Instruction Manual CHLORALERT TM Series T17CA4000 Chlorine Gas Leak Detector CAPITAL CONTROLS

Similar documents
Electrochemical Chlorine Sensor Module Addition or Replacement Series 1610B And 1620B

Instruction Manual Series 1650/1660 Gas Detector Sensor CAPITAL CONTROLS

Instruction Manual - Anti-Siphon Ejector Chlorine & Sulfur Dioxide 500 PPD (10 kg/h) Maximum Capacity

ASPIRE Laboratory Aspirator

Models TX-KE and TX-KP Toxic Gas Transmitters

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR OWNER S MANUAL

MST, INC. MODEL 2002 CARBON MONOXIDE MONITOR (AMBIENT-TYPE) OWNER S MANUAL

CAPITAL CONTROLS SERIES WP70CV3 GAS FEEDERS

MST, INC. MODEL 2002 OXYGEN MONITOR (AMBIENT-TYPE) OWNER S MANUAL

MODEL A-316 CARBON MONOXIDE (CO) MONITOR AND ALARM FOR COMPRESSED AIR TESTING

Upholstery and Drapery Cleaner. Operator and Parts Manual. Model No.: gal Extractor. MNL32506 Rev. 00 (08-98)

I N S T R U C T I O N S

Section K CHLORINATION EQUIPMENT

Nuclear Associates

TABLE OF CONTENTS FOR YOUR SAFETY

Damp Locations Wall & Ceiling Mount BEFORE YOU BEGIN

MST MODEL S1 AIRLINE "CO" MONITORING AND LOW PRESSURE WARNING SYSTEM MANUAL

BRITA. Water Filter & Chiller II. Chilled water filter. Instruction Booklet WF6100

FIREPLACE HEATER FOR THE FOLLOWING MODELS: EE-8075W EE-8075O EE-8075R

Installation & Operating Instructions

2-IN-1 Warm Mist Humidifier with Steam Inhaler

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

ULTRASONIC HUMIDIFIER With Night Light and Fragrance Diffuser

OWNER S MANUAL. FLAT PANEL FIREPLACE HEATER Model Number: Serena (EF202A) WARNING CAUTION

CONSTANT SUCTION UNIT

Capital Controls Series 480 All vacuum gas feeders designed for manual or semi-automatic gas regulation.

ELECTRIC STOVE HEATER Models: FS2213B

TQ Series. Electronic Control Pump Instruction Manual. 50Hz. ISO 9001 Certified Walrus Pump Co., Ltd.

Ionizer (Static Remover) STABLO-EX Instruction Manual

CERAMIC TOWER HEATER

- 1 - Updated on 18 March, 2010

MODEL 7000 SUCTION UNIT

METS MODEL E-112. (115 VAC, 50/60 Hz) USER S MANUAL INSTALLATION OPERATION MAINTENANCE TABLE OF CONTENTS

CAPITAL CONTROLS SERIES 480

WARNING: Warns of health hazards and identifies possible risks of injury. CAUTION: Indicates possible dangers to the machine or other objects.

T1 SCRUBBER DRYER OPERATOR MANUAL

INSTALLATION INSTRUCTIONS

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

ADORABLE Ultrasonic Cool Mist Humidifier

PERSONAL Ultrasonic Cool Mist Humidifier

RK/ RK Oxygen Detector Operator s Manual

Explorer CleaningPartsDirect.com Carpet Extractor. Operator and Parts Manual. Home Find... Go To.. Model No.: Can.

Decon Scrubber Owner s Manual

INDUSTRIAL ELECTROSTATIC AIR CLEANER MODEL: SH-SSP-300

RLD-H10G Refrigerant Leak Detectors

CERAMIC TOWER HEATER Instruction Manual

Original instructions JSNY7 Magnetic switch

One-Touch Dispense. Multi-temp selections. 208 F For tea, coffee, instant noodle. 194 F Keep warm around 194 F. 176 F Keep warm around 176 F.

LED IMPORTANT SAFETY INSTRUCTIONS

DIGITAL ULTRASONIC HUMIDIFIER WITH HUMIDISTAT

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0036A

EVAPORATIVE COOL MIST HUMIDIFIER Instruction Manual

FOR THE FOLLOWING MODELS: If you have any questions about the operation of your ultrasonic humidifier, please contact Crane Customer Care.

3M Benchtop Air Ionizer 963E. User s Guide

TECHNICAL INSTRUCTIONS

P.N ELECTRIC KETTLE User Manual Series MODEL BVSTKT308

SERIES VAC Microprocessor-Based Direct Spark Ignition Control with Inducer Blower Relay FEATURES APPLICATIONS SPECIFICATIONS DESCRIPTION

Enteral Pump Instructions for Use To Assemble When the low

USER GUIDE. DRENA 2 - User Manual ELECTRICAL PANEL FOR 2 MOTORS - WASTE WATER -

A measured step forward

RLD-H10P Refrigerant Leak Detector

Dr. Infrared Heater Elite Series

Antari Z-300 ORDERCODE 60661

Victoreen Primalert 35 Area Radiation Monitor

OWNER S MANUAL ELECTRIC FLAT PANEL FIREPLACE. AC V 60Hz. Model# : /60 - Built-in & Wall Mounted Electric Fireplace WARNING CAUTION

RADARAY XLi INSTALLATION & SERVICE MANUAL

Advisory Bulletin. TB-106 Cleaning and Maintenance For Bobrick B-850, B-855, B-856 or B-858 Automatic Counter-Mounted Soap Dispensers

INSTALLATION & OPERATION Rev 1.00A DISTRIBUTED EXCLUSIVELY BY CMA DISHMACHINES KNOTT AVENUE GARDEN GROVE, CALIFORNIA 92841

! "! # Please read these installation and operating instructions completely and carefully before starting.

Safety Information Contents List Odorox Boss XL3 Unit Operational Guidelines Operating The Odorox Boss XL3... 4

Reacti-Therm I and III Heating Modules

Instruction Manual. Cheese Warmer with Pump

Instruction Manual Series 17CA3000 Chloralert TM Plus Multi-Gas Detector CAPITAL CONTROLS

Instruction Manual Vaporizer for Chlorine or Sulfur Dioxide Series VAX4600

OWNER S MANUAL ELECTRIC FIREPLACE INSERT. Model No.: WESTERN (EF42D-45D) WARNING CAUTION

Owner's Manual. WS Series. Water Softener

EP8020 STEAM GENERATOR IRON PROFESSIONAL TYPE IRONING SYSTEM WITH DETACHABLE WATER TANK FOR HOUSEHOLD USE ONLY

Model 405 AIRGARD Fume Hood Monitor

Ultrasonic Humidification System Model 5462 & Model 5472 OPERATING MANUAL

IMPORTANT SAFETY INFORMATION! WARNING ALWAYS keep electric cords, home furnishings, drapes, clothing, papers, or other combustibles at least 3 feet (0

SAFETY OPERATION & MAINTENANCE MANUAL X-612 CARPET EXTRACTOR

Instructions for use. 301 plus

V. (Versus ) Cordless Bare Floor Vacuum

OPERATION MANUAL. CLASSIC 10 and CLASSIC 18 SERIAL NUMBER FROM JANUARY 2009 (0109) TO PRESENT

OWNER S MANUAL. Models SCU-H11, SCU-H9, SCU-H7, SCU-H5, SCU-H3

RK/ RK Carbon Monoxide Detector Operator s Manual

Infrared Portable Space Heater

Operating Instructions. For. Alarm Control Module. Model SSR 500

IMPORTANT SAFEGUARDS

COOPER POWER SERIES. Reclosers MN280015EN

METS MODEL E-106. (115 VAC, 60 Hz) USER S MANUAL INSTALLATION OPERATION MAINTENANCE

PRE INSTALLATION PROCEDURES

Instruction Manual. Bucket Oil Pump

SAFETY OPERATION & MAINTENANCE MANUAL X-405 CARPET EXTRACTOR This unit is intended for commercial use.

User Manual GV25 GV35 GV702. Company information: Original instructions GV12066 (1)

CERAMIC HEATER MODEL # EE If you have any questions about the operation of your ceramic heater, please contact Crane Customer Care.

CLEAN MIST ULTRASONIC HUMIDIFIER. Instruction Manual and Warranty Information READ AND SAVE THESE INSTRUCTIONS IM0040A

Super Runner Series Electric Fryers

NO VENTING REQUIRED 120V AC;

7911A Constant Temperature Ice Bath. User s Manual

Transcription:

Instruction Manual CHLORALERT TM Series T17CA4000 Chlorine Gas Leak Detector CAPITAL CONTROLS - 1-325.6030.7

These instructions describe the installation, operation and maintenance of the subject equipment. Failure to strictly follow these instructions can lead to an equipment rupture that may cause significant property damage, severe personal injury and even death. If you do not understand these instructions, please call De Nora Water Technologies for clarification before commencing any work at +1 215-997-4000 and ask for a Field Service Manager. De Nora Water Technologies, Inc. reserves the rights to make engineering refinements that may not be described herein. It is the responsibility of the installer to contact De Nora Water Technologies, Inc. for information that cannot be answered specifically by these instructions. Any customer request to alter or reduce the design safeguards incorporated into De Nora Water Technologies equipment is conditioned on the customer absolving De Nora Water Technologies from any consequences of such a decision. De Nora Water Technologies has developed the recommended installation, operating and maintenance procedures with careful attention to safety. In addition to instruction/operating manuals, all instructions given on labels or attached tags should be followed. Regardless of these efforts, it is not possible to eliminate all hazards from the equipment or foresee every possible hazard that may occur. It is the responsibility of the installer to ensure that the recommended installation instructions are followed. It is the responsibility of the user to ensure that the recommended operating and maintenance instructions are followed. De Nora Water Technologies, Inc. cannot be responsible deviations from the recommended instructions that may result in a hazardous or unsafe condition. De Nora Water Technologies, Inc. cannot be responsible for the overall system design of which our equipment may be an integral part of or any unauthorized modifications to the equipment made by any party other that De Nora Water Technologies, Inc. De Nora Water Technologies, Inc. takes all reasonable precautions in packaging the equipment to prevent shipping damage. Carefully inspect each item and report damages immediately to the shipping agent involved for equipment shipped F.O.B. Colmar or to De Nora Water Technologies for equipment shipped F.O.B Jobsite. Do not install damaged equipment. De Nora Water Technologies, Colmar Operations Colmar, Pennsylvania, USA is ISO 9001: 2008 Certified READ THE ENTIRE MANUAL BEFORE OPERATING USE ONLY IN ACCORDANCE WITH INSTRUCTION MANUAL WARNING: HAZARDOUS VOLTAGES 325.6030.7-2 -

INSTRUCTION BULLETIN CHLORINE GAS LEAK DETECTOR "CHLORALERT" Model T17CA4000 Rev. 1.2 IMPORTANT The instructions given herein cover generally the description, installation, operation and maintenance of subject equipment. DNWT reserves the right to make engineering refinements that may not be reflected in this Bulletin. Should any question arise which may not be answered specifically by these instructions, they should be directed to De Nora Water Technologies for further detailed information and technical assistance. DNWT takes all possible precautions in packing each equipment item to prevent damage during shipment. Carefully inspect each item and if damage has occurred report it immediately. Do not install any equipment if damage is such that faulty operation is likely to result. If plant engineering drawing are available at the installation site, locate, mount, pipe and wire thesubject equipment in accordance with these drawings. If plant engineering drawings are not available, refer to the Installation section for information concerning these requirements. Carefully inspect all packing material before discarding it to prevent loss of mounting hardware, accessories, spare parts or instructions. All instructions given on any attached tag should be followed. WARNING To assure safe operation this equipment carefully follow use and installation instructions and recommendations illustrated in this Manual. Improper use of the equipment may damage the equipment and endanger the safety of the operating personnel. - 3-325.6030.7

PAGE INTENTIONALLY LEFT BLANK 325.6030.7-4 -

TABLE OF CONTENTS 1 INTRODUCTION... 6 1.1 General Description... 6 2 TECHNICAL SPECIFICATIONS... 7 3 MODEL NUMBER BREAKDOWN... 9 3.1 Standard accessories... 9 4 INSTALLATION... 10 4.1 Location...10 4.2 Mounting...10 4.3 Sampling line...11 4.4 Electrical wiring...11 5 OPERATION, MAINTENANCE AND TESTING... 12 5.1 Electrolyte filling at start up...12 5.2 Electrolyte renewal after operation...12 5.3 Sample flow rate...12 5.4 Gas Test 12 5.5 Circuit performance check...13 6 MAINTENANCE PROCEDURES... 14 6.1 Measuring cell maintenance...14 6.2 Flowmeter...14 6.2.1 For Model T17CA4100...14 6.2.1 For Model T17CA4200...14 6.3 Electronic circuit board...14 6.4 Optical sensor (Only for Model T17CA4200)...14 7 CE DECLARATION... 15 FIGURES 1 CHLORALERT Internal View... 4 2 Outline Dimensions... 8 3 Typical Installation...10 4 Electrical Wiring...11-5 - 325.6030.7

1 INTRODUCTION 1.1 General Description The De Nora Water Technologies chlorine gas leak detector, CHLORALERT T17CA4000, continually monitors air samples for the presence of chlorine. It is usually installed in chlorine usage and storage areas to protect personnel and equipment. A small internal air blower continually draws a sample of air into the Chloralert. The flowmeter and valve enable a fixed amount of air to be passed over the surface of the special electrolyte, where the chemical reaction takes place in presence of chlorine. Two platinum electrodes, half immersed in the electrolyte, pick up the electrochemical current thus generated which is measured by the electronic circuit and used to trigger an alarm system. The trip point may be factory set to 1 or 3 ppm chlorine in air by volume. The power on/alarm lamp commences flashing in the alarm condition. In the Model T17CA4100, a relay is also energized to enable external alarms, extractor fans etc. to be activated. The alarm condition is generated, in addition to the presence of chlorine in air, also by power failures, when the power is returned to the system. In the Model T17CA4200, a relay is also de-energized to enable external alarms, extractor fans etc. to be activated. The electrical circuits are designed such that power supply failure causes the alarm relay to go into the alarm state. A flow sensor detects when there is no sample flow and causes the alarm relay to go into the alarm state. The sample fan is stopped when the alarm condition is actuated, to prevent saturation of electrolyte with chlorine and contaminated air to be discharged in safe area through the Chloralert itself. The unit has an integral electrical test facility. The alarm needs operator's acknowledgment to be reset. Figure 1 - CHLORALERT Internal View 325.6030.7-6 -

2 TECHNICAL SPECIFICATIONS Power supply: 120 Vac, 60 Hz 220 Vac, 50 Hz 110 Vac, 50 Hz Detection level: 1 ppm (3 mg/m 3 ), or 3 ppm (9 mg/m 3 ) Relay contacts: DPDT, 10 A. 240 Vac, resistive load or 28 Vdc Cell response time: Instantaneous Warm up time: Max 3 minutes Sampling rate: 0.05 m3/min Ambient temperature: -20 C (-4 F) to +65 C (149 F) Sample inlet connection: 1" NPT Weight: approx. 2.8 kg Mounting: Wall Dimensions: 215 H x 162 W x 235 D (mm) Material of construction: Housing: ABS Cell: ABS Electrodes: Platinum - 7-325.6030.7

CHLORALERT WHEN FLASHING S E V E T R E N T SERVICES Figure 2 - Mounting and Outline Dimensions in inches (mm) A L A R M C C T E S T R N R E S E T 325.6030.7-8 -

3 MODEL NUMBER BREAKDOWN T17CA4 B Basic CHLORALERT Alarm Design 1 - Standard 2 - Fail Safe Electrical Requirement 10-120 Vac, 60 Hz 20-220 Vac, 50 Hz 30-110 Vac, 50 Hz Design Level Alarm Level 3-1 ppm, Fixed 4-3 ppm, Fixed 3.1 Standard accessories 1 kit P/N 614S071U0l including: 1 electrolyte bottle 1 spare float 1 set of spare O-Ring 1 insect screen - 9-325.6030.7

A L A R M WHEN FLASHING S E V E R N C C T R E N T SERVICES T E S T R E S E T 4 INSTALLATION 4.1 Location The Chloralert should ideally be mounted outside the area to be monitored. However it can be mounted in the contaminated area if it is considered absolutely necessary. The location should be such that the ambient temperature never falls below -20 C (-4 F) nor rises above 65 C (149 F). It should never be mounted in direct sunlight and should be protected from the weather if mounted outside. 4.2 Mounting Securely mount the Chloralert at a convenient working height (approx. 1,3-1.5 m from the floor). 4.3 Sampling line As the Chlorine weight is higher then the air, we recommend that the sample inlet point is located at about 200 mm (8") from the floor. Use of frigid or flexible 1 PVC tubing is recommended. It is recommended to install the proper insect screen (supplied) at the terminal point of the sample line to avoid dust suction. WARNING NEVER USE TUBES WITH A DIAMETER LOWER THAN 1" WHEN THE SAMPLE PIPE LENGTH EXCEEDS 4 M CHLORALERT Figure 3 - Typical Installation 325.6030.7-10 -

4.4 Electrical wiring Cable gland on the Chloralert will accept a cable of diameter between 5 to 10 mm. Refer to the following drawings to connect the power supply and the external signals to the Chloralert. Figure 4 - Electrical Wiring - 11-325.6030.7

5 START-UP, OPERATION AND MAINTENANCE 5.1 Electrolyte filling at start up Refer to figure 5. Switch off power supply or unplug instrument. Remove cover rising it. Disconnect cell cable from terminal block noting carefully position of each cable. Disconnect exhaust tubing from the cell. Unscrew the four retaining screws and drop bottom half of cell from the upper for filling. The electrolyte bottle contains a dropper. Fill dropper stem to mark 1ml- Carefully transfer electrolyte from dropper to cell, taking care not to allow it to touch the top surface of the cell. Remove any electrolyte on this surface with a tissue then reassemble cell together taking care not to tilt the bottom half too much. Reassemble exhaust tubing, reconnect the cell cable, replace lid and finally reconnect supply cable and switch on. Hold RESET switch down until alarm lamp ceases to flash then wait further 5-10 sec. before releasing switch (this operation can take up to 3 minutes.) The unit is now in operation. 5.2 Electrolyte renewal after operation Approximately once a year or as necessary the electrolyte should be renewed. Proceed as for start up but wash the electrolyte thoroughly from the cell with water and gently clean the electrode faces with a non abrasive chlorine-free cleaner. Again thoroughly wash cell with water, dry with a clean tissue and refill with electrolyte as described under Section 5.1. 5.3 Sample flow rate At start up set the sample flow so that the flow indicator sits between the two graduation marks on the flowmeter tube. The flow may be adjusted by inserting a broad bladed screwdriver in the rate valve and gently twisting until the float is properly positioned. This should be done with the air control knob in the wide open position. If the required cell flow cannot be obtained, the air control knob should be gradually closed, until the adjustment of the float position is possible. The position of the float in the flowmeter tube should be checked from day to day (during the first week) and adjusted if necessary. 5.4 Gas test Verify the instrument operation in presence of chlorine gas after the start up. The test suggested hereafter is only qualitative and can be performed using solutions commonly available such as bleach and vinegar. Mix 6 parts of bleach and one part of vinegar in a beaker. Place the beaker under the sampling fan. The alarm LED should start flashing and any remote device connected to the alarm relay should be actuated. The alarm is reset by pressing RESET switch till the alarm LED stops flashing (this operation may require up to 3 minutes). 325.6030.7-12 -

5.5 Circuit performance check Simultaneously press and hold down both TEST and RESET switches. The alarm LED must change from the steady condition to flashing and the sampling fan stays activated. This checks that the electronic circuits are operating properly. If it is desired to check external circuits the RESET switch must be released: any remote device connected to the relay will be actuated. If it is required that external devices are not actuated during this test, release the TEST switch first wait for the LED to assume a steady state, (this operation may require up to 3 minutes), then release the RESET switch. - 13-325.6030.7

6 MAINTENANCE PROCEDURES Note: disconnect power supply before removing the instrument cover. 6.1 Measuring cell maintenance Refer to Section 5.1 and 5.2 of this manual. 6.2 Flowmeter 6.2.1 For Model T17CA4100 Clean tube and float whenever dirt has accumulated causing malfunction. Using a screwdriver loosen and slide the top tube retainer ring towards the enter of the tube. Remove tube by pulling forward. WARNING TAKE CARE NOT TO LOSE THE O-RING AND THE FLOATS! Clean all parts in mild detergent, rinse and wipe dry using clean tissue paper. To reassemble, place the tube retainer ring over the tube, replace the O-ring, drop the float in the tube and re-position tube between inlet and outlet connections. Gently slide the tube retainer ring back into position, making sure it is properly located by carefully rotating (red mark on the tube should be on the front side). 6.2.2 For Model T17CA4200 Clean tube and float whenever dirt has accumulated causing malfunction. Remove the plastic flow tube retainer by unclipping from the flow tube. Using a screwdriver loosen and slide the top tube retainer ring towards the enter of the tube, also lift the optical sensor housing towards the center of the tube. Remove tube by pulling forward. WARNING TAKE CARE NOT TO LOSE THE O-RING AND THE FLOATS! Clean all parts in mild detergent, rinse and wipe dry using clean tissue paper. To reassemble, place the tube retainer ring and optical sensor over the tube, replace the O-ring, drop the float in the tube and re-position tube between inlet and outlet connections. Gently slide the tube retainer ring and sensor housing back into position, making sure they are properly located by carefully rotating (red mark on the tube should be on the front side). 6.3 Electronic circuit board It is recommended that in the event of failure of the electronic board a replacement is fitted and the faulty board is returned to factory for repair. 6.4 Optical sensor (Only for Mod. T17CA4200) The flow sensor is housed in a heat shrunk sleeve. For this reason it is recommended that the faulty unit is replaced with a spare item. 325.6030.7-14 -

7 CE DECLARATION Chlorine Gas Leak Detector Model T17CA4000 is in conformity with the following standards: EN 56681-2 (Generic emission standard, industrial environment) EN 56682-2 (Generic emission standard, industrial environment) Following the provisions of directives: Electromagnetic compatibility 89/336/CEE, modified by directive 92/31/CEE". - 15-325.6030.7

Design improvements may be made without notice. Represented by: De Nora Water Technologies 3000 Advance Lane Colmar, PA 18915 ph +1 215 997 4000 fax +1 215 997 4062 web: www.denora.com mail: info.dnwt@denora.com DEC 2018 325.6030.7-16 - Registered Trademark. 2018. All Rights Reserved.