INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

Similar documents
INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Cleaning & Sterilization

I.F.U Re/Processing Reusable Medical Devices

CLEANING INSTRUCTIONS

Cleaning and Sterilisation for Burs, accessories and filling material

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Cleaning and Sterilization for Instruments & Posts

Operating / User Manual

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

RECOMMENDATIONS FOR CLEANING DEVICES

Aesculap Implant Systems

Instructions for Use Spinal Punches IFU

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Aesculap Implant Systems

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

O-PEP Hygiene-Beileger

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Processing instructions for the Ligamys suture forceps

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED

When you sterilize an instrument set, you know what the standards are.

Recommendations for cleaning, decontamination and sterilisation. devices

THE BASICS OF DECONTAMINATION

FLASH PAK STERILIZATION CONTAINER SYSTEM

Processing Instructions. Cleaning so easy.

Prolystica Ultra Concentrate Instrument Cleaning Chemistries Now less is more. NEW generation instrument cleaners

Dr. Weigert - New Standards in Cleaning and Disinfection of Medical Devices

Classification of Instruments -

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

RAP-hip Hip stem extraction set

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

The Hydrim M2 washer disinfector. Perfectly prepares large volumes of instruments for sterilisation.

HTM made easy. Testing Explained... in fact everything you need to know about HTM but were too afraid to ask!

The Hydrim C51wd Washer Disinfector. Perfectly prepares instruments for sterilisation in as little as 19 minutes.

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

STERIS Instrument Cleaning Chemistries

Setting The Standards For A Safer Future

CENTRAL STERILISATION. Innovative, hardworking Sterilisation & Infection Control products perfect for your Central Sterile Supply Department

Elaborating of a Reprocessing Instruction According to ISO Klaus Roth

Manual Cleaning? SOBECC Round Table: Manual and Automated Cleaning Methods th July 2012

PINNACLE MONITOR FOR AUTOMATED ENZYMATIC CLEANING (AEC) PROCESS

Reusable Laryngoscope Blades and Handles

Instructions for Cleaning, Sterilization, Inspection and Maintenance. of Trauma & Extremities Medical Devices

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Getinge WD14 Tablo Tabletop washer-disinfector. Fast and with high-throughput

REUSABLE ENDOSCOPY INSTRUMENTS (thermo resistant)

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

Your Health, our Technology AWD655. Instrument Washer Disinfector

GETINGE WD14 TABLO A GOOD-LOOKING TABLETOP WASHER-DISINFECTOR. Always with you

STERITITE INSTRUCTIONS FOR USE

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

Lower Your Costs. Expand Your Flexibility.

The future is bright. And clean. And fully automated.

CODE OF PRACTICE. No 11 CLEANING OF EQUIPMENT AND PREMISES

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

Technical Data Monograph. Prolystica HP Enzymatic Automated Detergent Instrument Protection Properties (North America)

Technical Data Monograph. Reliance Vision Single Chamber Washer/Disinfector

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

Washer disinfectors. > Small - Medium range. devoted to hygiene. For the Environmentally conscious

Presenter: Malcolm McLaughlin Date: 04/04/2019. A Review of Medical Device and Optical Critical Cleaning

BioSonic Ultrasonic Cleaning Systems. biosonic.coltene.com

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Additionally, consider the legal provisions valid for your country as well as to the hygienic instructions of the doctor s practice or hospital.

Getinge WD14 Tablo Tabletop washer-disinfector

Technical Data Monograph. AMSCO 3052 Single-Chamber Washer/Disinfector AMSCO 5052 Single-Chamber Washer/Disinfector

2015 Myco Industries, Inc Fax:

HYDRIM. Powerful cleaning the first step to effective sterilization. Instrument Washers A HIGHER STANDARD

Call our hotline at for all your instrument care and detergent questions. Medline Super Concentrates

8/18/2014. Welcome! How do I get a CE Certificate?

AMITY INTERNATIONAL (U.K.)

The Do's and Don ts in the Instrument Sterilisation Process

PERFORM Operating Document

Fiona Collins Liliana Donatelli Andrew Smith

Technical Data Monograph. The Instrument Protection Properties of Prolystica Enzymatic Presoak and Cleaner and other Enzymatic Products

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

Washer & Disinfector Systems

Instrument Cleaning Chemistries Automated Washing Process

Lower Your Costs. Expand Your Flexibility.

TECHNICAL SPECIFICATIONS FOR PLUSHER WDC 8 SERIES INSTRUMENT WASHER DISINFECTORS

Advance Autoclave. Designed to give you more for your money with a high capacity for faster throughput: More features and benefits...

Instructions for Use: Fiberscope and LED Light Source

AF2 PROFESSIONAL LINE. Washer disinfector for human waste containers. Your Health, our Technology

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

Washer disinfectors Small - Medium range

B+N 18/23L LED. Ritter CleanTec with LED Display BB 18 + BB 23. Sterilizable Products:

Transcription:

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

Contents: Pagina 1 Introduction...2 2 Treatment instructions...2 3 Cleaning...2 3.1 Place of use...4 3.2 Transport to the treatment area...4 3.3 Preparation before cleaning...4 3.4 Manual cleaning...4 3.5 Automated cleaning with washer disinfector...4 3.6 Inspection...6 4 Sterilisation...6 4.1 Preparation before sterilisation...6 4.2 Sterilisation process...6 5 Storage for use...7 6 References...7 Appendix Cleaning Agents...7 1. Introduction This brochure is intended as a general manual for cleaning and reusing medical equipment by Humeca instruments. The equipment, operators, cleaning agents and procedures all contribute to the effectiveness of the treatment. The healthcare institution should ensure that the effectively used combination results in a safe medical device. Equipment with the label single use only may not be cleaned for reuse. Single-use devices that are delivered unsterile must be sterilised before use. For these devices the appropriate sections from these instructions can be applied unless other specific instructions are provided in the product documentation. 2. Treatment instructions The adjacent table describes the requisite steps in preparing the medical equipment for reuse or in preparing a new device for first-time use. 3. Cleaning There are two methods for cleaning electric medical equipment from Humeca Instruments: 1. manual method 2. automated cleaning/washer disinfector Use the automated method whenever possible. The automated cleaning process is easier to repeat and more reliable. Moreover, staff is less exposed to contaminated equipment and the cleaning agents that are used. Whichever method is applied, staff should always use suitable personal protective equipment (PPE). Always read the safety and product information carefully for the proper handling and use of the product. Only specially designed cleaning agents (detergents) can be used to clean medical equipment. For manual cleaning, Suma Bac D10 is preferred, an agent with both bactericidal and fungicidal properties. An alkaline agent (ph 12.5) may be used for automated instrument cleaning. A cleaning agent with protective action (inhibitor) that does not harm the materials should be used. 2

FOLLOW THE DETAILED INSTRUCTIONS STEP BY STEP 3.1 Process items - Remove gross soil 3.2 Transport to treatment area - Prevent drying in - Minimal time between use and cleaning 3.3 Preparation before cleaning - Pre-cleaning using ultrasound, if required 3.4 Manual 3.5 Automated 3.4.2 Manual cleaning - Remove visible contamination - Mechanical action using soft brush - Rinse with clean running water - Evaluate 3.5.2 Automated cleaning Preliminary rinsing & disinfection - Chemical cleaning - Neutralisation - Intermediate rinsing - Disinfection - Drying 3.6 Inspection during set-up - Check for soil - Check for use - Check for damage 4.1 Preparation before sterilisation - Suitable packaging necessary 4.2 Sterilisation - Steam sterilisation 4.3 Storage - Inspect surroundings dry dust-free - Inspect time of storage For automated cleaning (with washer disinfector), preference is given to ph-neutral enzymatic or alkaline agents (with a maximum ph of 12.5). A cleaning agent with protective action (inhibitor) that does not harm the materials should be used. Only use agents that are designed for use in these machines. In any case, follow the indications, instructions and warnings by. Pay special attention to the concentration used and the total distribution of the cleaning agent. A recommended dose is included in the appendix of this brochure. Pay sufficient attention to the water quality that is used to dilute cleaning agents and rinse medical equipment. Water hardness, salts, chloride, silicates, iron and mineral residues from hard water can lead to discoloration or harm to the device or hinder effective cleaning and disinfecting. (The chloride content of the water level should not be greater than 120 mg/litre.) 3

3.1 Place of use Use absorbent clothes to remove the worst soil. 3.2 Transport to the treatment area Avoid mechanical damage (for example, do not combine heavy devices with delicate ones). Watch out for sharp elements to avoid both injury and damage to the medical device. Bring the medical devices as quickly as possible to the place where they need to be cleaned. (If they need to be moved to the treatment area, they can be covered with a damp cloth so that the soil does not dry up.) 3.3 Preparation before cleaning Disassemble the device to the extent possible. Consult the Humeca instructions. Most devices for heavy use cannot be disassembled beyond removing attachments. Consult product instructions for the exact procedures. 3.4 Manual cleaning 3.4.1 Pre-cleaning & cleaning Required equipment: Brushes soft; Personal protective equipment (PPE) as recommended by when using the cleaning agents (minimum of overalls, gloves, face mask and protective eyewear); Absorbent cloths. Procedure: Use absorbent cloths moistened with cleaning agent to remove gross soil; Make sure that all surfaces are well moistened; Prepare a cleaning solution from Suma Bac D10 (see product information for dosage); Clean the medical device thoroughly with suitable brushes and pay special attention to rough surfaces and places where the soil is protected against the brushes; Activate hinged or moving parts; Rinse under clean running water until all traces of the cleaning solution have been removed; Make use of ultrasonic cleaning with the agent Suma Med Enzyme, if necessary; Pay special attention to hinges, blind holes and connections between interconnecting parts. Activate all jacking devices or clamping devices for blades and knives to remove all residuals from the cleaning solution; Check to see whether any soil has remained behind and repeat these steps if necessary; Drain on absorbent cloths. 3.5 Automated cleaning with washer disinfector (preferred) 3.5.1 Pre-cleaning Required equipment: Personal protection equipment as recommended by (minimum of overalls, gloves, face mask and protective eyewear); Absorbent cloths. Procedure: Use absorbent cloths moistened with cleaning agent Suma Light D1 to remove the worst soil; Make sure that all surfaces are well moistened; Rinse with running water; Check to see whether any soil has remained behind and repeat these steps if necessary; Drain on absorbent cloths. 3.5.2 Cleaning Required equipment: Washing machine or disinfection device (see remark below); Cleaning and rinsing agent as required for the washer disinfectors. Procedure: Place the medical devices in the washer disinfector: - Avoid contact between devices (due to movement during washing, damage may be caused and the washing process may be obstructed). Use the specially designed sterilisation basket; - Place the medical devices so that the cannulas do not lie horizontally and are aimed downwards (so they empty out); Start the cleaning and disinfecting programme on the washer (see remark below); Empty the washing machine at the end of the disinfectant programme. Visually check each device to make sure that there is no residual soil and that everything is dry. Repeat the cleaning process if soil remains. Remaining moisture can be removed with medical compressed air. 4

Remark: Cleaning and washer disinfectors must meet the requirements of NEN 15883. There are two types of washing programmes: 1. Enzymatic washing programme This washing programme must include the following steps: Pre-washing preferably with softened or demineralised water (2-4 minutes at approximately 18 C). The detergent may be omitted. Wash with an enzymatic detergent (Suma Med Enzyme) for at least 6 minutes at a temperature between 37 C and 48 C. See table for dosage. Wash for at least 2 minutes. Disinfection phase at 93 C and time according to the guidelines. Dose Suma Med Rinse Plus. (Use RO or demineralised water) See table below for dosage. Dry for 15 minutes at 110 C. 2. Alkaline washing programme This washing programme must include the following steps: Pre-washing preferably with softened or demineralised water (2-4 minutes at approximately 18 C). The detergent may be omitted. Wash with an enzymatic detergent (Suma Med Enzyme) for at least 6 minutes at a temperature between 37 C and 48 C. Wash with alkaline detergent (Suma Med Alu Free) for at least 6 minutes at a temperature between 55 C and 60 C. See table below for dosage. Neutralisation (if use is made of alkaline agents) recommendation in dosing Suma Neutral. See table for dosage. Intermediate rinsing of at least 2 minutes. Softened or demineralised water. Disinfection phase at 93 C and time according to the guidelines. Dose Suma Med Rinse Plus. See table below for dosage. (use RO or demineralised water) Dry for 15 minutes at 110 C Automated cleaning and disinfection using Suma Med Enzyme Washing programme for Humeca Pre-rinse CW 1st cleaning Dosage Rinsing Thermal disinfection Dosage Drying Water < 25 C Demineralised / Demineralised/ softened softened and WW ½ - ½ ½ - ½ Demineralised Products and Suma Med Heat to 90 C and Suma 0.2 to dose adjustments Enzyme dose 2 to 6 m/l Med Rinse Plus dose 0.4 ml/l 110 C Wash > 37 and < 48 C Cold 93 C 2 min. Cleaning time 6 min. 3 min. 5 min. 15 min. Automated cleaning and disinfection using alkaline cleaner Suma Med Alu Free Washing programme for Pre-rinse CW Cleaning Dosage Rinsing Intermediate Thermal Humeca rinsing disinfection Dosage Drying Water Demineralised / Demineralised Demineralised softened WW /softened and ½ - ½ WW ½ - ½ Products and dose Suma Med Alu Free 1.5 to Neutral dose Heat to 90 C and Suma 0.2 to 110 C adjustments dose at 40 C 4 m/l 0.2 to 2 ml/l Med Rinse Plus dose 0.4 ml/l Wash > 55 C and < 60 C Cold Cold 93 C 2 min. Cleaning time 6 min. 2 min. 2 min. 5 min. 15 min. * With use of Ultrasonic Suma Med Enzyme 3 ml/l at 37 C Product information and safety sheets can be downloaded via www..nl 5

3.6 Inspection All medical devices must be inspected before preparation for sterilisation. Visual inspection without magnifying glass under good lighting conditions will generally suffice. All parts of the devices must be checked for visible soil, damage and/or corrosion. Special attention should be paid to: Places where soil continues to stick, such as interconnecting surfaces, hinges, shafts, etc. Recesses (holes, etc.); Elements where soil can be pressed against the device; Check for damage to the extent that the device no longer works or blemishes are formed that can damage tissue or surgical gloves. Function checks should be performed where possible: A check should be made to see whether interconnecting devices have been properly assembled; Medical devices with moving parts should be activated to check whether they still work properly (a lubricant for medical applications can be used that is suitable for steam sterilisation if required and only according to the instructions for use). 4. Sterilisation Steam sterilisation is recommended. This is suitable to sterilise Humeca medical instruments. The design and performance of the autoclave, however, may affect the efficiency of the process. Healthcare institutions should validate their procedure using the materials and the operators who usually handle the instruments. 4.1 Preparation before sterilisation Use special containers from Humeca where applicable. 4.2 Sterilisation process Steam sterilisation (moist heat) in an autoclave using a pre-vacuum (forced air) is recommended. Autoclaves must satisfy the requirements of and must be validated in accordance with ISO 13683. Humeca instruments has validated autoclave cycles for the sterilisation of complete medical devices. The following process parameters are validated by Humeca instruments and recommended for sterilisation. Method: steam sterilisation in accordance with ANSI/AAMI/ISO 11134. Cycle: Exposure time 1 : Temperature: Pressure: Drying time 2 : saturated steam with vacuum / fractional forced air. 4 minutes. The exposure time can be extended to 18 minutes in order to comply with the recommendations from WHO, RKI, etc. Humeca medical instruments can withstand such sterilisation cycles. 134 C - 137 C (273 F - 279 F). 2 bar (30 PSI) or 3 bar (45 PSI). Varies depending on the load, packaging method and materials. A drying time of 8 minutes is required for the electric devices for heavy use of Humeca instruments. 1 Exposure time: period in which the load and the complete chamber are kept at the sterilisation temperature. 2 Drying time: the period in which steam is removed from the chamber and the chamber pressure is reduced to let the condensate of the load evaporate, either through sustained drainage or through the injection and extraction of hot air or other gases. REMARK: final responsibility for the validation of sterilisation techniques and equipment lies with the healthcare institution. To guarantee optimal treatment, all cycles and methods must be validated for different sterilisation chambers, packaging methods and/or all sorts of configurations for the load. 6

5. Storage for use Storage life is dependent on the sterile barrier that is applied as well as storage, environmental and handling conditions. Each healthcare institution should determine a maximum storage life for use for sterilised medical equipment. 6. References ISO 11134: Sterilisation of healthcare products Requirements for validation and routine inspection Industrial steam sterilisation ISO 13683: Sterilisation of healthcare products Requirements for validation and routine inspection of steam sterilisation in healthcare institutions. APPENDIX 1 CLEANING AGENTS USED IN THE VALIDATION OF THE TREATMENT INSTRUCTIONS. A cleaning agent (detergent) specially developed to clean medical devices must be used. For manual cleaning, a combined cleaning and disinfection agent with both bactericidal and fungicidal properties is preferred. For automated cleaning (with washer disinfector), an enzymatic or alkaline agent (with a maximum ph of 12.5) with inhibitor is preferred. However, only use agents that are recommended for these machines. In all cases, observe the indications, instructions and warnings of the supplier of the cleaning agent. Remarks: The above table shows the cleaning agents that Humeca has used in validating the handling instructions in this document. Personal protection equipment must be provided for the operators in accordance with the instructions of the supplier and the safety information sheets. The suitability of alternative agents must be monitored based on information from the supplier and/or physical tests. Agents used in the validation of manual and automated (with washer disinfector) cleaning. Supplier Product Packaging Use for Suma Med Enzyme 2 x 5 litres Suitable for manual/ultrasonic and automated washing Suma Med Alu Free 10 litres Automated cleaning (alkaline cleaner with protector) Suma Med Neutral 2 x 5 litres Neutralisation of alkaline residues Suma Med Rinse Plus 2 x 5 litres Rinsing agent (can also be used for soaking) Suma Light D1.2 6 x 1 litres Hand detergent Suma Bac D10 6x 2 litres Combined cleaning and disinfectant agent for manual cleaning 7