VERIFICATION OF IONISING RADIATION MONITORS FOR PRACTICAL USE IN HOSPITALS. C. Montes, F. Sáez, C. Martín, P. Collado, C. J. Sanz, E.

Similar documents
RAM GENE-1 ERK. Contamination & Radiation Meter. Operating Manual. Document #PRIR90N2.DOC Version 2.2 October 2009

PORTABLE ISOTOPE IDENTIFIER Search Tool / Sample Counting System

Using the GammaRAE II Responder Personal Radiation Detector and Dosimeter in One. Firmware Version 3.00 D August 2006

Co-60 irradiator for therapy level calibrations at SSDLs

QUALITY ASSURANCE AND SAFETY AT A TID RADIATION TEST LABORATORY

Quality Assurance of Radiation Protection Monitoring Instruments in India

Fundamentals is subdivided into Sources of Radiation, Biological Effects, Mathematics, Chemistry, Physics, and Units and Terminology.

Battery-operated, hand-held alarm and measuring instruments for radiation protection

Nuclear Associates

Monitoring of Radiation Exposure Adjacent to a Linear Accelerator Treatment Bunker with a Highlight Window

A FIRST RESPONDERS GUIDE TO PURCHASING RADIATION PAGERS

HEALTH AND SAFETY PROGRAM 406 IONIZING/NONIONIZING RADIATION PROTECTION

Telephone: +44 (0)

RAM GAM-1C Gamma Survey Meter TYPE TEST

Environmental Protection Agency, Office of Radiation Protection and Environmental Monitoring.

RAM GENE / RAM GENE-1

Dosimeters/Survey meters

Personal Radiation Alarm Dosimeter LK Operating Instruction Manual. Contents

NUCLEAR INDUSTRY STANDARD PROCESS Radiological Protection. Level 3 Information Use

Ionisation Chambers Containing Boron as Neutron Detectors in Mixed Radiation Fields

Using the GammaRAE II Personal Radiation Detector

Department of State Health Services

Radioactivity dosimeter RD01. User's Manual

DIGITAL AREA RADIATION MONITOR

University of Maryland Baltimore. Radiation Safety Procedure

Radiation Safety. Product Catalog 2011/2012. Better products. More choices. One company. Fluke Biomedical.

Consultation Draft. Issue 2. Good Practice Guide No. 29. The Examination, Testing and Calibration of Installed Radiation Protection Instruments

California State University Bakersfield Radiation Safety Program Revised November 2017

REPUBLIC OF NAMIBIA MINISTRY OF HEALTH AND SOCIAL SERVICES REPORT BY: MS VERA UUSHONA TITLE: INSPECTION OF CONVENTIONAL DIAGNOSTIC X-RAY FACILITIES

RAD-60 USER GUIDE (ALSO FOR THE RAD-50 MODEL AND ISOTRAK DOSIMETERS)

APPLICATION FOR AUTHORISATION - PART 1

COUNTING ROOM EQUIPMENT RCT STUDY GUIDE

Analytical X-Ray Safety Program

Portable Survey Instruments NISP-RP-01

stationary dose rate measuring systems

Area Monitor/Frisker Count Rate Meter Victoreen

Scientific Production Company Doza. Radiation monitoring equipment

HEALTH AND SAFETY MANUAL

Laboratory Impex Systems Ltd

Lead-Lined Cabinetry. 7 Vreeland Road, Florham Park, NJ or

Lead-Lined Cabinetry. All Cabinets Feature: Pre-Configured Cabinet Models. Specialized Cabinet Models

Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N

Forward-scattered radiation from the compression paddle should be considered when average (or mean) glandular dose is estimated

TECHNICAL LEAFLET. The proposed instrument is an extended version of the computerized Geiger-Muller indicator PT

Passive Whole Body Monitoring Nuclear Power Industry Application Introduction to the Argos -5AB Zeus (Gamma Option) and the GEM

Dual-Diode Dosimeter Patient Dose Monitor

Application Note. Application Note for BAYEX

Identify the following features and specifications for the ESP1

Radioactive Contamination Monitor

CONTAMINATION MONITORING INSTRUMENTATION RCT STUDY GUIDE LEARNING OBJECTIVES:

Radiation Emitting Devices Research X-ray Safety Program

Advanced Radiation Detection Equipment

DMC 2000 Personal Electronic Dosimeter

MODEL 23 AND 23-1 ELECTRONIC PERSONAL DOSIMETER. August 2017

User s Manual. Electronic Personal Dosimeter (For Gamma(X)-ray) NRF30. Fuji Electric Co., Ltd. TN514629i 1/31

Setting the new standard for personal radiation detection.

Radiation Safety. Product Catalog 2007/2008. Fluke Biomedical. The accuracy you need from a company you can trust.

REGISTRY OF RADIOACTIVE SEALED SOURCES AND DEVICES SAFETY EVALUATION OF DEVICE (Amends In Its Entirety)

Radon Detection Specialists, Inc.

Electronic Personal Dosemeter

Use of control and analytical X-ray apparatus

Real-Time Remote Monitoring System Utilizing New Electronic Personal Dosimeter

GEOPHYSICAL ENVIRONMENTAL ANALYSIS EQUIPMENT AND SERVICES

INTERNATIONAL INTERCOMPARISON EXERCISE ON NATURAL RADIATION MEASUREMENTS UNDER FIELD CONDITIONS (IFC11) *

Competency: Critically evaluate the function of the ionisation chamber in the Linear Accelerator and its importance for correct treatment delivery

Radiation Alert Frisker INTERNATIONAL

Comprehensive applications of the gas flow proportional counters for Radiological Surveillance

User Manual. Advanced Markus Chamber Ionization Chamber Type 34045

Pico Envirotec Inc., CA & ENVINET a.s., CZ NEW APPROACH TO GLOBAL MONITORING OF THE RADIOLOGICAL SITUATION; INDUSTRIAL, ENVIRONMENTAL & EMERGENCY

Kata products Kata DGM-1500 Turva

Protecting our Borders while Ensuring Radiation Safety

Instrument Standards Used in Security Applications Presented by: Leticia Pibida, Ph.D.

This is a preview - click here to buy the full publication TECHNICAL REPORT. Nuclear medicine instrumentation Routine tests

Frequently Asked Questions

CALIBRATION OF A GAMMAMED 12i 192 Ir HIGH DOSE RATE SOURCE

Radiation Monitoring Systems

NEX-BETA-ABG FACILITY Drinking Water Safety Monitor

TRACERCO T402 and T402HR USER MANUAL

product catalogue

Victoreen & 660-8

PET Safety Manual. Mary K. W. Susselman, BS, CNMT, RT (N)(MR)

PM1703MO-1A PM1703MO-1B PERSONAL COMBINED RADIATION DETECTOR/DOSIMETER. Operation Manual

David F. Metzger 121 N. Leavitt Rd. #313, Amherst, OH Certified Radon Tester Ohio License #: RS340 & RC213; NRPP ID#: RT

LAUNDRY MONITORING SYSTEM Model # LIM-64S

DMC 3000 User s Guide

A Continuous Air Monitoring Sampler for 125 I - RIS-125

American Nuclear Society (ANS)

A new passive integrating detector for measurement of individual radon exposure at working places.

RadEye B20 RadEye B20-ER Multi-Purpose Survey Meter

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines. for Safety Systems at Radiation Treatment Centres

Radiation Detection And Measurements By G F Knoll Solution Manual Book

CURRENT INDOOR RADON SITUATION IN LITHUANIA

Highlights from a recent test campaign in Finland with scalable neutron detectors

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

INSTRUCTIONS IN CASE OF A RADIATION HAZARD

RAMSES: THE LHC RADIATION MONITORING SYSTEM FOR THE ENVIRONMENT AND SAFETY

The Modular System for Radioactivity Measurements FH 40 G/GL FHT 6020

Dynamic Radon Detection Over Measurement

22 Radiological and Environmental Monitoring Systems Functions, Basic Configuration, and Interfaces

The Modular System for Radioactivity Measurements FH 40 G/GL FHT 6020

Transcription:

VERIFICATION OF IONISING RADIATION MONITORS FOR PRACTICAL USE IN HOSPITALS C. Montes, F. Sáez, C. Martín, P. Collado, C. J. Sanz, E. De Sena Servicio de Radiofísica y Protección Radiológica. Hospital Universitario de Salamanca. Pº San Vicente, 58-182. 37007. Salamanca (Spain). Phone: +34 923 291 180. E-mail: mfbordes@usal.es Abstract. Regulations for Radiation Protection require to guarantee adequate monitoring of radiation and contamination levels in controlled or supervised areas. Surfaces in working areas, clothing or skin must be regularly checked in connection with radioactive contamination by using a contamination monitor; on the other hand, radiation levels must be measured in a variety of situations, for example, behind shielding or in the vicinity of therapy patients before releasing, by using a survey meter, and the radiation exposition to the staff members must be estimated using personal radiation monitors. In this paper, control procedures to verify performance of monitoring instruments of radiation and contamination levels for practical use in hospitals in a proper way are described. Regarding survey monitors, evaluations of zero setting, alarm threshold, battery test and response to calibrated check sources of 137 Cs, 60 Co and 133 Ba, were made. We also made sources with radionuclides used in the hospital which have been calibrated in a controlled dose calibrator. These sources were used to check the response of personal dosemeters, both pen sized and electronic ones. In order to check contamination monitors, plane sources with known activities from radionuclides usually used in our hospital were used, and the calibration factors (s -1 per Bq cm 2 ) for such radionuclides and geometry were obtained. In this way, we profit from having calibrated instruments for radionuclides used in the hospital workplace to determine whether working areas, protective clothing and the body surface have been contaminated. Measurements with calibrated plane sources ( 14 C and 36 Cl) were also made. A comparison of these values with the calibration factors of the instruments supplied by CIEMAT, the Spanish Secondary Standard Laboratory, was made. These procedures do not intend to replace calibration procedures at secondary standard laboratories, but they are complementary ones to check the response to radionuclides used in the hospital workplace as well as to have available measurements for constancy check to make sure detectors do not work wrongly. 1. Introduction Spanish Regulations for Radiation Protection [1] require to guarantee adequate monitoring of radiation and contamination levels in controlled and supervised areas. It is also required that monitoring instruments shall be calibrated and verified. In other countries there are regulations or recommendations in the same sense [2,3,4,5]. In this work, control procedures to verify performance of monitoring instruments of radiation and contamination levels for practical use in hospitals in a proper way are described. Furthermore, it is discussed if sources prepared by the user can be used. Finally, we have used this procedure in instruments in our hospital. 2. Kinds of detectors and tests 2.1. Kinds of detectors Radiation protection instruments usually are gas-filled detectors, although pen and electronic dosemeters can also be used. Main applications in the medical field are: - Survey instruments. Usually energy compensated GM detector type [6]. There are two types: portable and non-portable. n-portable are used for the evaluation of gamma dose rates in radiation therapy units and detect possible source return failure. Portable instruments are used to evaluate dose rates of therapy patients containing radiopharmaceuticals or permanent implants, to find lost gamma sources or for shielding assessment. 1

- Contamination monitors: Usually GM detector type. They are used for checking surfaces with respect to radioactive contamination in working areas as well as personal contamination after having worked in these areas. - Portable personal dosemeters: Usually pen or electronic dosemeters. They are direct-reading dosemeters, and are used for the evaluation of doses in special cases, such as pregnant workers or emergency situations, i.e. high dose rate source that do not return to the container. 2.2. Kinds of tests The reference manual should be read before first use, following the instructions regarding batteries, warm up period, etc. 2.2.1. Calibration tests Calibration must be done by accredited laboratories. It can be defined as the production of a correction factor to allow a better measurement of some quantity. Calibration certificate must contain information about the method followed in the calibration, what kind of sources were used and an estimation of the uncertainty associated with the calibration. These sources should be traceable to primary standards. 2.2.2. Verification tests It can be defined as a functional test to be sure detectors do not work wrongly. It could be done by the user in his workplace and it does not need any specific material. Verification tests should include, as a minimum: - Testing its response to a check source with well-known physical characteristics. This source of radiation should give a quality of radiation similar to that ones that are measured in the daily work. - Checking alarm setting. Some radiation detectors have a source attached for checking purposes. The reading with this source only will be taken as a constancy check with previous readings. Usually, sources used for the calibration of a radiation protection instrument are different from those we want to evaluate in the medical field. So we can profit from verification tests to check response of the instrument to sources used in the hospital workplace. 2.2.3. General maintenance Before each use, the following inspections and tests should be performed: - The instrument should be free of radioactive contamination. - Inspect the instrument for physical damage and repair minor damage if necessary. - Check the batteries or the power supply to verify proper operational voltages. - Avoid extreme temperatures or fast changes. - Turn on the instrument and allow it to warm up according to manufacturer specifications. - Zero setting should be adjusted in an area of known low background if necessary. 2

3. Verification of survey monitors 3.1. Description of the method Verification tests should include: - Looking for visible signs of damage. - Checking batteries, and change them if necessary. - Warming up before measurements. - Measuring background. - Checking response against sources. These sources are used in quality control for dose calibrators. We evaluated dose rate at 0,5 m and 1,0 m. Portable instruments were measured in a clear room to minimize scatter. - Testing alarm setting, visual and audible ones, bringing sources closer to the detectors. We also put the sources in touch with the instruments to verify performance under overload conditions. It also should be recorded last calibration date, resolution and measurement units and range. 3.2. Results 11 survey meters, 5 portables and 6 installed, were checked out. The non-portable ones are placed in 3 Teletherapy bunkers, one in a brachytherapy bunker and two in the Nuclear Medicine facilities. In Table I, there is a description of the equipments, their measurement units, range and if they have alarms and check sources. Visual inspection was correct in all of them. In one of the monitors, battery level was low, so the battery was changed before the measurements were made. Measurements of dose rate at 0.5 m and 1 m were made. For one monitor, the readings were oscillating. An inspection of the monitor was done and it was found that one electric junction was damaged. This monitor was repaired. After that it worked correctly, resulting in correct and stable readings. 3

Model/Type B.S Electrónica Zaragoza Table I. Survey meters. Placement Measurement units Radiotherapy mr/h Bunker LINAC c/s Range 0.01-50 mr/h 0.2-1000 c/s Alarm Check source B.S Electrónica Zaragoza MR870 Radiotherapy Bunker LINAC µsv/h, c/s 0.1-50 µsv/h 0.2-1000 c/s Minialarm MC-10/7-R Portable Sv/h 0.5 µsv/h 5 msv/h Nardeux Babyline 81 E-793 Portable Rad Rad/h 0.02 mrad/h 100 rad/h Nardeux Babyline 81-1836 Portable Gy Gy/h 0.2µGy/h-1Gy/h Radiametre ARA Duk 806 m.1435/6 Radiotherapy Bunker Co-60 mr/h N/A Texas Nuclear LOG Series Nuclear Medicine mr/h 0.05-200 mr/h Victoreen 451 P Portable Sv/h 0-50 msv/h Victoreen 813 m. 05/423 Nuclear Medicine mr/h 0.1 100 mr/h Victoreen PRIMALERT 35 m. 05/37 Brachytherapy Bunker high dose rate. N/A N/A Victoreen 290 SI Portable Sv/h 0-10 msv/h 4. Verification of contamination monitors 4.1. Description of the method As a minimum, a verification should contain the following steps: - Visual inspection of the monitor, looking for cracks, damages, etc. - Battery check. - Background measurement. If possible, integration modes should be used for a better accuracy. - In order to assess the response of the monitor, planar sources should be used. We have used two kinds of planar sources. The first ones are two certified area sources for calibrating contamination monitors with 14 C and 36 Cl. The second ones are constructed 99m Tc planar sources. The 4

readings with certified sources will be taken to assessment of performance over time and the readings with constructed sources will be used to evaluate real-life contaminations. The constructed sources used consisted of three PMMA sheets with 2 mm of thickness. In one of them has been done a 15 cm diameter hole, so if this sheet is put between the two others, we can get a quasi-planar cavity that can be filled in with solution. One must regard the following steps: - Get an activity of 99m Tc measured with the dose calibrator. - Fill in the phantom with the solution. - Calculate activity concentration in the phantom (Bq/cm 2 ) - Let the 99m T decay until the levels of surface contamination we want to measure are reached. Several days may be necessary, depending upon the starting activity, the volume of liquid and the phantom s active surface. - Scattering conditions should be reproducible. Certain care must be taken in account if we want to measure low levels of surface contaminations: because when an elution of 99m Tc is made with a typical generator, there is always a small amount of 99 Mo that passes into the elution. This is usually called Molybdenum breakthrough [7]. NRC limits for 99 Mo breakthrough is 10-4 Bq 99 Mo /Bq 99m Tc. When a generator is sent to a Nuclear Medicine Department, the level of 99 Mo Breakthrough must be measured. When the initial activity decays, since half-life of 99 Mo is longer (66 h) than that of 99m Tc, it may happen that Molybdenum s contribution to the reading of the surface monitor would be of the same order, or even bigger, than 99m Tc contribution. As an example, if an activity of about 5 mci (185 MBq) filled into a phantom of 177 cm 2, the surface contamination is about 1 MBq/cm 2. 96 hours later, surface contamination due to 99m Tc is about 16 MBq, while Molybdenum surface activity is about 38 MBq/cm 2. The variation of detector response with contamination area has not been considered. In other publications this effect may be up to factor of 4 [8]. 4.2. Results Four surface monitors were checked, three of them portables and one non-portable. Results are listed in Table II. Table II. Contamination monitors (obtained factors are expressed in s -1 per Bq/cm 2 ) Monitors Effective area Window thickness C-14 Cl-36 99mTc Berthold LB 123 12x19 cm 2 5 mg cm -2 (Titan) 4.0 41 1.1 Berthold LB 1210 100 cm 2 5 mg cm -2 (Titan) 7.9 49 0.88 Eberline ESP 15.5 cm 2 5 mg cm -2 mica 2.7 10 --- Victoreen 290 SI 4.5 cm 2 5 mg cm -2 (Steel) 1.2 4.3 0.015 5. Checking of personal dosemeters 5.1. Description of the method 35 pen dosemeters were checked out. A 99m Tc source of 22.0 mci was made, in a 20ml vial with 10 ml of elution. Pen dosemeters were put in a circumference of radius 40 cm, and the vial was put in its 5

centre. Dosemeters were irradiated for 21.5 hours. Readings of the pen dosemeters were taken. A source strength of 2.24 10-5 msv h -1 MBq m 2 was used for 99m Tc [9]. Three electronic dosemeters were checked out. A high dose rate brachytherapy unit, microselectron, was used. It has a source of 192 Ir, with a source strength of 2.32 cgyh -1 m 2. Dosemeters were irradiated for 107 seconds at 50 cm. Three irradiations were made. For each detector, an average of the three readouts was taken. Measurements were compared against a calibrated energy-compensated portable radiation detector at the same distance. 5.2. Results For pen dosemeters, the values of measurements were 15% below the real value of the dose. This underestimation may be due to the fact that there was any material just behind the dosemeters, while pen dosemeters are usually calibrated in a phantom that simulates a person, so there is an additional contribution of back-scattered radiation. For electronic detectors, there is a good agreement between the electronic ones and the ionization chamber, and the measurements are within tolerances. 6. Conclusion Several verification methods of different kinds of radiation detectors widely used in hospitals and health facilities have been described. The purpose of these protocols is not to replace calibration protocols, which must be done regularly, but to detect abnormalities in the performance of the equipment, as well as to keep it in working order for the isotopes that are usually managed. References 1. Real Decreto 783/2001, de 6 de Julio, por el que se aprueba el Reglamento sobre protección sanitaria contra radiaciones ionizantes. 2. U.S. Department of Energy, Instrument Calibration for Portable Survey Instruments, Implementation guideg10 CFR 835/E1 rev1 (1994). 3. National Physical Laboratory, The Examination, Testing and Calibration of Installed Radiation Protection Instrument, Measurement good practice guide nº 29 (2001). 4. National Physical Laboratory, The Examination, Testing and Calibration of Portable Radiation Protection Instrument, Measurement good practice guide nº 29 (1999). 5. American National Standards Institute, American National Standard Radiation Protection Instrumentation Test and Calibration, Portable Survey Instruments, ANSI N323A-1997. 6. Radiation Detection And Measurement. Third Edition. John Wiley & Sons, Inc. (2000) 7. Chandra, R. Nuclear Medicine Physics. The Basics. 5 th Edition. Williams & Wilkins, p. 36, 40, (1998) 8. Macdonald, J., Gibson, C.J. Calibration of radioactive contamination monitors for practical use in hospitals. Nuclear Medicine Communications, 23, 573-580, (2002) 9. Delacroix, D. Guerre, J.P, Leblanc, P., Hickman, C., Radionuclide and Radiation Protection Data Handbook 2002. Radiation Protection Dosimetry. Vol. 98. 1 (2002). 6