Standard Operating Manual. PEL Wet Station X Clean Organic Solvent Station

Similar documents
Standard Operating Manual. Ultra Fab Wet Station F Non-standard Processing Station

Standard Operating Manual. Ultra Fab Wet Station C Oxide/ Nitride Etch Station

Standard Operating Manual. Semitool Spin Rinser Dryer SRD

Standard Operating Manual

Standard Operating Manual

R I T Title: RCA Bench #2

Amerimade RCA Wet Bench SOP Page 1 of 13 Rev /22/2014

R I T Title: Manual Processing Bench #2 Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 08/05/2008

Wet Bench Standard Operating Procedures

Operation Procedure for SMiF Wet Hoods (Solvent, Spin Coat, Acid, and Develop Hoods)

Title: Nitride and Silicon Wet Etch Semiconductor & Microsystems Fabrication Laboratory Revision: I Rev Date: 01/20/2010

Standard Operating Manual

Polypro Acid Wet Bench UCLA Nanolab Yellow Room. Primary contacts: Max Ho, Wilson Lin. PREPARED BY: Steve Franz DATE Rev 2/01/12

Liftoff/Strip Solvent Bench SOP

Revised: January 8, Goal: Deposit, expose, and develop standard positive photoresist films as an etch mask, liftoff mask, or sacrificial layer.

Emergency Response Cart Guide

Standard Operating Manual

General Acids/Bases Bench Right and Left SOP

GENERAL WET BENCH SAFETY AND USES SOP. October 2013 GENERAL SAFETY RULES

CNS Cleanroom Safety Orientation Form

Ultrasonic and Megasonic Cleaners

The safety refresher are mandatory for ALL users Cleanroom access and instrument usage will be denied for users who:

The University of Washington Nanofabrication Facility (WNF) at Fluke Hall 125 Fluke Hall Seattle, WA 98195

Emergency Response Guide for Laboratories

Teaching Lab Headway Spinner SOP

University of Minnesota Nano Fabrication Center

EMERGENCY RESPONSE GUIDE IMMEDIATE EMERGENCY PROCEDURES

Application Data Sheet

Experiment #6 Photolithography: Microprocessing Technology Fabrication of the microstructures with SPR photoresist

Haz Waste Management

OPERATION MANUAL Model 800 Nanoimprint Controller

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Plasma-Therm 790 RIE (Reactive Ion Etcher) Basic Operation Guide

Operation of SemiTool Spin Rinser Dryers

LABORATORY SAFETY. Discipline of Metallurgy Engineering and Materials Science Indian Institute of Technology (IIT) Indore DECLARATION

ROTOR COATING WET BENCH OPERATING PROCEDURE

MIT Emergency Response Guide

Emergency Response Procedures

HI COD REACTOR and Test Tube Heater 2008 Series

Standard Operating Procedure: Spinner

SAFE WORK PROCEDURE. List the Hazards and risk controls as per risk assessment. assessment reference:

General Procedures: Materials Synthesis Lab 1358 CNAM/Toll

Acid / Base Bench Procedures and Operation Manual

Green Chemistry Safety Training PSE/CH 335. Departments of Forest Biomaterials and Chemistry North Carolina State University

Revised: July 18, 2017

Massachusetts Institute of Technology Kavli Institute for Astrophysics and Space Research (MKI)

The Cameron Clean Room

Protocol for Chemical Spills. By Gordon Krueger Environmental Safety Specialist University of Northern Iowa Environmental Health and Safety

Rules for Safe Operation

Application: Response procedures for major and minor spills, including: organic solvents acids bases dry chemicals mercury hydrofluoric acid

The Do's and Don ts in the Instrument Sterilisation Process

MODEL 7000 SUCTION UNIT

UNSIGNED HARDCOPY NOT CONTROLLED

PDS 2010 Parylene Coater SOP

Eggo Egg Cooker and Poacher

USER MANUAL ICE MAKER

CHM 130LL: Lab Safety Name: Answer Key _ Section:

HyProCoat 330. Organo-mineral finish for chromated zinc Excellent corrosion resistance Free from tears after drying Excellent ductility.

A. General Safety LAB/ROOM: INSPECTED BY: DATE:

Chapter 3: Emergency Planning for Laboratories

LSU Inter-Institutional Biological and Recombinant DNA Safety Committee (IBRDSC)

Air Fryer. Model: AF1050 HOUSEHOLD USE ONLY

Regulations - guidance

LABORATORY CHEMICAL HYGIENE AND SAFETY PLAN

INDUCTION COOKER MODEL: EIC-G1810(BK) Owner s Manual. Please read this manual carefully before operating your set. Retain it for future reference.

SUBFAB DOC PDOC 0152 ROGER SHILE 7/11/97. SCIENCE MISSION SQUID CARRIER SUBSTRATE (Part# ) FABRICATION 1. Document Revision Record

NRF Lithography Processes SOP 8/3/2015 Rev 17 Page 1 of 24. NRF Lithography Processes SOP

Inverter Split-type Room Air Conditioner

Aluminum Heating Mantles

Portable Air Conditioner. Instruction Manual

OWNER S MANUAL DLFCAB / DLFCHB / DLFDAB / DLFDHB High Wall Ductless System Sizes 09 36

Clean! Clean working by Qdel. Qdel laminating supplies Hamburgstraat 9 NL-7575 EG Oldenzaal T: E: I :

Safety in the Chemical Laboratory

Facility Safety Rules Hazard Communication Emergency Procedures Fire Safety. Safety Orientation

Laser Class 4 Laser Type Nd-YAG Laser Manufacturer New Wave Research Laser Model SoloII-30 PIV Laser Serial / Id # Serial number 16582

1.0 Denton Thermal Evaporator

Photolithography SOPs

HOT FOOD DISPLAY Technical Service Manual

EGG COOKER AND POACHER

Schuco byallied MEDICAL ASPIRATOR. USER'S MANUAL A Caution

OPERATING INSTRUCTIONS AND SAFETY PRECAUTIONS WASHER-EXTRACTOR WARNING

EMERGENCY RESPONSE PLAN FOR

Owner's Manual TABLE OF CONTENTS

Countertop Clear Ice Maker

Laboratory Inspection Checklist

Fume Hood Safety TYPES OF FUME HOODS

Safety and Rules of the Lab

i!iifziw AUTOMATIC WASHER Large Capacity Model LA6150XS and Surface Units. Ranges, MIcrowave Ovens, Trash Compactors. Room Au Condltloners, Oehumli

Tankless Water Heater

SCIF SOP for BSL 2 samples on Aria3

ENDOSCOPE REPROCESSOR OER-Pro Operation Manual

OPERATIONS MANUAL. Hotplate Stirrer/Hotplate/Magnetic Stirrer Model HTS-1003/HTP-1002/MGS Certified

User Manual Watt Drop-In Induction Range. Model: DC /2017. Please read and keep these instructions. Indoor use only.

Bench-Top Vacuum Oven with 28 segments programmable controller

Go to the Lab! Lab Safety Begins Before You

User s Information Manual

Safety instructions and routines on lab Faculty of Odontology Malmö University

HEALTH POT MODEL: EHP-E1881G(WH) EHP-E1882G(OR) Owner s Manual. Please read this manual carefully before operating your set.

Seaward Products. OWNER S MANUAL Installation Operation - Maintenance. Princess Gourmet Model 345. Built-in Electric Oven ELECTRIC WALL OVEN

Personal Protective Equipment

Transcription:

Standard Operating Manual PEL Wet Station X Clean Organic Solvent Station Version 1.1 Page 1 of 11

Contents 1. Picture and Location 2. Process Capabilities 2.1 Cleanliness Standard 2.2 Substrate Size 3. Contact List and How to Become a Qualified User 3.1 Contact person 3.2 Training to Become a Qualified User 4. Operating Procedures 4.1 System Description 4.2 Safety Warnings 4.3 Operation Precautions and Rules 4.4 Initial Status checks 4.5 Initial System Checks 4.6 Personal Protective Equipment (PPE) 4.6.1 Sequence for Donning PPE 4.6.2 Sequence for Removing PPE 4.7 Quick Dump Rinse Operation 4.8 Aspirator Operation 4.9 EKC265 Polymer Remover Operation Procedure 4.10 Clean up 4.11 Check Out Version 1.1 Page 2 of 11

PEL Wet Station X Clean Organic Solvent Station 1. Picture and Location This tool is located at NFF Room 2240 Cleanroom Class 100. 2. Process Capabilities 2.1 Cleanliness Standard: Clean 2.2 Substrate Size: Fragments, 2 or 4 3. Contact List and How to Become a Qualified User 3.1 Emergency Response and Communications Security Control Center: 2358-8999 (24hr) & 2358-6565 (24hr) Safety Officer: Mr. Wing Leong CHUNG 2358-7211 & 64406238 Deputy Safety Officer: Mr. Man Wai LEE 2358-7900 & 9621-7708 NFF Senior Technician: Mr. Henry YEUNG 2358-7896 NFF Technician: Mr. Michael KWOK, Yiu Ming KAN or Charles TANG 23587896 3.2 Training to Become a Qualified User Please follow the procedure below to become a wet station qualified user. Version 1.1 Page 3 of 11

1. Read through the on-line equipment operating manual of the equipment; http://www.nff.ust.hk/equipment-and-process/equipment-operation-manual.ht ml 2. Attend the equipment hand-on operation training either by peer or NFF staff. 3. If training is provided by NFF staff, user must log in NFF equipment reservation system, and register these trainings. 4. Send an e-mail to Mr. Henry YEUNG requesting wet station qualified exam. 5. Pass the examination for the equipment operation and security. 4. Operating Procedures 4.1 System Description The Wet Station X is for post Deep Silicon Etch polymer residue removal for Silicon wafer(s). This Wet Station provides two sets Stainless Steel Heated bath with EKC 265 for post Deep Silicon Etch polymer residue removal X1 (Right), general purpose bath X2 (left) and two quick dump rinse baths (QDR). Exhaust Meter POWER On/Off EMO CONTROL PANEL 2 Dipper Basket General Purpose BUZZER ASPIRATOR CHEMICAL RESISTED GLOVE MS2001 QDR QDR Version 1.1 Page 4 of 11

EMO Red button for emergency stop of the wet station Power On/Off Main power on/off for wet station Manual Aspirator Aspirates FHD-5 to Neutralization Tank 4.1.1 Mitsubishi F940GOT Temperature and QDR Controller There are two controllers at this wet station. Each The MITSUBISH F-940GOT Controller is for one Stainless Steel Heated bath and one Quick Dump Rinser (QDR). QDR Control Temperature Control Heat On/Off Activate and deactivate the heater output. Timer Run Start or Restart the timer. Timer Stop/Reset Stop and Reset the timer. Alarm Mute Silence the bath alarm. Aspirator Manual Dispose FHD-5 to Neutralization Tank. PROG Enter program mode to adjust parameters. EXIT Exit this page. START Activate the dump rinser. RESET Deactivate the dump rinser, automatically reset itself in preparation for another run. HOLD Halt the operation temporarily, Manually dump the bath. 4.2 Safety Warnings 1. Follow NFF Health and Safety Manual. 2. The Emergency Off Button (EMO Button) shuts down the machine immediately. Only use the EMO in emergency situations. Emergency situations are where injury of personnel or serious damages of the system impends Version 1.1 Page 5 of 11

immediately. 3. According to the general fire emergency procedure of HKUST, please report the accident to the Security Control Center first. The nature of other emergencies will determine whether you will call police, staff, or both. If someone is injured, the 9-999 emergency number should be called before calling staff. If there is a facilities problem, such as a flood or a utility problem that does not present a danger to lab users but may result in damage to equipment, the staff or EMO need to be called. The 9-999 emergency should not be called for facility or equipment problem. Always call 9-999 when a potentially life threatening situation might exist (injury, fire, gas leak, etc.) 4. When working at the wet station, always use appropriate personal protective equipment (PPE) apron, face shield, chemical resisted gloves. After complete the process, rinse off gloves at the rinse bath, dry with wiper and put it back. 5. Do not walk away from wet station while wearing PPE. If you need to step away from the wet station at any time, rinse off gloves at the rinse bath, dry with wiper and put it back. 6. Even you wearing PPE, never put your hands or fingers into a chemical bath. 7. Never touch any surface while wearing chemical-resisted gloves that other lab members may come into contact with, such as table tops, door handles, phone, computer keyboards, face shield, apron, etc. Gloves shall be removed before touching other surfaces. 8. All spills on the wet bench during process should be cleaned up immediately. 9. Check the leakage of chemical-resisted gloves by N2 Gun every time. 10. Tanks are breakable, carefully place wafers in them, never bang a cassette on the side. 11. During process, don t remove cassette handler from cassette. 12. Never operate the heated tanks without the proper fluid level. 13. Read and understand the MSDS for the chemical being used. 14. Know the location of the nearest safety shower and eyewash station. 15. Make sure the wet station having sufficient exhaust. 16. Improper disposal of waste could result in explosion and injury. Take time to consider how to dispose in your process or contact NFF staff. 17. No acid allowed at this wet station. Version 1.1 Page 6 of 11

18. Don t dispose hot chemicals. The temperature should be below 30 C before disposing. 19. If the machine failure while being used, never try to fix the problem by yourself. Please contact NFF staff. 20. When Neutralization system has the problem, the buzzer (there is one over between Wet Station D and E) will sound. If this happens, none of the wet station may be used. 4.3 Operation Precautions and Rules 1. Do not operate the machine unless you are properly trained and approved by NFF staff. 2. All wet-processing baths are designed and arranged according to their cleanliness levels and purpose; users are not allowed to use those solutions for different purpose. 3. The Teflon cassettes, cassette handle, 2 dipper basket with two green button label for EKC, green and white button label for General Purpose are only allowed in this wet station. Clean (EKC) Clean (General Purpose) 4. All container and mask handle should be reserved in the yellow room container booking system. Cleaning and drying the container and mask handle after you have finished using it, check out and return it back to original location. 5. Do not store wafer(s) in Yellow room Wet station cassette. To store wafer(s) in your own plastic cassettes and boxes. 6. After prepare the solvent in a container, please always cover and label the Version 1.1 Page 7 of 11

container. Dispose acetone or IPA to acetone or IPA waste collector in the Wet Station Z, manual aspirate FHD-5 and AZ400K developer to Neutralization tank if you are not using it. 7. Liftoff processes are not allowed at this bath. 8. Don t leave any unnecessary things on the wet station. 9. If a wafer falls out or breaks, do not attempt to retrieve it. Contact NFF staff. 10. Close the cover after each use. 11. Process Temperatures are set and should not be changed without NFF Staff approval. 12. Please fill all the details of the log sheet attached. 4.4 Initial Status Checks 1. Please check the status of shutdown notice posted in the NFF reservation website 2. Before operate the machine, please make sure you have read check list and fill the log sheet. 3. Please check-in the equipment via Equipment Card Reader. 4.5 Initial System Checks Before starting, verify the machine is in the correct idle state: 1. If system power is off, notify NFF staff to turn on power switch on the controller. Leave system ON, after your process. 2. Check the Bath Temperature is 65 C at Right-Heated Bath, Room Temperature at Left-Heated Bath. 3. Verify that the chemical you want to use is filled to the proper level 4. Verify that the wet station has DI wafer, Nitrogen. 5. Verify that the exhaust level is normal. 6. Check log book that last user had no problem. 4.6 Personal Protective Equipment (PPE) Users are required to use additional Personal Protective Equipment (PPE) when working in the wet station. 4.6.1 Sequence for Wearing PPE Version 1.1 Page 8 of 11

PPE consists of three items that should be worn in the following order: an apron, a face shield, and chemical resisted gloves. Check all items for damage before each use, inspect the apron and face shield for damage (e.g. cracks, scratches). Gloves are to be inspected for damage and contamination. Use Nitrogen Spray Gun injects the Nitrogen to the chemical resisted glove and immerse to the DI water Bath for leakage check. Face Shield places over face and eyes, and then adjust to fit. 4.6.2 Sequence for Removing PPE PPE should be removed in the following order: chemical resisted gloves, a face shield and an apron. Rinse and dry the chemical resisted gloves, remove them and hang them up. Hang face shields and avoid damage it. Lastly remove the apron and be careful to hang it. If apron has any liquid, use a wipe to dry it. Do not leave inside out. Apron Face Shield Chemical resisted Gloves 4.7 Quick Dump Rinse Operation The purpose of QDR is to rapidly wash surface of the water and leave the wafer in a clean condition. 1. Place the wafer in the quick dump rinse (QDR) bath, initially should be full of DI water and close the lid. 2. Press the START button of UFT 48-8 quick dump rinser controller to activate the QDR. It will cycle down from 4 to 0 in the display window. Version 1.1 Page 9 of 11

3. After completion of the cycle the warning sound will be audible until the RESET button are pressed. 4. Press RESET button to automatically reset itself in preparation for another run. 5. Press HOLD button to manually dump or fill the bath. 4.8 Aspirator Operation 1. The aspirator can be activated by pressing the ASPIRATOR button located on the control panel of the wet station. The aspirator timer is set by default to thirty seconds. 2. Use the aspirator to dispose the water-soluble alkali to Neutralization Tank. 4.9 EKC265 Polymer Remover Operation Procedure 1. DuPont PlasmaSolv EKC265 TM is for post Deep Silicon Etch polymer residu Check out the container in the NFF yellow room container reservation system immediately after use. 2. Check out the equipment via Equipment Card Reader after use. 3. Check out the container in the NFF yellow room container reservation system immediately after use. 4. Check out the equipment via Equipment Card Reader after use. 5. Check out the container in the NFF yellow room container reservation system immediately after use. 6. Check out the equipment via Equipment Card Reader after use. 7. Check out the container in the NFF yellow room container reservation system immediately after use. 8. Check out the equipment via Equipment Card Reader after use. e removal. 1. Wear Personal Protective Equipment (PPE) following the Section 4.6.1 2. Open the wet station Cover. 3. Place the wafer(s) to designated Teflon Cassette. Use designated Teflon Cassette handles to attach to the notches on both end walls of Teflon cassette. Both of them are marked with two green button label. 4. Place wafer(s) in the EKC265 Polymer Remover bath (X1). Make sure that the wafer(s) are completely covered by the solvent. Press Timer Run to start the cleaning process. 5. The process time normally is 40min at 65 C. Version 1.1 Page 10 of 11

6. When the process is completely, transfer wafer(s) to QDR bath to rinse the wafer(s) with DI Water. 7. Place the cassette in the quick dump rinse bath to start the DI rinse 4 cycles following the section 4.7. 8. Take off PPE, chemical resisted gloves, face shield and then apron following the section 4.6.2. 9. Remove Cassette Handle and load cassette with wafer(s) into Spin Rinser and Dryer (SRD) for final rinse and dry cycle (Refer to SRD operation Manual). 10. You can use Nitrogen Gun to drying your wafer(s). 11. Use oven to hydrating bake to remove water moisture, if necessary. 4.10 Clean up 1. Clean up the working area of wet station after using because the chemicals may spill out during your process. Ensure the bench top is clean and dry. 2. Wash and dry all the container, and then return them to original location. 3. Leave the Quick Dump Rinser filled with D.I. Water and the lids closed. 4. Put the chemicals back to their original location after using it. 5. Do not leave cleanroom wipers on the bench top. 6. Close Wet Station Cover. 7. Fill out the log sheet. 4.11 Check Out 1. Check out the container in the NFF yellow room container reservation system immediately after use. 2. Check out the equipment via Equipment Card Reader after use. Version 1.1 Page 11 of 11