Alaris Syringe Module v9

Similar documents
Alaris Syringe Module v9

Alaris Syringe module FAQs

Syringe Pump. User Manual

Directions for Use. Alaris System. Alaris Products. Supports Guardrails Suite MX (v8) August RATE (ml/h) RATE (ml/h) CHANNEL SELECT PAUSE

Administration of AMBISOME Using the GemStar Pump

M E D I C A L D E V I C E S

Directions for Use Alaris System (with PC Unit, Model 8000)

Mini Rythmic PN+ Set-up Guide

Directions for Use. Alaris System (with Alaris PC Unit, Model 8015) September 2009

Enteral Pump Instructions for Use To Assemble When the low

VARIABLE RATE INFUSION PUMP

SINGLE SPEED SYRINGE PUMP

ALLERGEN MIXING ASSISTANT

Sapphire TM. Bring Home The Future Of Infusion Care. Ambulatory Infusion System. > Intuitive, easy-to-use touch screen design


Section. 3: CADD Pump. Updated: June 20, Section Author(s): sstafford & jhval

DS9400 Series. Release Notes for Firmware V2.07. Fire Alarm Control Panel

CONTENTS. III. OPERATING INSTRUCTIONS... pg 7 LOADING DISPOSABLE SET (Figure 4)... pg 9. IV. INDICATORS AND ALARMS... pg 10


Clinician Information

bibag System User s Guide

Description of the CADD-Prizm VIP Pump

Refrigeration Controller Operator s Manual (HRC) PO Box 6183 Kennewick, WA

User s Manual. TIGER S EYE E-Series Mark V Jockey. TIGERFLOW Systems, Inc Mint Way Dallas, Texas

ZX1e ZX2e ZX5e. Document No Issue 01 user manual

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

Pioneer-R16 Gas Monitor Operator s Manual

H-10PM. Refrigerant Leak Detector

Rocket Quick Guide. Rocket PSU Portable Suction Unit

SN-50 Serial Syringe Pump User Manual. Version: 2.6

Operator s Manual DigiPump LP8 Infusion Pump - Digicare Biomedical Technology Inc CONTENTS. A. About this manual...2

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

Standard Operating Manual

Ambulatory infusion pump for continuous, intermittent, TPN or PCA delivery.

Syringe Pump. Syringe Pump. Product Model: ZD-50C6 Syringe Pump. ZD-50C6-H Syringe Pump

PR-L2466W- PA. Operating Instructions. High Performance Refrigerator PR-L2466W-PA

! WARNING To avoid risk of electrical shock, personal injury or death; disconnect power to range before servicing, unless testing requires power.

DS User Manual

MANUAL. Type ETO2 Controller for ice and snow melting. Warmup Inc. The world s best-selling electric floor heating brand

i.c³ User Guide For Helmer i.series Ultra-Low Freezers A/A

Thopaz+ Quick Reference Guide. Precious life Progressive care

PF1000 CONTROLLER. for the. Heateflex Corporation 405 E. Santa Clara St. Arcadia, CA TEL: (626) ; FAX: (626)

SCAN200E USER S MANUAL

Installation & Calibration Guide

O PE R AT I ON HO851G REVOLVING TRAY OVEN. MODEL HO851G GAS ML Pan ML Pan ML Pan ML Pan

GLD-30 Gas Leak Detector

v.a.c.simplicity TM Quick Reference Guide Power Connection Canister Battery Charging Light Control Panel

ACME DIAGNOSTICS STANDARD OPERATING PROCEDURE

Ion OneTouch Template Kit Publication Part Number Revision E Date 31 October 2011

IMPORTANT. PLEASE NOTE: The infrared beam path MUST be kept clear of obstructions at all times!

Original Perfusor & Infusomat

Tempco Part Number PCT30006 Temperature Control Enclosure with Relay Output for Tote Tank Heating Applications

GREASE INTERCEPTORS. Z1192 GREASE RECOVERY APPLIANCE (GRA) INSTALLATION and OPERATION INSTRUCTIONS

ToxiRAE II. PGM-1100 Series Personal Toxic Gas Monitor. User Manual

Passport 12m and Passport 17m BEDSIDE MONITOR GUIDELINE 5.21 Software Version NAME: HOSPITAL: DATE: VALIDATOR: Performed

MacDonald Steel Limited

OPERATOR S MANUAL. The Belmont Rapid Infuser, FMS Boston Road Billerica, MA 01821, USA US/Canada Worldwide

APC BC300 Series 40kW 208/450/480V User Guide

MPT Electric Fire Pump Controller

Beacon 200 Gas Monitor Operator s Manual. Part Number: RK Released: 6/6/08

Jetstream Series HPLC - PELTIER COLUMN-THERMOSTATS 1. GENERALITIES

Technical Update TAA.TU rev. 1

KD-CLN-LP200 Ultrasonic Vinyl Record Cleaner User s Manual English v1.2 Printed in Korea

(Place Logo from Lab here)

Mitocell: Instructions for Use

ZSC100 Gas Detection and Alarm System Controller

ModSync Sequencing System Installation & Operation Manual. For use with Fulton Steam Boilers.

GASGUARDIAN Channel Controller OPERATING & INSTALLATION MANUAL

one Operating instructions SenTRI ONE panel-based Fire detection and alarm system

GG-2 2-CHANNEL GAS DETECTION CONTROL PANEL. Installation and Operation Manual

Infusion System. Infusion and Syringe Pumps

Troubleshooting your BAXJECT II Needle-less Transfer device

Mark 25 Ultrapure Water Conductivity Analyzer

Series DCT1000DC Dust Collector Timer Controller Specifications Installation and Operating Instructions

TMC. Installation and Operation Manual TMC. Temperature and Pressure Monitoring for Heating and Cooling Applications. Temperature Monitoring Control

Laptop / PC Programming Manual

PRESSURA ROOM PRESSURE CONTROLLER MODELS 8630-SC 8630-PC

RPM1600 Series Room Pressure Monitors

User s Manual. THE iq Drive SYSTEM USER S INFORMATION IMPORTANT. Variable Speed Ultra-High SEER Split System IMPORTANT NOTE

Operation Manual Fighter ProVision Software. Version: 0.0 Revision: 1

User s Manual. CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP

Macurco DVP-120 Detection and Ventilation Control Panel

TECH SHEET DO NOT DISCARD PAGE 1. If unsuccessful entry into diagnostic mode, actions can be taken for specific indications:

Tempco PCT-3000 Series Temperature Control Console with Relay Output for Heating or Cooling Applications

LXG-300 Series. Models with LXG-300 Series Instruments

Sensor Control Panel

Sentry LIQUID LEVEL GAUGE MODEL 200 or 200C OWNER MANUAL REV 1.7 SEPT08 PAGE 1 OF 12

USER MANUAL Digital CO Monitor Model CO30

Tri-Tech Medical Inc.

Analog Room Pressure Monitor RPC Series

PONY 2/4 User Guide 2017

BeneVision TM80 Telemetry Monitor

SINGLE PATIENT DIALYSIS MACHINE

Motorised Infrared Optical Beam Smoke Detector. User Guide

For technical support call:

Undercounter High Temperature Sanitizing Dishwasher (USA Version)

IT801 Thermostat. User s Manual. The complete guide to the set up and operation of your new smart Wi-Fi thermostat.

Installation, operation & service manual. AlarmScout 961/962 line powered

Installation, operation & service manual. AlarmScout 961/962 loop powered

ExStik DO600 USER GUIDE. Dissolved Oxygen Meter. 99 Washington Street Melrose, MA Phone Toll Free

Transcription:

Cool Gray 7 U Text - PMS 294U QUICK REFERENCE GUIDE WARNINGS: To ensure proper operation of the Alaris System, you must be familiar with related features, disposables, administration sets, set-up, and programming. This guide is not intended to be comprehensive instructions for the set-up and operation of the Alaris System. For complete instructions along with Warnings and Cautions, refer to Alaris System User Manual (v9). Programming an Infusion using Guardrails Suite MX Protection The programming user interface is identical for the Alaris Pump Module and the Alaris Syringe Module. Set Priming, Loading and Removal Manual Priming 1. Ensure set is not connected to patient and that device is level with patient's heart. 2. If used, hold pressure sensing disc between two fingers and invert disc so patient side is up. 3. Slowly prime set while gently massaging disc to ensure disc does not become over or under filled and all air is expelled. 4. When priming is complete, clamp set. 5. Load syringe into device and install pressure sensing disc (for proper placement information, refer to the Alaris System User Manual). 6. Program infusion. 7. Release clamp and press START. 8. Attach set to patient. Note: Starting infusion before attaching set to patient minimizes any potential bolus that could be released from pressure buildup in set due to normal priming and syringe loading. Device Priming 1. Ensure set is not connected to patient and that device is level with patient's heart. 2. Load and select syringe type and size. 3. Program infusion parameters but do not start infusion. 4. Press OPTIONS. 5. Select Prime Set with Syringe. 6. If used, hold pressure sensing disc between two fingers and invert disc so patient side is up. 7. Press and hold PRIME while gently massaging disc to ensure disc does not become over or under filled and all air is expelled. 8. Release PRIME when fluid flow is observed at end of set. 9. Install pressure sensing disc. San Diego, CA, 800.854.7128 2003-2016 CareFusion Corporation or one of its affiliates. All rights reserved.

10. Press EXIT. 11. Press START. 12. Attach set to patient. Note: Using the Prime Set with Syringe feature engages the mechanical components of the pump and decreases the syringe's friction and compliance. If manual priming, when complete, use Prime Set with Syringe feature to minimize startup delays and delivery inaccuracies, especially at low infusion rates. Loading: Infusions 1. Open syringe barrel clamp until it clears syringe chamber. 2. Twist gripper control clockwise and raise drive head to fully extended position. 3. Insert syringe barrel flange between barrel flange grippers. 4. Lock syringe in place by closing barrel clamp. 5. Twist gripper control clockwise, lower drive head, lock plunger in place with plunger grippers. 6. Insert pressure disc (if used) with membrane facing device. Apply upward pressure until disc snaps into disc housing. Note: At the start of an infusion program, the system prompts to select and confirm syringe type and size. Ensure syringe manufacturer and size correctly identify the installed approved syringe. Mismatches may cause under-infusion or over-infusion to the patient. Syringe Barrel Clamp Drive Head Gripper Control/ Drive Head Release (closed position) Barrel Flange Grippers Syringe Barrel Clamp Plunger Grippers (closed position) 2 oz 60 WARNING: To prevent free flow, close set tubing clamp before loading or unloading syringe. 55 50 45 40 35 30 25 20 15 10 5 1½ 1 oz ½

Removal: 1. Close tubing clamp. 2. Open plunger grippers and syringe barrel clamp (silence safety alarm). 3. Remove syringe, apply downward pressure to remove disc from housing. Near End of Infusion When infusion is near completion, Near End of Infusion is enabled: Main display alternates between NEAR END and remaining VTBI. Yellow standby status indicator flashes. An audio prompt sounds which requires being silenced once, and will not reoccur following initial silencing. All Mode Enabled If the ALL option is enabled for the VTBI, entire contents of the syringe will be delivered (unless your facility has established a KVO volume). Note: KVO is not available for intermittent infusions. All Mode Disabled If the ALL option is disabled for the VTBI, a syringe content volume will be displayed after the VTBI soft key is selected. Selecting Pressure Limits Pressure Sensing Disc Installed 1. Press CHANNEL SELECT key. 2. Press OPTIONS key. 3. Press Pressure Limit soft key. 4. To enter a new pressure limit value, press Change Value soft key and use numeric keys. If Auto Pressure feature is enabled and desired, press Auto Pressure soft key. Note: If Auto Pressure is selected and current pressure is: 100 mmhg or less system adds 30 mmhg to current pressure, to create a new alarm limit. Greater than 100 mmhg system adds 30% to current pressure, to create a new alarm limit. 5. Press CONFIRM soft key. Pressure Sensing Disc Not Installed 1. Press CHANNEL SELECT key. 2. Press OPTIONS key. 3. Press Pressure Limit soft key. 4. Press appropriate soft key: Low, Med, High. 5. Press CONFIRM soft key.

Alarms and Troubleshooting: Alarms Alarm Meaning Response Channel Disconnected Check Syringe Occlusion Channel(s) disconnected while in operation or there is a communication problem. Plunger grippers opened during infusion and then closed. Infusion stops on affected module. Syringe barrel clamp opened during infusion and then closed. Infusion stops on affected module. Syringe plunger not captured while in idle state. System alarms after 30 seconds to indicate potential siphoning condition. Increased back pressure sensed while infusing. Infusion stops on affected module. To silence alarm and clear message from screen, press CONFIRM soft key. Reattach channel if desired, ensuring it is securely clicked into place at Channel Release Latch. If alarm is still present, replace module with an operational instrument. Securely lock plunger grippers, press CHANNEL SELECT key, and reselect syringe. Securely lock syringe barrel clamp and press RESTART key. Check for potential siphoning. Ensure administration set clamp is in closed position. Securely lock plunger grippers over syringe plunger. Clear occlusion. Press RESTART key, or press CHANNEL SELECT key and then START soft key. Pressure Disc Installed Pressure sensing disc installed during an infusion. Infusion stops on affected module. Press CONFIRM soft key and restart infusion.

Alarms and Troubleshooting: Alarms (Continued) Alarm Meaning Response Pressure Disc Removed Pressure sensing disc removed. Infusion stops on affected module. Reinsert pressure sensing disc and press RESTART key. Syringe Empty Syringe is empty. If syringe is not empty, other possibilities are: Pressure sensing disc inappropriate/defective. Syringe plunger travel impeded. Pressure transducer defective. Set up new infusion or press CHANNEL OFF key. Verify appropriate pressure sensing disc is in use and functioning properly. Verify syringe plunger movement is not impeded. If syringe is not empty and above actions do not correct alarm, contact qualified service personnel.

Errors Alarm Meaning Response Channel Error Error detected. Operation stops on affected module. To silence alarm and continue operation of unaffected modules, press CONFIRM soft key. Replace module with an operational instrument, as required. Service by qualified personnel is required. Syringe Calibration Required Syringe Driver Head Error Error on infusing module indicating calibration is required. Infusion stops on affected module. CALIBRATE scrolls in Message Display. Noninfusing module, with plunger grippers open, senses excessive pressure being applied downward on Drive Head. OCCLUSION scrolls in Message Display. To silence alarm and continue operation of unaffected modules, press CONFIRM soft key. Replace instrument with an operational instrument, as required. Service by qualified personnel is required. To silence alarm and continue normal operation, press CONFIRM soft key. P00000205