F Operating Manual. ESS0460-H119 PV Module Solar Soaking Chamber
|
|
- Hollie Davidson
- 5 years ago
- Views:
Transcription
1 F Operating Manual ESS0460-H119 PV Module Solar Soaking Chamber 12/19/2008
2 Thank you for your purchase. We are confident you will be completely satisfied with your new system. If for any reason you need assistance, please contact us: EYE Lighting International / Iwasaki Electric 9150 Hendricks Road Mentor, OH support@eyeappliedoptics.com In order to operate this system properly and prevent potential problems, please read the following operation instructions. Table of Contents 1. System Components 2. Installation 3. Control Panel 4. System Operation 5. Solar Lamps 6. Irradiation Intensity Control 7. Troubleshooting
3 1. System Components: 1 Solar Lighting Array Housing (60) 400W solar lamp array 2 Solar Lamp Power / Inverter Cabinet includes (5) inverter units (12 modules per unit; one module per lamp) and solar lamp starters (60 units; one starter per lamp) 3 Operation Control Panel control and failure alert 4 Main Power Breaker 3 phase main power switch for system 5 Solar Light Chamber Shields roll down curtains to shield excess light from work area 6 Solar Irradiation Chamber 7 Cooling Air Ducts dual axial blowers direct air over PV panels under test 8 PV Panel Support Trays interchangeable; (4) separate frames for standard thin-film modules (60cm x 120cm) and (4) free-format frames for all other size PV modules; trays provide for easy insertion of PV panels and allow cooling air to flow above and below panels
4 2. Installation A. When installing this chamber, please avoid placing it in any area where corrosive gas or explosive gas is present. It is also highly recommended that the area be as dust free as possible with controlled humidity. B. The chamber should be placed as close to the mains power source (power transformer) and mains breaker as possible. C. Due to the heat generated by the solar array, keep flammable materials away from the top of the chamber. D. Please refer to the following diagrams to ensure proper clearance around the chamber for proper operation and maintenance / inspection 600mm (24 in) 600mm (24 in) Rear Top 600mm Side 600mm 600mm Side 600mm (24 i ) Overhead View (24 i ) (24 i ) Frontal View (24 i ) This spacing is required for proper operation and routine inspection / maintenance Electrical / Power Characteristics System Power: 3 Ø; 208V AC; 60Hz; 35kVA (approximately 100 A) Cooling Blower Fans: 0.75 kw 2 units Solar Lamps: 208V; 400W 60 lamps Irradiation Power Control: 1500W solar lamp control circuit (approximately 3 A) Mains Power Connection: 3 Ø; 4 wire; RED, WHITE, BLACK power phase connections and GREEN ground connection; single phase solar lamp circuits are equally divided to balance the power over all three phases. To protect against unbalanced or open phase power failures, the system uses a voltage detecting (APR-S) phase reversal power breaker. Red White Black Chamber Power Connections Mains Power Normal electrical connection. If system does not function it may be due to phase connection mismatch with (APR-S) breaker. Please reconnect mains power as illustrated below. Alternative connections Red White Black Green Red White Black Electrical safety is provided through a dedicated earth connection. It is vital to ensure that the system is connected to earth ground. Chamber Power Connections Mains Power Alternative mains power phase connections. Red White Black Green
5 3. Control Panel MAIN POWER INDICATOR LAMP Indicator lamp will illuminate when the Main Power Breaker is ON and power is supplied to the system 2 IRRADIATION AREA ZONE SWITCHES The solar array is arranged in 5 zones of 12 lamps each. Each zone is controlled independently and energized by its associated button. When a zone is energized, its switch is illuminated. NOTE: cooling fans must be turned on prior to switching solar zones on. The solar array zone switches will not function until the cooling fans are turned on. 3 CUMULATIVE OPERATION HOUR METERS A separate meter is assigned to each zone and records the cumulative operation hours for the lamps in that zone. Recording the lamp operating time is extremely beneficial for preventative maintenance. If less than all zones are often used for particular test applications, please try to periodically move the PV panels within the chamber to equalize the operating time across all zones. 4 TEST OPERATION HOUR CONTROLLER This counter timer provides the capability to program a desired test duration in hours and minutes. To initiate the test, press the TEST START button. At the conclusion of the test, the lamps will automatically turn off. The cooling fans will continue to run for ten (10) minutes after test shutdown, at which time the fans will automatically turn off.
6 5 TEST START SWITCH This button initiates the programmed test period. When initiated, this switch will be illuminated. At the conclusion of the programmed test time, the alarm buzzer will sound and the solar lamps will be turned off. The buzzer will turn off automatically after one (1) minute or you can turn the buzzer off manually by pressing the buzzer stop switch. 6 TROUBLE (SYSTEM FAILURE) INDICATOR LAMP If a system error occurs, the system will automatically shut down, the TROUBLE lamp will be illuminated and the alarm buzzer will sound. Under some error conditions the solar lamps will immediately shut off but the cooling fans will continue to operate for ten (10) minutes. Defined failure modes include: Power Failure phase / power abnormality; APR-S breaker may trip off; lamps and fans will shut down Fan (TH1, TH2) Failure cooling fan failure or thermal protection shutdown; fans and lamps will shut down System Overheat system chamber temperature exceeds default setting of 100 ºC; lamps will shut down, fans will continue to operate Irradiation Power Over-limit irradiation control circuit maximum voltage (10.4V) exceeded; lamps will shut down, fans will continue to operate Solar Lamp / Array Failure lamp or inverter failure; lamps will shut down, LAMP FAILURE light will be illuminated, fans will continue to operate 7 LAMP FAILURE INDICATOR LAMP When a solar lamp circuit failure is detected, this indicator lamp will be illuminated and the alarm buzzer will sound. Cause of failure could be the lamp, inverter, starter, or wiring defect. Solar lamps will shut down but fan will continue for 10 minutes. 8 COOLING FAN SWITCH This switch initiates the cooling fans and the switch will be illuminated. Pressing the switch again will turn the cooling fans off. At completion of a programmed test period, the switch indicator lamp will flash on and off and the fans will continue to operate for ten (10) minutes. The fans can be manually stopped at any time during this period by pressing this switch. NOTE: cooling fans must be turned on prior to switching solar zones on. The solar array zone switches will not function until the cooling fans are turned on. 9 BUZZER STOP SWITCH The system alarm buzzer will sound due to the completion of a programmed test period or because of a system error. This switch allows you to manually turn the buzzer off. 10 EMERGENCY STOP SWITCH Pressing this switch will immediately shut down the lamps and cooling fans. It will also initiate the system alarm buzzer. The switch will remain latched in the OFF position until manually released. After corrective measures have been performed, the emergency stop switch is released and reset by turning the switch clockwise. 11 IRRADIATION POWER (DIMMING) CONTROLLER This controller adjusts the solar array lamp power. The digital indicator displays the controller voltage (0 10V). It provides the capability to adjust the solar array output from 1000W/m 2 to 800W/m 2. Additionally, as the solar lamps age the lamp power can be increased, if required, to maintain an average 1000W/m 2. NOTE: controller voltage may not exceed 10.4V.
7 4. System Operation Full Chamber Manual Operation a) Ensure main power breaker is ON (Main Power Lamp will be illuminated). b) Turn cooling fans ON (solar array lamps will not function until fans are turned on). c) Turn on solar array by pressing LIGHTING AREA ZONE switches 1-5 sequentially. Allow ten (10) minutes for solar lamps to reach full power and spectrum stabilization. d) At completion of test, turn off solar array by pressing LIGHTING AREA ZONE switches 1-5 sequentially. e) Turn cooling fan OFF f) Turn OFF Main Power Breaker if desired. Programmed Test Operation a) Follow steps a) c) above. b) Program desired test duration time in hours and minutes on the TEST OPERATION HOUR controller. c) Initiate test by pressing TEST START switch d) At completion of test, solar array will automatically shut down, alarm buzzer will sound, cooling fans will continue to operate for ten (10) minutes and cooling fan switch lamp will flash. e) Cooling fans may be turned OFF by pressing the COOLING FAN switch. f) Turn OFF Main Power Breaker if desired. Partial Chamber Operation a) Follow steps a) and b) from Full Chamber Manual Operation section above. b) Turn on solar array segment by pressing the desired LIGHTING AREA ZONE switches 1-5 sequentially. Allow ten (10) minutes for solar lamps to reach full power and spectrum stabilization. To ensure minimum IEC class C uniformity over entire target irradiation area please observe the following guidelines: For QUARTER chamber (650mm x 650mm) target irradiation area, place PV panel(s) in chamber at far left (facing chamber) side and turn on LIGHTING AREA ZONE switches 1-3 sequentially. For HALF chamber (650mm x 1300mm) target irradiation area, place PV panel(s) in chamber starting at far left (facing chamber) side and turn on LIGHTING AREA ZONE switches 1-4 sequentially. Refer to Irradiation Intensity Distribution profiles in Section 2 of this manual for specific details c) If programmed test is desired, follow steps b) f) from Programmed Test Operation section above. d) If manual test is desired, follow steps d) f) from Full Chamber Manual Operation section above. 5. Solar Lamps A. Please allow ten (10) minutes after initial startup for solar lamps to reach full power and spectrum stabilization. B. If solar array has been in operation and is shut down (hot re-start), please allow about 20 minutes to re-start lamps and for solar lamps to reach full power and spectrum stabilization. C. For preventative maintenance, the solar lamp life is specified at 3,000 hours of operation. Although the lamps will typically last much longer, the lamp life is specified to ensure irradiation output and spectrum remain within required IEC class C specifications.
8 6. Irradiation Intensity Control The system irradiation intensity was set to an average 1000W/m 2 (+/-15%; class C uniformity) at the time of installation. This intensity distribution test is included in Section 3 of this manual. We recommend that the irradiation intensity is verified approximately every 500 hours of operation or once per month. To adjust the irradiation intensity, please follow the steps below. a) Refer to initial installation irradiation intensity distribution diagram included in Section 3 of this manual. b) Select at least two (2) distributed test locations from distribution grid (these positions are marked by tape on the bottom of the chamber roller shelf). c) Turn on full solar array and allow at least ten (10) minutes for the lamps to stabilize. d) Place a full spectrum pyranometer on the first of these selected positions and measure irradiation intensity, then repeat at the second selected position. e) Compare these measurements against the initial installation irradiation intensity measurements, if the difference is more than five (5) percent it is recommended that a full 18 point irradiation intensity grid measurement is performed. This will provide the minimum irradiation intensity point and allow the calculation of the average irradiation intensity and full target uniformity. f) If the minimum irradiation intensity is below 850W/m 2 the irradiation intensity power should be adjusted using the DIMMING CONTROL calibrated dial. This circuit outputs a 0-10V signal that controls the power output (0V = Max power; 10V = Min power). A typical value for a new installation (set to 1000W/m 2 ) will be approximately 4.2V. g) Place the pyranometer at the minimum measured irradiation intensity grid point and ensure the reading is stable. Adjust the controller voltage lower while reading the intensity level from the pyranometer (allow sufficient time after any adjustment for the intensity level to stabilize). h) Adjust the power such that the intensity level is increased to a value between W/m 2. i) Re-measure the full 18 point grid irradiation intensity and re-calculate the Max, Min, and Average values to ensure they are within specification. If, for a particular application, you desire to reduce the average irradiation intensity to a value lower than 1000W/m2 (such as 850W/m2), please follow steps g) i) above and adjust the power output proportionally to the desired average irradiation intensity level. NOTE: To ensure solar lamp irradiation and spectrum stability, the average irradiation intensity (power output) should not be reduced below 800W/m 2.
9 7. Troubleshooting Cooling Problem Blower Lamp Probable Cause Action Phase Error X X Temporarily power phase interruption or wiring error. Check power connections and mains power breakers. Re-set breakers and power on system. If error still exists, ensure balanced Emergency Stop Cooling Fan Overload X X Emergency Stop switch has been activated and is still latched in the STOP position. X X Thermal relay or abnormal current error from cooling fan motor. 208V 3 phase power is being supplied to system. If power input and connections are correct, please contact EYE Lighting technical support. Reset the Emergency Stop switch by rotating switch clockwise. Power on system and ensure there are no system errors (power, lamps, fans, etc.). Troubleshoot any specific failures that are identified. If no failure mode is identified by control circuits, please contact EYE Lighting technical support. Ensure there are no obstructions restricting rotation of fan blowers. Rotate shaft and check for deformation, abnormal noise, or binding. If no problems are identified, please contact EYE Lighting technical support. Overheat Continues for 10 X Occurs when the chamber temperature Check the chamber temperature reading at the control panel to ensure minutes after exceeds the maximum allowable an overheat condition does not exist. If the temperature reading is occurrence of temperature (set at the factory to 100 below 100 ºC, troubleshoot the thermocouple controller. If the failure ºC). This is continuously measured by a temperature reading is above 100 ºC, ensure the cooling fans are type K thermocouple located near the operating correctly. If fans are operating properly, please contact EYE top of the chamber. Lighting technical support. Lamp Failure Continues for 10 X Lamp failure to start, or failure during From a cold start, a solar lamp should start within approximately 40 Detection minutes after operation. Starter failure at startup. seconds. If a lamp does not start within this timeframe, the LAMP occurrence of FAILURE indicator lamp will be illuminated and the system will shut failure down. Replace the affected lamp and try to re-start system. If lamp still doesn't start, troubleshoot the lamp starter (Power/Control cabinet) and wiring from starter to lamp socket. If no problems are identified, please contact EYE Lighting technical support. Abnormal Continues for 10 X Solar lamp output power (DIMMING Irradiation intensity modulated voltage control (DIMMING CONTROL) Voltage minutes after CONTROL) set beyond maximum limit. must be set between 0V - 10V. Re-set irradiation intensity and re-start Limit occurrence of system. If failure persists, please contact EYE Lighting technical failure support. NOTE: X = System component will cease operation immediately on failure
LABORATORY OVENS MODELS INCLUDE: TO-152G & TOH-152G TO-152F & TO-235F DEHYDRATING OVENS MODELS INCLUDE: TD-150F TD-330F TD-500F
0 LABORATORY OVENS MODELS INCLUDE: TO-152G & TOH-152G TO-152F & TO-235F DEHYDRATING OVENS MODELS INCLUDE: TD-150F TD-330F TD-500F WITH WEST 6100 CONTROL 1 IMPORTANT INFORMATION This manual contains operating
More informationEBAC MODEL CD425 ( ) INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL
EBAC MODEL CD425 (1018110) INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL Ebac Industrial Products 704 Middle Ground Boulevard Newport News, VA 23606 Tel: 757 873 6800 Fax: 757 873 3632 Website: www.ebacusa.com
More informationVACUUM OVEN. Model: VPX9-2. Installation and Operation Manual
VACUUM OVEN Model: VPX9-2 Installation and Operation Manual Sheldon Manufacturing, Inc. P.O. Box 627, 300 N 26 th Avenue, Cornelius, Oregon 97113 EMAIL: tech @ Shellab.com; www.shellab.com 1-800-322-4897
More informationEBAC MODEL CD425 ( ) INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL
EBAC MODEL CD425 (1018110) INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL Ebac Industrial Products, Inc. 700 Thimble Shoals Blvd, Suite 109 Newport News, VA. 23606-2575 Tel: (757) 873 6800 Fax: (757) 873 3632
More informationINSTALLATION AND OPERATION MANUAL
MICROPROCESSOR CONTROLLED OVEN MODELS: CE3F & CE5F 11/11 4861579 INSTALLATION AND OPERATION MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@shellab.com INTERNET: http://www.shellab.com/~shellab
More informationEquipment specifications (D6)
Equipment specifications (D6) This report has been produced for the Cyprus Research Promotion Foundation in fulfilment of deliverable number D6 of the project Advanced Photovoltaics Research and Testing
More informationNOTICE . SAFE SERVICING PRACTICES. Electric Wall Oven with Electronic Oven Control
SERVICE DATA SHEET 318047418 (0504) Rev. A Electric Wall Oven with Electronic Oven Control NOTICE This service data sheet is intended for use by persons having electrical and mechanical training and a
More informationFOR SERVICE TECHNICIAN S USE ONLY
FOR SERVICE TECHNICIAN S USE ONLY NOTE: This sheet contains important Technical Service Data. Tech Sheet Do Not Remove Or Destroy DANGER Electrical Shock Hazard Only authorized technicians should perform
More informationHT-2 / 9600 Series Control Contents
HT-2 / 9600 Series Control Contents Tools & Parts Tools Required Parts Required Error Messages 3 Flashing Dots Pressure or Flow Switch Not Activated Pressure or Flow Switch Activated Temperature Sensor
More informationSERVICE AND INSTALLATION MANUAL MODELS HDO(H) OIL FOR YOUR SAFETY
Bousquet Technologies Inc. 2121, Nobel, Ste Julie, Quebec, Canada, J3E1Z9 SERVICE AND INSTALLATION MANUAL MODELS HDO(H) OIL Oil-Fired air heater for industrial and commercial use. FOR YOUR SAFETY Do not
More informationINCUBATORS MODELS: 1525, 1535, 1545, 1555, 1565 GENERAL PURPOSE INCUBATORS INSTALLATION AND OPERATION MANUAL
INCUBATORS MODELS: 1525, 1535, 1545, 1555, 1565 GENERAL PURPOSE INCUBATORS INSTALLATION AND OPERATION MANUAL These units are UL listed general purpose air incubators for professional, industrial or educational
More informationFigure 1-1 Furnace Front Elevation
FURNACE EQUIPMENT 1.1 Furnace Description The LA-306 is a compact, near-infrared, conveyor belt furnace for laboratory and general purpose thermal processing in the range of 100-980 ⁰C in a controlled
More informationRefrigerated Incubator Model and Operating Instructions
Refrigerated Incubator Model 165000 and 165000-2 Operating Instructions N2400379 - Rev. 1 08May2018 1 Contents 1. SAFETY...3 1.1. EMF INTERFERENCE...4 1. PRODUCT INFORMATION...5 1.1 INTRODUCTION...5 2.
More informationBUILT-IN OVEN MODEL: EBO-D7080D(SS) Owner s Manual Please read this manual carefully before operating your set. Retain it for future reference.
BUILT-IN OVEN MODEL: EBO-D7080D(SS) Owner s Manual Please read this manual carefully before operating your set. Retain it for future reference. Record model number and serial number of the set. See the
More informationUser Manual 120/240/300 Outdoor Motion Detector
User Manual 120/240/300 Outdoor Motion Detector Swiss Garde 4000 Warning! - Any work on the main power supply must only be completed by trained specialists. - The product must be installed in accordance
More informationINSTRUCTIONS FOR THD-U SERIES HOT AIR MATERIAL DRYERS
INSTRUCTIONS FOR THD-U SERIES HOT AIR MATERIAL DRYERS Copyright Plastic Process Equipment, Inc 2015 Our PPE Material Dryers are a PPE exclusive only available from PPE. Do not confuse these with any other
More informationAHPO-6/18 GOLD PROOFER OVEN
AHPO-6/18 GOLD PROOFER OVEN For information or technical assistance, call: TOLL FREE 1(800) 735-DUKE (3853) or 1(314) 231-1130 PN 156081R 1 of 25 TABLE OF CONTENTS MANUFACTURER S INTRODUCTION...3 INSTALLATION...4
More informationRefrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA
Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA 99336 www.jmcvr.com 1-509-586-9893 Table of Contents TABLE OF FIGURES...1 OVERVIEW OF THE HRC CAPABILITIES...2 INSTALLATION AND
More informationOperational Overview and Controls Guide
DOCUMENT: ECSEQ2-2 EFFECTIVE: 02/26/10 SUPERSEDES: 02/14/07 Operational Overview and Controls Guide Standard Two or Three Pump Type VFD Booster Controls 6700 Best Friend Road. Norcross, GA 30071. (770)
More informationOPERATING INSTRUCTIONS MANUAL FLE-150 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE)
OPERATING INSTRUCTIONS MANUAL FLE-150 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE) Suitable for indoor or outdoor installation / Unvented / Unattended Type 26 Benfield
More informationFOR SERVICE TECHNICIAN S USE ONLY
F SERVICE TECHNICIAN S USE ONLY NOTE: This sheet contains important Technical Service Data. Tech Sheet Do Not Remove Or Destroy DANGER Electrical Shock Hazard Only authorized technicians should perform
More informationShaking Hot Tubs. Orbital / Reciprocating Shaking Water Bath Liter 115V S 28 Liter 115V Liter 230V
Shaking Hot Tubs Orbital / Reciprocating Shaking Water Bath 290400 28 Liter 115V 290400S 28 Liter 115V 290400-2 28 Liter 230V Operating Instructions 1 Table of Contents 1. Safety..3 2. Getting Started
More informationIndirect gas-fired air heater
Indirect gas-fired air heater SERIES HD INSTALLATION AND SERVICE MANUAL MANUFACTURED BY : BROTHERS LIMITED WARNING Improper installation, modification, adjustment or maintenance may cause damage, injury
More informationSERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN - NOTICE -
SERVICE MANUAL VC3ED FULL SIZE ELECTRIC CONVECTION OVEN VC3ED ML-137013 - NOTICE - This Manual is prepared for the use of trained Vulcan Service Technicians and should not be used by those not properly
More informationInstallation and Operation Manual
SENTRY Protect Plus QUADPLEX PANEL Installation and Operation Manual For Hardwired Pumps Environment e Corporation Table of Contents 1 Overview...3 2 Sentry Protect Plus Quadplex Menu Flowchart...4 3 Wiring
More informationChapter 8 Troubleshooting
Chapter 8 Troubleshooting This chapter explains equipment trouble and how to remedy it. When trouble is detected by one of the self-check features, the trouble is displayed on the instrumentation panel
More informationChapter 8 Troubleshooting
Chapter 8 Troubleshooting This chapter explains equipment trouble and how to remedy it. When trouble is detected by one of the self-check features, the trouble is displayed on the instrumentation panel
More informationFOR SERVICE TECHNICIAN S USE ONLY
F SERVICE TECHICIA S USE OY TE: This sheet contains important Technical Service Data. Tech Sheet Do ot Remove Or Destroy DAGER Electrical Shock Hazard Only authorized technicians should perform diagnostic
More informationE25 CONVECTION OVEN E25MS CONVECTION OVEN SERVICE MANUAL
E25 CONVECTION OVEN E25MS CONVECTION OVEN SERVICE MANUAL E25 Convection Oven -- WARNING: ALL INSTALLATION AND SERVICE REPAIR WORK MUST BE CARRIED OUT BY QUALIFIED PERSONS ONLY. E25 Convection Oven -2-
More informationEBAC MODEL WM150 INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL
EBAC MODEL WM150 INDUSTRIAL DEHUMIDIFIER OWNER S MANUAL WM150 OWNERS MANUAL Page 1 of 9 INTRODUCTION Designed for a wide range of applications, the WM150 is a rugged, industrial unit, which utilizes an
More informationOPERATING INSTRUCTIONS MANUAL FLE-60 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE)
OPERATING INSTRUCTIONS MANUAL FLE-60 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE) Suitable for indoor or outdoor installation / Unvented / Unattended Type 26 Benfield
More informationLABORATORY ILLUMINATED FLOTATION BATH
LABORATORY ILLUMINATED FLOTATION BATH MODELS INCLUDE: TFB-1 Thermoline Scientific Equipment Pty. Ltd. ABN 80 000 859 129 10-12 Ross Place Wetherill Park NSW 2164. P.O. Box 6862 Wetherill Park DC NSW 1851
More informationPart 3 Troubleshooting
Part Troubleshooting What is in this part? This part contains the following chapters: Chapter See page Troubleshooting 2 Error Codes: Hydro-box 7 Error Codes: Outdoor Units Error Codes: System Malfunctions
More informationMD-MINI Dry Bath. Instruction Manual
MD-MINI Dry Bath Instruction Manual Catalog No. MD-MINI www.majorsci.com service@majorsci.com Version 01B Revision on: 2013.07.15 Packing list MD-MINI: 1x MD-MINI Dry Bath 1x MD-MINI Lid 1x Power Cord
More informationHORIZONTAL MULTISTAGE CENTRIFUGAL PUMP
HORIZONTAL MULTISTAGE CENTRIFUGAL PUMP WWPPCHLFT260 Instructions WWPPCHLFT260_Horizontal Multistage Centrifugal Pump_IB.indd 1 READ THIS MANUAL CAREFULL BEFORE INSTALL, START THE PUMP 1. Suction 2. Plug
More informationBench Series Ovens Models 21 / 31 / 51 Operating Manual
Heat cycle HEATER SWITCH RECIRC. FAN Quincy Lab, Inc. Bench Series Ovens Models 21 / 31 / 51 Operating Manual Standard Contents (1) Bench Series Oven (2) Adjustable chrome wire shelf (4) Shelf brackets
More informationSERVICE INSTRUCTION. DYNAMIC COOLING with 0 zone REFRIGERATION. Fridge/freezer with. Publication number
SERVICE INSTRUCTION REFRIGERATION DISTRIPARTS AB BOX 501 S-562 80 NORRAHAMMAR Phone 036-31 80 00 Telefax 036-31 81 10 Telefax 036-31 80 88 Publication number 599 51 96-71 Rev. 04-04-20 SV/SERVICE DT/MA/AN
More informationReacti-Therm I and III Heating Modules
INSTRUCTIONS Reacti-Therm I and III Heating Modules TS-18822 TS-18824 Number TS-18822 TS-18824 Description Reacti-Therm I Heating Module (single block) Reacti-Therm III Heating Module (triple block) 2101.1
More informationFisher Scientific Isotemp Muffle Furnace 550 Series. Instruction Manual
Fisher Scientific Isotemp Muffle Furnace 550 Series Instruction Manual Model 14 Catalog No: 10-550-14 & 14A Model 58 Catalog No: 10-550-58 Model 126 Catalog No: 10-550-126 Service Division (Repairs):
More informationINSTALLATION AND OPERATION MANUAL
REACH-IN DRY CO 2 INCUBATOR SCO31 SCO31-2 SCO40 SCO41-2 PREVIOUSLY DESIGNATED AS 2428 2428-2 2440 2440-2 Revised 12/2013 4861504 INSTALLATION AND OPERATION MANUAL Sheldon Manufacturing Inc. P.O. Box 627
More informationSSV-DC user manual. Phason
Phason SSV-DC user manual The SSV-DC has a 0 to 10 VDC signal output and a disconnect relay that allows you to easily and effectively control up to 10 variable frequency drives, slave units, modulating
More informationCONTENTS CONSIDERATIONS. General Plumbing Component Connection ILLUSTRATION. Control System CONFIGURATION. Voltage Verification CONNECTION
AIR SERIES SYSTEM INSTALLATION MANUAL CONTENTS CONSIDERATIONS General Plumbing Component Connection 2 2 2 ILLUSTRATION Control System 3 CONFIGURATION Voltage Verification 4 CONNECTION Component Connection
More informationBUILT-IN OVEN MODEL: EBO-E7081D(SS) Owner s Manual Please read this manual carefully before operating your set. Retain it for future reference.
BUILT-IN OVEN MODEL: EBO-E7081D(SS) Owner s Manual Please read this manual carefully before operating your set. Retain it for future reference. Record model number and serial number of the set. See the
More informationDRI-AIR INDUSTRIES, INC. HA750B B HOT AIR DRYER OPERATING MANUAL
HA750B - 1500B HOT AIR DRYER OPERATING MANUAL Page 1 DRI-AIR INDUSTRIES, INC. 16 THOMPSON ROAD P.O. BOX 1020 EAST WINDSOR, CT 06088-1020 Tel. (860) 627-5110 FAX (860) 623-4477 Internet http://www.dri-air.com
More informationOPERATING INSTRUCTIONS MANUAL FLE-150 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE)
OPERATING INSTRUCTIONS MANUAL FLE-150 INDUSTRIAL / COMMERCIAL ELECTRIC HEATER (OPERATOR MUST RETAIN FOR FUTURE REFERENCE) Suitable for indoor or outdoor installation / Unvented / Unattended Type 26 Benfield
More informationMotion Sensor Oscillating Heater with Digital Control
Motion Sensor Oscillating Heater with Digital Control 33553 Part# NFE20C INSTRUCTION MANUAL Please Read And Save These Instructions 1 IMPORTANT SAFETY INSTRUCTIONS Please read these instructions carefully
More informationInstallation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro
Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro 2006/01 Content Important Information 1. Notes... 3 1.1 CE - Labeling... 3 1.2
More informationReacti-Therm I and III Heating/Stirring Modules
INSTRUCTIONS Reacti-Therm I and III Heating/Stirring Modules TS-18821 TS-18823 Number TS-18821 TS-18823 Description Reacti-Therm I Heating/Stirring Module (single block) Reacti-Therm III Heating/Stirring
More informationUser Manual HEAT RECOVERY VENTILATORS 1000 HE HE HE HE HE HE COMPACT 5585 COMPACT READ AND SAVE THESE INSTRUCTIONS.
User Manual HEAT RECOVERY VENTILATORS MODELS: 1000 HE HE 1.3 2000 HE HE 1.8 3000 HE HE 2.6 3055 COMPACT 5585 COMPACT VB0021 READ AND SAVE THESE INSTRUCTIONS. VVI, 550 Lemire Blvd., Drummondville, QC, Canada
More informationTTT Digital Incubator
TTT Digital Incubator 135000 135000-2 Operating Instructions N2400342 Rev. 1 08/13 Table of Contents 1.0 Safety... 3 2.0 Intended Use of Product... 4 3.0 Installation... 4 3.1 Unpacking the Unit... 4 3.2
More informationsmart mounting Induction Heater IH 070/090 Bedienungsanleitung Instructions for use Mode d emploi Manuale d istruzioni Manual de instrucciones
Induction Heater IH 070/090 Bedienungsanleitung Instructions for use Mode d emploi Manuale d istruzioni Manual de instrucciones 2 Induktionsheizgerät IH 070/090 simatec ag Table of contents EU Declaration
More informationSERVICE MANUAL REFRIGERATION
SERVICE MANUAL REFRIGERATION ELECTROLUX HOME PRODUCTS S.p.A. Publication no. Spares Operations Italy 599 37 75-07 Corso Lino Zanussi, 30 060824 I - 33080 PORCIA / PN (ITALY) ITZ/SERVICE/AA Fax +39 0434
More information3500 SERIES CONVECTION STEAM COOKER PARTS AND SERVICE MANUAL
3500 SERIES CONVECTION STEAM COOKER PARTS AND SERVICE MANUAL EFFECTIVE JULY 30, 2014 Superseding All Previous Parts Lists. The Company reserves the right to make substitution in the event that items specified
More informationRFD OVEN SERIES WITH PROTOCOL 3 TM CONTROLLER OWNER S MANUAL C-250 PN VERSION 1 11/2012
RFD OVEN SERIES WITH PROTOCOL 3 TM CONTROLLER OWNER S MANUAL C-250 PN 320253 VERSION 1 11/2012 PREFACE RFD Oven Series Owner s Manual 2 Version 1 Revision History Revision Date Author Description 1 11/2012
More informationOperation Manual. (Version 1.5.1)
VWR symphony TM Gravity Convection General Incubator Operation Manual (Version 1.5.1) VWR symphony Gravity Convection General Incubator 414004-610, 414004-612, 414004-614, 414004-616 414004-611, 414004-613,
More informationInduction Fryer. Instruction manual. Model Modèle Modell Modello Modelo Malli: CP793
Induction Fryer Instruction manual Model Modèle Modell Modello Modelo Malli: CP793 UK Safety Tips Position on a flat, stable surface. A service agent/qualified technician should carry out installation
More informationSERVICE MANUAL MULTI DECK 60 HOT MERCHANDISER MULTI DECK 100 HOT MERCHANDISER MULTI DECK 120 HOT MERCHANDISER CLASSIC DECK HOT MERCHANDISER
SERVICE MANUAL MULTI DECK 60 HOT MERCHANDISER MULTI DECK 100 HOT MERCHANDISER MULTI DECK 120 HOT MERCHANDISER CLASSIC DECK HOT MERCHANDISER Classic Deck Hot Multi Deck 60 Multi Deck 100 Multi Deck 120
More informationFreeze Protection Thermostat TRACON MODEL GPT 3 Installation and Operation Manual
MANUAL We manage heat Freeze Protection Thermostat TRACON MODEL GPT 3 Installation and Operation Manual Environmental Technology, Inc. 1850 N Sheridan Street South Bend, Indiana 46628 http://www.networketi.com/
More informationYCIV Hz & Hz
YCIV 0590-1500 50Hz & 0157-0397 60Hz Start-up Checklist SERVICE POLICY & PROCEDURES Supersedes: Nothing Form 201.23-CL1 (309) Commissioning PREPARATION Commissioning of this unit should only be carried
More informationProducts documentation (REVISION DATE: 03/10/2011) OMFP6010 (60cm PIROLITIC OVEN)
Products documentation (REVISION DATE: 03/10/2011) OMFP6010 (60cm PIROLITIC OVEN) Ovens Service Manual Models OMFP6010 CONTENTS This document has been published to be used for service only. The contents
More informationE32M CONVECTION OVEN SERVICE MANUAL. Revision 1/F
E32M CONVECTION OVEN SERVICE MANUAL -- WARNING: ALL INSTALLATION AND SERVICE REPAIR WORK MUST BE CARRIED OUT BY QUALIFIED PERSONS ONLY. -2- CONTENTS This manual is designed to take a more in depth look
More informationREACH-IN INCUBATOR MODEL: RI28/RI40 2/ INSTALLATION AND OPERATIONAL MANUAL
REACH-IN INCUBATOR MODEL: RI28/RI40 2/13 4861573-1 INSTALLATION AND OPERATIONAL MANUAL Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113 EMAIL: tech@shellab.com INTERNET: http://www.shellab.com/~shellab
More informationFS Series Forced Air Ovens. Operation and Service Manual
FS Series Forced Air Ovens Operation and Service Manual TestEquity LLC 6100 Condor Drive Moorpark, CA 93021 Support: 877-512-3457 Toll Free 805-480-0638 Corporate: 800-732-3457 805-498-9933 http://www.testequity.com
More informationSpa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert)
Spa Touch Control Panel with BP2100, BP6013 spa controllers. (Spa Owner s Manual insert) P.N. 7876C (export) February 12, 2015 For Spas equipped with BP2100, BP6013 controllers and Spa Touch panel. Spa
More informationU-CLASS SPA SYSTEMS. Dealer TechBook by Gecko Alliance. Technical Specifications, Programming, Installation & User's Instructions!
U-CLASS SPA SYSTEMS Dealer TechBook by Gecko Alliance Technical Specifications, Programming, Installation & User's Instructions! UC-CF TABLE OF CONTENTS Specifications 2 Description System features Electrical
More informationKQ Series. Multistage Booster Pump System. Installation Manual. ISO 9001 Certified KQ200/400 KQ800
KQ Series Multistage Booster Pump System Installation Manual KQ200/400 KQ800 ISO 9001 Certified Description The KQ series booster pump is an all-in-one compact and reliable automatic multistage centrifugal
More informationDRI-AIR INDUSTRIES, INC. ARID-X10C FLOOR MOUNT DRYER OPERATING MANUAL
ARID-X10C FLOOR MOUNT DRYER OPERATING MANUAL Page 1 DRI-AIR INDUSTRIES, INC. 16 THOMPSON ROAD P.O. BOX 1020 EAST WINDSOR, CT 06088-1020 Tel. (860) 627-5110 FAX (860) 623-4477 Internet http://www.dri-air.com
More informationCeramic Oscillating Heater. Instruction Booklet KCE95
Kambrook - Australia Building 2, Port Air Industrial Estate 1A Hale Street Botany NSW 2019 Australia Customer Service Line 1300 139 798 Customer Service Fax 1800 621 337 www.kambrook.com.au Kambrook -
More informationService Manual Model 3163
Service Manual Model 3163 Contents Important Safety Information.......... 1 Specifications.................. 2 General Information.............. 2 Direct Vent Requirements........... 2 Propane System................
More informationPackage Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents
- English... 1 Instruction Manual Analog, Dry Block Heater, 1 Block, HB1AL Analog, Dry Block Heater, 2 Block, HB2AL Analog, Dry Block Heater, 4 Block, HB4AL Analog, Dry Block Heater, 6 Block, HB6AL Digital,
More informationBrown University Revised June 29, 2012 Facilities Design & Construction Standards SECTION ELECTRICAL DESIGN CRITERIA
PART 1 - GENERAL 1.1 Background SECTION 26 00 10- ELECTRICAL DESIGN CRITERIA A. Brown University maintains it own campus electrical distribution system which serves the majority of the buildings and facilities
More informationLXG-300 Series. Models with LXG-300 Series Instruments
I.O.M. #147 updated : 08/16/2018 INSTRUCTION MANUAL INSTALLATION OPERATION MAINTENANCE LXG-300 Series Models with LXG-300 Series Instruments Model: Serial Number : 525 East Stop 18 Road Greenwood, IN 46142
More informationOWNER S MANUAL HIGH WALL INVERTER. (English) (BSHVD1S SERIES)
OWNER S MANUAL HIGH WALL INVERTER (English) (BSHVD1S SERIES) IMPORTANT As with any product that has moving parts or is subject to wear and tear, it is VERY IMPORTANT that you maintain your air conditioner
More informationSpa Touch Control Panel with 2000, 2100 controllers. (Spa Owner s Manual insert)
Spa Touch Control Panel with 2000, 2100 controllers (Spa Owner s Manual insert) P.N. 7876B February 11, 2015 For Spas equipped with BP2000, BP2100 controllers and Spa Touch panel. Spa Touch Control Panel
More informationSpeed and Frequency Seite 1 von 7
Speed and Frequency Seite 1 von 7 E16 Systems for High Safety Speed ing, all with Triple Modular Redundancy. A choice of versions to meet various demands. Compliant with SIL3 / IEC 61508 and/or API 670.
More informationSPECIFICATIONS Power Requirements: V, 9.5A, 60Hz V, 5.8A, 50Hz Including 2A for internal convenience outlet.
StableTemp Low Temperature Incubator Model 146E Catalog Number 39063-00(115VAC) 39063-05(230VAC StableTemp Low Temperature Incubator offers laboratories precise temperature control over the -10 C to 60
More informationPioneer-R16 Gas Monitor Operator s Manual
Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas
More informationINSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 870 SERIES INDUSTRIAL CONTROL PANELS
INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS FOR 870 SERIES INDUSTRIAL CONTROL PANELS GENERAL INDEECO Industrial Control Panels are designed to provide years of trouble free operation if properly
More informationSERVICE MANUAL MULTI DECK 35 HOT MERCHANDISER. Multi Deck 35 Pret a Manger
SERVICE MANUAL MULTI DECK 35 HOT MERCHANDISER Multi Deck 35 Pret a Manger - NOTICE - This manual is prepared for the use of trained Service Technicians and should not be used by those not properly qualified.
More informationSplit Portable Air Conditioner
SP-IOM-1 Please read and save this manual. Read carefully before attempting to assemble, install, operate or maintain the product described. Protect yourself and others by observing all safety information.
More informationINSTALLATION INSTRUCTIONS
Loop Alarm Model SC1290 and SCL1290 INSTALLATION INSTRUCTIONS LOVE LOVE CONTROLS a Division of Dwyer Instruments, Incorporated PO Box 338 Michigan City, IN 46361-0338 (800) 828-4588 (219) 879-8000 FAX
More informationARUBA SUN SERIES - LIGHTING SYSTEM USER MANUAL
ARUBA SUN SERIES - LIGHTING SYSTEM USER MANUAL 1 INTRODUCTION Thank you and congratulations for choosing Hamilton products. For over 20 years Hamilton has been providing aquarium lighting and leads the
More informationSERVICE DATA SHEET (9704) Rev. D
SERVICE DATA SHEET 318047401 (9704) Rev. D Electric Wall Ovens with Electronic Oven Control (ERC III - 318010700 & 318010900) NOTICE This service data sheet is intended for use by persons having electrical
More informationOPERATING MANUAL AIR CONDITIONER DUCT TYPE. ART Series KEEP THIS MANUAL FOR FUTURE REFERENCE FUJITSU GENERAL LIMITED P/N
OPERATING MANUAL AIR CONDITIONER DUCT TYPE ART Series KEEP THIS MANUAL FOR FUTURE REFERENCE FUJITSU GENERAL LIMITED P/N9374379293-02 CONTENTS SAFETY PRECAUTIONS... 1 NAME OF PARTS... 2 PREPARATORY OPERATION...
More information40KMC KMQ
40KMC------301 40KMQ------301 OWNER S MANUAL Split system Global cassette indoor unit IR Remote Control Room Controller Zone Manager The unit can be used with infrared Remote Control, with the Carrier
More information1.1 REFERENCES.1 American National Standards Institute/National Fire Protection Association (ANSI/NFPA).1 ANSI/NFPA 20, Centrifugal Fire Pumps.
Issued 2006/08/01 Section 13920 Packaged Fire Pumps and Controllers Page 1 of 7 PART 1 GENERAL 1.1 REFERENCES.1 American National Standards Institute/National Fire Protection Association (ANSI/NFPA).1
More informationUser Operation Manual
APPROVED BY: JBJ PRESCIENT III FIRE ALARM & GAS EXTINGUISHING CONTROL PANEL User Operation Manual PAGE 1 of 6 1 CONTENTS 1. FIRE - GENERAL... 3 2. FIRE - ZONES 1 OR 2 - AUTOMATIC MODE... 3 3. FIRE - ZONES
More informationSERVICE Manual FREE JOINT MULTI AIR CONDITIONER
FREE JOINT MULTI AIR CONDITIONER INDOOR UNIT MH020FPEA MH023FPEA MH026FPEA MH035FPEA MH052FPEA MH8VP2-09 MH9VP2-07 MH9VP2-2 MH026FKEA MH035FKEA MH052FDEA OUTDOOR UNIT MH8VP2X MH9VP2X MH052FXEA2 MH068FXEA4
More informationFire Risks and Safety Standards
Fire Risks and Safety Standards Florian Reil TÜV Rheinland Group Renewable Energies 51105 Cologne, Germany Phone: + 49 221 806 27 94 Email: Florian.Reil@de.tuv.com Workshop BIPV The Increasing Role of
More informationINSTRUCTION MANUAL. Laboratory Equipment Pty Ltd Ph: Fax:
INSTRUCTION MANUAL Laboratory Equipment Pty Ltd email: sales@labec.com.au Ph: 02 9560 2811 Fax: 02 9560 6131 www.labec.com.au 2014 Shaking Water Bath Instruction Manual Model : J-SWB19, J-SWB35, J-SWB60
More informationCompact Ultra Low Temperature Chamber
Compact Ultra Low Temperature Chamber MC-712R 812R CAT.NO.E14140-Z1804 High performance and reliability come in a compact package, covers wide range of temperature testing needs, from ultra-low temperature
More informationPR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA
Operating Instructions High Performance Refrigerator PR-L2466W- PA PR-L2466W-PA Please read these instructions carefully before using this product, and save this manual for future use. See page 11 for
More informationSUPER HIGH EFFICIENCY WATER HEATERS SUPPLEMENT TO INSTRUCTION MANUAL P/N (Replaces pg. 2 in instruction manual.) CONGRATULATIONS!
SUPER HIGH EFFICIENCY WATER HEATERS SUPPLEMENT TO INSTRUCTION MANUAL P/N 238-44219-00 (Replaces pg. 2 in instruction manual.) CONGRATULATIONS! You have just purchased one of the finest water heaters on
More informationUPRIGHT FREEZER WHS-502FWEW1(E) WHS-502FWEB1(E) WHS-502FWESS1(E)
UPRIGHT FREEZER WHS-502FWEW1(E) WHS-502FWEB1(E) WHS-502FWESS1(E) TABLE OF CONTENTS NAMES OF THE PARTS.2 IMPORTANT SAFETY INSTRUCTIONS..3 INSTALLATION INSTRUCTION BEFORE USING YOUR FREEZER...3 INSTALLING
More informationOperation & Maintenance Manual
Operation & Maintenance Manual for Model 2100 Fire+Smoke systems from Smoke Guard, Inc. 287 N. Maple Grove - Boise, ID 83704 Phone: (800) 574-0330 Fax: (208) 639-7851 E-Mail: info@smokeguard.com Website:
More informationELECTRIC CONVECTION HEATER WITH FAN AND TIMER
ELECTRIC CONVECTION HEATER WITH FAN AND TIMER FOR THE FOLLOWING MODELS: EE-8068 EE-8068R EE-8068W If you have any questions about the operation of your convection heater, please contact Crane Customer
More informationInternet Version for Reference Only INDUCED DRAFT COMMERCIAL WATER HEATERS SUPPLEMENT INSTRUCTIONS TO PART #
INDUCED DRAFT COMMERCIAL WATER HEATERS SUPPLEMENT INSTRUCTIONS TO PART #238-39387-00 THIS INSTRUCTION SUPPLEMENT IS ONLY INTENDED TO GIVE INSTALLATION INSTRUCTIONS AND INFORMATION RELATED TO THE INDUCED
More informationEmerson Inspire 1HDEZ Installation Instructions. Thermostat/Interface Equipment Control TROUBLESHOOTING
Emerson Inspire 1HDEZ-1521 Installation Instructions Thermostat/Interface Equipment Control TROUBLESHOOTING FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL
More informationOperation Manual SCT14B and SCT18B. Inspection. 3 General Description. 3 General Requirements. 3 Standard Features.
Spot Cooling Systems, Inc. 120 Century Drive Suite 00 Carrollton, TX 7006 00-6-776 Operation Manual SCT1B and SCT1B Warning! Improper installation, adjustment, alteration, service, or maintenance can cause
More informationInstallation & Operating Guide
5-036 HOT WATER TANK Installation & Operating Guide Read all instructions thoroughly. Keep this guide for future reference. Proof of purchase is required for Warranty. Staple receipt or proof of purchase
More information