The Heidelberg Integrated 14 CO 2 Sampler Institute of Environmental Physics, ICOS CRL

Similar documents
Multi-Mode Electrode pro

HAVE A SAMPLING AND ANALYSIS PLAN (SAP)

OPERATING INSTRUCTIONS KSG 40/60-2 KSG 50/70-2 KSG 50/80-2 KSG 50/95-2

HydroFlux Welder Torch System T0-102 / T0-103

LECTURE-18 Vapour Absorption Refrigeration System

1. General Description

Buehler Cast N Vac 1000 Embedding System

BOUMATIC ROBOT AND THE LELY SAMPLE

AIC Brazed Plate Heat Exchanger Operating and Instruction Manual

Shop. Item Information. Rotavapor R-300 B-305, SJ24/40, C, P+G, WB, 110V. Send item information. Show item in shop. Add item to cart.

FAULT FINDING. Water Gas Cassette Toilet FAULT FINDING

Safety Lab Goggles are required! Apron is optional!

Glovebox Usage Manual v1.4. When you enter the lab walk straight to the log book. Enter all information including N2 and O2/H2O levels.

WET & DRY GARAGE VAC 20L CAPACITY WATT 1.5M MOTOR TANK STEEL HOSE KP702 ON BOARD ACCESSORY STORAGE SUCTION STAINLESS

Manual Page 1 of 8. Manual. Laboratory autoclave

Assembly of a Microbial Fuel Cell

Installation, Operation and Maintenance Instructions ThermaFlex Gasketed Plate Heat Exchanger OM010

64 Beverage dispensing

NFT. basic steps to successful growing. nutriculture ltd. Do you want. Find out more. Hydroponic

OPERATING INSTRUCTIONS FOR THE TP18SX-HP

Installation and Operating Instructions DÜRR Regeneration Unit for X-ray developers XR 24, XR24 II, XR 24 Nova, XR 24 Pro

REMOVE AND INSTALL AIR CLEANER ASSEMBLY

PLATE HEAT EXCHANGER

Installation, Operation and Maintenance Instructions for ThermaFlex Gasketed Plate Heat Exchangers

TECHNOSKLO Ltd. Držkov

INSTRUCTION MANUAL FOR Freeze Dry Systems PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Laboratory Techniques

DEUTSCH Bio-Master Bio-Max

BASIC SYSTEM LOCATION INDEX

GAS CHROMATOGRAPHY PURDUE UNIVERSITY INSTRUMENT VAN PROJECT. DEMYSTIFICATION ACTIVITY FOR THE GAS CHROMATOGRAPH (Revised: ) INTRODUCTION

Short Path Distillation Unit. KDL 1 BASIC Version 2015

WALLPAPER STEAMER. Model ASSEMBLY AND OPERATING INSTRUCTIONS

SX-300 SX-500 SX-700

ISSUED: FEB. 23, 1998 REVISED: OCT. 15, 1999 HOSHIZAKI MODULAR FLAKER MODEL F-450MAF(-C)

Operation Manual R205BV, R209BV

Chapter 5 Checks and Maintenance

ASSEMBLY & OPERATING INSTRUCTIONS

PROLUX AIR PURIFIER / HUMIDIFIER MODEL: PL-2 Operation Instruction

Reliability is a beautiful thing TM MODULAR FLAKER F-800MAH(-C) F-800MWH(-C) INSTRUCTION MANUAL

Radiator Additive. ZetaLube. Description

English Manual for Jewelry Casting and Investing The professional choice

Rotary Evaporator Replacement Glassware

50 Disassembly and assembly Cab and its attachments

Model 1151-M Mini-Dust-R User s Manual

1. Plug electric cord to appropriate 110V 220V electrical outlet. Push the off/on switch to fill position.

Installation, Operation and Maintenance Instructions for ThermaFlex Brazed Plate Heat Exchangers

Komline-Sanderson Paddle Dryer. Drying technology for biosolids, sludges and by-products

BAXTER: FACTOR VII 600 IU (Plasma Derived)

Rotary Evaporator Replacement Glassware

Altra Series Dampener

Hoshizaki America, Inc.

USED OIL RE-REFINING UNIT PRESENTATION. Manufactured by GADGIL & COMPANY PUNE

U S E R G U I D E. Home Respiratory Equipment Cleaning Guide. Health Information

Operation of heat exchangers

P.O.D. Pressure On Demand

IMPORTANT WARNINGS Therefore, to avoid breaking it, do not turn it manually in order to seek the correct position.

Dr. Infrared Heater Elite Series

2006G2 OPERATING INSTRUCTIONS. 2006G2 Pressure-Vacuum Hand Pump July 2011 SOILMOISTURE

INSTALLATION INSTRUCTIONS GEO PRIME TANK. (Patent Pending) GPC

AUC Cell Washer. User Manual

INSTRUCTION MANUAL W-400

Determination of the enthalpy of vaporization of liquids

Version /08/2018 USER MANUAL VACUUM FILTER PRESS. All rights reserved Les Équipements d Érablière CDL inc.

Owners Manual MINI-CLASSIC II. Pure Water. Water Distiller

Methods of solid phase extraction. Methods of solid phase extraction


Hoshizaki America, Inc.

UF424 and UN324. Technical Service

QTFX-LF900 Low-Level Fog Machine User Manual

Training Papers Distillation with a Rotary Evaporator

INSTALLATION GUIDE. soft duomatik evoline

Hoshizaki America, Inc.

FCT-CE SERIES ELECTRIC TILTING KETTLE W/HAND CRANK PARTS AND SERVICE MANUAL

Rycroft Supapac Plate Heat Exchangers

Before using your machine, you must familiarize yourself with all of its components.

BMW E36 Thermostat Removal And Coolant Flush

Stenner Pump PCM Proportional Feed Chlorinator Installation & Start Up Guide

Owners Manual for the AV18AX

Chemistry Materials Separation Processes

COMPACT GARMENT STEAMER

Daily care manual. Table of contents

Tanis Engine Dehydrator. User s Guide

VaporMax PC800 Steam Cleaner

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

FC 5 FC 5 Premium. OPERATOR MANUAL Floor Cleaner...2 MANUAL DE USUARIO Limpiador de piso...22 MANUEL D UTILISATION Nettoyant pour sols...

Installation instructions Air drying unit for Bubble Sensor OTT CBS. English

Operating Manual RS6 Pro

Plant-A-Plant Carbon Dioxide Laboratory guide

Using the Lely Robot Sampler CanWest DHI Robot Manual

OPERATING & SERVICE PARTS MANUAL 700ES ENERGY SMART WRAPPER SERIES C

1 L Rotary Vacuum Evaporator Instruction Manual

7MGST9679PW0 GABINETE

Operator Guide. Publication No. OI /97-Eng. KODAK Automixer II Plus

How to care for your water storage tank

music hall wcs-2 wet vacuum record cleaner music hall audio llc

Adjustable Back Pressure Flow Controller Instructions

7MMSL7959TW0 GABINETE

Laboratory Techniques

WASHER/DRYER CONTROL PANEL PARTS For Model: MET3800TW2 (White)

Rotary Evaporator Replacement Glassware

Transcription:

The Heidelberg Integrated 14 CO 2 Sampler Institute of Environmental Physics, ICOS CRL 1. Description of the sampling technique and operation mode With the 14 CO 2 sampler atmospheric CO 2 is being collected over about two weeks to be analysed for its 14 C/ 12 C ratio. The CO 2 is absorbed in sodium hydroxide solution (NaOH), which is sent to the Central Radiocarbon Laboratory for analysis. This is the schematic set up of the sampler: 1air inlet pump flow meter gas meter wash bottle Motor with Raschig tube Atmospheric air is being pumped from the air inlet through a flow meter, where one can adjust the airflow with a needle valve. For a conventional 14 C analysis, CO 2 from about 25 m 3 air is necessary. Therefore, the mean flow rate for a sample integration time of two weeks must be about 75 l/h. After the flow meter, the air passes a gas meter where the amount of the already passed air is monitored. Then the air goes through a wash bottle to get humidified. This prevents the NaOH solution to get too concentrated. Now the air flows through the glass Raschig tube where the CO 2 is absorbed by the NaOH in the tube. The tube is slowly rotated with an evaporator drive to renew the absorbing NaOH film on the Raschig rings. The motor acts as a vacuum rotary feedthrough. 2. Installation of the sampler and start of first sampling The above described parts of the sampler have to be arranged as shown in figure 1. The metal plate with the pump must be screwed on the four rubber buffers (figure 2). Also, the flow meter has to be fixed at its small pedestal with the two screws (figure 3). The gas meter is put on its holding and fixed with the elastic bands (figure 4). The wash bottle stands in the ringshaped bottle fitting (figure 5). 1

Figure 1: View on the 14 CO 2 sampling system Figure 2: Installation of the pump Figure 3: Flow meter (back) 2

Figure 5: Wash bottle Figure 4: Gas meter Now stick up the black attachments from the motor on the hollow glass shaft as shown in figure 6 and screw it together. Put the hollow shaft (see figure 7) into the motor, then from the other side the gasket ( Drive side must show to the motor), the glass adapter for the flexible hose and the metal spring. Finally, fix all parts with the remaining black plastic ring. For easy assembling one can lock the motor by the lock button. Figure 6: Hollow glass shaft with attachments 1 2 Figure 7: Adjustment of the hollow glass shaft at the motor 1 2 3

Put the metal rack for the motor on the plate as shown in figure 8. The motor is orientated over the pedestal to stand safely. Now put a Raschig tube on the glass hollow shaft to adjust the position of the motor. When the tube lies horizontally on all four rolls one can fix the clip at the backside of the motor with the four screws. Figure 8: Mounting of the motor Now fix the power supply and control box at the top of the metal rack with the screws (figure 9), and slide the power supply unit into the provided attachment. Then plug in the wire of the power supply and put the wires from the motor and the control box together. Figure 9: Control box with its wires From pump via flow meter to the gas meter the thinner tubing with 6 mm inner diameter is used. From the gas meter via wash bottle to the motor one has to use the thicker tubing with inner diameter of 8 mm. Use the hose clip to seal the transition from the pump to the hose (figure 2). For the first filling of the Raschig tube with NaOH solution and for taking first samples see below (Section 3B). 4

3. Sample Change Note: For the chemical absorption of atmospheric CO 2 we use highly concentrated sodium hydroxide solution. This strong alkaline solution destroys clothes and human skin. Please use gloves and safety glasses and protect your clothes. Be very careful in all handling procedures when changing the samples! If, by accident, it happens that some solution gets on the skin, first use a dry tissue to remove the droplet and then wash carefully with tap water. After two weeks of absorbing the CO 2 from about 25 m 3 air the sodium hydroxide has to be changed. First, the NaOH solution is being filled backed into the glass bottle and the same Raschig tube will be filled with new sodium hydroxide solution: A. Emptying the Raschig tube 1.) Switch off pump and motor at the multiple socket outlet. 2.) Fix the empty sample bottle with help of the white blue plastic fitting and the metal clip at the end of the Raschig tube 3.) Take off the Raschig tube from the hollow motor glass shaft. Attention: The tube can get off suddenly. Be careful not to lose NaOH solution from the tube. The Raschig tube with the fixed empty bottle now has to be put vertically on the plastic pedestal. Now the sodium hydroxide (carbonate )solution is running back into the glass bottle. Keep the tube for at least 5 minutes in this vertical position so that all the solution drains back into the bottle. Figure 10: Plastic stand with Raschig tube and fixed sample bottle 5

4.) Usually there are about 50 ml of solution remaining in the tube. Because of this fill about 50 ml distilled water from the top into the Raschig tube to dilute the remaining solution and some crystals. This solution drains into the bottle additionally. Be careful not to add too much water. 5.) Now remove the bottle, close it and label it with the following data: - name of the station & number of the sample - sample time start and end - gas meter start and end 6.) Remove the plastic fitting from the Raschig tube and clean it with water or diluted acid. B. Filling the Raschig tube with new NaOH 1. Carefully clean the hollow glass shaft sticking in the motor and the glass shafts of the Raschig tube with a dry tissue. Then grease the ground in stopper of the hollow glass shaft with the dosing syringe (figure 11). Good greasing is important to be able to remove the Raschig tube after two weeks of sampling. Then put the Raschig tube horizontally on the shaft and turn it slightly. Fix it with the black attachment (figure 9). Figure 11: Greasing of the hollow glass shaft Figure 12: Fixation of the Raschig tube 2. In this position use the funnel with plastic tubing at the outlet of the tube to refill it with the content of one new glass bottle of sodium hydroxide solution (200 ml). Immediately close the bottle. 3. Restart the motor and make sure that the Raschig tube is rotating on all four rolls exactly horizontally and slowly (about 40 revolutions per minute). Attention: Turn control knob just a little to the right! 4. Refill the wash bottle with distilled water (if distilled water is not available you can also use clean rain water or demineralized water; but it should be carbonate free). Grease the glass stopper and fix it with a metal clip. 5. Switch on pump and motor and check the flow rate at the flow meter. Flow rate can be regulated at the needle valve and should be ca. 75 l/h for a two week integration period. 6

6. Note the sample change in the log sheet with: sample No. start date & time gas meter start value flow rate Please also note special changes. 7. If possible please check the actual flow rate and the gas meter value and note it together with date and time regularly. If necessary, adjust the flow rate. If the air is very dry, the wash bottle should be refilled during one sampling interval so that the sodium hydroxide solution does not become too concentrated or clumped together. Please send the samples about every 2 3 months to the following address: Dr. Samuel Hammer Institut für Umweltphysik Universität Heidelberg Im Neuenheimer Feld 229 D 69120 Heidelberg Tel.: +49 6221 546357 Fax: +49 6221 546405 Email: Samuel.Hammer@iup.uni heidelberg.de 7