Sensitivity to Thoron on Passive Radon Detectors

Size: px
Start display at page:

Download "Sensitivity to Thoron on Passive Radon Detectors"

Transcription

1 Sensitivity to Thoron on Passive Radon Detectors S. Tokonami 1, H. Yonehara 1, T. Sanada 2, M. Yang 3, M. Furukawa 1 and Y. Yamada 1 1 National Institute of Radiological Sciences, Chiba , Japan 2 Japan Chemical Analysis Center, Chiba , Japan 3 Research Institute of Uranium Mining, Hunan, China INTRODUCTION Although a lot of studies on radon have been done for a long time, there are few reports on thoron. It was considered that the presence of thoron could be negligible because of its own quantity compared to that of radon. However, recent studies have shown high thoron levels in our living environment (1-4). In particular, high thoron concentrations were occasionally observed in some areas, Japan (1,2). This fact made it clear that some of passive radon detectors were sensitive to the presence of thoron. They are commonly used for indoor radon surveys. It is possible that such detectors will give false values unless they are placed properly. Therefore, it is important to understand the detector response to thoron before practical use. A compact thoron chamber system was set up for the purpose. In this study, the followings are shown: configuration of the thoron chamber and sensitivity to thoron of several passive radon detectors. MATERIALS AND METHODS The thoron chamber system consists of two parts: an exposure chamber and a thoron gas generator. They are connected through an inline filter holder, a flow meter and an external pump. Figure 1 illustrates a sectional view of the thoron chamber system when thoron exposure tests are carried out for passive radon detectors. The exposure chamber is a 150 liters cylindrical vessel (565 mm diameter and 600 mm height), made of stainless steel. Four metal tubes are attached to the lid of the chamber. They are used to supply/exhaust thoron/radon gas and to take air samples. A fan is mounted on the lid to make the internal concentration uniform. The generator is a 10 liters stainless steel cylindrical vessel (240 mm diameter and 240 mm height). The vessel is filled with thorium-rich ceramics as the solid source. The ceramics were crushed into small pieces so as to promote thoron exhalation. After connecting the exposure chamber and generator, thoron gas circulates through the system with the external pump. The thoron concentration in the exposure chamber depends on the flow rate of the circulation. The flow rate was set at around 20 L/min in this system. Thoron and radon concentrations are measured with scintillation cells after taking an air sample through the metal tube. The measurement technique is based on two measurements using the difference of the half life between radon and thoron. Details on the technique are described in another paper (5). 1

2 sampling tube fan to scintillation cell pump flow meter 20 L/min O-ring inline filter holder air outlet thoron + radon gas air inlet 600 mm 565 mm passive device thorium-rich ceramics exposure chamber Figure 1. metal wire screen thoron gas generator A sectional view of the thoron chamber system when thoron exposure tests are carried out for passive radon detectors. RESULTS AND DISCUSSION In order to check the stability of the thoron chamber system, the radon concentrations in the exposure chamber were measured after radon gas was injected. Air samples were occasionally taken through the sampling tube. They were measured with the standard technique using the scintillation cell. The radon concentrations measured follow the half life of radon as shown in Figure 2. Since the exposure condition seems to be fairly stable, the configuration as the exposure chamber is satisfactory. This chamber system can also be used for the radon exposure test. As another test, thoron and radon concentrations in the exposure chamber were measured after connecting the exposure chamber with the thoron gas generator through the external pump, as shown in Figure 1. Air samples were occasionally taken while the thoron gas circulated through the system at the flow rate of 19.5 L/min. Their concentrations were shown in Figure 3. This performance test lasted 4 days. The average thoron and radon concentrations during the period were estimated to be 2530 Bq/m 3 and 230 Bq/m 3, respectively. In order to confirm the reliability of the proposed technique, the radon concentrations using the proposed technique were compared with those using the standard technique. The result was shown in Figure 4. There was a good agreement between the two. 2

3 10 6 Rn conc. (Bq/m 3 ) 10 5 Actual Rn Estimated Rn Elapsed time (day) Figure 2. Time variation of the radon concentration in the exposure chamber after radon gas injection. In the present study, the following passive radon detectors were actually examined: German monitors (CR-39 installed) (6), RADTRAK, R-T discriminative dosimeters (7) and electrostatic integrating radon monitors (EIRM) (8). The result of the conversion factor on the radon and thoron concentration is shown in Table 1. The radon conversion factors on the RADTRAK and R-T discriminative dosimeter were determined with the authorized facilities in advance. In terms of the RADTRAK, the conversion factors have changed compared to the previous result (9). The German monitor and the RADTRAK were highly sensitive to thoron. The R-T discriminative dosimeter consists of two radon detectors with two different ventilation rates. Since the upper hemisphere has a high ventilation rate, thoron concentrations are mainly measured. The lower hemisphere is designed to detect radon only. However, it was found that even this lower hemisphere was affected by thoron in this exposure test. On the other hand, the EIRM had no sensitivity to thoron. Although most of passive radon detectors are designed to minimize the entry of thoron, it can be realized that they are sensitive to the presence of thoron except one device in the present study. Table 1 Conversion factors on radon and thoron concentrations. Measuring device Radon Thoron German monitor (CR-39) x x 10-4 RADTRAK 6.74 x x 10-4 Radon-Thoron discriminative dosimeter (upper hemisphere) 2.40 x x 10-4 Radon-Thoron discriminative dosimeter (lower hemisphere) 2.82 x x 10-4 Electrostatic integrating radon-222 monitor x 10-4 No sensitivity unit: tracks/mm 2 /(Bq/m 3 -d) 3

4 Concentration (Bq/m 3 ) 10 3 Thoron Radon Elapsed time (day) Figure 3. Thoron and radon concentrations while the thoron gas circulates through the chamber system. 3 ) Rn conc. with standard tech. (Bq/m Rn conc. with prompt tech. (Bq/m 3 ) Figure 4. Comparison of the radon concentration with two different techniques. CONCLUSION In order to check the sensitivity to thoron on passive radon detectors, the thoron chamber system has been set up. Subsequently, performance tests of the chamber system were carried out so as to confirm its stability and reliability. Although some passive radon detectors were examined, most of them were sensitive to thoron. Although the presence of thoron can be negligible in the environment, it is important to check the sensitivity to thoron with the proposed test before practical use. 4

5 ACKNOWLEDGEMENTS The authors thank Ms. Ikumi Ishikawa, National Institute of Radiological Sciences, for her technical assistance on the exposure test. REFERENCES 1. M. Doi and S. Kobayashi, Characterization of Japanese Wooden Houses with Enhanced Radon and Thoron Concentrations. Health Phys., 66, (1994). 2. J. Ma, H. Yonehara, T. Aoyama, M. Doi, et al., Influence of Air Flow on the Behavior of Thoron and its Progeny in a Traditional Japanese House. Health Phys., 72, (1997). 3. W.H. Chung, S. Tokonami and M. Furukawa, Preliminary Survey on Radon and Thoron Concentrations in Korea. Radiat. Prot. Dosim. 80, (1998). 4. L.F. Toussaint, S. Tokonami, M. Doi, et al., The Measurement of Thoron Concentrations in Australia Using the Japanese Passive R-T Dosimeter. In: Proceedings of the 7 th Tohwa University International Symposium, , Fukuoka (1998). 5. S. Tokonami, T. Sanada, H. Yonehara, et al., Detection response to the presence of thoron on passive radon detectors. Radioisotopes, 49, 1-8 (2000) [in Japanese]. 6. M. Urban and E. Piesch, Low level environmental radon dosimetry with a passive track etch detector device. Radiat. Prot. Dosim., 1, (1981). 7. M. Doi and S. Kobayashi, The passive radon-thoron discriminative dosimeter for practical use. Hoken Butsuri, 29, (1994). 8. T. Iida, Y. Ikebe, T. Hattori, et al., An electrostatic integrating 222 Rn monitor with cellulose nitrate film for environmental monitoring. Health Phys., 54, (1988). 9. M.D. Pearson and R.R. Spangler, Calibration of alpha-track monitors for measurement of thoron ( 220 Rn). Health Phys., 60, (1991). 5

Radon and Thoron Measurements at Special Underground Circumstances

Radon and Thoron Measurements at Special Underground Circumstances Radon and Thoron Measurements at Special Underground Circumstances Tibor Kovács a*, Norbert Kávási b, Csaba Németh c, János Somlai a, Tamás Vigh d, Gábor Szeiler a, and Shinji Tokonami e, Hiroyuki Takahashi

More information

Development of high sensitivity radon detectors

Development of high sensitivity radon detectors Nuclear Instruments and Methods in Physics Research A 421 (1999) 334 341 Development of high sensitivity radon detectors Y. Takeuchi *, K. Okumura, T. Kajita, S. Tasaka, H. Hori, M. Nemoto, H. Okazawa

More information

Design and Fabrication of New Radon Chamber for Radon Calibration Factor of Measurement

Design and Fabrication of New Radon Chamber for Radon Calibration Factor of Measurement International Journal of Scientific & Engineering Research, Volume 3, Issue 10, October-2012 1 Design and Fabrication of New Radon Chamber for Radon Calibration Factor of Measurement Murtadha S. Al-Nafiey,*

More information

MetroRADON project and its potential impact on mitigation practices Dobromir Pressyanov Sofia University St. Kliment Ohridski Bulgaria

MetroRADON project and its potential impact on mitigation practices Dobromir Pressyanov Sofia University St. Kliment Ohridski Bulgaria MetroRADON project and its potential impact on mitigation practices Dobromir Pressyanov Sofia University St. Kliment Ohridski Bulgaria Wolfgang Ringer Austrian Agency for Health and Food Safety (AGES)

More information

A study of Indoor Radon / Thoron Levels in Some Dwellings by using Solid State Nuclear Track Detectors

A study of Indoor Radon / Thoron Levels in Some Dwellings by using Solid State Nuclear Track Detectors International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 6, Number 2 (2010), pp. 157 164 Research India Publications http://www.ripublication.com/ijpap.htm A study of Indoor Radon / Thoron

More information

ENVIRONMENTAL RADON MONITORING IN DWELLINGS NEAR THE RADIOACTIVE SITES

ENVIRONMENTAL RADON MONITORING IN DWELLINGS NEAR THE RADIOACTIVE SITES ENVIRONMENTAL RADON MONITORING IN DWELLINGS NEAR THE RADIOACTIVE SITES B.S BAJWA and H.S. VIRK Deptt. of Physics, Guru Nanak Dev University Amritsar-143005, India. In view of the of the fact that radon

More information

CURRENT STATE OF THE ART IN MEASURING ENVIRONMENTAL RADON. Abstract

CURRENT STATE OF THE ART IN MEASURING ENVIRONMENTAL RADON. Abstract CURRENT STATE OF THE ART IN MEASURING ENVIRONMENTAL RADON A.C. George and N. Bredhoff Radon Testing Corporation of America Elmsford, New York Abstract According to the US EPA, radon is the leading cause

More information

A new passive integrating detector for measurement of individual radon exposure at working places.

A new passive integrating detector for measurement of individual radon exposure at working places. A new passive integrating detector for measurement of individual radon exposure at working places. G. Sciocchetti, G. Cotellessa, E. Soldano, M. Pagliari and A. Ceccatelli (*) ENEA-ION-IRP, Centro Ricerche

More information

INTERNATIONAL INTERCOMPARISON EXERCISE ON NATURAL RADIATION MEASUREMENTS UNDER FIELD CONDITIONS (IFC11) *

INTERNATIONAL INTERCOMPARISON EXERCISE ON NATURAL RADIATION MEASUREMENTS UNDER FIELD CONDITIONS (IFC11) * INTERNATIONAL INTERCOMPARISON EXERCISE ON NATURAL RADIATION MEASUREMENTS UNDER FIELD CONDITIONS (IFC11) * BOTOND PAPP 1, ALEXANDRA CUCOŞ (DINU) 1*, MIRCEA MOLDOVAN 1, ROBERT BEGY 1, TIBERIU DICU 1, DAN

More information

RADON CONCENTRATIONS AFFECTED BY DIFFERENT FACTORS IN TWO OFFICE BUILDINGS. Raimo Halonen, Pirjo Korhonen, Pentti Kalliokoski and Helmi Kokotti

RADON CONCENTRATIONS AFFECTED BY DIFFERENT FACTORS IN TWO OFFICE BUILDINGS. Raimo Halonen, Pirjo Korhonen, Pentti Kalliokoski and Helmi Kokotti Radon in the Living Environment, 25 RADON CONCENTRATIONS AFFECTED BY DIFFERENT FACTORS IN TWO OFFICE BUILDINGS Raimo Halonen, Pirjo Korhonen, Pentti Kalliokoski and Helmi Kokotti Department of Environmental

More information

Detectors used in the intercomparison

Detectors used in the intercomparison Detectors used in the intercomparison The second radon-in-field international intercomparison for passive measurement devices: dwellings and workplaces 21 22 september 2017 Campus Bovisa Politecnico di

More information

Quality and Technology - Made in Germany -

Quality and Technology - Made in Germany - Quality and Technology - Made in Germany - TRACERLAB GmbH Aachener Str. 1354 50859, Koeln, Germany Tel.: +49 (0) 2234-942397 Fax.: +49 (0) 2234-942398 E-mail: asktracerlab@tracerlab.com Internet: www.tracerlab.com

More information

Sergey Kiselev. SRC Federal Medical Biophysical Center

Sergey Kiselev. SRC Federal Medical Biophysical Center Sergey Kiselev SRC Federal Medical Biophysical Center Uranium Thorium 1940s -uranium/thorium geological maps. Epidemiological studies U-mine cohort studies Place Zheltoreche nsk Uchkuduk Country Type of

More information

Outdoor radon concentration in China

Outdoor radon concentration in China NUKLEONIKA 2016;61(3):373 378 doi: 10.1515/nuka-2016-0062 ORIGINAL PAPER Outdoor radon concentration in China Qifan Wu, Ziqiang Pan, Senlin Liu, Chunhong Wang Abstract. A nationwide survey was conducted

More information

A new personal dosemeter for the individual monitoring of exposure to radon gas

A new personal dosemeter for the individual monitoring of exposure to radon gas A new personal dosemeter for the individual monitoring of exposure to radon gas Rosabianca Trevisi a*, Paolo Orlando b, Giuseppe Arcovito b, Francesco Cardellini c, Elisabetta Fonti b, Claudia Orlando

More information

RADON SURVEY IN KALAMATA (GREECE) Medicine, London SW7 2AZ U.K. Greece. Republic

RADON SURVEY IN KALAMATA (GREECE) Medicine, London SW7 2AZ U.K. Greece. Republic Radon in the Living Environment, 132 RADON SURVEY IN KALAMATA (GREECE) A. Geranios 1, M. Kakoulidou 1, Ph. Mavroidi 2, M. Moschou 3, S. Fischer 4, I. Burian 5 and J. Holecek 5 1 Nuclear and Particle Physics

More information

Radon Measurements in Argentina

Radon Measurements in Argentina AUTORIDAD REGULATORIA NUCLEAR (ARN) Radon Measurements in Argentina Juan Pablo Bonetto a, Analia Canoba, Fabio López Gerencia de Apoyo Científico técnico ARN a E-mail: jbonetto@arn.gob.ar Introduction

More information

THE RELIABILITY OF RADON REDUCTION TECHNIQUES. CB Howarth

THE RELIABILITY OF RADON REDUCTION TECHNIQUES. CB Howarth Radon in the Living Environment, 052 THE RELIABILITY OF RADON REDUCTION TECHNIQUES CB Howarth National Radiological Protection Board Chilton, Didcot, Oxon, OX11 0RQ, UK Tel: +44 1235 822796, Fax: +44 1235

More information

INTERNATIONAL JOURNAL OF RESEARCH SCIENCE & MANAGEMENT

INTERNATIONAL JOURNAL OF RESEARCH SCIENCE & MANAGEMENT [Kim* et al., 4(2): February, 17] ISSN: 234-97 PERFORMANCE IMPROVEMENT OF RADON COUNTERS 1 Yong-Cheol Kim, *2 Gyu-Sik Kim 1,2 Professor, Dept. of Electrical and Computer Eng., University of Seoul, Korea

More information

Radon in thermal waters and radon risk in chosen thermal water spas in V4 countries - preliminary results

Radon in thermal waters and radon risk in chosen thermal water spas in V4 countries - preliminary results Radon in thermal waters and radon risk in chosen thermal water spas in V countries - preliminary results Karol Holý, Pavol Blahušiak, Dominik Grządziel, Tibor Kovács, Krzysztof Kozak, Jadwiga Mazur, Monika

More information

Radiation Monitoring Equipment

Radiation Monitoring Equipment Radiation Monitoring Equipment Sharon Paulka Regional Training Course RAF 3007-1 Namibia, 17-21 August 2009 International Atomic Energy Agency Types of Monitoring Both Occupational & Environmental Require

More information

Guide for Radon Measurements in Residential Dwellings (Homes)

Guide for Radon Measurements in Residential Dwellings (Homes) Guide for Radon Measurements in Residential Dwellings (Homes) Health Canada is the federal department responsible for helping Canadians maintain and improve their health. We assess the safety of drugs

More information

Paul Kotrappa, Ph.D. and Carolyn Allen. Rad Elec Inc., 5714-C Industry Lane, Frederick, MD USA

Paul Kotrappa, Ph.D. and Carolyn Allen. Rad Elec Inc., 5714-C Industry Lane, Frederick, MD USA 177 REVIEW OF QA\QC ASPECTS OF ELECTRET ION CHAMBERS- MANUFACTURING PRACTICES AND PERFORMANCE IN WORLDWIDE INTER-COMPARISON EXERCISES IN THE PAST EIGHT YEARS Paul Kotrappa, Ph.D. and Carolyn Allen Rad

More information

Spatial variation of radon concentration in a room ZHANG ZHENGGUO*, XIAO DETAO**, ZHANG LIANG*, LI CHUNXIU*,

Spatial variation of radon concentration in a room ZHANG ZHENGGUO*, XIAO DETAO**, ZHANG LIANG*, LI CHUNXIU*, note technique Spatial variation of radon concentration in a room ZHANG ZHENGGUO*, XIAO DETAO**, ZHANG LIANG*, LI CHUNXIU*, (Manuscrit reçu le 6 octobre 1992) RÉSUMÉ ABSTRACT On a mesuré la concentration

More information

Assessment of Radon-222 Concentrations in Buildinps, Building Materials, Water and Soil in Jordan

Assessment of Radon-222 Concentrations in Buildinps, Building Materials, Water and Soil in Jordan Assessment of Radon-222 Concentrations in Buildinps, Building Materials, Water and Soil in Jordan Mahmoud Kullab Physics Department, Yarmouk University, Irbid 2 1 1-63, Jordan E-mail: mkullab@yu.edu.jo

More information

BRUNILDA DACI INSTITUTE OF APPLIED NUCLEAR PHYSICS TIRANA-ALBANIA. KOZETA BODE Responsible person for Radon Measurement in Albania

BRUNILDA DACI INSTITUTE OF APPLIED NUCLEAR PHYSICS TIRANA-ALBANIA. KOZETA BODE Responsible person for Radon Measurement in Albania BRUNILDA DACI INSTITUTE OF APPLIED NUCLEAR PHYSICS TIRANA-ALBANIA KOZETA BODE Responsible person for Radon Measurement in Albania Tallinn, Estonia, 23-27 May 2016 1 CONTENT OF THE PRESENTATION INTRODUCTION

More information

Measurement Devices. Topic 5 - Audio 56

Measurement Devices. Topic 5 - Audio 56 Topic 5 - Audio 56 Other Classifications Passive devices: Air diffuses in No power required for sampling Active devices: Requires power Pumps air through filter Key Points - Devices What lab sees: Different

More information

Radon Survey Summary Report

Radon Survey Summary Report GWASANAETHAU IECHYD A DIOGELWCH - HEALTH AND SAFETY SERVICES ADRODDIAD REPORT Radon Survey Summary Report September 2010 Gwasanaethau Iechyd a Diogelwch Health and Safety Services, Penbre, Lon y Coleg.

More information

Dynamic Radon Detection Over Measurement

Dynamic Radon Detection Over Measurement Dynamic Radon Detection Over Measurement Nuclear Engineer, specialised in detection of nuclear particles, instrumentations and radioprotection. Person in charge of radioprotection. Worked for nuclear industries

More information

EFFECTS OF SMALL PRESSURE DIFFERENCES BETWEEN THE STRUCTURE AND SURROUNDINGS ON RADON ENTRY

EFFECTS OF SMALL PRESSURE DIFFERENCES BETWEEN THE STRUCTURE AND SURROUNDINGS ON RADON ENTRY EFFECTS OF SMALL PRESSURE DIFFERENCES BETWEEN THE STRUCTURE AND SURROUNDINGS ON RADON ENTRY 1,2 Akis M.C., 1 Stadtmann H., 2 Kindl P. 1 ARC Seibersdorf research GmbH, Division of Health Physics /Radiation

More information

Abstract. Introduction/Background. New Radon guideline for Canada

Abstract. Introduction/Background. New Radon guideline for Canada AN UPDATE ON THE NEW CANADIAN RADON GUIDELINE AND ITS IMPLEMENTATION Naureen M. Rahman, Renato Falcomer, Jing Chen and Deborah Moir Radiation Protection Bureau, Health Canada, Ottawa, Ontario K1A 1C1,

More information

THE NATIONAL RADON SURVEY IN IRELAND. S.G. Fennell, G.M. Mackin, J.S. Madden and A.T. McGarry

THE NATIONAL RADON SURVEY IN IRELAND. S.G. Fennell, G.M. Mackin, J.S. Madden and A.T. McGarry Radon in the Living Environment, 136 THE NATIONAL RADON SURVEY IN IRELAND S.G. Fennell, G.M. Mackin, J.S. Madden and A.T. McGarry Radiological Protection Institute of Ireland, 3 Clonskeagh Square, Clonskeagh

More information

Electronic Personal Dosemeter

Electronic Personal Dosemeter Electronic Personal Dosemeter Tomoya Nunomia Hideshi Yamauchi Tetsuo Shibata 1. Introduction Since developing an electronic personal dosemeter that uses a semiconductor detector in 1983, Fuji Electric

More information

Use of MCNPX for Alpha Spectrometry Simulations of a Continuous Air Monitor

Use of MCNPX for Alpha Spectrometry Simulations of a Continuous Air Monitor Page 1 DOE/NV/XXXXX--XX Use of MCNPX for Alpha Spectrometry Simulations of a Continuous Air Monitor Robert B. Hayes, Ph.D., CHP, PE Senior Scientist Remote Sensing Laboratory PO Box 98521, Mail Stop RSL-47

More information

COMPACT RADIOACTIVE AEROSOL MONITORING DEVICE FOR EARLY WARNING NETWORKS

COMPACT RADIOACTIVE AEROSOL MONITORING DEVICE FOR EARLY WARNING NETWORKS COMPACT RADIOACTIVE AEROSOL MONITORING DEVICE FOR EARLY WARNING NETWORKS Denis GLAVIČ-CINDRO Toni PETROVIČ Drago BRODNIK Matjaž VENCELJ Jožef Stefan Institute Ljubljana, Slovenia Steven James BELL Lindsey

More information

Via Umberto Giordano, Padova Tel Fax VERSIONS ACCURACY

Via Umberto Giordano, Padova Tel Fax VERSIONS ACCURACY DOWNLOADS DESCRIPTION BATTERY LIFE Approximately 1.5 years of continuous monitoring 3 replaceable AAA batteries DISPLAY 1-day, 7-day, and long-term average radon level FUNCTIONS Reset button Mode button

More information

- For Continuous Radon Monitoring (CRM) Systems - Ultra Low Power Requirement

- For Continuous Radon Monitoring (CRM) Systems - Ultra Low Power Requirement RN53 Radon Detector RN53 - For Continuous Radon Monitoring (CRM) Systems - Ultra Low Power Requirement Description The RN53 Radon Detector operates on the method of electrostatic collecting of the radon

More information

HISTORY OF SOIL-GAS RADON CONCENTRATION MEASUREMENTS IN THE CZECH REPUBLIC

HISTORY OF SOIL-GAS RADON CONCENTRATION MEASUREMENTS IN THE CZECH REPUBLIC HISTORY OF SOIL-GAS RADON CONCENTRATION MEASUREMENTS IN THE CZECH REPUBLIC Martin Neznal & Matěj Neznal RADON v.o.s. Novákových 6, 180 00 Praha 8, Czech Republic radon@comp.cz www.radon.eu LATIN AMERICAN

More information

Radon Reference Facility at Bowser-Morner, Inc. AMUG Meeting Las Vegas May 4, 2010 Phillip H. Jenkins, PhD, CHP

Radon Reference Facility at Bowser-Morner, Inc. AMUG Meeting Las Vegas May 4, 2010 Phillip H. Jenkins, PhD, CHP Radon Reference Facility at Bowser-Morner, Inc. AMUG Meeting Las Vegas May 4, 2010 Phillip H. Jenkins, PhD, CHP Radon Reference Facility What We Do Calibration of Rn and RDP measuring devices Spiking of

More information

MEDICAL GEOLOGY/GEOCHEMISTRY

MEDICAL GEOLOGY/GEOCHEMISTRY MEDICAL GEOLOGY/GEOCHEMISTRY PILLALAMARRI ILA Earth Atmospheric & Planetary Sciences Neutron Activation Analysis Laboratory Massachusetts Institute of Technology Cambridge, MA 02139 IAP 2006: 12.091 Credit

More information

PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR

PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR Hiroshi Miyazaki 1, Hiromasa Tanjyoh 2, Kengo Suzuki 3, Tomoaki Iwami 4 1 New cosmos electric, co., ltd., Osaka, 543-0036,

More information

Radon in (Greek) workplaces GREECE

Radon in (Greek) workplaces GREECE Radon in (Greek) workplaces GREECE Maria Kolovou Department of Environmental Radioactivity Control www.gaec.gr Measurement of radon Radon measurements are performed according to the terms of ISO 11665

More information

SEASONAL VARIATION OF RADON CONCENTRATION - MEASUREMENTS IN SOME EUROPEAN COUNTRIES

SEASONAL VARIATION OF RADON CONCENTRATION - MEASUREMENTS IN SOME EUROPEAN COUNTRIES Radon in the Living Environment, 169 SEASONAL VARIATION OF RADON CONCENTRATION - MEASUREMENTS IN SOME EUROPEAN COUNTRIES Gilbert Jönsson a, Carmen Baixeras b, Lluis Font b, Diego Albarracin b, Giuseppe

More information

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points Topic 4 - Audio 46 Key Points : Measures taken to increase confidence in results. Quality Assurance: Program that assures that measures are being done and in a timely manner. Precision: How close measurements

More information

RADON IN SCHOOL WELL WATER: CASE STUDIES AND MITIGATION IMPLICATIONS

RADON IN SCHOOL WELL WATER: CASE STUDIES AND MITIGATION IMPLICATIONS RADON IN SCHOOL WELL WATER: CASE STUDIES AND MITIGATION IMPLICATIONS Alan J. Siniscalchi, Zygmunt F. Dembek, Sarah J. Tibbetts, Xaviel Soto, and Amjad Mahmood State of Connecticut Department of Public

More information

ASSESSEMENT OF INDOOR RADON, THORON AND THEIR PROGENY IN DWELLINGS OF BAREILLY CITY OF NORTHERN INDIA USING TRACK ETCH DETECTORS

ASSESSEMENT OF INDOOR RADON, THORON AND THEIR PROGENY IN DWELLINGS OF BAREILLY CITY OF NORTHERN INDIA USING TRACK ETCH DETECTORS ASSESSEMENT OF INDOOR RADON, THORON AND THEIR PROGENY IN DWELLINGS OF BAREILLY CITY OF NORTHERN INDIA USING TRACK ETCH DETECTORS DEEPAK VERMA *, M. SHAKIR KHAN Department of Applied Physics, Z. H. College

More information

What you need to know

What you need to know What you need to know What RADON is The health effects How to test your home How to reduce your risk Protect your health. Test your home. www.regionofwaterloo.ca/radon RADON information What is radon?

More information

adsw worth Center, New York State Department of Health, Albany, NY 12201

adsw worth Center, New York State Department of Health, Albany, NY 12201 American Association of Radon Scientists and Technologists, Inc. RADON IN NEW YORK STATE'S SCHOOLS Kitto, Michael ly2, Jerry ~ollins~ and Adda ~alamc-~lfie~ adsw worth Center, New York State Department

More information

Měření difúze, ultranízké koncentrace a emanace radonu pro podzemní laboratoř

Měření difúze, ultranízké koncentrace a emanace radonu pro podzemní laboratoř Institute of Experimental and Applied Physics CTU in Prague Měření difúze, ultranízké koncentrace a emanace radonu pro podzemní laboratoř Fadahat Mamedov, Ivan Štekl, Karel Smolek 1 Outline Introduction

More information

A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE

A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE A STUDY ON THE BEHAVIOR OF STEAM CONDENSATION IN U-SHAPED HEAT TUBE Satoru Ito 1, Keisuke Tsukada 1, Nobuyoshi Tsuzuki 2, Takao Ishizuka 3 and Hiroshige Kikura 3 1 Department of nuclear Engineering, Graduate

More information

Feasibility study of the passive solar room dehumidifying system using the sorption property of a wooden attic space through field measurement

Feasibility study of the passive solar room dehumidifying system using the sorption property of a wooden attic space through field measurement International Conference Passive and Low Energy Cooling 651 Feasibility study of the passive solar room dehumidifying system using the sorption property of a wooden attic space through field measurement

More information

AC, AT, LS Measurement Devices

AC, AT, LS Measurement Devices Group Designations for Measurement Devices The following is a list of devices approved by the NRSB. These devices were either approved by the EPA Radon Proficiency Program prior to its closure in 1998

More information

Introduction...1 Specifications... 1 Limitations... 1

Introduction...1 Specifications... 1 Limitations... 1 Table of Contents Introduction...1 Specifications... 1 Limitations... 1 New Algorithms...2 Regions of Interest... 3 New Algorithm ROIs... 3 Calibration... 5 Update Interval... 5 K Factor... 5 Subscripts...

More information

their original bags during the test period determine what background reading they might accumulate. The actual

their original bags during the test period determine what background reading they might accumulate. The actual AARST ANNUAL RADON CONFERENCE OCTOBER 4,5,6 1990 COMPARISON OF LONG TERM TESTING DEVICES INDOOR AND OUTDOOR Bill Brodhead WPB Enterprises, Inc. 2844 Slifer Valley Rd. Riegelsville, Pa., 18077 610-346-8004,

More information

Situation on Radon Assessment in Latvia

Situation on Radon Assessment in Latvia Situation on Radon Assessment in Latvia Žanna Martinsone Institute of Occupational Safety and Environmental Health, Riga Stradins University, LATVIA e-mail: Zanna.Martinsone@rsu.lv Regional workshop on

More information

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories

C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories C-NRPP Quality Control and Quality Assurance Manual for Radon Sampling and Analysis conducted by Radon Measurement Professionals and Laboratories October 2018 C-NRPP/PNCR-C 1 (855) 722-6777 c-nrpp.ca Canadian

More information

Radon in elementary schools in Tunisia

Radon in elementary schools in Tunisia Article published by EDP Sciences and available at http://www.radioprotection.org or http://dx.doi.org/10.1051/radiopro/2010003 Radioprotection 2010 DOI: 10.1051/radiopro/2010003 Vol. 45, n 2, pages 209

More information

Radon Survey Final Report

Radon Survey Final Report GWASANAETHAU IECHYD A DIOGELWCH - HEALTH AND SAFETY SERVICES ADRODDIAD REPORT Radon Survey Final Report April 2011 Gwasanaethau Iechyd a Diogelwch Health and Safety Services, Penbre, Lon y Coleg. LL57

More information

Intercomparison s working group:

Intercomparison s working group: Radon passive detectors intercomparison in the field: radiometric characterisation of the sites Enrico Chiaberto - ARPA Piemonte Intercomparison s working group: Filippo Berlier, Francesco Cardellini,

More information

The development of a plastic scintillator for radiotherapy dosimetry

The development of a plastic scintillator for radiotherapy dosimetry University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2008 The development of a plastic scintillator for radiotherapy

More information

DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR

DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR Katsuhito Ito, Ryohei Yanagishima, Daisuke Inui*, Tomoya Minagawa Fuji Electric Systems Co., Ltd. Fuji-machi

More information

INTERNATIONAL INTERCOMPARISON EXERCISE OF ACTIVE RADON DEVICES AND PASSIVE DETECTORS AT THE FIRST EAST EUROPEAN RADON SYMPOSIUM (FERAS 2012)

INTERNATIONAL INTERCOMPARISON EXERCISE OF ACTIVE RADON DEVICES AND PASSIVE DETECTORS AT THE FIRST EAST EUROPEAN RADON SYMPOSIUM (FERAS 2012) INTERNATIONAL INTERCOMPARISON EXERCISE OF ACTIVE RADON DEVICES AND PASSIVE DETECTORS AT THE FIRST EAST EUROPEAN RADON SYMPOSIUM ( 2012) BOTOND PAPP *, CONSTANTIN COSMA, ALEXANDRA CUCOŞ (DINU) Environmental

More information

CR-39 SAMPLING OF INDOOR RADON IN SOUTHERN ROMANIA *

CR-39 SAMPLING OF INDOOR RADON IN SOUTHERN ROMANIA * CR-39 SAMPLING OF INDOOR RADON IN SOUTHERN ROMANIA * ANGELA VASILESCU Horia Hulubei National Institute for Nuclear Physics and Engineering (IFIN), P.O.Box MG-6, RO-077125 Bucharest-Magurele, Romania, E-mail:

More information

National Atomic Energy Commission ( NATEC ) Radiation Exposure and Laboratories General Directorate

National Atomic Energy Commission ( NATEC ) Radiation Exposure and Laboratories General Directorate National Atomic Energy Commission ( NATEC ) Radiation Exposure and Laboratories General Directorate Environmental and Food Measurements Lab & Individual monitoring lab Yemen Republic Population : 22 million

More information

Radon Testing Results. City Hall

Radon Testing Results. City Hall Radon Testing Results January April 2016 City Hall LOCATION Radon Concentration Bq/m3 Basement elevator room 152 Boiler room 67 Mail room 60 Training room 78 Electrical room 71 Server room 83 Basement

More information

International Requirements for the Control of Radon

International Requirements for the Control of Radon Tenth Irish National Radon Forum Thursday 17 January 2013 International Requirements for the Control of Radon Trevor Boal, -NSRW International Atomic Energy Agency Content Safety Standards Requirements

More information

European Commission approach with regard to radon

European Commission approach with regard to radon European Commission approach with regard to radon Åsa Wiklund National Expert in DG TREN/H.4 Radiation Protection EUROPEAN COMMISSION 1 Radon at the European Commission DG TREN DG RTD DG SANCO Joint Research

More information

EC - RADON Latest strategies and draft regulations

EC - RADON Latest strategies and draft regulations EC - RADON Latest strategies and draft regulations European Commission Radiation Protection Participation in EARST WORKSHOP 2013, International and National RADON Regulations & Strategies EC-DG-ENER-D3

More information

C-NRPP Quality Assurance Guidance for Radon Test Devices

C-NRPP Quality Assurance Guidance for Radon Test Devices C-NRPP Quality Assurance Guidance for Radon Test Devices Table of Contents: 1. Purpose 2. Definitions 3. Quality Assurance Plan 4. Calibration 5. Spikes/Performance Tests 6. Duplicates 7. Blanks 8. Cross-Checks

More information

CURRENT INDOOR RADON SITUATION IN LITHUANIA

CURRENT INDOOR RADON SITUATION IN LITHUANIA CURRENT INDOOR RADON SITUATION IN LITHUANIA Rima Ladygienė, PhD Radiation Protection Centre LEGISLATION National policy on radon is based on the appropriate recommendations of International Commission

More information

Evaluation on Combustion Characteristics of Finishing Materials for Exterior Walls

Evaluation on Combustion Characteristics of Finishing Materials for Exterior Walls Journal of Mechanics Engineering and Automation 5 (2015) 465-471 doi: 10.17265/2159-5275/2015.08.007 D DAVID PUBLISHING Evaluation on Combustion Characteristics of Finishing Materials for Exterior Walls

More information

Environmental radioactivity surveillance system of ANPA

Environmental radioactivity surveillance system of ANPA NUKLEONIKA 2001;46(4):137 141 PROCEEDINGS Environmental radioactivity surveillance system of ANPA Anna Alonzi, Sergio Mancioppi, Massimo Notaro, Paolo Zeppa Abstract No nuclear power station are operative

More information

Replacement of the Criticality Accident Alarm System in the Tokai Reprocessing Plant

Replacement of the Criticality Accident Alarm System in the Tokai Reprocessing Plant Replacement of the Criticality Accident Alarm System in the Tokai Reprocessing Plant Yukihisa SANADA a*, Makoto NEMOTO b, Kei SUZUKI c, Keiichi KAWAI d and Takumaro MOMOSE a a Nuclear Fuel Cycle Engineering

More information

film flow, ultrasonic, and microwave. The engineering development effort builds upon the

film flow, ultrasonic, and microwave. The engineering development effort builds upon the 9. e,.-..- n FLUID-BASED RADON MITIGATION TECHNOLOGY DEVELOPMENT FOR INDUSTRIAL APPLICATIONS K. V. tiu, J. D. Gabor, R. E. Hoitz, and K. C. Gross Argonne National Laboratory 9700 S. Cass Ave. Argonne,

More information

ISSN: [Agcaoili S. O., 7(3): March, 2018] Impact Factor: 5.164

ISSN: [Agcaoili S. O., 7(3): March, 2018] Impact Factor: 5.164 IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY EVALUATION OF A HOME-MADE SOLAR BOX HEATER FOR DOMESTIC HEATING Shirly O. Agcaoili 1/ 1/ Researcher and Faculty Member of the

More information

Portable Alpha/Beta Continuous Online Air Monitor (CAM)

Portable Alpha/Beta Continuous Online Air Monitor (CAM) Aer 5000 Desktop Portable Alpha/Beta Continuous Online Air Monitor (CAM) The Aer5200 monitors the ambient air continuously to detect airborne radioactive aerosols (LLRD). Typical application fields are

More information

A bright alpha numeric display and the simplified keypad (three buttons) allow the operation even under harsh conditions.

A bright alpha numeric display and the simplified keypad (three buttons) allow the operation even under harsh conditions. Personal Alpha/Beta Continuous Air Monitor (CAM) The pocamon (personal online continuous air monitor) monitors continuously the breathing air to detect airborne radioactive aerosols (LLRD). Typical application

More information

Development of the Groundwater Radon Detector

Development of the Groundwater Radon Detector Development of the Groundwater Radon Detector Japan-Taiwan International Workshop on Hydrological and Geochemical Research for Earthquake Prediction Date Sep.-24,2002 Place AIST (Tsukuba) Educational Research

More information

Fire propagation over combustible exterior facades exposed to intensified flame in Japan

Fire propagation over combustible exterior facades exposed to intensified flame in Japan MATEC Web of Conferences 46, 02002 (2016) DOI: 10.1051/matecconf/20164602002 C Owned by the authors, published by EDP Sciences, 2016 Fire propagation over combustible exterior facades exposed to intensified

More information

RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN

RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN , Volume 6, Number 4, p.248-254, 04 RADIATION BLOCKAGE EFFECTS BY WATER CURTAIN C.L. Choi Department of Building Services Engineering, The Hong Kong Polytechnic University, Hong Kong, China ABSTRACT Water

More information

Radiation Safety issues for the PF-AR in KEK

Radiation Safety issues for the PF-AR in KEK Radiation Safety issues for the PF-AR in KEK H. Nakamura, S. Ban, K. Iijima,Y. Namito and K. Takahashi High Energy Accelerator Research Organization(KEK), 1-1 Oho, Tsukuba-shi, Ibaraki-ken, 305-0801, Japan

More information

THE RISK OF LEGIONELLA DEVELOPMENT IN SANITARY INSTALLATIONS

THE RISK OF LEGIONELLA DEVELOPMENT IN SANITARY INSTALLATIONS Proceedings of the REHVA Annual Meeting Conference 23 April 208, Brussels, Belgium https://www.rehvam208atic.eu/ THE RISK OF LEGIONELLA DEVELOPMENT IN SANITARY INSTALLATIONS B. Bleys *, O. Gerin 2, K.

More information

Aer 5200/5300. Data Sheet Aer 5200/5300. Portable Alpha/Beta Continuous Online Air Monitor (CAM) Closer to your application

Aer 5200/5300. Data Sheet Aer 5200/5300. Portable Alpha/Beta Continuous Online Air Monitor (CAM) Closer to your application Aer 5200/5300 Portable Alpha/Beta Continuous Online Air Monitor (CAM) The Aer5200 monitors continuously the ambient air to detect airborne radioactive aerosols (LLRD). Typical application fields are nuclear

More information

Radon Measurements at Australian Antarctic Stations

Radon Measurements at Australian Antarctic Stations Radon Measurements at Australian Antarctic Stations Brendan Tate Australian Radiation Protection and Nuclear Safety Agency Why Measure Radon in Antarctica? Indoor radon Request from AAD to find out radon

More information

ABSTRACT FOR AARST CONFERENCE OCTOBER 1990 INSTRUMENTATION FOR RADON MEASUREMENT AND DIAGNOSIS

ABSTRACT FOR AARST CONFERENCE OCTOBER 1990 INSTRUMENTATION FOR RADON MEASUREMENT AND DIAGNOSIS ABSTRACT FOR AARST CONFERENCE OCTOBER 1990 INSTRUMENTATION FOR RADON MEASUREMENT AND DIAGNOSIS Knowledge of existing protocols and guidelines, coupled with equipment acumen can make for a successful endeavor

More information

Radon Environmental Health Program Health Canada, Alberta Region. Bradley Hameister October 4, 2011

Radon Environmental Health Program Health Canada, Alberta Region. Bradley Hameister October 4, 2011 Radon Environmental Health Program Health Canada, Alberta Region Bradley Hameister October 4, 2011 Question What is radon? 2 RADON Naturally occurring radioactive gas that comes from uranium in the ground

More information

RADON IN SOIL GAS VERSUS RADON FLOW IN CHARACTERIZATION OF URANIUM TAILING SITES

RADON IN SOIL GAS VERSUS RADON FLOW IN CHARACTERIZATION OF URANIUM TAILING SITES RADON IN SOIL GAS VERSUS RADON FLOW IN CHARACTERIZATION OF URANIUM TAILING SITES Michael Buzinny 1, Tatiana Lavrova 2, Maxim Romanchenko 1 and Victor Sakhno 1 1 SE The Marzeev Institute of Hygiene and

More information

THE EFFECT OF VENTILATION AND FILTRATION ON RADON DECAY PRODUCT MEASUREMENTS

THE EFFECT OF VENTILATION AND FILTRATION ON RADON DECAY PRODUCT MEASUREMENTS THE EFFECT OF VENTILATION AND FILTRATION ON RADON DECAY PRODUCT MEASUREMENTS Bill Brodhead WPB Enterprises, Inc., 2844 Slifer Valley Rd., Riegelsville, PA USA wmbrodhead@hotmail.com www.wpb-radon.com ABSTRACT

More information

Comprehensive applications of the gas flow proportional counters for Radiological Surveillance

Comprehensive applications of the gas flow proportional counters for Radiological Surveillance Comprehensive applications of the gas flow proportional counters for Radiological Surveillance D.A.R.Babu, Anand Raman, Ashokkumar P and D.N.Sharma Radiation Safety Systems Division, Bhabha Atomic Research

More information

SOIL GAS RADON MONITORING IN ODO ONA AS AN INDIRECT PROBE FOR LOCATION OF GEOLOGICAL FAULT ZONES Adesunloro, Gbenga M1.

SOIL GAS RADON MONITORING IN ODO ONA AS AN INDIRECT PROBE FOR LOCATION OF GEOLOGICAL FAULT ZONES Adesunloro, Gbenga M1. GSJ: Volume 5, Issue 9, September 2017 5 GSJ: Volume 5, Issue 9, September 2017, Online: ISSN 2320-9186 SOIL GAS RADON MONITORING IN ODO ONA AS AN INDIRECT PROBE FOR LOCATION OF GEOLOGICAL FAULT ZONES

More information

Brochure More information from

Brochure More information from Brochure More information from http://www.researchandmarkets.com/reports/2411894/ Radiation Detection, Monitoring & Safety Market (Geiger Counter, Ionization Chamber, Scintillator, Terahertz detector,

More information

Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E

Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E Thermo Fisher Scientific Messtechnik GmbH Frauenauracher Straße 96 91056 Erlangen Germany www.thermo.com info.rmp.erlangen@thermofisher.com

More information

Radiation Control and Monitoring System on the HTTR

Radiation Control and Monitoring System on the HTTR Radiation Control and Monitoring System on the HTTR Y. Minowa, T. Nakazawa, H. Kikuchi, T. Nomura and N. Sato Japan Atomic Energy Research Institute Oarai Research Establishment, Oarai-machi, Ibaraki-ken

More information

Environmental Radiation Monitoring System

Environmental Radiation Monitoring System Environmental Radiation Monitoring System Toshihiro Takagi Osamu Kimura Atsushi Minagoshi 1. Introduction In order to ascertain conditions near the perimeter of a supervised area at a nuclear power plant

More information

Comprehensive Short-Term Radon Survey Report

Comprehensive Short-Term Radon Survey Report Environmental Works, LLC 2634 SE Steele St. Portland, OR. 97202 Phone: 503-719-6715 Fax: 503-972-1866 Email: Sebastian@eworksnw.com Comprehensive Short-Term Radon Survey Report Test Performed At: Trillium

More information

MULTI-POINT DENSITY ARRAY FIXED-POINT INTERFACE PROFILING MEASUREMENT

MULTI-POINT DENSITY ARRAY FIXED-POINT INTERFACE PROFILING MEASUREMENT MULTI-POINT DENSITY ARRAY FIXED-POINT INTERFACE PROFILING MEASUREMENT Why Multi-Point DENSITY ARRAY? Leadership in Radiation-Based Measurement Ohmart/VEGA is proud to offer the ultimate in fixed-point

More information

Instrumentation. Post RDD Event Urban Reoccupancy. James P. Menge PE. The world leader in serving science

Instrumentation. Post RDD Event Urban Reoccupancy. James P. Menge PE. The world leader in serving science The world leader in serving science Instrumentation Post RDD Event Urban Reoccupancy James P. Menge PE Perception In the aftermath of a radiological emergency the public will see radiation and its potential

More information

RADON LAZERWEB SERVICES

RADON LAZERWEB SERVICES RADON LAZERWEB SERVICES WWW.LAZERWEBSITES.COM Contents Overview...2 Radon is estimated to cause thousands of lung cancer deaths in the U.S. each year...2 Radon is a cancer causing, radioactive gas...2

More information

Forward-scattered radiation from the compression paddle should be considered when average (or mean) glandular dose is estimated

Forward-scattered radiation from the compression paddle should be considered when average (or mean) glandular dose is estimated Forward-scattered radiation from the compression paddle should be considered when average (or mean) glandular dose is estimated Bengt Hemdal Department of Medical Radiation Physics, Lund University Skåne

More information

Technologies, Solutions, and Applications. Continuous Level Coke Drum Measurement System

Technologies, Solutions, and Applications. Continuous Level Coke Drum Measurement System Technologies, Solutions, and Applications Continuous Level Coke Drum Measurement System Leadership in Continuous Coke Drum Level Refiners are processing heavier crude, which increases the amount of resid

More information