam Plastic recycling Protection of doors and racks PE foam Plastic recycling Bag for operating manual PE foil Plastic recycling Insulating air cushion foil (packing of optional accessories) PE foil Plastic recycling If recycling is not possible, all packing parts can also be disposed of with normal waste. KBF + KMF (E6) 07/2017 page 124/155

125 21.2 Decommissioning Turn off the chamber at the main power switch (1) and disconnect it from the power supply. Close the tap used for the water supply. Turn off humidity control (chap. 6.3). Remove the water installation. Temporal decommissioning: See indications for appropriate storage, chap Final decommissioning: Dispose of the chamber as described in chap to Disposal of the chamber in the Federal Republic of Germany According to Annex I of Directive 2012/19/EU of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as monitoring and control instruments (category 9) only intended for professional use. They must not be disposed of at public collecting points. The chambers bear the symbol for the marking of electrical and electronic equipment manufactured / placed on the market in the EU after 13 August 2005 and be disposed of in separate collection according to Directive 2012/19/EU on waste electrical and electronic equipment (WEEE) and German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG). WEEE marking: crossed-out wheeled bin with solid bar under. A significant part of the materials must be recycled in order to protect the environment. At the end of the device s service life, have the chamber disposed of according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739) or contact BINDER service who will organize taking back and disposal of the chamber according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739). Violation against existing law. CAUTION Do NOT dispose of BINDER devices at public collecting points. Have the device disposed of professionally at a recycling company which is certified according to the German national law for electrical and electronic equipment (Elektround Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739). or Instruct BINDER Service to dispose of the device. The general terms of payment and delivery of BINDER GmbH apply, which were valid at the time of purchasing the chamber. Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling according to Directive 2012/19/EU. The devices must be free from toxic, infectious or radioactive substances in order to eliminate any health hazards to the employees of the recycling companies. Prior to handing the chamber over to a recycling company, it is the user s responsibility that it is free from toxic, infectious or radioactive substances. Prior to disposal, clean all introduced or residual toxic substances from the chamber. Prior to disposal, disinfect the chamber from all sources of infection. Be aware that sources of infection may also be located outside the inner chamber. If you cannot safely remove all toxic substances and sources of infection from the chamber, dispose of it as special waste according to national law. Fill out the contamination clearance certificate (chap. 26) and enclose it with the chamber. KBF + KMF (E6) 07/2017 page 125/155

126 WARNING Contamination of the device with toxic, infectious or radioactive substances. Danger of intoxication. Danger of infection. NEVER take a chamber contaminated with toxic substances or sources of infection for recycling according to Directive 2012/19/EU. Prior to disposal, remove all toxic substances and sources of infection from the chamber. A chamber from which all toxic substances or sources of infection cannot be safely removed must be considered as special waste according to national law. Dispose of it accordingly. The refrigerant used R 134A (1,1,1,2-tetrafluorethane) is not inflammable at ambient pressure. It must not escape into the environment. In Europe, recovery of the refrigerant R 134A (1300) is mandatory according to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements regarding qualification of staff, disposal, and documentation Disposal of the chamber in the member states of the EU except for the Federal Republic of Germany According to Annex I of Directive 2012/19/EU of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as monitoring and control instruments (category 9) only intended for professional use. They must not be disposed of at public collecting points. The chambers bear the symbol for the marking of electrical and electronic equipment manufactured / placed on the market in the EU after 13 August 2005 and be disposed of in separate collection according to the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). WEEE marking: crossed-out wheeled bin with solid bar under. At the end of the device s service life, notify the distributor who sold you the device, who will take back and dispose of the chamber according to the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE). Violation against existing law. CAUTION Do NOT dispose of BINDER devices at public collecting points. Have the device disposed of professionally at a recycling company that is certified according to conversion of the Directive 2012/19/EU into national law. or Instruct the distributor who sold you the device to dispose of it. The agreements apply that were agreed with the distributor when purchasing the chamber (e.g. his general terms of payment and delivery). If your distributor is not able to take back and dispose of the chamber, please contact BINDER service. Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling according to Directive 2012/19/EU. The devices must be free from toxic, infectious or radioactive substances in order to eliminate any health hazards to the employees of the recycling companies. KBF + KMF (E6) 07/2017 page 126/155

127 Prior to handing the chamber over to a recycling company, it is the user s responsibility that it is free from toxic, infectious or radioactive substances. Prior to disposal, clean all introduced or residual toxic substances from the chamber. Prior to disposal, disinfect the chamber from all sources of infection. Be aware that sources of infection may also be located outside the inner chamber. If you cannot safely remove all sources of infection and toxic substances from the chamber, dispose of it as special waste according to national law. Fill out the contamination clearance certificate (chap. 26) and enclose it with the chamber. WARNING Contamination of the device with toxic, infectious or radioactive substances. Danger of intoxication. Danger of infection. NEVER take a chamber contaminated with toxic substances or sources of infection for recycling according to Directive 2012/19/EU. Prior to disposal, remove all toxic substances and sources of infection from the chamber. A chamber from which all toxic substances or sources of infection cannot be safely removed must be considered as special waste according to national law. Dispose of it accordingly. The refrigerant used R 134A (1,1,1,2-tetrafluorethane) is not inflammable at ambient pressure. It must not escape into the environment. In Europe, recovery of the refrigerant R 134A (1300) is mandatory according to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements regarding qualification of staff, disposal, and documentation Disposal of the chamber in non-member states of the EU Alteration of the environment. CAUTION For final decommissioning and disposal of the constant climate chamber, please contact BINDER service. Follow the statutory regulations for appropriate, environmentally friendly disposal. The main board of the constant climate chamber includes a lithium cell. Please dispose of it according to national regulations. The refrigerant used R 134A (1,1,1,2-tetrafluorethane) is not inflammable at ambient pressure. It must not escape into the environment. In Europe, recovery of the refrigerant R 134A (1300) is mandatory according to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements regarding qualification of staff, disposal, and documentation. KBF + KMF (E6) 07/2017 page 127/155

128 22. Troubleshooting Fault description Possible cause Required measures General No power supply. Check connection to power supply. Wrong voltage. Check power supply for correct voltage (chap. 4.4). Check chamber fuse and replace Chamber without function. Chamber fuse has responded. it if appropriate. If it responds again, contact BINDER service. Controller defective. Nominal temperature exceeded by 10 due to chamber failure. Contact BINDER service. Over temperature protective device (class 1) responds. Heating Controller defective. Chamber heating permanently, Semiconductor relay defective. Contact BINDER service. set-point not maintained. Controller not well adjusted, or adjustment interval exceeded. Calibrate and adjust controller. Pt 100 sensor defective. Chamber doesn t heat up. Heating element defective. Semiconductor relay defective Contact BINDER service. Chamber doesn t heat up when turned on. Safety controller responds. Mechanical safety device class 3.1 responds (with option safety device class 3.3). Mechanical safety device class 3.2 responds (with option safety device class 3.3). Inner chamber temperature has reached the safety controller setpoint. Safety controller set too low. Safety controller (chap. 12.2) defective. Limit temperature reached. Too much external heat load. Controller defective. Safety device defective. Semi-conductor relay defective Limit temperature reached. Controller or safety device defective. Acknowledge the alarm on the controller. Check temperature setpoint setting. If appropriate, select suitable safety controller setpoint (chap. 12.2). Contact BINDER service. Acknowledge the alarm on the controller. Check setting of temperature set-point and safety device class 3.1. If appropriate, select suitable limit value. Reduce heat load. Contact BINDER service. Acknowledge the alarm on the controller. Check setting of temperature set-point and safety device class 3.2. If appropriate, select suitable limit value. Contact BINDER service. KBF + KMF (E6) 07/2017 page 128/155

129 Fault description Possible cause Required measures Refrigerating performance Ambient temperature > 25 C / 77 Select cooler place of installation. F (chap.3.4). Low or no refrigerating performance. Humidity Humidity fluctuation: Control accuracy of ± 3 % r.f. is not reached. Humidity fluctuation, together with temperature fluctuation > 1 C with a set-point ca. 3 C above ambient temperature. Low or no dehumidification. Icing at the evaporator plates. Condensation at the walls of the inner chamber. Combination of temperature/humidity values not in the optimum range (see temperature humidity diagram, Figure 19). Compressor not turned on. Electro-valves defective. No or not enough refrigerant. Too much external heat load. Door gasket defective. Door opened very frequently. Place of installation too hot. Capillary tube blocked Not enough refrigerant. Humidity control turned off. Set-point was too long below ambient temperature. Combination of temperature/humidity values not in the optimum range (see temperature humidity diagram, Figure 19) Set-point was too long below ambient temperature, icing in the preheating chamber. Select combination of temperature/humidity values in the optimum range (chap. 17). Contact BINDER service. Reduce heat load. Replace door gasket. Open doors less frequently. Select cooler place of installation or contact BINDER service. Contact BINDER service. Turn on humidity control (chap. 6.3, 7.3). Defrost the unit (chap. 18). Select combination of temperature/humidity values in the optimum range (chap. 17). Defrost the chamber (chap. 18) Low humidity and temperature accuracy Fan speed has been reduced. Set fan speed to 100%. Controller No chamber function Display in standby mode. Press on touchscreen. (dark display). Main power switch turned off. Turn on the main power switch. Menu functions not available. Menu functions not available with Log in with the required higher current authorization level. authorization. No access to controller Password incorrect. Contact BINDER service. Chart recorder function: measured-value memory cleared; information lost. Controller does not equilibrate to setpoints entered in Fixed value operation mode Controller does not equilibrate to program set-points. Program duration longer than programmed. New setting of storage rate or scaling (minimum and/or maximum) (chap. 16.2). Controller is not in Fixed value operation mode. Controller is not in program operation mode, or program delay time is running. Tolerances have been programmed. Change the storage rate or scaling ONLY if the previously registered data are no longer required. Change to Fixed value operation mode. Start the program again. If appropriate, wait for the program delay time. For rapid transition phases, do NOT program tolerance limits in order to permit maximum heating, refrigerating, or humidification speed. KBF + KMF (E6) 07/2017 page 129/155

130 Fault description Possible cause Required measures Controller (continued) Program keeps the last program setpoint constant while in setting ramp. Ramp temperature transitions are only realized as steps. Humidity alarm when operating without water connection. Acknowledging the alarm does not cancel the alarm state. Alarm message: or <-<-< or >->-> Miscellaneous Impaired valve function of hose burst protection. Program line with setting ramp is incomplete. Setting step has been selected. Humidity control turned on. Cause of alarm persists. Sensor rupture between sensor and controller or Pt 100 sensor defective. Short-circuit. Calcification. When programming with setting ramp, define the end value of the desired cycle by adding an additional section with a section time of at least one second. Select setting ramp. Turn off humidity control (chap. 6.3). Remove cause of alarm. If the alarm state continues, contact BINDER service. Contact BINDER service. Remove calcifications by citric acid or acetic acid solutions (chap ). Have a plumber inspect the valve. Only qualified service personnel authorized by BINDER must perform repair. Repaired chambers must comply with the BINDER quality standards. 23. Technical description 23.1 Factory calibration and adjustment This chamber was calibrated and adjusted in the factory. Calibration and adjustment were performed using standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certified since December 1996 by TÜV CERT). All test equipment used is subject to the administration of measurement and test equipment that is also a constituent part of the BINDER QM DIN EN ISO 9001 systems. They are controlled and calibrated to a DKD-Standard at regular intervals. Repeated calibrations are recommended in periods of 12 months Over current protection The devices are equipped with an internal fuse not accessible from outside. If this fuse is blown, please contact an electronic engineer or BINDER service. KBF + KMF (E6) 07/2017 page 130/155

131 23.3 Definition of usable volume The usable volume illustrated below is calculated as follows: A, B, C = internal dimensions (W, H, D) C b c B a, b, c = distance to wall a = 0.1*A b = 0.1*B c = 0.1*C a a V USE = (A - 2 * a) * (B - 2 * b) * (C - 2 * c) c b Figure 28: Determination of the useable volume A The technical data refers to the defined usable volume. Do NOT place samples outside this usable volume. Do NOT load this volume by more than half to enable sufficient airflow inside the chamber. Do NOT divide the usable volume into separate parts with large area samples. Do NOT place samples too close to each other in order to permit circulation between them and thus obtain a homogenous distribution of temperature and humidity KBF Technical Data Chamber size Exterior dimensions Width, net mm / inch 880 / / / / Height, gross (incl. feet/castors) mm / inch 1050 / / / / Depth, net mm / inch 650 / / / / Depth, gross (including door handle, I- triangle, connection and 30 mm for cable) mm / inch 730 / / / / Wall clearance rear (minimum) (spacer) mm / inch 100 / / / / 3.94 Wall clearance side (minimum) mm / inch 160 / / / / 6.29 Doors Quantity of doors Quantity of inner glass doors Interior dimensions Width mm / inch 600 / / / / Height mm / inch 483 / / / / Depth mm / inch 351 / / / / Interior volume l / cu.ft. 102 / / / / 36.0 Steam space volume l / cu.ft. 156 / / / / 45.2 KBF + KMF (E6) 07/2017 page 131/155

132 Chamber size Racks Quantity of racks (regular) Quantity of racks (max.) Maximum load per rack kg / lbs. 30 / / / / 99 Maximum permitted total load kg / lbs. 100 / / / / 331 Weight Weight (empty) kg / lbs. 129 / / / / 805 Temperature data (without humidity) Temperature range C / F 0 to +70 / 0 to +70 / 0 to +70 / 0 to +70 / 32 to to to to 158 Temperature fluctuation ± K ,5 Temperature uniformity (variation) at 25 C / 77 F ± K ,2 0,2 at 40 C / 104 F ± K ,2 0,2 Max. heat compensation at 40 C / 104 F W Climatic data (with humidity) Temperature range C / F +10 to +70 / +10 to +70 / +10 to +70 / +10 to +70 / 50 to to to to 158 Temperature at 25 C / 77 F / 60% r.h. ± K ,1 fluctuation at 40 C / 104 F / 75% r.h. ± K ,1 Temperature at 25 C / 77 F / 60% r.h. ± K ,2 uniformity (variation) at 40 C / 104 F / 75% r.h. ± K ,2 Humidity range % r.h. 10 to to to to 80 Humidity fluctuation at 25 C / 77 F / 60% r.h. ± % r.h ,5 at 40 C / 104 F / 75% r.h. ± % r.h ,5 Recovery time after at 25 C / 77 F / 60% r.h. minutes doors were open for 30 s at 40 C / 104 F / 75% r.h. minutes Electrical data (model versions KBF V, KBF V, KBF V, KBF V) IP system of protection acc. to EN Nominal voltage (+/-10%) at 50 Hz power frequency V at 60 Hz power frequency V Current type 1N~ 1N~ 1N~ 1N~ Power plug shock proof plug Nominal power kw Installation category acc. to IEC II II II II Pollution degree acc. to IEC Over-current release category B, 2 poles Amp Different electrical data for KBF constructed for the USA and Canada (model versions KBF V, KBF V, KBF V, KBF V) Nominal voltage (+/-10%) at 50 Hz power frequency V at 60 Hz power frequency V Current type 2~ 2~ 2~ 2~ Power plug NEMA 6-20P 6-20P 6-20P 6-20P KBF + KMF (E6) 07/2017 page 132/155

133 Chamber size Environment-specific data Noise level (mean value) db (A) Energy consumption at 40 C / 104 F and 75 % r.h. Wh/h Filling weight of refrigerant R 134A (GWP 1300) kg 0,180 0,170 0,380 0,410 All technical data is specified for unloaded chambers with standard equipment at an ambient temperature of +22 C +/- 3 C / 71.6 F +/- 5.4 F and a power supply voltage fluctuation of +/-10%. Technical data is determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007. All indications are average values, typical for chambers produced in series. We reserve the right to change technical specifications at any time. If the chamber is fully loaded, the specified heating up and cooling down times may vary according to the load. Bringing a source of humidity into the inner chamber will affect the minimum humidity specification and may affect the humidity accuracy KMF technical data Chamber size Exterior dimensions Width, net mm / inch 880 / / / Height, gross (incl. feet/castors) mm / inch 1050 / / / Depth, net mm / inch 650 / / / Depth, gross(including door handle, I-triangle, connection, and 30 mm for cable) mm / inch 730 / / / Wall clearance rear (minimum) (spacer) mm / inch 100 / / / 3.94 Wall clearance side (minimum) mm / inch 160 / / / 6.29 Doors Number of doors Number of inner glass doors Interior dimensions Width mm / inch 600 / / / Height mm / inch 483 / / / Depth mm / inch 351 / / / Interior volume l / cu.ft. 102 / / / 24.7 Steam space volume l / cu.ft. 156 / / / 32.4 Racks Quantity of racks (regular) Quantity of racks (max.) Maximum load per rack kg / lbs. 30 / / / 99 Maximum permitted total load kg / lbs. 100 / / / 331 Weight Weight (empty) kg / lbs. 127 / / / 650 KBF + KMF (E6) 07/2017 page 133/155

134 Chamber size Temperature data (without humidity) Temperature range C / F -10 to +100 / 14 to to +100 / 14 to to +100 / 14 to 212 Average heating up time acc. to IEC K/min Average cooling down time acc. to IEC K/min Heating up time from -10 C / 14 F to +100 C / 212 F minutes Cooling down time from +100 C / 212 F to -10 C / 14 F minutes Max. heat compensation at 25 C / 77 F W Climatic data (with humidity) Temperature range C / F +10 to +90 / 50 to to +90 / 50 to to +90 / 50 to 194 Temperature fluctuation *) ± K 0.1 to to to 0.5 Temperature uniformity (variation) *) ± K 0.2 to to to 1.0 Humidity range % r.h. 10 to to to 98 Humidity fluctuation *) ± % r.h. 2,5 2 2 Dew point temperature range C +5 to to to +90 Electrical data (model versions KMF V, KMF V, KMF V) IP system of protection acc. to EN at 50 Hz power frequency V Nominal voltage (+/-10%) at 60 Hz power frequency V Current type 1N~ 1N~ 1N~ Power plug Shock-proof plug Nominal power kw Installation category acc. to IEC II II II Pollution degree acc. to IEC Over-current release category B, 2 poles Amp Deviant electrical data KMF for the USA and Canada (model versions KMF V, KMF V, KMF V) at 50 Hz power frequency V Nominal voltage (+/-10%) at 60 Hz power frequency V Current type 2~ 2~ 2~ Power plug NEMA 6-20P 6-20P 6-20P Environment-specific data Noise level (mean value) db (A) Energy consumption at 85 C / 185 F and 85 % r.h. Wh/h Filling weight of refrigerant R 134A (GWP 1300) kg *) Depending on the set-point. All technical data is specified for unloaded chambers with standard equipment at an ambient temperature of +22 C +/- 3 C / 71.6 F +/- 5.4 F and a power supply voltage fluctuation of +/-10%. Technical data is determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007. All indications are average values, typical for chambers produced in series. We reserve the right to change technical specifications at any time. Refrigerating performance decreases while operating the chamber at temperatures < 0 C / 32 F due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once a week. If the chamber is fully loaded, the specified heating up and cooling down times may vary according to the load. KBF + KMF (E6) 07/2017 page 134/155

135 Bringing a source of humidity into the inner chamber will affect the minimum humidity specification and may affect the humidity accuracy Equipment and options (extract) To operate the constant climate chamber, use only original BINDER accessories or accessories / components from third-party suppliers authorized by BINDER. The user is responsible for any risk arising from using unauthorized accessories. Regular equipment Microprocessor display program controller with 2-channel technology for temperature and humidity Ethernet interface for computer communication Temperature safety device class 3.1 acc. to DIN 12880:2007 Inner glass door with gasket DCT refrigerating system with refrigerant R134a Microprocessor controlled humidifying and dehumidifying system *) (humidity range, see diagram) Sizes 240, 720 and 1020: Four castors (2 lockable) 1 rack (KMF), 2 racks (KBF), stainless steel Access port 30 mm with silicone plug *) A water supply (1 to 10 bar) is necessary for the installation of the humidifying and de-humidifying system (chap. 4.3). If no suitable house water connection is available, you can manually supply water by filling a freshwater can (option, chap. 19.9). Furthermore, a water drain in a max. distance of 3 meters / 9.8 ft. and a max. height of 1 meter / 3.3 ft. is required (chap. 4.2). Options / accessories Additional rack, stainless steel Perforated shelf,stainless steel Reinforced rack with rack lockings Securing elements for additional fastening of racks (4 pieces) KMF: Reinforced inner chamber with 2 reinforced racks Temperature safety device class 3.3 acc. to DIN 12880:2007 Zero-voltage relay alarm outputs for temperature and humidity with DIN plug 6-poles Lockable door Access ports 30 mm or 50 mm or 100 mm with silicone plug Analog outputs 4-20 ma for temperature and humidity with 6 pole DIN socket, DIN plug included Flexible Pt 100 temperature sensor, output to DIN socket (BINDER INDIVIDUAL customized solutions) Object temperature display with flexible Pt 100 temperature sensor Communication interface RS485 BINDER Data Logger kit for temperature / humidity: TH 70 for KBF / TH 100 for KMF (chamber values) or TH 70/70 for KBF / TH 100/70 for KMF (chamber and ambient values) External freshwater and wastewater cans (20 liters / 0.71 cu.ft. each) BINDER Pure Aqua Service Exchange cartridge for BINDER Pure Aqua Service Safety kit for water connection with hose burst protection device and reflux protection device, premounted assembly (BINDER INDIVIDUAL customized solutions) KBF: Voltage changer for operation at 115 Volt KBF: Water protected internal socket 230 V AC Calibration of temperature and humidity including certificate Spatial temperature and humidity measurement including certificate Spatial temperature and humidity measurement acc. to DIN 12880:2007 including certificate Qualification folder KBF + KMF (E6) 07/2017 page 135/155

136 23.7 Spare parts and accessories (extract) BINDER GmbH is responsible for the safety features of the chamber only, provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts. The user is responsible for any risks arising from using unauthorized accessories/components. Chamber size Description Art. no. Rack, stainless steel Perforated rack, stainless steel Reinforced rack with rack lockings Securing elements for additional fastening of racks (4 pieces) Door gasket for glass door Door gasket silicone (kettle) Door gasket silicone (outer door) Intermediate door gasket, silicone Description Art. no. Plug for silicon access port d External freshwater and wastewater cans (20 liters / 0.71 cu.ft. each) BINDER Pure Aqua Service Exchange cartridge for BINDER Pure Aqua Service Safety kit for water connection with hose burst protection device and reflux protection device BINDER Individual Voltage changer for operation at 115 Volt (for KBF) Data Logger Kit TH Data Logger Kit TH 70/ Data Logger Kit TH Data Logger Kit TH 100/ Neutral cleaning agent, 1 kg Validation service Art. no. Qualification folder IQ-OQ (KBF) Qualification folder IQ-OQ (KMF) Qualification folder IQ-OQ-PQ (KBF) Qualification folder IQ-OQ-PQ (KMF) Execution of IQ-OQ DL Execution of IQ-OQ-PQ DL Calibration service Art. no. Calibration of temperature and humidity including certificate (1 measuring point) DL Spatial temperature and humidity measurement including certificate (9 measuring points temperature, 1 measuring point humidity) DL Spatial temperature and humidity measurement including certificate (18 measuring points temperature, 1 measuring point humidity) DL Spatial temperature and humidity measurement including certificate (27 measuring points temperature, 1 measuring point humidity) DL For information on components not listed here, please contact BINDER Service. KBF + KMF (E6) 07/2017 page 136/155

137 23.8 Dimensions size 115 KBF + KMF (E6) 07/2017 page 137/155

138 23.9 Dimensions size 240 KBF + KMF (E6) 07/2017 page 138/155

139 23.10 Dimensions size 720 KBF + KMF (E6) 07/2017 page 139/155

140 23.11 Dimensions size 1020 KBF + KMF (E6) 07/2017 page 140/155

141 24. Certificates and declarations of conformity 24.1 EU Declaration of Conformity for KBF KBF + KMF (E6) 07/2017 page 141/155

142 KBF + KMF (E6) 07/2017 page 142/155

143 KBF + KMF (E6) 07/2017 page 143/155

144 24.2 EU Declaration of Conformity for KMF KBF + KMF (E6) 07/2017 page 144/155

145 KBF + KMF (E6) 07/2017 page 145/155

Translation of the original operating manual

Translation of the original operating manual Operating Manual Translation of the original operating manual KBF-S / KBF-S-UL (E6) Constant climate chambers with RD4 controller Model Model version Art. No. KBF-S 240 KBFS240-230V 9020-0366, 9120-0366

More information

Translation of the original operating manual. Alternating climate chambers with program control

Translation of the original operating manual. Alternating climate chambers with program control Operating Manual Translation of the original operating manual MK (E4) Alternating climate chambers with program control Model Model version Art. No. MK 56 MK056-230V 9020-0316 MK056-240V 9020-0395 BINDER

More information

Model Model version Art. No. Model Model version Art. No. BD V 9010/ FD V 9010/ BD 56

Model Model version Art. No. Model Model version Art. No. BD V 9010/ FD V 9010/ BD 56 Operating Manual BD (E3) Incubators with natural convection BF (E3) Incubators with forced convection ED (E3) Drying and heating ovens with natural convection FD (E3) Drying and heating ovens with forced

More information

Translation of the Original Operating Manual. Safety drying ovens for drying of limited quantities of solvents

Translation of the Original Operating Manual. Safety drying ovens for drying of limited quantities of solvents Operating Manual Translation of the Original Operating Manual FDL 115 Safety drying ovens for drying of limited quantities of solvents with microprocessor program controller RD3 Model Model version Art.

More information

Constant climate chambers

Constant climate chambers Constant climate chambers KBF series Environmental simulation chambers for constant climatic conditions The KBF series was particularly designed for absolutely reliable stability tests and precise maintenance

More information

Constant climate chambers

Constant climate chambers KBF series (E5.2) Constant climate chambers Constant climate chambers The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions.

More information

Constant climate chamber with humidity

Constant climate chamber with humidity KBF series (E5.2) Constant climate chambers Constant climate chamber with humidity The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate

More information

Constant climate chamber with ICH-compliant illumination

Constant climate chamber with ICH-compliant illumination KBF P series (E5.2) Constant climate chambers Constant climate chamber with ICH-compliant illumination The BINDER KBF P constant climate chamber, equipped with ICH-compliant illumination, is the expert

More information

New Generation KBF 240 series

New Generation KBF 240 series Constant climate chambers New Generation 240 series The constant climate chamber is the expert for absolutely reliable stability tests and precise maintenance of constant climatic conditions. With its

More information

Constant climate chambers. KBF 240 (E5.2) - Constant climate chamber. Performance features and equipment:

Constant climate chambers. KBF 240 (E5.2) - Constant climate chamber. Performance features and equipment: Constant climate chambers KBF 240 (E5.2) - Constant climate chamber The KBF series was particularly designed for absolutely reliable stability tests and precise maintenance of constant climatic conditions.

More information

Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range

Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range BENEFITS Homogeneous climate conditions thanks to APT.line technology Automatic

More information

Model KMF 115 Constant climate chambers with expanded temperature / humidity range

Model KMF 115 Constant climate chambers with expanded temperature / humidity range Model KMF 115 Constant climate chambers with expanded temperature / humidity range BENEFITS Homogeneous climate conditions thanks to APT.line technology Automatic water and wastewater management Humidification

More information

Constant climate chambers

Constant climate chambers KBF series (E5.2) Constant climate chambers Constant climate chambers The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions.

More information

Constant climate chambers

Constant climate chambers Constant climate chambers KBF ICH series Environmental simulation chambers for constant climatic conditions with light As regards compliance with standards, performance and functionality, the characteristics

More information

Model KBF 1020 Constant climate chambers with large temperature / humidity range

Model KBF 1020 Constant climate chambers with large temperature / humidity range Model KBF 1020 Constant climate chambers with large temperature / humidity range BENEFITS Homogeneous climate conditions thanks to APT.line technology Automatic water and wastewater management Pressure

More information

Model KMF 720 Constant climate chambers with expanded temperature / humidity range

Model KMF 720 Constant climate chambers with expanded temperature / humidity range Model KMF 720 Constant climate chambers with expanded temperature / humidity range BENEFITS Homogeneous climate conditions thanks to APT.line technology Automatic water and wastewater management Humidification

More information

Environmental simulation chamber for complex alternating climate profiles

Environmental simulation chamber for complex alternating climate profiles MKF series (E3.1) Environmental simulation chambers Environmental simulation chamber for complex alternating climate profiles A BINDER environmental simulation chamber of the MKF series is ideally suited

More information

Model Model version Art. No. C 170 C V-R , C 170-UL C V-R ,

Model Model version Art. No. C 170 C V-R , C 170-UL C V-R , Operating Manual C 170 (E3) CO 2 Incubator with FPI-sensor system and display controller RP1 Model Model version Art. No. C 170 C170-230V-R 9040-0112, 9140-0112 C 170-UL C170-120V-R 9040-0113, 9140-0113

More information

Climate chambers with illumination and humidity The BINDER KBWF series shines with its

Climate chambers with illumination and humidity The BINDER KBWF series shines with its KBWF series (E5.2) Climate chambers with illumination Climate chambers with illumination and humidity The BINDER KBWF series shines with its homogeneous light distribution with natural illumination from

More information

Constant climate chambers

Constant climate chambers KBF series (E5.2) Constant climate chambers Constant climate chambers The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climatic conditions.

More information

BINDER Scientific Product Catalog

BINDER Scientific Product Catalog BINDER Scientific Product Catalog 36 Vacuum drying ovens Vacuum drying ovens 37 VD VDL VDL EX-Zone 2 Absolutely reliable drying Drying with no residues, no danger of incrustation, no oxidation all under

More information

Constant climate chambers

Constant climate chambers Constant climate chambers KBF P 240 (E2) - Constant climate chamber for photostability tests Specialized equipment for photostability tests. Integrated cooling of fans consistently compensates for the

More information

Constant climate chambers

Constant climate chambers Constant climate chambers KBF 115 (E2) - Constant climate chamber The KBF series was particularly designed for absolutely reliable stability tests and precise maintenance of constant climatic conditions.

More information

Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range

Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range The BINDER MKFT series environmental simulation chamber for low temperature

More information

KMF Environmental test chamber for constant climate

KMF Environmental test chamber for constant climate SHENZHEN YIHUAXIN ELECTRNICS C.,LTD Tel:+86-755-86218000 Fax:+86-755-86218111 www.yihuaxin.net yhx@yihuaxin.net Environmental test chambers KMF 240 - Environmental test chamber for constant climate The

More information

Model KBF 240 Constant climate chambers with large temperature / humidity range

Model KBF 240 Constant climate chambers with large temperature / humidity range Model KBF 240 Constant climate chambers with large temperature / humidity range BENEFITS Homogeneous climate conditions thanks to APT.line technology Automatic water and wastewater management Pressure

More information

Vacuum drying oven for non-flammable solvents

Vacuum drying oven for non-flammable solvents VD series 53 Vacuum drying ovens Vacuum drying oven for non-flammable solvents A BINDER vacuum drying oven of the VD series is impressive while gently drying with its homogeneous temperature distribution.

More information

Constant climate chambers

Constant climate chambers Constant climate chambers Always the right climate KBF KBF-S KBF P KBF LQC KMF series EN PRODUCT SELECTION CONSTANT CLIMATE CHAMBERS Features KBF Page 6 KBF-S Page X KBF P Page 9 KBF LQC Page 12 KMF Page

More information

Constant climate chambers. Always the right climate KBF KBF P KBF LQC KMF series

Constant climate chambers. Always the right climate KBF KBF P KBF LQC KMF series Constant climate chambers Always the right climate KBF KBF P KBF LQC KMF series EN1 TABLE OF CONTENTS Advantages 4 Constant climate chambers with large temperature / humidity range KBF 6 Constant climate

More information

testo 316-EX gas leak detector Instruction manual

testo 316-EX gas leak detector Instruction manual testo 316-EX gas leak detector Instruction manual 2 1 Safety and the environment 1 Safety and the environment 1.1. About this document Use > Please read this documentation through carefully and familiarize

More information

Ultra low temperature freezers for long-term storage at -86 C UF V 500, UF V 700

Ultra low temperature freezers for long-term storage at -86 C UF V 500, UF V 700 Ultra low temperature freezers for long-term storage at -86 C UF V 500, UF V 700 02 Ultra low temperature freezers for long-term storage at -86 C 03 LONG-TERM STORAGE OF SAMPLES Always the right design

More information

Plant growth chambers

Plant growth chambers Plant growth chambers KBWF 240 (E5.1) - Plant growth chamber with optimal climatic conditions The closest thing to natural conditions. Making use of the multifaceted programming options, we achieve perfect

More information

Safe Operation Manual. Honeywell IQ Force Gas Detector

Safe Operation Manual. Honeywell IQ Force Gas Detector Safe Operation Manual Honeywell IQ Force Gas Detector Honeywell Analytics 800-663-4164 403-248-9226 Fax 403-575-3708 03JUN2012 P/N 50105843-047 Version 01 http://www.honeywell.com HONEYWELL IQ Force PERSONAL

More information

Model KBWF 240 Growth chambers with light and humidity

Model KBWF 240 Growth chambers with light and humidity Model KBWF 240 Growth chambers with light and humidity The BINDER climate chamber with illumination of the KBW series achieves homogeneous light distribution with its natural illumination. That allows

More information

Product Highlights. BINDER Scientific with New Avantgarde.Line Products

Product Highlights. BINDER Scientific with New Avantgarde.Line Products Product Highlights BINDER Scientific with New Avantgarde.Line Products Growth Storage 3 In vivo and In vitro cultures Sample storage Growth chambers with illumination KBW and KBWF series APT.line Advanced

More information

Telephone dialing device TWG 01

Telephone dialing device TWG 01 Operating Manual Telephone dialing device TWG 01 Art. No. 9052-0005 BINDER GmbH Address Post office box 102 78502 Tuttlingen, Germany Tel. +49 7462 2005 0 Fax +49 7462 2005 100 Internet http://www.binder-world.com

More information

Safety vacuum drying oven for flammable solvents

Safety vacuum drying oven for flammable solvents VDL series 53 Vacuum drying ovens Safety vacuum drying oven for flammable solvents A BINDER safety vacuum drying oven of the VDL series ensures maximum safety when drying organic solvents standard with

More information

PolyGard Single Point Controller SPC-X3-34XX for Combustible Gases

PolyGard Single Point Controller SPC-X3-34XX for Combustible Gases PolyGard Single Point Controller SPC-X3-34XX for Combustible Gases DESCRIPTION Gas measuring and monitoring controller based on stateof-the-art microtechnology with integrated sensor for continuous monitoring

More information

Quality. Water. Water Quality. Testing. Testing. Toxic Gas Detection. Toxic Gas. Detection. Air Quality. Monitoring. Monitoring. Monitoring.

Quality. Water. Water Quality. Testing. Testing. Toxic Gas Detection. Toxic Gas. Detection. Air Quality. Monitoring. Monitoring. Monitoring. Water Quality Labo- Soil/ Soil/ Sedi- Sam- & Air / Dust Toxic gases are present in a wide variety of industrial and civil installations and their detection and monitoring is crucial for the safe working

More information

Digital Heat Block User Manual

Digital Heat Block User Manual Digital Heat Block User Manual Quidel Digital Heat Block Page 1 of 10 General Information Quidel Contact Information Contact Quidel Technical Support from 8:00 a.m. to 5:00 p.m. EST Tel: 800.874.1517 (in

More information

Model KBWF 720 Growth chambers with light and humidity

Model KBWF 720 Growth chambers with light and humidity Model KBWF 720 Growth chambers with light and humidity The BINDER climate chamber with illumination of the KBW series achieves homogeneous light distribution with its natural illumination. That allows

More information

Operating instructions

Operating instructions Operating instructions Capriz 2 24c 28c GB, IE Contents Contents 1 Safety... 3 1.1 Action-related warnings... 3 1.2 Intended use... 3 1.3 General safety information... 4 2 Notes on the documentation...

More information

EN - english. Instructions for installation and operation. Mobile humidity meter DP 207

EN - english. Instructions for installation and operation. Mobile humidity meter DP 207 EN - english Instructions for installation and operation Mobile humidity meter DP 207 Dear customers, Thank you very much for deciding in favour of the mobile humidity measuring instrument DP 207. Please

More information

Operating instructions

Operating instructions The energy you need Operating instructions Betacom 3 24c -A (H-GB) 30c -A (H-GB) GB, IE Contents Contents 1 Safety... 3 1.1 Action-related warnings... 3 1.2 Intended use... 3 1.3 General safety information...

More information

Operating instructions

Operating instructions Operating instructions For the operator Operating instructions ecotec plus Gas-fired wall-hung high-efficiency boiler GB, IE Publisher/manufacturer Vaillant GmbH Berghauser Str. 40 D-42859 Remscheid Telefon

More information

Operating instructions Page 12

Operating instructions Page 12 Operating instructions Page 12 Refrigerator with explosion-proof interior container Read the operating instructions before switching on for the first time 7082 271-00 LKEXv 910 Disposal notes The appliance

More information

Operating instructions

Operating instructions Operating instructions For the operator Operating instructions HOME SYSTEM GB, IE Publisher/manufacturer Vaillant GmbH Berghauser Str. 40 D-42859 Remscheid Tel. +49 21 91 18 0 Fax +49 21 91 18 28 10 info@vaillant.de

More information

Technical Description. Laboratory Climate Test Chamber. System Weiss. Mat. No.: Type WKL 34/40

Technical Description. Laboratory Climate Test Chamber. System Weiss. Mat. No.: Type WKL 34/40 Technical Description Laboratory Climate Test Chamber System Weiss Mat. No.:67592625 Type WKL 34/40 FACTS The test chambers types WTL/WKL with compact exterior dimensions were especially developed for

More information

Standard-Incubators. Avantgarde.Line BD BF series. Classic.Line B BD BF series. Solid.Line BD-S series

Standard-Incubators. Avantgarde.Line BD BF series. Classic.Line B BD BF series. Solid.Line BD-S series Standard-Incubators Avantgarde.Line BD BF series Classic.Line B BD BF series Solid.Line BD-S series EN PRODUCT SELECTION INCUBATORS Classic.Line Classic.Line Avantgarde.Line Solid.Line Classic.Line Avantgarde.Line

More information

VD 53 - Vacuum drying oven for non-flammable solvents

VD 53 - Vacuum drying oven for non-flammable solvents Vacuum drying ovens VD 53 - Vacuum drying oven for non-flammable solvents The VD series offers safe drying with homogenous temperature distribution, thanks to its APT.line air jacket system. ptimum heat

More information

Refrigerated Incubator Model and Operating Instructions

Refrigerated Incubator Model and Operating Instructions Refrigerated Incubator Model 165000 and 165000-2 Operating Instructions N2400379 - Rev. 1 08May2018 1 Contents 1. SAFETY...3 1.1. EMF INTERFERENCE...4 1. PRODUCT INFORMATION...5 1.1 INTRODUCTION...5 2.

More information

Climatic chamber for constant conditions with program control. with ICH compliant illumination and adjustable light cassettes

Climatic chamber for constant conditions with program control. with ICH compliant illumination and adjustable light cassettes Operating Manual APT.line KBF (E5.2) Climatic chamber for constant conditions with program control Model Art. No. KBF 115 (E5.2) 9020-0139, 9120-0139 KBF 115-UL (E5.2) 9020-0184, 9120-0184 KBF 240 (E5.2)

More information

Drying and heating chambers

Drying and heating chambers Drying and heating chambers Avantgarde.Line ED FD FED series Classic.Line E ED FD FED FP M series Solid.Line ED-S FD-S series EN PRODUCT SELECTION DRYING AND HEATING CHAMBERS Classic.Line Avantgarde.Line

More information

MD-MINI Dry Bath. Instruction Manual

MD-MINI Dry Bath. Instruction Manual MD-MINI Dry Bath Instruction Manual Catalog No. MD-MINI www.majorsci.com service@majorsci.com Version 01B Revision on: 2013.07.15 Packing list MD-MINI: 1x MD-MINI Dry Bath 1x MD-MINI Lid 1x Power Cord

More information

SILENT compact /

SILENT compact / SILT compact 2934 0000 / 2934 1000 TRANSLATION OF THE ORIGINAL INSTRUCTIONS FOR USE Made in Germany Ideas for dental technology 21-2245 21052015 / A Contents 1. Introduction...3 1.1 Symbols...3 2. Safety...3

More information

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number

GMA 301. Operation Manual. Worldwide Supplier of Safety Solutions. Part Number Worldwide Supplier of Safety Solutions GfG Instrumentation 1194 Oak Valley Drive #20 Phone: 734-769-0573 Fax: 734-769-1888 E-Mail: info@gfg-inc.com Internet: www.gfg-inc.com GMA 301 Operation Manual Part

More information

Vacuum drying VD VDL series

Vacuum drying VD VDL series Vacuum drying VD VDL series Extremely superior: BINDER Vacuum drying ovens u Extremely fast u Direct heat transfer to the sample material through thermal conducting plates u Fast, condensation-free drying

More information

User Manual. GPD 3000 e. gas detector

User Manual. GPD 3000 e. gas detector User Manual GPD 3000 e gas detector Table of contents Application...3 Unit ppm, Vol.%...4 Operation elements GDP 3000 e...5 Gas sensor...6 Important Information off sensor!...6 Measurement range...7 Turning

More information

Operation Manual for Drying Storage Cabinet with Drying Unit U-5002

Operation Manual for Drying Storage Cabinet with Drying Unit U-5002 Operation Manual for Drying Storage Cabinet with Drying Unit U-5002 (software version 4-09 and above) Aplicable to the following cabinets: MSD-601-52, MSD-1202-52, MSD-1222-52 HSD-241-52, HSD-351-52, HSD-472-52,

More information

External Dimensions: B = 600 mm D = 630 mm H = 2000 mm. Internal Dimensions: B = 451 mm D = 430 mm H = 1598 mm

External Dimensions: B = 600 mm D = 630 mm H = 2000 mm. Internal Dimensions: B = 451 mm D = 430 mm H = 1598 mm External Dimensions: B = 600 mm D = 630 mm H = 2000 mm Internal Dimensions: B = 451 mm D = 430 mm H = 1598 mm Capacity: 328 l Temperaturerange: -15 C to -30 C Housing galvanized sheet steel with white

More information

Combustible Gas Detection and Control System

Combustible Gas Detection and Control System Combustible Gas Detection and Control System DESCRIPTION Gas monitor with built-in combustible gas sensor, wall-mounted, accepts inputs from remote devices such as other gas sensors, temperature or humidity

More information

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6 Digital Gas Transmitters with Auxiliary s & Control Specifications subject to change without notice. USA 181011 Page 1 of 5 DESCRIPTION Digital RS-485 communicating, addressable toxic and combustible gas

More information

Vacuum Controller with Control Unit VC 900

Vacuum Controller with Control Unit VC 900 Operating Instructions Read and observe these operating instructions! Vacuum Controller with Control Unit VC 900 KNF Neuberger GmbH Alter Weg 3 79112 Freiburg Germany Tel. +49 (0)7664 / 5909-0 Fax +49

More information

Manual v1 Page 1 of 19. Manual. Laboratory autoclave

Manual v1 Page 1 of 19. Manual. Laboratory autoclave Manual 20.2.2015 v1 Page 1 of 19 Manual Laboratory autoclave Autoclave using steam for the treatment of material and laboratory processes e.g. sterilization, as an aid in research and development CertoClav

More information

6 POINTS TO CONSIDER WHEN PURCHASING A CONSTANT CLIMATE CHAMBER IN THE PHARMACEUTICAL INDUSTRY BUYER S GUIDE

6 POINTS TO CONSIDER WHEN PURCHASING A CONSTANT CLIMATE CHAMBER IN THE PHARMACEUTICAL INDUSTRY BUYER S GUIDE 6 POINTS TO CONSIDER WHEN PURCHASING A CONSTANT CLIMATE CHAMBER IN THE PHARMACEUTICAL INDUSTRY BUYER S GUIDE 6 points to consider when purchasing a constant climate chamber in the pharmaceutical industry

More information

ICP450. Compressor-cooled incubator. Product specification

ICP450. Compressor-cooled incubator. Product specification Product specification Compressor-cooled incubator ICP450 Ideal for ramp operation with rapid temperature changes during incubation, breeding or storage. With the help of our model selection, with dimensioned

More information

CMF, CMX & CMA series

CMF, CMX & CMA series Features Electrochemical sensor element 4-20 ma or 2-10 Vdc output CMF- IP 44 protection CMX- IP 54 protection CMA- IP 65 protection Wide supply voltage range (18-28 Vdc) Overload and short cicuit protected

More information

Vacuum drying chambers. VD VDL series

Vacuum drying chambers. VD VDL series Vacuum drying chambers VD VDL series EN Vacuum drying chambers Series VD Page 6 Series VDL Page 8 Options and Accessories Page 10 Both BINDER vacuum drying chamber models enable effective drying without

More information

WITH OMRON E5CC CONTROL

WITH OMRON E5CC CONTROL LABORATORY OVENS, FAN FORCED & GRAVITY CONVECTION. MODELS: TO-152G & TOH-152G TO-152F, TOH-152F & TO-235F TO-330F & TO-500F WITH OMRON E5CC CONTROL Thermoline Scientific Equipment Pty. Ltd. T/A Thermoline

More information

External Dimensions: B = 750 mm D = 700 mm H = 2100 mm. Internal Dimensions: B = 600 mm D = 580 mm H = 1270 mm. Temperaturerange: 2 C to 20 C

External Dimensions: B = 750 mm D = 700 mm H = 2100 mm. Internal Dimensions: B = 600 mm D = 580 mm H = 1270 mm. Temperaturerange: 2 C to 20 C External Dimensions: B = 750 mm D = 700 mm H = 2100 mm Internal Dimensions: B = 600 mm D = 580 mm H = 1270 mm Capacity: 600 l Temperaturerange: 2 C to 20 C Housing galvanized sheet steel with high quality

More information

Climatic chamber for constant conditions with program control

Climatic chamber for constant conditions with program control Operating Manual APT.line KBF (E5.2) Climatic chamber for constant conditions with program control Model Art. No. KBF 115 (E5.2) 9020-0139, 9120-0139 KBF 115-UL (E5.2) 9020-0184, 9120-0184 KBF 240 (E5.2)

More information

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time

Operating instructions Page 14. Refrigerator Read the operating instructions before switching on for the first time Operating instructions Page 14 Refrigerator Read the operating instructions before switching on for the first time 7085 039-00 LKv 5710 Content Disposal notes... 14 Description of the appliance... 14 Safety

More information

Laboratory Deep Freeze Storage Cabinet TC 230 Air circulated

Laboratory Deep Freeze Storage Cabinet TC 230 Air circulated External Dimensions: B = 600 mm D = 650 mm (at -30 C/-40 C: 760 mm) H = 1640 mm (at -30 C/-40 C: 1720mm) Internal Dimensions: B = 500 mm D = 500 mm H = 890 mm Capacity: 300 l Temperaturerange: Optional:

More information

Drying oven VWR DRY-Line with Natural Convection DL 53, DL 115

Drying oven VWR DRY-Line with Natural Convection DL 53, DL 115 Drying oven VWR DRY-Line with Natural Convection DL 53, DL 115 User Manual Issue 11-2015_en valid from device serial number 15-03000 DRY-Line DL 11-2015_en page 2/39 DRY-Line DL 11-2015_en page 3/39 Content

More information

Inc Series Digital Gas Detector/Controller. Operation Manual

Inc Series Digital Gas Detector/Controller. Operation Manual 5 Series Digital Gas Detector/Controller Operation Manual TABLE OF CONTENTS 1. 2. 3. 4. 5. 6. 7. General Description 1.1 Applications 1.2 Features Model Selection Guide Characteristics Installation 4.1

More information

Electronic Temperature Controller 70304,

Electronic Temperature Controller 70304, Electronic Temperature Controller 70304, 70304.2 Instruction Manual Version 1.00.01 List of content 1 General information... 4 2 Declaration of conformity... 4 3 Safety instructions... 5 4 Warranty...

More information

Electron Microscopy Sciences

Electron Microscopy Sciences Electron Microscopy Sciences INSTRUCTIONAL MANUAL CAT. 72341-10 Explosion Proof Hot Plate Stirrer Model SF87325 P.O. Box 550 s1560 Industry Road s Hatfield PA 19440 1 Safety Symbols Used in this Instruction

More information

Operational Instructions. Glue Applicator MS Translation from German to English

Operational Instructions. Glue Applicator MS Translation from German to English Operational Instructions Glue Applicator MS 200 Translation from German to English Table of Contents Product Description 1 General Safety Instructions 2 Safety Measures to be taken during Operation 3 Start-up

More information

Dry Bath Heating Systems

Dry Bath Heating Systems Dry Bath Heating Systems Instruction Manual N2400-4001 Dry Bath Heating System, Single Block, Digital N2400-4002 Dry Bath Heating System, Dual Block, Digital www.starlabgroup.com DECLARATION OF CONFORMITY

More information

Two-Channel Gas Controller

Two-Channel Gas Controller Two-Channel Gas Controller Specifications subject to change without notice. USA 09 Page of DESCRIPTION Highly configurable, UL 0 performance-tested and -certified, and wall-mounted gas monitor; continuously

More information

Product Manual Glue Applicators MS 80 and MS 200

Product Manual Glue Applicators MS 80 and MS 200 Product Manual www.pakarte.com Glue Applicators MS 80 and MS 00 Translation from German to English Table of Contents Product Description General Safety Instructions Safety Measures to be taken during Operation

More information

Medical Air circulated Refrigerator TC (2 independent temperature ranges), also in accordance with DIN 58345

Medical Air circulated Refrigerator TC (2 independent temperature ranges), also in accordance with DIN 58345 External Dimensions: B = 530 mm D = 640 mm H = 1600 mm Inside Dimension: Useful space on top Useful space at the bottom B = 450 mm B = 450 mm D = 535 mm D = 535 mm H = 530 mm H = 530 mm Capacity: 280 l

More information

REFRIGERATED RECIRCULATORY COOLER

REFRIGERATED RECIRCULATORY COOLER REFRIGERATED RECIRCULATORY COOLER MODEL INCLUDES: TRC-3000 Thermoline Scientific Equipment Pty. Ltd. T/A Thermoline Scientific ABN 80 000 859 129 10-12 Ross Place. Wetherill Park N.S.W. 2164. Australia.

More information

IMR IX176 Portable Gas Detector User Manual

IMR IX176 Portable Gas Detector User Manual IMR Portable Gas Detector User Manual Read this manual carefully before using this device. (727) 328-2818 / (800) RING-IMR Fax: (727) 328-2826 www.imrusa.com Ver. 1.0A4 CONTENTS SERVICE GUIDELINES... 3

More information

Micro III (Models G203, G203S, and G213) Operation Manual

Micro III (Models G203, G203S, and G213) Operation Manual Micro III (Models G03, G03S, and G13) Operation Manual Worldwide Manufacturer of Gas Detection Solutions Table of contents Page Introduction... 3 For your safety... 3 Application and use... 3 General

More information

CO₂ incubators. C CB series

CO₂ incubators. C CB series CO₂ incubators C CB series EN CO₂ incubators Series C Page 6 Series CB Page 9 Options and Accessories Page 12 Every CO₂ incubator from BINDER guarantees optimum, reproducible growth conditions for cell

More information

SAFETY MANUAL. Intelligent Sensors for H 2 S Gas Applications

SAFETY MANUAL. Intelligent Sensors for H 2 S Gas Applications SAFETY MANUAL Intelligent Sensors for H 2 S Gas Applications The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically

More information

Residential Gas Condensing Boiler Greenstar ZBR16/21/28/35/42-3A... ZWB28/35/42-3A...

Residential Gas Condensing Boiler Greenstar ZBR16/21/28/35/42-3A... ZWB28/35/42-3A... 70 80 99-00-O Residential Gas Condensing Boiler ZBR//8/35/4-3A... ZWB8/35/4-3A... 70 80 993 (03/03) CA/US Operating Instructions Contents Contents Key to symbols and safety instructions............................

More information

Operating Instructions for Reka Hot-Melt Glue Applicators

Operating Instructions for Reka Hot-Melt Glue Applicators Operating Instructions for Reka Hot-Melt Glue Applicators TR 700 TR 70 TR 80 TR 60 TR 50.4 TR 55 Table of Contents 1. Product Description 3 2. General Safety Instructions 4 3. Safety Measures to be taken

More information

Controllers. Instruction Manual WARNING

Controllers. Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

portable gas analyser

portable gas analyser www..com portable gas analyser The largest of s analysers equipped with electrochemical cells. It can fit even up to 7 EC cells and up to 3 sensors. GA-60 has a large (320*240), graphical LCD with backlighting.

More information

Undercounter Refrigerator Operation Manual i.series and Horizon Series

Undercounter Refrigerator Operation Manual i.series and Horizon Series Undercounter Refrigerator Operation Manual i.series and Horizon Series Blood Bank Models i.series: ib105 (Version B) Horizon Series: HB105 (Version B) Laboratory/Pharmacy Models i.series: ilr105 (Version

More information

Instruction Manual WARNING

Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANAUL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

SP-220. Features. Combustible handheld gas leakage detector

SP-220. Features. Combustible handheld gas leakage detector SP-220 Combustible handheld gas leakage detector The SP-220 gas leakage detector is a simple to operate unit with high sensitivity to a variety of gases. The SC version (SP-220SC) is able to detect super-toxic,

More information

Reference Manual. Ionizer Module for XPR Analytical Balances

Reference Manual. Ionizer Module for XPR Analytical Balances Reference Manual Ionizer Module for XPR Analytical Balances Ionizer Module 3 1 Introduction Your Ionizer Module is an accessory for METTLER TOLEDO analytical balances. The ionizer produces a corona of

More information

UFM-T100H US ES CA Power Alarm Test TM

UFM-T100H US ES CA Power Alarm Test TM Alarm Power TM UFM-T100H US ES CA A E F I J K G B C D H L M N O P Q R S T U V W X US 1 3 2 Power Alarm 4 6 5 Power Alarm 7 9 8 Po we Ala r rm Power Alarm 10 11 Po we Po we Ala rm r 12 Pow Ala Ala rm r

More information

PolyGard Single Point Controller SPC-X3-11XX for Toxic Gases and Oxygen

PolyGard Single Point Controller SPC-X3-11XX for Toxic Gases and Oxygen PolyGard Single Point Controller SPC-X3-XX for Toxic Gases and Oxygen DESCRIPTION Gas measuring, monitoring and warning controller based on state-of-the-art micro-technology with integrated sensor and

More information

testo Leakage detector for gas Instruction manual

testo Leakage detector for gas Instruction manual testo 316-2 Leakage detector for gas Instruction manual 2 Safety and environment Safety and environment On this document > Read this document through carefully, and familiarize yourself with the product

More information

ICH260C. Climate Chamber. Product specification

ICH260C. Climate Chamber. Product specification Product specification Climate Chamber ICH260C Unparalleled homogeneity of temperature and humidity! For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP GCCP. On this page, you can find

More information

Dräger X-am 7000 Multi-Gas Detection

Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 Multi-Gas Detection Dräger X-am 7000 is the solution for the simultaneous and continuous measurement of up to five gases. It is the ideal companion in a variety of applications where the

More information