TGM 8 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412)

Size: px
Start display at page:

Download "TGM 8 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412)"

Transcription

1 TGM 8 GAS Monitor instruction manual THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of complex equipment, this product will perform as designed only if installed, used and serviced in accordance with the manufacturer s instructions. OTHERWISE, IT COULD FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR DEATH. The warranties made by Mine Safety Appliances Company with respect to these products are voided if the products are not installed, used and serviced in accordance with the instructions in this manual. Please protect yourself and others by following them. We encourage our customers to write or call regarding this equipment prior to use or for any additional information relative to use or repair. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412) Mine Safety Appliances Company All Rights Reserved Manufactured by MSA Canada Inc., TORONTO, ONTARIO M3N 1X8 Distributed by MINE SAFETY APPLIANCES COMPANY PITTSBURGH, PENNSYLVANIA REV

2 World Wide Product Support Locations Headquarters MSA International P.O. Box 426 Pittsburgh, PA TEL: (412) FAX: (412) Australia MSA (Aust.) Pty. Ltd. 137 Gilba Road P.O. Box 43, Wentworthville Sydney. Girraween NSW 2145, Australia TEL: (02) FAX: (02) Brazil MSA do Brasil, Ltda. Av. Roberto Gordon. 138 Caixa Postal Diadema Sao Paulo, Brazil TEL: (11) FAX: (11) Canada MSA Canada Inc. 148 Norfinch Drive North York, Ontario M3N 1X8, Canada TEL: (416) FAX: (416) Chile MSA de Chile Ltda. Avda. Salvador 1434 Casilla 16647, Correo 9 Santiago, Chile TEL: (2) FAX: (2) China Wuxi-MSA Safety Equipment Co. Ltd. 305 Xicheng Road Wuxi City. Jiangsu Province Peoples Rep. Of China TEL: (510) FAX: (510) France MSA de France 13 rue de la Guivernone BP Cergy Pointoise Cedex TEL: (1) FAX: (1) Germany Auergesellschaft GmbH P.O. Box 620 D Berlin Germany TEL: (30) FAX: (30) Great Britain MSA (Britain) Ltd. East Shawhead Coatbridge, ML5 4TD Scotland, Great Britain TEL: (236) FAX: (236) Hungary MSA Auer Hungaria Biztonsagtechikai Kft. Gyomroi ut 140 H Budapest X. Hungary TEL: (1) FAX: (1) India MSA (India) Ltd. P-25 Transport Depot Rd Calcutta India TEL: (33) FAX: (33) Italy MSA Italiana S.p.A. P.O. Box Milano, Italia TEL: (2) FAX: (2) Japan MSA Japan Ltd Roppongi 6-Chome Minato-Ku Tokyo 106. Japan TEL: (3) FAX: (3) Mexico MSA de Mexico S.A. de C.V. Francisco 1. Madero 84 Apartado Naucalpan de Juarez (Edo de Mexico) TEL: (5) FAX: (5) Middle East MSA Middle East Bin Hamoodah Bldg. Khalifa St. P.O. Box Adu Dhabi United Arab Emirates TEL: (2) FAX: (2) Netherlands MSA Nederland B.V. P.O. Box AA Hoorn Netherlands TEL: (2290) FAX: (2290) Peru MSA del Peru S.A. Los Telares 139 Urb. Vulcano. Atc. Lima 3 Casilla 1933 (Mail) TEL: (14) FAX: (14) Russia Sibertech Monocrystal Institute Novosubirsk, Russia TEL: (7-3832) FAX: (7-3832) Southeast Asia MSA S.E. Asia Pte. Ltd. 51 Ayer Rajah Crescent #02-03 Singapore 0513 TEL: FAX: South Africa Boart MSA (Pty) Ltd. CNR. Kelvin St. & Steel Rd Spartan. P.O. Box 1710 Kempton Park 1620 TEL: (11) FAX: (11) 5648 Spain MSA Española S.A. Narciso Monturiol 7 Poligono Industrial Apartado de Correos Sant Just Desvern Barcelona TEL: (3) FAX: (3) Sweden AB Tegma Kopparbergsgatan 29 S Malmö TEL: (40) FAX: (40) Switzerland MSA (Switzerland) Ltd. Grubenstrasse CH-3072 Ostermundingen TEL: (31) FAX: (31) Zimbabwe MSA Zimbabwe (Pvt.) Ltd. P.O. Box AY280, AMBY Harare Tel: (4) FAX: (4)

3 1. The monitor described in this manual should be installed, operated, and maintained in strict accordance with any labels, cautions, warnings, drawings, instructions, and within the limitations stated. 2. The TGM Series must not be installed in areas or locations where explosive concentrations of combustible gases or vapours might occur in the atmosphere: Class I, Groups A, B, C, and D areas, as defined by the NEC. Because these monitors are not explosion-proof, they must be located in a non-hazardous area. 3. Use genuine MSA replacement parts when performing any maintenance on the TGM Series. Failure to do so may seriously impair instrument performance. Repair or alteration of the TGM Series beyond the scope of these instructions, or by anyone other than a certified MSA service person, could cause the product to fail to perform as designed and persons who rely on this product for their safety, could sustain severe bodily injury or death. 4. For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the instruction manual completely before operating or servicing a TGM Series.

4

5 TGM8 Table of Contents Table of Contents Section 1 General Information Introduction LED Indicators, Relays & Audible Alarm Warning and Alarm Condition, Sensor Failure Condition Section 2 Installation and Specifications Installation of Monitor Operating Specifications External Connection: Remote Sensors, Relays, Power Set-Up & Calibration Active Points Warning & Alarm Delays, Alarm Latch, Display Mode Audible Alarm Device Section 3 General Operation Scanning, Auto-Ranging Warning & Alarm Sensor Failure Section 4 Start-Up Procedure for TGM Calibration Log Calibration Log Parts List TOC-1

6 TGM8 Table of Contents Section 5 Figure 1: TGM8 Enclosure Drawing Figure 2: TGM8 Wiring Diagram Figure 3: TGM8 Panel Layout/Connections Figure 4: TGM8 Series Field Connection Diagram for RS-485 Series Communications Type Sensor Figure 5: TGM8 Field Connection Diagram for 4-20 ma Analog Signal Type Sensor Figure 6: TGM ma Sensor Wiring Section 6 CSA Certification FCC Compliance Appendix A Figure 7A: Troubleshooting Guide for TGM8... A-1 Figure 7B: Troubleshooting Guide for TGM8... A-2 Figure 7C: Troubleshooting Guide for TGM8... A-3 Figure 7D: Troubleshooting Guide for TGM8... A-4 TOC-2

7 TGM8 Section 1 General Information Section 1 General Information Introduction This manual contains instructions for the MSA TGM8 Series. Description: The TGM8 Series is a microprocessor-based monitoring system designed to interface with a wide range of MSA Sensors. A total of eight (8) Sensors can be connected to the TGM8. The system comes complete with audible and visual status indication, relay outputs, and adjustable warning and alarm levels, complete with reset. The alarm relay is selectable for non-latching or latching with H1-9. Standard factory configuration is the non-latching mode. Sensor Failure Condition: The TGM8 Monitor has Sensor OK LEDs. The Sensor OK LED (Green) goes out if anyone of the remote sensor(s) fail and the sensor fail relay is energized. The sensor fail relay is non-latching. LED Indicators & Relays & Audible Alarm Warning Condition: When the gas concentration exceeds the warning setpoint, the AMBER warning LED turns on and the warning relay is activated. The warning relay is non-latching. When the Warning condition disperse below the setpoint, the relay will re-set. Alarm Condition: When the gas concentration exceeds the alarm setpoint, the RED Alarm LED turns on, the common alarm relay is activated and the audible alarm sounds. 1-1

8

9 TGM8 Section 2 Installation and Specifications Section 2 Installation and Specifications Installation of Monitor Install the TGM8 on a wall or structure that is free from shock or vibration. It should be protected against physical damage, water, heat and direct sunlight. Install the Monitor so that its door can be opened and accessed for service. Operating Specifications Power Requirements: Standard 110 volts AC 50/60 Hz ± 0.5 Amps or (Optional) 220 volts AC 50/60 Hz ± 0.25 Amps or (Optional) 24 volts DC ± 1.5 Amps max. Signal Input: 4 to 20 milliamps, 2 or 3-wire connection, OR RS-485, 4-wire connection. Temperature: Operating: Storage: Sensor Power: 0 to 40 C (32 to 104 F) -10 to 50 C (14 to 122 F) 24 Volts 200 milliamps Relay Contact Ratings: 10 Amps 1/8 H.P., 125 Vac 6 Amps 1/8 H.P., 277 Vac 5 Amps 30 Vdc Number of Relays: WARNING: 8 Relay SPDT, ALARM: 1 Relay SPDT, TROUBLE: 1 Relay SPDT Digital Readout Scale (Optional): 4 Digit Display (Point # and Gas Level) Status Display: Discrete LED for WARNING, ALARM, and SENSOR OK for each point. Alarm & Warning Setpoints: Field Selectable: Warning = 25% or 35% of Full Scale Alarm = 50% or 100% of Full Scale (unless otherwise specified when ordering). Sensor Fail Setpoint: Non-adjustable Time Delays: Fixed delay of 5 minutes OFF delay for Warning and fixed delay of 5 minutes ON delay for Alarm. Field Selectable feature. 2-1

10 TGM8 Section 2 Operating Specifications Audible Alarm: Solid ON-Standard Factory Setting ( ft.) OR Variable Sound Pulse Calibration: All units are tested and pre-calibrated at the factory. Enclosure: Standard - Metal NEMA 1 Optional - Fiberglass NEMA 4 Dimensions: Height: 15.0 (380 mm) Width: 12.0 (305 mm) Depth: 3.75 (95 mm) 2-2

11 TGM8 Section 2 External Connections External Connections Perform all wiring and conduit installation in accordance to the Canadian Electrical Code or the National Electrical Code or to local electrical requirements. Use separate conduits for power wires, relay wires and signal wires. Remote Sensors: A total of eight (8) remote sensors can be connected to the TGM8 Monitor. Sensor may be a 4-20 ma analog signal type, 2- or 3-wire connector, or RS-485 serial communications type. The RS-485 type sensors may be of different gas type and range. The TGM8 Monitor will automatically adjust itself to read and interpret the RS-485 sensors properly. Use shielded cable for wiring installation. Signal wiring must not be installed in the same conduit with power or relay wiring. Refer to appropriate sensor wiring diagram (pp. 5-4, 5-5, 5-6) and sensor instruction manual for more details. WHEN CONNECTING THE REMOTE SENSORS, MAKE SURE THAT ALL WIRING IS CORRECT AND THE FOUR LEADS OF THE RS-485 BUS ARE NOT INTERCHANGED, OR PERMANENT DAMAGE TO THE SENSOR MAY RESULT. NOTE MAKE SURE THAT EACH RS-485 TYPE SEN- SOR IS GIVEN A UNIQUE ADDRESS ON ITS AD- DRESS JUMPERS, OR THE TGM8 MONITOR WILL NOT BE ABLE TO COMMUNICATE WITH IT. Relays: Perform all wiring and conduit installation in accordance to the Canadian Electrical Code or the National Electrical Code or to local electrical requirements. Use separate conduits for power wires, relay wires and signal wires. Remote Sensors: A total of eight (8) remote sensors can be connected to the TGM8 Monitor. Sensor may be a 4-20 ma analog signal type, 2- or 3-wire connector, or RS-485 serial communications type. The RS-485 type sensors may be of different gas type and range. The TGM8 Monitor will automatically adjust itself to read and interpret the RS-485 sensors properly. Use shielded cable for wiring installation. Signal wiring must not be installed in the same conduit with power or relay wiring. Refer to appropriate sensor wiring diagram (pp. 5-4, 5-5, 5-6) and sensor instruction manual for more details. WHEN CONNECTING THE REMOTE SEN- SORS, MAKE SURE THAT ALL WIRING IS CORRECT AND THE FOUR LEADS OF THE RS-485 BUS ARE NOT INTERCHANGED, OR PERMANENT DAMAGE TO THE SENSOR MAY RESULT. NOTE MAKE SURE THAT EACH RS-485 TYPE SENSOR IS GIVEN A UNIQUE ADDRESS ON ITS ADDRESS JUMPERS, OR THE TGM8 MONITOR WILL NOT BE ABLE TO COMMUNICATE WITH IT. 2-3

12 TGM8 Section 2 Setup & Calibration Setup & Calibration Active Points: The number of active scanning points (same as number of remote sensors) can be set by the first three jumpers (H1-1, H1-2, H1-3). The jumpers are binary encoded. If no jumpers are inserted, then only one point (address 1) is scanned. To scan more points, insert jumpers such that the decimal equivalent of the binary code on H1-1, H1-2 and H1-3 plus one, equals the number of points to be scanned. (Refer to Figure 3, Page 5-3). Warning & Alarm Delays: There is a fixed delay of five minutes available for both warning and alarm. This feature is selected with jumper H1-5. When the jumper is out, there is no delay on the warning and alarm relays. When the jumper is inserted, a fixed fiveminute delay is imposed on the warning and alarm. In the case of the warning, this delay is an OFF delay meaning that the warning relay(s) will remain activated for five minutes after the warning condition has been diminished. In the case of the alarm, this delay is an ON delay, meaning that it takes five minutes for the alarm relay (and audible device) to activate after the first alarm condition occurs. (Refer to Figure 3, Page 5-3). Alarm Latch: Normally, all relays return to their inactive state once their condition has cleared. The alarm relay can be made to latch in the activated position, even after all alarms have cleared. By inserting jumper H1-9, the alarm relay will remain latched after all alarms clear until the reset button is pressed. (Refer to Figure 3, Page 5-3). NOTE Pressing the reset button during an alarm condition does not affect the latching. The reset button must be pressed after the alarm condition has cleared in order to de-activate the alarm relay. Display Mode: The display of the TGM8 is auto-ranging for all RS-485 type serial sensors. The display mode jumpers have no effect on any RS-485 type sensors. The display mode jumpers take effect only if analog sensor inputs are used. The incoming analog signals are scaled to the full scale setting of the display mode jumpers. There are two sets of display ranges available: Range A and Range B. Jumper H1-4 selects which display range is active. If jumper H1-4 is OUT, Range A is selected; if jumper H1-4 is IN, Range B is selected. The following table shows the various jumper positions for the different display modes: 2-4

13 TGM8 Section 2 Setup & Calibration Range A: Full Scale H2-2 H out out 100 out in 10.0 in out 10.0/100* in in Range B: To select the solid-on audible alarm mode, jumper H1-6 must be inserted. To select the variable-on audible alarm mode, jumper H1-6 must be removed (Refer to Figure 3). In both modes, pressing the reset button will silence the audible alarm, but will not have any effect on the alarm LED on the front panel or the alarm relay. Full Scale H2-2 H out out 500 out in 250 in out 75 in in * This is a special Display Range that displays the first four channels (1-4) as ppm and the second four channels (5-8) as ppm. This is useful for NO 2 / CO applications. Audible Alarm Device: The TGM8 has two audible alarm modes. The first is a solid-on mode that sounds the audible device continuously while the TGM8 is in an alarm state. The second is a variable-on mode that sounds the audible device in an eight second interval pulsed fashion. The time between pulses is fixed, but the time of the audible alarm is determined by the amount that the alarm sensor is exceeding the alarm setpoint. For example, if the setpoint is at 50% and an alarm comes in at 55%, the audible alarm would give short beeps every eight seconds. As the alarm level increases from 55% to 100%, the beeps become longer in duration until a solid tone appears at 100%. 2-5

14

15 TGM8 Section 3 General Operation Section 3 General Operation Scanning: The TGM8 will scan each active point in sequence, starting with point one. It will scan up to the last active point and then repeat from point one. As each point is scanned, the point number and the sensor input value are displayed for two seconds. Auto-Ranging: The TGM8 automatically detects the type of sensor connected on an RS-485 bus and determines its full scale value. All setpoints and the display are adjusted accordingly to match the various sensors full scale levels. If an analog sensor is connected, the display must be manually selected for the appropriate sensor full scale level using the display mode jumpers (Refer to Section 2 Display Mode on page 2-4). Warning and Alarm: The alarm function on the TGM8 is common to all points. Each point has individual warning relays and discrete LED s for Sensor OK, Warning, and Alarm. All points ar etied to a common alarm relay. The audible alarm will sound every time a new alarm occurs. It may be silenced by pressing the reset button on the front door of the monitor. LED for that point will turn on and after a delay defined by jumper H1-5, the relay will activate. At the same time the relay activates, the audible alarm will sound. The audible alarm may be silenced by pressing the reset button. If all active points drop below the alarm setpoint, the alarm will silence (if not silenced already by reset button) and the alarm level relay will de-activate if jumper H1-9 is not inserted. If jumper H1-9 is inserted for latching function, then the alarm relay will not de-activate automatically once all active points drop below the alarm setpoint. The relay will only deactivate once the reset button is pressed. Sensor Failure: If the TGM8 loses communications with one or more sensors, or if one or more sensors fail, that sensor OK LED will turn off and the sensor fail relay will activate. If an analog sensor is used and the input is between 1 ma and 3 ma, the sensor OK LED will turn off and the sensor fail relay will activate. Any analog input below 1 ma causes the TGM8 to ignore that analog input and treat that sensor as an RS-485 type sensor. If the level of any or more of the active points exceed(s) the warning setpoint, the warning LED for that point will turn on and the warning relay for that point will activate. If all active points drop below the warning setpoint, the warning LED will turn off and after a delay defined by jumper H1-5, the relay will return to its inactive state. If the level of any or more of the active points exceed(s) the alarm setpoint, the alarm level 3-1

16

17 TGM8 Section 4 Start-Up Procedure Section 4 Start-Up Procedure for TGM8 A B (A = Remote Sensor(s) B = TGM8 Monitor) Loosen the locking screws on front cover to open door. Check wiring and locate the power switch on main circuit board inside enclosure. Loosen the locking screws on front cover and open the sensor. Check wiring. * Ensure tha the binary coded jumper on H1-1 is properly inserted to reflect number of remote sensor(s) connected to TGM8 Monitor. Ensure that the binary coded jumper on H1-1 is properly inserted to address each individual sensor. (If RS-485, 4-wire connection to TGM8 Monitor). Turn power ON. * Green LED (POWER) Indicator on front panel should be ON. Green LED (Sensor OK) Indicator on front panel should be ON. * Red LED (SENSOR FAIL) Indicator on front panel should be OFF. * Amber LED (WARNING) Indicator on front panel should be OFF. * Red LED (ALARM) Indicator on front panel should be OFF. Note: Warning LED and Alarm LED may come on during initial power up. If audible alarm sounds, press RESET button to silence. IMPORTANT: Allow the gas monitor and sensor** to stabilize. * Applicable only to MSA UNTGS Sensor/Transmitter. ** Refer to Sensor/Transmitter manual for stabilization time. 4-1

18 TGM8 Section 4 Start-Up Procedure A B (A = Remote Sensor(s) B = TGM8 Monitor) Ensure that delay function selection jumper on main circuit board is NOT inserted. * Insert Test Jumper to 100% test position. Red LED (SENSOR FAIL), Amber LED (WARNING), Red LED (ALARM) on the Remote Sensor shall come on. (See Note 1). The audible horn sounds, Amber LED (WARNING) and Red LED (ALARM) on TGM8 Monitor shall also come on. (Press the RESET BUTTON to silence horn). Note: Warning and Alarm Relays will energize. Any remote equipment which may be connected to the associated relay contacts should now be activated. Note: TGM8 Monitor is factory preset as follows: WARNING Level: ALARM Level: Audible Signal: Delay Function: 35% of Full Scale 50% of Full Scale Activates on at ALARM level No Delay (jumper not inserted) The MSA TGM8 Monitor allows for field adjustment from factory settings. Please refer to Section 2 of this manual to change settings in order to meet the required sequence of operation. * * Remove Test Jumper from 100% test position. Repeat this alarm simulation test on other sensor(s) if needed. The initial function test and start-up is now completed. Close front door and sensor cover and secure locking screws. Note 1: Remote Sensor, other than MSA UNTGS Series Sensor, must use calibration check gas to simulate alarm condition. * Applicable only to MSA UNTGS Sensor/Transmitter. 4-2

19 TGM8 Section 4 Calibration Log Calibration Log DATE READING RESET DATE READING RESET 4-3

20 TGM8 Section 4 Calibration Log Calibration Log DATE READING RESET DATE READING RESET 4-4

21 TGM8 Section 4 Parts List Parts List * When ordering replacement parts, Please State MSA s P/N and S/N of unit. Item TGM8 Mother Board 25/35 & 50/100 (110 Vac Version) TGM8 Display Board Complete TGM8 Display Board, Less Display Ribbon Cable Assembly MSA Part No Use only genuine MSA replacement parts when performing any maintenance on the TGM8 Monitor. Failure to do so may seriously impair instrument performance. Repair or alteration of the TGM8 Series Monitor beyond the scope of the instruction manual could cause the product to fail to perform as designed. Disconnect all power source(s) to the TGM8 before removing or changing any components. 4-5

22

23 TGM8 Section 5 Figure 1 Section 5 Figure 1: TGM8 Enclosure Drawing 5-1

24 TGM8 Section 5 Figure 2 Section 5 Figure 2: TGM8 Wiring Diagram 5-2

25 TGM8 Section 5 Figure 3 Section 5 Figure 3: TGM8 Panel/Layout Connections 5-3

26 TGM8 Section 5 Figure 4 Section 5 Figure 4: TGM8 Series Field Connection Diagram for RS-485 Series Communications Type Sensor 5-4

27 TGM8 Section 5 Figure 5 Section 5 Figure 5: TGM8 Field Connection Diagram for 4-20 ma Analog Signal Type Sensor 5-5

28 TGM8 Section 5 Figure 6 Section 5 Figure 6: TGM ma Sensor Wiring 5-6

29 TGM8 Section 6 CSA Certification & FCC Compliance Section 6 CSA Certification Certificate of Compliance No. LR FCC Compliance This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: 1. Device may not cause harmful interference. 2. Device must accept any interference received, including interference that may cause undesired operation. NOTE This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at his or her own expense. Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment. 6-1

30

31 TGM8 Appendix A Figure 7A Appendix A Figure 7A: Troubleshooting Guide for TGM8 A - 1

32 TGM8 Appendix A Figure 7B Appendix A Figure 7B: Troubleshooting Guide for TGM8 A - 2

33 TGM8 Appendix A Figure 7C Appendix A Figure 7C: Troubleshooting Guide for TGM8 A - 3

34 TGM8 Appendix A Figure 7D Appendix A Figure 7D: Troubleshooting Guide for TGM8 A - 4

35

36 Manufactured by MSA Canada Inc. TORONTO, ONTARIO, CANADA M3N 1X8 Call Toll 1 (800)

TGM 1 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412)

TGM 1 GAS Monitor. instruction manual. INSTRUMENT DIVISION MSA-INST or FAX (412) MSA INTERNATIONAL (412) or FAX (412) TGM 1 GAS Monitor instruction manual THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of complex

More information

Controllers. Instruction Manual WARNING

Controllers. Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

Instruction Manual WARNING

Instruction Manual WARNING Controllers Instruction Manual WARNING THIS MANAUL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like any piece of

More information

Remote Relay Module (RRM)

Remote Relay Module (RRM) Remote Relay Module (RRM) Instruction Manual WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING OR SERVICING THIS PRODUCT. Like

More information

SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM. User Instructions For The Model 5200 Gas Monitor System

SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM. User Instructions For The Model 5200 Gas Monitor System SPECIFICATION MODEL 5200 GAS MONITOR SYSTEM User Instructions For The Model 5200 Gas Monitor System To completely customize the specifications to your exact application, modifications to the following

More information

IT COULD FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR DEATH.

IT COULD FAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR DEATH. Chillgard LC Control Module Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR INSTALLING, USING, MAINTAINING OR SERVICING

More information

SPECIFICATION CHEMGARD INFRARED GAS MONITOR SERIES

SPECIFICATION CHEMGARD INFRARED GAS MONITOR SERIES SPECIFICATION CHEMGARD INFRARED GAS MONITOR SERIES USER INSTRUCTIONS FOR THE CHEMGARD INFRARED GAS MONITOR SERIES To completely customize the specification to your exact application, modification to the

More information

SAFETY MANUAL. UltraSonic IS-5 Ultrasonic Gas Leak Detector

SAFETY MANUAL. UltraSonic IS-5 Ultrasonic Gas Leak Detector SAFETY MANUAL UltraSonic IS-5 Ultrasonic Gas Leak Detector The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically

More information

Operating Manual ALTAIR QuickCheck Station ALTAIR 4 QuickCheck Station

Operating Manual ALTAIR QuickCheck Station ALTAIR 4 QuickCheck Station Operating Manual ALTAIR QuickCheck Station ALTAIR 4 QuickCheck Station Order No. 10079853/01 MSA AUER GmbH Thiemannstrasse 1 D-12059 Berlin Germany MSA AUER GmbH. All rights reserved MSA Declaration of

More information

STT 3000 Series STT250 /150 Surge Protection Device Specifications

STT 3000 Series STT250 /150 Surge Protection Device Specifications STT 3000 Series STT250 /150 Surge Protection Device Specifications Part Number: 46188660-501 EN2I-6061 June 2011 Introduction This device is designed to give the Smart Temperature Transmitter maximum protection

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING, MAINTAINING OR SERVICING THIS PRODUCT.

THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING, MAINTAINING OR SERVICING THIS PRODUCT. Chillgard RT Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING, MAINTAINING OR SERVICING THIS PRODUCT. Like any piece

More information

Instruction Manual PCS-400-XY. Pump Controller. Y=2: 4-20 ma Y=6: 0-10 VDC X=5: 1-5 VDC. 4 Pump Control Relays 1 Alarm Relay 1 Signal Fail

Instruction Manual PCS-400-XY. Pump Controller. Y=2: 4-20 ma Y=6: 0-10 VDC X=5: 1-5 VDC. 4 Pump Control Relays 1 Alarm Relay 1 Signal Fail Manufacturers of Process Controls and Instrumentation Instruction Manual Model: Function: Input: PCS-400-XY Pump Controller X=2: 4-20 ma X=5: 1-5 VDC X=6: 0-10 VDC X=7: Output: Power: 4 Pump Control Relays

More information

FF-SR Series. FF-SR25933 Two-hand Safety Control Module FF-SR25933

FF-SR Series. FF-SR25933 Two-hand Safety Control Module FF-SR25933 Two-hand Safety Control Module FF-SR Series FEATURES Complies with the Machinery Directive 98/37/EC and UL 508 Type III C per EN 574, Category 4 per EN 954-1 Inputs for two dual-contact pushbuttons 0,5

More information

Nitrogen Dioxide (NO2) Single-Point Gas Detection System

Nitrogen Dioxide (NO2) Single-Point Gas Detection System Nitrogen Dioxide (NO) Single-Point Gas Detection System DESCRIPTION Wall-mounted gas monitor with built-in nitrogen dioxide (NO)/diesel fume gas sensor, accepts one analog remote device such as a secondary

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

FUNCTIONAL SAFETY MANUAL

FUNCTIONAL SAFETY MANUAL FUNCTIONAL SAFETY MANUAL UltraSonic EX-5 Ultrasonic Gas Leak Detector The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent

More information

Dräger VARIOx-2. Electronical measuring device for the gas concentration measurement

Dräger VARIOx-2. Electronical measuring device for the gas concentration measurement Dräger VARIOx-2 Electronical measuring device for the gas concentration measurement The Dräger VARIOx-2 is used to test and adjust furnaces and auxiliary heating. Optimising combustion to save energy and

More information

1022 R/T Remote Monitor

1022 R/T Remote Monitor 1022 R/T Remote Monitor INSTRUCTIONS AMC-1022R/T WITH REMOTE SENSOR OR TRANSMITTER INSTALLATION AND OPERATING INSTRUCTIONS FOR THE AMC-1022R/T UNIT IMPORTANT: Please read these installation and operating

More information

Two-Channel Gas Controller

Two-Channel Gas Controller Two-Channel Gas Controller Specifications subject to change without notice. USA 09 Page of DESCRIPTION Highly configurable, UL 0 performance-tested and -certified, and wall-mounted gas monitor; continuously

More information

SAFETY MANUAL. FlameGard 5 MSIR Multi-Spectral Infrared Flame Detectors

SAFETY MANUAL. FlameGard 5 MSIR Multi-Spectral Infrared Flame Detectors SAFETY MANUAL FlameGard 5 MSIR Multi-Spectral Infrared Flame Detectors The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent

More information

Safe T Net 210. Instruction Manual Part Number Mar2010

Safe T Net 210. Instruction Manual Part Number Mar2010 Safe T Net 210 Instruction Manual Part Number 71-0012 31Mar2010 2007 Thermo Fisher Scientific Inc. All rights reserved. Specifications, terms and pricing are subject to change. Not all products are available

More information

Dräger PointGard 2100 Detection of toxic gases and vapours

Dräger PointGard 2100 Detection of toxic gases and vapours Dräger PointGard 2100 Detection of toxic gases and vapours The Dräger PointGard 2100 series is a self-contained gas detection system for the continuous area monitoring of toxic gases in ambient air PointGard

More information

Model 9010 and 9020 Monitoring Systems

Model 9010 and 9020 Monitoring Systems Model 9010 and 9020 Monitoring Systems Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like

More information

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd. F-2000 PC and AO OUTPUT BOARDS INSTRUCTION MANUAL Blue-White R Industries, Ltd. 500 Business Drive Huntington Beach, CA 92649 USA Phone: 714-89-8529 FAX: 714-894-9492 E mail: sales@blue-white.com or techsupport@blue-white.com

More information

Oxygen (O2) Single-Point Gas Detection System

Oxygen (O2) Single-Point Gas Detection System Oxygen (O) Single-Point Gas Detection System DESCRIPTION Wall-mounted gas monitor with built-in oxygen (O) sensor, accepts one analog remote device such as a secondary gas sensor, temperature or humidity

More information

Combustible Gas Detection and Control System

Combustible Gas Detection and Control System Combustible Gas Detection and Control System DESCRIPTION Gas monitor with built-in combustible gas sensor, wall-mounted, accepts inputs from remote devices such as other gas sensors, temperature or humidity

More information

POSITIONERS CONTROLS ELECTRO-PNEUMATIC INTELLIGENCE FOR DOUBLE & SINGLE ACTING PNEUMATIC ACTUATORS SERIES 6A. The High Performance Company CELEBRATING

POSITIONERS CONTROLS ELECTRO-PNEUMATIC INTELLIGENCE FOR DOUBLE & SINGLE ACTING PNEUMATIC ACTUATORS SERIES 6A. The High Performance Company CELEBRATING POSITIONERS ELECTRO-PNEUMATIC INTELLIGENCE CONTROLS SERIES 6A FOR DOUBLE & SINGLE ACTING PNEUMATIC ACTUATORS CELEBRATING The High Performance Company Y E A R S INTELLIGENCE Series 6A Intelligent Electro-Pneumatic

More information

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual ! WARNING MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2 TM, and MiniClO 2 TM Responder Detectors Instruction Manual "! WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE

More information

Dräger X-zone 5500 Portable gas detection system for flexible area monitoring

Dräger X-zone 5500 Portable gas detection system for flexible area monitoring Portable gas detection system for flexible area monitoring State-of-the-art area monitoring the in combination with the Dräger X-am 5000, 5100 or 5600 gas detection instruments can be used for the measurement

More information

Dräger P4000 Pressure and Leak Testing

Dräger P4000 Pressure and Leak Testing Dräger P4000 Pressure and Leak Testing The multifunction Dräger P4000 for testing natural gas lines, liquefied petroleum gas lines and drinking water systems is particularly handy and light. The ready-made

More information

GasSens. Self Checking Gas Sensors. Modular Gas Detector... Gas & Range Availability

GasSens. Self Checking Gas Sensors. Modular Gas Detector... Gas & Range Availability GasSens Modular Gas Detector... Self Checking Gas Sensors GasSens is a flexible component system providing a variety of options to meet individual gas detection and alarm requirements. From chemical and

More information

Dräger Polytron 8310 Stationary gas detector (DSIR)

Dräger Polytron 8310 Stationary gas detector (DSIR) Stationary gas detector (DSIR) SIMPLE INSTRUMENT MANAGEMENT POISON-RESISTANT AND FAIL-SAFE THANKS TO DIGITAL COMMUNICATION The DrägerSensor IR offers several benefits: technological advantages such as

More information

Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20

Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20 Solid Core and Split Core 4-20 ma Output Current Sensors CTS-20; CTP-20 PRODUCT DATA FEATURES Solid or split core loop-powered current transmitters Fast response time Integral DIN rail mounting flange

More information

GasScanner 4C. Four Channel Monitor. Operator s Manual. MINT-0280-XX Rev. A 02/25/08

GasScanner 4C. Four Channel Monitor. Operator s Manual. MINT-0280-XX Rev. A 02/25/08 GasScanner 4C Four Channel Monitor Operator s Manual MINT-0280-XX Rev. A 02/25/08 Product Warranty Matheson Tri-Gas., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship,

More information

Accuracy counts! Sensors with analog output

Accuracy counts! Sensors with analog output Accuracy counts! Sensors with analog output Company Contrinex is a leading international high-tech sensor company with headquarters in Switzerland. Contrinex offers quality and highly innovative sensing

More information

BUILD the clinic you ve always imagined.

BUILD the clinic you ve always imagined. BUILD the clinic you ve always imagined. With the Varian VitalBeam radiotherapy system, now you can affordably take your treatment capabilities further. Discover how you can build upon the technology itself

More information

Operation Manual. Rev. E GEM-MRI. Oxygen Sampling System.

Operation Manual. Rev. E GEM-MRI. Oxygen Sampling System. Operation Manual Rev. E2 2015.03 GEM-MRI Oxygen Sampling System www.critical-environment.com GEM-MRI - Operation Manual Rev. E2 2015.03 Table of Contents 1 POLICIES... 4 1.1 Important Note...4 1.2 Warranty

More information

Model 9010 and 9020 Monitoring Systems

Model 9010 and 9020 Monitoring Systems Model 9010 and 9020 Monitoring Systems Instruction Manual " WARNING THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like

More information

FUNCTIONAL SAFETY MANUAL. Gassonic Observer-H and Observer-i Ultrasonic Gas Leak Detector

FUNCTIONAL SAFETY MANUAL. Gassonic Observer-H and Observer-i Ultrasonic Gas Leak Detector FUNCTIONAL SAFETY MANUAL Gassonic Observer-H and Observer-i Ultrasonic Gas Leak Detector The information and technical data disclosed in this document may be used and disseminated only for the purposes

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

HPLC Column Heater Instruction Manual

HPLC Column Heater Instruction Manual HPLC Column Heater Instruction Manual For Technical Service Call 1-800-4BIORAD (1-800-424-6723) Warranty Unless otherwise specified, instruments sold by Bio-Rad Laboratories are warranted for 1 year against

More information

Thermo Scientific Ramsey TM Series Sensing Speed Conditions on Rotating Shafts and Machinery

Thermo Scientific Ramsey TM Series Sensing Speed Conditions on Rotating Shafts and Machinery Thermo Scientific Ramsey TM Series 60-200 Sensing Speed Conditions on Rotating Shafts and Machinery The Thermo Scientific Ramsey Series 60-200 monitors under-speed, overspeed and zero-speed conditions

More information

Dräger Polytron 5100 EC Detection of toxic gases and vapors

Dräger Polytron 5100 EC Detection of toxic gases and vapors Dräger Polytron 5100 EC Detection of toxic gases and vapors The Dräger Polytron 5100 EC is a cost-effective explosion proof transmitter for the detection of toxic gases or oxygen It uses a high performance

More information

AMC-RAM-4. Refrigerant Alarm Module USER MANUAL. Please read the installation and operating instructions completely and carefully before starting.

AMC-RAM-4. Refrigerant Alarm Module USER MANUAL. Please read the installation and operating instructions completely and carefully before starting. AMC-RAM-4 Refrigerant Alarm Module USER MANUAL IMPORTANT: Please read the installation and operating instructions completely and carefully before starting. Filename: 3310405B, DOC, AMC_RAM4_Manual.doc

More information

T7984 A,B,C Electronic Modulating Control Thermostats

T7984 A,B,C Electronic Modulating Control Thermostats T7984 A,B,C Electronic Modulating Control Thermostats FEATURES PRODUCT DATA APPLICATION These microprocessor-based thermostats provide proportional - integral (PI) individual room temperature control in

More information

Beacon 200 Gas Monitor Operator s Manual

Beacon 200 Gas Monitor Operator s Manual Beacon 200 Gas Monitor Operator s Manual Part Number: 71-0059RK Revision: J Released: 1/31/18 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

Operation Manual Part-No

Operation Manual Part-No Version 2.2, 08/2016 VISATRON Remote Indicator II For Oil Mist Detectors Operation Manual Part-No. 180050 VNxxx/87plus VNxxx/93 VNxxx/87-EMC VN115/87plus EX VN301plus (EX) Part-No. 150998 Page 2 of 20

More information

Dräger Polytron 8700 IR Detection of flammable gases and vapors

Dräger Polytron 8700 IR Detection of flammable gases and vapors Dräger Polytron 8700 IR Detection of flammable gases and vapors The Dräger Polytron 8700 IR is an advanced explosion proof transmitter for the detection of combustible gases in the lower explosion limit

More information

Type 2 safety light curtain Type 2 according to the IEC/EN parts 1 and 2 Compact profile: 31 x 32 mm 30 mm resolution for hand detection Wide r

Type 2 safety light curtain Type 2 according to the IEC/EN parts 1 and 2 Compact profile: 31 x 32 mm 30 mm resolution for hand detection Wide r Type 2 safety light curtain FF-SLD Series FEATURES Active Optoelectronic Protective Device compliant with the requirements of the IEC/EN 61496-parts 1 and 2 European Norms 30 mm resolution for hand detection

More information

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08 GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,

More information

Dräger Polytron 5720 IR Detection of carbon dioxide

Dräger Polytron 5720 IR Detection of carbon dioxide Dräger Polytron 5720 IR Detection of carbon dioxide The Dräger Polytron 5720 IR is a cost effective explosion proof transmitter for the detection of carbon dioxide in percent volume or ppm It uses a high

More information

SEC 2000 Millenium Infrared Gas Detector

SEC 2000 Millenium Infrared Gas Detector SEC 2000 Millenium Infrared Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email: sales@sensorelectronic.com

More information

Dräger Polytron 5200 CAT Detection of flammable gases and vapours

Dräger Polytron 5200 CAT Detection of flammable gases and vapours Dräger Polytron 5200 CAT Detection of flammable gases and vapours The Dräger Polytron 5200 CAT is a cost-effective explosion-proof transmitter for the detection of flammable gases in the lower explosion

More information

PRODUCT INFORMATION. GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL. In-situ Gas Analyzers

PRODUCT INFORMATION. GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL. In-situ Gas Analyzers PRODUCT INFORMATION GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL In-situ Gas Analyzers MONITORING CARBON MONOXIDE EMISSIONS INNOVATIVELY AND EFFICIENTLY

More information

Ramsey Series Motion Monitoring Systems Accurate Sensing of Speed Conditions on Rotating Shafts and Machinery

Ramsey Series Motion Monitoring Systems Accurate Sensing of Speed Conditions on Rotating Shafts and Machinery Ramsey Series 60-200 Motion Monitoring Systems Accurate Sensing of Speed Conditions on Rotating Shafts and Machinery Ramsey Series 60-200 Motion Monitoring Systems offer versatile and reliable packages

More information

CLEANROOM MONITOR CR3A Network - Installation Instructions

CLEANROOM MONITOR CR3A Network - Installation Instructions CLEANROOM MONITOR CR3A Network - Installation Instructions INTRODUCTION The CR3 Series Cleanroom Monitor, was developed specifically to allow for monitoring of confined spaces with accuracy and reliability.

More information

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL

HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL HALOGUARD TM IR MONITOR (OPTIONAL) FOUR or EIGHT POINT SCANNER INSTRUCTION MANUAL SE R IA L N O. M O D E L N O. Tem p. Range 1 - = > 60 o F 2 - = 40-60 o F 3 - = < 40 o F G as Type 1 - R-11 2 - R-12 3

More information

Dry Bath Heating Systems

Dry Bath Heating Systems Dry Bath Heating Systems Instruction Manual N2400-4001 Dry Bath Heating System, Single Block, Digital N2400-4002 Dry Bath Heating System, Dual Block, Digital www.starlabgroup.com DECLARATION OF CONFORMITY

More information

MD Munters Drive. Instruction Manual. Munters Drive * Motor and Controller Replacement Kit. Motor & Controller Replacement kit for AX and VX Fans

MD Munters Drive. Instruction Manual. Munters Drive * Motor and Controller Replacement Kit. Motor & Controller Replacement kit for AX and VX Fans Instruction Manual MD Munters Drive * Motor & Controller Replacement kit for AX and VX Fans *Patents Pending MD Munters Drive Motor and Controller Replacement Kit Models: MD-AX51xx-Hx MD-VX36xx-Hx MD-VX48xx-Hx

More information

Model OEM-2 INSTRUCTIONS FOR USE

Model OEM-2 INSTRUCTIONS FOR USE ECO SENSORS, INC. 3-03.2 OEM OZONE CONTROLLER Model OEM-2 INSTRUCTIONS FOR USE GENERAL The model OEM-2 is a system to control ozone generators and alarms based on an adjustable ozone concentration set

More information

maxium smart Vac The innovative system for efficient smoke evacuation

maxium smart Vac The innovative system for efficient smoke evacuation maxium smart Vac The innovative system for efficient smoke evacuation www.klsmartin.com Enhanced visibility and maximum safety at the operating site surgical smoke is a trouble of the past! maxium smart

More information

Beacon 110 Gas Monitor Operator s Manual

Beacon 110 Gas Monitor Operator s Manual Beacon 110 Gas Monitor Operator s Manual Part Number: 71-0110RK Revision: H Released: 12/5/17 RKI Instruments, Inc. www.rkiinstruments.com Product Warranty RKI Instruments, Inc., warrants gas alarm equipment

More information

MR4PMUHV Electronic. Temperature/Defrost Control with Relay Pack

MR4PMUHV Electronic. Temperature/Defrost Control with Relay Pack Master Catalog 125 Temperature Controls Section A Product/Technical Bulletin Issue Date 1098 MR4PMUHV Electronic Temperature/Defrost Control with Relay Pack The MR series temperature controls are designed

More information

METEOROLOGICAL INSTRUMENTS

METEOROLOGICAL INSTRUMENTS METEOROLOGICAL INSTRUMENTS INSTRUCTIONS WIND TRACKER MODEL 06206 R.M. YOUNG COMPANY 2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA TEL: (231) 946-3980 FAX: (231) 946-4772 WEB: www.youngusa.com

More information

MODEL QTS-1800 SERIES WALL MOUNT DIGITAL AND ANALOG TRANSMITTER/SENSOR

MODEL QTS-1800 SERIES WALL MOUNT DIGITAL AND ANALOG TRANSMITTER/SENSOR MODEL QTS-1800 SERIES WALL MOUNT DIGITAL AND ANALOG TRANSMITTER/SENSOR INSTALLATION OPERATION AND MAINTENANCE MANUAL QUATROSENSE ENVIRONMENTAL LTD. 5935 OTTAWA STREET, PO BOX 749 RICHMOND, ONTARIO CANADA

More information

Rev D 3/20/17

Rev D 3/20/17 Rev D 3/20/17 Installation and Operating Instructions ADAEZ Wireless Interface Module (WIM) Note Changes or modifications not expressly approved by the party responsible for compliance could void the user

More information

Dual Point General Purpose Heat Trace Control TRACON MODEL GPT 230 Installation and Operation Manual

Dual Point General Purpose Heat Trace Control TRACON MODEL GPT 230 Installation and Operation Manual We manage heat MANUAL Dual Point General Purpose Heat Trace Control TRACON MODEL GPT 230 Installation and Operation Manual 1850 N Sheridan Street South Bend, Indiana 46628 (574) 233-1202 or (800) 234-4239

More information

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR QAS-10000 SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR INSTALLATION, OPERATION AND MAINTENANCE MANUAL QUATROSENSE ENVIRONMENTAL LTD. 5935 OTTAWA STREET, PO BOX 749, RICHMOND, ONTARIO, CANADA. K0A

More information

Annunciator Relay MR 11 and MRE 11 MR 21 and MRE 21. Automation Equipment. Power Station Control Systems. Control Room Technology

Annunciator Relay MR 11 and MRE 11 MR 21 and MRE 21. Automation Equipment. Power Station Control Systems. Control Room Technology Annunciator Relay MR 11 and MRE 11 MR 21 and MRE 21 Power Station Control Systems Power Distribution Control Systems Control Room Technology Automation Equipment The products described in this brochure

More information

EN: This Datasheet is presented by the m anufacturer. Please v isit our website for pricing and availability at ore.hu.

EN: This Datasheet is presented by the m anufacturer. Please v isit our website for pricing and availability at   ore.hu. EN: This Datasheet is presented by the m anufacturer. Please v isit our website for pricing and availability at www.hest ore.hu. PE bistable - small, strong, smart Miniature PCB Relay PE bistable 1pole

More information

HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL

HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL HALOGUARD IR MULTI-POINT, COMPOUND SPECIFIC MONITOR INSTRUCTION MANUAL S E R I A L N O. M O D E L N O. T e m p. R a n g e 1 - = > 6 0 o F 2 - = 4 0-6 0 o F 3 - = < 4 0 o F - G a s T y p e 1 - R -1 1 2

More information

Dräger X-am 5000 Multi-Gas Detection

Dräger X-am 5000 Multi-Gas Detection Dräger X-am 5000 Multi-Gas Detection The Dräger X-am 5000 belongs to a generation of gas detectors, developed especially for personal monitoring applications. This 1- to 5-gas detector reliably measures

More information

TS-RA1, TS-RA2 and TS-RK Tank Overfill Alarms

TS-RA1, TS-RA2 and TS-RK Tank Overfill Alarms TS-RA1, TS-RA2 and TS-RK Tank Overfill Alarms Installation Guide franklinfueling.com 3760 Marsh Rd. Madison, WI 53718, USA Tel: +1 608 838 8786 Fax: +1 608 838 6433 Tel: USA & Canada +1 800 225 9787 Tel:

More information

T6984F,G Electronic Floating Control Thermostats

T6984F,G Electronic Floating Control Thermostats T6984F,G Electronic Floating Control Thermostats FEATURES PRODUCT DATA P+I control action provides accurate, stable room temperature control. Used with ML644 Series 60 Direct Coupled Damper Actuators.

More information

Dräger Polytron 8100 EC Detection of toxic gases and vapors

Dräger Polytron 8100 EC Detection of toxic gases and vapors Dräger Polytron 8100 EC Detection of toxic gases and vapors The Polytron 8100 EC is Dräger s top of the line explosion proof transmitter for the detection of toxic gases or oxygen It uses a high performance

More information

MR3CCUHV Temperature/Defrost Control

MR3CCUHV Temperature/Defrost Control Master Catalog 125 Temperature Controls Section A Product/Technical Bulletin Issue Date 0401 MR3CCUHV Temperature/Defrost Control The MR3CCUHV Temperature/Defrost Control is designed to control the temperature

More information

Dräger Polytron 8100 EC Detection of toxic gases and oxygen

Dräger Polytron 8100 EC Detection of toxic gases and oxygen Dräger Polytron 8100 EC Detection of toxic gases and oxygen The Polytron 8100 EC is Dräger s top of the line explosion-proof transmitter for the detection of toxic gases or oxygen It uses a high performance

More information

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR

1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR 1040 Gas Monitor INSTALLATION AND OPERATING INSTRUCTIONS AMC-1040 WITH INTEGRAL ELECTROCHEMICAL SENSOR IMPORTANT: Please read these installation and operating instructions completely and carefully before

More information

Solid Core and Split Core 0-5, 0-10 Vdc Output Current Sensors CTS-05,10; CTP-05,-10

Solid Core and Split Core 0-5, 0-10 Vdc Output Current Sensors CTS-05,10; CTP-05,-10 Solid Core and Split Core 0-5, 0-10 Vdc Output Current Sensors CTS-05,10; CTP-05,-10 DESCRIPTION FEATURES PRODUCT DATA Solid and Split Core Vdc Output Current Transmitters. Fast Response Time. Integral

More information

Dräger Polytron 8720 IR Detection of carbon dioxide

Dräger Polytron 8720 IR Detection of carbon dioxide Dräger Polytron 8720 IR Detection of carbon dioxide The Dräger Polytron 8720 IR is an advanced explosion proof transmitter for the detection of carbon dioxide in percent volume or ppm It uses a high performance

More information

PCMini52 Digital RH & Temperature Mini Probe User s Manual

PCMini52 Digital RH & Temperature Mini Probe User s Manual PCMini52 Digital RH & Temperature Mini Probe User s Manual 97300 Issue 1, March 2012 Michell Instruments Inside front cover (blank) PCMini52 2009-2012 Michell Instruments This document is the property

More information

Model D12-IR. Digital Gas Transmitter...

Model D12-IR. Digital Gas Transmitter... Model D12-IR Digital Gas Transmitter... ATI s Series D12 gas transmitter line now includes a versatile Infrared system that can be configured for LEL or select toxic gases. The D12-IR utilizes a compact

More information

Lab-Quality Results for Point-of-Care Testing

Lab-Quality Results for Point-of-Care Testing Lab-Quality Results for Point-of-Care Testing ph pco 2 po 2 Na + K + Ca ++ PT APTT ACT ACT-LR Hematocrit Immediate Care Diagnostics IL GEM Premier TM 3000 Lab-Quality Results for Point-of-Care Testing

More information

Dräger Polytron 5720 IR Toxic Gas Detection

Dräger Polytron 5720 IR Toxic Gas Detection Dräger Polytron 5720 IR Toxic Gas Detection The Dräger Polytron 5720 IR is a cost-effective explosion proof transmitter for the detection of carbon dioxide in volume percentage or ppm It uses a high-performance

More information

BA2702-P Linear Actuator

BA2702-P Linear Actuator Instruction Manual Linear Actuator with Potentiometer BA2702-P Linear Actuator Model: BA2702-P QM1246r0 1 BA2702-P Linear Actuator Instructions for Use and Maintenance Thank You: Thank you for purchasing

More information

Analog Room Pressure Monitor RPC Series

Analog Room Pressure Monitor RPC Series Description The Room Pressure Monitor is used to measure differential pressure in the range of 0.125 to 1"wc or 30 to 250 Pa. It combines precision high sensitivity silicon sensing capabilities and the

More information

DESCRIPTION FEATURES WARNINGS APPLICATIONS

DESCRIPTION FEATURES WARNINGS APPLICATIONS FEATURES Long mechanical life Single Pole Double Throw Power switching loads to Amp. Wide range of actuators and terminal design High precision with good repeatability High vibration and shock resistance

More information

Alarm Tone Generator Model AG17

Alarm Tone Generator Model AG17 Alarm Tone Generator Installation & Operation P005089 Rev. C 150930 9/30/2015 12:25 PM Ph: 403.258.3100 \ email:info@guardiantelecom.com \ www.guardiantelecom.com Table of Contents Package Contents...

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

ELECTRONIC FLAME SUPERVISION MULTI-BURNER

ELECTRONIC FLAME SUPERVISION MULTI-BURNER MULTI-BURNER MODEL: SENS-A-FLAME II Revision: 0 7112 DESCRIPTION Model 7112 Sens-A-Flame II is a second generation solid-state programmable combustion safeguard for supervising multiple burner ovens, furnaces,

More information

Operating Manual. Gasgard 100 Control System. Order No.: /00. MSAsafety.com

Operating Manual. Gasgard 100 Control System. Order No.: /00. MSAsafety.com Operating Manual Gasgard 100 Control System Order No.: 10153908/00 MSAsafety.com Warning! THIS MANUAL MT BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR ING OR SERVICING

More information

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system

INSTRUCTION MANUAL T119 1MN0110 REV. 0. operates with ISO9001 certified quality system INSTRUCTION MANUAL 1MN0110 REV. 0 operates with ISO9001 certified quality system TECSYSTEM S.r.l. 20094 Corsico (MI) Tel.: +39-024581861 Fax: +39-0248600783 http://www.tecsystem.it R. 1.1 25/08/16 ENGLISH

More information

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual

Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual We manage heat MANUAL Single Point Freeze Protection Heat Trace Control TRACON MODEL FPT 130 Installation and Operation Manual 1850 N Sheridan Street South Bend, Indiana 46628 (574) 233-1202 or (800) 234-4239

More information

Dräger QuadGard Control System

Dräger QuadGard Control System Dräger QuadGard Control System The Dräger QuadGard is a standalone, self contained control system for the detection of Toxic, Oxygen and Ex hazards The system can be configured for up to four independent

More information

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL Installation and Operation Manual 2 GG-2 Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified

More information

MS4PMUHVT Multi-Stage Electronic Temperature Control with Relay Pack

MS4PMUHVT Multi-Stage Electronic Temperature Control with Relay Pack Product/Technical MS4PMUHVT Issue Date 09/30/02 MS4PMUHVT Multi-Stage Electronic Control with Relay Pack The MS4PMUHVT Multi-Stage Electronic Control is intended for multi-stage temperature control applications.

More information

RAM3 Remote Alarm Module

RAM3 Remote Alarm Module RAM3 Remote Alarm Module INSTRUCTIONS Installation and Operation of the AMC-RAM3 with AMC Monitors IMPORTANT: Please read this installation and operating instructions completely and carefully before starting.

More information

GasScanner 1C. Single Channel Monitor. Operator s Manual. MINT-0278-XX Rev A

GasScanner 1C. Single Channel Monitor. Operator s Manual. MINT-0278-XX Rev A GasScanner 1C Single Channel Monitor Operator s Manual MINT-0278-XX Rev A Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship,

More information

GG-EM ENTRANCE MONITOR. Installation and Operation Manual

GG-EM ENTRANCE MONITOR. Installation and Operation Manual GG-EM ENTRANCE MONITOR Installation and Operation Manual 2 GG-EM Warning Use this product only in the manner described in this manual. If the equipment is used in a manner not specified by Calibration

More information