AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Similar documents
AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

I.F.U Re/Processing Reusable Medical Devices

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Classification of Instruments -

Processing instructions for the Ligamys suture forceps

Processing Instructions. Cleaning so easy.

O-PEP Hygiene-Beileger

REUSABLE ENDOSCOPY INSTRUMENTS (thermo resistant)

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Cleaning and Sterilisation for Burs, accessories and filling material

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

Instructions for Use Spinal Punches IFU

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

RAP-hip Hip stem extraction set

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Cleaning and Sterilization for Instruments & Posts

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Elaborating of a Reprocessing Instruction According to ISO Klaus Roth

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Data of 57 hospitals** Estimated for England and Wales

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060)

Instructions for Use (GA16)

Table of Contents. English

Aesculap Implant Systems

LipoCollector II Operating Instructions

Reusable Laryngoscope Blades and Handles

Electric steam steriliser

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Söring Ultrasonic Generator Sonoca 300

Instruction of use. Flexible Naso-Pharyngo- Laryngofiberscope,

Cleaning & Sterilization

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

RECOMMENDATIONS FOR CLEANING DEVICES

CLEANING INSTRUCTIONS

Operating / User Manual

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

MODEL 7000 SUCTION UNIT

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Aesculap Implant Systems

ASPIRE Laboratory Aspirator

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Reprocessing of surgical instruments

Recommendations for cleaning, decontamination and sterilisation. devices

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

ENDOILLUMINATION. in accordance with EN ISO 17664

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED

Hydraulic crimping tool. HO-16 Art QZERTIFIKAT DIN EN ISO convincing solutions

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Instructions for use Vacuum regulation valve with water trap

Manual. Reprocessing. RICHARD WOLF Heat-Stable Instruments. GA-J020 / en / V17.0 / PDI

Hydraulic crimping tool. HO-6 Art convincing solutions

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

Instructions for Cleaning, Sterilization, Inspection and Maintenance. of Trauma & Extremities Medical Devices

2015 Myco Industries, Inc Fax:

STERITITE INSTRUCTIONS FOR USE

TIH 300 S / TIH 400 S / TIH 500 S / TIH 700 S / TIH 900 S / TIH 1100 S

EN Bagged Vacuum Cleaner

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

X99 Series. Dental Handpiece. Instruction For Use

Basic SURGICAL SUCTION PUMP. EN Instructions for use

2CKA001473B System Manual Busch-Infoline. Handicapped toilet signal set 1510 UC

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

THE BASICS OF DECONTAMINATION

Manual Cleaning? SOBECC Round Table: Manual and Automated Cleaning Methods th July 2012

Operating Instructions. Accessory Units Melitta Cafina XT Series. Melitta Professional Coffee Solutions

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

Device Reprocessing Checklist Scoring Matrix 1=low 5=high

AIRGOCLEAN 10 E OPERATING MANUAL AIR CLEANER TRT-BA-AIRGOCLEAN10E-TC-001-EN

PICNet CONFERENCE 2013

HKF 8180 Operating instructions

When you sterilize an instrument set, you know what the standards are.

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily.

Operating Instructions. Rotating Vacuum Filter Type VRF

Clinical Use Only. ResMed HumidAire 2i. Disinfection and Sterilisation Guide. English

Steam Trap BK 45 BK 45-U BK 45-LT BK 46

FLASH PAK STERILIZATION CONTAINER SYSTEM

WORK IN THE DISINFECTION ROOM WITH WASHER-DISINFECTORS

Care instructions ESTETICA E80. Always be on the safe side.

STERIS Instrument Cleaning Chemistries

TTV 1500 / TTV 3000 OPERATING MANUAL CONVEYING FAN TRT-BA-TTV TC EN

Dr. Weigert - New Standards in Cleaning and Disinfection of Medical Devices

Crimping tool HHs-6. Art convincing solutions. Video

Manual Page 1 of 8. Manual. Laboratory autoclave

Additionally, consider the legal provisions valid for your country as well as to the hygienic instructions of the doctor s practice or hospital.

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

INSTRUCTION MANUAL W-400

RESISTENT EXTRA 3592, 3602 RESISTENT EXTRA PLUS 3592P, 3602P

Transcription:

HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks 65-227-01-65-240-12 1 Safety and responsibility Before using the instrument Read and follow the Instructions for Use. Keep these instructions in a place where they can be easily seen for reference at a later date. AS Medizintechnik GmbH does not assume any liability for damage as a result of improper use or incorrect preparation and care. In such cases the warranty is rendered void. Indication This instrument is designed for use in minimally invasive surgery and, in particular, arthroscopy. Only to be used by qualified specialist personnel. Contraindication This instrument is not designed for use on the central nervous and circulatory system. Risk of infection Prepare before each use. Prepare before returning to the manufacturer. Risk of injury Only use original or compatible thirdparty accessories. Do not use damaged instruments and do not repair. 2 Scope of supply After unpacking, check the instrument for any damage caused during transit. In the event of complaints, contact the manufacturer. For the manufacturer s address see the front page. 3 Product description The telescope is clamped to the arthroscopy sheath using the locking mechanism. Irrigation and suction is possible via two stop cocks. 1 Union nut 2 Irrigation openings 3 Telescope lock with locking mechanism 4 Cover ring with two stop cocks and LL connection GA_Arthroskopieschaft HIGH-FLOW_EN_C Page 1 of 3

HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks 4 Technical data Storage and transport conditions Designation Value Temperature -30 C +40 C Relative humidity 90 % Air pressure 700 1200 hpa Product lifetime 1000 cycles and 5 years 5 Assembly and disassembly of the instrument 5.1 Assembly of the instrument Fit the O-rings. Position the cock plugs in the cock body so that the pin is located in the guide. GA_Arthroskopieschaft HIGH-FLOW_EN_C Page 2 of 3

HIGH-FLOW Arthroscopy Sheath with one or two rotating Stopcocks Screw the spring caps onto the cock plugs. Pull the cover ring over the sheath tube. Fit the union nut. 5.2 Disassembly of the instrument Unscrew the union nut. Remove the cover ring. Unscrew the spring caps and remove the cock plugs. Remove the O-rings. 6 Before use 6.1 Important information Use components which are technologically compatible. Example: For instruments with a diameter of 2 mm an obturator of the same diameter is ideal. 6.2 Preparing the intervention Risk of infection from non-sterile instruments. Prepare the instrument before using for the first time and then prior to each subsequent use. Insert the obturator into the instrument and clamp securely into place using the locking mechanism. Make the incision. Release the locking mechanism and remove the obturator. Insert the telescope into the instrument and clamp securely into place using the locking mechanism. Perform the intervention. Prepare the instrument. See chapter 7 Preparation. 7 Preparation Risk of infection from non-sterile instruments. Prepare the instrument before use. LL The instrument must be dismantled prior to preparation. You will find more detailed information on preparation and care on the following pages. 8 Storage You will find more detailed information on preparation and care on the following pages. 9 Disposal Environmentally sound disposal enables valuable raw materials to be recycled. Dispose of the device in an environmentally friendly manner in accordance with the valid hospital guidelines. AS MEDIZINTECHNIK GMBH DOES NOT ACCEPT RESPONSIBILITY IF THIS CUSTOMER INFORMATION HAS BEEN VIOLATED PROVABLY. GA_Arthroskopieschaft HIGH-FLOW_EN_C Page 3 of 3

Contents Page 1 Safety instructions 2 Overview of reprocessing methods 2.1 Validated reprocessing methods 2.2 Material compatibility 3 Scope 4 Explanation of symbols 4.1 Warnings 4.2 Symbols used 5 Preparing decontamination 6 Pre-cleaning 7 Cleaning and disinfection 7.1 General information 7.2 Manual cleaning (enzymatic) and chemical disinfection 7.3 Mechanical cleaning (alkaline) and thermal disinfection 7.4 Mechanical cleaning (enzymatic) and thermal disinfection 8 Checking and care 9 Packaging 10 Sterilising 11 Storing sterile devices 12 Repair 13 Disposal 14 Information on validation 1 Safety instructions These are not a substitute for the Instructions for Use. Read and follow the Instructions for Use. Before reprocessing instruments Read and follow the. Keep these instructions in a place where they can be easily seen for reference at a later date. The manufacturer does not assume any liability for damage as a result of incorrect reprocessing and care. The methods described for manual and mechanical reprocessing have been validated by the manufacturer. These are recommendations from the manufacturer. If using different and/or modified reprocessing methods, the user must provide proof of the effectiveness of the method. The basic risks identified for the reprocessing process are summarised below based on the nature of the hazard. Warnings which relate to specific actions can be found directly next to the relevant action together with a classification of the risk. See section Explanation of symbols. Risk of infection Reprocess the instrument before initial use. Reprocess the instrument before each use. Reprocess the instrument before returning it to the manufacturer. Follow the instructions for use of the cleaning agents and disinfectants used as well as of the cleaning and sterilisation devices used. Wear personal protective equipment during reprocessing. Page 1 of 11

Discard disposable components after initial use. Note the special reprocessing requirements if there is a suspicion of prions and Creutzfeldt-Jakob disease. Risk of injury Do not use damaged instruments and do not repair. Only use original accessories. 2 Overview of reprocessing methods 2.1 Validated reprocessing methods In the following synoptic table, the validated reprocessing methods are shown itemised according to the instrument groups. The individual instruments are assigned to the different reprocessing groups in the respective Instructions for Use. This synoptic table is not a substitute for the following sections of these. ++ Validated reprocessing method 0 Process not validated by the manufacturer - Incompatibility More detailed information on reprocessing is available from the Robert Koch Institute (RKI) and the Instrument Reprocessing Working Group (AKI): RKI: AKI: Hygiene Requirements for Reprocessing Medical Devices (www.rki.de, as last amended) Proper Maintenance of Instruments (www.a-k-i.org, as last amended) Page 2 of 11

2.2 Material compatibility The following synoptic table shows for which instrument groups and which reprocessing methods materials are compatible. The user must validate these methods! The individual instruments are assigned to the different reprocessing groups in the respective Instructions for Use. This synoptic table is not a substitute for the following sections of these. + Method with verified material compatibility 0 Process not validated by the manufacturer - Incompatibility More detailed information on reprocessing is available from the Robert Koch Institute (RKI) and the Instrument Reprocessing Working Group (AKI): RKI: AKI: Hygiene Requirements for Reprocessing Medical Devices (www.rki.de, as last amended) Proper Maintenance of Instruments (www.a-k-i.org, as last amended) Page 3 of 11

3 Scope These are valid for all products where explicit reference is made to the in the Instructions for Use. 4 Explanation of symbols 4.1 Warnings DANGER Indicates a danger which results in death or serious injury if not avoided. Indicates a danger which can result in death or serious injury if not avoided. CAUTION Indicates a danger which can result in injuries if not avoided. IMPORTANT! Indicates measures in order to prevent damage to property. The safety alert symbol in accordance with ANSI Z535.6 is used in these Instructions for Use to indicate warnings of personal injury. 4.2 Symbols used Follow the Instructions for Use. Manufacturer of the medical device. Important! Refer to the Instructions for Use for important safety information such as warnings and precautionary measures. Specifies measures to reduce risks. You are requested to take action here. You find out the result of the action taken here. This symbol indicates additional information. 5 Preparing decontamination To prevent surgical residue from drying on, the following steps must be performed directly after surgery. Rinse the instrument with cold water. Remove coarse dirt with cold water. Rinse out cavities with cold water. If it is not possible to rinse with cold water, the instrument must be wrapped in a moist cloth to prevent any residues from drying on. Always transport the instrument to the reprocessing site in a closed container to prevent product damage and contamination of the environment. Page 4 of 11

6 Pre-cleaning Pre-cleaning prevents surgical residue from drying on. It must therefore be carried out directly after surgery. Pre-cleaning was validated with the cleaning agent Cidezyme from Johnson & Johnson: Cleaning Dosage ph value Enzymatic 0.8% 7.8-8.8 (diluted) Risk of infection and pyrogenicity from residues if unsuitable cleaning agents are used. Do not use fixing agents. Do not rinse with hot water. IMPORTANT! Avoid damaging product. Do not use abrasive brushes or scourers. Only use the cleaning agents which are listed in this section. Use disinfectant with corrosion protection. TIP: Remove caked-on tissue residues with a plastic brush. Immerse the instrument in a cold water bath with 0.8% cleaning solution: >5 min. Brush the instrument under cold water until all visible signs of soiling have been removed. Dismantle the instrument (if possible). See instrument s Instructions for Use. Open stop cocks (if relevant). Rinse the instrument under the surface of the water. This prevents contamination of the surrounding area. Brush the outside and inside under cold water with a round brush until no more residue is visible. Rinse out cavities, drill holes and threads (if relevant) with a cleaning gun: >10 s at 3-5 bar. Remove from the water bath and rinse off with cold water. Immerse in combined cleaning and disinfectant solution until subsequent cleaning to prevent any residue fromdrying on. 7 Cleaning and disinfection 7.1 General information Unless stated otherwise in the instrument s Instructions for Use, cleaning is performed with the instrument disassembled. Unless specified otherwise in the description of the reprocessing method, use water of the following quality: Pre-rinsing: Tap water Cleaning and rinsing: Demineralised water Final rinsing: Demineralised water Page 5 of 11

Risk of infection due to insufficient reprocessing. Remove protective caps (if relevant). IMPORTANT! Avoid damaging product. Do not use abrasive brushes or scourers. Only use the cleaning agents which are listed in the individual sections. With plastic instruments avoid contact with hydrogen peroxide (H2O2). Choose between manual and mechanical cleaning. Risk of infection due to insufficient reprocessing. Special reprocessing requirements must be observed if there is a suspicion of prions and Creutzfeldt-Jakob disease. To this end, note Annex 7 of the recommendation Hygiene Requirements for Reprocessing Medical Devices from RKI and BfArM. 7.2 Manual cleaning (enzymatic) and chemical disinfection Manual cleaning was validated with the cleaning agents Cidezyme/Enzol from Johnson & Johnson and Mucadont Zymaktiv from Merz Hygiene GmbH: Cleaning agent Cleaning Dosage Cidezyme/Enzol Enzymatic 0.8% Mucadont/Zymaktiv Enzymatic 0.8% Rinse the instrument under the surface of the water. This prevents contamination of the surrounding area. Preparation Dismantle the instrument (if possible). See instrument s Instructions for Use. Open stop cocks (if relevant). Cleaning Place the instrument in cold water: >10 min. Brush the instrument under cold water until all visible signs of soiling have been removed. Rinse out cavities, drill holes and threads with a cleaning gun with cold water: >20 s at 3-5 bar. Unless specified otherwise in the instrument s Instructions for Use, clean the components in the ultrasonic bath with 0.8% cleaning agent with the following settings: Temperature Frequency Duration 40-45 C 35 khz 10-15 min. Turn and move the components several times during cleaning in the ultrasonic bath. Rinse out cavities, drill holes and threads with a cleaning gun with cold water: >20 s at 3-5 bar. Immerse the instrument in deionised water and rinse through the cavities several times with deionised water. Page 6 of 11

Also clean the inner chambers of the instrument below water using a cleaning brush. Drying Dry on the inside and outside: >10 min. at 50-100 C and/or blow through with sterile compressed air. Disinfecting Disinfection was validated with the disinfectant Mucocit-T from Merz Hygiene GmbH: Disinfect with disinfectant: >10 min. See manufacturer s Instructions for Use. Disinfectant Dosage ph value Mucocit-T 4% 10.5 (diluted) 7.3 Mechanical cleaning (alkaline) and thermal disinfection Mechanical cleaning was validated with the cleaning device G 7735 CD from Miele. Mechanical alkaline cleaning was validated with the cleaning agent Neodisher FA from Dr. Weigert: Cleaning Dosage ph value Alkaline 0.5% 12.2-14 (diluted) Preparation Dismantle the instrument (if possible). See instrument s Instructions for Use. Open stop cocks (if relevant). Place instruments in a sieve tray on the cleaning device MIS mobile unit so that the cleaning agent can reach all inner and outer surfaces. Connect flush opening (if any) to MIS mobile unit. Cleaning Unless specified otherwise in the instrument s Instructions for Use, additionally clean the components prior to or in combination with mechanical cleaning in an ultrasonic bath with the following settings: Temperature Frequency Duration 40-45 C 35-45 khz 10-15 min. Turn and move the components several times during cleaning in the ultrasonic bath. Proper cleaning depends on the right cleaning program. Compare the cleaning program with that of your cleaning system and save where necessary: Start the cleaning program. 1. Pre-rinsing with cold water: 1 min. 2. Emptying 3. Repeated pre-rinsing with cold water: 3 min. 4. Emptying 5. Cleaning with 0.5% alkaline cleaning agent: at 55 C for 5 min. 6. Emptying 7. Neutralising with deionised water: 3 min. 8. Emptying Page 7 of 11

Drying 9. Rinsing with deionised water: 2 min. 10. Emptying Dry inside and outside in the cleaning and disinfection device: 15-25 min. at 90-110 C. The cleaning and disinfection device cycle must include a drying phase. Remove the product from the cleaning and disinfection device immediately at the end of the cycle. If necessary, also blow through the product with sterile compressed air until it is completely dry. Disinfecting Disinfection was validated with the device Getinge 88 series: Disinfectant Temperature Dwell time Demineralised water* 90+3 C 5 min Observe the national requirements as regards the A0 value (see ISO 15883) * Deionised/demineralised water 7.4 Mechanical cleaning (enzymatic) and thermal disinfection Mechanical cleaning was validated with the cleaning device G 7735 CD from Miele. Mechanical enzymatic cleaning was validated with the cleaning agent deconex 23 Neutrazym from Borer: Cleaning Dosage ph value Enzymatic 0.5% 8.8 (diluted) Preparation Dismantle the instrument (if possible). See instrument s Instructions for Use. Open stop cocks (if relevant). Place instruments in a sieve tray on the cleaning device MIS mobile unit so that the cleaning agent can reach all inner and outer surfaces. Connect flush opening (if any) to MIS mobile unit. Cleaning Unless specified otherwise in the instrument s Instructions for Use, clean the components prior to mechanical cleaning or, if possible, in the cleaning device together with ultrasound with the following settings: Temperature Frequency Duration 40-45 C 35-45 khz 10-15 min. Turn and move the components several times during cleaning in the ultrasonic bath. Proper cleaning depends on the right cleaning program. Compare the cleaning program with that of your cleaning system and save where necessary: Start the cleaning program. 1. Pre-rinsing with cold water: 1 min. 2. Emptying 3. Repeated pre-rinsing with cold water: 3 min. Page 8 of 11

4. Emptying 5. Cleaning with 0.5% enzymatic cleaning agent: at 45 C for 5 min. 6. Emptying 7. Rinsing with deionised water: 3 min. 8. Emptying 9. Rinsing with deionised water: 2 min. 10. Emptying Drying Dry inside and outside in the cleaning and disinfection device: 15-25 min. at 90-110 C. The cleaning and disinfection device cycle must include a drying phase. Remove the product from the cleaning and disinfection device immediately at the end of the cycle. If necessary, also blow through the product with sterile compressed air until it is completely dry. Disinfecting Disinfection was validated with the device Getinge 88 series: Disinfectant Temperature Dwell time Demineralised water* 90+3 C 5 min Observe the national requirements as regards the A0 value (see ISO 15883) * Deionised/demineralised water 8 Checking and care The right care of instruments will lengthen their service life and should therefore be carried out after every cleaning process. Risk of injury from faulty or damaged components Do not use damaged instruments and do not repair. Check to ensure they are clean and, if necessary, repeat cleaning. Check for damage (e.g., sharp edges, rough surfaces). Replace brittle and cracked seals (if relevant). Lubricate moving parts (e.g., joints, rotating stop cocks) with medical oil. Remove any excess oil. Assemble instruments (if possible) and check to ensure they are in perfect working order. See instrument s Instructions for Use. 9 Packaging The instrument must be packed appropriately prior to sterilisation to ensure that the sterile barrier remains intact after removal from the steriliser. Package the instrument to comply with ISO 11607 and EN 868. 10 Sterilising Unless stated otherwise in the instrument s Instructions for Use, sterilisation is performed with the instrument assembled. See instrument s Instructions for Use. Sterilisation was validated with the sterilisation device Selectomat S 3000 from MMM Group and Varioclav 400 E from Fisher Scientific. Page 9 of 11

IMPORTANT! Avoid damaging product. Observe the device s maximum load. See manufacturer s Instructions for Use. Risk of infection due to insufficient reprocessing. Special reprocessing requirements must be observed if there is a suspicion of prions and Creutzfeldt-Jakob disease. To this end, note Annex 7 of the recommendation Hygiene Requirements for Reprocessing Medical Devices from RKI and BfArM. Sterilisation was validated at 134 C and 4 minutes. The sterilisation time must therefore be at least 4 minutes. The following countries make different stipulations which must be observed: Country Germany France Switzerland Sterilisation time 4-30 min. 18-30 min. 18-30 min. Dwell times of 4-30 min. do not have a negative influence on the instrument/material. Open stop cocks (if relevant). Place in the sterilisation device so that the components are not touching each other and the steam can circulate freely. Set the sterilisation parameters (3-fold fractionated pre-vacuum): Temperature Pressure Duration 134-137 C 3 bar 44 psi See details in this section. Start the sterilisation process. 11 Storing sterile devices To avoid reducing durability and forfeiting any resistance to bacteria the following storage conditions must be observed: Store the sterile device sealed in a clean, dust-free and dry sterile container. Protect from direct light. Store the sterile container in a clean and dry environment with controlled humidity at room temperature. Do not store the sterile container in the vicinity of aggressive substances (e.g., alcohols, acids, bases, solvents and disinfectants). Also observe your internal storage standards for sterile devices. 12 Repairs Risk of injury from improper repairs. Only allow repairs to be performed by the manufacturer or by persons authorised by the manufacturer. Page 10 of 11

Risk of infection from non-sterile instruments. Reprocess the instrument before returning it to the manufacturer. Send the instrument back to the manufacturer in a reprocessed state and in its original packaging. For the manufacturer s address see the front page. 13 Disposal Environmentally sound disposal enables valuable raw materials to be recycled. Dispose of the device in an environmentally friendly manner in accordance with the valid hospital guidelines. 14 Information on validation The following materials and machines were used for validation: Cleaning agent: Cidezyme/Enzol from ASP Mucadont Zymaktiv from Merz Hygiene GmbH Neodisher FA from Dr. Weigert deconex 23 Neutrazym from Borer Disinfectant: - Mucocit-T from Merz Hygiene GmbH Cleaning and disinfection devices: G 7735 CD from Miele Sterilisation devices: Selectomat S 3000 from MMM Group Varioclav 400 E from Fisher Scientific Sterilising agent: Moist heat AS MEDIZINTECHNIK GMBH DOES NOT ACCEPT RESPONSIBILITY IF THIS CUSTOMER INFORMATION HAS BEEN VIOLATED PROVABLY. Page 11 of 11