TMS filter system with automatic backwashing

Similar documents
DOC JAN06 LT 200

MODEL 7000 SUCTION UNIT

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

Glass and Dishwashers AMX / AUX Series

Oxiperm C 164 Fully automatic chlorine dioxide processing plant C, -6000C, -7500C, C. Service manual. Oxiperm 164C 4-10 kg Service en

PJ130S High Pressure Cleaner

AUTOMATIC MODULAR ICE-CUBE MAKER WITH VERTICAL EVAPORATOR SYSTEM

Glass and Dishwashers AMX / AUX Series

Operating Instructions. Accessory Units Melitta Cafina XT Series. Melitta Professional Coffee Solutions

Chemical resistant Vacuum Pumps

HKF 8180 Operating instructions

Instruction Sheet. DANGER Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.

AUTOMATIC ICE-CUBE MAKER

ASPIRE Laboratory Aspirator

APW-VAP-150P High Pressure Cleaner Operator Manual

AUTOMATIC ICE-CUBE MAKER - INSTRUCTIONS AND WARNINGS

AUTOMATIC MODULAR ICE-CUBE MAKER

READ AND UNDERSTAND THESE INSTRUCTIONS BEFORE OPERATING THE MACHINE

Specifications. Vacuum motor power consumption(w/hp) 1200 / 1.6. Exhaust water pump power consumption(w/hp 800 / 1.1

rev3 INSTALLATION & OPERATION MANUAL OIL CIRCULATING HEATING SYSTEM MODEL OSM

HOT WASHER MODEL NO: KING150

INSTALLATION AND MAINTENANCE GUIDE TRENDY 1000 T ELECTRONIC LAVATORY FAUCET

Model # VWK-2 INSTRUCTION MANUAL

Glass and Dishwashers 402/452/502. (original instructions) (incl. Australian /502) Starting from Serial No.:

AUTOMATIC GRANULAR ICE FLAKER

Operating Instructions. Rotating Vacuum Filter Type VRF

Tornado Operations & Maintenance Manual

K 3.97 M. Specifications or visit our website: In USA call: Part No

K 2.90 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS

K 2.97 M. Specifications Operating pressure, max PSI Water volume GPM Voltage V Amp draw AMPS

Micro/Ultra Filtration System StamoClean CAT 430

EcoMaster - Residential Pools up to 40,000 Gallons

OPERATING INSTRUCTIONS

AIRGOCLEAN 10 E OPERATING MANUAL AIR CLEANER TRT-BA-AIRGOCLEAN10E-TC-001-EN

Operator s Manual CAYENNE HEAT N SERVE 4/3 RECTANGULAR RETHERMALIZER ENGLISH

DELUXE COLLECTION Executive Filtered Water Cooler AQP-WCM-FBOT4

TIH 300 S / TIH 400 S / TIH 500 S / TIH 700 S / TIH 900 S / TIH 1100 S

READ AND UNDERSTAND THESE INSTRUCTIONS BEFORE OPERATING THE MACHINE

PRESSURE WASHER MODEL NO: JETSTAR 1750 OPERATION & MAINTENANCE INSTRUCTIONS. WARNING Read the instructions before using the machine PART NO:

MODULAR AUTOMATIC GRANULAR ICE FLAKER

Manual. Above-counter dispenser for hot drinks OTHG /12

K Specifications. Max. Water Volume

GRUNDFOS INSTRUCTIONS. Sololift2 C-3. Installation and operating instructions

INSTALLATION & OPERATION MANUAL

INSTRUCTIONS FOR TAL SERIES VACUUM HOPPER LOADER

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Because you re not like everyone else. PORTABLE CLEAR ICE MAKER IM200SS OWNER S MANUAL

Operator s Manual. Histology Bath

Model K 520 M Part No

EXTREME 1 Reverse Osmosis System

Fisher Scientific Model 281A Vacuum Oven. Installation and Operating Manual Rev. 0

One-Touch Dispense. Multi-temp selections. 208 F For tea, coffee, instant noodle. 194 F Keep warm around 194 F. 176 F Keep warm around 176 F.

Friwa Compact. Operating instructions EN

Instruction Manual Suppository Penetration Tester. Type PM 30

User Manual GV25 GV35 GV702. Company information: Original instructions GV12066 (1)

Because you re not like everyone else. PORTABLE CLEAR ICE MAKER IM200SS OWNER S MANUAL. Manual v1.0

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

A measured step forward

Operating Manual. Reverse Osmosis Water Purifier. MELAdem 55

- 1 - Updated on 18 March, 2010

K Specifications. Max. Water Volume

Electric steam steriliser

EXTREME 4 Reverse Osmosis System

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

BlueHeat AirTop 2000 Heater

Cavitator Ultrasonic Cleaner Instruction Manual

INSTRUCTION MANUAL HETO CBN 8 / 18 / C Cooling Baths ID: IM.doc Revision 1

SILENT compact /

PRE INSTALLATION PROCEDURES

USER MANUAL SAVE THESE INSTRUCTIONS

INSTALLATION AND OPERATION MANUAL STEAM COIL BASE CONVECTION STEAMER MODEL SCX-16

PHOSPHAX sc, PHOSPHAX indoor sc

Ultraviolet Systems UV-C 50 INSTALLATION INSTRUCTIONS & PRODUCT MANUAL

GB Operating instructions

Pump assembly Instruction manual

EXTREME HP-1 Reverse Osmosis System

Installation manual surface mounting plugs and socket outlets, panel mounting plugs and socket outlets with screw connection (16/32 A)

Model K 240 plus Part No

VADA - Rain2Main Automatic Rainwater Controller

OWNER S MANUAL FOR AIR ADMIRAL VACUUM/PRESSURE STATION MODELS:

OPERATIONAL & INSTRUCTION MANUAL. Model: 2400FA (120v 1Ø) Model: 3200FA (220v-240V 1Ø) Voltage Supply to Machine:

Installation, operating and servicing instructions

Operating instructions

TCH 2000 E OPERATING MANUAL DESIGNER GLASS CONVECTOR TRT-BA-TCH2000E-TC-001-EN

TTV 1500 / TTV 3000 OPERATING MANUAL CONVEYING FAN TRT-BA-TTV TC EN

EXTREME HC-8 Reverse Osmosis System

Tornado Operations & Maintenance Manual

TrueClean ToteCleaner

Millo / Millo pro. Nr x000 / 1805-x000. Ideen für die Dentaltechnik A

BioMaster 250/350/600 BioMaster Thermo 250/350/600

Cleaning unit for coolant. :_decftez`_>r_fr] Book No.: V2

Z-800II / Z-1000II / Z-1020 Fog Machine User Manual

Operator s Manual. IP-100 Immersion Probe Cooler

ROTARY VACUUM EVAPORATOR RVO 004

CAUTION UM120W (RIGHT TANK) FOR WATER BORNE PAINTS ONLY UM120W (RIGHT TANK) SECTION IS NOT DESIGNED FOR USE WITH SOLVENT BASED PAINT

SPECIFICATION, INSTALLATION AND OPERATION MANUAL. Heat Lamp Assemblies. Models: W.HLA900, W.HLA1200 & W.HLA1500

Data Sheet. >>AMTAX inter 2

Glass and Dishwashers. BarAid 500/800. Installation and. Starting from Serial No.:

FINO JETSTEAM PRO 52072

Installation manual surface mounting plugs and socket outlets, panel mounting plugs and socket outlets QUICK-CONNECT (63 A)

Transcription:

DOC023.52.90470 TMS filter system with automatic backwashing User manual 02/2018, Edition 5 Distributed by Hach Lange GmbH. Manufactured by ETL Verfahrenstechnik GmbH. The manufacturer approved the translation of this manual.

Table of contents Section 1 Technical data... 5 Section 2 General information... 7 2.1 Safety information... 7 2.1.1 Danger information in this manual... 7 2.1.2 Warning labels... 7 2.2 Chemical and biological safety... 8 2.3 Overview of product... 9 2.4 Scope of delivery... 10 2.4.1 Optional scope of delivery... 11 2.4.2 Indoor variants... 11 2.5 Front and rear view... 12 Section 3 Installation... 15 3.1 Installation of the probe... 15 3.2 Installation of the control cabinet... 16 3.2.1 Installation of the tube heating... 17 3.2.2 Installation of the compressor... 18 3.2.3 Installation of the mini pump... 19 3.3 Open-loop control of the TMS system... 19 3.3.1 Factory settings of the operating times...21 3.4 Cleaning solution... 21 3.5 Filling of the overflow vessel... 21 Section 4 Start up... 23 4.1 Switch-on... 23 4.2 Valve settings... 23 Section 5 Maintenance... 25 5.1 Maintenance schedule... 25 5.2 Manual cleaning of the probe... 25 5.2.1 Mechanical cleaning... 26 5.2.2 Chemical cleaning... 26 5.2.2.1 Suitable cleaning solutions for manual cleaning tasks... 27 5.3 Probe change... 27 5.3.1 Change of the probe body... 28 5.3.2 Change of the probe... 28 5.4 Storage of the membrane inserts and the complete probes... 28 5.5 Topping up of the cleaning canister... 29 5.6 Cleaning of the permeate tubes... 29 5.7 Manual deaeration of the permeate pump... 30 5.8 Drain condensate in the optional compressor...31 Section 6 Troubleshooting... 33 Section 7 Replacement parts and accessories... 35 7.1 Accessories... 35 3

Table of contents 7.2 Replacement parts...35 Section 8 Warranty and liability...37 Appendix A Installation of the overflow vessel outside of the control cabinet...39 Appendix B Assignment plan of the open-loop control system...41 4

Section 1 Technical data Subject to change without notice. Standard TMS system Power supply System power, complete Permissible ambient temperature without heating Without cabinet heater and heating strips 230V/50Hz (or 115 V/50 Hz version) 85 W + 5 C to + 40 C (+ 41 F to + 104 F) Switch cabinet fan Switch-on temperature (factory setting) 35 C (95 F) Compressed air supply (by customer) Oil-free, particle-free and condensate-free minimum 4 bar Active membrane surface 220 cm 2 (34.1 in 2 ) Pore size Permeate performance in standard operation Delivery height Permeate pump Depends on membrane type 0.2 to 0.5 L/h depending on membrane type and length of the suction line With 5 m suction line: 4 m With 8 m suction line: 3 m 230V/50Hz, 38W Chemical cleans per 24 hours Default settings 1 Intermediate cleans per 24 hours Default settings 5 Contents of cleaning canister Contents of overflow vessel 2.5 liters ph of the raw fluid to be filtered ph 2 to 10 Cabinet dimensions (W x H x D) Weight of control cabinet without tubes Options Top chamber 20 ml (single analyzer supply) or 40 ml (optional, dual analyzer supply) Bottom chamber 100 ml (backwashing) 400 x 600 x 200 mm (15.75 x 23.62 x 7.87 in.) 18 kg Compressor, oil-free Mini pump Power supply (by customer) Motor power Sound power level Delivery amount (standard amount) Pressure Dimensions (W x L x D) Weight of compressor Power supply (internal wiring) Delivery amount 230V/50Hz (or 115 V/50 Hz version) 150 W 47 db 20 to 25 L/min minimum 4 bar/maximum 6 bar 260 x 210 x 140 mm (10.24 x 8.27 x 5.51 in.) 5 kg 230V/50Hz, 4W (or 115 V/50 Hz version) 220 ml/min 5

Technical data Heater of TMS cabinet Heater of compressor cabinet Power supply (internal wiring) Heating power Switch-on temperature of heater (Default settings) Heating strips for external permeate tubes and drain tube Permissible ambient temperature with heating Power supply (by customer) Heating power Switch-on temperature of heater (Default settings) Switch-on temperature of cabinet fan (Default settings) Permissible ambient temperature with heating Cabinet dimensions (W x H x D) Weight of compressor cabinet 230 V/50 Hz (or 115 V/50 Hz version) 100 W 25 C (77 F) 15 W/m -25 C to +40 C (- 13 F to +104 F) 230 V/50 Hz (or 115 V/50 Hz version) 100 W 25 C (77 F) 35 C (95 F) -25 C to +40 C (- 13 F to +104 F) 400 x 400 x 200 mm (15.75 x 15.75 x 7.87 in.) 13 kg 6

Section 2 General information 2.1 Safety information Please read this entire manual before unpacking, setting up, or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment. To prevent damage to or impairment of the device's protection equipment, the device may only be used or installed as described in this manual. 2.1.1 Danger information in this manual 2.1.2 Warning labels WARNING It is forbidden to remove, bypass or override safety devices, safety functions and monitoring devices. WARNING The manufacturer and the dealer are not responsible for any damage that may occur through incorrect use or misuse of this product, and decline to settle any such damage, including direct, indirect and consequential damage to the extent that this is permitted by the applicable law. The user is solely responsible for the identification of critical application issues and risks and for the installation of appropriate mechanisms to protect processes during a possible equipment malfunction. DANGER Indicates a potentially or imminently hazardous situation that, if not avoided, can result in death or serious injury. WARNING Indicates a potentially or imminently dangerous situation that, if it is not avoided, can lead to death or to serious injuries. CAUTION Indicates a possible dangerous situation that can have minor or moderate injuries as the result. NOTICE Indicates a situation that, if it is not avoided, can lead to damage to the device. Information that requires special emphasis. Note: Information that supplements points in the main text. Read all labels and tags attached to the instrument. Failure to do so could results in personal injury or damage to the instrument. This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information. 7

General information This symbol may be found on an enclosure or barrier within the product and indicates a risk of electric shock and/or death by electrocution. This symbol can be attached to the instrument and references a general prohibition. This symbol identifies chemicals that have a skin-corrosive effect (category 1) and can lead to serious eye damage (category 1). This symbol identifies chemicals that have an irritating effect. This symbol requires the wearing of protective gloves. This symbol requires the wearing of eye protection. This symbol requires the wearing of protective clothing. Electrical equipment marked with this symbol may no longer be disposed of in unsorted domestic or industrial waste in Europe after August 12, 2005. In conformity with the applicable provisions, consumers in the EU must return old electrical equipment to the manufacturer for disposal from this date. This is at no cost to the consumer. Note: You obtain instructions on the correct disposal of all (marked and not marked) electrical products that were supplied or manufactured by Hach-Lange at your relevant Hach-Lange sales office. 18-32 kg (39.7-70.5 lbs) When carrying or transporting the instrument components or instrument and if the total weight is more than 18 kg, make sure that suitable lifting equipment is used and/or that the instrument components or instrument are/is carried by two people. 2.2 Chemical and biological safety DANGER Potential danger with contact with chemical/biological substances. Working with chemical samples, standards and reagents can be dangerous. Make yourself familiar with the necessary safety procedures and the correct handling of the chemicals before the work and read and follow all relevant safety data sheets. Normal operation of this instrument may involve the use of hazardous chemicals or biologically harmful samples. Before handling these substances observe all danger notes and safety information printed on the containers of the original solutions and in the safety data sheet. Dispose of all consumed solutions in accordance with the national regulations and laws. Select the type of protective equipment suitable to the concentration and quantity of the dangerous material being used. 8

2.3 Overview of product General information The TMS system, consisting of control cabinet and filter probe, is designed for sample preparation for online analysis instruments. With an automatic backwashing of the filter probe and various membrane materials, this sample preparation system is particularly suitable for use in the supply of wastewater treatment plants and industry. The filter probe (Figure 1, point 10) is immersed via a basin edge holder directly in the fluid to be measured and only removed for maintenance work. The raw fluid to be filtered flows around the filter probe without pressure. A permeate pump (Figure 1, point 12) in the control cabinet generates a negative pressure in the filter probe in order to suck in permeate (filtrate). In the control cabinet, the permeate is delivered to the overflow vessel (Figure 1, point 2 and 3). From the drain of the overflow vessels (Figure 1, point 5), filtered sample is made available to the connected analyzer. The capacity of the overflow vessel is designed so that there is enough permeate available to the analyzers for a measurement, including during the backwash time of the filter probe. Cleaning of the filter probe is automatically performed with cleaning solution and compressed air. The open-loop control times for cleaning, backwashing, pauses and displacing with permeate can be individually adjusted for each application. The cabinet fans and the heater for the control cabinet, permeate lines and drain line of the overflow vessel are switched on and off depending on temperature. The preset temperatures can be individually adjusted. An optional mini pump is necessary for analyzers that do not suck in the sample independently (for example, NITRATA in a flowthrough cell). An optional compressor is necessary if no compressed air supply is provided by the customer. 9

General information Figure 1 Procedure schematic 1 Open-loop control 7 Mini pump (optional) 2 Overflow vessel, 20-mL top chamber 8 Filtrate/compressed air outlet 3 Overflow vessel, 100-mL bottom chamber 9 Cleaning solution 4 Overflow line 10 TMS probe 5 Filtrate inlet 11 Chain holder 6 Compressed air inlet 12 Permeate pump 2.4 Scope of delivery Filter probe Special brush for cleaning the filter probe Ventilated and heated control cabinet completely assembled from glass-fiber reinforced plastic and wiring with Open-loop control unit with permeate pump, Siemens LOGO!open-loop control, non-return valve, optical negative pressure monitoring, solenoid valves, cleaning container, overflow vessel, visual pressure monitor for compressed air Compressed air connection for supply by customer (d=6 mm (0.24 in.)) with integrated pressure-reducing valve (approximately 4 bar) Permeate line (connection of filter probe - permeate pump), PE tube (d=4 mm (0.16 in.)), blue, completely installed in protection tube (d=25 mm (0.98 in.)) with heating strip Permeate line (connection of permeate pump - overflow vessel), PE tube (d=4 mm (0.16 in.)) 10

2.4.1 Optional scope of delivery 2.4.2 Indoor variants General information Permeate line (connection of overflow vessel - analyzer), PE tube (d=4 mm (0.16 in.)), blue, length 1.5 m (4.92 ft) Cleaning solution (concentrate), 5 L Note: Only included in deliveries within Germany! 60 ml suction syringe with tube piece (d=6 mm (0.24 in.)) and tube piece (d=6 mm (0.24 in.)) with transition (to d=4 mm (0.16 in.)) Small compressor with separate power supply, installed in heated and ventilated protection cabinet, connection tube (d=6 mm (0.24 in.)) Compressor output - TMS system input, length 2 m (6.56ft) Mini pump for the active delivery of the sample to the analyzer, installed in control cabinet Adapter for the grid assembly of the filter probe The indoor variants are installed in a heatable instrument housing. Small compressor with separate power supply, connection tube (d=6 mm (0.24 in.)) Compressor output - TMS system input, length 2 m (6.56ft) Ventilated control cabinet completely assembled from glass-fiber reinforced plastic and wiring, open-loop control unit with permeate pump, Siemens LOGO!open-loop control, non-return valve, optical negative pressure monitoring, solenoid valves, cleaning container, overflow vessel 11

General information 2.5 Front and rear view Figure 2 Inside view of control cabinet 1 1 Cabinet heater (outdoor) 10 LED status (power ON/OFF) 2 Heater fan (outdoor) 11 System switch ON/OFF 3 Overflow 12 Thermostats for heater with heater fan (red) and housing fan (blue) 4 Tube connection (d=4 mm) for analyzer (plug connection) 13 Permeate pump 5 Overflow vessel, top chamber 20 ml, bottom chamber 100 ml. 14 Compressed air display 6 Inner housing insulation 15 Valve block 7 Negative pressure display 16 Canister cap 8 Open-loop control 17 Cleaning canister, 2.5 L 9 Optional mini pump switch 18 PVC-tray for cleaning canister 1The 12 control cabinet is available as a 115 V version.

General information Figure 3 Back view of control cabinet 1 Wall mount 3 Air slit (directed downward) 2 Housing fan cover 13

General information Figure 4 Bottom view of control cabinet 1 Power supply 230 V/50 Hz 6 Cabinet ventilation (or 115 V/50 Hz version) 2 Hose feed for overflow vessel drain 7 Free connection (for example, for 2nd analyzer) 3 Hose feed on suction side 8 Free out-/input, for example for data line 4 Hose feed on pressure side (to the analyzer) 9 Compressed air connection (d=6 mm (0.24 in.)) 5 Output of cleaning canister ventilation (no connection) Figure 5 Filter probe 1 Probe head 3 End cap 2 Probe body 14

Section 3 Installation 3.1 Installation of the probe DANGER The installation described in this section of the user manual should only be performed by qualified specialized personnel. WARNING The instrument may not be used in dangerous environments. The manufacturer and its suppliers reject any express or indirect guarantee for use with high-risk activities. NOTICE Any use other than use in accordance with the defined intended purpose in the user manual will result in any warranty claims being rendered invalid and may lead to personal injury and property damage, for which the manufacturer and its suppliers assumes no liability. Unpack all supplied parts carefully, as they are highly sensitive in part to shock and impact. Read the user manual prior to installation and proceed exactly as described. CAUTION Always lay cables and hoses so that they are straight and do not pose a tripping hazard. NOTICE The membrane surface of the filter probe is very sensitive to mechanical effects and therefore is to be handled with utmost care. The filter probe is installed with a basin edge mount. The installation of the basin edge mount is self-explanatory and is delivered without an additional user manual. Always observe the following prior to installation: Place the probe at a distance of at least 200 mm (7.87 in.) from the tank wall. Make sure that the probe can swing freely and there are no fittings or basin wall in the movement range of the probe. The permeate output of the probe is always at the top due to the design. Make sure that the probe is fully immersed. With a probe installation through a grid, use angle adapters to ensure secure tube laying. Make sure of a secure assembly of all tube connections by inserting the tube until it stops due to resistance from the coupling. The insertion depth is approximately 15 mm (0.59 in.) (refer to Figure 6). Make sure that the compressed air made available by the customer fulfils the following requirements for the probe cleaning: Oil-free and condensate-free Pressure at least 4.0 bar 15

Installation Figure 6 Tube connection A Mark the tube at 15 mm (0.59in.) B Insert the tube into the coupling part against any resistance until the mark is met. 1 Coupling part 2 Tube 1. Install the basin holder or channel holder (stainless steel) at a suitable point. 2. Install the clamp connectors at a suitable point on the basin edge. 3. Guide the pre-assembled protection tube (d=25 mm (0.98 in.)) from the raw water side into the stainless steel tube (d=42 mm (1.65 in.)). 4. Connect the permeate line (blue, d=4 mm (0.16 in.)) approximately 15 mm (0.59 in.) into the tube coupling of the probe (Figure 6). Shorten the line if necessary. 5. Connect the probe to the protection tube until hand-tight by use of a union nut (d=25 mm (0.98 in.)). 6. Connect the free end of the permeate line (blue, d=4 mm (0.16 in.)) with the suction side connection of the control cabinet (Figure 4, page 14). 7. Immerse the probe in the raw water. We recommend to secure the tube with a tube clamp (not included in the scope of delivery). Note: The screw connection of the probe should be immersed approximately 100 mm (3.94 in.) deep. Note: The probe should be at least 100 mm (3.94 in.) away from the basin floor. 3.2 Installation of the control cabinet DANGER Danger of an electric shock. Make sure that the supply voltage for the control cabinet and the optional compressor is 230 V/50 Hz, FI-secured by customer. CAUTION Always lay cables and hoses so that they are straight and do not pose a tripping hazard. 1. Install the control cabinet in the immediate proximity of the basin holder or in an instrument housing next to the analyzers. 2. Select the installation height of the control cabinet so that the drain of the overflow vessel lies either at the same level or above the supply of the analyzer provided by the customer. 16

Installation 3. Connect the customer compressed air supply with a coupling connection for the tube (d =6 mm (0.24 in.)) (Figure 4, page 14). A pressure reduction is already installed in the control cabinet and set to approximately 4 bar at the factory. Note: Compressed air must be particle-free, condensate-free and oil-free and be available at a minimum of 4 bar 4. Connect the already installed mains cable (length 2 m (6.56 ft)) of the control cabinet to an existing customer socket (230 V/50 Hz, FI-secured by customer.). Note: For installation in an instrument housing, an indoor variant is also available. Figure 7 Installation of the overflow vessel 1 Overflow vessel, bottom chamber 100 ml. 5 Connection to fill (d=4mm (0.16 in.)) and for mini pump (optional) 2 Overflow (d=10mm (0.39 in.)) 6 Overflow vessel, top chamber 20 ml 3 Permeate for analyzer 1 (d=4mm (0.16 in.)) 7 Connection for cleaning and displacing solution 4 Permeate line pressure-side supply (d=4mm (0.16 in.)) 1 As a standard the overflow vessel gives a volume of 20 ml in the top chamber for the measurement. This is sufficient for one analyzer. If two analyzers are connected to the filter system, a volume of 40 ml is necessary in the top chamber. Insert the supplied plastic tube as an adapter and upgrade the filter system for a 40-mL volume. 3.2.1 Installation of the tube heating 1. In the control cabinet, remove the overflow vessel and the cover of the cable channel lying underneath. 2. Connect the heating strips via the connection as follows: Heating strip 1 with suction side 17

Installation Heating strip 2 with pressure side Heating strip 3 with drain overflow 3. Close the cable channel with the cover. 4. Screw the overflow vessel back into the control cabinet. The cabinet heating including heater fan and the housing fan are switched on and off depending on temperature. The heating strips of the tubes are self-regulating and switch on and off as required. 3.2.2 Installation of the compressor If no customer compressed air is present, the optional compressor must be installed. The compressor is pre-assembled in a separate cabinet and is connected to the control cabinet with the accompanying connection tube (d=6 mm (0.24 in.)) (connections appropriate for tube couplings with input d=6 mm (0.24 in.)). 1. Install the compressor cabinet underneath the control cabinet so that the supply lines located underneath are freely accessible. Both cabinets are equal in width and depth and can be installed on a mounting device. 2. Connect the mains cable of the compressor with the inside socket (230 V/50 Hz, FI-secured) (Figure 8, point 1). 3. Connect the mains cable of the cabinet to the nearest socket (230 V/50 Hz, FI-secured by customer). The compressor is independent of the power supply of the control cabinet. 4. Connect the compressed air supply with a coupling connection to the control cabinet (Figure 4, page 14). Note: For installation in an instrument housing, an indoor variant is also available. 18

Installation Compressor 1 Figure 8 1 Inside socket 5 Compressed air tank 2 Thermostats for heater with heater fan (red) and housing fan (blue) 6 Dewatering valve 3 Compressor 7 Cabinet heater 4 Inner housing insulation 8 Electric distributor 3.2.3 Installation of the mini pump For non-sucking analyzers (for example, NITRATA in flowthrough cell) an optional mini pump is necessary, which supplies the analyzer with enough permeate. 1. Connect the output of the mini pump with the input of the analyzer overflow vessel. 2. Connect the output of the analyzer overflow vessel with the upper input of the overflow vessel (return) (Figure 2, point 4). 3.3 Open-loop control of the TMS system The TMS system is operated via the LOGO!open-loop control unit. 1The compressor is also available as a 115 V version. 19

Installation Figure 9 Display of the open-loop control unit 1 Display 4 TH: Membrane cleaning time with cleaning solution (5 min) 2 Interface cover 5 TL: Cycle time of the membrane cleaning with cleaning solution (23 h : 55 min) 3 Example: block number 50 6 Ta: Current time until the subsetquent membrane cleaning with cleaning solution (2 min : 8 sec) Figure 10 Keypad of the open-loop control unit 1 ESC: back, change, cancel 2 OK: advance, confirm A Left C Right B Up D Down The display shows the operating status of the system. Table 1 Displays Operating status Normal operation Sample delivery Cleaning Display Weekday, time, date Sample delivery Membrane cleaning or compressed air flushing 20

Installation 3.3.1 Factory settings of the operating times 3.4 Cleaning solution NOTICE The operating times cannot be changed by the user. If you need to change the operating times, contact the supplier's service team. Table 2 Factory settings of the operating times Operating status Display Default settings Total cycle time 24 h Backwashing with cleaning solution Membrane cleaning 1x daily Intermediate cleaning with compressed air Compressed air flushing every 4 h DANGER Potential danger with contact with chemical/biological substances. Working with chemical samples, standards and reagents can be dangerous. Make yourself familiar with the necessary safety procedures and the correct handling of the chemicals before the work and read and follow all relevant safety data sheets. NOTICE Observe the specified concentrations exactly. Stronger concentrations of solution can cause damage to the analyzers and faulty measurements. NOTICE The finished cleaning solutions must have a ph value of 1 to 13. For the automatic membrane cleaning, the cleaning canister (Figure 2, page 12, point 17) must be filled with cleaning solution. 1. Remove the empty cleaning canister from the control cabinet. 2. Pour the necessary amount of the supplied cleaning concentrate corresponding to the use instruction (stuck to canister) into the cleaning canister. 3. Top up the canister with distilled water. 4. Close the cleaning canister and mix the cleaning solution. 5. Check the ph value of the finished solution. 6. Insert the cleaning canister back into the PVC-tray in the control cabinet. 7. Connect the tubes with the closure lid of the canister. Note: Alternatively, chlorine bleach (NaOCl) diluted with distilled water with a maximum free chlorine proportion of 0.5% (for PHOSPHA sc) or 0.3% hydrochloric acid (HCl) (for Amtax SC) can also be used as a cleaning solution. 3.5 Filling of the overflow vessel Before the first use, the overflow vessel must be filled by hand with the supplied syringe. 1. Fill the syringe with distilled water. 2. Remove the upper closure of the overflow vessel (Figure 2, page 12, point 5). 21

Installation 3. Guide the tube of the syringe into the overflow vessel. 4. Fill the overflow vessel (20-mL and 100-mLvolume) until water comes out at the overflow of the bottom chamber (100 ml) (Figure 4, page 14, point 2). 5. Put back on the upper closure of the overflow vessel. 6. Repeat this procedure if in later operation, after renewed switch-on, the overflow vessel is not completely filled. 22

Section 4 Start up 4.1 Switch-on 4.2 Valve settings 1. Start the compressed air supply (from customer or with optional compressor). 2. Switch on the system at the system switch (Figure 2, page 12, point 11). A cleaning cycle (7:30 min) is started. The cleaning cycle is composed of: Backwashing with cleaning solution Compressed air flushing Displacement of the cleaning solution and compressed air Then the filter system switches into normal operation and the filtration process begins. Permeate is delivered into the overflow vessel. Note: The cleaning cycle always starts when the system was powered off and is switched back on with the system switch. Red light-emitting diodes on the valves show their operation in the program sequence, refer to Table 3. Table 3 Valve position in the program sequence (-) Power off (light-emitting diode off) (+) Power on (light-emitting diode on (red)) V3 V1 V2 V4 Permeate delivery cycle - - - - Cleaning cycle (chemicals) - + + + Cleaning cycle (compressed air) + - - - Displacement cycle (permeate) - + + - 23

Start up 24

Section 5 Maintenance 5.1 Maintenance schedule Table 4 Maintenance schedule Maintenance measure Time interval Every week Every month Every 6 months Every year Every 2 years Manual cleaning of the probe (application-dependent) Check probe for damage Probe change 1 Filling of the cleaning canister (application-dependent) Sealing check Check automatic membrane cleaning Cleaning of the permeate tubing Check assembly General function check of the system Check tubes Exchange connection tube to the analyzer Exchange overflow vessel drain tube Check permeate pump function Exchange permeate pump head Check cabinet fan Check cabinet heater Check tube heating strips Drain condensate in the compressor Check compressor Replace compressor 2 Check compressor cabinet fan Check compressor cabinet heater 1 In typical operating conditions, a different interval may be required depending on the specific application and local conditions. 2 Compressors are wearing parts and are not covered by the instrument warranty. 5.2 Manual cleaning of the probe NOTICE The membrane surface of the filter probe is very sensitive to mechanical effects and therefore is to be handled with utmost care. If the negative pressure gauge shows -0.8 bar during normal operation, the probe must be mechanically and/or chemically cleaned in the laboratory. Note: In the case of application in the wastewater treatment plant supply, the probe must be cleaned at a negative pressure of -0.7 bar to -0.9 bar (approximately 1x per week). A pulsing of the negative pressure gauge indicates a normal operating state of the system. Note: If after cleaning there is still a high negative pressure, check whether the suction line is dirty. 25

Maintenance 5.2.1 Mechanical cleaning 1. Switch off the system at the system switch. 2. Pull the probe out of the raw water. 3. Rinse away coarse impurities with low pressured water. 4. Carefully clean the membrane with the supplied brush until the light membrane surface is visible. Note: Do not press the brush too hard on the membrane surface and do not change the movement direction. 5.2.2 Chemical cleaning 5. Check the membrane for damage. 6. Immerse the probe back in the raw water. 7. Clean the brush thoroughly under running water. Let the brush dry and store it dust-free. 8. Switch on the system at the system switch. DANGER Potential danger with contact with chemical/biological substances. Working with chemical samples, standards and reagents can be dangerous. Make yourself familiar with the necessary safety procedures and the correct handling of the chemicals before the work and read and follow all relevant safety data sheets. NOTICE The cleaning causes a shift of the ph value in the raw water. This can cause loosened components to be settled. Be aware that after the cleaning, short-term measurement value disruptions can occur. 1. Switch off the system at the system switch. 2. Pull the probe out of the raw water. 3. Rinse away coarse impurities with low pressured water. 4. Carefully clean the membrane with the supplied brush until the light membrane surface is visible. Note: Do not press the brush too hard on the membrane surface and do not change the movement direction. 5. Unscrew the end cap from the probe. 6. Pull the probe body from the headpiece of the probe. Note: Be aware of both sealing rings and lift these off. 7. Clean and grease the screw connections and gaskets. 8. Insert the probe body carefully into the supplied cleaning container installed to floor stand. 9. Fill the cleaning container in the laboratory with cleaning solution. 26

Maintenance 10. Close the cleaning container and leave the probe body for up to a week in the cleaning bath. Note: Then rinse the probe body with drinking water. Note: If impurities continue to be on the membrane surface, repeat the chemical cleaning. 11. Insert a new or already chemically cleaned probe body with the gaskets between head- and end cap. 12. Screw the probe together hand-tight. 13. Check the membrane for damage. 14. Immerse the probe back in the raw water. 15. Switch on the system at the system switch. 5.2.2.1 Suitable cleaning solutions for manual cleaning tasks NOTICE The finished cleaning solutions must have a ph value of 1 to 13. NOTICE After each chemical cleaning, the probe body must be rinsed with drinking water. Note: The chemical cleaning of the probe body can be improved and the time taken can be shortened with the cleaning set LZH235. Depending on the membrane type and type of the impurity, another cleaning solution is required. Table 5 Suitable cleaning solutions for the standard membrane type porospoly Cleaning solution 1 Concentration Type of impurity MemClean-Cl 50 % Organic impurities MemClean peroxide 3 % Organic impurities Chlorine bleach NaOCl, 13 % Cl Maximum 10 % Organic impurities Caustic soda NaOH 0.5 % Organic impurities Hydrochloric acid HCl 0.3 % Inorganic impurities Citric acid 10 % to 25 % Inorganic impurities 1 Make sure that you use phosphate-free cleaning solutions if a phosphate analyser is connected. 5.3 Probe change If two cleaning steps are required, first perform cleaning with an alkaline solution and then a cleaning with an acidic solution. NOTICE The membrane surface of the filter probe is very sensitive to mechanical effects and therefore is to be handled with utmost care. 27

Maintenance 5.3.1 Change of the probe body 1. Switch off the system at the system switch. 2. Pull the probe out of the raw water. 3. Rinse away coarse impurities with low pressured water. 4. Carefully clean the membrane with the supplied brush until the light membrane surface is visible. Note: Do not press the brush too hard on the membrane surface and do not change the movement direction. 5.3.2 Change of the probe 5. Unscrew the end cap from the probe. 6. Pull the probe body from the headpiece of the probe. Note: Be aware of both sealing rings and lift these off. 7. Clean and grease the screw connections and gaskets. 8. Insert a new or already chemically cleaned probe body with the gaskets between head and end cap. 9. Screw the probe together hand-tight. 10. Check the membrane for damage. 11. Immerse the probe back in the raw water. 12. Switch on the system at the system switch. 1. Switch off the system at the system switch. 2. Pull the probe out of the raw water. 3. Rinse away coarse impurities with low pressured water. 4. Loosen the union nut on the protection tube (d=25 mm (0.98 in.)). 5. Disconnect the permeate line (blue, d=4 mm (0.16 in.)). 6. Clean and grease the screw connections and gaskets. 7. Connect the permeate line (blue, d=4 mm (0.16 in.)) approximately 15 mm (0.59 in.) to the tube coupling of the new or already chemically cleaned probe (Figure 6, page 16). 8. Connect the probe to the protection tube until hand-tight by use of a union nut (d=25 mm (0.98 in.)). 9. Immerse the probe in the raw water and tighten the tube retention on the stainless steel tube end. 5.4 Storage of the membrane inserts and the complete probes Used and cleaned membrane inserts are stored in approved cleaning solutions, for example, diluted chlorine bleach (NaOCl, approximately ph 11). Probes and probe bodies of the "porospoly" membrane type are stored dry and dust-free at ambient temperature. New probes are stored dry. 28

Maintenance 5.5 Topping up of the cleaning canister DANGER Potential danger with contact with chemical/biological substances. Working with chemical samples, standards and reagents can be dangerous. Make yourself familiar with the necessary safety procedures and the correct handling of the chemicals before the work and read and follow all relevant safety data sheets. NOTICE Observe the specified concentrations exactly. Stronger concentrations of solution can cause damage to the analyzers and faulty measurements. NOTICE The finished cleaning solutions must have a ph value of 1 to 13. For the automatic membrane cleaning, the cleaning canister (Figure 2, page 12, point 17) must be filled with cleaning solution. 1. Switch off the permeate pump (Figure 2, page 12, point 13). 2. Remove the empty cleaning canister out of the PVC-tray in the control cabinet. 3. Pour the necessary amount of the supplied cleaning concentrate corresponding to the use instruction (stuck to canister) into the cleaning canister. 4. Top up the canister with distilled water. 5. Close the cleaning canister and mix the cleaning solution. 6. Check the ph value of the finished solution. 7. Insert the cleaning canister back into the PVC-tray in the control cabinet. 8. Connect the tubes with the closure lid of the canister. 9. Switch on the permeate pump. Note: Alternatively, chlorine bleach (NaOCl) diluted with distilled water with a maximum free chlorine proportion of 0.5% or 0.3% hydrochloric acid (HCl) can also be used as a cleaning solution. 5.6 Cleaning of the permeate tubes NOTICE Make sure that all connected analyzers are running in the service mode. Observe the respective user manuals of the analyzers. NOTICE The cleaning causes a shift of the ph value in the raw water. This can cause loosened components to be settled. Be aware that after the cleaning, short-term measurement value disruptions can occur. All tubes that deliver permeate are rinsed with cleaning solution. 1. Switch all analyzers to service mode. Observe the respective specifications in the corresponding user manuals. 29

Maintenance 2. Check the remaining runtime of the cleaning cycle. If the system is in the cleaning cycle, wait until this is complete. 3. Switch off the permeate pump (Figure 2, page 12, point 13). 4. Pull the probe out of the raw water. 5. Rinse away coarse impurities with low pressured water. 6. Fill a container with cleaning solution. 7. Immerse the probe completely in the cleaning solution. 8. Switch on the permeate pump. Cleaning solution is sucked in via the probe and rinses all permeate delivery tubes. 9. Switch off the permeate pump after 30 minutes and leave the cleaning solution 1 2 hours to take effect. 10. Immerse the probe completely in a container with distilled water. 11. Switch on the permeate pump. Distilled water is sucked in via the probe and rinses all permeate-delivery tubes. 12. Switch off the permeate pump after 30 minutes. 13. Check the membrane for damage. 14. Immerse the probe back in the raw water. 15. Switch on the permeate pump. 5.7 Manual deaeration of the permeate pump If no permeate is delivered into the overflow vessel, perform a manual deaeration of the permeate pump. 1. Loosen the orange rotary knob. Air comes out of the orange bleed nipple. 2. Turn the orange rotary knob closed as soon as permeate comes out of the orange bleed nipple. If no permeate comes out: 1. Connect the tube of the syringe with the orange bleed nipple (Figure 11, page 31). 2. Suck air with the syringe until permeate comes out. 3. Turn the orange rotary knob closed. Note: It may take a few minutes until permeate is transported bubble-free into the overflow vessel. 30

Maintenance Figure 11 5.8 Manual deaeration of the permeate pump Drain condensate in the optional compressor NOTICE The condensate must be removed every month, otherwise the compressor can be damaged and the warranty invalidated. The compressor must be maintained in accordance with the accompanying user manual from the dealer. Empty the condensate from the compressor every month. 1. Open the dewatering valve on the underside of the compressor (Figure 12, page 32). The condensate flows out. 2. When no more water comes out, close the dewatering valve. 31

Maintenance Figure 12 Compressor 1 1 Dewatering valve 1The compressor is also available in a 115V version. 32

Section 6 Troubleshooting Problem Definition Solution System does not work Power supply interrupted Actuate system switch (Figure 2, page 12, point 11) Permeate pump is at a standstill but has voltage Permeate pump is running but is not delivering Negative pressure display (0 bar) Negative pressure display high (greater than -0.8 bar) Negative pressure too high, delivery amount too small Service life of the probe too short Permeate pump is running but sometimes does not deliver System is running, analyzer shows malfunction Air ingress in the overflow vessel Permeate pump switched off Various causes, depending on negative pressure display Permeate pump filled with air Actuate permeate pump switch (Figure 2, page 12, point 13) Check negative pressure display (Figure 2, page 12, point 7) Manual deaeration of the permeate pump (section 5.7, page 30) Probe body blocked Exchange and clean probe body (section 5.2, page 25) Permeate tubes kinked or blocked Probe contaminated Suction line leaky Cleaning of the probe insufficient Cleaning cycle defective Floating of the probe (at high flow speed) Overflow vessel empty Overflow empty or insufficiently filled at restart Tube connection between overflow vessel and analyzer defective Probe blocked, negative pressure too high Compressed air cannot ease via the membrane probe Immerse cleaned probe in drinking water and switch on permeate pump. If negative pressure is still too high: Remove probe and immerse loose tube end in drinking water. Switch on permeate pump Remove permeate pump head and check non-return valves on suction side and pressure side Exchange and clean probe (section 5.2, page 25) Switch system switch off and back on: manual starting of the cleaning cycle Remove probe and immerse loose tube end in drinking water. Switch on permeate pump Perform chemical cleaning: acid, alkali or a combination of both. Check cleaning cycle (steps and times) Change cleaning solution (other cleaner type) Check location; are, for example, precipitants or flocculation agents added in the surroundings of the sampling When was the last cleaning (current drain time of block 50 should be at least 30 min) Select another location Check delivery amount of the permeate pump Install optional mini pump Manual filling of the overflow vessel (section 3.5, page 21), then start cleaning cycle Check tube connection for blockages or kinks Exchange and clean probe (section 5.2, page 25) Compressed air goes after valve switching into the pump head and then into the overflow vessel Analyzer shows malfunction Mini pump defective Check mini pump Not enough compressed air for flushing Water in the compressor Drain condensate water section 5.8, page 31 Peaks in the measurement values after cleaning interval Concentration of the cleaning solution too high Dilute cleaning solution 33

Troubleshooting 34

Section 7 Replacement parts and accessories 7.1 Accessories Description Cleaning set for laboratory cleaning of the probe body (membrane insert), complete (floor stand with clamping tool, syringe with quick coupling, glass cylinder) Syringe with quick coupling (male+female part) Cleaning container and storage container for probe body (membrane insert) Probe holder for installation on the basin edge including mounting material (stainless steel/aluminum) Probe holder for grid installation (PVC) Mini pump including connections, 230 V, length 1.5 m (4.92 ft), tube d=6 mm (0.24 in.) Mini pump including connections, 115 V, length 1.5 m (4.92 ft), tube d=6 mm (0.24 in.) "Connector" service kit, all connections for the TMS system TMS flowthrough pot with lid and connections (hose fitting d25) Floor stand for cleaning set Weather shield Cat. No LZH235 LZH256 LZH212 LZH183 LZH185 LZH186 LZH289 LZH236 LZH226 LZH383 LZH384 7.2 Replacement parts Description Cleaning canister, 2.5 L Cleaning replacement canister, 5 L, with suction fitting and air bleed in the lid Cleaning canister, 2.5 L, with PVC-tray and fittings, conversion set Compressor, TC-20T, 230 V, complete with length 2 m (6.56 ft) PE tube d=6 mm (0.24 in.) Compressor, TC-20T, 115 V, complete with length 2 m (6.56 ft) PE tube d=6 mm (0.24 in.) Compressed air tube, length 5 m (16.4 ft), d=6 mm (0.24 in.), blue Drain line, length 2 m (6.56 ft), unheated (protection tube) Drain line, length 2 m (6.56 ft), heated, 115 V, complete (protection tube, heating strip with connection (male+female part)) Drain line, length 2 m (6.56 ft), heated, 230 V, complete (protection tube, heating strip with connection (male+female part)) Filter probe, complete (housing including probe body) Housing of the filter probe including plug connection Permeate line, length 25 m (82.02 ft), d=4 mm (0.16 in.), blue Pressure line, length 1.5 m (4.92 ft), heated, 230 V, complete (protection tube, permeate line, heating strip with connection (male+female part)) Pressure line, length 1.5 m (4.92 ft), heated, 115 V, complete (protection tube, permeate line, heating strip with connection (male+female part)) Pressure line, length 1.5 m (4.92 ft), unheated, complete (protection tube, permeate line) Probe body (insert), membrane type porospoly-t-025-12-0050 including O-rings PVC-tray for cleaning canister, 2.5 L Suction line, length 5 m (16.4ft), heated, 230 V, complete (protection tube, permeate line, heating strip with connection (male+female part)) Suction line, length 5 m (16.4 ft), heated, 115 V, complete (protection tube, permeate line, heating strip with connection (male+female part)) Cat. No LZH374 LZH192 LZH386 LZH210.xx LZH255 LZH294 LZH268 LZH264 LZH263 LZH160 LZH295 LZH269 LZH261 LZH262 LZH267 LZH211 LZH375 LZH257 LZH258 35

Replacement parts and accessories 7.2 Replacement parts Description Suction line, length 5 m (16.4 ft), unheated, (protection tube, permeate line, filter probe connection) "Tubes - open-loop control unit" service kit, all tubes for a TMS open-loop control unit, including 8 m (26.25 ft) suction side and 1.5 m (4.92 ft) pressure side "Tubes" maintenance kit, 0.5 m (1.64 ft) overflow tube d=10 mm (0.39 in.), 2 m (6.56 ft) permeate pressure side d=4 mm (0.16 in.) with plug transition and approximately 0.1 m (3.94 in.) tube d=6 mm (0.24 in.) Cat. No LZH265 LZH290 LZH291 36

Section 8 Warranty and liability The supplier guarantees that the supplied product is free of material and manufacturing defects and undertakes to repair or replace any defective parts at no cost to the owner. The warranty period is 24 months. If a maintenance contract is taken out within 6 months of purchase, the warranty period is extended to 60 months. With the exclusion of further claims, the supplier is liable for defects, including the lack of assured properties, as follows: all parts that, within the warranty period calculated from the day of the transfer of risk, can be demonstrated to have become unusable or that can only be used with significant limitations owing to circumstances prior to transfer of risk, in particular due to incorrect design, substandard materials or inadequate finish, shall be repaired or replaced at the supplier's discretion. The identification of such defects must be reported to the supplier in writing as soon as possible, but no later than 7 days after the discovery of the fault. If the customer fails to notify the supplier, the product is considered approved despite the defect. Further liability for indirect or direct damages is not accepted. If device-specific maintenance- or inspection work prescribed by the supplier is to be performed within the guarantee period by the customer (maintenance) or by the supplier (inspection) and these requirements are not met, claims for damages that result from non-observance of these requirements are void. Further claims, in particular for consequential damages, cannot be made. Wear parts and damages that arise through improper handling, incorrect installation or non-regulatory use are excluded from this clause. The reliability of the supplier's process instruments has been tried and tested in many applications and these instruments are therefore often used in automatic control loops to provide the most economical method of operation for the relevent process. To avoid or limit consequential damage, it is therefore recommended that the control loop be designed such that an instrument malfunction results in an automatic changeover to the backup control system. This guarantees the safest operating condition both for the environment and the process. 37

Warranty and liability 38

Appendix A Installation of the overflow vessel outside of the control cabinet The overflow vessel is installed by the factory in the control cabinet, however you can also remove it from the control cabinet and secure it next to the analyzer. CAUTION The pressure line must be no longer than 2 m. A 1.5 m line is supplied as standard. 1. Connect the pressure-side permeate line (d=4 mm (0.16 in.)) to the left top-side plug connection (Figure 13, point 4). 2. Connect the side connected overflow pipe (100 ml chamber) to an overflow tube (d=10 mm (0.39 in.)) (Figure 13 point 2). Connect the overflow tube bottom-side of the control cabinet to a heated (optional) drain tube, ensure the tube leads to an open drain. Note: The overflow tube (d=10 mm (0.39 in.)) and the drain tube (d=25 mm (0.98 in.)) must not be submerged. 3. Connect the side connected overflow pipe (20 ml area) to the analyzer (tube d=6 mm (0.24 in.)) (Figure 13, point 3). 4. Select the installation height of the overflow vessel so that the drain of the overflow vessel lies either at the same level or above the supply of the analyzer provided by the customer. 5. The optional mini pump for the permeate line makes sure that the analyzer is reliably supplied with sufficient permeate when the difference is small in height between the analyzer inlet and the TMS overflow vessel. 6. Connect the output of the mini pump to the inlet of the internal overflow vessel of the analyzer. Connect the outlet of this overflow vessel to the second upper inlet of the overflow vessel of the TMS system (plug connection 6 mm). The construction creates a permeate circuit between the TMS system and the analyzer. 39

Installation of the overflow vessel outside of the control cabinet Figure 13 Overflow vessel 1 Overflow vessel, lower compartment 100 ml. 5 Connection to fill (d=4mm (0.16 in.)) and for mini pump (optional) 2 Overflow (d=10mm (0.39 in.)) 6 Overflow vessel, upper compartment 20 ml 3 Permeate for analyzer 1 (d=4mm (0.16 in.)) 7 Connection for cleaning and displacing solution 4 Permeate line pressure-side supply (d=4mm (0.16 in.)) 1 As a standard the overflow vessel provides a volume of 20 ml in the upper compartment for the measurement. This is sufficient for one analyzer. If two analyzers are connected to the filter system a volume of 40 ml is required in the upper compartment. This upgrade can be made by inserting a plastic tube as an adapter (included in the box). 40

Appendix B Assignment plan of the open-loop control system Figure 14 Assignment plan DANGER Risk of electric shock! When performing maintenance work, always switch off the main switch at the right-hand side on the open-loop control housing and pull out the power plug. 1 Mini pump switch (optional) 7 Mini pump (optional) 2 System switch ON / OFF 8 Signal 3 Heating strip 1 suction side 9 Permeate pump 4 Heating strip 2 pressure side 10 Permeate pump pulse sender 5 Heating strip 3 drain-overflow 11 Thermostat 6 Cabinet heater 12 Outer fan 41

Assignment plan of the open-loop control system 42

HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info-de@hach.com www.de.hach.com HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499 Hach Company/Hach Lange GmbH, 2014, 2017 2018. All rights reserved. Printed in Germany 02/2018 Edition 5