Measuring Couch Motion Accuracy with X-ray Beam Fluence

Similar documents
Independent validation of machine performance check for the Halcyon and TrueBeam linacs for daily quality assurance

Understanding total measurement uncertainty in power meters and detectors

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

Doc. No. SP R0. CXI Detector stage. Prepared by: Signature Date Jean-Charles Castagna Design Engineer

Development of the CMS Phase-1 Pixel Online Monitoring System and the Evolution of Pixel Leakage Current

HPF 1.2 NEW INTERCONNECTION SYSTEM FOR EXTREME VIBRATION REQUIREMENTS IN THE VEHICLE

When Precision is not good enough

An investigation of the reproducibility and usefulness of automatic couch motion in complex radiation therapy techniques

Dosimetric verification of intensity modulated radiation therapy

SYNCHROTRON RADIATION ABSORPTION AND VACUUM ISSUES IN THE IR*

Case Study. A Geothermal Energy Generation Facility Embraces Infrared. Polymer Window Inspections. Overview: iriss.com

FEE Diagnostics and Commissioning. June 17, 2008

VIBRA-series: VIBRA, VIBRA +

Victoreen & 660-8

Analisi del Focus-shift su sorgenti multikw. Luca Porcelluzzi Ophir Spiricon Europe GmbH

MapCHECK 3. The New Benchmark for 2D IMRT QA. Your Most Valuable QA and Dosimetry Tools

ISO Acceptance tests for Nd:YAG laser beam welding machines Machines with optical fibre delivery Laser assembly

Hands-On Laser Beam Path Alignment Practical. Environment, Safety, Health & Quality Office SERF E141 Last updated: 12/28/2017

Application Note. Application Note for BAYEX

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Customized MCPs for Analytical Instruments. An MCP for every application

Doc. No. SP R0. CXI Reference Laser. Prepared by: Jean-Charles Castagna Signature Date Design Engineer

Physics Requirements for the CXI Detector Stage

Laser Distance Sensor Type M Analog-Output

WHAT IS LASER ULTRASONIC TESTING?

FEATURE ARTICLE. The PLCA-800 Series Inline Particle Sensor for Chemical Solutions. Yoshihito Yuhara. Abstract

On-line Side Stream Oil in Water Analyser - Model OIW-EX 100 or OIW-EX 1000

Application Note. Which characterization method is best for your application? The Future of Thermal Imaging is Here!!!

Link loss measurement uncertainties: OTDR vs. light source power meter By EXFO s Systems Engineering and Research Team

INTERNATIONAL STANDARD

Design Considerations of LED based Visible Notification Appliances. Johnny K. W. Siu, Ph.D. - Light Engine Limited I. INTRODUCTION

9 W, 15 W and 25 W Industrial Fiber Laser Systems IFL Series

PROFILER Family IC PROFILER, IC PROFILER - MR & SRS PROFILER. Your Most Valuable QA and Dosimetry Tools

Establishing a Center of Excellence for. Security Science and Engineering

NoFlex Linear Drive Technology Ensures Accuracy in Dispensing of Adhesives and Solder Paste

Vehicle Detection: How To Choose the Right Technology... 3

SECTION Process and Immersion Turbidity and Total Suspended Solids Sensor

Chapter 1. Introduction

This is to certify that the optical output of the: SENSORNET LR-DTS Temperature Sensing System: (Variants Mark 1, Mark 2a and Mark 2b)

COMMISSIONING OF THE DOSE DELIVERY SYSTEM AT MEDAUSTRON

Rework: IR Technology vs. Convection Air

BACK SCATTER ELECTRON (BSE) Rev. 0313

XGLAB. X and Gamma Ray Electronics. Looking for new frontiers...

Active Scanning Beam 2: Controlling Delivery

The focused solution for YAG laser procedures

OptoYag M. clear vision

25 TEXAS ADMINISTRATIVE CODE Radiation Safety Requirements for Analytical and Other Industrial Radiation Machines

LCLS Instruments. Jerry Hastings June 20, 2008 LCLS

clear & preserve vision OptoYag&SLT

High-Power Q-Switched Diode-Pumped UV Laser Q-Series, Q305

LHCb Rich Detectors Control and High Voltage Systems

Bunch Length Interferometer System

Conductive Heating Technologies for Medical Diagnostic Equipment. Russell Strehlow, Manager of Research and Development, Minco

CT398 Heaterstat TM Temperature Controller. Technical User Guide

PHOTOELECTRIC DUAL BEAM DETECTOR R4222 RECEIVER INDICATORS 1

Development of a variable output LED-365 irradiance standard

INTERNATIONAL STANDARD

POWER Selecting Shell Side Design Temperature for Feedwater Heaters

Summary of Comments (Washington Revisions November 7, 2000) Update November 27, 2000

SECTION Fluorescent Dissolved Oxygen Measuring System

Force and Bungee Cord Length: Will Longer Length Un-Stretched of Cord result in a Greater or Lesser Force than a Shorter Length of Un-Stretched Cord

Miniature Infrared Temperature Sensors with Optional Touch Screen Display. PSC-SSS-PM with touch screen display (-CRT version)

RS Pro Infrared Temperature Sensor

THEORY OR OPERATION 2 SENSOR UNIT - MECHANICAL 6 SENSOR UNIT - ELECTRICAL 8 CONTROL UNIT - MECHANICAL 9 CONTROL UNIT - ELECTRICAL 9 OPTIONS 11

MX1A and MX1B: Laser Displacement Detection Sensors. 85% to 110% of the rated voltage MX1A: 15VA, MX1B: 25VA

Minnesota Rules, Chapter 4732 X-Ray Revision

state of the art methane leak detection CHARM and GasCam 2011 October 13 th Dr. Axel Scherello

for Photonics Advanced phosphor and glass technologies for high performance alignment, detection, imaging and spectroscopy applications

TacBio: A UV-based Biological Aerosol Detector. TacBio Overview Page 1

FLIGHT UNLOADING IN ROTARY SUGAR DRYERS. P.F. BRITTON, P.A. SCHNEIDER and M.E. SHEEHAN. James Cook University

Provläsningsexemplar / Preview TECHNICAL REPORT. Nuclear medicine instrumentation Routine tests

The following report is prepared and published by:

Guidelines for the Evaluation of Direct Reading Monitors

Constant climate chambers

Spectrum Illumination Machine Vision Lighting

/Mar/2014 Rev A

PRODUCT CONFORMITY CERTIFICATE

SCAN200E USER S MANUAL

Lecture 8 OPENINGS &ORGANIZATION OF FORM AND SPACE

LASER SOLUTIONS. Ensure Accurate Setup from Simulation to Treatment. FIXED: Micro // MOVABLE: CT Sim+

CONsOlIDATOR 4 & 8. MulTI- C h ANNEl CONTROllERs. ConsoliDator 4 Model PD940 ConsoliDator 4 Features. ConsoliDator 8 Features.

LFE50. UV Flame Safeguard. Building Technologies Division DETACTOGYR

Preparation of response measurements for crossover simulation with VituixCAD

DH400. Accessories. NOTE: Set up for this sensor should be performed by an AAADM-certified installer.

CLEAN-LASERSYSTEME GMBH

A Study on the 2-D Temperature Distribution of the Strip due to Induction Heater

Development of a Transient Simulation Model of a Freezer Part II: Comparison of Experimental Data with Model

R&D for the improvement of O&M in CSP plants. Dr. Marcelino Sánchez. - November,

Today s Outline - September 12, C. Segre (IIT) PHYS Fall 2016 September 12, / 21

I am Rick Jeffress and I handle sales for Fike Video Image Detection. We thank the AFAA for coordinating the venue for this presentation, and we

Instruction Manual. Self-Leveling Combination Cross-Line Laser and Five-Beam Laser Dot Model No , &

Gaseous detectors. 1

in mobility Gentle & precise cleaning of sensitive Very quiet air-cooled system Large operating distance (up to 250 mm) Laser class 4 product

Audible warning devices for external alarm

Technologies, Solutions, and Applications. Continuous Level Coke Drum Measurement System

Sensors and Systems for the Steel Industry

Laser Weld Keyhole Dynamics and Control. M. H. Cho, D.F. Farson Welding Engineering Department, The Ohio State University, Columbus, Ohio

TPC BASED IMP. Limin Duan, Gas Detector Group. Institute of Modern Physics, CAS TPC tracker detector technology mini-workshop

Stony Brook University Hospital Environmental Health & Safety Policy & Procedure Manual. EH&S 6-8 Original : 1/00 Revision Date: 3/16/15 Pages 6

Transcription:

Measuring Couch Motion Accuracy with X-ray Beam Fluence Rudi Labarbe, M.S. and Brett Nelson, M.S. IBA Particle Therapy, Belgium and Logos Systems, Scotts Valley, CA Introduction The accuracy of radiosurgery systems can be verified and maintained with quality assurance (QA) procedures performed on a regular schedule. Typically these procedures focus on the operational health of the mechanical delivery system for the radiation (gantry or robot), the beam shaping multi-leaf collimators, and the energy profile of the X-ray or proton beam radiation source. To ensure that radiation is delivered accurately to the patient, it is also essential that the patient treatment couch and the electro-mechanics be examined periodically. This report examines the use of an imaging system with innovative scintillator geometry to record the position of an IBA proton therapy treatment couch as it was moved through a series of linear displacements. Figure 1 X-ray or Proton Beam Interaction with the XRV-100 Scintillator The heart of the XRV-100 phantom (Logos Systems, Scotts Valley, CA) consists of an imaging cone (Figure 1) laminated with Gadox X-ray scintillator phosphor. As ionizing radiation passes through the cone, two spots of visible light are formed by Compton scattering at the entry and exit points of the beam. These two spots precisely define the path of the radiation beam through three-dimensional (3D) space. By calculating the beam vector and position from these spots, the XRV software can track the gantry or robot in real-time. Alternatively, the gantry can be stationary while the imaging system monitors the patient couch as it moves through various positions during the course of a simulated treatment. Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 1 of 8

Materials and Methods The XRV phantom was placed on the IBA treatment couch as shown in Figure 2 such that movement along the couch in the negative Y direction could be easily captured. The IBA gantry was rotated to an overhead position (0 degrees) and an X-ray source of approximately 150 kv within the aperture enclosure was used to project photons through the MLC forming a square beam approximately 8.4 mm on a side. Position measurement tests were performed with the beam stationary and the couch moving along its Y axis. The negative Y axis movement of the couch corresponds to negative Z axis movement of the beam spots upon the XRV imaging cone. The imaging cone dimensions allowed the X-ray beam position to be captured and measured along a distance of 60 mm. The pattern of the test was to expose the phantom with X-rays for approximately 1 second, move the couch in the negative Y direction a random distance, and then repeat. The X-ray beam was manually energized by push button, so the actual duration of the beam fluctuated around the 1 second target. Figure 2 - Phantom Positioning on Couch with Corresponding Coordinate Systems The XRV imaging cone is manufactured with reference holes precisely machined at 10 millimeter Z axis intervals along the top, bottom, and sides of the cone. These holes can be backlit to form a cross shaped pattern that is used for calibration at the factory or in the field. These holes can be seen at the bottom of Figure 3 and range from 40 millimeters to 110 millimeters from the cone base flange on the left side of the image. Factory calibration also involves the use of a radiographic measuring bar that can be temporarily fastened to the imaging cone. The measuring bar has X-ray opaque fiducials mounted at 10 millimeter intervals along its length. The bar is shown attached to the cone base flange at the top of Figure 3. Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 2 of 8

Figure 3 - Imaging Cone with Radiographic Measuring Bar Attached Figure 4 shows the pattern of dots formed by the reference holes when the cone is backlit. The software uses the position and spacing of the dots on this 2D camera image to calibrate 3D locations on the inner cone surface. The outer surface of the cone is usually covered with a shroud in order to darken the phantom interior. This optical calibration system of backlighting the cone and updating the software parameters is used each time the phantom is transported in its carrying case. Figure 4 - Calibration Reference Hole Measurements Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 3 of 8

Data Representation The initial XYZ couch position during the test was -.3, -.03, -13.58 centimeters according to the IBA data monitors in the treatment room. The corresponding XRV imaging cone Z axis position was 102.9 millimeters. The couch was moved in an increasingly negative Y axis IBA position (decreasing Z axis XRV position) 44 times until it reached a maximum excursion at -6.37 cm. The couch was then moved in the positive Y direction several times until it was back near its starting location. Figure 5 shows the XYZ positions of the couch at beam 29 and beam 47. Figure 5 - Proton Therapy Couch Position Monitor While the X-ray beam was pulsed at each new couch position, the XRV software was monitoring the beam spot activity on the inside of the imaging cone, detecting the start of new beams, calculating beam metrics, and archiving beam images and data to disk. The XRV digital camera used a frame rate of 10 frames per second, a shutter period of 100 milliseconds, and a gain of 7.2 db to produce beam spots of optimal brightness for analysis. Each captured beam was integrated over approximately 7-8 frames depending on how long the X-ray control button was pushed. Beam data was saved in a comma separated variable (CSV) format that was later input to a spreadsheet. Figure 6 shows the square beam spots captured at four couch positions superimposed on a single image. The outer blue ring corresponds to a distance of 110 millimeters from the cone base flange and the inner blue ring corresponds to a distance of 40 millimeters. The spacing between each of the concentric rings is 10 mm. The width of the top beam spot along the arc is measured to be 9.01 mm. The measurement in the vertical direction (also called radial) was 9.13 mm. The beam enters the cone vertically yielding a brighter entry spot at the top and a dimmer exit spot at the bottom. As shown above in Figure 2, the XRV positive X axis maps onto the IBA couch negative X axis and the XRV positive Y axis maps onto the IBA negative Z axis. Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 4 of 8

Figure 6 - Superimposed Beam Spots from 4 Couch Positions The four couch positions in Figure 6 starting from the top are at beams 1, 48, 47, and 46. These correspond to couch Y axis positions of -.03 cm, -.42 cm, -2.14 cm, and -5.02 cm. It can be seen from the overlap of the entry beam spot for beam 1 and 48 that there is a several millimeter drift in the X position of the beam from the start of the test to the end. This is also evident in the change of X location from -.32 cm to -.67 cm (delta of 3.5 mm) in the Current columns of Figure 5. The cause for this was that the buttons that control the couch X axis were briefly pressed in the process of positioning the couch Y axis for beams 46, 47, and 48. This unplanned movement along the X axis does not interfere with measurements along the Y axis. Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 5 of 8

Test Results and Analysis The couch Y axis IBA position readings were: -.03, -.08, -.39, -.42, -.46, -.84, -1.36, -1.42, -1.44, -1.48, -1.50, - 1.61, -1.65, -2.13, -2.15, -2.18, -2.31, -2.45, -2.50, -2.77, -2.86, -3.09, -3.32, -3.36, -3.38, -3.45, -3.59, -3.84, - 4.17, -4.23, -4.26, -4.39, -4.47, -4.58, -4.69, -4.89, -5.00, -5.07, -5.13, -5.22, -5.30, -5.32, -5.40, -5.75, -6.37, - 5.02, -2.14, and -.42 centimeters. These positions were also captured by the XRV software and archived to disk starting at the initial cone Z position of 102.9 millimeters. Shown in Figure 7 is a linear regression of all the couch and XRV phantom measurement points. The absolute value of the couch positions are used in order to facilitate a more convenient representation of the data. The fitted slope of.9991 demonstrates excellent agreement between the two datasets indicating that the non-linearity between the XRV measurements and the internal measurements of the treatment couch electro-mechanics is less than.1%. IBA Treatment Couch vs XRV-100 Positions 70 Couch Positions (mm) 60 50 40 30 20 y = -0.9991x + 103.14 R 2 = 1 Series1 Linear Best Fit 10 0 0 20 40 60 80 100 120 XRV-100 Measurements (mm) Figure 7 Linear Regression of the Couch positions versus the XRV-100 Measurements The XRV measurements can now be used as the x value input to the fitted line equation to produce idealized y value couch positions. As an example, using the second XRV reading of 102.307 mm, the resulting couch measurement according to the fitted line would be 102.21 103.14 or -.925 mm giving a delta of.1 mm from the actual reading of -.08 cm (-.8 mm). Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 6 of 8

Couch Variations from Linear Regression (mm) 0.14 0.12 0.1 0.08 0.06 0.04 0.02 0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 Figure 8 Variation of Couch Measurements from Linear Regression equation Figure 8 contains the absolute value of all the deltas between the couch measurements and the fitted line couch values when using the XRV measurements for input as X into the equation for the line. The largest delta is.1 mm and is located at the second XRV reading. The average of the values in Figure 8 is.05 mm and represents the combined reporting error of both the couch electro-mechanics and the XRV hardware/software. Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 7 of 8

Conclusion The measurement data demonstrates that the average variation of the IBA couch positions versus the XRV-100 measured positions along a single axis of motion was.05 mm with the maximum variation being.1 mm. These values represent the combined measurement error of the two systems and it is beyond the scope of this study to further isolate the individual accuracy of the couch or the XRV system. Therefore, the XRV-100 phantom maximum measurement error along the imaging cone Z-axis was less than.1 mm and the average error less than.05 mm. These error values are well within the published.2 mm margin of the XRV-100 device. Contact Information Rudi Labarbe, M.S. System Engineer IBA Particle Therapy Chemin du Cyclotron 3 1348 Louvain-la-Nueve Belgium Brett Nelson, M.S. Director of Engineering Logos Systems Int l 175 El Pueblo Road Scotts Valley, CA 95066 www.logosvisionsystem.com Phone: 831-600-6101 02/15/12 Measuring Couch Motion Accuracy with X-ray Beam Fluence Page 8 of 8