INSTALLATION INSTRUCTIONS DÜRR MOTORSEPA S 300

Similar documents
GB Installation instructions DÜRR Combi-Separator 7110

DÜRR SUCTION SYSTEMS. High demands reliably exceeded DÜRR AMALGAM SEPARATORS

Top suction machines DÜRR VS SUCTION SYSTEMS FOR THE DENTAL PRACTICE. VS Combination Suction Units VS 300 S VSA 300 S VS 600 VS 900

VS suction systems for the dental practice Top suction machines. VS Combination Suction Units

ECO MST 1. Separierautomatik MULTI SYSTEM TYP 1 ECO Gerätedokument Einbau, Betrieb und Wartung

METASYS. Central Suction System EXCOM hybrid 2/5. Equipment Logbook Assembly, operation and maintenance

CENTRAL SUCTION SYSTEMS. Dry / semi-wet suction For 1-15 dental units Underpressure of 180 mbar

We ve invented suction. Again. The radial Tyscor V/VS suction machine COMPRESSED AIR SUCTION IMAGING DENTAL CARE 50 % HYGIENE. potential energy saving

Amalgam separator CA 4

Diaphragm pump 4 - headed

Cleaning unit for coolant. :_decftez`_>r_fr] Book No.: V2

Lyovapor L-300 Technical data sheet

hinge right B 1 85 cm + 6,5-7 cm max. height of the device feet B cm B 3 85 cm D 2 75 cm (with a maximum opening angle of 170 )

INTELLIGENT SUCTION. Upgrade-AtoB

Suction Pump F-30.xx

HEAT RECOVERY AIR HANDLING UNIT

The heart of the practice Dürr Dental compressors

Illustration shows item no Z Original operating manual. Operating manual. BlueMobil 60. Item no: Z-3285

CE Unit is suitable for material transport. The transported gas may be up to 80 C.

Preserving the value of your suction unit. Systematic hygiene from Dürr Dental

Installation manual W365H, W375H, W3105H, W3130H, W3180H, W3240H, W3300H Exacta & Clarus Control

AS 50 operating manual, 2013

PRESSURE WASHER MODEL NO: JETSTAR 1750 OPERATION & MAINTENANCE INSTRUCTIONS. WARNING Read the instructions before using the machine PART NO:

ORIGINAL INSTRUCTIONS

Operating Instructions and Spare Parts List. ZR Monocyclone. Issued 09/01. ZR Monocyclone 9

K Operating Instructions. Before first use of the unit read these operating instructions and act in accordance with them.

METASYS. Amalgam Separator COMPACT A8 / A16. Installation, operation and maintenance

Diaphragm pumps. Operation Manual. Models: 2019C-02 (230V / 50Hz) 2019B-01 (115V / 60Hz) 2019C-50 (230V / 50Hz)

Operation Manual. Trash Water Pump ARTICLE Image similar

Hako. The power package for inviting cleanliness: Hako-Citymaster 1800TDI EURO 3. Superior technology for outdoor cleaning.

Illustration shows item no Z Original operating manual. Operating manual. BlueMobil eco. Item no: Z-3278

OK, OKA, OKAF; ELD; ELDM; ELH; OKC; SC; SCA; SCAF; OK-LN; OKA-LN; OKAF-LN.

**BVC professional G

Operation Manual. Diaphragm pumps 2-headed. Types MP 101 Zp, MP 201 Ep MPC 101 Zp, MPC 201 Ep

Monitoring Devices 11/1

Specifications. Vacuum motor power consumption(w/hp) 1200 / 1.6. Exhaust water pump power consumption(w/hp 800 / 1.1

Installation, Operating and Servicing Instructions

AirStar Clinical Air Compressor AS60, AS90 and AS120. Installation & Operating Instructions

Cabinet Refrigerators/Freezers Instruction manual

Technical Ventilation

MEDIST s.r.o., Petrusovskeho 4, Humenne, Slovakia Tel.: , Fax: ,

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

Operating Instruction Centrifugal fans without scroll (Translation of the Original) BA-RLE-AC /2014 RLE RLA

Manual. Above-counter dispenser for hot drinks OTHG /12

**BVC control G. * With 4l PP collection bottle for general aspiration tasks, completely autoclavable together with a 0.2 µm protection filter

O P E R A T I N G M A N U A L

Description and Operating Instructions. ACCUVAC Basic. Aspirator WM WM 10709

NT 65/2 Tact² Tc. Tilting chassis with container lock. Suction tubes and floor nozzles can be stored in any direction in the accessory holders

HOT WASHER MODEL NO: KING150

Operation Manual. Diaphragm pumps. MP 054 Zp, MPC 054 Zp MP 104 Ep, MPC 104 Ep

Installation plan Commercial Washing Machine PW 413 PW 418

NT 75/2 Ap Me Tc. On-board storage. ApClean (semi-automatic filter cleaning system) Integrated drain hose. Hose and bend attachment

NT 75/2 Ap Me Tc. On-board storage. ApClean (semi-automatic filter cleaning system) Integrated drain hose. Hose and bend attachment

Instruction manual ASPIMAX 16

REGISTER YOUR PRODUCT. for warranty claims INSTALLATION PROOF. Please complete and return

Operating instructions

HYDROCYCLONE ISO 18 INDEX. Page

INSTALLATION, OPERATION & MAINTENANCE MANUAL

METASYS. Amalgam separator ECONOMY SYSTEM TYPE 2 ECO II. Installation, operation and maintenance

RONDA 2-58 Duo Heavy Duty Industrial Vacuum Cleaner For Liquid Filtration and Transport

IC-45AV Vertical Clean Centrifuge

INSTRUCTION MANUAL (English version) Water still

Instructions. Portable Absorb Phlegm. Apparatus

CRU-S Series Stainless Steel Condensate Recovery Unit

A m e r i c a s D e n t a l E q u i p m e n t C o m p a n y. Rev 2/15. D a n s e r e a u

Operating instructions. Web Cleaner 1 / 2. Air Line

VC 45 Surgical Suction Unit

NT 65/2 Eco Me. On-board storage. Integrated drain hose (oil-resistant)

WET & DRY VACUUM CLEANERS

Special Requirements For Health Care Facilities

STORAGE, ERECTION, OPERATION AND MAINTENANCE MANUAL FOR CENTRIFUGAL FANS MODEL - MXE

Read Only Document Not For Distribution

Test Unit GTL 100 for Fire Detectors

ASSEMBLY & OPERATING INSTRUCTIONS

MP 2401 E, MP 1801 Z, MP 1201 T

Operating Manual BDC 1216

Translation of the original manual. Vacuum Unit SVAGG. Assembly and Operating manual. Superior Clamping and Gripping

Katherm NE - Natural convection with electric heating element 2.45

Z-350 Fazer Machine User Manual English

RECESSED WARM CEILING HEATERS MODELS HSRCH3000RF, HSRCH4500RF INSTALLATION AND OPERATING MANUAL

INDUSTRIAL VACUUM CLEANER WITH AUTO START/STOP

GRANULE GASTRO EXTRA POWERFUL FOR KITCHENS USING A LARGE NUMBER OF GN TRAYS. * tabling not supplied by GRANULDISK

Installation manual. Tumble dryers T4130. Compass Control

Industrial Solid State Contactor TE10S. Static switching of resistive loads and shortwave infrared heating tubes up to 20 kw

Installation & Operators Manual. The Classeq Pass Through Range

ASPIRE Laboratory Aspirator

Installation and Operation Instructions Domestic hot water module FriwaMini with circulation - DN 15

Grohtherm Wireless! Installation Instructions Pumped

SMHA Hatch High temperature fan HT-fan Installation and Maintenance

Instructions for use. BA Optima. Air motors BA602 (BA640081) BA604 (BA640060)

Propane Flame Projector Safety instructions and operating manual

Service Manual - Series 450/451 IG IGV IG IGV IG IG IG IGV IG

INSTALLATION MANUAL. Split-type Air Conditioner (Cooling and Heating) Outdoor Unit UQB09JJWC UQB12JJWC. Indoor Unit AQB09JJWC AQB12JJWC

Electronically controlled instantaneous water heater. MCX: 27300, and models. Installation instructions

Carpet washer. Vax careline V-135B. Vax model number:

User Manual GV25 GV35 GV702. Company information: Original instructions GV12066 (1)

K 5.80 M ( ) Caution! Do not use the unit without first having read the operating instructions.

brands you trust. Pumps for hazardous materials Type GUP 3-1,5 & GP 20/10 Ex

isense & isense-ht 230V Flush Mounted Domestic Continuous Extract Fans

Silent TS. Nr / Ideas for dental technology / A. Made in Germany

Transcription:

INSTALLATION INSTRUCTIONS DÜRR MOTORSEPA S 300 9

10

INSTALLATION INSTRUCTIONS DÜRR MOTORSEPA S 300 CONTENTS 1. Advice... 12 1.1 Safety advice... 12 1.2 General advice... 12 1.3 Electrical safety check... 12 2. Delivery schedule... 13 3. Purpose... 13 4. Model survey... 14 5. Technical data... 14 5.1 Hose connections on the Motorsepa S 300... 14 6. Function... 15 7. Installation... 16 8. Pipe material... 16 9. Important advice for pipe-/hose laying... 16 10. Electrical connection... 17 11. Final procedures... 18 12. Maintenance... 18 Spare parts list... 20-21 11

1. ADVICE 1.1 Safety advice Before commencing installation PLEASE NOTE: The Dental Unit, to which the Dürr Motorsepa S 300 is to be connected to, must be switched off when installing and again checked when switching on after completion! Work on electrical equipment may only be carried out by qualified electrician's (per VBG4) or through our trained service technicians. With the combination of equipment, as well as the installation or repair of electrical parts, the relevant safety and testing regulations must be observed. (VDE 0100, 0105, 0107, 0750, 0751, IEC 364, 601-1, UVV- VBG 1,4,5,103 and MedGV, abroad the relevant regulations must be followed e.g. IEC,CEE). Medical product guideline 93/42/EWG German medical product law (MPG) With repairs only use original spare parts otherwise the quarantee and test certification will be invalidated. The DÜRR Motorsepa S 300 according to medical equipment regulations (MedGV) is a Group III apparatus, an energy driven medico-technical unit. removed, otherwise large particles (pieces of teeth, fillings etc.) can cause faults in the Motorsepa S 300. 1.3 Electrical safety check According to IEC 601-1, VDE 0751 an electrical safety check must be carried out after installation and commissioning, servicing, maintenance and repair of electrical equipment The test results must be documented and filed. 1.2 General advice Please note whether the drain of the Dental Unit has been laid correctly. Where deviations occur, necessary connection parts must be obtained from a sanitary specialist store or ordered from DÜRR DENTAL. The drain of the Motorsepa S 300 must be carried out according to the Installation Instructions. The Motorsepa S 300 must not be connected to the drain of hot water equipment. (Max. drainwater temp. 40 C) As on older Dental Units the drain hose can be very dirty or even brittle, the drain hose should be checked before the installation of the Motorsepa S 300 and if necessary changed for a new one. Check also whether the coarse filter e.g. in the spittoon is in position. The coarse filters must not be 12

2. DELIVERY SCHEDULE Motorsepa S 300 230 V~, 50-60 Hz (w/o transf.)... 7102-01 230 V~, 50-60 Hz (with transf.)... 7102-11 3. PURPOSE The Motorsepa S 300 is designed for one treatment place. The motor driven automatic separator S 300 has the task to separate secretion from the airstream. The advantage of the S 300 in comparison with ordinary separator systems is that the separation stage is driven b y a motor and thus ensures, irrespective of volume flow, constant separation of the air-water-secretion mixture. The S 300 separates the aspirated fluids and solids from the suction air via a two stage separation system with cyclonic separation and separation turbine. Without interruption of the suction performance the aspirated liquid is, via an integrated pump, safely sluiced through and led towards the drain. The air is fed back to the suction machine and then as exhaust air led outside. Even with a high liquid yield the suction is not interrupted - one can continue working continuously. The S 300 stands on rubber buffers, which reduce vibration and running noise, so that the noise level is reduced to a minimum. The Motorsepa is expecially suitable for prophylaxis workplaces, e.g. in the 12 o'clock position. The apparatus does not replace the suction machine and cannot separate amalgam particles. The S 300 offers the following advantages: 2 stage separation system with cyclonic separation and turbine separator, through which there is high safety against foam and liquid sweeping back into the suction machine. No interruption of the suction flow through overfilling or bloodfoam Little maintenance Simple installation 13

4. MODEL SURVEY Motorsepa S 300 Model 7102-01 (230 V, 50-60 Hz) w/o transf. Model 7102-11 (230 V, 50-60 Hz) with transf. 5. TECHNICAL DATA Model 7102-01 / 7102-11 Current (A) 0,7 Start Current (A) 2,1 Voltage (V) 230 Elect. output (kw) 0,14-0,17 Frequency (Hz) 50-60 Speed (rpm) 2840-3360 Duty rating (%DR) 100 Max. flowrate fluid (l/min) 4 Air flowrate (l/min) 300 Max. permissible neg. pressure (mbar) 180 Size (HxWxD, cm) 29 x 23 x 27 Weight (kg) 4,8 5.1 Hose connection A Suction-Connection Suction nozzle Ø 32 mm B C Vacuum-connection Vacuum nozzle Ø 36 mm Drain-Connection Drain nozzle Ø 22 mm 14

D I B A E C H G F 6. FUNCTION A Suction-connection B Vacuum-connection C Drain-connection D Place selection valve Everytime the suction hose is removed from the holder, the suction machine and the Motorsepa S 300 is switched on. Thus the air-water secretion mixture can be aspirated from the patient's mouth and via the secretion entrance A led towards the Motorsepa. The S 300 separates the aspirated liquid and solids from the suction air via a two-stage separation system with cyclonic separator and separation turbine. During the suction procedure the aspirated liquid is accelerated in the secrete entrance A and in the cyclone separator H turned into a spiral motion. The E Drain valve F Waste water pump G Separation turbine H Cyclone separator I Air channel created centrifugal forces fling the aspirated matter onto the outer wall. In this 1. stage a "coarse separation" of the liquid occurs in the cyclone separator H. The centrifuged liquid is pumped into the main drain via the drain valve E and drain connection C. In the 2. stage a "fine separation" of the remaining liquid matter, such as blood foam, occurs through the centrifugal forces created in the separation turbine G. The air then flows via the place selection valve D into the vacuum line to the suction machine. 15

A B C Suction-connection Suction nozzle Ø 32 mm Vacuum-connection Vacuum nozzle Ø 36 mm Drain-connection Drain nozzle Ø 22 mm 7. INSTALLATION 7.1 Drain check Prior to the installation the drain must be checked: At least 1 litre/20 secs. With a clear drain, a backflow can be prevented. 7.2 Installation room The room temperature must not sink below 5 C in winter and not rise above 40 C in summer. 7.3 Installation site We recommend that the Motorsepa be installed in a well ventilated cabinet. Please ensure good accessability of the hose connections! With easy access the safety sieve (pos. 11, p.21) can easily be cleaned every 4 weeks. 8. PIPE MATERIAL Only the following pipe materials should be used: HT-Drainpipes made of polypropylene (PP), DIN V 19 560/09.92 (marked red, low flammable) or Geberit-pipes, Type PEh. Not to be used.: HT-Stabil pipes, DIN V 19 561/09.92 (marked yellow, normal flammable). These pipes are made of ABS/ASA-material and therefore not resistant against some of the medicines and solutions used in the Dental surgery. 9. IMPORTANT ADVICE FOR LAYING OF PIPES-/HOSES Laying drain pipelines must be carried out according to DIN 1986, pt.1 and 2 or according to the relevant regulations of the country concerned. The connection between the pipeline and the Motorsepa S 300 connection must be made with a flexible hose, at least 0.5 m long. Through this, vibration of the pipe system is prevented. Hoses must be laid stress free! 16

10. ELECTRICAL CONNECTION If the connection to the mains supply is made via the floor-or ceiling installation, an all-polar isolating device (an all-pole switch or all-pole line protection switch - fuse- with >3 mm contact opening width) must be built into the circuit. On the floor plate an earthing connection is present, which during an electrical malfunction prevents possible current flow across the housing endangering life and limb! Circuit fuse: max. 10 A Details for the 230 V cable: H05VV-F 3G 0,75 (3-pole, 0,75²) 5 4 3 2 10.1 Control via control panel motor terminal box of S 300 Pict. A: w/o transformer, Model 7102-01 Pict. B: with transformer, Model 7102-11 1a Control box 1b Manifold 2 Motor terminal box S 300 3 Control panel in motor terminal box 4 Rectifier 5 Relay 6 Fuse1A inert as per IEC 127-2/V 7 Condenser 8 Place selection valve 9 Motor S 300 10 Transformer 230V, 12V, 1,5VA 1a Pict. A: w/o transformer Model 7102-01 6 9 7 8 5 3 2 1b Pict. B: with transformer Model 7102-11 4 10 6 9 7 8 17

11. FINAL PROCEDURES Switch on Unit or Surgery main switch. Carry out electrical safety check as per VDE 0751. Check machine for function and connections for possible leakage. Fill-in guarantee card. Instruct personnel. 12. MAINTENANCE Every week change fine filter in the manifold. Every 4 weeks clean the protective sieve 11 at the secretion entrance. 11 18

19