TM-802 Formaldehyde Meter User s Manual

Similar documents
Intelligent combined sensor for Air Quality (VOC) + Humidity (rh) + Temperature (T) with analogue outputs, Modbus and LCD display. Features.

USER GUIDE. Formaldehyde (HCHO or CH 2 0) and Total Volatile Organic Compounds (TVOC) Tester. Model VFM200

Precision Temperature Humidity Meter DTH184DL User s Manual

NGD8800 USER S MANUAL

APPLICATION DATA SHEET

Wall Hung Boiler Room Thermostat OR30 OT Thermostat

Gas detector tubes are most effective as a reference device for conducting spot checks, which gives an idea of present gas at that moment in time.

6-1 Installing Batteries 6-2 Automatic Circuit/Reset Feature 6-3 Feature Sensitivity Adjustment

SP-220. Features. Combustible handheld gas leakage detector

ESD3000 ESD200 Combustible / Toxic Gas Detector

Model R9400. Instruction Manual. CO Meter with Temperature. wwwreedinstruments. com

Quality. Water. Water Quality. Testing. Testing. Toxic Gas Detection. Toxic Gas. Detection. Air Quality. Monitoring. Monitoring. Monitoring.

User Manual R1.0 CO2Meter, Inc. Personal CO2 Monitor SAN-10

MiniCO, MiniH 2 S, MiniOX, MiniOX Remote, MiniCl 2. , and MiniClO 2. Responder Detectors Instruction Manual "! WARNING

Digital Gas Transmitters with Auxiliary Inputs & Control. PolyGard 2 DC6

Wöhler CM 220. Carbon Monoxide Meter. The Measure of Technology. Contents. 1. Important information Specification... 13

MULTI-GAS MONITOR. Eastern Energy Co., Ltd. 40/4 Vitoondumri Rd., Banbueng Banbueng, Chonburi, Thailand

Model:AS8900 MULTI-GAS MONITOR O 2 %VOL H 2S PPM LEL%VOL CO PPM. Version number:as

ESD500. ESD500 Combustible / Toxic Gas Detector

LD910. User Manual. Combustible Gas Leak Detector. Detects all Hydrocarbon and other Combustible Gases. Made in U.S.A. MSHA Approved Intrinsic Safety

Model VFM200. User Manual. Formaldehyde (HCHO or CH 2 0) and Total Volatile Organic Compounds (TVOC) Tester

CD100A COMBUSTIBLE GAS DETECTOR

EAGLE 2. 6 Channel Capacity. Inlet fitting. Loud buzzer (95dB) IrDA communication port Low flow shutoff and alarm Alarm LED s with wide visibility

LS1. Operator Manual. Kane International Ltd

GS100M Gas Detection Controller

Freightliner Refrigerator Troubleshooting Guide For (TJ18F) (TJ22F) (TJ18FP3)

Refrigerant Leak Detector

CO2 & Oxygen Monitor Operating Instructions Model: RAD

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

Electron Microscopy Sciences

Manual Vane Anemometer PCE-VA 20

Material Safety Data Sheet

MD-MINI Dry Bath. Instruction Manual

KANE-ALERT-CO 2 Monitor with Temperature

compact detector Detection of more than 200 flammable and toxic gases iso 9001:2008

Users Manual. Air Velocity Meter

Using the MiniRAE 2000 & ppbrae PID

P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification

E1000 Portable gas detector

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

User's Guide. CO 2 Monitor. Model CO200

MATERIAL SAFETY DATA SHEET

CLETEF-786. Infrared Thermometer Instruction Manual

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

Micro IV Operation Manual

Installation and Wiring 1.

4P75 CiTipeL Combustible Gas Sensor Part Number: PM

GasAlertMicroClip X3 Specifications

C0-210 Carbon Monoxide Probe

Setting the alarms in electrochemical sensor equipped toxic gas instruments

Air-Quality Detection Module. User s Manual

R9900. Model. Instruction Manual. Indoor Air Quality Meter. reedinstruments. www. com

Operate the Mutli-Gas Monitor PID

Model GX-2001 Canadian Version Operator s Manual Part Number: RK Revision: 0 Released: 1/17/05

G450. Multi-gas Detector. Field Operation Manual Oak Valley Dr, Ste 20, Ann Arbor MI USA (800) (734)

Air-Quality Detection Module. User s Manual

Dual-Mode Infrared Thermometer Instruction Manual

Indoor Air Quality Meter Rev. 2.x Meters. The Value Leader TM

COMPUTHERM. CO1400 Carbon Monoxide Alarm OPERATING INSTRUCTIONS

BST-MG08/09 Multi-gas Detecting Alarm. Manual Instruction

SENSIT HXG-2. Instruction Manual For use with combustible gases Gas Leak Detector

III. 99 Washington Street Melrose, MA Phone Toll Free Visit us at OPERATING INSTRUCTIONS

USER MANUAL Digital CO Monitor Model CO30

Inc Series Digital Gas Detector/Controller. Operation Manual

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

User Manual. Non-contact Infrared Thermometer.

Portable Combustible Gas Detector GP-1000 Operating Manual (PT0-120)

SBS-H2 Hydrogen Gas Detector Kit

IMR EX610. Portable Gas Detector. Operation Manual

PolyGard Single Point Controller SPC-X3-34XX for Combustible Gases

GX-2009 Portable Gas Monitor Operator s Manual

MATERIAL SAFETY DATA SHEET

GX-2009 Portable Gas Monitor Operator s Manual

ALTAIR 5X Multigas Detector Product Specification

Portable Gas Leak Detector SP-220 Series TYPE FUM TYPE SC Operating Manual

Safety Data Sheet. P.O. Box St. Louis MO Phone: Emergency telephone number: CHEMTEL GHS02 GHS04 GHS07 GHS08

Low mg/m

PT0E Portable Gas Leak Detector. SP-220 Series. Operating Manual (PT0E-139)

4P-90 CiTipeL Combustible Gas Sensor Part Number: PM

Safety Data Sheet. acc. to OSHA HCS. Printing date 09/25/2014 Revised On 09/25/2014

CMF, CMX & CMA series

Selecting Gas Detectors for Confined Space Entries

Analog Room Pressure Monitor RPC Series

PCD110X - Premium Choice with DC Motor (Ceiling Mounted Ventilation Fans)

GasAlert. Introduction. WWarning. Table 1. GasAlert Models. Gas Monitored

Red Alarm Indicator Light

Room thermostat with Auto Timer, independent DHW

Instruction Manual. Self-Leveling Rotary Laser Level Model No &

Combustible Gas Detection and Control System

EST User Manual. (Please read this manual carefully before using)

IMR IX176 Portable Gas Detector User Manual

Model GD Instruction Manual. Combustible Gas Leak Detector. reedinstruments. www. com

Thermo Scientific TVA2020 Introduction

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

99 Washington Street Melrose, MA Phone Toll Free Visit us at

GX-2009 MSHA Certified Version CH 4 /O 2 /CO/H 2 S Portable Gas Monitor Operator s Manual


MICROpeL 75 Combustible Gas Sensor Part Number: PM

Micro IV. Single Gas Detector. Operations Manual Oak Valley Dr, Ste 20, Ann Arbor MI USA (800) (734)

UV Sentry. Cost Effective Perimeter & LDAR Monitoring. ...Precise, Species Specific Fugitive Emissions Quantification

Transcription:

Formaldehyde Meter User s Manual HB2TM802000

Content 1 INTRODUCTION 2 2 ACCESSORY 2 3 SAFETY PRECAUTION 2 4 METER DESCRIPTION 3 5 OPERATION 3 5.1 Switch Unit 4 5.2 Auto power off 4 5.3 LED back light 5 5.4 Record values of the maximum, the minimum and the average 5 5.5 Switch the SETL or TWA Expression of Formaldehyde 5 5.6 Read value holding function 6 5.7 Alarm setting 6 5.8 Manual recording 7 5.9 Record reading 7 5.10 Clear record 8 5.11 Calibration 8 6 FORMALDEHYDE HAZARDS 12 7 GENERAL SPECIFICATIONS: 15 8 ELECTRICAL SPECIFICATIONS: 17 8.1 Formaldehyde 17 8.2 Humidity 17 8.3 Temperature 17 9 SAFETY AND MAINTENANCE STANDARDS 17 10 BATTERY REPLACEMENT 18 11 END OF LIFE 18 1

2 TM-802 1 INTRODUCTION Measure the formaldehyde density, temperature, humidity in the environment simultaneously Monitor the occupational safety in TWA (8 hours) and STEL (15 min) Use the high-precision electrochemical sensor for formaldehyde measurement 2 ACCESSORY 1 Meter 1 User manual 6 UM-4(AAA)1.5V battery 1 Carrying case 3 SAFETY PRECAUTION Note! Please refer to this manual. Improper use may damage the meter and its components. Complies with European Directive. Do not operate in environments with flammable gas or humid environments. Operating altitude: up to 2000M. Operating environment: Indoor use; Pollution degree 2. Clean with soft cloth when dirty, such as glasses cloth. Do not clean with chemicals and other solvents. EMC: EN61326-1:CISPR 11:Group 1, Class B Class B Equipment for use in all establishments other than domestic. Group 1 RF energy generated is needed for internal functioning.

4 METER DESCRIPTION Fig. 1 Description of meter: 1. Formaldehyde sensor 7. Maximum hold and data hold button 2. Temperature and humidity sensor 8.Left button/rec record button 3. LCD display 9. Settings function button 4. Power switch button 10. DC voltage input jack 5. Data hold button 11. Tripod 6. Back light button 12. Battery cover 5 OPERATION 1. Press button for more than 2 seconds to turn power on. 2. After the warm-up time about 5 minute, the measurement screen will display. 3

5.1 Switch Unit Press or mg/m 3 ; to select the formaldehyde unit of PPM Press for 2 seconds to select or F. 5.2 Auto power off When any operation stopped for about 15 minutes, it will shut down automatically. Turn off the auto power off function: at shutdown mode, press button and then press button to boot, LCD will display n-sl, indicating that the automatic shutdown function has been turned off. Restart to restore the automatic shutdown function. 4

5.3 LED back light Press button to turn back light on or off. When turned on, the back light will be turned off automatically after 15 seconds. 5.4 Record values of the maximum, the minimum and the average Press button to start the function, press button once again to select the modes of maximum (MAX), minimum (MIN) or average (AVG) displayed on the LCD sequentially. Press off. button for 2 seconds to turn this function 5.5 Switch the SETL or TWA Expression of Formaldehyde Press or normal. button to select the modes of SETL, TWA 5

TWA: the average total value in 8 hours (Summarize and average the values every 3 minutes, and update every 3 minutes) STEL: the total average value in 15 minutes (Summarize and average the values every 15 minutes, and update every 15 minutes) 5.6 Read value holding function Press function. button to start or stop the data holding 5.7 Alarm setting Press mode. button for 2 seconds to enter the alarm 6

Press button to select the alarm turn on (A-on) or turn off (A-oF). After selected, press normal mode. button to return to the After start the alarm mode as (A-on), press button to select the alarm value, including 0.1/0.3/0.5/0.75/1.00/1.50/2.00/3.00/5.00/0000, press button after the alarm value determined, finish the setting of alarm and turn on it. Under the alarm mode, the automatic shutdown function will be canceled. 5.8 Manual recording In the normal state, press and button simultaneously to record a data, the screen will display the icon and the number of records; 5.9 Record reading In the normal state, press and button simultaneously to enter the reading mode. The screen will display the icon and the number of records, and 7

then display the record data. Press button to select the number of records; if to exit the reading mode, press button to exit, or wait without pressing any button for 15 seconds to exit automatically. 5.10 Clear record In the manual reading state, press button for 2 seconds to clear the records. 5.11 Calibration As calibrating, please keep away from the organic gases such as alcohol to avoid to interfere the concentration of formaldehyde. At shutdown status, press and button simultaneously, then press calibration mode. button to enter the As entering the calibration mode, you may decide whether to reset as the factory defaults by pressing or button. 8

Then, press button, verify and access the calibration screen of each point; or press exit. Press button, you may select the items sequentially: formaldehyde 0PPM calibration, formaldehyde 0.5PPM calibration, 33% RH calibration, 75% RH calibration, temperature calibration; or press button to exit. to Formaldehyde 0PPM calibration: Press button to select the 0PPM calibration screen. 9

Place the meter in a clean environment, Press button to start the formaldehyde 0PPM calibration, and finish the calibration after 8 minutes. Or press button to exit 0PPM calibration state without calibration. Formaldehyde 0.5PPM calibration: Press button to select the 0.5PPM calibration screen. Press button, and input formaldehyde of 0.5PPM with flow rate 0.8 l/min to the meter. The detection will stop after 8 minutes. 10

Press button or to adjust the value of main window to 0.5ppm. After adjusted, wait for 30 seconds and complete the calibration. Humidity 33%RH calibration Press screen. button to select the 33%RH calibration Place the meter in the environment of 33%RH. After the humidity stable, press button, wait for 8 seconds and complete the calibration. Humidity 75%RH calibration Press screen. button to select the 75%RH calibration Place the meter in the environment of 75%RH. After the humidity stable, press button, wait for 8 seconds and complete the calibration. 11

Temperature calibration Press button to select the temperature calibration screen. Place the meter and the standard part in the same environment for 30 minutes. After the temperature stable, press button. Press button or button to adjust the value of main window to be same as the standard part. After adjusted, wait for 30 seconds without pressing any button to complete the calibration. 6 FORMALDEHYDE HAZARDS Formaldehyde Human concentration response 0.03 ppm General outdoor environment value 0.08 ppm Tasteless, no sensation for the human 0.1 ppm Standard recommended value 12 International standard Limiting value of ASHRAE, ANSI, EPA,

0.15-0.25 ppm Breathing for a long time may easily cause the allergies on children s skin. 0.25-0.3 ppm It may cause the symptoms of asthma, allergies, cough, etc 0.5 ppm It may cause abnormal immune function and oncogenic risk. 0.75ppm It may cause Chromosomal abnormalities and affecting fertility, and cause cancer easily. >2.0ppm It may cause cancer and chronic respiratory 13 NIOSH for STEL residence, the recommended value from Taiwan Environmental Protection Administration. Limiting value of OSHA for workplace Limiting value of OSHA and TWA for workplace Limiting value of OSHA and TWA for workplace

disease. 14

7 GENERAL SPECIFICATIONS: TM-802 Easy to monitor: LCD can display the three values of formaldehyde, temperature and humidity at the same time. Low battery indication The manual recording can be up to 200 readings Back light display. The maximum / minimum / average hold. Formaldehyde maximum and minimum alarm value setting. Monitoring and displaying the occupational environmental safety for TWA (8 hours), STEL (15 minutes) Battery life is about 75 hours under continuous use. Power: LR03(AAA)1.5V *6 batteries. AC to DC Adaptor:(Optional) External AC 100~240V to DC 9V/0.5A power supply. About the DC plug, the inner side of the extension lead shall be connected to the positive electrode, and the outer is the negative electrode. The inside diameter (I.D.) of the extension lead is 5.5mm, and the outside diameter (O.D.) is 2.1mm. Operating temperature and humidity: -10 ~ 40 (14 to 104 F); less than 80% Storage temperature and humidity: -20 ~ 50 (-4 to 122 F); less than 70% RH Dimension: 191x73x35mm (LxWxH) Weight: 250g(without the batteries weight) 15

The substances may interfere the sensors: Substance Relative sensitivity formaldehyde 100 ammonia 0.0 water vapor 0.0* carbon dioxide 0.0 acetone 0.0 ethyl methyl ketone 0.0 benzene 0.0 terephthalate 0.0 toluene 0.0 methane 0.0 ethyl acetate 0.0 hydrogen 0.1 chlorine -3 carbon monoxide 1 phenol 8 sulfur dioxide 12 ethanol, methanol 50 Note: within the specified range. The step change of %RH will cause a short-term transient response. 16

8 ELECTRICAL SPECIFICATIONS: Accuracy of the environmental conditions: 23 ± 5 and RH <80% 8.1 Formaldehyde Measurement Range Resolution Accuracy Response time (T90) 8.2 Humidity 0.00-5.00PPM(mg/m 3 ) 0.01ppm ±5% of reading + 0.03ppm(baseline shift) 30 seconds Range 5.0% to 95.0%RH Resolution 0.1%RH Accuracy ±5.0%RH (20 to 80%) ±8.0%RH (<20%,>80%) 8.3 Temperature Range -20 to 50 (-4 to 122 o F) Resolution 0.1 (0.1 F) Accuracy ±1.0 (±1.8 F) 9 SAFETY AND MAINTENANCE STANDARDS 1. When the LCD shows the symbol, it means that the electric quantity is low, please replace with batteries immediately to ensure accuracy. 2. When the meter is dirty, please wipe it with a soft cloth, such as glasses cloth, and do not wipe it with 17

chemical solvents. 3. If not using for a long time, please remove the batteries to prevent the leakage of battery electrolyte solution which will corrode the internal components. 4. If fault on the meter, can only be sent to the service suppliers who are authorized, or sent back to the original factory for maintenance. 5. When testing high concentrations of formaldehyde (about 1PPM above), the period shall not be more than ten minutes; after testing,it shall be placed in the environment of clean air for more than half an hour before and then shut down. If not, it will accelerate the aging of the sensor. 10 BATTERY REPLACEMENT 1. Turn off the power. 2. Open the battery cover on the back of meter and remove the batteries. 3. Install new batteries and comply with the polarity positions of positive and negative. 4. Install the battery cover back. 11 END OF LIFE Caution: This symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal. 18

Professional Electrical and Environment Test & Measurement Instruments: Battery Capacity / Impedance Tester, TACHO Meter,LED light meter, Temperature & Humidity meter,infrared Thermometer, Sound level meter,light meter, EMF meter, UV Light meter, RF meter, Hot wire Anemometer, CO meter Anemometer, Lan cable tester, CO 2 meter, Solar power meter, Radiation meter, Clamp meter, Multimeter, Phase Rotation test, Digital Insulation tester Our products of high quality are TENMARS ELECTRONICS CO., LTD 6F, 586, RUI GUANG ROAD, NEIHU, TAIPEI 114, TAIWAN. E-mail: service@tenmars.com http://www.tenmars.com