PARTNERING IN ESTHETIC DIGITAL DENTISTRY SM. Odyssey 3Watt. Owners Manual

Similar documents
Odyssey 2.4G. Owners Manual

OPERATORS MANUAL DIODE LASER

Laser Safety, Utilization and Credentialing Policy

This slide presentation has been prepared for Temple University Health System and is intended only for use by it s faculty, staff and students.

LASER SAFETY. Class 2 and 2a - low power, low risk Class 2 and 2a lasers have low power and emit visible light. They will cause harm if viewed

Lasers in Veterinary Practice. Safe Use Guidelines

page 1 BWF Dental Laser User Manual

Laser Safety Management

Title: Laser Safety Effective Date: 9/91 Revision: 2/97 Number of Pages: 7

Important Safety Notice

Eickemeyer Vetrix Laser

INSTRUCTIONS FOR USE SOL LASER

LASER SAFETY PROGRAM

Content. Laser physics Laser treatment parameters Laser safety. by Fotona d.o.o., 2017

West Virginia University Laser Safety Program

OPERATOR MANUAL. MDL 15 Medical Laser Unit. OPERATOR MANUAL MDL-15 / March 2005

Laser & Ultraviolet Light Sources Safety

Rutgers Environmental Health and Safety (REHS)

Laser Safety. This Laser Safety Program applies to the acquisition, use, transfer, and disposal of all Class 3B and 4 lasers at UNC Asheville.

cannot be overexposed to laser radiation if there is an effective laser safety program in place!!!

Stony Brook University Hospital Environmental Health & Safety Policy & Procedure Manual. EH&S 6-8 Original : 1/00 Revision Date: 3/16/15 Pages 6

Laser Safety Part III Administration

Meticulous Structured CREATIVE INCREDIBLY Innovative TECHNOLOGY

Policy Redesign Project. Laser safety. Table of Contents

Rice University Laser Safety Manual

*For inquiries and questions about lasers, see the Yellow-Page.

*smith&nephew VERSAJET HYDROSURGERY SYSTEM

Steril Air mobil. Owner s Manual. GENTRA Rıhtım cd. Yuva sk. No: ¾ Beyoğlu İstanbul Turkey Tel : Fax:

Safety Manual Section 15 Radiation Safety

SECTION 1: STARTING 1.1 UNPACKING AND INSPECTION

Laser Safety Citations

Medical Lasers and Laser Safety

G4 Pulsed Fiber Laser

Laser Safety Program Manual

North Dakota State University Laser Safety

Administration of AMBISOME Using the GemStar Pump

Laser Use Authorization

OPERATING MANUAL. 1. Model: Pinnacle Pro: 980nm, 30W

Model CWA200 Owner s Manual

American National Standard

MODEL 7000 SUCTION UNIT

Laser Safety Part III Administration

UL U TR LTRASONIC S ONIC SCALE ALER PIEZO MINI

SYNRAD Laser Safety 101

PURPOSE OF THE PROCEDURE

Approved by Principal Investigator: Date: Approved by Laser Safety Officer: Date:

Laser Safety Program

Laser Use at Subaru Telescope (NAOJ) Laser Safety Policy

Laser Use Registration & Application (L.U.R.)

BIOVAC 500 SMOKE EVACUATOR

POLICY. Number: Title: Laser Safety. Authorization

Laser Standard Operating Procedure

LASER SAFETY TRAINING CATALOG LASER LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY.

Operating Microscope Adapter. Operator Manual

DUQUESNE UNIVERSITY. Laser Safety Program

MicrOTDR. Optical Fault Locator. User Manual

HEALTH AND SAFETY MANUAL

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

Table of Contents. English

This document is a preview generated by EVS

A. Administrative. B. Labeling and Posting. C. Control Measures. Laser Safety Self-Audit Checklist. Building Room Principal Investigator Date

G4 Pulsed Fiber Laser

LASER SAFETY BARRIER

ML562G84 APPLICATION. Display system. Symbol Parameter Conditions Ratings Unit. Pulse (Duty 40%)

R100 Oil-Less Refrigerant Recovery Unit

Design Manual Installation Operation Maintenance

9 W, 15 W and 25 W Industrial Fiber Laser Systems IFL Series

THE UNIVERSITY OF TEXAS RIO GRANDE VALLEY LASER SAFETY MANUAL. August 2016

GENERAL PRECAUTIONS. Do NOT dismantle.

NEED FOR LASER SAFETY

Order code: EQLA21 user manual

X99 Series. Dental Handpiece. Instruction For Use

HERCULES (110V) (220V) USER S MANUAL

LASER SAFETY TRAINING CATALOG LASER INSTITUTE OF AMERICA S MAKE LASER SAFETY YOUR PRIORITY LASER

OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE

COMBINATION ELECTRIC FIREPLACE/HEATER Model #FP23-1A-AKM

Approved by Principal Investigator: Date: Approved by Laser Safety Officer: Date:

Columbia St. Mary s Laser Credentialing

Directive on Laser Safety. NRC Occupational Safety and Health Manual

12 VELOCITY OWNER S MANUAL OPERATING INSTRUCTIONS - MAINTENANCE - SAFETY - TROUBLESHOOTING

Technical Details. Important Safety Instructions READ ALL INSTRUCTIONS BEFORE USING THIS MACHINE

CALIFORNIA INSTITUTE OF TECHNOLOGY LASER SAFETY MANUAL. January, 1998

Laser Safety. The George Washington University Office of Laboratory Safety Ross Hall, Suite B

Homeowner s Installation Instructions & Operating Manual

CONTRACTOR WHS HAZARD STANDARD LASER SAFETY EXTERNAL USE ONLY

order code: EQLA08 user manual

Contact Information. Procedure Title. Procedure Author

Owner's Manual. Please read this document carefully before installing and/or using your vacuum cleaning system.

Micro-Computer Induction Cooktop SR-1885SS

Enteral Pump Instructions for Use To Assemble When the low

Chargebuster Ion Gun Installation, Operation and Maintenance

PROCEEDINGS OF SPIE. ANSI laser standards, education (Z136.5), research, development or testing (Z136.8)

OPERATING INSTRUCTIONS AND MAINTENANCE MANUAL

PART V: WORKING WITH LASERS

OPERATION and MAINTENANCE INSTRUCTION MANUAL. ADU-20B CompriCart

Operator Manual. Xeo Series THE XEO SERIES OPERATOR MANUAL D0075, REV L, 02/12

Chargebuster Ion Gun Installation, Operation and Maintenance

w w w. p r o l i g h t. c o. u k USER MANUAL Order code: EQLA15

2 Ensure that all personnel in the area are wearing the appropriate protective eyewear.

INSTRUCTION MANUAL DIRL-GL

Transcription:

PARTNERING IN ESTHETIC DIGITAL DENTISTRY SM Odyssey 3Watt D I O D E L A S E R Owners Manual

Table of Contents Section 1 - Introduction...........................1 Section 2 -Specifications 2.0 Laser Specifications........................2 2.1 Delivery System Specifications................2 2.2 Warranty.................................2 2.3 Parts List.................................2 Section 3 Laser Assembly Instructions 3.0 Removing the Laser from the Packaging........3 3.0.1 Instructions on Unpacking & Dealer Assistance.3 3.0.2 Shipping Container Information.............3 3.0.3 Contents of the Shipping Container..........3 3.1 Assembling the Laser......................3 3.1.1 Power Cord Installation....................3 3.1.2 Power/Fan Switch - Check..................3 3.1.3 Laser Key Switch Check..................3 3.1.4 Remove Interlock.........................3 3.1.5 Foot Pedal Installation.....................3 3.1.6 Emergency Shutdown Switch...............3 3.1.7 Installing the Disposable Fiber Cassette........4 3.1.8 Summary: Laser Assembly Procedure.........4 Visual Assembly Aides Figure 3.0 Back Panel of the Laser................5 Figure 3.1 Front Panel of the Laser................5 Figure 3.1a Key Switch Activation................5 Figure 3.2 Laser Control Pad.....................6 Figure 3.2a Laser Display Indicators...............6 Figure 3.5 Emergency Shutdown Button - Activating..6 Figure 3.6 Emergency Shutdown Button - Releasing..6 Figure 3.7 Top View - Emergency Shutdown Button and Fiber Retraction Switch.......7 Figure 3.7a Rocker Switch Activation................7 Figure 3.8 Laser Aperture and Cassette Power Receiver Location on the Chassis.........7 Figure 3.9 Disposable Fiber Cassette Base with a 15 Pin Power Coupler............8 Figure 3.10 Removable End Cap on the Fiber Cassette..8 Figure 3.11 Disposable Fiber Cassette without an End Cap........................8 Figure 3.12 Proximal End of the Fiber as it Exits the Cassette................8 Figure 3.13 Locking Hub........................8 Figure 3.14 Alignment Slot......................8 3.2 Facility Requirements.......................9 3.2.1 Power Requirements......................9 3.2.2 Heating and Ventilation....................9 3.2.3 Lighting................................9 3.2.4 Combustible Chemicals and Gases...........9 3.2.5 High Speed Vacuum Systems...............9 3.2.6 Access and Visual........................9

Section 4 - Safety Considerations 4.0 Food and Drug Administration................9 4.1 Statutory Licensure.........................9 4.2 OSHA and its Provisions.....................9 4.3 Laser Safety Program.......................9 4.4 Continuing Education......................12 4.5 In-office Safety Issues......................12 4.5.1 Lighting...............................12 4.5.2 Safety Eyewear.........................12 4.5.3 Test Firing the Laser......................13 4.5.4 Power Changes with Fiber Changes.........13 4.5.5 Fiber Preparations.......................13 4.5.6 Danger Laser in Use Signage.............13 4.5.7 Sharps Disposal.........................13 4.5.8 Plume Evacuation.......................13 4.5.9 Key Switch and Mode Selection............13 4.5.10 Safety Education.......................13 4.5.11 Laser Security..........................13 4.5.12 Emergency Shutdown Options............13 4.5.13 Hard Tissue Procedures..................13 4.6 ANSI Standards...........................13 Section 5 Operating the Laser 5.0 Standby and Ready Status..................13 5.1 Continuous Wave Mode...................14 5.2 Pulsed Energy Mode.......................14 5.3 Tissue Responses to Laser Energy.............14 Figure 5.1 Poor Cleave on the Fiber..............14 5.4 Fiber and Handpiece Care..................14 5.5 Warning (Fiber Cassette Care)...............15 5.6 System Procedures........................15 5.6.1 Selecting the Treatment Center.............15 5.6.2 Checking Foot Pedal Installation............15 5.6.3 Checking the Disposable Fiber Cassette Installation.....................15 5.6.4 Checking Fiber Preparation................15 5.6.5 Checking the Emergency Shutdown Button - Release Instructions......15 5.6.6 Checking the Key Activation and Control Panel Display....................15 5.6.7 Setting Parameters......................15 5.6.8 Setting Power..........................15 5.6.9 Setting Aiming Beam Parameters.........15 5.6.10 Avoiding Use of a Defective Fiber Tip (Shards) 15 5.6.11 Aiming Beam Activation.................15 5.6.12 Test Firing the Laser Before Beginning Procedures...................15 5.6.13 Using Quick/Short Strokes with the Fiber Tip.15 5.6.14 Cleaning the Fiber Tip During a Procedure...15 5.6.15 Preventive Measure Placing the Laser in Standby..............16 5.6.16 Cleave Used Fiber Tip and Discard.........16 5.6.17 Preventive Measure - Cleaning the Fiber Before Retracting It......16 5.6.18 Preventive Measure Turning the Key Off Between Procedures....16 5.6.19 Recording Power, Mode and Time Parameters in the Patients Chart...........16 5.6.20 Self Diagnostic and Monitoring............16 Section 6.0 Systems Components: Preparation, Care and Maintenance 6.0 Disposable Fiber Cassette...................16 6.0.1 Replacing the Fiber Cassette...............16 6.1 Fiber Preparation..........................16 6.1.1 Jacket.................................17 6.1.2 Cladding..............................17 6.1.3 Quartz/Silica Fiber.......................17 6.1.4 Stripping the Fiber.......................17 6.1.5 Cleaving the Fiber.......................17 6.1.6 Initiating the Fiber.......................18 6.1.7 Fiber Disinfection........................18 6.2.0 Maintenance...........................18 6.2.1 Laser Chassis Disinfection.................18 6.2.2 Calibration.............................18 6.3.0 Handpiece Preparation...................19 6.3.1 Preparation for Autoclaving................19 6.3.2 Components of the Handpiece.............19 6.3.3 Providing Adequate Fiber to Cleave, Strip and Initiate..................19 6.3.4 Disposable Tips.........................19 6.3.5 Using the Stripper.......................19 6.3.6 Mounting the Disposable Tip..............19 Section 7.0 Labels, Signs, Warnings and Information 7.1 Federal Compliance.......................19 7.2 Danger Laser in Use Signage...............20 7.3 Laser Classification Aiming Beam...........20 7.4 Laser Classification Treatment Laser.........20 7.5 Caution Changes Not Approved............20 7.6 Caution Avoid Skin and Eye Contact........20 7.7 Caution No User Serviceable Parts Within the Chassis........................20 7.8 Caution Sale or Use of a Laser by Other than a Physician or Dentist..........20 7.9 WARNING Do Not Use Eyewear for Other than the 810 nm Wavelength.........20 7.10 Manufacturer s General Information.........20 7.11 Nominal Ocular Hazard Distance............20 Section 8.0 Servicing 8.1 Warranty Card...........................21 8.2 Repairs / Customer Service Address...........21 Section 9.0 Glossary of Laser Terminology............21 Section 10.0 Selected References on Laser Dentistry....22 Section 11.0 Troubleshooting......................23

Section 1 - Introduction As dentists strive to create the perfect smile, they are often compromised by the technology they use in their practice. Respected leaders from the dental profession and dental equipment manufacturers have sought to identify the most practical and least invasive technology available to deliver restorative and preventive care. Today, thanks to continuing efforts by these industry leaders, we have seen the introduction of many new devices that have advanced the dentist s ability to perform at the highest standards. The Odyssey 3Watt DIODE LASER from Ivoclar Vivadent, Inc. represents the latest solid state diode laser technology available for soft tissue modification and preventative care. Now featuring wireless power control. Unlike solid state lasers that utilize a man-made rod of elements such as yttrium, aluminum, and garnet, doped with a rare earth compound like Erbium, the diode has components that have become known for their durability, dependability and longevity. We are, of course, referring to semi-conductor crystal technology like that found in televisions, DVD players, telephones and many more of those household products that we have learned to rely on each day. The major components of the Odyssey 3Watt are semiconductor chips made from Aluminum, Gallium and Arsenide, together commonly referred to as AlGaAs. They are activated or pumped by passing an electrical current through the diode to produce an elliptical shaped display of monochromatic light that can be focused into a very small point and placed into the delivery fiber. The wave length produced by the diode is approximately 810 nanometers (nm) and produces invisible non-ionizing thermal radiation that does not create changes in cellular DNA. The diode is air cooled and highly efficient when used correctly. For safety, the diode features several ways to stop energy flow if the operator wishes to deactivate the laser. The safety system includes a choice of an emergency shutoff switch, a key, a power switch, a power cord or an electrical plug. Any of these items can be used to shut down the laser. The design and technology used in the Odyssey 3Watt allows the dentist or hygienist* to transport the laser between different operatories. It has a lightweight and durable chassis that is designed to use 110 120 V electricity found in most dental offices. Training is recommended and opportunities for such are available through such outlets as Ivoclar Vivadent, Inc., please visit our web site for training dates and locations at www.getodysseylaser.com, the Academy of Laser Dentistry, dental schools and many dental continuums. You should also ask your authorized dealer representative for the names of dentists in your area who have a laser and who could help you in a mentoring capacity. There are many applications for using this laser system and you will be amazed of the results and wonder how you ever practiced dentistry without the Odyssey 3Watt. Laser safety is paramount in importance and each office should quickly develop and implement a laser safety program and appoint a laser safety officer to be responsible for the laser. Their duties include management of the laser and all accessories as well as training office personnel in all aspects of laser safety. More duties are outlined in Section 4, subsection 4.4 of this manual. Remember: Always test fire the laser outside the mouth before using it on a patient. The doctor or hygienist, the patient and any staff member present in the operatory should be wearing the appropriate safety eyewear whenever the laser is being operated. Strict adherence to protocols for safe laser use is essential. * In States / Provinces where the Dental Practice Act allows hygienists to utilize a laser. 1

Section 2 Specifications 2.0 Laser Specifications Weight 5.5 lbs 2.1 Delivery System Specifications 2.1.1 Quartz silica fiber 1 cassette approximately 20 each (6 meters) Fiber Diameter 400 microns Dimension in inches (H X W X L) 5.5 x 6.25 x 8.75 Autoclavable No Laser Classification (Per 60825) Laser Diode Wavelength Laser Aiming Beam Beam Divergence Class 4 Laser Device 810 nm ±20 nm 630 nm 660 nm ±15 nm 9 degrees ± 1 degree Retractable fiber delivery cassette (non-autoclavable) 1 2.2.2 Handpiece - Autoclavable 2 2.2.3 Tips for the Handpiece Autoclavable and disposable 50 per box 2.2.4 Laser Aperture -ST Adapter Type Yes Power Range Hertz Rate in Pulsed Mode - fixed 100 mw to 3 Watts 1.0 Hz 2.2 Warranty (See also Section 8) Laser Fiber and Cassette 1 year Parts and Labor 90 days Parts and Labor Pulse Duration - fixed 0.5 seconds Duty Cycle pulsed mode 50% continuous wave 100% Aiming Beam (3 mw ) Audible Notification Visual Notification Power Requirements Amperage Yes Yes Yes 110-120 VAC @ 60 Hertz 220-240 VAC @ 50 Hertz 1.5 Amps @ 110-115 V AC 0.75 Amps @ 210-230 V AC 2.3 Parts List Part Identification number Fiber Stripper Tool 579031 Fiber Cleaver 579032 Protective Glasses Solid Sides 598558 Protective Glasses Plain Sides 598559 Protective Glasses Window Sides 598560 400 Micron Fiber Cartridge System 579036 Package of (20) Handpiece Tips - 60 degree angle 603502 Handpiece 579038 Handpiece Fiber Lock (inside of handpiece) 595477 Power Cord 579041 Odyssey 3Watt Owner s Manual 603505 Odyssey 3Watt Clinical Guide 603499 Laser Safety Sign 603504 2

Section 3 - Laser Assembly Instructions 3.0 Removing the Laser from the Packaging 3.0.1 - Instructions on Unpacking & Dealer Assistance: Your local authorized Ivoclar Vivadent dealer can provide a representative to assist you when you are ready to remove the laser from its shipping container. You may remove it yourself if you wish to get a head start on the set-up, but please do not attempt to unpack the laser and install the various components without reading this section first. If you are unsure about any aspect of the assembly, call your authorized dealer representative for assistance. 3.0.2 - Shipping Container Information: Though highly unlikely, you may need to return the laser for service or repair and the shipping container you received with your laser has been especially designed to transport the laser. 3.0.3. - Contents of Shipping Container: The contents of the shipping container should include the following: (1) Odyssey 3Watt Laser Unit with 400 Micron Fiber Cartridge System (1) Fiber Stripper Tool (1) Fiber Cleaver (1) Protective Glasses - Window Sides Style (1) Protective Glasses - Solid Sides Style (1) Protective Glasses - Plain Style (2) Packages of (20) Handpiece Tips - 60 degree angle (2) Handpiece (autoclavable) (1) Laser Key (1) Foot Pedal (1) Power Cord (1) Odyssey 3Watt Owner s Manual (1) Odyssey 3Watt Clinical Guide (1) Laser Safety Sign (1) Warranty Information Please check all items sent with your laser to insure that all components are accounted for. 3.1 Assembling the Laser Each of the following items should be inspected, inserted into the appropriate receptacle, and when applicable, locked using the locking hub. 3.1.1 Power Cord Installation: Remove the power cord from the Odyssey 3Watt package and plug the power cord into the appropriate receptacle on the back of the laser. See Figure 3.0 on page 5. To prevent power surges due to electrical storms or spikes in line voltage, you should use a power strip with a circuit breaker or unplug the laser when you are not present. Plug the power cord into a 110 Volt AC outlet rated at 60Hz. 3.1.2 Power/Fan Switch: The power/fan switch for the laser and fan is the first item you turn on each day. The switch is located on the rear panel of the laser near the lower right hand corner. See Figure 3.0 on page 5. 3.1.3 Laser Key Switch: The laser key switch is the major circuit breaker for your laser. It will be the second item you turn on when activating the laser each day. Place the laser key into the key receptacle located near the lower right corner on the front of the laser. See Figure 3.1 and Figure 3.1a on page 5. Check the key switch by turning the key clockwise approximately 1/4 of a turn. This is the (ON) operating position for the key. The fan will start when the power/fan switch is on and the key is turned. Prior to leaving the office, the laser safety officer should check to see that the key switch has been turned off and the key removed and stored in a safe place. 3.1.4 Odyssey 3Watt Remote Interlock: The Odyssey 3Watt Laser is equipped with a Remote Interlock Jack. The Remote Interlock Jack is provided so that a clinician may install the laser in a dedicated laser treatment room such that the laser will be interlocked with the entrance door of the room. In such an interlocked installation, the laser would shut off anytime the door is opened, hypothetically, to protect the person s eyes who is entering the room. It is recognized that such installment is not facilitated nor required in many operatories or clinics. To that end, the Remote Interlock is available to any practitioner that requires it. The Remote Interlock Jack is located and clearly labeled on the rear of the laser. The miniphono jack is wired in the normally closed position; meaning that no further action is required to operate the laser without the interlock loop. If the interlock loop is desired you may purchase the loop from a local electronics store. You need only inform the local electronics store that you require a mini (1/8 ) mono-phono jack wired into a normally closed momentary switch and select the switch design that best suits your needs. To install the loop, install the switch on the door and simply plug the mini phono jack into the Remote Interlock Jack on the rear of the laser. 3.1.5 Foot Pedal Installation Procedure: Locate the foot switch receptacle on the back of the laser and insert the proximal end of the foot switch cord into the marked receptacle on the back of the laser. Then check to see if it is securely locked. 3.1.6 Emergency Shutdown Switch: Before you can activate the laser, you must first check to see if the emergency shutdown switch has been depressed prior to shipment and locked in the off position. The switch is the red button located on the top of the laser. Release the switch by grasping it between your thumb and index finger, and gently depress it as you turn the button 1/4 turn clockwise. You will feel it click or release and the spring loaded button will pop 3

back up. (See Figure 3.5 and Figure 3.6 on page 6). The display on the control panel should now be lighted. If the cartridge lock plate is not fully engaged, this will prevent the display from lighting up. If you find that the display is still not operational, check all attachments, keys and switches to see that they are securely installed and that you have an active wall plug for electricity. (See Troubleshooting Section pg. 22). If the laser can not be activated, please contact your authorized distributor who can help you to get a replacement. If the control panel does light up when you release the shutdown switch, you should test the switch again by depressing it to turn the laser off. If the shutdown switch is locked down, the laser will not turn on. 3.1.7 Installing the Disposable Fiber Cassette: When it comes time to install a new cassette, carefully remove the fiber cassette from the packing. Remove the end cap from the old cassette and remove the cassette (do not dispose of the end cap). To install the new cassette, locate the end of the new fiber. Avoid placing your finger on the proximal end of the new fiber. See Figure 3.12 on page 8. Human oils on the fiber can burn and diminish the effective transmission of radiant energy once the fiber is installed. While opening the door on the laser aperture with one hand, insert the proximal end of the new fiber into the laser aperture. See Figure 3.13 on page 8. Be careful to align the slot on the fiber end with locking hub of the aperture. See Figure 3.14 on page 8. Once inserted, turn the fiber locking hub to the right (clockwise) to lock the fiber securely into the ST connection. Once the fiber is connected, align the dove tail of the cassette with the dove tail receiver on the chassis wall. See Figure 3.8 on page 7. This will help assure that the power coupler (a serial port) on the underside of the fiber cassette will be aligned with the receiver on the laser chassis. Slowly press the cassette into place. See Figures 3.8, 3.9 & 3.11 on pages 7-8. Once connected, replace the end cap to secure the fiber cassette. The laser will not operate until the end cap is firmly latched in place. Turn the key switch and power/fan switch to the on position. Test your power connection by advancing the fiber, using the rocker switch on the top of the cassette. Depress the switch at its most anterior point to see if it moves the fiber forward. initiated fiber for this check because the energy would stop at the tip. 3.1.8 Summary: Laser Assembly Instructions: 1. Attach the laser s power cord and place the plug into the wall receptacle. 2. Attach the foot pedal. 3. Attach remote interlock, if desired (not required). 4. Check the emergency shutdown switch to see that is has been released. 5. Install the fiber cassette and attach the fiber to the ST connector and lock the locking hub. 6. Place the key into the key switch receptacle (see Figure 3.1a on page 5) and turn the key to the right. The control console should light up. 7. The light on the console indicating the operating status should be in the standby mode. NOTE: When the power cord is plugged in, the fan/ power motor switched on, the wireless foot pedal receiver attached and the emergency shutdown released, the key will turn the laser on. Next, select a low power to check transmission of energy (1 Watt 1.5 Watt). Put on the safety eyewear to protect your eyes. Place the laser in ready mode and depress the foot pedal to activate the laser while holding the fiber approximately 2-4 mm away from a piece of paper with printing on it. Aim at the printing, the paper will begin to burn in 1-2 seconds. See section 6.1.6. on page 15. You should be able to see the aiming beam on the paper and the fiber tip should produce enough heat to let you know it is installed properly and the fiber is sound. You do not want to use an 4

Manufacturer s Specifications, Regulatory and General Information Emergency Shutdown Visual Assembly Aides Fiber Cassette End Cap Release Disposable Fiber Cassette Figure 3.0 Back Panel of the Laser Power/Fan Switch Fuse Receptacle Remote Interlock Connector Power Cord Receptacle Foot Pedal Connection Odyssey 3Watt Digital Display Figure 3.1 - Front Panel of the Laser LED Screen Fiber Extend and Retraction Switch Label required by Safety Standards Fiber Aperture Control Panel Key Switch Handpiece Holder Label required by Safety Standards Figure 3.1a Key Switch Activation Turn the key 1/4 of a turn to the right to turn on the laser 5

Figure 3.2 - Control Pad Settings Laser On: This illuminates when the foot pedal is depressed. This indicates that the working beam (810nm) energy is being emitted. Ready Indicator: This illuminates when the READY key is pressed. It will blink for 3 seconds, then remain steady. Once it is steady on, the aiming beam will be activated. Continuous and Pulse Mode: This illuminates when the unit is in continuous mode. Use the MODE key to toggle back and forth between Continuous and Pulse. Working Beam Setting: Indicates the working beam power output setting. Mode: Press this key to select Continuous or Pulse mode. Aiming Beam: Use these to adjust the aiming beam power. Increase and Decrease: Use these to adjust the power settings from 0 to 3 watts. Hold the desired key to rapidly change the value. Adjustable Aiming Beam: The Odyssey 3Watt diode laser is actually two lasers in one, the infrared 810 nanometers wavelength laser which performs the actual treatment of the soft tissue and the 630 nanometers laser pointer which illuminates the direction where the laser will be used, allowing the operator to aim prior to and during laser activation. In addition the aiming beam control allows for five levels of intensity of the aiming beam for guidance in the operatory field if desired. Each bar represents 20% of maximum intensity. Thus 0 bars means the aiming beam is shut off. Figure 3.2a - Display Indicators Laser On Indicator Working Beam Power Setting Aiming Beam Setting Check of the Emergency Shutdown System Ready Indicator 1. Depress to shutdown (feel the click) Power is OFF in the down position Emergency Shutdown Switch Continuous Mode Pulse Mode 2. Depress slightly and turn right 1/4 of a turn turn to release. The emergency stop will pop up and the LED screen will light. Laser On: This illuminates when the foot pedal is depressed. This indicates that the working beam (810nm) energy is being emitted. Figure 3.5 Shutdown Figure 3.6 Release 6

Aluminum handle Fiber extend and retract rocker switch (See below) Depress the RETRACT or EXTEND rocker switch to advance or retract the fiber RETRACT EXTEND Figure 3.7 Top View Figure 3.7a Rocker Switch Activation Receiver for the Fiber Cassette end cap locking device Laser Aperture with ST Connection: The proximal end of the fiber inserts here and is locked by turning the locking hub to the right Dove tail Alignment guide Receiver for the Power Coupling Device Figure 3.8 Laser Aperture (ST adapter) and power receiver for the power coupling device on the base of the cassette. 7

15 pin Power Coupling Device Depress at the base of the arrow to release the cover for the fiber cassette Figure 3.9 Base of the Fiber Cassette and the 15 Pin Power Coupling Device Figure 3.10 Removable End Cap of the Fiber Cassette Proximal end Figure 3.11 Disposable Fiber Cassette without the End Cap Figure 3.12 Fiber exit from the Disposable Fiber Cassette The proximal end of the fiber extends from the internal wall of the fiber cassette and must be attached to the laser by inserting it into the Laser Aperture (ST adapter). Alignment slot Figure 3.13 Locking Hub Figure 3.14 Alignment Slot - Frontal View Once it is inserted into place, it is locked by turning the locking hub approximately 1/8 turn to the right. 8

3.2 Evaluating the Facility and Environmental Considerations (United States) In order to insure the safe use of the laser in your facility, please check to make sure that the proposed location has the following: 3.2.1 Power Requirements: 110-120 V AC ±10 % at 60 Hz 3 Amps Frequency range 45-63 Hz 9 volt Lithium battery 3.2.2 Heating and Ventilation: The room where the laser is used should have good cooling and heating system so that the laser can be operated within the optimum range of 20º - 30ºC (68º - 86º F). Avoid storing or transporting the laser in temperatures below 0º Celsius (32º F). 3.2.3 Lighting: Overhead lighting and or dental unit light should provide enough illumination to allow good operator vision when activating the laser intra-orally. 3.2.4 Combustible Chemicals and Gases: All gases that are combustible or support combustion and are used in the operatory area where the laser is in use must be turned off during the procedure. Cleaning supplies or other flammable chemical compounds should be stored in an area away from the surgical site in order to avoid possible combustion. 3.2.5 High Speed Vacuum Systems: Plume evacuation is a priority when vaporizing tissues. The Clinician or operator, and their chair-side assistants should keep themselves and the patient safe by using a high volume vacuum system and high filtration masks that are suitable for virus and bacterial control. 3.2.6 Access and Visual: Access to the treatment area should allow the dental team to restrict entry while the laser is in use. There should be a Laser In Use Safety Sign placed in a designated area adjacent to the entry into the treatment area. See Figure 7.1 on page 15. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. (Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense). 4. Safety Considerations The safe use of the Odyssey 3Watt is the responsibility of the entire dental team including the doctor, the laser safety officer appointed from the dental office team. Protocols for the safe use of lasers have been developed by a combination of medical and dental professionals working in concert with educators at the university level, scientists and laser manufacturers. Dental professionals have had to develop protocols and guidelines for using the laser on oral soft tissues. Sound judgment and the concern for patient safety should be the basis of all laser care. The following entities have influence over laser use. 4.1 Statutory Licensure Usually, states or provinces do not have a specific licensure requirement for use of a laser by a dentist. Most states require a hygienists to attend licensure training that includes both a lecture and hands-on training. Prior to using the laser, the hygiene applicants are required to pass a proficiency test for certification. These courses are usually taught by members of the Academy of Laser Dentistry who possess instructor credentials. Some US States and Canadian Provinces have implemented specific rules governing the use of lasers. You should check your State or Province s website to determine applicability of any requirements to your office. 4.2 OSHA and its Provisions Worker safety is the responsibility of the employer and is regulated by OSHA (Occupational Safety and Health Administration), a division of the U.S. Department of Labor. OSHA has issued no specific standard for safe use of lasers but recognizes ANSI standard Z136.1 as a source for analyzing safety with respect to medical lasers. For more information see OSHA Technical Manual (TED 1-0.15A) Section III, Chapter 6, 1999. Ivoclar Vivadent Inc. recommends implementation of a Laser Safety Program for the safety of your patients and office staff in connection with the use of the Odyssey 3Watt Diode Laser. 4.3 Laser Safety Program We recommend implementation of a Laser Safety Program appropriate for your dental office. The plan may include the following: Delegation of authority and responsibility for supervision and control of the laser to a designated Laser Safety Officer; Minimum Training requirements for users of the laser Laser security against unauthorized use of the laser Standard operating procedures to regulate the work environment in order to protect the patient and office staff from laser hazards. The safe use of a laser is the responsibility of the Laser Safety Officer (LSO) who can be a full or part-time employee, or the laser operator. It is their responsibility to train the staff, main- 9

tain records concerning training and the laser s performance, perform safety checks and prepare the laser for use on a daily basis. The LSO must keep records of any incidents that relate to the failure of the laser or any adverse effects related to laser therapy and report such incidents as prescribed by law. The laser safety officer assures that a medical follow-up has been sought or has occurred following any adverse incident during treatment. The LSO is responsible for the training of all office personnel who are involved with the laser preparation and use. Daily checks of the facility and equipment are also the LSO s responsibility. The LSO should test fire the laser each day prior to beginning each treatment procedure. For more information on the contents of a Laser Safety Plan, you can review ANSI Standard Z136.3 for Safe User of Lasers in Health Care Facilities or TR IEC 60825-8 Guidelines for the Safe Use of Medical Laser Equipment. 4.4 Continuing Education The laser safety officer should insure that the operator and staff attend laser courses taught by qualified laser educators. Ongoing reviews of laser safety procedures should be a part of normal office routine. 4.5.5 Fiber Preparations: After cleaving and stripping the fiber, photo-initiation of the fiber tip will allow the operator to remove tissue more rapidly when contact procedures are indicated. Gingival debris on the tip will retain the heat and it should be removed. The tip will also begin to blacken and deteriorate as it retains the heated debris and can break if not removed by cleaving it. Clean the tip often using a 2 X 2 gauze sponge moistened with water. Do not use combustible liquids to moisten the 2 X 2. 4.5.6 Danger - Laser In Use Signage: Each operatory where the Odyssey 3Watt is used should have a laser in use sign placed at the operatory entrance when a procedure is in progress. This signage will help to eliminate eye damage caused by inadvertent exposure to laser energy. See Figure 7.1 on page 17. 4.5.7 Sharps Disposal and Sponge Removal: Remove cleaved fiber remnants and place them into a sharps container for disposal. All sponges used for cleanup of lasers and fibers should be disposed of in a bag for contaminated soft products. 4.5 In-office Safety Issues 4.5.1 Lighting: Always use the Odyssey 3Watt in a well lighted and ventilated area. Make certain that chemicals or gases capable of supporting or causing combustion are not present when using the laser. Use high volume vacuum to remove the laser plume and provide a high filtration masks for all people present in the treatment area during lasing. 4.5.2 Safety Eyewear: While using the Odyssey 3Watt laser, doctors, auxiliary staff, patients, and anyone attending them in the operatory must wear the appropriate safety eyewear that has been designed for use with the 810 nm wavelength. Never point the laser tip directly at the face, eyes or skin of anyone while emitting energy. The aiming beam is also capable of causing eye damage 4.5.3 Test Firing the Laser: Always test-fire the Odyssey 3Watt prior to using it intra-orally. Using a power of 1 Watt continuous wave or less, place the laser in the ready mode. Then, activate the laser for 1-2 seconds while aiming the fiber onto a 2X2 gauze sponge wetted with water. Do not use alcohol or any other combustible material to wet the 2X2 sponge as it may ignite. 4.5.4 Power Changes With Fiber Changes: Switching to a smaller diameter fiber will increase the density of the power at the fiber tip. As a result, you may need to adjust your power downward. Increasing the power may be required when switching to a larger diameter fiber. In order to achieve the same rate of work after changing fiber diameters, remember this: a smaller diameter fiber will require less power and conversely, a larger diameter will require more power. 10 4.5.8 Plume Evacuation: Use high volume evacuation suction during procedures to remove laser smoke or plume debris. Use masks suitable for viral filtration. Caution - laser plume may contain viable tissue particulates. 4.5.9 Key Switch and Mode Selection: When the key switch is in the ON position (turn to the right), the laser has been enabled and can be activated while in the READY status. When not in use, insure that the key has been turned off or that the laser is placed in the STANDBY status. 4.5.10 Safety Education: Provide comprehensive safety procedure training for all office personnel and include the staff in all outside laser courses when possible. Be certain that all members of the dental team understand how the laser works and can advise patients as to their safety and advantages over conventional procedures. 4.5.11 Laser Security: To prevent the unauthorized use of the laser while not in use, the key should be removed from the unit and maintained by the Laser Safety Officer. 4.5.12 Emergency Shutdown Options: Any of these mechanisms can be used to shut down the emission of laser energy in a real or perceived emergency. 1. Depress the emergency shutdown button 2. Foot Pedal remove your foot to stop lasing 3. Key turn off the key 4. Switch the Power/Fan to the off position (O) 5. Power Cord unplug from the wall outlet 4.5.13 Hard Tissue Procedures: The Odyssey 3Watt diode is not an appropriate laser for hard tissue procedures. The diode laser is attracted to melanin, hemoglobin and to some extent to water and oxygenated hemoglobin. Avoid prolonged

exposure of the energy when working in and around the cervical areas of the tooth. Due to the thin layer of enamel in this area, the laser s energy may be absorbed by the hemoglobin in the pulp and pulpal hyperemia may occur. Extended exposure to laser energy could lead to pain and possible pulpal necrosis. 4.5.14 State/Provincial Certifications: Check your state or provincial government website to determine if there are any laws pertaining to the use of lasers. Some governments require a separate certification, inspection and fees in order to use a laser of this type. 4.6 American National Standards Institute (ANSI) - Safety Standards ANSI is a non-governmental, non-profit agency that has established guidelines and safety standards for the use of lasers and other electro-optics. The provisions of ANSI Z 136.3 outlines standards for lasers used in dentistry and the assessment of laser risks. ANSI also establishes guidelines for safety eyewear and classifies all lasers based on their potential for damage to eyes or tissue. 5. Operating the Laser The Odyssey 3Watt will deliver energy in either a continuous wave (CW) mode or in a pulsed mode which are called temporal emission modes (time related modes). Selecting the appropriate mode is a factor of controlling target tissue temperatures and the efficiency of energy delivered. The pulse duration (.05 seconds) and the number of pulses per second (1) have been fixed by the manufacturer using a 50 % duty cycle and you will therefore need only to change the power and mode. 5.0 Standby and Ready Status Once you have turned on the power /fan switch on the back of the laser and the key switch on the front, you will see the laser LED screen light up. The status of the laser will be in Standby which is a non-active status. The laser will not emit energy while in Standby, even if you depress the foot pedal. On the left of the control panel you will see the ready button. See Figure 3.2. Press this key to place the unit into the ready status. When this key is pressed whatever settings are displayed on the unit will be saved into the unit s memory for the program setting shown. The settings saved into the three programmable modes will be retained even if the laser is turned off. Any adjustments to the manual settings after the laser is in the ready status will not be saved to memory. The laser will now be ready to emit energy as you depress the foot pedal. 5.1 Continuous Wave (CW) Mode In setting up the laser while in the CW mode, you will deliver the amount of power in one second that you have set the laser for. i.e., set the laser for 2 Watts CW and while in the ready mode, the laser will deliver 2 Watts per second as long as you have the foot pedal depressed. The CW mode is generally the fastest way to ablate tissues but heat can build up and cause 11 collateral damage to the target and adjacent tissues. Cool the tissues being lased by using periodic blast of air from a triplex syringe and high speed suction. You may use water to cool in areas where there is prolonged exposure to the laser s beam. Avoid using the air syringe when you have an opening in soft tissue adjacent to or within the surgery site. An air embolism may occur as a result of air captured within the tissue during the cooling process. 5.2 Pulsed Energy Mode Pulsing the laser energy will allow some cooling of the tissue in between emissions of energy. The duty cycle is the percentage of the time in each second that the laser is emitting energy. The pulses per second, the duty cycle and the energy intensity per pulse will determine your average power. In the pulsed mode, the Odyssey 3Watt is programmed to deliver 1.0 pulse per second with each pulse lasting for 0.5 seconds. The duty cycle is set for 50% so you will have 1 energy pulse with 1 period of rest with no energy between each pulse. If the laser is producing energy at 1 time per second for 0.5 seconds per pulse, you will be producing energy for a total of 0.5 seconds. The result will be an average power per second that will be 50% of what you have set the laser for. Therefore, when using pulsed energy, you will have to adjust your power upward in order to achieve the same rate of work as the same power set in CW. 2 Watts of Pulsed energy will be the same average power output as 1 Watt CW. 5.3 Tissue Responses to Laser Energy Maximum results will be achieved by regulating the power and the speed that the operator moves the fiber tip. Tissue Charring is an undesirable after effect of too much power or the tip moving too slowly. Always use the least amount of power necessary to complete your procedure. The ideal tissue response will show little or no discoloration after lasing and there will be less residual damage and faster healing. Avoid penetrating or damaging the periosteum, and do not use the laser on alveolar bone. Because the laser energy is attracted to melanin and hemoglobin, power must be reduced when treating patients with darker soft tissue. Always begin lasing with the lowest power you can use to remove or modify the target tissues. Avoid damage to the gingival sulcus by moving the fiber tip quickly and using low power settings. Check to make sure you have a good cleave of the fiber so that no shard is present on the tip. A shard may act as a miniature scalpel and damage the small blood vessels, thus preventing hemostasis and coagulation. (See Fig. 5.1) Figure 5.1 Fiber with a shard (a bad cleave) 5.4 Fiber and Handpiece Care Do not continue to use the fiber tip once you have observed that the tip has a blackened appearance that is greater in

length than 2-4 mm from the previous cleave spot. See (Figure 6.1) The protein debris of gingival tissue accumulates on the tip during surgery and retains extreme heat that can cause rapid tip deterioration and subsequent breakage. This is especially important when using the laser for periodontal pocket debridement. During surgery, clean the tip often using a 2 X 2 sponge moistened with water. Do not use alcohol or other combustible liquids to moisten the 2 X 2 gauze sponge and do not use the sponge while the tip is hot. Always use a cold disinfectant solution like BIREX to wipe off the fiber jacket before retracting the fiber. The fiber can be advanced or retracted by using the rocker switch on the top of the fiber cassette. The fiber cassette is a removable assembly with a plug-in capability that provides power for the internal retraction device. This cassette is disposable after all fiber has been used. This cassette is not autoclavable. In vitro studies performed at the University at Buffalo validate that the infection control measures indicated by Ivoclar Vivadent can prevent the transmission of oral microorganisms between sites in a patient's mouth and between patients. Dr. Violet L. Haraszthy Assistant Professor of Restorative Dentistry Dr. Joseph J. Zambon Professor of Periodontics and Endodontics University at Buffalo, School of Dental Medicine Be advised about the potential hazards when inserting, steeply bending or improperly securing the fiberoptics to the chassis. Radiation exposure may occur in these instances which could be harmful to yourself, your staff and your patient. Special care should be taken not to break or snap the fiber. As the Aiming Beam passes down the same delivery system as the Working Beam, it provides a good method of checking for integrity of the delivery system. If the aiming beam spot is not present at the distal end of the delivery system, its intensity is reduced or it looks diffused, this is a possible indication of a damaged or not properly working delivery system. The handpiece should be autoclaved after each patient and a new disposable tip applied. Recommended autoclave cycle is 132 C at 27 psi for 15 min. Be sure to remove the fiber locking clasp and disposable tip prior to autoclaving as these parts will melt in the autoclave. 5.5 WARNING-!!!! DO NOT PLACE THE FIBER AND FIBER CASSETTE IN THE AUTOCLAVE TO STERILIZE THE SYSTEM. USE COLD DISINFECTING SOLUTIONS TO WIPE DOWN THE FIBER AND THE EXTERIOR OF THE LASER FIBER CASSETTE. 5.6 Systems Procedures 5.6.1 Treatment Area Requirements: The laser should be placed in an area with good ventilation and lighting. The electrical service required is a 110 Volt AC outlet - 60 Hz. The area where the laser is placed should be free of standing water. Combustible gases or those that support combustion should be turned off and all flammable materials or chemicals stored in the area should be removed. 5.6.3 Fiber Cassette: Before using the laser, check the fiber cassette to see that it is firmly seated. Depress the extend/ retract rocker switch on the cassette to see that it is functioning properly. After Checking the Fiber Cassette Power, extend some fiber from the cassette. 5.6.4 Fiber Preparation: You will want to have approximately 3 feet (1 meter) of fiber available to strip the fiber, cleave the fiber, initiate the fiber and disinfect the fiber (See Section 6.1). When the fiber is prepared, the handpiece and tip should be attached (See Section 6.3). Once prepared place the handpiece into the holder on the side of the laser. See Figures 3.7 and 3.7a on page 7. The rocker switch for fiber cassette operation should not be depressed during laser operation. 5.6.5 Emergency Shutdown Button: Check the Emergency Shutdown Button to see that it has not been depressed. If it has, release it by depressing it slightly as you turn it 1/4 turn to the right. 5.6.6 Key Switch: Turn the Key Switch on the front of the laser to the on position by turning it approximately 1/4 turn to the right (clockwise). The control panel should light up and show the laser is in Standby Mode. 5.6.7 Setting Parameters: Review your power and mode requirements and then depress the mode button to select either Continuous (CW) or pulsed mode. The mode you have selected will be displayed just below the LED screen and will be designated by a small light in the corner of either the pulse or CW panel display. See Figure 3.2a on page 6. 5.6.8 Select Your Power by pressing the up or down arrow until you have reached the desired Wattage. Beginning with a low of 0.1 Watts, the power increases in increments of 100 mw up to a maximum of 3 Watts (CW ). By holding the up or down arrow, you can have an un-interrupted increase until you reach your desired power. 5.6.9 Aiming Beam: The aiming beam can be turned on and off by pressing the aiming beam button on the laser control panel. Press this button to progressively increase the aiming beam intensity from 0 (off) to full intensity. Press the key once more to cycle back to 0. Each bar represents approximately 20% of the aiming beam output. Note: that adjusting the intensity of the aiming beam, has no effect on the output power of the primary laser. 5.6.10 Examine the Fiber Tip to insure that you have not created a fiber tip shard during the cleaving process. The shard can act like a miniature scalpel and cause damage while diffusing the light beam and lowering the laser s power. See Section 6.1 for fiber preparation. 12

5.6.11 Depress the Ready Button and the aiming beam should light after 2 seconds. Review your power and mode requirements and then depress the mode button to select either CW or pulsed mode. 5.6.12 Test Fire the Laser outside the mouth by activating the laser into a 2 X 2 gauze sponge that has been wetted with water to prevent combustion. Do not use flammable liquids to wet the sponge. 5.6.13 Depress the Foot Pedal and make short quick strokes at the lowest power that you can to remove the target tissues while lightly contacting it. 5.6.14 Remove Your Foot from the foot pedal and use a clean 2 X 2 gauze sponge wetted with water to remove debris from the fiber tip. Do not use flammable liquids to wet the sponge. 5.6.15 Place the Laser in Standby Mode until you are ready to start another procedure. 5.6.16 Cleave used fiber tip and discard in suitable biowaste disposal. 5.6.17 Wipe the Outside of the Fiber using a disinfectant or sterilization solution and then retract the fiber by depressing the back of the rocker switch. Do not retract the distal end of the fiber into the cassette. 5.6.18 Turn the Key Off if you are not going to be starting another procedure. 5.6.19 Record the Powers and total lasing times used for each procedure in the patient s chart. Example: Patient Name: Mary Jones Procedure: Gingivectomy # 6 and # 7 #6 Lasing time 90 seconds @ 2.0 Watts CW air cooled #7 Lasing time 60 seconds @ 1.5 Watts CW air / water spray 5.6.20 Odyssey 3Watt Self Diagnostic and Monitoring: When the Odyssey 3Watt Laser s microprocessor detects an issue with performance it will immediately notify you by way of an audible beep. There are two different ways in which the Odyssey 3Watt will alert you to any issues: 2. Continuous audible beep when the unit is turned on. If you are operating the Odyssey 3Watt and it emits a constant audible beep whether or not the foot pedal is engaged, the microprocessor has determined that the laser has either low power or a general fault has occurred. In this event the Odyssey 3Watt Laser should be turned off allowed to sit for 5 minutes and turned on again. If the Odyssey 3Watt then performs without beeping the microprocessor has been able to make operational adjustments to the laser and the unit will perform its functions. If, however, upon restart the unit continues to beep, the microprocessor was unable to adjust the unit enough and the unit will need to be sent in for adjustment. (8.2 Repairs & Returns to Ivoclar Vivadent, Inc.) 6. System Components: Preparations, Care and Maintenance The fiber optic element of a laser is responsible for carrying the light from the diode array to the tissue being treated. The dental laser fibers are usually made of quartz, sapphire, silica or a combination of those elements. Quartz/silica is the most popular product used in diode lasers. Be advised about the potential hazards when inserting, steeply bending or improperly securing the fiberoptics to the chassis. Radiation exposure may occur in these instances which could be harmful to yourself, your staff and your patient. Special care should be taken not to break or snap the fiber. As the Aiming Beam passes down the same delivery system as the Working Beam, it provides a good method of checking for integrity of the delivery system. If the aiming beam spot is not present at the distal end of the delivery system, its intensity is reduced or it looks diffused, this is a possible indication of a damaged or not properly working delivery system. 6.0 Disposable Fiber Cassette The fiber cassette is a removable assembly with a plug-in capability that provides power for the internal retraction device. This cassette is disposable after all fiber has been used. This cassette is not autoclavable. 1. Continuous audible beep when foot pedal is engaged. If you are operating the Odyssey 3Watt with the foot pedal engaged and the Odyssey 3Watt emits a constant audible beep and stops the beep when you release the foot pedal, the microprocessor has determined that the laser power output has fallen below the set level. In this event the Odyssey 3Watt Laser should be turned off and allowed to sit for 5 minutes then turned on again. If the Odyssey 3Watt then performs without beeping, the microprocessor has been able to make operational adjustments to the laser and the unit will perform its functions. If, however, upon restart the unit continues to beep when the foot pedal is engaged, the microprocessor was unable to adjust the unit enough and the unit will need to be sent in for adjustment. (8.2 Repairs & Returns to Ivoclar Vivadent, Inc.) 13 The retractable fiber also provides an opportunity to use a variation of the single use disposable concept infection control. Each time the laser fiber is prepared for use, the previously used terminal end is cleaved and discarded. The newly exposed, never used, portion of the fiber is then inserted into a sterilized handle and disposable tip. In my opinion the Odyssey Laser complies with CDC recommendations, and OSHA expectations for critical (Spaulding classification) items used in dentistry. Terrence J. Thines, D.D.S., M.S. Chair-Infection Control Committee Department of Oral Dionastic Sciences School of Dental Medicine State University of New York at Buffalo

6.0.1 Replacing the Fiber Cassette: A black mark on the fiber indicates 6 of fiber remaining on the cassette. At this point, a new fiber cassette should be ordered. When the cassette is empty, it should be released from the laser aperture and gently removed. Slowly lift the cassette up using the dove tail as a guide. The fiber cartridge can be thrown away. Note: Retain the end cap, do not discard the end cap. Review the installation procedure again on page 5. 6.1 Fiber Preparation The fiber cassette contains approximately 20 (6 meters +) and is wound onto a spool. The fiber itself has three components: Jacket Cladding 400 micron quartz/silica fiber 6.1.1 Jacket: This is the protective cover for the fiber system and usually is made of a synthetic material that is clear or white in color. There can be other colors used but there are no standardized color systems to denote the diameter of the fiber or its use. Figure 6.0 Adjustable Fiber Stripper Tool 6.1.2 Cladding: This is the material on the outside of the quartz/silica fiber that is used to block the lateral escape of laser energy as it traverses the fiber. During stripping, you may nick the cladding and you will likely see the red aiming beam light as it escape the site of the damage. This is not a danger if all people in the area have the appropriate safety eyewear. Figure 6.1 Remove 1/4 to 3/8 of the jacket with Stripper Tool 6.1.3 Quartz/Silica Fiber: The fiber is fairly flexible but can be broken if bent into a small circle or bent at an angle of 90 degrees. The cladding will burn as protein from the gingiva accumulates on the fiber and will deteriorate the tip. It can fracture if not cleaved once the blackened area has reached 3-4 mm. Stop lasing and wipe off the tip regularly as you work to avoid accumulation of protein debris. Use water on a 2X2 gauze sponge to clean the tip. Do not use flammable materials like alcohol products when cleaning a hot tip. Dispose of all small fiber remnants after you have cleaved the fiber. They should be kept in a small box with a lid until they can be properly disposed of in the sharps container. 6.1.4 Stripping the Fiber: The fiber s jacket is removed using a stripper. Once a fiber tip is initiated and begins retaining debris from the tissues during lasing, you will get a deterioration of the fiber tip. When you have a blackened tip that extends 3-4 mm up the fiber shaft, it is time to cleave the fiber and strip the jacket to prepare for the next procedure. Begin by selecting enough jacket so that when it is removed, you will have approximately 3/8 of bare fiber exposed. Place the fiber in the stripper and grasp that portion of the fiber that will have the jacket removed between your thumb nail and index finger. See Figure 6.0. Grasp the fiber with the stripper by applying pressure to the handles. With a slow steady force, remove the jacket by pulling the fiber away from the stripper. See Figure 6.1. 14 Figure 6.1.1 Pull fiber out of Stripper Tool Figure 6.1 Stripping the jacket Jacket Cladding Fiber

6.1.5 Cleaving the Fiber: As the tip deteriorates, it is more likely to fracture and could fall into the sulcus or a deep periodontal pocket. To avoid this problem, it is prudent to periodically cleave the discolored tip. The cleave is made after stripping off the jacket to expose approximately 1-4 - 3/8 of bare fiber. The cleave should be made at a point approximately 1/2 (8-10 mm) from the site of the previous cleave so that there is no visible discoloration of the tip. The operator should place their index finger at the spot of the proposed cleave so as to stabilize the fiber. Using the pen style cleaver, draw the cleave blade across the top of the fiber with enough pressure to score the fiber. See Figure 6.2. With the thumb and index finger of each hand holding the fiber at a spot approximately 1/2 on either side of the cleave mark, break fiber against cleave. See Figure 6.2.1. After you have cleaved the fiber, point it perpendicular to a white piece of paper and place the laser in READY mode. Hold the tip approximately 1/2 from the paper and you should see a near perfect circle of red light. If you have a large comet effect radiating from the circle, you have not obtained a good cleave. WARNING: Always dispose of the fiber tip remnants in the Sharps container immediately after cleaving the fiber. Figure 6.3 Figure 6.4 6.1.6 Initiating the Fiber: The tip of the fiber should be cleaved to provide a flat surface that can be prepared to retain heat by introducing it to a dark material like ink, blood or water color paints. The easiest way to prepare the tip is to lightly move the flat surface of the tip across a piece of articulating paper at about 1 Watt CW. The tip will retain the ink and the ink will make the tip glow if you exceed 1-2 seconds while in contact with the paper. Prolonged heating will accelerate the deterioration of the tip. See Figure 6.5.1 Move the tip left or right while lightly touching the paper. Articulating paper Warning: Using too much power can start a fire on the articulating paper. Figure 6.5.1 Make one pass using light but steady pressure. Figure 6.2 Scoring & Cleaving 6.1.7 Fiber Disinfection: The fiber and spool remain in the fiber cassette and are not autoclavable while in the cassette. Once stripped, cleaved and initiated, the tip will reach temperatures of several hundred degrees centigrade, thus, as the laser emits energy, it will rid the tip of pathogens. Additionally, the tip can transfer heat up the shaft of the fiber to the edge of the jacket which may melt slightly. After your procedure, always cleave the blackened tip and strip the damaged jacket. Wipe the jacket down using BIREX, CIDEX or a comparable product that can disinfect the jacket. Dry the jacket using a clean 2X2 sponge prior to retracting the fiber into the fiber cassette. Also see section 6.0 on page 15. 6.2 Laser Maintenance Figure 6.2.1 Break the fiber against the cleave. If the circle has just a small comet effect, it will probably work. See Figures 6.3 and 6.4. 6.2.1 Laser Chassis Disinfection: The exterior of the laser should be cleaned using a liquid disinfectant similar to BIREX or CIDEX. Do not spray the disinfectant directly on the chassis. Apply with a gauze sponge or wipe. Do not use abrasive materials to clean the system. Place a barrier material similar to cellophane over the control panel and LED screen prior to treating the next patient 15

6.2.2 Calibration: The Odyssey 3Watt Laser uses solid-state circuitry to continuously monitor the power output, and adjusts the power supplied to the laser module to keep the output consistent with the user defined setting. If output levels are more than ± 20% of the set value, the unit is designed to shut down power to the laser, and an audible alarm will sound. If this happens, the unit should be turned off and allowed to sit for 5 minutes and turned on again. If the laser then performs without beeping, the microprocessor has been able to make operational adjustments and the unit will perform its functions. If, upon restart, the unit continues to beep, the unit will need to be sent in for adjustment by a Ivoclar Vivadent, Inc., Warranty Department. We suggest that your practice establish an internal calibration program for your laser. Recalibration is recommended a minimum of once per year based on average usage. Recalibration may be performed by the manufacturer by returning the unit. In the alternative, you may purchase a calibrated hand held power meter approved for use with 810nm devices to check power output. The laser should be set at 1, 2 and 3 Watts with output checked at each level. The output display should be within 20% of the meter reading. If not, recleave the fiber and re-check. If the output display is still outside the 20% tolerance, return the unit to the manufacturer for recalibration. There are no methods available for the user to adjust the calibration of the unit and the unit chassis must not be removed by the user for any reason. 6.3 Handpiece Preparation 6.3.1 Autoclavable: The handpiece for the Odyssey 3Watt is an autoclavable material that should be cleaned with warm soapy water, rinsed, bagged and autoclaved after each patient. Recommended autoclave cycle is 132 C at 27 psi for 15 min. *CAUTION remove plastic fiber locking cap prior to sterilization. Wipe off with cold disinfectant solution. Use it with handpiece to secure fiber. 6.3.2 Handpiece Components: The handpiece has three (3) components: a. Body b. Disposable Tip c. Fiber locking cap see Figure 6.6 below 6.3.3 Rocker Switch: Using the rocker switch on the top of the fiber cassette, press the switch to the extend position (toward the front of the laser). Extend 3-4 feet of fiber and feed it through the fiber locking collar after you have loosened the collar. The fiber should extend about 1 foot beyond the distal end of the handpiece so that you can strip the fiber and place the disposable tip. 6.3.4 Disposable Tips: The disposable tips can be shaped to provide the clinician any degree of access required. 6.3.5 Prepare Fiber: Using your stripper, remove approximately 1/4-3/8 inch of jacket from the distal end of the fiber. Cleave the fiber and examine the results by turning the laser to the Ready status. The aiming beam should create a near perfect circle when directed onto a white surface from about 1/4-3/8 inch above the paper. See Figures 6.3 and 6.4 on page 15. Once cleaved, the fiber tip should be disinfected (See Section 6.1.7). 6.3.6 Mounting the Tip: After selecting your tip, straighten it using your thumb and index finger. Begin at the proximal end of the tip and feed the stripped end of the fiber through the straightened tip until only bare fiber is extending beyond the disposable tip. Gentle slide the disposable tip onto the distal end of the handpiece and pull the fiber from the proximal end until the disposable tip is almost seated on the handpiece body. Seat the disposable tip securely on the handpiece body and tighten the locking collar. You may now shape the disposable tip to the desired form. Press the rocker switch to make sure the fiber moves freely. You should now initiate the fiber tip if your procedure calls for it. 7.0 Labels, Signs, Warnings and Manufacturer s Information 7.1 Federal Compliance: Various labels are included on the laser and the wireless foot switch as evidence of conformity to U.S. and International safety standards. The labels on the units are required under these standards for safety purposes and should not be removed. Please review all labels prior to using the laser. See Figures 7.0.1, 7.0.2, 7.0.3 and 7.0.4. *Fiber Locking Clasp Disposable Tip Body Fiber Locking Cap Figure 6.6 handpiece with disposable tip and fiber locking cap. *CAUTION remove plastic fiber locking cap prior to sterilization. Wipe off with cold disinfectant solution. Use it with handpiece to secure fiber. Figure 7.0.1 Label required under FDA Laser Notice 50 16

7.4 Aiming Beam laser 630-660 nm 3 mw Figure 7.0.2 Label required under FDA Laser Notice 50 7.5 CAUTION Changes or modifications not expressly approved by Ivoclar Vivadent, Inc. could void the user s authority to operate the equipment. 7.6 CAUTION Laser Radiation - Avoid exposure to the eyes or skin from direct or scattered radiation Figure 7.0.3 Laser Aperture Label 7.7 CAUTION This product contains no user serviceable components within the chassis. Visible and invisible radiation may be present when the cover is removed. 7.8 CAUTION US Federal law restricts this device to sale by or on the order of a licensed dentist. 7.9 CAUTION Eyewear that protects your eyes from wavelengths other than 810 nm does not provide proper protection for use with this laser. Damage to the retina or cornea may be irreparable if exposed to direct, reflected or scattered radiation. Figure 7.0.4 Rating Label with Serial Number 7.2 Danger Laser in Use: Each treatment area should have a laser in use warning sign posted at the entrance to the treatment area. This signage serves to warn people to not enter the treatment area without proper safety eyewear an protective clothing when the laser is in use. See Figure 7.1. Figure 7.1 Label to be posted at entrance to treatment area 7.3 Class 4 Laser Product - (Treatment laser) 810 nm 3 W 7.10 CAUTION Use of controls or adjustments, or performance of procedures other than those specified herein may result in hazardous radiation exposure. 7.11 Nominal Ocular Hazard Distance (NOHD) The NOHD is the distance beyond which the exposure during normal operation is not to exceed the appropriate Maximum Permissible Exposure (MPE). The NOHD for persons wearing safety glasses is shown in Table 1 below. The Nominal Hazard Zone (NHZ) is the area where the laser source within which exposure levels exceeds the MPE (which is the highest level of laser radiation to which a person may be exposed without hazardous effects or adverse biological changes in the eyes or skin). The outer limit of the NHZ is the NOHD. Eye Protection within the NHZ is mandatory. Source of Radiation MPE (mw/cm 2 ) Divergence angle NOHD (inches) (degrees) No Eye wear With Eye wear inch/cm (OD=4) inch/cm Fiber optic devices 1.66 9 60.63/154 0.6087/1.546 Reflection from Tissue 1.66 N/A 0.37/0.96 0.00378/0.0096 Assumptions: Maximum Laser Power = 5 Watts Direct viewing angle=0, reflectance viewing angle=20, reflectance coefficient of tissue=0.001 Table 1 - Nominal Ocular Hazard Distance (NOHD) for various viewing conditions while wearing eye protection. 17