Chapter 7 Checks and Maintenance

Similar documents
Chapter 5 Checks and Maintenance

Chapter 5 Inspection and maintenance

Daily maintenance and checks

Chapter 5 Inspection and maintenance

Chapter 8 Troubleshooting

Chapter 8 Troubleshooting

Dehumidifier. Instruction Manual. Model DH-320/A

Owner s Manual CMC100 Dehumidifier

Owner s Manual CMC100 Dehumidifier

INDOOR UNIT/OUTDOOR UNIT

WINDOW TYPE ROOM AIR CONDITIONER OPERATION AND INSTALLATION MANUAL

Chapter 7 Troubleshooting

KSD-35 DR11 KUE-35 DVN11

INDOOR UNIT/OUTDOOR UNIT

I N S T R U C T I O N S

SYSTEM Inverter Air Conditioners

INDOOR UNIT/OUTDOOR UNIT

DEHUMIDIFIER USER MANUAL 11 PINT FOR MODEL: 3PAD11 COMFORT...BUILT TO LAST

Owner s Manual Super-Slim Four-Way Cassette

Maintenance 50 Serie CAUTION. Before resetting your electronic card that displays an error code. OVERHEAT MESSAGE

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

Owner s Manual. Middle Static Pressure Duct Type MEU-18MPH2 MEU-24MPH2 MEU-36MPL2 MEU-48MPL2 MIDDLE STATIC PRESSURE DUCT TYPE AIR CONDITIONER

Portable Air-conditioner Use and Care Manual

ANTIBACTERIAL HAND VAC

OPERATING INSTRUCTIONS

GCG-10. Instruction Manual. G-Series Cooler. Manual is for the following models: GCG-10-N33EB G-10-N33EB UPRIGHT COOLER

Air Conditioner. Instruction Manual. For Models WA-1010E, WA-1010H, WA-1010M WA-1230E, WA-1230H P O R T A B L E

Portable Air-conditioner

English. Important Read this user manual carefully before you use the appliance and save it for future reference.

DH 100 P. Always Dry dehumidifier. with. Condensation Removal System

READ MANUAL BEFORE OPERATING SYSTEM Read the owner s manual thoroughly before operating to ensure the most efficient use of the system.

INSTALLATION MANUAL. Split-type Air Conditioner (Cooling and Heating) Outdoor Unit UQB09JJWC UQB12JJWC. Indoor Unit AQB09JJWC AQB12JJWC

Window-Type Air Conditioner

INDOOR UNIT/OUTDOOR UNIT

TCB-DP22CUL. Drain Pump Kit. Model name : AIR CONDITIONER (SPLIT TYPE) Installation Manual. Installation Manual 2 Air Conditioner (Split type) English

OPERATING INSTRUCTIONS

SPLIT TYPE. Instruction manual

SYSTEM Inverter Air Conditioners

INDOOR UNIT/OUTDOOR UNIT

Series of all-purpose physical and chemical environmental cabinets meeting the diverse needs of laboratories and research facilities.

PUREPOWER SERIES CENTRAL VACUUM POWER UNITS PP500, PP600 & PP650

AIR CONDITIONER. Please read this owner, s manual carefully before using the air conditioner, and keep this manual for the future reference.

MODEL MC-UL592 MC-UL594 POWER SOURCE. 230V - 240V ~ 50 Hz MAX INPUT 1400 W 1500 W NOMINAL INPUT DIMENSIONS (W x L x H)

INDOOR UNIT/OUTDOOR UNIT

INDEX. Granita Machine. Operation a nd Instruction Manual

Full Size Canister Service Manual Riccar Models 1700 / 1800 Power Nozzles RPB-100 / RPB-220 / RPB-224 / RPB-250

HOW TO OPERATE THE REFRIGERATOR COMPARTMENT

DOUBLE DOOR COMPACT REFRIGERATOR. User Manual MODEL:KSTRC312BW KSTRC312BB

SYSTEM Inverter Air Conditioners

EVAC Commercial Marine Equipment TOILET TECHNICAL DATA EVAC 90, SQUATTING TOILET. To main vacuum line, connection Ø50.

RCM-77. Instruction Manual. G-Series Cooler. Manual is for the following model: RCM-77-N23EB. U.S. Patent No. 8,215,125 RECHARGEABLE COLD MERCHANDISER

TECHNICAL INFORMATION Touchtronic Clothes Dryers

USER S MANUAL. Bottom-Load Water Cooler MODEL NO. CLBL420

ANTIBACTERIAL HAND VAC WITH UV LIGHT

Portable Room Air Conditioner and Heat Pump

RCM-77. Instruction Manual. G-Series Cooler. U.S. Patent No. 8,215,125 RECHARGE COLD MERCHANDISER

Due to continual improvement in design or otherwise, the product you purchase may differ slightly from the illustration in this book.

Owner s Manual CPU-1QD CPU-1 CPU-1T CPU-2 CPU-2T HEAVY DUTY UPRIGHT VACUUMS. Save These Instructions

Installation Instructions

USER MANUAL PORTABLE AIR-CONDITIONER 2011 FE A4320A BR14081R-1

OPERATING INSTRUCTIONS

AquaReef Cube. Inspiring Excellence in Fish Keeping!

WINDOW TYPE COOLING MODEL AKT7F AKT9F REVERSE CYCLE MODEL AKT7U AKT9U FUJITSU GENERAL LIMITED

15,000 BTU Portable Air Conditioner

INSTALLATION MANUAL. Split-type Air Conditioner (Cooling and Heating) Indoor Unit AQB18J6WC AQB24J2WC. Outdoor Unit UQB18J6WC UQB24J2WC

GMSG OWNER S MANUAL SPLIT-SYSTEM ROOM AIR CONDITIONER CONTENTS

Portable Air Conditioner Instruction Manual

AH-PNX11 AU-PNX11 AH-PNX14 AU-PNX14 EN-1 SPLIT TYPE ROOM AIR CONDITIONER OPERATION MANUAL ENGLISH

WARNING USER S MANUAL. Model iq 2000 Series Dust Collector with manual shaker

GM 100 Sprint Plus GM 100 Sprint Plus Pet Pack INSTRUCTIONS FOR USE

DUCTED AIR CONDITIONER. Owner s Manual. KD Series KD24. Kaden Owner s Manual 1

WINDOW TYPE ROOM AIR CONDITIONER OPERATION AND INSTALLATION MANUAL

STORM. Bottom-Load. Water Dispenser User Manual

Operating Instructions. Accessory Units Melitta Cafina XT Series. Melitta Professional Coffee Solutions

OPERATING INSTRUCTIONS

Split Air Conditioner

I N ST R UC T I ON MODEL 2612 & 2712 SLICERS MODELS 2612 ML ML FORM (12-98) 2612 SLICER

Owner s Manual. For Fuller Brush Tidy Maid Vacuum. Model FBTM-PW

Air Conditioner Features...8 Operating Instruction Care and Cleaning...17 Before You Call...18 Major Appliance Limited Warranty...

Refrigerator. Installation Guide. Free Standing Appliance

ICED CAPPUCCINO MACHINE Taylor MODEL: 390

INSTALLATION MANUAL COMFORT...BUILT TO LAST. 9,000, 12,000 and 18,000 BTU SINGLE-ZONE DUCTLESS MINI-SPLIT SYSTEM Heat Pump

Technical Data TYPE T14 & T14D TEMPERATURE PILOT SPENCE ENGINEERING COMPANY, INC. 150 COLDENHAM ROAD, WALDEN, NY SD 4511A T14 PILOT

M650-38A-OAK ELECTRIC FIREPLACE HEATER NDF-65NB-1 INSTRUCTION MANUAL

ULTRA LOW TEMPERATURE FREEZER. User Manual

Free Standing Water Cooler

Service Manual & Installation Manual

OPERATING MANUAL AIR CONDITIONER DUCT TYPE. ART Series KEEP THIS MANUAL FOR FUTURE REFERENCE FUJITSU GENERAL LIMITED P/N

Owner's Manual TABLE OF CONTENTS

T1 SCRUBBER DRYER OPERATOR MANUAL

INVERTER SPLIT - TYPE

OWNER S MANUAL. R 410A Ductless Split System Air Conditioner and Heat Pump

PORTABLE AIR CONDITIONER

How to Use and Care for Your Sleek

CASSETTE TYPE. AUT Series FUJITSU GENERAL LIMITED AIR CONDITIONER OPERATING MANUAL P/N KEEP THIS MANUAL FOR FUTURE REFERENCE AM PM

SuperKlean Washdown Products

INSTALLATION MANUAL. Decoration Panel Provided with Filter Auto Cleaning Function Self Cleaning decoration panel BYCQ140D7GW1 BYCQ140D7GFW1

Room Air Conditioner Service and Parts Manual

Reliability is a beautiful thing TM CUBELET ICE DISPENSER DCM-500BAF DCM-500BWF DCM-750BAF DCM-750BWF INSTRUCTION MANUAL

THROUGH-WALL AIR-TO-AIR HEAT PUMP AND AIR CONDITIONER. Instruction Manual. Model AMB-12H

Transcription:

Chapter 7 Checks and Maintenance This chapter explains equipment checks and maintenance. To keep the chamber in good working condition, perform checks and maintenance periodically. 97

7.1 Check and Maintenance Lists Check list For an explanation on each check item, see "7.2 Checks". If any of the following checks result bad, contact the place of purchase or ESPEC CORP. Check item Table 7.1 Check list When to check Main power switch (leakage breaker) trip test Overheat protector trip test Water level check for humidifying tray and humidifying tray water level regulator Once monthly Before long test runs Before every test Once every 3 months After moving chamber (See "4.5 Water Level Check".) Maintenance list For an explanation on each maintenance item, see "7.3 Maintenance". Table 7.2 Maintenance list Maintenance item When to perform Condenser filter cleaning Water tank (auxiliary and main) and tank filter cleaning (Not necessary with temperature-only chambers) Water supply pump filter cleaning (Not necessary with temperature-only chambers) Humidifying tray cleaning Test area cleaning Electric parts compartment and water circuit (heat exhaust) compartment cleaning Take-down before long periods of disuse Water supply pump filling (Not necessary with temperature-only chambers) Once monthly Once monthly Once monthly Once monthly Before every test Once yearly Before long periods of disuse Prior to use after long periods of disuse 98

7.2 Checks Main Power Switch (Leakage Breaker) Trip Test Once a month and before long test runs, test-trip the main power switch (leakage breaker). With the main power switch in the ON position, press the test button. If the switch's lever falls to the middle position, the breaker is working properly. Test button Test button 200, 220V AC spec. 380V AC spec. Fig. 7.1 Test button Reference When the leakage breaker in the main power switch trips, the switch's lever falls halfway between the ON and OFF positions. To turn power back ON again, first set the switch to the OFF position and then to the ON position. Overheat Protector Trip Test Before every test, test-trip the overheat protector. 1. Check the main power switch is in the ON position. 2. Press the POWER key on the operating panel. When the display comes on, the current temperature (& humidity) inside the chamber will appear on it. 3. Press the OPER./STOP key. The chamber will start up. 99

4. Set the overheat protector about 5 C lower than chamber temperature. If the overheat protector is working properly, a buzzer will sound and the Alarm screen will be appear on the display when temperature inside the chamber reaches the overheat protector setting. All digits in the display of the overheat protector will flash. If a buzzer is not sounded, something is wrong with the equipment. Contact the place of purchase or ESPEC CORP. Display key key Fig. 7.2 Overheat Protector 5. To silence the buzzer, press the BUZZER key. Return the overheat protector to its usual setting. 100

7.3 Maintenance Condenser Filter Cleaning Clean the condenser filter so as to prevent dust from adhering to the condenser. 1. Remove the condenser filter. With chamber types 1 to 3, pull the filter out from the left side of the chamber as shown below. On type 4 chambers, there are two filters, both of which are located to the backside of the water tank drawer door. Pull both of them out. Condenser filter Fig. 7.3 How to remove the condenser filter (Types 1 to 3) 2. Rinse the filter clean of dust and dirt with plain water. 3. Drip off water and dry in a dark place. The condenser filter is made of plastic. Direct exposure to sunlight can deform it. 4. Reinstall the filter as before. 101

Water Tank and Tank Filter Cleaning (Not necessary with temperature-only chambers) Water can stagnate if left inside the water tank for 1 month or longer. Stagnant water will shorten the service-life of the humidifying heater and wet-bulb wick. Therefore, clean the water tank and tank filter once a month. There are two water tanks; the main tank is stationary while the auxiliary tank is portable. The main water tank has a filter and frame on the water supply port. To remove the filter, hold it by the projecting stub and jiggle it in all directions while lifting upward. Auxiliary water tank (Portable) Tank filter Filter frame Main water tank (Stationary) Fig. 7.4 Water tanks and filter CAUTION Return all parts to their original positions after cleaning. Improperly installed parts or missing parts can cause leaks or disable water supply. Snap the bottom edges of the filter frame onto the ledges of the water supply port on the main water tank. Filter frame Main water tank Snap onto ledges on all sides. Fig. 7.5 Filter frame Press the filter fully to the bottom of the filter frame. Improper pressed filter can cause filter cover damage or disable water supply. 102

To clean the auxiliary water tank Chapter 7 Checks and Maintenance 1. Remove the auxiliary water tank from the water tank drawer. 2. Open the lid and add about 2 L of water. 3. Shake the tank in all directions to clean the inside. 4. Press the valve on the bottom of the tank and drain the water. Valve Fig. 7.6 Draining the auxiliary water tank 5. Repeat steps 2 to 4 two or three times. 6. Return the auxiliary water tank to its original position. To clean the main water tank CAUTION Stop the chamber before cleaning the main water tank. Also, supply water to the humidifying tray after cleaning the tank. Water supply automatically purges the water circuit of air which serves to properly maintains water supply rate. 1. Remove the auxiliary water tank from the water tank drawer, and remove the tank filter and filter frame from the main water tank. 2. Clean the main tank on the inside with a brush or cloth. 3. Add some water to the main tank. 103

4. Free the hose from the hose clip, disconnect it at the coupling and drain the tank. The hose is located toward the rear of the water tank drawer. Plug side Hose Hold here and pull downward. Socket side Slide downward to free hose. Water tank drawer Hose clip Fig. 7.7 Disconnecting the hose 5. Blot up any excess water from inside the main tank with a piece of cloth. 6. Reinstall the filter frame, tank filter and the auxiliary water tank as before. To clean the tank filter 1. Remove the auxiliary water tank from the main water tank, and remove the tank filter. 2. Rinse the filter clean of dirt with plain water. 3. Reinstall the tank filter and the auxiliary water tank as before. 104

Water Supply Pump Filter Cleaning (Not necessary with temperature-only chambers) The water supply pump is fitted with a filter so as to prevent dirt from infiltrating the wick pan and humidifying tray water supply pumps. Clean the filter once a month. Tank filter Filter frame Pipe Filter Main water tank (Stationary) Fig. 7.8 Water supply pump filter Water tank drawer CAUTION Stop the chamber before cleaning the water supply pump filter. Also, after cleaning, supply water to the humidifying tray so as to purge the supply line of air. This air purge helps keep water in the tray to the proper level. 1. Check the main power switch is in the OFF position. 2. Remove the auxiliary (portable) water tank from the water tank drawer. 3. Detach the tank filter. 4. Pull the filter off the pipe inside the stationary water tank and remove. (If there is water still in the water tank, drain the tank as explained in step 4 of To clean the main water tank before detaching the filter.) 5. Rinse the filter clean of dirt with plain water. 6. Reassemble parts as before. 7. Supply water to the humidifying tray. (Run a temperature-humidity test.) 105

Humidifying Tray Cleaning CAUTION SHARP EDGES! Be careful of protrusions or sharp edges inside the chamber. For your safety, always wear gloves when working inside the chamber. HOT ON THE INSIDE! Cool down the test area before cleaning the humidifying tray. During and shortly after operation above 55 C, the chamber is HOT and HUMID on the inside (specimens, shelves, door gasket, test area walls). During operation, dirt and foreign matter stick to the humidifying tray and humidifying heater. To ensure long lasting use, clean the tray and heater of dirt once a month. A good way to inhibit dirt from accumulating is to drain the humidifying tray after every test. 1. Open the chamber door. 2. Lift the protective grille outwards from the bottom and then upwards to detach. 2 1 Fig. 7.9 How to detach the protective grille 3. Clean the surfaces of the humidifying tray and humidifying heater with a brush or other means. 4. Reattach the grille and close the chamber door. 106

Test Area Cleaning Dirt and foreign matter inside the test area can throw test results off. Clean the test area before every test. 1. Open the chamber door. 2. Wipe walls and parts clean with a soft cloth. 3. Close the door. Electric Parts Compartment and Water Circuit (Heat Exhaust) Compartment Cleaning Dust buildup inside the electric parts compartment or the water circuit (heat exhaust) compartment can lead to trouble. Clean inside both compartments once a year. 1. Check the main power switch is in the OFF position. 2. Detach the electric parts compartment and water circuit (heat exhaust) compartment doors. 3. Clean dirt from both compartments with a vacuum cleaner or by other means. 4. Reattach the compartment doors. Take-Down Before Long Periods of Disuse Before long periods of disuse, do the following. Failure to do so can affect testing and shorten equipment service-life. Replace the water in the water circuit. (Not necessary with temperature-only chambers) Drain the wick pan and humidifying tray of water. (Not necessary with temperature-only chambers) Dry the test area (run the chamber). Set the main power switch in the OFF position and shut OFF primary power supply. 107

To replace the water in the water circuit (Not necessary with temperature-only chambers) Set test area temperature and humidity to 70 C and 90%rh and run the chamber in the constant mode for approx. 10 min. 1. Check the main power switch is in the ON position. 2. Set the constant mode with temperature at 70 C and humidity at 90%rh. 3. Press the OPER./STOP key to start the chamber. 4. Run the chamber for about 10 min with the door closed. To drain water (Not necessary with temperature-only chambers) 1. Check the main power switch is in the ON position. 2. Press the POWER key to activate control power. 3. Hold the DRAIN switch in the "MANUAL" position for about 2 seconds. Water will drain from the humidifying tray and wick pan. To dry the test area The chamber is run to dry the test area. Turn the refrigerator and humidity control OFF, and run the chamber at a minimum 70 C for about 60 minutes. After that, crack the chamber door slightly and run the chamber under the same conditions for about 15 minutes. As necessary, drain the humidifying tray manually (not necessary with temperature-only chambers). 1. Check the main power switch is in the ON position. 2. Turn refrigerator control OFF. For an explanation on how to set refrigerator control, see "5.6 Handy Features". 3. Change settings so that the operation is not interrupted or an alarm generated if the chamber is run with the door cracked slightly. Make the following settings. For details on settings, see "Pause Control" and "Door Alarm ON/OFF" in "Chapter 6 System Setup" of the User's Manual - Reference -. 108

Turn the pause control OFF. Turn the door alarm OFF. 4. Set target temperature to a minimum 70 C and turn OFF humidity control (not necessary with temperature-only chambers). 5. Press the OPER./STOP key to start the chamber. Run the chamber for about 60 minutes with the door closed, then for 15 minutes with the door slightly cracked. 6. Return settings made in step 3 to their original settings. Note Drain the humidifying tray manually as needed during operation (not necessary with temperature-only chambers). Power OFF Set the main power switch in the OFF position, then shut OFF the primary power supply. 109

Water Supply Pump Filling After long periods of disuse, if there is no water in the tubes, internal parts of the wick pan pump and humidifying tray pump can stick and prevent water supply by the pumps. In such case, supply water to the primary side of the two pumps (lower half) so that they fill with water. If there is enough water in the tubes, this procedure is unnecessary. 1. Check the main power supply switch is in the OFF position. 2. Open the water circuit compartment. Detach the crescent clamps from the wick pan water supply tube and the humidifying tray water supply tube (connected to elbow), using a pair of pliers. Humidifying tray water supply pump Toalon tube Crescent clamp Elbow Crescent clamp Wick pan water supply pump Toalon tube Fig. 7.10 Detaching crescent clamps 3. Disconnect the toalon tubes indicated by the arrow from the elbow by pulling carefully. (The tube may be difficult to detach. Be careful not to damage the elbow.) 110

4. Fill the disconnected toalon tubes with water. (A convenient way to do this is with a small container or funnel.) Fig. 7.11 Filling supply tube with water (1) CAUTION Be careful not to wet electric parts of the water supply unit. 5. Add water to the tube on the bottom of the pump. The tube can be more quickly filled by gently shaking it to free trapped air. Add water to here. Wick pan water supply pump Fig. 7.12 Filling pump tube with water (2) Humidifying tray water supply pump 111

6. Once filled, fit the toalon tubes over the elbow as far as possible and reattach the crescent clamps as before. Note Check the toalon tubes do not come off when tugged slightly. 7. Close the water circuit compartment, set the main power switch in the ON position and run a humidity operation. 8. Water is supplied to the humidifying tray at the rear of the test area. While water is being supplied, the wick pan pump starts up. After that, water overflowing the wick pan drips onto the humidifying tray. If you cannot confirm that water is dripping, stop the chamber and run another humidity operation. Fig. 7.13 Checking for water dripping 9. If water does not drip onto the humidifying tray after 4 or 5 attempts, contact the place of purchase or ESPEC CORP. 112