COMMANDER 1900 Series Circular Chart Recorders. Programming Guide. Recorder Versions COMMANDER deg F ABB Instrumentation

Similar documents
Circular Chart Recorder

COMMANDER 1900 Series Circular Chart Recorders. Operating Guide. Recorder Versions COMMANDER deg F ABB Instrumentation

Operating Guide IM/C1900-OGR Rev. J. C1900 Series Circular Chart Recorders

ABB Instrumentation. Specifications. COMMANDER 150 Universal Process Indicator. the 1 /8 DIN indicator to match all your display requirements

COMMANDER 150. User Guide Universal Process Indicator A1

Hazardous Area Zirconia Oxygen Systems

User Guide. DATUM L150 Panel-Mounted Universal Level Indicator A1. ABB Instrumentation

Installation Guide. EXGP Oxygen Analyzer System (ATEX) Interface Electronics Unit

Electromagnetic Flowmeters

ABB. The Company. Electrical Safety. Symbols EN ISO 9001:2000

SM3000 Videographic Recorder. User Guide. Modbus (RTU) Communications Option

COMMANDER 160 Wall/Pipe-mounted Universal Process Indicator. User Guide. ABB Automation

/8 DIN Universal Process Indicator

User Guide IM/C160_8. Wall-/Pipe-mounted Universal Process Indicator C160

User Guide IM/L160_6. Wall-/Pipe-mount Level Indicator L160

/8 DIN Level Indicator

Zirconia Oxygen Analyzer

I/A Series 716C 1/16 DIN Temperature Controller

ControlMaster CM30 and CM50 Universal process controllers, 1 /4 and 1 /2 DIN Custom configuration sheet

Zirconia Oxygen Analyzer

PROCESS & TEMPERATURE UNIVERSAL INPUT DIGITAL METERS

Process & TeMPerATUre UniversAl input DigiTAl MeTers

Analog Input Module IC670ALG630

ABB MEASUREMENT & ANALYTICS DATA SHEET. C50 1/16 DIN controller / alarm unit

Low Temperature Zirconia Oxygen Analyzer

Installation Manual Tracker 221

eed quality electronic design USER s MANUAL

Analyze IT Zirconia Oxygen Analyzer for Small Boiler Applications

NOVUS AUTOMATION 1/5. Range: -35 to 720 C (-31 to 1328 F) Range: -90 to 1300 C (-130 to 2372 F) Range: 0 to 1760 C (-32 to 3200 F)

CELLTROL II BIOREACTOR CONTROL SYSTEM OPERATIONS MANUAL

Temperature Controllers

ZDT GP Series High Temperature Zirconia Oxygen Analyzer. Superior technology and quality from the world leader in oxygen measurement

VT4810 SINGLE / DUAL ZONE CONTROLLER INSTALLATION MANUAL

TEMPERATURE CONTROL MODULE MODEL R3-TC2

Table of Contents SECTION PAGE SECTION 1 INTRODUCTION Description Features Models... SECTION 2 RS-232 COMMUNICATIONS...

User Guide IM/SM3000 EN Rev. L. SM3000 Multipoint Videographic Recorder

Series Temperature Controller Instruction Sheet

4600 Series Operating Instructions. Models 4623 and 4628 Conductivity Transmitters To Meet United States Pharmacopœia (USP 23) Regulations

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25T EN0l-6091 February 2009

UDC1000 and UDC1500 MICRO-PRO SERIES UNIVERSAL DIGITAL CONTROLLERS

VERTEX VT10 SERIES PID OPERATION MANUAL MICROPROCESSOR BASED PID CONTROLLER

ABB Kent-Taylor. Model 6553 Digital Display (SP13 x 2) Series of Intrinsically Safe Gas Analyser Systems for Hydrogen & Purge Gas Purity Measurement

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications

ABB Instrumentation. Instruction Manual. Model 6553 Series of Intrinsically Safe Gas Analyser Systems for Hydrogen & Purge Gas Purity Measurement

IntelliPack : 800 Series

Carbon Monoxide Transmitter

B-40/B-41 Modulating Temperature Controller

Endura AZ40 Oxygen and carbon monoxide equivalent (COe) analyzer

User Guide IM/AX4CO4 Rev. K. AX413, AX430, AX433, AX436 & AX438 Single and dual input analyzers for high level conductivity

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

AutomationDirect PM24. Microprocessor - Based Process/Temperature Limit Controller. Operator s Manual. Technical Support.

Nov 08 PRODUCT SPECIFICATION SHEET

Module Features are-configurable, no module jumpers to set

02/11/2015

PACSystems* RX3i. Thermocouple Input Module, 12 Channels, IC695ALG412. GFK-2578B October 2011

4/8/12/16 Channel Temperature Scanner with FREE Data Logging PC Software RS485

02/11/2015

Advanced Test Equipment Rentals ATEC (2832)

CONsOlIDATOR 4 & 8. MulTI- C h ANNEl CONTROllERs. ConsoliDator 4 Model PD940 ConsoliDator 4 Features. ConsoliDator 8 Features.

Micro-controller X C C1 C2 AL1 AL2 AL3. Model: PXR4/5/9. Operation Manual SEL PXR. ECNO:406e

UDC 700 Universal Digital Controllers and Indicator Specifications

Refrigerated air dryers

CT224 Setup Tutorial

Model 96VTR ¼ DIN Process Controller

99 Washington Street Melrose, MA Phone Toll Free Visit us at

F4 Process Controller Installation and Operation Guide

INSTRUCTION MANUAL FOR MICROCOMPUTER BASED TEMPERATURE INDICATING CONTROLLER FCS-23A

Table of Contents SECTION PAGE

Serie Rugghölzli 2 CH Busslingen. Tel. +41 (0) Fax +41 (0)

DSGH. Radiation-Based Detector with GEN2000 Electronics for Density Measurement QUICK REFERENCE GUIDE

User Manual. Digi-Sense TC9500 Advanced Multiparameter Temperature Controller with Thermocouple, Thermistor, and RTD Inputs

DIGITAL METERS Large Display Temperature Meters Instruction Manual

Table of Contents 1. OVERVIEW SYSTEM LAYOUT SPECIFICATIONS FUNCTION... 11

ORP (mv) Controller Model: ORP-XP2. UP DOWN ENTER Time & Date Main Menu Alarm Reset

Series Digital Controller Instruction Sheet

UDC100 Universal Digital Controller Specifications

Oakton TEMP 9500 Advanced Multiparameter Controller

Oxygen / Carbon Dioxide Analyser / Transmitter. Model 1637

DIMENS. MIN TYPICAL MAX A 71.0 (2.795) 71.0 (2.795) 71.8 (2.826) B 29.0 (1.141) 29.0 (1.141) 29.8 (1.173)

Analog Room Pressure Monitor RPC Series

STT 3000 Series STT250 Smart Temperature Transmitters Specifications Models STT25H, STT25S, STT25M, STT25D EN0I-6031 March 2010 Rev.

Two-Channel Gas Controller

Warning Caution INSTRUCTION MANUAL DIN RAIL MOUNTING TYPE INDICATING CONTROLLER

SA100. General Description. Features. Temperature Controller SA100. Corresponding to Various Applications. Simple Mounting on DIN Rail

AutomationDirect. PC35 Configuration Sheet

Frequently asked questions: Intelligent Transmitter Series

JIR-301-M No.JIR31E To prevent accidents arising from the misuse of this instrument, please ensure the operator receives this manual.

Programmable Output Dual Relay Quad Relay. (Trip Points, High Alarm, Low Alarm, Failsafe, Non-Failsafe, Normally Open, Normally Closed)

VAV Thermostat Controller Specification and Installation Instructions. Model TRO24T4XYZ1

BA374C Intrinsically safe field mounting indicating temperature transmitter

F PC and AO OUTPUT BOARDS INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

ENERGY LIGHT USER S GUIDE ENERGY LIGHT USER S GUIDE

Instruction manual for DIN rail mounting type indicating controller DCL-33A

Instruction Manual Model A

CLEANROOM MONITOR CR3A Network - Installation Instructions

M2500 Engine Controller Installation Manual

Instruction Manual. Model 6553 Series (ATEX) Gas Analyser Systems for Intrinsically Safe Hydrogen & Purge Gas Purity Measurement

User s Guide. CN616 Series 1 4 DIN Economical 6-Zone PID Temperature Controllers. Shop online at omega.com SM

Pump-Down Controller MODEL 4052

Software Version 2.01 LEVEL MONITOR MODEL 220

Transcription:

COMMANDER 900 Series Circular Chart Recorders Programming Guide Recorder Versions COMMANDER 900 00. 200.5 70.5 deg F ABB Instrumentation

ABB INSTRUMENTATION The Company ABB Instrumentation is an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications. As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support. The quality, accuracy and performance of the Company s products result from over 00 years experience, combined with a continuous program of innovative design and development to incorporate the latest technology. The NAMAS Calibration Laboratory No. 0255(B) is just one of the ten flow calibration plants operated by the Company, and is indicative of ABB Instrumentation s dedication to quality and accuracy. BS EN ISO 900 R E G IS T E R E D FIR M St Neots, U.K. Cert. No. Q5907 Stonehouse, U.K. Cert. No. FM 206 EN 2900 (ISO 900) Lenno, Italy Cert. No. 9/90A Stonehouse, U.K. Cert. No. 0255 Use of Instructions Warning. An instruction that draws attention to the risk of injury or death. Note. Clarification of an instruction or additional information. Caution. An instruction that draws attention to the risk of damage to the product, process or surroundings. Information. Further reference for more detailed information or technical details. Although Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage, it must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded process system performance leading to personal injury or death. Therefore, comply fully with all Warning and Caution notices. Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of Technical Communications Department, ABB Instrumentation. Health and Safety To ensure that our products are safe and without risk to health, the following points must be noted:. The relevant sections of these instructions must be read carefully before proceeding. 2. Warning labels on containers and packages must be observed.. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given.. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/or temperature. 5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures must be used. 6. When disposing of chemicals ensure that no two chemicals are mixed. Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be obtained from the Company address on the back cover, together with servicing and spares information.

CONTENTS Section Page INTRODUCTION... 2 GENERAL PROGRAMMING... 2 2. Preparation for Changes to the Parameters... 2 2.2 Security System... 2 INTRODUCTION The COMMANDER 900 series of documentation is shown in Fig... The Standard Manuals, including the specification sheet, are supplied with all instruments. The Supplementary Manuals supplied depend on the specification of the instrument. BASIC CONFIGURATION LEVEL... 2. Set Up Input (Process Variable)....2 Set Up Pen Range/Event Source... 8. Set Up Chart... 9. Set Up Alarms... 0.5 Set Up Relay Output....6 Set Up Digital Output... 6.7 Set Up Analog Output... 8.8 Digital Inputs... 20.9 Access Page... 2.0 Scale Adjust... 22 ADVANCED CONFIGURATION LEVEL... 25. Set Up Function Keys... 25.2 Set Up Logic... 26. Set Up Pen Functions... 28 5 CONNECTIONS & LINKS... 29 INSTALLATION Product Identification Siting Mounting Electrical Connections Installation Record Part No. IM/C900 INS PROGRAMMING General Programming Basic Config. Level Advanced Config. Level Connections & Links Part No. IM/C900 PGR OPERATION Setting Up Displays & Controls Operating Level Simple Fault Finding Part No. IM/C900 OGR SPECIFICATION SHT. Full Specification Part No. SS C900 A Standard Manuals ADVANCED SOFTWARE OPTIONS Flow Totalisation Ramp/Soak Profile Maths Functions Timer Functions Part No. IM/C900 ADV Modbus (RTU) Serial Adaptors Serial Connections Programming Pages ASCII Tables Part No. IM/C900 MOD B Supplementary Manuals Fig.. COMMANDER 900 Documentation

2 GENERAL PROGRAMMING BASIC CONFIGURATION LEVEL The programming procedures are used to make changes to the operating parameter values and for scale adjustment see Fig..2. The programming of all channels is performed using faceplate see Fig.. When changing the input type it may be necessary to reposition the input selector links accordingly see Section 5, CONNECTIONS & LINKS. 2. Preparation for Changes to the Parameters Ensure that the external alarm/control circuits are isolated if inadvertent operation during programming is undesirable. Any change to the operating parameters are implemented using the or switches see Section of the Operating Guide. Note. The instrument responds instantly to parameter changes which are saved automatically when leaving the current frame. 2.2 Security System A security system is used to prevent tampering with the programmed parameters by restricting access to programming levels, other than the OPERATOR LEVEL; all users have access to this level. A security password is used to give access to the programming pages. The password can be set to any value from 0 to 9999. The instrument is despatched with the password set to '0' see Section.5 of Operating Guide.. Set Up Input (Process Variable)... Input types Linearization Electrical ranges Engineering ranges Fault detection Digital filtering.2 Set Up Pen Range... 8 Chart ranges Event pen sources. Set Up Chart... 9 Chart duration (speed) Chart stop function Auto pen drop Pen lift. Set Up Alarms... 0 Acknowledge type Global alarm acknowledge Alarm type Trip/hysteresis/time hysteresis.5 Set Up Relay Output... Relay sources Relay polarity.6 Set Up Digital Output... 6 Digital output source Digital output polarity.7 Set Up Analog Output... 8 Retransmission sources Retransmission ranges Current output ranges.8 Digital Inputs... 20 Input polarity.9 Access Page... 2 Configurable password Internal security link.0 Scale Adjust... 22 Process variable offset adjustment Process variable span adjustment Pen calibration Mains filter Pen Linearity Check Faceplate Fig.. Location of Faceplate 2

OPrtOr LEVEL Operator Level basic CONFIG Basic Config AdVNCd CONFIG Advanced Config Set Up Inputs Section. Page SEtUP INPUt Set Up Input (PV) Select Channel IntYP Input Type (PV) LNtYP Linearizer Type rng HI _. Input Range High rng LO _. Input Range Low UNItS Temperature Units ENG HI Engineering High dec Pt Decimal Point ENG LO Engineering Low bspd Broken Sensor Drive FdLP-I _. Fault Detection Level PrGFLt _. Programmable Filter Set Up Pen Range Section.2 Page 8 SEtUP PENrNG Set Up Pen Range Select Pen PEN HI Pen Range High PEN LO Pen Range Low IN SrC In Source OUt.SrC Out Source Setup Chart Section. Page 9 SEtUP CHArt Set Up Chart CHArt Chart Time CH StP Stop Chart Signal AUtdrP Auto Pen Drop LIFt.EN Pen Lift Enable PENLFt Pen Lift/Pen Status Set Up Alarms Section. Page 2 SEtUP ALArMS Set Up Alarms ACKtYP Acknowledge Type ACKSrC Alarm Ack Source Select Alarm type Alarm Type trip Trip Level HYSt Hysteresis t HYSt Time Hysteresis Set Up Relay Output* Section.5 Page SEt UP relays Set Up Relay Select Relay SOUrCE Relay Source POLrtY Polarity Fig..2 Basic Configuration Level Set Up Digital Output Section.6 Page 6 digtal OUtPtS Digital Outputs Select Output SOUrCE Output Source POLrtY Polarity Set Up Analog Output* Section.7 Page 8 SEt UP ANALOG Set Up Output Select Output OP SrC Output Source rng HI Retrans. Range Hi rng LO Retrans. Range Low HI OP Output Range High LO OP Output Range Low * These pages do not appear on the non-upgradeable version (90J) Digital Inputs* Section.8 Page 20 digtal INPUtS Digital Inputs Select Digital I/P POLrtY Polarity Access Page Section.9 Page 2 ACCESS PAGE Access Page C PASS Configure Password Scale Adjust Section.0 Page 2 SCALE AdJUSt Scale Adjust Select PV/Pen/Filter reset Scale Reset OFFSEt Offset Adjustment SPAN Span Adjust SEtPEN At00 Cal. Pen At 00% SEtPEN At 0 Cal. Pen At 0% CHECK._ Check Pen Cal. FILtEr Hrt Mains Filter BASIC CONFIGURATION LEVEL

BASIC CONFIGURATION LEVEL. Set Up Input (Process Variable) Information. Universal inputs mv, ma, V, THC, RTD and resistance. Internal cold junction compensation. Linearization of temperature sensors to allow use of non-linearizing transmitters or any electrical input. Programmable fault levels and actions. Digital filter to reduces the effect of noise on inputs. Example A setting up: a current input of to 20 ma displaying a range of 0 to 200psi a fault detection level 0% above 200psi (engineering/display range) and 0% below 0psi (engineering/display range) in the event of a fault being detected and/or the fault detection level being exceeded the process variable is driven downscale. Input Type Linearizer Type Temp. Units Electrical Range Engineering Range (Display Range) Broken Sensor Protection Drive Programme Filter RTD THC Current Voltage Millivolts Low resistance High resistance 5/2 /2 RTD THC N THC B THC E THC L THC J THC T THC S THC R THC K None F C None 220 20.0 (Input Range High).0 (Input Range Low) Fault Detection Levell0 % Fault Detection Level 0 % Upscale Downscale Value set to 0 Value set low Value set high Example B setting up: a Type K thermocouple displaying temperature in F displaying a range of 0 to 2000 F a fault detection level 0% above 2000 F (engineering/display range) and 0% below 0 F (engineering/display range) in the event of a fault being detected and/or the fault detection level being exceeded the process variable is driven upscale. 20 200 0 Input Type Linearizer Type Temp. Units Engineering Range (Display Range) Broken Sensor Protection Drive Programme Filter RTD THC Current Voltage Millivolts Low resistance High resistance 5/2 /2 RTD THC N THC B THC E THC L THC J THC T THC S THC R THC K None F C None 2200 200 Fault Detection Levell0 % 2000 0 Fault Detection Level 0 % Upscale Downscale Value set to 0 Value set low Value set high

BASIC CONFIGURATION LEVEL. Set Up Input (Process Variable) SEt UP INPUt PV PV PV 2 PV Page Header Set Up Input (Process Variable) To advance to Set Up Pen Range Page press the Select Channel Select the channel to be programmed: PV Channel PV 2 Channel 2 PV Channel PV Channel switch. INtYP rtd tcpl VOLt LO OHM HI OHM MAMP MVLt LNtYP 5/2 /2 SQrt rtd tc b tc N tc E tc J tc t tc S tc r tc K Note. In the remaining frames press the Input Type (Process Variable) switch to view the channel selected. Caution. Ensure the correct input link positions are selected and the input is wired correctly see Section 5, CONNECTIONS & LINKS. Select the input type required: rtd Resistance thermometer tcpl Thermocouple VOLt Voltage LO OHM Low resistance ( 750Ω) HI OHM High resistance (>750Ω) MAMP Current MU.Lt Millivolt ( 50mV) None Linearizer Type Select the linearizer type required: 5/2 x 5/2 /2 x /2 Open channel flow applications SQrt Root Square rtd Resistance thermometer tc b Type B thermocouple tc N Type N thermocouple tc E Type E thermocouple tc J Type J thermocouple tc t Type T thermocouple tc S Type S thermocouple tc r Type R thermocouple tc K Type K thermocouple No linearizer Continued on next page. 5

BASIC CONFIGURATION LEVEL. Set Up Input (Process Variable) Input Type rtd or tcpl rng HI 20.0 Input Range High Set the maximum electrical input value required (in electrical units). Note. The value set must be within the limits detailed in the table below. Input Type Range Low Min. Range High Max. Min. Range (Low to High) Millivolts 0 50 5.0 Volts 0 5 0. Milliamps 0 50.0 Resistance Low 0 750 20 Resistance High 0 9999 00 rng LO.0 UNItS deg F deg C ENG HI 000 Input Range Low Set the minimum electrical input value required (in electrical units). Note. The value set must be within the limits detailed in the above table. Temperature Units Select units required. Engineering Range High Set the maximum engineering (display) value required. Note. The value set must be within the limits detailed in the tables below. Linearizer Degrees Fahrenheit Degrees Celsius Type Min. Max. Min. Span Min. Max. Min. Span Type B 0 272 278 8 800 70 Type E 8 652 8 00 900 5 Type J 8 652 90 00 900 50 Type K 8 272 7 00 00 65 Type N 28 272 62 200 00 90 Type R & S 0 092 576 8 700 20 Type T 8 572 08 250 00 60 RTD 28 2 5 200 600 25 Performance accuracy is not guaranteed below 725 F/00 C for types B, R and S thermocouples. Minimum span below zero Type T 26 F/70 C Minimum span below zero Type N 89 F/05 C THC standard DIN 70 IEC 58 RTD standard DIN 760 IEC 75 5/2 /2 Square Root None Linearizer Type Continued on next page. Engineering Range High and Low Min. Max. 9999 +9999 6

BASIC CONFIGURATION LEVEL. Set Up Input (Process Variable) decpt 000 ENG LO 0 bspd UP dn FdLP Decimal Point Set the decimal point position required for both the engineering range high and engineering range low values. Engineering Range Low Set the minimum engineering (display) value required, Note. The value set must be within the limits detailed in Engineering Range High tables opposite. Broken Sensor Protection Drive In the event of a fault being detected on the input and/or if the Fault Detection Level Percentage is exceeded (see next frame), the process variable is driven in the direction of the drive selected. Select the broken sensor drive required: No drive UP Upscale drive dn Downscale drive. Fault Detection Level Percentage A fault level percentage can be set to detect a deviation above or below the display limits. For example, if set at 0.0%, then if an input goes more than 0% above Engineering Range High or more than 0% below Engineering Range Low, a fault is detected. On some ranges the input circuitry may saturate before the fault level set is reached. In this case an error is detected below the level set. Set the level required, between 0.0 and 00.0% of engineering span (range low to high) in 0.% increments. Note. If an input exceeds the minimum or maximum value for the linearizer selected an error is detected regardless of any fault level. PrGFLt Programmable Filter Filters the process variable input, i.e. if the input is stepped it smooths the transition between steps and may also be used for some degree of cleaning of noisy inputs. The filter time represents the time a step in the input takes to change the displayed process variable from 0 to 90% of the step. Set the value required, between 0 and 60 in second increments. Return to Select Channel frame. 7

BASIC CONFIGURATION LEVEL.2 Set Up Pen Range/Event Source Information. Trend pens have an independent chart range allowing a selected part of the engineering (display) range to be used for extra resolution on the chart. Three position event pen function can be driven by digital inputs, alarms, logic equation results and real time events (when timer option is fitted). Select Pen Range (in engineering units) Select 'In' Source Select 'Out' Source Event Pen Chart Position *Pen at 80% 000 (Eng. Range High) 700 (Pen Range High) *Source on Source on Pen at 60% Pen 2 at 0% Pen at 20% Source off Source off 00 (Pen Range low) 0 (Eng. Range Low) *In source takes priority if both sources enabled Record Function Event Function *With Real Time Event Pen option fitted, Pen is above 00% SEt UP PENrNG Page Header Set Up Pen Range To advance to Set Up Chart Page press the switch. RECORD EVENT PEN PEN PEN 2 PEN PEN HI PEN LO IN SrC EQN NONN OUtSrC EQN NONN Select Pen Select the pen to be programmed Note. In the remaining frames press the switch to view the pen selected. Record (trend) or event pen function is set in the ADVANCED CONFIGURATION LEVEL (if True Time Event Pen option is selected, the fourth pen is fitted with a special pen arm and is set automatically for event pen function) see Section., Set Up Pen Functions. Pen Range High Set the maximum value required on the chart, in engineering units (the value must be within the engineering range set in Set Up Input Page see Section.). Pen Range Low Set the minimum value required on the chart, in engineering units (the value must be within the engineering range set in Set Up Input Page). In Source Select a source to move the pen inwards on the chart. For a description of sources see Table. on page 5. Out Source Select a source to move the pen outwards on the chart. For a description of sources see Table. on page 5. Return to Select Pen frame.

BASIC CONFIGURATION LEVEL. Set Up Chart Information. Programmable chart duration between and 67 hours or 7 and 2 days. Chart stop function the chart can be stopped by an alarm, digital input, logic equation result or a real time event (if timer option is fitted). Auto pen drop automatically drops the pen(s) onto the chart after a 5 minute delay to ensure recording is not left disabled inadvertently. SEt UP CHArt Page Header Set Up Chart To advance to Set Up Alarms Page press the switch. CHArt 2dY Hr Chart Duration Select the chart duration required per revolution of the chart; between and 67 hours or 7 and 2 days. CH StP EQN Stop Chart Source Select the source required for stopping the chart. For a description of sources see Table. on page 5. SEt UP CHArt AUtdrP LFt EN Auto Pen Drop Select 'YES' to enable or 'NO' to disable. If 'YES' selected, pen(s) drop automatically onto the chart 5 minutes after they are lifted. If 'NO' selected, the pen(s) remain lifted until they are manually dropped by the operator. Pen Lift Enable/Disable The switch can be disabled if required. Select 'YES' to enable or 'NO' to disable. PENLFt record drop LIFt return PArK At ref Pen Lift/Pen Status To raise pen(s) press switch. The following status displays are shown: record pen records on chart LIFt pen lifts off chart PArK pen moves to park position At ref pen at reference position To lower pen(s) press switch. The following status displays are shown: return pen returns to record position drop drops (lowers) onto chart record pen records on chart Return to top of Set Up Chart Page.

BASIC CONFIGURATION LEVEL. Set Up Alarms Information. Four alarms per channel identified A to D (for channel ) up to A to D (for channel ). Three operator acknowledge options. Global alarm acknowledgment by digital input, alarm, logic equation result or real time event (if option fitted). High/low process alarms. Fast/slow rate of change of process variable alarms. Adjustable hysteresis value to prevent oscillation of alarm state. Time hysteresis to allow delayed triggering of alarms. Hysteresis Trip Point Hysteresis High Process Low Process Process Variable Alarm On Alarm Off Alarm On Alarm Off Fig.. High and Low Process Alarm with Hysteresis Alarm Trip Point Output 70 0 Alarm On 0 Time in seconds 0 0 Alarm Off Counter Started Counter Reset Counter Started Hysteresis Time Elapsed Counter Reset Time Hysteresis Status Example shows time hysteresis set to 70 seconds used with a high process alarm Fig.. Time Hysteresis Alarm 0

BASIC CONFIGURATION LEVEL. Set Up Alarms 9.5 hour The maximum time it takes to detect an alarm condition is present (T), in seconds, is calculated as follows: 800 T = 0.8 + x 2 Trip Value The time it takes for the alarm state to be cleared once the alarm condition has been removed is also equal to T. 0. 0. hour hour hour 9.5 Alarm On Alarm On Alarm Off Alarm Off Falling Fast Rate T T Rising Fast Rate T T T 0. The maximum time it takes to detect an alarm condition is present (T), in seconds, is calculated as follows: 800 T = 0.8 + x 2 Trip Value 0. hour The time it takes for the alarm state to be cleared once the alarm condition has been removed is also equal to T. hour 9.5 hour hour 9.5 Alarm On Alarm Off Falling Slow Rate T T T Alarm On Alarm Off Rising Slow Rate Examples shown are for a trip value of 0%/hour on a PV engineering range of 0.0 to 00.0 T = 0.8 + 800 x 2 0 T = 82 seconds Fig..5 Slow Rate Alarms with Hysteresis Examples shown are for a trip value of 0%/hour on a PV engineering range of 0.0 to 00.0 T = 0.8 + 800 x 2 0 T = 82 seconds Fig..6 Fast Rate Alarms with Hysteresis

BASIC CONFIGURATION LEVEL. Set Up Alarms SEt UP ALArMS ACKtYP LAtCH NOrMAL Page Header Set Up Alarms To advance to Set Up Relay Output page press the switch. Alarm Acknowledge Type Alarms may be acknowledged while they are displayed. Select the alarm acknowledge type: no acknowledge facility. If the cause of the alarm no longer exists, the alarm state and display are cleared automatically. Alarm cause L.E.D. Alarm State Present Flashing Active Not Present Off Inactive NOrMAL and LAtCH if the cause of the alarm no longer exists, the alarm display remains until it has been acknowledged. Alarm cause Acknowledge L.E.D. Alarm State Present No Flashing Active Present Yes Steady Active Not Present Previously acknowledged Off Inactive Present No Flashing Active Not Present No Flashing Active/Inactive* Not Present Yes Off Inactive *Alarm state is active if LAtCH is selected or inactive if NOrMAL is selected ACKSrC ALM d Global Alarm Acknowledge Source Select the alarm acknowledgment source required. For a description of sources see Table. on page 5. ALM d Select Alarm Select the alarm to be programmed. Ch. Ch. Ch. 2 Ch. ALM A ALM d ALM A ALM d2 ALM A2 ALM d ALM A Note. In the remaining frames press the switch to view the alarm selected. Continued on next page. 2

BASIC CONFIGURATION LEVEL. Set Up Alarms OFF type S rate F rate LO PrC HI PrC OFF Alarm Type Select the alarm type required for the alarm selected. HI PrC high process LO PrC low process F rte fast rate (rate of change of process variable) S rte slow rate (rate of change of process variable) OFF alarm off trip Trip Level Set the trip value required for the alarm selected. The following are displayed in engineering units: HPrC, LPrC. The following are displayed as a percentage of the engineering span (engineering range high engineering range low) per hour between ±0.5 and ±500%: FrtE and SrtE. HYSt t HYSt Hysteresis Hysteresis is operational when the alarm is active. Set the hysteresis value required for high/low process, in engineering units (within the engineering range) or in 0.% increments for rate alarms. The alarm is activated at the trip level but is only turned off after the alarm variable has moved into the safe region by an amount equal to the hysteresis value. For rate alarms this setting is a percentage of the trip rate see 'FrtE' and 'SrtE' in previous frame. Time Hysteresis Set the time hysteresis value required between 0 and 9999 seconds. Note. The alarm condition must be present continually for the time set, before the alarm becomes active. If a hysteresis level is also set, the alarm condition remains active until the process variable moves outside the hysteresis band. When the alarm condition no longer exists the alarm becomes inactive, i.e. time hysteresis does not affect turning off of alarm states. Return to Select Alarm frame.

BASIC CONFIGURATION LEVEL.5 Set Up Relay Output Information. Relay Output omitted on 90J (non-upgradeable version). Relays can be energized by alarms, logic equation results, digital inputs, real time events (timer option ) and totalizer wrap signal (totalizer option). External Totalizer count function external counter can only be driven by module type ( relays module) fitted in module positions, 5 and 6. Polarity to allow failsafe settings. Select Relay Output Relay Source Polarity Selection Source State Polarity Relay State Relay Contacts Relay 5. Relay 5.2 Relay 5. Relay 5. Module Position Relay No. Alarm Acknowledge Logic Equation Logic Equation 8 Digital Input Digital Input 2 Alarm A Active Alarm A Active Positive Negative Energized De-energized NC C NO NC C NO Alarm A Alarm D Alarm A Inactive Positive De-energized NC C NO None Alarm A Inactive Negative Energized NC C NO SEt UP relays Page Header Set Up Relays To advance to Set Up Digital Output Page press the switch. rly. Module Position Relay No. Select Relay Output Select the output to be programmed. The selections in this frame relate to the number of fitted modules with relays and their relative module positions. Example for a type (four relays) module fitted in position five the following selections are also programmable: relay 5. (position 5, relay ) relay 5.2 (position 5, relay 2) relay 5. (position 5, relay ) relay 5. (position 5, relay ) Note. In the remaining frames press the switch to view the relay selected. SOUrCE EQN Relay Source Select the source required to activate the selected relay. For a description of sources see Table. on page 5. Note. To drive an external counter COUNt.x must be selected. Continued on next page

BASIC CONFIGURATION LEVEL.5 Set Up Relay Output POLrtY POStVE NEGtVE Polarity The polarity selection is used to invert the effect of the digital source state on the relay state as shown in the following table: Source State Polarity Relay State Active Positive Negative Energized De-energized Non-active Positive Negative De-energized Energized Select the polarity required Caution. Check connections before operating see Section 5, CONNECTIONS & LINKS. Return to Select Relay Output frame. Source AL_ACK Description Alarm Acknowledge Unacknowledged process alarm anywhere in the unit timer.2 timer. Real time event 2 Real time event Real time events (only available if timer option fitted see Advanced Software Options Manual). EQN EQN EQN 2 EQN rap * COUNt. rap * COUNt. dig 6.8 Programmable logic equation Programmable logic equation Programmable logic equation 2 Programmable logic equation Wrap around on total Total external counter drive Wrap around on total Total external counter drive Digital Input 6.8 Programmable logic equations see Section.2, Set Up Logic Wrap around and count (only available if totalizer option fitted) dig. AL d AL C AL b AL A AL d AL C AL b AL A Digital input. Alarm D Alarm C Alarm B Alarm A Alarm D Alarm C Alarm B Alarm A Digital Input number Module number Channel Alarms (if applicable) Channel Alarms (if applicable) AL d2 AL C2 AL b2 AL A2 AL d AL C AL b AL A Alarm D Alarm C Alarm B Alarm A Channel 2 Alarms (if applicable) Alarm D Alarm C Channel Alarms Alarm B Alarm A No source required * Only available on -relay and 8-digital output modules (types and 5), fitted in module positions,5 and 6. Table. Description of Sources 5

BASIC CONFIGURATION LEVEL.6 Set Up Digital Output Information. This page is not displayed if there are no digital outputs fitted. Up to 2 digital outputs are available depending on the module types fitted. Digital outputs can be energized by alarms, logic equations results, digital inputs, real time events (timer option) and totalizer wrap signal (totalizer option). External Totalizer count function external counter can only be driven by module type 5 (8 digital outputs module) fitted in module positions, 5 and 6. Polarity inverts the effect of the selected source on the output state. Select Digital Output Page 6 Header Output 5. Output 5.2 Output 5. Output 5. Output 5.5 Output 5.6 Output 5.7 Output 5.8 Module Position Output No. Digital Source Alarm Acknowledge Logic Equation Logic Equation Digital input Digital input 2 Alarm A Alarm D None Polarity Selection Source State Polarity Output State Digital Input Active Positive Energized Digital Input Active Digital Input Inactive Negative Positive De-energized De-energized Digital Input Inactive Negative Energized 6

BASIC CONFIGURATION LEVEL.6 Set Up Digital Output digtal OUtPtS OUt. Module Position Output No. SOUrCE EQN POLrtY POStVE NEGtVE Page Header Set Up Digital Outputs to advance to Set Up Analog Output page press the switch. Select Digital Output Select the output to be programmed the selections in this frame relate to the number of fitted digital output modules and their relative module positions. Example for a type 5 (eight digital outputs) module fitted in position five the following selections are also programmable: OUt 5. (position 5, output ) OUt 5.2 (position 5, output 2) OUt 5. (position 5, output ) OUt 5. (position 5, output ) OUt 5.5 (position 5, output 5) OUt 5.6 (position 5, output 6) OUt 5.7 (position 5, output 7) OUt 5.8 (position 5, output 8) Note. In the remaining frames press the Output Source Select the source required to activate the selected digital output. For a description of sources see Table. on page 5. switch to view the output selected. Note. To drive an external counter COUNt.x must be selected. Polarity The polarity selection is used to invert the effect of the source state on the output as shown in the following table: Active Source State Polarity Output State Non-active Positive Negative Positive Negative Energized De-energized De-energized Energized Select the polarity required Caution. Check connections before operating see Section 5, CONNECTIONS & LINKS. Return to Select Digital Output frame. 7

BASIC CONFIGURATION LEVEL.7 Set Up Analog Output Information. Analog Output omitted on 90J (non-upgradeable version). Fitted analog outputs assignable to retransmit any process variable. Selectable retransmission range allows maximum resolution on range of interest. Adjustable output range for non-standard and reversed outputs. Note. The example below shows analog output set to retransmit part of process variable 's engineering range (250 to 750 C) as a.0 to 20.0mA current output. Select Analog Output Select Output Source Setting Output Ranges 000 C (Engineering Range High) Output Output 6 PV PV2 PV PV 750 C (Retransmission Range High) Range to be transmitted 250 C (Retransmission Range Low) 20.0mA (Output Range High).0mA (Output Range Low) 0 C (Engineering Range Low) 8

BASIC CONFIGURATION LEVEL.7 Set Up Analog Output SEt UP ANALOG OP OP x OP SrC PV PV rng HI Page Header Set Up Analog Output To advance to Digital Inputs Page press the switch. Select Analog Output Select the analog output to be programmed. The selections in this frame relate to the number of fitted modules with analog output. Example Output is the analog output in position (fitted on the main board), output is the analog output fitted in module position. Note. In the remaining frames press the switch to view the analog output selected. Output Source Select output source required. The selections in this frame correspond to the channels on the instrument (as available) PV (channel ), PV2 (channel 2) etc. Retransmission Range High Set the engineering range value (in engineering units) at which maximum output is required. rng LO HI OP. Retransmission Range Low Set the engineering range value (in engineering units) at which minimum output is required. Output Range High Set the maximum current output required for the Retransmission Range programmed between 2.0 and 20.0mA. LO OP. Output Range Low Set the minimum current output required for the Retransmission Range programmed between 2.0 and 20.0mA. Return to Select Analog Output frame. 9

BASIC CONFIGURATION LEVEL.8 Digital Inputs Information. Digital Input omitted on 90J (non-upgradeable version). Up to 0 digital inputs are available depending on the module types fitted. Volt-free contacts or TTL levels. Polarity sets the logic state (unchanged or inverted) for the module position(s). Select Digital Input Input State Polarity Selected Logic State Position Position 2 Position Position Position 5 Position 6 Switch Input (volt-free) or or 5V 0V 5V 0V Logic Input (TTL) Negative Positive Negative Positive Input Active Input non-active Input non-active Input Active digtal INPUtS Page Header Digital Inputs To advance to Access Page press the switch. POSN Select Digital Input Select digital module position to be programmed. POSN x Note. In the remaining frames press the switch to view the module selected. POLrtY POStVE NEGtVE Polarity Select the polarity required for the module position selected above: POStVE logic input state unchanged NEGtVE logic input state inverted Return to Select Digital Input frame. 20

BASIC CONFIGURATION LEVEL.9 Access Page Information. Configurable password protection of PROGRAMMING LEVELS. Internal security link enable/disable password protection. ACCESS PAGE C PASS Page Header Access Page. To advance to Scale Adjust Page press the Configuration Password Prevents access to the Programming Pages. Set the required password, between 0 and 9999. switch. Return to top of Access Page. Without Correct Security Code With Correct Security Code Operating Pages SECODE Enter Security Code (programmed innaccess Page) OPrtOr LEVEL LK Enable Security position, allows access to CONFIGURATION LEVELS with correct security code. basic LEVEL AdVNCd LEVEL Position Position 2 Position Position Position 5 Position 6 LK Disable Security position, allows unprotected access to CONFIGURATION LEVELS. Fig..7 Use of Security Code in Operator Level Fig..8 Location of Security Link 2

BASIC CONFIGURATION LEVEL.0 Scale Adjust Information. Analog Inputs do not require re-calibrating when the input type or range is changed. Process variable adjust reset removes any previously programmed offset or scale adjustment settings. System offsets errors can be removed using process variable scale offset adjustment. System scale errors can be removed using process variable span adjustment. Process variable offset/span adjustment can be used to perform spot calibration Pen(s) can be independently calibrated and checked across the full range of the chart. Mains filter selectable for maximum noise rejection. Pen Linearity Check automatically draws a pen linearity test pattern. Scale Adjustment Chart Display PV (x) Span Adjust (+) Offset PV AND 200. CH AL AL2 AL AL CH2 CH CH Offset Adjustment Engineering Range Display Offset Adjustment Display 250.0 C OFFSEt 99.8 CH AL AL2 AL AL OFFSEt 00.0 CH AL AL2 AL AL CH2 CH2 CH CH CH CH 00 C 50.0 C Span Adjustment Engineering Range Display Span Adjustment Display 250.0 C SPAN 225.5 CH AL AL2 AL AL SPAN 225.0 CH AL AL2 AL AL CH2 CH2 CH CH CH CH 225 C 50.0 C Note. As a general rule: use Offset adjustment for spot calibration at <50% of engineering range span. use Span adjustment for spot calibration at >50% of engineering range span. 22

BASIC CONFIGURATION LEVEL.0 Scale Adjust SCALE AdJUSt Page Header Scale Adjust To advance to BASIC CONFIGURATION LEVEL frame use the switch. LINCHK FILtEr PEN PEN PV PV reset YES NO OFFSEt Select Process Variable/Pen Select linearity check, process variable or pen required: LINCHK the pens automatically draw a test pattern to check pen linearity. done is displayed on completion FILtEr mains frequency filter PEN x pens to PV process variable on channel PV process variable on channel PV 2 process variable on channel 2 PV process variable on channel None Note. In the remaining frames press the selected. switch to view the process variable or pen Process Variable Scale Adjustment Reset Set YES to reset the process variable offset and span values to their nominal values (values are reset when frame is exited). Process Variable Offset Adjustment Electrical and resistance thermometer inputs: apply the correct input for the spot calibration required. RTD inputs: use resistance values obtained from standard tables. Thermocouple Inputs: measure the ambient temperature at the output terminals of the signal source (calibrator). From thermocouple tables obtain the millivolt equivalent of this temperature (a) and that for the spot calibration temperature (b). Subtract (a) from (b) and set the signal source to the resultant value. (The voltage is negative if the spot calibration temperature is below the measured ambient temperature). Note. The displayed units are engineering units. Set the value required. The decimal point position is set automatically. Example If the display range is 50.0 to 250.0 and a spot calibration is required at 00 and 225, inject a signal equivalent to 00 and set the display to 00.0 using the and switches SPAN Span Adjust Proceed as for Offset Adjustment above and apply the correct input for the spot calibration required. The displayed units are engineering units. Set the value required. The decimal point is set automatically. For the example above, inject a signal equivalent to 225 and the set the display to 225.0. Continued on next page. 2

BASIC CONFIGURATION LEVEL.0 Scale Adjust PEN PEN SEtPEN At 00 SEtPEN At 0 CHECK Calibrate Pen At 00% Drives the pen automatically to the full scale position on the chart. Use the and switches to set pen to 00% on the chart. Calibrate Pen At 0% Drives the pen automatically to the zero position on the chart. Use the and switches to set pen to 0% on the chart. Check Pen Calibration The pen calibration can be checked at any point on the chart. Use the and switches to move the selected pen from the zero point up to the 00% position on the chart. FILtEr FILtEr 60 Hrt 50 Hrt Note. If the true time event option is fitted the red pen does not move beyond the 9% position on the chart. Select Filter Select the mains frequency of the supply used to ensure maximum noise rejection on analog inputs. FILtEr Return to Select Process Variable/Pen frame. 2

ADVANCED CONFIGURATION LEVEL. Set Up Function Keys... 25.2 Set Up Logic... 26. Set Up Pen Functions... 28 OPrtOr LEVEL Operator Level basic CONFIG Basic Config AdVNCd CONFIG Advanced Config Set Up Functions Keys Section. Page 25 Set Up Logic Section.2 Page 26 Set Up Pen Functions Section. Page 28 SEt UP F KEYS Set up Function Keys F KEY Function Key F KEY 2 Function Key 2 SEtUP LOGIC Set Up Logic ECN. Select Equation EQN. _. Configure Logic EQN. _.7 Configure Logic PEN FUNCtN Pen Functions PEN _ Pen Function PEN _ Pen Function Fig.. Advanced Configuration Level. Set Up Function Keys Information. Programmable function key on each faceplate Home function returns the instrument display to the start of the operating page when at the top of any page. Global alarm acknowledge function acknowledges any unacknowledged alarms on all channels. SEt UP F KEYS Page Header Set Up Function Keys To advance to the Set Up Logic press the switch. F KEY HOME ALMACK Function Key Select function required. HOME Home (return to Operating Page in OPERATING LEVEL) ALMACK Acknowledge alarm F KEY2 ALMACK Function Key 2 Select function required (if applicable). Return to Set Up Function Keys frame. 25

ADVANCED CONFIGURATION LEVEL.2 Set Up Logic Information. logic equations 7 elements per equation OR/AND operators Can combine internal and external digital signals i.e. alarms, digital inputs, other logic equation results and real time events (timer option). For each equation, the logic elements to 7 are arranged sequentially, as shown below. Odd numbered elements are used for logic inputs and even numbered elements for logic gates. Logic inputs must be set to one of the digital sources listed in Table. on page 5. Logic gates must be set to ANd, Oror End. Setting an element to End terminates the equation. Logic Inputs EQN. EQN. EQN.5 EQN.7 EQN.2 ANd/Or EQN. ANd/Or EQN.6 ANd/Or Result Note. Elements on each equation are calculated sequentially, i.e. elements, 2 and are evaluated first and this result is then combined with elements and 5. Similarly, this resultant is then combined with elements 6 and 7 to give the logic equation result. Example Reservoir level monitoring using: process variable with an engineering range 0 to 00 feet logic equation result assigned to relay. which is used to operate the control valve. Manual Overide Control Valve Relay Output Alarm B 80 feet 50 feet Alarm A Level Rate of change > 0ft/h/ Time Level Sensor Alarm C Flow Conditions Input Elements Entering the Logic Equation Close reservoir control valve if: Alarm A set to high process trip at 50 ft Reservoir level >50 feet AND rate of change >0 ft/hr OR Reservoir level >80 ft OR Manual override switch operated Alarm B set to high process trip at 80 ft Alarm C set to fast rate trip at 0% of range per hour (0 ft/hr) Manual override switch: Connected to digital input. Digital input number Module number Negative polarity Volt-free switching AL A AL C AL B dig. EQN. EQN. EQN.5 EQN.7 EQN.2 ANd EQN. Or EQN.6 Or 26

ADVANCED CONFIGURATION LEVEL.2 Set Up Logic SEt UP LOGIC EQN EQN2 EQN EQN EQN - EQN Equation No. Element No. Page Header Set Up Logic To advance to Set Up Pen Functions Page press the Select Equation Select equation to be constructed. In the remaining frames press the Equation n/element Select the source required for element. switch. switch to view the equation selected. For a description of sources see Table. on page 5. EQN-2 Or ANd ENd END EQN - EQN Equation n/element 2 Select the operator required to combine elements and : Or Or ANd And ENd Ends equation Equation n/element Repeat previous two steps for elements to 7. Odd numbered elements = sources Even numbered elements = operators Return to Select Equation frame. 27

ADVANCED CONFIGURATION LEVEL. Set Up Pen Functions Information. Any fitted pen can be assigned to a trend or an event function. PEN FUNCtN Page Header Pen Functions To advance to Advanced Configuration frame press the switch. PEN FUNCtN PEN trend EVENt Pen Select pen function required: trend Trend pen EVENt Event pen Note. The event pen and true time line event pen are separate functions and only the event pen can be selected in this page. The true time line event pen option allows event marking on the same time line as the red pen and requires a special pen arm and motor assembly. Refer to the order code in the Specification Sheet. PEN 2 PEN Pen 2 to Repeat as for Pen (if applicable). Return to top of Set Up Pen Functions Page. 28

Main Input, Standard Input & Analog + Relay Relays Module 8 Digital Inputs/Outputs Module + NC Common Common Analog Output* 2 2 NO Relay 2 Input Output C Input 2 Output 2 NC Input Output Analog Input 5 5 NO Relay 2 5 Input Output 6 6 C 6 Input 5 Output 5 7 8 9 Common Logic Logic 2 Logic Inputs* 7 8 9 NC NO C Relay 7 8 9 Input 6 Input 7 Input 8 Output 6 Output 7 Output 8 0 Normally Open 0 NC 0 Common Common 2 Common Normally Closed Relay Output 2 NO C Relay 2 2 2 * Not fitted on Analog + Relay Module Digital I/Ps Digital O/Ps Digital I/O Module PL2 5 CONNECTIONS & LINKS 5 6 6 THC & mv + 6 Voltage + Current + RTD (-lead) Red White 5 Red 6 RTD (2-lead) & Resistance 5 6 White Red Link 2-wire Transmitter Tx 6 + * 5 5 5 Main Board Position Position 2 Position Position Position 5 Position 6 8 PL 8 PL 8 PL 2 PL8 2 PL8 2 PL8 I/O Module 5 8 PL 5 8 PL 5 8 PL 2 PL 2 PL 2 PL 29

PRODUCTS & CUSTOMER SUPPORT A Comprehensive Instrumentation Range Analytical Instrumentation Transmitters On-line ph, conductivity, and dissolved oxygen transmitters and associated sensing systems. Sensors ph, redox, selective ion, conductivity and dissolved oxygen. Laboratory Instrumentation ph and dissolved oxygen meters and associated sensors. Water Analyzers For water quality monitoring in environmental, power generation and general industrial applications including: ph, conductivity, ammonia, nitrate, phosphate, silica, sodium, chloride, fluoride, dissolved oxygen and hydrazine. Gas Analyzers Zirconia, katharometers, hydrogen purity and purge-gas monitors, thermal conductivity. Controllers & Recorders Controllers Digital display, electronic, pneumatic. Discrete singleloop and multi-loop controllers which can be linked to a common display station, process computer or personal computer. Recorders Circular and strip-chart types (single and multi-point) for temperature, pressure, flow and many other process measurements. Electronic Transmitters Smart & Analog Transmitters For draft, differential, gauge and absolute pressure measurement. Also, liquid level and temperature. Customer Support ABB Instrumentation provides a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre. United Kingdom ABB Kent-Taylor Limited Tel: + (0)80 752 Fax: + (0)80 70787 United States of America ABB Instrumentation Inc. Tel: + 76 292 6050 Fax: + 76 27 6207 Italy ABB Kent-Taylor SpA Tel: +9 (0) 58 Fax: +9 (0) 58278 Client Warranty Prior to installation, the equipment referred to in this manual must be stored in a clean, dry environment, in accordance with the Company's published specification. Periodic checks must be made on the equipment's condition. In the event of a failure under warranty, the following documentation must be provided as substantiation:. A listing evidencing process operation and alarm logs at time of failure. 2. Copies of operating and maintenance records relating to the alleged faulty unit. I to P Converters and Field Indicators Flow Metering Magnetic Flowmeters Electromagnetic, insertion type probes and watermeters. Turbine Flowmeters Wedge Flow Elements Mass Flow Meters Transmitters, sensors, controllers and batch/display units. Level Control Submersible, Capacitance & Conductivity. Pneumatic Instrumentation Transmitters Indicating Controllers Recording Controllers

The Company s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. ABB 998 Printed in UK (0.98) IM/C900 PGR Issue 6 ABB Kent-Taylor Ltd. St. Neots Cambs. England, PE9 EU Tel: + (0) 80 752 Fax: + (0) 80 2798 ABB Instrumentation Inc. PO Box 20550, Rochester New York 602-0550 USA Tel: + 76 292 6050 Fax: + 76 27 6207 ABB Kent-Taylor SpA 2206 Lenno Como Italy Tel: +9 (0) 58 Fax: +9 (0) 58278