Isotemp Hotplates, Stirrers and Stirring Hotplates

Similar documents
StableTemp Hotplates, Stirrers and Stirring Hotplates

StableTemp Digital Hotplates, Stirrers and Stirring Hotplates

Cimarec Stirring Hot Plates

Instruction Manual. Mini Hotplate Mini Stirrer Mini Hotplate Stirrer

Instruction Manual Mini Hotplate Mini Stirrer Mini Hotplate-Stirrer Mini Auto Stirrer Mini Fixed Temperature Hotplate

HOT PLATE OPERATION MANUAL AND PARTS LIST Series 64

Instruction Manual. For All Corning Hot Plates, Stirrers, Stirrer/Hot Plates with Digital Displays, and External Temperature Controller (6795PR)

Isotemp Explosion-Proof Refrigerators and Freezers

Instruction Manual. Model AccuPlate Hotplate AccuPlate Stirrers AccuPlate Hotplate Stirrer 12/10 CLS-EQ-LHP01REV

Isotemp Plus General Purpose Laboratory Freezers

Reacti-Therm I and III Heating/Stirring Modules

OPERATIONS MANUAL. Hotplate Stirrer/Hotplate/Magnetic Stirrer Model HTS-1003/HTP-1002/MGS Certified

Reacti-Therm I and III Heating Modules

SUPER-NUOVA Digital Plates

INSTRUCTION MANUAL. stirring hotplate. VMS-C7 advanced series. Brand :

Instruction Manual Analog Vortex Mixer with Fixed Cup Head

Isotemp Low Temperature Incubator

60 Liter Infinity Storage Reservoir

Type SP87300 SAFE-T SHP9 Explosion Proof Stirring Hot Plate

SAVE THESE INSTRUCTIONS

Type SP87300 SAFE-T SHP9 Explosion Proof Stirring Hot Plate

Installation and Operation Instructions Luxury Towel Warmer Free Standing or Wall Mounted

Professional Series. Instruction Manual Hotplates, Stirrers, Hotplate-Stirrers

Package Contents. Dry Block Heater Power Cord Instruction manual Warranty Card. Serial Number: Date of Purchase: Supplier: Table of Contents

CFC-FREE CB-60/CB-80 Cryobath Low Temperature Bath. Thermo NESLAB Manual P/N U00551 Rev. 04/03/00. Instruction and Operation Manual

Owner & Operator s Manual. Table of Contents

USER MANUAL SAVE THESE INSTRUCTIONS

Operator s Manual. The Bullet Blender 50 BB50-AU, BB50-DX. Contents. Congratulations!

Electron Microscopy Sciences

YOUR PORTABLE AIR PURIFIER

INSTRUCTION MANUAL. Stirring hotplate (with probe and stand) Brand : (Dragon laboratory)

Portable Induction Ranges

Aluminum Heating Mantles

Operator s Manual. The Quietek

FAN-FORCED UTILITY HEATER

6 Sheet Strip-Cut Shredder for Home/Office

TTT Digital Incubator

Magnetic Stirring Systems with Heat

Store Name: Location: Date Purchased: Serial Number (located on back of unit):

555 Hotplate 565 Magnetic Stirrer 575 Hotplate-Stirrer

75 Sheet Micro-Cut Shredder With CD/DVD Slot

PLEASE DO NOT RETURN THIS PRODUCT TO THE STORE!

NESLAB CB 60/CB 80 Cryobaths. Operation. Product Service Information, Applications Notes, MSDS Forms, .

20 High velocity Air

YOUR AROMATHERAPY DIFFUSER

Wrap-Around TOTE Tank / IBC Heaters (TOTE and TOT Series)

FOOD WARMING/MERCHANDISING CABINET MODEL

MaxLite LED MICRO-T PANEL

Hi-Temp Bath Installation and Operation Manual

1500W STOVE HEATER INSTRUCTION MANUAL. Item No: 050-HA-50090

Installation and Operation Manual. 445 Procedures Cart. Page 3. Page 5. Page 8. Page 10. Important Information Page 2. Installation.

16 Stand Fan with Remote Control Model No: FS3-40R-30

HR1-08R-21 Reflective Heater Operating Instructions. Model No. HR1-08R Soleus Air International

Instruction Manual M6627 / Version 7.4. Slide Drying Bench MH6616 MH6616X1

Owner s Manual. PIR-1 IR Learner

BASEBOARD MICATHERMIC HEATER

Indoor/Outdoor Electric Patio Heater Model No: HP Operating Instructions HP Soleus Air International

MS-11 Flat Panel Reflective Heater Operating Instructions. Model No. MS Soleus Air International

User Manual. MS7-H550-Pro MS7-H550-S MS7-Pro MS7-S

PRACTICAL USER S GUIDE FOR THE ELECTRIC STOVE

Digital Heat Block User Manual

Series: MBC1-TC Mini Benchtop Temperature Controller

ELECTRIC STOVE HEATER Models: FS2213B

IAQ Series. Bosch IAQ Photo Catalytic Oxidizer (PCO) Residential Application. Installation Manual and Owner s Guide

Fisher Scientific Model 281A Vacuum Oven. Installation and Operating Manual Rev. 0

User s Manual & Operating Instructions

Operator s Manual. IP-100 Immersion Probe Cooler

ROTATING WAFFLE MAKER

Analog Melting Point Apparatus

Please consider the option to recycle the packaging material and donate any appliance you are no longer using.

HC Ceramic Tower Heater Operating Instructions HC Soleus Air International

RK-05 Carbon Monoxide Detector Operator s Manual

Stand Fan. English. Owner s Guide. Model F R

- Oscillating Table Model Fan

INSTALLATION AND OPERATING INSTRUCTIONS FOR THE VEHICLE-MOUNTED RADIATION DETECTION SYSTEM

RECESSED LIGHT CONVERSION KIT PENDANT FITTER

Dry Bath Heating Systems

Instruction Leaflet. Heater with Adjustable Thermostat PLEASE READ AND SAVE THESE IMPORTANT INSTRUCTIONS

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

FOOD WARMING/MERCHANDISING CABINET

Corning LSE Benchtop Shaking Incubator

Superdryer Dehumidifiers

Digital 7 Square Plate Magnetic Stirrer

Remote Relay Module (RRM)

Ion Genesis II Pump Controller Digital Level Control with Pump Alternation and High Water Alarm

FRESHAIR USER S MANUAL. CAUTION: Read manual carefully for proper procedures and operation.

MS-H-Pro-T Magnetic Hotplate-Stirrer with Timer

Conserve Socket Power Timer. User Guide

Users Manual. LAURUS Systems, Inc. - Ph: Fax:

Silicone Rubber Heating Tape with Adjustable Thermostat Control (HSTAT Series) Instruction Manual

Please consider the option to recycle the packaging material and donate any appliance you are no longer using.

INSTRUCTIONS AND WARRANTY

Instruction Leaflet. Read instructions before operating. Retain for future reference.

HGW-308R Wall Mountable Micathermic Heater w/ Remote. Model No. HGW-308R Soleus Air International

BIONAIRE HEATER. Instruction Leaflet. pure indoor living MODEL: BCH4130. Read instructions before operating. Retain for future reference.

OWNER S MANUAL ELECTRIC FIREPLACE INSERT. Model No.: WESTERN (EF42D-45D) WARNING CAUTION

10 Table/Wall Mount Fan Model No: FTY-25 Operating Instructions FTY Soleus Air International

IMPORTANT SAFETY INFORMATION:

C L E A N I N G / M A I N T E N A N C E WA R R A N T Y I N F O R M AT I O N CLEANING/MAINTENANCE FAN STORAGE 3 YEAR LIMITED WARRANTY INSTRUCTIONS

FOOD WARMING/MERCHANDISING CABINET

Transcription:

Isotemp Hotplates, Stirrers and Stirring Hotplates OPERTION MNUL ND PRTS LIST LT892X 2/28/09 Catalog # Type Top Plate Size and Material Voltage 0000H Hotplate 0x0 Ceramic 20 0000S Stirrer 0x0 Ceramic 20 0000SH Stirring Hotplate 0x0 Ceramic 20 006H Hotplate 4x4 Ceramic 20 006S Stirrer 4x4 Ceramic 20 006SH Stirring Hotplate 4x4 Ceramic 20 0049H Hotplate 7x7 Ceramic 20 0049S Stirrer 7x7 Ceramic 20 0049SH Stirring Hotplate 7x7 Ceramic 20 20049H Hotplate 7x7 luminum 20 20049S Stirrer 7x7 luminum 20 20049SH Stirring Hotplate 7x7 luminum 20 06H Hotplate 4x4 Ceramic 00 049H Hotplate 7x7 Ceramic 00 049S Stirrer 7x7 Ceramic 00 049SH Stirring Hotplate 7x7 Ceramic 00 2049S Stirrer 7x7 luminum 00 2049SH Stirring Hotplate 7x7 luminum 00 000H Hotplate 0x0 Ceramic 00 000S Stirrer 0x0 Ceramic 00 000SH Stirring Hotplate 0x0 Ceramic 00 026H Hotplate 4x4 Ceramic 230 026S Stirrer 4x4 Ceramic 230 026SH Stirring Hotplate 4x4 Ceramic 230 0249H Hotplate 7x7 Ceramic 230 0249S Stirrer 7x7 Ceramic 230 0249SH Stirring Hotplate 7x7 Ceramic 230 20249H Hotplate 7x7 luminum 230 20249S Stirrer 7x7 luminum 230 20249SH Stirring Hotplate 7x7 luminum 230 0200H Hotplate 0x0 Ceramic 230 0200S Stirrer 0x0 Ceramic 230 0200SH Stirring Hotplate 0x0 Ceramic 230

Table of Contents Safety Information...3 Warnings...3 Specifications...5 Heating Specifications...6 Stirring Speed Specifications...6 Environmental Conditions...7 Declaration of Conformity...7 Unpacking and Installation...8 General Usage...8 Unpacking...8 Installation...8 Operation...9 Setting the Stirring Speed...9 Setting the Temperature...9 Heating Metal Vessels and Sand Baths...9 Maintenance and Troubleshooting...0 General Cleaning Instructions...0 Error Codes...0 Replacement Parts... ccessories... Warranty...2 2

Safety Information lert Signals Warning Warnings alert you to a possibility of personal injury. Note If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired. Your Fisher Isotemp Hotplate, Stirrers or Stirring Hotplate has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes. For safe operation, please pay attention to the alert signals throughout the manual. Warning: These products should be used only under the operating conditions specified in the Operating Manual. lways use safe laboratory practices and do not leave the hotplate in operation while unattended as product funcionality or laboratory practice failures could occur that might lead to uncontrolled or excessive heating of the top surface. Safety procedures (including, but not limited to, unplugging when not in use) and response plans should be put in place to address the worst case possibility. If an over-temperature failure occurs, the top surface temperature could rise to the maximum temperature (300-540 C depending on your model s specification) and remain at that temperature indefinitely. Under these conditions, the material being heated on the surface of the hotplate could reach levels in excess of the maximum temperature. This manual contains important operating and safety information. The user must carefully read and understand the contents of this manual prior to the use of this equipment. Warnings To avoid electrical shock, always:. Use a properly grounded electrical outlet of correct voltage and current handling capacity. 2. Disconnect from the power supply prior to maintenance and servicing. To avoid personal injury:. Do not use in the presence of flammable or combustible materials fire or explosion may result. This device contains components which may ignite such materials. Not rated for use in hazardous atmospheres. 2. Use caution when heating volatile materials; 3

SFETY INFORMTION top surface and element can reach the Flash Point Temperature of many chemicals. These stirring hotplates are not explosion proof. Fire or explosion may result. Unit contains components which may ignite such materials. 3. Keep top surface clean. Use a non-abrasive cleaner. lkali spills, hydrofluoric acid spills or phosphoric acid spills may damage top and lead to thermal failure. Unplug unit and remove spills promptly. Do not immerse unit for cleaning. 4. Replace the top immediately if damaged by etching, scratching or chipping. damaged top can break in use. 5. Do not use metal foil on hotplate which may block air flow. Overheating will result. 6. Check and tighten the removable cord periodically making sure it is secure. If loosened, the cord could become hot and/or spark and be a potential fire hazard. If cord appears damaged, replace immediately. 7. Do not remove or modify grounded power plug. Use only properly grounded outlets to avoid shock hazard. 8. Use appropriate hand and eye protection when handling hazardous chemicals. 9. Gross weight of items placed on top of stirring hotplates should not exceed 35 lbs. (5.9 kg.) on the 0 x 0 models, 25 lbs. (.3 kg) on the 7 x 7 models, and 5 lbs. (6.8 kg.) on the 4 x 4 models. 0. The top plate of the unit can remain hot for some time after use. CUTION - HOT TOP light will remain on until top plate temperature cools to below 50 C.. Note that the exterior housing will be hot during and for a period of time after use. 2. Refer servicing to qualified personnel. 4

Specifications Catalog # Electrical (Volts/mps/Watts/Freq./Phase) Temperature Range* Speed Range () Top Plate Capacity 0000H 20/.7/40/60/ 5-400 C N / 6L flask / 35 lbs. 0000S 20/0.37/20/60/ 60-200 6L flask / 35 lbs. 0000SH 20/2/440/60/ 5-400 C 60-200 6L flask / 35 lbs. 006H 20/3.2/385/60/ 5-540 C N / L flask / 5 lbs. 006S 20/0.37/20/60/ 60-200 L flask / 5 lbs. 006SH 20/3.3/395/60/ 5-540 C 60-200 L flask / 5 lbs. 0049H 20/8.8/060/60/ 5-540 C N / 0049S 20/.037/20/60/ 60-200 0049SH 20/8.9/070/60/ 5-540 C 60-200 20049H 20/5.5/660/60/ 5-300 C N / 20049S 20/0.5/35/60/ 60-200 20049SH 20/5.7/680/60/ 5-300 C 60-200 06H 00/3.8/385/50-60/ 5-540 C N / L flask / 5 lbs. 049H 00/0.7/070/50-60/ 5-540 C N / 049S 00/0.47/25/50-60/ 60-200 049SH 00/.0/095/50-60/ 5-540 C 60-200 2049S 00/0.6/35/50-60/ 60-200 2049SH 00/6.8/680/50-60/ 5-300 C 60-200 000H 00/.2/25/50-60/ 5-400 C N / 6L flask / 35 lbs. 000S 00/0.47/25/50-60/ 60-200 6L flask / 35 lbs. 000SH 00/.5/50/50-60/ 5-400 C 60-200 6L flask / 35 lbs. 026H 230/.7/420/50-60/ 5-540 C N / L flask / 5 lbs. 026S 230/0.7/20/50-60/ 60-200 L flask / 5 lbs. 026SH 230/.8/440/50-60/ 5-540 C 60-200 L flask / 5 lbs. 0249H 230/4.8/50/50-60/ 5-540 C N / 0249S 230/0.7/20/50-60/ 60-200 0249SH 230/4.9/70/50-60/ 5-540 C 60-200 20249H 230/2.9/690/50-60/ 5-300 C N / 20249S 230/0.3/35/50-60/ 60-200 20249SH 230/3.0/720/50-60/ 5-300 C 60-200 0200H 230/6.4/530/50-60/ 5-400 C N / 6L flask / 35 lbs. 0200S 230/0.7/20/50-60/ 60-200 6L flask / 35 lbs. 0200SH 230/6.5/550/50-60/ 5-400 C 60-200 6L flask / 35 lbs. * This stirring hotplate does not cool. The minimum temperature is 5 C if used in a cold room. 5

Specifications Heating Specifications Top Plate Surface - Solid Ceramic Temperature range ll 4" x 4" and 7" x 7" models: ll 0" x 0" models: Heat-up time to maximum temperature (unloaded top plate). Temperature stability at the center of the top plate surface (@ 00 C). 4 F to 004 F (5 C* - 540 C) 4 F to 752 F (5 C - 400 C) 8 minutes ± 2.0 C ccuracy of the temperature display vs the actual average temperature of a 2 diameter of settingarea at the center of the top plate (setpoint 00 C). (Top plate temperature was verified with an infrared camera mounted approx. 26" above the center of the top plate surface; temperature accuracy at the edges of the top plate may vary from the stated accuracy.) ± 5.0 C Stirring Speed Specifications Speed Range 60 to 200 RPM (Maximum speed is dependent on the viscosity of the solution) Speed Dial Position Estimated Stirring Speed 2 60 4 25 6 350 8 700 0 00 2 200 Stability of the stirring speed setpoint (600 ml of water in a 000 ml glass flask) ± 5.0% Top Plate Size Max Recommended Flask Size Max Weight on Top Plate 4" x 4" liter 5 lbs 7" x 7" 4 liters 25 lbs 0" x 0" 6 liters 35 lbs * This stirring hot plate does not cool. The minimum temperature is 5 C if used in a cold room. 6

SPECIFICTIONS Environmental Conditions Operating: 7 C to 27 C; 20% to 80% relative humidity, non-condensing. Installation category II (overvoltage) in accordance with IEC 664. Pollution degree 2 in accordance with IEC 664. ltitude Limit: 2,000 meters. Storage: -25 C to 65 C 0% to 85% relative humidity Declaration of Conformity (for 230 volt) We hereby declare under our sole responsibility that this product conforms with the technical requirements of the following standards: EMC: EN 6000-3-2 Limits for harmonic current emissions EN 6000-3-3 Limits for voltage fluctuations and flicker EN 6326- Electrical equipment for measurement, control, and laboratory use; Part l: General Requirements Safety: EN 600- Safety requirements for electrical equipment for measurement, control, and laboratory use; Part I: General Requirements EN 600-2-00 Part II: Particular requirements for laboratory equipment for the heating of materials (Hotplates and Stirring Hotplates) EN 600-2-05 Part II: Particular requirements for laboratory equipment for mixing and stirring (Stirrers and Stirring Hotplates) per the provisions of the Electromagnetic Compatibility Directive 2004/08/EC, and per the provisions of the Low Voltage Directive 2006/95/EC. The authorized representative located within the European Community is: Thermo Fisher Scientific Electrothermal House Unit 2 Purdeys Industrial Estate Purdeys Way Rochford, Essex, SS4 ND United Kingdom Tel: +44(0)702 303350 Copies of the Declaration of Conformity are available upon request. 7

Unpacking and Installation General Usage Do not use this product for anything other than its intended usage. Warning Use a properly grounded electrical outlet of correct voltage and current handling capacity. Unpacking Remove your Isotemp hotplate, stirrer or stirring hotplate from the carton. Inspect to ensure that the unit has not been damaged during shipment. If the unit appears to have sustained shipping damage contact Fisher Scientific at 866-984-3766. The following items are included in the shipment: Isotemp Hotplate, Stirrer or Stirring Hotplate Cord Stir Bar (Stirring models only) dapter (00V models only) Knob Operator s Manual Installation Set the unit on a flat stable surface at least 2" away from combustible materials, and plug the cordset into a properly grounded electrical outlet of correct voltage and current handling capacity. 8

Operation Warning Use caution when heating volatile materials; top surface and element can reach the Flash Point Temperature of many chemicals. These hotplates and stirring hotplates are not explosion proof. Fire or explosion may result. Unit contains components which may ignite such materials. Use appropriate hand and eye protection when handling hazardous chemicals. Caution: Hot Top. void Contact. The top plate of the unit can remain hot for some time after use. CUTION - HOT TOP light will remain on until top plate temperature cools to below 50 C. Setting the Stirring Speed To set the speed, turn the STIR knob to the desired setting. Turn the knob clockwise to increase the speed or counter-clockwise to decrease the speed. To discontinue stirring, turn knob counter-clockwise to OFF. Setting the Temperature To set the temperature, turn the HET knob to the desired setting. Turn the knob clockwise to increase the temperature or counter-clockwise to decrease the temperature. To discontinue heating, turn knob counter-clockwise to OFF. The temperature will display in C. When choosing a setpoint, the display will indicate the setpoint for 5 seconds, after which the display will flash until the desired setpoint is reached. The CUTION - HOT TOP indicator illuminates when the plate reaches 50 C. When the heating is turned off, the indicator will continue to flash until the plate temperature drops below 50 C. Heating Metal Vessels and Sand Baths Metal vessels and sand baths may be heated safely without the danger of the ceramic top breaking. Use the lowest temperature setting possible for applications to limit thermal stress to the ceramic top. Using a metal vessel or sand bath may reduce the life of the stirring hotplate. 9

Maintenance and Troubleshooting General Cleaning Instructions Wipe exterior surfaces with lightly dampened cloth containing mild soap solution. Error Codes - Hotplates and Stirring Hotplates only The following errors should not be addressed by the user. If any of the errors appear, contact Fisher Scientific Technical Service at 866-984-3766. Displayed Message Intended to Detect Cause Solution E0 Thermocouple Thermocouple not connected. Replace element Call Tech. Service out of range. assembly (attached to thermocouple). Thermocouple open circuit. E02 Excessive heat-up time. Thermocouple short circuit. Remove short. Call Tech. Service Failure of element. Replace element assembly (attached to thermocouple). Failure of element circuit. Replace Element. Replace Control Board. E lternating current Failure to detect alternating Replace Control Board. Call Tech. Service measurement failure. current. E2 Motor system failure. Locked rotor condition. Free locked rotor. Call Tech. Service Failure of motor. Replace motor. Failure of motor circuit. Failure of motor sensor. Replace Control Board. Replace motor sensor. Notes: Errors E0 and E02 are heating errors. Heating functions will be locked out. Stirring functionality is unaffected. Errors E and E2 are stirring errors. Stirring functions will be locked out. Heating functionality is unaffected. 0

Replacement Parts 4x4 Ceramic 7x7 Ceramic 7x7 luminum 0x0 Ceramic Part No. Part No. Part No. Part No. Description 26575 26575 26575 26575 PC Board Fuse - 220-240V PCX8 PCX8 PC895X PC886X Control Board - 00V, 20V (SHP) PCX9 PCX9 PC895X2 PC886x2 Control Board - 220-240V (SHP) PC887X PC887X PC893X PC884X Control Board - 00V, 20V (HP) PC887X2 PC887X2 PC893X2 PC884X2 Control Board - 220-240V (HP) PC888X PC888X PC888X PC885X Control Board - 00V, 20V (S) PC888X2 PC888X2 PC888X2 PC885X2 Control Board - 220-240V (S) 266058 266058 266058 266058 PC Board Fuse - 00V, 20V PCX3 PCX3 PCX3 PCX3 Display Board PC363X3 PC363X3 PC363X3 PC363X3 Speed/Heat djustment Board (2) KBX KBX KBX KBX Knob (2) CRX06 CRX06 CRX06 CRX06 Cord Set - 00V, 20V CRX04 CRX04 CRX04 CRX04 Cord Set - 220-240V KBX78 KBX78 KBX78 KBX78 Knob ZSX22 ZSX65 ZSX65 ZSX65 Stir Bar MT33X3 MT33X3 MT33X3 MT36X3 Motor - 20V, 00V MT33X4 MT33X4 MT33X4 MT36X4 Motor - 220-240V ELX4 ELX44 ELX77 ELX47 Heating Element - 20V - w/thermocouple ELX42 ELX45 ELX78 ELX48 Heating Element - 00V - w/thermocouple ELX43 ELX46 ELX79 ELX49 Heating Element - 220-240V - w/thermocouple 540-0033 70-07 CT420X 79-0073 Ceramic or luminum Top EL30X EL33X EL895X EL36X Hot Plate Top ssembly - 20V EL30X2 EL33X2 EL895X2 EL36X2 Hot Plate Top ssembly - 00V EL30X3 EL33X3 EL895X3 EL36X3 Hot Plate Top ssembly - 220-240V ---- ----- TC895X ----- Thermocouple ccessories Part No. Description 0-666-0GQ Support Rod - 2 4-666-0BQ Support Rod - 8 4-666-0CQ Support Rod - 24 Please refer to the Specification Plate for the complete model number, serial number, and series number when requesting service, replacement parts or in any correspondence concerning this unit. ll parts listed herein may be ordered from Fisher Scientific by calling 866-984-3766

Warranty Laboratory instruments and equipment manufactured by Fisher Scientific Company L.L.C. Laboratory Equipment Division (hereinafter called the Company ) are warranted only as stated below. Subject to the exceptions and upon the conditions specified below, the Company agrees, at its election, to correct by repair, by replacement, or by credit to the purchaser, any defect of materials or workmanship which develops within one year (3 months for refrigerator and freezer products) from the date of purchase by the original purchaser by the Company or by an authorized dealer of the Company provided that investigation or factory inspection by the Company discloses that such defect developed under normal and proper use The exceptions and conditions mentioned above are the following: a. The Company makes no warranty concerning components or accessories not manufactured by it, such as tubes, batteries, etc. However, in the event of the failure of any component or accessory not manufactured by the Company, the Company will give reasonable assistance to the purchaser in obtaining from the respective manufacturer whatever adjustment is reasonable in the light of the manufacturer s own warranty. b. The Company shall be released from all obligations under its warranty in the event repairs or modifications are made by persons other than its own service personnel or authorized dealer personnel unless such repairs by others are made with the written consent of the Company. c. THE COMPNY MKES NO WRRNTY OF MERCHNTBILITY, FITNESS FOR NY PRTICULR PURPOSE, OR NY OTHER WRRNTY, EXPRESS OR IMPLIED, EITHER IN FCT OF BY OPERTION OF LW, STTUTORY OR OTHERWISE. d. The above warranty and the above obligations to repair, replace, or credit are complete and exclusive and the Company expressly disclaims liability for lost profits or for special, indirect, incidental, consequential, or exemplary damages of any nature whether attributable to contract, warranty, negligence, strict liability, or otherwise even if the Company has been advised of the possibility of such damages. e. Representations and warranties made by any person, including dealers and representatives of the Company, which are inconsistent or in conflict with the foregoing warranty shall not be binding upon the Company unless reduced to writing and signed by an officer of the Company. For the name of the authorized Thermo Scientific product dealer nearest you or any additional information, contact us: 308 Ridgefield Court, shville, NC, 28806 US Phone: -866-984-3766 Fax: -828-665-407 Web: www.thermo.com 2