Neonatal Ventilator NV8
Product Dimension Dimension Weight Caster Caster Display Screen Type Size Resolution Brightness LED Indicator Audio Indication Interface Power interface 1300mm x 440mm x 520mm 26kg (excluding the air compressor) 4 inch TFT LCD color touch screen 8 inch 800 600 (pixel) Adjustable AC power indicator Battery indicator Work indicator Alarm indicator Speaker Buzzer One AC power interface One 12V 24V DC power interface Equipotential terminal One equipotential grounding terminal One RJ45 network port Port One serial port One USB port CPU Type Dual CPU Patient Type Neonate to Pediactric Ventilator Specification Modes of ventilation(standard) NCPAP NIPPV SNIPPV HFNC Modes of ventilation(optional) Environmental Specifications Main Unit Item Temperature(ºC) Relative Humidity Atmospheric (non condensing) Pressure (kpa) Working 5~40 0%~95% 70~106 Transport and storage 20 ~60 (oxygen battery: 20 ~50 ) 0%~95% 50~106 Performance Specifications Gas Circuit
Gas Supply Gas Gas connection Gas inlet pressure Gas Supply Fresh Gas Display range and accuracy Gas Circuit System Interface and Joint Oxygen, air NIST 280~600kPa Range of air: Range of oxygen: Accuracy: 0~20L/Min 0~20L/Min ±5% of the displayed value Insp Fresh Gas Port YY1040.1 2003 15mm inner cone joint Pressure sampling port Luer taper System Principle and Parameter Specifications Parameter Specifications Adjustable Range of Respirator Parameters Parameter Adjustable Range Step Size Work Mode CPAP 1~13 cmh2o 1~3 cmh2o: 0.2cmH2O NCPAP Peep 1~13 cmh2o 3~13 cmh2o: 0.5cmH2O NIPPV, SNIPPV Pinsp 3~15 cmh2o 3~8cmH2O: 0.5cmH2O NIPPV, SNIPPV Papnea 3~15 cmh2o 8~15cmH2O: 1cmH2O NCPAP Manual vent 3~15 cmh2o 8~15cmH2O: 1cmH2O NCPAP 0.5~20L/Min 0.5~2L/Min: 0.1L/Min HFNC 0.5~20L/Min 2~10L/Min: 0.5L/Min HFNC 0.5~20L/Min 10~20L/Min: 1L/Min HFNC Manual vent 3~25L/Min 3~10L/Min: 0.5L/Min 3~25L/Min 10~25L/Min: 1L/Min O2% 21~100% 1% NCPAP, NIPPV Flush O2 23~100% 1% Ti 0.1~15s 0.01s SNIPPV Rate 1~120bpm 1 NIPPV Tapnea OFF, 10~30s 5 NCPAP SNIPPV Rb 1~120 bpm 1 SNIPPV Manual ventilation time 1~15s 1 Flush oxygen ventilation 30~120s 30 time Respirator Performance Drive pressure 280~600kPa Fresh gas flow 40L/Min or higher (gas supply pressure: 280KPa) Parameters Monitored by Respirator O2% 0~100% CPAP / PEEP
Ppeak Pmean I:E 1:1~1:10 Te 0.4~30s Rsp 0~200 bpm 0~40L/min Parameter Accuracy Control Parameters CPAP: ±0.2cmH2O or ±2% of the set value, whichever is greater. Pinsp: ±0.2cmH2O or ±2% of the set value, whichever is greater. Pressure PEEP: ±0.2cmH2O or ±2% of the set value, whichever is greater. O2% ±3% of the set value Manual vent: ±0.2cmH2O or ±2% of the set value, whichever is greater. Papnea: ±0.2cmH2O or ±2% of the set value, whichever is greater. Flush O2 Rate Rb Ti ±3% of the set value ±0.5 bpm or ±1% of the set value ±0.5 bpm or ±1% of the set value set value ±0.005s : ±0.2L/Min or ±5% of the set value, whichever is greater. Manual ventilation: ±0.2L/Min or ±5% of the set value, whichever is greater. Tapnea Error: ±0.1s Measuring Parameters O2% ±2% of the actual reading CPAP / PEEP ±0.2cmH2O or ±2% of the actual reading, whichever is greater. Ppeak Pmean ±1 respiration/min or ±5% of the Rsp actual reading, whichever is greater. ±0.2L/Min or ±5% of the set value, whichever is greater. Alarm Settings Parameter Adjustable Range Remark Pressure Upper alarm limit 2.5~20cmH2O The upper alarm limit should exceed the lower alarm limit. Lower alarm limit 0.5~(upper alarm limit 2)cmH2O Oxygen concentration Upper alarm limit 20~105% The upper alarm limit should exceed the lower alarm limit. Lower alarm limit 18~(upper alarm limit 2)% Alarm mute 120s
Alarm delay 1~10s (step size: 1s) Specifications of Oxygen Sensor Oxygen Sensor Output Oxygen output: 9 13mV (at 210hPa) Range and accuracy 0~100%, ±2% 100% oxygen signal deviation 100±1% Resolution 1hPa O2 Expected service life 1.5x10 6 % measurement hours (at 20 C) 0.8x10 6 % measurement hours (at 40 C) Response time (21% air 100% oxygen) <15s Linearity Linear 0 100% O2 Working temperature 20 C~+50 C Temperature compensation 0 40 C (±2% fluctuation) Atmospheric pressure 50~200KPa Relative humidity 0~99% 100% oxygen concentration output drift Typical value beyond one year: <5% Material White ABS Package Sealed package Validity 13 months (or shorter if any of our requirements is violated) as from the package removing date Alarm Specifications Alarm Alarm sound pressure 45dB 85dB Maximum sound pressure for high alarm 82dB Maximum sound pressure for medium alarm 80dB Maximum sound pressure for low alarm 79dB