FastGene Plate Centrifuge. Instruction Manual. Catalog No. NG

Similar documents
MD-MINI Dry Bath. Instruction Manual

SAFEVIEW-MINI2 TRANSILLUMINATOR

Dry Bath Heating Systems

Digital Heat Block User Manual

Corning LSE Digital Dry Bath Heater

CR-1730R User s Manual

Pasto Sous vide water bath

Reacti-Therm I and III Heating Modules

Instruction Manual. For All Corning Hot Plates, Stirrers, Stirrer/Hot Plates with Digital Displays, and External Temperature Controller (6795PR)

INSTRUCTION MANUAL MNBX14

Instruction Manual. Model AccuPlate Hotplate AccuPlate Stirrers AccuPlate Hotplate Stirrer 12/10 CLS-EQ-LHP01REV

Reacti-Therm I and III Heating/Stirring Modules

Operation Manual Vortex mixer unitexer 1

OPERATIONS MANUAL. Hotplate Stirrer/Hotplate/Magnetic Stirrer Model HTS-1003/HTP-1002/MGS Certified

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1

ICE MAKER. 26 Lbs. Capacity HME010019N. Before using your Ice Maker, please read this manual.

Instruction Manual for the Revolutionary Science RS-IF-203 Incufridge

TTT Digital Incubator

Analog Melting Point Apparatus

Low Speed Centrifuge. Model XC-2415 Model XC-2450 Operating Manual

Centrifuge/Vortex for PCR plates CVP-2

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Corning LSE 6L Digital Water Bath

Instructions For Use ENG. LED Magnifier Lamp

12,000 BTU Evaporative Portable Air Conditioner

Magnetic Stirring Systems with Heat

Please Keep This Manual For Future Reference REV.01

Instruction Manual for the Revolutionary Science RS-PB-200 Dual Poly Pro Bath

BEFORE FIRST USE: TABLE OF CONTENTS

Tissue-Tek TEC Plus Cryo Console OPERATING MANUAL

INSTRUCTION MANUAL. Stirring hotplate (with probe and stand) Brand : (Dragon laboratory)

User Manual. MS7-H550-Pro MS7-H550-S MS7-Pro MS7-S

Aluminum Heating Mantles

OPERATION MANUAL. StableTemp Economy Mechanical Convection Oven. Model : ,-90,-91,-92,-93,-94

Feel Perfect. Heated Throw. Instruction Booklet. TR6300 Luxe Faux Fur. Please read these instructions carefully and retain for future reference.

MICRO-KJELDAHL EQUIPMENT ( MM SERIES ) OPERATING AND SAFETY INSTRUCTIONS

ELECTRIC FIREPLACE HEATER MODEL EF-BLT09 OWNER S MANUAL

General safety precautions English

Package Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents

NO VENTING REQUIRED 120V AC;

HI COD REACTOR and Test Tube Heater 2008 Series

BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

Refrigerated Incubator Model and Operating Instructions

Operator s Manual. The Bullet Blender 50 BB50-AU, BB50-DX. Contents. Congratulations!

Note To The User: This product has updated features not yet mentioned in the product manual.

PORTABLE AIR CONDITIONER OWNER S MANUAL

1 HP Air Mover. Owner s Manual

CORDLESS HAND VACUUM INSTRUCTION MANUAL

Isotemp Plus General Purpose Laboratory Freezers

English. Instruction and operation manual S 530. Leak detector for pneumatic systems

INSTRUCTION MANUAL. Laboratory Equipment Pty Ltd Ph: Fax:

General safety precautions English

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

CHEST FREEZER INSTRUCTION MANUAL. Model No.: EWCF5WBX EWCF7WBX

English. Instruction and operation manual S 530. Leak detector for pneumatic systems

Corning LSE Benchtop Shaking Incubator

FAN-FORCED UTILITY HEATER

pure indoor living Remote Control TOWER FAN MODEL: BTF4011AR-CN

OWNER S MANUAL ELECTRIC FIREPLACE INSERT. Model No.: WESTERN (EF42D-45D) WARNING CAUTION

Academy The Straightener IM.indd 1 24/06/ :10. Instruction Manual LSHS16

Operator s Manual. Histology Bath

OWNER S MANUAL. FLAT PANEL FIREPLACE HEATER Model Number: Serena (EF202A) WARNING CAUTION

English 6. Copyright DEWALT

INSTRUCTION MANUAL PUMP R103H UK / R Made in China.

INSTRUCTIONS AND WARRANTY

DEHUM3 INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE

ES-20 Orbital Shaker-Incubator

Handling or using the product improperly and in disregard of the instructions with this mark might result in serious bodily injury or death.

Operator s Manual. IP-100 Immersion Probe Cooler

StableTemp Hotplates, Stirrers and Stirring Hotplates

PORTABLE AIR CONDITIONER OWNER S MANUAL

Portable Refrigerator

275HVD NON-CONTACT, HIGH VOLTAGE DETECTOR DETECTOR DE ALTA TENSIÓN SIN CONTACTO

C L E A N I N G / M A I N T E N A N C E WA R R A N T Y I N F O R M AT I O N CLEANING/MAINTENANCE FAN STORAGE 3 YEAR LIMITED WARRANTY INSTRUCTIONS

Operator s Manual. The Quietek

Thermal Electric Wine & Beverage Chiller Model: WC-2463W. Instruction Manual

PROLUX AIR PURIFIER / HUMIDIFIER MODEL: PL-2 Operation Instruction

PRIMUS IMPORTANT. User Manual and Product Specifications. Portable Fridge/Freezer. Part No. PRI18LP - 18 Litre. Read instructions before operation.

OWNER'S GUIDE. Gas Continuous Flow Water Heaters. Owner's Information Warranty

BIONAIRE HEATER. Instruction Leaflet. pure indoor living MODEL: BCH4130. Read instructions before operating. Retain for future reference.

Installation Manual September Opticom Infrared System. Model 798 Emitter Bezel Mount Kit

Electron Microscopy Sciences. INSTRUCTIONAL MANUAL CAT & Step Up Slide Warmer

i-cube Microplate Incubator E10030

Instruction Manual for the Revolutionary Science RS-PB-100 Poly Pro Bath TABLE OF CONTENTS

23-IN Electric Logset

ELECTRIC STOVE HEATER Models: FS2213B

HSGM Heißschneide-Geräte

WALL MOUNT ELECTRIC FIREPLACE OWNERS MANUAL

1500 Watt OIL FILLER RADIATOR INSTRUCTION MANUAL. pure indoor living MODEL BOF2001-CN

Viper 2.6 Viper deluxe

Dehumidifier. User Manual

PORTABLE ICE MAKER. Model: IM-124S INSTRUCTION MANUAL

Micro-Computer Induction Cooktop SR-1885SS

ELECTRIC FIREPLACE INSERT MODEL EF-BLT10 OWNER S MANUAL

Digital 7 Square Plate Magnetic Stirrer

Sintrol Snifter A1+ Manual. Version 1.2.2

SFH136-CN TRUSTED FOR OVER 100 YEARS. Sunbeam Heater. Instruction Leaflet PLEASE READ AND SAVE THESE IMPORTANT INSTRUCTIONS

INSTRUCTIONS MANUAL. 1.8L Electric Kettle with Multi-Temp Feature SK-1800R (Red) SK-1800SS (Stainless Steel)

Operating Manual HD01

Introduction. Table of Contents

Transcription:

FastGene Plate Centrifuge Instruction Manual Catalog No. NG040 www.nippongenetics.eu info@nippongenetics.eu Version 01 20.01.2015

Content -1x Plate Centrifuge -2x Adapter for 96 well plates -1x Power Cord -1x Instruction Manual 1

Table of contents Content... 1 Warning... 3 Section 1 Introduction... 8 Section 2 Product Specifications... 8 Section 3 Installation Instructions... 9 Section 4 Device Operation Instructions... 9 4.1 Controls and Features... 9 4.2 Turn On the Instrument... 10 4.3 Operation Mode: Quick Start Mode... 10 4.4 Operation Mode: Time mode... 10 Section 5 Troubleshooting Guide... 11 5.1 Problem and Solution... 11 5.2 Maintenance... 11 Section 6 Ordering information... 11 Section 7 Warranty... 12 2

Warning FastGene Plate Centrifuge has been tested and found to comply with safety limits for the CE regulation. Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment. It is strongly recommended for the user to read the following points carefully before operating this equipment. 1. Read and follow carefully the manual instructions. 2. Do not alter the equipment. Failure to follow these directions could result in personal and/ or laboratory hazards, as well as invalidate equipment warranty. 3. Use a properly grounded electrical outlet with correct voltage and current handing capacity. 4. Disconnect from power supply before maintenance and servicing. Refer servicing to qualified personnel. 5. Never use this instrument series without having the safety cover correctly in position. 6. Do not use the unit if there is any sign of damage to the external tank or cover. Replace damaged parts. 7. Do not use in the presence of flammable or combustible material; fire or explosion may result. This device contains components which may ignite such materials. 8. Refer maintenance and servicing to qualified personnel. 9. Ensure that the system is connected to electrical service according to local and national electrical codes. Failure to properly connection may create fire or shock hazard. 10. Use appropriate materials and operate correctly to avoid possible hazards of explosion, implosion or release of toxic or flammable gases arising from overheated materials. 11. The unit shall be operated only by qualified personnel. 3

Safety Information Use high level of precaution against any electrical device. Before connecting the electrical supply, check to see if the supply voltage is within the range stated at the rating label, and see to it that the device be seated firmly. Place the unit in a safe and dry location; it must NOT touch the surrounding. Follow the safety precautions for chemicals / dangerous materials. If needed, please contact qualified service representative or info@nippongenetics.eu Environmental Conditions Ensure the instrument is installed and operated strictly in the following conditions: 1. Indoor use only 2. 95% RH (non-condensing) 3. 75 kpa 106 kpa 4. Altitude must not exceed 2000 meters 5. Ambient to 40 C operating temperature 6. Pollution degree: 2 7. Mains supply voltage fluctuations up to ±10% of the normal voltage Avoiding Electrical Shock Follow the guidelines below to ensure safe operation of the unit. FastGene Plate Centrifuge has been designed to use with shielded wires thus minimizing any potential shock hazard to the user. Nippon Genetics Europe recommends against the use of unshielded wires. To avoid electrical shock: 1. In the event of solution accidentally spilled into the instrument, it must be dried out for a period of time, at least 2 hours, and restored to NORMAL CONDITION before each operation. 2. NEVER connect or disconnect wire leads from the power jacks when the power is on. 3. WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus. 4. ALWAYS make sure that hands, work area, and instruments are clean and dry before making any connections or operating the equipments. 5. ONLY connect the power cord to a properly grounded AC outlet. Avoiding Damage to the Instrument 4

1. Do not attempt to operate the device if it is damaged. 2. Protect this unit from physical damage, corrosive agents and extreme temperatures (direct sunlight, etc.). 3. For proper ventilation and safety concerns, keep at least 10 cm of space behind the instrument, and at least 5 cm of space on each side. 4. Use high level of precautions against the damages on the unit. 5. Do not operate the unit out of environmental conditions addressed above. 6. Prior to apply any cleaning or decontamination method other than manufacturer s recommendation, users should check with the manufacturer s instruction to see if the proposed method will damage the equipment. Equipment Operation Follow the guidelines below to ensure safe operation of the unit: 1. Check the displayed temperature figure and external temp. Probe to see if it is overheating, and check if it will function in the case of a single fault at least once per day. 2. NEVER access dangerous chemicals or other materials to prevent possible hazard of explosion and damage. 3. A temporary conductivity caused by condensation might occur even though this series is rated Pollution Degree 2 in accordance with IEC 664. Battery Replacement Notice The product may contain an internal manganese dioxide. There is risk of fire and burns if the battery pack is not handled properly. To reduce the risk of personal injury: 1. Do not attempt to recharge the battery. 2. Do not expose to temperatures higher than 40 (104 ). 3. Do not disassemble, crush, puncture, short external contacts, or dispose of in fire or water. 4. Risk of explosion if battery is replaced by an incorrect type. Dispose of used batteries according to the instructions. 5. Replace only with the spare designated for this product. 6. Please understand the positive and negative mark. Do not use the battery positive and negative wrong. Symbols The symbols used on the FastGene Plate Centrifuge are explained below. 5

Indicates an area where a potential shock hazard may exist. Consult the manual to avoid possible personal injury or instrument damage. ATTENTION: Hot surface! Indicates disposal instruction. DO NOT throw this unit into a municipal trash bin when this unit has reached the end of its lifetime. To ensure utmost protection of the global environment and minimize pollution, please recycle this unit. 6

Potential Risk and Preventive Measure 1. Risk assessment table Potential Risk Frequency Bruise Burn Scald Drop Frequent Likely Possible Rare Unlikely Power cord plug wrong 2. Preventative measures of risk Potential Risk Bruise Burn Scald Drop Power cord plug wrong Preventive measures Do not put the machine near the table edge. Cover the lid or wear insulated gloves. Cover the lid to avoid spray water or wear insulated gloves. Place the unit on a flat and dry surface Observe correct adapter plug. 7

Section 1 Introduction The FastGene Plate Centrifuge is designed to meet the requirement of accurate and simply centrifugation of 96 or 384 well plates, small foot print and ease of use. FastGene Plate Centrifuge is capable of up to 480 x g. The microprocessor controller helps providing precise spinning control, whether you need a PCR or an ELISA plate being centrifuged. Small foot print Only a small lab bench space is needed Quick start Allows a one hand operation Section 2 Product Specifications Display Maximum Centrifugation force Timer Dimension Weight Rated Voltages Weight Lid LCD (illuminated) 480 x g 0-10 min L360 x W290 x H140mm 3,6 kg AC input 200V - 240V~; 2A, 50/60 Hz / DC input: +12V / 5A, 60W max. Approx. 1.3 kg Transparent Lid 8

Section 3 Installation Instructions The FastGene Plate Centrifuge is actually a pre-installed instrument. As long as it is placed on a sturdy and level surface in a safe, dry place, and is inserted with equally heavy plates, it is ready for operation. Section 4 Device Operation Instructions 4.1 Controls and Features Please refer to below figures on the following page for the location of the different keys. 9

4.2 Turn On the Instrument 1. Place Programmable FastGene Plate Centrifuge on a sturdy, level surface in a safe, dry place away from laboratory traffic. 2. Ensure that the AC power switch is OFF, and then plug the power cord into a grounded AC outlet of the appropriate voltage. 3. Switch the main power ON. 4.3 Operation Mode: Quick Start Mode 1. Switch the main power ON. 2. Press the RUN/STOP button to initiate the centrifugation 3. Stop pressing the RUN/STOP button to finish the centrifugation 4.4 Operation Mode: Time mode 1. Switch the main power ON. 2. Press the Set button to adjust to the desired time 3. Press the RUN/STOP button to initiate the centrifugation 4. Centrifugation will be stopped after finishing the countdown. 10

Section 5 Troubleshooting Guide Many operating problems may be solved by carefully reading and following the instructions in this manual accordingly. Some suggestions for troubleshooting are given below. Should these suggestions not resolve the problem, contact our SERVICE DEPARTMENT or a distributor in your region for assistance. If troubleshooting service is required, please include a full description of the problem. 5.1 Problem and Solution Problem Recommendations 1. Check the FUSE No signal on the screen Fan Error 2. Ensure that the AC power switch is ON 3. Check the power cord is properly plugged into a grounded AC outlet with the appropriate voltage Contact our service department or a distributor in your region. 5.2 Maintenance Centrifuge chamber and rotors may be cleaned with a moist cloth containing a mild soap solution. Section 6 Ordering information Cat. No. NG-040 Description FastGene Plate Centrifuge 11

Section 7 Warranty Nippon Genetics Europe warrants apparatus of its manufacture against defects in materials and workmanship, under normal service, for one year from the shipping date to purchaser. This warranty excludes damages resulting from shipping, misuse, carelessness, or neglect. Nippon Genetics Europe s liability under the warranty is limited to the receipt of reasonable proof by the customer that the defect is embraced within the terms of the warranty. All claims made under this warranty must be presented to Nippon Genetics Europe within one year following the date of delivery of the product to the customer. Manufacturer Nippon Genetics Europe GmbH Address: Binsfelder Straße 77 52351 Düren Germany Tel: +49 2421 2084690 Fax: +49 2421 2084 691 12