Justin Douglas KU NMR Labs v1 10/29/09

Similar documents
DCIF Training Guide Variable Temperature Bruker B400 and B401 (last update )

Temperature Control and VT Experiments on Varian Spectrometers

TECHNICAL NOTE. Response Factors for Flame Ionization Detector Operation FLAME IONIZATION DETECTOR OPERATION BASICS

FACT SHEET Program: Lab Fire Safety

vwr.com. L a b o r a t o r y P u m p s + S y s t e m s

Intelligent combined sensor for Air Quality (VOC) + Humidity (rh) + Temperature (T) with analogue outputs, Modbus and LCD display. Features.

Operating manual ELECTRIC ASPIRATOR 00) : 1) Page 1

LD910. User Manual. Combustible Gas Leak Detector. Detects all Hydrocarbon and other Combustible Gases. Made in U.S.A. MSHA Approved Intrinsic Safety

JAG group microwave manual. (updated October ) Table of contents. 1 General information 2. 2 General safety rules 2. 3 The microwave reactor 3

Low mg/m

Multivapor TM P-6 / P-12 Operation Manual

Transmission Photometer UV / VIS / NIR / IR. Reflection Photometer VIS / NIR / IR

CentriVap DNA Systems. Models

Choosing the best vacuum concentrator for your laboratory. Tips and best practices from industry pioneers

Electron Microscopy Sciences

Q Trap LC/MS/MS TurboIonSpray Ion Source Manual. About This Manual... 1 Conventions Introduction... 3 Features... 4

Silicon-Lined Canister Cleaning Practices and Blank VOC Concentrations. Jason S. Herrington, Gary Stidsen, Trent Sprenkle, Steve Kozel

API 2000 LC/MS/MS TurboIonSpray Ion Source Manual

ULTIMA XIR ULTIMA X 3 ULTIMA XE. ULTIMA X Series. [ State-of-the-Art Gas Monitoring]

4P75 CiTipeL Combustible Gas Sensor Part Number: PM

PROTOCOL FOR FREE LIPID SOP

Operation Manual. Vacuum Controller V-850/ en

Thermo Finnigan LCQ Advantage Getting Started

4P-90 CiTipeL Combustible Gas Sensor Part Number: PM

To receive important product updates, complete your product registration card online at register.labconco.com

CD100A COMBUSTIBLE GAS DETECTOR

Electron Microscopy Sciences

Autoclave Operations Manual

Shepherd FTIR. Laboratory Quality Results. Continuous, real time portable monitoring of 385 individual species of gas: 100+ HAPs. ...

SENTRY-MS Mono-static Multi-gas Open Path Analyzer

ROTARY VACUUM EVAPORATOR RVO 400

Application Note CBRNE Stand-Off Detection of Toxic Industrial Chemicals in Industrial Complexes Using RAPIDplus

AirSentry FTIR Multi-gas Open Path FTIR Analyzer

PCD110X - Premium Choice with DC Motor (Ceiling Mounted Ventilation Fans)

Early Warning Solutions

SEC-MALS Noise Assessment and Cleaning Guide

Solvent Evaporation. Fast, Reliable and Affordable

159 UV-VIS Detector. User s Guide. Part Number

QAQC LABS ANALYTICAL DRYING

NGD8800 USER S MANUAL

Type SP87300 SAFE-T SHP9 Explosion Proof Stirring Hot Plate

Haz Waste Management

Heated tools. Semiconductor equipment

NMR SAMPLING ACCESSORIES CLEANING NMR SAMPLE TUBES

Explosion Proof Speaker, Ringer & Speaker/Ringer

HERMS (Heat Exchanger Recirculating Mash System) Controller

Mariner Microspray and Low-Flow Electrospray Fused Silica Spray Tip and Capillary Kit. Operating Instructions. 1 Product Description.

Security + Automation Made Easy QUICK START GUIDE

DYE QUALITY CONTROL REFERENCE MATERIAL KIT

LFL Monitoring. MEGTEC Systems

PARTS CLEANING AND DEGREASING

TM-802 Formaldehyde Meter User s Manual

Laboratory Diaphragm Pumps for Air Gases Liquids

Each burner shall be equipped with a Micro-Processor Based Burner Management Flame Safeguard and Parallel Positioning Control System.

Vanquish. Variable Wavelength Detectors VF-D40. Operating Manual

OVEN INDUSTRIES, INC.

ESD3000 ESD200 Combustible / Toxic Gas Detector

CryoFit. Handling of Flow Cell and Capillaries. User Guide. Version 002

=Classic range= Classic range technologies. Magnesium anode. Sheathed heating element

What s So Special About TFA?

On-line SPE Enrichment of Trace Organic Contaminants in Water and Juices

ESD500. ESD500 Combustible / Toxic Gas Detector

Analysis of Pharmaceuticals and Personal Care Products in Groundwater by Online SPE and LC/MS/MS

Safety Lab Goggles are required! Apron is optional!

Standard Operating Procedure Glovebox and Thermal Evaporator

Installation and Wiring 1.

User manual and installation guide

Vacuum Pump V-100 Operation Manual

CEM MARS5 Microwave Updated November 13, 2017

MICROpeL 75 Combustible Gas Sensor Part Number: PM

Explosion Proof Speaker, Ringer & Speaker/Ringer

OPERATIONAL MANUAL TUBE FURNACES

API LC/MS/MS TurboIonSpray Ion Source Manual Table of Contents

An Automated SPE Procedure and Analysis for the Extraction of Chlorpromazine and Thioridazine from Plasma Application Note 213

MICROPLATE EVAPORATORS

Portable Combustible Gas Detector GP-1000 Operating Manual (PT0-120)

Process and Air Monitoring. Online GC systems for continuous monitoring of organic compounds in air and gaseous samples

10,000 prize draw. EXTRA year guarantee FREE rewards gallery. instructions. details. Register online now!

APPLICATION DATA SHEET

User Manual FLOWMAX-90. for model. Condensing water heater 85,000 BTU. Installation, operating, commissioning and maintenance instructions.

Explosion-proof Duct Heaters

800 MHz US 2 NMR system

New Features. IES Virtual Environment 6.4

HITACHI CHROMASTER. Clarity Control Module. Code/Rev.: M173/70I Date: 6/23/2018

Solid Phase Extraction using the 96-well Format ISOLUTE -96 Fixed Well Plates and ISOLUTE Array 96-well Plates

OPSIS LiquidLINE SoxROC Extraction Unit

Temperature Control for Research and Industry. Model 270 Manual

10,000 prize draw. instructions. EXTRA year guarantee FREE rewards gallery. details. Register online now!

99 Washington Street Melrose, MA Phone Toll Free Visit us at

Agilent 1200 Series Variable Wavelength Detector

Operation manual [LOW TEMP. INCUBATOR] Model : IL-11 / 11A / 21 / 21A. Manual No : 00HAA ( Version : 5.0 )

Virginia St. Jean, CIH Industrial Hygienist/DPH Green Programs Manager SF Hazardous Materials Unified Program Agency

GENERAL WET BENCH SAFETY AND USES SOP. October 2013 GENERAL SAFETY RULES

Soft Ionisation Techniques, Terry Whitmore, Hiden Analytical

Customised HPLC Systems for Highest Performance

Security + Automation Made Easy QUICK START GUIDE

Nanoscale Fabrication & Characterization Facility. Thermionics E-Beam Evaporator Users Guide

Thank You for Attending Today s Webinar:

Thermo Scientific Flammable Material and Explosion-Proof Refrigerators and Freezers. Safe cold storage for Hazardous material

Leading Safety Standards Superior Ease of Use Reduced Cost of Ownership. Intelligent. Evaporation

Transcription:

High Temperature NMR Justin Douglas (justindouglas@ku.edu) KU NMR Labs v1 10/29/09 Special Note: This guide is intended to supplement proper training and formal check out by the NMR facility manager. Improper operation or decision-making when doing variable temperature experiments can have disastrous consequences for the instrument and other users. How the NMR regulates the temperature: Bruker uses a feedback loop to regulate the temperature. Figure 1 highlights key components of this system. The temperature is measured by a thermocouple sitting below the sample. The entire probe is heated using an electrical heater located at the bottom of the probe. This heat is then transfer using air flow. Figure 1. The key components of the temperature regulation system for Bruker NMRs. Note that this diagram is not to scale. DOs and DO NOTs: DO DO NOT Reserve at least 2 hours of time to elevate the temperature to 90 C, record a 1H 1D and cool the probe back to room temperature. Use the white ceramic spinners for temperatures above 50 C Get in a hurry Use the blue plastic spinners for temperatures above 50 C

Increase the temperature in 10 C steps and allow it to equilibrate Increase the gas flow up Know the boiling point of your solvent (see Appendix 1 for boiling points of common solvents) Keep temperature below 90 C Know that your compound is stable at the temperature you sample Increase the temperature too rapidly. Increase the maximum heater power to more than 10% Approach the boiling point of your solvent Go above 90 C (see Justin if you need higher temperatures. Above 90 C we switch to N2 gas instead of air to prevent oxidation of the heating element) Explode a sample in the probe. You will be held financially accountable for any damage to the probe. How to set up high temperature NMR experiments: #-1) Make sure you ve gone over this procedure with Justin or Sarah before you begin. #0) Use less solvent. You want a short sample (~25 mm or ~ 350 µl). This will make it harder to shim, but easier to maintain temperature across the sample. If you need help shimming Justin or Sarah will be happy to help. #1) Insert your sample into the spinner and use the depth gauge to set the sample depth. Use the white ceramic spinner for temperatures above 50 C. The spinners are usually located in the second cabinet from the right on the south wall in Malott B042 and the top drawer on the right in SBC. The blue plastic spinners will warp at higher temperatures. The ceramic spinners have a tighter fit to the o-rings, so be careful inserting your tube. #2) Log into the workstation. Follow the normal steps to set up your dataset, load acquisition parameters and lock. #3) Type edte at the command line to open the temperature control window. Figure 2 shows the main display tab of this window.

Figure 2. The main display tab of the edte window. In this tab we monitor the current sample temperature, the target temperature, the output of the probe heater and the gas flow. #4) In order to accelerate the response of the temperature regulation system, we increase the gas flow. If going to high temperatures, you can increase to 670 l/h (Fig. 3), although it is OK to use less. Click the + button once and see if that is enough to get to the desired temperature in a reasonable time. Figure 3. Increasing the gas flow. #5) Click the change button to the right of the target temperature. Enter 35 C. Click on the monitoring tab to observe the sample temperature (green), the target temperature (yellow) and the heater output (red) as a function of time (Fig. 4). It will take ~ 5 minutes for the system to equilibrate (Fig. 5).

Figure 4. Monitoring the response to an increase in temperature. Figure 5. The temperature has reached equilibrium. #6) After the system equilibrates, you can continue to increase the temperature in 10 C steps until you reach your target temperature. The reason we increase temperature so deliberately is that rapid temperature changes could damage the quartz inserts inside the probe. These inserts insulate and protect the coil, but they also isolate the gas flow from the thermocouple.

If the heater power (red line in monitoring tab) is at the maximum for more than 2 minutes and the temperature is not increasing, then you can increase the maximum output of the heater. Click the set max. button next to the Probe Heater percentage in the main tab (Fig. 6). Increase the maximum power by 1-2%. Go back to the monitoring tab and check if the temperature increases. Under no circumstances should you ever set the maximum power higher than 10% Figure 6. Set Sample Heater Max Power window. Under no circumstances should this be set above 10%. #7) When you reach your target temperature and the sample has equilibrated you should shim. It is very important to wait for the temperature to equilibrate because thermal gradients will wreak havoc with shimming. Hence, it is a waste of time to shim before the sample equilibrates, because you ll just have to reshim once it equilibrates. Also, if you have been trained to tune the probe you can tune it at this step. #8) Record a spectrum as you normally would. #9) The final step is to cool the probe back to room temperature. First, reset the heat power to 5% and reduce the gas flow back to 270 l/h. Reduce the temperature by 10 C and monitor the temperature in the monitoring window (Fig. 7). Note that you don t need to let it fully equilibrate at each step. Give it ~1 minute before lowering the temperature another 10 C. Then final step (from 35 to 25 C) will take >5 minutes. Once the temperature reaches 30 C you can turn the probe heater off by clicking the button to the left of the heater percentage in the main tab.

Figure 7. Monitoring cool down to room temperature. What could go wrong: If the temperature starts to increase out of control it is important not to panic. It will take several minutes for your sample to boil. Turn off the heater. Once the heater is off, then eject your sample. Please let Justin or Sarah know about any problems or issues.

Appendix 1. Boiling points of common NMR solvents: solvent boiling point ( C) acetic acid-d4 118 acetone-d6 56.5 acetonitrile-d3 81.6 benzene-d6 80.1 chloroform-d 61-62 cyclohexane-d12 80.7 deuterium oxide 101.42 N,N-dimethyl-formamide-d7 153 dimethyl sulfoxide-d6 189 1,4-dioxane-d8 101.1 ethanol-d6 78.5 methanol-d4 64.7 methylene chloride-d2 39.75 pyridine-d5 115-116 1,1,2,2-tetrachloroethane-d2 147 tetrahydrofuran-d8 66 toluene-d8 110.6 trifluoroacetic acid-d 72.4 trifluoroethanol-d3 75