Thermal Conductivity Detector on a 6820 GC, Accessory G4313A

Similar documents
TCD-to-FID Series Connection on a 6820 GC, Accessory G4319A

Preparing the 6850 Series II GC for 597x MSD Installation

Installing a Fast Oven Upgrade Kit

Installation. Leveling

Installation Guide. For Models: E-MLV36BNK4LK1 E-MLV36ESP4LK1 E-MLV36WW4LK1 READ THESE INSTRUCTIONS AND AND SAVE THEM FOR FUTURE USE

Installation Instructions

Installation Instructions

INSTALLATION. Glass Panel Doors (select models) CAUTION

Multi-Function Cooktop

Dacor Technical Service

Keysight 85025/85026 Detector Exchange Program - All Serial Numbers

Installation Guide. For Models: E-DCF52BNK5C3 E-DCF52FBZ5C3 E-DCF52MWW5C3 READ THESE INSTRUCTIONS AND SAVE THEM FOR FUTURE USE

Installation Guide. For Models: E-PD52ABZ5C4 E-PD52BP5C4 E-PD52WW5C4 READ THESE INSTRUCTIONS AND SAVE THEM FOR FUTURE USE

Installation Instructions

ST. KITTS CEILING FAN

52 BAYTOWNE CEILING FAN

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

22 The Flame Ionization Detector

TECHNICAL INSTRUCTIONS

Installation. Built-in Full Height Wine Cellar VCWB301

Torrena 42 Ceiling Fan

Camarillo 52 Ceiling Fan

Installation Guide. For Model: COS52OSB5CR READ THESE INSTRUCTIONS AND SAVE THEM FOR FUTURE USE. net weight of fan: lb (9.9 kg) PRINTED IN CHINA

ELSTON 52 CEILING FAN

52 BARSTOW CEILING FAN

C-IV 60 CEILING FAN READ AND SAVE THESE INSTRUCTIONS. FAN RATING AC 120V. 60Hz

COLD WEATHER START KIT Ceiling Mount 4000

DUVAL 52 CEILING FAN MODELS #50201, Español p. 19 LISTED FOR DAMP LOCATION

Installation Manual PS-225 & PS-275

Installation Instructions

Installation Instructions

Wide-Bore Heat Exchanger Installation Kit ( )

Installation and Operation Manual For Hunter Ceiling Fans

Installation Instructions

WARNING: ELECTRICAL GROUNDING INSTRUCTIONS

Step 1: Identify the parts of your shade (parts shown are not to scale)

Installation and Operation Manual For Hunter Ceiling Fans /16/2004

INSTALLATION INSTRUCTIONS

52 STRATHMERE CEILING FAN

GAS COOKTOP INSTALLATION INSTRUCTIONS

1320 Detectors. 1 of 30. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Installation Instructions

INSTALLATION INSTRUCTIONS

1. SAFETY RULES. 8. Avoid placing objects in the path of the blades.

Installation Manual PS-200 & PS-201

ANYWHERE. U.S. Patent(s) Pending INSTRUCTION MANUAL WARRANTY CERTIFICATE

48 TAYLOR CEILING FAN

MAYFIELD CEILING FAN LISTED E ITEM # MODEL #BTH44ABZC5C BTH44BNK5C Español p. 20 ATTACH YOUR RECEIPT HERE.

Installation Instructions

SUTTON 52 CEILING FAN

Table of Contents What to Expect with Your Installation. Ceiling Plate. Tools Needed.

TOUCHDOWN 48 CEILING FAN

BK AND BKT BLOWER ACCESSORY INSTALLATION INSTRUCTIONS

Installation Instructions

Installation Instructions

ITEM # SAIL STREAM CEILING FAN MODEL #40048

INSTALLATION INSTRUCTIONS

Platinum Split 4000 & Platinum Split 4000 Fully Ducted COLD WEATHER START KIT Installation Manual

Installation Instructions

F.P.O. Installation Guide. For Placement Only

Installation Instructions

Heater for BRP Commander. Caution: Before using this product, read this manual and follow all safety instructions. Owner s Manual

CEILING FAN OWNER S MANUAL

Menumaster. Commercial Microwave Oven Owner's Manual GCSAND7RR

INSTALLATION INSTRUCTIONS

SPRINGHILL 44 CEILING FAN

Cooktop Chimney Ventilation Hoods

52 CEILING FAN. Owner s Manual Models #50336, 50337

52 SAN LUCAS CEILING FAN

Installation Instructions

Owner s Guide and Installation Manual

BAY BRIDGE CEILING FAN

Viking Installation Guide

INSTALLATION GUIDE Dual Fuel Ranges

WAILEA OWNER S MANUAL

Installation Instructions

Dryer. User manual DV22K6800** DV22K A-00_EN (US)_ indd :15:41

Gas Cooktop Installation, User and Service Instructions GMS 955.1

1. SAFETY RULES. 1. To reduce the risk of electric shock, insure electricity has been turned off at the circuit breaker or fuse box before beginning.

Viking Installation Guide

ITEM # , CARATUK RIVER CEILING FAN MODEL #40046, 40303

Carbon Rack Enclosure

PRO 48 INSTALLATION INSTRUCTIONS

CEILING FAN OWNER S MANUAL

Menumaster. Commercial Microwave Oven Use and Care Manual CFSP70 GSAND-7D SAND-7D SAND-700

ITEM # SAIL STREAM CEILING FAN MODEL #41695

Undercounter Refrigeration

ALUMA INSTRUCTION MANUAL WARRANTY CERTIFICATE

Mechanical seals external, single or double, to DIN EN 12756

TILGHMAN CEILING FAN. LISTED For Damp Location E ITEM # MODEL #WCK52LMW5N WCK52NWZ5N. Español p. 20 ATTACH YOUR RECEIPT HERE

CEILING FAN OWNER S MANUAL

Advanced Test Equipment Rentals ATEC (2832) Installation Note

ARMITAGE CEILING FAN ITEM # MODEL #CC52WW5L. Español p. 17 ATTACH YOUR RECEIPT HERE. Purchase Date

Ceiling Fan Ow n e r s Manual

Installation Instructions. Instructions de montage. Instrucciones de instalación HC

CEILING FAN OWNER'S MANUAL

BUILT-IN MICROWAVE OVEN INSTALLATION INSTRUCTIONS 24" (61.0 cm), 27" (68.6 cm), 30" (76.2 cm)

Installation Instructions

Vari-Cyclone Ceiling Fans

Transcription:

Thermal Conductivity Detector on a 6820 GC, Accessory G4313A Installation Guide This installation guide provides procedures for installing a Thermal Conductivity Detector (TCD) on an Agilent 6820 Gas Chromatograph (GC). Before following these procedures, refer to safety information at the end of this document. Parts list Description Quantity T-20 Torx screw 4M 1 Screw M4 12 mm 3 Cable tie 100 3 mm 4 3-way switching valve cable 1 TCD sticker - English 1 TCD sticker - Chinese 1 Nutwarmer cup assembly 1 Insulation 1 Bottom insulation cup 1 Top insulation cup 1 Capillary inlet sticker for flow module - English 1 Capillary inlet sticker for flow module - Chinese 1 TCD body 1 TCD BD/D part 1 TCD ship kit 1 Agilent Technologies

Tools Configuration considerations Steps Electrostatic protection such as a grounded wrist strap T-20 and T-10 Torx screwdrivers Diagonal cutter Small flat-blade screwdriver 7/16-inch wrench To achieve a standard factory configuration, it may be necessary to move an already existing detector. For a two-detector configuration, factory placement rules are as follows: A TCD is always placed in the front location. A flame ionization detector (FID) is always placed in the back location. 1 Preparing the GC 2 Positioning and Securing the Detector 3 Connecting the Detector 4 Installing the Flow Control Module 5 Installing the Switching Valve Extension Cable 6 Restoring the GC to Operating Condition 2 Installation Guide: Thermal Conductivity Detector (TCD)

Preparing the GC WARNING Hazardous voltages are present in the mainframe when the GC is connected to electrical power. Avoid potentially dangerous shock hazards by disconnecting the power cord before removing the side panels. 1 Switch off electrical power to the GC and disconnect the power cord. Allow time for the oven and heated zones to cool. Then switch off supply gases at their sources. 2 Remove column(s) and any associated hardware from inside the column oven. 3 Lift the hinged GC top cover at its front edge to expose the detector area. Remove the cover by raising it to vertical, lifting its left hinge pin from its bracket, and then sliding the cover to the left to free its right hinge pin. 4 Remove two screws along the lower edge of the left side (flow control) panel. Slide the panel slightly to the rear and lift it off. Electronic components can be damaged by static electricity: use a properly grounded static control wrist strap when removing the electronics panel. 5 In the same manner, remove two screws along the lower edge of the right side (electronics) panel. Slide the panel slightly to the rear and lift it off. 6 Remove the back panel by removing two screws at its bottom edge, loosening two screws at its top edge, and the lifting off the two top screws. Installation Guide: Thermal Conductivity Detector (TCD) 3

Positioning and Securing the Detector 7 Finally, remove the rear top cover by removing four screws, two at each end. 8 At this time, if necessary to achieve a standard factory configuration, completely remove and reinstall an existing detector into its required location. If an existing detector is being completely replaced by this new detector, carefully store the old detector assembly and all associated parts in a safe place for possible future use. It is neither necessary nor advisable to separate a detector from its flow control module, as doing so may cause leaks. Although handling the detector and connected flow module as a unit is awkward, it can be managed. 1 If necessary, remove the round metal cutout on the oven top in the detector position to be used. Cut the metal circle with diagonal cutters so its nibs are connected to the piece removed. Discard the cutout. GC insulation is made of refractory ceramic fibers. Ventilate your work area. Wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator. Dispose of unneeded insulation in a sealed plastic bag. 4 Installation Guide: Thermal Conductivity Detector (TCD)

2 If necessary, lift out the die-cut insulation plug from the detector position to be used. Be careful to remove only insulation within the scribed circle. 3 Carefully remove the scribed circle of insulation from the oven top to create an opening into the oven. Method 1: Use a sharp knife to cut out insulation using the scribed circle as a guide. Method 2: Pierce the insulation with a screwdriver. Rotate the screwdriver around the edge of the scribed circle to remove excess insulation. Remove any pieces of insulation which fall inside the oven. 4 While positioning the detector into the cavity, use both supplied insulation and removed oven insulation to pack around the bottom and sides of the detector. The goal is to fill Installation Guide: Thermal Conductivity Detector (TCD) 5

all voids around the detector body as it is placed into the cavity. Pallet 5 Partially tighten the pallet screws with a T-20 Torx screwdriver, then tighten them evenly. Connecting the Detector Printed circuit board components can be damaged by static electricity; use a properly grounded static control wrist strap when handling the detector board. 1 Remove the PC board from its antistatic bag and slide it fully into the slot and main circuit board connector associated 6 Installation Guide: Thermal Conductivity Detector (TCD)

with the installed detector. The top of the board should be aligned with the slot on the TCD pallet. Insert into main board TCD board Front detector slot Back detector slot 2 Attach filament leads to the TCD board into the two left-most positions of the connector block. Insert a small flat-blade screwdriver into the rectangular slot and push downward Installation Guide: Thermal Conductivity Detector (TCD) 7

while inserting each filament lead fully into the correct round hole. Polarity does not matter. Delta PRT leads (thin, with pins) Filament leads (thick, black) Connect filament Connect Delta PRT Connector block TCD board If Delta PRT sensor leads are not properly connected to the TCD connector block, filament burnout may occur. 8 Installation Guide: Thermal Conductivity Detector (TCD)

3 In the same manner, connect the Delta PRT leads to the two right-most positions in the connector block. Polarity does not matter. Detector cell 4 Connect the heater/sensor plug to the square receptacle closest to the installed detector ("FD" for front detector, or "BD" for back detector). Heater/sensor plug Heater/sensor receptacle ("BD", back detector) Heater/sensor receptacle ("FD", front detector) Installation Guide: Thermal Conductivity Detector (TCD) 9

Installing the Flow Control Module This section describes installation of the detector flow control module. It is neither necessary nor advisable to separate a detector from its flow control module, as doing so may cause leaks. Although handling the detector and connected flow module as a unit is awkward, it can be managed. 1 Determine the correct location for the detector flow control module: Detector in the front position use the upper detector flow control module location. Detector in the back position use the lower detector flow control module location. Handle the module carefully to avoid damaging its components and/or connected tubing. 2 Starting from the detector, the tubing between the detector and its switching valve manifold assembly should remain located within the front oven top area. This makes it possible to remove the detector cell itself for servicing and/or replacement at a later time without having to remove instrument side and back panels. 3 Secure the switching valve manifold assembly to the support bracket on top of the oven using two T-10 screws and plastic washers. If necessary, temporarily remove the bracket from 10 Installation Guide: Thermal Conductivity Detector (TCD)

the oven top to secure the manifold block, then resecure the bracket. Switching valve manifold assembly T-10 screws and plastic washers Support bracket View from right (electronics) side 4 Route gas supply tubing from the switching valve manifold assembly through either slot at the back of the oven top, then alongside existing tubing and wiring located behind the oven top, and finally into the flow control (left) portion of the GC. Avoid making sharp bends in the tubing. 5 Route the flow control module to the appropriate location at the front panel in the flow control portion of the GC. 6 Remove the existing detector label plate from the module location to be used and install the new one provided. 7 Verify that the two valves on the flow module are in their respective midway positions (rotated fully clockwise to a stop, then rotated two to three turns counterclockwise), and then slide the module onto two studs provided at the rear of the flow panel. Installation Guide: Thermal Conductivity Detector (TCD) 11

8 Secure the module with a 1-3/4-inch screw. The screw passes through a hole at the center of the rear of the flow module. General representation Installing the Switching Valve Extension Cable 9 Use cable ties to secure installed tubing to existing wiring and tubing running behind the oven top. 10 Refer to the 6820 Getting Started manual both for factory standard plumbing configurations and for proper swage techniques to connect hydrogen and air supply gases to the installed detector flow module. 1 Route the switching valve leads and connector into the wiring and tubing area behind the oven top. 2 Connect the switching valve extension cable connector to its mating connector on the TCD board by pushing them fully together. Note that connectors are "keyed" to be joined in only 12 Installation Guide: Thermal Conductivity Detector (TCD)

one orientation. After making the connection, verify the TCD board remains firmly seated into its main board connector. Switching valve leads TCD board 3 Route the switching valve extension cable to the switching valve connector and join them fully together. Note that connectors are "keyed" to be joined in only one orientation. Cable tie TCD switching valve Switching valve and extension cable connected (behind oven top) View from right (electronics) side 4 Use cable ties as needed to secure installed switching valve wiring to existing wiring and/or to other support points. Installation Guide: Thermal Conductivity Detector (TCD) 13

Restoring the GC to Operating Condition 1 Restore gas supplies and, following leak test procedures described in your Maintenance and Troubleshooting manual, check for leaks, particularly at all new connections. 2 Reinstall GC covers and panels. 3 Restore GC electrical power. 4 Switch off electrical power, then switch it on again to ensure the new configuration is properly retained in GC memory. Do not switch on the detector at this time. Filament damage may occur since, at this point, proper gas flow rates have not yet been established to the detector and no column has been installed. 5 Press [ 前检测 ] or [ 后检测 ] ( [ Front Det ] or [ Back Det ], respectively). If the detector has been properly installed, you will see the following display: 前检测器 (TCD) 温度 45 关热丝关输出 ( 关 ) 0.0 FRONT DET (TCD) Temp 45 Off Filament Off Output (Off) 0.0 6 Verify detector operation by following the checkout procedure described in your Getting Started manual. 14 Installation Guide: Thermal Conductivity Detector (TCD)

Note that, when the TCD is operating, you should hear a "ticking" sound from the TCD switching valve. If you do NOT hear this sound when the detector is on, switch it off immediately to prevent filament damage and then troubleshoot and fix the problem (see your Maintenance and Troubleshooting manual for information). NOTE The new TCD is packaged with insulation cups that can be assembled and installed over the column fitting, if desired. Installation Guide: Thermal Conductivity Detector (TCD) 15

Agilent Technologies Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G4313-90000 Edition First edition, July 2003 Printed in China Agilent Technologies, Inc. 412 Ying Lun Road Waigaoqiao Free Trade Zone Shanghai, 200131 P. R. China Safety Notices A notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. G4313-90000