The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors.

Size: px
Start display at page:

Download "The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors."

Transcription

1 The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors. A dissertation submitted to the University of Manchester for the degree of Master of Science by Research in the Faculty of Engineering and Physical Sciences Charles Robert Bradbury School of Physics and Astronomy

2 Contents List of figures 4 Abstract 6 Declaration 7 Copyright 8 Acknowledgements 9 1. Introduction Ionisation Chambers Detector properties Design considerations Experimental Setup Gas choice Detector Design Signal Processing Results Gas deterioration Pressure dependence Range Determination Electric field dependence

3 CONTENTS Electron combination effects Conclusions. 51 Bibliography 53

4 List of Figures Figure 2.1: Layout of a generic transversal gas-filled ionisation chamber p16. Figure 2.2: SRIM track simulations of 107MeV Pd 106 fragments travelling through a 30µm mylar layer p17. Figure 2.3: Variation of ionisation density as a function of distance travelled by an α particle in air p18. Figure 2.4: Layout of a generic Bragg curve spectroscopy setup p19. Figure 3.1: Schematic diagram of chamber setup used p24. Figure 3.2: The first image shows the field shape that is generated by a planar voltage drop. The second shows the homogenising effect of including field shaping rings p26. Figure 3.3: Circuit diagram of the detector p27. Figure 3.4: Typical preamplifier output as seen on scope p28. Figure 3.5: Observed energy spectrum of Cf 252 in 85mbar freon p30. Figure 3.6: Diagram of signal analysis setup p31. Figure 3.7: All four signal stages shown simultaneously on true timescale p32. Figure 4.1: Variation of signal size with age of gas p34. Figure 4.2: Variation of rise time with age of gas p36. Figure 4.3: Variation of rise time with pressure p38. Figure 4.4: Rise time pressure against reduced electric field p39.

5 LIST OF FIGURES 5 Figure 4.5: SRIM calculated ranges for 99MeV Pd 106 fragments into gas p40. Figure 4.6: Variation of signal amplitude with pressure p41. Figure 4.7: Observed energy spectrum of Cf 252 in 70mbars of isobutane p42. Figure 4.8: SRIM calculated ranges for 99MeV Pd 106 fragments into gas alongside calibrated ranges p43. Figure 4.9: Rise time pressure against reduced electric field p44. Figure 4.10: Variation of signal amplitude with reduced electric field strength p45. Figure 4.11: Table showing reduced field strength required for various signal amplitudes p46. Figure 4.12: Variation of electron drift velocity with reduced electric field strength p46. Figure 4.13: Variation of electron drift velocity with reduced electric field strength plotted alongside data from literature p47. Figure 4.14: Variation of signal amplitude with electron drift time p48. Figure 4.15: Variation of signal amplitude with electron drift time (longer timescale) p49.

6 Abstract A Bragg curve spectrometer was built and run using both isobutane and freon independently as fill gases. Analysis of the signals produced by light fission fragments emitted from a Cf252 source allowed detailed comparison of the gases characteristics when used in detector chambers. Isobutane was found to provide larger signal amplification, however this appeared to deteriorate with the age of the gas, with an observed loss of 0.824% (±0.049%) per hour. Freon provided considerably faster signals however the rise time increased over time, at a rate of 0.967% (±0.106%). Attempts to determine the range of fission fragments did not agree with simulated predictions in the case of either gas. Furthermore the electron drift velocity shows a different trend against reduced electric field strength for the two gases. Signal amplitude losses were observed as the drift time increased as a result of recombination and addition effects. A combination coefficient was calculated for both gases and found to be (±0.0069)µs-1 for isobutane and (±0.0154)µs-1 for freon. From this the mean electron lifetime was found to be 12.0(±1.0)µs for isobutane and 3.06(±0.14)µs for freon. However below reduced field strengths of Vcm-1mbar-1 isobutane was found to deviate from the expected exponential decay trend, possibly indicating the presence of a new recombination phenomena.

7 Declaration No portion of the work referred to in the dissertation has been submitted in support of an application for another degree or qualification of this or any other university or other institute of learning. Charles Robert Bradbury.

8 Copyright i. The author of this dissertation (including any appendices and/or schedules to this dissertation) owns any copyright in it (the Copyright ) and s/he has given The University of Manchester the right to use such Copyright for any administrative, promotional, educational and/or teaching purposes. ii. Copies of this dissertation, either in full or in extracts, may be made only in accordance with the regulations of the John Rylands University Library of Manchester. Details of these regulations may be obtained from the Librarian. This page must form part of any such copies made. iii. The ownership of any patents, designs, trade marks and any and all other intellectual property rights except for the Copyright (the Intellectual Property Rights ) and any reproductions of copyright works, for example graphs and tables ( Reproductions ), which may be described in this dissertation, may not be owned by the author and may be owned by third parties. Such Intellectual Property Rights and Reproductions cannot and must not be made available for use without the prior written permission of the owner(s) of the relevant Intellectual Property Rights and/or Reproductions. iv. Further information on the conditions under which disclosure, publication and exploitation of this dissertation, the Copyright and any Intellectual Property Rights and/or Reproductions described in it may take place is available from the Head of School of Physics and Astronomy.

9 Acknowledgements First and foremost I d like to thank the whole nuclear department I ve thoroughly enjoyed being part of the group. No matter what time in the afternoon (or indeed morning) I ve reached my physics saturation point, there has almost always been at least one person equally willing to call it a day and go to the pub. For this I am grateful. Naturally however there are some people who need singling out. I d like to thank Gavin Smith, my supervisor for accepting me onto this course and helping and guiding me throughout the year. I d also like to thank Joe Dare who has been on hand to help me virtually every day throughout the whole year, and has been consistently patient, helpful and insightful. Also Andy MacFarlane who has provided invaluable, friendly help with my experiment and has managed to fix all the things in the lab I ve managed to break. There also many other members of the department, in particular Paul Campbell and Andy Pollitt, who have spent much of the year being bombarded with questions and always somehow finding the time to help. Finally I d like to thank my parents for helping me to fund this course, supporting me on this course and indeed supporting me throughout my entire academic career, despite being absolutely clueless on physics from about GCSE level upwards. Thanks mum and dad, I really appreciate it

10 1. Introduction Identification and analysis of radiation and radioactive sources is an important part of both experimental nuclear physics and commercial radiation usage and in many cases their capabilities and reliability are of paramount importance. Whilst a Geiger- Muller tube merely counts activity levels, more advanced detectors may be required to record the energy, the type of radiation, the atomic mass and even the atomic number of a particle. Unsurprisingly, gathering such precise information on such a scale requires a carefully considered experimental setup. The type of nuclear detector I have worked with in this project is a Bragg chamber, a type of ionisation chamber in which irradiated particles pass through a gaseous medium, ionising gas molecules. My work studies the experimental effects of using different gas types in the chamber. As will be discussed this is a complex issue to address theoretically, with many factors to consider. As a result most research of this subject has adopted a predominantly experimental approach, my work being no exception. This requires building a suitable working detector, and the design, development and implementation of this detector has formed the bulk of my studies. This has allowed me to run tests with different gases to compare the differences in energy resolution, rise-time and signal amplitude in order to highlight operational advantages and disadvantages of different gases. Specifically I have set up my experiment to compare two gases isobutane and freon. Both gases possess desirable properties for use in an ionisation chamber and as a result have frequently been used in experiments. Yet whilst their individual suitability has been proven, and their superiority over other gases has been shown, there appears to be no direct comparison [OGG + 82], [Bre76], [SWH + 81]. By doing so and analysing the different characteristics of both gases my results would aid

11 SECTION 1: INTRODUCTION 11 others when designing a chamber as well as potentially helping to optimise existing detectors. I have begun this report by outlining the workings of a Bragg chamber and addressing the key considerations faced to someone building one. From this I have detailed the design of my chamber and the experimental method employed. Finally I have presented the data and subsequent results and made conclusions on the properties and suitability of the gases.

12 2. Ionisation chambers Irradiated particles entering an ionisation chamber travel through a (usually) gaseous medium, subject to an electric field. The gas molecules are ionised by these particles creating positive ions and electrons which travel to the cathode and anode respectively generating a current. The concept is simple, and prior to the 1960s ionisation chambers were widely used in nuclear spectroscopy before becoming somewhat superseded by solid state and scintillation detectors. Keeping the detector fill gas isolated means that the particles being studied are inevitably required to traverse an entrance window (unless the source itself is actually in the chamber), and despite the use of extremely thin windows the energy straggling experienced hinders the energy resolution of the chamber [SDH + 74]. Furthermore, collecting the charged particles in a gas is a relatively slow process [Kno99]. For studies of heavy ions however, ionisation chambers remain the most practical type of detector for the following reasons: [i] Heavy ions are more capable of causing radiation damage. However ionisation chambers are not especially subject to radiation damage and if necessary a constant gas flow can be used to keep the gas pure. [ii] Physical size is less of a concern; chambers can be built to appropriate dimensions. This gives ionisation detectors more flexibility than other types and allows the construction of larger chambers, with large cross sections. This is especially useful for the study of fission fragments where events may be relatively rare. [iii] Gaseous detectors allow for variation in stopping distance by varying the pressure, meaning a single detector can be used for a large range of different particles [SDH + 74], [Kno99]. Gaseous chambers became even more advantageous with the introduction of Bragg Curve Spectroscopy (BCS) by Gruhn et. al. and also Schiessl et. al. in the early eighties [GBL + 82], [SWH + 81]. By setting up the chamber such that the incoming particles travel parallel to the electric field the specific ionisation (de/dx) along the

13 SECTION 2: IONISATION CHAMBERS 13 particle s track can be observed as a variation in charge collection on the anode. Such an approach allows for improved energy resolution and in theory the determination of the atomic charge number, Z. In the following section I have outlined the general properties of a detector and discussed which characteristics are of greatest significance for a comparison of fill gas types. 2.1 Detector Properties. W.R.Leo outlines seven key properties of a detector which have varying importance depending on the detector s purpose [Leo94]. I have briefly outlined these and discussed their relevance to my work Sensitivity The sensitivity of a detector is its ability to detect radiation of a specific nature and energy and is dependent on a number of different factors. A thick entrance window for instance would prevent some particles from entering, resulting in poor sensitivity for low energy or heavily ionising particles. The properties of the fill gas may also affect the sensitivity oxygen for example has a high electron affinity and will soak up electrons as they travel to the anode, reducing the size of the signal. A detector must always be configured such that it offers maximum sensitivity to the radiation it is expected to detect. This allows large signals to form, resulting in a good signal-to-noise ratio and improved energy resolution [Leo94] Response A detector creates a current pulse as charged particles are collected. More energetic particles will cause more ionisation resulting in more charge being collected, however the relationship between energy and pulse size may not be linear. If the chamber is set up such that incoming particles are fully stopped within the detector

14 and all the charge is collected (i.e. limiting effects such as recombination of charged particles and electron attachment) an inherent lack of linearity is avoided. This can be difficult to achieve however, as some recombination and/or attachment is unavoidable. Other effects such as ballistic deficit (the loss of signal from capacitance decay while the charge is still being collected) may also become significant [Leo94] Energy Resolution In an ideal world a beam of monoenergetic particles (such as from a good alpha source) will produce identical signals. However in reality these particles will be measured as having a distribution of energies. This distribution is described by its full width at half maximum (FWHM) and by minimising this, a detector can provide more reliable information about a particle s energy. When examining fission fragments the energy spectrum should exhibit a characteristic double peak structure, which becomes poorly defined with increased peak overlap if the energy resolution is large [Leo94]. Resolution is often a vital consideration in experiments, particularly when particle identification is an objective, and previous work has shown significant differences when using different gases [OGG + 82]. However, the quality of the spectrum is somewhat dominated by the signal processing method employed, making it a difficult and unreliable means of comparison between gases. It is more practical to instead compare errors on raw signal energy measurements Response function Related to the energy resolution, the response function is the shape of the distribution shown by a detector receiving a monoenergetic source. This might be expected to be Gaussian; however unwanted effects may result in distortion of the curve shape. This is generally more of a problem when detecting lighter particles as effects such as bremsstrahlung may become significant [Leo94] Response time

15 SECTION 2: IONISATION CHAMBERS 15 The signal produced by a particle will not be created instantaneously as the charge must be collected. The rate at which the signal forms can be important, particularly if the detector is required for timing measurements or if the signal is to be digitalised. A fast response time will result in a clear, sharp signal, and minimises effects such as recombination and ballistic deficit. The choice of fill gas has been observed to make significant differences [OGG + 82] Efficiency The intrinsic efficiency is the ratio of particles that enter the detector and are subsequently detected. Provided the gas pressure is such that all particles are stopped, and the electric field is large enough to collect all the charge, all gases should provide the same high efficiency [Leo94] Dead Time This is a measure of how long a detector is insensitive to further radiation after detecting an event. Like the response time this is inherently linked to the drift speed of the ionised particles and provided the collection time is considerably shorter than the average time between events, this is not an important consideration [Leo94]. An experimental setup allowing assessment of detector sensitivity, energy resolution and response time is therefore required to allow firm conclusions to be drawn about the suitability of each gas for different contexts. By varying the field voltage and gas pressure a more thorough comparison can be made, allowing observations of the ideal conditions and the flexibility of the gases. 2.2 Design Considerations Prior to the introduction of BCS, an ionisation chamber might typically have had the layout shown in Figure 2.1. Incoming radiation passes through a thin entrance window, required in order to keep the detector gas separate. Before reaching the

16 window the radiation should only travel through a high vacuum to ensure all energy loss occurs within the chamber. The region between the cathode and the Frisch grid can be referred to as the active detector region. Figure 2.1: Layout of a generic transversal gas-filled ionisation chamber Entrance Window A similar chamber built by H. Sann et. al. in their 1974 paper A Position Sensitive Ionisation Chamber [SDH + 74] used a 2.5μm Hostaphan (a trade name of polyethylene terephthalate or PET ) window, chosen; because of its stability even after many cycles of pressurizing and depressurizing and despite the fact that it is rather inhomogenous ( d/d ±6%) and much thicker than required by the gas pressure actually used. Indeed the entrance window material must satisfy a particularly demanding set of conditions. As well as featuring stability and homogeneity it must be sufficiently robust and airtight, even when only micrometres thin. As will be shown, even a small change in gas pressure can have a significant effect upon the radiation stopping distance, particularly when dealing with heavy fission fragments. Unsurprisingly then, an obstructive solid can bring them to an abrupt halt. The thickness of the window is therefore of paramount importance to minimise the energy straggling which occurs, potentially ruining the sensitivity of the detector and broadening the energy resolution. Figure 2.2 shows simulated tracks of a typical

17 SECTION 2: IONISATION CHAMBERS 17 fission fragment, in this case a 107 MeV Palladium atom, travelling through PET. Most are stopped after approximately 20μm and the deviations in direction suggest that significant but unpredictable energy losses occur well before this. The simulation used is the Stopping and Range of Ions in Matter (SRIM) computer program [ZBL85]. Figure 2.2: SRIM track simulations of 107MeV Pd106 fragments travelling through a 30µm mylar layer. Approximately 100 tracks are shown. [ZBL85]. Clearly, were the source positioned in the chamber there would be no need for an entrance window and these problems would be avoided. However this would allow no opportunity for secondary detectors such as would be required to digitalise the signal or for a time of flight (TOF) measurement, used to determine the atomic mass, A [Dar09]. It is often desirable to have the source positioned well away from the active detector region to collimate radiation and reduce the event rate, minimising pile-up. If the region between source and detector is not separated from the active detector region then the entire system will be filled with detector gas. Energy straggling would then result in a much wider range window which coupled

18 with the increased perpendicular deviation, would negatively affect the detector s response linearity, energy resolution and efficiency. It is generally preferable therefore, for longer detector setups, to deal with the problems associated with using an entrance window. Doing so allows the other sections to be pumped down to a high vacuum Ionisation and Charge Collection Having passed the entrance window the radiation travels through the chamber and ionises gas molecules. The specific ionisation, equivalent to the rate of energy loss of the particle de/dx, varies along the track in accordance with the Bethe-Bloch equation, the nature of which is shown in Figure 2.3. The ionised particles and electrons then drift to the cathode and anode respectively creating a current. Note that in Figure 2.1 the electric field is perpendicular to the particle trajectories making this a transversal chamber [OGG + 82]. Note also the presence of a Frisch grid, a fine mesh of wires held at a high positive voltage to generate the electric field. When the electrons reach the grid they then drift to the anode which is held at an even higher voltage. Up to this point however the grid shields the anode from the motion of the electrons. As a result the positional dependence of the signal size is removed and all electrons appear as if they were freed level with the grid [Lil01].

19 SECTION 2: IONISATION CHAMBERS 19 Figure 2.3: Variation of ionisation density as a function of distance travelled by an α particle in air (the Bragg curve) [Lil01]. It should be observed that provided an incoming particle is fully stopped within the chamber the initial energy of the particle, the total number of ionisation events and the current generated across the electrodes are all theoretically proportional. Furthermore since ionisation creates charged pairs, half the current is a result of ion motion and half is due to electron motion. It is therefore only necessary to measure signals on the anode. This is important as the positive ions drift approximately 1000 times slower than the electrons and if collected would greatly hinder the rise time of the signal [Lil01] Bragg Curve Spectroscopy BCS allows more information to be gathered from incident radiation. Instead of a transversal setup, a chamber can be built with axial orientation such that the radiation travels parallel to the field direction. A typical BCS chamber is shown in Figure 2.4. Figure 2.4: Layout of a generic Bragg curve spectroscopy setup [SWH + 81].

20 The layout allows field shaping rings to encircle the particle trajectories. These are held at uniformly increasing voltages to improve the homogeneity of the electric field [SWH + 81]. As discussed, a dedicated cathode is not in fact necessary. The grounded metal wall and entrance window at the front of the chamber serve this purpose. Superimposed upon the image is the Bragg curve of a radiated particle, fully stopped after travelling a distance x through the gas. The electrons produced continue in this direction towards the anode. As the drift speed of the electrons is relatively slow the ionisation events may be treated as occurring simultaneously along the entire trajectory. Consequentially the signal shape will match that of the Bragg curve. Since the shape of the Bragg curve is predominantly dependent on the particle s atomic number [SAD + 93], it can be used to partially identify the radiation. However this is difficult to accurately achieve. Complete particle identification would in addition require mass analysis such as that achieved through a TOF setup. This requires a timing detector at a known distance from the ionisation chamber so that velocity measurements can be made. The energy is determined from the ionisation chamber and the mass can then be calculated. Axial chambers also offer a better energy resolution than transversal chambers due to their symmetry about the fragment trajectories. Transversal chambers will necessarily distort the electric field about the entrance window, and this effect becomes more pronounced with larger windows [OGG + 82].

21

22 3 Experimental Setup 3.1 Gas Choice Gas properties As mentioned, both isobutane and freon, have both been shown to be suitable detector fill gases. This means that they adequately provide the requirements outlined in section 2.1 and do not present any major undesirable effects. Such effects could include corrosivity such as exhibited by isopentane [OGG + 82]. It is difficult to predict how gases will perform in an ionisation chamber because of the complexity of the process they are involved in. The gas stops the incoming radiation (and in doing so forms ionised charge pairs) then provides a medium for the charged particles to drift through the electric field during which time it will absorb some electron through combination reactions. Particularly important properties include the stopping power and ionisation energy of the gas. The stopping power determines the pressure required to stop the radation over a certain distance and this in turn affects the drift speed of the electrons. The ionisation energy affects how much energy is required to create a charged pair, and thus how much charge is collected per fission fragment. A high ionisation energy also increases the likelihood of recombination (i.e. the electron recombining with a positive ion before reaching the anode). The electron affinity of the gas determines the liklihood of a neutral molecule collecting an additional electron [Kno99] Isobutane

23 SECTION 3: EXPERIMENTAL SETUP 23 Isobutane (fully known as methylpropane) is an alkane with the chemical formula C 4 H 10. It is therefore an example of an organic compound, specifically a hydrocarbon. Isobutane has a molecular mass of 16 and offers a relatively low ionisation energy of 10.68eV [Nis11a] Freon Freon is the DuPont trade name for chloroflourocarbons [DuP11]. The name is therefore used trivially for a wide range of chemicals but throughout this work it refers specifically to freon 14 (R-14) formally known as tetrafluoromethane or carbon tetrafluoride, and with the chemical formula CF 4. It is much heavier than isobutane with a molecular mass of 88g/mol and also has a higher ionisation energy of 16.2eV [Nis11b]. 3.2 Detector Design Overview The design of the detector used primarily follows features outlined in section 2, most fundamentally the axial orientation. This allows Bragg curve spectroscopy and renders the chamber suitable for further work involving mass measurement although this has not formed part of this experiment. Without a secondary detector signal digitalisation is not possible, instead a high frequency oscilloscope has been used for analysis. The detector layout is shown in Figure 3.1.

24 Figure 3.1: Schematic diagram of chamber setup used. Sections have been cut away to show internal detail Source The source used is Californium-252 with an activity of approximately 100μCi. The age of the source and the half life of Californium (approximately 18 months and 32 months respectively), were taken into account when determining this activity level [Kra88]. The long beam tube reduces this high level of activity to a more manageable event rate. It also collimates the radiated particles to ensure no trajectories pass outside the field shaping rings. Were this to happen the particle would only deposit some of its energy within the detectable area. This structure deteriorates with the age of the Californium. An older source was originally used but this provided a poor spectrum with almost complete overlap of the peaks, making it harder to assess the performance of the detector and harder to identify the radiation represented by the spectrum. It would also have caused problems for energy resolution comparisons Beam Tube and Intermediate Section

25 SECTION 3: EXPERIMENTAL SETUP 25 This region must be completely vacated to prevent any ionisation taking place before particles reach the window. A rotary pump and turbo pump are used to achieve this. There is a bypass tube which allows these pumps to also evacuate the Bragg chamber. When the turbo pump is activated there is a slight pressure difference between gauges 1 to 2 (with typical readings of 1 x 10-6 mbars and 6 x 10-4 mbars respectively), due to their proximities to the turbo pump. SRIM tests indicate that even a pressure as high as 1 x 10-2 mbars would scarcely affect the fission fragments over the metre or so that they travel. In the set up used, the central section acts as merely an extension to the beam tube. However its inclusion provides the potential inclusion of a secondary detector in further experimentation, either for TOF measurements or as a trigger for signal digitalisation Entrance Window A 0.5μm mylar window was initially used ( mylar is a trade name of PET), however this proved too thin to hold the required pressure gradient. With a chamber pressure above about 50mbars the pressure in the central section was observed to rapidly rise to unacceptably high levels. This was therefore replaced with a 0.9μm window which was observed to hold the central section below 2 x 10-6 mbars even with chamber pressures as high as 85mbars. As shown in Figure 3.1 a metal grid is used to support the window at these relatively high pressure gradients. This is kept as thin as possible to maximise the window s transparency. In addition pressure changes are applied at only 1mbar per second to further minimise the stress on the mylar. SRIM simulations predict that a 107MeV Palladium-106 fragment loses approximately 8.11MeV in the window The Electric Field and Chamber Casing The electric field is required to collect the electrons. As discussed in section 2 there is no need to collect the positive ions, they are simply allowed to drift to the front of the chamber. As demonstrated by Figure 3.2, field shaping rings are included to

26 homogenise the electric field. Without the rings the electron drift speed would have positional dependence. Figure 3.2: The first image shows the field shape that is generated by a planar voltage drop. The second shows the homogenising effect of including field shaping rings. The Frisch grid acts as the anode while the electrons are being collected. However once the electrons reach the grid it must be ensured that they then proceed to the anode. This is achieved by firstly making the grid as transparent as possible and secondly by making the electric field between the anode and grid (E AG ) larger than the field between the grid and cathode (E GC ). Specifically, the following condition must be met in order to make the amount of charge lost to the Frisch grid negligible; Equation 3.1 [SAD + 93] Where r is the radius of the grid wire and g is the inter-wire spacing.

27 SECTION 3: EXPERIMENTAL SETUP 27 The grid is constructed from 50μm gold-plated tungsten wires with a separation of 2mm. This provides a geometric transparency of 97.5% and from the above equation requires E AG /E GC to be greater than This is achieved through the following setup: Figure 3.3: Circuit diagram of the detector. Typical HVs ranged from 250V to 1200V. Based on the resistor values shown in Figure 3.3 this gives an E AG /E GC value of 2.2 which is sufficient to minimise the collection of electrons at the grid. The resistor chain in the grid circuit also steps down the voltage evenly between the shaping rings. There are 12 rings spaced 1cm apart. There are also 1cm gaps between the window and the first ring, the last ring and the grid and the grid and the anode, giving a total detector length of 14cm. The window to grid distance is therefore 13cm. The internal diameter of the rings is 15cm, which means there is a total active detector volume of 2297cm 3. The total volume of the chamber, including any freely connected piping, is approximately 14800cm 3. This volume is significant because it determines the mass of detector gas subjected to deterioration Pre-Amplifier The pre-amplifier used is an ev-5091 single charge-sensitive hybrid preamplifier made by ev products and modified in the lab to increase the gain from 1mV/pC to

28 1.5mV/pC. The capacitance fall time is 25µs [End11]. This relatively fast decay time minimises the dead time of the detector and reduces the likelihood of signal pile-up. However it also increases the ballistic deficit. The extent of this effect is examined in section 4. The pre-amplifier is housed inside the chamber for two reasons. Firstly by minimising its proximity to the anode there is less opportunity for charge to dissipate and/or pick up noise. Secondly the metal walls of the chamber act as a Faraday cage and help to shield the pre-amplifier from external noise sources such as radio waves. 3.3 Signal Processing Raw signals The output signal from the preamplifier is a few hundred mv high and rises rapidly over a microsecond scale. A typical signal as seen on the oscilloscope is shown in Figure 3.4. Figure 3.4: Typical preamplifier output as seen on scope. Each square represents a height of 200mV and a width of 4µs.

29 SECTION 3: EXPERIMENTAL SETUP 29 The signal is the rapidly rising section of the trace. The tail which follows is simply the capacitor decay of the pre-amplifier. Identifying the rise time of the peak is made more difficult by the smoothing of the signal at the start and end of the pulse (a consequence of component capacitance). A reasonable measurement to make is therefore the time taken for the amplitude to go from 10 to 90 percent of its full height. This is the part of the signal identified in Figure 3.4 by two dotted lines. It was observed in section 2 that the signal should form the mirror image of the fission fragment s Bragg curve as it ionises the detector gas. This is indeed the case, however fission fragments constitute relatively low energy radiation and rather than seeing the characteristic curve shown in Figure 2.3 we instead simply see the final part of the curve as the particle is brought to an abrupt halt. Studying such a small section of the particle s Bragg curve makes Z identification well beyond the scope of this experiment Signal Amplification In Figure 3.4 the scope has been triggered on a single raw signal, straight from the preamplifier. This is then further processed by a shaping amplifier which differentiates the signal to iron out low frequency noise, and integrates it to negate the effects of high frequency noise. This is important as the signals are subjected to significant voltage variations resulting from mains interference and radio signals penetrating the chamber casing. Selecting an appropriate shaping time ensures the amplification is linear Energy Spectra Amplified signals can then be measured and counted by a data acquisition card connected to a PC. This can be presented graphically as a fission fragment energy spectrum such as shown in Figure 3.5.

30 Figure 3.5: Observed energy spectrum of Cf 252 in 85mbar freon. Tests run with an anode voltage of 1000V. This shows the nature of the Cf 252 fission fragment radiation. Without calibration the channel allocation is arbitrary, but comparison with an equivalent spectrum in the literature [Kno99] reveals that the higher of the two peaks represents Palladium- 106 nuclei. Fission in the Californium source creates pairs of smaller nuclei which are emitted in opposite directions with opposite momentum. Fission events favour emission of a heavy particle and light particle rather than emission of equal fragments [Kra88]. With equal momentum the lighter particles will carry more kinetic energy, resulting in the dual peaked energy distribution shown in Figure 3.5. The two peaks correspond to the most commonly emitted light particles and heavy particles respectively. Studying the energy spectrum is a useful way of checking that the detector is working as expected Signal Identification and Measurement As previously mentioned, without a TOF setup to measure mass or sufficient information about the Bragg curve to identify Z, the Bragg chamber is not capable of identifying fission fragments. However to make comparisons of signal quality

31 SECTION 3: EXPERIMENTAL SETUP 31 across different gases and conditions the same particle must be observed each time. This is achieved through the setup shown below. Figure 3.6: Diagram of signal analysis setup. The oscilloscope used is a Tekronix TDS 3054C model with a 500 MHz bandwidth and a 5.0GS/s sample rate. The amplifier is an Ortec 571 model, the PSA is an ORTEC 552 model and the gating unit is a Canberra 1454 model. Raw signals such as that shown in Figure 3.4 are amplified. The amplified signals are then sent to both a Pulse Shape Analyser unit and a linear gate unit. If the amplitude of an incoming signal is within a user-defined range then the PSA will emit a pulse which then serves as a coincidence trigger for the linear gate unit. This unit is set to coincidence gating and therefore only transmits input signals if the PSA has deemed it to be within the acceptable range. By observing the developing spectrum on the PC, the PSA can be tuned such that only the light particle peak is transmitted. By also triggering the oscilloscope on this signal, the raw signal can be observed with the knowledge that it is Pd 106. The stages of this signal identification process are shown as they appear on the scope in Figure 3.7.

32 Figure 3.7: All four signal stages shown simultaneously on true timescale. Signal 1 is the raw signal, signal 2 is the amplified signal, signal 3 is the PSA output signal used as the coincidence gate and signal 4 is the accepted signal used as the scope trigger. Note that signal 4 preserves the amplitude of 2. The scope can then be used to measure both the rise time and amplitude of the raw signal.

33

34 4. Results 4.1 Gas deterioration Initial tests sought to analyse the deterioration of the detector gases over time. This was observed through repeated measurements of both amplitude and rise time with a constant electric field and pressure. Figure 4.1: Variation of signal size with age of gas. Tests run with isobutane at 80mbars and an anode voltage of 1000v and freon at 85mbars and an anode voltage of 1000v. Figure 4.1 shows the amplitude of measured signals in relation to the age of the gas, i.e. how long it has been in the chamber and subjected to radiation. With isobutane there is a clear, apparently linear reduction in signal size over the timeframe shown, to the extent that around a fifth of the signal is lost over a 24

35 SECTION 4: RESULTS 35 hour period. More specifically the signal size is reduced by (±0.0047) mv per minute, or 4.87mV per hour. Extrapolating backwards indicates the detector is capable under these conditions of achieving signal sizes of 589.1(±5.0) mv with completely fresh gas. The percentage losses over a minute and an hour are therefore % (±0.0008%) and 0.824% (±0.049%) respectively. Experiments which use isobutane for long periods of time generally employ a continuous gas flow system. Typically this might ensure the isobutane is refreshed on average every 3 hours, which from these results would mean a percentage signal loss of approximately 2.47% (±0.15%). These results are comparable to those found by A. Oed et. al. [OGG + 82] who determined a deterioration rate of approximately 1% signal amplitude loss per hour. It must be considered that the observed change in signal size may not be entirely down to gas deterioration. If there exists a small leak in the chamber then air may enter and reduce the purity of the gas. However, close monitoring of the pressure shows no change throughout each experimental run. Furthermore, the observation that freon does not lose signal height suggests that the changes seen are characteristic of the isobutane. In further tests (i.e. tests observing characteristics of the gas other than its deterioration with time) the isobutane used during this experiment has invariably been at different levels of freshness for each measurement. The resulting signal losses have therefore been accounted for by corrections to each measurement, i.e. the corrected amplitude is given by: Equation 4.1 (Where T is the time (in minutes) for which the gas has been in the chamber.) This formula is well supported by its effect on further data sets. When applied to other plots (such as signal size vs. reduced field strength), corrected data points fit significantly better than measured data points, particularly those which used older

36 isobutane. However since there is an error associated with this correction results are optimised by keeping T as low as possible i.e. regularly refreshing the gas. Freon meanwhile demonstrates good stability, showing far less deterioration over time. However the initial amplitude is inferior to that seen with isobutane, suggesting that provided the gases are less than approximately 23 hours old, isobutane offers larger signals and hence a better signal to noise ratio. Furthermore, in Figure 4.2 we see an apparent increase in rise time when using freon, an effect not seen with Isobutane. Figure 4.2: Variation of rise time with age of gas. Tests run with isobutane at 80mbars and an anode voltage of 1000v and freon at 85mbars and an anode voltage of 1000v. The limited data makes this difficult to analyse, but again the observed trend is fitting with other data sets where rise times have been adjusted accordingly. Using corrections based on Figure 4.2 improves the strength of observed correlations in the data. The rise time increase is measured as (± ) µs per minute, or (± ) µs per hour. Taking the fresh rise time value ( ) as (±0.007) µs these correspond to percentage increases of % (±0.0018%) per minute and 0.967% (±0.106%) per hour. The correction is therefore:

37 SECTION 4: RESULTS 37 Equation 4.2 The size of the associated error makes it important to keep the freon reasonably fresh. The 106μCi activity rate of the source equates to approximately 7800 particles reaching the detector every second. Approximately 3% of these will be alpha particles (REFERENCE nuclear info) which, with the chamber at the pressures used, will be stopped in the anode rather than the gas. However, even assuming that all the particles deposit all their energy in the gas they would only ionise something in the region of molecules over a 24 hour period less than % of the gas. It would be naïve then to assume that the changes to the signals are consequences of radiation damage. It is likely that the deterioration is in fact due to contamination of the gas from the chamber itself, either from dust or material out-gassing. Even in very small quantities this may significantly affect the gas properties, depending on how sensitive the ionisation, recombination and mobility mechanisms are to impurities. To gain further insight into this phenomenon a wider range of gases might be studied to determine how this deterioration depends upon properties such as ionisation potential, electron affinity and molecular mass. Repeating the experiment in a different chamber with different components could show the extent to which the changes are dependent on the gas surroundings. 4.2 Pressure dependence Changing the pressure in the chamber has two effects on fragment detection. Firstly the stopping power of the gas is increased, reducing the range of the particles. Secondly the drift speed of the electrons is reduced. These changes have opposite effects upon the rise time a reduced range shortens the window over which electrons must be collected reducing the rise time, whilst a slower drift speed lengthens the rise time. Figure 4.3 shows that the two gases have varying sensitivity

38 to these effects, with the rise time increasing with pressure in the case of isobutane and decreasing with pressure in the case of freon. Figure 4.3: Variation of rise time with pressure. Tests run with an anode voltage of 1000V. To observe the effect of pressure changes on just the drift speed, the range variation must be accounted for. To achieve this, a linear relationship between pressure and range is assumed and the product of rise time and pressure is now plotted. This can be plotted against the reduced electric field (E/P) to generalise the results as shown in Figure 4.4.

39 SECTION 4: RESULTS 39 Figure 4.4: Rise time pressure against reduced electric field. Tests run with an anode voltage of 1000V. We see that isobutane demonstrates a greater sensitivity to the reduced field size over this range of conditions (i.e mbars at 66Vcm -1 ). It is likely however that over a wider range the trend becomes non-linear and the gradients in Figure 4.4 are only representative of the gases under these reduced fields. 4.3 Range determination To calculate the drift velocity of the electrons we must first determine the range of the particles in the gas. With the setup used there is no way of directly identifying this, and as such the range values rely somewhat on SRIM calculations. However it is possible to determine the minimum pressure, (the pressure below which the range exceeds the chamber length). It should be considered that whilst the full chamber length is 13cm there is an additional 1cm gap between the Frisch grid and the anode. This region should be avoided as electrons ionised here will create less charge on the anode than those which have drifted the full 1cm gap. However,

40 since this would only affect the tail of the signal where the ionisation density is lowest, noticeable changes to the signal are not seen until the pressure is low enough for particles to reach the anode itself. Range simulations using SRIM predict the pressure dependences shown in Figure 4.5. Figure 4.5: SRIM calculated ranges for 99MeV Pd 106 fragments into gas. Note that fragment energy has been reduced to account for energy losses in entrance window. From this we should expect significant changes to the signals to occur at pressures of 52mbars and 66mbars for isobutane and freon respectively, with Frisch grid related deviations potentially occurring at 56mbars and 71mbars. Figure 4.6 shows the changes seen to signal size.

41 SECTION 4: RESULTS 41 Figure 4.6: Variation of signal amplitude with pressure. Tests run with an anode voltage of 1000V. In both cases we see the signal drop at pressures well above that predicted by the SRIM simulations below approximately 78mbars for isobutane and 82mbars for freon, suggesting the fragments are exhibiting a longer range than predicted. There are two pieces of evidence to support this finding. Firstly, studying the energy spectra of tests conducted below these pressures shows a shift in percentage yield from light to heavy particles. An example is shown in Figure 4.7.

42 Figure 4.7: Observed energy spectrum of Cf 252 in 70mbars of isobutane. Tests run with an anode voltage of 1000V. Unlike in Figure 3.5, Figure 4.7 shows a higher ratio of heavy to light particles, strongly suggesting that many of the lighter particles (with greater range than the heavy particles) are not depositing all their energy before reaching the anode. Secondly, at these lower pressures there is an observed lack of amplitude dependence on signal rise times. Although the electronics setup is designed to restrict the signals to just that of the energy corresponding to the light particle peak, there is still a small tolerance (this variation constitutes the most significant contribution to the experimental energy resolution.) Plotting the individual signal rise times against their amplitudes shows a positive correlation, with more energetic particles creating a longer track and hence the detector taking more time to collect the ionised charge. However at pressures below 78mbars and 82mbars for isobutane and freon respectively this trend ceases to exist and the signal rise times appear constant and independent of the amplitude. This indicates that more energetic particles are ionising the same track-length as less energetic particles, which would be the case if they are were reaching the anode.

43 SECTION 4: RESULTS 43 Assuming then that this data is correct, the predicted ranges are 67% (in the case of isobutane) and 80% (in the case of freon ) of that observed. This discrepancy may be due to problems with the SRIM model under the conditions used in this experiment. By using these as reference points the SRIM predictions can be calibrated allowing range estimates to be made. Figure 4.8 shows the predicted SRIM ranges alongside the same data calibrated to my results. Figure 4.8: SRIM calculated ranges for 99MeV Pd 106 fragments into gas alongside calibrated ranges. The pressures actually used for observing the electric field dependence and drift speed patterns were 80mbars for isobutane and 85mbars for freon. From Figure 4.8 the actual range of the particles can be estimated to be approximately 13.5cm in isobutane and 13.6cm in freon at these pressures. This means that in both cases some electrons will actually be ionised behind the Frisch grid, but the charge loss will be small and will only affect the electrons ionised at the very end of the particle s track. The particle s ionisation rate is low in this region, making any overall signal loss insignificant. 4.4 Electric field dependence

44 Increasing the anode (and grid) voltage raises the electric field, resulting in faster collection of the electrons and hence faster rise times. This is shown in Figure 4.9. It is plotted as the product of rise time and pressure against the reduced field to allow comparison with Figure 4.4. Figure 4.9: Rise time pressure against reduced electric field. Tests run with either a fixed anode voltage of 1000V or fixed pressure of 80mbars (isobutane)/85mbars (freon ). The results match that seen in Figure 4.4, confirming that the trend is non-linear and that the drift speed is dependent on both E and P. In addition, the faster electron collection reduces the likelihood of charge being lost through effects such as recombination and electron attachment resulting in larger signals. This is shown in Figure 4.10.

The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors.

The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors. The Effect of Different Fill Gases on Bragg Curve Spectrometer Detectors. A dissertation submitted to the University of Manchester for the degree of Master of Science by Research in the Faculty of Engineering

More information

Technical Manual Ionisation Chamber MUSIC80

Technical Manual Ionisation Chamber MUSIC80 Technical Manual Ionisation Chamber MUSIC80 Entwicklungsbüro Dr. R. Schneider Lärchenstraße 9 85640 Putzbrunn Tel: 089 608 533 71 Fax: 089 608 533 72 E-Mail: robert@schneider.net Entwicklungsbüro Andreas

More information

Gaseous detectors. 1

Gaseous detectors.  1 Gaseous detectors Very important class of detectors with many applications charged particle, x- & gamma rays, visible light detection applications in many areas of research, & commercial particle & nuclear

More information

Antoine Drouart DSM/DAPNIA/SPhN. Köln, 5 th of December,

Antoine Drouart DSM/DAPNIA/SPhN. Köln, 5 th of December, Köln, 5 th of December, 2007 1 Basic CATS/BTD mechanics Köln, 5 th of December, 2007 2 Beam Tracking Today -Detector able to cope with a few 10 5 pps -Spatial resolution (FWHM) : 0.7mm -Time resolution

More information

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg.

February 12, PHY357 Lecture 11. Experimental Methods. Accelerators. Particle Interactions. Particle Detectors. Full experiment (eg. PHY357 Lecture 11 Experimental Methods Accelerators Particle Interactions Particle Detectors Full experiment (eg. ATLAS) Introduction! Several different general classes of particle detectors (sensors)!

More information

Experimental Particle Physics PHYS6011 Joel Goldstein, RAL

Experimental Particle Physics PHYS6011 Joel Goldstein, RAL Experimental Particle Physics PHYS6011 Joel Goldstein, RAL 1. Introduction & Accelerators 2. Particle Interactions and Detectors (1/2) 3. Collider Experiments 4. Data Analysis Charged Particle Detectors

More information

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21 Today s Outline - September 12, 2016 C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016 1 / 21 Today s Outline - September 12, 2016 Detectors C. Segre (IIT) PHYS 570 - Fall 2016 September 12, 2016

More information

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery Trainee: TRAINEE Module: Radiotherapy Physics Competency:

More information

Aging Analysis of Micromegas Detectors for ATLAS New Small Wheel

Aging Analysis of Micromegas Detectors for ATLAS New Small Wheel Aging Analysis of Micromegas Detectors for ATLAS New Small Wheel Melissa Quinnan August 15, 2015 Supervisor: Michele Bianco Abstract In preparation for the coming High Luminosity Large Hadron Collider

More information

LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S)

LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S) LARGE VOLUME HEMISPHERICAL NUCLEAR RADIATION DETECTOR CZT/500(S) Ganibu Dambis 26, P.O.Box 33 Riga, LV-1005, LATVIA Tel. (+371)6738-3947 Fax:(+371)6738-2620 office@bsi.lv www.bsi.lv CONTENTS INTRODUCTION...

More information

Texas active target (TexAT) detector part 1: Design and construction progress

Texas active target (TexAT) detector part 1: Design and construction progress Texas active target (TexAT) detector part 1: Design and construction progress E. Koshchiy, G.V. Rogachev, E. Uberseder, and E. Pollacco 1 1 IRFU, CEA Saclay,Gif-Sur-Ivette, France Construction of a general

More information

TIMS TIMS. Thermal Ionisation Mass Spectrometry.

TIMS TIMS. Thermal Ionisation Mass Spectrometry. TIMS Thermal Ionisation Mass Spectrometry www.nu-ins.com Incorporating our field-proven Zoom lens multi-collector technology, the TIMS improves the versatility and overall performance of this long established

More information

Computer Modelling and Simulation of a Smart Water Heater

Computer Modelling and Simulation of a Smart Water Heater Computer Modelling and Simulation of a Smart Water Heater Maria Kathleen Ellul University of Malta ellul_maria@yahoo.com Adrian Muscat University of Malta adrian.muscat@um.edu.mt Abstract A computational

More information

Aging measurements on triple-gem detectors operated with CF 4 based gas mixtures

Aging measurements on triple-gem detectors operated with CF 4 based gas mixtures IEEE, Roma, October 2004 Aging measurements on triple-gem detectors operated with CF 4 based gas mixtures M. Alfonsi 1, G. Bencivenni 1, W. Bonivento 2,A.Cardini 2, P. de Simone 1, F.Murtas 1, D. Pinci

More information

Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency.

Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency. Thick-Film Heater achieves Superior Performance in Thermal Response, Uniformity and Efficiency. by Mary Ruggiero, P.Eng., PhD and John Stockton, P.Eng Abstract IntegrAL thick-film heaters by Datec Corporation

More information

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110 TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110 APPLICATION NOTE CC-107 Introduction This purpose of this document is to detail the advanced, state of the art features TSI has incorporated in the design

More information

Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment

Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use in the JET First Trace Tritium Experiment EFDA JET CP(04)05-07 L.B.C. Worth, R.J.H. Pearce, J. Bruce, J. Banks, S. Scales and JET EFDA Contributors Development of a Novel Contamination Resistant Ionchamber for Process Tritium Measurement and use

More information

Temperature dependent Components within the Pyroelectric Detector

Temperature dependent Components within the Pyroelectric Detector 1 Introduction This application note is offered to provide some insight into how temperature affects the various components of a typical pyroelectric detector and how the effects of ambient temperature

More information

Development of Motor Fan Noise Prediction Method in Consideration of Operating Temperature during Engine Idling

Development of Motor Fan Noise Prediction Method in Consideration of Operating Temperature during Engine Idling New technologies Development of Motor Fan Noise Prediction Method in Consideration of Operating Temperature during Engine Idling Yasuhito Suzuki* Masahiro Shimizu* Abstract In these years there is an increasing

More information

Alpha Attenuation Due to Dust Loading. Amber E. Dailey Francis Marion University Savannah River Site

Alpha Attenuation Due to Dust Loading. Amber E. Dailey Francis Marion University Savannah River Site Alpha Attenuation Due to Dust Loading Amber E. Dailey Francis Marion University Savannah River Site Purpose The purpose of this experiment was to properly correct for the attenuation of the energy of alpha

More information

PHYSICS OF FOIL HEAT GAIN/LOSS IN BUILDINGS

PHYSICS OF FOIL HEAT GAIN/LOSS IN BUILDINGS PHYSICS OF FOIL HEAT GAIN/LOSS IN BUILDINGS There are three modes of heat transfer: CONDUCTION, CONVECTION, and RADIATION (INFRA-RED). Of the three, radiation is the primary mode; conduction and convection

More information

The LHCb Outer Tracker: Production & Ageing studies

The LHCb Outer Tracker: Production & Ageing studies The LHCb Outer Tracker: Production & Ageing studies Kaffeepalaver MPI-K Physikalisches Institut Physikalisches Institut 1 LHC at CERN p-p collisions beam energy 7 TeV 8.6km Four experiments: Atlas, CMS,

More information

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS Steve Glander: Senior Sales Engineer All Weather, Inc. 1165 National Dr. Sacramento, CA

More information

Measurement system for detection of Radon in building

Measurement system for detection of Radon in building Measurement system for detection of Radon in building Outline Properties of Radon Detection of ionised particles Measurement system for radon Properties of Radon Radon is a gaseous isotope ( 222 Rn) with

More information

ICONE DRAFT Proceedings of the 16th International Conference on Nuclear Engineering ICONE16 May 11-15, 2008, Orlando, Florida, USA

ICONE DRAFT Proceedings of the 16th International Conference on Nuclear Engineering ICONE16 May 11-15, 2008, Orlando, Florida, USA DRAFT Proceedings of the 16th International Conference on Nuclear Engineering ICONE16 May 11-15, 2008, Orlando, Florida, USA ICONE16-48187 THERMAL CYCLE EVALUATION FOR FEEDWATER HEATER OUT OF SERVICE CONDITION

More information

Reducing the Carbon Footprint of Existing Domestic Heating: A Non-Disruptive Approach

Reducing the Carbon Footprint of Existing Domestic Heating: A Non-Disruptive Approach EEDAL 2009 16-18 June 2009 Reducing the Footprint of Existing Domestic Heating: A Non-Disruptive Approach Martin O Hara Danfoss Randall Limited Abstract There is insufficient time between today and 2020

More information

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS EML-624 A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS FOR HOMELAND SECURITY PURPOSES Paul Bailey Environmental Measurements Laboratory U.S. Department of Homeland Security 201 Varick Street,

More information

Off Design Operation of Hybrid Noncondensable Gas Removal Systems for Flash Steam Cycle Geothermal Power Plants

Off Design Operation of Hybrid Noncondensable Gas Removal Systems for Flash Steam Cycle Geothermal Power Plants GRC Transactions, Vol. 40, 2016 Off Design Operation of Hybrid Noncondensable Gas Removal Systems for Flash Steam Cycle Geothermal Power Plants Jason Project Engineer, Vooner FloGard Corporation, Charlotte

More information

Pyroelectric Infrared Sensor PIR

Pyroelectric Infrared Sensor PIR Pyroelectric Infrared Sensor PIR Infrared radiation is the electromagnetic waves in the wavelength region longer than the visible light wavelengths, lying from 0.75 μm to 1000 μm. The wavelength region

More information

Important Considerations When Selecting a Fan for Forced Air Cooling. By: Jeff Smoot, CUI Inc

Important Considerations When Selecting a Fan for Forced Air Cooling. By: Jeff Smoot, CUI Inc Important Considerations When Selecting a Fan for Forced Air Cooling By: Jeff Smoot, CUI Inc Designing an appropriate thermal management solution requires a systemic approach; each component on a circuit

More information

Conductive Heating Technologies for Medical Diagnostic Equipment. Russell Strehlow, Manager of Research and Development, Minco

Conductive Heating Technologies for Medical Diagnostic Equipment. Russell Strehlow, Manager of Research and Development, Minco Conductive Heating Technologies for Medical Diagnostic Equipment Russell Strehlow, Manager of Research and Development, Minco Table of Contents Abstract.............................................................................

More information

Status Report about the TPC Detector and Module at CEPC

Status Report about the TPC Detector and Module at CEPC Status Report about the TPC Detector and Module at CEPC Huirong QI On behalf of CEPC TPC subgroup, IHEP, CAS 2016.01.19, IAS, HK Content Goals and options Some considerations Preliminary simulation Hybrid

More information

Drying principles and general considerations

Drying principles and general considerations Drying principles and general considerations Drying Mechanisms In the process of drying heat is necessary to evaporate moisture from the grain and a flow of air is needed to carry away the evaporated moisture.

More information

THERMAL PROCESSING THEORY

THERMAL PROCESSING THEORY THERMAL PROCESSING THEORY 9.1 Infrared Waves Infrared waves form part of the electromagnetic spectrum. Electromagnetic waves with wavelengths from 0.78 µm to 1000 µm are called infrared waves. You are

More information

CL4001 HEAT TRANSFER OPERATIONS

CL4001 HEAT TRANSFER OPERATIONS CL4001 HEAT TRANSFER OPERATIONS MODULE V Lecture Notes: Debasree Ghosh HEAT EXCHANGER Lecturer, Department of Chemical Engineering, Birla Institute of Technology, Mesra Heat Exchangers: Introduction and

More information

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points Topic 4 - Audio 46 Key Points : Measures taken to increase confidence in results. Quality Assurance: Program that assures that measures are being done and in a timely manner. Precision: How close measurements

More information

HYDROGEN DETECTION INSTRUMENTATION FOR THE SIAC HYDROGEN RECOMBINERS. Summary

HYDROGEN DETECTION INSTRUMENTATION FOR THE SIAC HYDROGEN RECOMBINERS. Summary SLAC -TN-67-21 D. R. Walz L. R. Lucas August 1967 not to be &hractad, quoted or rurther dl9c10sedwithout approval HYDROGEN DETECTION INSTRUMENTATION FOR THE SIAC HYDROGEN RECOMBINERS Summary We have examined

More information

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.

Humidity sensors. Behavior in different environments. Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9. Humidity sensors Behavior in different environments Supervisor: Vaclav Vacek Collaborators: Martin Doubek, Martin Janda and Michal Vitek 28.9.29 CZECH TECHNICAL UNIVERSITY IN PRAGUE CERN Content Content...

More information

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE

Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE 1 Omnisens DITEST TM FIBER OPTIC DISTRIBUTED TEMPERATURE & STRAIN SENSING TECHNIQUE Introduction Omnisens DITEST (Distributed Temperature and Strain sensing) is a distributed temperature and/or strain

More information

Optical Time Domain Reflectometry for the OMEGA EP Laser

Optical Time Domain Reflectometry for the OMEGA EP Laser Optical Time Domain Reflectometry for the OMEGA EP Laser Adam DeJager Greece Odyssey Academy Advisors: Samuel Morse and Richard Kidder Laboratory for Laser Energetics University of Rochester Summer High

More information

INTERNATIONAL JOURNAL OF RESEARCH GRANTHAALAYAH A knowledge Repository

INTERNATIONAL JOURNAL OF RESEARCH GRANTHAALAYAH A knowledge Repository DISTANT INFRARED SPECTRAL ANALYSIS OF INDUSTRIAL HOT GAS EJECTION IN ATMOSPHERE R.S. Asatryan *1, N.R. Khachatryan 2, H.S.Karayan 3 *1, 2, 3 Department of theoretical metrology, National institute of Metrology,

More information

OVERVIEW AND FUTURE DEVELOPMENT OF THE NEUTRON SENSOR SIGNAL SELF-VALIDATION (NSV) PROJECT

OVERVIEW AND FUTURE DEVELOPMENT OF THE NEUTRON SENSOR SIGNAL SELF-VALIDATION (NSV) PROJECT OVERVIEW AND FUTURE DEVELOPMENT OF THE NEUTRON SENSOR SIGNAL SELF-VALIDATION (NSV) PROJECT Jean-Christophe Trama, Alain Bourgerette, Eric Barat, Bernard Lescop LETI (CEA - Advanced Technologies) CEA/Saclay

More information

50(394) IEC. ore content meter ore sorting equipment overvoltage (of a Geiger-Müller county tube)...

50(394) IEC. ore content meter ore sorting equipment overvoltage (of a Geiger-Müller county tube)... INDEX A (absorbed) dose ratemeter... 394-11-17 absorption band (of a scintillator)... 394-18-08 acceptance test... 394-20-09 accident monitor... 394-12-27 activator... 394-10-08 actuated equipment... 394-14-05

More information

DFE T ECHNOLOGY. At Conceptronic, the design and management process for reflow soldering systems is called Dynamic Flow Engineering, or DFE.

DFE T ECHNOLOGY. At Conceptronic, the design and management process for reflow soldering systems is called Dynamic Flow Engineering, or DFE. DFE T ECHNOLOGY As components become smaller, component mixes become more diverse, and board densities continue to increase, the processes of packaging, interconnection, and assembly are challenging the

More information

FLIGHT UNLOADING IN ROTARY SUGAR DRYERS. P.F. BRITTON, P.A. SCHNEIDER and M.E. SHEEHAN. James Cook University

FLIGHT UNLOADING IN ROTARY SUGAR DRYERS. P.F. BRITTON, P.A. SCHNEIDER and M.E. SHEEHAN. James Cook University FLIGHT UNLOADING IN ROTARY SUGAR DRYERS By P.F. BRITTON, P.A. SCHNEIDER and M.E. SHEEHAN James Cook University Paul.Britton@jcu.edu.au, Phil.Schnieder@jcu.edu.au, Madoc.Sheehan@jcu.edu.au Keywords: Drying,

More information

Hygrometry. The art or science of humidity observation

Hygrometry. The art or science of humidity observation Hygrometry The art or science of humidity observation Humidity Under normal atmospheric conditions, the amount of water vapor any sample of air can hold depends primarily on its temperature Various terms

More information

Conceptual Design of a Better Heat Pump Compressor for Northern Climates

Conceptual Design of a Better Heat Pump Compressor for Northern Climates Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 1976 Conceptual Design of a Better Heat Pump Compressor for Northern Climates D. Squarer

More information

Press Release. How can the efficiency of the dryer section be increased? Dryer Section All Paper Grades. Heimbach wherever paper is made.

Press Release. How can the efficiency of the dryer section be increased? Dryer Section All Paper Grades. Heimbach wherever paper is made. Dryer Section All Paper Grades Press Release How can the efficiency of the T. Bock (Dipl.-Ing.), Manager Application & Technical Service, Heimbach GmbH & Co. KG, thomas.bock@heimbach.com I. Durniok (Dipl.-Ing.),

More information

PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR

PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR PERFORMANCE EVALUATION OF THE MINIATRIZED CATALYTIC COMBUSTION TYPE HYDROGEN SENSOR Hiroshi Miyazaki 1, Hiromasa Tanjyoh 2, Kengo Suzuki 3, Tomoaki Iwami 4 1 New cosmos electric, co., ltd., Osaka, 543-0036,

More information

Optimizing Chemistry and Coater Settings for Faster Drying

Optimizing Chemistry and Coater Settings for Faster Drying Optimizing Chemistry and Coater Settings for Faster Drying David Keely, Field Tech Service Manager, The Dow Chemical Company, Louisville, KY Dr. Bill Griffith, Principal Scientist, The Dow Chemical Company,

More information

A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2. Abstract

A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2. Abstract A GEM-based detector for detection and imaging sparks and flames G. Volpe 1 and V. Peskov 2 1 Dipartimento Interateneo di Fisica \M. Merlin" and Sezione INFN, Bari, Italy 2 CERN and Institute for Chemical

More information

CCD and CID solid-state detectors

CCD and CID solid-state detectors CCD and CID solid-state detectors Technical Overview 5110 ICP-OES Introduction Many of today s ICP-OES instruments have progressed from traditional photomultiplier tube (PMT) detection systems to solid-state

More information

Building and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes

Building and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes Building and Characterizing 14GHz InGaAs Fiber Coupled Photodiodes Gabrielle Inglis Advisor: Robert Boni Laboratory for Laser Energetics University of Rochester Summer High School Research Program 200

More information

UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation

UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation 15 th International Conference on Environmental Science and Technology Rhodes, Greece, 31 August to 2 September 2017 UVC Detection as a Potential for Alpha Particle Induced Air Fluorescence Localisation

More information

A THIN GOLD COATED HYDROGEN HEAT PIPE -CRYOGENIC TARGET FOR EXTERNAL EXPERIMENTS AT COSY

A THIN GOLD COATED HYDROGEN HEAT PIPE -CRYOGENIC TARGET FOR EXTERNAL EXPERIMENTS AT COSY 6 th Conference on Nuclear and Particle Physics 17-21 Nov. 2007 Luxor, Egypt A THIN GOLD COATED HYDROGEN HEAT PIPE -CRYOGENIC TARGET FOR EXTERNAL EXPERIMENTS AT COSY M. Abdel-Bary a, S. Abdel-Samad a,

More information

Commissioning of the Silicon Drift Detector of the ALICE experiment at the LHC

Commissioning of the Silicon Drift Detector of the ALICE experiment at the LHC Silicon Drift Detector of ALICE Commissioning of the Silicon Drift Detector of the ALICE experiment at the LHC Emanuele Biolcati for ALICE collaboration Dipartimento di Fisica dell Università di Torino

More information

Gas Detection in Process Industry The Practical Approach

Gas Detection in Process Industry The Practical Approach Gas Detection in Process Industry The Practical Approach Murtaza Gandhi, Jef Rowley, Derek Nelson, Casey Campbell (Baker Engineering and Risk Consultants, Inc., Houston, TX, USA) There are multiple gas

More information

BEKA ceiling heating versus traditional underfloor heating

BEKA ceiling heating versus traditional underfloor heating BEKA ceiling heating BEKA ceiling heating versus traditional underfloor heating 0. Preliminary remark In the future, rooms will be heated and cooled by surfaces providing radiant heating and cooling. The

More information

LHCb Rich Detectors Control and High Voltage Systems

LHCb Rich Detectors Control and High Voltage Systems LHCb Rich Detectors Control and High Voltage Systems Mario Sannino On behalf of LHCb RICH Group Rich 2007 Trieste 19.10.2007 1 Rich 2007 Trieste 19.10.2007 LHCb Rich Detectors Control and High Voltage

More information

BUILDING MANAGEMENT SYSTEMS. Introduction to Building Management Systems

BUILDING MANAGEMENT SYSTEMS. Introduction to Building Management Systems BUILDING MANAGEMENT SYSTEMS Introduction to Building Management Systems The Building Management System is used for all the functions carried out by the building including energy services security monitoring,

More information

Portable Survey Instruments NISP-RP-01

Portable Survey Instruments NISP-RP-01 NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection NISP-RP-01 This is an industry document for standardizing radiation protection processes used by supplemental radiation protection technicians.

More information

EVAPORATION NEEDS ENERGY

EVAPORATION NEEDS ENERGY Basic: EVAPORATION NEEDS ENERGY Drying cloth involves two processes. Energy has to be provided to change the water from liquid to vapour and an air stream is needed to remove the vapour. We discuss the

More information

Paper No. : 04 Paper Title : Unit Operations in Food processing Module 11 : Principles of Refrigeration

Paper No. : 04 Paper Title : Unit Operations in Food processing Module 11 : Principles of Refrigeration Paper No. : 04 Paper Title : Unit Operations in Food processing Module 11 : Principles of Refrigeration 11.1 Introduction Preservation of foods is a vital processing step in food processing. There are

More information

Music Sculptures of Glass and Gold

Music Sculptures of Glass and Gold Music Sculptures of Glass and Gold Studies on the Structural Vibration and Cavity Resonance Behavior of Various Loudspeaker Enclosures and Baffles Written by Kolja Willimzik RESEARCH AND DEVELOPMENT MANAGER

More information

THERMAL DIAGNOSIS OF MV SWITCHBOARDS: A COST-EFFECTIVE, DEPENDABLE SOLUTION BASED ON AN OPTICAL SENSOR

THERMAL DIAGNOSIS OF MV SWITCHBOARDS: A COST-EFFECTIVE, DEPENDABLE SOLUTION BASED ON AN OPTICAL SENSOR THERMAL DIAGNOSIS OF MV SWITCHBOARDS: A COST-EFFECTIVE, DEPENDABLE SOLUTION BASED ON AN OPTICAL SENSOR Christian PETIT Anticipation manager Schneider Electric - Network protection and control unit -Marketing

More information

Eliminating Condensation and Resulting Corrosion. Sulphur Tail Gas and De-Gas Lines

Eliminating Condensation and Resulting Corrosion. Sulphur Tail Gas and De-Gas Lines Eliminating Condensation and Resulting Corrosion In Sulphur Tail Gas and De-Gas Lines By David R. Hornbaker, Controls Southeast, Inc. and Graham M. MacPherson, P.Eng., Millenia Resource Consulting Abstract

More information

Griffin 824 Operational Suitability: Evaluation of Clear-Down Time for Mass Spectrometry- Based Trace Detector WHITE PAPER. Philip J. Tackett, Ph.D.

Griffin 824 Operational Suitability: Evaluation of Clear-Down Time for Mass Spectrometry- Based Trace Detector WHITE PAPER. Philip J. Tackett, Ph.D. WHITE PAPER Griffin 824 Operational Suitability: Evaluation of Clear-Down Time for Mass Spectrometry- Based Trace Detector INTRODUCTION Philip J. Tackett, Ph.D. 1 SYNOPSIS The following white paper describes

More information

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF

SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF 2015 SIEMENS ULTRAMAT 6 IR CARBON MONOXIDE ANALYZER METHOD VALIDATION FOR TESTING CARBON MONOXIDE IN NITROGEN, NF FINAL APPROVAL REVISION #: 0 ISSUE DATE: December 4, 2015 PAGE: 1 of 13 This Methods

More information

Basics. Conversion principle and parameters of pyroelectric detectors

Basics. Conversion principle and parameters of pyroelectric detectors Basics Conversion principle and parameters of pyroelectric detectors 1 Basic Principles of Pyroelectric Detectors The pyroelectric effect was already known in ancient times (Theophrastos of Eresos 4th

More information

AIR CONDITIONING. Carrier Corporation 2002 Cat. No

AIR CONDITIONING. Carrier Corporation 2002 Cat. No AIR CONDITIONING Carrier Corporation 2002 Cat. No. 020-016 1. This refresher course covers topics contained in the AIR CONDITIONING specialty section of the North American Technician Excellence (NATE)

More information

HYDROCARBONENGINEERING June June 2015

HYDROCARBONENGINEERING June June 2015 www.hydrocarbonengineering.com Controlled Combustion Arthur Groenbos, Yokogawa Europe, discusses tunable diode laser gas analysis for combustion management in fired heaters. While fired heaters are used

More information

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR

MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR MODEL-BASED OPTIMIZATION OF AN INFRARED GAS SENSOR ABSTRACT Ingo Sieber Forschungszentrum Karlsruhe P.O. Box 3640 76021 Karlsruhe Germany sieber@iai.fzk.de Manufacturing test structures of microsensors

More information

C-NRPP Quality Assurance Guidance for Radon Test Devices

C-NRPP Quality Assurance Guidance for Radon Test Devices C-NRPP Quality Assurance Guidance for Radon Test Devices Table of Contents: 1. Purpose 2. Definitions 3. Quality Assurance Plan 4. Calibration 5. Spikes/Performance Tests 6. Duplicates 7. Blanks 8. Cross-Checks

More information

Development of high sensitivity radon detectors

Development of high sensitivity radon detectors Nuclear Instruments and Methods in Physics Research A 421 (1999) 334 341 Development of high sensitivity radon detectors Y. Takeuchi *, K. Okumura, T. Kajita, S. Tasaka, H. Hori, M. Nemoto, H. Okazawa

More information

INFLUENCE OF SOLAR RADIATION AND VENTILATION CONDITIONS ON HEAT BALANCE AND THERMAL COMFORT CONDITIONS IN LIVING-ROOMS

INFLUENCE OF SOLAR RADIATION AND VENTILATION CONDITIONS ON HEAT BALANCE AND THERMAL COMFORT CONDITIONS IN LIVING-ROOMS INFLUENCE OF SOLAR RADIATION AND VENTILATION CONDITIONS ON HEAT BALANCE AND THERMAL COMFORT CONDITIONS IN LIVING-ROOMS Staņislavs GENDELIS, Andris JAKOVIČS Laboratory for mathematical modelling of environmental

More information

PRELIMINARY ANALYSIS OF THE NUMBER OF OCCUPANTS, FIRE GROWTH, DETECTION TIMES AND PRE-MOVEMENT TIMES FOR PROBABILISTIC RISK ASSESSMENT

PRELIMINARY ANALYSIS OF THE NUMBER OF OCCUPANTS, FIRE GROWTH, DETECTION TIMES AND PRE-MOVEMENT TIMES FOR PROBABILISTIC RISK ASSESSMENT PRELIMINARY ANALYSIS OF THE NUMBER OF OCCUPANTS, FIRE GROWTH, DETECTION TIMES AND PRE-MOVEMENT TIMES FOR PROBABILISTIC RISK ASSESSMENT DAVID CHARTERS DEAN MCGRAIL Arup Fire, 78 East Street, Leeds, LS9

More information

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland Available on CMS information server CMS CR -2016/298 The Compact Muon Solenoid Experiment Conference Report Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland 28 October 2016 (v2, 25 November 2016)

More information

Particle Detectors. Summer Student Lectures 2007 Werner Riegler, CERN, History of Instrumentation History of Particle Physics

Particle Detectors. Summer Student Lectures 2007 Werner Riegler, CERN, History of Instrumentation History of Particle Physics Particle Detectors Summer Student Lectures 2007 Werner Riegler, CERN, werner.riegler@cern.ch History of Instrumentation History of Particle Physics The Real World of Particles Interaction of Particles

More information

DSP EC Pro. Advanced, Digital Gamma-Ray Spectrometer for HPGe Detector Systems

DSP EC Pro. Advanced, Digital Gamma-Ray Spectrometer for HPGe Detector Systems Advanced, Digital Gamma-Ray Spectrometer for HPGe Detector Systems DSPEC Pro: Advancing the Art of Digital Gamma-Ray Spectroscopy The immense versatility of DSPEC Pro, brought about by the number of unique

More information

Smoldering Propagation Characteristics of Flexible Polyurethane Foam under Different Air Flow Rates

Smoldering Propagation Characteristics of Flexible Polyurethane Foam under Different Air Flow Rates Smoldering Propagation Characteristics of Flexible Polyurethane Foam under Different Air Flow Rates Zhou Y., Fei Y.*, Hu S. Q. Nanjing Tech University, College of Fire Safety Science and Engineering &

More information

Unit THE NATURE OF HEAT

Unit THE NATURE OF HEAT Unit 5.0 - THE NATURE OF HEAT Heat is a form of energy, in the form of infrared radiation. Heat from the sun travels through space at the speed of 300,000,000 m/s. Upon arriving on earth, much of the radiant

More information

Unit 2 Fire Safety (Zone 2)

Unit 2 Fire Safety (Zone 2) IFE Level 3 Diploma in Fire Safety and Fire Science Unit 2 Fire Safety (Zone 2) Examiner Report March 2017 Introduction 21% of the candidates that sat the examination attained a pass. Candidates did not

More information

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection

Title: Standard Operating Procedure for Measurement of Total Hydrocarbon Using a Flame Ionization Detection Procedure No: SOP-020 Revision No: 1.0 (January 21, 2011) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, and to

More information

L14 ENVIRONMENTAL CONTROL TECHNICAL DATA

L14 ENVIRONMENTAL CONTROL TECHNICAL DATA L14 ENVIRONMENTAL CONTROL Ever since components have been made to control electrotechnical tasks, heat loss has been a subject to take into consideration. Sometimes more sometimes less. Major problems

More information

PHYSICS FORM 5 TRANSFER OF THERMAL ENERGY

PHYSICS FORM 5 TRANSFER OF THERMAL ENERGY Heat energy is transferred from one place to the next by three mechanisms: 1. Conduction 2. Convection 3. Radiation Conduction This is the process of heat transfer from one place to another using the movement/vibration

More information

SILICON DRIFT DETECTORS

SILICON DRIFT DETECTORS SILICON DRIFT DETECTORS PRODUCT BROCHURE 2 JUNE 2018 About PNDetector PNDetector is well-known as a manufacturer of high performance radiation detectors for microanalysis, quality assurance and material

More information

Compression of Fins pipe and simple Heat pipe Using CFD

Compression of Fins pipe and simple Heat pipe Using CFD Compression of Fins pipe and simple Heat pipe Using CFD 1. Prof.Bhoodev Mudgal 2. Prof. Gaurav Bhadoriya (e-mail-devmudgal.mudgal@gmail.com) ABSTRACT The aim of this paper is to identify the advantages

More information

PoS(ICHEP2012)516. The status of the CMS pixel upgrade detector

PoS(ICHEP2012)516. The status of the CMS pixel upgrade detector University of California Riverside E-mail: Ferdinando.Giordano@cern.ch The present CMS pixel detector was designed for a maximum luminosity of 34 cm s. Following the Phase upgrade of the LHC, the peak

More information

SPECTRAL INSTRUMENTATION

SPECTRAL INSTRUMENTATION SPECTRAL INSTRUMENTATION The typical setup of an IR experiment includes a light source, a sample, (a microscope), the spectral apparatus, a detector and a computer. IR experiment A light source emits polychromatic

More information

An Introduction to Mass Spectrometry

An Introduction to Mass Spectrometry An Introduction to Mass Spectrometry Guy Wilson UPC 2 & Purification Business Development, Northern Europe, Waters Ltd 2013 Waters Corporation 1 What is a Mass Spectrometer? Data System Sample Introduction

More information

ZONE MODEL VERIFICATION BY ELECTRIC HEATER

ZONE MODEL VERIFICATION BY ELECTRIC HEATER , Volume 6, Number 4, p.284-290, 2004 ZONE MODEL VERIFICATION BY ELECTRIC HEATER Y.T. Chan Department of Building Services Engineering, The Hong Kong Polytechnic University, Hong Kong, China ABSTRACT Selecting

More information

NTC THERMISTOR SENSOR PERFORMANCE

NTC THERMISTOR SENSOR PERFORMANCE NTC THERMISTOR SENSOR PERFORMANCE Accuracy, Interchangeability, Beta Tolerance, Tolerance Comparison, Stability, Drift, and Moisture Induced Failure When designing the best temperature probe for a specific

More information

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team INTRODUCTION The OTDR is a very efficient tool for characterizing the elements on

More information

The FEL Facility Optical Diagnostics and FEL Characterization

The FEL Facility Optical Diagnostics and FEL Characterization The FEL Facility Optical Diagnostics and FEL Characterization M. Shinn, C. Behre, S. Benson, J. Coleman, R. Evans, A. Grippo, J. Gubeli, D. Hardy, K. Jordan, G. Neil, S. Zhang G. Williams Jefferson Lab

More information

Finned Heat Sinks for Cooling Outdoor Electronics under Natural Convection

Finned Heat Sinks for Cooling Outdoor Electronics under Natural Convection Finned s for Cooling Outdoor Electronics under Natural Convection Lian-Tuu Yeh, Ph D & PE Thermal Consultant, Dallas, TX, 75252 USA Abstract For tower or poled mounted electronics, the heat sink weight

More information

J. R. Qualey III, L. Desmarais, J. Pratt Simplex Time Recorder Co., 100 Simplex Drive, Westminster, MA 01441

J. R. Qualey III, L. Desmarais, J. Pratt Simplex Time Recorder Co., 100 Simplex Drive, Westminster, MA 01441 J. R. Qualey III, L. Desmarais, J. Pratt Simplex Time Recorder Co., 100 Simplex Drive, Westminster, MA 01441 Response-Time Comparisons of Ionization and Photoelectric/Heat Detectors 1. Introduction Despite

More information

A Review of Hydronic Balancing of Cooling Water Circuit

A Review of Hydronic Balancing of Cooling Water Circuit A Review of Hydronic Balancing of Cooling Water Circuit Abhilash Shetty Department Of Mechanical Engineering SIES Graduate School of Technology, Nerul, Navi Mumbai-400706, Maharashtra, India Deepak Pandey

More information

LECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry

LECTURE 11. Dr. Teresa D. Golden University of North Texas Department of Chemistry LECTURE 11 Dr. Teresa D. Golden University of North Texas Department of Chemistry C. Detectors Properties of Detectors -quantum counting efficiency -linearity -energy proportionality -resolution C. Detectors

More information

CHAPTER I INTRODUCTION. In the modern life, electronic equipments have made their way

CHAPTER I INTRODUCTION. In the modern life, electronic equipments have made their way 1 CHAPTER I INTRODUCTION In the modern life, electronic equipments have made their way in to practically every part, which is from electronic gadgets to high power computers. Electronic components have

More information

TESTS OF ADSIL COATING

TESTS OF ADSIL COATING TESTS OF ADSIL COATING Test B - Long Term Test FSEC-CR-1259-01 July 11, 2001 Prepared for: Bob Suggs Florida Power & Light Company 9250 W. Flagler Street Miami, Florida 33174 Principal Investigator Dr.

More information