Available online Journal of Scientific and Engineering Research, 2017, 4(10): Research Article

Similar documents
Humidity Generation and Humidity Measurement The Complete Guide

An Introduction of Below-Freezing and Humidity Environment Controlling Technology Equipped on Brake Test Equipment

Constant climate chamber with humidity

Constant climate chambers

Constant climate chambers

Introduction Portable Meters Transmitter Probes RELATIVE HUMIDITY

Constant climate chambers

Lab 1b Cooling Tower Performance and Sensors for Thermal-Fluid Systems

Constant climate chamber with ICH-compliant illumination

Understanding total measurement uncertainty in power meters and detectors

Available online Journal of Scientific and Engineering Research, 2017, 4(12): Research Article

Constant climate chambers

Chapter 8 Balances on Nonreactive Processes PSYCHROMETRIC CHART

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000

3.4 Humidity. Types of instruments

CHAPTER 2 EXPERIMENTAL APPARATUS AND PROCEDURES

Constant climate chambers. KBF 240 (E5.2) - Constant climate chamber. Performance features and equipment:

PLANT GROWTH CHAMBER With CO2, Temperature & HumidityControl Technology

Climate chambers with illumination and humidity The BINDER KBWF series shines with its

Compression of Fins pipe and simple Heat pipe Using CFD

EXPERIMENTAL STUDY OF WATER DESALINATION BASED ON HUMIDIFICATION- DEHUMIDIFICATION PROCESS VIA HEAT PUMP

6 POINTS TO CONSIDER WHEN PURCHASING A CONSTANT CLIMATE CHAMBER IN THE PHARMACEUTICAL INDUSTRY BUYER S GUIDE

4th International Conference on Sensors, Measurement and Intelligent Materials (ICSMIM 2015)

Walk in Steam Curing Chamber for Concrete Application. Detailed Working Specifications

ME Mechanical Engineering Systems Laboratory. Experiment 3 - Mass and Energy Balances in Psychrometric Processes

Supertron Sensing Pte Ltd Certificate No. : LA C Blk 16 Kallang Place #07-04 Issue No. : 4 Singapore Date : 04 February 2015

Psychrometrics. Outline. Psychrometrics. What is psychrometrics? Psychrometrics in daily life and food industry Psychrometric chart

Constant climate chambers

Hygro and hygrothermo transducers (capacitive) for HVAC monitoring

Constant climate chambers

LEAKAGE MONITORING OF SEAL PLUG FOR INDIAN PRESSURIZED HEAVY WATER REACTOR BY PSYCHOMETRICS ANALYSIS METHOD

Experimental Investigation of a Multi Effect Membrane Based Regenerator for High Concentration Aqueous LiCL Solution as Desiccant

Experimental Study of Alternatives to Sand in Zeer Pot Refrigeration Technique

An experimental study of the impact of tunnel suppression on tunnel ventilation

For an ideal gas mixture, Dalton s law states that the sum of the partial pressures of the individual components is equal to the total pressure.

Introduction to humidistats

Hygrometry. The art or science of humidity observation

GE General Eastern Instruments

Performance Rating of Commercial and Industrial Humidifiers

ME 410 MECHANICAL ENGINEERING SYSTEMS LABORATORY MASS & ENERGY BALANCES IN PSYCHROMETRIC PROCESSES EXPERIMENT 3

DRYING STUDY OF EUCALYTPTUS STAIGERIANA LEAVES BY MONITORING THE HUMIDITY OF THE DRYER DISCHARGE

THE EFFECT OF HEAT LOAD ARRANGEMENT ON THE PERFORMANCE OF RADIANT PANELS

Solar Still - Improved Distillation

New Generation KBF 240 series

Available online Journal of Scientific and Engineering Research, 2018, 5(10): Research Article

Model MKFT 115 Dynamic climate chambers for rapid temperature changes with humidity control and extended low temperature range

D7660 ULTRASONIC HUMIDIFIER

Open and Closed Door Moisture Transport and Corresponding Energy Consumption in Household Refrigerator

AND AIR-CONDITIONING. Dr Ali Jawarneh Department of Mechanical Engineering Hashemite University

Available online at I-SEEC Proceeding - Science and Engineering (2013) 19 26

Optimization of Hybrid Noncondensable Gas Removal System for a Flash Steam Geothermal Power Plant

REVIEW ON CONDENSER AIR LEAK TEST

Model CB 220 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

CFD Analysis of temperature dissipation from a hollow metallic pipe through circular fins using Ansys 14.5

HATCHERY. ROSS TECH Hatchery Maintenance. August 2011

COOLING SYSTEMS FOR CONTINOUS GALVANIZING LINE Miroslav Raudensky a, Milan Hnizdil a, Jaroslav Horsky a, FrédéricMarmonier b

Energy Efficient Heat Exchangers for HVAC Applications

Numerical Stability Analysis of a Natural Circulation Steam Generator with a Non-uniform Heating Profile over the tube length

Performance Evaluation of a Plug-In Refrigeration System Running Under the Simultaneous Control of Compressor Speed and Expansion Valve Opening

Model 5190-F Temperature/Humidity Probe

Meteorology 432. Hygrometry Spring 2013

ZONE MODEL VERIFICATION BY ELECTRIC HEATER

573 Dew Point Mirror. Industrial Chilled Mirror Hygrometer. Typical applications:

DP 300 mobile dew point instrument

FSTC Equipment Test Report

An Experiment of Heat Exchanger Produces Hot Water Using Waste Heat Recovery from Air Conditioning

A660 Air Conditioning Laboratory Unit

HP 101A Series Humidity Temperature Probes INSTRUCTION MANUAL

Available online Journal of Scientific and Engineering Research, 2018, 5(11): Research Article

Specific Energy Consumption comparative study of Hot Air dryer and Heat Pump dryer for highland drying process Sayompon Srina

14 Drying. I Basic relations and definitions. Oldřich Holeček, Martin Kohout

Electric radiator tests

2. HEAT EXCHANGERS MESA

973 Dew Point Mirror. Portable Industrial Chilled Mirror Hygrometer. Typical applications: REFLECTING YOUR STANDARDS

Bench Series Ovens Models 21 / 31 / 51 Operating Manual

A study of Transient Performance of A Cascade Heat Pump System

MP 601A Humidity Probe INSTRUCTION MANUAL

Huminator II. Precision Humidity Generator. ww.testo.in

Fitoclima Pharma. Stability and Photostability Chambers for the pharmaceutical industry

ME 410 MECHA ICAL E GI EERI G SYSTEMS LABORATORY

Dew Point Mirror 373. High Precision. Fast Response. Wide Measuring Range. Laboratory Reference. Touch Screen. Full Color High Resolution LCD

Hygro Transducer, Hygrothermal Transducer, and CO 2 Measuring Probe for Building Management and Climate Monitoring

Infrared Moisture Balance (FD-660)

Model 5190-E Temperature/Humidity Probe

DFE T ECHNOLOGY. At Conceptronic, the design and management process for reflow soldering systems is called Dynamic Flow Engineering, or DFE.

CONVECTION OVENS FOR FOOD SERVICE APPLICATIONS BLODGETT SHO-E

A study on performance improvement of corrugated type total heat exchanger considering the structure of flow passage on surface

FORM ACCL-PC1 ACCL WITNESS TEST PROCEDURE AND CHECKLIST

Some Modeling Improvements for Unitary Air Conditioners and Heat Pumps at Off-Design Conditions

Humidity Measurement Technology

Experimental Study of Initial Process of Frost on Heat Exchanger Surface of Refrigerated Transport Vehicle

,Chaojian He6, f. School of Guangxi University, Nanning , China.

Experimental Study of Direct Contact Condensation of Steam on Water Droplets

PERFORMANCE ANALYSIS OF A RE-CIRCULATING HEAT PUMP DRYER

Environmental testing Part 3-6: Supporting documentation and guidance Confirmation of the performance of temperature/humidity chambers

Performance of R-22, R-407C and R-410A at Constant Cooling Capacity in a 10

Application of two hybrid control methods of expansion valves and vapor injected compression to heat pumps

APC APPLICATION NOTE #119

Internal flow Patterns of the Horizontal Heat Mode Closed-Loop Oscillating Heat Pipe with Check Valves (HHMCLOHP/CV)

S.A. Klein and G.F. Nellis Cambridge University Press, 2011

Transcription:

Available online www.jsaer.com, 2017, 4(10):305-311 Research Article ISSN: 2394-2630 CODEN(USA): JSERBR Low Cost Humidity/Temperature Calibration System Yasser A. Abdelaziz 1 Central Laboratories Department, Qatar Standards, Doha, Qatar 1 Researcher at the National Institute for Standards (NIS-Egypt) and deputation to the central laboratories, Qatar standards (CLD-QS) during this research work. Abstract A simple and low cost humidity/temperature calibration system is described in this article. This system basically is consists of a temperature controlled cabinet (refrigeration system), humidifier, temperature/humidity controller and reference hygrometer as a simple climatic chamber. It can also provide temperature calibration for built-in digital thermometer (without external temperature sensor) or thermo hygrometers. The working temperature range of the designed calibration system is started from 5 to 50 and from 50 to 98 for humidity. The designed calibration system was tested and calibrated for the temperature and humidity in its full range and an accuracy of 0.3 and 1.5 was achieved. The purpose of this paper is to describe the assembly of a low cost system to calibrate thermo-hygrometers. Keywords Soft set, operations, soft ring, soft ring ideal, idealistic soft ring 1. Introduction In the current years we noted, an increasing interest in the calibration of thermo-hygrometers over the world due to most laboratories, working zones and storage areas have a critical processes sensitive to ambient temperature and relative humidity. In this work an easy assembly for a humidity calibration system is described using some basic equipment as; temperature chamber, humidifier, temperature/humidity controller and reference hygrometer. The biggest advantage of this system is simplicity and low cost that makes this system attractive for calibration of thermohygrometers in the small calibration laboratories of developing countries. This simple design will allow anyone who has basic information about thermo-hygrometers and climatic chambers to try it with a very reasonable cost. 2. Developed Calibration System The individual components and techniques of the developed humidity calibration system will be discussed consists of the following items; 2.1. Temperature controlled cabinet (refrigeration system) A temperature controlled cabinet over the range + 5 ºC to + 50 ºC are using. This cabinet is supported with a digital temperature controller with resolution 0.1 ºC. Air (air circulation system) is forced over the heater and cooling fins and through the air duct in the upper surface to ensure a uniform temperature throughout the cabinet. The front doors are an internal double glass door and external metal door to ensure that the temperature and humidity are stable enough. When calibrating thermo-hygrometer good temperature control is essential. 305

Abdelaziz YA, 2017, 4(10):305-311 A temperature controlled refrigerators (mini refrigerator) can be used instead of the temperature controlled cabinet, but with an important condition is the presence of a built in temperature controller for these refrigerators. 2.2. Ultrasonic humidifier An ultrasonic humidifier is used to generate the humidity in this system. This ultrasonic humidifier uses a ceramic diaphragm vibrating at an ultrasonic frequency to create water droplets that silently exit the humidifier in the form of cool fog. The mist gets forced out by a tiny fan. The ultrasonic humidifiers use a piezoelectric transducer to create a high frequency mechanical oscillation in a film of water. Ultrasonic humidifier should be cleaned regularly to prevent bacterial contamination from being spread throughout the air. The amount of minerals and other materials can be greatly reduced by using distilled water. The ultrasonic humidifier is provided with an adjustable mist control star from 0 up to 100 % with 10 % subdivision. 2.3. Humidity controller A digital humidity controller with wide humidity measuring range and controlling range is used control the steam produced by the ultrasonic humidifier device with accuracy 1% for humidification and dehumidification modes. This controller comes with a thin film humidity sensor probe. Humidity measuring range is 1% ~ 100% RH and control range is 1% ~ 100% RH. 2.4. Reference Thermo-hygrometer A temperature/relative humidity recorder (Rotronic Hygrolog HL-NT data logger) was used as a reference thermo-hygrometer for monitoring and recording the temperature and humidity of the climatic chamber. This recorder is comprised of a display unit and one or two sensor units, each with both a temperature sensing element (thermistor) and relative humidity sensing device. The sensor units are connected directly through long cables. Consequently the instrument is able to monitor the temperature and relative humidity at two separated locations. The working range of this thermo-hygrometer with its external probe is -30 to 70 C for temperature and from 0 to 100 %rh for humidity with accuracy ±1.5 %rh for humidity and 0.2 for temperature. 2.5. Monitor Camera A low cost video camera, Web Cam is used in conjunction with a PC to follow and read your devices under test during the calibration process that can easily record the measurements manually, where the used chamber has an opaque metallic external door. 3. Design of the calibration system The design of the developed calibration system used to calibrate the hygrometer, temperature/relative humidity recorders and data logger will be shown in the following figures (Fig. 1 and Fig. 2). Figure 1: Illustration diagram of the developed humidity calibration system Figure 2: Actual photo for the developed humidity calibration system 306

Abdelaziz YA, 2017, 4(10):305-311 4. How does the chamber work? As it s known, relative humidity is the ratio of partial pressure of water in the air to the equilibrium vapor pressure of water at that temperature (measured in rh%). H R = (p w/p s) 100 Where; H R is the relative humidity () p w is the partial vapor pressure of water p s is the saturation vapor pressure of water (at the same temperature) To generate a constant flow of air at a specified temperature and absolute humidity, there are different methods, where the flow rate of water per unit time is then given by [1]: qw = (pw HR qa) / 1000 = (CV M qa) / (22.4 T 273 (P/Patm)) Where; q w is the water flow rate (g/min) H R is the fractional relative humidity q A is the flow rate of air (L/min) M is the molar mass (g/mol) C V is the concentration (volume per volume) P is the experimental pressure T is the temperature in Kelvin. Note; at RTP (25 C and 1atm) this reduces to: qw= CV M qa24.5 Controlling the humidity of the chambers with accuracy and repeatability is not possible without controlling its temperature. Set the temperature using the themperature controller and set the humidity using the humidity controller Monitor the chamber stability Monitor the reference thermohygrometer stablity Record the readings of the reference thermo-hygrometer and UUT Figure 2 The higher in the chamber temperature, it means that the air is capable of handling more moisture at a given temperature. Then, a stable temperature helps ensure a stable humidity level. So refrigeration system is suitable for doing this. The humidity within the chamber can easily decrease using the built in refrigeration system in the chamber attracts moisture in the air, causing it to condense into water, which can then be drained away, then the humidity will decrease until reaches the adjusted setting value of the humidity controller. Also to increase the chamber humidity we using a controlled ultrasonic humidifier to generate stable water vapor through a small orifice. The water vapor will spread out into the chamber until the humidity reaches the 307

Abdelaziz YA, 2017, 4(10):305-311 adjusted setting value of the humidity controller, then the humidity controller in turn immediately stops the humidity generator. If the chamber humidity decreased less than the setting value, the humidity controller will restart the ultrasonic generator again until reaching the setting value. 5. Results and discussion of the characterization of the chamber The main objectives of characterization and calibration of unloaded/empty chamber is to determine the uniformity, stability of temperature and humidity within the chamber, temperature/humidity indication error and measurement uncertainty by comparison technique [2]. All measurements were carried out within working zone the chamber according to DIN EN 60068 part 3-5 and DIN 12880 (parts 1 and 2) [3 and 4], through fixing the reference sensors in the corner points and center of working zone the chamber using nine thermo-hygrometer sensors as shown in Fig. 3. Figure 3: Set up diagram of reference thermo-hygrometer sensors (from No.1 up to No.9) locations red numbers within the working zone of the chamber All calibrations of the chamber were carried out in the range from 15 up to 30 for temperature and from 40 rh% up to 70 rh% for humidity. After reference sensors were placed in their positions according to Fig. 3 and ensure closing the chamber doors, we can set the temperature and humidity respectively, with 15 minutes time interval to avoid transient saturation conditions within the chamber. We can determine the following characteristics for the climatic chamber; Temperature/Humidity gradient Temperature/Humidity gradient expressed as the difference between the highest and the lowest temperature value of the temperature sensors on different locations within the working space after achieving stability [2]. The maximum value of temperature gradient was found to be 0.4 at 30 as shown in table (1). The maximum value of humidity gradient was found to be 1.1 at 70 as shown in table (2). Temperature/Humidity fluctuation It is the change of the temperature measured at the center of the working space of the chamber after achieving stability [5]. Its maximum temperature fluctuation value was found to be 0.3 at 30, as shown in table (1). The maximum humidity fluctuation value was found to be 0.8 at 70 values as shown in table (2). 308

Abdelaziz YA, 2017, 4(10):305-311 Temperature/Humidity stability It is the stability of temperature or relative humidity over a period of time after achieving steady-state conditions. Steady-state conditions are considered to be achieved when systematic variations of temperature or relative humidity are no longer measured [5]. About 30 min of measurements is recorded using a calibrated data logger. All stability measurements are carried out at the center of the working space of the chamber (Location No. 9, see Fig. 3). The temperature stability was found to be 0.2 at 20, as shown in table (3). The humidity stability was found to be 0.4 at 60 values as shown in table (3) and Fig. 4. Table 1: Temperature characterization (gradient & fluctuation) of the chamber Setting point, Temperature gradient, Temperature fluctuation, 15 0.3 0.2 20 0.3 0.3 30 0.4 0.3 Table 2: Humidity characterization (gradient & fluctuation) of the chamber Setting point, Humidity gradient, Humidity fluctuation, 40 0.9 0.7 60 0.7 0.6 70 1.1 0.8 Table 3: Temperature/Humidity stability of the chamber Setting point Temperature/Humidity @ temporal variation 60 %rh ± 0.9 20 ± 0.5 Figure 4: Temperature and Humidity stability over more than hour within the chamber 6. Temperature/Humidity calibration uncertainty analysis The uncertainty of measurement includes several components. The temperature and humidity calibration uncertainty budget for this new calibration system is shown in Table 4 and 5, in accordance with JCGM 100:2008, Guide to the Expression of Uncertainty in Measurement (GUM) [6]. Individual components are listed with their estimated standard uncertainties. 309

Chamber Reference Instrument Chamber Reference instrument Abdelaziz YA, 2017, 4(10):305-311 The estimation of the uncertainty includes the uncertainty of the used standards which contains; the calibration, the drift and the resolution. But the uncertainty arising from the performance of the chamber includes local homogeneity, instability, loading effect and resolution of the chamber indicator [5]. The measurement uncertainty stated below tables (4, 5) is expanded uncertainty which is obtained from the standard uncertainty by multiplication by the expansion factor k=2 and probability of 95 % confidence level. Table 4: Combination of temperature uncertainties Uncertainty source Value Distribution Divisor Standard uncertainty Calibration 0.2 Normal 2 0.1 Repeatability 0.15 Normal 1 0.15 Temperature effect 0.1 rectangular 1.73 0.06 Drift 0.1 Rectangular 1.73 0.06 Linearity 0.3 Rectangular 1.73 0.17 Resolution 0.1 Rectangular 1.73 0.06 Temperature gradient 0.4 Normal 1 0.4 Temperature 0.3 Normal 1 0.3 fluctuations Resolution 0.1 Rectangular 1.73 0.06 Instability 0.5 Rectangular 1.73 0.29 Local inhomogeneity 0.1 Rectangular 1.73 0.06 Combined Uncertainty = 0.64 Expanded Uncertainty at C. L. 95% (k=2) = 1.3 Table 5: Combination of humidity uncertainties Uncertainty source Value Distribution Divisor Standard uncertainty Instrument 0.5 Normal 2 0.25 Calibration Repeatability 0.9 Normal 1 0.9 Hygrometer drift 0.1 Rectangular 1.73 0.06 Linearity 0.1 Rectangular 1.73 0.06 Resolution of hygrometer 0.1 Rectangular 1.73 0.06 humidity gradient 1.1 Normal 1 1.1 humidity fluctuations 0.8 Normal 1 0.8 Resolution 1.0 Rectangular 1.73 0.58 instability 0.9 rectangular 1.73 0.52 Local 0.3 Normal 1 0.3 inhomogeneity Combined Uncertainty = 1.85 Expanded Uncertainty at C. L. 95% (k=2) = 3.7 Conclusion A simple low cost system has been built for humidity/temperature calibration with working temperature range from 5 to 50 and from 50 to 98 for humidity. The characterization results have shown that the gradient within chamber is 0.4 for temperature and 1.1 humidity in the working range. This system achieved a good accuracies and stability for humidity/temperature calibration. The results have shown also that the system calibration capability is 0.8 for temperature and 3.7 for humidity in the working range. 310

Abdelaziz YA, 2017, 4(10):305-311 Further on the basis of cost and simplicity of construction is a very important advantage for this system. Acknowledgments This work is carried during my deputation to the central laboratories, Qatar standards (CLD-QS). Saleh Rumaihi, Abbas Elkhayyat, Omer Elsheeb and Nasser Salama and all CLD-QS staff are gratefully acknowledged for their support and contribution. References [1]. The Complete guide humidity generation and humidity measurement, published by OWLSTONE NANOTECH, OW-002349-MC, 2004, UK, Page No. 11-14. [2]. Nabila I. El-sayed, Faten M. Megahed and Yasser A. Abdelaziz (2011). Determination of Uncertainty and Performance of NIS Humidity Standards, NCSL conference. [3]. Environmental testing - Supporting documentation and guidance; confirmation of the performance of temperature chambers, DIN EN 60068, Parts 3-5. [4]. Electrical laboratory devices - Heating ovens and incubators, DIN 12880, Parts 1 and 2. [5]. Guideline for calibration of climatic chamber, DKD-R- 5-7 Edition (07/2004). [6]. Guide to the expression of uncertainty in measurement (GUM) JCGM 100:2008. 311