Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits

Similar documents
Installation & Operating Instructions for Med Touch Series Area, Master & Combination Alarm Systems

Installation & Operating Instructions for Integrated Series Area Alarm Zone Valve Box Systems

Tri-Tech Medical Inc.

Installation & Operating Instructions for ZT and EZT Integra Touch Series Area Alarm Zone Valve Box Systems

Installation Instructions for Med Touch Master Alarms - Wireless

Installation & Operating Instructions for Area, Master & Combination Alarm Systems

DC Series Area, Master and Combination Alarm Systems Opreator s Manual

Installation, Operation, and Maintenance Instructions

Med Touch Combination Alarm Conversion Kits

BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS

Tri-Tech Medical Inc.

MEGA. cal electronic gas alarm. Installation, Operation, and Maintenance Instructions. Part No Rev. D Pg.

Installation Instructions

Aqua-Aire Sit-Bath System 6300 Installation / Assembly Instructions

INSTALLATION INSTRUCTIONS FOR 6532 SERIES PACKAGE HEAT PUMP

Installation & User Guide

Med Touch Master Alarm

Handshower, Hose & Handshower Accessories

INSTALLATION INSTRUCTION

A PROUD HERITAGE OF EXPERIENCE & QUALITY DISHWASHER BR- DWSH01- S

Hanson LED C e i l i n g F a n

Installation Instructions Built-In Dishwasher

INSTALLATION INSTRUCTIONS - POWER SENTRY CONTROL POWER SENTRY CONTROL UNIT PSBCEB2 INSTALLATION INSTRUCTIONS

Installation Instructions. For the 18 Built-In Dishwasher and Front Color Panels

Sunburst. Instruction Manual. Includes our new Wall Control System. A Kichler Décor ceiling fan

Tri-Tech Medical Inc.

Dishwasher Installation Instructions DW 24XT/DW 24XV

INSTALLATION INSTRUCTION

62" LED SPYRA. Product images may vary slightly from actual product. INSTRUCTION MANUAL

INSTALLATION INSTRUCTIONS FOR 6636 SERIES PACKAGE AIR CONDITIONER

Classic Twelve O clock Cabinet Installation Instructions

EZT Series Integra Med Touch Area Alarm & Zone Valve Box with EZ Backfeed

Single stage operation oil burner

Armstrong J Series Float, Thermostatic Steam Traps, Condensate Controllers & Liquid Drainers Installation and Maintenance Manual

INSTALLATION INSTRUCTIONS FOR SERIES TWO TON HIGH EFFICIENCY PACKAGED HEAT PUMP

Installation Instructions Built-In Dishwasher

UNITY CEILING RECESSED MOUNT CM1900 CM1901 CM1906 CM1912

A PROUD HERITAGE OF EXPERIENCE & QUALITY DISHWASHER BR-DWSH01-S

Duct and Rough Service Carbon Monoxide Sensor

ChicagoTM. Instruction Manual. Includes our new CoolTouch TM Control System Looks permanent, but goes wherever you go! U.S.

Installation Instructions

DRAIN WATER TEMPERING INSTALL

COPPER FREESTANDING TUB

JOHN DEERE X465, X475, X485, X495, X575, X585

Tripod Ceiling Fans. Installation Instructions Models: CT52, D52, FR52, ME52 & OL52 (800) BETTER BY DESIGN MEJOR DISEÑO BETTER BY DESIGN

INSTALLATION INSTRUCTIONS

CABS/BLADES/ SPREADERS/ACCESSORIES

Installation Instruction. Puritan Bennett. Wall Air Water Trap For 800 Series Ventilator Basic Cart. Pla

52 StarkkTM. Instruction Manual. A Kichler Select ceiling fan

60" Hatteras BayTM. Patio. Instruction Manual. Includes our new Wall Control System. A Kichler Décor ceiling fan

Cascade Premier and Elite Premier Bathing Systems with Aqua-Aire Installation / Assembly Instructions

Built-In Dishwasher. Installation Instructions. BEFORE YOU BEGIN Read these instructions completely and carefully. IMPORTANT The dishwasher MUST be

Manifold Installation Controller MIC-180. Installation and Operation Instructions

PARTS & ACCESSORIES INSTALLATION AND SAFETY INSTRCUTIONS ITEM NO.:60006PC SAFETY PRECAUTION. Canopy. Downrod. Housing. Blade. Transmitter CR2032/3V

Wiring Instructions. Single Doors 99 Checkmate Series Use one or Master Unit each door or one or Support unit each door

1. SAFETY RULES. 8. Avoid placing objects in the path of the blades.

INSTALLATION AND MAINTENANCE

INSTALLATION INSTRUCTIONS FOR 8330*5511 MOUNTING KIT

Installation Instructions

Innes. Instruction Manual A Kichler Decor ceiling fan. Includes our new TM

Installation Instructions

Aquapeutics GENERAL INSTALLATION GUIDE. Toll Free Customer Service line 1 (800)

52 Lacey LED. Instruction Manual. 6 Speed DC Wall Control System

INSTALLATION INSTRUCTIONS FOR 6536 SERIES TWO TON PACKAGED HEAT PUMP

Acrylic Claw Foot Tub

DH100ACDCLP Air Duct Smoke Detector with Extended Air Speed Range

PARTS & ACCESSORIES INSTALLATION AND SAFETY INSTRCUTIONS ITEM NO.:60010BZGTGLD SAFETY PRECAUTION. Canopy. Downrod. Housing. Transmitter CR2032/3V

1 ESMA, Inc. P. O. BOX 734 * SOUTH HOLLAND, IL * (800) * FAX (708)

INSTALLATION INSTRUCTIONS

ColorMax Cyclone Cleaning Kit

DPM-G DPM-CR DPM-W DPM-B INDOOR DOME PENDANT MOUNT INSTALLATION MANUAL

Fillmore Small Pendant Assembly and Installation Instructions

Top Control Dishwasher

GE Monogram. Installation. Instructions. Professional Ranges. 48" Natural Gas Models ZDP48N4GWSS ZDP48N6RWSS ZDP48N6DWSS

SAVE THESE INSTRUCTIONS

52 DorsetTM. Instruction Manual. Basic Function Wall Control System Included. A Kichler Decor ceiling fan

Installation Instructions Built-In Dishwasher

Installation Instructions. Instructions de montage. Instrucciones de instalación HC

CEILING FAN OWNER'S MANUAL

Eccotemp L5 and L10 Comprehensive Troubleshooting Guide

CEILING FAN OWNER'S MANUAL

Installation Instructions

42 Kevlar. Instruction Manual. Kichler Lighting 7711 East Pleasant Valley Road P.O. Box Cleveland, Ohio

VPMmini. Operator's Manual

60" Tulle PatioTM. Instruction Manual. A Kichler Select ceiling fan

60" Lyndon Patio. Instruction Manual Customer Service :30 AM to 5:00 PM EST, Monday - Friday A Kichler Decor ceiling fan

JOHN DEERE GATOR HPX/XUV 2 PASSENGER HEATER INSTALLATION INSTRUCTIONS (p/n: 9PH20S30)

Handshower, Hose & Handshower Accessories

LED. 60 StarkkTM. Instruction Manual. A Kichler Select ceiling fan

1. SAFETY RULES WARNING WARNING. 8. Avoid placing objects in the path of the blades.

CURTIS/KUBOTA BX-1500, 1800, 2400, 1850, 2350, 25 BX1870, BX2370, BX25D & BX HEATER INSTALLATION (p/n: 9PH20S18)

54" Skye. Instruction Manual Customer Service :30 AM to 5:00 PM EST, Monday - Friday A Kichler Decor ceiling fan

INSTALLATION INSTRUCTION for ceiling fan

3MNLR72XXD Series Fan. Owner s Guide and Installation Manual. UL Model NO. : 3MNLR72

GRANDE OWNER S MANUAL

Installation Instructions

INSTALLATION OF DURAVENT POLYPRO SINGLE WALL POLYPROPYLENE VENTING SYSTEMS

Installation Instructions

Ligature Resistant Sensor Faucet Patent # US D635,386 #SAL-5011/SF370

Installation Instructions

Transcription:

Manufacturer of Medical Gas Pipeline Equipment Installation Instructions for Med-Touch Series LCD Medical Gas Area, Master & Combination Alarms Conversion Kits Master Alarm Conversion Combination Alarm Conversion Area Alarm Conversion e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 1

Contents Table of Contents Alarm Conversion Installation ----- 3 Existing Rough-In Box Disassembly --- 4 Installation of DISS valves & Transducers ---------------------------------- 5 Installation of Conversion Plate ---------- 5-6 Wiring/AC --------------------------------- 6 Installing the alarm front panel --------- 7 Wiring alarm front panel ---------------- 7-8 Installing Ground Screw ------------------ 7-8 Installing the transducers (not remote) - 8 Power Supply Switch ----------------------- 9 Restore Power & Gases -------------------- 9 Master Alarm Conversion Section ---- 10 Master Alarm Conversion Installation 11 Installation of Conversion Plate ---------- 12 Wiring/AC- -------------------------------- 13 Installing the alarm front panel ---------- 13 Wiring Master Alarm front panel ------ 13-14 Installing Ground Screw ------------------ 14 Wiring Remote Signal Master (Dry Contacts) ------------------------------ 15 Wiring Remote Signal Master Transducer ----------------------------------- 16 Wiring Remote Signal Master (4-20 ma Transmitter) --------------------- 17 Power Supply Switch ---------------------- 18 Restore Power & Gases -------------------- 18 See Med Touch Alarm Installation & Operating Instruction Manual #99-0500 For Testing and operation e6322rir 10/28/16, 35401 Avon Commerce Pkwy, Avon, Ohio 44011 No. 99-0520 - 2 -

Area Alarm Conversion Kit This product has been designed to convert an existing area alarm to a area alarm. Installation of this kit involves the removal of the existing alarms front panel, power supply, and transducer(s). Installation also involves installing a flange, power supply, transducers and alarm front panel and making the necessary plumbing and electrical connections. All installation and testing should be done in accordance with NFPA 99 or CSA Z7396.1. WARNING: Installation of this product requires the temporary shut down of medical gases to the affected area of the medical facility. All medical gases supplying the alarm to be converted must be shut down until the conversion is complete. It is the responsibility of the installer to obtain approval from the proper facility personnel before beginning this conversion. WARNING: Electrical power intended for the alarm to be installed should be disconnected prior to installation. WARNING: This device should only be installed by qualified personnel. Installation should not be attempted by anyone not having general experience with the installation of devices of this nature. Locate the medical gas zone valve box providing service to the area alarm to be converted. After approval has been granted by the proper facility personnel, close all gas valves providing service to the area alarm to be converted. Disconnect electrical power to the area alarm to be converted. Note: This manual illustrates the conversion of an Ohio area alarm. There are many other models of area alarms that may also be converted and there will be differences in the kits used to convert them. Remove the existing front panel, power supply and transducers. Note: Gas labels corresponding to the gas services being monitored have been placed on the wall adjacent to the alarm before disassembly. This is to help prevent a possible mismatching of gas services during the conversion. These labels will be covered when the alarm is converted. Note: the transducers may be located in the back box or remotely in the ceiling or remotely located in Medical zone valve boxes as shown here). Note: If the transducers are to be installed remotely, see the Med Touch Installation & Operating Instruction Manual #99-0500 for the proper techniques to wire remote transducers. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 3 -

Area Alarm Conversion Kit The Ohio area alarm (pictured here) must be disassembled from the bottom module up. Note there is a short hose connected to a Hanson quick connect coupling that must be disconnected in each gas module. There is also an electrical plug connector on each gas module that must be disconnected. Remove existing power supply from back box If the gas connection fittings are not DISS gas specific demand valves, they must be removed. In this conversion the Hanson couplings should be removed. When the fittings are removed the remaining fitting in most cases will be either a 1/8 or a 1/4 NPT Female thread. In some conversions, the threads are flare fittings and adaptors are provided as part of the conversion kit. Special adaptors provided with conversion kit. The demand check valve fittings are stamped with either the abbreviated gas name or the chemical symbol of the gas. Take special note as to which gas service is being provided by each gas riser as they may not be labeled on the back box as shown below. If they are not labeled or marked, do so now. The special adaptors which have been included must be used. In some cases, there is insufficient clearance for the poppet in the demand check to open allowing gas to flow. These adaptors provide sufficient clearance to allow the demand valves to operate properly when engaged with mating fittings. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 4 -

Area Alarm Conversion Kit In this photo the adaptors and demand valves are installed. Install the transducers onto the demand valves. This photo shows all of the transducers installed. The power supply comes pre-installed on the conversion plate. It may not be installed on the conversion plate in the correct position to allow it to fit into the back box. If necessary, the power supply may be removed from the conversion plate (2 screws) and repositioned on any of the 4 sides of the conversion plate. Pre-drilled and tapped holes have been provided. Position the conversion plate with the power supply into the back box. The conversion plate must cover the back box and the power supply must recess into the back box. As long as these requirements are met, the conversion plate may still provide extra coverage and allow for centering as desired. Warning: The wall prints should be checked for power, water or other service lines that may be installed. After the desired position of the conversion plate has been chosen, holding the conversion plate assembly in place, mark the four holes positions on the wall. The four holes are located near the corners of the conversion plate. It is recommended that a level be used. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 5 -

Area Alarm Conversion Kit Temporarily hold the conversion plate in the desired position and mark the wall (thru the holes that are located near the corners of the conversion plate). Remove the conversion plate and drill the 4 holes using a 3/8 bit. CAUTION: The molly bolts provided can be used for wall thicknesses of 5/8 to 1 ¼. If wall thickness is outside this range another type of wall anchor must be used. Install the molly bolts (provided) into the four drilled holes. Tighten screws until snug. DO NOT OVERTIGHTEN. Remove the screws & washers. Position the flange assembly onto the back box lining up the conversion plate mounting holes with the mollies. Thread the molly screws with washer thru the conversion plate into the mollies and tighten. The power supply has been pre-wired for 110 VAC. Connect the three wires from the power supply (Black = Line, White = Neutral & Green = Ground) to the proper incoming 110 VAC wires using wire nuts. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 6 -

Area Alarm Conversion Kit The alarm front panel snaps onto the two hinge pins. If the hinge gets detached, simply snap it back on. Attach the two wire lanyards provided to the screw mounts on the sides of the conversion plate. Attach the CPU wiring harness (the one with the black corrugated plastic cover) to the white plug in connector on the leftmost circuit board. Note it is very important that this connection be made properly, with the pins and holes in proper alignment and the latching mechanisms on the white connectors mated properly. Attach the green ground wire, which is in the wiring harness, to the ground screw on the left corner of the front panel just in front of the power supply. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 7 -

Area Alarm Conversion Kit The CPU wiring harness connection will look like this when properly connected. Using the self-taping screw provided, attach the green ground wire harness originating from the ground lug terminal on the alarm front panel and power supply mount, to an appropriate grounding point in the back box. The wire terminal connector on the gas board module has six wire connection slots. The two wires from the transducer should be installed in the BLK & WHT SENSOR slots. These are the two slots closest to the top edge of the gas board (as shown). The transducer plug may be removed from the gas module to make it easier to install the wires. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 8 -

Area Alarm Conversion Kit The power supply has been provided with both a removable fuse and an on/off switch to provide for ease of any possible future service work to the alarm or changes to the medical gas piping system. Make sure the fuse is installed and the switch is in the on position. When closing the alarm front, be careful to not pinch any wires and ribbon cable between the alarm front panel and the conversion plate. Note: Also verify that the ribbon cable is not touching or blocking the latching panel screw. Restore electrical power and medical gases to the converted area alarm. Before the facility restores this portion of the medical gas piping system to patient use, the alarm must be tested for high and low line pressure alarms for all gases (except vacuum & WAGD/evacuation which are tested for low line pressure only) and tested for possible crossconnection to ensure each gas module is monitoring the correct gas. It is recommended that this testing be done by an independent third party medical gas certification company. See the Medical Med Touch Alarm Installation & Operating Instruction Manual #99-0500 for complete instructions on testing and operating the new alarm panel. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 9 -

Master Alarm Conversion Kit Installation Instruction TU Series This Page Left Intentionally Blank e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 -10-

Master Alarm Conversion Kit This product has been designed to convert an existing master alarm to a master alarm. Installation of this kit involves the removal of the existing alarms front panel and power supply. Installation also involves installing a flange, power supply, and alarm front panel and making the necessary electrical connections. All installation and testing should be done in accordance with NFPA 99 or CSA Z7396.1 WARNING: Installation of this product requires the temporary shutdown of the medical gas master alarm panel being converted, but should not effect any other master alarm(s) in the facility. It is the responsibility of the installer to obtain approval from the proper facility personal before beginning this conversion. WARNING: Electrical power intended for the alarm to be installed should be disconnected prior to installation. WARNING: All alarms for medical gas sources monitored by the alarm to be converted will be shut down until the conversion is complete. Arrangements should be made to monitor these alarms from another master alarm panel and/or manually. WARNING: This device should only be installed by qualified personnel. Installation should not be attempted by anyone not having general experience with the installation of devices of this nature. Locate the electrical breaker providing service to the master alarm to be converted. After approval has been granted by the proper facility personnel, shut off the breaker providing 120 VAC electrical service to the master alarm to be converted. Remove the front panel of the alarm to be converted. Label each common & signal wire with the alarm signal name before disconnecting them from the original circuit board. Disconnect all of the common & signal wires from the original front panel. Remove the existing power supply and fuse. Position the conversion plate with the power supply into the back box. The conversion plate must cover the back box and the power supply must recess into the back box. As long as these requirements are met, the conversion plate may still provide extra coverage and allow for centering as desired. Warning: The wall prints should be checked for power, water or other service lines that may be installed. After the desired position of the conversion plate has been chosen, holding the conversion plate assembly in place, mark the four holes positions on the wall. The four holes are located near the corners of the conversion plate. It is recommended that a level be used. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 11

Master Alarm Conversion Kit The power supply comes pre-installed on the conversion plate. It may not be installed on the conversion plate in the correct position to allow it to fit into the back box. If necessary, the power supply may be removed from the conversion plate (2 screws) and repositioned on any of the 4 sides of the conversion plate. Pre-drilled and tapped holes have been provided. Holding the flange assembly in place, mark the four hole positions on the wall. The four holes are located near the corners of the flange. Remove the flange with power supply assembly. Drill the four holes positions using a 3/8 bit. Install molly bolts into the four holes drilled. Tighten screw until snug. DO NOT OVERTIGHTEN. Remove screw and washer. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 -12-

Master Alarm Conversion Kit The power supply has been pre-wired for 110 VAC. Connect the three wires from the power supply (Black = Line, White = Neutral & Green = Ground) to the proper incoming 110 VAC wires using wire nuts. The alarm front panel snaps onto the two hinge pins. If the hinge gets detached, simply snap it back on. Attach the two wire lanyards provided to the screw mounts on the sides of the conversion plate. Attach the CPU wiring harness (the one with the black corrugated plastic cover) to the white plug in connector on the leftmost circuit board. Note it is very important that this connection be made properly, with the pins and holes in proper alignment and the latching mechanisms on the white connectors mated properly. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 -13-

Master Alarm Conversion Kit Attach the green ground wire, which is in the wiring harness, to the ground screw on the left corner of the front panel just in front of the power supply. Using the self-taping screw provided, attach the green ground wire harness originating from the ground lug terminal on the alarm front panel and power supply, to an appropriate grounding point in the back box. When completed the ground and 120 VAC wiring connections should look like this. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 -14-

Master Alarm Conversion Kit Wiring the Remote Devices to Remote Signal Master Board (dry contacts) Locate the two banks of 8 terminal connectors on the back of the remote signal (master) board. Inputs 1 thru 4 are on the right edge of the board and 5 thru 8 are on the left edge of the board There are a total of 16 connection points two connection points for each remote signal. The pairs are labeled 1 IN thru 8 IN on the circuit board. The NC (normally closed) labeled terminal of each pair is where the signal wire should be landed. The C (common) labeled terminal of each pair is where the common wire should be landed. Note: it is required that polarity be maintained between the switch being monitored and the master alarm. The terminal connector may be unplugged from the circuit board to simplify installation of the remote signal wires. Land the corresponding signal on the NC (normally closed) & common wires on the C (common) terminal of each pair of terminal connectors. It is imperative that the signal wires are landed in the NC position and that the common wires are landed in the C position. Remote signal #1 Remote signal #2 Common wire signal #1 Common wire signal #2 e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 15 -

Master Alarm Conversion Kit Wiring the Remote Devices to Remote Signal Master Board (transducer sensors) Locate the bank of 8 terminal connectors on the right edge of the back of the remote signal (master) board. There are a total of 8 connection points two connection points for each transducer sensor. The pairs are labeled 1 IN thru 4 IN on the circuit board. The white wire from the transducer should be landed in the WHT position and the black wire from the transducer should be landed in the BLK position. Note: only Medical D and T series alarm transducers are compatible with the Med Touch alarm system. The terminal connector may be un-plugged from the circuit board to simplify installation of the remote signal wires. Transducer # 1 white Transducer #2 white Transducer #1 black Transducer #2 black The sequence in which the transducers are connected to the board correspond with the order in which they will appear on the display. The #1 transducer will be displayed on top and the #4 transducer will be displayed on the bottom. There will be up to two display positions (high line pressure and low line pressure) for each transducer. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 16 -

Master Alarm Conversion Kit Wiring the Remote Devices to Remote Signal Master Board (4-20mA transmitters) Locate 3 port terminal connector on the right edge of the back of the remote signal (Master) board. The terminal connector will be labeled 4-20mA. The terminal connector may be unplugged from the circuit board to simplify installation of the wires. For 4-20mA transmitters with loop power supplied by an external source. Land the 4-20mA signal wire in the Number 2 terminal port. Land the (-) ground wire to the Number 3 terminal port For 4-20mA transmitters with loop power supplied by the Alarm Circuit Board (35-0037) Land the 4-20mA signal wire to the Number 2 terminal port. Land the +12VDC wire to the Number 1 terminal port. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 17 -

Master Alarm Conversion Kit The power supply has been provided with both a removable fuse and an on/off switch to provide for ease of any possible future service work to the alarm or changes to the medical gas piping system. Make sure the fuse is installed and the switch is in the on position. When closing the alarm front, be careful to not pinch any wires and ribbon cable between the alarm front panel and the conversion plate. Note: Also verify that the ribbon cable is not touching or blocking the latching panel screw. Restore electrical power and medical gases to the converted area alarm. Before the facility restores this portion of the medical gas piping system to patient use, the alarm must be tested for high and low line pressure alarms for all gases (except vacuum & WAGD/evacuation which are tested for low line pressure only) and tested for possible crossconnection to ensure each gas module is monitoring the correct gas. It is recommended that this testing be done by an independent third party medical gas certification company. See the Medical Med Touch Alarm Installation & Operating Instruction Manual #99-0500 for complete instructions on testing and operating the new alarm panel. e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520 - 18 -

Manufacturer of Medical Gas Pipeline Equipment TU Series e6322rir 10/28/16, 35401 Avon Commerce Pkwy., Avon, Ohio 44011 No. 99-0520