1 CNAP MONITOR 500 HD QUICK GUIDE This Quick Guide DOES NOT REPLACE the user manual! Version 07/2016 V1.5.5 EN

2 2 CONTENT SET UP MEASUREMENT NAVIGATION FAQs FUNCTIONS: FAST ACCESS

3 Hemodynamic Parameters...5 Set Up Checklist for fast Set Up...6 Power Up CNAP Monitor... 7 Prepare CNAP Monitor for Measurement - Components...8 Prepare CNAP Monitor for Measurement - Connections...9 Connecting to a Patient Monitor...10 Zeroing of CNAP and patient monitor...11 Overview CNAP Monitor Patient Set Up...12 Choose correct CNAP finger sensor size...13 Place CNAP finger sensor and CNAP Controller...14 Place NBP Cuff...15 Measurement Select Measurement Mode / Start Measurement...16 Started Measurement: Perfusion Index / Signal Quality...17 Choose Patient Category / Enter Biometric Data...18 Stop / Resume Measurement...19 Navigation Navigation Frames...20 Function of Navigation Frames

4 4 Functions - Fast Access CONTENT Main Keys and Status Indicators The SETUP Key...25 Calibration Modes: Function / Benefits Triggering A Manual NBP - Fast Access...27 When is it recommended to trigger a manual NBP?...28 Change Finger manually...29 FAQs: 1. How can I change alarm settings for current measurement? How can I change alarm settings permanently? Enabling hemodynamic parameters for a measurement? How can I improve measurement quality? How can I adjust trend, signal and time settings? The battery? Service intervals for the CNAP Monitor? What to do if CNAP needs service? SERVICE HOTLINE? Where can I get additional information? How can I contact CNSystems directly?...41

5 HEMODYNAMIC PARAMETERS CNAP Monitor Measurement Type BP BP + PPV BP + HD CNAP NBP continuous noninvasive arterial pressure, sys, dia, mean noninvasive oscillometric blood pressure sys, dia x x x x x x Pulse pulse rate x x x PPV pulse pressure variation x x SVV stroke volume variation x CO continuous noninvasive cardiac output x CI cardiac index x SV stroke volume x SVI stroke volume index x SVR systemic vascular resistance x SVRI systemic vascular resistance index x 5

6 SET UP 6 SET UP CHECKLIST CHECKLIST STEP 1 STEP 2 STEP 3 STEP 4 STEP 5 STEP 6 STEP 7 Start the device and connect the CNAP Controller Cable (orange connectors) to CNAP Monitor and CNAP Controller (see also p.7,9) Use the indicator on the CNAP Controller to find the correct CNAP finger sensor size for the patient. (see p. 13) Slide the finger sensor over two neighbouring fingers, and fix the CNAP Controller on the forearm using the Velcro fixation cuff. (see p. 14) Be sure that the CNAP Controller is in the center of the forearm cuff so that the patient can freely bend his / her wrist. (see p.14) Select the right NBP cuff size for the patient using the indicator on the inside of the NBP cuff and connect the black hose to the CNAP Monitor. (see p.15) Fit the NBP cuff on the patient s upper arm. The NBP cuff can be placed ipsilaterally or contralaterally to the CNAP finger sensor. (see p.15) Start measurement by clicking Start New Measurement. Choose patient category Adult or Pediatric. (see p.16, 18)! ATTENTION During CNAP initialisation the patient needs to be in a stable position. During general anesthesia start CNAP before or after the induction phase!

7 POWER UP CNAP MONITOR 1 Press POWER ON / OFF button 2 When alarm sounds, confirm by pressing the yellow alarm button. SET UP 7

8 SET UP 8 PREPARE CNAP MONITOR FOR MEASUREMENT COMPONENTS 1 CNAP Monitor 1 2 CNAP Controller Cable 3 CNAP Controller CNAP Finger sensor NBP Cuff 4 6 NBP Air hose

9 PREPARE CNAP MONITOR FOR MEASUREMENT CONNECTIONS 1 Connect the controller cable to the orange connector of the CNAP Controller. 3 Connect NBP air hose to NBP Cuff 4 2 Connect finger sensor to black connection of CNAP Controller.! ATTENTION Plug in straight, not tilted! Connect the NBP air hose and the CNAP Controller cable to the CNAP Monitor. SET UP 9

10 SET UP 10 CONNECTING TO A PATIENT MONITOR Interface cable 1 Prepare patient monitor Interface Cable and connect it to the arterial port of your patient monitor. 2 Transducer Cable Select right color from list in user manual (Chapter 4-5-2) and connect it to the BP wave out port on the CNAP Monitor. 3 connect to Connect Transducer cable and patient monitor interface cable.! ATTENTION Zeroing neccessary! (see next page) BP Wave Out port

11 ZEROING OF CNAP AND PATIENT MONITOR 1 Go to parameter frame on CNAP Monitor: *click* 2 Select IBP: Zeroing Start - *click* then zero transducer on patient monitor *click* *click* Zeroing of patient monitor *click* 3 Stop Zeroing on CNAP : Select IBP: Zeroing Stop - *click* TIP CNAP Zeroing is automatically active when the CNAP Monitor is on and connected to the patient monitor but no measurement has been started so far. SET UP 11

12 SET UP 12 OVERVIEW: PATIENT SET UP Connections NBP Cuff CNAP Controller CNAP Finger sensor

13 STEP 1. CHOOSE CORRECT CNAP FINGER SENSOR SIZE AVAILABLE SIZES: small, medium or large 1 B2 Choose the sensor size indicated on the shortest bar which is still visible. Put proximal phalanx of index finger on indicator of CNAP Controller align with the grey line on the left. 2! ATTENTION Always choose the larger sensor if you are in doubt about the size! SET UP 13

14 SET UP 14 STEP 2. PLACE CNAP FINGER SENSOR AND CNAP CONTROLLER! ATTENTION 1 Slide the finger sensor over two neighbouring fingers, make sure that it is not placed on the finger joints! 2 slide back fully! Make sure that the CNAP Controller is placed in central position of the slide and fix the strap. NOTE 12 hours before lifetime expiration of the finger sensor a message appears. It automatically disappears and does not affect measurements. Please make sure to contact your local distributor in time for a new sensor. See also: Operators Manual chapter 16.1

15 STEP 3. PLACE NBP CUFF 1 Choose the right NBP cuff from 4 different sizes by using the size indicator on the inside of the cuff. See also: Operators Manual chapters and Fasten NBP cuff on the upper arm (ipsilaterally or contralaterally to arm with fingersensor). Align marker (white arrow) with the brachial artery in the crook of the arm. 3 SET UP 15

16 MEASUREMENT 16 SELECT MEASUREMENT MODE / START MEASUREMENT 1 2 Once the patient set up is completed, select the measurement mode in the pop-up menu from three options: > BP only > BP + PPV > BP + HD (full hemodynamics) Confirm by clicking the wheel. *click* 3 To start the measurement, select Start New Measurement and confirm with the click wheel. NOTE BP + PPV and BP + HD must be enabled to be selectable. Contact your distributor for a licence key if you cannot select these options.

17 STARTED MEASUREMENT 1 Once measurement has started, finger sensor checks perfusion index (PI): TIP no signal good perfusion If perfusion is poor (=red), try pushing the finger sensor to the back of the finger, warming the hand (use exam glove or towel), changing finger sensor size, or changing hands. 2 At the same time menu window opens for choosing patient category and for entering biometric data (see next page). Calibration to NBP reference value also takes place.! ATTENTION Keep patient still, do not move hand! No changes in body posture! MEASUREMENT 17

18 MEASUREMENT 18 CHOOSE PATIENT CATEGORY AND ENTER BIOMETRIC DATA Choose patient category (Adult or Pediatric) in the pop-up menu and enter biometric patient data (for full hemodynamics). NOTE This is important as alarm ranges and NBP inflation pressure are different! Confirm with ok by clicking the wheel. Calculation of hemodynamic parameters starts; display after a few seconds. *click* Before starting any! ATTENTION interventions, make sure that the calibration has been completed: numerical values should not have gray background anymore! *click*

19 STOP / RESUME MEASUREMENT 1 Press the Start/Stop Button to STOP measurement (e.g. when finished or to readjust set-up).! ATTENTION Do not remove the CNAP finger sensor when measurement is still in progress! STOP MEASUREMENT FIRST! 2 RESUME a measurement with the Start/Stop Button (e.g. after readjusting set-up, patient transfer, etc.) Patient data remain displayed! MEASUREMENT 19

20 NAVIGATION 20 NAVIGATION FRAMES The monitor has a clickwheel and 5 different frames on the screen to access all menus / functions. Alarm frame Signal & Trend frame *turn* Signal and Trend View selector frame *click* clickwheel Parameter frame (= default) Hemodynamic frame

21 FUNCTION OF NAVIGATION FRAMES Parameter frame Signal & Trend frame Alarm & Status frame This is the default frame! > manual NBP triggering > manual finger change > finger change interval setting > alarm settings > zeroing of IBP port > NBP measurement type > intervention setting > change of trend view (graphic numeric) > scaling of trend curve > selection of trending parameters (up to 2 additional) > change of time range > display of alarm and status messages > direct access to alarm settings as long as alarm message is visible > display of battery status, date, next finger change, next NBP measurement NAVIGATION 21

22 NAVIGATION 22 FUNCTION OF NAVIGATION FRAMES Signal & Trend View Selector Hemodynamic frame > switch between signal view and parameter trends > activating/ deactivating HD > manual entry or change of patient data > discharging of patient (data of current measurement are deleted for measurement with new patient) > changes between absolute and indexed values > manual entry of external CO value for calibration > selection Fluid Parameter PPV or SVV > manual entry of CVP

23 FAST ACCESS KEYS and STATUS INDICATORS MAIN KEYS FOR FAST ACCESS MAIN SCREEN: return to main screen from any submenu POWER ON / OFF: hold for 2 seconds to power on/off. power on indicator: > green: switched on > orange: system error SETUP: access to configuration menu FUNCTIONS 23

24 FUNCTIONS 24 MAIN KEYS FOR FAST ACCESS ALARM PAUSE/ OFF: alarm functions control: > Press once: alarm acknowledgement > Press twice: set alarm to pause / off > Press three times: re-activate alarm function PRINTING START/ STOP: > press once: print blood pressure curve > press and hold for one second: print trend view START/ STOP: > stop and resume measurement Battery charge status: > green = ok > orange = less than 25% > red = less than15 minutes left!

25 1. Monitor Settings THE SET UP KEY > Change Language / Time / Date > En-/ disable device features: PPV, HD, Record, Analog Out, enter licence keys, etc. > Recording on/off (if activated) > Switch between Metric / US customary units > Access PPU measurement statistics 3. Service Settings For service purposes only! 2. Measurement Settings > NBP Mode: select calibration mode (see next page) > NBP Interval > CNAP Calibr. Interval > Hemodynamic (on/off, change settings) > Display Options: Interventions, Views, Scales,etc. > Trend Parameters: select parameters to be displayed in the trend view > Print options > Parameter averaging > Patient category FUNCTIONS 25

26 FUNCTIONS 26 CALIBRATION MODES Function AUTOMATIC NBP CALIBRATION: INTERNAL (= DEFAULT) Calibration to automatic NBP measurements triggered by the integrated NBP module in adjustable intervals between 5-60 minutes. Can be stopped any time by pressing Start/Stop Button. Benefit This is the default mode delivering absolute values automatically as long as no other blood pressure reference value (NBP or IBP) is entered manually (see also: external calibration mode). Function Function Calibration to a manually entered NBP, IBP or height offset reference value from external devices. Automatic calibration to a heart level value based on an emipric database for bridging the short time until CNAP is calibrated to a measured or externally entered NBP value. MANUAL BP CALIBRATION: EXTERNAL Benefit TRENDING MODE: EMPIRIC CALIBATION Benefit CNAP can be operated without internal NBP measurement calibrating to a manually entered value for absolute BP information. Useful for special applications. Blood pressure curve is displayed only 10 secs after start-up for a quick trending prior to calibration to the directly measured reference value.

27 TRIGGERING A MANUAL NBP NOTE Triggering a manual NBP is possible at ANY TIME in any mode! 1 Go to parameter frame: 2 Choose NBP: Start - *click* *click* *click* FUNCTIONS 27

28 FUNCTIONS 28 WHEN IS IT RECOMMENDED TO TRIGGER A MANUAL NBP? 1. Upon hand movement or patient repositioning Trigger a manual NBP after moving or repositioning the patient or the patient s hand or whenever hydrostratic change/ offset happens.! ATTENTION 2. In situations that quickly require updated calibration After an intervention or whenever you feel that an update to the NBP reading may be beneficial.

29 CHANGE FINGER MANUALLY NOTE Switching a measurement to the other finger is possible at ANY TIME except during a NBP measurement! 1 Go to parameter frame: 2 Choose CNAP: Change Finger - *click* *click* TIP FAST ACCESS: Parameter frame is the default frame. If not changed, just click wheel 2 times to change finger! *click* FUNCTIONS 29

30 FAQs 30 FREQUENTLY ASKED QUESTIONS FAQs: 1. How can I change alarm settings for current measurement? How can I change alarm settings permanently? Enabling hemodynamic parameters for a measurement? How can I improve measurement quality? How can I adjust trend, signal and time settings? The battery? Service intervals for the CNAP Monitor? What to do if CNAP needs service? SERVICE HOTLINE? Where can I get additional information? How can I contact CNSystems directly?...41

31 1. HOW TO CHANGE ALARM SETTINGS: CURRENT MEASUREMENT ALARMS OFF/ON AND CHANGE LIMITS 1 Go to parameter frame with click-wheel and *click* *click* 2 Select Adult alarm limits and turn values to Off or change limits TIP *click* Whenever a physiological alarm is active, the settings can be directly changed by clicking on the alarm frame. FAQs 31

32 2. HOW TO CHANGE ALARM SETTINGS: PERMANENTLY ALARMS OFF/ON AND CHANGE LIMITS FAQs 32 1 Press Setup key. 2 Change the alarm default settings in the submenu Service Alarm Defaults (Password: 8737)

33 3. ENABLING HEMODYNAMIC PARAMETERS FOR A MEASUREMENT 1 If HD is not enabled, hemodynamic frame shows question marks: select with click-wheel - *click* 2 Choose Measurement Mode: BP+HD and confirm with ok. If BP + HD is active, BP + PPV is automatically enabled.! NOTE TIP ATTENTION *click* Hemodynamic parameters visible now! Hemodynamic parameters can also be set permanently in Set Up Measurement Hemodynamic If it is not possible to enable the Hemodynamic parameters, please request a license key from CNSystems or your distributor. See also: Operators Manual chapters 4.4 and 12.2 FAQs 33

34 FAQs HOW TO IMPROVE MEASUREMENT QUALITY RED: no signal can be detected Reasons / limitations: > poor peripheral perfusion (shock, severe Raynaud s, hypothermia, vasoconstriction) > edema TIP for improving signal quality: > Check position of CNAP finger sensor: Do not place on finger joints! > Check if CNAP finger sensor size is correct > Warm up fingers to improve blood flow (use disposable exam glove or towel) > Try to measure on the other hand GREEN: measurement starts automatically. See also: Operators Manual chapter 3.7 no signal good perfusion

35 5. HOW TO ADJUST TREND SETTINGS, SIGNAL AND TIME SCALES 1 Go to Signal & Trend frame with click-wheel and *click* *click* TREND DISPLAY For switching > graphical, > numerical, > alarm history, > Intervention views. 2 Trend Menu: Trend adjustments TREND PARAMETERS > Parameter 1+2: You can choose up to 2 additional parameters shown in Trend view (SVV, PPV, CO, SV, SVR). > Adjust MEAN and AMPLITUDE NOTE of BP and Pulse in case trends are not visible within scale range. NOTE TIME SCALE / TIME SHIFT > Adjust displayed time range according to your needs. > Trend navigation in time back and forth with arrows. Switching parameters here also causes changes in the Hemodynamic frame! FAQs 35

36 FAQs THE BATTERY HOW LONG CAN I MEASURE RUNNING ON BATTERY POWER? up to 120 minutes Battery charge status is indicated with Battery-status-LED. Battery charge status is also shown in right edge of alarm & status frame in steps of 25%. HOW CAN I IMPROVE THE LIFECYCLE OF THE BATTERY? green = ok orange = 25% left red = 15 minutes left!! ATTENTION Never completely discharge the battery! Immediately connect to power supply if battery charge status is below 25%! Send device to CNSystems for maintenance every 24 months. See also: Operators Manual chapter 3.3.2

37 7. CNAP MONITOR SERVICE INTERVALS WHAT ARE THE DEFINED SERVICE INTERVALS? EVERY 12 months SAFETY CONTROL: is usually done according to medical directives. Please contact your biomedical department. See Service Manual for Users for more details. EVERY 24 months CALIBRATION/ METROLOGICAL CONTROL: must be done by CNSystems or an authorized service partner. MAINTENANCE: must be done by CNSystems or an authorized service partner. See also: Operators Manual chapter 17.4 FAQs 37

38 FAQs WHAT TO DO IF CNAP MONITOR NEEDS SERVICE! ATTENTION Metrological control or repair work must be performed by CNSystems or an authorized service partner! 1 For all service requests let us know serial number and software version (displayed on start-up screen). 2 Contact CNSystems via Website: support/service-order service@cnsystems.com Online service order QR-Code:

39 9. SERVICE HOTLINE SERVICE HOTLINE: Service hotline QR-Code: NOTE Service work is usually finished within five working days. Express-Service is possible on request! FAQs 39

40 FAQs WHERE TO GET ADDITIONAL INFORMATION CNSYSTEMS INFO LIBRARY CNAP Monitor - Operators Manual - Comes in hard copy with every CNAP Monitor - Available as PDF-file by CNSystems Website - Applications - Downloads Website - Newsletter - Regular news about our products - Register online: Website - Description of the technology - Still didn t find what you are looking for? Contact us directly! support/contact office@cnsystems.com

41 11. HOW TO CONTACT CNSystems DIRECTLY i GENERAL REQUESTS Contact form: office@cnsystems.com Phone: Fax: REGISTER FOR OUR NEWSLETTER Online: news-events/newsletter feedback@cnsystems.com Newsletter registration FAQs 41

42 42

43 CNAP Monitor 500 HD Quickguide CNSystems Medizintechnik AG, 2016 Version: 07/2016, V1.5.5 EN 43

44 Backcover: Currently using your CNAP Monitor for blood pressure monitoring only? Upgrade to the full set of hemodynamic parameters: BP I PR I CO I SV I SVR I PPV I SVV with Software Upgrade V5.2: Full hemodynamics available after 90 secs upon startup Noninvasive HD for all low & intermediate risk surgeries as an addition to arterial line Perfect match with all other monitoring equipment Improved accuracy through calibration to manual CVP & CO/CI

according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC

according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC The CNAP Monitor 500 meets the requirements of -mark 0408 according to the European standard for medical devices 93/42/EWG, annex IX as amended by 2007/47/EC The product starting with the serial number

More information

General Skills. yes. Device controls, buttons, indicators, and connectors

General Skills. yes. Device controls, buttons, indicators, and connectors HeartStart MRx Skills Checklist This checklist assesses qualified medical personnel s understanding of MRx operation after instructor-based training. Perform a quick skills assessment using the General

More information

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version This document is a guideline only to be used as an aid to comprehensive Inservice training. NAME: HOSPITAL: DATE: VALIDATOR:

More information

T1 BEDSIDE MONITOR GUIDELINE

T1 BEDSIDE MONITOR GUIDELINE APPENDIX A T1 BEDSIDE MONITOR GUIDELINE 5.23 Software Version This document is a guideline only to be used as an aid to comprehensive Inservice training. NAME: HOSPITAL: DATE: VALIDATOR: A. OVERVIEW 1.

More information

INSTRUCTIONS FOR USE. IntelliVue X2. Multi-Measurement Module Release G.0 with Software Revision G.0x.xx. Patient Monitoring

INSTRUCTIONS FOR USE. IntelliVue X2. Multi-Measurement Module Release G.0 with Software Revision G.0x.xx. Patient Monitoring INSTRUCTIONS FOR USE IntelliVue X2 Multi-Measurement Module Release G.0 with Software Revision G.0x.xx Patient Monitoring Printed in Germany 09/08 *M3002-9001B* Part Number M3002-9001B 4512 610 28261 S

More information

OMNI TOUCH SCREEN PATIENT MONITOR

OMNI TOUCH SCREEN PATIENT MONITOR OMNI TOUCH SCREEN PATIENT MONITOR OMNI Intuitive Designed for a fast paced work environment, the Infinium Omni patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N

Nellcor Bedside SpO 2 Patient Monitoring System, PM100N Nellcor Bedside SpO 2 Patient Monitoring System, PM100N In-Service Presentation Nellcor Bedside SpO 2 Patient Monitoring System is CE marked and commercially av ailable in European Union countries. Nellcor

More information

BeneVision Central Monitoring System. Quick Reference Guide

BeneVision Central Monitoring System. Quick Reference Guide BeneVision Central Monitoring System Quick Reference Guide NOTE: The Quick Reference Guide is not intended as a replacement to the Operating Instructions. Prior to operating equipment, the user must be

More information

Getting Started. Genesis Touch System

Getting Started. Genesis Touch System Getting Started Genesis Touch System Getting Started P4810EN.02 2012 Honeywell HomMed. All rights reserved. 1/27/12 Table of Contents Indications for Use...1 Contra-Indications...1 Purpose of this Document...1

More information

Connex Integrated Wall System 2.2 Software Version

Connex Integrated Wall System 2.2 Software Version Connex Integrated Wall System 2.2 Software Version Clinical Inservice Presentation MC12931 1 Table of Contents Device Overview & Intended Use Power ON Power OFF Home Screen Navigation Spot Profile Intervals

More information

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis)

Quick Reference Guide STAN S31 fetal heart monitor. Power ON/OFF. ST Analysis (Fetal ECG analysis) Quick Reference Guide STAN S31 fetal heart monitor Important! Read the STAN S31 User Manual. 1. 2. 3. 1 2 3 4 5 6 Transducer connectors 4. Function buttons Power switch 5. Information panel Fetal ECG (FECG)

More information

BEDSIDE MONITORING Dash 3000/4000 V5

BEDSIDE MONITORING Dash 3000/4000 V5 BEDSIDE MONITORING Dash 3000/4000 V5 Quick Reference Guide 2004 General Electric Company PN: 0304-CS-QRGD Rev. A 2018408-010 BEDSIDE MONITORING Dash 3000/4000 V5 Quick Reference Guide Components Display

More information

Omni (K) PATIENT MONITOR

Omni (K) PATIENT MONITOR Omni (K) PATIENT MONITOR Omni (K) Intuitive Designed for a fast paced work environment, the Infinium Omni (K) patient monitor offers an extremely simple and adaptable user interface. Patient information

More information

Avalon Fetal Monitor

Avalon Fetal Monitor Configuration Guide Avalon Fetal Monitor FM20, FM30, FM40, FM50 For monitor release J.3 with software revision J.3x.xx Patient Monitoring 1Table of Contents 1 Release-Specific Information 5 What's New

More information

INSTRUCTIONS FOR USE. IntelliVue Patient Monitor. MP5 Release G.0 with Software Revision G.0x.xx. Patient Monitoring

INSTRUCTIONS FOR USE. IntelliVue Patient Monitor. MP5 Release G.0 with Software Revision G.0x.xx. Patient Monitoring INSTRUCTIONS FOR USE IntelliVue Patient Monitor MP5 Release G.0 with Software Revision G.0x.xx Patient Monitoring Part Number M8105-9001C Printed in Germany 09/08 4512 610 29031 *M8105-9001C* M8105-9001C

More information

Transport Pro Patient Monitor Operator s Manual Software Version 2.1

Transport Pro Patient Monitor Operator s Manual Software Version 2.1 GE Healthcare Transport Pro Patient Monitor Operator s Manual Software Version 2.1 Transport Pro English 2024598-003 (CD) 2024579-020A (paper) 2007 General Electric Company. All rights reserved. The information

More information

Multiparameter. er Monitor DM

Multiparameter. er Monitor DM d e n t a l a n d m e d i c a l e q u i p m e n t User manual Multiparameter er Monitor DM ENGLISH Contents 1 Introduction...1 1.1 General Information...1 1.2 Screen Display...2 1.3 Button Functions...5

More information

Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report. Emily Thunstrom

Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report. Emily Thunstrom Preventing ECMO Patient Harm While Using Remote Monitoring: A Case Report Emily Thunstrom Disclosure I have no disclaimers Outline Alarms Perfusion alarm management ECMO Alarms and Rounding Case Report

More information

QAM.L L550 series. Patient Monitor Manual QAM.L

QAM.L L550 series. Patient Monitor Manual QAM.L QAM.L550.1112.2 L550 series Patient Monitor Manual QAM.L550.0915.3 Preface This manual has been provided to give detailed descriptions on DARAY s L550 patient monitor, covering the performance, operation

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series Copyright 2005 Welch Allyn. All rights are reserved. No one is permitted to reproduce

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2014 Welch Allyn. All rights are reserved. To support the intended use of the product

More information

Installer Manual KNX Touchscreen Thermostat

Installer Manual KNX Touchscreen Thermostat Installer Manual 02952 KNX Touchscreen Thermostat Index GENERAL FEATURES AND FUNCTIONALITY from page 5 ETS PARAMETERS AND COMMUNICATION OBJECTS from page 7 COMMUNICATION OBJECTS GENERAL FEATURES AND FUNCTIONALITY

More information

Operation Manual Fighter ProVision Software. Version: 0.0 Revision: 1

Operation Manual Fighter ProVision Software. Version: 0.0 Revision: 1 Operation Manual Fighter ProVision Software Version: 0.0 Revision: 1 TABLE OF CONTENTS 1. Introduction 5 2. Software Installation 5 3. PC Users 6 3.1 Introduction 6 3.2 Default Code 6 3.3 Edit PC User

More information

Vitalmax 4000 CL Product Specifications

Vitalmax 4000 CL Product Specifications Page 1 of 5 Vitalmax 4000 CL 510 Garden Avenue North, Clearwater, FL 33755 Phone: (727) 442-8118 Fax: (727) 443-7257 A. Mechanical Description Size 10" H x 13" W x 8"D (25 cm x 33 cm x 20 cm) Weight 18

More information

Vital Signs Monitor 300 Series

Vital Signs Monitor 300 Series Vital Signs Monitor 300 Series Directions for Use Software version 1.2X ii Welch Allyn Vital Signs Monitor 300 Series 2009 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate,

More information

Monitor Alarms and Events

Monitor Alarms and Events This chapter contains the following topics: What Are Alarms and Events?, page 1 How are Alarms and Events Created and Updated?, page 2 Find and View Alarms, page 3 Set Alarm and Event Management Preferences,

More information

Clinical Information Center Pro v5 (CIC Pro)

Clinical Information Center Pro v5 (CIC Pro) GE Healthcare Clinical Information Center Pro v5 (CIC Pro) Participant Reference Guide Clinical Systems Clinical and Educational Development Welcome We would like to take a moment to thank you for choosing

More information

Trident User s Manual

Trident User s Manual Labkotec Oy Myllyhaantie 6 33960 Pirkkala FINLAND Tel. +358 (0)29 006 260 18.05.2017 Fax +358 (0)29 006 1260 Internet: www.labkotec.fi 34 pages Trident Copyright 2017 Labkotec Oy 1/34 TABLE OF CONTENTS

More information

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models:

User s Guide. Cardell Touch Veterinary Vital Signs Monitor. For Models: Cardell Touch Veterinary Vital Signs Monitor For Models: 8013-001 8013-002 8013-003 8013-004 User s Guide TP200 Rev. A 003-2840-00 Rev. AA2 (6/20/18) Software Version V4.0.X Product Information Dealer:

More information

CELLTROL II BIOREACTOR CONTROL SYSTEM OPERATIONS MANUAL

CELLTROL II BIOREACTOR CONTROL SYSTEM OPERATIONS MANUAL Operation Manual Celltrol II Bioreactor Control System Page 1 of 33 Table of Contents 1) Introduction... 3 1.1) Scope of Document... 3 1.2) Control System Overview... 3 1.3) Introduction to Celltrol II...

More information

AGC 200 Advanced Gen-set Controller OPERATOR S MANUAL

AGC 200 Advanced Gen-set Controller OPERATOR S MANUAL Advanced Gen-set Controller OPERATOR S MANUAL Display readings Push-button functions Alarm handling Log list Document no.: 4189340607A SW version 3.5X.X or later Table of contents 1. ABOUT THIS DOCUMENT...3

More information

Instructions for Use Vista 120

Instructions for Use Vista 120 Instructions for Use Vista 120 To properly use this medical device, read and comply with these Instructions for Use. Patient Monitor Software 1.n Release date April 25 2011 Screen images Schematic renderings

More information

Q-tag CLm doc family Type: doc / doc L / doc LR / doc D

Q-tag CLm doc family Type: doc / doc L / doc LR / doc D Q-tag CLm doc family Type: doc / doc L / doc LR / doc D Q-tag CLm doc The clever one Q-tag CLm doc LR The profitable one Q-tag CLm doc L The durable one Q-tag CLm doc D The cool one To monitor your transport

More information

Guide, Quick Start, Short, X-100 System

Guide, Quick Start, Short, X-100 System Rev. CR/CO Date Orig. A 9950 2/15/16 DH Guide, Quick Start, Short, X-100 System P/N LANGUAGE NONIN ELECTRONIC FILE FINALSIZE 10885-001 ENGLISH 10885-001.01.INDD 5.5 X 8.5 NOTE 1: a. MATERIAL: 100lb cover

More information

Model 900. Pulse Oximeter User s Manual

Model 900. Pulse Oximeter User s Manual Model 900 Pulse Oximeter User s Manual MEDIAID INC. 17517 Fabrica Way Suite H, Cerritos, CA 90703 USA Phone: 714-367-2848; Fax: 714-367-2852 Website: www.mediaidinc.com POX010-900 Contents Table of Contents

More information

Quick start guide. English

Quick start guide. English Quick start guide English Read your entire Astral User Guide or Clinical Guide before use. The Astral device 1. Adapter port Can be fitted with single limb adapter, single limb leak adapter or double limb

More information

SECTION L MONITORING EQUIPMENT

SECTION L MONITORING EQUIPMENT BREATH-ALERT Y1090 BREATH-ALERT MK4 MONITOR The Breath-Alert Mk4 respiratory monitor is a simple yet sensitive and reliable device for monitoring breathing. A heat-sensitive sensor is placed in the breath

More information

Operating manual. Wascator FOM71 CLS

Operating manual. Wascator FOM71 CLS Operating manual Wascator FOM71 CLS Operating manual in original language 438 9200-91/EN 2010.05.03 Register now And be able to join the Wascator FOM 71 CLS web site where you can find: General information

More information

User Manual. Dryer Controller M720

User Manual. Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 1.00 Software version 1.00 Preliminary version Manual M720 Dryer controller Page 1 of 42 Document history Preliminary version: - Created in April, 2009

More information

DPM Summary of Features and Benefits

DPM Summary of Features and Benefits Summary of Features and Benefits PAT I E N T M O N I TO R 12.1" TFT display, configurable up to 8 waveforms, enables maximum data analysis of your most critical parameters Standard features include 3/5-lead

More information

FCD-wire Contents. List of Figures

FCD-wire Contents. List of Figures FCD-wire Contents FCD-X21 Configuration 1 Introduction... 1 2 Opening the FCD Application... 1 3 FCD Window... 2 4 FCD LEDs... 3 5 Configuration Operations... 4 FCD Info...4 FCD System Info...5 FCD Interface

More information

Platinum Touch/Rotem One Touch

Platinum Touch/Rotem One Touch User Manual Platinum Touch / Rotem One Touch Platinum Touch/Rotem One Touch Climate Controllers Ag/MIS/UmGB-2583-05/18 Rev 1.0 P/N: 116556 Platinum Touch/Rotem One Touch User Manual Revision: N.1.0 of

More information

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006 Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One Firmware Version 3.00 D August 2006 GammaRAE II Responder Features Getting Started Replacing the batteries Turn the unit

More information

Section 5. Control Systems

Section 5. Control Systems Section 5 There are several Kaleidescape options available for controlling a Premier system. Kaleidescape Remote Kaleidescape App for ipad Kaleidescape Child Remote Browser interface control panels A Premiere

More information

Alarm Client. Installation and User Guide. NEC NEC Corporation. May 2009 NDA-30364, Revision 9

Alarm Client. Installation and User Guide. NEC NEC Corporation. May 2009 NDA-30364, Revision 9 Alarm Client Installation and User Guide NEC NEC Corporation May 2009 NDA-30364, Revision 9 Liability Disclaimer NEC Corporation reserves the right to change the specifications, functions, or features,

More information

Sensor Control Panel

Sensor Control Panel USER S MANUAL PU SENS 01 Sensor Control Panel V55-6EN-03(SENS).indd 1 18.08.2015 10:37:16 2 PU SENS 01 CONTENTS Safety Requirements 2 Main Technical Data 3 Control Panel Mounting 3 Control Panel Operation

More information

Table of Contents. i-vu CCN Standard 4.2

Table of Contents. i-vu CCN Standard 4.2 i-vu CCN Standard 4.2 Owner's Guide CARRIER CORPORATION 2009 A member of the United Technologies Corporation family Stock symbol UTX Catalog No. 11-808-381-01 7/13/2009 Table of Contents Accessing your

More information

SCAN200E USER S MANUAL

SCAN200E USER S MANUAL SCAN200E USER S MANUAL Code No. 2071 1052 rev. 1.4 Code No. 2071 1052 Rev. 1.4 Page 2/16 SCAN200E User s Manual Foreword This manual is for SCAN200E Controller running software version 2.03 or later. We

More information

PowerLogic ION Setup Meter Configuration Software Configuration Guide

PowerLogic ION Setup Meter Configuration Software Configuration Guide PowerLogic ION Setup Meter Configuration Software Configuration Guide 70002-0293-03 12/2010 Conventions Used in this Manual This section describes the symbols and terminology used in this guide. Symbols

More information

Ion Gateway Cellular Gateway and Wireless Sensors

Ion Gateway Cellular Gateway and Wireless Sensors Page 1 of 9 Account & Network Setup If this is your first time using the Ion Gateway online system site you will need to create a new account. If you have already created an account you can skip to the

More information

Genesis Touch Installation Readiness

Genesis Touch Installation Readiness Genesis Touch Installation Readiness Step by Step Readiness 1 2015 Honeywell International All Rights Reserved CP412.01 Overview Who should attend this presentation? Installers of Genesis Touch monitors

More information

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X

Vital Signs Monitor. 300 Series. Directions for Use. Software Version 1.0X Vital Signs Monitor 300 Series Directions for Use Software Version 1.0X Copyright 2003, Welch Allyn Protocol, Inc. Welch Allyn is a registered trademark of Welch Allyn, Inc. Protocol is a trademark of

More information

SureSigns VS4 Vital Signs Monitor

SureSigns VS4 Vital Signs Monitor SureSigns VS4 Vital Signs Monitor INSTRUCTIONS FOR USE Release A.04 English Notice Proprietary Information This document contains proprietary information, which is protected by copyright. Copyright Copyright

More information

PULSE OXIMETER PACKAGE

PULSE OXIMETER PACKAGE instrumentation and software for research PULSE OXIMETER PACKAGE CANL-425SV-A USER S MANUAL DOC-095 Rev. 1.6 Copyright 2012 All Rights Reserved Med Associates Inc. P.O. Box 319 St. Albans, Vermont 05478

More information

No part of this publication may be reproduced, stored in an automated data file or made public in any form or by any means, whether electronic,

No part of this publication may be reproduced, stored in an automated data file or made public in any form or by any means, whether electronic, No part of this publication may be reproduced, stored in an automated data file or made public in any form or by any means, whether electronic, mechanical, by photocopying, recording or in any other manner

More information

Contents 1 Set Up 2 Gateway information 3 Operation of the App 4 Troubleshooting Description of sensors. 1 Set Up. 1.1 Connect the Gateway

Contents 1 Set Up 2 Gateway information 3 Operation of the App 4 Troubleshooting Description of sensors. 1 Set Up. 1.1 Connect the Gateway Contents 1 Set Up 2 Gateway information 3 Operation of the App 4 Troubleshooting Description of sensors 1 Set Up After downloading the Weatherhub app, follow these steps: 1.1 Connect the Gateway Connect

More information

Manual# User s Manual. 200E Series. DCU 210E/208E Diesel Engine Control Unit RP 210E Remote Panel

Manual# User s Manual. 200E Series. DCU 210E/208E Diesel Engine Control Unit RP 210E Remote Panel Manual# 1006494 User s Manual 200E Series DCU 210E/208E Diesel Engine Control Unit RP 210E Remote Panel User's Manual Rev. 1.0 Marine Pro 200E Series ~~~ DCU 210E Diesel Engine Control Unit DCU 208E Diesel

More information

i-vu CCN 4.0 Owner s Guide

i-vu CCN 4.0 Owner s Guide i-vu CCN 4.0 Owner s Guide CARRIER CORPORAION 2007 A member of the United echnologies Corporation family. Stock symbol UX. 11-808-377-01 07/07 able of Contents ACCESSING YOUR SYSEM... 3 YOUR SYSEM DEAILS...

More information

IDEAL E-NERGY Thermostat Touch type User Manual

IDEAL E-NERGY Thermostat Touch type User Manual 67800 03/18 (JRK) IDEAL E-NERGY Thermostat Touch type 203250 User Manual WWW.MFH-SYSTEMS.COM Contents Menu Overview............... 3 Introduction................. 4 Startup Wizard 1/2.............. 5 Startup

More information

M9500 Patient Monitor User s Manual

M9500 Patient Monitor User s Manual M9500 Patient Monitor User s Manual Guangdong Biolight Meditech Co., Ltd. Address: Innovation First Road, Technology Innovation Coast, Jinding, Zhuhai, P.R.CHINA Tel: +86-756-3399900 Fax: +86-756-3399989

More information

Rotem One Touch. User Manual PRELIMINARY DOCUMENT. Ag/MIS/UmGB /18 Rev 1.0 P/N:

Rotem One Touch. User Manual PRELIMINARY DOCUMENT. Ag/MIS/UmGB /18 Rev 1.0 P/N: User Manual Rotem One Touch Rotem One Touch Ag/MIS/UmGB-2603-06/18 Rev 1.0 P/N: 110741 Rotem One Touch User Manual Revision: N.1.0 of 02.2018 Product Software: 7.14 This manual for use and maintenance

More information

USER MANUAL S203. Controller for three circuits. - control for 2 heating circuits - 1 domestic hot water control. Saving energy, creating comfort

USER MANUAL S203. Controller for three circuits. - control for 2 heating circuits - 1 domestic hot water control. Saving energy, creating comfort USER MANUAL S203 Controller for three circuits - control for 2 heating circuits - 1 domestic hot water control Saving energy, creating comfort This user manual consists of two parts. Issues that are intended

More information

ECL Comfort 110, application 116

ECL Comfort 110, application 116 Operating Guide ECL Comfort 110, application 116 (valid as of software version 1.08) English version www.danfoss.com How to navigate? Adjust temperatures and values. Switch between menu lines. Select /

More information

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual

Veterinary Portable Multi-parameter Patient Monitor. Operation Manual Veterinary Portable Multi-parameter Patient Monitor Operation Manual Content Chapter 1 Introduction... 1-1 1.1 General Information... 1-3 1.2 Screen Display... 1-4 1.3 Button Functions and Basic Operation...

More information

EXPERT TRI-STAR. Temperature controller. User s Manual

EXPERT TRI-STAR. Temperature controller. User s Manual Temperature controller r s Manual WARNINGS The warranty can be void if this product is used in a manner not specified by the manufacturer. Every effort has been made to ensure that this manual is complete,

More information

PERFUSION APPARATUS MODEL No: PER-5.2 USER MANUAL. Page 1 of 21

PERFUSION APPARATUS MODEL No: PER-5.2 USER MANUAL. Page 1 of 21 PERFUSION APPARATUS MODEL No: PER-5.2 USER MANUAL Page 1 of 21 TABLE OF CONTENTS 1 GETTING STARTED... 3 1.1 SYMBOLS USED IN THIS MANUAL... 3 1.2 MACHINE INFORMATION... 3 1.3 CONTACT INFORMATION... 3 1.4

More information

Avigilon Control Center System Integration Guide

Avigilon Control Center System Integration Guide Avigilon Control Center System Integration Guide with Velocity INT-HIRSCH-A-Rev3 Copyright 2013 Avigilon. All rights reserved. No copying, distribution, publication, modification, or incorporation of this

More information

Operating Display SCREENS

Operating Display SCREENS I-SV-310 Operating Display SCREENS for SUNNEN SV-SERIES VERTICAL HONING MACHINES SUNNEN, THE SUNNEN LOGO, GENERAL HONE, AND THE GENERAL HONE LOGO ARE REGISTERED TRADEMARKS OF SUNNEN PRODUCTS COMPANY. SUNNEN

More information

IntelliVue Cableless Measurements

IntelliVue Cableless Measurements Instructions for Use IntelliVue Cableless Measurements CL SpO2 Pod CL NBP Pod CL Respiration Pod Release B.02 Patient Monitoring 1Table of Contents 1 Introduction and Basic Operation 5 Introducing the

More information

Avigilon Control Center System Integration Guide

Avigilon Control Center System Integration Guide Avigilon Control Center System Integration Guide with Picture Perfect 4 INT-PP4-A-Rev1 Copyright 2012 Avigilon. All rights reserved. No copying, distribution, publication, modification, or incorporation

More information

Halton SAFE / 7.14 user guide and installation instructions

Halton SAFE / 7.14 user guide and installation instructions Halton SAFE / 7.14 user guide and installation instructions VERIFIED SOLUTIONS BY H A LTO N Enabling Wellbeing Table of contents 1 System description 3 2 User Accounts 4 3 Main menu 7 3.1 Main menu - Change

More information

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English

Operator s Manual. Avant Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor English Operator s Manual Avant 2120 Pulse Oximeter and Noninvasive Blood Pressure (NIBP) Monitor 0123 English CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

More information

Adaptive CyCLO Technical and HMI User Guide. CyCLO User Guide. Version th December 2017 REV

Adaptive CyCLO Technical and HMI User Guide. CyCLO User Guide. Version th December 2017 REV CyCLO User Guide Version 2.00 19 th December 2017 REV 2.00 1 Contents 1. Hardware... 3 1.1. Introduction... 3 1.2. Electrical Specification... 3 1.3. Board Overview... 4 1.4. Electrical Installation...

More information

USER MANUAL USB Multi-Function Datalogger Model RHT35

USER MANUAL USB Multi-Function Datalogger Model RHT35 USER MANUAL USB Multi-Function Datalogger Model RHT35 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech multi-function, easy-to-use, portable

More information

Micro-VPAC IIT LSC - Lift Station Controller. User Manual. Version 1.0.2

Micro-VPAC IIT LSC - Lift Station Controller. User Manual. Version 1.0.2 Micro-VPAC IIT LSC Micro-VPAC IIT LSC - Lift Station Controller User Manual Version 1.0.2 Table of Contents ii CHAPTER: Table of Contents Contents Introduction 5 General 6 Control Description 6 Manual

More information

G4S SMARTalarm User Guide

G4S SMARTalarm User Guide G4S SMARTalarm User Guide CONGRATULATIONS WITH YOUR NEW SECURITY SYSTEM! We are glad that you have chosen G4S SMARTalarm. G4S is the largest supplier of security solutions in the world. We have invested

More information

LineGuard 2300 Program User Manual (FloBoss 107)

LineGuard 2300 Program User Manual (FloBoss 107) Form A6251 Part Number D301346X012 November 2012 LineGuard 2300 Program User Manual (FloBoss 107) Remote Automation Solutions Revision Tracking Sheet November 2012 This manual may be revised periodically

More information

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor

GE Healthcare. Quick Guide. Dash 2500 Patient Monitor GE Healthcare Quick Guide Dash 2500 Patient Monitor Hardkeys Main Menu Hide menu and view entire waveform screen Display the large numeric vital signs Review or change alarm limits for all operating parameters

More information

Installation Guide. ECL Comfort 210, application A Table of Contents

Installation Guide. ECL Comfort 210, application A Table of Contents 1.0 Table of Contents 1.0 Table of Contents... 1 1.1 Important safety and product information..................... 2 2.0 Installation... 4 2.1 Before you start.....................................................

More information

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry

User Guide. Cardell Insight Veterinary Monitor. For Models: 8014 Blood Pressure Blood Pressure, Pulse Oximetry Cardell Insight Veterinary Monitor For Models: 8014 Blood Pressure 8015 Blood Pressure, Pulse Oximetry User Guide 003-2981-00 Rev. B (2/21/17) Software Version 1.2V Product Information Dealer: Date of

More information

Dryer Controller M720

Dryer Controller M720 User Manual Dryer Controller M720 Hardware version 2.00 Software version 2.00 Manual M720 Dryer controller Page 1 of 60 Document history Preliminary version: - Created in April, 2009 Hardware Version 2.00,

More information

Instructions for Use. Infinity Vista. Infinity Patient Monitoring Series

Instructions for Use. Infinity Vista. Infinity Patient Monitoring Series Instructions for Use Infinity Vista Infinity Patient Monitoring Series Infinity Configured Monitoring Series Infinity Vista Instructions for Use WARNING: For a full understanding of the performance characteristics

More information

Vital Signs Monitor PC-900. User Manual

Vital Signs Monitor PC-900. User Manual Vital Signs Monitor PC-900 User Manual AM1000A 神经功能重建治疗仪说明书 Shenzhen Creative Industry Co., Ltd. II This Manual is written and compiled in accordance with the IEC 60601-1(Medical electrical equipment Part1:

More information

Manage Alarms. Before You Begin. NTP-E57 Document Existing Provisioning CHAPTER

Manage Alarms. Before You Begin. NTP-E57 Document Existing Provisioning CHAPTER CHAPTER 9 Manage Alarms This chapter provides procedures required to view and manage Cisco ONS 15600 alarms and conditions. Cisco Transport Controller (CTC) detects and reports SONET alarms generated by

More information

Manage Alarms. Before You Begin CHAPTER

Manage Alarms. Before You Begin CHAPTER CHAPTER 9 Manage Alarms This chapter provides procedures required to view and manage Cisco ONS 15600 alarms and conditions. Cisco Transport Controller (CTC) detects and reports SONET alarms generated by

More information

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain

OPERATOR S MANUAL. Patient Monitor M20. EU representative TECNOMED 2000 S.L. Valencia, Madrid Spain OPERATOR S MANUAL Patient Monitor M20 EU representative TECNOMED 2000 S.L. Valencia, 25-28012 Madrid Spain Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1 Donghwa-ri, Munmak-eup, Wonju-si,

More information

Autoclave Operations Manual

Autoclave Operations Manual UNIVERSITY OF ILLINOIS AT URBANA/CHAMPAIGN - COMPOSITES MANUFACTURING LAB Autoclave Operations Manual Version 1 Written by KRH 2/21/2011 Contents ACCS... 4 Alarms... 7 Main Display Tabs... 7 Process Control

More information

THX-DL Data Logger USER & INSTALLATION MANUAL V

THX-DL Data Logger USER & INSTALLATION MANUAL V THX-DL Data Logger USER & INSTALLATION MANUAL V1.2012 www.thermomax-refrigeration.com Contents PRESENTATION Summary of Features 2 INSTALLATION Safety Precautions 4 THX Unit 4 Sensors 4 Alarm Relay 4 Power

More information

Compact Genset Controller, CGC 400 Push-buttons LEDs Display and menu structure Display readings Alarm handling and log list

Compact Genset Controller, CGC 400 Push-buttons LEDs Display and menu structure Display readings Alarm handling and log list OPERATOR'S MANUAL Compact Genset Controller, CGC 400 Push-buttons LEDs Display and menu structure Display readings Alarm handling and log list DEIF A/S Frisenborgvej 33 DK-7800 Skive Tel.: +45 9614 9614

More information

THE CLINICAL CONNECTION

THE CLINICAL CONNECTION GE Healthcare THE CLINICAL CONNECTION Training Choices. Anytime. Anywhere. Dash Series V5 Basic Bedside Monitoring Mentor Applications Class Workbook (Full Hemodynamics) PN:0304-CS-BUDFH Rev. A 2018408-009

More information

Product Specifications

Product Specifications 1.Safety Specifications 1.1Product Classification Class Type of protection against electric shock Specification Class I, with internal power supply Degree of protection against electric shock Anti-defibrillation:

More information

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium.

HBP Patient Monitor. Operator s Manual. EU representative OBELIS S.A Bd. Général Wahis, Brussels, Belgium. Patient Monitor HBP-2070 Operator s Manual EU representative OBELIS S.A Bd. Général Wahis, 53 1030 Brussels, Belgium Local distributor Manufacturer Mediana Co., Ltd. Wonju Medical Industry Park, 1650-1

More information

System Galaxy Quick Guide

System Galaxy Quick Guide System Galaxy Quick Guide CONFIGURATION AND OPERATION Integrating CCTV with System Galaxy JAN 2019 SG 11.1.0.2 System Galaxy Quick Guide For CCTV Integration Configuration & Operation Information in this

More information

D3D Wi-Fi GSM Smart Alarm System -User Manual

D3D Wi-Fi GSM Smart Alarm System -User Manual D3D Wi-Fi GSM Smart Alarm System -User Manual D3D Wi-Fi / GSM Smart Alarm system (Model : D10). Please read all instructions carefully & follow steps for easy home installation. 1 P a g e D3D Wi-Fi / GSM

More information

Laptop / PC Programming Manual

Laptop / PC Programming Manual Laptop / PC Programming Manual Doc. # Fire PC Program rev B 01.07 This Document is property of Evax Systems, Inc. The Evax Fire Solutions Programmer Components 2 1.0 System Setup 4 1.1 Interface Setup

More information

Avigilon Control Center System Integration Guide

Avigilon Control Center System Integration Guide Avigilon Control Center System Integration Guide with Gallagher Command Centre INT-CARDAX-C-Rev3 Copyright 2013 Avigilon. All rights reserved. No copying, distribution, publication, modification, or incorporation

More information

mach 5 100kHz Digital Joulemeter: M5-SJ, M5-IJ and M5-PJ Joulemeter Probes

mach 5 100kHz Digital Joulemeter: M5-SJ, M5-IJ and M5-PJ Joulemeter Probes mach 5 100kHz Digital Joulemeter: M5-SJ, M5-IJ and M5-PJ Joulemeter Probes Measure every pulse at pps! Our new mach 5 Digital Joulemeter is designed to measure laser pulse energy up to 130,000 pps with

More information

ARGUS LCM. Patient monitor ARGUS LCM and ARGUS LCM PLUS. Operating Instruction. Art. no.: rev.: d

ARGUS LCM. Patient monitor ARGUS LCM and ARGUS LCM PLUS. Operating Instruction. Art. no.: rev.: d ARGUS LCM Patient monitor ARGUS LCM and ARGUS LCM PLUS Operating Instruction Sales and Service Information The SCHILLER sales and service centre network is world-wide. For the address of your local distributor,

More information

Temperature Controller CC24-7 UNIBOX USER S MANUAL Legion Dr. Mason, MI USA October 2010 Ph. (517) Fax (517)

Temperature Controller CC24-7 UNIBOX USER S MANUAL Legion Dr. Mason, MI USA October 2010 Ph. (517) Fax (517) 3 4 Temperature Controller USER S MANUAL Aerotech, Inc. FORM: QM1386 4215 Legion Dr. Mason, MI 48854-1036 USA October 2010 Ph. (517) 676-7070 Fax (517) 676-7078 FOR CUSTOMER USE Enter the serial number

More information

Operating, Installation and Service Manual W3400H, CLC 41 Clarus Control

Operating, Installation and Service Manual W3400H, CLC 41 Clarus Control Operating, Installation and Service Manual W3400H, CLC 41 Clarus Control 438 9202-01 05.26 Washer extractor, drum volume 400 litres 3 Contents Machine operation... 7 General description... 7 Automatic

More information