FOW FOA COUNTERPRESSURE STERILIZERS

Similar documents
FOF INDUSTRIAL SATURATED STEAM STERILIZERS

Load type and process/autoclave selection

FOB THIRD-GENERATION BENCH-TOP SATURATED-STEAM STERILIZERS

PASSION THAT DRIVES PERFORMANCES. WASHINGSOLUTIONS

cgmp STEAM STERILIZERS

LABOKLAV. Autoclaves Made in Germany

Faster UltraSafe Class II Microbiological Safety Cabinets. Protection, safety, reliability. And more.

SPECIALIZED EQUIPMENT

Mink MV. Claw Vacuum Pumps for Rough Vacuum Industrial Applications. Claw Vacuum Technology by Busch

OVENS LINE. Pocket 4060 rack oven for bread and pastry

ZETA MAGNETIC AGITATORS

STEAM STERILIZER SERIE S1000 (RBE)

A WORLD OF QUALITY, SAFETY AND HYGIENE STERILCLAVE. 18 litres-bhd/b/s

CENTRIFUGAL PUMPS. STATE the purposes of the following centrifugal pump components:

Faster BH-EN and BHG Class II Microbiological Safety Cabinets

International Reactor Bulletin 1100 Corporation. Standard and Custom Reactors for the Chemical Process Industries

Protection, safety, reliability. And more.

BARRIER TECHNOLOGY RABS & ISOLATORS FOR ASEPTIC PROCESSING

PAPER INDUSTRY PROCESS DESCRIPTION

PRODUCT INFORMATION. Liquid ring vacuum pumps Liquid ring compressors. Series LVP

SBM Schoeller-Bleckmann Medizintechnik Product Sterilization

Modern vacuum and overpressure generation for pneumatic conveying

SECTION STEAM CONDENSATE PUMPS

LN series STERLING PUMPS. Horizontal Split Case Pumps. Double Volute. Single Stage.

BLUE GALAXY DEPYROGENATION TUNNELS. Solid Dose Sol

Specialized in the design and the tailor-made production of dryers, coolers and thermal processing equipment for industrial use.

Goulds Model 3500 Medium Consistency Pump Systems. Pump System Advantage PP-1D. Standpipe Arrangement

Vacuum drying oven for non-flammable solvents

OCEANO BOTTLE WASHER

CLS-1450 Series Spinner Flask Instruction Manual

EQUIPMENT FOR OIL/PAPER DRYING AND STABILIZATION

MEDIACLAVE. Fast, reproducible and safe media sterilization

WINE, BREWERY & SPIRITS SUPPLY CATALOG

Safety vacuum drying oven for flammable solvents

BASIL 6000 TUNNEL CAGE WASHER

OPTIMIZING PERFORMANCE IN SPRAY OPERATIONS

Radio Frequency Drying Systems for the Textile Industry

Fedegari Aseptic Technology Integration: Case Study. Sergio Mauri, Director, Global Integrated Projects

MSMCR. Counter-Rotating Blenders. I Application. I Principle of operation

KRAL Screw Pumps. L Series.

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

MIX Type Blenders. I Application. I Principle of operation

ASEPTIC SAMPLING VALVES SANITARY SAMPLING VALVES SANITARY SAMPLING BOTTLE TANK BOTTOM ASEPTIC VALVE

Combi oven Metos SelfCookingCenter 61 Sous Vide

providing the perfect mix

LOW STATIC HYDRONIC DUCTED UNIT

Echelpoelweg KA Weerselo The Netherlands. Tel. (+31) Fax. (+31)

Getinge HS33. Article List. General Specification

TLA-1245 EFFICIENT HEAT CYCLE SHRINK TUNNEL OPERATION MANUAL

GESTRA Information A 1.7

Mixing Division. Industrial Mixers. MIX S.r.l. MIXING SYSTEMS AND COMPONENTS FOR PLANTS. ATEX RANGE AVAILABLE CM/172

Incubators. with BioCote antimicrobial protection

High-efficiency Turbo Chiller (NART Series)

Cleaning and sterilization solutions

The Logistics of Filling and Freeze Drying using Multiple Isolators. The use of isolator technology to fill and lyophilize pharmaceuticals

Manual. Atomizer Air Humidifiers. Type DG4 / DG8

ENVAIR lab CO2 Tri-Gas Incubator

ENVAIR lab CO2 Incubator

COLD STORE AIR CURTAIN MAT

HOW TO SELECT AN INDIRECT THERMAL TECHNOLOGY FOR INDUSTRIAL MATERIALS PROCESSING

OPERATION OF LAFER SOLVENT SCOURING PLANT

SIZE REDUCTION TECHNOLOGY

W+ Pumps. Minimize Life Cycle Costs

DS series STERLING PUMPS. Industrial End Suction Pumps ISO/DP

Compact. Cost effective. Productive. GSW-V Vacuum Gripper

OFFICINE MINUTE VILLORBA - TREVISO

Methods of solid phase extraction. Methods of solid phase extraction

Centrifugal Pumps acc. to EN / ISO 2858 with Shaft Sealing Type NCL

DIVISION 15 MECHANICAL

KIC Air Curtains. Technical Guide

RENZ FOR QUALITY PRODUCT CATALOGUE.

MST sterilizer overview

Drying and heating chambers

EcoPure TM HYGIENIC MAGDRIVE CENTRIFUGAL PUMP

we make air move Double Inlet Centrifugal Fans (Forward Curved) FBD Series (Square Outlet Metric Size)

DESCRIPTION OF COOLING TOWERS SERIES TAT

Waste-heat recovery: Weighing in the environmental Factor

TMCI Padovan Evaporators

DAE Systems. Fan Selection Guide. Section 1: Terminology

DRYING TECHNOLOGY A high efficiency dryer with a unique design. Presented by Mohamed LAMMI

Ergonomic handling solutions. Vacuum Handling Systems Branch Solutions Metal

ovens line elettrodrago for bread and pastry

CombiChem. Centrifugal pumps according to ISO 5199 and ISO 2858 / EN (DIN 24256)

ATLANTIS WASHING STATION

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Self-priming centrifugal pumps

Centrifugal Pumps acc. to DIN EN ISO 2858 with Shaft Sealing Type NCL

FOOD - PHARMACEUTICAL Industrial Vacuum Accessories

CombiBloc Close-coupled centrifugal pumps

Technical roller systems Rubber roller coverings Spreader rollers

MCF 500 CENTRIFUGAL FAN FOR AIR, FOR BIOGAS, LANDFILL AND NATURAL GAS

High-powered and efficient thanks to excellent technology.

APPLICATIONS > KORUND

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves

GIRTON CABINET WASHER MODEL 794A

Optimized Terminal Sterilization. Always with you

Air distribution for clean rooms. Clever and flexible.

GFD. GVD V-shape coil condenser. Güntner Info. 8 additional basic models for container transport!w. V-shape coil drycooler

Conical Mills (Under-Driven)

BASIL 6700 TUNNEL CAGE WASHER

Transcription:

WHY AND HOW TO STERILIZE IN COUNTERPRESSURE The technology of moist-heat sterilization in counterpressure has been developed specifically for the terminal treatment of solutions in sealed containers. Its primary aim is to "neutralize" the effects of the overpressure which unavoidably forms inside containers when they are sterilized using conventional saturated-steam autoclaves. This overpressure tends to make glass containers explode and to lift their stoppers, irreversibly deforms flexible or rigid plastic containers, expels the plungers of prefilled-syringes, etc. The collateral but not at all secondary aim is to perform load cooling with a sterile fluid, either water or air, thus complying with the CGMP of the FDA for LVP. Counterpressure autoclaves currently in use belong to two types: superheated water spray autoclaves steam + air mixture autoclaves To a certain extent, the two types are complementary and Fedegari has built both for several years. They are also presented together in this brochure, highlighting also the advantages of the two types. The figure below shows some typical products which can be sterilized with these autoclaves. This presentation is absolutely non-limitative, and new packages which require careful use of counterpressure sterilization are constantly being introduced in the market. Our Customers should contact us during the development of their new products/packages, since often modest adaptations of these products/packages allows optimization of the results of the sterilization processes. Our Research & Development Service is avaible to our Customers for this purpose. CONSTRUCTION FEATURES The chamber is cylindrical, usually single-wall, with thick insulation and an aluminum-sheet outer casing. The construction materials are class AISI 316 stainless steel. Either one or two rectangular doors, inscribed in the circular cross-section. These doors may be of two kinds: laterally sliding and automatic laterally hinged and semiautomatic; the doors engage and disengage their clamping devices with a diagonal movement automatically produced by a gearmotor, but the opening and closing action relative to the vertical hinges is manual.

TEMPERATURE/TIME PRINTOUT In both cases, the silicone gasket slides in a special slot formed by NC milling in the frame of the chamber and it is activated by compressed air. The gasket and the slot have specially designed dovetail cross-sections to ensure perfect operation and tightness without requiring lubrication and vacuum for gasket retraction. All valves and the entire piping are made of stainless steel. AUTOMATIC LOADING/UNLOADING SYSTEMS When it is necessary to sterilize large and heavy loads, it is sensible to resort to PRESSURE/TIME PRINTOUT palletization systems featuring automatic handling. The handling solution is extremely simple and efficient. It consists of a loading bench located in front of the autoclave, where the train of pallets is prepared. The bench has two rows of free rollers on which the pallets rest, and it is also provided with an underlying rod with a rack-like movement which pushes the pallets into the autoclave. The autoclave is also internally equipped with two rows of free rollers and therefore inside the chamber there are no sealed shafts, chains or gear: this means that all the risks of leaks and all lubrication problems have been eliminated. Usually, a second bench, arranged in front of the unloading door of the autoclave, removes the pallets by means of an extractor rack-rod. PROCESS CONTROL As all the other sterilizers, also the counterpressure autoclaves are equipped with the Thema3 process controller, specifically developed for monitoring and controlling sterilization processes. Also for these sterilizers, in order to ensure top process and safety performance, Fedegari offers its standard process controller (for its description, refer to the relevant brochure). EXPANDED STERILIZATIONTEMPERATURE/TIME PRINTOUT

Distribution system for the circulation water. Equipped with nozzles producing a "solid" spray cone, mounted so as to avoid water stagnation in the ducts of the sparger. FOW SUPERHEATED WATER SPRAY AUTOCLAVES FOW sterilizers are based on an extremely straightforward principle which is implemented with a highly reliable and mechanically very simple unit. When the program begins, after the load to be sterilized has been introduced, only the lower part of the chamber is filled with treated water of adequate microbiological quality. The chamber has a circular cross-section with a single wall, but has rectangular doors. The air inside the chamber is not removed. The water, which is drawn from the bottom of the chamber by a sanitarytype centrifugal pump, is circulated through a moving-plate heat exchanger, also of the sanitary type, which is indirectly heated in countercurrent with mains steam. The water is then fed back into the upper part of the chamber and is distributed onto the load by a system of solid-cone spray nozzles. Water redistribution onto the lower layers of the load is ensured by appropriate shelves. Heating of the circulation water and thus of the load is gradual but quite fast. The sterilization temperature of 121 C is reached within approximately 25-30 minutes inside 500 ml containers. After the sterilization time has elapsed, the circulation water, which by then is sterile as well, continues to circulate through the exchanger, which however in this step is fed with cold cooling water. Roller system for automatic load handling. The rollers are mounted on special roll-bearings with a cage made of plastic material accepted by F.D.A. All metal parts are made of stainless steel and require no lubrication. Within approximately 10 minutes, the temperature inside the containers gradually drops to approximately C, which is the ideal temperature to also obtain rapid and spontaneous drying of the load, when extracted from the autoclave. The entire process therefore lasts approximately 60 minutes. Throughout the process, an adequate counterpressure of sterile air is maintained inside the chamber in order to counterbalance, in the most appropriate and automatic manner, the overpressure within the product. Moving-plate heat exchanger made of stainless steel, sanitary type, double function (heating/cooling). Combines maximum efficiency and maximum safety. Plate cleaning is spontaneously ensured by the turbulence, speed and alternation of the circulation of fluids (steam and cooling water). Sanitary-type centrifugal pumps. They ensure an extremely high water circulation speed, which is essential to obtain optimum temperature uniformity.

FOA STEAM-AIR MIXTURE AUTOCLAVES FOA sterilizers have a chamber with a circular cross-section, generally with a single wall, and with rectangular doors. In these autoclaves, too, the air inside the chamber is not removed. The process begins by introducing steam directly in the chamber by means of the sparger located at the bottom of the chamber. One or more fans are located in the upper sector, based on the length of the chamber. Fedegari s widely tested magnetic coupling is used in this case as well. The action of the fans, combined with the action of the flow deflectors, generates a continuous and rapid homogenizing and circulation of the steam-air mixture, which forms in the chamber (see the functional diagram). At equal temperature and pressure, air is in fact on the average 1.5 times denser than steam and would tend to stratify on the bottom of the chamber, producing unacceptable temperature gradients. Once the heating and sterilization phases have ended, the cooling phase begins and develops as follows: the steam-air mixture in the chamber is replaced with sterile compressed air at controlled pressure; this air continues to circulate by virtue of the fans cold cooling water circulates in the batteries of cooling plates located on the sides of the chamber. The pictures at the bottom of the page refer to a FOA sterilizer with two peculiarities. It is a mixed machine, i.e suitable to run not only counterpressure processes with steamair mixture but also saturated steam processes. To this end it is also equipped with a vacuum pump and with a sparger in the upper part of the chamber for the load water spray cooling. The flow deflectors. Ensure efficient and uniform circulation of the steam-air mixture during heating/sterilization and of the air during cooling. The battery of stainless- steel hollow plates for circulating the cooling water. These are definitely preferable to tube-type exchangers, since they produce high turbulence in the circulating water and thus improving heat exchange. They are also easy to disassemble and inspect. The plates are hidden by the flow deflector. The internal impeller of the fan (the protection grid is removed). The vanes are made of a plastic material: they reduce the weight of the impeller and ensure extremely high efficiency and perfect static and dynamic balancing. Motor fans with magnetic coupling: this solution removes any tightness problem deriving from seal wear, and minimizes maintenance, without requiring any lubrication. However, the cooling rate is unavoidably much lower than in FOW autoclaves, since in this case it is based on two solid/gas heat exchanges, which notoriously have a low efficiency. Nonetheless, we try to improve their efficiency by keeping the air pressure in the chamber at a high value (within the limits set by the characteristics of the product) and by maintaining its circulation rate at high level. Throughout the phases of the program, air counterpressure in the chamber is adjusted in the same manner as in FOW sterilizers.

FUNCTIONAL DIAGRAMS FOW SUPERHEATED WATER SPRAY STERILIZERS FOA STEAM-AIR MIXTURE STERILIZERS

CRITICAL COMPARISON FOW FOA Temperature uniformity in time Temperature uniformity in space } easly in ± 0,5 C limits requested by FDA for LVP } easly in ± 0,5 C limits requested by FDA for LVP Total pressure uniformity in time Conterpressure management flexibility Excellent Excellent Consumption of high-microbiological-quality water Consumption of top water for cooling Consumption of compressed air Consumption of industrial steam Consumption of ultraclean steam Condensate recovery Yes, modest, for initial filling No Possible and easy No Yes, approx. 3 - times higher than FOW No Not possible. Cooling water recovery Autoclave price Total process duration Autoclave productivity/price Operating principle Mechanical construction Qualification/Validation Operating flexibility according to type of load Possibility of combination with satured steam processes Possible, recovered water is initially very hot. Acceptable Short High Very simple and straightforward Simple Normal Suitable for any kind of container with the following remarks:.upward concavities collect water.product is unloaded wet.pvc bags can produce blushing phenomena Possible, but involves significant and rather expensive solutions Possible, recovered water is initially very hot. Approx. 10% higher than FOW Approx. 30% higher than FOW Approx. 30% lower than FOW Plus FOW complex Plus FOW complex Normal Suitable for any kind of container with the following remarks:.upward concavities collect condensate only.other kind of container can be unloaded lighty damp. blushing phenomena of PVC bags is limited Feasible and moderately expensive

Testing of a cluster of three counterpressure sterilizers, with individual capacity of 20 m 3 each. MODEL USEFUL CHAMBER DIMENSIONS WIDTH (cm) HEIGHT (cm) DEPTH (cm) CAPACITY ( l ) CYLINDER(Ø cm) OPER. PRESSURE (abs bar) OPER. TEMPERATURE ( C) 50 830 118 FOW 3 FOA 3 FOW FOA 720 900 1.0 1. 1.0 1.0 2.160 210 1.350 118 250 1.600 118 310 1.9 118 165 2.010 210 2.560 250 3.050 310 3.7 2.700 370.510 MODEL USEFUL CHAMBER DIMENSIONS WIDTH (cm) HEIGHT (cm) DEPTH (cm) CAPACITY ( l ) CYLINDER(Ø cm) OPER. PRESSURE (abs bar) OPER. TEMPERATURE ( C) 2.160 220.300 2.700 260 5.000 FOW 5 FOA 5 3.20 3.600 325 6.300 350 6.900.500 30 8.300 5.00 530 10.000 6.300 600 11.700 7. 6 13.300 8.100 760 1.0 The above mentioned data are to be intended as indicative. The manufacturer reserves itself the right to change them without notice. These tables list the basic characteristics of standard models. Overall dimensions are not stated as they depend on the type door that is chosen; consumption figures are also not listed because they vary according to the type of load that is processed. The full identification of the autoclave Model is e.g. FOW..10. The last number (10) is the approximate number of 500-ml glass bottles (according to DIN standard) that forms a complete load for that autoclave. Albuzzano (PV) - Italia - tel. +39 0382 3111 - fax +39 0382 3 - www.fedegari.com