TP500 / TP500C Heat Therapy System STANDARD (NORME) C22.2 NO 125 RISK CLASS (CATEGORIE DE RISQUES) NO 2G OPERATOR'S MANUAL PIN /98

Similar documents
T/PUMP TP400 & TP401 Heat Therapy System

T/PUMP. Series Localized Temperature Therapy System TP600 & TP700. Service Manual. 2014/06 E REV E

INSTRUCTIONS FOR OPERATION AND CARE OF

Installation & Operating Guide

Installation & Operating Guide

T/Pump Series Localized Temperature Therapy System TP700 TP700C. Maintenance Manual.

SCHWAN INSTALLATION GUIDE

Instruction Manual Item No.: V~, 60Hz

SOUP COOKER/WARMER 120V, 240V MODEL 6423 INSTRUCTIONS. Installation Operation Maintenance

Quest PowerHeat HFC 100 Pro

Instruction Manual Item No.: V~, 60Hz

Instruction Manual. Bucket Oil Pump

OPERATOR S MANUAL & PARTS LIST

OPERATOR'S MANUAL. IMPORTANT: READ OPERATOR'S MANUAL CAREFULLY Please fill out & return your warranty card! DP80405

Navigator Upright Vacuum NV70 Series

INSTALLATION & OPERATION MANUAL

INSTALLATION & OPERATION MANUAL

Cool Flow & CoolOR PERSONAL COOLING SYSTEMS POLAR PRODUCTS polarproducts.com

INSTALLATION & OPERATION MANUAL

Owner s Manual Phoenix Aquadry TX 200

INSTALLATION & OPERATION MANUAL

WIDE AREA VACUUM MODEL PF2030 WARNING: OPERATOR MUST READ AND UNDERSTAND THIS MANUAL COMPLETELY BEFORE OPERATING THIS EQUIPMENT.

INSTANT HOT WATER DISPENSER

RTE-5/5B & RTE-9/9B Refrigerated Bath Circulators. NESLAB Manual P/N Installation, Operation and Service Manual

RCM-77. Instruction Manual. G-Series Cooler. Manual is for the following model: RCM-77-N23EB. U.S. Patent No. 8,215,125 RECHARGEABLE COLD MERCHANDISER

Bag-In-A-Box Oil Pump System

Installation Manual CARBONATOR With Plain-Water Booster

User Manual AIR HOG. Vacuum Booster REV. 3/20/ Stowe Dr. Poway, CA P: (858) F: (858)

Instruction Manual. Cheese Warmer with Pump

DELUXE ALTERNATING PRESSURE SYSTEM USER GUIDE

INSTALLATION & OPERATION MANUAL

INSTRUCTIONS FOR OPERATION AND CARE OF

RCM-77. Instruction Manual. G-Series Cooler. U.S. Patent No. 8,215,125 RECHARGE COLD MERCHANDISER

Operating Instructions

WET/DRY VACUUM. QUEST for Continuous Improvement Windsor s Quality Management System is Certified ISO MODEL: T1. Operating Instructions (ENG)

INSTALLATION & OPERATION MANUAL

WIRING DIAGRAM. This manual describes the unit features and explains how to set-up, operate and maintain these AirPac Models. COOLIT2600 COOLIT2900

OSD10, OSD20 & OSD30 DRINK DISPENSER

INSTALLATION & OPERATION MANUAL

STEAM MOP INSTRUCTION MANUAL CATALOG NUMBER BDH1725SM PLEASE READ BEFORE RETURNING THIS PRODUCT FOR ANY REASON. SAVE THIS MANUAL FOR FUTURE REFERENCE.

π H-6621 INDUSTRIAL DEHUMIDIFIER WARNINGS SPECIFICATIONS uline.com WATER REMOVAL ELECTRICAL REQUIREMENTS BUILT-IN ELECTRICAL SAFETY

SAFETY PRECAUTIONS. 2) Before operating machine: - Make sure all safety devices are in place and operate properly.

SOLUS-500 EXTRACTOR 120V

OWNER'S MANUAL. FRY Holding Unit (FHU) MODELS FHU1 FHU2. Duke Manufacturing Company

Instruction Manual. Double Candy Apple Cooker

5300, 5300CH, 5300EX, 5300FN

INSTALLATION & OPERATION MANUAL

SCHWAN INSTALLATION GUIDE

Model 7501/7501T/7501TH/7501T240/7501T240H Portable Eyewash/Portable Eyewash with Heater

CAUTION: DO NOT OPERATE MACHINE UNTIL YOU HAVE READ ALL SECTIONS OF THIS INSTRUCTION MANUAL

8070 Mytee Lite. Instructions for. Please read before use. Register your product at support/register.

Nexus FS Filtration Installation Guide. September 19, 2011

OWNER S MANUAL & WARRANTY

Use & Care Manual. Electric Tankless Water Heaters. With Installation Instructions for the Installer AP15447 (10/10)

STEAM MOP INSTRUCTION MANUAL CATALOG NUMBER BDH1765SM PLEASE READ BEFORE RETURNING THIS PRODUCT FOR ANY REASON. SAVE THIS MANUAL FOR FUTURE REFERENCE.

Free Standing Water Cooler

As part of the V.I.P. family, you are entitled to the best protection by one of the most comprehensive warranties in the industry.

INSTRUCTION MANUAL BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING INSTRUCTIONS.

Safety, Operation and Maintenance Manual with Parts List

Mobile Evaporative Cooler. Use & Care Guide MC18M INDEX

PUREPOWER SERIES CENTRAL VACUUM POWER UNITS PP500, PP600 & PP650

AquaStream. ½ HP Floating Fountain. Please Keep for Your Records. Model #AS05 6A, 120V/60Hz Owner s Manual UNIT IDENTIFICATION

CFC-FREE CB-60/CB-80 Cryobath Low Temperature Bath. Thermo NESLAB Manual P/N U00551 Rev. 04/03/00. Instruction and Operation Manual

Model HC-2. Crathco Whipped Hot Chocolate Dispenser. Table of Contents. Operation and Instruction Manual. for

One-Touch Dispense. Multi-temp selections. 208 F For tea, coffee, instant noodle. 194 F Keep warm around 194 F. 176 F Keep warm around 176 F.

Water Boilers ME10EN, ME15EN. Table of Contents

SM088 STEAM MOP INSTRUCTION MANUAL. For Authentic Monster Replacement Parts Call Model #: SM088 G-B-R

B.I.C.A Built-In Coffee Appliance

Electronic Service Manuals

WARMING AND MERCHANDISING CABINET

17 GALLON PROFESSIONAL RESTROOM CLEANER

Industrial Dehumidifier Operating Instructions. Model: VG 1500

Professional Warming Drawer and Warming Drawer Cabinet

K Specifications. Max. Water Volume

CONDENSATE PUMP INSTALLATION MANUAL OFFICE PRO 10

DISHWASHER INSTALLATION GUIDE SPECIFICATIONS, INSTALLATION, AND MORE

NT1-P CONDENSATE NEUTRALIZATION TANK WITH PUMP INSTALLATION, OPERATION, AND MAINTENANCE INSTRUCTIONS

BEFORE OPERATING THE MACHINE: WARNING

Outdoor Beverage Center

MODEL 7000 SUCTION UNIT

Installation. 324 Series Built-In Dishwashers U L. Viking Range, LLC 111 Front Street Greenwood, Mississippi USA (662)

REFRIGERATED PREP TABLES Installation, Operation and Maintenance Instructions

Table of Contents SPECIFICATIONS Page 2. FOR THE INSTALLER Page 3. FOR THE PLUMBER Page 4. INSTALLATION Page 5. INITIAL START UP Page 6

Operator s Manual. 2.5 Gallon Model No. VOM205P FOR YOUR SAFETY. Read and understand this manual before use. Keep this manual for future reference.

ELECTRIC STOVE MODEL CGESS SAFETY INFORMATION AND OPERATION MANUAL

Operator s Manual. Wet/Dry Vacuum with Detachable Blower 12 Gallon Model No. VBV1210 FOR YOUR SAFETY

Wagner Product Training

SolidFlo G2 Solid Handling Pumps Product Manual

Operation and Maintenance Manual

RecoverX Oil-Filled Hermetic Refrigerant Recovery System. Operation and Maintenance Manual

Operations and Safety Manual

6 Oz. Twin Pop. Popcorn Machine Instruction Manual Model #2666EX. Cincinnati, OH USA. Part No EX Revised June 1996

I N S T R U C T I O N S

AllClear G2 Series Pressurized Filters Product Manual

DANGER Any appliance is electrically READ ALL INSTRUCTIONS BEFORE USING

INSTRUCTIONS FOR OPERATION AND CARE OF

Flood Pumper Plus. Water Transfer Machine. Operator and Parts Manual. Model No.: Rev. 00 (11-99)

G-10s. Instruction Manual. G-Series Cooler UPRIGHT COOLER. Part No.11IPA

OWNER S MANUAL MODELS AP8, AP8B, AP8C, AP8D DUMP STATION HEATER

IMPORTANT SAFETY INSTRUCTIONS EC-AG1-25 EC-AG1, EC-AG2 SAVE THESE INSTRUCTIONS.

Transcription:

TP500 / TP500C Heat Therapy System @ LISTED 304L STANDARD (NORME) C22.2 NO 125 RISK CLASS (CATEGORIE DE RISQUES) NO 2G OPERATOR'S MANUAL PIN 10569-000 12/98

OPERATOR'S MANUAL TP500/TP500C TIPUMP CONTENTS Safety Precautions... I Introduction... 2 FeatureslSpecifications... 3 Operating Instructions... 5 StoragelCleaning... 7 Troubleshooting... 8 BEFORE YOU BEGIN... Read and understand this TIPUMP OPERATOR'S MANUAL and all PRECAUTlONS (see page I) prior to using the TIPump. RECEIVING INSPECTION Check the shipping carton for damage immediately upon receipt. If package damage is discovered, the device should be unpacked with the carrier's agent present. Any claims for shortage or damage must be filed with the delivering carrier by the purchaser. D o not return pumps damaged in shipment to GAYMAR without contacting our Technical Service Department for advice (see phone numbers below). If damaged goods are returned to GAYMAR without notifying the carrier, GAYMAR will assume the repairs will be made at the customer's expense. Should you desire, repairs can be made in your facility by your own qualified personnel. Contact the GAYMAR Customer Service Department to request a TIPUMP SERVICE MANUAL. Call GAYMAR for spare parts required. TIPUMP, TIPAD, Mul-T-Pad, Cl~k-Tite, and Gayrnar are registered trademarks of Gaymar Indusrr~es. Inc. U. S. PATENT 4.068.870 O 1998. Gaymar Industries. Inc. All rights reserved.

OPERATOR'S MANUAL TP5001TP500C TIPUMP SAFEeTY PRECAUTIONS SAFETY PRECAUTIONS Risk of explosion. Do not use in the presence of flammable anesthetics. Risk of electric shock. Disconnect power before servicing the TIPump. This device pumps warmed water through a pad. Set pad temperature only as prescribed by and under the guidance of a physician. Monitor the patient's temperature and skin condition every 20 minutes or as directed by a physician. Failure to adhere to these warnings could result in patient injury. The following GroupslConditions require additional surveillance: Pediatric patients GroupICondition at risk Potential injury Hyperthermial hypothermia Patients with impaired circulation Areas of application are under pressure Ischemia lschem~a In combination with topical solutions whose toxicity Chemical may be affected by the application of heat injury In combination with other heat sources Thermal injury Only qualified medical service personnel should repair the TIPump. lmproper repair may result in death or serious injury, equipment damage, or malfunction. Always perform the FUNCTIONAL CHECK AND SAFETY INSPECTION as specified in the SERVICE MANUAL after making repairs and before returning the TIPump to patient use. lmproper repair may result in death or serious injury, equipment damage, or malfunction.

INTRODUCTION OPERATOR'S MANUAL TP5001TP500C TiPUMP Single pod Multiple pods (TP500 only) Figure I-TIPurnp Heat Therapy System INTRODUCTION Heat therapy is effective in the dilation of blood vessels, thereby increasing the blood flow to the heated area. Heat therapy has a variety of uses, the most common being treatment of aches and pains in joincs and muscles. The GAYMAR TIPumpz Heat Therapy System provides a means of applying heat therapy by supplying temperature-controlled water through a connector hose to a Gaymar T/PadB. The hose is terminated in easy-to-use Clik-TiteB or Colder-style connectors. The TIPad provides the interface for delivering the heat therapy. The unique button design allows water to flow and provides trouble-free operation when the pad is folded. This reduces the number of pads your facility must keep in inventory. The pads are applied to the part of the body requiring heat therapy, and the circulating water maintains the pad at the set point temperature. The temperature set point is key-operated to prevent tampering. The TIPads can be interconnected (on model TP500 only) to provide therapy to more than one body site at a time.

OPERATOR'S MANUAL TP5001TP500C TIPUM P FEATURES CordIHose Strap Serial Number \ Vented. Leak~roof Carrv Handle (pullout-proof) lnktructi&s \\ ted Switch Power Cord w~th Molded Hosp~tal Grade A\ \\ Tw~n, 10' long Leakproof Pump Durable Thermoplastic Case Circuit Breaker Extra Capacity Reservoir 1500 ml (51 oz) (see f~gures 3 & 4) TEMPERATURE / '.A' 1 1. : Operating Water Level Indicator (key-operated) / / \ Nonrnarking Suction Cups TEMPERATURE SETTING Key Dual Temperature Set Point Scale ("CPF) \ OVER TEMP Light Figure 2-TIPump FeaturesISpecifications FEATURES Attached Hose Tip-over Switch Hospital ID Label I0 ft (305 cm) dual hose. Connectors allow pads to be connected to the pump (see figs. 3 and 4, p. 5). Turns heater off if pump is tipped. NOTE: This does not activate the OVER TEMP light. A label is provided on the back of the TlPump for your convenience. Use a felt tip or ball point pen to add any required hospital identification. 0 VER TEMP Indicates the pump and heater have been turned off. Light This light is activated by the two OVER TEMP safety thermostats. Refer to the TROUBLESHOOTING section. Circuit Breaker OVER TEMP Safety Thermostats When the circuit breaker trips, the small button in the center will extend l/4", exposing a white band. Refer to the TROUBLESHOOTING section. Either of two limit thermostats will shut off pump and heater if the high temperature limit is exceeded. The OVER TEMP light will glow. Refer to'the TROUBLESHOOTING section.

SPEClFlCATlONS OPERATOR'S MANUAL TP500lTP500C TIPUMP SPEClFlCATlONS Size (approx.) a-1~8'' S-~/E" 6-'14'' (20.6 cm x 14.3 cm x 15.9 cm) Weight (empty) 5 Ib 2 oz (2.3 kg) Reservoir capacity 5 1 oz (1 500 ml) maximum Flow rate Ambient operating temperatures Storage temperatures (empty) 9 gph (34 Iph) minimum with pad attached 60 F to 90 F (15.6"C to 32.2"C) -30 F to 160 F (-34.4"C to 7 1. I "C) Temperature set point 85 F to 107 F (29.4"C to 4 1.7"C) range Average operating temperature accuracy Power cord Circuit breaker Current leakage Ground resistance +2"F at 107 F setting 18 AWG, 3 conductor, 9 ft (274 cm), Type SJT with molded, hospital grade plug 3 amperes I CO microamperes maximum 0.5 ohm maximum Electrical requirements Voltage (VAC) 1 20 Frequency (Hz) 60 Current (amps) 1.8 Power req'd (watts) 200 Safety approvals Tested to UL 544 and CSA C22.2, No. 125

OPERATOR'S MANUAL TP5001TP500C TIPUMP OPERATING INSTRUCTIONS TIPumps are supplied with one of two types of hose connectors: TP500 TIPumps have Clik-Titem connectors; TP500C TIPumps have Colder-style connectors. CLIK-TIT @ CONNECTORS (on TP500): To attach Clik-Tite connectors from hose to pad: I. Insert male fittings into female fittings with a twisting motion (figs. 3A and 3B). 2. When fittings are fully inserted, snap locking ring into place (figs. 3C and 3D). 3. To disconnect, simply reverse the procedure. NOTE: Refer to figure I (p. 2) and item 9 (p. 6) when connecting multiple pads. Figure 3-Clik-Titem Connectors COLDER-STYLE CONNECTORS (on TPSOOC): I. To attach Colder-style connectors to a pad, push the male coupling onto the female coupling. When you hear an audible "click", the connectors are joined (fig. 4A). 2. To open or close the hose pinch clamps: Open the clamp by pushing the serrated end (fig. 4B). Close the clamp by pressing the clamp together (fig. 4C). 3. To disconnect Colder-style connectors, press down on the thumb tab of the female coupling. The couplings will partially disconnect. Pull the male coupling out fully to disconnect (fig. 4D). Figure +Colder-style Connectors

OPERATING INSTRUCTIONS OPERATOR'S MANUAL TP500lTP500C TIPUMP START-UP PROCEDURE I. Before filling the pump, always attach a pad to the connector hose or close the clamps on the connector hose ends (see figs. 3 and 4, p. 5). Then, unkink pad and hose. Open hose clamps. 2. Open the fill cap on top of the pump. Fill the pump with room temperature (i.e., not hot) distilled water to the operating level indicated on the side of the pump. NOTE: Using tap water will decrease the life of the pump. 3. Insert temperature setting key. Set temperature as prescribed by the physician. Remove key to prevent tampering. 4. Plug TIPump into a properly grounded outlet. 5. Turn on the power switch. The selected water temperature will be reached in approximately 20 minutes. 6. If the water level drops below the operating level, add water. Do not overfill. Overfilling can result in reduced pump motor life. 7. Apply TJPad to the patient. Follow TJPad instructions. 8. For specified performance, keep the pump at or above the level of the pad. 9. If the pump is placed below the pad(s), water will drain into the pump when it is shut off. If the pump has been overfilled or if multiple pads are connected, excess water can leak out onto the floor or cause reduced motor life. SHUTDOWN PROCEDURE I. Turn off pump before disconnecting pad. Close all hose clamps. To prevent water spillage, always disconnect pad from pump with connectors raised above the level of the pad and pump. 2. Connect the pump and TJPad Clik-Tite connectors together, where applicable.

OPERATOR'S MANUAL TP5001TP500C TIPUMP STORAGE I CLEANING Storage (Short term) Disconnect pad. Connect ends of the connector hoses together, where applicable. Open hose clamps. Leave water in the reservoir. To prevent hose kinks, coil the hose rather than folding it. Fasten the hose and cord with strap. Storage (Long term) Connect pad. Open hose clamps. Add 114 ounce AirKem A-33, GAYMAR MTA33 germicidal, or equivalent to water already in reservoir. Run for two (2) minutes. Drain pump. Close hose clamps. Disconnect pad. To prevent hose kinks, coil the hose rather than folding it. Fasten the hose and cord with strap and store pump. Draining Disconnect the TIPump from AC power. Disconnect the pad or hoses from one another, keeping hoses at or above the level of the TlPump. Remove the fill cap and invert the TlPump over a sink. When all fluid has drained from the hoses and reservoir, replace the fill cap and connect the hoses together, where applicable. Cleaning Clean outer surfaces of the TIPump with: a damp cloth and soapy water, Fantastik spray cleaner, or a mild abrasive such as Soft Scrub cleanser without bleach. To clean the fluid system, prepare a germicidal solution according to the manufacturer's instructions. Use AirKem A-33 or equivalent, available from Ecolabs, Inc., 370 Wabasha, St. Paul, MN 35 102 (phone: 1 800 247-5362), or from GAYMAR, product catalog MTA33. Drain the pump. Fill the reservoir to the operating level indicated on the side of the pump. Set the temperature indicator to its lowest setting (fully counterclockwise), start the TIPump, and circulate the solution for one hour. D.rain the solution and refill the pump with distilled water. Using distilled water will not promote algae growth or mineral buildup. Change the distilled water monthly or more often depending upon use. Pads I Accessories For best results use only GAYMAR T/Padsm or Mul.T*Pads@, The unique button design allows water to flow and provides trouble free operation when the pad is folded. This reduces the number of different sizes of pads your facility must keep in inventory. The TIPads can be interconnected (on model TP500 only) to provide therapy to more than one body site at a time (see fig. I, p. 2). For a brochure listing the various pads, contact the GAYMAR Customer Service Department (see next page for telephone numbers). An optional bed bracket (model TP20A) is available to mount the TlPump on the footboard of a bed.

TROUBLESHOOTING I MAINTENANCE OPERATOR'S MANUAL TP5001 TP500C TIPUM P PROBLEM "OVER TEMP" light is on. TlPump will not heat. I TIPurnp will not pump. TlPump will not turn on - - Water leaks from connector: I Clik-Tite connector 1. Pad or hose is kinked. Hose clamps are closed. Turn switch off, unplug pump, and dump water out. 2. Water level is low, or reservoir is empty. Unkink pad or hose. Open hose clamps. Refill with room temperature distilled water to proper level. Plug TIPump in, 3. TlPump is filled with water that is too hot turn switch on, and wait 40 minutes (it must be below 75 F). I 1. TlPump is tipped. POSSIBLE CAUSE 2. "OVER TEMP" lrght is on. 1. Water level rs low, or reservoir is empty. 2. "OVER TEMP" lrght is on. 1. The "ON" switch may not be fully engaged REMEDY Place TIPump on level surface. Refer to "OVER TEMP' section above. Refill with room temperature distilled water to proper level. Refer to "OVER TEMP" section above. The light on the switch should be Irt. Verify that the switch is fully engaged in the "ON" position. 2. The electrical cord is not plugged Into a grounded Insert the plug fully into the receptacle electrical receptacle. 3. The c~rcuit breaker has tripped 1. Damaged 0 ring. 2. Locking ring on Clik-Tite connector is not snapped into place (see fig. 3). If the circuit breaker has tripped, turn the power switch to "OFF". Wait two (2) minutes. Press the circuit breaker button in. Turn TIPump on. If the circuit breaker trips again, remove TlPump from use. If defective, replace Clik-Tite connector. Snap Clik-Tite locking ring shut. Colder-style connector 3. The couplings are not securely connected. Push the male and female ends together until you hear a "click." If it still leaks, replace connector. PREVEN JIVE MAINTENANCE I SERVICE I I Risk of electric shock. Disconnect power before servicing. I Only qualified medical service personnel should repair the TIPump. Improper repair may result in death or serious injury, equipment damage, or malfunction.