Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Similar documents
detectors UltiMate 3000 Variable Wavelength Detectors

Thermo Scientific Rocket Evaporator. Revolutionary solvent evaporation complete laboratory workflow

High Volume Injection up to 5 ml with the Agilent 1200 Infinity Series Online SPE Solution for Highest Sensitivity

Enhancing Throughput of Online SPE Analysis of Pesticides in Wastewater The Agilent 1200 Infi nity Series Online SPE Solution

Thermo Scientific Dionex SolEx SPE Cartridges

Tool-free LC connections for HPLC, UHPLC and nano LC systems

Thermo Scientific Dionex UltiMate 3000 Fluorescence Detectors

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples

Application Note. The Agilent 1200 Infinity Series Online SPE Solution for the Highest Sensitivity. Author. Abstract.

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Optimizing the Performance of the Agilent 1290 Infinity Evaporative Light Scattering Detector

GB National food safety standard Determination of benzoic and sorbic acid in milk and milk products

Thermo Scientific Dionex AutoTrace 280 SPE Instrument. Automate Large-Volume. Solid-Phase Extractions

ISO INTERNATIONAL STANDARD

An Advanced Heat Exchanger for the Agilent 1290 Infinity Thermostatted Column Compartment

Thermo Scientific Dionex Integrion HPIC System

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

Rapid Quantification of Polar and Semipolar Pesticide Metabolites with Combined Online SPE and Direct Injection

Application Note. Authors. Abstract

ELISA NIMBUS. Automated ELISA Processing System

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

Automated Fractionation System that Increases Recovery and Optimize Precision of Solid Phase Extraction Methodology

On Line SPE LC/MS/MS Applications

AppNote 6/2008. Automated SPE and Fast GC-ECD Analysis of PCBs in Waste Oil KEYWORDS ABSTRACT INTRODUCTION. MACH, Fast GC-ECD, SPE, PCB, waste oil

SEC-MALS Noise Assessment and Cleaning Guide

An Automated SPE Procedure and Analysis for the Extraction of Chlorpromazine and Thioridazine from Plasma Application Note 213

This document is a preview generated by EVS

AppNote 3/2009. Automation of Solid Phase Extraction Methods using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

Customised HPLC Systems for Highest Performance

How Important is Performance to your SPE Method Development?

Direct Injection and Online SPE LC/MS/MS for the Determination of Pharmaceuticals and Personal Care Products (PPCPs) in Water Samples

Rain Forest Quick Guide

Nexera-i Specifications Prominence-i Specifications

get mass. without massive work. [ 3100 MASS DETECTOR ]

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

[ ACQUITY QDa DETECTOR ] SEPARATING BEYOND QUESTION BRINGING THE POWER OF MASS DETECTION TO YOU

PRODUCT INFORMATION HISIC450 OVERHEIGHT DETECTORS RELIABLE OVERHEIGHT DETECTION OF VEHICLES. Traffic sensors

Agilent InfinityLab Online SPE Solution

Agilent 1290 Infinity II Online SPE Solution

Improved HPLC Separations at Elevated Temperatures Using a New Stand-alone Preheater

Auto SPE-06Plus Solid Phase Extraction System. Reeko Instrument USA

Agilent 1290 Infinity II Vialsampler Data Sheet

Spark Holland introduces

ice 3000 Series Atomic Absorption Spectrometers

BENEFITS. QUATERNARY HPLC SYSTEMS Quaternary and Other Gradients. The Complete Range of HPLC. Convenience of Maintenance. Systems Configuration

Product Guide for Ludger Clean Post-endoglycosidase and Post-exoglycosidase Clean-up Plate

HydroFlex 3-in-1 microplate strip washing. Flexible and reliable microplate strip washing and vacuum filtration in a compact, modular platform

A Guaranteed Direct Replacement to Hypersil

The Complete and Easy Guide to Configuring Your Specific Thermo NanoLC for Mass Spec Analysis

HydroFlex FLEXIBLE AND RELIABLE MICROPLATE STRIP WASHING AND VACUUM FILTRATION IN A COMPACT, MODULAR PLATFORM. 21 CFR Part 11* 98/79/EC IVD-D*

FIXED GAS ANALYZER (FGA)

To request a quote or locate your nearest dealer, contact Customer Care toll-free in the US at or

VICOTEC450 EXTRACTIVE VISIBILITY MEASURING DEVICES EXTRACTIVE VISIBILITY MEASUREMENT FOR INCREASED SAFETY IN TUNNELS

HydroSpeed plate washer for 96- and 384-well formats

Accuracy counts! Sensors with analog output

PURITY GRADES OF LAB REAGENTS

The KingFisher Family

KNX The worldwide STANDARD for home and building control. KNX Association International

CALIDUS tm the Modular micro Gas Chromatograph

Agilent 1200 Series Variable Wavelength Detector

300 series HPLC autosamplers

HydroSpeed plate washer for 96- and 384-well formats

Agricultural. drugs of abuse. Food & Beverage. Pesticides SOLID PHASE EXTRACTION. Inject. Wash Elute. Cartridges. Concentrate. Silica.

English Version. This Technical Specification (CEN/TS) was approved by CEN on 25 October 2009 for provisional application.

ice 3000 Series Atomic Absorption Spectrometers Flame and furnace AAS analysis for routine laboratories

ice 3000 Series Atomic Absorption Spectrometers Flame and furnace analysis for routine laboratories

Fundamentals of Solid-Phase Extraction (SPE): Principles and Practical Tips for Developing Sample Preparation Methods

General Animal Injections

Precision Temperature Control. Pulse Free Pump. Column Oven. Superb Performance Ion Chromatography. Leak Detector. Modular Flexibility

An Automated Method for the Selective Solid Phase Extraction of Ochratoxin A from Wheat Using Molecularly Imprinted Polymers

Installing the Agilent CE ESI-MS Sprayer Kit (G1607A)

PRODUCT INFORMATION. GM901 Carbon Monoxide Gas Analyzers CO MEASUREMENT FOR EMISSION MONITORING AND PROCESS CONTROL. In-situ Gas Analyzers

Customers should provide all necessary operating supplies upon request of the engineer.

Purge and Trap Troubleshooting Made Easy?

GMS800 FIDOR TOTAL HYDROCARBON ANALYZER CONTINUOUS MONITORING OF HYDROCARBON EMISSIONS

Laser Analyzers for NH3 Safety Monitoring and Leak Detection

HPLC Column Ovens and Temperature Controllers

KNX The worldwide STANDARD for home and building control. KNX Association International

ProStar 325 UV-Vis Detector

Automação de SPE utilizando sistema robótico Triplus RSH autosampler. Nova Analitica

Common Features of DX-600 Modules

SRI 8610C GC for PCB in Transformer Oil Analysis November 2014

Agilent 81663A DFB Laser 8165xA Fabry-Perot Lasers

Return loss measurement of fiber optic components

APPLICATION NOTEBOOK FOR AFFINIMIP SPE

NEW! TSX Series high-performance undercounter refrigerators

OZ Optics Limited. Fiber Optic Attenuators. June 2018

Who is GBC Scientific Equipment?

Installation Instructions

FREESTYLE SPE UNIQUE POSSIBILITIES FOR SOLID PHASE EXTRACTION

NEW! TSX Series high-performance undercounter refrigerators

Agilent 1200 Series Diode Array and Multiple Wavelength Detectors

Recent Improvements in Process GC Instrumentation

Gagnez avec «David» Mighty MEPS contre les cartouches «Goliath» SPE

Gilson 151/152 UV/VIS Detectors User s Guide

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

Thermo Scientific Heratherm Heating and Drying Ovens. efficient safe easy. Part of Thermo Fisher Scientific

PRODUCT INFORMATION. SureSense THE SURE WAY TO DETECT ANY OBJECT. Photoelectric Sensors

Transcription:

Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE Richard F. Jack, 1 Jeff Rohrer, 1 Xu Qun, 2 and Lina Liang 2 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Shanghai, People s Republic of China

Overview Purpose: Waterblooms of cyanobacteria (blue-green algae) can produce potent toxins that have become a severe problem for eutrophic aquatic environments. Hepatotoxins are among the primary toxins produced by these species, which grow in lakes, ponds, and rivers used as drinking water sources. Microcystin contamination of drinking water at low nanomolar concentrations is considered a risk factor for cancer, and microcystin-lr has been associated with most of the incidents of toxicity involving microcystins. Therefore, the World Health Organization (WHO) has proposed a provisional guideline concentration of 1.0 µg/l for microcystin-lr in drinking water. 1 Methods: Described here is a simple, fast, and effective target-cut on-line SPE method followed by HPLC with UV detection. This on-line SPE method differs from typical protocols in that the bound analyte on the SPE column is selectively eluted using a mobile phase gradient, just as the first dimension of a two-dimensional chromatography system is performed. This reduces the number of interferences for sample analysis. Here, the target-cut online SPE method followed by HPLC with UV detection was applied to the determination of three microcystins (-LR, -RR, and -YR) in drinking, tap, and lake water. The three target analytes were coeluted from the first column using chromatographic conditions that eliminated as many interferences as possible. The analytes were then sent to the analytical flow path and separated on the second column using the same type of stationary phase under different chromatographic conditions. Results: This design takes advantage of the separation power of both columns and may eliminate interferences more efficiently than typical on- and off-line SPE methods. Sub-µg/L concentrations of microcystins-lr, -RR, and -YR spiked in water samples were determined, which exceeds the WHO requirement. Introduction The analytical approaches commonly used for microcystins include bioassay, chemical, and biochemical methods. Bioassays have been used in screening, but were found to be nonspecific and/or more time consuming. Biochemical methods, such as enzyme-linked immunosorbent assay (ELISA) and protein phosphatase inhibition assay (PPIA), are advantageous as screening methods due to their high sensitivity and ability to quickly treat a large number of samples, although they provide poor identification and have the potential for false positives. Reversed-phase high-performance liquid chromatography (HPLC) with UV detection, liquid chromatography tandem mass spectrometry (LC-MS), and capillary electrophoresis are chemical methods that have been used for the identification and quantification of microcystins. 2 Methods 2 Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Methods Liquid Chromatography Dionex UltiMate 3000 HPLC system including: DGP-3600A Dual Gradient Analytical Pump SRD 3600 Integrated Solvent and Degasser Rack WPS-3000TSL Thermostatted Semipreparative Autosampler (with 2500 µl sample loop) TCC-3200 Thermostatted Column Compartment equipped with two 2p 6p valves VWD-3400RS Four Channel Variable Wavelength Detector (Without Flow Cell) Thermo Scientific Chromeleon Chromatography Data System software Thermo Scientific Orion 420A+ ph Meter Standards 100 µg each of microcystins-lr (CAS 101043-37-2), -RR (CAS 111755-37-4), and -YR (CAS 101064-48-6), 95% (HPLC), Alexis Corporation. Prepare stock standard solutions with 50 µg/ml concentrations by dissolving the standards with 2000 µl of methanol. Prepare the standard solutions used for the calibration curve by making appropriate dilutions of the stock standard solutions with water. Samples Tap water samples were collected at the Thermo Scientific Shanghai Applications Lab. The lake water sample was collected at Zhangjiang High Science and Technology Park located in the Pudong District of Shanghai. Bottled spring water samples were purchased from a supermarket in Shanghai. These samples were filtered through a 0.45 µm membrane (Millex-HN) prior to injection. FIGURE 1. Structures of microcystins Chromatographic Conditions Thermo Scientific Poster Note PN70078_e 05/12S 3

Chromatographic Conditions On-Line SPE Column: Thermo Scientific Acclaim RSLC PA2, 3 µm Analytical (3.0 33 mm, P/N 066276) Mobile Phase: A: 22.5 mm KH 2 PO 4 2.5 mm K 2 HPO 4 buffer (dissolve ~ 3.1 g of KH 2 PO 4 and 0.44 g of K 2 HPO 4 in 1 L of water) B: CH 3 CN in gradient 3 Separation Column: Acclaim PA2, 3 µm Analytical (3.0 150 mm, P/N 063705) Mobile Phase: A: 0.05% (v/v) H 3 PO 4 (dilute 0.6 ml of 85% H 3 PO 4 to 1 L with water) B: CH 3 CN in gradient 3 Valve Switching 3 Flow Rate: 0.7 ml/min for both SPE and separation Inj. Volume: 2500 µl on the SPE column Temperature: 40 C UV Detection: Absorbance at 240 nm FIGURE 2. Flow schematics for A) traditional and B) target-cut on-line SPE methods equipped with one 2p 6p valve for sample preparation and analysis 4 Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Comparison of Traditional and Target-Cut On-Line SPE Methods The commonly used on-line SPE flow scheme (Figure 2A) couples the SPE column directly with the analytical HPLC column using one six-port (2p 6p) column valve. The filtered sample is injected directly onto the system and delivered to the SPE column. The SPE column is then switched into the analytical flow path to elute the bound analytes (6-1 position). The analytes are then separated on the analytical column and detected by the UV detector. For the target-cut on-line SPE method, 3 a small change in the flow scheme of the traditional on-line SPE mode reverses the flush direction on the SPE column (Figure 2B). The bound analyte on the SPE column is selectively eluted using a mobile phase gradient, Just before the front portion of the analyte peak elutes from the SPE column, the SPE column is switched into the analytical flow path. Results Microcystins Separation Using Target-Cut On-Line SPE Figure 3 shows chromatograms of three types of water samples spiked with 1.0 µg/l each of microcystin-rr, -YR, and -LR standard using the traditional and target-cut on-line SPE methods, respectively. FIGURE 3. Chromatograms of a) bottled spring water, b) tap water, and c) lake water, all spiked with 1 µg/l each of microcystin-rr, -YR, and -LR standard using A) traditional and B) target-cut on-line SPE methods Thermo Scientific Poster Note PN70078_e 05/12S 5

FIGURE 4. Chromatograms of a tap water sample spiked with 0.5 µg/l each of microcystin-rr, -YR, and -LR standard using different target-cut modes: A) Valve-switching starts from microcystin-rr and ends at microcystin-lr, when they are eluted from the SPE column. B) The three microcystins elute together from the SPE column. As shown in Figure 4A, with the target-cut method, a large amount of interferences were still cut to the analytical flow path, which resulted in interference with the determination of microcystins at sub-µg/l concentrations. Figure 4B shows the target-cut method with a CH 3 CN phosphate buffer (ph 6.0) mobile phase to elute analytes in one peak from the SPE column, and the analytical column using CH 3 CN 0.05% H 3 PO 4 (v/v, ph 2.2) mobile phase. Conclusion 6 Sensitive Determination of Microcystins from Environmental Waters Using On-Line SPE

Conclusion This work describes a target-cut on-line SPE method that can fully recover low concentrations (< 1 µg/l) of three microcystins (-RR, -YR, and -LR) when added to three different water samples. These concentrations are less than the maximum concentrations recommended by WHO. This method is fully automated and easily configured on an UltiMate 3000 2 Dual HPLC system. References 1. Cyanobacterial Toxins: Microcystin-LR in Drinking-Water; Background Document for Development of WHO Guidelines for Drinking- Water Quality: World Health Organization, 2003. 2. Lee, H.S.; Jeong, C.K.; Lee, H.M.;Choi, S.J.; Do, K.S.; Kim, K.; Kim, Y.H. On-Line Trace Enrichment for the Simultaneous Determination of Microcystins in Aqueous Samples Using High-Performance Liquid Chromatography with Diode- Array Detection. J. Chromatogr. A 1999, 848, 179. 3. Dionex (now part of Thermo Scientific) Application Note 261: Sensitive Determination of Microcystins in Drinking and Environmental Waters, 2010 [Online] www.dionex.com/en-us/ webdocs/88494-an261-hplc-microcystins- Water-07Oct2010-LPN2607.pdf (accessed Apr. 19, 2012). www.thermofisher.com/dionex 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 3420 8600 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655 PN70078_E 07/16S