Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

Similar documents
Instructions for Use Spinal Punches IFU

V. Mueller. V. Mueller Reusable Devices: Cleaning and Sterilization Guide USA. Rx Only

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

I.F.U Re/Processing Reusable Medical Devices

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

Reusable Laryngoscope Blades and Handles

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Operating / User Manual

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Cleaning & Sterilization

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 1. Perform reprocessing immediately after each use.

STERITITE INSTRUCTIONS FOR USE

Processing instructions for the Ligamys suture forceps

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

IMPORTANT INFORMATION ON THE DEPUY SYNTHES

EMAX 2 PLUS AND XMAX CLEANING AND STERILIZATION BASKET. Mechanical/Automated Cleaning and Sterilization Guide in a Variety of Applications

Recommendations for cleaning, decontamination and sterilisation. devices

RECOMMENDATIONS FOR CLEANING DEVICES

Cleaning and Sterilization for Instruments & Posts

Caution: Federal U.S. law restricts this device to sale by or on the order of a physician.

idrive Ultra Powered Stapling System

Aesculap Implant Systems

Operator s Guide. Autoclavable Internal Handles with Integrated Paddles Rev. M

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

AS Medizintechnik GmbH Sattlerstraße Tuttlingen Germany

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC wrenches

Instructions for Use: Fiberscope and LED Light Source

Visao HIGH-SPEED OTOLOGIC DRILL CATALOG

CLEANING INSTRUCTIONS

INSTRUCTIONS FOR USE & STERILIZATION FOR INSTRUMENTS

When you sterilize an instrument set, you know what the standards are.

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

Cleaning and Sterilisation for Burs, accessories and filling material

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Aesculap Implant Systems

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

Processing (cleaning, disinfection and sterilization) of instruments from Smith & Nephew Orthopaedics AG

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

Manual Pre-disinfection/Cleaning and Sterilization Instructions for SATELEC Tips and Files

STERIS Instrument Cleaning Chemistries

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

INSTRUCTIONS FOR CLEANING AND STERILISING HUMECA MEDICAL EQUIPMENT FOR BOTH ELECTRIC AND MANUAL DEVICES

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

INSTRUCTIONS EVIS EXERA II DUODENOVIDEOSCOPE

REUSABLE ENDOSCOPY INSTRUMENTS (thermo resistant)

FLASH PAK STERILIZATION CONTAINER SYSTEM

HANDS-ON MODULE 5. CLEANING FLOW-i V3.0

X99 Series. Dental Handpiece. Instruction For Use

SteriTite - The Container System of Choice SteriTite

SteriTite - The Container System of Choice SteriTite

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

Operator s Manual. Model G17-EU (International) Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

THE BASICS OF DECONTAMINATION

2015 Myco Industries, Inc Fax:

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

INSTRUCTIONS EVIS EXERA GASTROINTESTINAL VIDEOSCOPE

BIOVAC 500 SMOKE EVACUATOR

Cavitator Ultrasonic Cleaner Instruction Manual

All Truphatek's products:

O-PEP Hygiene-Beileger

UV Lamp and Quartz Sleeve Replacement/Maintenance

Operator s Manual. Model G32-S Model G32-E Disinfection Soak Stations

Operator s Manual. Model G17 Disinfection Soak Station for Hysteroscopes, Cystoscopes and ENT scopes

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

DRAFT. Module 4. Decontamination & Environmental Cleaning

Institute of Decontamination Sciences Blackburn House, Redhouse Road, Seafield, Bathgate, West Lothian EH47 7AQ

RAP-hip Hip stem extraction set

H5i disinfection guide

Prolystica Ultra Concentrate Instrument Cleaning Chemistries Now less is more. NEW generation instrument cleaners

THE ULTRASONIC DISRUPTION OF ORGANIC SOIL ON COMPLEX INSTRUMENTATION

Call our hotline at for all your instrument care and detergent questions. Medline Super Concentrates

WORLD LEADERS IN ADVANCED SURGICAL INSTRUMENT REPROCESSING TECHNOLOGY

Cavitator Ultrasonic Cleaners

CryoPen Cryosurgical System OPERATOR S MANUAL

PICNet CONFERENCE 2013

SteriTite - The Container System of Choice SteriTite Instructions for Use

Cavitator Ultrasonic Cleaner Instruction Manual Model 20G

Veterinary Autoclave

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

Flexible Nasopharyngoscope. Operator Manual. Manufactured by Henke-Sass, Wolf Distributed by Anthony Products, Inc.

Instructions for Use and Reprocessing Instructions

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

GETINGE CLEAN DETERGENT RANGE GET MAXIMUM PERFORMANCE VALIDATED

Instrument Cleaning Chemistries Automated Washing Process

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

A new wave in instrument washing and disinfection Introducing the new HYDRIM c61wd G4 series

Technical Data Monograph. The Instrument Protection Properties of Prolystica Enzymatic Presoak and Cleaner and other Enzymatic Products

Processing Instructions. Cleaning so easy.

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

EVALUATION REPORT DISPOCLEAN ENDOSCOPE CLEANING BLADE. ALKAPHARM UK Ltd

Important information

READ AND UNDERSTAND THESE INSTRUCTIONS BEFORE OPERATING THE MACHINE

Transcription:

Fiber Optic Cables Instructions for Use and Processing Instructions DFU-950-0031-00 Fiber Optic Cables Made in USA Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904) 737 7611 Distributed by: Arthrex, Inc. 1370 Creekside Blvd. Naples, FL 34108, USA (800) 934-4404 Arthrex GmbH Erwin-Hielscher-Strasse 9 81249 München, Germany +49 89 909005-0 RMS UK Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 Page 1 of 7

Device Symbols Used: Manufacturer See Instructions for Use Warning: Instructions for preventing personal injury Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician Non sterile Authorized Representative in the European Community Page 2 of 7

INSTRUCTIONS FOR USE 1. About this document Arthrex fiber-optic cables are designed to deliver maximum light when coupled to a medical grade fiber-optic light source. Arthrex fiber-optic cables can be used with quartz halogen, metal halide, LED, or xenon light sources. They are compatible with virtually all endoscopes, medical instruments, and microscopes. Arthrex fiber-optic cables are reusable devices that are provided non-sterile, but must be sterilized before their initial use and after each use. This document describes the correct handling and recommended methods for their processing. Users of these fiber-optic cables are encouraged to contact their Arthrex representatives if, in their professional judgment, they require more comprehensive information on its use and care. The current version of this document can be found on the internet at http://www.arthrex.com. You can also request this document from Arthrex. 2. Intended Use Arthrex fiber-optic cables are designed to illuminate a surgical site by relaying light from a fiber-optic light source onto the desired site. Arthrex fiber-optic cables are medical grade, high transmittance, peak efficiency. 3. Contraindications These fiber-optic cables have no contraindications of which we have knowledge. 4. Warnings and Precautions Users of these fiber-optic cables should be thoroughly familiar and trained in the use and care of the product. Users are encouraged to contact their Arthrex representatives if, in their professional judgment, they require more comprehensive information. The Arthrex website (www.arthrex.com) also provides detailed information. Arthrex fiber-optic cables are provided non-sterile and must be sterilized before each use and after each use. See instructions for cleaning and sterilization. Read, observe and store these instructions and any other applicable instructions. After receipt of the fiber-optic cable, and prior to each use, always inspect for any evidence of damage. Pay particular attention to optical surfaces, looking for scratches or dings. Use fiber optic cable only as intended. Use caution to treat Arthrex fiber-optic cables as you would any fine optical device. Do not look into the end of the fiber optic cable while it is connected to a light source and the light source is ON. Eye injury or blindness will result. Page 3 of 7

WARNING: Risk of burns! The optical fibers emit high-energy light at the distal end of the fiber-optic cable. This can cause the temperature of the body tissue to rise to 106 F (41 C). Avoid direct contact of the distal end with body tissue or flammable materials, such as drapes, as it can cause burns and fires. Reduce the light intensity of the light source when working near body tissue or flammable materials. Bring the fiber optic cable to the decontamination area as soon as possible after use. Observe valid protective measures to prevent contaminating the environment. 5. Cleaning and Disinfection CAUTION Blue Light Guide cables are not intended to be cleaned using alkaline cleaners (ph above 10). These Blue Light Guide cables must be cleaned with enzymatic cleaners and will be damaged by alkaline cleaners. 5.1 Inspection and Preparation for Cleaning, Disinfection and Sterilization Arthrex fiber-optic cables are delicate medical instruments and must be used and handled with care. It is recommended that fiber-optic cables are reprocessed as soon as is reasonably practical following use. Observe valid protective measures to prevent contaminating the environment. When properly performed, cleaning, disinfection and/or sterilization do not compromise the use and mechanical performance of these fiber-optic cables. These fiber-optic cables are used with patients who may harbor both recognized and unrecognized infections. To prevent the spread of infection, all fiber-optic cables must be thoroughly cleaned, disinfected, and sterilized after use on each patient. Fiber-optic cables for endoscopes, microscopes, and surgical instruments are high quality optical devices. Arthrex recommends the following Steam Sterilization (Autoclave) guidelines. 5.2 Manual cleaning Fiber-optic cables require similar care to that taken for any precision optical component. After each use, and prior to disinfection or sterilization, the fiberoptic cable should be washed and cleaned of all debris. Scrub fiber-optic cable with a soft brush and mild detergent until all visible contamination has been removed, paying particular attention to any crevices or seams. Always avoid any harsh materials or detergents that can scratch or in any way damage the optical surfaces on each end of the fiber-optic cable. Page 4 of 7

5.3 Automated cleaning (full cycles) Phase The devices shall be run through a hospital grade washer full cycle. The minimum parameters for the full cycle are as follows. Motor Speed High [USA Parameters] Recirculation Time [Minutes] Temperature Detergent Type and concentration Pre-Wash 02:00 Cold Tap Water 1 Enzol, 1oz/gal Enzyme Wash 03:00 Hot Tap Water (Neutral ph 7.8-8.8) Hot Tap Water Rinse 1 00:15 60 C Drying 06:00 90 C CAUTION Blue Light Guide cables, except for those equipped with ergonomic molded silicone rubber handles, are not intended to be cleaned using alkaline cleaners (ph above 10). They will be damaged by alkaline cleaners and must be cleaned with enzymatic cleaners per the instructions above. Phase Motor Speed High [European Parameters] Recirculation Time [Minutes] Temperature Detergent Type and concentration Pre-Wash 02:00 Cold Tap Water 1 Neodisher Enzyme Wash 03:00 Hot Tap Water Mediclean Forte ¼ oz / gal (ph 10.5-11) Rinse 1 00:15 Hot Tap Water 60 C Drying 06:00 90 C Page 5 of 7

6. Sterilization 6.1 Steam Sterilization The cycle selected is dependent on equipment and hospital protocol. General guidelines are: Method Steam (Wrapped) Steam (Wrapped) Steam (Un-Wrapped) STERILIZATION PARAMETERS Minimum Exposure Cycle Temperature Exposure Time Pre-vacuum 132 C (270 F) 4 Minutes Gravity 132 C (270 F) 15 Minutes Gravity 132 C (270 F) 10 Minutes 6.2 Chemical Disinfection It is recommended that the guidelines below, published by Advanced Sterilization, a Johnson & Johnson company, be followed. CHEMICAL DISINFECTION PARAMETERS Product High Level Disinfection Sterilization Cidex Activated Dialdehyde Solution 45 minutes @ 25 C (77 F) Cidex Plus 28 Day Solution 20 minutes @ 25 C (77 F) 6.3 Hydrogen Peroxide sterilization (STERRAD method) Follow Sterrad (a Johnson & Johnson Company) Guidelines. 6.4 Special Precaution Transmissible Spongiform Encephalopathy Agents It is outside the scope of this document to describe in detail the precautions that should be taken for Transmissible Spongiform Encephalopathy Agents. The agents for transmission of Creutzfeldt-Jakob disease (CJD) are believed to be resistant to normal processes of disinfection and sterilization and therefore the normal processing methods of decontamination and sterilization as described above may not be appropriate where CJD transmission is a risk. Page 6 of 7

In general, the tissues that come into contact with equipment are those of low TSE infectivity. However, particular precautions should be taken when handling equipment that has been used on known, suspected, or at-risk patients. 6.5 Limited Warranty Your fiber-optic cable has a one (1) year warranty from the date of shipment on workmanship and all defects of material except for broken fiber. Should your product prove to have such defects within one (1) year of shipment, Arthrex will repair or replace the product or component part without charge. Should your fiber-optic cable need servicing under this warranty, please contact your distributor or your customer support specialist for return authorization documentation. You should carefully pack the product in a sturdy carton, including a note describing the defects, your name, your company name, telephone number and a return address. Warranty does not cover equipment subject to misuse, accidental damage, and normal wear and tear. This warranty gives you specific legal rights, and you may also have other rights that vary from state to state. 6.6 Post Warranty Repair Please contact your distributor or your customer support specialist for return authorization documentation. 6.7 Storage Fiber-optic cables should be stored in a clean, dry environment. 6.8 Disposal Observe country-specific regulations and laws for the disposal of medical products. Page 7 of 7