Application Note. Application Note for BAYEX

Similar documents
Passive Whole Body Monitoring Nuclear Power Industry Application Introduction to the Argos -5AB Zeus (Gamma Option) and the GEM

CONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:

Instrumentation. Post RDD Event Urban Reoccupancy. James P. Menge PE. The world leader in serving science

Utilization of advanced clutter suppression algorithms for improved standoff detection and identification of radionuclide threats

Portable Survey Instruments NISP-RP-01

Understanding total measurement uncertainty in power meters and detectors

Fundamentals is subdivided into Sources of Radiation, Biological Effects, Mathematics, Chemistry, Physics, and Units and Terminology.

A bright alpha numeric display and the simplified keypad (three buttons) allow the operation even under harsh conditions.

NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use

RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System

Thermo Scientific Aerosol Monitor FHT 58 S Technical Specification ZT-221E

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

LAUNDRY MONITORING SYSTEM Model # LIM-64S

Use of MCNPX for Alpha Spectrometry Simulations of a Continuous Air Monitor

WinAMS AMS-4 Monitoring System for Windows

Portable Alpha/Beta Continuous Online Air Monitor (CAM)

OPERATION OF RADIATION PORTAL MONITORS OF SPECIAL NUCLEAR MATERIALS AT NUCLEAR SITES IN TERMS OF RELIABILITY

Quality Assurance/ Quality Control. Topic 4 - Audio 46. Key Points

COMPANY PROFILE. DETECTION & MEASUREMENT OF IONIZING RADIATIONS RADIOLOGICAL MONITORING SYSTEMS HEALTH AND ENVIRONMENTAL SAFETY

Contamination Meter Users Manual

API MANUAL OF PETROLEUM MEASUREMENT STANDARDS

INTERNATIONAL STANDARD

Identify the following features and specifications for the ESP1

Ramon 2.2 Radon-Monitor

Failure Modes, Effects and Diagnostic Analysis

AFMG. Design for Speech Intelligibility Using Software Modeling. Stefan Feistel, Bruce C. Olson AHNERT FEISTEL MEDIA GROUP

Nuclear Associates

AM-806. Waste/Door Radiation Monitor OPERATIONS MANUAL

Detector Configurations for the MAJORANA Demonstrator

Aer 5200/5300. Data Sheet Aer 5200/5300. Portable Alpha/Beta Continuous Online Air Monitor (CAM) Closer to your application

Using the GammaRAE II Personal Radiation Detector

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Alpha Attenuation Due to Dust Loading. Amber E. Dailey Francis Marion University Savannah River Site

How to Use Fire Risk Assessment Tools to Evaluate Performance Based Designs

ILNAS-EN 61649:2008. Weibull analysis. Analyse de Weibull. Weibull-Analyse

Introduction...1 Specifications... 1 Limitations... 1

Assessment of factors causing false alarm signals from detection equipment, and ways to mitigate this challenge

Ranger EXP RADIATION A L E R T

LUDLUM MODEL 26 FRISKER USER S MANUAL. December 2016 Serial Number PF and Succeeding Serial Numbers

Battery-operated, hand-held alarm and measuring instruments for radiation protection

TSI AEROTRAK PORTABLE PARTICLE COUNTER MODEL 9110

Uncertainty Analysis For Input Parameters Of Electrical Cabinet Fire Simulation By Coupling Latin Hypercube Sampling And CFAST

ISO 3999 INTERNATIONAL STANDARD. Radiation protection Apparatus for industrial gamma radiography Specifications for performance, design and tests

ISO 7183 INTERNATIONAL STANDARD. Compressed-air dryers Specifications and testing. Sécheurs à air comprimé Spécifications et essais

RC4000 Radiation Detection System

Battery Performance Alert: A TOOL FOR IMPROVED PATIENT MANAGEMENT FOR DEVICES UNDER BATTERY ADVISORY

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

CZT/NaI Hybrid Hand-Held Performance Evaluation

DEVELOPMENT OF LARGE AREA PLASTIC SCINTILLATION DETECTOR FOR RADIOACTIVE CONTAMINATION MONITOR

Troubleshooting Thermo Scientific LevelPRO Gauges

HEALTH AND SAFETY MANUAL

ATOMLAB XENON TRAP MONITOR OPERATION AND INSTALLATION MANUAL FN: /12

Advanced Radiation Measurement Solutions

Environmental Radiation Monitoring System

icam -HD High Flow Rate Alpha/Beta Air Monitor Radiation Safety. Amplified. DESCRIPTION KEY FEATURES BENEFITS OPTIONS

Benefits of Enhanced Event Analysis in. Mark Miller

MOVING FILTER AIR MONITOR ALPHA, BETA/GAMMA PARTICULATES

Setting the new standard for personal radiation detection.

Standard ECMA th Edition - December Declared Noise Emission Values of Information Technology and Telecommunications Equipment

Consultation Draft. Issue 2. Good Practice Guide No. 29. The Examination, Testing and Calibration of Installed Radiation Protection Instruments

NEX-BETA-ABG FACILITY Drinking Water Safety Monitor

Study and development of an innovative 3G/4G wireless network analysis tool

Measuring Couch Motion Accuracy with X-ray Beam Fluence

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006

Y. ORMIERES. Fire risk analysis method for nuclear installations

Statistical Testing of Electric Meters

International comparison of validation methods for dust concentration measurement

Computer Modelling and Simulation of a Smart Water Heater

ISO AIR FILTERS FOR GENERAL VENTILATION: DETERMINING FRACTIONAL EFFICIENCY

LUDLUM MODEL 334AB ALPHA/BETA PARTICULATE MONITOR. December 2016

Battery Performance Alert

Comprehensive applications of the gas flow proportional counters for Radiological Surveillance

Reliability and Safety Assessment in Offshore and Process Industries

NEXT GENERATION DRINKING WATER RADIATION SAFETY MONITOR

MetroRADON project and its potential impact on mitigation practices Dobromir Pressyanov Sofia University St. Kliment Ohridski Bulgaria

Radiation Safety and Equipment Considerations. Bonnie Meilner Regional Sales Manager LAURUS Systems Inc.

Impacts of an Energy Star Promotion

User s Manual. YTA110, YTA310, YTA320, and YTA710 Temperature Transmitters. Manual Change No

AN ANALYSIS OF THE PERFORMANCE OF RESIDENTIAL SMOKE DETECTION TECHNOLOGIES UTILIZING THE CONCEPT OF RELATIVE TIME

Real-time neural network application to mine fire nuisance emissions discrimination

False Alarm Suppression in Early Prediction of Cardiac Arrhythmia

False Alarm Analysis of the CATM-CFAR in Presence of Clutter Edge

Radiation Monitoring Equipment

LUDLUM MODEL 26-1 FRISKER USER S MANUAL. April 2017 Serial Number PF and Succeeding Serial Numbers

FlameGard 5 UV/IR HART

Relevant Radon properties

Installer Manual KNX Touchscreen Thermostat

ISO/TR TECHNICAL REPORT. Fire-safety engineering Technical information on methods for evaluating behaviour and movement of people

MOVING FILTER AIR MONITOR ALPHA, BETA/GAMMA

PURPOSE AND SPECIFICATIONS

RAM GENE / RAM GENE-1

THE NEXT GENERATION IN VISIBILITY SENSORS OUTPERFORM BOTH TRADITIONAL TRANSMISSOMETERS AND FORWARD SCATTER SENSORS

Radioactivity dosimeter RD01. User's Manual

WM 00 Conference, February 27 March 2, 2000, Tucson, AZ RELEASE SURVEYING OF SCRAP METALS WITH THE IONSENS CONVEYOR A CASE STUDY

Environmental radioactivity surveillance system of ANPA

A Continuous Air Monitoring Sampler for 125 I - RIS-125

AIR or STACK MONITOR NOBLE GAS, GROSS BETA-GAMMA PARTICULATE, & IODINE. Optional Alpha, Tritium, Carbon-14

Chapter 17, Initiating Devices

Transcription:

Application Note Application Note for BAYEX Preface This application note provides the user a more detailed description of the Bayesian statistical methodology available in Version 8.05 and above, of the Monitor software. This document is intended to be a useful reference for the customer in setting statistical parameters and decision points for making a clean or contaminated decision during personnel monitoring. The user has a choice in selecting the Traditional or Bayesian methodology. The purpose of the application note is to highlight the similarities and differences between the two methodologies, with particular attention to their advantages and limitations in radiation measurement. Introduction Scope Effective radiation monitoring systems are essential in radiation protection and security. All radioactivity measurements are subject to uncertainties. The background due to ambient radiation and other factors need to be considered when determining the radioactivity level of a sample/personnel monitoring result. Statistical false alarms may occur as a result of fluctuations in the number of recorded events. In order to interpret the measured data correctly, and draw valid conclusions, the uncertainty must be indicated and dealt with properly as is described in the JCGM 100:2008 Guide to the Expression of Uncertainty in Measurement (GUM). Radioactive decay follows a Poisson distribution, which can be approximated by a normal distribution (Gaussian) when the number of counts over a period of time is greater than 20. This approximation breaks down when the number of counts is below this number and Gaussian statistics is not valid in calculating statistical errors. So in comes the Bayesian statistics to the rescue for the determination of statistical uncertainty at low count rates /alarm levels. The Bayesian methodology is particularly beneficial for low alarm levels and count rates for alpha contamination. Bayesian methodology vs. Traditional methodology Similarities and differences between the two methods are summarized in Table 1, below. Table 1 Similarities and Differences Traditional and Bayesian Methodology Traditional Methodology Probability is measurable frequency of events determined from repeated experiments. Parameters are fixed but unknown quantities. Is objective and relies only on statistical data. Well-developed methodology with no need of prior data. Poisson distribution is approximated by Gaussian. Only uncertainties due to counting statistics are taken in account. Permit all values. Lower Limit of Detection (LLD). Clean /contaminated decision is made using net count rate vs. ATP (Alarm Trip Point) derived from user entered Alarm Activity. When the ATP is lower than LLD then the LLD (Lower Limit of Detection) is used instead. Bayesian Methodology Probability is a direct measure of uncertainty and might or might not represent a long-term frequency. Parameters are random variables with distributions attached to them. Provides probabilistic interpretations (1-g). Uses prior knowledge. Poisson distribution is approximated by Gaussian in defining detection limit and decision threshold; however calculation of best estimate and confidence intervals are done through Bayesian approximation. Combine all uncertainty contributions, Type A and Type B uncertainties as per the JCGM 100:2008 guide. Prior distribution of activity, best estimate of activity and confidence interval are always positives. Characteristic limits are defined (ISO 11929:2010) in a way that they don t differ significantly from those of traditional statistics. Clean /contaminated decision is made using upper limit of confidence interval of activity vs. user entered Alarm Activity. When the Alarm Activity is lower than the Decision Threshold then the unit will go Out of Service. 1

The most important input and output parameters of the Monitor software, Ver. 8.05 are displayed in Table 2. Traditional Methodology Alarm activity Bkg False Alarm (K a ) Extend Confidence (K b1 ) Alarm Confidence (K b2 ) Table 2 Input and Output Parameters Bayesian Methodology Alarm activity False alarm probability (a) Detection probability (1-b) Confidence interval probability (1- g) Bkg Reset Level (K D ) Bkg Reset Level (K D ) Background Average Period Target Uncertainty Detection Efficiency (Calibration) Count time Lower Limit of Detection (LLD) Alarm Level Set Point (ALS) Contaminated Trip Point (CTP) Alarm Trip Point (ATP) Used Trip Point (UTP) Net Count Rate Background Average Period Measurement Process Uncertainty Certificate Emission Rate Uncertainty Confidence Coverage Factor (k c ) Detection Efficiency (Calibration) Detection Efficiency Uncertainty (Calibration) Count time Decision Threshold Detection Limit Best Estimate Activity Uncertainty Best Estimate Activity Confidence Interval INPUT PARAMETERS OUTPUT Default Setting The default settings for the Bayesian methodology were established using empirical data to achieve enhanced performance and balance between the False Alarm Rate (FAR), Detection Probability (DP) and Count Time while keeping the default settings close to the traditional methodology s default settings. The Bayesian methodology provides the confidence interval of activity with the attached probability (1- g ), which is in addition to FAR and DP and it makes difficult the direct comparison of the two methodologies. Based on industry norms and our test results, CANBERRA recommends the following default settings to be used (see Table 3) in any type of contamination monitors, in detecting alpha, beta and gamma radiation, for both Traditional and Bayesian methodologies. Note that the alarm levels for your facility, location, or equipment may be dependent upon different guidance /regulations. Please consult your local authority to select levels appropriate for your use. Table 3 Default setting for Traditional and Bayesian Methodology US units in parenthesis Traditional Methodology Bkg False Alarm K a = 3 Extend Confidence K b1 = 2 Alarm Confidence K b2 = 2 Bkg Reset Level K D = 4 Background Average Period Alarm Activity (Alpha) Alarm Activity (Beta) Alarm Activity (Gamma) 300 s Bayesian Methodology 16.67 Bq (1000 dpm) 83.33 Bq (5000 dpm) 2775 Bq (75 nci Cs-137) False alarm probability (a) a = 0.135%, (k 1-a = 3) Detection probability (1-b) 1-b = 97.725%, (k 1-b = 2) Confidence interval probability (1- g ) Bkg Reset Level K D = 4 Background Average Period Alarm Activity (Alpha) Alarm Activity (Beta) Alarm Activity (Gamma) 1- g = 97.725% 300 s 16.67 Bq (1000 dpm) 83.33 Bq (5000 dpm) 2775 Bq (75 nci Cs-137) 2

Test results Bayesian and Traditional Methodology The Bayesian methodology is described in details in Supplementary Bayesian Methodology User s Manual. The measurements were conducted on the Sirius -5AB contamination monitor. Alpha and beta radiation were measured with a gas flow detector, LFP-579, and gamma radiation was measured with a plastic scintillation detector, TPS-BG-579. All the other detectors and therefore all sum zones were switched off during the measurements. The testing conditions were kept as constant as possible during the tests in order to minimize the number of test variables. The activities of Am-241 and Tc-99 radioactive sources were attenuated by limiting the effective source area, which it was possible by partially closing the shutter attached to the source until the desired activity was achieved: 16.67 Bq for alpha and 83.33 Bq for beta. The activity of the Co-60 gamma source was not attenuated; this source had an activity of 836 Bq when the test was done. No radioactive source was used for the FAR test. The graphical presentation of the measured data can be found in Figure 1, Figure 2, and Figure 3 with the following notations: AA alarm activity e detection efficiency t b background count time R b background count rate indexes a, b and g refer to type of radiation indexes T and B refer to Traditional and respective Bayesian methodology The count time was calculated at the point where the alarm activity (red line) intersects the detection limit curve (see Figure 1). During the measurement, three different scenarios may happen for the measured activity: Activity is smaller than decision threshold (in this case the methodology is not valid for the measurement purpose) Activity is between the decision threshold and detection limit (DP<1-b) Activity is bigger than the detection limit (DP 1-b) Activity (Bq) Count time (s) Figure 1 Detection Limit and Decision Threshold Bayesian Methodology. Alpha alarm set point shown. 3

Test results Input Parameters Traditional methodology Bayesian methodology False Alarm Probability Detection Probability Confidence Interval Probability K a = 3 K b1 = K b2 = 2 a = 0.135%, (k 1-a = 3) 1-b = 97.725%, (k 1-b = 2) 1-g = 97.725% AA a = 16.67 Bq, AA b = 83.33 Bq, AA g = 836 Bq, e a = 4.85%, e b = 4.95%, e g = 2.5%, t b = 300 s, R bta = 0.053 cps, R bba = 0.04 cps, R btb = 6.08 cps, R bbb = 6.08 cps, R btg = 69.75 cps, R bbg = 69.31 cps Count time (s) Figure 2 Count Time for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology For the Bayesian methodology, the count time was calculated as described above, and was rounded up to the closest integer. 4

Figure 3 False Alarm Rate for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology DP (%) FAR (%) Figure 4 Detection Probability for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology As shown in Figure 2, Figure 3 and Figure 4 respectively, the Bayesian methodology provides better results than the Traditional methodology, with slightly lower count times for the alpha radioactivity, some improvements in detection probability, and lower false alarm rates for low levels of alpha radioactivity. 5

In order to reduce the count time, the detection and confidence interval probabilities were decreased to 90%, while keeping the false alarm probability at 0.135%, (see table below): Input Parameters Traditional methodology Bayesian methodology False Alarm Probability Detection Probability Confidence Interval Probability K a = 3 K b1 = K b2 = 1.28 a = 0.135%, (k 1-a = 3) 1-b = 90%, (k 1-b = 1.28) 1-g = 90% AA a = 16.67 Bq, AA b = 83.33 Bq, AA g = 836 Bq, e a = 4.85%, e b = 4.95%, e g = 2.5%, t b = 300 s, R bta = 0.046 cps, R bba = 0.033 cps, R btb = 6.09 cps, R bbb = 6.08 cps, R btg = 67.85 cps, R bbg = 67.68 cps The compared test results can be found in Figure 5, Figure 6 and Figure 7, respectively. Count time (s) Figure 5 Count Time for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology 6

Figure 6 False Alarm Rate for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology DP (%) FAR (%) Figure 7 Detection Probability for Alpha, Beta and Gamma Radiation for the Traditional and Bayesian Methodology As shown in Figure 2 and Figure 5 the count time for the alpha radiation has been reduced from 10 s to 5 s for the Bayesian methodology and from 11 s to 6 s for the Traditional methodology; less significant count time reduction was observed for higher level of beta and gamma radioactivity. The performance degradation in detection probability was more significant for the Traditional methodology and for low radioactivity (89.4% vs. 98.2%), see Figure 4 and Figure 7. No significant change in FAR was observed. 7

Interaction of input parameters and their effect on output results Interactions of different input parameters and their effect on the measurements were studied, in simulation mode for alpha, beta and gamma radiation on a single detector and by using the Bayesian algorithm as implemented in the Monitor software, Ver. 8.05. The test results are graphically presented in Figure 8 to Figure 25. Figure 8 Variation of Count Time in Function of Detection Probability for Different False Alarm Probabilities (Alpha Radiation) Input parameters: Alarm Activity=16.66 Bq, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps. Observation: Significant increase in count time is expected for 1-b>97%. 8

Figure 9 Variation of Count Time in Function of Detection Probability for Different False Alarm Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps. Observation: Significant increase in count time is expected for 1-b>97%. Figure 10 Variation of Count Time in Function of Detection Probability for Different False Alarm Probabilities (Gamma Radiation) Input parameters: Alarm Activity=2775 Bq, e = 5%, u re = 0, S f = 1, t b = 300 s, R b = 70 cps. Observation: Significant increase in count time is expected for 1-b>98%. 9

Figure 11 Variation of Count Time in Function of False Alarm Probability for Different Detection Probabilities (Alpha Radiation) Input parameters: Alarm Activity=16.66 Bq, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps. Observations: Steeper slope of the curves for a <0.5 results in increase of count time. Figure 12 Variation of Count Time in Function of False Alarm Probability for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps. 10

Figure 13 Variation of Count Time in Function of False Alarm Probability for Different Detection Probabilities (Gamma Radiation) Input parameters: Alarm Activity=2775 Bq, e = 5%, u re = 0, S f = 1, t b = 300 s, R b = 70 cps. Figure 14 Variation of Count Time in Function of Detection Efficiency for Different Detection Probabilities (Alpha Radiation) Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps. 11

Figure 15 Variation of Count Time in Function of Detection Efficiency for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps. Figure 16 Variation of Count Time in Function of Detection Efficiency for Different Detection Probabilities (Gamma Radiation) Input parameters: Alarm Activity=2775 Bq, a = 0.135%, u re = 0, S f = 1, t b = 300 s, R b = 70 cps. 12

Figure 17 Variation of Count Time in Function of Background Count Time for Different Detection Probabilities (Alpha Radiation) Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, u re = 0, S f =1, t b = 300 s, R b = 0.05 cps. Figure 18 Variation of Count Time in Function of Background Count Time for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps. 13

Figure 19 Variation of Count Time in Function of Background Count Time for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=2775 Bq, a = 0.135%, u re = 0, S f = 1, t b = 300 s, R b = 70 cps. Figure 20 Variation of Count Time in Function of Relative Uncertainty of Detection Efficiency for Different Detection Probabilities (Alpha Radiation) Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, e = 8%, S f = 1, t b = 300 s, R b = 0.05 cps. 14

Figure 21 Variation of Count Time in Function of Relative Uncertainty of Detection Efficiency for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, e = 8%, S f = 1, t b = 300 s, R b = 6 cps. Figure 22 Variation of Count Time in Function of Relative Uncertainty of Detection Efficiency for Different Detection Probabilities (Gamma Radiation) Input parameters: Alarm Activity=2775 Bq, a = 0.135%, e = 5%, S f = 1, t b = 300 s, R b = 70 cps. 15

Figure 23 Variation of Upper Limit of Confidence Interval in Function of Measured Gross Count Rate for Different Detection Probabilities (Alpha Radiation) Note: The graph also shows the variation of the calculated best estimate of activity (green dashed line) in function of measured gross count rate; the red line on the graph is the alarm activity. When the detected gross count rate is above the alarm activity the contaminated decision will be made; the minimum gross rate at which contaminated decision is made, is at the intersection of alarm activity with the curves (not the best estimate curve) of different 1- g probabilities; the minimum gross rate is lower for higher 1- g. Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, 1-b = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps. Figure 24 Variation of Upper Limit of Confidence Interval in Function of Measured Gross Count Rate for Different Detection Probabilities (Beta Radiation) Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, 1-b = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps. 16

Figure 25 Variation of Upper Limit of Confidence Interval in Function of Measured Gross Count Rate for Different Detection Probabilities (Gamma Radiation) Input parameters: Alarm Activity=2775 Bq, a = 0.135%, 1-b = 97.725%, e = 5%, u re = 0, S f =1, t b = 300 s, R b = 70 cps. Figure 26 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1-β (Alpha Radiation) Note: For fixed count time, but different 1-b, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, 1- g = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps, R g = 1 cps. 17

Figure 27 Variation of Best Estimate and Upper Limit of Confidence Interval vs. α (Alpha Radiation) Note: For fixed count time, but different a, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=16.66 Bq, 1-b = 97.725%, 1- g = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps, R g = 1 cps. Figure 28 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1- γ (Alpha Radiation) Note: The Count Time and Best Estimate don t depend on 1-g. Input parameters: Alarm Activity=16.66 Bq, a = 0.135%, 1-b = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 0.05 cps, R g = 1 cps. 18

Figure 29 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1-β (Beta Radiation) Note: For fixed count time, but different 1-b, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, 1- g = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps, R g = 7 cps. Figure 30 Variation of Best Estimate and Upper Limit of Confidence Interval vs. α (Beta Radiation) Note: For fixed count time, but different a, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=83.33 Bq, 1-b = 97.725%, 1- g = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps, R g = 7 cps. 19

Figure 31 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1- γ (Beta Radiation) Note: The Count Time and Best Estimate don t depend on 1-g. Input parameters: Alarm Activity=83.33 Bq, a = 0.135%, 1-b = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 6 cps, R g = 7 cps. Figure 32 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1-β (Gamma Radiation) Note: For fixed count time, but different 1-b, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=2775 Bq, a = 0.135%, 1-g = 97.725%, e = 8%, u re = 0, S f = 1, t b = 300 s, R b = 70 cps, R g = 71 cps. 20

Figure 33 Variation of Best Estimate and Upper Limit of Confidence Interval vs. α (Gamma Radiation) Note: For fixed count time, but different a, the Best Estimate and Upper Limit of Confidence Interval are constant. Input parameters: Alarm Activity=2775 Bq, 1-b = 97.725%, 1-g = 97.725%, e = 8 %, u re = 0, S f = 1, t b = 300 s, R b = 70 cps, R g = 71 cps. Figure 34 Variation of Best Estimate and Upper Limit of Confidence Interval vs. 1- γ (Gamma Radiation) Note: The Count Time and Best Estimate don t depend on 1-g. Input parameters: Alarm Activity=2775 Bq, a = 0.135%, 1-b = 97.725%, e = 8 %, u re = 0, S f = 1, t b = 300 s, R b = 70 cps, R g = 71 cps. 21

Table 4 summarizes of how the change of input parmeters will affect the detection probability, false alarm rate and count time for the traditional and Bayesian methodology. For fixed count time variation of a and /or 1-b will not affect the FAR & DP. 1-g is not used in the calculation of the count time. Traditional methodology Ka Kb FAR DP Table 4 Interaction of Input and Output Parameters Count time Bayesian methodology a 1-b 1- g FAR DP Input parameters: K a, K b, a, 1-b, 1-g Output parameters: FAR, DP, Count Time Legend: increase, decrease, no change Count time 22

Summary The implementation of the Bayesian statistics into analysis of monitoring data has enhanced the detection sensitivity and ability of our contamination monitoring systems, in detecting and confirming the presence of radioactive contamination efficiently and reliably. The Bayesian methodology is better at lower count rates where the Gaussian approximation for the frequency distribution of net count rate breaks down. Since alarm level set-points are lower and the background is lower, the best performance is achieved for alpha. BAYEX is based on Bayesian statistics and is fully compatible with the ISO11929:2010 standard. It is applicable to all types of radiation and provides a better approach in the detection of radiation since it takes into account uncertainties that are currently not propagated using the traditional methodology. The Bayesian methodology implemented in all of CANBERRA s current contamination monitors gives meaningful positive estimates of net rates and more accurate calculation of decision threshold and detection limit. BAYEX is particularly useful for detecting low activity levels of radioactivity with count rates below background level, where it is hard to distinguish a real signal from noise. It has the potential to reduce the number of false positives without compromising the detection sensitivity. 23

References 1. ISO 11929-1:2000, Determination of the detection limit and decision threshold for ionizing radiation measurements. 2. ISO 11929-1:2010, Determination of the characteristic limits (decision threshold, detection limit and limits of the confidence interval) for measurements of ionizing radiation. 3. JCGM 100:2008, Evaluation of measurement data Guide to the expression of uncertainty in measurement (GUM). Terms and definitions Unless otherwise specified, terms and definition used in this document can be found in the Supplementary Bayesian Methodology User s Manual. Quantities and symbols Unless otherwise specified, quantities and symbols used in this document can be found in the Supplementary Bayesian Methodology User s Manual". Sirius is a trademark and/or registered trademark of Mirion Technologies, Inc. and/or its affiliates in the United States and/or other countries. All other trademarks are the property of their respective owners. Radiation Safety. Amplified. www.canberra.com Part of Mirion Technologies 24 C40547-03 /2014