AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

Similar documents
STERITITE INSTRUCTIONS FOR USE

FLASH PAK STERILIZATION CONTAINER SYSTEM

SterilContainer S for use in the Steris V-PRO 60 Low Temperature Sterilization System

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

Open the Door to Increased Productivity and Cost Efficiency

Best Practice Guidelines For Cleaning, Disinfection and Sterilization of Critical and Semi-critical Medical Devices

Control Operating Costs. Gain Process Efficiency.

CLEANING INSTRUCTIONS

The Reliance 400/500 Glassware Washers

VHP ARD Mobile Biodecontamination System

Introducing The Reliance 200 and 250 Glassware Washing Systems.

Instrument Cleaning Chemistries Automated Washing Process

Prolystica Ultra Concentrate Instrument Cleaning Chemistries Now less is more. NEW generation instrument cleaners

Cleaning & Sterilization

AMSCO C Series Small Steam Sterilizers 16 x 16" (406 x 406mm) 20 x 20" (508 x 508 mm)

Accessories. Maybachstraße. Symmetry Fax: Revised 10/15

SteriTite - The Container System of Choice SteriTite

AMSCO Washer/ Disinfectors. Central Sterile Services DISCOVER POWERFUL WASHING WITH INCREASED EFFICIENCIES. One Integrated Approach to Healthcare.

Operator s Manual. Video Laryngoscope. ( Instructions for use ) Doc no: Revision 1.1 8th January 2013

Technical Data Monograph. The Instrument Protection Properties of Prolystica Enzymatic Presoak and Cleaner and other Enzymatic Products

Lower Your Costs. Expand Your Flexibility.

V-PRO LOW TEMPERATURE STERILIZATION SYSTEMS

Purpose. Team Roles and Responsibilities. System Requirements. Instrument Processing Training

VHP M100 Biodecontamination Systems

Lower Your Costs. Expand Your Flexibility.


Instructions for Use Spinal Punches IFU

SteriTite - The Container System of Choice SteriTite

CLEANING AND HANDLING OF WRIGHT INSTRUMENTS

Operating / User Manual

DIRECTIONS FOR USE. Keeping you insafehands to avoid needlestick injuries. see for demo video

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Instructions for care, maintenance, cleaning and sterilization of Smith & Nephew orthopaedic devices

CONVENTUS CAGE TM - DR Procedure Instruments Instrument Tray. Equipment Cleaning and Sterilization

Device Reprocessing Checklist Scoring Matrix 1=low 5=high

Christina Bradley Laboratory Manager Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Central Sterile Services Department will receive, decontaminate, process, store and/or dispense sterile and non-sterile to departments requesting.

Clinical Use Only. ResMed HumidAire 2i. Disinfection and Sterilisation Guide. English

SteriTite - The Container System of Choice SteriTite Instructions for Use

CONVENTUS CAGE TM PH Procedure Instruments Instrument Tray Cleaning and Sterilization

Reusable Laryngoscope Blades and Handles

I.F.U Re/Processing Reusable Medical Devices

Integra Salto Talaris and Integra XT Non-Sterile Device Reprocessing. Instructions for Use. 4. Manual Cleaning Instructions

AMSCO Washer-Disinfectors. Central Sterile Services DISCOVER POWERFUL WASHING WITH INCREASED EFFICIENCIES. One Integrated Approach to Healthcare

SteelcoSure Independent performance verification of your equipment and processes

Operating Manual "Mark II" Solid-State Digital Automatic Controlled Sterilizers

MARSHALLTOWN MEDICAL AND SURGICAL CENTER LAUNDRY/LINEN SERVICES. SUBJECT: Infection Control POLICY NO: 106

Cleaning and Sterilization for Instruments & Posts

AMSCO V-PRO MAX LOW TEMPERATURE STERILIZATION SYSTEM (For Use Outside U.S.A. Only)

Cleaning and Sterilisation for Burs, accessories and filling material

SUBJECT: Monitoring the performance of a sonic cleaners and/or automatic washers cleaning ability for lumens/cannulated items daily.

INSTALLATION INSTRUCTIONS. WOOD BLINDS 1-3/8, 2 and 2-3/8 SLAT SIZES. COMPOSITE FAUX WOOD BLINDS 2 and 2-1/2 SLAT SIZES

Made in USA. Sunoptic Technologies 6018 Bowdendale Avenue Jacksonville, FL 32216, USA (904)

SBS Design Considerations for Autoclave Sterilization

STERIS Instrument Cleaning Chemistries

SUGGESTED POLICY FOR THE MONITORING AN ULTRASONIC CLEANER DAILY

Commit 2 Clean TM/MC. Bloodborne Pathogens Awareness Program

idrive Ultra Powered Stapling System

PINNACLE MONITOR FOR AUTOMATED ENZYMATIC CLEANING (AEC) PROCESS

When you sterilize an instrument set, you know what the standards are.

SKLAR COATED SURGICAL INSTRUMENTS RECOMMENDED CARE & CLEANING INSTRUCTIONS

Autoclaves. Steam Autoclaves

Fiona Collins Liliana Donatelli Andrew Smith

INSTRUCTIONS FOR CLEANING AND STERILIZATION OF ALL B.BRAUN AESCULAP HAND HELD SURGICAL INSTRUMENTS AND POWER PRODUCTS (UNITED KINGDOM REQUIREMENTS)

Technical Data Monograph. AMSCO 3052 Single-Chamber Washer/Disinfector AMSCO 5052 Single-Chamber Washer/Disinfector

STANDARD OPERATING PROCEDURES

ASPIRE Laboratory Aspirator

Infection Control Manual Residential Care Part 3 Infection Control Standard IC9: Environmental

THE BASICS OF DECONTAMINATION

STERILIZATION IS A KEY RESPONSIBILITY OF THE INSTRUMENT

Processing Instructions. Cleaning so easy.

AS Medizintechnik GmbH Sattlerstrasse 15, Tuttlingen, Germany Tel 07461/ Fax 07461/

DEKO 190 Ward washer-disinfector

Penner Stationary Patient Lift Safe Operation & Daily Maintenance Instructions

PERFORM Operating Document

RECOMMENDATIONS FOR CLEANING DEVICES

Application Equipment

Rhima Deko 190 Washer-disinfector for human waste containers

RELIANCE VISION SINGLE-CHAMBER WASHER/DISINFECTOR

ComPact Air Drive II USER S MANUAL

Accessible. Durable. Smart.

VHP M100 Biodecontamination Systems: Closed Loop

OPERATOR MANUAL Amsco Century TM Medium Steam Sterilizer 26" x 26" (660 x 660 mm) Scientific. (07/16/07) P Rev. 0

KC22/32 SERIES Sealless Non-Metallic Centrifugal Pumps Installation and Maintenance Instructions

VHP M100 BIODECONTAMINATION SYSTEM

OWNER S MANUAL COMMERCIAL COLD PRESS JUICER

2015 Myco Industries, Inc Fax:

Integrity 270 Integrity 270VP INTEGRITY STEAM STERILIZER INTRODUCTION

A new wave in automated instrument washing. Introducing the new HYDRIM G4 series

Jennings Street School

EUTECTIC EC-10DV Series

Steam Sterilization and Sterility Assurance in Health Care Facilities ANSI/AAMI ST79. Suzanne Latta, CSPDT. Education Materials Coordinator

User Guide for Reprocessing of the SUPER VIEW Wide Angle Viewing System Sterilizable Components and Accessories

A new wave in automated instrument washing and disinfection Introducing the new HYDRIM G4 WD series

A new wave in instrument washing and disinfection Introducing the new HYDRIM c61wd G4 series

Instructions for Use: Endoplus Modular Handle Assemblies, Multi-Use Scissors Inserts, Reusable Inserts

Safe Use of a Class II Biological Safety Cabinet

OPERATOR MANUAL Amsco Eagle 3000 Series FLJ Floorloading Vacamatic Sterilizers. (10/05/01) P Rev. 7

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

Transcription:

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide

.

AMSCO STERILIZATION CONTAINER SYSTEM User s Guide TABLE OF CONTENTS 1. Unpacking...2 2. Container transport and handling...2 3. General cleaning, decontamination, and inspection...2 4. Unit assembly...3 5. Sterilization...4 6. Drying...5 7. Instructions for use in the operating room...6 8. Instructions for use in flash sterilization...7 9. Decontamination after use...7 10. Chemical/biological indicators...8 11. How to evaluate chemical indicators...8 LATEX Product and packaging are latex free -1-

1. UNPACKING For aseptic reasons it is recommended that unpacking of the container system be done in an area other than where they will be used. 1. Open corrugated carton and carefully remove the container system and/or components. 2. Remove any plastic bags and twist ties on the side clamps of the container. Also remove and read package inserts. 3. Inspect the container system for nicks, gouges, or damage that may have occurred during shipping. (If there is damage, notify your STERIS Customer Service Representative.) 4. Check contents of the bag against the packing list. 5. Transport container system and/or components to decontamination area. 2. CONTAINER TRANSPORT AND HANDLING To effectively handle the weight of the containers, you should: Understand container flow through your institution. Predetermine how the container is to be transported. Train all personnel to ensure the correct methods for handling and transport are being followed. Preparation Decontamination Sterilization Storage OR 1. Minimize or eliminate hand carrying. Utilize carts when possible. (Photos 1 & 2) CAUTION: Do not lift the container using the side clamps; use the handles located at the ends of the container. -2-2. Slide containers when possible to avoid lifting. 3. Store heaviest containers on open wire racks at a height between the mid-chest to mid-thigh area. (Photo 3) 4. Ensure that an appropriate location is available for container staging in the Operating Room. A prep stand is preferrable over a mayo or ring stand. (Photo 4) 5. Remove the container lid and/or base from the room prior to entry of the patient, to avoid having to decontaminate these components. Containers and components should be routinely cleaned following facility established policies and guidelines. 3. GENERAL CLEANING, DECONTAMINATION, AND INSPECTION CAUTION: Do not use boiler additives such as cyclohexylamine or morphine in steam in excess of FDA limits. 1. Process dissassembled lid, tray, base, and retainer rings in automated washing equipment, a washer disinfector, or another approved institutional protocol for cleaning. 2. Clean ceramic filters frequently to prevent clogging. Remove any excessive foreign material by soaking the filter in water until the foreign material is loose, then brush, and rinse. Flush the filter from the backside with running water. The filter may also be processed in an ultrasonic cleaner. 3. Use heat resistant protection on the hands or allow components to cool prior to removal from washing equipment. 4. Visually inspect each component to ensure that it is clean and that there is no damage to the component. For ceramic filters, crack lines on the top surface of the filter are considered evidence of damage and the ceramic filter should not be used. If components or filters show evidence of damage, do not use. 5. Transport components to preparation area.

4. UNIT ASSEMBLY Select the Appropriate Filter System There are three different filter systems. Select a filter system that is appropriate for the sterilization cycle to be used. (Refer to the text and the chart below for guidance.) A. Cartridge Filter System 1) The cartridge is made of plastic with the filter media permanently attached. Every cartridge filter has an internal chemical indicator attached to the handle which will indicate exposure to a specific sterilization process. The cartridge is for single use only. (Photo 5) 2) The reusable retainer ring is made of composite material. B. Disc Filter System 1) The disc filter consists of filter media without an attached plastic cartridge. The disc system does not have an attached internal chemical indicator. It is for single use only. 2) The reusable retainer ring is made of a composite material. A silicone gasket circles the retainer ring filter sealing surface. (Photo 6) C. Ceramic Filter System 1) The reusable ceramic filter system does not have an internal chemical indicator. 2) The ceramic filter is encased in a retainer ring. A silicone gasket circles the retainer ring filter sealing system. (Photo 7) GUIDANCE FOR FILTER SELECTION STERILIZATION CYCLE FILTER SYSTEM PREVAC GRAVITY FLASH* EO GAS STEAM STEAM CARTRIDGE - Yes Yes Yes N/A (Steam-Enhanced for Drying) CARTRIDGE (EO) N/A N/A N/A Yes DISC - Yes Yes Yes N/A (Steam-Enhanced for Drying) DISC (EO) N/A N/A N/A Yes CERAMIC Yes No Yes Yes *Gravity or prevac Load Filters A. Gently place new disposable filters (cartridge or disc) in all filter access ports located in the lid and base units. Place a retainer ring on top of each filter and rotate clockwise until it has been totally engaged. Ensure that all filter access ports in the lid and base units are enclosed with a filter. B. The ceramic filter is encased by a retainer ring. Place a ceramic filter in each filter access port and rotate clockwise until it has been totally engaged. Ensure that all filter access ports are enclosed with a filter. Load Tray A. It is not recommended to place any absorbent material in the tray. This type of material may hold water (condensation) and extend the drying time. B. Load inner tray, making sure instruments are arranged in such a way as to not interfere with the lid when closing. C. As appropriate, place chemicals and/or biological indicators in the tray following AAMI Practices for Steam Sterilization and Sterility Assurance. D. Place loaded instrument tray into container base. Use the chart below as the guideline for maximum weights to be loaded. Instruments, especially complex ones such as air powered drills, multihinged instruments, or multi-part instruments must be prepared and sterilized according to the recommendations of the instrument manufacturer. RECOMMENDED MAX INSTRUMENT LOAD WT CONTAINER PREVAC GRAVITY FLASH EO SIZE STEAM STEAM 22.50cmx35.56cm 9" x 14" 6 lbs. (2.7 kg) 6 lbs. (2.7 kg) 2 lbs. (0.9 kg) 6 lbs. (2.7 kg) 22.50cmx71.12cm 9" x 28" 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) NA** 16 lbs. (7.3 kg) 27.94cmx40.64cm 11" x 16" 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 27.94cmx48.26cm 11" x 19" 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 30.48cmx30.48cm 12" x 12" 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 30.48cmx58.42cm 12" x 23" 26 lbs. (11.8 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) 16 lbs. (7.3 kg) **the 9" x 28" container should not be used in flash sterilization. -3-

Install Lid A. Put lid on container. If lid does not seat properly, check to see if instruments are preventing closure. B. Engage and close container side clamps, after making sure there are no restrictions. (Photo 8) Install Information Data Card/Chemical Indicator A. Data card is slid into place in the data card holder. (Photo 12) Install Tamper Evident Security System A. For steam sterilization, place tamper evident bands on each side clamp. Make sure the band is on both the front and back parts of the side clamp. (Photo 9) NOTE: After steam sterilization, the tamper evident bands will shrink and have to be broken to disengage the side clamp. B. For ethylene oxide sterilization, place tamper evident locking tags on each side clamp. Make sure each device is placed through both parts of the side clamp and locked into place. (Photo 10) NOTE: To remove tamper evident locking tags and disengage side clamps, turn tag firmly clockwise to snap it off. Install Identification System A. Two nameplates are available. 1) Customized aluminum engraved nameplates in a variety of colors. 2) Plastic nameplates, which can be marked with a permanent marker. (Photo 11) B. Nameplate placement and removal. 1) Either type of nameplate can be easily inserted or removed. 2) Push the small center tab on the nameplate holder, and the nameplate can be easily inserted by sliding up, or removed by sliding down. -4- B. Pertinent sterilization information should be recorded on the external data card per institutional protocol. The data card includes an external chemical indicator that will change color with exposure to the respective steam or ethylene oxide sterilization processes. NOTE: Do not attach any type of tape or adhesive backed labels to the container. Such adhesives will bake on and are difficult to remove. 5. STERILIZATION Load Sterilization Containers A. Remove all toweling material or loose linen from sterilizer rack to aid in steam penetration and accelerated drying. B. Load prepared containers onto sterilizer carriage in the horizontal position. C. The 23" or longer containers may require placement lengthwise on the sterilizer carriage to avoid touching the wall of the sterilizer jacket. Check before loading. D. Testing confirms that containers may be stacked five high on a sterilizer carriage. However, it is recommended that containers be stacked no more than two high. NOTE: Same size containers only may be stacked on top of each other. Determine Sterilization Cycle Parameters A. Sterilize containers using instructions provided by the manufacturer of instruments/devices to be processed within the parameters recommended by the manufacturer of the sterilizer and container, and institutional protocol. B. The following chart recommends cycles that have been validated for use with Amsco containers. Verification that selected cycles are operating properly following AAMI standards is required. Each sterilization cycle must be verified using chemical and/or biological indicators according to your institutional policies, standards, and

RECOMMENDED PARAMETERS FOR STEAM STERILIZATION CYCLES (Unless otherwise specified by the manufacturer of the sterilizer and/or instrument) Steam Sterilization Loads Steam Sterilization Loads < 16 lbs. (7.3kg) of Instruments 16 lbs. (7.3kg) of Instruments a. Prevacuum Cycle a. Prevacuum Cycle Temperature Exposure Dry Time Temperature Exposure Dry Time Minimum Minimum Minimum Minimum 270-274 F (132-134 C) 4 minutes 20 minutes 270-274 F (132-134 C) 4 minutes 30 minutes Gravity Cycle Gravity Cycle Temperature Exposure Dry Time Temperature Exposure Dry Time Minimum Minimum Minimum Minimum 270-274 F (132-134 C) 20 minutes 30 minutes 270-274 F (132-134 C) 20 minutes 45 minutes 250-254 F (121-123 C) 45 minutes 30 minutes 250-254 F (121-123 C) 45 minutes 45 minutes procedures. This should be done for each sterilizer in your facility since the sterilization process can be affected by work practices, sterilizer performance, and the utilities supplying the sterilizer. Users of the container system also have the responsibility to ensure items placed in a container are suitable for the sterilization process to be employed. C. Following sterilization, remove containers from the sterilizer and place in cool-down area away from drafts. D. After containers have cooled, they may be placed in storage. 6. DRYING There are a number of factors that can contribute to achieving a consistently dry container and instrument set. Listed below are the factors that may influence drying with suggested recommendations to ensure dry instruments. 1. Loading Containers AMSCO containers are designed to be used without absorbent material inside the container. Absorbent materials hold condensation, which increases drying time and may cause condensation on the instruments. Instruments should be arranged so that the sterilant will contact all instrument surfaces. Overloading instrument trays may inhibit drying. 2. Loading Sterilizer Carriage A. Dedicated loads, that is, loads consisting of only AMSCO containers, are recommended. If it is necessary to sterilize a load of containers with wrapped or packaged items, place the containers on the bottom of the sterilizer carriage. The filter access ports of the containers must not be blocked by wrapped or packaged items. (Photo 13) -5- It is recommended that containers be stacked no more than two high. B. Place containers flat on the sterilizer carriage. AMSCO containers will not function properly if turned on their sides or ends. Containers may be stacked on top of each other, but it is recommended that same size containers be stacked no more than two high. C. The sterilizer must be kept free of towels and drapes, which may impede radiant heat transfer from sterilizer walls to containers, preventing thorough drying. NOTE: Same size containers only may be stacked on top of each other. Sterilization Cycle 3. Prevacuum steam sterilization is the preferred steam sterilization cycle. 4. Weight Load Increased instrument weight load may require increased drying time. 5. Instrument Dry Time An increased dry time may remove condensate from instruments. 6. Instrument Processing Instruments should be dry prior to entering the sterilizer. Wet instruments will increase the moisture content within the sterilizer. 7. Sterilizer Function Observe preventative maintenance schedule for the sterilizer as recommended by the sterilizer manufacturer. 8. Pre- and Post-Conditioning Use a pre-conditioning or a post-conditioning time. Pre-conditioning warms the instruments

prior to sterilizing helping to lessen excess moisture. Post-conditioning results in a slower cool down time, allowing radiant heat from the walls of the sterilizer to help dry the instruments. 9. Quality of Steam Steam that arrives at the sterilizer should be at least 97% vapor quality steam. 1 This is a critical factor for having consistently dry loads. Clogged sterilizer and steam line filter traps can be a source of low quality steam. 10. Environmental Considerations This includes the temperature, humidity, and air exchange rate of the cool-down area. Refer to the AAMI Steam Sterilization/Sterility Assurance in Healthcare Facilities document for environmental guidelines. 11. Use a Vacuum Steam Sterilizer With the exception of flash sterilization, all containers should be consistently dry upon completion of a cycle. 1 American National Standards Institute, Inc. Association for Advancement of Medical Instrumentation - ANSI/AAMI 46:2002 - Steam Sterilization and Sterility Assurance in Healthcare Facilities, 3.3.4.2, p. 10. 7. OPERATING ROOM HANDLING 1. Circulating Nurse A. Place sterilized container on convenient surface that helps facilitate aseptic opening. The table surface should be stable and of the proper height. B. Check that all tamper evident bands/devices are intact. If they are not, contents should be considered contaminated and should not be used. (Photo 14) D. To disengage side clamps, hold side clamps firmly and pull the bottom out, away from the container base. Disengage side clamps from lid and open to its full extended position. Make sure all side clamps are completely open and extended before lid removal. (Photo 16) E. Aseptically remove the lid from the base by placing fingers directly on the lid removal tabs. Place thumb directly on the lid itself. Lift lid up and away from the container toward your body. (Photo 17) F. If cartridge filter is used, visually confirm internal chemical indicator has turned the appropriate color. (See How to Evaluate Chemical Indicators. ) G. Check to be certain that filters have been used in all filter access ports and that the filters are securely in place. 2. Scrub Nurse A. Remove inner tray by firmly grasping the inner tray handles, pulling the inner tray up in one continuous motion until clear of container. CAUTION: Avoid touching the outer part of the container base with gloves or sterile attire. (Photo 18) C. Confirm external indicator has turned the appropriate color and sterilization information is correct. (See How to Evaluate Chemical Indicators. ) (Photo 15) -6- B. If using a cartridge filter and/or internal chemical indicator, visually confirm that the internal indicator has turned the appropriate color prior to adding instrument tray to the sterile field. (See How to Evaluate Chemical Indicators. ) Process biological indicators per institutional protocol.

8. SPECIAL INSTRUCTIONS FOR USE IN FLASH STERILIZATION 1. Circulating Nurse or Operating Room Technician A. Place new disposable filters and their appropriate retainer rings in all filter access ports or ensure the permanent ceramic filters are in place. B. Load inner tray, making sure instruments are not arranged in such a way as to interfere with the lid sliding shut. C. Place loaded instrument tray into base. D. Set the lid of the container on the base and slide the lid open approximately three inches. (Photo 19) E. Place the container in the sterilizer with the lid extending toward the operator. (Photo 20) F. For gravity displacement cycles, a minimum of 10 minutes of steam exposure is required for flash sterilization. For pre-vacuum cycles, a minimum of three minutes of steam exposure is required for flash sterilization. More complex instruments (e.g., with lumens), or porous items, require four minutes of steam exposure. G. Sterilize per AAMI recommended practices and institutional protocol using the maximum load weights in the following chart. H. After sterilization, open sterilizer door and slide the container lid closed with appropriate device to prevent burning (e.g. a heat resistant glove). I. To prevent burns, caution should be used when removing container and instruments from flash sterilization cycles. Carefully remove hot container, using a heat resistant glove, from -7- sterilizer utilizing the handles at ends of container, and place on transport cart or transfer directly to where it will be opened. CONTAINER SIZE FLASH STERILIZATION: MAX INSTRUMENT LOAD WEIGHTS 22.50cm x 35.56cm 9" x 14" 2 lbs. (0.9 kg) 22.50cm x 71.12cm 9" x 28" Not For Flash 27.94cm x 40.64cm 11" x 16" 16 lbs. (7.3 kg) 27.94cm x 48.26cm 11" x 19" 16 lbs. (7.3 kg) 30.48cm x 30.48cm 12" x 12" 16 lbs. (7.3 kg) 30.48cm x 58.42cm 12" x 23" 16 lbs. (7.3 kg) 9. DECONTAMINATION AFTER USE 1. Workers should wear appropriate personal protective equipment (e.g. gowns, masks, eye protection, gloves, head coverings and shoe covers) according to institutional protocol. 2. Open the side clamps and remove lid from container. 3. Remove all filter retainer rings. 4. Rotate retainer ring counterclockwise and lift. 5. Remove and discard all disposable filters. 6. Remove and inspect ceramic filters. 7. Refer to Section on General Cleaning, Disinfection and Inspection for processing instructions. 8. Keep all retainer rings. 9. Remove inner tray with the instruments and process according to institutional protocol. 10. Remove and discard all pieces of tamper evident bands/locks from side clamps. Do not remove clear plastic tab cover on back part of side clamp. 11. Remove external data card. 12. Unit is now ready for decontamination. Refer to General Cleaning, Decontamination and Inspection Section. 13. Following decontamination, periodic inspection of the unit is suggested to ensure that all parts are in good working order.

10. CHEMICAL/BIOLOGICAL INDICATORS 1. Biological Indicators Biological indicators should be used in accordance with AAMI Guidelines and institutional protocols. Only biological indicators can determine if sterilization has taken place. 2. Chemical Indicators Chemical indicators DO NOT verify sterility. They provide the user with a visual indication that the parameters for the sterilization process exist. 11. HOW TO EVALUATE CHEMICAL INDICATORS (For External Data Cards And Internal Filter Cartridge Indicators) 1. Steam Sterilization Indicators Steam indicators are a cream color in their unexposed state, turning to a shade of dark brown after processing. Ensure that the indicator has changed to the appropriate color as directed on the legend of the indicator. The indicator must be completely turned across the bar and must be a shade of dark brown. NOTE: In a failure mode, ink turns no darker than tan. 2. Ethylene Oxide (EO) Sterilization Indicators EO Indicators are a shade of cream in their unexposed state, turning to a shade of orange after processing. The indicator must be completely turned across the bar and must be a shade of orange. NOTE: In a failure mode, ink turns to light yellow. -8-

.

Publication ID #M2506EN.2004-03, Rev. C GPSI Printed 4/2006, 5000 2003-2006 by STERIS Corporation. All rights reserved. Printed in the USA. STERIS OFFICES WORLDWIDE Canada 800 661 3937 France 33 155 176 000 Germany 49 2203 890 6969 Italy 39 02 2130341 Japan 81 3 5777 1580 Korea 82 2 517 1517 Latin America 305 442 8202 Singapore 65 68 41 7677 Spain 34 91 658 5920 Switzerland 41 32 376 0200 United Kingdom 44 1256 840400 STERIS Corporation 5960 Heisley Road Mentor, OH 44060-1834 USA 440-354-2600 800-548-4873 www.steris.com Technologies to Prevent Infection and Contamination