User s Guide HHAQ-109. Portable Multi-gas Detector. Shop online at omega.com SM

Similar documents
OM-CP-Cryo-Temp Ultra Low Temperature Data Logger. Shop online at. omega.com For latest product manuals: omegamanual.

E1000 Portable gas detector

IMR EX610. Portable Gas Detector. Operation Manual

IMR IX176 Portable Gas Detector User Manual

User s Guide. Shop online at. Made in China. Two-wire Display and Control Unit

LVCN-302 Set-Up Manual for Sump Model

RECON/4a Portable Multi-Gas Detector Operation and Maintenance Manual

ENMET Corporation. PO Box 979 Ann Arbor, MI RECON. Manual. Manual Part Number MCN-410, 12/03/08

Macurco Single-Gas XL Series Monitor, CM-1XL Carbon Monoxide (CO), HS-1XL Hydrogen Sulfide (H2S) User Instructions

User s Guide TOTE SERIES. Wrap-Around Tote Tank Headers. TOTE SERIES Wrap-Around Tote Tank Heaters. Shop online at omega.com

User s Guide. OSXL685 and OSXL689. High Performance Infrared Thermometer. Shop online at

User s Guide. FLSW Series ROTAMETER WITH ALARM. Shop online at omega.com. For latest product manuals:

User s Guide. Setpoint Option Card. Shop online at. omega.com For latest product manuals: omegamanual.

TC100N. Fixed Gas Detector Operation Manual

GasAlertMicroClip X3 Specifications

MGC Dock User s Manual

QD6310. Fixed Gas Detector Operation Manual

User s Guide. CN606 / 612 AC Series & CN606 / 612 DC Series CONTRONAUTICS INCORPORATED. Table of Contents.

User s Guide FST1000 SERIES. Air Flow Switch with Two SPST Relay Contact Closures. Shop online at omega.com

Duct Mount. Installation Instructions

Humidity Monitor model 01080

ALTAIR 5X Multigas Detector Product Specification

Safe Operation Manual. Honeywell IQ Force Gas Detector

OI-2400-DOCK Multi Gas Docking Station. Operation Manual trevision 2.2w

MGC Dock User s Manual

For the Safety of Your

Warranty Registration

E C O L O G I C A L C O O L E R S INTELLIGENT

Confined Space Monitoring. Leak Investigation. Functions 4. Bar Hole Sampling. Line Purge Testing. Actual Size GX-2012

Room Humidity/Temp Sensor

FRESHAIR USER S MANUAL. CAUTION: Read manual carefully for proper procedures and operation.

G450. Multi-gas Detector. Field Operation Manual Oak Valley Dr, Ste 20, Ann Arbor MI USA (800) (734)

Series: MBC1-TC Mini Benchtop Temperature Controller

CO2 RESPONDER Portable Carbon Dioxide Detector QUICK REFERENCE GUIDE

Vanguard. Toxic & combustible gas detector

User Manual. Humidity-Temperature Chart Recorder. Model RH520

User s Guide OS210-C4. IR Temperature Sensors. Shop online at omega.com. For latest product manuals: omegamanual.

Operate the Mutli-Gas Monitor PID

YOUR PORTABLE AIR PURIFIER

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

5 Operating Modes GX Smallest 6 gas sample draw PID library of over 600 VOC s 2 Interchangable smart sensor slots

Duct Humidity/Temp Sensor

1, 2, 3, and 4-Gas Detector. Technical Reference Guide

Thermometer with Probe model 00891A

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

User's Guide. Pinless Moisture Meter. Model MO257

QAS SERIES PLUG-IN CONTROLLERS c/w SOLID STATE SENSOR

Thermometer with Wired Sensor

INSTALLATION AND OPERATING INSTRUCTIONS FOR THE VEHICLE-MOUNTED RADIATION DETECTION SYSTEM

EST User Manual. (Please read this manual carefully before using)

INSTRUCTION MANUAL FOR RIKEN PORTABLE TOXIC GAS MONITOR

SEC Signature Process Gas Analyzer

MODEL OX-07 Type A INSTRUCTION MANUAL FOR RIKEN PORTABLE OXYGEN MONITOR PT0E-0860

RK Multi Point Detector Operator s Manual

For Masonry & Factory Built Fireplace Retro-Fits Revised 5 December 2013 Healthy Hearth Riverside Dr. #214 Lake Elsinore, CA 92530

Users Manual. LAURUS Systems, Inc. - Ph: Fax:

QAS-101 SERIES SINGLE CHANNEL CONTROLLERS

Single Gas Clip and SGC Plus User s Manual Rev 2.1 READ FIRST BEFORE OPERATION WARNING STATEMENTS / AVERTISSEMENT Activating tthe detector for the

YOUR AROMATHERAPY DIFFUSER

RK-02 Multi Point Detector Operator s Manual

BST-MG08/09 Multi-gas Detecting Alarm. Manual Instruction

RK-05 Carbon Monoxide Detector Operator s Manual

MODEL: SP12C7. (Portable 4 gas detector) Operating Manual

OPERATION & INSTALLATION MANUAL

GasAlertMicro 5. 1, 2, 3, 4, and 5 Gas Detectors. Quick Reference Guide

Outside Air Humidity & Temperature Sensor

Room Monitor SAVE THIS MANUAL FOR FUTURE REFERENCE.

Owner s Manual. PIR-1 IR Learner

RK Carbon Monoxide Transmitter Operator s Manual

QUICK REFERENCE GUIDE

Clip SGD. Portable single Gas Detector. User manual

OPEARATING GUIDELINES FOR AERO-252 (Pages 1 to 3 contain general information on CO Guardian. Pages 4 to 6 contain information specific to Aero 252)

RK/ RK Carbon Monoxide Detector Operator s Manual

Digital Refrigerator/Freezer Thermometer model 00986

Remote Relay Module (RRM)

IMPORTANT. World-wide Patents. VIGIA is a trademark of Col-Ven S.A.

Document No

BLT-999W-2 Item # 60757

Because you re not like everyone else. PORTABLE CLEAR ICE MAKER IM200SS OWNER S MANUAL

Duct Temperature Sensors

Instruction Manual. AcuRite Atlas. Indoor Display model 06061

Falcon-II Next Generation, Air Quality Monitor CO2 & Temperature

Operator s Manual. 1, 2, 3, and 4-Gas Detector

BST-MG07 Multi-gas Detecting Alarm Manual Instruction

VX SERIES Wireless Thermostat with Occupancy Sensor

ELECTRIC FIREPLACE HEATER WITH SINGLE GLASS DOOR

Mold Alert... Operation Manual P0270

Classic & Deluxe Pants Press SPP-0669 SPP-0670

Single Gas Monitor 777

Thermometer model 02059

COMBUSTIBLE GAS LEAK DETECTOR PEN

HC12W COOL MIST ULTRASONIC HUMIDIFIER OWNER S MANUAL. Read and save these instructions.

Dräger X-am 5000 Multi-Gas Detection

Because you re not like everyone else. MISTING FAN MF18W OWNER S MANUAL

Single Gas Clip and SGC Plus User s Manual Rev 2.3

Dual Laser InfraRed (IR) Thermometer

User s Guide RH650. Shop online at omega.com SM. 9-In-1 Thermo-Hygrometer/ Moisture Meter

Controllers. Instruction Manual WARNING

This instruction manual explains the use of three models listed on the left.

Ion Endeavor Pump Controller Digital Level Control with Pump Alternation and High Water Alarm

Transcription:

User s Guide Shop online at omega.com SM e-mail: info@omega.com For latest product manuals: www.omegamanual.info HHAQ-109 Portable Multi-gas Detector

User Guide Thanks for our using our products. Before using this device, please read this manual carefully. Safety Notice Before using this device, please read the below information: Do not use if the device is damaged. Before using, please check for cracks or parts missing. If the device is damaged or uncompleted, please contact Omega Engineering.. We highly recommend you do the "bump test" described in Section 4.6 everyday to check performance of the device. And please calibrate it by following Section 4.8 if the test value is beyond specific range. 1

Bump test is to confirm the good response ability of the sensor, and to ensure that the audio, visual and vibration alarms work properly. Use only accessories specially designed for HHAQ-109. Use only the charger supplied with the device. Do not recharge the device in a hazardous area. If devices with catalytic or semiconductor sensor, are exposed to the target gas with concentration beyond its detecting range, the increase working load will badly affect its function or damage the device. Devices with catalytic or semiconductor sensor will be poisoned if exposed to environment containing lead chromate, sulfur compound, phosphorus compound or silicon; do not use the device in such environments. Please don't expose the device to the environment which consists of hydrogen sulfide, halogenated hydrocarbon or high corrosive gases for a long time. Otherwise, it will slow the response of the gas sensor and reduce the sensitivity. If the device has to be used in the above environment, please follow Section 4.3 to carry out the Bump test before using it. 2

Please don't expose the device to an environment with electric shock, strong magnetic field or serious continuous mechanic shocking. Disposal of the Lithium battery inside should be handled by the qualified persons or dangerous goods handling operators. Unauthorized disassemble, adjustment or repair of the gas device is forbidden. Avoid dropping or severe vibration of the device. For any problem beyond the description of this manual, please contact Omega Engineering. 1. Brief introduction HHAQ-109 is a compact and lightweight multi gas detector that continuously measures combustibles, O2, CO, H2S and other toxic gases in ambient air. Its functional and watertight design (IP 66) incorporates a Bump proof, rubberized housing to meet the toughest requirements of harsh environments like underground tunnels, mines etc. 3

Main features: Friendly human-machine interface Ultra-wide angle LCM screen Self adjusting function to lower the testing error Adjustable 2-level alarm levels; STEL and TWA alarm Audible, visual and vibrative alarm signals Self protection design for combustible gas sensor Battery low voltage alert function With real-time clock Intrinsically safe design 2. Technical data Detection method: Diffuse naturally Detecting gas: See the Table 1 and Table 2 on the last page Response time: O2, CO: T 90 <30s Indication error: ±5% FS (O2) / ±5ppm (CO) Working condition: Temperature: -20 C~50 C Humidity: <95%RH (no condensation) Power source: Lithium battery, DC3.7V 1800mAh Charging time: 6 hours Battery working time: 200 hours (no alarming status) Explosion proof: Exia II CT3 Ga Ingress protection: IP66 Dimension: L*W*H 120mm 68mm 30mm 4

Weight: About 160g 3. Structure drawing and display information Button function: 5

Power on: Hold button for 1 second and then release it Power off: Hold button for 3 seconds till the screen is off Open backlight: Press button once Mute and cancel vibration: During alarming status, press button once Check device status: In the status of power on, backlight is on and detection status, press button once and the screen will show the device status including the max. value, the min. value, STEL value, TWA value, present time, serial number and version number. Calibration: After the device entering detection status, hold button for more than 20 seconds. The screen will first turn off and then turn on. When the screen shows calibration status, release the button. 6

4. Operation instruction 4.1 Power on When the device is powered off, hold the button for 1 second and then release it. The buzzer gives sound once and the device is on. The screen will in turn display power one interface, warm-up interface, self test information, high alarm level, low alarm level, TWA value, STEL value etc. And then, the device enters into detection mode. On the screen, it shows the value of the target gases. Note: The device is initially set as auto zero calibration after power on, always power on the device in the clean air. Otherwise, the device will not work properly. 4.2 Power off In the power on mode, hold the button for 3 seconds. Buzzer will give 3 long sounds and 2 short sounds. Then the device is off. 4.3 Alarming When the gas concentration in the air reaches or exceeds the 7

preset alarm levels, the device will give audible, visual and vibrative alarming signals. If the user cannot leave this environment, he can cancel the audible and vibrative signals by pressing the button, so as to save the battery voltage. If the target toxic gas concent ration reaches or exceeds the preset value, the device will also give STEL and TWA alert signals. The device will also give sensor fault alert signal and low voltage alert signal. On the screen, it will show the relative signal indication. Note: The functions of STEL alert, TWA alert, sensor fault alert and low voltage alert must be activated in advance when using. 4.4 Device status checking When powering on, backlight is on and detection status, press the button once and the screen will show the device status including the max. value, the min. value, STEL value, TWA value, present time, serial number and version number. 8

4.5 Calibration With the device in detection mode, hold the button for more than 20 seconds. The screen will firstly turn off and then turn on. When the screen shows calibration status, release the button. After warm up, the device will begin zero calibration automatically. Please put the device in clean air. When the screen shows the calibration gas point, please follow the drawing on next page to input a standard calibration gas. If you do not have a standard calibration gas, do not attempt to calibrate this instrument yourself. Doing so will compromise its accuracy. The device will automatically analyze the input gas and the complete the input gas and the complete the calibration of the sensors. After successful calibration, the screen will show the calibrated sensor name and relative information. 9

4.6 Bump test In order to make sure the device is working normally, it is suggested to do a bump test once every day before using. Test method: a. After the device is powered on, put it in to the target gas or standard gas environment which is above the high alarm level. If the device works normally, then user can carry it for detection purpose. b. If the detector reading is above the normal indication error range, then please follow the above Section 4.5 to re-calibrate the device. c. If the device doesn t response to the gas, please contact Omega Engineering. 10

5. Charging When the device cannot be powered on due to low voltage, please charge the device through the supplied charger. After correctly connecting the device with the AC100~240V power source, the device will be automatically power on and show the charging icon. When the screen icon shows the battery voltage is full, please unplug the device, and use normally. Warning: In charging status, the device has no detection function. It s forbidden to charge the device on working site, so as to avoid any fire or explosion occurring. Please charge the device when it is power off, so as to ensure the charging efficiency. Note: Please charge it once every 3 months if the device is not used for long time. It s forbidden to charge the device in the environment which temperature is below than 0 C, because it will possibly damage the battery. 11

6. Sensor using and replacement The device adopts sensor module. During using, please note the life span of the sensor modules. When the sensor module is expired, contact Omega Engineering to purchase a new module. In order to assure the accuracy, it s suggested to re-calibrate the sensor modules once every 6 month. If replacement modules are purchased, the package includes the sensor module and calibration parameter file. After plugging the new sensor module onto the device, please import the parameter file into the device through the software on the PC. 7. Trouble shooting Fault Reason Solution Cannot power on No response to gas Value displayed is not correct Time display is not correct Gas value displayed is negative Sensor fault displayed on the screen Too low voltage Dead lock Circuit fault Warming up Circuit fault Sensor overdue Not calibrated for long time Battery is used up Electromagnetic interference Sensor drift Sensor fault Charge device Contact Omega Contact Omega Wait till it ends Contact Omega Contact Omega Calibrate the sensor Charge the device and set the time Reset the time Make zero calibration Contact Omega 12

8. Using notice Avoid dropping or shocking the device. If using in high concentration gas environment, the device can not work normally. Please strictly follow this manual to use the device. Otherwise, it will cause the incorrect detection result or damage the device. Do not use or store the device in any corrosive environment or harsh environment (like exceeding high or low temperature, high humidity, electromagnetic field, strong sunshine etc.). After long term use, if there is dust on the interface, please use clean soft cloth to clean it. Otherwise, surface will be scratched or damaged. In order assure the detection accuracy, recommended re-calibration period is every 6 months and cannot be more than 1 year. Dispose of Lithium battery in accordance with local requirements. For any problem not described in this manual, please contact Omega Engineering. Disassembling, modifying and repairing the device should be carried out by authorized personnel. Do not charge the device or upload the data to PC in hazardous environments. 13

Attached Table 1 Gas Detection Range Preset Low Alarm Preset High Alarm TWA STEL O 2 0-30%vol 19.5%vol 23.5%vol / / CO 0-1000ppm 35ppml 200ppml 35ppml 200ppml 14

omega.com info@omega.com Servicing North America: U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047 Stamford, CT 06907-0047 USA Toll-Free: 1-800-826-6342 (USA & Canada only) Customer Service: 1-800-622-2378 (USA & Canada only) Engineering Service: 1-800-872-9436 (USA & Canada only) Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail: info@omega.com For Other Locations Visit omega.com/worldwide The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA s control. Components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs. OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages. CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a Basic Component under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner. RETURN REQUESTS / INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product. FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA: 1. Purchase Order number to cover the COST of the repair, 2. Model and serial number of the product, and 3. Repair instructions and/or specific problems relative to the product. OMEGA s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. Copyright 2016 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. M5500/0116