Medical electronic suction unit Properties and uses Suction units MEVACS 40, MOD-1 present a new generation of suction units. Their operation is very easy. The whole suction process is controlled by microprocessor and the choice of the mode of suction and parameters are set on the touch screen. This makes them suitable for general use in surgery, neurosurgery, plastic surgery, genecology, and thorax draining, at AR-UIC departments, as well as for suction from air passages. There is no service-needed vacuum pump into the suction unit. Besides classical suction have special modes - drainage suction, pulse suction, vacuum extraction and STAND-BY function. Ordering number: MEVACS 40, MOD-1 No:72000.02 Page 1 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 1 / 11 JS
Main advantages of the suction unit are the following: 1. Oil-free vacuum pump, maintenance-free. 2. High level of underpressure - 93 kpa / 700 mmhg (93% of vacuum). 3. High suction output - MEVACS - 40 l/min 4. Long-term operation - several days. 5. Simple operation on the touch screen and a visual indication of the amount of vacuum achieved. 6. Very low noise level. 7. Microprocessor guaranteed precise height control vacuum. 8. STANBY function - standby economical mode (frequent unit switching on and off during the operation is not necessary). 9. Reliable overflow protection system: safety jar for foaming secretions with safety valve in the lid safety valve against overflow in the lid of collecting jar contactless electronic scanning of filling collection jar (after filling the jar is suction stopped automatically - light and sound signalling) hydrophobic microbiologiccal filter 10. The suction mode (NORMAL MODE) in the range 0 to max. vacuum (0 to -93 kpa). 11. The soft suction mode "DRAINAGE" between 0 and -20 kpa: with protection of sensitive tissue with the possibility of thorax jars interconnection with the possibility of long-term drainage suction by means of accurate limit water vacuum meter 12. Pulsed extraction mode "PULSE" in the whole range of vacuum values. 13. The vacuum extraction mode "VACEXTR" with automatic vacuum rise. 14. Protection of the unit, personnel and patient against infection by means of fitted microbiological filter. 15. Memory function of adjusted unit values during blackout 16. The possibility control all functions with optional mode of foot controller: a/ as ON/OFF switch b/ as foot vacuum regulator - MOMENTARY 17. Wide range of additional accessories: unbreakable autoclavable collection jars with volume 0.5L, 1L, 2L, 4L universal trolley with a basket, holders... suction adapters, single-use suction bags, cylinders... various diameters of silicone suction hoses vacuum-extractors, curettage adapters Areas of use a) Surgery - Neurosurgery - Cardio surgery - Plastic surgery - Accident surgery - Orthopaedics b) Anaesthetic/Resuscitation Dpt. (soft suction, long-term drainage suction) c) Gynaecology (suction curettage, vacuum extraction, secretion suction of new-born children and children, pulse suction of breast milk) d) Ward of internal diseases - Intensive Care Unit - Endoscopy e) Ambulances, other - in the range of technical parameters Page 2 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 2 / 11 JS
Product description 1 2 3 4 5 1) main switch ON/OFF 2) - touchscreen 3) hydrophobic filter 4) collection jar 5) jar holder 6) suction hose 6 7 7) safety jar 8) safety jar holder 9) - connector for foot control 10) - electrical socket with fuses 8 9 10 11) - contactless level sensor 11 Page 3 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 3 / 11 JS
Vacuum control and parameter settings On control and setting the required parameters is intended "Touch Screen". Digital vacuometer Value of vacuum setting Suction modes The choice of value vacuum by touching Adjust values modes Control foot switch Control of level sensor Contactless sensor of overfilling jar 25 26 If the level of secretion would reach contactless sensor (25), suction stops and the icon will appear (26). The operator must empty the jar conventional way by local ordinances or is replaced by a new empty jar. Page 4 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 4 / 11 JS
Foot switch and selecting the method to use Foot switch can operate in two modes: a) as a foot vacuum regulator - MOMENTERY b) as an ON / OFF switch 32 32 Mode foot controller - MOMENTERY: by switch on suction unit and by pressing and holding the suction foot controller suction unit works (increasing vacuum to the set value) and releasing the suction foot controller stops working Mode ON / OFF: by switch on suction unit and by pressing the foot control switch ON the suction as if we have turned on the main switch ON / OFF repeatedly pressing the suction unit is switched off. Touch the icon (32) "Footswitch" (the visible points ON) mode is set to "FOOT CONTROLLER - MOMENTERY". Repeatedly touch switches to the mode (the visible points OFF) "ON / OFF". Suction modes The suction unit MEVACS has the following suction modes: 1. Suction "NORMAL MODE" in the range 0 to max. vacuum (0 to -93 kpa) with automatic "STANDBY". 2. Soft suction DRAINAGE in the range -5 to -20 kpa. 3. Pulse suction PULSE in the range 0 to max. vacuum (0 to -93 kpa) 4. Vacuum extraction VACEXTR with controlled vacuum rise. Suction NORMAL MODE The basic suction mode between 0 and -93 kpa. Size vacuum selected easily - by touching the desired value. STANBY mode - power saving mode. This mode is recommended in cases with frequent switching on and off. The choice of value vacuum by touching Page 5 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 5 / 11 JS
Soft suction DRAINAGE" Range of suction 0-20 kpa. Touch to select the value of the of vacuum of 4 pre-set values 5; 10; 15 or 20 kpa or vacuum can also select continuously from 1 kpa to 20 kpa by touching the screen. Mode soft suction "DRAINAGE" serves primarily to protect the soft tissue in long term suction surgical wounds. By leech onto tissue to suction cannula the suction unit automatically interrupts the process of suction holding time and then vacuum drops to zero. Values holding time at max. vacuum and zero vacuum can be set from 0 to 12.5 seconds. After the pre-set time will be suction process automatically resumes. Pulse suction PULSE Range of suction 0-93 kpa. The "PULSE" is used mainly to protect the delicate tissue, so its use is for heavy operations in the brain, heart, intestine.... By leech onto tissue to suction cannula the suction unit after reaching the set maximum vacuum automatically interrupts the process of suction holding time and then vacuum drops to a minimum value set. After the pre-set time will be suction process automatically resumes. Values holding time the "MAX" vacuum can be set from 0 to 6.2 sec. and for "MIN" vacuum in the range from 0 to 12.5 sec. Vacuum extraction VACEXTR The suction unit can automatically manage the startup of vacuum. By selecting the "VACEXTR" and the closed end of the extraction cup device automatically creates a vacuum of 20 kpa. The vacuum was gradually increased step by 2 kpa and a pause between each step is factory-set to 6 seconds. The vacuum is gradually increased up to the set value. After reaching the maximum set value, a short acoustic signal. If it is necessary above (or below) the vacuum, its value we can continue to increase to a maximum of 93 kpa. With repeated run-up to the desired final value of vacuum MEVACS climb on previously attained value of vacuum it without delay and continues in automatic mode. Page 6 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 6 / 11 JS
Check functions The suction unit MEVACS has the following check functions: 1. Full secretion jar check 2. Clogged filter check 3. Suction unit adjustment Clogged filter alamr Filling jar alarm The emergency status alarm Adjustment of suction with entering the PIN code. This mode is only for service technicians. Microbiological filter Suction microbiologic filter eliminated risk of contamination with bacteria, viruses and infected particle from patient to vacuum pump or central vacuum distribution. Suction microbiological filter is hydrophobic with very high bacterial efficiencies up to 99,99999% particles bigger than 0,027 micron (which is smaller than Hepatitis A, B and C). Page 7 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 7 / 11 JS
Technical data Typology (MDD 93/43/EEC) Medical Device CLASS IIa Model MEVACS 40, MOD-1 Specification by EN ISO 10079-1 High Vacuum - High Flow Electrical safety Class I Direct/indirect contact safety Type B Substances penetration protection IP 20 Safety level in presence of anaesthetic/explosive mixture Not protected device Power consumption 220-240V ± 10% / 50-60 Hz Imput power 110 VA + 15% Electric protection 2 fuses Ø 5 x 20 T - 1A Vacuum pump Maintenance-free diaphragm vacuum pump Air flow rate of pump 40 L / min Air flow rate by Drainage with limited water column 1,5-4 l/min Maximum vacuum up to - 93 kpa (700mmHg) Fine-control suction up to - 20 kpa (150 mmhg) Suction Drainage suction with limited water column up to - 40 cm/h2o Pulse suction up to - 93 kpa (700mmHg) Jars (portable version) 1 x 2,0 LT polycarbonate jar with overflow valve Jars (with trucks version) 1or 2 x 1,0 / 2,0 / 4,0 LT polycarbonate jars with overflow valve Suction hose Silicone - Ø 8 x 3 mm - 1,5 m Operation Continuous Hydrophobic suction microbiological filter with Filtration very high bacterial efficiencies up to 99,99999% particles bigger than 0,027 micron Weight Portable version 10,4 kg On trolley version 19,6 kg Dimension (L x W x H) Portable version apr. - 525 x 255 x 420 mm On trolley version apr. - 565 x 455 x 1065 mm Supply cord 4 m Control of vacuum on touch screen Precision of control vacuum (precision ± 1,6%) Polycarbonate jar ml. graduated - autoclavable Special features (optional): - Trolley MEVOZ - Operation with foot control (IP 68) - Limited water column up to 40 cm/h2o Working condition Surrounding envir. temperature 5 35 o C Relative humidity 30 75% Storage condition Surrounding envir. temperature - 40 70 o C Relative humidity 10 100% International standards: MDD 93/42/EEC; EN 60601-1; EN 60061-1-2; EN ISO 10079-1; EN 14971 Technical life: 10 years Page 8 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 8 / 11 JS
Alternative sets Suction unit MEVACS with this accessories: suction microbiologic hydrophobic filter 300 ml safety jar with outlet 2.0 L polycarbonate jar TM with silicone lid (on suction unit) 2.0 L polycarbonate jar TM with metal lid (on trolley MEVOZ) suction catheter container 50/400 mm TM trolley MEVOZ MV7 with basket Suction unit MEVACS with this accessories: suction microbiologic hydrophobic filter 300 ml safety jar with outlet 2.0 L polycarbonate jar TM with silicone lid (on suction unit) 2.0 L polycarbonate jar TM with metal lid (on trolley MEVOZ) 4.0 L polycarbonate jar TM (on trolley MEVOZ) suction catheter container 50/400 mm TM trolley MEVOZ MV7 with basket Page 9 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 9 / 11 JS
Suction unit MEVACS with this accessories: suction microbiologic hydrophobic filter 300 ml safety jar with outlet 2.0 L polycarbonate jar TM with silicone lid (on suction unit) 2 x 4.0 L polycarbonate jar TM (on trolley MEVOZ) suction catheter container 50/400 mm TM trolley MEVOZ MV7 with basket Suction unit MEVACS with this accessories: suction microbiologic hydrophobic filter 300 ml safety jar with outlet 2.0 L polycarbonate jar TM with silicone lid (on suction unit) 2.0 L polycarbonate jar CM for thoracic suction (on trolley MEVOZ) Limiting water manometer TM 40/600 mm for thoracic suction trolley MEVOZ MV7 with basket Page 10 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 10 / 11 JS
Optional accessories Each suction unit MEVACS can be equipped with additional accessories according to user needs. MEDIST, s.r.o. is ready to realize your requirements. Please contact us. Page 11 / 11 Medical electronic suction unit MEVACS 40, MOD-1, TOUCHSCREEN, M-VP-0011.2 ENG 30.7.2017 11 / 11 JS