23 ISD Series Integrating Detectors for Laser Power Integrating Detectors The ISD series Laser power detectors are set-up with Silicon or Indium-Gallium-Arsenide photodiode detectors mounted to integrating spheres. The ISD detectors are suitable for measurement of both narrow and divergent beam Lasers. The measurement port size must be selected so that the complete Laser output signal enters the measurement port of the integrating sphere. Wide Range of Stock Items Standard ISD integrating sphere detectors are supplied from stock with diameters from 8mm to 300mm and measurement port sizes from 2 to 100mm diameter. Integrating spheres with Barium Sulfate coating and with the synthetic ODM98 coating are offered. Individual Custom Configuration and Design ISD-0.8-SiLP 8mm Diameter Integrating Laser Detector The Gigahertz-Optik GmbH s UM series modular integrating spheres and accessory components makes custom solutions easy. Select a base sphere at the required diameter and coating type then lay out a port plan and add port adapters, detectors, baffles and stands as stock items. Please see Light Measurement System Components on our website for more integrating sphere information. OEM Design and Production Gigahertz-Optik GmbH has many years of experience in OEM design and production of integrating spheres and integrating sphere detectors. Please contact our technical sales support to discuss your requirements. ISD-30-SiIGA 300mm Diameter Integrating Laser Detector with Si and InGaAs Detector Calibration of luminous flux lm is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities.
2 Outline Diameter Diameter ISD-0.8P-SiLP 8mm 2mm ODM98 00-1100nm SiLP Photodiode ISD-1.6-Si 16mm 3mm ODP97 00-1100nm ISD-3P-Si 30mm 5mm ODM98 00-1100nm ISD-3P-IGA 30mm 5mm ODM98 800-1800nm InGaAs Photod. ISD-5P-Si 50mm 12.5mm ODM98 00-1100nm ISD-5P-SiUV 50mm 12.5mm ODM98 250-1100nm ISD-5P-IGA 50mm 12.5mm ODM98 800-1800nm InGaAs Photod. ISD-10-Si 100mm 15mm ODP97 00-1100nm ISD-15-Si 150mm 25mm ODP97 00-1100nm ISD-30-Si-IGA 300mm 101mm ODP97 00-1800nm Si & InGaAs ISD-SUB ISD-SMA DP-11 K-ISDSi-SD K-ISDSiUV-SD K-ISDIGA-SD 33/W@633nm 6/W@900nm 0.3/W@633nm 27/W@900nm 2.9µA/W@633nm 28µA/W@900nm 0.17/W@1300nm 0.2/W@1550nm /W@633nm 5/W@900nm 0.7/W@250nm /W@633nm 5.5/W@900nm 0.1/W@1300nm 0.12/W@1550nm 1.6/W@633nm 1.9/W@900nm 60µA/W@633nm 70µA/W@900nm 0.2/W@633nm 0.25/W@900nm 25µA/W@1300nm 18µ A/W@1550nm 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- 1 5-0 C 2m -1,-2,- Optional UMPF-0.5 substitution port with UMPP-0.5 port plug. Offers the possibility for substitution correction with optional auxiliary lamp. For UM based ISD models only Optional UMPF-0.5_UMPA-0.5/11_UFC-11-SMA detector port with SMA connector. For UM based ISD models only Optional detector port for UPK based ISD models only Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm Calibration of spectral responsivity in A/W in 10nm steps from 250 to 1100nm Calibration of spectral responsivity in A/W in 10nm steps from 800 to 1800nm K-ISDSiIGA-SD Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1800nm
25 ISD-0.8P-SiLP: 8mm Ø Integrating Detector for Laser Power Compact Integrating for High and Fast Light Throughput The ISD-0.8P-SiLP integrating sphere detector employs an 8 mm diameter ODM coated integrating sphere. The small size integrating sphere ensures high light level throughput through low attenuation as well as a fast light throughput by short transit time of the light signal (speed of light divided by the product of sphere diameter and sphere factor). 1 2 3 Low Profile Si-Photodiode with Fast Rise Time The low profile Si photodiodes (identical to model PD- 1106) offer fast rise time for longer wavelengths as compared to crystalline type silicon photodiodes. Measurement or Fiber Adapter An UFC-11/SMA fiber connector adapter can be mount to the ISD-0.8P-SiLP integrating sphere detector s measurement port. Calibration of spectral radiant power sensitivity in W is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities. 1) 2mmØ Measurement 2) 8mmØ Integrating with ODM98 3) Low Profile Silicon Photodiode ISD-0.8P-SiLP Laser Power Detector and its Principal Design 200 00 600 800 1000 1200 ISD-0.8P-SiLP actinic curve none ISD-0.8P-SiLP
26 Specifications Max Power Peak/CW ISD-0.8P-SiLP 8mmØ 2mmØ ODM98 00-1100nm SiLP Diode K-ISDSi-SD 33/W@633nm 6/W@900nm 30mW / *) 22mW / *) Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm *) The max CW or average power is limit by the max operation temperature Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description ISD-0.8P-SiLP 10282-1 - - Integrating Detector, Calibration Certificate Re-Calibration Item No. K-ISDSi-SD 300285 Dimensions: 100929_ISD-0.8-SiLP.pdf Purchasing Information Detector Heads Options Description UFC-11-SMA 100929 SMA connector adapter PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
27 ISD-1.6-Si and ISD-1.6-Si-FC: 16mm Ø Integrating Detector for Laser Power Compact Integrating Detector The ISD-1.6-Si integrating sphere detector employs a 16 mm diameter Barium Sulfate (ODP97) coated integrating sphere with 3mm measurement port and a Silicon photodiode. Direct Irradiation of Fiber Connector The ISD-1.6-Si-FC is designed for applications where both the power and spectral distribution of Laser diodes must be measured. High resolution spectrometers are needed to measure the narrow spectra Laser diodes so the FC fiber connector is placed opposite the measurement port of the integrating sphere to allow direct irradiation of the fiber for maximum radiation input to the spectrometer. Calibration of spectral radiant power sensitivity in W is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities. ISD-1.6-Si-FC Integrating Laser Detector 1) 2) 3) ) 1) 3mmØ Measurement 2) 16mmØ ODP97 Coated 3) ) FC Fiber Connector with Fiber to High Resolution Diode Array Spectrometer ISD-1.6-Si-FC Principal Design FC Fiber Connector at the Back Side 200 00 600 800 1000 1200 ISD-1.6-Si actinic curve none ISD-1.6-Si
28 Specifications ISD-1.6-Si 16mm 3mm ODP97 00-1100nm ISD-1.6-Si-FC 16mm 3mm ODP97 00-1100nm 0.3/W@633nm 0.27/W@900nm 0.3/W@633nm 0.27/W@900nm Max Power Peak/CW 3.3W / *) 3.7W / *) 3.3W / *) 3.7W / *) K-ISD1.6-SD Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm *) The max CW or average power is limit by the max operation temperature Dimensions: 102619_ISD-1.6-Si.pdf 102618_ISD-1.6-Si-FC.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration Item No. ISD-1.6-Si 102619 102619-1 102619-2 Integrating Detector, Calibration Certificate ISD-1.6-Si-FC 102618 102618-1 102618-2 Integrating Detector, Calibration Certificate K-ISD1.6-Si-SD 300286 K-ISD1.6-Si-SD 300286 Purchasing Information Detector Heads Options Description - - -
29 ISD-3P-Si and ISD-3P-IGA: 30mmØ Integrating Detector for Laser Power Compact Integrating with Unique Collar Baffle The ISD-3P integrating sphere detectors employ a 30 mm diameter ODM98 coated integrating sphere with 5mm diameter measurement port. The unique collar baffle limits direct irradiation of the detector from dispersed light. The collar baffle enables optional detector ports to be added for additional detectors, fiber connectors or auxiliary use. Synthetic for Best Light Distribution Gigahertz-Optik s synthetic ODM98 coating offers almost perfect hemispherical diffuse reflection for best light distribution inside the sphere. Available with Si and InGaAs Photodiode The ISD-3P-Si features a Si photodiode detector for use in the spectral range from 00 to 1100nm where as the ISD-3P-IGA an InGaAs photodiode for the 800 to 1800nm spectral range.. Calibration of spectral radiant power sensitivity in W is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities. ISD-3P Integrating Laser Power Detector 200 00 600 800 1000 1200 ISD-3P-Si actinic curve none ISD-3P-Si 800 1000 1200 100 1600 1800 ISD-3P-IGA actinic curve none ISD-3P-IGA ISD-3P in Custom Set-up with InGaAs Detectors
30 Specifications Max Power Peak/CW ISD-3P-Si 30mm 5mm ODM98 00-1100nm Si diode 2.9µA/W@633nm 28µA/W@900nm 30W / *) 35W / *) ISD-3P-IGA 30mm 5mm ODM98 800-1800nm InGaAs diode UP-DP-11 K-ISDSi-SD 0.17/W@1300nm 0.2/W@1550nm Optional detector port for UPK based ISD models only 5.8W / *) 5W / *) Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm K-ISDIGA-SD Calibration of spectral responsivity in A/W in 10nm steps from 800 to 1800nm *) The max CW or average power is limit by the max operation temperature Dimensions: 10283_ISD-3P-Si.pdf 1028_ISD-3P-IGA.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration Item No. ISD-3P-Si 10283-1 10283-2 10283-3 Integrating Detector, Calibration Certificate ISD-3P-IGA 1028-1 1028-2 1028-3 Integrating Detector, Calibration Certificate K-ISD3PSi-SD 300287 K-ISD3PIGA-SD 300288 Purchasing Information Detector Heads Options Description UP-DP-11 101219 Additional detector port UFC-11-SMA 100929 SMA connector adapter PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
31 ISD-5P-Si, ISD-5P-SiUV and ISD-5P-IGA: 50mmØ Integrating Detector for Laser Power Compact 50mm Diameter Integrating Detector The ISD-5P integrating sphere detectors employ a 50 mm diameter ODM98 coated integrating sphere. The unique collar baffle limits direct irradiation of the detector from dispersed light. The collar baffle enables optional detector ports to be added for additional detectors or fiber connectors. Synthetic for Best Light Distribution Gigahertz-Optik s synthetic ODM98 coating offers almost perfect hemispherical diffuse reflection for best light distribution inside the sphere. Available with Si, Si-UV and InGaAs Photodiode The ISD-5P-Si features a Si photodiode detector covering the spectral range from 00 to 1100nm. The ISD -5P-Si is calibrated for the extended wavelength range from 250 to 1100nm to cover UV applications. The ISD-5P-IGA integrating sphere detector uses an In- GaAs photodiode for the 800 to 1800nm spectral range. Calibration of spectral radiant power sensitivity in W is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities. ISD-5P Integrating Laser Power Detector 200 00 600 800 1000 1200 ISD-5P-Si actinic curve none ISD-5P-Si 200 00 600 800 1000 1200 ISD-5P-SiUV actinic curve none ISD-5P-SiUV 800 1000 1200 100 1600 1800 ISD-5P-IGA actinic curve none ISD-3P-IGA
32 Specifications Max Power Peak/CW ISD-5P-Si 50mm 12.5mm ODM98 00-1100nm /W@633nm 5.5/W@900nm 0.25W / *) 0,18W / *) ISD-5P-SiUV ISD-5P-IGA DP-11 K-ISDSi-SD 50mm 12.5mm ODM98 250-1100nm 50mm 12.5mm ODM98 800-1800nm InGaAs Photod 0.7/W@250nm /W@633nm 5.5/W@900nm 0.1/W@1300nm 0.12/W@1550nm Optional detector port for UPK based ISD models only 1.W / *) 0.25W / *) 0.18W / *) 10W / *) 8.3W / *) Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm K-ISDSiUV-SD Calibration of spectral responsivity in A/W in 10nm steps from 250 to 1100nm K-ISDIGA-SD Calibration of spectral responsivity in A/W in 10nm steps from 800 to 1800nm *) The max CW or average power is limit by the max operation temperature Dimensions: 10285_ISD-5P-Si.pdf 10286_ISD-5P-SiUV.pdf 10287_ISD-5P-IGA.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration ISD-5P-Si 10285-1 10285-2 10285-3 Integrating Detector, Calibration Certificate ISD-5P-SiUV 10286-1 10286-2 10286-3 Integrating Detector, Calibration Certificate ISD-5P-IGA 10287-1 10287-2 10287-3 Integrating Detector, Calibration Certificate K-ISD5PSi-SD 300289 K-ISD5PSiUV-SD 300290 K-ISD5PIGA-SD 300291 Purchasing Information Detector Heads Options Description UP-DP-11 101219 Additional detector port UFC-11-SMA 100929 SMA connector adapter PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
33 ISD-10-Si: 100mmØ Integrating Detector for Laser Power Compact Integrating Detector The ISD-10-Si integrating sphere detector employs an 100 mm diameter Barium Sulfate (ODP97) coated integrating sphere with 15mm measurement port and a Silicon photodiode. Detector with Diffuser Window Because of the diffuser window in front of the Si photodiode the detector integrates reflected radiation from almost the complete integrating sphere surface. This minimizes any hot-spot effects caused by the first reflection from the Laser. Detector in Shadow Position to the Measurement ISD-10-Si Integrating Laser Power Detector The detector is positioned close to the measurement port for a large acceptance angle within the sphere without direct irradiation of the detector. Optional Fiber Connector An optional SMA connector is available to add a spectrometer to the sphere for relative spectral distribution measurements. Optional Auxiliary In applications where substitution errors caused by back reflected light from test samples placed at the measurement port should be compensated an auxiliary port is optionally available. Calibration of spectral radiant power sensitivity in W is provided by Gigahertz-Optik s Calibration Laboratory.. 200 00 600 800 1000 1200 ISD-10-Si actinic curve none ISD-10-Si
3 Specifications Max Power Peak/CW ISD-10-Si 100m m 15mm ODP97 00-1100nm 1.6/W@633nm 1.9A/W@900nm 0.6W / *) 0.5W / *) ISD-AUX Optional UMPF-0.5 auxiliary port with UMPP-0.5 port plug. ISD-SMA Optional UMPF-0.5_UMPA-0.5/11_UFC-11-SMA detector port with SMA connector K-ISDSi-SD Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm *) The max CW or average power is limit by the max operation temperature Dimensions: 10288_ISD-10-Si.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration Item No. ISD-10-Si 10288-1 10288-2 10288-3 Integrating Detector, Calibration Certificate K-ISD10Si-SD 300292 Purchasing Information Detector Heads Options Description ISD-AUX 102831 Auxiliary port with port plug ISD-SMA 102833 SMA detector port PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
35 ISD-15-Si: 150mmØ Integrating Detector for Laser Power 6 Inch Integrating Detector The ISD-15-Si integrating sphere detector with 150mm diameter integrating sphere represents the most common 6 inch class of integrating spheres. The sphere is Barium Sulfate (ODP97) coated with a 25.mm diameter measurement port and a Silicon photodiode detector. Detector with Diffuser Window Because of the diffuser window in front of the Si photodiode the detector integrates reflected radiation from almost the complete integrating sphere surface. This minimizes any hot-spot effects caused by the first reflection from the Laser. Detector in Shadow Position to the Measurement The detector is positioned close to the measurement port for a large acceptance angle within the sphere without direct irradiation of the detector. Optional Fiber Connector An optional SMA connector is available to add a spectrometer to the sphere for relative spectral distribution measurements. Optional Auxiliary In applications where substitution errors caused by back reflected light from test samples at the measurement port should be compensated an auxiliary port is optionally available. ISD-15-Si Integrating Laser Power Detector 200 00 600 800 1000 1200 ISD- 15- Si actinic curve none ISD-15-Si Calibration of spectral radiant power sensitivity in W is provided by Gigahertz-Optik s Calibration Laboratory..
36 Specifications Max Power Peak/CW ISD-15-Si 150mm 25mm ODP97 00-1100nm 60µA/W@633nm 70µA/W@900nm 16W / *) 1W *) ISD-AUX Optional UMPF-0.5 auxiliary port with UMPP-0.5 port plug ISD-SMA Optional UMPF-0.5_UMPA-0.5/11_UFC-11-SMA detector port with SMA connector K-ISDSi-SD Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1100nm *) The max CW or average power is limit by the max operation temperature Dimensions: 10089_ISD-15-Si.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration Item No. ISD-15-Si 10289-1 10289-2 10289-3 Integrating Detector, Calibration Certificate K-ISD15Si-SD 300293 Purchasing Information Detector Heads Options Description ISD-AUX 102831 Auxiliary port with port plug ISD-SMA 102833 SMA detector port PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
37 ISD-30-Si-IGA: 300mmØ Integrating Detector for Laser Power 100mm Diameter Measurement The ISD-30-SiIGA integrating sphere detector with 300mm diameter integrating sphere features a large 100mm diameter measurement port making it suitable for use with larger diameter Laser beams like Laser range finders. Its two detectors, Si and InGaAs photodiodes, cover the spectral range from 00nm to 1800nm. The sphere is Barium Sulfate (ODP97) coated. Detectors with Diffuser Windows Diffuser windows mounted in front of the detectors enable the detectors to integrate reflected radiation from almost the complete integrating sphere surface minimizing hot-spots caused by the first beam reflection. Detector in Shadow Position The detector is positioned close to the measurement port for a large acceptance angle within the sphere without direct irradiation of the detector. Optional Fiber Connector An optional SMA connector is available to add a spectrometer onto the sphere for relative spectral distribution measurements. Optional Auxiliary In applications where substitution errors caused by back reflected light from test samples at the measurement port should be compensated an auxiliary port is optionally available. Calibration of spectral radiant power sensitivity in W is provided by the Gigahertz-Optik calibration lab.. ISD-30-SiIGA Integrating Laser Power Detector 200 00 600 800 1000 1200 ISD-30-Si ISD-30-Si actinic curve none 800 1000 1200 100 1600 1800 ISD-30-IGA actinic curve none ISD-30-IGA
38 Specifications Max Power Peak/CW ISD-30-SiIGA 300mm 100mm ODP97 00-1100nm 800-1800nm InGaAs Photod. 0.2/W@633nm 0.25/W@900nm 25µA/W@1300nm 18µA/W@1550nm 5W / *) W / *) 0W / *) 55W / *) 1 1 5-0 C 5-0 C 2m 2m -1,-2-1,-2 ISD-AUX Optional UMPF-0.5 auxiliary port with UMPP-0.5 port plug ISD-SMA Optional UMPF-0.5_UMPA-0.5/11_UFC-11-SMA detector port with SMA connector K-ISDSiIGA-SD Calibration of spectral responsivity in A/W in 10nm steps from 00 to 1800nm *) The max CW or average power is limit by the max operation temperature Dimensions: 100850_ISD-30-SiIGA.pdf Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description Re-Calibration Item No. ISD-30-SiIGA 102850-1 102850-2 102850-3 Integrating Detector, Calibration Certificate K-ISD30SiIGA-SD 30029 Purchasing Information Detector Heads Options Description ISD-AUX 102831 Auxiliary port with port plug ISD-SMA 102833 SMA detector port PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information
18 ISD-5-VISNIR Integrating Detector for Total Radiant Flux of 2 Sources Integrating Detector for 2 Light Sources The ISD-5-VISNIR integrating sphere detector for total radiant flux measurement over the wavelength range from 00 to 1000nm is set-up with a radiometric detector mounted to a compact size integrating sphere. All light flux must enter the measurement port of the integrating sphere for total radiant flux measurement. The ISD-5-VISNIR detector is suitable for measurement of directional type light sources, e.g. cold light sources with light guide and other narrow beam sources. Compact Integrating The ISD-5VISNIR detector features a 50mm diameter integrating sphere with ODP97 Barium Sulfate coating and a 12.5mm diameter measurement port. Full Custom Configuration and Design Gigahertz-Optik GmbH s UM series modular integrating spheres and accessory components makes custom solutions easy. Select a base sphere at the required diameter and coating type then lay out a port plan and add port adapters, detectors, baffles and stands as stock items. Please see Light Measurement System Components on our website for more integrating sphere information. ISD-5-VISNIR 200 00 600 800 1000 1200 ISD-5-VISNIR actinic curve 00-1000nm ISD-5-VISNIR Calibration of luminous flux lm is performed at Gigahertz-Optik s Calibration Laboratory for Optical Radiation Quantities. Specifications Diameter Diameter ISD-5-VISNIR 50mm 12.5mm ODP97 00-1000nm 0.1/W K-ISD5VISNIR-I Calibration of integral luminous flux sensitivity in A/lm K-Si-SR Calibration of the relative spectral responsivity within the detectors spectral range Purchasing Information Detector Heads and Re-Calibrations -1 Con. -2 Con. - Con. Description ISD-5-VISNIR 10281-1 10281-2 10281-3 Integrating Detector, Calibration Certificate Re-Calibration Item No. K-ISD5VISNIR-I 30028 Purchasing Information Detector Heads Options Description K-Si-SR 301209 Calibration of relative spectral responsivity from 250-1100nm or the detectors spectral responsivity range PMS- *) Post-Mount Stand components *) Please check chapter Light Measurement System Component for detail information