Installation & Operating Instructions for Area, Master & Combination Alarm Systems

Size: px
Start display at page:

Download "Installation & Operating Instructions for Area, Master & Combination Alarm Systems"

Transcription

1 Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for Area, Master & Combination Alarm Systems Andersen Medical Gas 12 Place Lafitte Madisonville, LA Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

2 Contents Tri-Tech Introduction Features & Benefits Major components Alarm Panel Installation Rough-In Box Installation Mounting Wiring Plumbing Front Panel Installation Wiring alarm front to power supply Installing the Transducers Wiring the transducers Wiring the Remote Devices Labeling the Alarm Front Alarm Displays & Functions Component identification CPU module displays & functions Power on indicator Silence button Test button Program mode indicator Program button Module button Units button Alarm button History button View Alarm History Clear Alarm History Identification (ID) button Status LED CPU reset Table of Contents Gas (Area) module displays & functions Gas service Pressure LED Digital Display Pressure LED Hi/Normal/Low Display- 10 Units of measure LED display Remote Signal (Master) module displays & functions Normal status LED Abnormal status LED Remote signal description label Alarm Operation Component identification Remote signal (master) module Gas (area) module Silencing the alarm Testing the alarm Programming the Alarm Keypad identification Accessing the Program Mode Hi/Low pressure limits Remote Signal Alarm Points Alarm Repeater Delay Units of Measure Display Zone Identification # Adding/Removing Modules System Alert Codes Appendix A Glossary of Terms Appendix B Alarm Specifications Appendix C Signal Wire Color Code Log Technical Assistance or Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

3 Introduction Color coded gas identification label Gas (Area) Module Remote signal (master) description label Audible alarm location Digital pressure display Slot for future expansion with label Alarm test button Alarm silence indicator Alarm silence button Power on indicator Programming keypad Program mode indicator Status code display Units of measure indicator PSI and Hg or Bar or kpa Remote signals abnormal status LEDs - Red Remote signals normal status LEDs - Green Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

4 Introduction Introduction The Tri-Tech Medical gas alarm system monitors the status of the medical gas distribution system and provides audible and visual indicators. The alarm can be used in conjunction with the Tri-Tech Medical MCP100A or ACP-100-A Interface Panel to communicate all alarm/status information to the building management system. The Tri-Tech alarm system monitors the status of the medical gas sources in accordance with NFPA 99 and CSA Z305.1, and are C.U.L. listed. Features & Benefits Microprocessor controlled State of the art maintenance free electronics provide excellent reliability. Remote signal modules Provides monitoring of 16 separate remote devices. Self test program alarm code display LED display reveals the nature of the malfunction and reduces maintenance time. Three year PC board guarantee A quality product you can buy with confidence. Independent multiple master alarms When service is interrupted to one alarm the other(s) continue to operate independently. Transient signal filter Prevents or reduces nuisance (false) alarms signals created by RFI interference. Transient alarm notification Indicates which parameter experienced a transient (brief intermittent) signal. Audio & visual signal indicators Audible alarm and visual display of both normal and abnormal status of each signal monitored assures prompt and informative indication of a problem. Programmable remote signals Configure each remote signal independently (either normally closed, normally open or turn unused signals off). Programmable gas module high and low set points (Programmed from factory at 60/40 psig and 12 in Hg). Compact unit Requires minimal wall space. Optional building management interface panel Allows communication with building management system. Alarm history recall Can recall previous alarm signals even after the alarm condition has been corrected and the alarm panel has been cleared. CUL Listed Alarm repeat feature Adjustable from 5 minutes to 4 hours (factory programmed for 10 minutes). Easy to install and service Hinged frame with lanyards for easy accessibility.. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

5 Introduction / Components Rough-in box or back box Front panel or front Transducer Assemblies Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

6 Introduction Introduction / Components The Tri-Tech Medical gas alarm system monitors the status of the medical gas distribution system and provides audible and visual indicators. The alarm panel can be interfaced with a Tri-Tech Medical MCP100-A or ACP-100-A Building Management interface panel for remote monitoring of the entire facility. The Tri-Tech Medical gas alarm system is a two or three section assembly comprised of a rough-in box, a front panel and transducers (only if it is an area alarm or combination alarm). Rough-in box The rough-in box houses the power supply, fuse, on/off switch, and a terminal strip for electrical wiring. An isolated transformer reduces the 110V or 220V AC input to low voltage DC. Front panel The front panel includes enclosed printed circuit boards with programming circuitry. The CPU module includes a power on indicator, programming buttons, an audible alarm, and an LED status display. The Remote Signal module(s) can monitor up to 16 signals per circuit board. Multiple remote signal modules can be ordered in a single alarm. The signals can be configured to display an abnormal condition on either a normally closed (NC) or a normally open (NO) circuit. Each signal may also be turned off if it is not being utilized. The Gas (Area) module(s) on the front panel are identified with gas specific, color coded labels (per NFPA 99 or CSA Z305.1). The gas displays include LED s which indicate high/normal/low pipeline pressure. A digital LED display shows the actual gas pressure. The gas pressure may be displayed in PSI and Hg, or BAR or kpa. The unit is pre-programmed to display PSI / Hg from the factory, but may be re-programmed in the field to display BAR or kpa. Transducers The transducer converts pressure to an electrical signal and supplies the electrical signal to the front panel. After the initial 24 hour 150 psi standing pressure test (required per NFPA 99) has been completed the pressure/vacuum transducers are mounted in the rough-in box. The transducers may be remotely attached to the piping system at distances no further than 100 feet using Beldon 8777 shielded cable. Tri-Tech Medical recommends mounting the transducers in the alarm back box not remotely. Should a transducer require service or replacement is considerably more time consuming to locate and replace or service remotely located transducers Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

7 Alarm Displays & Functions Color coded gas identification label Gas (Area) Module Remote signal (master) description label Audible alarm location Digital pressure display Slot for future expansion with label Alarm test button Alarm silence indicator Alarm silence button Power on indicator Programming keypad Program mode indicator Status code display Units of measure indicator PSI and Hg or Bar or kpa Remote signals abnormal status LEDs - Red Remote signals normal status LEDs - Green Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

8 Alarm Displays & Functions Alarm Displays & Functions CPU Module Power on Indicator The power on indicator (green LED) is illuminated whenever electrical power (120 or 240 VAC) is connected to the alarm. Test Button When the Test button on the front panel is pressed, the alarm undergoes a self-test checking the lights and displaying the gas settings. If errors are detected, they are displayed on the status LED and/or on the area module gas LED displays. Status Display The status LED displays system alert codes when a problem occurs (these are numeric). See System Alert Codes on page 34 for details. The status LED also displays a P whenever the alarm is in the program mode. Alarm Silence In the event of an alarm condition an audible alarm sounds. The audible alarm can be silenced by pressing the alarm silence button. Whenever there is an alarm condition and the alarm has been silenced, the Alarm Silence Indicator LED will be illuminated. The high or low pressure LED or the remote signal LED will remain illuminated until the alarm condition is rectified. Program Mode Indicator The Program Mode Indicator LED is only illuminated when the alarm has been placed in the program mode (see Programming the Alarm page 25). It should be noted that when this LED is illuminated and the alarm is in the program mode, it is not monitoring or reporting alarm conditions for the medical gas systems. History Button The History button may be pressed and held at any time to view alarm history. Viewing alarm history is only active when the History button is depressed, releasing the button returns the alarm to normal operation. When depressed H will be displayed in the Status Display LED. Gas(Area) Module - If there was an alarm condition for any gas (area) module, the High and/or Low Pressure LEDs will be illuminated. If both the High and Low Pressure LEDs are illuminated, the gas has had both a High and Low alarm. Remote Signal (Master) Module If there was an alarm condition for any Remote Signal the Red LED will be illuminated. All other LED s will be off. Clear Alarm History To clear alarm History you simply press the History button, hold it down and press Silence button (while the history button is still depressed. Program Button See Programming the Alarm pages Module Button See Programming the Alarm pages Units Button See Programming the Alarm pages Repeater Button See Programming the Alarm pages ID (Identification) Button See Programming the Alarm pages CPU Reset Button See Programming the Alarm pages Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

9 Alarm Displays & Functions Gas Displays (Area Module) Pressure Reading Display The LED pressure reading displays the pressure as indicated by the transducer. Whichever of the PSI or BAR or kpa lights is illuminated reveals the unit of measure displayed on the LED pressure reading. The Nitrogen gas display indicates one tenth of the actual pressure when displayed using kpa units (i.e kpa is displayed as 110 and the kpa x 10 LED is illuminated). High / Normal / Low Status Lights In addition to digitally displaying the pressure of a gas, the pressure is displayed by a series of green, yellow & red LEDs reflecting the pressure as either normal, high or low. These indications are relative to the high and low pressure set points which have been programmed into the alarm. These high and low set points should be set in accordance with NFPA 99 and CSA Z305.1 at ± 20% of the normal operating pressure. As the pressure climbs above normal line pressure to 20% above normal line pressure, the LED s on the upper side of normal will illuminate, then the uppermost normal & yellow High LED s will illuminate, finally the two red High LED s will illuminate. Simultaneously with the Red LED s illuminating, the audible alarm will sound. As the pressure falls to below normal line pressure to 20% below normal line pressure, the LED s on the lower side of normal will illuminate, then the lowermost normal & yellow low LED s will illuminate, finally the two red Low LED s will illuminate. Simultaneously with the Red LED s illuminating, the audible alarm will sound. Units Button & PSI / BAR / Kpa LED s By placing the alarm in the program mode (see Programming the Alarm page 25 ) the units button toggles the LED pressure readings on the gas displays between PSI (inches Hg for vacuum & evacuation BAR / kpa LEDs. The corresponding PSI, BAR or kpa LED is illuminate indicating which units are being displayed on the gas display. Note: Vacuum & evacuation are actually displayed as inches Hg in the PSI mode. The Nitrogen gas display indicates one tenth of the actual pressure when displayed using kpa units (i.e kpa is displayed as 110 and the kpa x 10 LED is lit. The power on LED is illuminated whenever electrical power is connected to the alarm. Note Alarm settings are maintained even if power is interrupted. Remote Signal Status LED s (Master Module) Normal Status Light (Green) The normal green status light when illuminated signifies that the equipment being monitored is not in an alarm condition. A flashing green light indicates a normal condition, but a transient (see glossary of terms for definition) signal from source equipment has been detected. Abnormal Status Light (Red) The abnormal status light when illuminated signifies that the equipment being monitored is in an alarm condition. Note: If no LED is illuminated the remote signal status LEDs have probably been set in the unused or off position. This can be verified by having the alarm run a self-diagnostic check. Simply press the Test button. As the alarm runs thru the test cycle, observe the Status Display LED when the Remote Signal in question is blinking. A d displayed on the Status Display LED indicates that the Remote Signal has been disabled. Remote Signal Description Label The remote signal description label identifies the equipment that is being monitored on the normal and abnormal LED s to the left and right of the label. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

10 Operation Color coded gas identification label Gas (Area) Module Remote signal (master) description label Audible alarm location Digital pressure display Slot for future expansion with label Alarm test button Alarm silence indicator Alarm silence button Power on indicator Programming keypad Program mode indicator Status code display Units of measure indicator PSI and Hg or Bar or kpa Remote signals abnormal status LEDs - Red Remote signals normal status LEDs - Green Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

11 Alarm Operation Alarm Operation Refer to the Alarm Installation section for the procedure necessary to install the alarm. The Programming The Alarm section covers the procedures to follow in order to configure the alarm if the preprogrammed settings are not appropriate. After installation has been completed and the alarm has been properly configured, it is ready for operation. Remote Signal (Master) Module With electrical power applied to the alarm, the system will monitor the status of the medical gas distribution system. The remote signals (master) module displays (via LED indicators) the normal/abnormal status of the remote equipment connected to the alarm. In normal operation all active signals will be green. Whenever an abnormal status signal is detected the corresponding green LED will be extinguished and the red LED will light. Simultaneously an audible alarm will sound. Pressing the Silence button silences the audible alarm (the Red LED continues to be lit). The appropriate personnel should be notified immediately of the alarm condition. When the alarm condition has been rectified the Red LED is automatically extinguished and the Green LED illuminates. Whenever an intermittent Transient (see Glossary of Terms for definition) abnormal status signal is detected from remote equipment by the alarm, the Green or Red LED will start to flash. Transient alarms indicate a potential problem. The appropriate personnel should be notified of the alarm condition. Pressing the Test or Repeater button returns the flashing LED to normal operation. Note The Remote Signal (Master) alarm is set to Ignore transient signals that are less than 0.6 seconds in duration. Gas Modules With the electrical power applied to the alarm and the gas systems adequately pressurized, the following indicators are illuminated: 1) the Power On LED, 2) the pressure readings of the gas on each gas display, 3) the Normal LED (green) on each gas display. If the pressure of one of the gases drops below the low limit setting, the following events take place simultaneously: 1) the Normal LED will be extinguished, 2) the Pressure Low LED (red) will light up, 3) an audible alarm will sound. If the pressure of one of the gases rises above the high limit setting, the following events take place simultaneously: 1) the Normal LED will be extinguished, 2) the Pressure High LED (red) will light up, 3) an audible alarm will sound. Silencing the Alarm Press the Silence button when the alarm is sounding and the alarm will be silenced. This alarm is equipped with a Repeater Delay feature which monitors only the Gas Module (Area) alarms. The Repeater Delay has been factory programmed to make the alarm re-sound every 11 (eleven) minutes as long as the alarm condition exists. (Note: Remote Signal (Master) alarms are not monitored by the Repeater Delay). Pressing the Test button initiates a self-test of the alarm. Each LED will illuminate in succession for 2.5 seconds. The displays are then cleared and the system circuitry is checked for proper voltages. If no errors are found, the alarm returns to normal operation. If an error is found, the error is indicated on the Status Code Display. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

12 Alarm Operation Testing the Alarm Pressing the Test button initiates a self-test of the alarm, resets any system status alerts that may be active and re-initiates the alarm-repeat function. Pressing any key will abort the test displays but not the internal tests. The test commences in the following order: 1. The alarm will sound, the status display will display 8 and the 1 st active module will be displayed with all lights illuminated for 2.5 seconds. This is the only time the alarm will sound and the Program & Silence LED indicators will illuminate during the test. 2. Each remaining active module will be displayed with all lights illuminated for 2.5 seconds, in sequence. The Gas (Area) Modules will be displayed first (if gas modules exist). The Remote Signal (Master) Modules will be displayed second (if Remote Signal modules exist). Each existing active module will be displayed for 2.5 seconds. 3. After all existing modules have completed the lamp verify test above, any existing Gas (Area) Module set points will be displayed for 2.5 seconds. If no Gas Modules exist, these displays will be bypassed. All the high set points will be displayed for 2.5 seconds first with the high alarm LEDs indicating which set point you are observing. Any Master modules that exist will not be illuminated during this time. If a gas is not active, i.e. the module exists but an inactive shunt instead of a transducer is connected to the gas, the gas display will be blanked. 4. Next the active Master modules will be displayed, one at a time, displaying the current alarm programming and the contact state of each alarm point. Each of the Master module displays will consist of the following: The module will be illuminated. The top parameter will be flashing indicating it is the first alarm point under test (the active alarm point). The status display will display the current contact state of the active alarm point. This display will consist of: C = closed, O = open, d = disabled. If the active alarm point is disabled, both alarm point LED s (both the green & the red) will be illuminated. This is only for the test display. In normal operation, both of the LED s will be turned off. After 2.5 seconds, the next lower alarm point will flash and the status display will show it s contact state for 2.5 seconds. This sequence will continue until the last (bottom) alarm point has been displayed. This action will continue until all active Master module alarm points have been displayed. 5. When all active modules have been displayed (Gas and/or Master), the test will be complete and the system will return to normal operation. Remember, you can exit the test at any time by pressing any key. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

13 Alarm Installation Alarm Installation Installation of the Tri-Tech Medical alarm involves installing the rough-in box, the transducers (if it is an area alarm or combination alarm) and front panel and making the necessary conduit, plumbing and electrical connections. All installation and testing should be done in accordance with NFPA 99 or CSA Z WARNING: All electrical power should be disconnected prior to installation. WARNING: This device should only be installed by qualified personnel. Installation should not be attempted by anyone not having general experience with the installation of devices of this nature. Rough-In Box Installation This is a rough-in box for a combination alarm. Your rough-in box should look the same or similar to this unit. Refer to the building plans to determine the location of the alarm. The contractor is to provide rigid mounting that will support the alarm box on both sides. The metal flanges provided on both sides of the rough-in box are to rest against the rigid mounting brackets. Screws (contractor provided) are to be driven thru the holes in the metal flanges into the mounting brackets. Mount alarm rough-in box so it will be flush or just below the finished wall surface using the adjustment feature on the flanges. (Side/end view of rough-in back box) Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

14 Alarm Installation The power supply hole is located in the bottom (left side) of the rough-in box. Remove the plaster cover and panel covering the power supply. Remove the plug from the hole. Make conduit connections for wiring from the facility emergency power source. To remove the power supply cover, loosen the two screws at the top of the cover and slide the cover to the right, then lift the cover over the screw heads. Slide the wiring harness strain relief to the left until it is free from the cover. Wiring Use the ½ conduit knock-out provided on the bottom left side of the rough-in box to route conduit to supply either 110 or 220 V AC to the power supply. Note: Separate conduit should be used for low voltage wires (use knock outs provided on the top of the box). Neutral Hot Field Ground Route wires through the power supply conduit installed in the bottom and left side of the roughin box. Connect the 110 or 220V AC facility emergency power source electrical wiring to the terminal strip provided on the lower left side of the box. (N = neutral, H = hot, FG = field ground) Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

15 Alarm Installation The wiring harness will be provided pre-wired to the power supply. If by chance it is not prewired, the single green ground wire should be fastened to the screw on the top left corner of the power supply board. The wiring plug in connector (which should contain the rest of the wires), plugs into the connector terminal on the upper left corner of the power supply board. The black wire will be on the top. This is how the power supply should look when the wiring is completed. Replace the power supply cover and plaster cover (to protect the unit while the drywall and wall covering work are being done) after wiring is complete Plumbing the Riser Extension Tubes Riser extension tubes After removing the protective plastic caps, connect the riser extension tubes (only included if this is an area or combination alarm) to the gas piping system. Make sure drop is for the proper gas service. Using a purge gas to insure cleanliness in the tubes, silver braze the joints. Do not use soft solder. Conduct heat away from the check valves. (A purge kit is required for purge equipment connection to the system). The 150 psi standing pressure test of the piping system must be successfully completed and all pressure removed from the system before connecting the transducers to the system. Attaching transducers without depleting the 150 psi standing pressure first will result in damage to the transducers! Note: It is recommended that gas piping system tubing be connected from the extension tube to the ceiling for a future or blank gas display even though a transducer is not being installed. Cap the unused tube at the ceiling. This will simplify future expansion. Note: It is also possible to install the transducers remotely (not in the alarm box). In this type of installation the riser extension tubes should be removed from the alarm back box and connected to gas piping system at a distance not to exceed 100 feet (wiring harness length) from the alarm back box. Belden 8777shielded cable is required. Note: Tri-Tech Medical recommends mounting the transducers in the alarm back box not remotely. Should a transducer require service or replacement is considerably more time consuming to locate and replace or service remotely located transducers. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

16 Alarm Installation Installing the Hinges & Clips After the walls are finished, the hinges and clips may be installed onto the rough-in box. The two hinges should be fastened to the lower front edge of the rough-in box using the screws provided. The hinge should rest flat against the surface of the drywall. The clips should be fastened to the upper front edge of the rough-in box opposite the hinges using the screws provided. The clip should rest flat against the surface of the drywall. Installing the Alarm Front Panel Slide the holes cut into the bottom edge of the front panel over the front catch of the hinges. Support the front panel in the open position to attach the lanyards. Attach the two (left & right side) lanyard cables to the alarm front panel using the screws provided. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

17 Alarm Installation Wiring the Alarm Front Panel to the Power Supply Attach the green ground wire, which is in the wiring harness, to the ground screw on the left corner of the front panel just in front of the power supply. Attach the plug connector at the end of the wiring harness to the appropriate connector located at the top left corner of the CPU circuit board. The plug should lock into place. The plug can only be inserted one way. This is how the wiring harness plug connector should look when properly installed on the back of the CPU board. Installing the Transducers After depressurizing the 150 psi standing pressure from the piping system, install the appropriate transducers to the appropriate riser extension tube connectors inside the rough-in box. Hand tighten then torque lightly with a wrench. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

18 Alarm Installation Connecting the Transducer Wiring to the Alarm Front Circuit Board Insert the plug(s) from the transducer(s) to the appropriate connector(s) on the back of the gas module circuit board(s). The plug on the transducer and the panel connector are both labeled by gas symbol at the factory for proper installation. The plug can only be inserted one way. Visually verify that the appropriate transducer plug is being attached to the appropriate gas module by looking at the front of the alarm panel. The transducer plug will look like this when it is connected properly to the gas module. Insert the transducer cable underneath the cable clip as shown using the clips provided. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

19 Alarm Installation Wiring the Remote Devices to the Remote Signal (Master) Board Locate the two banks of 16 terminal connectors on the back of the remote signal (master) board. There are a total of 32 connection points two connection points for each remote signal. The pairs are labeled P1 thru P16 on the circuit board. The top terminal of each pair is where the signal wire should be landed. Remote signal #1 signal wire Remote signal #1 common wire Remote signal #2 signal wire Remote signal #2 common wire There are two options for landing the common wire(s). Option #1 you can land the corresponding common wire on the bottom terminal of each pair of terminal connectors. This option is recommended for maximum interference suppression. Option #2 you can connect all of the common wires together and land just one common wire on the bottom terminal of P16. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

20 Alarm Installation Labeling the Alarm Front After all remote signal (master) wires have been landed, identified & cataloged (using appendix C in this manual), you are ready to apply the remote signal labels. Remove by unthreading the two nylon nuts shown here from the back of the alarm front panel. The panel will now lift out of the black frame. Provided in the packaging with the alarm front panel is a long white card and a pre-printed sheet of labels for the remote alarm signals. Wash your hands before handling these labels and cards! Apply the labels to the appropriate position on the card using the guide lines on the card. When complete, insert the card into the Monitored Signals label pocket - between the stainless steel plate and the front of the Monitored Signals label. The gas modules are shipped pre-labeled from the factory. If a gas service is changed to a different gas or added to the alarm, it will be necessary to insert the appropriate label in the pocket of the gas module. The new gas label supplied by Tri- Tech Medical will slide into the gas module label pocket between the stainless steel plate and the front of the gas module label. When the labeling is completed, install the stainless plate into the black frame and re-attach with the two nylon nuts. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

21 Alarm Installation You are now ready to supply power to the alarm. The toggle switch on the front of the power supply should be placed in the ON position. Check the green power LED indicator on the front of the CPU module. It should be illuminated. You are now ready to press system to normal operating alarm is shown here with O normal operating pressures Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

22 Progamming the Alarm Alarm silence indicator Alarm silence or down button Alarm test or up button Power on indicator Program mode indicator Program button Module button Status code display Repeater button CPU Reset button History or left/right button Gas ID (identification) button Units button Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

23 Programming the Alarm Programming the Alarm The alarm has been programmed at the factory prior to shipment. Programming of the alarm may be necessary if: a) the high or low pressure limits for a gas need to be modified, b) if a blank gas module is being put into service for an added gas service, c) if a gas service is being deleted, d) if a gas service is being changed, e) if a remote signal alarm point needs to be re-configured, f) if a remote signal module is being added in the place of a blank module, g) if the alarm identification number needs to be changed. Note: Only authorized personnel should program the alarm! It is important to note that while the alarm is in the program mode, it is not monitoring the medical gas system and alarm conditions will not trigger an alarm. Accessing the Alarm Program Mode The programming buttons are located on the front, lower left corner of the alarm panel see photo on page 24. To enter the programming mode, the Program button must be pressed and held. While the Program button is pressed the ID button must also be pressed. When this sequence is performed correctly, the program mode will be entered and a P will be displayed on the Status Code Display. All other displays will be blanked. If this sequence is not performed correctly, an O may be displayed on the Status Code Display indicating a key press error. Pressing the Silence Key will remove this error indication. When you have successfully accessed the program mode, it is important to note that some of the buttons revert to their sub-functions: The Test button becomes the UP key ( ). The Silence button becomes the DOWN key ( ). The History button becomes the TOGGLE key ( Note: In order to perform any of the following programming features you must first set the alarm in the program mode. Programming the High & Low Gas Pressure set points While in the program mode, press the Module button until the desired Gas Module is illuminated. With the desired Gas Module illuminated, press the TOGGLE key (History button) to view the high & low set points. The high or low set point pressures will be displayed on the Digital Pressure Display. Simultaneously, the corresponding High or Low Pressure Indicator LED s will illuminate. With the high or low set point pressure displayed, press either the UP key (Test button) or the DOWN key (Silence button) to change the set point to the desired pressure. When high and low set points have been changed to the desired settings, you must press the Program button twice to save the changes and return the alarm to the normal monitoring mode. Note: The alarm is designed with a safety feature so that the high and low set points must be at least 2 to 5 increments (psi, bar, kpa, in Hg) apart. The high set point will not be able to be set below the low set point and visa versa. Note: Depressing the UP or DOWN keys will accelerate the rate of change up to 12 increments per second. Note: An abbreviated name of the gas service will be one of the displayed items along with the high and low pressure alarm set points. The full names of the gases and their corresponding abbreviations are per the chart below: n2 = nitrogen o2 = oxygen n20 = nitrous oxide Co2 = carbon dioxide UAC = vacuum EUA = evacuation Air = medical air hpa = high pressure air hpo = hyperbaric oxygen hpc = high pressure carbon dioxide gas = gas mixture Err = error Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

24 Programming the Alarm Note: If Err should be displayed on the Gas Pressure Display of any gas module, this indicates a problem. Some possible problems and corrective actions are: The transducer is not connected to the Gas Module. To correct, check the transducer connection to the back side of the Gas Module. A transducer for a different gas service has been connected to the Gas Module. To correct, check the transducer and the Gas Module gas identification labels and make sure they match. If the above corrective actions do not correct the problem, contact the factory for assistance. Programming the Remote Signal (Master) Alarm Points While in the program mode, press the Module button until the desired Remote Signal (Master) Module is illuminated. It s LED will display the current programming for the contact state that defines a normal condition. With the desired Remote Signal (Master) Module flashing, press the UP or DOWN key to view the next Remote Signal Alarm Point. All contacts are programmed as closed contact equals a normal condition from the factory. The contact point that is flashing is capable of being programmed. The Status Code Display will display the current state for the contact point that is flashing. The Status Code Display will read either: O for open or C for closed or d for disabled Note: Any contact point that has been disabled will have both the green (normal) and red (alarm) LED illuminated only when in the programming mode, to make it easy to identify them. In the normal mode of operation, both LED s of a disabled parameter will be blanked (off). With the desired contact point flashing, press the TOGGLE key (history button) to change the state of the selected contact point. The sequence of the TOGGLE key is as follows: Closed = normal Open = normal Disabled Each individual contact point may be programmed to any of the three (open, closed or disabled) states. Using the UP or DOWN keys you can go on to the next contact point. When programming of all of the contact points is complete press the Program button twice to save the changes and return the alarm to the normal mode. Programming the Gas Alarm Repeater Delay While in the program mode, press the Alarm button. A number (0 7) will be displayed on the Status Code Display. The first gas display will display A-r which is an abbreviation for Alarm Repeat. All other modules will no longer be illuminated. If a 0 is displayed on the Status Code Display, this means that the Alarm Repeat feature has been disabled. (This setting is ideal during the construction phase of a project). (Note disabling or enabling the Repeater will not change the delay setting). To enable this feature press the TOGGLE key (History button) until a number 1 7 is displayed on the Status Code Display. When the Alarm Repeat feature has been enabled, the UP or DOWN keys may be used to change the setting. The chart below represents the approximate length of silence time before the silenced alarm will repeat: 1 = 5 minutes 2 = 10 minutes 3 = 15 minutes 4 = 30 minutes 5 = 1 hour 6 = 2 hours 7 = 4 hours The alarm will be pre-programmed from the factory at ten minutes. This is the recommended setting. To save the change and return the alarm to the normal mode, press the Program button twice. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

25 Programming the Alarm Programming the Units of Measure Displays While in the program mode, press the Units button. (This key is only active if a Gas Module exists in the alarm.) The Status Code Display will display a U and the first Gas Module will display E-U on the Gas Pressure Display. The Units of Measure Indicators will display the selected engineering units. The High / Low Pressure LED s will be off. All other module displays will be off. Use the TOGGLE key to switch between engineering units PSI, BAR or kpa. When you have selected the engineering units that you would like for the pressure display, press the Program button twice to save the change and return the alarm to the normal mode. Note: All Gas Modules will display pressure in the same units as selected. Note: Vacuum & Evacuation Gas Modules will automatically display in in/hg when PSI is selected. Note: The kpa and x10 LED s will both illuminate on all high delivery pressure Gas Modules (i.e. Nitrogen and Hyperbaric Oxygen) when kpa is selected. Because the Gas Pressure LED Display is only able to display three digits and high delivery pressures viewed in kpa are four digits, the Gas Pressure LED must be read as a four digit number by multiplying the displayed number by ten. I.E. the Gas Pressure LED Display is displaying 125 in kpa. The pressure should be read as 1,250 kpa (125 x 10). Programming the Zone Identification # Note: This feature is only used when a Tri-Tech Medical ACP-100 Building Management Interface Panel is added to the alarm system. Note: Each alarm must have a unique Zone Identification Number no two can share the same number. While in the program mode, press the I.D. button. The Zone Identification Number will be displayed on the Status Code Display. If a Gas Module is present in the alarm, Id will be displayed on the Gas Pressure LED Display. Valid identification numbers are As the Status Code Display is only a single character display, it will flash the current Zone Identification Number. I.E. let s assume the current Zone Identification Number is 16, the Status Code Display will flash: Blank, Blank, 1, Blank 6, Blank, Blank, 1, Blank, 6, etc. Use the UP and DOWN keys to change the Zone Identification Number. Press the Program button twice to save the Zone Identification Number selected and return to the normal mode. Note: Valid identification numbers are The alarm is pre-programmed at the factory with 0 (zero) as the zone identification number. 0 (zero) should be used whenever there is not a Tri-Tech Medical ACP-100 Building Management Interface Panel installed in the alarm system. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

26 Adding or Deleting Modules Adding or Deleting Modules This section covers the procedure for adding or deleting a module to the alarm. This section also covers the procedure for removing an active Gas Module from service or placing a currently inactive gas module into service. Note: Before attempting any wiring or electrical component changes, the power to the alarm must be turned off! Failure to turn the power off the alarm during wiring or electrical component changes may result in damaged or destroyed components. Use the ON/OFF switch that has been provided in the lower left hand corner of the rough-in or back box. Adding / Removing a Remote Signal (Master) Module In order to add a Remote Signal (Master) Module, the alarm must have a future expansion position (as shown on right of photo) available. The first step will be to remove the future expansion label and replace it with a remote signal label. Open the alarm front by pulling the top of the front panel away from the wall. Next, remove the front panel from the front frame by unthreading the two nylon nuts shown here from the back of the alarm front panel. The panel will then lift out of the frame. Set the panel upright with its bottom resting on the frame. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

27 Adding or Deleting Modules Peel off the existing future expansion label and remove any remaining adhesive with rubbing alcohol. After the surface is clean and dry, install the remote signal label. Match up the holes on the top and bottom of the label with the holes on the plate for proper alignment. Provided in the packaging with the alarm front panel is a long white card and a pre-printed sheet of labels for the remote alarm signals. Wash your hands before handling these labels and cards! Apply the labels to the appropriate position on the card using the guidelines on the card. When complete, insert the card into the Monitored Signals label pocket - between the stainless steel plate and the front of the Monitored Signals label. Reinstall the front panel into the front frame. Attach the remote signal circuit board to the alarm plate using the screws and stand-off s provided. Attach the short ribbon cable to the open position on the bottom of the neighboring circuit board and the other end of the short ribbon cable to the innermost receptacle on the newly added remote circuit board. This will leave the outermost receptacle vacant on the new circuit board. Note: The male connector on the ribbon cable will only connect one way to the receptacle. Remove the existing ribbon cable running across the top of the circuit boards. Install the new longer ribbon cable provided to connect all of the circuit boards. Note: The male connector on the ribbon cable will only connect one way to the receptacle. Note: The top and bottom ribbon cable male connectors and female receptacles are not interchangeable. When finished : Press the Reset button Press and hold the Program button and press the ID Button while still holding the Program button ( P should appear in the Status Code Display) Press the Program button twice. Turn the power off Turn the power back on Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

28 Adding or Deleting Modules Wiring the Remote Devices to the Remote Signal (Master) Board Locate the two banks of 16 terminal connectors on the back of the remote signal (master) board. There are a total of 32 connection points two connection points for each remote signal. The pairs are labeled P1 thru P16 on the circuit board. The top terminal of each pair is where the signal wire should be landed. Remote signal #1 signal wire Remote signal #1 common wire Remote signal #2 signal wire Remote signal #2 common wire There are two options for landing the common wire(s). Option #1 you can land the corresponding common wire on the bottom terminal of each pair of terminal connectors. This option is recommended for maximum interference suppression. Option #2 you can connect all of the common wires together and land just one common wire on the bottom terminal of P16. Unused remote signals may simply be disabled (see page 26 Programming the Remote Signal Alarm Points). When finished : Press the Reset button Press and hold the Program button and press the IDButton while still holding the Program button ( P should appear in the Status Code Display) Press the Program button twice. Turn the power off Turn the power back on Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

29 Adding or Deleting Modules Adding / Removing a Gas (Area) Module In order to place an unused Gas (Area) Module into service, the alarm must have a future expansion position available Open the alarm front by pulling the top of the front panel away from the wall. Next, remove the front panel from the front frame by unthreading the two nylon nuts shown here from the back of the alarm front panel. The panel will then lift out of the frame. Set the panel upright with its bottom resting on the frame. Remove the future label from the window of the unused Gas Module. Insert the proper gas service label into the window of the unused Gas Module. Re-install the front panel into the front frame. If the proper riser tube assembly is not installed, it will need to be installed and piped at this time. Knockouts have been provided in the plates on the top of the alarm box. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

30 Adding or Deleting Modules Install the appropriate transducer(s) to the appropriate riser extension tube connectors inside the rough-in box. Hand tighten then torque lightly with a wrench. Note: If this gas service is being added to the system, it must be installed and tested in accordance with NFPA 99 guidelines. It is important that the transducers are not connected to the piping system until after depressurizing the 150 psi standing pressure from the piping system. Insert the plug(s) from the transducer(s) to the appropriate connector(s) on the back of the gas module circuit board(s). The plug on the transducer and the panel connector are both labeled by gas symbol at the factory for proper installation. The plug can only be inserted one way. The transducer plug will look like this when it is connected properly to the gas module. Install the strain relief(s) provided to the circuit board stand off(s) using the screws provided. When finished : Press the Reset button Press and hold the Program button and press the IDButton while still holding the Program button ( P should appear in the Status Code Display) Press the Program button twice. Turn the power off Turn the power back on Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

31 Adding or Deleting Modules Inactivating Gas Module Displays If a Gas Module is not used or to be removed from service, jumper connectors must be installed in order to blank the displays and to prevent the audible alarm from sounding. If the Gas Module to be removed from service is on the bottom of the circuit board (labeled Transducer-B on the circuit board), jumper connectors must be attached to the 1 st & 2 nd and to the 8 th & 9 th (when counted from the top of the circuit board to the bottom) transducer connector pins as shown here. jumper connectors If the Gas Module to be removed from service is on the top of the circuit board (labeled Transducer-A on the circuit board), jumper connectors must be attached to the 4 th & 5 th and to the 11 th & 12 th (when counted from the top of the circuit board to the bottom) transducer connector pins as shown here. jumper connectors When finished : Press the Reset button Press and hold the Program button and press the IDButton while still holding the Program button ( P should appear in the Status Code Display) Press the Program button twice. Turn the power off Turn the power back on Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

32 System Alert Codes The following is a list of applicable error codes and corrective actions: 7) System Error A power supply or other system hardware fault was detected. To correct, try pressing the Test button. If the alert continues, there is a hardware malfunction. 6) NV Memory Error - The NV memory was found to be corrupt at power-up, power return or Reset. The monitor has come-up with its defaults active. To correct, try turning the power off and on, if the alert disappears, the power may have comeup in an unstable fashion. If the alert continues, press the CPU Reset button, re-program all functions and modules, exit the programming mode and then turn the power off and on. If the alert disappears, there may have been a bad save to the NV-memory or a power brown-out at the time of the last save. If the alert continues, the NV-memory chip or other hardware is defective. System Alert Codes 4) RS485 Communications Error An error has occurred while communicating with other devices on the network. To correct, press the Test button to clear the alert. If the alert repeats, check if other monitors on the network are having the same problem. If yes, there is a problem with the network wiring or alarm network device. If the other devices are not having problems and the alert keeps reoccurring, there is a hardware problem with the communications circuits of this monitor. 0) Key Board Error An error has occurred while reading the key pad data. To correct, press the Silence button. If the alert keeps reoccurring, there is a hardware problem with this monitor. 5) Configuration Error An error was made while configuring the modules and/or transducers or the monitors hardware configuration is intentionally being changed. This error will also appear if a module malfunctions. (See Monitor Configuration below). Anytime Err is displayed this means that an error was made when configuring that gas module. Tri-Tech, 810 Center Road, Bldg.-E, Avon, Ohio

DC Series Area, Master and Combination Alarm Systems Opreator s Manual

DC Series Area, Master and Combination Alarm Systems Opreator s Manual Opreator s Manual Ver. 2012 - Rev. 1004 Class 1 Inc. design manufacture installation service world class innovation Call us for your next project. Introduction The Tri-Tech Medical gas alarm system monitors

More information

Installation & Operating Instructions for Med Touch Series Area, Master & Combination Alarm Systems

Installation & Operating Instructions for Med Touch Series Area, Master & Combination Alarm Systems Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for Med Touch Series Area, Master & Combination Alarm Systems Master Alarm Area Alarm Andersen Medical Gas 12 Place

More information

Installation & Operating Instructions for Integrated Series Area Alarm Zone Valve Box Systems

Installation & Operating Instructions for Integrated Series Area Alarm Zone Valve Box Systems Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for Integrated Series Area Alarm Zone Valve Box Systems e5536rir 6/15/12 Tri-Tech, 35401 Avon Commerce Pkwy., Avon,

More information

Zone Valve Alarm Combination Systems Installation and Operating Manual

Zone Valve Alarm Combination Systems Installation and Operating Manual Zone Valve Alarm Combination Systems Installation and Operating Manual Ver. 2012 - Rev. 1004 Class 1 Inc. design manufacture installation service world class innovation Call us for your next project. Table

More information

Installation & Operating Instructions for ZT and EZT Integra Touch Series Area Alarm Zone Valve Box Systems

Installation & Operating Instructions for ZT and EZT Integra Touch Series Area Alarm Zone Valve Box Systems Manufacturer of Medical Gas Pipeline Equipment Installation & Operating Instructions for ZT and EZT Integra Touch Series Area Alarm Zone Valve Box Systems e6565rc 05/11/18, 35401 Avon Commerce Pkwy., Avon,

More information

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Submittal Data Sheet Project Information Project Number Approval Features The Tri-Tech Area and Master Alarm Panel digitally displays gas pressure (1 psi increments) monitors and displays normal and alarm

More information

Med Touch Combination Alarm Conversion Kits

Med Touch Combination Alarm Conversion Kits Submittal Data Sheet Features The Powerex Combination Alarm Panel conversion kits are designed to upgrade or retro-fit existing panels produced by several major brands. The conversion kit replaces all

More information

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg. WARD OXYGEN Installation, Operation, and Maintenance Instructions MEGA medical electronic gas alarm MEGA WARD MEGA medical electronic gas alarm WARD POWER SUPPLY Medical Electronic Gas Alarm OXYGEN SYSTEM

More information

Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits

Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits Manufacturer of Medical Gas Pipeline Equipment Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits Master Alarm Conversion Combination Alarm

More information

EZT Series Integra Med Touch Area Alarm & Zone Valve Box with EZ Backfeed

EZT Series Integra Med Touch Area Alarm & Zone Valve Box with EZ Backfeed Submittal Data Sheet Features The Powerex Med Touch area alarm zone valve panel digitally displays gas pressure (1 psi increments) and provides alarm conditions as required by the latest edition of NFPA

More information

Installation, Operation, and Maintenance Instructions

Installation, Operation, and Maintenance Instructions Installation, Operation, and Maintenance Instructions MEGA 2 Medical Electronic Gas Alarm Andersen Medical Gas 12 Place Lafitte Madisonville, LA 70447 http://www.themedicalgas.com 1-866-288-3783 MAN 01-027

More information

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Submittal Data Sheet Project Information Project Number Approval Features The Master Alarm Panel conversion kits are designed to upgrade or retro-fit existing panels produced by several major brands. The

More information

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm Operating & Maintenance Manual Alert-4 Ethernet LCD Master Alarm w w w. a m i c o. c o m Contents User Responsibility 4 Introduction 4 Features 5 Description of the Alarm 5 Shipment Details 5 The Alarm

More information

Med Touch Master Alarm

Med Touch Master Alarm Submittal Data Sheet Features The Powerex Med Touch Master Alarm Panel monitors and displays normal and alarm conditions from up to 128 remote medical gas source signals and provides alarm conditions as

More information

Alert-3 LCD Alarm v2.0

Alert-3 LCD Alarm v2.0 Operating & Maintenance Manual Alert-3 LCD Alarm v2.0 1ST FLOOR ROOM 12 2ND FLOOR ROOM 25 3RD FLOOR ROOM 37 MAINTEN EXT 302 4TH FLOOR ROOM 47 5TH FLOOR ROOM 58 w w w. a m i c o. c o m Contents User Responsibility

More information

Installation Instructions for Med Touch Master Alarms - Wireless

Installation Instructions for Med Touch Master Alarms - Wireless Installation Instructions for Med Touch Master Alarms - Wireless Master Alarm Panel Local Sending Master Alarm Panel - Master Alarm Panel - Repeater Page 1 of 7 Common Master Alarm Back Box Assembly Assembling

More information

Pioneer-R16 Gas Monitor Operator s Manual

Pioneer-R16 Gas Monitor Operator s Manual Pioneer-R16 Gas Monitor Operator s Manual Edition 7/2/97 RKI INSTRUMENTS, INC RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587 (510) 441-5656 Chapter 1: Description About the Pioneer-R16 Gas

More information

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

ZX1e ZX2e ZX5e. Document No Issue 01 user manual ZX1e ZX2e ZX5e Document No. 996-130 Issue 01 user manual MORLEY-IAS ZX2E/ZX5E Fire Alarm Control Panels Table of Contents 1 INTRODUCTION... 4 1.1 NOTICE... 4 1.2 WARNINGS AND CAUTIONS... 4 1.3 NATIONAL

More information

Product Manual SZ1145

Product Manual SZ1145 Product Manual SZ114 General Purpose Monitor Communicating Controls Description The SZ114 is a microprocessor-based monitoring and alarm interface designed to monitor up to four 1000 Ω platinum temperature

More information

AXS Strikemaster Engineering Manual

AXS Strikemaster Engineering Manual AXS Strikemaster AXS Strikemaster Programming and Operation Manual Jan 2014 Page 1 of 23 Table of Contents Important Note... 3 Features and Benefits... 3 Ordering Details... 4 Dimensions... 5 Installation...

More information

User Manual. Dryer Controller M720

User Manual. Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 1.00 Software version 1.00 Preliminary version Manual M720 Dryer controller Page 1 of 42 Document history Preliminary version: - Created in April, 2009

More information

Integra Med Touch Series LCD Area Alarm & Zone Valve Box Systems

Integra Med Touch Series LCD Area Alarm & Zone Valve Box Systems Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product described. Protect yourself and others by observing all safety information.

More information

Dryer Controller M720

Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 2.00 Software version 2.00 Manual M720 Dryer controller Page 1 of 60 Document history Preliminary version: - Created in April, 2009 Hardware Version 2.00,

More information

Caliber Series LED Edgelit Exit Sign EXPORT AC Only or Emergency Operation READ AND FOLLOW ALL SAFETY INSTRUCTIONS

Caliber Series LED Edgelit Exit Sign EXPORT AC Only or Emergency Operation READ AND FOLLOW ALL SAFETY INSTRUCTIONS Caliber Series LED Edgelit Exit Sign EXPORT AC Only or Emergency Operation INSTALLATION AND OPERATING INSTRUCTIONS IMPORTANT SAFEGUARDS When using electrical equipment, basic safety precautions should

More information

Product Manual SZ1144

Product Manual SZ1144 Product Manual SZ1144 Refrigeration Temperature Monitor Communicating Controls Description The SZ1144 is a microprocessor-based monitoring and alarm interface designed to monitor up to four 1000 Ω platinum

More information

CORNELL Emergency Response Systems

CORNELL Emergency Response Systems CORNELL Emergency Response Systems Door Monitor Systems Series 1000 CORNELL Communications, Inc. Milwaukee, WI USA 800-558-8957 - www.cornell.com rev 6/04 A-1000 SERIES DOOR MONITOR OPERATION AND WIRING

More information

Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions

Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions Bulletin E-97DC Thank you for purchasing the Dwyer DCT1000DC Dust Collector Timer Controller. You

More information

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd. F-2000 PC and AO OUTPUT BOARDS INSTRUCTION MANUAL Blue-White R Industries, Ltd. 500 Business Drive Huntington Beach, CA 92649 USA Phone: 714-89-8529 FAX: 714-894-9492 E mail: sales@blue-white.com or techsupport@blue-white.com

More information

Installation and Maintenance Manual. Alarm Valve Combo Unit Alert-1 v1.6

Installation and Maintenance Manual. Alarm Valve Combo Unit Alert-1 v1.6 Installation and Maintenance Manual Alarm Valve Combo Unit Alert-1 v1.6 Table of Contents User Responsibility 4 Introduction 5 Features 5 Installation of Valve Boxes 6 Maintenance and Annual Test 7 Description

More information

Contents. Glossary

Contents. Glossary Contents Glossary ------------------------------------------------------------------------------------------------------ 6 1. Introduction to the IDS 1632 -------------------------------------------------------------

More information

NX-148 LCD CODE PAD TABLE OF CONTENTS

NX-148 LCD CODE PAD TABLE OF CONTENTS NX-148 LCD CODE PAD TABLE OF CONTENTS Glossary Of Terms... 4 Understanding The Lights... 5 Code Pad Functions Arming In The ON Mode... 6 Making The System Ready To Arm... 7 Using Quick Arm... 7 Arming

More information

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08

GasScanner 8C. Eight Channel Monitor. Operator s Manual. MINT-0281-XX Rev. A 01/29/08 GasScanner 8C Eight Channel Monitor Operator s Manual MINT-0281-XX Rev. A 01/29/08 Product Warranty Matheson Tri-Gas, Inc., warrants gas alarm equipment sold by us to be free from defects in materials,

More information

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08 Beacon 200 Gas Monitor Operator s Manual Part Number: 71-2102RK Released: 6/6/08 Table of Contents Chapter 1: Introduction.................................................3 Overview.............................................................3

More information

Digiplex LED Keypads User s Manual

Digiplex LED Keypads User s Manual KLEDEU03.fm Page -1 Friday, May 4, 2001 11:25 AM Digiplex LED Keypads User s Manual KLEDEU03.fm Page 0 Friday, May 4, 2001 11:25 AM KLEDEU03.fm Page 1 Friday, May 4, 2001 11:25 AM TABLE OF CONTENTS 1.0

More information

Contents. Glossary Introduction to the IDS Notes Understanding the Keypad Indicators Operation of the Keypad...

Contents. Glossary Introduction to the IDS Notes Understanding the Keypad Indicators Operation of the Keypad... 2 Contents Glossary...7 1. Introduction to the IDS805...8 1.1 Notes...8 2. Understanding the Keypad Indicators...8 3. Operation of the Keypad...9 4. System Information...10 4.1 Programmed Functions...10

More information

The most user friendly Security Alarm System L S Section 1 Overview of System Section 2 Planning your Installation

The most user friendly Security Alarm System L S Section 1 Overview of System Section 2 Planning your Installation The most user friendly Contents Section 1 Overview of System 1.1 Kit Contents 1.2 Tools Required 1.3 System Features Security Alarm System L S 4 0 0 Section 2 Planning your Installation 2.1 Location of

More information

SPECIFICATION: AlarmLine TM INTELLIGENT LINEAR HEAT DETECTION

SPECIFICATION: AlarmLine TM INTELLIGENT LINEAR HEAT DETECTION 1. GENERAL SPECIFICATION 1.1 The contractor shall provide an Intelligent AlarmLine Linear Heat Detection System to perform the following functionality: A. Fire Alarm, Supervisory and Trouble-Event Initiation

More information

Touchpad Exit Controller Administration Guide

Touchpad Exit Controller Administration Guide Touchpad Exit Controller Administration Guide 2018 RF Technologies, Inc. All specifications subject to change without notice. All Rights Reserved. No Part of this work may be reproduced or copied in any

More information

OPERATION & INSTALLATION MANUAL FOR ALARM PANEL M2AP01

OPERATION & INSTALLATION MANUAL FOR ALARM PANEL M2AP01 OPERATION & INSTALLATION MANUAL FOR ALARM PANEL M2AP01 Table of Contents Safety Instructions 4 Owner/Operator Responsibility 4 Specifications 5 Introduction 6 Installation Instructions 7 Setup 7 Wiring

More information

IDS800 USER MANUAL. Summary of Operation. + [ ] 2 IDS800 USER MANUAL NO K ISSUED APR 2003 VER 1.44

IDS800 USER MANUAL. Summary of Operation.   + [ ] 2 IDS800 USER MANUAL NO K ISSUED APR 2003 VER 1.44 Summary of Operation A rm/ disarm [#] + [USER CODE] Quick Quick Quick Away Arm Stay Arm Stay Arm & Go H old down [ 1] for 1 second H old down [ 5] for 1 second H old down [ 6] for 1 second Panic Fire Medical

More information

MODEL 5100 VOTING LOGIC MODULE

MODEL 5100 VOTING LOGIC MODULE DESCRIPTION DESCRIPTION The SST Model 5100 Multi-Input Voting Logic Module is used to monitor the status of up to 14 different points in a hazard zone, and report when a selectable number of these points

More information

Watchguard WGAP864 User Manual

Watchguard WGAP864 User Manual Watchguard WGAP864 User Manual v1.0 Issued September 2016 1 2 Table of Contents Glossary... 5 1. Introduction to your Watchguard WGAP864... 6 2. Before Operating your Alarm System... 6 3. Understanding

More information

Technical Specification

Technical Specification Technical Specification Surface Area Valve Service Unit Module Semi Flush Area Valve Service Unit Module Product Description The CPX AVSU Module is used in conjunction with area valve service units (AVSU).

More information

IDS S E C U R I T Y IDS816. User Manual. MANUAL NO A ISSUED November 2004 VERSION 1.00

IDS S E C U R I T Y IDS816.  User Manual. MANUAL NO A ISSUED November 2004 VERSION 1.00 INHEP DIGITAL IDS S E C U R I T Y IDS816 User Manual MANUAL NO. 700-283-02A ISSUED November 2004 VERSION 1.00 Contents 1. Introduction to the IDS816... 4 2. Understanding the Keypad Indicators... 4 3.

More information

OHIO MEDICAL CORPORATION. Medical Gas Alarms. Installation and Maintenance Manual. Area Alarm. Master Alarm. Combo Alarm. P/N (Rev.

OHIO MEDICAL CORPORATION. Medical Gas Alarms. Installation and Maintenance Manual. Area Alarm. Master Alarm. Combo Alarm. P/N (Rev. OHIO MEDICAL CORPORATION Medical Gas Alarms Installation and Maintenance Manual Area Alarm Master Alarm Combo Alarm P/N 255098 (Rev.8) 07/08 Page 2 NOTES Andersen Medical Gas 12 Place Lafitte Madisonville,

More information

Crypto-Lock. Model CC-8521B Updated 7/17. Access Control System. MONITEQ, Inc. INSTRUCTION MANUAL

Crypto-Lock. Model CC-8521B Updated 7/17. Access Control System. MONITEQ, Inc. INSTRUCTION MANUAL Crypto-Lock Model CC-8521B Updated 7/17 Access Control System INSTRUCTION MANUAL MONITEQ, Inc. 2 TABLE OF CONTENTS 1. INTRODUCTION..4 2. SPECIFICATIONS.....6 3. SUPPLIED EQUIPMENT.. 6 4. FUNCTIONS OF CONTROLS

More information

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11 Table of Contents 1. OVERVIEW... 3 2. SYSTEM LAYOUT... 4 3. SPECIFICATIONS... 8 3.1 SYSTEM COMPONENTS...9 3.2 PLC INPUTS AND OUTPUTS...9 3.3 FUNCTION KEYS...10 3.4 DEFAULT SET POINTS AND TIMERS...10 4.

More information

Carbon Monoxide Transmitter

Carbon Monoxide Transmitter Introduction The CO Transmitter uses an electrochemical sensor to monitor the carbon monoxide level and outputs a field-selectable 4-20 ma or voltage signal. The voltage signal may also be set to 0-5 or

More information

Understanding the Code Pad lights...4. Code Pad tones...5. Fully arming the system On MODE...6. Fully arming the system - Quick Arm MODE...

Understanding the Code Pad lights...4. Code Pad tones...5. Fully arming the system On MODE...6. Fully arming the system - Quick Arm MODE... TABLE OF CONTENTS...Glossary of terms...2...code Pad Diagram...3 Understanding the Code Pad lights...4 Code Pad tones...5 Fully arming the system On MODE...6 Fully arming the system - Quick Arm MODE...6

More information

CommStat 6. Controller for Redundant HVAC Systems PRODUCT DATA SHEET

CommStat 6. Controller for Redundant HVAC Systems PRODUCT DATA SHEET CommStat 6 Controller for Redundant HVAC Systems PRODUCT DATA SHEET General Description The CommStat 6 HVAC controller is designed for controlling up to six redundant air conditioners in an E-House or

More information

Security System. User Guide for the LED Command Center

Security System. User Guide for the LED Command Center Security System User Guide for the LED Command Center National Security Systems Inc (800)457-1999 MY SECURITY COMPANY IS: CALL BEFORE TEST: THIS SECURITY SYSTEM IS CONNECTED TO TELEPHONE NUMBER: THE SECURITY

More information

Solution 880 Operators Manual. Issue 1.00

Solution 880 Operators Manual. Issue 1.00 Solution 880 Operators Manual Issue 1.00 Solution 880 Operators Manual Copyright 1998 by, SYDNEY, AUSTRALIA Document Part Number MA408O Document ISSUE 1.00 Printed 15 June 1998 This documentation is provided

More information

Installation, Operating and Maintenance Manual

Installation, Operating and Maintenance Manual STATUS ZONES CONTROLS FIRE FAULT DISABLED FIRE 1 2 3 4 5 6 7 8 TEST FAULT DISABLED 1 5 BUZZER SILENCE RESET 1 2 TEST 2 6 LAMP TEST 3 SUPPLY 3 7 SYSTEM FAULT 4 8 SOUNDERS ACTIVATE/ SILENCE 4 FAULTS INSTRUCTIONS

More information

Steamist Installation Guidelines

Steamist Installation Guidelines Steamist Installation Guidelines The Heavy Commercial steam generator (HC-9 thru HC-18) comes from the factory assembled, carefully wired, and tested. Please read all instructions before installing or

More information

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge

Replaceable LED modules. Sleep or unattended mode. Auto-silence and auto-acknowledge Replaceable LED modules 11 Alarm Sequences as per ISA-18.1 standard Each channel/window fully field programmable RS232 or RS485 MODBUS-RTU communication Repeat relay for each window and multifunction relays

More information

Tri-Tech Medical Inc.

Tri-Tech Medical Inc. Check Valves with Extensions Project Information Project Number Approval Hinged double disc check valve Removable center section (3 piece design) allows for easy field replacement (no cutting into pipe

More information

Public Safety DAS Annunciator Panel

Public Safety DAS Annunciator Panel Public Safety DAS Annunciator Panel 120 VAC Models: 1221-A, 1221-B, 1221-C Revision D 91117 48 VDC Models: 1221-A-48, 1221-B-48, 1221-C-48 24 VDC Models: 1221A-24, 1221-B-24, 1221-C-24 CAUTION: (Read This

More information

Flopurge TS. Operation Manual

Flopurge TS. Operation Manual Flopurge TS Operation Manual Part Number 079-0204 Spectron Gas Control Systems United Kingdom Unit 4, Herald Court, University of Warwick Science Park, Coventry, CV4 7EZ +44 (0)24 7641 6234 sales@spectron-gcs.com

More information

Beacon 800 Gas Monitor Operator s Manual

Beacon 800 Gas Monitor Operator s Manual Beacon 800 Gas Monitor Operator s Manual Part Number: 71-0037RK Revision: F Released: 4/18/17 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be

More information

Digital Control Unit- Installation, Operation & Maintenance IMPORTANT INFORMATION

Digital Control Unit- Installation, Operation & Maintenance IMPORTANT INFORMATION Environment IMPORTANT INFORMATION It is not anticipated that this equipment will be exposed to adverse environmental conditions without additional protection. Site the equipment in a frost free area. Ensure

More information

CA5000 Liquid CO 2 Freezer Backup System

CA5000 Liquid CO 2 Freezer Backup System CA5000 Liquid CO 2 Freezer Backup System OPERATING INSTRUCTIONS MANUAL HAMPSHIRE CONTROLS CORPORATION ONE GROVE STREET / P.O. BOX 516, DOVER, NEW HAMPSHIRE USA 03821 TEL. (603) 749-9424 FAX (603) 749-9433

More information

EW 40 Wireless Fan Control

EW 40 Wireless Fan Control Installation & Operating Manual EW 40 Wireless Fan Control USA CAN Product Information... Chapters 1 + 2 Mechanical Installation... Chapter 3 Electrical Installation... Chapter 4 Start Up and Configuration...

More information

SEC 2000 Millenium Infrared Gas Detector

SEC 2000 Millenium Infrared Gas Detector SEC 2000 Millenium Infrared Gas Detector Instruction and Operation Manual Sensor Electronics Corporation 5500 Lincoln Drive Minneapolis, Minnesota 55436 USA (952) 938-9486 Fax (952) 938-9617 Email: sales@sensorelectronic.com

More information

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm

Operating & Maintenance Manual. Alert-4 Ethernet LCD Master Alarm Operating & Maintenance Manual Alert-4 Ethernet LCD Master Alarm Contents User Responsibility 4 Introduction 4 Features 5 Description of the Alarm 5 Shipment Details 5 The Alarm Back Box 5 The Frame/Module

More information

IDS816 User Manual H Issued January 2009

IDS816 User Manual H Issued January 2009 1 Contents Glossary-------------------------------------------------------------------------------------------------------------------6 1. Introduction to the IDS 816---------------------------------------------------------------------------7

More information

G34 AU1B (External Ultrasonic Sensor) Version 3

G34 AU1B (External Ultrasonic Sensor) Version 3 Car Alarm Series 3 B 4 Buttons G34 AU1B (External Ultrasonic Sensor) Version 3 24 CAR ALARM GENIUS Series 3B 4 Buttons G34 AU1B (External Ultrasonic Sensor) Module controlled using Micro-Processor 2 Transmitters

More information

Refer to Bulletin E-1101 for detailed information on the FLAME-MONITOR System.

Refer to Bulletin E-1101 for detailed information on the FLAME-MONITOR System. The Fireye EP260, EP270 (early spark termination), or EP265 (pilot stabilization) programmer modules are used with the FLAME-MONITOR Burner Management Control System (P/N E110). Several operational characteristics

More information

Security System. User s Guide for the Text Command Center

Security System. User s Guide for the Text Command Center User s Guide for the Text Command Center MY ALARM COMPANY IS: CALL BEFORE TEST: THIS SECURITY SYSTEM IS CONNECTED TO TELEPHONE NUMBER: THE SECURITY CONTROL PANEL IS CONNECTED TO THE PHONE JACK LOCATED:

More information

RANGER 8600 DOWNLOADABLE CONTROL COMMUNICATOR INSTALLATION MANUAL

RANGER 8600 DOWNLOADABLE CONTROL COMMUNICATOR INSTALLATION MANUAL RANGER 8600 DOWNLOADABLE CONTROL COMMUNICATOR INSTALLATION MANUAL TABLE OF CONTENTS GENERAL DESCRIPTION... 2 STANDARD AND OPTIONAL PARTS LIST... 2 PARTS DIAGRAM... 3 TERMINAL DRAWING AND SPECIAL NOTES...

More information

Intruder alarm panel Terxon MX Operating instructions

Intruder alarm panel Terxon MX Operating instructions Intruder alarm panel Terxon MX Operating instructions Operating instructions Perfect Security for home and office These operating instructions are an important product accessory. They contain important

More information

Alert-3 LCD Alarm v2.5

Alert-3 LCD Alarm v2.5 Operating & Maintenance Manual Alert-3 LCD Alarm v2.5 1ST FLOOR ROOM 12 2ND FLOOR ROOM 25 3RD FLOOR ROOM 37 MAINTEN EXT 302 4TH FLOOR ROOM 47 5TH FLOOR ROOM 58 Contents User Responsibility 4 Introduction

More information

Models NFPA 1221-A, NFPA 1221-B Public Safety DAS Annunciator Panel. Revision E 61117

Models NFPA 1221-A, NFPA 1221-B Public Safety DAS Annunciator Panel. Revision E 61117 Models NFPA 1221-A, NFPA 1221-B Public Safety DAS Annunciator Panel Revision E 61117 CAUTION: (Read This First) This panel has been designed to make it nearly bullet proof to mistakes made when wiring

More information

U ser's Guide PC6010

U ser's Guide PC6010 User's Guide PC6010 Quick Reference Guide This manual is for Basic and Advanced users. Each of these types of user can access a different set of functions. The and symbols next to the title of each procedure

More information

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family

NexSysLink. 2 CAN Display Operation Manual. CAN Instruments Product Family NexSysLink CAN Instruments Product Family 2 CAN Display Operation Manual Contact Beede Beede Electrical Instrument Company, Inc. 88 Village Street Penacook, NH 03303 (603) 753-6362 Toll-free 800-451-8255

More information

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS. HZM-Series-Class I, Div 2 Battery Unit. Instalation instructions

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS. HZM-Series-Class I, Div 2 Battery Unit. Instalation instructions HZM-Series-Class I, Div 2 Battery Unit HZM-Series-Class I, Div 2 Battery Unit Instalation instructions IMPORTANT SAFEGUARDS 1 4 5 6 7 13 When using electrical equipment, basic safety precautions should

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS The SensorNet I/O Module provides attachment for up to 32 alarms and up to 16 switched devices. An RS-232 or RS-422 communications port is also provided for receiving serial alarms.

More information

The Information Contained in this Document is Proprietary and should only be used for Service or Training of Authorized Blodgett Servicers who will

The Information Contained in this Document is Proprietary and should only be used for Service or Training of Authorized Blodgett Servicers who will The formation Contained in this Document is Proprietary and should only be used for Service or Training of Authorized Blodgett Servicers who will be working on the Blodgett XR8 Oven. All other uses are

More information

IDS S E C U R I T Y IDS816. User Manual MANUAL NO B ISSUED DEC 2004 VERSION 2.00

IDS S E C U R I T Y IDS816. User Manual MANUAL NO B ISSUED DEC 2004 VERSION 2.00 INHEP DIGITAL IDS S E C U R I T Y IDS816 User Manual MANUAL NO. 700-283-01 B ISSUED DEC 2004 VERSION 2.00 Contents 1. Introduction to the IDS816... 4 2. Understanding the Keypad Indicators... 4 3. Programmable

More information

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS. NXM Series - Survive-All. Emergency Light Unit

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS. NXM Series - Survive-All. Emergency Light Unit Emergency Light Unit IMPORTANT SAFEGUARDS When using electrical equipment, basic safety precautions should always be followed including the following: READ AND FOLLOW ALL SAFETY INSTRUCTIONS 1. Do not

More information

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA 99336 www.jmcvr.com 1-509-586-9893 Table of Contents TABLE OF FIGURES...1 OVERVIEW OF THE HRC CAPABILITIES...2 INSTALLATION AND

More information

Lubrication cycle IP Enclosure Rating. 110 VAC, 220/230 VAC (50/60 Hz) Alarm Fault Relay Contacts. IP-55 (Liquid tight connector)

Lubrication cycle IP Enclosure Rating. 110 VAC, 220/230 VAC (50/60 Hz) Alarm Fault Relay Contacts. IP-55 (Liquid tight connector) SMAC Controller Industrial Lubrication Systems Operation The SMAC Controller is a multi-purpose programmable controller used with industrial lubrication systems. Controller settings are saved whenever

More information

EXTINGUISHING AGENT RELEASE MODULE

EXTINGUISHING AGENT RELEASE MODULE EXTINGUISHING AGENT RELEASE MODULE Operation, Installation & Programming Manual Revision 3.00 Distributors For: 18-20 Brookhollow Ave telephone 02 8850 2888 www.firesense.com.au Baulkham Hills NSW 2153

More information

ACCURATE ELECTRONICS INC PO BOX SW HALL BLVD BEAVERTON OR USA FAX

ACCURATE ELECTRONICS INC PO BOX SW HALL BLVD BEAVERTON OR USA FAX Page 1 of 10 Model 10807800 January 2014 ACCURATE ELECTRONICS INC PO BOX 1654 97075-1654 8687 SW HALL BLVD 97008 BEAVERTON OR USA 503.641.0118 FAX 503.646.3903 WWW.ACCURATE.ORG Practice Section 10807800

More information

INSTALLATION GUIDE Dual Fuel Ranges

INSTALLATION GUIDE Dual Fuel Ranges INSTALLATION GUIDE Dual Fuel Ranges Contents Wolf Dual Fuel Ranges......................... 3 Safety Instructions............................ 4 Dual Fuel Range Specifications.................. 5 Dual Fuel

More information

Installation Instructions. Fire Command Center/Fire Alarm Annunciators D1256/D1257

Installation Instructions. Fire Command Center/Fire Alarm Annunciators D1256/D1257 Instructions Fire Command Center/Fire Alarm Annunciators D1256/D1257 74-06925-000-H Page 2 2004 Bosch Security Systems Contents 1.0 Introduction...5 1.1 Before You Begin...5 1.2 Type Styles Used Here...5

More information

Operating instructions 2-channel code lock. Item No

Operating instructions 2-channel code lock. Item No Operating instructions 2-channel code lock Item No. 1560346 Table of contents Page 1. Introduction...4 2. Explanation of symbols...4 3. Intended use...5 4. Delivery content...5 5. Safety instructions...6

More information

Security System With Scheduling. User Guide. N5943-8V4 7/04 Rev A

Security System With Scheduling. User Guide. N5943-8V4 7/04 Rev A ADEMCO VISTA-120 Security System With Scheduling User Guide N5943-8V4 7/04 Rev A Your Honeywell security system is designed for use with devices manufactured or approved by Honeywell for use with your

More information

INTELLIGENT FIRE TECHNOLOGY. Twinflex and Multipoint V3. User Guide (TO BE RETAINED BY USER) Issue 3

INTELLIGENT FIRE TECHNOLOGY. Twinflex and Multipoint V3. User Guide (TO BE RETAINED BY USER) Issue 3 INTELLIGENT FIRE TECHNOLOGY Twinflex and Multipoint V3 User Guide (TO BE RETAINED BY USER) 26-0340 Issue 3 Rafiki Protection Limited Rafiki policy is one of continual improvement and the right to change

More information

ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS

ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS II 1400-3 ACSI MODEL 1410 POWER SUPPLY INSTALLATION INSTRUCTIONS Features: For use with doors or groups of doors that must interlock with each other to regulate control of accessing one area from another,

More information

Protégé Eclipse LED Keypad User Manual PRT-KLES

Protégé Eclipse LED Keypad User Manual PRT-KLES Protégé Eclipse LED Keypad User Manual PRT-KLES The specifications and descriptions of products and services contained in this manual were correct at the time of printing. Integrated Control Technology

More information

D1255RB/D1256RB/D1257RB

D1255RB/D1256RB/D1257RB D1255RB/D1256RB/D1257RB EN Installation Instructions Fire Keypads and Fire Alarm Annunciator D1255RB/D1256RB/D1257RB Installation Instructions Listings and Approvals Listings and Approvals UL 365 UL 609

More information

PNC 1000 SERIES 2, 4, 8 Zone Fire Alarm Control Panel

PNC 1000 SERIES 2, 4, 8 Zone Fire Alarm Control Panel PNC 1000 SERIES 2, 4, 8 Zone Fire Alarm Control Panel INSTALLATION, OPERATION AND MAINTENANCE MANUAL Version: CN-PM-1000.VER1.1-12/2012 EN54 INFORMATION In accordance with EN 54-2 clause 13.7, the maximum

More information

Installation Instructions

Installation Instructions 8GXM / 0GXM Multi---Split High---Wall Duct Free Split System 8GXM --- Size 18k, k, and 0k 0GXM --- Size 9k, 1k, and 18k Installation Instructions NOTE: Read the entire instruction manual before starting

More information

Programmable Thermostat

Programmable Thermostat Set & $ave Programmable Thermostat Installation and Operation Manual English Model 44360 Form# 42710-01 20091204 2009 Hunter Fan Co. 2 Table of Contents Important Information... 5 Tools... 6 Uninstalling

More information

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS. 3LERHZ Series Class I, Div 2 Combo Unit. Installation instructions

IMPORTANT SAFEGUARDS READ AND FOLLOW ALL SAFETY INSTRUCTIONS. 3LERHZ Series Class I, Div 2 Combo Unit. Installation instructions 3LERHZ Series Class I, Div 2 Combo Unit Installation instructions 3LERHZ Series Class I, Div 2 Combo Unit IMPORTANT SAFEGUARDS 1 3 When using electrical equipment, basic safety precautions should always

More information

Macurco DVP-120 Detection and Ventilation Control Panel

Macurco DVP-120 Detection and Ventilation Control Panel Macurco DVP-120 Detection and Ventilation Control Panel User Instructions IMPORTANT: Keep these User Instructions for reference. TABLE OF CONTENTS 1 Introduction 4 1.1 General Information 4 1.2 Features

More information

TABLE OF CONTENTS GENERAL... 2 A. DEFINITIONS DIGITAL, ADDRESSABLE FIRE ALARM SYSTEM... 2

TABLE OF CONTENTS GENERAL... 2 A. DEFINITIONS DIGITAL, ADDRESSABLE FIRE ALARM SYSTEM... 2 Division 28 ELECTRONIC SAFETY AND SECURITY TABLE OF CONTENTS 28 0000 GENERAL... 2 A. DEFINITIONS...2 28 3111 DIGITAL, ADDRESSABLE FIRE ALARM SYSTEM... 2 A. DESIGN REQUIREMENTS...2 B. CONSTRUCTION REQUIREMENTS...2

More information

BURGLAR ALARM PANEL BS-468

BURGLAR ALARM PANEL BS-468 BURGLAR ALARM PANEL BS-468 Contents 1. Description... 3 2. Instructions for the user... 4 2.1Basic operations... 4 Complete system.... 4 Split system.... 4 2.2 Armed system indication... 5 2.3 Advanced

More information

Matrix2000. User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 Zones. Version: 2.0 Revision: 1

Matrix2000. User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 Zones. Version: 2.0 Revision: 1 Matrix2000 User Guide - CPD Advanced Analog Fire Alarm Control Panel 4 24 s Version: 2.0 Revision: 1 IMPORTANT INFORMATION Limitation of liability It is mandatory, Matrix2000 panel to be installed in accordance

More information

IDS S E C U R I T Y IDS816. User Manual MANUAL NO C ISSUED APRIL 2005 VERSION 2.00

IDS S E C U R I T Y IDS816. User Manual MANUAL NO C ISSUED APRIL 2005 VERSION 2.00 INHEP DIGITAL IDS S E C U R I T Y IDS816 User Manual MANUAL NO. 700-283-01C ISSUED APRIL 2005 VERSION 2.00 Contents 1. Introduction to the IDS816... 4 2. Understanding the Keypad Indicators... 4 3. Programmable

More information